Hach DR/820, DR/850, DR/890 Instrument Manual

Page 1
DR/820, DR/850, and DR/890
Portable Datalogging Colorimeter
Instrument Manual
48440-18
© Hach Company, 1997–2009. All rights reserved. Printed in the U.S.A. jk/te/dk 2/09 2 ed
Page 2
Page 3
TABLE OF CONTENTS
SAFETY PRECAUTIONS ........................................................................................................3
OPERATION
SECTION 1 GENERAL DESCRIPTION .............................................................................7
1.1 Instrument Description.........................................................................................................7
1.2 Unpacking the Instrument ....................................................................................................8
1.2.1 Standard Accessories.................................................................................................. 9
1.2.2 Optional Accessories ..................................................................................................9
1.3 Description of the Keypad....................................................................................................9
1.4 Display Screen in Function and Numeric Modes...............................................................12
1.5 Icon and Display Screen.....................................................................................................12
SECTION 2 INSTRUMENT SETUP...................................................................................15
2.1 Battery Installation .............................................................................................................15
2.2 Turning the Instrument On.................................................................................................17
2.3 Setting the Date and Time..................................................................................................17
2.3.1 Entering the Correct Date .........................................................................................17
2.3.2 Entering the Correct Time ........................................................................................18
2.4 Sample Cell Insertion.........................................................................................................19
2.5 Sample Cell Adapter Installation .......................................................................................19
2.5.1 Using the 16-mm COD/Test ‘N Tube Vial Adapter.................................................19
2.6 Using the Instrument Cap as A Light Shield......................................................................20
SECTION 3 INSTRUMENT OPERATION .......................................................................21
3.1 Menus.................................................................................................................................21
3.1.1 Setup Menu...............................................................................................................21
3.1.2 Recall Menu..............................................................................................................22
3.2 Performing an Analysis.....................................................................................................
3.2.1 Colorimeter Setup for Sample Analysis ...................................................................23
3.2.2 Sample Preparation...................................................................................................24
3.2.3 Zeroing the Colorimeter ...........................................................................................24
3.2.4 Measuring the Prepared Sample ...............................................................................24
3.2.5 Alternate Chemical Forms........................................................................................25
3.2.6 Using the Timer ........................................................................................................25
3.2.6.1 Using the Timer in Manual Mode................................................................25
3.2.6.2 Stopping the Timer ...................................................................................... 26
3.3 Reagent Blank Correction ..................................................................................................26
3.4 Adjusting the Standard Curve ............................................................................................27
3.5 Using a Programmed Method ............................................................................................29
.22
3.6 Quality Assurance ..............................................................................................................30
1
Page 4
TABLE OF CONTENTS
SECTION 4 CREATING USER-ENTERED PROGRAMS.............................................. 31
4.1 User-Entered Programs...................................................................................................... 32
4.2 Calibration Curves ............................................................................................................. 32
4.3 User-entered Program Information for Bleaching Chemistries......................................... 32
4.4 Creating a New User-entered Program.............................................................................. 33
4.5 Reviewing and Editing User-Entered Programs................................................................ 36
4.6 Erasing User-entered Programs ......................................................................................... 37
SECTION 5 DATA RECALL AND STORAGE ................................................................. 39
5.1 Recalling Data.................................................................................................................... 39
5.2 Erasing All Stored Data .....................................................................................................40
SECTION 6 PRINTING AND DATA TRANSFER............................................................ 43
6.1 Data Transfer Adapter Basics............................................................................................ 43
6.1.1 Attaching the Data Transfer Adapter ....................................................................... 43
6.2 RS232 Connections............................................................................................................ 44
6.2.1 Setup and Use of the Printer..................................................................................... 45
6.2.2 Connecting to a Personal Computer......................................................................... 46
6.3 Sending Data to a Printer or Computer.............................................................................. 47
6.3.1 Sending Currently Displayed Data........................................................................... 47
6.3.2 Sending Recalled Data ............................................................................................. 48
6.3.3 Sending All Stored Data........................................................................................... 48
SECTION 7 MAINTENANCE............................................................................................. 51
7.1 Cleaning the Colorimeter................................................................................................... 51
7.1.1 Cleaning the Data Transfer Adapter......................................................................... 51
7.1.2 Sample Cells............................................................................................................. 51
7.2 Replacement Instructions................................................................................................... 51
7.2.1 Battery Replacement ................................................................................................ 51
SECTION 8 TROUBLESHOOTING .................................................................................. 53
8.1 Introduction........................................................................................................................ 53
8.1.1 Error Codes .............................................................................................................. 53
8.1.2 Beeper/Error Icon..................................................................................................... 54
8.1.3 Concentration Out of Range..................................................................................... 54
8.1.4 Low Battery.............................................................................................................. 54
GENERAL INFORMATION
REPLACEMENT PARTS AND ACCESSORIES ................................................................. 57
How To Order.......................................................................................................................... 59
Repair Service.......................................................................................................................... 60
Warranty .................................................................................................................................. 61
CERTIFICATION ...................................................................................................................63
2
Page 5
SAFETY PRECAUTIONS
Before attempting to unpack, set up or operate this instrument, please read this entire manual. Pay particular attention to all warnings, cautions and notes. Failure to do so could result in serious injury to the operator or damage to the equipment.
To ensure the protection provided by this equipment is not impaired, this equipment MUST NOT be installed or used in any manner other than that which is specified in this manual.
Use of Hazard Information
If multiple hazards exist, the signal word corresponding to the greatest hazard shall be used.
DANGER
Indicates either a potentially or an imminently hazardous situation which, if not avoided, could result in either death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate injury.
Precautionary Labels
Electrical equipment marked with this symbol may not be disposed of in European public dis­posal systems after 12 August of 2005. In conformity with European local and national regula­tions (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential injury. If on the instrument, refer to the instruction manual for operation or safety information.
NOTE
Information that requires special emphasis.
Please pay particular attention to labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed.
The DR/800 Series Colorimeters are Class 1 LED products. A Class 1 LED product has insufficient energy to be considered an eye hazard.
Section 2.1 Battery Installation
SECTION 4 CREATING USER-ENTERED PROGRAMS
3
Page 6
SPECIFICATIONS
Specifications subject to change without notice.
Wavelength Range(s):
Model DR/890: 420, 520, 560, 610 nm
Model DR/850: 520, 610 nm
Model DR/820: 520 nm
Wavelength Accuracy: ±1 nm
Wavelength Selection: Automatic
Photometric Linearity: ±0.002 A (0-1 A)
Photometric Reproducibility: ±0.005 A (0-1 A)
Photometric Accuracy: ±0.005 A @1.0 ABS Nominal
Source Lamp: Light Emitting Diode (LED)
Detector: Silicon Photodiode
Data Readout: 4-digit LCD, 1.5-cm Character Height
Readout Modes: % Transmittance, Absorbance, Concentration
External Outputs: IR (Infrared to RS-232 Serial using the Data
Transfer Adapter)
Battery Power: (4) AA alkaline cells
Instrument Dimensions: 23.6 x 8.7 x 4.7 cm (9.3 x 3.4 x 1.9
inches)
Instrument Weight: 470 g (1 lb.)
Photometric Range: 0-2 A
Stray Light: <1.0% at 400 nm
Battery Life: 6 months (typical)
Temperature Range:
Operating Range: 0 to 50 °C (32 to 122 °F) Storage Range: -40 to 60 °C (- 40 to 140 °F)
Humidity: 90% at 50 °C
Environmental: Designed to meet IP67 Standard; dustproof and
waterproof
4
Page 7
OPERATION
DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the necessary Material Safety Data Sheets and become familiar with all safety procedures before handling any chemicals.
DANGER
La manipulation des échantillons chimiques, étalons et réactifs peut être dangereuse. Lire les Fiches de Données de Sécurité des Produits (FDSP) et se familiariser avec toutes les procédures de sécurité avant de manipuler tous les produits chimiques.
PELIGRO
La manipulación de muestras químicas, estándares y reactivos puede ser peligrosa. Revise las fichas de seguridad de materiales y familiarícese con los procedimientos de seguridad antes de manipular productos químicos.
GEFAHR
Das Arbeiten mit chemischen Proben, Standards und Reagenzien ist mit Gefahren verbunden. Es wird dem Benutzer dieser Produkte empfohlen, sich vor der Arbeit mit sicheren Verfahrensweisen und dem richtigen Gebrauch der Chemikalien vertraut zu machen und alle entsprechenden Materialsicherheitsdatenblätter aufmerksam zu lesen.
PERIGO
A manipulação de amostras, padrões e reagentes químicos pode ser perigosa. Reveja a folha dos dados de segurança do material e familiarize-se com todos os procedimentos de segurança antes de manipular quaisquer produtos químicos.
