![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg1.png)
Programmable DC Power Supply
ISO-9001 CERTIFIED MANUFACTURER
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg2.png)
This manual contains proprietary information, which is protected by
copyright. All rights are reserved. No part of this manual may be
photocopied, reproduced or translated to another language without
prior written consent of Good Will company.
The information in this manual was correct at the time of printing.
However, Good Will continues to improve products and reserves the
rights to change specification, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg3.png)
Table of Contents
3
Table of Contents
SAFETY INSTRUCTIONS .................................................. 4
GETTING STARTED .......................................................... 8
PSU Series Overview .............................. 9
Appearance .......................................... 14
Configuration Settings ......................... 22
REMOTE CONTROL ....................................................... 32
Interface Configuration ........................ 34
Command Syntax ................................. 63
Command List ..................................... 67
Status Register Overview ................... 127
Error List ........................................... 140
APPENDIX .................................................................... 149
PSU Default Settings ......................... 149
Error Messages & Messages .............. 151
LED ASCII Table Character Set ........... 153
INDEX .......................................................................... 154
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg4.png)
PSU Programming Manual
4
SAFETY INSTRUCTIONS
This chapter contains important safety
instructions that you must follow during
operation and storage. Read the following before
any operation to insure your safety and to keep
the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the
instrument.
Warning: Identifies conditions or practices that
could result in injury or loss of life.
Caution: Identifies conditions or practices that
could result in damage to the PSU or to other
properties.
Attention Refer to the Manual
Protective Conductor Terminal
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg5.png)
Do not dispose electronic equipment as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased.
General
Guideline
CAUTION
Do not place any heavy object on the PSU.
Avoid severe impact or rough handling that
leads to damaging the PSU.
Do not discharge static electricity to the PSU.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening.
Do not disassemble the PSU unless you are
qualified.
(Measurement categories) EN61010-1:2010 and EN61010-2-030
specifies the measurement categories and their requirements as
follows. The PSU falls under category II.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
0 is for measurements performed on circuits not directly
connected to Mains.
AC Input voltage range: 85Vac~265Vac
Frequency: 47Hz to 63Hz
To avoid electrical shock connect the protective
grounding conductor of the AC power cord to
an earth ground.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg6.png)
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: 20%~ 85% (no condensation)
Altitude: < 2000m
Temperature: 0°C to 50°C
(Pollution Degree) EN61010-1:2010 and EN61010-2-030 specifies
the pollution degrees and their requirements as follows. The PSU
falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric
strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused
by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to
condensation which is expected. In such conditions, equipment
is normally protected against exposure to direct sunlight,
precipitation, and full wind pressure, but neither temperature
nor humidity is controlled.
Location: Indoor
Temperature: -25°C to 70°C
Relative Humidity: ≤90%(no condensation)
Do not dispose this instrument as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased. Please make sure
discarded electrical waste is properly recycled to
reduce environmental impact.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg7.png)
SAFETY INSTRUCTIONS
7
Power cord for the United Kingdom
When using the power supply in the United Kingdom, make sure
the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
As the colours of the wires in main leads may not correspond with
the coloured marking identified in your plug/appliance, proceed
as follows:
The wire which is coloured Green & Yellow must be connected to
the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal
which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the
terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment
or contact the supplier.
This cable/appliance should be protected by a suitably rated and
approved HBC mains fuse: refer to the rating information on the
equipment and/or user instructions for details. As a guide, a cable
of 0.75mm2 should be protected by a 3A or 5A fuse. Larger
conductors would normally require 13A types, depending on the
connection method used.
Any exposed wiring from a cable, plug or connection that is
engaged in a live socket is extremely hazardous. If a cable or plug is
deemed hazardous, turn off the mains power and remove the cable,
any fuses and fuse assemblies. All hazardous wiring must be
immediately destroyed and replaced in accordance to the above
standard.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg8.png)
PSU Programming Manual
8
GETTING STARTED
This chapter describes the power supply in a
nutshell, including its main features and front /
rear panel introduction. After going through the
overview, please read the theory of operation to
become familiar with the operating modes,
protection modes and other safety considerations.
