Section 2 General Information .............................................................................................................................. 9
2.1 Safety information .............................................................................................................................................. 9
2.1.1 Use of hazard information ......................................................................................................................... 9
3.3 Power connections ........................................................................................................................................... 12
3.6 Mobile use of the DR 2800 for field analysis .................................................................................................... 19
3.6.1 Position the Protective Cover .................................................................................................................. 20
Section 4 Start Up ................................................................................................................................................ 23
4.1 Power the instrument on and off ...................................................................................................................... 23
4.2 Language selection .......................................................................................................................................... 23
5.1.1 Tips for the use of the touch screen ........................................................................................................ 25
5.1.2 Use of the alphanumeric keypad ............................................................................................................ 25
5.1.3 Main Menu .............................................................................................................................................. 26
5.2.1 Operator ID ............................................................................................................................................. 27
5.2.2 Sample ID ............................................................................................................................................... 28
5.2.3 Date and time .......................................................................................................................................... 29
5.2.4 Display and sound preferences .............................................................................................................. 29
5.2.5 Power Management ................................................................................................................................ 29
5.2.6 PC and printer ......................................................................................................................................... 31
5.2.6.2 Print data .................................................................................................................................... 33
5.2.6.3 HACH Data Trans ....................................................................................................................... 33
5.3 Store, recall, send and delete data .................................................................................................................. 37
5.3.1 The data log ............................................................................................................................................ 37
5.3.1.1 Auto/manual data storage ........................................................................................................... 37
5.3.1.2 Recall stored data from the data log ........................................................................................... 37
5.3.1.3 Send data from the data log ....................................................................................................... 38
5.3.1.4 Delete stored data from the data log .......................................................................................... 39
5.3.2 Time Course ........................................................................................................................................... 40
5.3.2.1 Data Storage from Time Course ................................................................................................. 40
5.3.2.2 Recall Stored Data from Time Course ........................................................................................ 40
5.3.2.3 Send Data from Time Course ..................................................................................................... 41
5.3.2.4 Delete Stored Data from Time Course ....................................................................................... 42
5.4.1 Select a saved test/method; enter user-specific basic data .................................................................... 43
5.4.2 Stored program options .......................................................................................................................... 43
5.4.3 Use of program timers ............................................................................................................................ 45
5.4.4 Set the dilution factor .............................................................................................................................. 46
5.4.5 Run a standard adjust ............................................................................................................................. 46
5.4.6 Set the chemical form ............................................................................................................................. 47
5.4.6.1 Change of the default setting of the chemical form .................................................................... 48
5.4.7 Run a reagent blank ................................................................................................................................ 48
5.4.8 Analysis of samples ................................................................................................................................ 49
5.4.9 Add stored programs to the favorite programs list .................................................................................. 50
5.5.1 Complete a barcode 13 mm test/vial ...................................................................................................... 51
5.5.2 Select the measuring range .................................................................................................................... 52
5.5.3 Select the chemical evaluation form ....................................................................................................... 52
5.5.3.1 Change of the default setting of the chemical form .................................................................... 53
5.5.4 Basic test-specific and sample-specific data settings ............................................................................. 53
6.1 User Programs ................................................................................................................................................. 61
6.1.1 Program a user method .......................................................................................................................... 61
6.1.1.1 Single wavelength settings ......................................................................................................... 63
6.1.1.2 Multi wavelength settings ............................................................................................................ 64
6.1.1.3 Calibration settings for single and multi wavelength mode ......................................................... 66
6.1.1.4 Store a user program .................................................................................................................. 70
6.1.1.5 Additional user-defined parameters and functions ..................................................................... 7
6.1.2 Free programming program type ............................................................................................................ 72
6.1.2.1 Measurement process ................................................................................................................ 72
6.1.2.2 Enter a new element of a measuring sequence ......................................................................... 73
6.1.2.3 Enter the calibration formula (evaluation formula) ...................................................................... 75
6.1.2.4 Enter variables ............................................................................................................................ 78
6.1.2.5 Save a free programming user program ..................................................................................... 78
6.1.3 Select a user program ............................................................................................................................. 79
6.1.4 Add, edit and delete user programs from the favorites list ...................................................................... 79
6.1.4.1 Add to Favorites .......................................................................................................................... 80
6.2.1 Recall a favorite program ........................................................................................................................ 81
6.2.2 Delete a favorite program ....................................................................................................................... 81
6.3 Standard Addition – monitoring/checking results ............................................................................................. 82
6.3.1 Complete a standard addition ................................................................................................................. 83
6.4 Single Wavelength (absorbance, concentration and transmittance measurements) ....................................... 87
6.4.1 Set up single wavelength mode .............................................................................................................. 87
6.4.2 Take single wavelength measurements .................................................................................................. 89
6.5 Multi-Wavelength mode – measurements at more than one wavelength ........................................................ 90
6.5.1 Set the reading mode at different wavelengths ....................................................................................... 90
6.5.2 Complete a measurement in the multi wavelength mode ....................................................................... 93
6.6 Time course of absorbance/transmittance ....................................................................................................... 94
6.6.1 Time course setup parameters ............................................................................................................... 94
6.6.2 Time course scan reading ....................................................................................................................... 95
6.6.3 Analysis of time course data ................................................................................................................... 96
6.6.3.1 Navigation of a time scan or a time scan analysis ......................................................................97
6.7 System checks ................................................................................................................................................. 97
6.7.1 Instrument information ............................................................................................................................ 98
6.7.2 Upgrade of the instrument software ........................................................................................................ 98
6.7.5 Lamp history ......................................................................................................................................... 104
6.7.6 Factory service ...................................................................................................................................... 104
6.7.7 Service time .......................................................................................................................................... 105
7.2 Insert or change of the battery ....................................................................................................................... 110
7.2.1 Information about using the battery ...................................................................................................... 111
7.2.1.1 Dispose of the lithium battery ................................................................................................... 111
7.2.1.2 Optimal operation of the battery ............................................................................................... 112
7.2.1.3 Lifespan of the battery .............................................................................................................. 112
7.2.1.4 Load the battery/Operating time ............................................................................................... 112
Index .................................................................................................................................................................... 125
5
Table of Contents
6
Section 1Specifications
Specifications are subject to change without notice.
Performance Specifications
Operating ModeTransmittance (%), Absorbance and Concentration
Source LampGas-filled Tungsten (visible)
Wavelength Range340–900 nm
Wavelength Accuracy± 1.5 nm
Wavelength Reproducibility< 0.1 nm
Wavelength Resolution1 nm
Wavelength CalibrationAutomatic
Wavelength SelectionAutomatic, based on method selection
Spectral Bandwidth< 8 nm
Photometric measuring range± 3.0 Ext in Wavelength Range 340–900 nm
Photometric Accuracy
Photometric Linearity
Stray Light
Data storage500 measured values (result, date, time, sample ID, user ID)
5 mAbs at 0.0–0.5 Abs
1% at 0.50–2.0 Abs
< 0.5%–2 Abs
< = 1% at > 2 Abs
with neutral glass at 546 nm
< 0.2% T @ 340 nm with KV450/3
< 0.1% T @ 340 nm with NaNO
2
User programs50
Physical and Environmental Specifications
Width220 mm (8.6 in)
Height135 mm (5.3 in)
Depth330 mm (12.9 in)
Weight
Operating Requirements10–40 °C (50–104 °F), max. 80% relative humidity (non-condensing)
Storage Requirements-40–60 °C (-40–140 °F) max. 80% relative humidity (non-condensing)
Use only shielded cable with max. length of 3 m.
1 x USB type A
1 x USB type B
IP 41 with closed lid
IP 42 with Protective Cover in place
7
Specifications
8
Section 2General Information
2.1Safety information
Please read this entire manual before unpacking, setting up or
operating this equipment. Pay attention to all danger, warning and
caution statements. Failure to do so could result in serious injury to
the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not
impaired, do not use or install this equipment in any manner other
than that specified in this manual.
2.1.1 Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation
which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation
which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in
minor or moderate injury.
2.1.2 Precautionary labels
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC),
European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no
charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories and all auxiliary items for proper disposal.
This symbol indicates that the instrument contains a Class 1 LASER device.
LASER CLASS 1
Data: 0.3 mW; λ = 650 nm
Important Note: Indicates a situation which, if not avoided, may
cause damage to the instrument. Information that requires special
emphasis.
Note: Information that supplements points in the main text.
Read all labels and tags attached to the instrument. Personal injury
or damage to the instrument could occur if not observed. A symbol,
if noted on the instrument, will be included with a danger or caution
statement in the manual.
2.1.3 Class 1 LASER
A Class 1 LASER is installed in this instrument. Class 1 LASERS
are products where the radiant power of the LASER beam
accessible (the accessible emission) is always below the Maximum
9
General Information
2.1.4 Chemical and Biological Safety
Permissible Exposure value. Therefore, for Class 1 LASERS the
output power is below the level at which it is believed eye damage
will occur. Exposure to the beam of a Class 1 LASER will not result
in eye injury. Class 1 LASERS may therefore be considered safe.
However, Class 1 LASER products may contain LASER systems of
a higher Class but there are adequate engineering control
measures to ensure that access to the beam is not reasonably
likely. Examples of such products include LASER printers and
compact disc players. CDRH assession number 0510555-02.
Data: 0.3 mW; wavelength = 650 nm
DANGER
Potential Chemical/ Biological Exposure Hazards. Handling
chemical samples, standards and reagents can be dangerous.
Users of this product are advised to familiarize themselves
with safety procedures and the correct use of chemicals, and
to carefully read all relevant Material Safety Data Sheets.
Normal operation of this instrument may involve the use of
hazardous chemicals or biologically harmful samples.
2.2Overview of product
•The user must observe all cautionary information printed on the
original solution containers and safety data sheet prior to their
use.
•All waste solutions must be disposed in accordance with local
and national law.
•The type of protective equipment must be selected according to
the concentration and amount of the dangerous substance at
the specific workplace.
The DR 2800 spectrophotometer is a VIS spectrophotometer with a
wavelength range of 340 to 900 nm. The instrument comes with a
complete set of application programs and multi-language support.
The DR 2800 spectrophotometer contains the following application
modes: Stored Programs (pre-installed tests), Barcode Programs,
User Programs, Favorite Programs, Single Wavelength Mode,
Multi-Wavelength Mode and Time Course Mode.
The DR 2800 spectrophotometer provides digital readouts in direct
concentration units, absorbance, or percent transmittance.
When a user-generated or programmed method is selected, the
on-screen menus and prompts direct the user through the test.
10
This menu system also can be used to generate reports, statistical
evaluations of generated calibration curves, and to report
instrument diagnostic checks.
Section 3Installation
3.1Unpack the instrument
WARNING
Electrical and Fire Hazards. Use only the provided power
supply. Only qualified personnel should conduct the tasks
described in this section of the manual.
The DR 2800 Spectrophotometer comes packaged with the
following items:
•DR 2800 spectrophotometer
•Plug-in power supply, including 4 adapter for EU, GB, USA,
AUS/China
•Dust cover
•3 different cuvette/cell adapters (A, B and C)
•Light Shield belongs to standard configuration of DR 2800
•Protective Cover
•DR 2800 User Manual
•CD-ROM containing the HACH and HACH LANGE procedures
3.2Operating environment
•Quick start guide
Note: If any of these items are missing or damaged, contact the
manufacturer or a sales representative immediatly.
The following conditions are necessary to ensure correct
instrument operation and accurate results:
•Place the instrument firmly on an even surface. Do not push
any objects under the instrument.
•Maintain an ambient temperature of 10 to 40 ºC (50 to 104 ºF)
for proper instrument operation.
•The relative humidity should be less than 80%; moisture should
not condense on the instrument.
•Leave at least a 15 cm (6 in.) clearance at the top and on all
sides for air circulation to avoid overheating of electrical parts.
•Do not operate or store the instrument in extremely dusty,
damp or wet locations.