5
Page 8
6
Page 9
SECTION 1 GENERAL DESCRIPTION
1.1 Instrument Description
The Hach DR/800 Series Colorimeter shown in Figure 1 is a microprocessor-controlled, LED-sourced filter photometer suitable for colorimetric testing in the laboratory or the field. The instrument is precalibrated for common colorimetric measurements and includes convenient calibration capability for user-entered and future Hach methods. Instrument features include:
Test results are displayed in concentration, absorbance,
or percent transmittance.
Automatic wavelength selection and ranging in the
preprogrammed parameters.
Data storage and recall for datalogging in the field
or laboratory.
Conversion of results to alternate forms for many parameters
(i.e., PO
, P2O5, P).
4
Reagent Blank Correction and Standard Adjust features may
be used to compensate for lot-to-lot variations in reagents.
Icon prompts displayed during testing.
A built-in timer to monitor specific reaction times called for
in the test procedures. Appropriate times are programmed into the calibration data for specific tests. The timer also can be used manually by the operator independent of the stored methods.
IR output for RS232 interface capability allows an external
printer or computer to interface with the colorimeter.
Entry of user-entered methods or new Hach methods.
Error signals for procedural or instrument troubleshooting.
The colorimeter operates on battery power. The instrument holds four AA-size alkaline dry cells (batteries supplied) that power the instrument for at least six (6) months.
7
Page 10
SECTION 1, continued
Figure 1 DR/800 Series Colorimeter Standard Package*
1.2 Unpacking the Instrument
Remove the instrument and accessories from the shipping container and inspect each item for any damage that may have occurred during shipping. Verify that all items listed on the packing slip are included. If any items are missing or damaged, please contact Hach Customer Service, Loveland, Colorado for instructions.
* Carrying Case may be ordered separately.
8
Page 11
SECTION 1, continued
Hach’s toll-free number for customers within the United States is 800-227-4224. For customers outside the United States, contact the Hach office or distributor serving you.
1.2.1 Standard Accessories
Sample Cells — (2) round, 1-inch pathlength, glass, with
COD/TNT Adapter for use with 16-mm vials used in COD
Batteries — (4) AA alkaline
Documentation package — includes Instrument Manual and
In addition to these standard accessories, several other optional accessories are available from Hach Company (refer to Replacement Parts and Accessories).
10-mL, 20-mL, and 25-mL marks; (2) 1-cm pathlength, plastic
and Test ’N Tube methods.
Procedures Manual.
1.2.2 Optional Accessories
Data Transfer Adapter (for RS232 interface)
HachLink™ Software
Instrument case
DR/Check™ ABS Standard
1.3 Description of the Keypad
Figure 2 shows the colorimeter’s keypad. The description and function of each individual key is given in Table 1 .
9
Page 12
SECTION 1, continued
Figure 2 Keypad
Table 1 Keys and Descriptions
KEY DESCRIPTION
Allows the user to select a program. Also used as a numeric key function when the “#” icon is illuminated at the base of the display screen.
Accesses the SETUP menu (the SETUP icon illuminates in the upper left-hand corner of the display screen). The setup menu provides access to options such as reagent blank, standard adjust, user-entered programs, and instrument configurations.
Also used as a numeric key function when the “#” icon is illuminated at the base of the display screen.
Prints currently displayed data. In the RECALL menu, prints recalled data. Also used as a numeric key function when the “#” icon is illuminated at the base of the display screen.
Use this key to turn the instrument on and, when the instrument is on and the EXIT icon is not illuminated, press this key to turn the instrument off.
When the EXIT icon is illuminated at the base of the display screen, the the current entry or selection.
EXIT key cancels
10
Page 13
SECTION 1, continued
Table 1 Keys and Descriptions (Continued)
KEY DESCRIPTION
Displays the current date. In the RECALL menu, displays the date the recalled sample was stored. Also used as a numeric key function when the “ base of the display screen. Used to set the current date from the SETUP menu.
#” icon is illuminated at the
Displays the current time. In the was stored. Also used as a numeric key function when the “#” icon is illuminated at the base of the display screen. Used to set the current time from the
When performing an analysis, this key displays the concentration value of the reading. Used as a toggle key to access alternate chemical forms, if available. Also used as a numeric key function when the “
Scrolls up through selected menus or stored data.
When performing an analysis, this key allows the user to store a current reading in one of 99 sample locations. The user can store the reading as numbers 1-99 by pressing ENTER. Use the up and down arrow keys to find unused storage numbers or use numeric keys to enter a sample number. Also used as a numeric key function when the “#” icon is illuminated at the base of the display screen.
Begins the retrieval of stored sample readings ( left portion of the screen). Also used as a numeric key function when the “ illuminated at the base of the display screen.
Toggles between displaying Absorbance and % Transmittance. Also used as a numeric key function when the “
#” icon is illuminated at the base of the display screen.
RECALL menu, displays the time the recalled sample
SETUP menu.
#” icon is illuminated at the base of the display screen.
RECALL icon illuminates in the upper-
#” icon is
Scrolls down through selected menus or stored data.
Zeros the instrument on the current sample blank. Also used as a numeric key function when the “
When the READ icon is illuminated at the base of the display screen, this key reads and displays the sample concentration. Also used as a numeric key function when the “ is illuminated at the base of the display screen; the first press is a decimal, the second press toggles the value sign.
If using a Hach-stored program, the TIMER key automatically sets the appropriate reaction time. If not in a Hach-stored program, the TIMER key allows the user to set a timer. When the “ most recent level of action (deletes the whole entry, not just the last number).
Within a menu, selects the displayed menu item. During numeric entry, accepts the displayed value.
#” icon is illuminated at the base of the display screen.
#” icon
#” icon is illuminated at the base of the display screen, CE clears the
11
Page 14
SECTION 1, continued
1.4 Display Screen in Function and Numeric Modes
The main display operates in two modes: function mode and numeric mode. The user does not select the mode, changeover is automatic depending on the options selected, where the user is in an analysis, and what information the instrument needs from the user.
The main display shows action icons ( displayed below the horizontal line. This shows there are two available options to select (zero the instrument, or take a reading) at this point in the analysis.
The numeric mode is signified by the “ the horizontal line. In the numeric mode, some function keys act as numeric entry keys (corresponding to the number on the key).
1.5 Icon and Display Screen
Figure 3 shows the icons displayed by the DR/800 Series Colorimeters. Tabl e 2 provides a brief description of each display element.
Figure 3 Icon and Main Display Screen
ZERO and READ)
#” icon illuminated below
12
Page 15
SECTION 1, continued
Table 2 Main Display Screen Icons
ITEM NO. DEFINITION
1 Indicates the user is in the SETUP menu.
2 Indicates the user is in the
3 The sample concentration exceeds the limits of the selected program.
4 Sample cell icon. Indicates a reagent blank adjustment is in use for the current program.
5 Standard adjust icon. Indicates a standard adjust is in use for the current program.
6 Indicates a low battery condition - replace the batteries as soon as possible.
7 These three icons follow the sample reading and represent either absorbance, percent
transmittance, milligrams per liter, micrograms per liter, or grams per liter.
8 Illuminates whenever the numbers in the main display (22) or the sample display (11) refer
to a sample number.
9 Indicates the instrument is waiting for information from the user.
10 Depending on the arrow(s) illuminated, these icons indicate the available scroll direction
(using the ARROW keys) for accessing options.
11 In the RECALL menu or when storing data, these digits show the selected sample number.
12 Most icons displayed in this area are action icons. Action icons tell the user what actions
are acceptable options during an analysis.
13 Indicates an invalid key press was made. This icon flashes briefly accompanied by a short
beep.
14 Exit action icon - (when illuminated) tells the user that pressing the EXIT key to exit the
current level of action is an acceptable option.
RECALL menu.
15 Enter action icon - (when illuminated) tells the user that pressing the ENTER key to confirm
an action is an acceptable option.
16 Numeric entry action icon - (when illuminated) tells the user the numeric key pad is active.
17 Timer action icon - (when illuminated) tells the user that the instrument is presently running
a timer. This icon will flash while the timer is counting.
18 Read action icon - (when illuminated) tells the user that pressing the READ key to read the
sample cell is an acceptable option.
19 Zero action icon - (when illuminated) tells the user that pressing the ZERO key to zero the
instrument on a sample cell is an acceptable option.
20 Depending on the currently active menu, the series of alphabetical letters displayed here
gives information on a current reading, a stored reading, indicates the options available within a menu, or prompts the user for the next action.
21 Shows the active program number, (either a user-entered (101-110) or Hach-stored
program (1-100). The program number is displayed immediately below the PRGM icon.
