Hach POCKET COLORIMETER II, 59530-00 Instruction Manual

Page 1
59570-88
POCKET COLORIMETER II
ANALYSIS SYSTEMS
INSTRUCTION MANUAL
Chlorine (Cl
2
)
© Hach Company, 2006. All rights reserved. Printed in the U.S.A. te/dk 09/06 3ed
Page 2
1—2
Important Note
This manual is intended for use with the following Pocket Colorimeter™ II instrument:
Chlorine (Cl
2
) Cat. No. 59530-00
Page 3
1—3
Table of Contents
Safety Precautions............................................................................................... 1—7
Laboratory Safety............................................................................................... 1—7
Use of Hazard Information................................................................................1—7
Precautionary Labels........................................................ ....... ...........................1—8
Specifications......................................................................................................... 1—9
Instrument Keys and Display...........................................................................1—11
Instrument Cap Cord.........................................................................................1—12
Chlorine, Free and Total, LR...........................................................................1—15
Measuring Hints ...............................................................................................1—15
Accuracy Check................ ....... ....................................................... ....... ...........1—27
Interferences......................................................................................................1—29
Method Performance........................................................................................1—30
Chlorine, Free and Total, HR..........................................................................1—31
Measuring Hints ...............................................................................................1—31
............................................................................................................................1—38
Accuracy Check................ ....... ....................................................... ....... ...........1—39
Interferences......................................................................................................1—41
Page 4
1—4
Table of Contents, continued
Method Performance........................................................................................1—42
Spec™ Secondary Standards for DPD Chlorine.....................................1—43
Using the Spec
Standards................................. .............................................1—44
Summary of Method..........................................................................................1—45
Replacement Parts..............................................................................................1—46
Instrument Operation..........................................................................................2—3
Key Functions........................... ...........................................................................2—3
Menu Selections... ...... ....... ....................................................... ...........................2—4
Switching Ranges................................................................................................2—4
Setting the Time..................................................................................................2—4
Recalling Stored Measurements........................................................................2—5
Battery Installation.............................................................................................2—6
Error Codes..............................................................................................................2—9
Error Messages ....................................................................................................2—9
Page 5
1—5
Table of Contents, continued
Standard Calibration Adjust..........................................................................2—13
User-Entered Calibration ................................................................................2—15
Overview............................................................................................................2—15
Calibration Procedure Using Prepared Standards........................................2—17
Entering a Predetermined Calibration Curve..................................... ...... .....2 —2 0
Editing a User-entered or Factory Calibration Curve................ ....... ...... .....2 —2 2
Exiting the Calibration Routine......................................................................2—24
Deleting Calibration Points........................................................... ..................2—24
Retrieving the Factory Calibration.................................................................2—25
Maximum/Minimum Displayed Value....... ...... ....... .......................................2—26
Certification..........................................................................................................2—28
How to Order........................................................................................................2—35
Repair Service......................................................................................................2—37
Warranty................................................................................................................2—38
Page 6
1—6
Page 7
1—7
Safety Precautions
Please read this entire manual before unpacking, setting up, or operating this instrument. Pay particular attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
To ensure the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that which is specified in this manual.
Laboratory Safety
As part of good laboratory practice, pl eas e familiarize yourself with the reagents used in these procedures. Read all product labels and the m aterial safety data sheets (MSDS) before using them. It is always good practice to wear safety glasses when handling chemicals. Foll ow instructions carefully. Rinse thoroughly if contact occurs. If you have questions about reagents or procedures, please contact the manufacturer or distributor.
Use of Hazard Information
If multiple hazards exist, this manu al wi ll use th e sign al wo rd (Danger, Caution, Note) corresponding to the greatest hazard.
Page 8
1—8
Safety Precautions, continued
DANGER Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION Indicates a potentially hazardous situation that may result in minor or moderate injury.
NOTE Information that requires special emphasis.
Precautionary Labels
Please pay particular attention to labels and tags attached to the instrument. Personal in ju ry or da mage to the instrument could occur if not observed.
This symbol, if noted on the instrume n t, referen ces the instruct io n manual
for operational and/or safety inform ation.
Page 9
1—9
Specifications
Lamp: Light emitting diode (LED) Detector: Silicon photodiode Photometric precision: ± 0.0015 Abs Filter bandwidth: 15 nm
Wavelength: 528 nm
Absorbance range: 0–2.5 Abs Dimensions: 3.2 x 6.1 x 15.2 cm (1.25 x 2.4 x 6 inches) Weight: 0.2 kg (0.43 lbs)
Sample cells: 1 cm (10 mL), 25 mm (10 mL) Operating conditions: 0 to 50 °C (32 to 122 °F); 0 to 90% relative humidity
(noncondensing) Power supply: Four AAA alkaline batteries; approximate life is 2000 tests*
* Backlight usage will decrease battery life.
Page 10
1—10
OPERATION
DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the necessary Material Safety Data Sheets and become familiar with all safety procedures before handling any chemicals.
DANGER
La manipulation des échantillons chimiques, étalons et réactifs peut être danger euse. Lir e les Fiches de Données de Sécurité des Produits (FDSP) et se familiariser avec toutes les procédures de sécurité avant de manipuler tous les produits chimiques.
PELIGRO
La manipulación de muestras químicas, estándares y reactivos puede ser peligro s a. Revi se las fichas de seguridad de materiales y familiarícese con los procedimientos de seguridad antes de manipular productos químicos.
GEFAHR
Das Arbeiten mit chemischen Proben, Standards und Reagenzien ist mit Gefahren verbunden. Es wird dem Benutzer dieser Produkte empfohlen, sich vor der Arbeit mit sicheren Verfahrensweisen und dem richtigen Gebrauch der Chemikalien vertraut zu machen und alle entsprechenden Materialsicherheitsdatenblätter aufmerksam zu lesen.
PERIGO
A manipulação de amostras, padrões e reagentes químicos pode ser perigosa. Reveja a folha dos dados de segurança do material e familiarize-se com todos os procedimentos de segurança antes de manipular quaisquer produtos químicos.
PERICOLO
La manipolazione di campioni, standard e reattivi chimici può essere pericolosa. La preghiamo di prendere conoscenza delle Schede Techniche necessarie legate alla Sicurezza dei Materiali e di abituarsi con tutte le procedure di sicurezza prima di manipolare ogni prodotto chimico.
Page 11
1—11
Instrument Keys and Display
Item Description
1 POWER/BACKLIGHT Key 2
ZERO/SCROLL Key
3
MENU Key
4 Numeric Display 5 Range Indicator 6 Range Indicator 7 Menu Indicator 8 Calibration Adjusted Indicator 9 Battery Low Indicator
10
READ/ENTER Key
Page 12
1—12
Instrument Cap Cord
The instrument cap for the Pocket Colorimeter™ II doubles as a light shield. Accurate measurements cannot be obtained unless the sample or blan k is covered with the cap. Use the instrument cap cord to secure the cap to the body of the colorimeter and prevent loss of the cap. See Figu re 1 on page 1—13.
1. Loop the instrument cap cord through the ring on the cap.
2. Remove the battery compartment cover. Press the knotted end of the cord into
the hole indicated by the arro w.