•Keep the surface of the instrument, the cell compartment and
all accessories clean and dry at all times. Splashes or spills on
and in the instrument should be cleaned up immediately (see
section 7.1 on page 109).
Important Note: Protect the instrument from temperature
extremes, including heaters, direct sunlight and other heat sources.
11
Installation
3.3Power connections
Install the correct adapter plug on the supplied external power
supply (Figure 1) by sliding the adapter on until it "clicks" into
position. Correctly mounted, both housing of power supply and plug
are in line. Plug the external power supply cord into the connector
on the back panel of the instrument, then plug the supply into a
power outlet (100–240 V~ / 50–60 Hz). Press the power switch on
the back of the instrument to initialize power
(Figure 2 on page 13).
Figure 1 Power adapter
1Power supply with EU adapter plug installed3USA adapter plug
2UK adapter plug4AUS/China adapter plug
12
3.4Connection
Installation
The DR 2800 has two USB interfaces as a standard feature,
located on the back of the instrument (Figure 2).
The USB Type A interface is used for communications with a
printer, USB memory stick, barcode scanner, or keyboard. A USB
memory stick is used to update instrument software.
The USB Type B interface is used for communications with a PC.
The optional Hach Data Trans software (see section 5.2.6.3 on
page 33) must be installed on the PC for this use.
A USB hub may be used to connect several accessories at a time.
Note: USB cables must not be longer than 3 meters (10 feet).
These USB interfaces enable data and graphics to be output to a
Printer and a PC and upgrade instrument software (see
section 6.7.2 on page 98).
Figure 2 Interfaces
1On/Off switch4Cover
2USB type B5USB type A
3Plug in power supply
13
Installation
3.5Cell compartments, Cell adapters, Light Shield and Protective Cover
3.5.1 Cell compartments and adapters
The DR 2800 has two cell compartments (Figure 3). Only one
cuvette/sample cell type at a time can be used for a measurement.
Cell compartment #1
•13-mm and 16-mm round cuvettes/vials
Note: Cell compartment #1 contains a barcode reader for
cuvettes/vials.
Cell compartment #2
Cell compartment #2 uses adapters to accommodate different
cuvette/sample cell types.
•1-inch square or 50-mm rectangular cuvettes/cells (can be
inserted directly into the cell compartment without using an
adapter).
•Adapter A: 10-mm square cuvettes/cells
•Adapter B: Pour-Thru cells (refer to the instruction sheet
supplied with the Pour-Thru cell) and multi-path cuvettes/cells
Note: Pour-Thru Cell must be used with Adapter B, not Adapter C.
•Adapter C: One-inch round cuvettes/cells and AccuVac®
Ampules
Note: One-inch round cuvettes/cells and AccuVac Ampules must be
used with Adapter C, not Adapter B.
Figure 3 Cell compartments
1Cell compartment #12Cell compartment #2
14
Installation
Figure 4 Cuvette/Sample cell adapters
110 mm square cuvette/sample cell adapter (A) 31 inch round cuvette/sample cell adapter (C)
21 inch Pour-Thru adapter (B)
3.5.2 Installation of the cuvette/sample cell adapters
1. Open the cell compartment.
2. Select the correct adapter for the cuvette/sample cell type.
3. Insert the adapter so the arrow on top of the adapter points to
the left (Figure 5 on page 16) and the orientation tab fits the
groove in the compartment opening. The cuvette/sample cell
type imprint should be legible on the adapter (Figure 4).
Note: The arrow on top of the adapter indicates the direction of the
light beam path.
15
Installation
Figure 5 Installation of a cuvette/sample cell adapter
3.5.3 Use of the light shield for measurements
The light shield (Figure 6 on page 17) prevents light interference
when using 13 mm and 16 mm vial tests and must be in place
before measurements can be taken in cell compartment #1. The
light shield is required only when using 13 mm or 16 mm vial tests.
16
Installation
The DR 2800 is shipped with the light shield installed. Remove the
light shield before using cell compartment #2. The light shield can
be stored in the Protective Cover (Figure 8 on page 18)
Figure 6 Light Shield
Installation of the light shield
1. Open the cell compartment.
2. Insert the light shield so the arrow on the light shield points to
the left and the orientation tab fits the groove in the
compartment opening (Figure 7).
The inside of the Protective Cover is intended to house
the cuvette/sample cell adapters. The recesses for
holding the cuvette/sample cell adapters are marked with
the corresponding letters.
adapter A in position A.
18
3.6Mobile use of the DR 2800 for field analysis
If the DR 2800 spectrophotometer is used to carry out field
analyzes, it may be necessary to take a number of measures to
compensate for the fluctuating ambient light conditions.
You should take the measures described below if the following
warning is displayed after you take a measurement:
Error: Too much ambient light! Move device into shade or
close the lid!
General measure:
Shield the instrument from the sun by standing so that your shadow
falls on it, and carry out the measurement again. If the warning
appears again, take the following measures.
Measures:
Measurements in a 10 mL round cuvette/cell, a rectangular
cuvette/cell or AccuVac Ampules
Close the lid of the cuvette compartment and carry out the
measurement.
Measurements in a 13 or 16 mm round cuvette/vial or a 25 mL
square cuvette/cell
To obtain consistent measurement conditions in all weather
conditions, place the Protective Cover on the open cuvette/cell
compartment (see Figure 9 Position the Protective Cover on page
20) and carry out the measurement.
Installation
Important Note: When you carry out measurement in 25 mL
square cuvettes/cells, take care that there is no adapter in the
Protective Cover.
The enclosure rating of the photometer increases from IP41 to IP42
when the Protective Cover is in place.
19
Installation
3.6.1 Position the Protective Cover
Figure 9 Position the Protective Cover
1. After you have inserted the zero solution cell or the sample cell,
position the Protective Cover over the open cell compartment.
The lettering "Protective Cover" must be readable from the
display side of the instrument.
2. Press the Protective Cover lightly into the correct position until
the cell compartment is completely sealed and the Protective
Cover cannot slip off.
3. Carry out the measurement.
Note: The Protective Cover can simultaneously be used to house the
different cuvette adapters. The cuvette adapter recesses in the Protective
Cover are marked with the corresponding letters and arrows of the adapter.
The arrows in the Protective Cover and on the adapters indicate the
direction of insertion.
1. Plug in to the power supply (laboratory analysis) or insert the
battery (field analysis).
2. The push-button switch on the back of the instrument switches
the instrument on (press for about 1 second) and off (press for
about 3 to 5 seconds). An acoustic signal confirms that the
instrument has been switched off.
Note: Do not turn the instrument off and on in rapid succession. Always
wait about 20 seconds before turning the instrument on again, otherwise
the electronic and mechanical systems will be damaged.
4.2Language selection
The spectrophotometer software includes several language
options. The first time the instrument is turned on, the language
selection screen will appear.
4.3Self-Check
1. Select the desired language.
2. Press OK to confirm the language selection. The self check will
start automatically.
Change the language setting
The instrument functions in the selected language until the option is
changed.
1. While turning the instrument on, touch the screen at any point
until the list for selecting a language appears
(about 30 seconds).
2. Select the required language.
3. Press OK to confirm. The test program subsequently starts
automatically.
Each time the instrument is powered up, a series of diagnostic tests
are performed automatically to ensure operation of major system
components.
This procedure, which takes approximately two minutes, checks the
system, lamp, filter adjustment, wavelength calibration and voltage.
Each test which functions correctly is confirmed with a check mark.
The Main Menu is displayed when power up diagnostics are
completed.
23
Start Up
If the instrument detects any deviation relative to the last
calibration, a system check has to be carried out.
4. Remove the cuvette from the cell compartment and close the
lid.
5. Then press Start.
The system check is carried out (duration approx. 6 minutes).
Note: Further error messages during self check, see Section 8 on page
117.
24
Section 5Standard Operations
5.1Overview
5.1.1 Tips for the use of the touch screen
The entire screen is touch-activated. To make a selection, press the
screen with a fingernail, fingertip, pencil eraser or a stylus. Do not
press the screen with a sharp object, such as the tip of a ball
point pen.
•Do not place anything on top of the screen, to prevent damage
or scratching on the screen.
•Press keys, words or icons to select them.
•Use scroll bars to move up and down long lists very quickly.
Press and hold the scroll bar, then move up or down to move
through the list.
•Highlight an item from a list by pressing it once. When the item
has been successfully selected, it will be displayed as reversed
text (light text on a dark background).
5.1.2 Use of the alphanumeric keypad
This display is used to enter letters, numbers and symbols as
needed when programming the instrument. Unavailable options are
disabled (grayed out). The icons on the right and left of the screen
are described in Table 1.
The central keypad changes to reflect the chosen entry mode.
Press a key repeatedly until the desired character appears on the
screen. A space can be entered by using the underscore on the
YZ_ key.
Note: A USB keyboard (with US keyboard layout) or a USB Barcode
handset scanner can be used for input (see Section 9 on page 119).
Table 1 Alphanumeric keypad
Icon / keyDescriptionFunction
ABC/abcAlphabetic
# %SymbolsPunctuation, symbols and numerical sub- and superscripts may be entered.
123NumericFor entering regular numbers.
CEClear EntryClear the entry.
Left ArrowBackspace
Right ArrowAdvance
When entering alphabetic characters (ex. user-entered units), this key allows to
toggle between upper and lower case letters.
Moves back one position. This deletes the character previously entered in the new
position.
Moves to the next space in an entry when two adjacent characters occur on the
same key.
25
Standard Operations
5.1.3 Main Menu
OptionFunction
A variety of modes may be selected from the ”Main Menu“. The
following table briefly describes each menu option.
Table 2 ”Main Menu“ Options DR 2800
Stored Programs /
Barcode Programs
(HACH LANGE
Programs)
User Programs
Favorite ProgramsList of methods/tests created by the user to suit his own requirements.
Single Wavelength
Multi Wavelength
Time CourseThe time scan records the absorbance or % transmittance at a wavelength over a defined time.
System Checks
Recall DataStored data can be called up, filtered, transmitted and deleted.
Instrument Setup
Stored programs are pre-programmed methods that make use of HACH reagents and LANGE
cuvette tests and pipette tests.
The DR 2800 Procedures Manual contains illustrated, step-by-step procedures for analyzes
using HACH programs. The working procedures for LANGE tests are included in the test packs.
User programs make "made to measure analysis" possible:
- Users can program methods they have developed themselves
- Existing HACH and LANGE methods can be stored as user programs. The LANGE tests can
then be modified to suit the user's requirements.
Single wavelength measurements are:
Absorbance measurements: The light absorbed by the sample is measured in absorbance
units.
Transmittance measurements (%): The percentage of the light that passes through the sample
and reaches the detector is measured.
Concentration measurements: A concentration factor can be entered to enable the measured
absorbance values to be converted into concentration values.
In the multi-wavelength mode, absorbance (Abs) or percentage transmittance (%T) is measured
at up to four wavelengths, and absorbance differences and absorbance relationships are
calculated. Simple conversions into concentrations can also be carried out.
The system checks menu offers a number of options, including optical checks, output checks,
lamp history, instrument update, service time and instrument backup.
In this mode, user-specific or method-specific settings can be entered: Operator-ID, Sample-ID,
Date & Time, Display & Sound, Power Management, PC & Printer and Password.
26
5.2Instrument Setup mode
5.2.1 Operator ID
Standard Operations
1. Select Instrument Setup in the ”Main Menu“.
A selection of functions appears in order to configure the functions
of the instrument.
Use this option to enter up to 30 sets of operator initials (up to
5 characters each) into the instrument. This feature helps record
which operator measured each sample.
1. Press Operator ID in the Instrument Setup.
2. Press New to enter a new Operator ID.
3. Use the alphanumeric keypad to enter a new Operator ID.
4. Press OK to confirm.
5. The display shows the chosen Operator ID.
6. Press OK. The instrument will return to the Instrument Setup
screen and show the selected operator identifier.
7. The chosen Operator ID is activated.
Note: Press Delete to remove an Operator ID from the list.