22 Depending on the currently active menu, the numbers displayed here represent either the
sample reading, the clock timer, or the numeric characters entered by the user.
13
Page 16
14
Page 17
SECTION 2 INSTRUMENT SETUP
2.1 Battery Installation
Power is supplied by four AA-sized alkaline batteries. Typically, a set of batteries provides approximately six months of operation. The colorimeter lamp is an LED and is on only long enough for the measurement sequence to take place (approximately 2 seconds).
The instrument will automatically shut off if no keystrokes are made for 15 minutes when in normal mode and four hours when in user-entry mode.
Figure 4 provides an exploded view of the battery installation. When replacing discharged batteries, always replace the complete set of four.
Hach recommends using alkaline batteries in this instrument. Do
not use rechargeable Nickel Cadmium (NiCad) batteries.
Note: For performance reasons, never remove the battery cover from this product except when servicing the batteries.
DANGER Use of nickel-cadmium batteries under a fault condition creates a potential fire hazard.
The battery compartment is accessible from the underside of the instrument. Make sure the sample cell compartment is empty. Lay the instrument upside down on a padded surface, and install batteries as follows:
1. Disconnect the Data Transfer Adapter (if connected) from
the instrument.
2. Loosen the two battery compartment screws and remove the battery compartment door as shown in Figure 4.
3. Install four AA alkaline batteries in the battery holder as shown in Figure 4. Match the polarities of the batteries with the polarity markings in the battery compartment.
4. Replace the battery compartment cover, tighten the screws, and return the instrument to the upright position.
PELIGRO La utilización de pilas de níquel-cadmio en condiciones de falla crea el riesgo de incendio.
15
Page 18
SECTION 2, continued
PERIGRO O uso de baterias de níquel-cádmio em condição de falha cria a possibilidade de incêndio.
DANGER L'utilisation de batteries nickel-cadmium dans des conditions inappropriées crée un risque d'incendie.
GEFAHR Unter einer Störungsbedingung stellt die Verwendung von Nickel­Kadmium-Batterien eine Feuergefahr dar.
Figure 4 Battery Installation
16
Page 19
SECTION 2, continued
2.2 Turning the Instrument On
Once batteries are installed, turn the instrument on using the
EXIT/I/Okey (located on the top row, far-right column of the
instrument keypad).
Press the key once to power-on the instrument. The display will show the software version number, then will default to the last used program number. The instrument is now ready for operation.
2.3 Setting the Date and Time
Setting the instrument’s date and time allows sample readings to be stored and recalled with the proper date and time. Check the currently entered date or time by pressing the respective
TIME key.
To set the date and time, continue with Section 2.3.1 or
2.3.2 below.
DATE or
2.3.1 Entering the Correct Date
Check the current date by pressing the DATE key. If the date is incorrect, follow the procedure below to change it.
Enter the correct year, then the correct month and day as follows:
1. Press the
2. Access the
keypad. (The down display is shown.)
3. Press the down
4. Press the
5. Four horizontal lines (showing available spaces for numeric
entry) and corresponding to the correct year using the numeric keypad. For instance, if the year is 1997, press
ENTER key.
I/O key to turn the instrument on.
SETUP menu by pressing the SETUP key on the
ARROW icon on the right side of the
ARROW key until DATE is displayed.
ENTER key to select the date option.
YEAR ? will appear on the display. Enter the digits
1997 then the
If an incorrect number is entered, press the enter the information.
17
CE key and re-
Page 20
SECTION 2, continued
Next, the instrument prompts for the month and day.
6. Enter the correct month and day using the numeric keypad.
The month must be entered first, followed by the day. If an incorrect number is entered, press the the information.
Note: When entering a one digit month or day, always press the ZERO
key before the digit. For example: If the month and day to be entered is March 4, press
ENTER key to accept.
7. Press the ENTER key to accept the new information. Press the
EXIT key to return to the main menu
2.3.2 Entering the Correct Time
1. Press the I/O key to turn the instrument on.
CE key and re-enter
0304 on the numeric keypad, then press the
2. Access the
keypad. (The down
SETUP menu by pressing the SETUP key on the
ARROW icon on the right side of the
display is shown.)
3. Press the down
4. Press the
ARROW key until TIME is displayed.
ENTER key to select the time option.
5. Enter the time in 24-hour (military) notation using the
numeric keypad then press the For example, 9:00 a.m. is entered as p.m. is entered as entered, press the
1400 ENTER. If an incorrect number is CE key and re-enter the information.
6. The display returns to the setup menu. Press the
ENTER key to accept the entry.
0900 ENTER and 2:00
EXIT key to
return to the main menu.
18
Page 21
SECTION 2, continued
2.4 Sample Cell Insertion
Wipe the sample cell with a lint-free cloth or tissue and insert the cell into the sample cell compartment with the diamond-shaped marker toward the keypad.
Figure 5 Placing Samples in the Sample Cell Compartment
2.5 Sample Cell Adapter Installation
When installing an adapter into the sample cell compartment, insert the adapter into the cell compartment and rotate until it drops into the alignment slots. Finish installation by gently pushing down on the adapter until it snaps into position.
Place the appropriate vial or sample cell into the adapter - the vial or sample cell should fit well into the adapter. If it does not, double-check that the correct sample container (vial or sample cell) is being used and that the adapter is installed correctly.
2.5.1 Using the 16-mm COD/Test ‘N Tube Vial Adapter
The methods for chemical oxygen demand (COD) and Test ‘N Tube (TNT) determinations in the colorimeter procedures manual use 16-mm vials as the sample cell for the colorimetric measurement. This adapter also holds a standard 16-mm test tube.
Place the COD/TNT Adapter in the instrument's sample cell compartment as instructed in Section 2.5, above. Place the vial into the COD/TNT Adapter (see Figure 6).
Always place the instrument cap over the adapter when measuring in bright sunlight (see Figure 7).
19
Page 22
SECTION 2, continued
Figure 6 Installing the COD/TNT Adapter
2.6 Using the Instrument Cap as A Light Shield
The colorimeter cap is removable (by sliding it away from the keypad) to expose the sample cell compartment. The instrument cap also functions as a light shield. Samples may be analyzed without the light shield in place but using the light shield provides a good seal against stray light and contributes to more accurate results. Use of the light shield is especially important when performing an analysis in bright light or direct sunlight.
To use the instrument cover as a light shield during measurements, place the cap over the sample cell, into the grooved marks on the instrument. See Figure 7.
Figure 7 Installing the Light Shield
20
Page 23
SECTION 3 INSTRUMENT OPERATION
3.1 Menus
The instrument has two important menus which allow access to different options. The menus are:
Setup Menu
Recall Menu
Once the user has selected the desired menu, the arrow icons illuminate on the main display screen. The arrows indicate that additional options are available in this menu. Press the up
ARROW key (whichever arrow is illuminated) to scroll until
down the desired option is displayed. A press of the time selects the displayed option.
ENTER key at this
or
3.1.1 Setup Menu
When choosing options from menus, the up or down keys, within menu options. Use the Press
ENTER, and EXIT keys help navigate between menus and
ENTER key to select a menu option.
ENTER again to accept a new setting. Pressing the EXIT key
ARROW
leaves a menu or leaves the displayed item unchanged.
Enter the Setup Menu from the Main Menu by pressing the
SETUP key.
Refer to the arrow icons on the main display for an indication of which key (up
ARROW or down ARROW) to press on the keypad -
the direction of the illuminated icon on the main display indicates the available scrolling direction. The up and down
ARROW
(scroll) keys move from one menu option to the next.
Example: when the down arrow icon is illuminated on the display screen, the user only has the option to press the down
ARROW key
to move from one option to the next. Once the user scrolls down, the down arrow icon changes to a dual-direction arrow icon (up and down arrows) until the user reaches the last menu option available in that menu.
The following instrument operating features are available in the
Setup menu
BLANK - Used to compensate for color contributed by the
reagents in a reagent blank. Readjust for each new lot of
21
Page 24
SECTION 3, continued
reagents. Press the ENTER key to activate this option. The default setting is off. Chemistries which zero on a reagent blank do not have this option available.
STD - Standard Adjust option allows for entry of the value of
the prepared standard. Press the option. The default setting is off.
ENTER key to activate this
Note: The BLANK and/or STD options may or may not be available, depending on the selected Hach program; the BLANK and STD options are always available for user­entered programs.
PRINT (ALL) - Prints all stored data to a printer or
downloads it to a personal computer. The information printed includes: sample concentration reading, date of sample, time of sample, units, sample number, program number, Absorbance, and %T. Refer to Section SECTION 6 for detailed printing information and to Figure 10 for an example of a printout. Press the
ENTER key to activate this option.
USER - Allows access to the User-Entered Program Menu.