3. Slide the cord into the slot on the battery compartment cover. Snap the cover into place.
Page 13
1—13
Instrument Cap Cord, continued
Figure 1 Attaching the Instrument Cap Cord
Page 14
1—14
Page 15
1—15
Chlorine, Free and Total, LR (0.02 to 2.00 mg/L Cl
2
)
Method 8021 (Free)
For water, treated water, estuary water, and sea water
Method 8167 (Total)
For water, treated water, wastewater, estuary water, and sea water
DPD Method*
USEPA accepted for reporting drinking water analyses** (free and total chlorine) and wa stew ater an alys es** (total chlori ne only)
Measuring Hints
Analyze samples immediately . Do not use plastic containers to collect sample s.
For best results, dedicate a set of cells to each free chlorine and total chlorine test.
Note:
The Pocket Colorimeter II is designed to measure solutions contained in sample cells.
DO NOT
dip the meter in the sample or pour the sample directly into the cell holder.
* Adapted from Standard Methods for the Examination of Water and Wastewater. ** Procedure is equivalent to USEPA method 330.5 for wastewater and Standard Method 4500-Cl G for drinking water.
Page 16
1—16
Chlorine, Free and Total, LR, continued
Using Powder Pillows (USEPA accepted for reporting)
1.Fill a 10-mL cell with
sample (the blank). Cap.
Note:
Samples must be analyzed immediately and cannot be preserved for later analysis.
2.Press the POWER key to
turn the meter on. The arrow sh ould indicate
the low range channel (LR).
Note:
See page 2—4 for information on selecting the correct range channel.
3.Remove the meter cap.
Place the blank in the cell holder with the diamond mark facing the keypad. Fit the meter cap over the cell compartment to cover the cell.
Note:
Wipe excess liquid and finger prints off sample cells.
Page 17
1—17
Chlorine, Free and Total, LR, continued
4.Press ZERO/SCROLL.
The display will show “- - - -” then “0.00”.
Remove the blank fro m the cell holder.
5.Fill a second 10-mL cell
to the 10-mL line with sample.
Note:
Do not use the same sample cells for free and total chlorine analysis without thoroughly rinsing the cells with sample between free and total tests.
6.Ad d the contents of one
DPD Free Chlorine Powder Pillow or one DPD Total Chlorine Powder Pillow to the sample cell (the prepar ed sample).
Note:
SwifTest™ Dispensers for Free or Total Chlorine can be used in place of powder pillows. See Using the SwifTest™
Dispenser (USEPA accepted for reporting) on page 1—
23.
Page 18
1—18
Chlorine, Free and Total, LR, continued
7.Cap and sh ake gently for
20 seconds.
Note:
Shaking dissipates bubbles that may form in samples with dissolved gases.
Note:
A pink color will develop if chlorine is present.
8.For free chlorine, place
the prepared sample cell in the cell holder. Cover with the instrument cap and proceed to step 10 within one minute after adding the DPD Free Pillow.
Note:
Accuracy is not affected by undissolved powder.
Note:
Wipe off sample cells.
9.For total chlorine, place
the prepared sample in the cell holder and cov er the cell with the instrument cap. Wait three to six minutes after adding the DPD Total Pillow.
Proceed to step 10.
Page 19
1—19
Chlorine, Free and Total, LR, continued
10. Press READ/ENTER.
The instrument will show “- - - -” followed by the results in mg/L chlorine.
Note:
If the sample temporarily turns yellow after reagent addition, or if the display shows overrange (page 2—12) dilute a fresh sample and repeat the test. A slight loss of chlorine may occur because of the dilution. Multiply the result by the appropriate dilution factor.
Page 20
1—20
Chlorine, Free and Total, LR, continued
Using AccuVac® Ampuls (USEPA accepted for reporting)
1.Fill a 10-mL sample cell
with sample (the blank). Cap. Collect at least 40 mL of sample in a 50-mL beaker.
Note:
Empty AccuVac Ampuls are available for use as blanks. See
OPTIONAL REAGENTS on page 1—47.
2.Press the POWER key to
turn on the meter. The arrow in the di splay
should indicate the low range cha nnel (LR).
See page 2—4 for information on selecting the correct range channel.
3.Remove the instrument
cap. Place the blank in the cell holder , with the diamond mark facing the keypad. Cover the cell with the instrument cap.
Note:
Wipe liquid and finger prints off sample cells.
Page 21
1—21
Chlorine, Free and Total, LR, continued
4.Press ZERO/SCROLL.
The display will show “- - - -” then “0.00”.
Remove the blank.
5.Fill a DPD Free Chlorine
Reagent AccuVac® Ampul or a DPD Total Chlorine Reagent AccuVac Ampul with sample.
Note:
Keep the tip immersed until the ampule fills completely.
6.Quickly inv ert the ampule
several times to mix. Wipe off any liquid or fingerprints.
Note:
A pink color will
form if chlorine is present.
Note:
Accuracy is not affected by undissolved powder.
Page 22
1—22
Chlorine, Free and Total, LR, continued
7.For free chlorine, place
the prepared sample in the cell holder . Cover the ampule with the instrument cap and proceed to step 9 within one minute after filling the AccuVac Ampul.
Note:
Wipe liquid off the
AccuVac Ampul.
8.For total chlorine, insert
the ampule into the cell holder, then cover with the instrument cap. Wait three to six minutes after filling the AccuV ac Ampul. Proceed to step 9.
Note:
Wipe liquid off the
AccuVac Ampul.
9.Press READ/ENTER. The
instrument will show
- - - - followed by the results in mg/L chlorine.
Note:
If the sample temporarily turns yellow after reagent addition, or if the display shows overrange (page 2—12) dilute a fresh sample and repeat the test. A slight loss of chlorine may occur because of the dilution. Multiply the result by the appropriate dilution factor.
Page 23
1—23
Chlorine, Free and Total, LR, continued
Using the SwifTest™ Dispenser (USEPA accepted for reporting)
1.Fill a 10-mL cell with
sample (the blank). Cap.
Note:
Samples must be analyzed immediately and cannot be preserved for later analysis.
2.Press the POWER key to
turn the meter on. The arrow sh ould indicate
the low range channel (LR).
Note:
See page 2—4 for information on selecting the correct range channel.
3.Remove the meter cap.
Place the blank in the cell holder with the diamond mark facing the keypad. Fit the meter cap over the cell compartment to cover the cell.
Note:
Wipe excess liquid and finger prints off sample cells.
Page 24
1—24
Chlorine, Free and Total, LR, continued
4.Press ZERO/SCROLL.
The display will show “- - - -” then “0.00”.
Remove the blank fro m the cell holder.
5.Fill a second 10-mL cell
to the 10-mL line with sample.
Note:
Do not use the same sample cells for free and total chlorine analysis without thoroughly rinsing the cells with sample between free and total tests.
6.Use the SwifTest™
Dispenser to add one dispensation of DPD Free Chlorine reagent or one dispensation of DPD Total Chlorine reagent to the sample cell (the prepared sample).
Page 25
1—25
Chlorine, Free and Total, LR, continued
7.Cap and sh ake gently for
20 seconds.
Note:
Shaking dissipates bubbles that may form in samples with dissolved gases.
Note:
A pink color will develop if chlorine is present.
8.For free chlorine, place
the prepared sample cell in the cell holder and cover with the inst rument cap. Proceed to step 10 within one minute after adding the DPD Free Pillow.
Note:
Accuracy is not affected by undissolved powder.
Note:
Wipe off sample cells.
9.For total chlorine,
place the prepared sample in the cell holder and cover with the instrument cap. Wait three to six minutes after adding the DPD Total Pillow.
Proceed to step 10.
