1.1 General Information............................................................................................................................................9
1.4 User Access Levels...........................................................................................................................................11
2.2.1 Attaching the Titrant Bottle Holder .........................................................................................................15
2.2.2 Connecting the Electrode Head and Accessories.....................................................................................15
2.2.3 Connecting the Titrant Bottle...................................................................................................................16
2.2.4 Assembling the Burette ...........................................................................................................................17
2.2.5 Connecting the Temperature Sensor .......................................................................................................18
2.2.6 Installing the Electrode ...........................................................................................................................19
2.2.7 Power Connection....................................................................................................................................19
2.2.8 Serial Printer and PC Connections...........................................................................................................20
3.1 Instrument Display and Keypad .......................................................................................................................23
3.2 AutoCAT Main Menus......................................................................................................................................24
3.2.1 How to Access the Menus........................................................................................................................24
3.2.2 Methods Menu .........................................................................................................................................24
3.2.4 Electrodes Menu ......................................................................................................................................26
3.3 System Configuration .......................................................................................................................................27
3.3.1 Entering the Supervisor Code ..................................................................................................................27
3.3.2 Setting a User ID......................................................................................................................................27
3.3.3 Choose a Language ..................................................................................................................................27
3.3.4 Set the Time .............................................................................................................................................28
3.3.5 Set the Date ..............................................................................................................................................28
3.3.5.1 Select Optional PC Keyboard ............................................................................................28
3.3.6 Selecting the ac Power Frequency ...........................................................................................................28
3.3.9 Rules for Routine Access Users...............................................................................................................29
3.3.10 Customizing the Station.........................................................................................................................30
3.3.11 Reset to Factory Settings........................................................................................................................30
3.5 Using Sub-ID's..................................................................................................................................................31
5.2.1 Data Collection ........................................................................................................................................43
5.2.1.1 Current Range ....................................................................................................................43
5.2.2 End Point Determination..........................................................................................................................44
5.2.2.1 Manual End Point Determination (MEPD) .......................................................................45
5.2.2.2 Automatic End Point Determination (AEPD) ...................................................................45
5.2.2.3 Auto Detection ...................................................................................................................45
5.2.3 Parameters Used in the Calculation of the Analyte Concentration..........................................................45
7.4 Samples for Back Titration ...............................................................................................................................63
9.1.3 Cleaning the Instrument.........................................................................................................................139
9.1.4 Cleaning and Conditioning the Electrode ..............................................................................................139
9.3 Replacing the Fuse..........................................................................................................................................140
1.1 General Titration Theory ................................................................................................................................145
1.2 Errors and Interferences..................................................................................................................................147
1.2.5 Choice of Reductant...............................................................................................................................149
1.3 Errors Common to Total Chlorine Determinations.........................................................................................149
1.4 Shift of Endpoints Due to Sample Iodine Demand.........................................................................................151
1.5 Order of Reagent Addition .............................................................................................................................151
Appendix BPrinter Installation and Setup......................................................................................153
Appendix CReplacement Parts & Accessories..............................................................................157
Appendix EAutoCAT 9000 Validation Study ................................................................................161
General Information........................................................................................................................................167
How To Order ......................................................................................................................................................171
Please read this entire manual before unpacking, setting up, or operating this instrument.
Pay particular attention to all danger and caution statements. Failure to do so could result
in serious injury to the operator or damage to the equipment.
To ensure that the protection provided by this equipment is not impaired, do not
use or install this equipment in any manner other than that which is specified in this
manual.
Hazard Information
If multiple hazards exist, this manual will use the signal word (Danger, Caution, Note)
corresponding to the greatest hazard.
DANGER
Indicates a potentially or imminently hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate
injury.
NOTE
Information that requires special emphasis.
Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed.
!
This symbol, when noted on the instrument, references this instruction manual for
operational and/or safety information.
This symbol, when noted on the product, identifies the location of a fuse or current
limiting device.
50081_Safety.fmPage 5
Page 650081_Safety.fm
Specifications
Specifications are subject to change without notice.
Methods
Titration methods7 embedded methods
Titration stopsAutomatic and manual titration termination
Titrant addition techniqueIncremental
Titrant calibration4 embedded procedures
Back titration with manual reductant addition
Measuring Ranges
Current ranges2 µA, 20 µA, 200 µA, 1 mAmp
Temperature range–10 to +100 °C (14 to 212 °F)
Temperature resolution0.1 °C (0.18 °F)
Printout
Automatic, GLP compliant
Selectable; OFF or 80 columns
Detailed or condensed
Graphic or no graphic
Results
QC check on results with visual warning
Statistical calculations
Units
All standard units for samples/results
Titration Curve
Up to 2000 points can be stored
Storage Capacity
Global password protection for programming access
Non-volatile memory
Storage of 200 titration results
Stored parameters characterized by own ID, location, and calibration data
Embedded operating procedures for electrode and reagent exchange
Automatic electrode, titrant calibration, and QC prompt
Sample List
Up to 126 data with alphanumeric ID
QC sample definition
Stirrer Platform
Magnetic stirrer22 reproducible speeds (0 to 1100 rpm) in 50-rpm increments
Beaker volumes5- to 400-mL
Specifications
50081_Specifications.fmPage 7
Specifications
Burette
Embedded burette stand1
Burette volume5- or 10-mL
ISO/FDIS 8655-3 compliant
Burette step motor18,000 steps
UV-protected encapsulated glass syringe
Embedded operating procedures Burette exchange, air bubble removal (Flush), Rinse, Fill, and
Empty functions
Inputs/Outputs
Electrode input
Temperature input
Serial connections for printer/PC
PS/2 port for PC keyboard
Language Options
English and Spanish
General Specifications
CasingSplashproof ABS plastic.
Graphic 128 x 128-dot LCD and alphanumeric keypad
Dimensions (H x W x D)380 x 230 x 450 mm (15 x 9 x 17¾ inch) excluding tubing
Weight5 kg (11 lb) excluding reagent bottles
CE markingEMC directive 89/336/EED Compliant
LV directive 73/23/EEC Compliant
Power requirements50/60 Hz (±3 Hz)
100/240 V ac (±10%)
Pollution Degree II
Overvoltage Category II
FusesPrimary, User-serviceable. Slow-blow 1.0 A, 250 V
Secondary, factory-serviceable only.
Environmental operating conditions5 to 40 °C (41 to 104 °F) temperature range
20 to 80% relative humidity range
Specifications
Page 8
50081_Specifications.fm
Section 1Introduction
1.1 General Information
The AutoCAT 9000™ is an automated chlorine amperometric end point titration
instrument. Its biamperometric system uses a dual platinum electrode (DPE) probe.
Hach offers amperometric methods for determining total chlorine*, free chlorine, sulfite,
and chlorine dioxide in water. The chlorine dioxide procedure also provides a
measurement of chlorite concentration.
Follow the steps listed in Section 1.2, Instrument Startup Summary, on page 9. After
completing these steps, refer to the method calibrations and method procedures for
instructions on specific analyses.
Amperometric titration theory and interferences are detailed in Amperometric Titration Theory on page 145.
1.2 Instrument Startup Summary
When you receive your AutoCAT 9000, proceed in the following order:
1. Unpack the instrument and inspect all parts for any damage that occurred during
shipment. Refer to Table 1 and Figure 1 and Figure 2 on page 24.
2. Assemble the instrument following the instructions in Section 2 on page 13
3. Review Basic Operation on page 23 to familiarize yourself with the user interface
and the keypad, navigation, and command keys.
4. Set the software parameters and configure the interface following the instructions
given in Section 4 Software Setup.
5. Clean and condition the electrode according to section 9.1.4 on page 139.
6. The instrument is set up and ready to perform analyses. The default titrant
concentration is adequate when using a fresh bottle of titrant but you may choose to
calibrate the titrant using the procedures given in Calibration Procedures on page 49
or you can download the exact concentration of your titrant, see Obtaining the Certificate of Analysis for Hach Reagents on page 41.
7. At this time you may wish to run a few test samples to familiarize yourself with the
instrument. We recommend that you use tap water to perform the Total Chlorine
procedure in Titration Procedures on page 65.
8. If desired, you may choose to perform certain setup or configuration steps in Basic
Operation on page 23 but they are not essential for using the instrument.
*Forward and Back titration methods are offered for the determination of Total Chlorine.
50081_Introduction.fmPage 9
1.3 Unpacking
Remove the instrument and accessories from the shipping cartons and inspect each item for
damage that may have occurred during shipment. Refer to Tab l e 1 and Figure 1 on page 23 to
verify that your order is complete.
Table 1 Packing List
Illustration
Item Number
Not shown50083-001AutoCAT 9000 AmperometricTitration Workstation
Not shown18010-001125 V ac power cord
1—1Titrant pick-up tubing
250082-401Delivery tubing with plastic anti-diffusion tip
750082-501Bottle stopper for 500 mL and 1000 mL plastic titrant bottles
850085-001Magnetic stir bar
950082-601Bottle stopper for 1-L glass titrant bottle
1050088-001Bottle holder
11—1Stop ring for electrode head (may be pre-assembled)
12—1Metal key for burette plunger adjustment
1350082-301Metal rod for electrode head support, 20-mm, with stop ring
1450082-801Bayonet electrode head macro, with support
1550082-701Fuse, slow-blow 1.0 A, (5 x 20 mm)
16A25B6001Tubing from stopcock to glass cylinder
Catalog NumberQty.Product Description
Not shown500-46H1Beaker, 250-mL
Not shown1999-531Titrant, PAO standard solution, 0.00564 N, 1000-mL
Not shown28341-001KI SwifTest Dispenser
Not shown1077-601Refill vial for KI SwifTest Dispenser
Not shown14909-321Buffer, pH 4, 100-mL MDB
Not shown21553-321Buffer, pH 7, 100-mL MDB
Not shown50081-181User manual
If any items are missing or damaged, please contact Hach Company or your sales
representative immediately.
Page 1050081_Introduction.fm
1.4 User Access Levels
The AutoCAT offers two user access levels:
Supervisor allows users to edit titration and analysis parameters to fit specific needs. A
Supervisor-level user can also assign a password to protect the data and customized settings
from inadvertent changes.
Routine
analyses.
allows users to access only the routine functions when performing the embedded
50081_Introduction.fmPage 11
Page 1250081_Introduction.fm
Section 2Instrument Assembly
2.1Operating Environment
Maintain an ambient temperature of 5–40 °C (40–104 °F) for proper instrument
operation. The relative humidity should be between 20 and 80%; do not allow moisture
to condense on the instrument.
Note: Retain the original packaging material. Instruments returned for service should be shipped
in the original packaging material to protect against damage during transportation.
Figure 1Individual Parts (Refer to Table 1 for Part Descriptions)
5Dual Platinum Electrode15Bottle Stopper for Plastic Titrant Bottle
6Support Rod for Electrode Head16Pick-up Tubing
7Burette 17Bottle Holder
8Plastic Anti-diffusion Tip Tubing Connection18Stop Ring for Electrode Head
9Burette Tubing Connection19Electrode Head
10Rotating Valve20Keyboard PS/2 Connector
Page 1450081_Assembly.fm
2.2Instrument Assembly
Perform the assembly procedures in this section and then complete the steps in Section 3 on
page 23 to enter basic user information in preparation for performing analyses to comply with
Good Laboratory Practices (GLP). Refer to Section 4 on page 33 for the software setup
required prior to performing analyses. The software setup in Section 4 must be completed
before any analyses can be performed.
2.2.1Attaching the Titrant Bottle Holder
1. Slide the titrant bottle holder into the mounting slot located on the
bottom of the instrument housing.
Operating Environment
2. Affix holder into position by compressing the bottle holder slightly and
inserting the tip into the hole located under the rim of the instrument
housing.
Note: Never pick up or carry the titrator by the bottle holder.
2.2.2Connecting the Electrode Head and Accessories
1. Thread the support rod onto the threaded stud of the stirrer platform.
Note: Never pick up or carry the titrator by the support rod.
50081_Assembly.fmPage 15
Operating Environment
2. Clip the stop ring into the locating hole on the support rod.
Note: The stop ring may be pre-assembled on the support rod.
3. Compress the clips on each side of the electrode head and slide the
head onto the metal rod. Release the two clips to secure the head
in place.
2.2.3Connecting the Titrant Bottle
4. Drop the supplied magnetic stir bar into the titration beaker and place
the beaker on the titration platform.
Note: The provided stir bar has been specifically selected for use with the
AutoCAT 9000. A different stir bar may spin erratically and produce
unwanted turbulence and added noise in the titration curve.
1. Place the titrant bottle into the bottle holder.
2. Thread the desiccant column onto the supplied bottle stopper.
3. Thread the appropriate bottle stopper (for 500 mL or 1000 mL titrant
bottle) onto the new titrant bottle and place it into the holder.
Page 1650081_Assembly.fm
Operating Environment
4. Push the long section of exposed titrant pick-up tubing through the
remaining connection point on the bottle stopper and into the titrant
bottle. Thread the tubing connector into the bottle stopper.
5. Thread the opposite connector into the corresponding receptacle on the
rotating valve.
Note: Make sure that the tubing is connected to the correct point on the rotating
valve as indicated by the “bottle” icon.
6. The plastic anti-diffusion tip is pre-assembled onto the beaker supply
tubing. Thread the tubing connector into the corresponding receptacle
on the rotating valve.
7. Insert the plastic anti-diffusion tip into a small hole on the electrode
head. Turn the locking collar ¼-turn clockwise to lock it in place.
Note: Make sure that the tubing is connected to the correct point on the rotating
valve, indicated by the “beaker” icon.
2.2.4Assembling the Burette
Piston
Travel
1. Use the metal key to gently pull the piston plunger downward until it
protrudes slightly below the bottom of the burette.
Note: The bottom lip of the cylinder will offer only SLIGHT resistance. Do not pull
the plunger out completely. It it is inadvertently removed it may be carefully
pushed back into the cylinder. After reinsertion, closely monitor the
integrity of the seals after titrant is installed.
2. Place the burette on a flat surface and press the burette downward until
the piston shaft is flush with the bottom of the burette.
Note: Piston position is essential to proper instrument operation. Refer to the
illustration (left) for the correct position.
50081_Assembly.fmPage 17
Operating Environment
3. Place the burette on the burette stand in position A then slide the
burette from position A to position B.
Note: Inspect the alignment of the
piston with the plunger shaft.
Ensure that they are properly
engaged as displayed in the
illustration to the right.
4. Turn the burette ¼-turn clockwise to engage the locking tabs with the
burette stand and lock the burette into position.
5. Thread the connectors of the remaining section of tubing into the top
of the burette and into the corresponding receptacle on the
rotating valve.
Note: Ensure that the tubing is connected to the correct point on the rotating
valve as indicated by the “burette” icon.
2.2.5Connecting the Temperature Sensor
1. Remove the protective cap from the temperature sensor.
2. Insert the temperature sensor into a remaining small hole on the
electrode head.
3. Turn the sensor ¼-turn clockwise to lock the collar in place.
4. Connect the cable to the CINCH connector on the back of the
instrument (see Figure 3 on page 19).