Note: Alternatively, enter or change an Operator ID in measurement mode.
In the results screen, press Options>More>Instrument Setup or if an
Operator ID is already assigned, select the "Operator ID symbol
immediately in the results screen.
27
Standard Operations
5.2.2 Sample ID
Use this option to enter up to 100 Sample Identification tags (up to
13 characters each) into the instrument. Sample IDs can be used to
specify the sample location or other sample specific information.
1. Press Sample ID in the Instrument Setup.
2. Press New to enter a new Sample ID.
3. Use the alphanumeric keypad to enter a new Sample ID.
Note: If a USB Barcode handset scanner ( see Section 9 on page 119)
is connected, Sample IDs can also be scanned. Sample IDs can also
be entered with a USB keyboard.
4. Press OK to confirm.
5. To number the Sample IDs sequentially (e.g. Inflow (01 etc.)), select Add Number.
•Use the arrow keys to specify the first number of the
sequence.
•Use the key between the arrow keys to enter the first
number of the sequence using the alphanumeric keypad.
6. Press OK to return to "Instrument Setup".
28
7. The Sample ID is activated. Each Sample ID is automatically
numbered in ascending order after a measurement. The
number is shown in parentheses behind the Sample ID.
Note: To remove a Sample ID, highlight the ID and press Delete.
Note: A Sample ID can be entered or changed in measurement mode. In
the results screen, press Options>More>Instrument Setup. If a Sample
ID is already assigned, select the Sample ID symbol in the results screen.
5.2.3 Date and time
1. Press Date & Time in the Instrument Setup.
2. The date and time are subdivided over a number of fields.
3. Press OK to confirm. The instrument will return to
5.2.4 Display and sound preferences
1. Press Display & Sound in the Instrument Setup.
Standard Operations
Press the appropriate field and use the arrow keys to change
the value.
Instrument Setup.
5.2.5 Power Management
The following options will be displayed:
•Display Contrast—Adjusts the display brightness to suit
lighting conditions.
•Screen touch—Activates//Deactivates a short beep each time
the screen is pressed (Default:off).
•Reading done—Activates/Deactivates a sound when a
reading is complete (Default: short beep every time a reading is
complete).
•Timer—Adjusts the length of the timer sound. Select Short or
Long. Long beeps are recommended for noisy environments.
2. Press OK to confirm. The instrument will return to
Instrument Setup.
The DR 2800 can run on mains or battery power.
Note: The battery is not part of the standard scope of delivery.
1. Select Power Management in the ”Instrument Setup“ menu.
The battery symbol indicates the charge status of the battery
in %.
Note: The timer settings in the Power Management menu are only
active when the instrument is running on battery power.
29
Standard Operations
2. Select one of the options under Standby mode to set the
length of the idle period that can elapse before the instrument
switches to the Standby mode when running on battery power.
Note: In standby mode, the backlighting of the screen is switched off.
Touching the display causes the lighting to switch on again.
3. Select one of the options under Auto off to set the length of the
idle period that can elapse before the instrument automatically
switches off when running on battery power.
Note: After the instrument switches itself off automatically, you have to
press the push-button switch on the back of the instrument to start it
again.
30
5.2.6 PC and printer
USB InterfacesDescription
USB (Type B)
USB (Type A)
This USB interface is only intended to connect the instrument to a PC (with installation of the
HACH Data Trans Software).
This USB port can be used to connect a printer, a USB memory stick , a barcode scanner or a
keyboard.
Standard Operations
The instrument is provided with 2 USB interfaces, which are located
on the back of the instrument (see Figure 2 on page 13). These
interfaces can be used for exporting data and graphics to a printer,
updating data and for data communication to a personal computer.
These interfaces can be used for the connection of a USB stick, an
external USB keyboard or a USB Barcode handset scanner.
Note: A USB hub may be used to connect several accessories at a time.
A USB memory stick is used to upgrade data and software, see
section 6.7.2 on page 98.
Important Note: A screened USB cable must not be longer than
3m!
Table 3 USB connector
5.2.6.1 Printer setup
1. Press PC & Printer in the Instrument Setup.
A list with information about the connections opens.
For reasons of compatibility, the printer language must be
HP PCL 3.
2. Press Printer.
3. Press Setup to display the Printer Setup screen.
31
Standard Operations
Printer Setup:
•Resolution: Print quality
•Paper: Paper size
Note: If an optional Thermal Printer is connected, the function "Auto Send"
on/off is available.
4. Select Auto-Send: On to send all measured data automatically
to the Thermal printer.
Note: The option Auto-Send is not available for any other printer (e.g. ink
jet printer).
5. Press Resolution to select the print quality.
Select between
•100 dpi
•150 dpi and
•300 dpi
6. Press OK to confirm.
Note: Press OK again to return to the PC & Printer menu.
32
5.2.6.2 Print data
Standard Operations
7. Press Paper to select the paper size.
Select between
•Letter
•Legal
•Executive
•A4
8. Press OK to confirm.
Note: Press OK again to return to the PC & Printer menu.
1. Press Recall Data in the Main Menu.
2. Select the data source, where the data to be printed are stored.
5.2.6.3 HACH Data Trans
3. A list is displayed. Data can be filtered. For more information
see section 5.3.1.2 on page 37.
4. Press the PC & Printer icon to send the data (table, curve)
immediately to the printer.
5. Highlight Single point or Filtered data or All data and press
OK to confirm.
Sending Data... is displayed until the data have been printed.
The optional HACH Data Trans software must be installed on the
PC for the subsequent to process for measurement data.
1. Press PC & Printer in the Instrument Setup.
2. Select PC.
3. Press Setup to display the PC Setup screen.
For further installation instructions, refer to the HACH Data
Trans user manual.
33
Standard Operations
5.2.7 Password
4. Select Auto-Send: On to send all measured data automatically
to the PC.
Note: If Auto-Send: Off is selected, the PC & Printer icon must be
pressed, in order to send data to the PC.
Note: The remote function is only for monitoring the data transfer.
The Password menu contains a variety of security settings to
control access to various functions. For example, prevent
unauthorized changes to stored programs or instrument
configurations.
1. Press Password in the Instrument Setup menu.
2. In order to highlight the Security List assign a password. Press
Set Password.
3. Use the alphanumeric keypad to enter a new Password (up to
10 characters each) and press OK to confirm.
The access to the Security List is activated.
34
Standard Operations
4. Press Security List to lock various functions for unauthorized
users.
5. Highlight the desired functions to control.
6. Confirm the Security List with OK to return to the Password
menu.
7. Press On to highlight the new settings of the Security List.
5.2.7.1 Password deactivation
8. Enter the new Password again to confirm.
9. Press OK to return to Instrument Setup.
Note: The alphanumeric keypad to the Password inquiry appears when a
user tries to reach a locked setting.
1. Press Password in the Instrument Setup.
2. Use the alphanumeric keypad to enter the former Password and press OK to confirm.
35
Standard Operations
3. Press Off to deactivate the settings of the Security List.
4. Press OK to return to Instrument Setup.
Note: Use this function to delete the former Password or to enter a new
one.
36
5.3Store, recall, send and delete data
5.3.1 The data log
The Data Log will store up to 500 readings taken in the modes:
Stored Programs, Barcode Programs, User Programs, Favorite
Programs, Single Wavelength and Multi Wavelength. A complete
record of the analysis is stored, including the Date, Time, Results,
Sample ID, and Operator ID.
5.3.1.1 Auto/manual data storage
The data storage parameter indicates whether data are to be stored
automatically or manually (in which case the user has to decide
which data to store).
1. Press Store: On/Off in the Options menu.
•With the Store On setting, all measurement data are stored
automatically.
•With the Store Off setting, no measurement data are
stored. However, this setting can be changed to Store On
in the result display through configuration. The reading
currently shown in the display is then stored.
Standard Operations
Note: When the instrument's memory (data log) is full, the oldest data are
automatically deleted allowing the new data to be stored.
5.3.1.2 Recall stored data from the data log
1. Press Recall Data in the Main Menu.
2. Press Data Log.
3. Press Filter: On/Off.
A listing of the stored data is displayed.
37
Standard Operations
4. The function Filter Settings is used to search for specific
items.
5. Highlight On to turn on the filters to select data by
•Sample ID
•Operator ID
•Start Date
•Parameter
or any combination of the four.
6. Press OK to confirm the selection. The chosen items are listed.
7. Press View Details to get more information.
5.3.1.3 Send data from the data log
Data is sent from the data log as CSV (Comma Separated Value)
files through a USB memory stick to a file named DATALOG. The
file can then be processed using a spreadsheet program. The file
name will be formatted as:
DLYear_Month_Day_Hour_Minute_Second. CSV.
To send data to a Printer, see section 5.2.6.2 on page 33.
1. Plug in the USB device (Figure 2 on page 13).
2. Press Recall Data from the Main Menu. Press Options and then the PC & Printer icon.
3. Select the data to send to the memory stick and press OK.
Note: The number in parenthesis is the total number of data sets
assigned to this selection.
To send measurement data to a PC:
The optional HACH Data Trans software must be installed on the
PC, see section 5.2.6.3 on page 33.
38
5.3.1.4 Delete stored data from the data log
1. Press Recall Data in the Main Menu.
2. Press Data Log>Options>Delete.
3. Highlight Single Point or Filtered data or All data and press
Standard Operations
OK to confirm.
Note: The number in parentheses is the total number of data sets
assigned to this selection.
39
Standard Operations
5.3.2 Time Course
5.3.2.1 Data Storage from Time Course
The instrument can store 20 Time Course Data sets. The data can
be stored manually at the user's discretion after viewing the data.
1. Press the Store icon in the ”Options“ menu from Time Course
mode after a reading is taken.
The Store Data list opens.
2. Press Store to save the current scan to the highlighted
numbered line. A run can also be overwritten.
5.3.2.2 Recall Stored Data from Time Course
1. Press Recall Data in the ”Main Menu“.
a. Select Time Course to call up the required data.
b. If you are already working in this program, press Options,
More... and then Recall Data.
40
Standard Operations
2. Press Graph to look at details.
Note: Press View Summary to return to the Recall Data list.
3. Press Table to look at details.
Note: Press View Summary to return to the Recall Data list.
5.3.2.3 Send Data from Time Course
Option 1:
4. Press Recall Data in the ”Main Menu“ and then Time Course
to send the data to a USB memory stick or to a printer.
The files will be automatically send as CSV files (Comma
Separated Value) to a file ”TCData“ (Time Course Data).
The file name will be formatted as: ”TCData_X.csv”
X = number of scans (1–20)
For further processing use a spreadsheet program.
41
Standard Operations
Option 2:
5. Press Time Course and then Options, More... and Send Data
to send the data to a USB memory stick or to a printer.
The files will be automatically send as CSV files (Comma
Separated Value) to a file ”TCData“ (Time Course Data).
The file name will be formatted as:
”TCYear_Month_Day_Hour_Minute_Second.CSV“ (Time
Course Data)
For further processing use a spreadsheet program.
5.3.2.4 Delete Stored Data from Time Course
1. Press Recall Data from the ”Main Menu“ and then
Time Course or ”Options“ menu, More..., Recall Data:
A listing of the stored data is displayed.
2. Highlight any data you want to delete.
3. Press Delete in the ”Options“ menu and confirm with OK.
42
5.4Stored Programs
The instrument contains more than 200 programmed procedures.
They can be accessed through the Stored Programs menu.
5.4.1 Select a saved test/method; enter user-specific basic data
1. Press Stored Programs in the Main Menu to view an
alphabetical list of stored programs with program numbers.
The Stored Programs list will appear.
2. Highlight the required test.
Note: Select the program number by name or use the arrow keys to
scroll through the list quickly and highlight the program or press
Select by number to search for a specific program number. Use the
alphanumeric keypad to enter the test number and press OK.
3. Press Start to run the program. After a program is selected, the
screen for that parameter will appear.