PSU Series Overview .................................................. 9
Series lineup ........................................................................................ 9
Main Features ................................................................................... 10
Accessories ........................................................................................ 11
Appearance .............................................................. 14
PSU Series Front Panel ................................................................... 14
PSU Series Display and Operation Panel .................................... 17
Rear Panel .......................................................................................... 19
Configuration Settings ............................................. 22
Setting Normal Function Settings ................................................. 22
Setting Power On Configuration Settings ................................... 24
Configuration Table ......................................................................... 26
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg9.png)
GETTING STARTED
9
PSU Series Overview
Series lineup
The PSU series consists of 10 models, covering a number of
different current, voltage and power capacities:
1
Minimum voltage guaranteed to 0.2% of rating voltage.
2
Minimum current guaranteed to 0.4% of rating current.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bga.png)
PSU Programming Manual
10
Main Features
High power density: 1500W in 1U
Universal input voltage 85~265Vac, continuous
operation.
Output voltage up to 600V, current up to 200A.
Active power factor correction.
Parallel master/slave operation with active
current sharing.
Remote sensing to compensate for voltage drop
in load leads.
19" rack mounted ATE applications.
A built-in Web server.
OVP, OCP and OHP protection.
Preset memory function.
Adjustable voltage and current slew rates.
Bleeder circuit ON/OFF setting.
CV, CC priority start function. (prevents
overshoot with output ON)
Supports test scripts.
Built-in RS-232/485, LAN and USB interface.
Analog output programming and monitoring.
Optional interfaces: GPIB, Isolated Voltage (0-
5V/0-10V) and Isolated Current (4-20mA)
programming and monitoring interface.
(Factory options)
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bgb.png)
GETTING STARTED
11
Accessories
Before using the PSU power supply unit, check the package
contents to make sure all the standard accessories are included.
Qty.
Output terminal cover
Analog connector plug kit
Output terminal M8 bolt set
(6V~60V model)
User manual, Programming
manual
* CTC GW/INSTEK JAPAN
USE ,RoHS
Factory Installed
Options
Voltage programming isolated
analog interface
Current programming isolated
analog interface
Front Panel Filter Kit (Operation
Temperature is guaranteed to
40˚C)
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bgc.png)
PSU Programming Manual
12
Cable for 2 units of PSU-Series in
parallel mode connection
Bus Bar for 2 units of PSU-Series in
parallel mode connection
Joins a vertical stack of 2 PSU units
together. 2U-sized handles x2, joining
plates x2.
Cable for 3 units of PSU-Series in
parallel mode connection
Bus Bar for 3 units of PSU-Series in
parallel mode connection
Joins a vertical stack of 3 PSU units
together. 3U-sized handles x2, joining
plates x2.
Cable for 4 units of PSU-Series in
parallel mode connection
Bus Bar for 4 units of PSU-Series in
parallel mode connection
Joins a vertical stack of 4 PSU units
together. 4U-sized handles x2, joining
plates x2.
RS232 cable with DB9 connector kit.
It Includes RS232 cable with DB9
connector, RS485 used master cable
(gray plug), slave cable (black plug)
and end plug terminal.
RS485 cable with DB9 connector kit.
It Includes RS485 cable with DB9
connector, RS485 used master cable
(gray plug), slave cable (black plug)
and end plug terminal.
Rack-mount slides (General Devices
P/N: C-300-S-116-RH-LH)
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bgd.png)
USB Cable 2.0-A-B Type, Approx. 1.2M
Power Cord SJT 12AWG/3C, 3m MAX
Length, 105 ºC, RNB5-5*3P UL/CSA
type
Power Cord H05W-F 1.5mm2/3C, 3m
MAX Length, 105 ºC, RNB5-5*3P VDE
type
Power Cord VCTF 3.5mm2/3C, 3m
MAX Length, 105 ºC, RNB5-5*3P PSE
type
Certificate of traceable calibration
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bge.png)
PSU Programming Manual
14
Appearance
PSU Series Front Panel
Lock/Local PROT Function Test Set Output
Unlock ALM_CLR M 1 M 2 M 3
Shift
: Long Push
VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUN
C C
A
C V
V
VOLTAGE CURRENT
PSU 40- 38
DC Power Supply
0 -40V / 0 -38A
Voltage Current
2 3 4 5
Display Area
1
116 7 8 9 10 12 13
Used to turn the power on/off.
USB A port for data transfer,
loading test scripts etc.