Press the New User-entered Program on page 33 for more information.
ENTER key to activate this option. See Creating a
DATE - Allows the user to set the date. Press the ENTER key
to activate this option. See Section 2.3.1 for more information.
TIME - Allows the user to set the time. Press the ENTER key
to activate this option. See Section 2.3.2 for more information.
ERASE (ALL) - This clears all the data previously stored in
memory. Press the
3.1.2 Recall Menu
The recall menu allows access to stored data. Complete information and instructions for the use of this menu is presented in SECTION 5 DATA RECALL AND STORAGE.
3.2 Performing an Analysis
The procedures manual provides illustrated, step-by-step procedures for performing all the factory-entered methods. This instrument manual has supplemental information on how the instrument performs the necessary functions, and how to use the special operating features. Once you are familiar with the
ENTER key to activate this option.
22
Page 25
SECTION 3, continued
instrument, the instructions in the procedure manual should be sufficient to analyze your samples.
Colorimetric testing with preprogrammed calibrations can be divided into four general phases:
1. Colorimeter setup
2. Sample preparation
3. Zeroing the instrument
4. Measuring the prepared sample
The following sections describe each phase in detail.
3.2.1 Colorimeter Setup for Sample Analysis
The colorimeter setup using a Hach program begins by selecting the desired program number. Programs numbers can be found in the individual procedure. (See the DR/800 Series Procedures manual supplied with your instrument.) Prompting icons will appear to indicate which keys are acceptable for the user to select. After turning the instrument on, the main display shows information from the last program used before the instrument was turned off.
If a different program is desired, press the the desired program number with the numeric keypad. The instrument will recall that program.
If the number selected is not valid, an error signal sounds and the display momentarily flashes an error icon. The display returns to the prompt for the program number. Re-enter the correct program number. Only select chemistries are available on the DR/820 and DR/850. See APPENDIX A AVAILABLE PARAMETERS AND RANGES on page 65.
When the program number has been successfully entered, the display immediately prompts the user to zero the instrument by illuminating the
ZERO action icon.
23
PRGM key and enter
Page 26
SECTION 3, continued
3.2.2 Sample Preparation
For colorimetric tests, sample preparation is next. The zero solution (or blank) and sample are usually prepared at this time. Generally, sample preparation consists of adding the contents of a premeasured reagent pillow to a small volume of sample. Follow the instructions in the procedure specific to your analysis.
It is important to observe the waiting period specified in the test procedure to ensure the color from the reaction of the reagent(s) with the target analyte develops fully. Many procedures also give a maximum time limit after which the color may begin to fade.
The instrument has method-specific timers for color development times pre-programmed into the software. The user is notified when the time has elapsed by a series of short beeps.
3.2.3 Zeroing the Colorimeter
The instrument must be zeroed for each test or series of tests to establish a zero reference for the measurement. This is done by placing a solution recognized as the blank solution in the cell holder, covering the sample with the instrument cap, and pressing
ZERO key. The next prompt will display zeros and illuminate
the
READ action icon. The instrument is now ready to take the
the first sample reading.
Note: Once the zero reference point has been established, several
samples can be measured consecutively by placing each of them into the cell holder and pressing the be re-zeroed at any time by placing the zero solution (blank) into the instrument and pressing the
3.2.4 Measuring the Prepared Sample
When ready to take the reading, place the prepared sample in the sample compartment. For best results orient the sample cell consistently for each measurement; see Figure 5 on page 19. Place the instrument cap (light shield) over the sample cell and press the displayed.
Toggle between absorbance or percent transmittance values by successive presses of the restore the concentration display. Successive presses of the CONC
READ key. After a brief pause, the results will be
READ key. The instrument can
ZERO key.
ABS %T key. Press the CONC key to
24
Page 27
SECTION 3, continued
key toggles between alternate forms, if any. See Appendix A for currently available chemistries and their alternate forms.
3.2.5 Alternate Chemical Forms
Many Hach programs provide alternate chemical forms for the measured parameter. Press the alternate forms after the measurement is displayed. Each press of the
CONC key takes the user to the next alternate chemical form.
For example: In Program #1, mg/L Al may also be displayed as mg/L Al returns to its original form and reading.
3.2.6 Using the Timer
Many Hach test methods use one or more timers which are pre­programmed into the DR/800 Series Colorimeters. When the procedure instructs you to, press the interval. Press beeps will sound at the end of a timer interval. If the method requires additional timers, the instrument will automatically display the next timer when the first timer elapses. Press to start the next timer.
2O3
CONC key to scroll through these
. If alternate forms are not available, the instrument
TIMER key to display a timer
ENTER to start the timer count-down. Several
ENTER
To zero the instrument on the blank while the timer is running, press the
EXIT key. The timer icon will continue to flash,
indicating the timer is running. Press available functions within the method (such as blank correction) while the timer is running. To return to the timer display, press the
TIMER key.
3.2.6.1 Using the Timer in Manual Mode
The manual timer function allows the operator to use the timer independently from the method timer. Make sure the instrument is not in numeric entry mode and activate the timer with a press of
TIMER key. If using a pre-programmed timer, press the TIMER
the key again.
#” action icon is illuminated, to indicate that the numeric
The “ keypad is active. Enter the desired time using the numeric keypad. For example, to enter 2 minutes, press To enter 12 minutes, press momentarily show the entered time, then the timer countdown
ZERO, or perform other
200, then ENTER.
1200, then ENTER. The display will
25
Page 28
SECTION 3, continued
will begin. The display will show the remaining time. At the end of the elapsed time, the instrument sounds five beeps.
3.2.6.2 Stopping the Timer
Stop the timer at any point in the countdown by pressing the
ENTER key. If a pre-programmed timer was in use, the full timer
period will be displayed. Press the count down. If a manual timer was in use, the display for entering a new length of time will be shown, see Section 3.2.6.1 Using the Timer in Manual Mode. Press
3.3 Reagent Blank Correction
The Reagent Blank Correction can be used with some of the factory-entered methods. It subtracts the color absorbed when running the test with deionized water instead of sample. The blank value is subtracted from every result to correct for any background color due to reagents.
ENTER key again to resume
EXIT to leave the timer mode.
When using the Reagent Blank Correction feature, the blank correction should be entered before the Standard Adjust feature is used.
To enter a programmed correction for the reagent blank:
1. Run the test using deionized water with each new lot
of reagents.
2. Press
READ to obtain the blank value; record the value for use
in step 4..
3. Press
will show
SETUP, scroll to BLANK and press ENTER. The display
BLANK?.
4. Enter the blank value obtained in step 2..
5. Press
ENTER to accept the value as the blank to be subtracted
from each reading.
6. The display will show 0.00 mg/L (resolution and units vary
with the method) and the sample cell icon will be displayed, indicating that the reagent blank feature is enabled and the blank value will be subtracted from each reading. Repeat the reagent blank adjust for each new lot of reagents.
26
Page 29
SECTION 3, continued
Note: After entering a reagent blank adjust, the display may flash “limit”
when zeroing if the sample used for zeroing has a lower absorbance value than the reagent blank.
To disable the Reagent Blank adjust feature, press SETUP, scroll
BLANK and press ENTER twice. The concentration readings
to will be displayed without subtracting the blank. The sample cell icon will no longer appear in the display.
Do not attempt to use the Reagent Blank Adjust feature if the procedure uses a reagent blank for zeroing.
3.4 Adjusting the Standard Curve
The DR/890 Colorimeter has over 90 Hach Programs permanently installed in memory (other models have fewer programs). A program usually includes a preprogrammed calibration curve. Each curve is the result of an extensive calibration performed under ideal conditions and is normally adequate for most testing. Deviations from the curve can occur from using compromised testing reagents, defective sample cells, incorrect test procedure, incorrect technique, or other correctable causes. Interfering substances or other causes may be beyond the analyst’s control.
In some situations, using the preprogrammed curve may not be convenient:
Running tests where frequent calibration curve checks
are required.
Testing samples which give a consistent test interference.
Consider the following before adjusting the calibration curve:
1. Will future test results be improved by adjusting the curve?
2. Are interfering substances consistent in all the samples that
you will test?
Any precision and test range information provided with the procedure may not apply to an adjusted curve calibration.
You can adjust many of the calibration curves by following the steps found in the test procedures. Working carefully is
27
Page 30
SECTION 3, continued
important. After the adjustment, it is wise to run standard solutions of several concentrations to make sure the adjusted curve is satisfactory. Perform standard additions on typical samples to help determine if the adjusted curve is acceptable.
Think of the standard adjust measurement as a two-step process. First, the instrument measures the sample using the preprogrammed calibration. Second, it multiplies this measurement by an adjustment factor. The factor is the same for all concentrations. The instrument will remember the factor indefinitely and will display the standard adjustment icon when it is used.