Standard Operations
5.4.2 Stored program options
Note: All corresponding data (wavelength, factors and constants) are
already preset.
4. Follow the chemical procedures described in the corresponding
Procedures Manual.
1. From the Main Menu, select Stored Programs. Select the
necessary method and press Start.
2. Press Options for Parameter Setup. Refer to Table 4 for stored
program descriptions.
Table 4 Stored programs options
OptionsDescription
MoreFor further Options
43
Standard Operations
Table 4 Stored programs options (continued)
OptionsDescription
Store Off/On
% Trans/Conc/AbsTo switch to % transmittance, concentration or absorbance readings
With the Store On setting, all measurement data are stored automatically. With the Store Off
setting, no measurement data are stored.
Send Data icon /
Send Data
Timer icon
Dilution Factor Off/On
Standard Addition
Standard Adjust
Chemical FormSome of the stored tests/methods allow to select the chemical form and the measuring range.
Reagent Blank
Save as User ProgramTo store the selected parameters as a User Program, see section 6.1 on page 61.
Recall DataCall up saved measurement data or time courses, see section 5.3 on page 37.
To send Data to a printer, computer or USB memory stick (Type A)
This functions as a stopwatch. It helps to ensure that the steps of an analysis are correctly timed
(e.g. reaction times, wait times, etc., can be exactly specified). When the specified time has
elapsed, an acoustic signal is emitted. The use of the timer has no influence on the measurement
program.
A corrective dilution factor can be entered in order to take account of certain properties.
The number entered at the dilution factor prompt will be multiplied by the result to compensate for
the adjustment. For example, if the sample has been diluted by a factor of 2, enter 2. The default
setting of the dilution factor is turned off.
Note: When a dilution is in effect, the dilution icon will appear on the display.
This enables the accuracy of the measurements to be checked. The (working) procedure for a
test parameter contains a detailed explanation of how to use this function.
The (working) procedure for a test parameter indicates whether a standard adjustment is
necessary and, if so, how to proceed.
Some of the stored tests/methods include the "Reagent Blank" function. This enables the reagent
blank value to be added to or subtracted from, the subsequent readings. The reagent blank value
shifts the calibration curve along the y-axis, without changing the shape or gradient of the curve.
The effect corresponds to a y-axis intercept of the calibration straight line. This is made clear by
the following equation: Concentration = [(Conc. factor) * Abs] – (reagent blank value).
Instrument SetupBasic data of the instrument, see section 5.2 on page 27.
44
5.4.3 Use of program timers
Standard Operations
Some procedures do not require the use of timers. Other
procedures require several timers. These timers are
pre-programmed into each Stored Program, along with a
description of the activity to be performed during the timed period.
1. Press the Timer icon on the display.
2. Press OK to start the first timer.
The timer will count down on the screen.
3. To start the next timed activity for the Stored Program, press
the Timer icon and OK.
Note: Press Close to view the measurement screen while the timer is
running. The time will be shown left side bottom instead of the date.
Note: Press Cancel to stop the timer midway through the countdown.
Note: The Timer will beep when the end of the running time is
reached.
A general purpose timer is also available in many programs. When
the timer icon is visible, press the icon and select General Timer.
A new screen will appear. Enter the length of the timed interval and
press OK to start the timer. The timer will beep when the timed
interval ends.
45
Standard Operations
5.4.4 Set the dilution factor
The Dilution Factor function is used to calculate the original
concentration of a sample that has been diluted by a known ratio.
For example, dilution factor can bring the analyzed concentration
within the test range.
1. Press Options>More…>Dilution Factor.
The number entered at the dilution factor prompt will be
multiplied by the result to compensate for the adjustment.
For example, if the sample has been diluted by a factor of 2,
enter 2. The default setting of the dilution factor is turned off.
2. Press OK to confirm. Press again OK.
3. Press Return to return to the result display.
4. Confirm ’Store present reading again?’ to save the current
measured value if necessary.
Note: When a dilution is in effect, the dilution icon will appear on the
display.
Note: If using undiluted samples, set the dilution factor off again.
5.4.5 Run a standard adjust
46
The Standard Adjust function allows the calibration curve for a
stored program to be adjusted based on analysis of a known
standard solution. The Accuracy Check section of written
procedures often suggests a standard solution concentration for
this purpose.
Standard Operations
Read a standard before setting Standard Adjust to On.
1. Follow the entire procedure, using a known standard for the
sample.
2. After reading the concentration, press
Options>More>Standard Adjust.
3. If Standard Adjust is set to Off, turn it On.
The Current Reading will show the concentration. The box on
the right will show the default standard value for the test, as
mentioned in the procedure.
4. If the measurement used a standard concentration that is
different from the one displayed in the box, press the box on the
right to enter a different standard value and enter the new
value. Press OK to confirm.
5.4.6 Set the chemical form
5. Press Adjust to enable the Standard Adjust. The Standard
Adjust icon will appear.
Note: The adjustment must be within certain limits, which vary with
each program. The allowable percentage is shown after "Adjustment“.
Note: When a Standard Adjust is in effect, the Standard Adjust icon
will appear on the display.
Some Stored Programs allow a variety of chemical forms to
be selected.
Press the unit (e.g. mg/L) or the chemical representation of the
evaluation form (e.g. Al
3+
). A list of available evaluation forms is
displayed. Select the required form by pressing the corresponding
entry in the list.
Note: When the program is exited, the evaluation form reverts to the
standard setting.
An alternative way of changing the standard setting:
47
Standard Operations
1. Press Options>More>Chemical Form.
2. Select the Chemical Form.
Note: The stoichiometric conversion of the measurement result is
carried out automatically.
Note: The selected Chemical Form will appear on the display. Test
results will be calculated and displayed in this chemical form.
5.4.6.1 Change of the default setting of the chemical form
1. Insert the sample cuvette or blank (depending on the working
procedure) into the cell compartment.
2. In the result display, press Options>More>Chemical Form.
3. A list of available evaluation forms appears. Select the new
default setting.
5.4.7 Run a reagent blank
4. Press Save as Default.
The current result and all further measurements will be displayed in
the new chemical form.
Some of the stored tests/methods include the "Reagent Blank"
function. This enables the reagent blank value to be measured and
then taken into account in calculating the measurement result.
Measurement/analysis of a reagent blank:
1. Prepare the test/method in accordance with the (working)
procedure. Instead of a sample, deionized water is used to
determine the reagent blank value.
2. Select the test. If required by the (working) procedure, insert
the cell with deionized water into the cell compartment. Press
Zero.
3. Insert the prepared sample cell into the cell compartment. Press Read. The result is displayed.
48
Standard Operations
4. Press Options>More>Reagent Blank.
5. Press On to highlight the Reagent Blank function.
6. The concentration shown on the key is the measured value of
the reagent blank. To use this value for more analyses of this
parameter, press OK.
7. To enter a previously recorded reagent blank value, press the
key and use the alphanumeric keypad.
8. Press OK.
Note: The Reagent Blank function is deactivated when the measurement
program is left. To use the same blank value later for other tests using the
same reagent lot, enter the value per step 7.
Note: The results calculated using the reagent blank value must lie within
the limits of the measuring range of the test/method.
5.4.8 Analysis of samples
Note: The reagent blank icon is shown in the result display (see arrow)
when the function is active.
1. Press Stored Programs and select a program.
2. Insert the blank cuvette/cell into the cuvette/sample cell holder.
3. Press Zero.
49
Standard Operations
4. Remove zero solution and insert sample cuvette/cell into the
cell compartment.
5. Press Read. The result will be displayed.
6. For data storage, see section 5.3.1 on page 37.
5.4.9 Add stored programs to the favorite programs list
The Favorites menu simplifies test selection by creating a list of the
most frequently used tests from the Stored Programs and
User Programs.
1. Press Stored Programs in the Main Menu. The Stored
Programs list will appear.
2. Highlight the selection by pressing it or Select by Number to
search for the program by number.
3. Press Add to Favorites and press OK to confirm.
The program can now be selected from Favorite Programs
menu in the Main Menu.
50
5.5Barcode Programs
Standard Operations
A special barcode reader in cell compartment #1 automatically
reads the barcode on the 13 mm cuvette/vial as the cuvette/vial
completes a single rotation. The instrument uses the barcode
identification to automatically set the correct wavelength for the
analysis and calculates the result immediately with the help of the
stored factors.
In addition, measured values are recorded at 10 different positions
during a rotation. A special outlier-elimination program is run and
then the average of the measured values is calculated. Cuvette/vial
errors and soiling are recognized and highly precise results are
obtained.
Figure 11 Cell compartment #1 for barcode 13 mm cuvettes/vials
5.5.1 Complete a barcode 13 mm test/vial
1. Insert the light shield in cell compartment #2.
2. Prepare the barcode 13 mm cuvette/vial in accordance with the
working procedure and insert the cuvette/vial in cell
compartment #1.
•When a coded cuvette/vial is placed in cell compartment #1
(Figure 11), the corresponding measurement program is
automatically activated in the Main Menu.
51
Standard Operations
5.5.2 Select the measuring range
•Otherwise, press Barcode Programs in the Main Menu.
and insert the blank or sample cuvette/vial (depending on
the working procedure) in cell compartment #1.
The measurement is started automatically and the results are
displayed.
To evaluate other cuvette/vial tests and other parameters, insert the
prepared cuvette/vial into the cell compartment and read the result.
Some tests can be used for different measuring ranges. After the
sample cuvette/vial has been inserted, a list of the different
measuring ranges is displayed.
Select the required measuring range by highlighting the appropriate
line.
Press Start Permanent if this measuring range is to apply to all
subsequent measurements.
Changing the standard setting
In the result display, press Options>More>Permanent: On. The
key will change to Permanent: Off.
5.5.3 Select the chemical evaluation form
The chemical form of the test result of a number of parameters can
be selected individually.
In the result display, press the unit (e.g. mg/L) or the chemical
representation of the evaluation form (e.g. PO
possible evaluation forms is displayed, from which the required
form can be selected. Press OK to confirm.
Another way of changing the standard setting is:
1. In the result display, press Options>More>Chemical Form.
A list of available evaluation forms appears.
2. Select the required chemical form and press OK to confirm.
–
3
–P). A list of
4
52
Note: The selected chemical form is displayed, but does not become
the default. To change the default, see section 5.5.3.1.
5.5.3.1 Change of the default setting of the chemical form
1. Insert the blank or sample cuvette/cell (depending on the
working procedure) into the cell compartment.
2. In the result display, press Options>More>Chemical Form.
3. A list of available evaluation forms appear. Select the new
default setting.
4. Press Save as Default.
The current result and all further measurements will be displayed in
the new chemical form.
5.5.4 Basic test-specific and sample-specific data settings
Press Options to change test or sample specific settings.
Standard Operations
Table 5 Barcode program options
OptionsDescription
MoreFor further Options
Store Off/On
Abs % TransTo switch to % transmittance or absorbance readings
Send Data icon /
Send Data
Timer icon
Dilution Factor Off/On
Chemical FormSome of the stored tests/methods allow to select the chemical form and the measuring range.
EditTo modify an existing program
With the Store On setting, all measurement data are stored automatically. With the Store Off
setting, no measurement data are stored.
To send data to a printer, computer or USB memory stick (Type A)
This functions as a stopwatch. It helps to ensure that the steps of an analysis are correctly timed
(e.g. reaction times, wait times, etc., can be exactly specified). When the specified time has
elapsed, an acoustic signal is emitted. The use of the timer has no influence on the measurement
program.
A corrective dilution factor can be entered in order to take account of certain properties.
The number entered at the dilution factor prompt will be multiplied by the result to compensate for
the adjustment. For example, if the sample has been diluted by a factor of 2, enter 2. The default
setting of the dilution factor is turned off.
Note: When a dilution is in effect, the dilution icon will appear on the display.
Note: If undiluted samples are used, set the dilution factor off.
