Hach 3650 ATEX User Manual

Operator Quick Guide
ORBISPHERE 3650 ATEX
Revision D - 18/03/2008

Operating Information

About this Guide

The information in this guide has been carefully checked and is believed to be accurate. However, Hach Ultra assumes no responsibility for any inaccuracies that may be contained in this guide. In no event will Hach Ultra be liable for direct, indirect, special, incidental, or consequential damages resulting from any defect or omission in this guide, even if advised of the possibility of such damages. In the interest of continued product development, Hach Ultra reserves the right to make improvements in this guide and the products it describes at any time, without notice or obligation.

Before Taking Measurements

IMPORTANT: Before taking any measurements,
please ensure that the conditions of use correspond to the certification of the instrument and the installation was performed as described in the Operator’s Manual.

Instrument Controls

The front panel of the instrument has a three-digit liquid crystal display (LCD).
The LCD includes a right-side marker to distinguish between gas concentration and temperature display. This marker also indicates the measurement display units (ppm, ppb, or %) depending on the instrument model.
To the LCD's right is a label showing the measurement units configured at the factory for your application.
In addition to the controls indicated on the front panel, there is also a pressure relief valve switch on top of the instrument to enable atmospheric pressure equilibrium for sensor calibration, or for measurements in gaseous samples in % units.
The button in the bottom right corner backlights the LCD for approximately three minutes. The other push-button controls are:
POWER turns instrument power on or off. The instrument starts in measurement mode
MEAS places the instrument in measurement mode
CAL calibrates the analyzer against a reference sample. This button can be locked out from the WinLog97 PC program
STO stores a measurement value into memory
The × Ø keys toggle between gas concentration and temperature measurement displays in measurement mode, increases or decreases the storage number during storage or memory view, or sets a calibration value during calibration
POWER + CAL sensor calibration
POWER + STO starts automatic data acquisition.
POWER + × starts memory storage view
Page 2

Startup

To start the instrument, press the keyboard
POWER switch. When you turn power on, the
instrument displays its model number briefly, and then starts in measurement mode.
The instrument can be operated independently, making measurements as a portable analyzer. You may store these measurement values for later analysis via the WinLog97 program (on your personal computer) or the memory view mode (on the instrument).

Measurement

Taking Measurements

Connect the top-mounted inlet to accept your sample, typically this is accomplished by connection to a sampling valve. The sample flow can be regulated by the knurled stainless steel knob on top of the flow chamber.
The LCD includes a right-side marker to distinguish between gas concentration measurements and temperature. To switch between gas measurement and temperature measurement, press the × Ø buttons. To backlight the LCD for approximately three minutes, press the Backlight button.
The analyzer will store up to 500 gas measurement values, labeled by numbers 0 through 499, along with the current date and time of each measurement. You have the choice of acquiring this information manually or automatically, as described below.

Automatic Data Acquisition

Before starting automatic measurement storage, first select the sampling rate desired using the Sampling Rate menu of the WinLog97 program.
Gas concentration measurements are displayed for about two minutes. After two minutes the instrument displays the sample number (starting at 000), then the gas concentration measurement value followed by [---] to indicate the measurement is being stored.
This storage sequence repeats automatically, at the rate specified.
To end automatic storage, switch the instrument
OFF (by pressing the POWER key) while it is in
normal measurement mode. Switching ON again without holding down the STO button returns the instrument to measurement mode.