Adjust the calibration curve using the reading obtained with a Hach Standard Solution or carefully prepared standard made from a concentrated Hach Standard Solution. It is important to adjust the curve in the correct concentration range. For most purposes, Hach recommends adjusting the curve using a standard concentration that is 70 to 85% of the maximum concentration range of the test.
For example, the Hach preprogrammed method for fluoride has a range of 0-2.0 mg/L F. To adjust the calibration curve, use a standard with a concentration between 1.4-1.6 mg/L. Hach provides a 1.60 mg/L Fluoride Standard Solution (75% of the full range). This is a convenient standard to use for adjusting the calibration curve.
If the range of all your samples is known to be below a concentration that is less than 50% of the full range (50% of 2.0 is
1.0 mg/L), then adjust the standard curve with a standard that is within that range.
For example, if all the samples contain 0.6-0.9 mg/L F, you may use a 1.00 mg/L fluoride standard to adjust the curve. You may use the 1.00 mg/L standard because it is closer to the sample range you are working with.
If you are using a Reagent Blank Correction, the blank correction should be entered before the standard curve is adjusted.
To adjust the standard curve:
28
Page 31
SECTION 3, continued
1. Prepare the standard.
2. Use the standard as the sample in the procedure.
3. When the reading for the standard is obtained, press
4. Use the
5. Press
ARROW keys to scroll to the “STD” setup option.
ENTER to activate the standard adjust option.
SETUP.
6. Enter the concentration of the standard used.
7. Press
ENTER. The standard adjust icon will be displayed,
indicating that the curve has been adjusted with the standard.
Note: If the attempted correction is outside the allowable adjustment
limit, the instrument will beep and flash the error icon and the operation will not be allowed.
To eliminate the standard curve adjustment:
1. Press the
2. Enter the stored-program number and press
PRGM key.
ENTER.
3. Press the
4. Press an
Note: On will be displayed if an adjustment is in use.
SETUP key
ARROW key to display STD.
5. Press ENTER twice.
3.5 Using a Programmed Method
Hach programmed and user-entered program options are available on this instrument. The Procedures Manual supplied at the time of purchase contains all currently available Hach­programmed methods. Up to ten user-programmed methods may be entered into the instrument. See Creating a New User-entered Program on page 33 for instructions for this function. Tabl e 3 shows the components of typical factory programmed test method.
29
Page 32
SECTION 3, continued
Step Action/Keystroke Display
Tabl e 3
1. Turn power on. Press the EXIT/I/O
2. Select the Program
number to be used.
3. As needed, set, then start the TIMER.
4. ZERO the instrument using the sample blank.
Press the PGRM key, then enter the program number and press the ENTER
Press the TIMER key. In a Hach­stored program, the timer automatically defaults to the appropriate reaction time.
Press ENTER to start the timer.
Insert blank and press ZERO
key.
key. The instrument defaults to and
displays the initial screen of the last program used.
EXAMPLE: If program 20 was the last program in use, the instrument will automatically recall program 20.
After the PGRM key is pressed, a blinking cursor appears with a question mark. Enter the desired program number. Press ENTER to confirm this action and recall the desired program number.
The entered or programmed reaction time will display and count down to zero.
. After the ZERO key is pressed,
the instrument zeros on the sample cell.
5. Obtain reading in concentration, absorbance, or % transmittance.
Place the prepared sample into the cell holder. Press READ.
3.6 Quality Assurance
Verify the performance of your instrument in seconds using Hach DR/Check™ ABS Secondary Standards (Cat. No. 27639-00). These gel standards provide a measure of instrument absorbance for use anywhere, anytime, as often as laboratory quality assurance procedures suggest. The kit includes a blank and one standard each of a low-, mid-, and high range absorbance value from 0 to 2 ABS, for use with any DR/800 Series Colorimeter.
The instrument reads the sample and displays the results.
30
Page 33
SECTION 4 CREATING USER-ENTERED
PROGRAMS
DANGER This instrument is not intended for use with flammable samples or those containing hydrocarbons.
PELIGRO Este instrumento no está destinado para uso con muestras inflamables o que contengan hidrocarburos.
PERIGO Este instrumento não é feito com o fim de ser empregado com amostras inflamáveis ou aquelas que contêm hidrocarbonetos.
DANGER Cet instrument n’est pas conçu pour une utilisation avec des échantillons inflammables ou des échantillons contenant des hydrocarbures.
GEFAHR Dieses Gerät darf nicht für Tests mit brennbaren Proben oder Proben, die Kohlenwasserstoffe enthalten, benutzt werden.
The DR/800 Series Colorimeters can store the calibration information needed to read prepared samples from up to ten different user-entered programs. To create a new, user-entered program, you will need a blank and prepared standards or the correct absorbance readings for each standard. The prepared standards are made with standard solutions of the parameter (i.e. analyte). Up to twelve concentrations of standards, including a zero concentration standard, may be used.
The absorbances of the prepared standards must be different from one another. If the colorimeter detects a duplicate, it will beep and ignore the latest reading.
Using a previously entered method’s program number erases all the previously entered information stored under that program number.
While creating a user-entered program, the colorimeter remains on for four hours following any keypress. If more than four hours pass between keypresses, the instrument will power down. All data which was entered, but not yet stored, will be deleted. The user-entered program must be re-entered from the beginning.
31
Page 34
SECTION 4, continued
4.1 User-Entered Programs
The instrument allows storage of up to ten user-entered programs (101–110) and up to 113 Hach programs. A minimum of two data points are required for the instrument to recognize and accept a user-entered program.
Program numbers 101 through 110 are reserved for storing
user-entered programs.
The maximum number of data points that can be entered for a
method is 12. After the twelfth standard (1 through 12) is accepted, the instrument stores the method and will not accept any more data, but will allow the user to review the data already entered.
Before entering a calibration, determine the optimum wavelength, timing sequences (if any), and the workable range of the method.
4.2 Calibration Curves
Calibration curves may have positive or negative slopes, but they must be based on absorbance (% transmittance not allowed) and must pass through the origin that represents zero concentration.
It is important that the standards adequately describe the curve over the range of interest. Because this is largely dependent on the shape of the curve, it may be necessary to prepare a preliminary curve using extra data points to help select the appropriate standards.
If the curve is linear, only two concentration data points are needed. For example, standards with a zero absorbance and a standard with 1.000 absorbance are appropriate. If the curve is nonlinear, additional data points are needed to achieve good accuracy. Up to 12 data points can be entered for a single calibration curve.
4.3 User-entered Program Information for Bleaching Chemistries
Although the majority of colorimetric test procedures produce a higher absorbance (i.e. deeper color) as the concentration of the parameter being measured increases, some tests do the opposite. These bleaching chemistries (fluoride is one example) produce a
32
Page 35
SECTION 4, continued
lighter color at increasing concentrations. The zero concentration standard is usually produced by combining deionized water with the reagents. Commonly this solution is used to zero the instrument as in Step step 14 of Section 4.4 Creating a New User- entered Program.
Once the zero is entered, the prepared standards must be read from lightest to darkest. In the case of bleaching chemistries, the absorbance values reported by the colorimeter may be negative.
Even when your test produces a lower absorbance (lighter color) with increasing concentration, the prepared standards must be read by the colorimeter in the order of increasing absorbance (i.e., from colorless or the palest color, to the deepest color). The instrument will not accept standards read out of order.
4.4 Creating a New User-entered Program
Use the step-by-step instructions below to enter a new user­entered program into instrument memory. Terminate at any point (before the program is stored) by pressing the
EXIT key until the
display is blank. The colorimeter will not retain any of the entered data.
1. Press the
2. Press the
I/O key to turn on the instrument.
SETUP key. The display will show SETUP in the
upper-left and the down-arrow icon in the lower-right. Available action functions are also shown.
3. Press the down
4. Press the up
5. Press the
ARROW key until USER is displayed.
ARROW key if the display goes past USER.
ENTER key. Four horizontal lines (numeric entry
display) will be displayed.
6. Select a program number from 101 through 110 by pressing
the corresponding digit key. The number will appear in the display.
Note: Press CE to correct errors.
7. Press ENTER. A wavelength and nm will be displayed.
33
Page 36
SECTION 4, continued
If the wavelength is correct as displayed, skip to
Step step 8.
Some instrument models can use different wavelengths.
If a different wavelength is preferred, proceed as follows:
a. Press
ENTER. A flashing question mark will be displayed
in the lower-right corner.
b. Press either
ARROW key until the preferred wavelength is
displayed.
c. Press
ENTER to accept the displayed wavelength. The
down-arrow icon will be displayed.
