Programmable DC Power Supply
PSU Series
QUICK START GUIDE
GW INSTEK PART NO. 82SU-60250MD1
ISO-9001 CERTIFIED MANUFACTURER
![](/html/a3/a3ed/a3edc31593f7908b41046988cb4ebea17e72c4535f92e3a125da53b10df7fd6a/bg2.png)
This manual contains proprietary information, which is protected by
copyright. All rights are reserved. No part of this manual may be
photocopied, reproduced or translated to another language without
prior written consent of Good Will company.
The information in this manual was correct at the time of printing.
However, Good Will continues to improve products and reserves the
rights to change specification, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
![](/html/a3/a3ed/a3edc31593f7908b41046988cb4ebea17e72c4535f92e3a125da53b10df7fd6a/bg3.png)
Table of Contents
1
Table of Contents
SAFETY INSTRUCTIONS ................................................... 2
GETTING STARTED ........................................................... 6
PSU Series Overview .............................. 6
Appearance .......................................... 11
OPERATION .................................................................... 19
Set Up .................................................. 19
Basic Operation ................................... 22
APPENDIX ...................................................................... 31
PSU Dimensions .................................. 31
Declaration of Conformity .................... 32
![](/html/a3/a3ed/a3edc31593f7908b41046988cb4ebea17e72c4535f92e3a125da53b10df7fd6a/bg4.png)
PSU Series Quick Start Guide
2
SAFETY INSTRUCTIONS
This chapter contains important safety
instructions that you must follow during
operation and storage. Read the following before
any operation to insure your safety and to keep
the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the
instrument.
WARNING
Warning: Identifies conditions or practices that
could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that
could result in damage to the PSU or to other
properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
![](/html/a3/a3ed/a3edc31593f7908b41046988cb4ebea17e72c4535f92e3a125da53b10df7fd6a/bg5.png)
Do not dispose electronic equipment as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased.
General
Guideline
CAUTION
Do not place any heavy object on the PSU.
Avoid severe impact or rough handling that
leads to damaging the PSU.
Do not discharge static electricity to the PSU.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening.
Do not disassemble the PSU unless you are
qualified.
(Measurement categories) EN61010-1:2010 and EN61010-2-030
specifies the measurement categories and their requirements as
follows. The PSU falls under category II.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
0 is for measurements performed on circuits not directly
connected to Mains.
AC Input voltage range: 85Vac~265Vac
Frequency: 47Hz to 63Hz
To avoid electrical shock connect the protective
grounding conductor of the AC power cord to
an earth ground.
![](/html/a3/a3ed/a3edc31593f7908b41046988cb4ebea17e72c4535f92e3a125da53b10df7fd6a/bg6.png)
PSU Series Quick Start Guide
4
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: 20%~ 85% (no condensation)
Altitude: < 2000m
Temperature: 0°C to 50°C
(Pollution Degree) EN61010-1:2010 and EN61010-2-030 specifies
the pollution degrees and their requirements as follows. The PSU
falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric
strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused
by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to
condensation which is expected. In such conditions, equipment
is normally protected against exposure to direct sunlight,
precipitation, and full wind pressure, but neither temperature
nor humidity is controlled.
Location: Indoor
Temperature: -25°C to 70°C
Relative Humidity: ≤90%(no condensation)
Do not dispose this instrument as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased. Please make sure
discarded electrical waste is properly recycled to
reduce environmental impact.
![](/html/a3/a3ed/a3edc31593f7908b41046988cb4ebea17e72c4535f92e3a125da53b10df7fd6a/bg7.png)
SAFETY INSTRUCTIONS
5
Power cord for the United Kingdom
When using the power supply in the United Kingdom, make sure
the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
As the colours of the wires in main leads may not correspond with
the coloured marking identified in your plug/appliance, proceed
as follows:
The wire which is coloured Green & Yellow must be connected to
the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal
which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the
terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment
or contact the supplier.
This cable/appliance should be protected by a suitably rated and
approved HBC mains fuse: refer to the rating information on the
equipment and/or user instructions for details. As a guide, a cable
of 0.75mm2 should be protected by a 3A or 5A fuse. Larger
conductors would normally require 13A types, depending on the
connection method used.