Page 1850081_Assembly.fm
2.2.6Installing the Electrode
1. Remove the protective cap from the electrode.
2. Insert the electrode into a large hole (opposite the plastic anti-diffusion
tip) on the electrode head.
3. Turn the electrode ¼-turn clockwise to lock the collar in place.
4. Connect the cable to the BNC connector on the back of the instrument
(see Figure 3 on page 19).
Note: The electrode must be installed opposite the plastic anti-diffusion tip to
allow maximum mixing of the titrant before the reaction is sensed
by the electrode.
A UL/CSA-approved 125-V ac power cord (Cat. No. 18010-00) is supplied with the North
American AutoCAT 9000 models.
To power North American AutoCAT 9000 models with 230 V ac, replace the supplied 125-V
ac power cord with a UL/CSA approved 230-V ac power cord. See section Section 3.3.6 Selecting the ac Power Frequency on page 28 for configuration settings.
Plug the power cord into the back panel of the instrument and connect the power cord to the
proper outlet. See Figure 3.
2Local
3Electrode
4Temperature Sensor
5Power
6Fuse Access
7ON/OFF Switch
After assembly is complete, perform the steps in Section 3 on page 23 to enter basic user
information in preparation for performing analyses to comply with Good Laboratory
Practices. Refer to Section 4 on page 33 for the software setup required prior to
performing analyses. The software setup in Section 4 must be completed before any
analyses can be performed.
50081_Assembly.fmPage 19
Operating Environment
2.2.8Serial Printer and PC Connections
Note: Use of the specified or equivalent shielded cable is mandatory for proper protection from electromagnetic
interferences.
Connect the AutoCAT 9000 Workstation to a personal computer (PC) using the optional
computer interface serial cable (Cat. No. A95P201). The cable provides a direct connection
between the AutoCAT 9000 Workstation and the 9-pin “D” connector used for the serial port
found on most personal computers. If your computer has a 25-pin D connector, use a 9-pin to
25-pin adapter (available at many computer supply stores).
Connect the AutoCAT 9000 to an existing printer using the optional printer interface serial
cable (Cat. No. A95P201). The printer must have the following characteristics:
•80 characters
•RS232C: 9600 baud, 8 data bits, no parity, 1 stop bit
•Must allow printing of tables
Hach offers an optional printer (Cat. No. A70P021) for use with the AutoCAT 9000. See
Appendix BPrinter Installation and Setup on page 153.
Page 2050081_Assembly.fm
Operation
Operation
DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the
necessary Material Safety Data Sheets and become familiar with all safety procedures
before handling any chemicals.
DANGER
La manipulation des échantillons chimiques, étalons et réactifs peut être dangereuse.
Lire les Fiches de Données de Sécurité des Produits (FDSP) et se familiariser avec
toutes les procédures de sécurité avant de manipuler tous les produits chimiques.
PELIGRO
La manipulación de muestras químicas, estándares y reactivos puede ser peligrosa.
Revise las fichas de seguridad de materiales y familiarícese con los procedimientos de
seguridad antes de manipular productos químicos.
GEFAHR
Das Arbeiten mit chemischen Proben, Standards und Reagenzien ist mit Gefahren
verbunden. Es wird dem Benutzer dieser Produkte empfohlen, sich vor der Arbeit mit
sicheren Verfahrensweisen und dem richtigen Gebrauch der Chemikalien vertraut zu
machen und alle entsprechenden Materialsicherheitsdatenblätter aufmerksam zu lesen.
PERICOLO
La manipolazione di campioni, standard e reattivi chimici può essere pericolosa. La
preghiamo di prendere conoscenza delle Schede Techniche necessarie legate alla
Operation
50081_OperationStop.fmPage 21
Page 2250081_OperationStop.fm
Section 3Basic Operation
3.1 Instrument Display and Keypad
The instrument display and keypad form an interactive user interface for the AutoCAT
9000 allowing the user to view data, select options, and enter commands.
Individual screens define certain keystrokes at the bottom of the display. The horizontal
scroll bar indicates the relative status of the page being displayed. Use the
•The keypad is used to enter alpha and numeric characters and punctuation. Press and
hold a key to cycle through the series of characters assigned to it. Choose a character
by pressing and holding the key until the desired character is displayed, then release
the key. Press and hold to scroll from uppercase to lowercase letters. Delete the
character displayed above the screen cursor by pressing the
DEL key.
•The navigation keys include a CHECK MARK and four directional keys (LEFT, RIGHT,
, DOWN) Use these keys to navigate within menus, move cursors, and select
UP
between options. The
confirm displayed information. The
CHECK MARKkey is generally used to select an option or
CHECK MARK key is also used to stop the data
collection process during titration.
•The remaining keys on the keypad are labeled ESC, PRINT, and STOP.
•The ESC key is generally used to exit the currently displayed screen and return to
the previously displayed screen.
Basic Operation
50081_Operation.fmPage 23
Basic Operation
•The PRINT key sends titration and calibration analysis results to the printer.
•The STOP key is an emergency stop and is used to interrupt the current operation
3.2 AutoCAT Main Menus
Figure 2Main Menu
Results can be printed either during the method procedure or later from the GLP
archives.
Note: Curve and Line Fit printer parameters can only be utilized when printing during a
method procedure—not from the GLP archives.
or exit an embedded method. If an analysis is interrupted, the AutoCAT 9000
will prompt the user to resume or end the analysis. All data for the analysis will
be lost if it is ended prior to completion.
the
STOP key is also used to access the Setup menu from the Methods menu
(press and hold for approximately three seconds).
Legend
NumberItem Name
1Title Bar
2Selected Menu Item
3Available Menu Items
4Cell Tab
5Electrode Tab
6Reagents Tab
7Methods Tab
3.2.1 How to Access the Menus
•Press UPor DOWN key to scroll vertically and highlight various items within
currently displayed menus. To select, highlight the desired item and press the
CHECK MARK to confirm. Selections can also be made by pressing the corresponding
number (
1 – 6, if available), to the right of the selection; this feature provides quick,
one-step access.
•Individual menus define certain keystrokes at the bottom of the display. The
horizontal scroll bar indicates the relative status of the screen(s) being displayed.
Note:Press the RIGHTor LEFT key to tab horizontally between Methods, Titrants, Electrodes,
and Cell menus.
3.2.2 Methods Menu
The Methods menu provides access to the embedded procedures, adjustable parameters,
and stored titration analyses results. Major components are explained below.
•Title bar—displays the instrument name, current time, and current "method" name,
along with the Sub-ID.
•Run method—initiates the method currently selected.
•Select method—allows selection of the method(s) for analyses.
•Display method—(available in Routine Mode) allows the user to change the
Method or Sub-ID.
Basic Operation
Page 24
50081_Operation.fm
•Method library—(available in Supervisor Mode) contains all of the adjustable
•GLP-Archives—provides access to the Good Laboratory Practice (GLP) tables, the
•Method Tab—is an animated icon that becomes active when a titration/calibration is
Note:From the Methods menu, adjust the contrast of the display by pressing 7to lighten,
3.2.3 Titrants Menu
The Titrants menu provides access to the embedded calibration procedures, adjustable
parameters, stored calibration results, and burette functions.
•Title bar—displays Titrants
•Calibrate/Enter titrant conc.—determines the concentration of the titrant by
•Install titrants—allows the user to install or replace the titrants in the selected
•Check titrants—allows the user to verify the parameters of the titrant used in the
•Titrant library—contains all of the adjustable calibration parameters. Supervisors
•GLP-Archives—provides access to the Good Laboratory Practice (GLP) tables and
•Burette functions—allows the user to fill, empty, flush, rinse, and replace the
•Titrants tab—displays one of four titrant status icons, see Table 1.
Basic Operation
method parameters. Those with supervisor access can edit, reset, and delete method
parameters to suit specific needs.
stored method analysis results, and global variables.
being performed.
and 0 to darken.
running a calibration or calibration sequence. Alternatively, the concentration can be
input manually on the alphanumeric keypad.
method.
selected method.
can choose to have the titrant concentration input manually or determined through
embedded calibration.
display the stored titrant calibration results.
burette.
Table 1 Titrant Status Icons
Sunny Icon—the titrant calibration or manual entry of the titrant concentration has
been performed satisfactorily for the current titrant.
Cloudy Icon—either the titrant calibration should be performed within 24 hours, or
the titrant will expire in less than one week.
Stormy Icon—the titrant calibration or the titrant in the system has expired, the
calibration has failed to satisfy the user-programmed acceptance limits, or the
installed titrant and selected method are incompatible.
Question Mark—a problem exists within the method parameters for the titrant.
Supervisory level access is required.
3.2.3.1 Changing Titrants
The auto-flush routine takes 60 seconds and uses 10–12 mL of titrant. This function
provides a thorough flush and should be used when changing to a different titrant.
1. From the Titrant menu, press
2. Press
50081_Operation.fmPage 25
3 to begin the flush cycle. Allow the cycle to run to completion.
6 to access Burette Functions.
Basic Operation
Basic Operation
3.2.3.2 Daily Start-up
Even though the delivery tube has an anti-diffusion tip, titrant in the delivery line can be
slightly diluted or contaminated when electrodes are in storage solution (tap water, or
water with weak bleach solution). Bubbles may form due to titrant outgassing. Diluted
titrant and air bubbles can be purged with a manual empty/fill cycle prior to first run on a
daily basis, or after a lengthy idle period.
A manual flush cycle uses only 5 mL of titrant.
1. From the Titrant menu, press
2. Press
3. When the burette empties at least half of its contents, press
the burette.
3.2.4 Electrodes Menu
The Electrodes menu provides access to the electrode cleaning and conditioning
procedure, adjustable parameters, direct measurement, and electrode/temperature sensor
functions.
•Title bar—displays Electrodes.
•Method—indicates the method currently selected.
•Connect electrodes—allow the user to connect the electrode.
•Disconnect electrode—allows the user to disconnect the electrode.
•Replace electrode—allows the user to replace the electrode.
•Check electrodes—identifies and displays the parameters of the electrode used in a
method.
•Electrode cleaning—guides the user through the electrode cleaning and
conditioning procedure.
•Display measurement—performs current measurement or displays the temperature,
based on the selected electrode.
•Electrodes tab—displays one of three electrode status icons, see Ta b l e 2.
6 to access Burette Functions.
2 to empty the burette.
STOP, then press 1 to refill
Basic Operation
Page 26
Table 2 Electrode Status Icons
Sunny Icon—the electrode has been correctly installed.
Stormy Icon—there is a problem with the electrode installation.
Question Mark—a supervisory level problem exists within the method parameters
for the installed electrode.
50081_Operation.fm
3.2.5 Cell Menu
The Cell menu provides access to stirrer function and speed increment adjustment.
•Title bar—displays Cell.
•Stirring—allows the user to switch the stirrer On or Off.
•Speed—allows the user to adjust the internal stirring speed from 100 to 1100
revolutions per minute (rpm) in 50-rpm increments.
Note:The default stir speed is 250 rpm.
•Cell tab—displays an animated icon that indicates the stirrer status.
3.3 System Configuration
3.3.1 Entering the Supervisor Code
Note:There is no factory default code. To proceed in supervisory mode, a user-defined code
must be entered.
1. Turn the instrument ON. From the Methods Menu, press and hold the STOP key for
approximately three seconds to enter the Setup Menu.
2. Input the Supervisor Code input up to 10 digits consisting of any combination of
numbers, upper- and lower-case letters, or symbols. Press the
this code. All future access to the Setup menu will require the input of this casesensitive code.
Note:To establish the Supervisor code, see Accessing Supervisor Mode on page 33.
Basic Operation
CHECK MARK to record
3. Press the CHECK MARKto confirm.
Note:If already operating in Supervisor mode, the AutoCAT 9000 will not request the supervisor
code to access the Setup menu.
4. Select the Configuration Menu and proceed to Section 3.3.2.
3.3.2 Setting a User ID
Select whether or not a User ID will be required:
1. Highlight User ID.
2. Press the
3. Press the
No (to disable the User ID function).
4. Press the
Configuration Menu.
3.3.3 Choose a Language
1. Highlight Language.
2. Press the
3. Press the
French, or Spanish).
4. Press the
Configuration Menu.
CHECK MARKto confirm.The title bar will display User ID.
UP or DOWNkey to toggle between Yes (to enable the User ID function) and
CHECK MARKto validate the highlighted selection and return to the
CHECK MARK to confirm.The title bar will display Language.
UP or DOWN key to highlight one of the available languages (English,
CHECK MARK to accept the highlighted selection and return to the
Basic Operation
50081_Operation.fmPage 27
Basic Operation
3.3.4 Set the Time
3.3.5 Set the Date
1. Highlight Time.
2. Press the
3. Input the hour (00 to 23). Press the
4. Input the minutes (00 to 59). Press the
5. Input the seconds (00 to 59). Press the
CHECK MARKto confirm the selection.
RIGHT key to scroll to minutes.
RIGHT key to scroll to seconds.
CHECK MARKto validate and return to the
Configuration Menu.
Note:Press the LEFT arrow to access a previous field.
1. Highlight Date.
2. Press the
3. Input the day (1 to 31). Press the
4. Select the month using the
5. Input the year (2000 to 2069). Press the
CHECK MARKto confirm the selection.
RIGHT key to scroll to Month.
UP or DOWN key. Press the RIGHT key to scroll to Year.
CHECK MARKto validate and return to the
Configuration Menu.
Note:Press to access a previous field.
3.3.5.1 Select Optional PC Keyboard
Select an optional, external keyboard to connect to the 6-pin PS/2 receptacle located on
the right hand side, bottom edge of the instrument.
1. Scroll to highlight PC Keyboard.
2. Press the
3. Press the
CHECK MARKto confirm the selection.
UP or DOWN key Highlight English (US) for a standard, QWERTY-type
keyboard.
4. Press the
Note:When using the optional keyboard, press F12 to enter the Setup menu. This is done by
CHECK MARKto validate and return to the Configuration Menu.
pressing the Fn and F2 keys simultaneously.
3.3.6 Selecting the ac Power Frequency
Select the ac power supply frequency (50 Hz/230 V ac or 60 Hz/115 V ac) to optimize
instrument operation for regional power.
1. Highlight AC Power Frequency.
2. Press the
3. Press the
Note: The default, 60-Hz /125-V ac setting is required for operation in the U.S. and Canada.
CHECK MARKto confirm the selection.
UP or DOWN key to highlight the desired setting.
Basic Operation
Page 28
4. Press theCHECK MARK to validate and return to the Configuration Menu.
50081_Operation.fm
3.3.7 Printer Status
See Printer Installation and Setup on page 153 for the printer installation procedure.
Basic Operation
1. Select the printer status using the
Press the
2. Press the
the
Note:If the printer option is enabled, a printer must be connected to the instrument. If no printer
CHECK MARKto confirm the selection.
UP or DOWN key. With the desired setting highlighted, press
CHECK MARKto validate and return to the Configuration Menu.
is connected, an error message will be displayed and the instrument will try to detect a
printer for extended periods of time.