Page 26
1—26
Chlorine, Free and Total, LR, continued
10. Press READ/ENTER.
The instrument will show - -
- - followed by the results in mg/L chlorine.
Note:
If the sample temporarily turns yellow after reagent addition, or if the display shows overrange (page 2—12) dilute a fresh sample and repeat the test. A slight loss of chlorine may occur because of the dilution. Multiply the result by the appropriate dilution factor .
Page 27
1—27
Chlorine, Free and Total, LR, continued
Accuracy Check
Standard Additions Method
1. Use the ampule breake r to snap the neck off a low range Chlorine Standard Solution Ampule, 20–30 mg /L Cl
2
.
2. Use a TenSette® pipet to add 0.1, 0.2, and 0.3 mL of standard to three 10-mL samples. Swirl gently to mix. (For AccuVac Ampuls, use 0.2, 0.4, and 0.6 mL of standard and a 25-mL sample in a 50-mL beaker.)
3. Analyze a 10-mL aliquot of each sample as described in the procedure. Each
0.1 mL of standard will cause an incremental increase in chlorine. The exact value depends on the concentration of the ampule sta nd ard. Ch eck the certificate enclosed with the ampules for the chlorine concentration and calculation of the expected chlorine increase.
Standard Solution Method
Standard solutions for chlorine are difficult and time-consuming to prepare. Errors can occur if attention to detail is not addressed during preparation of the standards. The calibration curve is prepared under rigorous analytical laboratory conditions. Use the factory calibration for most no rm al testing.
Page 28
1—28
Chlorine, Free and Total, LR, continued
A user calibration or a user-prepared chlorine standard may be required by a regulatory official or agency. Two options are available on the Pocket Colorimeter II to meet this requirement.
A chlorine standard may be prepared and used to validate the calibration curve using the Standard Calibration Adjust feature (see page 2—13 for more information). The concentration of the prepared standard must b e determined with an alternate instrument such as a spectrophotometer, colorimeter, or by using an alternate method such as amperometric titr ation. The concentration of the chlorine standard for the LR procedure must be between 0.50 and 1.50 mg/L chlorine.
In addition, a user-generated calibration curve can be programmed into the Pocket Colorimeter™ II. See User-Ent ered Calib rati on on pa ge 2—15 for more information.
Page 29
1—29
Chlorine, Free and Total, LR, continued
Interferences
Interfering Substance Interference Levels and Treatments
Acidity Greater than 150 mg/L CaCO3. May not develop full color or
color may fade instantly. Neutralize to pH 6–7 with 1 N Sodium Hydroxide. Determine amount to be added on a separate 10-mL sample, then add the same amount to the sample being tested. Correct for the additional volume.
Alkalinity Greater than 250 mg/L CaCO
3
. May not develop full color or color may fade instantly. Neutralize to pH 6–7 with 1 N Sulfuric Acid. Determine amount to be added on a separate 10-mL sample, then add the same amount to the sample being tested. Correct for the additional volume.
Bromine, Br
2
Interferes at all levels
Hardness No effect at less than 1,000 mg/L as CaCO
3
Iodine, I
2
Interferes at all levels
Page 30
1—30
Chlorine, Free and Total, LR, continued
Method Performance
Estimated Detection Limit (EDL) = 0.02 mg/L Typical precision (95% confindence interval) = 1.00 ± 0.05 mg/L
Manganese, Oxidized (Mn
4+
, Mn7+) or Chromium,
Oxidized (Cr
6+
)
1. Adjust sample pH to 6–7.
2. Add 3 drops Potassium Iodide (30-g/L) (Cat. No. 343-32) to a
10-mL sample.
3. Mix and wait one minute.
4. Add 3 drops Sodium Arsenite (5-g/L) (Cat. No. 1047-32) and
mix.
5. Analyze 10 mL of the treated sample as described in the procedure.
6. Subtract the result from this test from the original analysis to obtain the correct chlorine concentration.
Monochloramine, LR method
Causes a gradual drift to higher readings. When read within 1 minute after reagent addition, 3 mg/L monochloramine causes less than a 0.1 mg/L increase in the reading.
Ozone Interferes at all levels.
Interfering Substance Interference Levels and Treatments
Page 31
1—31
Chlorine, Free and Total, HR (0.1 to 8.0 mg/L Cl
2
)
For water, treated water, estuary water, and seawater (Free Chlorine) For water, treated waters, wastewater, estuary water, and seawater (Total Chlorine)
DPD Method*
USEPA accepted for reporting drinking water analyses (fr ee and total chlorine) and wastewater analyses (total chlorine).
Measuring Hints
If the chlorine concentration is typically less than 2 mg/L, use the low range procedure.
Analyze samples immediately . Do not use plastic containers to collect samples. For best results, dedicate a set of sample cells to each test (free and total).
If the sample temporarily turns yellow after reagent ad dit io n or shows overrange (page 2—12), dilute a fresh sample and repeat the test. A slight loss of chlorine may occur. Multiply the result by the dilution factor.
* Adapted from Standard Methods for the Examination of Water and Wastewater.
Page 32
1—32
Chlorine, Free and Total, HR, continued
High range free chlorine determinations are subject to variable levels of interferences from monochloramine. See Interferences on page 1—41.
Note:
The Pocket Colorimeter II is designed to measure solutions contained in sample cells.
DO NOT
dip the meter in the sample or pour the sample directly into the cell holder.
Page 33
1—33
Chlorine, Free and Total, HR, continued
Using Powder Pillows
1.Fill a 1-cm/10-mL cell
with sample (the blank). Cap.
Note:
Samples must be analyzed immediately and cannot be preserved for later analysis.
2.Press the POWER key to
turn the meter on. The arrow sh ould indicate the
high range channel (HR).
Note:
See page 2—4 for information on selecting the correct range channel.
3.Remove the meter cap.
Place the blank into the cell holder, with the diamond mark facing the back of the cell holder. Cover the cell with the cap.
Note:
Wipe liquid off
sample cells.
Page 34
1—34
Chlorine, Free and Total, HR, continued
4.Press: ZERO/SCROLL
The display will show “- - - -” followed by “0.0”.
Remove the blank.
5.Fill another 1-cm/10-mL
sample cell to the 5-mL line with sample. Cap.
Note:
Do not use the same sample cells for free and total chlorine without thoroughly rinsing the cells between the free and total tests.
6.Ad d the contents of two
DPD Free Chlorine or two DPD Total Chlorine Powder Pillows to the sample cell (the prepare d sample). Cap the cell and shake gently for 20 seconds.
Note:
Gentle shaking dissipates bubbles which may form in samples containing dissolved gases.
Page 35
1—35
Chlorine, Free and Total, HR, continued
7.For free chlorine, place the
prepar ed sample cell in the cell holder and cover with the instrument cap within one minute after adding the DPD Free Pillow.
Proceed immediately t o step 9.
Note:
The SwifTest™ Dispenser can be used in place of the powder pillow (see page 1—36).
8.For total chlorine, place the
prepared sample in the cell holder and cover with the instrument cap. Wait three to six minutes after adding the DPD Total Pillows. Pr oceed to step 9.
Note:
Wipe off sample cells.
9.Press READ/ENTER.
The instrument will show “- - - -” followed by the results in mg/L chlorine (Cl
2
).
Page 36
1—36
Chlorine, Free and Total, HR, continued
Using the SwifTest™ Dispenser
1.Fill a 1-cm/10-mL cell
with sample (the blank). Cap.
Note:
Samples must be analyzed immediately and cannot be preserved for later analysis.