Any exposed wiring from a cable, plug or connection that is
engaged in a live socket is extremely hazardous. If a cable or plug is
deemed hazardous, turn off the mains power and remove the cable,
any fuses and fuse assemblies. All hazardous wiring must be
immediately destroyed and replaced in accordance to the above
standard.
![](/html/a3/a3ed/a3edc31593f7908b41046988cb4ebea17e72c4535f92e3a125da53b10df7fd6a/bg8.png)
PSU Series Quick Start Guide
6
GETTING STARTED
This chapter describes the power supply in a
nutshell, including its main features and front /
rear panel introduction. After going through the
overview, please read the theory of operation to
become familiar with the operating modes,
protection modes and other safety considerations.
PSU Series Overview
Series lineup
The PSU series consists of 5 models, covering a number of different
current, voltage and power capacities:
1
Minimum voltage guaranteed to 0.2% of rating voltage.
2
Minimum current guaranteed to 0.4% of rating current.
![](/html/a3/a3ed/a3edc31593f7908b41046988cb4ebea17e72c4535f92e3a125da53b10df7fd6a/bg9.png)
GETTING STARTED
7
Main Features
High power density: 1500W in 1U
Universal input voltage 85~265Vac, continuous
operation.
Output voltage up to 60V, current up to 200A.
Active power factor correction.
Parallel master/slave operation with active
current sharing.
Remote sensing to compensate for voltage drop
in load leads.
19" rack mounted ATE applications.
A built-in Web server that lets you monitor the
instrument directly from an internet browser on
your computer.
OVP, OCP and OHP protection.
Preset memory function.
Adjustable voltage and current slew rates.
Bleeder circuit ON/OFF setting. (to prevent
over-discharging of batteries)
CV, CC priority start function. (prevents
overshoot with output ON)
Supports test scripts.
Built-in RS-232/485, LAN and USB interface.
Analog output programming and monitoring.
Optional interfaces: GPIB, Isolated Voltage (0-
5V/0-10V) and Isolated Current (4-20mA)
programming and monitoring interface.
(Factory options)
![](/html/a3/a3ed/a3edc31593f7908b41046988cb4ebea17e72c4535f92e3a125da53b10df7fd6a/bga.png)
PSU Series Quick Start Guide
8
Accessories
Before using the PSU power supply unit, check the package
contents to make sure all the standard accessories are included.
Qty.
Output terminal cover
Analog connector plug kit
Output terminal M8 bolt set
User manual, Programming
Manual
* CTC GW/INSTEK JAPAN
USE ,RoHS
Factory Installed
Options
Voltage programming isolated
analog interface
Current programming isolated
analog interface
![](/html/a3/a3ed/a3edc31593f7908b41046988cb4ebea17e72c4535f92e3a125da53b10df7fd6a/bgb.png)
Cable for 2 units of PSU-Series in
parallel mode connection
Bus Bar for 2 units of PSU-Series in
parallel mode connection
Joins a vertical stack of 2 PSU units
together. 2U-sized handles x2, joining
plates x2.
Cable for 3 units of PSU-Series in
parallel mode connection
Bus Bar for 3 units of PSU-Series in
parallel mode connection
Joins a vertical stack of 3 PSU units
together. 3U-sized handles x2, joining
plates x2.
Cable for 4 units of PSU-Series in
parallel mode connection
Bus Bar for 4 units of PSU-Series in
parallel mode connection
Joins a vertical stack of 4 PSU units
together. 4U-sized handles x2, joining
plates x2.
RS232 cable with DB9 connector kit
RS485 cable with DB9 connector kit
Rack-mount slides (General Devices
P/N: C-300-S-116-RH-LH)
USB Cable 2.0-A-B Type, Approx. 1.2M
Power Cord SJT 12AWG/3C, 3m MAX
Length, 105 ºC, RNB5-5*3P UL/CSA
type
Power Cord H05W-F 1.5mm2/3C, 3m
MAX Length, 105 ºC, RNB5-5*3P VDE
type