8. Press the
down ARROW key to move to the RES (resolution)
option. One to four zeros, a decimal point if needed, and the units of concentration may be modified here.
If the displayed resolution and units are correct, skip to
step step 9.
If the displayed resolution or units are incorrect for your
test, proceed with the following:
a. Press
displayed.
ENTER. A flashing question mark will be
b. Press either
ARROW key until the preferred resolution
and concentration units are displayed. The available options are:
0.000 0.00 0.0 0
0.000 Pg/L 0.00 Pg/L 0.0 Pg/L 0 Pg/L
0.000 mg/L 0.00 mg/L 0.0 mg/L 0 mg/L
0.000 g/L 0.00 g/L 0.0 g/L 0 g/L
c. Press
9. Press the down
ENTER. The question mark will disappear.
ARROW key to scroll to STD. STD and the
number of the standard (i.e., 1 is shown for the first standard, 2 for the second, etc.) will be shown on the lower portion of the display.
34
Page 37
SECTION 4, continued
10. Press ENTER. Four horizontal lines (denoting numeric entry)
will be displayed.
11. Enter the standard’s concentration, using the numeric entry
keys (the # icon will be illuminated on the display).
Note: Press the CE key to correct errors.
12. Press the ENTER key. The concentration will be displayed.
Note: A beep means that the concentration is a duplicate of a previous
standard or the concentration is too high for the selected resolution. Repeat step step 11 with a different concentration and continue.
13. Press the down ARROW key. ABS will be displayed followed
by the number of the standard.
14. The colorimeter requires one zero be entered in this
procedure; the
ZERO action icon will appear in the lower
portion of the display. Place a blank into the cell holder and press the disappear, across the display. The
ZERO key. Four horizontal lines will appear, then
READ action icon will
appear in the lower portion of the display.
Note: If necessary, the colorimeter can be re-zeroed. The most recently
entered zero will be used for subsequent readings.
15. Prepare the standards using the same reagents and procedure
used to test samples.
16. Place the prepared standard into the cell holder.
17. Press the
Note: Or, press the ENTER key to input an absorbance value or change
the value read by the instrument. Use the numeric keys to enter the value then press the ENTER key.
Note: A beep indicates that the absorbance is a duplicate of a previously
entered standard or that it falls between two previous standards. Repeat steps step 15 through 17. with the correct standard, or press the up ARROW key and repeat steps 9. through 17. with a correct, prepared standard and blank.
READ key. An absorbance value will be displayed.
18. Press the down ARROW key to advance to the next standard.
19. Repeat step 9 through step 18 for all remaining standards.
35
Page 38
SECTION 4, continued
20. Press the EXIT key once. STORE ? will be displayed.
21. Press the
ENTER key to store the new method in the
instrument’s memory.
4.5 Reviewing and Editing User-Entered Programs
Note: When a user­entered program is edited and stored, all stored data associated with that program is erased.
All method information previously stored by the operator can be reviewed and changed to add, delete, or modify data points. At any point during the editing function, the operator can terminate the procedure and exit by pressing the
EXIT key. No changes to the
program will occur.
Because the standards must be read in the order of increasing absorbancy, data points may not be inserted into the middle of an existing user-entered program.
To review and edit previously stored user-entered programs:
1. Press the
2. Press the
I/O key to turn the instrument on.
SETUP key.
3. Scroll to the
USER option and press ENTER.
4. Enter the program number of the method to review or edit
and press the
5. Scroll through the calibration information using the
keys. To avoid making changes, press the
6. To edit the data shown on the display, press
necessary changes, then press the
ENTER key.
ARROW
EXIT key.
ENTER. Make
ENTER key to return to
reviewing the data.
7. Press
8. Press
EXIT once. STORE? will be displayed.
ENTER to store the program.
36
Page 39
SECTION 4, continued
4.6 Erasing User-entered Programs
Note: When a user­entered program is erased, all stored data associated with that program is also erased.
User-entered programs are automatically deleted when another user-entered program is entered and stored in the previously entered method’s storage number (101–110). They also may be erased from the instrument memory as follows:
1. Press the
2. Press the
3. Scroll to the
I/O key to turn the instrument on.
SETUP key.
USER option and press ENTER.
4. Enter the program number of the method to be erased and
then press
ENTER.
5. Scroll to the concentration data for STD 1 using the down
ARROW key. Press ENTER.
6. Press
7. Press the
CE. Press ENTER.
EXIT key. ERASE? will be displayed.
8. Press the
ENTER key to erase the method or the EXIT key to
retain it in memory.
37
Page 40
38
Page 41
SECTION 5 DATA RECALL AND STORAGE
To store sample data, press STORE after the sample measurement is displayed. Data must be stored if you wish to recall it later for review, downloading or printing. The following information is stored for each sample:
instrument model
instrument serial number
chemical form
concentration
units
absorbance
%T
5.1 Recalling Data
date
time
sample number
program number
After the shown in the lower-right portion of the display. In the center of the display, the next available storage number will appear. If this storage location is acceptable, press
To select an empty storage number (between 1 and 99), use the
ARROW keys to scroll to the desired number or enter the desired
number with the numeric keys. Press instrument will store the data, then revert back to the measurement display.
STORE key is pressed, a flashing question mark icon is
ENTER to accept it.
ENTER to accept. The
To recall data stored in the colorimeter, press the RECALL key.
Use the numbers containing stored sample reading data (of the available storage numbers 1-99) are displayed in the RECALL menu. For
ARROW keys to scroll through the stored data. Only data
39
Page 42
SECTION 5, continued
example, if the user has stored data into storage numbers 6, 10 and 15, those are the only accessible numbers in the Recall option. Available storage numbers are not shown as a Recall option because no data has been stored for recall purposes.
When a number of sample readings have been stored and a specific reading must be recalled, proceed as follows:
1. Press the
2. Press the
I/O key to turn the colorimeter on.
RECALL key to access the RECALL Menu.
3. A beep indicates that no data is presently stored.
4. Press the number key(s) or either
ARROW key on the keypad
to scroll to the desired sample number.
5. Press the
ENTER key. The stored reading will be displayed.
6. A beep indicates that no data is stored as that sample number.
7. While the stored reading is displayed, press the
TIME key to display the date or time the reading was
the stored. Press the
CONC key to display the concentration data.
8. If other stored data is desired, press either
DATE key or
ARROW key until
the data is displayed.
9. Press
EXIT to terminate data recall.
5.2 Erasing All Stored Data
Stored data can be erased and the memory of the instrument cleared using the procedure that follows.
1. Press the
2. Press the
3. Scroll using the down
appears on the display.
4. Press the
the instrument.
I/O key to turn the colorimeter on.
SETUP key to enter the SETUP menu.
ARROW key until ERASE and ALL
ENTER key to confirm this selection with
40
Page 43
SECTION 5, continued
A flashing question mark icon will appear in the lower-right portion of the screen as an additional step to avoid erasing all data if the choice was made erroneously.
5. Confirm this action by pressing the
not the desired action, press the
After the
ENTER key is pressed, the instrument automatically
ENTER key or, if this is
EXIT key.
erases all stored data and returns to the last used program.
41
Page 44
42
Page 45
SECTION 6 PRINTING AND DATA TRANSFER
6.1 Data Transfer Adapter Basics
The optional Data Transfer Adapter (DTA) utilizes infra-red technology to receive data from the DR/800 Series Colorimeter and then transfer the signal into RS232 format. The DTA then sends the RS232 signal to a printer or personal computer.
The sleeve design makes the adapter compact, easy to use, and robust. Samples may be tested while the adapter is in place and the information may be immediately printed or downloaded to a computer. Data stored in the instrument memory may also be printed or downloaded at any time.
6.1.1 Attaching the Data Transfer Adapter
The Data Transfer Adapter is designed to fit on the instrument in the same way the instrument cap does. To install the DTA, simply remove the instrument cap, then slide the DTA onto the colorimeter body until it snaps into place. The DTA design allows the instrument cap to be used as a light shield. See Figure 8 Installing the Data Transfer Adapter.
Figure 8 Installing the Data Transfer Adapter
43
Page 46
SECTION 6, continued
6.2 RS232 Connections
The RS232 receptacle on the Data Transfer Adapter connects with a 9-pin Sub-D connector (see Figure 9 RS232 Connection). A suitable RS232 cable is listed under Optional Accessories in REPLACEMENT PARTS AND ACCESSORIES on page 57.
The RS232 interface output is an eight-bit data word plus one stop bit and no parity with a baud rate of 1200. It can communicate with either a serial printer or a serial communication port on a computer.
Press the Section 6.3 Sending Data to a Printer or Computer for instructions.
With the use of a serial-to-parallel converter, the data string transmitted from the colorimeter prints on any compatible parallel printer of the type normally used with IBM compatible applications.