3.3.8 Beep ON/OFF
This selection toggles the ON/OFF status of the alarm beeper. If enabled, the AutoCAT
will emit an audible beep to alert the user to analysis completion, titrant expiration, etc.
1. Highlight Beep.
2. Press the
3. Press the
4. Press the
CHECK MARKto confirm the selection.
UP or DOWN key
CHECK MARKto validate and return to the Configuration Menu.
3.3.9 Rules for Routine Access Users
1. Access the Setupmenu.
2. Select Access Routine Mode.
DOWN key to highlight Printer.
Supervisor defines which operations can be ignored in routine mode. Configure
The
Demand and Alarm parameters for Titrant calibration by highlighting the desired
selection then pressing The configuration options are as follows:
Demand: Unlocked
The routine user can ignore the demand for titrant calibration and continue the
measurements.
Demand: Locked
The routine user must either calibrate the titrant or input a new titrant concentration
before continuing the measurements.
Alarm: Unlocked
The routine user can accept a titrant calibration even if the analysis result lies outside the
acceptance range specified by the method.
Alarm: Locked
The routine user must repeat the titrant calibration when the analysis result lies outside
the acceptance range specified by the method.
Basic Operation
50081_Operation.fmPage 29
Basic Operation
3.3.10 Customizing the Station
From the Setupmenu, select Customize.
1. Highlight Station.
2. Input the desired station information.
3. Press the
CHECK MARKto confirm.
4. Individually highlight text lines 1–4 (below Laboratory).
5. Input text (up to 32 characters x 4 lines). After the entry of each line, press
CHECK MARK to validate and move to the next line.
the
6. Press
Note:A typical workstation is defined with a name, user(s) name(s), address, contact information,
ESC to return to the Setup menu.
etc.
3.3.11 Reset to Factory Settings
This option will return the AutoCAT parameters to factory defaults. All user-installed
information, archived analysis results, method/titrant library changes, etc., will be
erased. Only the burette parameters will remain in memory.
1. From the Setup menu, select Reset to Factory Settings.
2. Press the
CHECK MARK or Press 1 to confirm the selection and return to the Setup
menu.
3.3.12 Exit
This selection allows the user to exit the Setup menu and return to the main menus with
either supervisor or routine access.
3.4 Edition Mode
1. From the Setupmenu, select Exit.
2. Highlight Return In Mode. Press the
3. Press the
UPor DOWN key. Press the CHECK MARK to validate.
CHECK MARKto confirm the selection.
4. Select Confirm. The Methods menu will be displayed.
The AutoCAT 9000 edition feature allows the user to multi-task by viewing/editing
parameters during embedded procedure timer and data collection periods. This can only
be accessed during data collection, timers, and on screens where Esc is not defined.
1. To access the Edition feature (during a titration/calibration):
a. Press
ESC.
b. Select Edition.
The user can now access parameters and make any desired changes.
2. To return to the embedded procedure:
a. Press
ESC.
b. Select Analysis Window.
Note:Access to parameters is governed by the access mode (Supervisor/Routine).
Basic Operation
Page 30
50081_Operation.fm
3.5 Using Sub-ID's
Basic Operation
Note:Any changes made with the Edition feature will take effect on subsequent procedures— not
on the current sample titration.
The AutoCAT 9000 allows the user to store optimized parameters for up to three Sub-IDs
for each method. The user can create these Sub-IDs with parameters optimized for a
specific point of the process they need to monitor.
This eliminates the need for operators to change the titration parameters for samples
from different points of their process: they need only to select an appropriate Sub-ID.
1. To change the Sub-ID in Routine Mode, select Display Method on the Main Menu
screen. In Supervisor Mode, select Method Library on the Main Menu screen.
2. From this screen, both the Method and the Sub-ID can be changed. To change the
Method or the Sub-ID select ID or Sub-ID respectively.
3. Initially these Sub-IDs have identical default parameters, and are simply named
A, B, C. The user can assign names to these Sub-IDs, such as Influent, Effluent, Low,
Medium, High, Tap 1, Tap 2, Pre-Dechlor, etc.
4. The individual parameters for a Sub-ID (Predose, Max. Volume, Increment, and
Current Range) can be changed by selecting any of the titrations listed for the method.
Note: More information on the effects of changing these parameters can be found in
Collecting and Accessing Data (page 43).
5. To return to the Main Menu press ESC.
Basic Operation
50081_Operation.fmPage 31
Basic Operation
Basic Operation
Page 32
50081_Operation.fm
Section 4Software Setup
The AutoCAT 9000 requires both assembly and software setup before any calibration
or titration analyses may be performed. The software setup labels the installed
hardware/titrant for use with the embedded procedures. In this section the
supervisor-level user will perform, in order, the following setup tasks:
1. Input the supervisor code and operate in supervisor mode.
2. Software setup for installing the electrode and temperature sensor.
3. Software setup for installing the burette.
4. Software setup for installing the titrant.
5. Input the titrant concentration value.
After the initial software setup, the user will seldom need to install the burette, electrode,
or temperature sensor. The titrant will need to be installed only when replacing the titrant
bottle or switching to a method requiring a different titrant.
Use the following procedures to set up the AutoCAT 9000 titrator software.
4.1 Accessing Supervisor Mode
1. Turn the instrument ON
or, if already in operation,
access the Methods menu
(see Section 3.2.1 on
page 24).
Press and hold the
for approximately 3 seconds
to enter the Setup menu.
Note: The Total Cl2 - Fwd.
procedure has been chosen for
demonstration purposes
because it utilizes the provided
PAO titrant.
50081_SoftwareSetup.fmPage 33
STOP key
2. Enter or set the
supervisor code by using the
key pad to input any
combination of numbers,
upper- and lower-case
letters, or symbols (up to
10 digits).
Note: There is no factory
default code. To proceed in
supervisory mode, a
user-defined code must be
entered.
Note: To initially set the
Supervisor Code, refer to the
instructions in Section 3.3.1 on
page 27. All future access to
the Setup menu will require the
user to input this case-sensitive
code.
3. Press the CHECK MARK
to validate, and enter the
Setup menu.
4. Press 5 to Exit.
Software Setup
Software Setup
5. Press the CHECK MARK
to change the
Return In Mode.
4.2 Software Setup for Electrode and Temperature Sensor Installation
6. Highlight Supervisor.
Press the
validate.
CHECK MARK to
7. Press 1 to select
Confirm Mode.
8. The AutoCAT is now
operating in Supervisor
mode. The Methods menu
will be displayed. Proceed
to Software Setup for
Electrode and Temperature
Sensor Installation.
1. From the Methods
menu (see How to Access
the Menus on page 24), use
the
RIGHT and LEFT keys to
tab to the Electrodes menu.
Software Setup
Page 34
2. Press 1 to select
Connect Electrodes.
3. Highlight S/N.
Press the
input or change.
CHECK MARK to
4. Use the alphanumeric
keypad to input the last
10 digits of the electrode
serial number (located on
the electrode cable).
Press the
validate.
CHECK MARK to
50081_SoftwareSetup.fm
Software Setup
5. Press 1 to select confirm
and proceed to temperature
sensor installation.
6. Highlight S/N.
Press the
input or change.
CHECK MARK to
7. Use the alphanumeric
keypad to input the last
10 digits of the sensor serial
number (located on the
electrode cable).
Press the
CHECK MARK to
validate.
4.2.1Software Setup for Disconnecting or Replacing Electrodes
To disconnect the temperature sensor and electrode:
1. Press
2. Confirm by pressing the
3. Remove the probes.
To replace the temperature sensor and/or electrode:
1. Press 3 to select Replace Electrode.
2. Highlight ID, press the
2 to select Disconnect Electrode.
CHECK MARK (for each probe).
CHECK MARK to confirm.
8. Press 1 to select
Confirm and complete the
installation and return to the
Electrodes menu.
Proceed to Software Setup for Burette Installation on
page 36.
3. Highlight either 50086-00 (electrode), or 50087-00 (temperature sensor).
4. Press the
5. Press
CHECK MARK to validate.
1 to confirm.
6. Remove the probe.
7. Install the replacement probe.
8. Input the serial number of the replacement probe.
9. Press
50081_SoftwareSetup.fmPage 35
1 to confirm.
Software Setup
Software Setup
4.3 Software Setup for Burette Installation
1. Use navigation keys to
tab to the Titrants menu.
5. Highlight
Burette Volume. Press the
CHECK MARK to input or
change.
2. Press 6 to select
Burette Functions.
6. Highlight the desired
burette.
Note: The default software
setting is 5-mL.
Press the CHECK MARK to
validate.
3. Press 5 to select
Install Burette.
7. Highlight S/N. Press the
CHECK MARK to input or
change.
4. The burette will
automatically move to the
install position.
8. Use the alphanumeric
keypad to input the burette
serial number (located on
the bottom edge of the
burette).
Press the
validate.
CHECK MARK to
Software Setup
Page 36
50081_SoftwareSetup.fm
Software Setup
9. Verify the burette
information.
Press
1 to confirm.
10. Install the burette if not
already installed. See
Section 2.2.4 Assembling
the Burette on page 17 for
11. The burette will
automatically reposition
after the install.
complete instructions.
Press the
CHECK MARK to
validate.
4.3.1Software Setup for Removing or Replacing the Burette
To remove the burette:
1. Press
2. Press the
3. Follow the embedded procedure (including the purge and rinse cycles).
4. Remove the burette.
To replace the burette:
1. Press
2. Press
3. Highlight Burette Volume and press the
4. Input the replacement burette volume and press the
6 to select Remove Burette.
CHECK MARK to confirm.
5 to select Replace Burette.
1 to confirm.
12. Press Esc to return to
the Titrants menu. Proceed
to Section 4.4 on page 38.
Note: When the burette
installation is complete the
Replace Burette and
Remove Burette options will be
displayed.
CHECK MARK to confirm.
CHECK MARK to validate.
5. Highlight Serial Number and press the
6. Input the replacement burette serial number and press the
7. Press
1 to confirm.
8. Remove the burette and press the
9. Install the replacement burette and press the
50081_SoftwareSetup.fmPage 37
CHECK MARK to confirm.
CHECK MARK to validate.
CHECK MARK to confirm.
CHECK MARK to confirm.
Software Setup
Software Setup
4.4 Software Setup for Titrant Installation
1. From the Titrantsmenu,
2 to select Install
press
Titrants.
5. Highlight Expiration.
Press the
input or change.
CHECK MARK to
2. Press 1 to select
Titrants Setup.
6. Use the alphanumeric
keypad to input the month
and year of the titrant
expiration date (located next
to the barcode on the titrant
bottle).
Press the
validate.
CHECK MARK to
3. Highlight Lot no.
Press the
input or change.
CHECK MARK to
7. Verify the titrant
information. Press
confirm.
1 to
4. Use the alphanumeric
keypad to input or change
the titrant lot number—
located next to the barcode
on the titrant bottle.
Press the
CHECK MARK to
validate.
8. Press 2 to select
No Rinse.
Note: Initial installation does
not require a burette rinse;
however, it is highly
recommended for titrant
replacement or exchange.
Software Setup
Page 38
50081_SoftwareSetup.fm
Software Setup
9. Connect the tubing and
install the titrant bottle if it
is not already installed. See
Section 2.2.3 Connecting
the Titrant Bottle on page 16
for complete instructions.
Place the 250-mL beaker
onto the stirrer platform.
Press the
confirm.
CHECK MARK to
10. The burette will
complete several cycles to
purge air bubbles in the
system.
Verify that connections are
not leaking and that air is not
present in the delivery
tubing.
Lightly tap the tubing during
the flush to dislodge and
11. The titrant has been
successfully installed.
Discard the waste titrant and
clean and pretreat the
beaker before performing
another titration.
Press
ESC to return to the
Titrants menu.
Proceed to Setting Titrant
Concentration on page 40.
displace any trapped
bubbles.
Note: DO NOT stop the
flushing process before it is
complete or the titrant
installation procedure will not
be completed. Flushing will
take approximately
60 seconds.
4.4.1Software Setup for Switching or Replacing Titrants
Note: AutoCAT will display
an error message if the
method and titrant are
incompatible. If this
message appears, the
user must switch titrants to
accommodate the method.
1. Access the Methods menu (see Section 3.2.1 on page 24)
2. Press
3. Highlight the desired method, press the
4. Use the
5. Press
6. Press
7. Press
8. Highlight Lot No. and press the
9. Input the lot number for the required titrant, press the
10. Highlight Expiration, press the
11. Input the Expiration date for the required titrant, press the
Note: If replacing depleted titrant, the embedded procedure will automatically begin here.
2 choose Select Method.
CHECK MARK to validate.
RIGHT and LEFT keys to move to the Titrants menu.
2 to select Install Titrants.
1 to select Titrants Setup to replace titrants (for a new method), or
press
2 to select Bottle Exchange to replace depleted titrant with a new bottle.
1 to confirm.
CHECK MARK to confirm.
CHECK MARK to validate.
CHECK MARK to confirm.
CHECK MARK to validate.
12. Press 1 to confirm.
13. Ensure that the correct change will be made.
Software Setup
50081_SoftwareSetup.fmPage 39
Software Setup
4.5 Setting Titrant Concentration
14. Press 1 to confirm.
15. Select Rinse Before Replacing and follow the embedded method procedure.
1. From Titrants menu,
1 to select
press
Calibrate/Enter
Titrant Conc.
5. Highlight User and
press the
to input or change the
user ID.
CHECK MARK
2. Highlight PAO 5.64mN
(or another method) and
press the
CHECK MARK to
view the list of available
titrants for the chosen
procedure.
Note: 5.64 mN is equivalent to
0.00564 N.
6. Use the alphanumeric
keypad to input the user ID.
Press the
validate.
Note: The user ID may be any
combination of alphanumeric
characters up to 16 characters
in length.
CHECK MARK to
3. Highlight the titrant
identifier (PAO 5.64mN in
this example) and press the
CHECK MARK to validate.
Note: The AutoCAT 9000
automatically stores installed
titrants for later concentration
updates or calibrations.
7. Verify the user ID and
press
1 to confirm.
4. Press 1 to select Enter
Titrant Concentration.
8. Highlight Titrant Conc.
and press the
to input or change the
concentration.
CHECK MARK
Software Setup
Page 40
50081_SoftwareSetup.fm
Software Setup
9. Use the alphanumeric
keypad and navigation keys
to input or change the titrant
concentration.
Press the
validate.
Note: Product lots differ slightly
in concentration. See Section
4.6 Obtaining the Certificate of
Analysis for Hach Reagents.
CHECK MARK to
10. Highlight Certificate
No. and press the
MARK to input/change.
CHECK
11. Use the alphanumeric
keypad to input the titrant
certificate number.
Press the
validate.
Note: This supports the
GLP archives for titrant
install/calibration, but is not
required for operation.
CHECK MARK to
4.6 Obtaining the Certificate of Analysis for Hach Reagents
Using an internet connection, download the “Certificate of Analysis” (COA) for any
Hach Company reagent or titrant to obtain the exact concentration of any unopened
bottle of Hach titrant or standard solution.