2.Press the POWER key to
turn the meter on. The arrow s hould indicate the
high range channel (HR).
Note:
See page 2—4 for information on selecting the correct range channel.
3.Remove the meter cap.
Place the blank into the cell holder, with the diamond mark facing the back of the cell holder. Cover the cell with the cap.
Note:
Wipe liquid off sample
cells.
Page 37
1—37
Chlorine, Free and Total, HR, continued
4.Press: ZERO/SCROLL
The display will show “- - - -” followed by “0.0”.
Remove the blank.
5.Fill another 1-cm/10-mL
sample cell to the 5-mL line with sample. Cap.
Note:
Do not use the same sample cells for free and total chlorine without thoroughly rinsing the cells between the free and total tests.
6.Use the SwifTest™
Dispenser to add two dispensations of DPD Free Chlorine reagent or DPD Total Chlorine reagent to the sample cell (the prepare d sample). Cap the cell and shake gently for 20 seconds.
Note:
Gentle shaking
dissipates bubbles.
Page 38
1—38
Chlorine, Free and Total, HR, continued
7.For free chlorine, place the
prepar ed sample cell in the cell holder and cover with the instrument cap within one minute after adding the DPD Free Pillow.
Proceed immediately to
step 9
.
Note:
Wipe liquid off sample
cells.
8.For total chlorine, place the
prepared sample in the cell holder and cover with the instrument cap. Wait three to six minutes after adding the DPD Total Pillows. Pr oceed to
step 9
.
Note:
Wipe liquid off sample
cells.
9.Press READ/ENTER.
The instrument will show “- - - -” followed by the results in mg/L chlorine (Cl
2
).
Page 39
1—39
Chlorine, Free and Total, HR, continued
Accuracy Check
Standard Additions Method
a. Use the ampule breake r to snap the neck off a high range Chlorine
Standard Solution Ampule, 50–70 mg/L Cl
2
.
b. Use a TenSette
®
pipet to add 0.1, 0. 2, and 0.3 mL of standard to three 5-mL
samples. Swirl gently to mix.
c. Analyze each sample as described in the procedure. Each 0.1 mL of
standard will cause an incremental in crease in chlorine. The exact value depends on the concentration of the ampule standard. Check the certificate enclosed with the ampules for calculation of the expected increase in the chlorine concentration.
Standard Solution Method
Standard solutions for chlorine are difficult and time-consuming to prepare. Errors can occur if attention to detail is not addressed during preparation of the standards. The calibration curve is prepared under rigorous analytical laboratory conditions. Use the factory calibration for most no rm al testing.
Page 40
1—40
Chlorine, Free and Total, HR, continued
A user calibration or a user-prepared chlorine standard may be required by a regulatory official or agency. Two options are available on the Pocket Colorimeter™ II to meet this requirement.
A chlorine standard may be prepared and used to validate the calibration curve using the Standard Calibration Adjust feature (see page 2—13 for more information). The concentration of the prepared standard must b e determined with an alternate instrument such as a spectrophotometer, colorimeter, or by using an alternate method such as amperometric titr ation. The concentration of the chlorine standard for the HR procedure must be between 4.5 and 7.0 mg/L chlo rine.
In addition, a user-generated calibration curve can be programmed into the Pocket Colorimeter™ II. See User-Ent ered Calib rati on on pa ge 2—15 for more information.
Page 41
1—41
Chlorine, Free and Total, HR, continued
Interferences
See additional Interferen ces on page 1—29.
Interfering Substance Interference Levels and Treatments
Monochloramine For conventional free chlorine disinfection (beyond the
breakpoint), monochloramine concentrations are very low. If monochloramine is present in the sample, its interference in the free chlorine test varies with the temperature, the relative amount of monochloramine to free ammonia, and the time required to do the analysis. Approximate interference levels of monochloramine in the free chlorine test are listed below (as mg/L Cl
2
).
NH2Cl Sample Temperature °C (°F)
(as Cl2) 5 (40) 10 (50) 20 (68) 30 (83)
1.2 0.2 0.2 0.3 0.3
2.5 0.4 0.5 0.6 0.6
3.5 0.5 0.6 0.7 0.8
Page 42
1—42
Chlorine, Free and Total, HR, continued Method Performance
Estimated Detection Limit (E DL) = 0.1 mg /L Typical precision (95% confidence interval) = 5.0 ± 0.2 mg/L
Page 43
1—43
Spec™ Secondary Standards for DPD Chlorine
Note:
Due to improvements in the optical system of the Pocket Colorimeter™ II, the tolerance ranges and values on the Certificate of Analysis of previously purchased Spec standards may no longer be valid. Obtain a new set of standards, or use the Pocket Colorimeter II to assign new values to existing standards.
The DPD Chlorine Spec Secondary Standards are avai lable to quickly check the repeatability of the Pocket Colorimeter™ II instrument (see OPTIONAL REAGENTS on page 1—47).
After initial measurements for the Spec standards are collected in the low range (LR) channel, the standards can be re-checked as often as desired to ensure the instrument is working consistently.
The standards do not ensure reagent quality nor do they ensure the accuracy of the test results. Analysis of real standard solutions using the kit reagents is required to verify the accuracy of the entire Pocket Colorimeter system. The Spec Standards should NEVER be used to calibrate the instrument. The certificate of analysis lists the expected value and tolerance for each Spec Standard.
Note:
Before proceeding, make sure the instrument is in the low
(LR)
range channel. See
Switching Ranges on page 2—4.
Page 44
1—44
Spec™ Secondary Standards for DPD Chlorine, continued Using the Spec Standards
1. Place the colorless Spec√ blank into the cell holder with the alignment mark facing the keypad. Tightly cover the cell with the instrument cap.
2. Press
ZERO. The display will show “0.00”.
3. Place the STD 1 cell into the cell holder. Tightly cover the cell with the instrument cap.
4. Press
READ/ENTER. Record the concentration measurement.
5. Repeat steps 3 and 4 with cells labeled STD 2 and STD 3.
6. Compare these measurements with previous measurements to verify the
instrument is performing consistently. (If these are the first measurements, record them for comparison with later measurements.)
Note:
If the instrument is user-calibrated, initial standard measurements of the Spec
Standards will need to be performed again for the user calibration.
Page 45
1—45
Summary of Method
Chlorine can be present in water as free chlorine and as combined chlorine. Both forms can coexist in the same solut ion and can be determined toget her a s total chlorine. Free chlorine is present as hypochlorous acid or hypochlorite ion. Combined chlorine represents a combination of chlorine-containing compunds including but not limited to monochloramine, dichloramine, nitrogen trichloride, and other chloro derivatives. The combined chlorine oxidizes triiodide ion (I
3
) to
iodine (I
2
). The iodine and free chlorine reacts with DPD (N,N-diethyl-p-phenylenediamine) to form a red solution. The color intensity is proportional to the total chlorine concentration. To determine the concentration of combined chlorine, run a free chlorine test and a total chlorine test. Subtract the results of the free chlorine test from the total chlorine test to obtain the combined chlorine concentration.
The range of analysis using the DPD method for chlorine can be extended by adding more indicator in proportion to sample volume. For example, two powder pillows of DPD Chlorine Reagent are added to a 5-mL sample portion to extend the range of analysis.
Page 46
1—46
Replacement Parts
REQUIRED REAGENTS
Description Unit Cat. No.