53
Standard Operations
Table 5 Barcode program options (continued)
OptionsDescription
Save as User ProgramTo store the selected parameters as a User Program, see section 6.1 on page 61.
Recall DataCall up saved measurement data or time courses, see section 5.3 on page 37.
Instrument SetupBasic data of the instrument, see section 5.2 on page 27.
5.5.5 Sample blank
Turbidity and color in the sample matrix can falsify the results of a
photometric analysis. The interference factors come from the
sample or are created by reactions with the reagents.
The influence of turbidity and/or color can be eliminated or reduced
by taking a sample blank reading.
In the barcode mode, a special cuvette/vial (TNT919) containing
the sample blank is placed in cell compartment #1 after the sample
reading has been taken and is automatically measured. The
sample reading is then corrected by adding or subtracting the blank
value. The final result is displayed, with the message "After blank
value corr.".
Some barcode tests do not require a sample blank value to be
determined, as turbidity and color are dealt with during the test
procedure. Other barcode tests use the reagent test vials to
determine the sample blank. Please refer to the Hach procedure for
method-specific instructions.
54
5.5.6 Update/edit barcode tests
5.5.6.1 Manual update of a barcode test
Using the data provided in the barcode, the instrument
automatically sets the measurement wavelength and factors. If a
discrepancy is detected between the barcode data and the stored
data or a new test is identified, the instrument requests an update.
Data types are described in Table 6.
Table 6 Program update options
Program pointDescription
NameName of the analysis parameter
Standard Operations
VersionAn abbreviation assigned by the user or the version number, etc. is entered here
Exact definition of the test: the number of wavelengths at which measurements are carried out,
Measurement process
FormulaDefinition of the evaluation formulas with which the test result is calculated
Variables
Timer 1, Timer 2,
Timer 3, Timer 4
the number of absorbance measurements needed, the keys to be used, any waiting periods
between measurements, etc.
The number of variables shown in the display depends on the definition of the measurement
process and the formulas. Input of the numerical values of the wavelengths, factors, constants,
etc.
This function can be used to enter abbreviations and defined times for up to four timers. Highlight
the appropriate line and press Edit.
The timers are activated or deactivated with the control boxes on the left of the display. In the
next column, a selection can be made from a list of names that describe the corresponding work
step. In the third column, the times for each active timer are entered.
Enter the Name
Check the working procedure in advance to determine whether this
point needs to be changed.
1. Highlight the line containing Name and press Edit. Use the
alphanumeric keypad to enter the name indicated in the
working procedure.
2. Press OK to confirm.
55
Standard Operations
Enter the Version
1. Highlight the line containing Versi on and press Edit. Use the
alphanumeric keypad to enter the name indicated in the
working procedure.
2. Press OK to confirm.
Enter the Measurement Process
For detailed information on entering the measurement process, see
section 6.1.2 on page 72.
Check the working procedure in advance to determine whether this
point needs to be changed.
1. Highlight the Measurement Process line in the data overview
and press Edit.
2. Press Edit, select the sequence to be edited and press Delete.
3. Press New and use the alphanumeric keypad to enter the
process indicated in the working procedure.
Enter the Formula, Concentration Units, Designation,
Measuring Ranges
For detailed information on entering the Formula, see section 6.1.2
on page 72.
Check the working procedure in advance to determine whether this
information needs to be changed.
1. Highlight the Formula line in the data overview and press Edit.
2. Select the formula, press Edit and use the alphanumeric
keypad to enter the data specified in the working procedure (for
C1=, C2=, units, name, measuring range limits, etc.).
56
Standard Operations
Enter the Variables (Factors, Wavelength and Coversion
Factors)
For detailed information on entering the Variables, see section 6.1.2
on page 72.
Check the working procedure in advance to determine whether this
point needs to be changed.
1. Highlight the Vari ables line in the data overview and press
Edit.
2. Select the variable, press Edit and use the alphanumeric
keypad to enter the data specified in the working procedure
(for F1, F2, λ1, U1 etc.). Confirm each entry with OK.
Abbreviation of Variables:
F1: Factor 1
F2: Factor 2
λ1: Wavelength 1
U1: Conversion Factor 1 for the first chemical form
U2: Conversion Factor 2 for additional chemical forms etc.
Timer 1, Timer 2, Timer 3 and Timer 4:
The timers are activated or deactivated with the control boxes on
the left of the display. In the next column, a selection can be made
from a list of names that describe the corresponding work step. In
the third column, the times for each active timer are entered.
57
Standard Operations
5.5.6.2 Update an existing barcode test
Note: Only in very few cases does the revision of a test require all test
specifications to be updated.
Option 1: Manual update of test data
1. Insert the light shield in cell compartment #2.
2. The instrument needs to be at the Main Menu or press
Barcode Programs in the Main Menu.
3. Insert the blank or sample cuvette/vial (depending on the
working procedure) in cell compartment #1.
4. After "Barcode controlnumber?", a data overview is displayed,
including the specifications of the test that is to be revised.
Note: The working procedures of the corresponding test contains the
new data (Wavelength, Factors, Measurement Ranges, Conversion
Factor for chemical Form, etc.).
5. Highlight the corresponding line and press Edit.
6. When finished programming, press Store and then Cancel to
return to the User Programs menu.
7. Press Start to begin the test with the new data.
Option 2: Manual check/revision of test data
If an incorrect entry was made during the data update and is not
recognized via the barcode of the cuvette/vial (e.g. an incorrect
measuring range), another option is available for checking and if
necessary revising, the test data.
1. Insert the light shield in cell compartment #2.
2. The instrument need to be at the Main Menu or press Barcode
Programs in the Main Menu.
3. Insert the blank or sample cuvette/vial (depending on the
working procedure) in cell compartment #1.
4. Press Options>More>Edit.
58
5. A data overview is displayed, including the specifications of the
test that is to be revised. Compare the displayed data with the
data in the working procedure and edit the displayed data in
accordance with the working procedure.
5.5.6.3 Program a new test
Standard Operations
1. The instrument needs to be at the Main Menu or press
Barcode Programs in the Main Menu.
2. Insert the sample cuvette/vial or blank (depending on the
working procedure) into the cell compartment.
3. After "Program not available", a data overview is displayed,
including the specifications of the test that is to be revised.
The working procedures of the corresponding test contains the
new data (Wavelength, Factors, Measurement Ranges,
Conversion Factor for chemical Form etc.).
4. Highlight the corresponding line including the specifications of
the test that is to be programmed and press Edit.
5. To finish programming, press Store and then Cancel to return
to the Main Menu.
6. Press Start to begin the test with the new data.
5.5.7 Upgrade of the instrument software
To obtain the software for the update from the Internet at
www.hach.com:
1. Go to http://www.hach.com.
2. On the DR 2800 product page, click Lab System
Software/Software Update Downloads under Downloads.
3. Locate the appropriate download and follow the prompts for
saving the file(s) to the USB memory stick.
4. On the DR 2800, press Instrument Update in the System
Checks menu.
5. Connect the USB stick to the USB interface on the DR 2800
(section 3.4 on page 13). Press OK.The link is established
automatically and the software is updated.
59
Standard Operations
6. Press OK to return to the System Checks menu.
Note: When the instrument software has been updated, a prompt to restart
the instrument is displayed.
60
Section 6Advanced Operations
6.1User Programs
User programs provide the opportunity to complete "made to
measure" analysis.
The User Programs database is empty when the instrument leaves
the factory and is used to accommodate programs created by users
specific needs. Here are a few examples of entries:
•Programming of user-created procedures. The analysis
procedure must be developed first, before it can be
programmed. The user must define or determine the program
sequences, calculation formulas, measurement wavelengths,
factors, measuring range limits, etc.
•Modified tests
•Assignment of user programs to the favorites menu for
frequently used tests.
•Creation of a specific selection of methods and tests.
Press User Programs in the Main Menu and then
Program Options. The Program Options menu contains several
input and editing options (Table 7):
Table 7 Program options user programs
OptionsDescription
Select New to program a new user program.
New
Add to favoritesSelect Add to favorites to add an existing user program to the list of frequently used programs.
EditSelect Edit to modify an existing program
Delete
Note: The first time Program Options is selected, only the New option is available. The other
options remain inactive (grey) until the first program has been created.
Select Delete to remove a program from the list of user programs. The program will be
simultaneously deleted from the Favorites list.
6.1.1 Program a user method
All input steps and their significance and options are explained in
the following sections.
1. Select New in the Program Options menu.
61
Advanced Operations
Program Number:
Specific test number, with which the program can subsequently be
called up from the selection list in the User Programs menu or the
Favorites menu.
2. Use the alphanumeric keypad to enter a program number
between 950 and 999. The lowest available number appears
automatically.
3. Press OK.
Note: If the program number is already assigned to another user
program, a message appears, asking whether the existing program
should be replaced. Press OK to overwrite the existing program.
Program Name:
4. Use the alphanumeric keypad to enter a program name. The
name can be a maximum of 28 characters long.
Program TypeDescription
5. Press Back to go back to the previous program point or press
Next to continue with the input of the program data.
Program Type:
6. Select the required option (Table 8) and press Next.
7. If the Single Wavelength (section 6.1.1.1 on page 63) or Multi
Wavelength (section 6.1.1.2 on page 64) is selected, define the
unit, wavelength, absorbance formula, wavelength λx,
concentration factor Kx, concentration resolution, chemical
form and calibration equation.
More information on Free Programming parameters see
section 6.1.2 on page 72.
Table 8 Program descriptions
Single WavelengthMeasurements at a defined wavelength
Multi Wavelength
Free Programming
In the Multi Wavelength mode, absorbance values can be measured at up to four wavelengths
and the results can be mathematically processed to obtain sums, differences and relationships.
This is an advanced form of programming for original user-developed methods. In Free
Programming, the user defines the measurement process, variables and calculations involved in
obtaining a reading.
62
6.1.1.1 Single wavelength settings
Advanced Operations
If the Single Wavelength mode is selected, the following
parameters can be defined:
Units:
Select the required unit from the list and press Next.
Note: Units of measure not included in this list can be added in the edit
program under Program Options, Edit. Select Units, Edit and then New.
Wavelength (single wavelength program type):
Use the alphanumeric keypad to enter the measurement
wavelength. The entered wavelength must be in the range from
340–900 nm.
Press Next to proceed.
Concentration resolution (number of decimal places)
Select the required number of decimal places from the displayed
list and press Next.
63
Advanced Operations
6.1.1.2 Multi wavelength settings
Chemical form:
Enter the chemical formula used in the display to represent the
analysis parameter.
Use the alphanumeric keypad to enter the chemical form and press
Next to enter Calibration settings.
If the Multi Wavelength mode is selected, the following parameters
can be defined:
Units:
Select the required unit from the list and press Next.
Note: Units of measure not included in this list can be added in the edit
program under Program Options, Edit. Select Units, Edit and then New.
Absorbance formula (multi-wavelength program type):
The Absorbance Formula menu is used to define the wavelengths
and the coefficients used in the formula. The absorbance formula
defines the calculation for the multi-wavelength measurement.
Press the appropriate key to edit the input.
Press the Formula key.
In the displayed list, select the formula for the program and press
OK to confirm.
64
Advanced Operations
List of available absorbance formulas
A
is the absorbance at wavelength 1,
1
A
is the absorbance at wavelength 2 and so on
2
K
is the factor at wavelength 1,
1
K
is the factor at wavelength 2 and so on
2
If a subtraction has to be completed, the factors can be entered
with a minus sign.
Wavelength λ
Press a λ
wavelength. Press another λ
:
x
key and use the alphanumeric keypad to enter a
x
key and enter the next wavelength. If
x
necessary, repeat until all the wavelengths for the formula have
been entered. The wavelengths must be in the range from
340–900 nm. Press OK.
Concentration Factor Kx
Multiplication factor for converting absorbance values into
concentration values.
Press a factor key and use the alphanumeric keypad to enter a
factor. If the formula includes more than one factor, press another
factor key and enter another factor and repeat until all the factors
have been entered. Press OK. When all the relevant data have
been entered, press Next.