Manual Data Acquisition

Switch the instrument OFF (by pressing the
POWER key). Then hold down the STO button
while switching the instrument back ON. The LCD displays the message Sto for about one second.
For the first measurement you wish to store, press the STO button once to display a sample number. The default sample number is 000 (for first time access), or the last used memory position where data was stored, incremented by a value of 1.
You can increase or decrease this number by pressing the × Ø buttons within three seconds.
Press STO a second time, within five seconds of the first. The instrument then displays the gas concentration measurement value for about three seconds, followed by a clearing message as the value is stored.
Repeat the above steps to store additional measurements.
Page 3

Measurement (cont)

Viewing Stored Measurements

Switch the instrument OFF (by pressing the
POWER key). Hold down the × button while
switching the instrument back ON. The LCD displays a sample location number.
Scroll through the numbered sample locations of all the stored values using the × Ø buttons.
To view the actual gas concentration measurement value at a particular sample number, press the STO button. The LCD now displays the stored value for that sample number.
Press STO a second time to return to the next numbered location display, to continue scrolling or view another stored value.
To return to the measurement mode, switch the instrument OFF and then back ON again without holding down any additional buttons.

Downloading Data

If you have made measurements and stored them in the instrument, you should be ready to bring them into the WinLog97 program for viewing, copying, saving and printing. To download stored data from the instrument to the PC, choose the
DownLoad data command from the Logger menu.
The window displays the stored measurements from the instrument showing the sample number, gas concentration, a date and time stamp, and a sample description.

Sample Description

For help in identifying the locations of various sampling points that are stored by the analyzer, you may choose the Sampling Point Description command from the Logger menu.
The measurement values to be placed in positions 0 through 499 can be described however you wish. Double-click on a particular position (or click Modify), then type a description in the box.

Copy Data

To copy the results to the Windows Clipboard, so that the data can be pasted into a spreadsheet, word processor or other Windows program that accepts tabular text information, choose the
Clipboard command from the Export menu.

Save Data

To save this list of measurements as a text file, capable of being recalled by the WinLog97 program or imported as a file into other Windows programs, choose the Save As command from the File menu and enter the device and file name.

Print Data

To place this list of measurements into a tabular format and send it to the printer, choose the Print command from the File menu.

Clear Data

To clear all the values stored in the instrument, choose the Clear Data command from the Logger menu. Since this action will clear the storage memory of the instrument, a warning appears which you must confirm. Choose Clear to start the memory clear operation.
Page 4

Measurement (cont)

Timebase Updating Rate Max.

Samples
30 Seconds/Division 5 Secs/Sample 60
1 Minute/Division 5 Secs/Sample 120
10 Minutes/Division 5 Secs/Sample 1,200
30 Minutes/Division 9 Secs/Sample 2,000
1 Hour/Division 18 Secs/Sample 2,000
2.5 Hours/Division 45 Secs/Sample 2,000

Monitor Measurements in Real-Time

You may wish to analyze a particular sampling point via the WinLog97 program's Monitoring menu. To use this Monitoring chart, the instrument must be connected to your PC.
Choose Monitoring from the WinLog97 menu to bring up a chart display.

The chart shows the gas concentration (in blue), temperature (in red), and pressure (in green) as the sample is being measured by the instrument. The chart is updated directly from instrument measurements, at a rate determined by the time scale set in the TIMEBASE box at the lower right corner of the chart.

Click the TIMEBASE up/
the time scale of the divisions of the chart. Each division mark along the baseline (1, 2, ...10) can be made to represent from 30 seconds to 2½ hours, providing from 5 minutes to 25 hours of continuously displayed samples (as illustrated in the table).
Click on the Continuous box to enable or disable continuous charting. When this box is checked, the chart scrolls continuously after reaching the 10 division, and the oldest samples are lost off the left of the chart.
When Continuous is not checked, the chart stops displaying new results after reaching the 10 division, and all subsequent measurements are lost.
Click the up/down pointers for each measurement variable (GAS,
TEMPERATURE and PRESSURE) to change the scaling of that value on the chart. The display of each measurement variable may be turned on or off by choosing the appropriate On or Off switch.
A running display of latest sample Gas, Temperature and Pressure is also shown in the bottom-right corner of the chart.
Use the buttons at the bottom of the chart to control real-time monitoring. Choose Go to clear the chart and start real-time monitoring, Stop to stop monitoring and Copy to copy the data from the chart as text information to the Windows Clipboard. This information can be pasted from the clipboard into any Windows application, such as a spreadsheet or word processor.
Choose Close to close the Monitoring window.
down pointers to change
Page 5

Instrument Configuration

The following commands are all available from the Configuration menu in the WinLog97 software installed on your PC.
The instrument must be connected to your PC in order to change its configuration.