Free Chlorine Tests
DPD Free Chlorine Reagent Powder Pillows..............................100/pkg .....21055-69
or
DPD Free Chlorine Reagent AccuVac
®
Ampuls
(low range test only) ..................................................... ....... ...... 25/pkg ....25020-25
SwifTest™ DPD Free Chlorine Reagent with dispenser..........250* tests ....28023-00
Total Chlorine Tests
DPD Total Chlorine Reagent Powder Pillows.............................100/pkg .....2 1056-69
or
DPD Total Chlorine Reagent AccuVac
®
Ampuls
(low range test only) ..................................................... ....... ...... 25/pkg ....25030-25
SwifTest™ DPD Total Chlorine Reagent with disp enser.........250* tests ....28024-00
REQUIRED APPARATUS (AccuVac® Ampuls)
Beaker, 50 mL ......................................................................................each ......500-41H
* 125 tests when performing the high range test
Page 47
1—47
Replacement Parts, continued
OPTIONAL REAGENTS
Description Unit Cat. No.
Chlorine Standard Solution, 25–30 mg/L, 2-mL ................. .......20/pkg.....26300-20
Chlorine Standard Solution, 50–75 mg/L, 2-mL ................. .......20/pkg.....14268-20
Chlorine Standards, secondary, Spec
™,
0.0, 0.2, 0.8 and 1.5 mg/L............................................................4/pkg.....26353-00
Empty AccuVac® Ampuls (for reading blanks)...........................25/pkg.....26779-25
Potassium Iodide Solution, 30 g/L....................................100 mL MDB*.........343-32
Spec
Secondary Standards, Chlorine .............................................each.....26353-00
Sodium Arsenite Solution, 5 g/L........................................ 100 mL MDB.......1047-32
Sodium Hydroxide Standard Solution, 1 N...................... 100 mL MDB.......1045-32
Sulfuric Acid Standard Solution, 1 N................................100 mL MDB.......1270-32
SwifTest™ DPD Free Reagent Replacement Vial...................250** tests.....21055-60
SwifTest™ DPD Total Reagent Replacement Vial..................250** tests.....21056-60
Wate r, deionized.............................................................................. .......4 L.........272-56
* Marked Dropper Bottle ** 125 tests when performing the high range test
Page 48
1—48
Replacement Parts, continued
OPTIONAL APPARATUS
Description Unit Cat. No.
AccuVac® Snapper Kit.........................................................................each ....24052-00
Ampule Breaker Kit..................................................... ........................ each ....24846-00
Batteries, AAA, alkaline ...................................................................4/pkg ....46743-00
Caps for 10-mL sample cells.......... ............................................... 12 /pkg ....24018-12
Cylinder, graduated, 25 mL. glass.....................................................each ........ 508-40
Cylinder, graduated, 100 mL, glass...................................................each ........ 508-42
sens
ion1 Basic Portable pH Meter, with electrode ......................each .....51700-10
Pipet, TenSette
®
, 0.1 to 1.0 mL ..........................................................each .....19700-01
Pipet Tips, For 19700-01 TenSette
®
Pipet.................................... 50/pkg .... 21856-96
Sample Cells, 10-mL with screw caps....................... ...... ................6 / pkg . ...24276-06
REPLACEMENT PARTS
Cap for 1-cm/10 mL sample cell .......................................................each . ...52626-00
Instrument Cap/light shield................... .............................................each ....59548-00
Instrument Manual................................. ....... ......................................each .... 59570-88
Sample Cells, 1-cm/10-mL.................................. .............................2/pkg ....48643-02
Page 49
2—1
Section 2
Instrument Manual
Page 50
2—2
Page 51
2—3
Instrument Operation
Key Functions
Key Description Function
POWER On/Off/Backlight
To turn on the backlight, turn on the instrument, then press and hold the power key until the backlight turns on. Press and hold again to turn off the backlight. This key functions the same in all instrument modes and ranges.
ZERO/SCROLL In measurement mode, sets the instrument to zero.
In menu mode, scrolls through menu options. Also scrolls numbers when entering or editing a value.
READ/ENTER In measurement mode, initiates sample measurement.
In menu mode, selects a menu option. When entering numbers, moves one space to the right and executes the function when the entry is complete.
Page 52
2—4
Instrument Operation, continued
Menu Selections
Press the MENU key to access the menu selections.
Switching Ranges
1. Press the MENU key. The display will sho w “SE L”. A flashing arrow indicates the current range.
2. Press the
READ/ENTER key to toggle between ranges.
3. Press
MENU again to accept and exit back to the measurement screen.
Setting the Time
1. Press the MENU key, then press the ZERO/SCROLL ke y un til th e di spla y shows a time in the “00:00” format.
MENU Enter/Exit the menu mode
Press and hold for approximately 5 seconds to enter user-entered method mode.
Key Description Function
Page 53
2—5
Instrument Operation, continued
2. Press READ/ENTER. The digit to be edited will flash.
3. Use the
ZERO/SCROLL key to change the entry, then press READ/ENTER to
accept and advance to the next digit. The time is entered in 24-hour format.
Recalling Stored Measurements
1. Press the MENU key, then press the ZERO/SCROLL key until the display shows
RCL. The instrument automatically stores the last 10 measurements.
2. In RCL, press
READ/ENTER to recall the stored measurements, beginning with
the most recent measurement taken. The meter stores the measurement number as 0 1 (most recent) through 1 0 (oldest), the time the measurement was taken, and the measurement value. The
ZERO/SCROLL key allows for selection
of a specific measurement by number. The
READ/ENTER key scrolls thro ugh all
stored data points.
Page 54
2—6
Instrument Operation, continued
Battery Installation
Figure 1 on page 2—7 provides an exp lod ed vi ew of battery installation.
1. Unhook the latch and remove the bat tery co mpartment cover. The polarities are shown on the battery holder.
2. Place the four batteries provided with the instrument in the holder as indicated and replace the battery compartment cover. The display will show the software version number (e.g., “P 1.6”) after correct battery installation.
When replacing discharged batteries, always replace the complete set of four alkaline batteries. Rechargeable batteries are not recommended and cannot be recharged in the instrument.
Note:
The Low Battery icon will appear on the display when the batteries have 10 % battery life remaining. The battery icon will flash when the batteries are too low to complete measurements. See Instrument Keys and Display on page 1—11.
Page 55
2—7
Instrument Operation, continued
Figure 1 Battery Installation
Page 56
2—8
Page 57
2—9
Error Codes
When the instrument canno t perfo r m th e funct io n in it iat ed by the operator, an error message will appear in the display. Refer to the appropriate message information below to determine what the problem is and how it can be corrected. Resolve error messages in the order that they appear on the display. Service Centers are listed in page 2—37.
Error Messages
1. E-0 No Zero (User mode)
Error occurs when trying to read a standard in the user calibration mode before setting the meter to zero.
Zero the instrument on an appropriate blank.
2. E-1 Ambient Light Er ro r
There is too much light present to take a valid measurement.
Verify instrument cap is correctly seated.
If the problem persists, contact a Service Center (page 2—37).
Page 58
2—10
Error Codes, continued
3. E-2 LED Error
The LED (light source) is out of regulation.
Replace batteries.
Verify LED lights up (inside the cell holder) when the READ/ENTER or
ZERO/SCROLL key is pressed.
If the problem persists, contact a Service Center (page 2—37).
Note: When an E-1 or E-2 error occurs on a measurement, the display will show “_.__”. (The
decimal place is determined by the chemistry.) If the E-1 or E-2 error occurs while zeroing the meter , the meter will require the user to re-zero.