Air inlet for cooling the inside of
the PSU series.
Used to set the voltage value or
select a parameter number in the
Function settings.
The display area shows setting values, output
values and parameter settings. The function LEDs
below show the current status and mode of the
power supply. See page 17 for details.
Displays the current or the value
of a Function parameter.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bgf.png)
Used to lock all front panel
buttons other than the Output
Button or it switches to local
mode.
(Long push) Used to unlock the
front panel buttons.
Used to set and display OVP, OCP
and UVL.
(Long push) Used to release
protection functions that have
been activated.
Used to configure the various
function.
(+Shift) Used to recall the M1
setup.
(+Shift and hold) Used to save the
current setup to M1.
Used to run customized scripts for
testing.
(+Shift) Used to recall the M2
setup.
(+Shift and hold) Used to save the
current setup to M2.
Used to set and confirm the output
voltage and output current.
(+Shift) Used to recall the M3
setup.
(+Shift and hold) Used to save the
current setup to M3.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg10.png)
PSU Programming Manual
16
Used to enable the functions that
are written in blue characters
below the button.
Used to turn output on and off.
Lights in green during output ON.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg11.png)
GETTING STARTED
17
PSU Series Display and Operation Panel
VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUN
C C
A
C V
V
VOLTAGE CURRENT
Display Area
14 1516 17
18 19 20 21 22 23 24
25
26 27 28
Displays the voltage or the parameter number of a
Function parameter.
Displays the current or the value of a Function
parameter.
Lights in green during constant voltage mode.
Lights in green during constant current mode.
The voltage slew rate enable.
Lights up when the LAN interface is connected.
Lights in green during remote control.
Lights in red when an error has occurred.
The output on/off delay enable.
Lights in red when a protection function has been
activated.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg12.png)
PSU Programming Manual
18
The current slew rate enable.
Lights in green when the memory value are being
recalled or saved.
Lights in green when the memory value are being
recalled or saved.
Lights in green when the memory value are being
recalled or saved.
Auto sequence has been activated.
Only the ERR and ALM LED’s are red. All the
others are green.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg13.png)
GETTING STARTED
19
Rear Panel
S LS SLSNC
REMOTE SENSE
ANALOG PROGRAMMING
RS485 / 232
OUT IN
LAN
DC OUTPUT
AC INPUT
L N
100 240V
2000VA MAX.
AC
47 63Hz
234 15678
9 10
87654321
0 -5V / 0 - 10V
ISOLATED PROGRAMMING
87654321
4 -20mA
ISOLATED PROGRAMMING
6-60V models:
PSU 6-200, 12.5-120,
20-76, 40-38, 60-25
100-600V models:
PSU 100-15, 150-10,
300-5, 400-3.8, 600-2.6
S LS SLSNC
REMOTE SENSE
ANALOG PROGRAMMING
OUT IN
RS485 / 232
LAN
L N
AC INPUT
100 240V
2000VA MAX.
AC
47 63Hz
Output terminals for 6V to 60V
models.
Output terminals for 100V to 600V
models.
USB port for controlling the PSU
remotely.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg14.png)
PSU Programming Manual
20
Ethernet port for controlling the
PSU remotely.
Two different types of cables can
be used for RS232 or RS485-based
remote control.
PSU-232: RS232 cable with DB9
connector kit.
PSU-485: RS485 cable with DB9
connector kit.
RJ-45 connector that is used to
daisy chain power supplies with
the Remote-IN port to form a
communication bus.
PSU-485S: Serial link cable with
RJ-45 shielded connector.
External analog control connector.
Compensation of load wire drop.
87654321
0 - 5V / 0 - 10V
ISOLATED PROGRAMMING
Blank sub-plate for standard units.
Isolated Analog connector for
units equipped with Isolated
Current and Voltage
Programming and Monitoring
option.
GPIB connector for units equipped
with IEEE programming option.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg15.png)
Connector for grounding the output (two
positions, shown in red).
L N
AC INPUT
100 240V
2000VA MAX.
AC
47 63Hz
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg16.png)
PSU Programming Manual
22
Configuration Settings
Setting Normal Function Settings
The normal function settings, F-01~F-61, F70~F-78, F-88~F-89 and F100~F122 can be easily
configured with the Function key.
Ensure the load is not connected.
Ensure the output is off.