INSTRUMENT - PC CONNECTION

The Serial port menu lets you choose one of four serial communication ports. Usually, COM1 is used to connect to a mouse, so try COM2 first.

SETTING SAMPLING INTERVALS

The analyzer can perform as a standalone data acquisition device, automatically recording gas measurements with the date and time, and storing up to 500 of these values.

MEMBRANE SELECTION

You may find it necessary to use a different type of membrane for different applications.
To re-configure the analyzer, choose Membrane to bring up the box which reveals the membrane models available.

CALIBRATION MODE

Use the Calibration mode command to select how the sensor is to be calibrated.
Choose In Air to allow calibration of the sensor in air, In Line to calibrate the sensor directly in the sampling line, or In a Span Gas to calibrate in a source of gas of known concentration.
Choosing Sampling Rate lets you select time intervals (acquisition rate) for this storage capability.
Page 6

CALIBRATION KEY STATUS

You can use the Calibration Key Status menu to prevent an accidental sensor re-calibration from the instrument keyboard.
Enable or Disable the instrument front panel
CAL button. Disabling the button will prevent
accidental sensor re-calibration.

Instrument Configuration (cont)

CALIBRATION RANGE CHECKING

When calibration is performed for In Air and In a Span Gas calibration modes, the sensor
measurement current is compared to an ideal current for the selected membrane to determine whether or not to complete the calibration. You can use the Calibration Range Checking menu to enable or disable sensor calibration range checking in the instrument.
Choose Disabled to calibrate without checking the value of the measurement current, within a range of 0% to 999% of the ideal current. When set to Enabled, at calibration the measurement current should be between 25% and 175% of the ideal current; if the value is outside of these limits, the calibration fails and Err is displayed on the instrument LCD.

INSTRUMENT CONFIGURATION REVIEW

To review if the instrument is set up as expected, choose the Configuration view command.

SPAN GAS

When calibrating the sensor in a span gas, use the Span Gas menu to enter the span gas concentration (as a %age) when using a span gas as a calibration medium.

DUAL USE (MODEL 3650Ex/113 ONLY)

Use the Dual Use menu to change the measurement phase (either dissolved or gaseous) for the model 3650Ex/113 dual-use analyzer.
Choose ppm (dissolved) to set the instrument for dissolved measurement in liquids, or % (gaseous) to set the instrument to gas phase measurement.
Page 7

Calibration

BAROMETRIC PRESSURE SENSOR

The instrument internal barometric pressure sensor is calibrated at the factory, and normally requires no further attention. However, you may wish to calibrate it against your own instrumentation, or simply check the instrument for accuracy.
Select Troubleshooting, Pressure Calibration from the WinLog97 program, and enter the current atmospheric pressure, in mbars, in the calibration pressure entry box.

GAS SENSOR

The sensor can be calibrated either in air or directly in line in a liquid sample. Additionally, oxygen sensors can be calibrated using a span gas.
To perform a calibration, you must first choose which method you wish to use. The mode of calibration may be selected using the WinLog97 program. Alternatively, you can select the mode of calibration using the instrument front panel buttons.
Switch the instrument power OFF. Switch the instrument on by holding down the CAL button, and then pressing the POWER button at the same time. The instrument display will show either Air for in air calibration, LI for in line calibration, or SPA for span gas calibration
Use the × Ø buttons to change the mode to your choice and press the STO button to set the selected mode.