Note: Up to 5 digits can be entered, including a maximum of 4 decimal
places to the right of the decimal.
Concentration resolution (number of decimal places)
Select the required number of decimal places from the displayed
list and press Next.
65
Advanced Operations
Chemical form:
Enter the chemical formula used in the display to represent the
analysis parameter.
Use the alphanumeric keypad to enter the chemical form and press
Next to enter Calibration settings.
6.1.1.3 Calibration settings for single and multi wavelength mode
A method is calibrated by determining the absorbance values of
several standard solutions of known concentration.
There are three ways to create and store a calibration curve.
Instructions for each method follows (Table 9).
Table 9 Calibration settings
ModeDescriptions
A calibration table is created by entering the concentration values and the absorbance values of
Enter values
Read Standards
Enter Formula
the analyte solution. The absorbance values are plotted versus standard concentrations and the
calibration curve is displayed as a graph (page 66).
A calibration table is created by entering the concentration values of the standard solutions and
then measuring the absorbance of the analyte solutions. The absorbance values are plotted
versus standard concentrations and the calibration curve is displayed as a graph (page 68).
If the calibration curve can be determined from the mathematical relationship between
concentration and absorbance by linear regression, etc., the corresponding formula can be
selected (linear, 2nd or 3rd order polynomial) from a list and the appropriate factors can be
entered (page 69).
Calibration by entering calibration values
Enter concentration/absorbance
1. Select the Enter Values and press Next.
66
Advanced Operations
2. To enter the standard concentrations and corresponding
absorbance values in the displayed table, press the "+" symbol.
Use the alphanumeric keypad to enter the values.
Press OK and enter the corresponding absorbance value.
Press OK.
The entered data are displayed in the table. Repeat the
sequence for each data point to enter it.
3. To change a value in the table, highlight the appropriate line,
press the unit key (e.g. mg/L) or Abs and enter the changed
value via the alphanumeric keypad.
4. When the data have all been entered, press Graph.
Note: The correlation coefficient (r2) is shown on the left below the
axes.
5. The linear equation corresponds to the standard setting. Press
Next Curve to display the polynomial 2nd order curve. Press
Next Curve again, to display the polynomial 3rd order curve.
6. Press Force 0 to change the setting from Off to On. The curve
now passes through the origin of the coordinate system.
Note: This may have an adverse effect on the correlation coefficient
2
).
(r
7. Press Table to display the table again.
8. When the table has been completed and the curve type has
been chosen, press Done when the graph is displayed or Exit
when the table is displayed. Go to section 6.1.1.4 on page 70.
67
Advanced Operations
Calibration by reading standards
1. Press Read Standards and press Next.
2. To enter the standard concentrations in the displayed table,
press the "+" symbol. Use the alphanumeric keypad to enter
the standard concentration. Press OK.
3. Press the "+" symbol again (see arrow) and enter the next
standard concentration. Repeat this sequence until all standard
conentrations (maximum of 24 solutions) have been entered.
4. Highlight the line with the appropriate concentration and insert
the cuvette with the corresponding standard solution.
5. Insert the zero solution into the cell compartment. Press Zero.
6. Insert the first standard solution into the cell compartment.
Press Read.
Insert the second standard solution into the cell compartment.
Press Read.
Repeat this sequence until all the standard solutions have been
measured (maximum of 24 solutions).
The entered and measured data are displayed in the table.
Note: To delete a standard concentration, highlight the appropriate line and
press the Delete icon.
The timer icon shown in the display helps to ensure, when
necessary, that the steps of an analysis are correctly timed (e.g.
reaction times, wait times, etc., can be exactly specified). When the
specified time has elapsed, an acoustic signal is emitted. The use
of the timer has no influence on the measurement program.
7. When the data have all been entered and the measurements
have all been completed, press Graph.
68
8. The linear curve corresponds to the standard setting. Press
Next Curve to display the polynomial 2nd order curve.Press
Next Curve again to display the polynomial 3rd order curve.
Advanced Operations
9. Press Force 0 to change the setting from Off to On. The curve
then passes through the origin of the coordinate system.
Note: This may have an adverse effect on the correlation coefficient
(r2).
10. Press Table to display the table again.
11. When the table has been completed and the curve type has
been chosen, press Done when the graph is displayed or Exit
when the table is displayed. Go to section 6.1.1.4 on page 70.
Calibration by entering the formula
1. Press Enter Formula and press Next.
2. Press the formula key.
A list of available formulas (linear and 2nd and 3rd order
polynomial) is displayed. Up to 4 coefficients can be entered,
depending on the selected formula. Press the required formula.
3. Depending on the selected formula, the required coefficients (a,
b, c...) are displayed. Press the coefficient keys and enter the
corresponding values via the alphanumeric keypad. After each
entry, press OK to confirm.
Note: The coefficients can have 5 digits and can have a positive or a
negative sign.
69
Advanced Operations
6.1.1.4 Store a user program
The input of the basic data is complete. An overview of the variable
program data is displayed.
1. To enter more specifications or change existing ones, highlight
the appropriate line and press Edit.
2. Select Store to save the user program.
3. Press the PC & Printer icon to send the program data to a
printer or to a USB memory stick (connect the USB memory
stick to the USB interface first).
Note: The test data will be formatted in the preinstalled PrgData folder
as a .csv and as a .lst file.
4. To transfer the program data from one instrument to another
create a new folder on the USB memory stick named dbhlc
and copy the .lst files into this folder. (see section 6.7.2 on
page 98).
6.1.1.5 Additional user-defined parameters and functions
In addition to the previously defined basic data, additional
parameters and functions can be defined for user programs:
•measuring range upper and lower limits
•timer functions
•chemical forms
Upper and lower limits of the measuring range
It is possible to enter a maximum (upper) and minimum (lower)
measurement value. An error message is displayed if a reading is
above the upper limit or below the lower limit.
1. Highlight the appropriate line in the overview of the program
data and press Edit.
2. Set On and press the 0.000 key to enter the measuring range
limit. Confirm the entry by pressing OK.
70
Advanced Operations
Timer 1 / Timer 2 / Timer 3 / Timer 4:
This function can define time intervals for up to four timers. Timer
designations such as Shake, Wait and Swirl can be assigned.
1. Highlight the appropriate line in the overview of the program
data and press Edit.
2. The timers are activated or deactivated with the check boxes in
the left part of the display. In the next column, a selection can
be made from a list of names that designate the corresponding
work step. In the third column, the times for each timer are
entered (in mm:ss).
Chemical form 2 / chemical form 3 / chemical form 4:
If a Chemical Form 1 has been defined, up to three additional
alternative forms can be entered here.
1. Highlight the appropriate line in the overview of the program
data and press Edit.
2. The chemical forms are activated or deactivated with the check
boxes in the left part of the display.
3. Press the left key to enter another chemical form with the
alphanumeric keypad and press OK to confirm. Press the right
key to enter the conversion factor to calculate the concentration
of the additional chemical form from the concentration of
Chemical Form 1 and press OK to confirm.
4. Press Store to save the program data. Press Cancel to return
to the Main Menu.
71
Advanced Operations
6.1.2 Free programming program type
Free Programming is an advanced option for entering original
user-developed methods. When the Free Programming option is
selected, an overview of the specifications of the programmed test
is displayed. Each input option can be modified to develop the user
method. Refer to Table 10 for more information. To modify an input
option, select the appropriate line and press Edit.
Important Note: Perform steps 1–6 of Programming a New User
Method (section 6.1.1 on page 61) before proceeding with Free
Programming.
Highlight the line containing the program point that is to be edited or
defined and press Edit.
Table 10 Definitions of the program points
Program pointDescription
NameName of the analysis parameter
VersionAn abbreviation or version number assigned by the user is entered here.
Exact definition of the test: the number of wavelengths at which measurements are made, the
Measurement Process
FormulaDefinition of the formulas with which the test result is calculated.
Variables
Timer 1, Timer 2,
Timer 3, Timer 4
number of absorbance measurements needed, the keys to be used, any waiting periods between
measurements, etc.
The number of variables shown in the display depends on the definition of the measurement
process and the formulas. Input of the numerical values of the wavelengths, factors, constants,
etc.
Used to enter abbreviations and defined times for up to four timers. Highlight the appropriate line
and press Edit. The timers are activated or deactivated with the control boxes on the left of the
display. In the next column, a selection can be made from a list of names that describe the
corresponding work step. In the third column, the times for each active timer are entered.
6.1.2.1 Measurement process
The measurement process defines the handling and the
measurements of the test:
•At which and how many wavelengths should measurements be
completed
72
•How many absorbance measurements must be completed?
•When should the zero measurement and the sample
measurement be completed?
•Are waiting times necessary between measurements?
•Should individual program sequences be repeated?
The elements of a measuring sequence, such as zero and sample
measurements and the timer(s) (reaction times, waiting times, etc.)
are individually defined.
6.1.2.2 Enter a new element of a measuring sequence
Important Note: Each component of the measurement process
must be entered in the order in which it will be completed.
1. Highlight the Measurement Process line in the data overview
and press Edit.
2. Press Edit again and then New.
Advanced Operations
Content and definition of the keys
[Z] key / Zeroing
1. Press the [Z] key to program a zero measurement. Confirm with OK.
2. Press New and then Zeroing... and use the alphanumeric
keypad to enter the wavelength at which the zero measurement
is to be completed. Press OK and confirm the input by pressing
OK again.
3. If zero measurements are to be carried out at a number of
wavelengths, repeat the above two steps for each wavelength.
Note: The entered measurement sequence is displayed.
73
Advanced Operations
Process Timer key
1. Press the Process Timer key to enter any waiting, reaction or
handling times that have to be taken into account. Use the
alphanumeric keypad to enter the time. Press OK and confirm
the input by pressing OK again.
Note: This time is integrated into the measurement process.
Note: The entered measurement sequence is displayed.
[R] key / Reading...
1. Press the [R] key to program a measurement of the substance that is to be analyzed. Confirm with OK.
2. Press New and then Reading... and use the alphanumeric
keypad to enter the wavelength at which the measurement is to
be completed. Press OK and confirm the input by pressing OK
again.
3. If measurements are to be completed at a number of
wavelengths, repeat the above two steps for each wavelength.
Note: The entered measurement sequence is displayed.
{ } key
Elements of the measurement sequence that are to be repeated
are placed in brackets.
The left bracket "{"marks the start of the sequence that is to be
repeated and the right bracket "}" marks the end.
Note: The key showing the right bracket remains inactive until a left bracket
is entered.
1. Press {.
2. Press the key that defines the sequence that is to be repeated:
[Z] or [R]. Confirm with OK.
3. Press New and then press Zeroing... or Reading... and use
the alphanumeric keypad to enter the wavelength at which the
measurement should be completed. Press OK and confirm the
input by pressing OK again.
74
4. Press } to end the sequence.
Note: If an action such as a zero measurement recurs at different
stages of a measurement sequence, the series of actions is numbered
sequentially (e.g. Z1, Z2, etc.).
Deleting an element of a measuring sequence
Select the appropriate line and press Delete. The element is
deleted.
Inserting an element of a measuring sequence
Select the line in the measuring sequence where the insertion is be
made and press New.
A new element can be entered at the selected position.
When the input is complete, press OK in the "Measurement
Process" display. The data overview is then displayed.
6.1.2.3 Enter the calibration formula (evaluation formula)
The calibration formula (evaluation formula) defines the calculation
and display of intermediate and final results. The previously defined
elements of the measuring sequence are the basis for calculating
the concentrations.
Advanced Operations
Enter calibration formula C1
1. Highlight the Formula line in the data overview and press Edit.
2. Highlight the line C1: Off and press Edit.
3. Select C1: Off again and press Edit. The display changes to
C1: On.
4. Highlight the next line C1 = to define the formula and press
Edit.
75
Advanced Operations
ScreenKeyDescription
Refer to Table 11 for detailed information on the Edit
formula keys.