1. Go to www.hach.com.
2. Click on “Information Central”.
3. Click on “Certificate of Analysis Download”.
12. Verify the titrant
concentration and certificate
number.
Press
1 to confirm and return
to the Titrants menu.
Initial software setup is now
complete for the Total Cl2 Fwd. method.
The Catalog Number and Lot Number, located on the label of your Hach reagent bottle,
will be requested.
Software Setup
50081_SoftwareSetup.fmPage 41
Page 4250081_SoftwareSetup.fm
Section 5Collecting and Accessing Data
5.1 Using the AutoCAT 9000 for Amperometric Methods
The following analysis procedures, based on Standard Methods, are embedded directly
in the instrument software:
•Forward titration for free chlorine
•Forward titration for total chlorine
•Back titration for total chlorine
•Forward titration for chlorine dioxide
Hach sells pre-standardized titrants, prepared according to Standard Methods, for both
forward and back titrations. The 0.00564 N phenylarsine oxide titrant (Cat. No. 1999-53)
has a shelf life of >1 year in the unopened bottle. The unopened 0.0282 N Iodine
Standard Solution titrant (Cat. No. 23333-53) is stable for 6 to 12 months.
Periodically recalibrate the titrants after opening to ensure accuracy. Dispense reagents
directly from the product bottle to eliminate contamination from transferring them to
separate containers.
5.2 AutoCAT Titration
AutoCAT 9000 functions can be separated into three categories: (1) data collection,
(2) end point determination, and (3) calculation. Each category is described in the
sections below.
5.2.1 Data Collection
The AutoCAT displays the titration curve (current (µA) vs. titrant volume (mL))
graphically in real time, as it dispenses titrant and measures the current. After the
titration is complete, the user can choose between a fully automated or manual routine to
determine the end point. Following end point determination, the AutoCAT 9000
automatically calculates the analyte concentration.
The titration curve is constructed by plotting the measured current as a function of the
volume of titrant added. Several parameters, which must be set prior to data collection,
are available to control the data collection process. These parameters include the
(1) current range, (2) volume increment, (3) predose volume, and (4) Auto-scaling.
5.2.1.1 Current Range
Current Range allows the user to set the maximum current (in µA) expected for a
particular titration. The AutoCAT 9000 allows four possible settings (2 µA, 20 µA,
200 µA, and 1 mA) for the current range. The default setting (20 µA) setting is adequate
for most purposes. The 2 µA setting is recommended when analyzing samples with
concentration less than 0.600 mg/L Total Chlorine.
Note:The 200-µA and 1-mA settings are available for experimental use, and are not
recommended for most titrations.
Collecting and Accessing Data
50081_AutomatedTitration.fmPage 43
Collecting and Accessing Data
5.2.1.2 Volume Increment
The AutoCAT 9000 delivers titrant to the sample solution in discrete volume increments.
During data collection, the AutoCAT 9000 injects one volume increment of titrant
solution into the sample once every 3 seconds. The volume increment is an adjustable
parameter that controls the volume (mL) of titrant dispensed for each addition. The
smallest increment setting is 0.0003 mL.
In general, for forward titrations, a reasonable volume increment can be calculated based
on the estimated analyte concentration (in mg/L) by dividing the end point volume in
milliliters (numerically equivalent to the chlorine concentration when using
0.00564 N PAO as a titrant) by the number of points to be collected before the end point.
For example, if you wish to collect 50 points before the end point then the appropriate
volume increment would be given by the following equation:
Volume Increment (mL)
A table is provided in each embedded method procedure to help determine the
appropriate volume increment to use for data collection.
5.2.1.3 Predose Volume
Predose defines a volume of titrant added to the sample prior to data collection.
If 1.50 mL of titrant is required to reach the end point, the entire titration can be run using
a 0.03 mL volume increment and no predose. Alternatively, a titration can run using a
0.01 mL volume increment and a 1.00 mL predose. In this case 1.00 mL of titrant is
added prior to data collection and the final 0.50 mL is delivered during data collection.
The same number of points will be collected and the same amount of time will be
required. In the first titration a 20 µA current range will be required to avoid “clipping”
the top of the titration curve. A 2 µA current range may be used in the second case.
5.2.1.4 Auto-scaling
The AutoCAT 9000 has an auto-scaling feature. When Auto-scaling is enabled the
AutoCAT 9000 will attempt to set the current range and volume increment automatically.
This feature is useful for titration of samples containing over 25 ppb total chlorine. Autoscaling is less useful for free chlorine titrations and cannot be used with back titrations.
Auto-scaling can be used with or without a titration predose.
Two associated parameters, No. points before EP and Fraction before EP, become
available when Auto-scaling is activated.
The No. points before EP sets the approximate number of points desired before the end
point.
The Fraction before EP sets the fraction of the total number of points collected. For
example if the fraction is set to “0.50” then approximately 50% of the data points are
collected before the end point and 50% afterwards.
The AutoCAT9000 incorporates two routines for end point determination. Both routines
(described in Section 5.2.2.1 and 5.2.2.2) attempt to determine the volume of titrant
needed to reach the break in the titration curve.
50081_AutomatedTitration.fm
Collecting and Accessing Data
5.2.2.1 Manual End Point Determination (MEPD)
MEPD requires the user to manually position the cursors on the instrument display to
define separate linear regions of the titration curve on each side of the end point (i.e., the
break in the titration curve). When selecting line segments, it is generally advisable to
select points closer to the end point of the titration, and to avoid, as much as possible,
“noisy” sections of the curve. Select a pair of lines that intersect as near as possible to the
break in the titration curve. After the linear regions have been selected, the linear least
squares regression lines are fit through all the points between each pair of cursors. The
intersection of the two regression lines determines the end point. More guidelines for
Manual End Point Determination can be found on page 135.
5.2.2.2 Automatic End Point Determination (AEPD)
AEPD allows the AutoCAT 9000 to search for a pair of intersecting lines that best fit the
titration curve. The end point is determined from the intersection of the two best-fit lines.
5.2.2.3 Auto Detection
Normally, when a user runs an analysis on the AutoCAT 9000, the instrument will go
directly from data collection to the zoom window. The instrument will stop in the zoom
window, allowing the user to select the desired range of data points and the mode (AEPD
or MEPD) to be used for end point determination.
If the auto-detection feature is enabled (set to Yes), then the instrument will proceed
directly from data collection to AEPD without the usual intervening zoom window.
Following AEPD, the results are displayed and the user can accept the results or
reprocess the data using either AEPD or MEPD.
5.2.3 Parameters Used in the Calculation of the Analyte Concentration
After the end point has been determined, the AutoCAT 9000 automatically calculates the
corresponding analyte concentration. The titrant concentration, sample volume, and
dilution factors are used to calculate the analyte concentration, regardless of the method.
In addition, the back titration method uses the concentration and volume of the reductant
solution used to fix the sample. Each parameter must be properly set prior to running
an analysis.
A Number of Decimals parameter is available in supervisor-level mode to set the number
of decimal places used to report the analysis. For example, if Number of decimals is set
to “4”, the experimental results will be reported as 0.3296 mg/L. If it is set to “2”, the
result would be displayed as 0.33 mg/L.
5.2.3.1 Titrant Concentration
The titrant concentration (mN) can be entered manually or determined experimentally
using the appropriate embedded titrant calibration routine. The exact concentration of the
titrant is needed when entering the titrant concentration. See Obtaining the Certificate of Analysis for Hach Reagents on page 41.
5.2.3.2 Sample Volume
The sample volume (mL) is the volume of the sample solution actually placed on the
instrument. This is a run-time parameter, allowing it to be changed each time a sample
analysis is performed. The default sample volume is 200 mL. If the sample volume is
changed, the new value is used as a default until the method defaults are reset or the
value changed again.
Collecting and Accessing Data
50081_AutomatedTitration.fmPage 45
Collecting and Accessing Data
5.2.3.3 Dilution
If a sample solution is diluted prior to analysis, the sample parameter Dilution should be
enabled (set to Yes), allowing access to three other parameters (Predilution vol., Final
dilution vol., and Sample vol.). These parameters allow the calculated analyte
concentration to be corrected for dilution and are only applicable when dilution is
enabled.
This is generally done to bring the analyte concentration into the proper range for
analysis. The AutoCAT 9000 can automatically correct any measured analyte
concentration for sample dilution and three parameters are used to calculate analyte
concentration:
•Predilution volume is the volume of the original sample
•Final dilution vol. is the volume of the sample after dilution
•Volume for Test is the volume of solution taken from the diluted sample
and analyzed.
Normally, AutoCAT returns the concentration of the sample actually presented to it (i.e.,
the concentration of the sample placed in the analysis beaker and analyzed by the
instrument). Using the dilution information, the AutoCAT can automatically convert the
concentration measured for the diluted sample into the concentration of the undiluted
sample.
5.2.3.4 Reductant Concentration
When performing a back titration the sample is fixed by spiking it with reductant (either
PAO or sodium thiosulfate) solution. Reductant concentration is the concentration (mN)
of the solution used to fix the sample. The exact concentration of a Hach reductant
solution can be obtained. See Obtaining the Certificate of Analysis for Hach Reagents on
page 41.
5.2.3.5 Reductant Volume
In a back titration the sample is fixed by spiking it with reductant (either PAO or sodium
thiosulfate) solution. Reductant Volume is the volume (mL) of the reductant solution
used to fix the sample.
5.2.4 Analysis Calculations
When two or more consecutive new tests (as opposed to new samples) are run, the
AutoCAT 9000 automatically calculates the mean and the standard deviation of the mean
for the results obtained from the individual tests. This is equivalent to making several
replicate measurements starting with identical samples for each measurement. The
standard deviation of the mean is different than the standard deviation of the replicate
measurements.
Collecting and Accessing Data
Page 46
50081_AutomatedTitration.fm
Collecting and Accessing Data
N
5.2.4.1 Calculation of Mean
The mean is calculated using the following equation:
R
i
∑
Mean
Where:
Ri is the result from the ith replicate measurement (i.e., Test)
N is the number of replicate measurements.
5.2.4.2 Calculation of Standard Deviation (<5 replicate analyses)
When the number of replicate measurements is between 2 and 5, the standard deviation
of the mean, S
i1=
-----------------=
N
is estimated using the following formula:
Mean
2
S
Mean
R
–
maxRmin
⎛⎞
---------------------------------
⎝⎠
Q
------------------------------------------=
N
Where:
R
is the maximum value obtained from the replicate measurements
max
is the minimum value obtained from the replicate measurements
R
min
Q is taken from the following table:
NQ
21.128
31.693
42.059
52.326
5.2.4.3 Calculation of Standard Deviation (>5 replicate analyses)
If more than five replicate measurements (i.e., Tests) are made, the standard deviation of
the mean, S
S
Mean
is calculated using the following formula:
Mean
N
R
∑
i1=
-------------------
N
2
i
i1=
--------------------------------------------=
N
∑
2
R
i
–
N1–()N
The final average is reported as: Mean (± S
Mean
)
Collecting and Accessing Data
50081_AutomatedTitration.fmPage 47
Collecting and Accessing Data
5.2.5 Retrieving Analysis Results from Archives
As part of the Good Laboratory Practices (GLP) capability incorporated into the
AutoCAT 9000, all analysis results are automatically stored in the archive and organized
according to method. For example, all results obtained using the free chlorine method are
listed together. Before a particular result can be viewed the associated method must first
be selected. The most recent result is always assigned to position “1” in the archive.
For example, if the user performs a series of 10 separate chlorine analyses, then the
results for the most recent analysis will be stored in position “1” of the archive while the
results of the least-recent analysis will be stored in position “10” of the archive. Each
time a new analysis is performed, the results are pushed to the next (higher number)
position in the archive. The AutoCAT 9000 can archive the results from 200 individual
analyses; if more than 200 results are archived, the oldest analysis result is displaced in
favor of the most recent result.
Important Note: All archived results are deleted when the system defaults are reset.
Collecting and Accessing Data
Page 48
50081_AutomatedTitration.fm
Section 6Calibration Procedures
Calibration Procedures
50081_CalibrationStopper.fmPage 49
Page 5050081_CalibrationStopper.fm
PAO/5.64 mN
Amperometric Forward Titrant Calibration
Titration Workstation
Scope and Application: The following sequence guides the user through the calibration procedure for the
0.00564 N (5.64 mN) Phenylarsine Oxide titrants.
Sample Tips and Techniques
• Sample glassware must be cleaned and pretreated to satisfy any chlorine demand. See Section 7.2 Container Pretreatment on
page 63 for additional information.
• Always use organic-free reagent water for sample dilution (Cat. No. 26415-49).
• Rinse the electrode, temperature probe, and anti-diffusion tip with deionized water before every titration.
Reagent Tips and Techniques
• Hach buffer reagents for chlorine titrations are highly recommended for this analysis.
• Download the “Certificate of Analysis” (COA) to obtain the exact concentration of any unopened bottle of Hach titrant or standard
solution. See Obtaining the Certificate of Analysis for Hach Reagents on page 41.
• Never substitute buffers designed for calibrating pH meters. They may contain dyes that interfere with amperometric titration.
• Never use buffers contaminated with mold or bacteria.
• It is recommended that titrations for titrant calibration be performed in triplicate.
Instrument Tips and Techniques
• Adjust the contrast of the display by accessing the Methods menu, and then pressing 7 to lighten or 0 to darken.
• A distinction is drawn between running a new test and a new sample. Each test is a replicate run of the current or previous
sample analysis. The AutoCAT automatically tracks the results of a series of tests, and automatically calculates the mean and
standard deviation for all the results. Each test is given a different number, all of which are listed under the same Sample ID.
When starting a new sample, a new ID is assigned.
• The AutoCAT can accommodate any sample size. Standard Methods recommends a default volume of 200 mL. The volume can
be adjusted to expand the test range.
• The AutoCAT calculates concentration based on the sample volume. Make sure that the sample volume is correct.
• Rinse the electrode, temperature probe, and anti-diffusion tip with deionized water between calibrations.
• The data for any individual sample calibration may be repeatedly analyzed in manual or automatic mode. The results are not
stored in the archives until they are accepted.
• Press the STOP key any time to interrupt instrument operation. The user will be given the option to resume, restart, or quit
completely.
• Press CHECK MARK
• Press the DEL key during any timer period (except the Start Timer) to immediately exit the timer and proceed to the next step in
the procedure.
• Flush the burette each day before the first sample test or calibration is performed. See Daily Start-up on page 26.
• Flush the burette when changing titrants. See Changing Titrants on page 25.
50081_PAONaThioCalibration.fmPage 51
key any time during data acquisition to halt data collection.
PAO/5.64 mN
PAO/5.64 mN
AutoCAT Calibration MethodPAO/5.64 mN Titrant
1. Turn the instrument on
if necessary and access the
Titrants menu.
Press
4 to select Titrant
Library.
Note: Supervisor-mode is
required to access calibrations.