4. E-3 Standard Adjust Error
The value obtained on the prepared standard exceeds the adjustment limits allowed for the standard concentration, or the concentration of the standard is outside the concentration range allowed for standard calibration adjust.
Prepare the standard and rerun according to the procedure.
Prepare a standard at or near the recommended concentrations given in the
procedure.
Verify that the concentration of the standard ha s been entered correctly.
Page 59
2—11
Error Codes, continued
If the problem persists, contact a Service Center (page 2—37).
5. E-6 Abs Error (User mode )
Indicates that the absorbance value is invalid, or indicates an attempt to make a curve with less than two points.
Enter or measure the absorbance value again.
If the problem persists, contact a Service Center (page 2—37).
6. E-7 Standard Value Error (User mode)
Standard concentratio n is equ al to another standard concent r ati on tha t is already entered.
Enter the correct standard concentration.
If the problem persists, contact a Service Center (page 2—37).
7. E-9 Flash Error
The meter is unable to save data.
If the problem persists, contact a Service Center (page 2—37).
Page 60
2—12
Error Codes, continued
8. Underrange—flashing number below stated test range
Verify instrument cap is correctly seated.
Check zero by measuring a blank. If error recurs, re-zero the instrument.
If the problem persists, contact a Service Center (page 2—37).
Note: See Maximum/Minimum Displayed Value on page 2—26 for more information.
9. Overrange—flashing number above stated test range
Note: Flashing value will be 10% over the upper test limit.
Check for light blockage.
Dilute and retest sample.
Note: See Maximum/Minimum Displayed Value on page 2—26 for more information.
Page 61
2—13
Standard Calibration Adjust
The Pocket Colorimeter™ II instrument is factory-calibrated and ready for use without user calibration. Use of the factory calibration is recommended unless the user is required to generate a calibration. The Standard Calibration Adjust can be used to meet regulatory requirements.
This feature allows the factory default calibration curve to be adjusted with a known standard. Use the standard described in the procedure.
1. Place a blank in the meter (in measurem ent mo de). Press
ZERO/SCROLL.
2. Place the reacted standard in the meter. Press
READ/ENTER.
3. Press
MENU, then press ZERO/SCROLL until the display shows “SCA”.
4. Press
READ/ENTER to display the standa rd calibration adjust valu e.
5. Press
READ/ENTER to adjust the curve to the displayed value. The meter will
return to the measurement mode and the Calibration Adjusted icon will appear in the display window.
If an alternate concentration is used, or if a standard concentration is not given:
6. Repeat steps 1–4.
Page 62
2—14
Standard Calibration Adjust, continued
7. Press ZERO/SCROLL to access the Edit function, then press READ/ENTER to begin editing. The digit to be edited will flash. Use the
ZERO/SCROLL key to
change the entry, then press
READ/ENTER to accept and advance to the
next digit. When the last digit is entered, press
READ/ENTER and the meter will adjust the
curve to the value entered. The meter will return to measurement mode and the Calibration Adjusted icon will appear in the display window.
To turn off Standard Calibration Adjust (SCA):
1. Press
MENU.
2. Press
ZERO/SCROLL until “SCA” app e ars in the displa y.
3. Press
READ/ENTER, then press ZERO/SCROLL until “Off” appears in the display.
4. Press
READ/ENTER to turn off SCA.
Note: Perform another standard calibration adjust to turn SCA on again. Note: For meters with factory-calibrated ranges or methods, Standard Calibration Adjust
(SCA) will be disabled when a user-entered method is programmed into the meter. To turn SCA back on, restore the meter to factory default calibration. See Retrieving the Factory Calibration on page 2—25.
Page 63
2—15
User-Entered Calibration
Overview
The Pock et Colorimeter™ II will accept a user-prepared calibration curve. The curve can extend from 0 to 2.5 absorbance. A user-prepared calibration curve may be entered into any channel that does not contain a factory-progr ammed curve. These channels are labeled “abs” on instruments having a single factory calibration or are labeled “1” and “2” on the uncalibrated single wavelength instruments. Any chemistry that can be run at the instrument wavelength may be user-entered in these channels.
Using prepared standard solutions that cover the range of interest, th e mete r generates a calibration curve by calculatin g th e stra igh t-l in e segments between each standard entered. A calibration curve may be entered using the keypad. Factory-entered calibration curves may also be recalculated or adjusted using the same procedure.
To enter the user-entered calibration mode, press the
MENU key and hold it down
until the display shows “USER” (about 5 seconds), followed by “CAL”. Press
ZERO/SCROLL to scroll through the options.
Page 64
2—16
User-Entered Calibration, continued
CAL—Used to enter and edit standard values and measure absorbance values,
or review the existing calibration.
Edit—Used to enter and edit standard values and ab sorbance values with the
keypad or review the existing calibration. Used to enter a predetermined calibration curve.
dFL—Used to return the instrument back to the default factory calibration.
User-entered calibrations are stored upon exit from the calibration or edit modes.
Note: To return to factory settings, following the instructions in Retrieving the Factory
Calibration on page 2—25.
If the instrument is shut off or loses power during data entry, all edits will be lost. Automatic shut-off in user-entered cali br at io n entry mode is 60 minutes.
CAL and Edit Submenus
In CAL mode, standard values are entered and absorbance values are measured. In Edit mode, standard and absorbance values are entered.
To select CAL from the User menu, press READ/ENTER.
To sel ect Edit from the User menu, press ZERO/SCROLL and READ/ENTER.
Page 65
2—17
User-Entered Calibration, continued
Once in the CAL or Edit optio n, press the READ/ENTER key to navigate through
each option.
Note: Press
ZERO/SCROLL
to quickly scroll through each option.
Calibration Procedure Using Prepared Standards
Note:
Deionized water or a reagent blank can be used to zero during the calibration procedure. Calibrations generated with deionized water as the zero will give less accurate results if the reagent blank is significantly more turbid or colored than deionized water. Use the deionized water or the reagent blank as the zero concentration point (S0) in the following calibration procedure.
1. Turn on the instrument and select the range to be calibrated. An arrow at the top of the display will point to the selected range. To change ranges, press the
MENU key, then use the READ/ENTER ke y to to gg le bet w een range s 1 and 2.
Press
MENU again to return to measurement mode.
2. Follow the procedure for the chemical method to be calibrated . Prepare a reagent blank (if needed) and a standard solution. Allow the color to develop fully.
Page 66
2—18
User-Entered Calibration, continued
3. Insert the reagent blank or deionized water into the meter and cover with the cap. Press the
ZERO/SCROLL key . The meter will display “- - - -”, followed by
“0.000”. This initializes (zeroes) the meter.
4. Press the
MENU key and hold it down until the display shows “USER”, followed
by “CAL”. Press
READ/ENTER to enter the calibration mode.
5. In factory-calibrated meters, S0 will appear in the display.
Note: When recalibrating a factory-calibrated meter or range, RES (resolution) cannot
be changed.
6. In uncalibrated meters or meters with ranges labeled Abs, “RES” will appear. Press
ZERO/SCROLL to review the current resolution (decimal placement). Press
ZERO/SCROLL again to accept the current resolution. To change the resolution,
press
READ/ENTER, then ZERO/SCROLL to change the resolution. Press
READ/ENTER to accept the new resolution. “S0” will appear on the display.
7. Press the
READ/ENTER key again, then enter the blank value.