Note: The evaluation formula is built up successively in the display in
accordance with the input.
Note: The arrow key deletes the most recently entered element of the
formula.
Table 11 Edit formula key descriptions
Press the Abs/Variables key to select, from the
displayed list, the required element of the defined
Abs/Variables
measuring sequence and therefore the corresponding
measurement wavelength, so that this can be taken into
account in the formula.
76
New NumberPress New Number to enter a new factor or constant.
Table 11 Edit formula key descriptions (continued)
ScreenKeyDescription
Press +– ÷x, to enter a mathematical operation.
Select the operation and press OK to confirm.
The available choice of mathematical operations
depends on the defined formula.
This means that functions such as "( )" or "ln"/"log" etc.
are only active if a term in parentheses or the calculation
of a logarithm is mathematically permissible in the
defined formula (this also applies to the basic
mathematical operations).
+–
÷x
The following mathematical operations are available:
•+ (Addition)
•– (Subtraction)
•÷ (Division)
•x (Multiplication)
•^ (Exponent)
•Ln (Natural logarithm)
Advanced Operations
•Log (Common logarithm)
Press >=< to include logic statements/links/conditions in
the formula. The following functions are available:
= (Equal to), < (Less than), > (Greater than), <= (Less
than or equal to), >= (Greater than or equal to), IF,
THEN, ELSE
>=<
When the evaluation formula C1 has been entered
completely, press OK to confirm. Press OK again to
return to the Formula display.
When the formula C1 has been entered and confirmed,
the parameter name, upper and lower limits of the
measuring range and display result (yes, no) can be
entered.
Enter the next calibration formula (C2 or Cn)
1. Highlight C2: Off and press Edit.
2. Select C2: Off again and press Edit. The display switches to
C2: On.
3. Highlight the next line C2 = to define the formula and press
Edit.
In addition to the keys described in Table 11, only one other
function is available:
77
Advanced Operations
6.1.2.4 Enter variables
Select Conc. key
If an already defined formula, in this case C1, is to be taken into
account in the formula for C2, press Select conc.
Enter the number of the formula (e.g. 1 for C1) and press OK to
confirm.
Cn can now be linked with a mathematical operation.
Note: The Cn concentrations that are to be calculated are numbered in
sequence: C1, C2, C3, etc.
Note: When the first Cn formula has been defined, the Formula list is
automatically extended by Cn+1.
1. Highlight the Variables line in the data overview and press
Edit.
2. Select the variable to be edited, press Edit and use the
alphanumeric keypad to enter the data specified in the working
procedure (for F1, F2, λ1, U1 etc.). Press OK to confirm each
entry.
Abbreviation of Variables:
F1: Factor 1
F2: Factor 2
λ1: Wavelength 1
U1: Conversion Factor 1 for the first chemical form
U2: Conversion Factor 2 for the further chemical form etc.
6.1.2.5 Save a free programming user program
1. Press Store to save the entered data. The data can be stored
under any data point (Measurement sequence, Formula, Timer,
etc.).
2. Press the PC & Printer icon to send the program data to a
printer or to a USB memory stick (connect the USB memory
stick to the USB interface first).
Note: The test data will be formatted in the preinstalled PrgData folder
as a .csv and as a .lst file.
3. To transfer the program data from one instrument to another
create a new folder on the USB memory stick named dbhlc
and copy the .lst files into this folder (see section 6.7.2 on
page 98).
78
6.1.3 Select a user program
1. Press User Programs in the Main Menu to view an
alphabetical list of user programs with program numbers.
The User Programs list will appear.
Note: Use the scroll bar to scroll through the list quickly.
Note: Highlight the selection by pressing it or press Select by
Number to search for the program by number.
2. Press Start to run the program.
6.1.4 Add, edit and delete user programs from the favorites list
The most frequently used tests/methods in the User Program menu
can also be added to the list of favorites to simplify their selection.
Advanced Operations
1. Press User Programs in the Main Menu.
The User Programs list will appear.
2. Highlight the selection by pressing it or press Select by
Number to search for the program by number.
Note: Use the scroll bar to scroll through the list quickly. Use the
alphanumeric keypad to enter the test number (program number) and
press OK to confirm.
3. Press Program Options.
4. Press Add to Favorites, Edit or Delete and press OK to
confirm.
Note: If the stored program is deleted in User Programs, it will also be
deleted in Favorites Programs.
79
Advanced Operations
6.1.4.1 Add to Favorites
6.1.4.2 Edit
1. Press Add to Favorites and press OK to confirm.
The program is added to the Favorites.
1. Press Edit and press OK to confirm.
6.1.4.3 Delete
An overview of the specifications of the programmed test is
displayed. More information about the input options is provided
in the section 6.1.2 on page 72.
1. Press Delete and press OK to confirm.
The program is deleted from the list of User Programs.
Note: If the stored program is deleted in User Programs, it will also be
deleted in Favorites Programs.
80
6.2Favorite Programs
6.2.1 Recall a favorite program
Advanced Operations
The most frequently used tests/methods in the Stored Programs
menu and the User Programs menu can also be added to the list
of favorites to simplify their selection.
To add Stored Programs and/or User Programs to the favorites
list or the favorite programs, see section 6.1.4 on page 79.
1. Press Favorite Programs in the Main Menu.
The Favorite Programs list will appear.
2. Highlight the selection by pressing it.
Note: Use the scroll bar to scroll through the list quickly.
Note: Press Select by Number to search for the program by number.
Use the alphanumeric keypad to enter the test number (program
number) and confirm the input by pressing OK.
6.2.2 Delete a favorite program
3. Press Start.
1. Press Favorite Programs in the Main Menu.
The Favorite Programs list will appear.
2. Highlight the selection by pressing it.
Note: Use the scroll bar to scroll through the list quickly.
Note: Press Select by Number to search for the program by number.
Use the alphanumeric keypad to enter the test number (program
number) and confirm the input by pressing OK.
3. Press Remove Program and press OK to confirm.
Note: If a Favorite Programs is deleted, it will stay in the User
Programs or Stored Programs.
Note: If the stored program is deleted in User Programs, it will also be
The accuracy of measured values (their correspondence with the
actual concentration of the analyte in the sample) and their
precision (correspondence of the measurement results obtained
from several samples containing the same concentration of the test
analyte) can be determined or improved using the standard addition
method.
This method (also referred to as spiking) serves to identify
sample-specific interference factors, e.g. substances in the sample
that falsify the analysis (sample matrix effect), a defective
measuring instrument or contaminated reagents.
Method:
A defined amount (concentration) of a standard solution of the test
substance is added to the sample. The detection rate should be
close to 100%.
Probability that the measurement results are correct
is high.
Assumption: The analysis was falsified by
substances in the sample (sample matrix effect)
Test to determine whether a sample matrix effect
is present:
Use distilled water instead of the sample. Add
standard solution as described in the procedure.
Detection rateConclusion
Ions in the sample are
100%
interfering with the
analysis, causing false
results to be obtained.
No interfering ions -
≠ 100%
consider other
interference factors
Measures to identify other interference factors:
Checklist:
1. Check if the procedure is completed correctly:
.
82
a. Are the reagents added in the correct order?
b. Is enough time allowed for color development?
c. Is the correct glassware in use?
d. Is the glassware clean?
Advanced Operations
e. Does the test require the sample to be at a certain
temperature?
f.Was the pH of the sample in the correct range?
g. Is the pipette volume correct?
2. Check the used reagents by repeating the standard addition
procedure with freshly prepared reagents.
Detection rateConclusion
The originally used reagents were defective.
100%
If none of these measures resolves the problem, please contact the
manufacturer or a sales representative.
Check the standard solution:
Repeat the standard addition procedure with a
freshly prepared standard solution.
Detection rateConclusion
The originally used
100%
standard solution was
defective.
6.3.1 Complete a standard addition
Carry out the standard addition in accordance with the
corresponding procedure.
There are two different methods:
Peak volume (Standard addition):
Defined volumes of a standard solution are added step by step to
an already analyzed sample. The sample is measured again after
each addition.
Sample volume:
A defined volume of a standard solution of known concentration is
added to the volume of sample specified in the procedure and the
sample is measured after each addition. In most cases three
different standard solutions are prepared and the procedure is
repeated for each of them.
Note: The units and chemical forms used for the sample are used for the
standard solutions. Be careful to ensure that to use the correct units for
subsequent entries.
83
Advanced Operations
Peak volume/sample volume methods
1. Select Stored Programs in the main menu. Select the required
program.
2. Press Start.
3. Analyze a sample without added standard solution in
accordance with the instructions in the Procedures Manual.
When the measurement is complete, leave the sample
cuvette/cell in the cuvette/sample cell holder.
4. Press Options>More>Standard Addition.
An overview of the data of the standard addition procedure is
displayed.
5. Press OK to accept the standard values for standard
concentration, sample volume (total) and standard addition
volume. Press Edit to change any of these values.
84
Advanced Operations
6. Press the key of the value to change it. Use the alphanumeric keypad to change the value. Press OK to confirm.
7. Press the keys to enter the standard addition volumes. Use the alphanumeric keypad to enter the new data and press OK.
Description of the table of readings
•The first column shows the standard addition volume. 0 mL
stands for a sample containing no added standard.
•The second column shows the reading of the samples with and
without added standard.
•The third column shows the detection rate of the added
standard.
Note: The black highlighted line is active.
8. The reading of the sample in the instrument, without added
standard, is automatically shown under 0 mL.
9. Prepare the standard addition solution as described in the
procedure.
10. Use the arrow keys to select the first standard addition volume
in the table and insert the cuvette with the corresponding
volume of added standard into the cell compartment.
Press Read.
Repeat the procedure from point 8 with all the other standard
addition solutions.
85
Advanced Operations
11. After all the standard addition solutions have been measured,
press Graph.
The regression line through the standard addition data points is
displayed.
The correlation coefficient r
2
indicates how close the data
points are to the line.
If the correlation coefficient = 1, the curve is linear.
The concentration shown above the curve is the estimated
concentration of the sample without the added standard.
Note: In the curve menu, the name on the Curve key switches to
Table. Press Table to display all the data in the table again.
12. Press Ideal line to display the relationship between the added
standard solutions and the ideal line (detection rate 100%).
86
Advanced Operations
6.4Single Wavelength (absorbance, concentration and transmittance
measurements)
The Single Wavelength mode can be used in three ways. For
sample measurements at a single wavelength, the instrument can
be programmed to measure the absorbance, % transmittance or
concentration of the analyte.
Absorbance measures the amount of light absorbed by the sample,
in units of absorbance.
% transmittance measures the percent of the original light that
passes through the sample and reaches the detector.
Turn the concentration factor on to select a specific multiplier for
converting absorbance readings to concentration. In a graph of
concentration versus the absorbance, the concentration factor is
the slope of the line.
6.4.1 Set up single wavelength mode
Press Single Wavelength in the Main Menu. Press Options for
Parameter Setup.
Table 12 Single wavelength setup options
OptionsDescription
MoreFor further Options
Store Off/On
% Trans/AbsTo switch to % transmittance, concentration or absorbance readings
λ Wavelength
Timer icon
Concentration FactorMultiplication factor for converting absorbance values into concentration values.
Concentration
Resolution
Save as User ProgramTo store the selected parameters as a User Program, see section 6.1 on page 61
Recall DataCall up saved measurement data or time courses, see section 5.3 on page 37.
With the Store On setting, all measurement data are stored automatically. With the Store Off
setting, no measurement data are stored.
To enter the measurement wavelength. Use the alphanumeric keypad to enter the measurement
wavelength. The entered wavelength must be in the range from 340–900 nm.
This functions as a stopwatch. It helps to ensure that the steps of an analysis are correctly timed
(e.g. reaction times, wait times, etc., can be exactly specified). When the specified time has
elapsed, an acoustic signal is emitted. The use of the timer has no influence on the measurement
program.
To select the position of the decimal point in the calculated concentration readings.
