Chlorine Test Kit
CN-70 (1454200), CN-70F (1454201), CN-70T (1454202)
DOC326.98.00003
Test preparation
CAUTION:
Use the recommended personal protective equipment.
NOTICE: This product has not been evaluated to test for chlorine and chloramines in
medical applications in the United States.
• Analyze samples immediately after collection.
• Put the color disc on the center pin in the color comparator box (numbers to the front).
Use sunlight or a lamp as a light source to find the color match with the color comparator box.
•
• Rinse the tubes with sample before the test. Rinse the tubes with deionized water after the test.
• If the color match is between two segments, use the value that is in the middle of the two
segments.
• If the color disc becomes wet internally, pull apart the flat plastic sides to open the color disc.
Remove the thin inner disc. Dry all parts with a soft cloth. Assemble when fully dry.
• The long-path adapter for the low range test shows the color in the tubes from top to bottom.
Make sure the light source is above the tubes during the color match.
• Undissolved reagent does not have an effect on test accuracy.
• For free chlorine, read the result immediately after the reagent is added to prevent interference
from monochloramine. If the sample contains 3.0 mg/L monochloramine, the free chlorine
result increases each minute by 0.1 mg/L.
• If the low range test result is more than the maximum limit, use the mid range test procedure
with a fresh sample.
Test procedure—Free or total chlorine, low range (0–0.68 mg/L Cl2)
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used.
Replacement items
Description Unit Item no.
DPD Free Chlorine Reagent Powder Pillows, 25 mL 100/pkg 1407099
DPD Total Chlorine Reagent Powder Pillows, 25 mL 100/pkg 1406499
Bottle, square, with 25-mL mark each 1704200
Color disc, DPD chlorine, 0–3.4 mg/L each 990200
Color comparator box each 173200
Long-path adapter each 2412200
Plastic viewing tubes, 18 mm, with caps 4/pkg 4660004
Optional items
Description Unit Item no.
Caps for plastic viewing tubes (4660004) 4/pkg 4660014
Glass viewing tubes, glass, 18 mm 6/pkg 173006
Stoppers for 18-mm glass tubes and AccuVac Ampuls 6/pkg 173106
Water, deionized 500 mL 27249
1. Install the long-
path adapter in the
color comparator
box.
2. Fill a tube to
the top line with
sample.
3. Put the tube
into the left
opening of the
color comparator
box.
4. Fill the bottle to
the 25‑mL mark
with sample.
5. Add one DPD
(Free or Total)
Chlorine Powder
Pillow. Swirl to
mix.
6. For free
chlorine, read the
result within
1 minute.
For total chlorine,
wait 3 minutes.
Read the result
within 6 minutes.
7. Fill a second
tube to the top line
with the prepared
sample.
8. Put the second
tube into the color
comparator box.
9. Hold the color
comparator box
below a light
source. Turn the
color disc to find
the color match.
10. Read the
value in the scale
window. Divide the
value by 5 to get
the result in mg/L.
Test procedure—Free or total chlorine, mid range (0–3.4 mg/L Cl2)
1. If installed,
remove the longpath adapter.
2. Fill a tube to
the first line (5 mL)
with sample.
3. Put the tube
into the left
opening of the
color comparator
box.
4. Fill the bottle to
the 25‑mL mark
with sample.
5. Add one DPD
(Free or Total)
Chlorine Powder
Pillow. Swirl to
mix.
6. For free
chlorine, read the
result within
1 minute.
For total chlorine,
wait 3 minutes.
Read the result
within 6 minutes.
7. Fill a second
tube to the first line
(5 mL) with the
prepared sample.
8. Put the second
tube into the color
comparator box.
9. Hold the color
comparator box in
front of a light
source. Turn the
color disc to find
the color match.
10. Read the
result in mg/L in
the scale window.
©
Hach Company/Hach Lange GmbH, 2002, 2011, 2013, 2016. All rights reserved. Printed in U.S.A. www.hach.com 800-227-4224 techhelp@hach.com *DOC326.98.00003* 01/2016, Edition 4