See Accessing Supervisor
Mode on page 33.
5. Press the CHECK MARK
to change the titrant
concentration input mode.
2. Highlight ID.
Press the
to change.
CHECK MARK
6. Highlight Calibrate.
Press the
to validate.
CHECK MARK
3. Highlight
PAO 5.64mN.
Press the
CHECK MARK
to validate.
Note: PAO 5.64mN has been
chosen for demonstration
purposes only.
7. The titrant
concentration input mode is
now set to Calibrate and
calibration parameters are
available by scrolling this
screen. Press
return to the Titrants menu.
ESC twice to
4. Press 1 to select
Edit titrant.
8. Press 1 to select
Calibrate/Enter titrant conc.
Note: Titrant calibration is
required before further
analyses may be run, as
indicated by the “Stormy” icon.
PAO/5.64 mN
Page 52
50081_PAONaThioCalibration.fm
PAO/5.64 mN
9. Press 1 to select.
Run calibration to start the
calibration sequence.
Note: Last Calibration will
indicate the date of the
previous calibration.
10. Press the CHECK MARK
to change the User or press
to confirm.
Note: The user ID will be
stored in the GLP archives for
any future reference to this
calibration.
Note: This screen will be
skipped If User ID is disabled.
See Section 3.3.2 Setting a
User ID on page 27.
11. Highlight
1
Concentration.
Press the
change the concentration or
proceed to step 12.
Note: Use a recent lot of
0.0282 N Iodine Standard
Solution for calibration.
Note: The default
concentration may be changed
to reflect standardization or the
“Certificate of Analysis” value
for a new bottle of titrant.
CHECK MARK to
12. Highlight Lot No.
Press the
change the titrant lot number
or proceed to step 13.
CHECK MARK to
13. Highlight Volume for
1.
Test
Press the
change the calibration
volume or proceed to
step 14.
Note: The default volume for
the standard solution has been
optimized for this embedded
calibration.
CHECK MARK to
14. Prepare a calibration
sample.
a. Add 200 mL of DI-water
to a 250-mL glass beaker.
b. Using a Class A pipet, add
0.50 mL of recently
standardized iodine
solution.
c. Add 1 SwifTest
dispensation (0.1 g) of
potassium iodide (KI), and
15. Rinse the electrode,
temperature probe, and
anti-diffusion tip with
deionized water.
Insert the specified magnetic
stir bar into the calibration
sample and place it onto the
stirrer platform.
Lower the electrode head
onto the beaker rim.
1 to confirm.
Press
16. Press 1 to confirm the
displayed increment or
press the
change the Increment.
Note: The default volume
increment has been optimized
for this embedded calibration;
it may be changed to suit
specific needs.
CHECK MARK to
1.0 mL of pH 4 acetate
buffer.
Press
1 to confirm.
PAO/5.64 mN
50081_PAONaThioCalibration.fmPage 53
PAO/5.64 mN
17. The calibration titration
starts by filling the burette
with titrant and
subsequently dispensing a
2.2-mL predose of PAO into
the calibration sample.
This will bring the electrical
signal into measurable
range.
18. The Start Timer will
begin.
This allows the electrode to
stabilize before data is
collected.
Once the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve in real
time.
19. The Zoom range
determination window will
display the titration curve.
Use the navigation keys to
select and move the cursor
(vertical lines) to define the
range of data points for
analysis.
Note: Defining the range of
data points eliminates
undesirable portions of the
titration curve.
Select type of end point
determination.
Press the
Automatic. Proceed to
step 22 or press
Manual and proceed to
step 20.
CHECK MARK for
DEL for
20. Select two line
segments, one on each side
of the titration end point, by
positioning cursors at the
beginning and end of each
segment.
Press the
confirm changes, press
to return to the previous
window.
CHECK MARK to
DEL
21. Press the CHECK MARK
to accept the analysis.
Proceed to step 24.
DELto return to the
Press
previous window (Zoom
range determination) to
adjust the linear segments
and reanalyze the data.
PAO/5.64 mN
Page 54
22. The AutoCAT will run
the analysis based on the
automatic selection of linear
segments.
Note: This algorithm requires
approximately 60 seconds to
complete.
23. Press the CHECK MARK
to accept the analysis.
Press DELto return to the
previous window (Zoom
range determination) to
adjust the linear segments
and reanalyze the data.
24. Press 1 continue.
Press
2 to select More details
to view analysis statistics
specific to the current
calibration.
50081_PAONaThioCalibration.fm
PAO/5.64 mN
25. Press 1 to select
New Test and run a replicate
calibration. Proceed to
step 26.
Press
3 to select
End Analysis and store the
titrant calibration value and
return to the Methods menu.
Note: If subsequent replicate
calibrations are run, the final
titrant concentration will
automatically be stored as the
mean value from all the
replicate titrant calibrations.
Safety
Summary of Method
26. The replicate analysis is
identical to the first
calibration.
Upon completion of the
analysis, and acceptance of
the data, subsequent replicate
analyses can be performed if
desired.
1 to continue.
Press
Press
2 to select
27. The mean and standard
deviation are calculated
from the replicate analyses.
Press the
accept the calibration value
and return to the
Titrants menu.
CHECK MARK to
In the Titrants menu, the
accepted calibration has
caused the Titrant tab icon
to become sunny. This
indicates that the titrant
concentration has been
properly calibrated and
analyses may resume.
More Details to view
analysis statistics specific to
the current calibration.
Use good safety practices and laboratory techniques throughout the procedure. Consult
the Material Safety Data Sheet (MSDS) for specific reagent(s) information.
The 0.00564 N reductant titrants, phenylarsine oxide (PAO) and sodium thiosulfate
(NaThio), are standardized by titrating them against a know volume of 0.0282 N Iodine
Standard Solution. The PAO/NaThio calibration procedure embedded in the
AutoCAT 9000 is adapted from procedure 4500-Cl C. a. in “Standard Methods for
Examination of Water and Wastewater“(20th Edition).
A calibration sample is prepared by adding 0.50 mL of 0.0282 N Iodine Standard
Solution to 200 mL of chlorine-free deionized water. Potassium iodide (KI) and pH 4
acetate buffer are then added to the calibration sample. The calibration sample is then
titrated using the desired reductant titrant. The embedded procedure calls for a 2.2-mL
predose prior to titration. The titration end point is around 2.5 mL using the embedded
method as defined. If the concentration of the Iodine Standard Solution is significantly
less than 0.0282 N, it may be necessary to reduce the predose volume prior to titration.
Titrate to an amperometric end point.
Pipet, Volumetric, Class A, 0.50 mL ...........................................................1 ....................... each.....................14515-34
Contact the HACH office or distributor serving you.
www.hach.comtechhelp@hach.com
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
Iodine
Amperometric Back Titrant Calibration
Titration Workstation
Scope and Application: The following sequence guides the user through the calibration procedure for the
0.0282 N (28.2 mN) Iodine Standard Solution titrant.
Sample Tips and Techniques
• Sample glassware must be cleaned and pretreated to satisfy any chlorine demand. See Section 7.2 Container Pretreatment on
page 63 for additional information.
• Always use organic-free reagent water for sample dilution (Cat. No. 26415-49).
• Rinse the electrode, temperature probe, and anti-diffusion tip with deionized water before every calibration.
Reagent Tips and Techniques
• Download the “Certificate of Analysis” (COA) to obtain the exact concentration of any unopened bottle of Hach titrant or standard
solution. See Obtaining the Certificate of Analysis for Hach Reagents on page 41.
• Hach buffer reagents for chlorine titrations are highly recommended for this analysis.
• Never substitute buffers designed for calibrating pH meters. They may contain dyes that interfere with amperometric titration.
• Never use buffers contaminated with mold or bacteria.
• It is recommended that titrations for titrant calibration be performed in triplicate.
Instrument Tips and Techniques
• Adjust the contrast of the display by accessing the Methods menu and pressing 7
• A distinction is drawn between running a new test and a new sample on the AutoCAT. Each test is a replicate run of the current or
previous sample analysis. The AutoCAT automatically tracks the results of a series of tests, and automatically calculates the
mean and standard deviation for all the results. Each test is given a different number, all of which are listed under the same
Sample ID. When starting a new sample, a new ID is assigned.
• Rinse the electrode, temperature probe, and anti-diffusion tip with deionized water between calibrations.
• The AutoCAT 9000 can accommodate any sample size. Standard Methods recommends a default volume of 200 mL. The
volume can be adjusted to expand the test range.
• The AutoCAT 9000 calculates concentration based on the sample volume. Make sure that the sample volume is correct.
• The data for any sample calibration may be analyzed repeatedly in manual or automatic mode. The results are not stored in the
archives until they are accepted.
• Press the STOP key any time to interrupt instrument operation. The user will be given the option to resume, restart,
or quit.
to lighten, and 0to darken.
• Press the CHECK MARK
• Press the DEL key during any timer period (except the Start Timer) to immediately exit the timer and proceed to the next step in
the procedure.
• Flush the burette each day before the first sample test or calibration is performed. See Daily Start-up on page 26.
• Flush the burette when changing titrants. See Changing Titrants on page 25.
50081_IodineCalibration.fmPage 57
key any time during data acquisition to halt data collection.
Note: Supervisor-mode is
required to access calibrations.
See Accessing Supervisor
Mode on page 33.
5. Press the CHECK MARK
to change the titrant
concentration input mode.
2. Highlight ID.
Press the
to change.
CHECK MARK
6. Highlight Calibrate.
Press the
validate.
CHECK MARK to
3. Highlight Iodine
28.2mN.
Press the
CHECK MARK
to validate.
7. The titrant
concentration input mode is
now set to calibrate and the
calibration parameters are
now available by scrolling
this screen.
ESC twice to return to
Press
the Titrants menu.
4. Press 1 to select
Edit Titrant.
8. Press 1 to select
Calibrate/Enter titrant conc.
Note: Titrant calibration is now
required before analyses may
be run as indicated by the
“Stormy” icon.
Iodine
Page 58
50081_IodineCalibration.fm
Iodine
9. Press 1 to select.Run
Calibration to start the
embedded calibration
sequence.
Note: Last Calibration will
indicate the date of the
previous calibration.
10. Press the CHECK MARK
to change the User.
Press
1 to confirm.
Note: The user ID will be
stored in the GLP archives for
future reference.
Note: This screen is skipped if
User ID is disabled. See
Section 3.3.2 Setting a User ID
on page 27.
11. Highlight
Concentration.
Press the
change the concentration or
proceed to step 12.
Note: The default
concentration may be changed
to reflect standardization or the
“Certificate of Analysis” value
for a new bottle.
CHECK MARK to
12. Highlight Lot No.
Press the
change the titrant lot number
or proceed to step 13.
CHECK MARK to
13. Highlight Volume for
Test 1.
Press the
change the calibration
volume or proceed to
step 14.
CHECK MARK to
14. Prepare a calibration
sample.
a. Add 200 mL of DI-water
to a 250-mL glass beaker.
b. Add 1.00 mL of Arsenite
Standard Solution.
c. Add 4 g of sodium
bicarbonate.
Press
1 to confirm.
15. Rinse the electrode,
temperature probe, and
anti-diffusion tip with
deionized water.
Insert the magnetic stir bar
into the calibration sample
and place it on the stirrer
platform.
Lower the electrode head
16. Press 1 to confirm.
Press the
change the increment.
CHECK MARK to
onto the beaker rim.
1 to confirm.
Press
Iodine
50081_IodineCalibration.fmPage 59
Iodine
17. The calibration titration
starts by filling the burette
with titrant and
subsequently dispensing a
3.4-mL predose of Iodine
into the calibration sample.
This will bring the electrical
signal into measurable
range.
18. The Start Timer will
begin to allow the electrode
to stabilize before data is
collected.
After the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve in real
time.
19. The Zoom Range
Determination window will
display the titration curve.
Select and move the vertical
lines (cursor) to define the
range of data points for
analysis.
Note: Defining the range of
data points eliminates
undesirable portions of the
titration curve.
Select type of end point
determination:
Press the
Automatic and proceed to
step 22 or press
Manual and proceed to
step 20.
CHECK MARK for
DEL for
20. Select two line
segments, one on each side
of the titration end point, by
positioning the cursors at
the beginning and end of
each segment.
Press the
confirm changes.
Press
previous window.
CHECK MARK to
DELto return to the
21. Press the CHECK MARK
to accept the analysis.
Proceed to step 24.
DELto return to the
Press
previous window (Zoom
range determination) to
adjust the linear segments
and reanalyze the data.
Iodine
Page 60
22. The AutoCAT will run
the analysis based on the
automatic selection of linear
segments.
Note: This algorithm requires
approximately 60 seconds to
complete the analysis.
23. Press the CHECK MARK
to accept the analysis.
Press DELto return to the
previous window (Zoom
range determination) to
adjust the linear segments
and reanalyze the data.
24. Press 1 to select
Continue.
Press 2 to select More Details
and view analysis statistics
specific to the current
calibration.
50081_IodineCalibration.fm
Iodine
25. Press 1 to select New
Test to run a replicate
calibration. Proceed to
step 26 or select
Press
2 to select End
Analysis and store the titrant
calibration value and return
to the Methods menu.
Note: If subsequent replicate
calibrations are run, the final
titrant concentration will
automatically be stored as the
mean value from all the
replicate titrant calibrations.
Safety
Summary of Method
26. The replicate analysis is
identical to the first
calibration.
Upon completion of the
analysis, and acceptance of
the data, subsequent replicate
analyses can be performed if
desired.
1 to select Continue
Press
27. The arithmetic mean is
calculated from the replicate
analyses.
Press the
accept the calibration value
and return to the
Titrants menu.
CHECK MARK to
In the Titrants menu, the
accepted calibration has
caused the Titrant tab icon
to become sunny. This
indicates that the titrant
concentration has been
properly calibrated and
analyses may resume.
and perform additional tests
and/or samples, or press
2 to
select More Details to view
analysis statistics for the
current calibration.
Use good safety practices and laboratory techniques throughout the procedure. Consult
the Material Safety Data Sheet (MSDS) for specific reagent(s) information.
The 0.0282 N Iodine solution is standardized by titrating it against 0.1000 N Arsenite
Standard Solution. The Iodine calibration procedure embedded in the AutoCAT 9000 is
adapted from procedure 4500-Cl C. g. in “Standard Methods for Examination of Water
and Wastewater” (20th Edition).
A calibration sample is prepared by adding 1.00 mL of standard 0.1000 N Arsenite
solution and 4.0 g sodium bicarbonate (NaHCO
) to 200 mL of chlorine free deionized
3
water. The calibration sample is then titrated against the Iodine titrant. The embedded
procedure calls for a 3.4-mL predose prior to titration. The titration end point is around
3.5 mL using the embedded method as defined. Perform a back titration to an
amperometric end point.
Iodine
50081_IodineCalibration.fmPage 61
Iodine
Required Reagents
DescriptionQty. required per testUnit Cat. No.