All RS232 connections are made using the serial DTA. This port uses an industry standard 9-pin connector. See Figure 9 RS232 Connection.
Figure 9 RS232 Connection
PRINT key to send data to the printer or computer. See
I/O port on the
Note: For optimum performance and ESD protection, use a five-
conductor shielded cable. Use a metal shell for the printer or CRT terminal connector, and connect the shield of the cable to the metal shell and to the sleeve (signal ground) of the RS232 plug.
44
Page 47
SECTION 6, continued
6.2.1 Setup and Use of the Printer
Follow the manufacturer's instructions when configuring the printer for compatibility with the colorimeter.
Pressing
PRINT manually starts the printing, and pressing the EXIT
key stops the printing (refer to Section 6.3 Sending Data to a Printer or Computer).
Connect the DTA to a printer using an appropriate cable.
Tabl e 4 and Table 5 show the proper pin connections for 9-pin computer cables and for 25-pin printer cables. Use of cables that do not match the pin information given may cause undesirable operation.
Table 4 Standard 9-pin to 9-pin Computer Cable
DR/800 Series 9-pin
Connector Socket
Pin Signal Name Pin Signal Name
2 RXD 3 TXD
3 TXD 2 RXD
4 DTR no connection
Computer 9-pin D Connector, plug
5 GND 5 GND
6 DSR no connection
7 RTS 8 CTS
8 CTS 7 RTS
Table 5 Standard 9-pin to 25-pin Printer Cable
DR/800 Series 9-pin
Connector Socket
Pin Signal Name
2 RXD no connection
3 TXD 3 RXD
4 DTR no connection
5 GND 7 GND
6 DSR 20 DTR
7 RTS no connection
Computer 9-pin D Connector, plug
Pin Signal Name
45
Page 48
SECTION 6, continued
Table 5 Standard 9-pin to 25-pin Printer Cable
DR/800 Series 9-pin
Connector Socket
Pin Signal Name
8CTS 20 DTR
To print, the communication parameters (baud rate, data bits and parity) of the instrument and the printer must match.
6.2.2 Connecting to a Personal Computer
Connect the colorimeter to a personal computer (PC) with the computer interface cable (Cat. No. 48129-00) listed under REPLACEMENT PARTS AND ACCESSORIES on page 57. The cable provides a direct link between the colorimeter and the 9-pin D connector used for the serial port on most personal computers. If your computer has a 25-pin D connector, use a 9-pin to 25-pin adapter (available at most computer supply stores).
Use a communications software, such as HachLink™ Software (Cat. No. 49665-00) to collect data from the instrument. HachLink is a Windows-based application that allows a personal computer to capture data from several Hach instruments, including the DR/800 Series Colorimeters.
Computer 9-pin D Connector, plug
Pin Signal Name
The captured data can be stored in a text file as a spread-sheet compatible format or as free-format text. Data captured in the spreadsheet format is easily transferred into most spreadsheet
®
programs (i.e., Excel
, Microsoft Works®, Lotus 123®) for
graphing and reporting.
To install and run HachLink™ Software, the computer and software must meet the following minimum requirements:
PC with a 486/100 MHz or higher processor
32 MB RAM
Hard drive with at least 20 MB free space
3½ inch floppy disk drive or CD-ROM drive
VGA graphics with 640 x 480 or higher resolution,
256 colors
46
Page 49
SECTION 6, continued
Mouse or other pointing device
9-pin serial port (or 25-pin serial port with 9-pin adapter)
Windows 95, 98, 2000, or NT 4.0 operating system
To transfer data, the communication parameters (baud rate, data bits and parity) of the instrument and the computer must match. Once the communication link is established, press data to the computer.
6.3 Sending Data to a Printer or Computer
A permanent record of test results may be obtained using the DTA RS232 serial output to drive a printer or by transferring the data to a computer for storage. Data displayed on the instrument screen may be sent to an accessory printer or computer by attaching the DTA to the instrument and pressing the The data can be recalled from memory or the current sample reading. Only the information shown on the display will be printed.
PRINT to send
PRINT key.
6.3.1 Sending Currently Displayed Data
To transfer currently displayed data:
1. Remove the instrument cap and slide the DTA onto the
instrument until snug.
2. Make sure the DTA is properly connected to the computer or printer. See Section 6.2 RS232 Connections.
3. Press the while the data is being transmitted to the DTA.
PRINT key. Four dashes and PRINT are displayed
47
Page 50
SECTION 6, continued
6.3.2 Sending Recalled Data
To transfer recalled data:
1. Remove the instrument cap and slide the DTA onto the
instrument until snug.
2. Make sure the DTA is properly connected to the computer or printer. (See Section 6.2 RS232 Connections.)
3. Recall the data to be transferred (see Section 5.1 Recalling Data on page 39).
4. When sample data is displayed, press the horizontal lines and transmitted to the DTA.
6.3.3 Sending All Stored Data
All data stored in memory may be sent to a printer or computer via the
1. Remove the instrument cap and slide the DTA onto the instrument until snug.
2. Make sure the DTA is properly connected to the computer or printer. (See Section 6.2 RS232 Connections.)
3. Press the
4. Scroll, using the down
5. Press the
PRINT. All stored data is now sent to the DTA. After the data
has been successfully sent, the instrument defaults to the last used program.
PRINT key. Four
PRINT are displayed while the data is
SETUP menu option as follows:
SETUP key.
ARROW key, to the PRINT option.
ENTER key when the word ALL is displayed above
Transferred data will include the following information:
instrument model number
instrument serial number
instrument software version
date
48
Page 51
SECTION 6, continued
time
program number
sample number
concentration
units
chemical form
Overrange errors (limit)
absorbance
%T
Figure 10 Printed Data Format
49
Page 52
50
Page 53
SECTION 7 MAINTENANCE
7.1 Cleaning the Colorimeter
Use a damp cloth to wipe the outside of the colorimeter enclosure. Wipe up spills promptly. Use cotton swabs to clean and dry the sample compartment if any spillage occurs.
Keep the colorimeter and sample cells clean at all times. Use a lens tissue or a soft, lint-free cloth (that will not leave an oil film) to wipe out the sample cell.
7.1.1 Cleaning the Data Transfer Adapter
Little cleaning is required of this adapter. Clean the outside and inside with a barely damp cloth. Wipe up spills promptly.
7.1.2 Sample Cells
Clean sample cells with detergent, rinse several times with tap water, and then rinse thoroughly with deionized water. Some cells may require acid washing or other special cleaning procedures. Refer to the Procedures Manual for additional information. Rinse sample cells used with organic solvents (chloroform, benzene, toluene, etc.) with acetone before the detergent wash, and again as a final rinse before drying.
7.2 Replacement Instructions
To prevent static electricity damage to the instrument, always turn the instrument off before removing the batteries.
7.2.1 Battery Replacement
When the LOW BATTERY icon appears in the display the batteries must be replaced or recharged as soon as possible to ensure proper instrument performance. Turn the instrument off before removing the battery compartment door.
The correct date and time may need to be reentered after battery replacement. See Section 2.3 Setting the Date and Time.
See Section 2.1 Battery Installation for complete installation instructions.
51
Page 54
52
Page 55
SECTION 8 TROUBLESHOOTING
8.1 Introduction
Correcting problem conditions with the DR/800 Series Colorimeters in the field is limited to responding to the error messages presented in the display. Other problems must be handled by a Hach technician at a service center. Refer to REPAIR SERVICE. Do not attempt to service anything other than the battery; there are no other field-serviceable parts. Opening the instrument case will void the warranty.
8.1.1 Error Codes
This feature identifies the problem area or areas when an error indication occurs during instrument operation. When an error occurs, which refers to a diagnostic error code. Refer to Table 6 below to determine the error cause and possible corrective actions. Turn the instrument off, then on, to restore the instrument to operation.
ERROR displays on the screen, followed by a number
ERROR Code
Number
1 Unconfigured instrument Contact Hach Instrument Service Department
2 Could not read program data Contact Hach Instrument Service Department
3 Could not write program data Contact Hach Instrument Service Department
4 Measurement battery error Replace instrument batteries
5 Measurement A/D error Contact Hach Instrument Service Department
6 Measurement offset error Check to be sure instrument light shield (cap) is
7 Measurement low light error Check for light path blockage
8 Measurement over-range
Table 6 Error Codes
Error Code Type Corrective Action
correctly installed
Zero is out of instrument range; dilute to within range
Contact Hach Instrument Service Department
Make sure instrument cap is properly installed
error
Contact Hach Instrument Service Department
53
Page 56
SECTION 8, continued
8.1.2 Beeper/Error Icon
When a key is pressed that calls for the instrument to perform a function that is not available at that time, the beeper is activated once and the error icon appears on the display. Each time the unacceptable command is entered, one beep sounds.