87
Advanced Operations
Table 12 Single wavelength setup options (continued)
OptionsDescription
Instrument SetupBasic data of the instrument, see section 5.2 on page 27.
Concentration factor:
1. Press Concentration Factor: Off in the Options menu. Press
On to highlight this feature.
2. Press the ”Factor“ key and use the alphanumeric keypad to
enter the factor by which absorbance readings are to be
multiplied. Press the ”Unit“ key to select the units for
concentration measurements or to create a new unit.
3. Press OK to confirm.
Concentration resolution:
1. Press Concentration Resolution in the Options menu.
2. Select the resolution and press OK to confirm.
88
6.4.2 Take single wavelength measurements
1. Insert the blank cuvette/cell into the cuvette/sample cell holder. Press Zero.
Note: The Read key is only active after the zero measurement has
been completed.
2. Insert the sample cuvette/cell into the cuvette/sample cell
holder. Press Read.
3. For data storage, see section 5.3.1 on page 37.
Advanced Operations
89
Advanced Operations
6.5Multi-Wavelength mode – measurements at more than one wavelength
In the multi-wavelength mode, absorbance values can be
measured at up to four wavelengths and the results can be
mathematically processed to obtain sums, differences and
relationships.
Absorbance measures the amount of light absorbed by the sample,
in units of absorbance.
% Transmittance measures the percent of the original light that
passes through the sample and reaches the detector.
Turning on the concentration factor allows selection of a specific
multiplier for converting absorbance readings to concentration. In a
graph of concentration versus the absorbance, the concentration
factor is the slope of the line. Concentration is calculated using a
single factor for each wavelength, which is input by the user.
6.5.1 Set the reading mode at different wavelengths
Press Multi Wavelength in the Main Menu. Press Options for
Parameter Setup.
Table 13 Multi-wavelength setup options
OptionsDescription
MoreFor further Options
Store Off/On
% Trans/AbsTo switch to % transmittance, concentration or absorbance readings
λ Wavelength
Timer icon
Concentration FactorMultiplication factor for converting absorbance values into concentration values.
Concentration
Resolution
Absorbance FormulaCalculation basis for evaluating samples
Save as User ProgramTo store the selected parameters as a User Program, see section 6.1 on page 61
With the Store On setting, all measurement data are stored automatically. With the Store Off
setting, no measurement data are stored.
To enter the measurement wavelength. Use the alphanumeric keypad to enter the measurement
wavelength. The entered wavelength must be in the range from 340–900 nm.
This functions as a stopwatch. It helps to ensure that the steps of an analysis are correctly timed
(e.g. reaction times, wait times, etc., can be exactly specified). When the specified time has
elapsed, an acoustic signal is emitted. The use of the timer has no influence on the measurement
program.
To select the position of the decimal point in the calculated concentration readings.
Recall DataCall up saved measurement data or time courses, see section 5.3 on page 37.
Instrument SetupBasic data of the instrument, see section 5.2 on page 27.
λ / Absorbance formula:
1. Press Absorbance Formula.
2. The formula selected in the top key determines the number of
wavelength and coefficent keys that will appear below. To
change the absorbance formula, press the top key, select a
formula from the displayed list and press OK. When a new
formula is selected, the number of variables below changes to
match.
The following formulas are available:
K
A 1 + K 2 A
1
K 1 A 1 + K 2 A 2 + K 3 A
K 1 A 1 + K 2 A 2 + K 3 A 3 + K 4 A
K 1 A 1 / K 2 A
(K 1 A 1 + K 2 A 2 ) / K 3 A
2
3
4
2
3
(K 1 A 1 + K 2 A 2 ) / (K 3 A 3 + K 4 A 4 )
A
refers to the absorbance at wavelength 1
1
A
refers to the absorbance at wavelength 2, etc.
2
K
refers to the coefficient at wavelength 1
1
K
refers to the coefficient at wavelength 2, etc.
2
Coefficients can be set negative where subtraction is required.
91
Advanced Operations
3. To change a wavelength, press one of the ”λx:“ keys. Enter the
desired wavelength coefficient into the numeric keypad.
Press OK to confirm.
4. To change a coefficient, press one of the ”K
:“ keys. Enter the
X
desired coefficient into the numeric keypad.
Press OK to confirm.
Note: The instrument allows entry of up to 5 significant digits, with a
maximum of 4 significant digits after the decimal point.
Concentration factor:
1. Press Concentration Factor: Off in the Options menu. Press
On to highlight this feature.
2. Press the ”Factor“ key to enter the factor by which
absorbance readings are to be multiplied. Press the ”Unit“ key
to select the units for concentration measurements or to create
a new unit.
3. Press OK to confirm.
Concentration resolution:
1. Press Concentration Resolution in the Options menu.
2. Select the resolution and press OK to confirm.
92
6.5.2 Complete a measurement in the multi wavelength mode
1. Insert the blank cuvette/cell into the cuvette/sample cell holder. Press Zero.
Note: The Read key does not become active until the zero
measurement has been completed.
2. Insert the sample cuvette/cell into the cuvette/sample cell
holder. Press Read.
3. For data storage, see section 5.3.1 on page 37.
Advanced Operations
93
Advanced Operations
6.6Time course of absorbance/transmittance
The Time Course Mode is used to collect data in either absorbance
or transmittance for a user-specified length of time. After the data
are collected, they can be displayed in either graphic or tabular
format.
6.6.1 Time course setup parameters
Press Time Course mode in the Main Menu. Press Options to
configure parameters.
Table 14 Time course setup options
OptionDescription
MoreFor further Options
Store iconTo store the scan data
Time & Interval
To input the total time for data collection and the time interval between the collection of data
points
λTo input the wavelength setting
View Table
Scale & Units
Send DataTo send Data to a printer, computer or USB memory stick (Type A)
Recall DataCall up saved measurement data or time courses, see section 5.3 on page 37.
Instrument SetupBasic data of the instrument, see section 5.2 on page 27.
To display readings in absorbance, transmittance or concentration. This can be changed after
sample data are collected.
Scale: In the automatic scaling mode, the y-axis is automatically adjusted so that the total scan is
displayed.
The manual scaling mode allows sections of the scan to be displayed.
Units: Choice of absorbance or transmittance.
94
Advanced Operations
Time & interval:
1. Press Time & Interval in the Options menu.
2. Input the total time and the reading time and press OK to
confirm.
Note: In total 500 measuring steps are possible. To select a total time and
a time interval that would cause this number of measurements to be
exceeded, the time interval is defined automatically and the OK key is
inactivated.
Scale & units:
1. Press Scale & Units in the Options menu.
2. Highlight Abs or %T as the required units.
6.6.2 Time course scan reading
3. Highlight Auto or Manual scaling on the graph´s y-axis.
Note: If manual scaling is selected, use the alphanumeric keypad to set the
min.
and y
limits y
selected range. If automatic scaling is selected, the instrument sets the
limits automatically so that the total range can be displayed.
. The graph is adjusted to display only the values in the
max.
4. Press OK to confirm.
5. Press Return to return to the scan mode.
After the parameters have been selected, the instrument must be
blanked, then the sample can be analyzed.
1. Insert the blank cuvette/cell into the cuvette/sample cell holder.
Press Zero. The blank reading is shown on the display.
2. Insert the sample cuvette/cell into the cuvette/sample cell
holder. Press Read. Start collecting time course (kinetic) data.
Note: During the measurement the Zero and Read keys change to
Mark and Stop.
95
Advanced Operations
6.6.3 Analysis of time course data
•Select Mark to mark the next data point collected. This mark is
not used by the instrument, but is available for the user and
may indicate a significant event, such as the addition of a
sample or other reagent. The mark is also shown in the table.
•Select Stop to stop taking sample readings.
After the data are collected, the following manipulations can be
done on the graphic data:
The Time Course Program is complete, if
•the sound is turned on, the instrument beeps when the
readings are done
•the graph is shown fullsize,
•the scaling of the x-axis occurs automatically,
•the Cursor functions in the vertical navigation bar are
highlighted.
96
6.6.3.1 Navigation of a time scan or a time scan analysis
After a time scan has been completed, the time and the
absorbance/transmittance data are displayed as a curve.
Where the cursor is positioned on the curve, the elapsed time up to
this point and the corresponding absorbance are highlighted.
Table 15 Navigating the time scan
Advanced Operations
Cursor Function/
Zoom Function
Curve Icon
(Choice of Cursor Mode)
Arrow keys
Zoom Icon
Description
Delta mode: A second cursor is highlighted. The position of the fixed cursor was previously
defined in Cursor Mode Single. Use the active cursor to select any point on the measurement
curve. The difference to the fixed cursor is shown on the curve. The delta values are
correspondingly highlighted and displayed on the x and y axes.
The gradient of the curve and the correlation coefficient (r
Delta mode are shown under the curve.
Cursor Mode Single: The cursor moves to each selected measurement point of the scan.
The arrow keys (right/left) are used to move the cursor (depending on the selected mode) to the
next data point. The data of the data point (wavelength/absorbance or transmittance value) are
highlighted on the x and y axes.
Note: Press any point on the curve to display the associated data.
This function is used to magnify the section of the curve in the vicinity of the cursor. The original
curve size can be restored by pressing the zoom icon again.
6.7System checks
2
) between the cursor points in the
1. Press System Checks in the Main Menu.
The System Checks menu contains instrument information and
various performance tests.
97
Advanced Operations
6.7.1 Instrument information
1. Press Instrument Information in the System Checks menu.
2. The model, serial number and software version are displayed.
6.7.2 Upgrade of the instrument software
To obtain the software for the update from the Internet at
www.hach.com:
1. Go to http://www.hach.com.
2. On the DR 2800 product page, click Lab System
Software/Software Update Downloads under Downloads.
3. Locate the appropriate download and follow the prompts for
saving the file(s) to the USB memory stick.
4. On the DR 2800, press Instrument Update in the System
Checks menu.
5. Connect the USB stick to the USB interface on the DR 2800
(section 3.4 on page 13). Press OK.The link is established
automatically and the software is updated.
6. Press OK to return to the System Checks menu.
Note: When the instrument software has been updated, a prompt to restart
the instrument is displayed.
98
6.7.3 Optical checks
6.7.3.1 Wavelength check
Advanced Operations
For each optical check, the measured results must be evaluated
against user-specific requirements. The check options do not
define tolerances.
1. Press Optical Checks in the System Checks menu.
The Optical Checks menu contains programs for checking the
wavelength accuracy, stray light and photometric accuracy.
An optional test filter set (Verification Kit) (Section 9 on page 119)
containing 6 precision glass filters, target values, tolerances and
instructions is available and is recommended as an aid for carrying
out comprehensive in-house instrument checks.
The Wavelength Check test is used to check wavelength accuracy
at 807 nm.
1. Press Wavelength Check in the Optical Checks menu.
2. Insert the adapter (A) for 10-mm rectangular cells in the cell
compartment #2 and insert the sample cuvette/cell (Neodym or
BG20/2) in the adapter. Close the cell compartment.
Press Start.
3. The result is displayed and is to be compared with the
nominal/standard data (given in the quality control certificate) of
the sample cuvette/cell.
4. Press Cancel to return to Optical Checks.
99
Advanced Operations
6.7.3.2 Stray light check
The Stray Light test is used to measure the stray light in the
instrument at 340 nm.
1. Press Stray Light Check in the Optical Checks menu.
2. Remove any cuvette/sample cell from the cell compartment.
3. Insert the adapter (A) for 10-mm rectangular cells in the cell compartment #2. Press Zero.
4. Insert the sample cuvette/cell or reference filter into the cell compartment #2. Close the cell compartment. Press Start.
100
5. The ”Average“ and ”Std Deviation“ are calculated from
100 successive absorbance measurements. The result is
displayed and is to be compared with the nominal/standard
data (given in the quality control certificate) of the sample
cuvette/cell.
Note: Failings and passings will be defined by the user.
6. Press Cancel to return to Optical Checks.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.