Note: Press the
READ/ENTER
key to move from digit to digit. Use the
ZERO/SCROLL
key to
change the number.
8. After completing entry of the blank value, press the READ/ENTER key. The display will show “A0”.
Page 67
2—19
User-Entered Calibration, continued
9. Insert the reagent blank or deionized water into the cell holder. Cover the
blank with the instrument cap.
10. Press the
READ/ENTER key . The meter will measure and display the absorbance
value for “S0”.
11. Remove the sample blank. Press the
ZERO/SCROLL key . “S1” will appear. Press
the
READ/ENTER key, then enter the fir st standard value.
Note: Press the
READ/ENTER
key to move from digit to digit. Use the
ZERO/SCROLL
key to
change the number.
12. After completing entry of the first standard value, press the READ/ENTER key.
The display will show “A1”.
13. Insert the first reacted standard solution into the cell holder. Cover the
prepared standard with the instrument cap.
14. Press the
READ/ENTER key . The meter will measure and display the absorbance
value for S1.
15. The calibration is complete with two points. If additional standards are
required, press
ZERO/SCROLL until “Add” appears on the display. Repeat
steps 1114 to enter additional standards.
Page 68
2—20
User-Entered Calibration, continued
16. Press the MENU key twice to exit and accept the changes. The instrument will use this calibration to determine the displayed concentration of future sample measurements.
Entering a Predetermined Calibration Curve
Note: Entering a predetermined calibration curve requires at least two data pairs. Each data
pair requires a concentration value and the absorbance value for the given concentration. Up to 10 data pairs may be entered. This procedure uses the Edit mode.
1. Turn on the instrument and select the range to be calibrated. An arrow at the top of the display will point to the selected range. To change ranges, press the
MENU key, then use the READ/ENTER key to to gg le bet ween range s 1 an d 2.
Press
MENU again to return to measurement mode.
2. Press the
MENU key and hold it down until the display shows “USER”, followed
by “CAL”. Press
ZERO/SCROLL to scroll to EDIT. Press READ/ENTER.
3. In uncalibrated meters or in Abs range, “RES” will appear. Press ZERO/SCROLL. To change the resolution (decimal placement), press
READ/ENTER. Press
ZERO/SCROLL to select the new resolution, then press READ/ENTER to accept.
“S0” will appear on the display.
Page 69
2—21
User-Entered Calibration, continued
4. Enter the concentration value and absorbance value of the first data pair
(S0, A0).
5. To enter the S0 value, press
READ/ENTER. Use the ZERO/SCROLL key to select
the numerical value, then press the
READ/ENTER key to accept the entry and
advance to the next decimal place. Repeat this sequence until the S0 concentration value is entered.
6. After editing the S0 value, press
READ/ENTER to accept. “A0” will appear on
the display.
7. To enter the absorbance value for S0, press the
READ/ENTER key to go to entry
mode. Use the
ZERO/SCROLL key to select th e numerical value, then press the
READ/ENTER key to accept the entry and advance to the next decimal place.
Repeat this sequence until the absorbance value for S0 is entered.
8. After enterin g A0, press
READ/ENTER to accept. “S1” will appear on the
display.
9. Repeat steps 5 through 8 for each standard value and absorbance value pair in
the calibration curve
Note: After A1 is entered, Add will appear in the display. If additional data pairs are to be
entered, press
READ/ENTER
and continue with step 9.
Page 70
2—22
User-Entered Calibration, continued
10. When all the calibration data has been entered, press MENU twice to return to the measurement mode.
Editing a User-entered or Factory Calibration Curve
1. Press the MENU key and hold it down until the display shows “USER”, followed by “CAL”. Press
ZERO/SCROLL until EDIT appears.
2. Press the
READ/ENTER key to enter Edit mode. In factory-calibrated meters,
“S0” will appear in the display.
Note: When editing a factory-calibrated meter or range, RES (resolution) cannot be changed. Note: When RES or S0 appears in the display, press
ZERO/SCROLL
to quickly scroll to the
data to be edited.
3. In uncalibrated meters or in Abs range, “RES” will appear . Press ZERO/SCROLL to review the current resolution. Press
ZERO/SCROLL again to accept the
displayed resolution. To change the resolution (decimal placement), press
READ/ENTER. Press ZERO/SCROLL to select the new resolution, then press READ/ENTER to accept. “S0” will appear on the di splay.
4. Press
READ/ENTER. The current concentration value for S0 will appear on the
display.
Page 71
2—23
User-Entered Calibration, continued
5. To edit the S0 value, press READ/ENTER. Use the ZERO/SCROLL k ey to select the
numerical value, then press the
READ/ENTER key to accept the entry and
advance to the next decimal place. Repeat this sequence until the S0 concentration value is entered.
6. After editing the S0 value, press
READ/ENTER to accept. “A0” will appear on
the display.
7. To edit the absorbance value for S0, press th e
READ/ENTER key to go to entry
mode. Use the
ZERO/SCROLL key to select th e numerical value, then press the
READ/ENTER key to accept the entry and advance to the next decimal place.
Repeat this sequence until the absorbance value for S0 is entered.
8. After editing A0, press
READ/ENTER to accept. “S1” will appear on the display .
9. Repeat steps 4 through 8 for each standard value and absorbance value pair in
the calibration curve.
10. When all calibration data has been reviewed or edited, “ADD” will appear in
the display.
11. Press
READ/ENTER to add more calibration points, or press MENU twice to
return to the measurement mode.
Page 72
2—24
User-Entered Calibration, continued
Note: When a factory calibration curve has been edited, the “calibration adjust” icon will
appear in the display.
Exiting the Calibration Routine
Exit the calibration routine by pressing th e MENU key to return to measurement mode. The instrument uses the last completed user-entered calibration or the factory calibration if no user-entered calibration has been completed.
Deleting Calibration Points
1. Select the range containing user-entered calibration points. See Switching
Ranges on page 2—4.
2. Press and hold the
MENU key until “USER”, then “CAL” appears. Press
READ/ENTER.
Note: Calibration points can also be deleted in Edit mode.
3. Press ZERO/SCROLL to select the point to delete (e.g., S0 or S1 or S2). Press
READ/ENTER.
4. The left digit will flash. Press
ZERO/SCROLL until “dEL” ap pea rs. (“dE L” wi ll
appear after the numeral 9.)
Page 73
2—25
User-Entered Calibration, continued
5. Press READ/ENTER to delete. Repeat for all points to be deleted.
Note: The minimum number of valid points is two. For example, if five points have been
entered, three can be deleted using this feature.
6. Press MENU to return to the measurement mode.
Retrieving the Factory Calibration
1. Select the range to restore factory default calibration. See Switching Ranges
on page 2—4.
2. Press and hold the
MENU key until “USER”, then “CAL” appears.
3. Press the
ZERO/SCROLL key to find dFL.
4. Press the
READ/ENTER key to select dFL and restore the instrument to the
factory default calibration.
Note: For meters with factory-calibrated ranges or methods, Standard Calibration Adjust
(SCA) will be disabled when a user-entered method is programmed into the meter. To turn SCA back on, restore the meter to factory default calibration.
Page 74
2—26
User-Entered Calibration, continued Maximum/Minimum Displayed Value
In meters with absorbance (Abs) ranges, the maximum displayed value and minimum displayed value is rel ated to the value of the standa rds entered in a user calibration.
Measurements that exceed the minimum or maximum standards entered in the user calibration will return a flashing number indicating “underrange” or “overrange”. See Error Codes (page 2—12) for more information.