Arsenite Standard Solution, 0.1000 N.........................................................1.00 mL............ 100 mL ...............28435-42
Sodium Bicarbonate ....................................................................................4 g .................... 454 g.......................776-01
Iodine Standard Solution, 0.0282 N ............................................................varies................ 1000 mL .............23333-53
Pipet, Volumetric, Class A, 1.00 mL ...........................................................1 ....................... each.....................14515-35
Optional Reagents
Dilution water, organic-free ........................................................................varies................ 500 mL ...............26415-49
Contact the HACH office or distributor serving you.
www.hach.comtechhelp@hach.com
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
Page 6250081_IodineCalibration.fm
Section 7Sampling
7.1 Chlorine
A common source of error when testing water samples for chlorine is the failure to
obtain a representative sample. Free chlorine is a strong oxidizing agent and is very
unstable in natural waters. It reacts readily with certain inorganic compounds and
oxidizes organic compounds more slowly. Various factors, including reactant
concentrations, pH, temperature, salinity, and sunlight influence the decomposition of
free chlorine in water. Monochloramine, on the other hand, is much more persistent in
the environment. Typically, the decay rate of monochloramine is tenfold slower than
the decay of free chlorine in natural waters
Ideally, samples should be analyzed for chlorine on site. Because the amperometric
method is usually performed in the laboratory, the delay between sample collection and
analysis should be minimized. If sampling from a tap, allow water to flow 4–5 minutes
to ensure a representative sample.
7.2 Container Pretreatment
Avoid plastic sample containers with a high chlorine demand. Pretreat glass sample
containers to remove any chlorine demand by soaking in a dilute bleach solution (1 mL
commercial bleach solution to 1 liter of water) for at least one hour. After soaking, rinse
thoroughly with deionized/distilled water. Rinse sample containers thoroughly with
deionized/distilled water after use to reduce the need for pretreatment.
*
.
7.3 Handling
Use separate, dedicated sample containers that are labelled for free and total chlorine
determinations. Do not use the same sample containers for free and total chlorine
analysis. Store your labelled sample containers for later use.
Avoid excess agitation and exposure to sunlight when sampling. Allow several
volumes of the container to overflow and cap the sample container to eliminate
headspace above the sample.
7.4 Samples for Back Titration
If the back-titration method is used for total chlorine determinations, preserve the
sample on site. Add 1.00 mL of 0.00564 N phenylarsine oxide (PAO) and 1.0 mL of
pH 4 acetate buffer to a clean, dry glass container with at least 250 mL capacity (such
as a BOD bottle). At the sampling site, measure 200 mL of sample and carefully
transfer it to the sample container. Swirl to mix. Do not overflow the sample container
with preservative in it. Headspace will remain above the sample.
The delay between sample fixing and analysis should be minimized (usually less than
1 hour) to prevent bacterial decomposition of excess PAO in the sample. It is important
that the entire contents of the sample container be transferred to the beaker used in the
titration. See 5.2.3.4 Reductant Concentration and Section 5.2.3.5 Reductant Volume on
page 46.
* Water Chlorination: Environmental Impact and Health Effects, Vol. 4, Book 1: Chemistry and Water Treatment, page 33.
Sampling
50081_Sampling.fmPage 63
Page 6450081_Sampling.fm
Section 8Titration Procedures
Titration Procedures
50081_ProcedureStopper.fmPage 65
Page 6650081_ProcedureStopper.fm
Chlorine Dioxide
Generator Yield
Amperometric Forward Titration
Titration Workstation
Chlorine Dioxide
Chlorite
Free Chlorine
100–5000 mg/L as Cl02
100–5000 mg/L as Cl0
100–2000 mg/L as Cl
2
2
Scope and Application: For yield calculation of a chlorine dioxide generator*
* Procedure equivalent to Standard Methods 4500-ClO2 E.
Sample Tips and Techniques
• Soak all glassware in a strong chlorine dioxide (ClO2) solution (300–500 mg/L) for at least 1 hour.
Rinse thoroughly with deionized water. Use the glassware for this method only.
• Chlorine dioxide and its by-products are volatile and can be easily lost from aqueous solution. Minimize air contact by collecting
the effluent sample into an amber glass bottle with minimum headspace.
• Minimize agitation when measuring sample volumes. Remove sample portions with a volumetric pipet. Always place the tip at the
bottom of the sample container.
• Always use organic-free water for effluent sample dilution (Cat. No. 26415-49).
• Rinse the electrode, temperature probe, and anti-diffusion tip with deionized water before every titration.
.
Reagent Tips and Techniques
–
• Download the “Certificate of Analysis” (COA) to obtain the exact concentration of any unopened bottle of Hach titrant or standard
solution. See Obtaining the Certificate of Analysis for Hach Reagents on page 41.
• Hach buffer reagents for chlorine titrations are highly recommended for this analysis.
• Never substitute buffers designed for calibrating pH meters. They may contain dyes that interfere with amperometric titration.
• Never use buffers contaminated with mold or bacteria.
Instrument Tips and Techniques
• Adjust the contrast of the display by accessing the Methods menu and pressing 7
• A distinction is drawn between running a new test and a new sample. Each test is a replicate run of the current or previous
sample analysis. The AutoCAT 9000 automatically tracks the results of a series of tests, and automatically calculates the mean
and standard deviation for all the results. Each test is given a different number, all of which are listed under the same Sample ID.
When starting a new sample, a new ID is assigned.
• The AutoCAT 9000 can accommodate any sample size. The default volume for this method is 5.0 mL and can be adjusted to
expand the test range.
• The AutoCAT 9000 calculates concentration based on the sample volume. Ensure that the sample volume is correct.
• Press STOP at any time to interrupt instrument operation; the user can resume, restart, or quit completely.
• Press the CHECK MARK at any time during data acquisition to halt data collection.
• Press DEL during any timer period (except Start Timer) to immediately exit the timer and proceed to the next step.
• Press ESC
during any titration/calibration data collection period or timer to access Edition mode. See Edition Mode on page 30.
to lighten or 0to darken.
• Flush the burette each day before the first sample test or calibration is performed. See Daily Start-up on page 26.
• Flush the burette when changing titrants. See Changing Titrants on page 25.
Chlorine Dioxide Generator Yield
50081_ChlorineDioxideGeneratorYield.fmPage 67
Chlorine Dioxide Generator Yield
AutoCAT Titration MethodClO2 Yield Titration 1
1. Turn the instrument on
and access the
Methods menu.
Press
2 to choose
Select Method.
2. Highlight
ClO
Gen. Yield
2
Press the
to validate.
CHECK MARK
3. Press 1 to select
Run method.
This method uses the 0.10 N
Sodium Thiosulfate titrant.
See Software Setup for Titrant Installation on
page 38.
4. Measure 1–5 mL of
generator effluent sample
with a pipet and transfer to a
250-mL beaker containing
the magnetic bar.
Note: All containers must be
pretreated for chlorine demand,
see Container Pretreatment on
page 63.
5. Dilute the sample to
approximately 200-mL with
Organic Free water.
Chlorine Dioxide Generator Yield
Page 68
6. Place the sample onto
the AutoCAT 9000 stirrer
platform.
7. Verify User.
Press the
enter a new user or press
confirm.
Note: This screen will be
skipped if the User ID is
disabled. See Setting a User ID
on page 27.
CHECK MARK to
1 to
8. Highlight Sample ID.
Press the
enter a new sample ID or
proceed to step 9.
50081_ChlorineDioxideGeneratorYield.fm
CHECK MARK to
Chlorine Dioxide Generator Yield
9. Highlight Predilution
volume.
Press the
enter the volume pipetted in
Step 4.
Press
Note: Final dilution volume and
Volume for Test default values
need not be changed.
CHECK MARK and
1 to confirm.
10. Add 1 dose of pH 7.0
Phosphate Buffer.
Press the
to confirm.
Note: Phosphate buffer is
added to adjust the sample pH.
The precise amount added is
not crucial for the accuracy or
precision of the analysis.
CHECK MARK
11. Use the measuring
spoon to add 1.0 g of
potassium iodide (KI).
Press the
to confirm.
Note: KI is added in excess.
The precise amount added is
not crucial for the accuracy or
precision of the analysis.
CHECK MARK
12. The timer will start,
allowing the reagents to mix
before starting the titration.
Chlorine Dioxide Generator Yield
50081_ChlorineDioxideGeneratorYield.fmPage 69
Chlorine Dioxide Generator Yield
13. Lower the electrode
head onto the beaker rim.
Press the
to confirm.
CHECK MARK
14. The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press the
CHECK MARK to change one
of the analysis parameters or
press 1 to confirm.
Note: Any changes made will
become the new default
parameters for this method
Sub-ID.
Note: These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note: This step will be skipped
if auto-scaling software is
enabled.
15. The Start Timer
will start to allow the
electrode to stabilize before
data collection.
Once the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve.
16. The Zoom Range
Determination window will
display the titration curve.
Use the navigation buttons to
select and move the cursors
to define the range of data
points for analysis.
Note: Defining the range of
data points excludes
undesirable portions of the
titration curve from
the analysis.
Chlorine Dioxide Generator Yield
Page 70
50081_ChlorineDioxideGeneratorYield.fm
Chlorine Dioxide Generator Yield
17. Select type of end point
determination.
Press
DEL for Manual.
Note: Automatic determination
attempts to identify linear
segments without user input
and is not recommended for
this method.
18. Use the navigation keys
to select two line segments,
one on each side of the
titration end point, by
positioning vertical lines
(cursors) at the beginning
and end of each segment.
Press the
confirm changes or press
DELto return to the Zoom
window.
Note: For more information on
manual EP determination, see
page 135.
CHECK MARK to
19. Press the CHECK MARK
to accept the analysis and
proceed to Titr. 2 or press
DELto return to the previous
window to adjust the linear
segments and reanalyze the
data.
Note: The analysis data may
be analyzed as many times as
needed from the Zoom Range
Determination window, in
manual and/or automatic
modes. The results are not
stored until accepted.
Note: The AutoCAT 9000
stores all titration results and
then automatically calculates
each analyte value after the
procedure is complete.
Chlorine Dioxide Generator Yield
50081_ChlorineDioxideGeneratorYield.fmPage 71
Chlorine Dioxide Generator Yield
AutoCAT Titration MethodClO2 Yield Tiration 2
1. Raise the electrode
head.
Add 2.0 mL of
2.5 N Hydrochloric acid and
stir for several seconds.
Carefully remove the sample
from the stirrer platform and
place in the dark.
Press the
to confirm.
CHECK MARK
2. The timer will start.
This timer allows a 5-minute
reaction period before
starting the titration.
3. Remove the sample
from the dark and carefully
place onto the AutoCAT
stirrer platform.
Lower the electrode head
onto the beaker rim.
Press the
to confirm.
CHECK MARK
4. The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press CHECK
MARK to change one of the
analysis parameters or press
1 to confirm.
Note: Any changes made will
become the new default
parameters for this method
Sub-ID.
Note: These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note: This step will be skipped
if autoscaling software is
enabled.
Chlorine Dioxide Generator Yield
Page 72
50081_ChlorineDioxideGeneratorYield.fm
Chlorine Dioxide Generator Yield
5. The Start Timer will
begin, allowing the
electrode to stabilize before
data is collected.
Once the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve.
6. The Zoom Range
Determination window will
display the titration curve.
Select and move the cursors
to define the range of data
points for analysis.
7. Select the type of end
point determination.
Press
DEL for Manual.
Note: Automatic determination
attempts to identify linear
segments without user input
and is not recommended for
this method.
8. Using the navigation
keys, select two line
segments, one on each side
of the titration end point, by
positioning vertical lines
(cursors) at the beginning
and end of each segment.
Press the
confirm changes.
Press
Zoom window.
Note: For more information on
manual EP determination, see
page 135.
CHECK MARK to
DELto return to the
9. Press the CHECK MARK
to accept the analysis and
proceed to Titration 3 or
press
DELto return to the
previous window to adjust
the linear segments and
reanalyze the data.
Chlorine Dioxide Generator Yield
50081_ChlorineDioxideGeneratorYield.fmPage 73
Chlorine Dioxide Generator Yield
AutoCAT Titration MethodClO2 Yield Titration 3
1. Press the CHECK MARK
to enter a new volume.
Press
1 to confirm.
5. Purge nitrogen gas
through the sample for
15 minutes.
When the timer beeps,
transfer the purged sample
into a 250-mL beaker. Add
the magnetic stir bar.
Note: If the nitrogen purge has
already been performed, press
DEL to skip this timer and
proceed to the next step.
2. Transfer 200 mL of
organic-free DI water to a
Gas Washing Bottle (GWB).
Add identical volume of
sample pipetted in Titration
1, Step 4.
Add 1 dose of pH 7.0
phosphate buffer, and swirl
to mix.
6. Use the measuring
spoon to add 1.0 g of
potassium iodide (KI).
Place the sample onto the
stirrer platform.
Press the
to confirm.
Note: KI is added in excess.
The precise amount added is
not crucial for the accuracy or
precision of the analysis.
CHECK MARK
3. Insert the purge tube
and dispersion tip into the
GWB. Connect the GWB
inlet to a tank of purified
nitrogen.
7. The timer will begin,
allowing the reagents to mix
before starting the titration.
4. Use a needle valve
to adjust the flow of
nitrogen to provide a steady
stream of bubbles through
the sample.
Press the
CHECK MARK
to continue.
8. Lower the electrode
head onto the beaker rim.
Press the
to confirm.
CHECK MARK
Chlorine Dioxide Generator Yield
Page 74
50081_ChlorineDioxideGeneratorYield.fm
Chlorine Dioxide Generator Yield
9. The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press the
CHECK MARK to change one
of the analysis parameters or
press 1 to confirm.
Note: Any changes made will
become the new default
parameters for this method
Sub-ID.
Note: These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note: This step will be skipped
if autoscaling software is
enabled.
10. The Start Timer will
begin, allowing the
electrode to stabilize before
data is collected.
After the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve in real
time.
11. The Zoom range
determination window will
display the titration curve.
Select and move the cursors
to define the range of data
points for analysis.
Proceed to step 12.
12. Select type of end point
determination.
Press
DEL for Manual.
Note: Automatic determination
attempts to identify linear
segments without user input
but is not recommended for
this method.
Chlorine Dioxide Generator Yield
50081_ChlorineDioxideGeneratorYield.fmPage 75
Chlorine Dioxide Generator Yield
13. Use the navigation
buttons to select two line
segments, one on each side
of the titration end point, by
positioning the cursors at
the beginning and end of
each segment (indicated by
arrows).
Press the
confirm changes or press
DELto return to the previous
window.
Note: For more information on
manual EP determination, see
page 135.
CHECK MARK to
14. Press the CHECK MARK
to accept the analysis and
proceed to Titration 4.
Press
DELto return to the
previous window to adjust
the linear segments and
reanalyze the data.
Chlorine Dioxide Generator Yield
Page 76
50081_ChlorineDioxideGeneratorYield.fm
Chlorine Dioxide Generator Yield
AutoCAT Titration MethodClO2 Yield Titration 4
1. Raise the electrode
head.