Function settings F-90~97 can only be
viewed.
Function setting F-89 (Show Version) can only be
viewed, not edited.
Configuration settings F-90~ F-97 cannot be edited
in the Normal Function Settings. Use the Power
On Configuration Settings. See page 24 for details.
1. Press the Function key. The
function key will light up.
2. The display will show F-01 on the left and the
configuration setting for F-01 on the right.
VOLTAGE CURRENT
VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUN
C V C C
V A
3. Rotate the voltage knob to change
the F setting.
F-00~F-61, F-70~F-78,
F-88~F-97, F100~F122
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg17.png)
4. Use the current knob to set the
parameter for the chosen F setting.
Press the Voltage knob to save the
configuration setting. ConF will be
displayed when it is configuring.
VOLTAGE CURRENT
VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUN
C V C C
V A
Press the Function key again to exit
the configuration settings. The
function key light will turn off.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg18.png)
PSU Programming Manual
24
Setting Power On Configuration Settings
The Power On configuration settings can only
be changed during power up to prevent the
configuration settings being inadvertently
changed.
Ensure the load is not connected.
Ensure the power supply is off.
1. Hold the Function key whilst turning the
power on.
Lock/Local PROT Function Test Set
Output
Unlock ALM_CLR M 1 M 2 M 3
Shift
: Long Push
VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUN
C C
A
C V
V
VOLTAGE CURRENT
PSU 40-38
DC Power Supply
0 - 40V / 0 - 38A
Voltage Current
2. The display will show F-90 on the left and the
configuration setting for F-90 on the right.
VOLTAGE CURRENT
VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUN
C V C C
V A
3. Rotate the voltage knob to change
the F setting.
4. Use the current knob to set the
parameter for the chosen F setting.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg19.png)
Press the Voltage knob to save the
configuration setting. ConF will be
displayed when it is configuring.
VOLTAGE CURRENT
VSR LAN RMT ERR DLY ALM ISR M 1 M 2 M 3 RUN
C V C C
V A
Cycle the power to save and exit the
configuration settings.
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg1a.png)
PSU Programming Manual
26
Configuration Table
Please use the configuration settings listed below when applying
the configuration settings.
V-I mode slew rate select
0 = CV high speed priority (CVHS)
1 = CC high speed priority (CCHS)
2 = CV slew rate priority (CVLS)
3 = CC slew rate priority (CVLS)
0.001~0.06V/msec (PSU 6-200)
0.001~0.125V/msec (PSU 12.5-120)
0.001~0.2V/msec (PSU 20-76)
0.001~0.4V/msec (PSU 40-38)
0.001~0.6V/msec (PSU 60-25)
0.001~1.000V/msec (PSU 100-15)
0.001~1.500V/msec (PSU 150-10)
0.001~1.500V/msec (PSU 300-5)
0.001~2.000V/msec (PSU 400-3.8)
0.001~2.400V/msec (PSU 600-2.6)
Falling voltage slew rate
0.001~0.06V/msec (PSU 6-200)
0.001~0.125V/msec (PSU 12.5-120)
0.001~0.2V/msec (PSU 20-76)
0.001~0.4V/msec (PSU 40-38)
0.001~0.6V/msec (PSU 60-25)
0.001~1.000V/msec (PSU 100-15)
0.001~1.500V/msec (PSU 150-10)
0.001~1.500V/msec (PSU 300-5)
0.001~2.000V/msec (PSU 400-3.8)
0.001~2.400V/msec (PSU 600-2.6)
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg1b.png)
0.001~2A/msec (PSU 6-200)
0.001~1.2A/msec (PSU 12.5-120)
0.001~0.76A/msec (PSU 20-76)
0.001~0.38A/msec (PSU 40-38)
0.001~0.25A/msec (PSU 60-25)
0.001~0.150A/msec (PSU 100-15)
0.001~0.100A/msec (PSU 150-10)
0.001~0.025A/msec (PSU 300-5)
0.001~0.008A/msec (PSU 400-3.8)
0.001~0.006A/msec (PSU 600-2.6)
Falling current slew rate
0.001~2A/msec (PSU 6-200)
0.001~1.2A/msec (PSU 12.5-120)
0.001~0.76A/msec (PSU 20-76)
0.001~0.38A/msec (PSU 40-38)
0.001~0.25A/msec (PSU 60-25)
0.001~0.150A/msec (PSU 100-15)
0.001~0.100A/msec (PSU 150-10)
0.001~0.025A/msec (PSU 300-5)
0.001~0.008A/msec (PSU 400-3.8)
0.001~0.006A/msec (PSU 600-2.6)
Internal resistance
setting
0~0.03Ω (PSU 6-200)
0~0.104Ω (PSU 12.5-120)
0~0.263Ω (PSU 20-76)
0~1.053Ω (PSU 40-38)
0~2.4Ω (PSU 60-25)
0~6.667Ω (PSU 100-15)
0~15.00Ω (PSU 150-10)
0~60.00Ω (PSU 300-5)
0~105.3Ω (PSU 400-3.8)
0~230.8Ω (PSU 600-2.6)
0 = OFF, 1 = ON, 2 = AUTO
Current Setting Limit
(I-Limit)
Voltage Setting Limit
(V-Limit)
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg1c.png)
PSU Programming Manual
28
Display memory
parameter when recalling
(M1, M2, M3)
Auto Calibration Parallel
Control
0 = Disable, 1 = Enable, 2 = Execute
Parallel Calibration and set to Enable.