Example 1
For a calibration with the following stan da rd s: S0=0.000
S1=1.000
Maximum Displayed Value 1.000 Minimum Displayed Value 0.000
Page 75
2—27
User-Entered Calibration, continued
Example 2
For a calibration with the following standa rds: S0=1.00
S1=2.00 S2=4.00
For Hach-calibrated programs, the maxi mu m and min im u m di sp lay e d va lu es always equal the factory-calibrated values and cannot be changed .
Maximum Displayed Value 4.00 Minimum Displayed Value 1.00
Page 76
2—28
Certification
Hach Company certifies this instrument was tested thoroughly, inspected, and found to meet its published specifications when it was shipped from the factory.
The Pocket Colorimeter™ II instrument has been tested and is certified as indicated to the following instrumentation standards:
EMC Immunity:
Per 89/ 336/ EEC EMC: EN 6 1326: 1998 (Electrical Equipment for measurement, control and laboratory use—EMC requirements). Supporting test records by Hach Company, certified compliance by Hach Company.
Standard(s) include:
IEC 1000-4-2: 1995 (EN 61000-4-2: 1995) Electro-Static Discharge Immunity (Criteria B)
IEC 1000- 4- 3: 1995 (EN 6 1000- 4- 3: 1996) Radi ated RF Elect ro- Magnet ic Field Immunity (Criteria A)
Additional Immunity Standard(s) include:
ENV 50204: 1996 Radiated Elect rom ag net ic Field from Digital Telephones
Page 77
2—29
Certification, continued
(Criteria A) Radio Frequency Emissions:
Per 89/ 336/ EEC EMC: EN 6 1326: 1998 (Electrical Equipment for measurement, control and laboratory use—EMC requirements) “Class B” emission limits. Supporting test records from Hach EMC Test Facility , certified compliance by Hach Company.
Additional Radio Frequency Emissions Standard(s) include:
EN 55022 (CISPR 22), Class B emissions limits. Canadian Interference-causing Equipment Regulation, IECS-003, Class A:
Supporting test records from Hach EMC Test Facility , certified compliance by Hach Company.
This Class A digital apparatus meets all requirement s of the Canad ian Interference-causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
FCC Part 15, Class “A” Limits: Supporting test records from Hach EMC Test Facility, certified compliance by Hach Company.
Page 78
2—30
Certification, continued
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and (2) This device must accept any interference received, includin g interference that may cause undesired operation. Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmfu l interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at his own expense. The following techniques of reducing the interference problems are applied easily.
Page 79
2—31
Certification, continued
1. Remove power from the Pocket Colorimeter instrument by removing one of its
batteries to verify that it is or is not the source of the interference.
2. Move the Pocket Colorimeter instrument away from the device receiving the
interference.
3. Reposition the receiving antenna for the device receiving the interference.
4. Try combinations of the above.
Page 80
2—32
Page 81
2—33
At Hach Company , customer service is an important part of every product we make.
With that in mind, we have compiled the following information for your convenience.
GENERAL INFORMATION
Page 82
2—34
Page 83
2—35
How to Order
By Telephone: By Mail:
6:30 a.m. to 5:00 p.m. MST Hach Company Monday through Friday P.O. Box 389 (800) 227-HACH (800-227-4224) Loveland, Colorado 80539-0389 U.S.A.
By FAX: For order information by E-mail:
(970) 669-2932 (Hach Loveland) orders@www.hach.com
Information Required:
Hach account number (if available) Purchase order number
Billing address Catalog number
Shipping address Brief description or model number
Your name and phone number Quantity
Page 84
2—36
How to Order, continued
Technical and Customer Service (USA only)
Hach Technical and Customer Service Department personnel are eager to answer questions about our products and their use and to take your orders. Specialists in analytical methods, they are happy to put their talents to work for you . Call 1-800-227-4224 or E-mail techhelp@hach.com.
International Customers
Hach maintains a worldwide network of dealers and distributors. To locate the representative nearest you, send E-mail to intl@hach. com or call (97 0) 669-3050.
In Canada
Hach Instrument Service Centre, Winnipeg, Manitoba, Canada Telephone: (204) 632-5598; (800) 665-7635 FAX: (204) 694-5134
Page 85
2—37
Repair Service
Authorization must be obtained from Hach Comp an y before sending any items for repair. Please contact the Hach Service Center serving your location.
In the United States: Canada:
Hach Company Hach Sales & Service Canada Ltd. 100 Dayton Avenue 1313 Border Street, Unit 34 Ames, Iowa 50010 Winnipeg, Manitoba R3H 0X4 (800) 227-4224 (USA only) (800) 665-7635 (Canada only) FAX: (515) 232-3835 Telephone: (204) 632-5598
FAX: (204) 694-5134 E-mail: canada@hach.com
Latin America, Caribbean, Africa, Europe, the Middle East, Far East, Indian Subcontinent: or Mediterranean Africa:
Hach Company World Headquarters HACH Company, c/o P.O. Box 389 Dr. Bruno Lange GmbH & CO. KG Loveland, Colorado 80539-0389 U.S.A. Willstätterstr. 11 Telephone: (970) 669-3050 40549 Düsseldorf, Germany FAX: (970) 669-2932 Telephone: +49/(0)211/52 88-0 E-mail: intl@hach. com. FAX: +49/(0)211/52 88-134
Page 86
2—38
Warranty
Hach Company warrants this product to the original purchaser against any defects that are due to faulty material or wo rkmanship for a period of two years from date of shipment.
In the event that a defect is discovered during the warranty period, Hach Company agrees that, at its option, it will repair or replace the defective product or refund the purchase price, excluding original shipping and handling char ges. Any product repaired or replaced under this warranty will be warranted only for the remainder of the original product warranty period.
This warranty does not apply to consumable products such as chemical reagents; or consumable components of a prod uct, such as, but not limited to, lamps and tubing.
Contact Hach Company or your distribu tor to initi ate warranty sup port. Product s may not be returned withou t aut horization from Hach Company.
Page 87
2—39
Warranty, continued
Limitations
This warranty does not cover:
damage caused by acts of God, natu ral disaster, labor unrest, acts of war
(declared or undeclared), terrorism, civil strife or acts of any governmental jurisdiction
damage caused by misuse, neglect, acciden t or improper application
or installation
damage caused by any repair or attemp ted repair not authorized by
Hach Company
any product not used in accordance with the instructions furnished by
Hach Company
freight charges to return merchandise to Hach Company
freight charges on expedited or express shipment of warranted parts
or product
travel fees associated with on-site warranty repair
Page 88
2—40
Warranty, continued
This warranty contains the sole express wa rran ty mad e by Hach Comp an y in connection with its products. All implied warranties, including without limitation, the warranties of merchantability and fitness for a particular purpose, are expressly disclaimed.
Some states within the United States do not allow the disclaimer of implied warranties and if this is true in your state the above limitation may not apply to you. This warranty gives you specific rights, and you may also have other rights that vary from state to state.
This warranty constitutes the final, complete, and exclusive statement of warranty terms and no person is authorized to make any other warranties or representations on behalf of Hach Company.
Limitation of Remedies
The remedies of repair , replacement or refu nd of purchase price as stated ab ove are the exclusive remedies for the breach of this warranty. On the basis of strict liability or under any other legal theory, in no event shall Hach Company be liable for any incidental or consequential damages of any kind for breach of warranty or negligence.
Loading...