In a Span Gas Calibration

The span gas calibration procedure may be used if you have a source of gas with a known concentration of O
type of calibration, the instrument must be set for calibration In a Span Gas. The Windows WinLog97 program must also be used in this procedure.
displayed.
When the reading is stable, press CAL again.
If the new calibration current is within 25% to 175% of the ideal current, the instrument displays
CAL and returns to the measurement mode. Calibration is now complete, the sensor is calibrated, and you can proceed with your measurements.
If the new calibration current is not within this range, the instrument displays Err and returns to
measurement mode. The system will not calibrate, and it is likely that a sensor service will be required.
(in% units). To perform this
2
Switch on the instrument, if necessary, and wait a minute or so for the displayed measurement to settle. Then expose the sensor to a span gas sample with a known oxygen concentration. Enter the percentage of oxygen in the span gas using the WinLog97 program.
Press the CAL button. Remember, this button may have been locked out to prevent an accidental reset.
A brief clearing [---] message appears. Press CAL again within a 3 second period. The percentage of the measured current to the ideal one is
Page 8

Calibration (cont)

In Line Calibration

The in line calibration procedure can be used to calibrate the sensor directly in line, against a liquid sample with a known dissolved oxygen concentration.
To perform this type of calibration, the instrument must be set for calibration In line. Switch on the instrument, if necessary, and wait a minute or so for the displayed measurement to settle.
Expose the sensor to a liquid sample with a known gas concentration.

In Air Calibration (oxygen sensors only)

The oxygen sensor can be accurately calibrated in air. To perform this type of calibration, the instrument must be set for calibration In air.
You can perform this procedure with the sensor connected to the instrument, but without the flow chamber mounted.
The membrane must be dry, and prior to calibrating it may be necessary to remove the screw-on protection cap and its optional grill to make sure that the components are not damp.
Switch on the instrument, if necessary, and wait a minute or so for the displayed measurement to settle.
Press the CAL button. Remember, this button may have been locked out to prevent an accidental reset.
A brief clearing [---] message appears. Press
CAL again within a 3 second period. A
measurement will flash on the LCD, showing the oxygen concentration of the calibration sample.
Assuming you know the gas content to be a certain value, you can adjust the displayed value with the × Ø keys.
When the reading is adjusted to the known concentration, press CAL again. The instrument displays CAL and returns to the measurement mode.
The process is then exactly the same as for calibration in a span gas, detailed previously.
Page 9

Troubleshooting

You may wish to use the Troubleshooting menu to make sure that the instrument is configured correctly for your application, and is in good working order.
The instrument must be connected to your PC to perform these tests.

SERIAL LINK TEST

Normally, the instrument will inform you of a disconnected RS-232 (serial) link when appropriate. However, you can confirm a good connection using the Serial Link Test option, and echoing a test message via the instrument.
Enter text characters in the Text to be sent box, then click Send. If the serial link is operating correctly, the exact same text will be displayed back from the instrument in the Echo box.

KEYBOARD TEST

The Keyboard Test option will reveal whether all the instrument buttons are functioning correctly.
Type a number in the Number box (you may also select one of three units positions for the LCD's rightmost indicator bar as well). Then choose Send. The number and indicator bar placement should appear on your instrument LCD.

CLOCK SETTINGS

Choose the Clock settings option to set the date and time in the instrument.

ANALOG VOLTAGES VIEW

The Analog Voltage View option gives a real­time look at voltages used by the system to transmit information about sensor current, temperature and pressure. This is useful when trying to identify an instrument problem with an Orbisphere service representative either on-site or over the phone.
Press any one of the instrument's buttons (except the on/off button) for a full second or more. The appropriate square on-screen should darken.

DISPLAY TEST

Choosing the Display Test option lets you perform a one-way communication between computer and instrument.
Page 10
The voltage limits for normal operation are:
Current channel: -1.5 V to +1.5 V
Temperature channel: +10 mV to +4 V
Pressure channel: -100 mV to +100 mV

MEASUREMENTS VIEW

The Measurements View option confirms, on your PC monitor, what your instrument should be displaying on the LCD for gas concentration and sample temperature.
Page 11
Loading...