Note: Must be a short between each
unit before starting.
Measurement Average
Setting
0 = Low, 1 = Middle, 2 = High
Alarm Recovery and
Output Status
0 = Safe Mode, 1 = Force Mode
0:Lock Panel, Allow Output OFF
1:Lock Panel, Allow Output ON/OFF
Show front panel USB
status
0 = None, 1 = Mass Storage
Show rear panel USB
status
0 = None, 1 = Linking to PC
0 = Disable USB, 1 = Full Speed, 2 =
Auto Detect Speed
0 = Disable GPIB, 1 = Enable GPIB
Show GPIB available
status
0 = No GPIB, 1 = GPIB is available
0 = GW Instek, 1 = TDK GEN, 2 =
Agilent 5700, 3 = Kikusui PWX
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg1d.png)
Socket Server
Enable/Disable
0 = Disable, 1 = Enable
Show Socket Server Port
Web Server
Enable/Disable
Web Password
Enable/Disable
0 = Disable, 1 = Enable
Web Enter Password
0 = Disable UART, 1 = RS232,
2 = RS485
0 = 1200, 1 = 2400, 2 = 4800,
3 = 9600, 4 = 19200, 5 = 38400,
6 = 57600, 7 = 115200
0 = None, 1 = Odd, 2 = Even
0 = SCPI, 1 = TDK (emulation mode)
UART Address
(For multi-unit remote
control)
00 ~ 30
UART Multi-Drop control
0 = Disable, 1 = Master, 2 = Slave,
3 = Display information
Displayed parameter: AA-S
AA: 00~30 (Address),
S: 0~1 (Off-line/On-line status).
0 = None
1 = Return to factory default settings
![](/html/f7/f76d/f76d39a114a453754a105f7e19519ee711bc67f6d93db6684d87600e8522da90/bg1e.png)
PSU Programming Manual
30
0, 1 = Version
2, 3, 4, 5 = Build date (YYYYMMDD)
6, 7 = Keyboard CPLD
8, 9 = Analog Board CPLD
A, B = Analog Board FPGA
C, D, E, F = Kernel Build
(YYYYMMDD)
G, H = Test Command Version
I, J, K, L = Test Command Build
(YYYYMMDD)
M, N = Reserved
O, P = Option Module
Power On Configuration Settings*
0 = Control by Local
1 = Control by External Voltage
2 = Control by External Resistor -
Rising
3 = Control by External Resistor -
Falling
4 = Control by Isolated Board
0 = Control by Local
1 = Control by External Voltage
2 = Control by External Resistor -
Rising
3 = Control by External Resistor -
Falling
4 = Control by Isolated Board
Output Status when
Power ON
0 = Safe Mode (Always OFF),
1 = Force Mode (Always ON),
2 = Auto Mode (Status before last
time power OFF)
Master/Slave
Configuration
0 = Independent
1 = Master with 1 slave in parallel
2 = Master with 2 slaves in parallel
3 = Master with 3 slaves in parallel
4 = Slave (parallel)
0 = 5V [5kΩ], 1 = 10V [10kΩ]