Hach DR/2000, DR/2010 Instruction Sheet

Page 1
DR/2000 and DR/2010 Spectrophotometer
Wavelength Calibration Adjustment
Purpose
These instructions provide a step-by-step procedure for performing a wavelength calibration adjustment for the DR/2000 Spectrophotometer (software versions 2.0, 2.2, 3.0 and greater) and DR/2010 Spectrophotometer. Do not software! Perform the adjustment any time the lamp assembly (Cat. No. 46644-00) is replaced and periodically for optimum performance. The adjustment assures the best possible accuracy and repeatability. Use the Calibration Filter Assembly (Cat. No. 46646-00) to make the adjustment.
Start by following the three steps below. Then choose the step-by-step instructions that apply to the specific instrument and software you are using to finish the calibration adjustment.
1. If lamp replacement is necessary, see the Lamp Replacement instructions shipped with
the lamp.
46644-88
perform this procedure if your instrument has 1.261 or 1.27
2. Determine which software version applies to the spectrophotometer being adjusted. For the
DR/2010 Spectrophotometer, the software version will briefly appear after the instrument is turned on. For the DR/2000 Spectrophotometer, different messages will appear, depending on the software in the instrument (see Table 1):
Table 1 DR/2000 Messages Indicating Software Versions
DR/2000 Software Version Display
2.0 or 2.2 If the first display is then enter 888 with the numeric keys. Press READ/ENTER.The display will show V2.0 or V2.2.
3.0 and greater
1.261 and 1.27 DO NOT perform the calibration adjustment. Monochromator
Self-Test V3.X
calibration could be permanently lost (Hach Service cannot restore the calibration).
Method #?,pressSHIFT CONFIG/METH,
Select the appropriate wavelength calibration adjustment procedure on the following pages and continue.
Page 2
Wavelength Calibration Adjustment, continued
DR/2010 Spectrophotometer Procedure
850 nm
1. Empty the cell
compartment and close the cover. Adjust the wavelength to about 850 nm.
2. Press SHIFT SETUP/2 to
enter the Setup Menu. Press
DOWN ARROW twice to
the access Momentary mode, press
ENTER then UP ARROW, then ENTER again. If in Constant-
On mode, go to step 3.
Lamp:.Ifin
3. Scroll to the Adjust nm
option with the key and press ENTER.The display will show:
Peak Filter:
UP ARROW
4. Press: ENTER.The
display will show:
Zero Required
5. Press: ENTER
The display will show:
100 %T Peak Filter
6. Insert the calibration
filter into the cell compartment with the tab facing right. Close the cover.
7. Slowly adjust the
wavelength by turning the control counterclockwise. Record the wavelength where the % transmittance reading is greatest.
2
8. Turn the wavelength
control clockwise so the wavelength is at least 10 nm higher than the wavelength value recorded in step 7.
Page 3
Wavelength Calibration Adjustment, continued
To us e fact ory
default setting, see
instructions below
this box.
9. Slowly turn the
10. Press: ENTER
11. Press: ENTER
wavelength control counterclockwise. Watch the %T values displayed. Stop exactlyonthe highest transmittance.
The display will show:
Are You Sure?
The calibration adjustment will occur automatically. If successful, the display will show Completed and then Adjust nm. If you cannot obtain a good adjustment, contact the Hach Service center serving you.
DR/2000 Spectrophotometer Procedure, Version 3.0 and Greater
12. To use the factory
default setting, press
SETUP
, then press UP
ARROW twice.
The display shows
.PressENTER.PressUP
nm:
ARROW
Default Setup:,thenENTER.
once to display
The display shows
Completed.PressEXIT twice
to get the
Enter Program #
prompt.
850 nm
SHIFT
Adjust
1. Press:
SHIFT CONFIG/METH
2. Press CONFIG/METH
twicetoreturntothe
METHOD #? prompt.
If CONSTANT ON is displayed, go to step 2.If
MOMENTARY is displayed,
READ/ENTER to toggle
press
CONSTANT ON.
to
1. Do not do this calibration if
your instrument has version
1.261 or 1.27 software!
3. Press SHIFT % T to select
the %T mode. The display will read:
%T
3
4. With the cell
compartment empty and the cover closed, rotate the wavelength dial until the small display shows:
850 nm
Page 4
Wavelength Calibration Adjustment, continued
5. Press: CLEAR/ZERO
The display will show:
100.0 %T
Note: If the display does not read within ±0.3 %T, press
CLEAR/ZERO again.
6. Place the Calibration
Filter into the cell holder. Close the cover.
Note: Alignthetabwiththe notch in the cell holder opening.
7. Slowly turn the
wavelength dial counter­clockwise (decreasing wavelength). Record the wavelength where the %T is highest.
8. Turn the wavelength
control clockwise so the wavelength is at least 10 nm higher than the wavelength value recorded in step 7.
9. Slowly turn the
wavelength control counterclockwise. Watch the %T values displayed. Stop exactly on the highest transmittance.
10. Press SHIFT
CONFIG/METH
with the Arrow keys until the display shows:
ADJUST nm
and scroll
11. Press: READ/ENTER
The display will show:
ARE YOU SURE?
READ/ENTER again.
Press If the wavelength calibration adjustment was successful, the display will show:
COMPLETED
with 808 in the nm field and then: ADJUST nm
If you cannot obtain a good adjustment, contact the Hach Service center serving you.
4
12. Press CONFIG/METH
three times to return the display to:
METHOD #?
Page 5
Wavelength Calibration Adjustment, continued
DR/2000 Spectrophotometer Procedure, Versions 2.0 and 2.2
850 nm
1. Press:
SHIFT CONFIG/METH
Scroll with the Arrow keys until the display shows:
CONSTANT ON
Note: Do not do this calibration if your instrument has 1.261 or
1.27 version software!
2. Press READ/ENTER three
times to accept the mode choice and return to the
METHOD #? prompt.
3. Press SHIFT %T to select
the %T mode. The display will show:
%T
4. With the cell
compartment empty and the cover closed, rotate the wavelength dial until the small display shows:
850 nm
5. Press:
6. Place the Calibration
Filter into the cell holder.
CLEAR/ZERO
The display will show:
100.0% T
Note: If the display does not read within ±0.3 %T, press
CLEAR/ZERO again.
Close the cover.
Note: Align the key with the notch in the cell holder opening.
7. Slowly turn the
wavelength dial counter­clockwise (decreasing wavelength). Record the wavelength where the %T is highest.
5
8. Turn the wavelength
control clockwise so the wavelength is at least 10 nm higher than the wavelength value recorded in step 7.
Page 6
Wavelength Calibration Adjustment, continued
9. Slowly turn the
wavelength control counterclockwise. Watch the %T values displayed. Stop exactlyonthe highest transmittance.
10. Press:
SHIFT CONFIG/METH
Enter 888 with the numeric keys.
11. Press READ/ENTER and
immediately key in
READ/ENTER.
If the calibration was successful, the display will
808 in the nm field
show with the
METHOD #?
prompt.
If you cannot obtain a good adjustment, contact the Hach Service center serving you.