If an unavailable program number is entered, the beeper also sounds. Program numbers must range from 101 to 110 (user programs) and 1 to 100 (Hach programs). Entering a number other than these results in the beeper sound and the error icon briefly illuminates. After this beep sounds, re-enter the proper number.
8.1.3 Concentration Out of Range
An out-of-range condition is indicated by the illuminated LIMIT icon. It means the sample concentration exceeds the range of the programmed calibration. Make sure the test procedure is followed correctly, dilute the sample (for over range samples) and rerun the test. Each Hach test has an upper concentration value that defines the program range. Measurements beyond that range are unreliable.
8.1.4 Low Battery
The instrument continuously monitors battery voltage. If the battery voltage falls to a level which indicates less than ten percent battery life remains, the instrument automatically warns the operator by displaying the batteries as soon as possible for most reliable instrument performance.
LOW BATTERY icon. Replace the
54
Page 57
GENERAL INFORMATION
At Hach Company, customer service is an important part of every product we make.
With that in mind, we have compiled the following information for your convenience.
55
Page 58
56
Page 59
REPLACEMENT PARTS AND ACCESSORIES
REQUIRED APPARATUS
Description Unit Cat. No.
Adapter, Assembly, COD............................................................................each ........... 4846400
Batteries, Alkaline AA .............................................................................pkg/4 ...........1938004
Manual Set, DR/890,
includes Instrument & Procedure Manual & Binder ...............................each ........... 4847077
Manual Set, DR/850,
includes Instrument & Procedure Manual & Binder ...............................each ........... 4845077
Manual Set, DR/820,
includes Instrument & Procedure Manual & Binder ...............................each ........... 4844077
Sample Cell, 25 x 95 mm 10-20-25 mL....................................................pkg/6 ........... 2401906
Sample Cell, 1 cm/10 mL, pk/2.................................................................pkg/2 ...........4864302
OPTIONAL ACCESSORIES
Adapter, Data Transfer, RS232, includes 48129-00 cable ..........................each ...........4849000
Cap, Sample Cell, for 25 x 95 mL cell....................................................pkg/12 ........... 2401812
Carrying Case, DR/800 Series Colorimeter, hard-sided .............................each ...........4942500
Carrying Case, DR/800 Series Colorimeter, soft-sided w/shoulder strap...each ...........2722000
Carrying Case, Portable Laboratory............................................................each ...........4943000
Computer Interface Cable, 6 ft., for use with the DTA ...............................each ...........4812900
DR/Check™ ABS Standards......................................................................... set ........... 2763900
HachLink Software .....................................................................................each ...........4966500
Instrument Cap ............................................................................................each ...........4943100
57
Page 60
58
Page 61
How To Order
By Telephone: By Mail:
6:30 a.m. to 5:00 p.m. MST Monday through Friday (800) 227-HACH (800-227-4224)
By FAX: (970) 669-2932
Ordering information by E-mail: orders@hach.com
Hach Company P.O. Box 389 Loveland, CO 80539-0389 U.S.A.
Information Required
• Hach account number (if available) • Billing address
• Your name and phone number • Shipping address
• Purchase order number • Catalog number
• Brief description or model number • Quantity
Technical and Customer Service (U.S.A. only)
Hach Technical and Customer Service Department personnel are eager to answer questions about our products and their use. Specialists in analytical methods, they are happy to put their talents to work for you. Call 1-800-227-4224 or E-mail techhelp@hach.com.
International Customers
Hach maintains a worldwide network of dealers and distributors. To locate the representative nearest you, send E-mail to
intl@hach. com or contact:
In Canada, Latin America, Africa, Asia, Pacific Rim:
Telephone: (970) 669-3050; FAX: (970) 669-2932
In Europe, the Middle East, or Mediterranean Africa: HACH Company, c/o
Dr. Bruno Lange GmbH Willstätterstr. 11 D-40549 Düsseldorf Germany Telephone: +49/[0]211.52.88.0 Fax: +49/[0]211.52.88.231
59
Page 62
Repair Service
Authorization must be obtained from Hach Company before sending any items for repair. Please contact the HACH Service Center serving your location.
In the United States:
Hach Company 100 Dayton Avenue Ames, Iowa 50010 (800) 227-4224 (U.S.A. only) Telephone: (515) 232-2533 FAX: (515) 232-1276
In Canada:
Hach Sales & Service Canada Ltd. 1313 Border Street, Unit 34 Winnipeg, Manitoba R3H 0X4 (800) 665-7635 (Canada only) Telephone: (204) 632-5598 FAX: (204) 694-5134 E-mail: canada@hach.com
In Latin America, the Caribbean, the Far East, the Indian Subcontinent, Africa, Europe, or the Middle East:
Hach Company World Headquarters P.O. Box 389 Loveland, Colorado, 80539-0389 U.S.A. Telephone: (970) 669-3050 FAX: (970) 669-2932 E-mail: intl@hach.com
60
Page 63
Warranty
Hach Company warrants this product to the original purchaser against
any defects that are due to faulty material or workmanship for a period of one year from date of shipment.
In the event that a defect is discovered during the warranty period, Hach Company agrees that, at its option, it will repair or replace the defective product or refund the purchase price, excluding original shipping and handling charges. Any product repaired or replaced under this warranty will be warranted only for the remainder of the original product warranty period.
This warranty does not apply to consumable products such as chemical reagents; or consumable components of a product, such as, but not limited to, lamps and tubing.
Contact Hach Company or your distributor to initiate warranty support. Products may not be returned without authorization from Hach Company.
Limitations
This warranty does not cover:
• damage caused by acts of God, natural disaster, labor unrest, acts of war (declared or undeclared), terrorism, civil strife or acts of any governmental jurisdiction
• damage caused by misuse, neglect, accident or improper application or installation
• damage caused by any repair or attempted repair not authorized by Hach Company
• any product not used in accordance with the instructions furnished by Hach Company
• freight charges to return merchandise to Hach Company
• freight charges on expedited or express shipment of warranted parts or product
• travel fees associated with on-site warranty repair
This warranty contains the sole express warranty made by Hach Company in connection with its products. All implied warranties, including without limitation, the warranties of merchantability and fitness for a particular purpose, are expressly disclaimed.
Some states within the United States do not allow the disclaimer of implied warranties and if this is true in your state the above limitation may not apply to you. This warranty gives you specific rights, and you may also have other rights that vary from state to state.
This warranty constitutes the final, complete, and exclusive statement of warranty terms and no person is authorized to make any other warranties or representations on behalf of Hach Company.
Limitation of Remedies
The remedies of repair, replacement or refund of purchase price as stated above are the exclusive remedies for the breach of this warranty. On the basis of strict liability or under any other legal theory, in no event shall Hach Company be liable for any incidental or consequential damages of any kind for breach of warranty or negligence.
61
Page 64
62
Page 65
CERTIFICATION
Hach Company certifies this instrument was tested thoroughly, inspected and found to meet its published specifications when it was shipped from the factory.
The DR/800 Series Colorimeter has been tested and is certified as indicated to the following instrumentation standards:
EN 60825-1: LEDs used in this product are Class 1
Immunity:
EN 50082-1 “1997”(Generic Immunity Standard) per 89/336/EEC EMC: Supporting test records by Hach Company, certified compliance
by Hach Company.
Required Standard/s include:
EN 61000-4-2 (IEC 1000-4-2) Electro-Static Discharge
EN 61000-4-3 (IEC 1000-4-3) Radiated RF Electro-Magnetic Fields
ENV 50204 Radiated Electro-Magnetic Field from Digital Telephones
Emissions:
Per 89/336/EEC EMC: Supporting test records by Intellistor O.A.T.S., (NVLAP #0369) certified compliance by Hach Company.
Required European Standard/s include:
EN 55011 (CISPR 11) Emissions, Class B Limits
Additional Emissions Standard/s include:
CANADIAN INTERFERENCE-CAUSING EQUIPMENT REGULATION, IECS-003, Class A:
Supporting test records by Intellistor O.A.T.S., certified compliance by Hach Company.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
63
Page 66
CERTIFICATION, continued
FCC PART 15, Class “A” Limits:
Supporting test records by Intellistor O.A.T.S., certified compliance by Hach Company.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at his own expense. The following techniques of reducing the interference problems are applied easily.
1. Remove power from the Colorimeter by removing one of its
batteries to verify that it is or is not the source of the interference.
2. Move the Colorimeter away from the device receiving the
interference.
3. Reposition the receiving antenna for the device receiving
the interference.
4. Try combinations of the above.
64
Page 67
65
Page 68