Add 2.0 mL of
2.5 N Hydrochloric acid and
stir for several seconds.
Carefully remove the sample
from the stirrer platform and
place in the dark.
Press the
to confirm.
CHECK MARK
2. The timer will start.
This timer allows a 5-minute
reaction period before
starting the titration.
3. Remove the sample
from the dark and place
onto the AutoCAT titration
platform.
Lower the electrode head
onto the beaker rim.
Press the
to confirm.
CHECK MARK
4. The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press the
CHECK MARK to change one
of the analysis parameters or
press 1 to confirm.
Note: Any changes made will
become the new default
parameters for this method
Sub-ID.
Note: These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note: This step will be skipped
if autoscaling software is
enabled.
Chlorine Dioxide Generator Yield
50081_ChlorineDioxideGeneratorYield.fmPage 77
Chlorine Dioxide Generator Yield
5. The Start Timer will
begin, allowing the
electrode to stabilize before
data is collected.
Once the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve in real
time.
6. The Zoom Range
Determination window will
display the titration curve.
Select and move the cursors
to define the range of data
points for analysis.
Proceed to step 7.
7. Select type of end point
determination.
Press
DEL for Manual.
Note: Automatic determination
attempts to identify linear
segments without user input
but is not recommended for
this method.
8. Use the navigation
buttons to select two line
segments, one on each side
of the titration end point, by
positioning the cursors at
the beginning and end of
each segment.
Press the
confirm changes or press
DELto return to the previous
window.
Note: For more information on
manual EP determination, see
page 135.
CHECK MARK to
9. Press the CHECK MARK
to accept the analysis.
Press
DELto return to the
previous window. Adjust the
linear segments and
reanalyze the data.
10. The analysis for this test
is complete.
Use the
the second page of results,
or:
Press
and analyze additional
samples, perform more tests,
or finish analyses.
Press
and view analysis statistics
specific to the current sample
ID.
Chlorine Dioxide Generator Yield
Page 78
RIGHT key to view
1 to select Continue
2 to select More Details
11. View Yield and Ratio
on this screen, and then
press the
to main results screen.
LEFT key to return
12. Press 1 to select
New Test to run a replicate
sample.
2 to select New Sample
Press
to initiate analysis of a new
sample with a new Sample
ID. See Instrument Tips and Techniques on page 67.
Press
3 to select
End Analysis and return to
the Methods menu.
50081_ChlorineDioxideGeneratorYield.fm
Summary of Method
Chlorine Dioxide Generator Yield
This method is used to determine the yield from a chlorine dioxide generator system and
to optimize the generator performance. Yield is defined as the ratio of chlorine dioxide
produced to the theoretical maximum. The percent yield is calculated as follows:
Most modern generators will have yields of 95% or better.
The AutoCAT gives results for chlorine dioxide (ClO
chlorine (Cl
). Four successive titrations are required:
2
chlorine dioxide concentration
Titration 1Cl
+→
2
), chlorite (ClO
2
1
---
ClO
2
5
100×% yield=
–
), and free
2
+→
ClO
–
2
4
→
---
5
ClO
Titration 2
Titration 3Cl
Titration 4ClO
2
(not volatilized by a nitrogen gas purge)→
2
–
2
The AutoCAT 9000 stores all titration results and, after completion, calculates each
analyte value.
This method can be greatly accelerated by purging a sample portion (used for
Titrations 3 and 4) with nitrogen while performing Titrations 1 and 2.
Another measurement used to gauge ClO
generator production is the ratio of Titration 2
2
to Titration 1 results. This provides an estimate of any untreated chlorite or chlorine
feedstock in the generator effluent. This method is only applicable to those generators
using chlorite and gaseous chlorine feeds. Ideally, the optimum ratio of Titration
2/Titration 1 results should lie between 3.9 and 4.05. See Tab l e 1.
Table 1 Generator Effluent Condition
Titr 2/Titr 1 RatioEffluent Condition
Less than 3.9Unreacted chlorine, possible chlorate contamination
3.9–4.05Optimum
Greater than 4.05Unreacted chlorite
Typically, a ratio of less than 3.75 indicates a generator yield of less than 95%.
Chlorine Dioxide Generator Yield
50081_ChlorineDioxideGeneratorYield.fmPage 79
Chlorine Dioxide Generator Yield
Required Reagents
DescriptionQty. required per testUnit Cat. No.
Hydrochloric Acid Standard Solution, 2.5 N ..............................................4 mL................. 100 mL .................1418-32
Phosphate Buffer pH 7, w/dropper..............................................................2 mL................. 100 mL ...............21553-32
Potassium Iodide, ACS................................................................................1.0 g ................. 100 g....................167-26H
Sodium Thiosulfate Solution, 0.10 N.......................................................... varies................ 1000 mL .................323-53
Gas Washing Bottle .....................................................................................1....................... each.....................26622-00
Pipet, Volumetric, Class A, 5.00-mL...........................................................1 ....................... each.....................14515-37
Support, Ring Stand.....................................................................................1 ....................... each.........................563-00
Support Ring, for GWB...............................................................................1 ....................... each.....................26563-00
Optional Reagents
Dilution water, organic-free ........................................................................varies................ 500 mL ...............26415-49
Pipet, Volumetric, Class A, 1.00 mL ...........................................................1 ....................... each.....................14515-35
Pipet, Volumetric, Class A, 2.00 mL ...........................................................1 ....................... each.....................14515-36
Pipet, Volumetric, Class A, 3.00 mL ...........................................................1 ....................... each.....................14515-03
Pipet, Volumetric, Class A, 4.00 mL ...........................................................1 ....................... each.....................14515-04
Contact the HACH office or distributor serving you.
www.hach.comtechhelp@hach.com
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
Chlorine Dioxide
w/ Chlorite and Free Chlorine
Amperometric Forward Titration
Titration Workstation
Chlorine Dioxide
Chlorite
Free Chlorine
0.10–5.00 mg/L as Cl0
0.10–5.00 mg/L as Cl0
0.10–2.00 mg/L as Cl
*
2
–
2
2
Scope and Application: For water and wastewater**
* Higher ranges can be determined through sample dilution. See Section 5.2.3.3 Dilution on page 46.
**Procedure equivalent to Standard Methods 4500-ClO
E.
2
Sample Tips and Techniques
• Soak all glassware in a strong chlorine dioxide (ClO2) solution (300–500 mg/L) for at least 1 hour.
Rinse thoroughly with deionized water. Use the glassware for this method only.
• Chlorine dioxide and its by-products are volatile and can be easily lost from aqueous solution. Minimize air contact by attaching a
flexible hose to a tap and placing the end at the bottom of a 1-L amber glass bottle. Turn on the tap and allow several volumes to
overflow, then slowly remove the sample line and cap the container with minimum headspace.
• Minimize agitation when measuring sample volumes. Remove sample portions with a volumetric pipet. Always place the tip at the
bottom of the sample container. If using 200-mL sample increments, use a 100-mL pipet to withdraw two portions of sample.
• Always use organic-free water for sample dilution (Cat. No. 26415-49).
• Rinse the electrode, temperature probe, and anti-diffusion tip with deionized water before every titration.
.
Reagent Tips and Techniques
• Download the “Certificate of Analysis” (COA) to obtain the exact concentration of any unopened bottle of Hach titrant or standard
solution. See Obtaining the Certificate of Analysis for Hach Reagents on page 41.
• Hach buffer reagents for chlorine titrations are highly recommended for this analysis.
• Never substitute buffers designed for calibrating pH meters. They may contain dyes that interfere with amperometric titration.
• Never use buffers contaminated with mold or bacteria.
Instrument Tips and Techniques
• Adjust the contrast of the display by accessing the Methods menu and pressing 7
• A distinction is drawn between running a new test and a new sample. Each test is a replicate run of the current or previous
sample analysis. The AutoCAT 9000 automatically tracks the results of a series of tests, and automatically calculates the mean
and standard deviation for all the results. Each test is given a different number, all of which are listed under the same Sample ID.
When starting a new sample, a new ID is assigned.
• The AutoCAT 9000 can accommodate any sample size. Standard Methods recommends a default volume of 200 mL. The
volume can be adjusted to expand the test range.
• The AutoCAT 9000 calculates concentration based on the sample volume. Make sure that the sample volume is correct.
• Press STOP any time to interrupt instrument operation; the user can resume, restart, or quit completely.
• Press the CHECK MARK any time during data acquisition to halt data collection.
• Press DEL during any timer period (except Start Timer) to immediately exit the timer and proceed to the next step.
• Press ESC
• Flush the burette each day before the first sample test or calibration is performed. See Daily Start-up on page 26.
during any titration/calibration data collection period or timer to access Edition mode. See Edition Mode on page 30.
to lighten or 0to darken.
• Flush the burette when changing titrants. See Changing Titrants on page 25.
Chlorine Dioxide
50081_Chlorine_ChlorineDioxide.fmPage 81
Chlorine Dioxide
AutoCAT Titration MethodChlorine Dioxide Titr. 1
1. Turn the instrument on
if necessary and access the
Methods menu.
Press
2 to choose
Select Method.
5. Place the sample onto
the AutoCAT 9000 stirrer
platform.
2. Highlight
Chlorine dioxide.
Press the
CHECK MARK
to validate.
6. Verify User.
Press the
enter a new user or press
confirm.
Note: This screen will be
skipped if the User ID is
disabled. See Setting a User ID
on page 27.
CHECK MARK to
1 to
3. Press 1 to select
Run method.
This method used the
5.64mN PAO titrant. See
Software Setup for Titrant
Installation on page 38.
7. Highlight Sample ID.
Press the
enter a new sample ID or
proceed to step 8.
CHECK MARK to
4. Measure 200 mL of
sample solution with a pipet
and transfer to a 250-mL
beaker with the specified
magnetic stir bar.
All containers must be
pretreated for chlorine
demand, see Container Pretreatment on page 63.
8. Highlight Volume for
Test 1.
Press the
enter a new volume.
Press
CHECK MARK to
1 to confirm.
Chlorine Dioxide
Page 82
50081_Chlorine_ChlorineDioxide.fm
Chlorine Dioxide
9. Add 1.0 mL of pH 7.0
Phosphate Buffer Solution.
Press the
CHECK MARK
to confirm.
Note: Phosphate buffer is
added to adjust the sample pH.
The precise amount added is
not crucial for the accuracy or
precision of the analysis.
13. The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press the
CHECK MARK to change one
of the analysis parameters or
1 to confirm.
press
Note: Any changes made will
become the new default
parameters for this method
Sub-ID.
Note: These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note: This step will be skipped
if auto-scaling software is
enabled.
10. Use the measuring
spoon to add 1.0 g of
potassium iodide (KI).
Press the
CHECK MARK
to confirm.
Note: KI is added in excess.
The precise amount added is
not crucial for the accuracy or
precision of the analysis.
14. The Start Timer
will start to allow the
electrode to stabilize before
data collection.
Once the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve.
11. The timer will start.
This allows a short period of
time for the reagents to mix
before starting the titration.
15. The Zoom Range
Determination window will
display the titration curve.
Use the navigation buttons to
select and move the cursors
to define the range of data
points for analysis.
Note: Defining the range of
data points excludes
undesirable portions of the
titration curve from
the analysis.
12. Lower the electrode
head onto the beaker rim.
Press the
CHECK MARK
to confirm.
16. Select type of end point
determination.
Press
DEL for Manual.
Note: Automatic determination
attempts to identify linear
segments without user input
and is not recommended for
this method.
Chlorine Dioxide
50081_Chlorine_ChlorineDioxide.fmPage 83
Chlorine Dioxide
17. Use the navigation keys
to select two line segments,
one on each side of the
titration end point, by
positioning vertical lines
(cursors) at the beginning
and end of each segment.
Press the
confirm changes or press
DELto return to the Zoom
window.
Note: For more information on
manual EP determination, see
page 135.
CHECK MARK to
18. Press the CHECK MARK
to accept the analysis and
proceed to Titr. 2 or press
DELto return to the previous
window to adjust the linear
segments and reanalyze the
data.
Note: The analysis data may
be analyzed as many times as
needed from the Zoom Range
Determination window, in
manual and/or automatic
modes. The results are not
stored until accepted.
Note: The AutoCAT 9000
stores all titration results and
then automatically calculates
each analyte value after the
procedure is complete.
Chlorine Dioxide
Page 84
50081_Chlorine_ChlorineDioxide.fm
Chlorine Dioxide
AutoCAT Titration MethodChlorine Dioxide Titr. 2
1. Raise the electrode
head.
Add 2.0 mL of
2.5 N Hydrochloric acid and
stir for several seconds.
Carefully remove the sample
from the stirrer platform and
place in the dark.
Press the
to confirm.
CHECK MARK
2. The timer will start.
This timer allows a 5-minute
reaction period before
starting the titration.
3. Remove the sample
from the dark and carefully
place onto the AutoCAT
stirrer platform.
Lower the electrode head
onto the beaker rim.
Press the
to confirm.
CHECK MARK
4. The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press the
CHECK MARK to change one
of the analysis parameters or
press 1 to confirm.
Note: Any changes made will
become the new default
parameters for this method
Sub-ID.
Note: These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note: This step will be skipped
if autoscaling software is
enabled.
Chlorine Dioxide
50081_Chlorine_ChlorineDioxide.fmPage 85
Chlorine Dioxide
5. The Start Timer will
begin to allow the electrode
to stabilize before data is
collected.
Once the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve.
6. The Zoom Range
Determination window will
display the titration curve.
Note: Select and move the
cursors to define the range
of data points for analysis.
7. Select the type of end
point determination.
Press
DEL for Manual.
Automatic determination
attempts to identify linear
segments without user input
and is not recommended for
this method.
8. Using the navigation
keys, select two line
segments, one on each side
of the titration end point, by
positioning vertical lines
(cursors) at the beginning
and end of each segment.
Press the
confirm changes.
Press
Zoom window.
Note: For more information on
manual EP determination, see
page 135.
CHECK MARK to
DELto return to the
9. Press the CHECK MARK
to accept the analysis and
proceed to Titr. 3 or press
DELto return to the previous
window to adjust the linear
segments and reanalyze the
data.
Note: For more information on
manual EP determination, see
page 135.
Chlorine Dioxide
Page 86
50081_Chlorine_ChlorineDioxide.fm
Chlorine Dioxide
AutoCAT Titration MethodChlorine Dioxide Titr. 3
1. Press the CHECK MARK
to enter a new volume.
Press
1 to confirm.
5. Purge nitrogen gas
through the sample for
15 minutes.
When the timer beeps,
transfer the purged sample
into a 250-mL beaker. Add
the specified magnetic stir
bar.
Note: If the nitrogen purge has
already been performed, press
DEL to skip this timer and
proceed to the next step.
2. Pipet 200 mL sample
into a Gas Washing Bottle
(GWB).
Add 1 mL of pH 7 Phosphate
buffer.
Swirl to mix.
6. Use the measuring
spoon to add 1.0 g of
potassium iodide (KI).
Place the sample onto the
stirrer platform.
Press the
to confirm.
Note: KI is added in excess.
The precise amount added is
not crucial for the accuracy or
precision of the analysis.
CHECK MARK
3. Insert the purge tube
and dispersion tip into the
GWB. Connect the GWB
inlet to a tank of purified
nitrogen.
7. The timer will start to
allow a short period of time
for the reagents to mix
before starting the titration.
4. Use a needle valve
to adjust the flow of
nitrogen to provide a steady
stream of bubbles through
the sample.
Press the
CHECK MARK
to continue.
8. Lower the electrode
head onto the beaker rim.
Press the
to confirm.
CHECK MARK
Chlorine Dioxide
50081_Chlorine_ChlorineDioxide.fmPage 87
Chlorine Dioxide
9. The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press the
CHECK MARK to change one
of the analysis parameters or
press 1 to confirm.
Note: Any changes made will
become the new default
parameters for this method
Sub-ID.
Note: These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note: This step will be skipped
if autoscaling software is
enabled.
10. The Start Timer will
begin to allow the electrode
to stabilize before data is
collected.
After the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve in real
time.
11. The Zoom range
determination window will
display the titration curve.
Select and move the cursors
to define the range of data
points for analysis.
Proceed to step 12.
12. Select type of end point
determination.
Press
DEL for Manual.
Note: Automatic determination
attempts to identify linear
segments without user input
but is not recommended for
this method.
Chlorine Dioxide
Page 88
50081_Chlorine_ChlorineDioxide.fm
Chlorine Dioxide
13. Use the navigation
buttons to select two line
segments, one on each side
of the titration end point, by
positioning the cursors at
the beginning and end of
each segment (indicated by
arrows).
Press the
confirm changes or press
DELto return to the previous
window.
Note: For more information on
manual EP determination, see
page 135.
CHECK MARK to
14. Press the CHECK MARK
to accept the analysis and
proceed to Titr. 4.
Press
DELto return to the
previous window to adjust
the linear segments and
reanalyze the data.
Note: For more information on
manual EP determination, see
page 135.
Chlorine Dioxide
50081_Chlorine_ChlorineDioxide.fmPage 89
AutoCAT Titration MethodChlorine Dioxide Titr. 4
1. Raise the electrode
head.
Add 2.0 mL of
2.5 N Hydrochloric acid and
stir for several seconds.
Carefully remove the sample
from the stirrer platform and
place in the dark.
Press the
to confirm.
CHECK MARK
2. The timer will start.
This timer allows a
5-minute reaction period
before starting the titration.
3. Remove the sample
from the dark and place
onto the AutoCAT titration
platform.
Lower the electrode head
onto the beaker rim.
Press the
to confirm.
CHECK MARK
4. The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press the
CHECK MARK to change one
of the analysis parameters or
press 1 to confirm.
Note: Any changes made will
become the new default
parameters for this method
Sub-ID.
Note: These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note: This step will be skipped
if autoscaling software is
enabled.
Chlorine Dioxide
5. The Start Timer will
begin.
This allows the electrode to
stabilize before data
is collected.
Once the electrode has
stabilized, the AutoCAT
9000 will begin data
acquisition and plot the
titration curve in real time.
6. The Zoom Range
Determination window will
display the titration curve.
Select and move the cursors
to define the range of data
points for analysis.
Proceed to step 7.
7. Select type of end point
determination.
Press
DEL for Manual.
Note: Automatic determination
attempts to identify linear
segments without user input
but is not recommended for
this method.
8. Use the navigation
buttons to select two line
segments, one on each side
of the titration end point, by
positioning the cursors at
the beginning and end of
each segment.
Press the
confirm changes or press
DELto return to the previous
window.
Note: For more information on
manual EP determination, see
page 135.
CHECK MARK to
9. Press the CHECK MARK
to accept the analysis.
Press DELto return to the
previous window. Adjust the
linear segments and
reanalyze the data.
10. The analysis for this test
is complete.
Press 1 to select Continue to
perform additional tests
and/or samples, or to finish
analyses.
2 to select More Details
Press
to view analysis statistics
specific to the current sample
ID.
11. Press 1 to select
New Test to run a replicate
sample.
2 to select New Sample
Press
to initiate analysis of a new
sample with a new Sample
ID. See Instrument Tips and
Techniques on page 81.
Press
3 to select
End Analysis to return to the
Methods menu.
Chlorine Dioxide
50081_Chlorine_ChlorineDioxide.fmPage 91
Chlorine Dioxide
1
)
Summary of Method
This procedure permits the determination of chlorine dioxide (ClO2), chlorite (ClO
and chlorine (Cl
) in the sample.
2
–
2
This is a multi-step procedure involving four amperometric titrations and two titration
samples. These titrations are as follows:
Sample 1
Titration 1 Cl
-- -
+→
ClO
2
2
5
),
Titration 2
4
-- -
ClO
5
2
+→
ClO
–
2
Sample 2
Titration 3 Cl
Titration 4 ClO
→
(not volatized by a nitrogen gas purge
→
2
–
2
The AutoCAT 9000 stores all titration results and, after completion, calculates each
analyte value.
This method can be greatly accelerated by purging a sample portion (used for
Titrations 3 and 4) with nitrogen while performing titrations 1 and 2.
Required Reagents
DescriptionQty. required per testUnit Cat. No.
Phenylarsine Oxide (PAO) Titrant, 0.00564 N............................................ varies................ 1000 mL ...............1999-53
Hydrochloric Acid Standard Solution, 2.5 N ..............................................4 mL................. 100 mL .................1418-32
Phosphate Buffer pH 7, w/dropper..............................................................2 mL................. 100 mL ...............21553-32
Potassium Iodide, ACS................................................................................2.0 g ................. 100 g....................167-26H
Gas Washing Bottle .....................................................................................1....................... each.....................26622-00
Pipet, Volumetric, Class A, 100-mL............................................................1 ....................... each.....................14515-42
Support, Ring Stand.....................................................................................1 ....................... each.........................563-00
Support Ring, for GWB...............................................................................1 ....................... each.....................26563-00
Contact the HACH office or distributor serving you.
www.hach.comtechhelp@hach.com
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
Free Chlorinec
Amperometric Forward Titration
Titration Workstation
0.100–5.000 mg/L as Cl2*
Scope and Application: For drinking water; USEPA accepted for reporting**
* Higher ranges can be determined through sample dilution. See Section 5.2.3.3 Dilution on page 46.
**Procedure equivalent to USEPA method 330.2 and Standard Methods 4500-Cl.D. for drinking water.
Sample Tips and Techniques
• To avoid loss of chlorine, be careful not to agitate the sample when measuring or pouring.
• Sample glassware must be cleaned and pretreated to satisfy any chlorine demand before the sample is introduced. See Section
7.2 Container Pretreatment on page 63 for additional information.
• Always use organic-free water for sample dilution (Cat. No. 26415-49).
• When sampling tap water, let the faucet run for at least 4–5 minutes prior to collecting the sample.
• A convenient test sample can be prepared by diluting Chlorine Standard Solution (Cat. No. 14268-10) with deionized (DI) water.
• Rinse the electrode, temperature probe, and anti-diffusion tip with deionized water before every titration.
Reagent Tips and Techniques
• Download the “Certificate of Analysis” (COA) to obtain the exact concentration of any unopened bottle of Hach titrant or standard
solution. See Obtaining the Certificate of Analysis for Hach Reagents on page 41.
• Hach buffer reagents for chlorine titrations are highly recommended for this analysis.
• Never substitute buffers designed for calibrating pH meters. They may contain dyes that interfere with amperometric titration.
• Never use buffers contaminated with mold or bacteria.
Instrument Tips and Techniques
• Adjust the contrast of the display by accessing the Methods menu and then pressing 7
• A distinction is drawn between running a new test and a new sample. Each test is a replicate run of the current or previous
sample analysis. The AutoCAT automatically tracks the results of a series of tests, and automatically calculates the mean and
standard deviation for all the results. Each test is given a different number, all of which are listed under the same Sample ID.
When starting a new sample, a new ID is assigned.
• The AutoCAT can accommodate any sample size. Standard Methods recommends a default volume of 200 mL. The volume can
be adjusted to expand the test range.
• The analysis data may be analyzed repeatedly from the Zoom Range Determination window, in manual (MEPD) and/or automatic
(AEPD) modes. The results are not stored until accepted.
• The AutoCAT calculates the chlorine concentration based on the sample volume. Make sure that the sample volume is correct.
• Press the STOP key any time to interrupt instrument operation. The user will be given the option to resume, restart, or quit.
• Press the CHECK MARK any time during data acquisition to halt data collection.
• Press the DEL key during any timer period (except Start Timer) to immediately exit the timer and proceed to the next step in
the procedure.
• Flush the burette each day before the first sample test or calibration is performed. See Daily Start-up on page 26.
• Flush the burette when changing titrants. See Changing Titrants on page 25.
to lighten or 0to darken.
Free Chlorine
50081_Chlorine_Free_Forward.fmPage 93
Free Chlorine
Amperometric TitrationFree Cl2 - Fwd
1. Access the
methods menu.
Press
2 to choose
Select Method.
5. Place the sample onto
the AutoCAT 9000 stirrer
platform.
2. Highlight Free
Cl2-Forward.
Press the
CHECK MARK
to validate.
6. Verify User.
Press the
enter a new user or press
confirm.
Note: This screen will be
skipped if the User ID is
disabled. See Setting a User ID
on page 27.
CHECK MARK to
1 to
3. Press 1 to select
Run method.
7. Highlight Sample ID.
Press the
enter a new sample ID or
proceed to step 8.
CHECK MARK to
4. Measure 200 mL of
sample solution into a clean
250 mL beaker.
Insert the specified magnetic
stir bar into the sample
beaker.
Note: Use only the provided
stir bar to minimize loss of
chlorine.
8. Highlight
Volume for Test 1.
Press
1 to confirm or press
the
CHECK MARK to enter a
new volume.
Free Chlorine
Page 94
50081_Chlorine_Free_Forward.fm
Free Chlorine
9. Add 1.0 mL of pH 7.0
Phosphate Buffer Solution.
Press the
confirm and start the timer.
Note: The Phosphate buffer is
added to adjust the sample pH.
The precise amount added is
not crucial for the accuracy or
precision of the analysis.
CHECK MARK to
10. The timer will start and
the reagents will mix for a
short period before titration
begins.
11. Lower the electrode
head onto the beaker rim.
Press the
to confirm.
CHECK MARK
12. The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time. (Verify
Increment from Table 1,
“Generator Effluent
Condition,” on page 79).
Highlight the appropriate
line and press the
CHECK MARK to change one
of the analysis parameters or
press 1 to confirm.
Note: Any changes made will
become the new default
parameters for this method
Sub-ID.
Note: These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note: This step will be skipped
if autoscaling software is
enabled.
Free Chlorine
50081_Chlorine_Free_Forward.fmPage 95
Free Chlorine
13. The Start Timer will
start.
The electrode will stabilize
before data collection.
14. The Zoom Range
Determination window will
display the titration curve.
Select and move the cursors
to define the range of data
points for analysis.
Note: Defining the range of
data points allows the user to
exclude undesirable portions
of the titration curve from
the analysis.
15. Select the type of end
point determination.
Press
CHECK MARK for
Automatic. and proceed to
step 18 or press
Manual.
Note: Automatic determination
attempts to identify linear
segments without user input.
Note: This screen is not
displayed if the Auto-detection
software is enabled.
DEL for
16. Use the navigation keys
to select two line segments,
one on each side of the
titration end point, at the
beginning and end of each
segment.
Press the
confirm changes or press
DELto return to the previous
window.
Note: For more information on
manual EP determination, see
page 135.
CHECK MARK to
17. Press the CHECK MARK
to accept the analysis.
Proceed to step 19 or press
DELto return to the previous
window (Zoom range
determination) to adjust the
linear segments and
reanalyze the data.
Free Chlorine
Page 96
18. The AutoCAT will
perform the analysis based
on the automatic selection
of linear segments.
Note: The automated analysis
requires approximately
60 seconds to complete.
19. Press the CHECK MARK
to accept the analysis or
press
DELto return to the
previous window (Zoom
range determination) to
adjust the linear segments
and reanalyze the data.
20. The analysis for this test
is complete.
1 to select Continue or
Press
2 to select More Details
press
to view analysis statistics
specific to the current
Sample ID.
50081_Chlorine_Free_Forward.fm
21. Press 1 to select
New Test to perform a
replicate sample.
Press
2 to select New Sample
to analyze a new sample with
a new ID.
3 to select
Press
End Analysis to return to the
Methods menu.
Free Chlorine
Method Performance
Table 1 Recommended Volume Increment Settings for Free Chlorine
Note: These recommendations apply to Free Chlorine titrations, correspond to data collection
times typically under 5 minutes, and provide adequate data points on both sides of the titration
end point for proper analysis. Adjusting the volume increment will reduce the data collection
times and number of data points collected. For example, doubling the volume increment will cut
the data collection time and the number of points collected in half.
Table 2 Precision Statements
At Concentration95% Confidence Limits
0.277 mg/L Cl
1.924 mg/L Cl
4.942 mg/L Cl
2
2
2
0.251–0.303 mg/L Cl
1.778–2.070 mg/L Cl
4.650–5.234 mg/L Cl
2
2
2
Free Chlorine
50081_Chlorine_Free_Forward.fmPage 97
Free Chlorine
Table 3 Accuracy Assessments
Safety
Summary of Method
SampleSpike AdditionAverage % Recovery
Drinking Water
Free Cl
Free Cl
Free Cl
at 0.060 mg/L92.6 (n=4)87.5–98.0
2
at 0.220 mg/L89.7 (n=6)89.2–92.7
2
at 0.530 mg/L91.0 (n=4)85.4–93.5
2
% Range of
Recovery
Lower Limit
Titration curve end points are not as sharp as those for Total Chlorine. Therefore, the
determination of Free Chlorine concentrations below 0.1 mg/L becomes problematic. As
a result, the lower limit for the AutoCAT 9000 has been arbitrarily set to 0.1 mg/L for
experimental reasons.
Use good safety practices and laboratory techniques throughout the procedure. Consult
the Material Safety Data Sheet (MSDS) for specific reagent(s) information.
This procedure follows method number 4500-Cl D. b. in “Standard Methods for the
Examination of Water and Wastewater” (20th Edition). This procedure permits
determination of the Free Chlorine concentration in a water sample. Free chlorine
corresponds to that portion of chlorine existing either as elemental chlorine (Cl
–
hypochlorous acid (HOCl), or hypochlorite (OCl
) ion.
),
2
In the procedure, the sample pH is adjusted to pH 7 by addition of Phosphate buffer. The
sample solution is then titrated amperometrically with standard phenylarsine oxide
solution.
Required Reagents
DescriptionQty. required per testUnit Cat. No.
Phenylarsine Oxide (PAO) Titrant, 0.00564 N............................................ varies................ 1000 mL ...............1999-53
Phosphate Buffer, pH 7, w/dropper .............................................................1 mL................. 100 mL ...............21553-32