GW Instek PCS-1000 User Manual

Precision Current Shunt Meter
PCS-1000
USER MANUAL
GW INSTEK PART NO. 82CS-1K000EB1
ISO-9001 CERTIFIED MANUFACTURER
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
Table of Contents
Table of Contents
SAFETY INSTRUCTIONS ................................................... 4
GETTING STARTED ........................................................... 8
PCS-1000 Overview ................................ 9
Appearance .......................................... 11
OPERATION .................................................................... 18
Set Up .................................................. 19
Basic Operation ................................... 25
COMMUNICATION INTERFACE ...................................... 42
Interface Configuration ........................ 44
Command Syntax ................................. 56
Command List ..................................... 59
Status Registers ................................... 86
Error Messages .................................... 87
APPENDIX ...................................................................... 88
PCS Default Settings ............................ 88
LED ASCII Table Character Set ............. 88
PCS-1000 Specifications ....................... 89
PCS Dimensions .................................. 91
Declaration of Conformity .................... 92
INDEX............................................................................. 93
3
PCS-1000 User Manual
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the instrument or to other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to insure your safety and to keep the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the instrument.
4
SAFETY INSTRUCTIONS
General Guideline
CAUTION
Do not place any heavy object on the
instrument.
Avoid severe impact or rough handling that
leads to damaging the instrument.
Do not discharge static electricity to the
instrument.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening. Do not disassemble the instrument unless you
are qualified.
(Measurement categories) EN 61010-1:2001 specifies the measurement categories and their requirements as follows. The instrument falls under category II (600VAC).
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed on
circuits not directly connected to Mains.
Power Supply
WARNING
AC Input voltage range:
100V/120V/220V/240V ±10% (selectable range)
Frequency: 50/60Hz To avoid electrical shock connect the protective
grounding conductor of the AC power cord to an earth ground.
Cleaning the Instrument
Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Safety Guidelines
5
PCS-1000 User Manual
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: Full accuracy to 80% RH, at
40°C
Altitude: < 2000m Temperature: 0°C to 50°C
(Pollution Degree) EN 61010-1:2001 specifies the pollution degrees and their requirements as follows. The instrument falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Storage environment
Location: Indoor Temperature: -40°C to 70°C Relative Humidity: <90%
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
6
SAFETY INSTRUCTIONS
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the instrument in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code:
As the colours of the wires in main leads may not correspond with the coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, a cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable, any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced in accordance to the above standard.
7
PCS-1000 User Manual
PCS-1000 Overview ........................................................... 9
Main Features .............................................................................................. 9
Accessories ................................................................................................ 10
Appearance ..................................................................... 11
Front Panel ................................................................................................ 11
Rear Panel .................................................................................................. 16
GETTING STARTED
This chapter describes the instrument in a nutshell, including its main features and front / rear panel introduction.
8
GETTING STARTED
Performance
Wide DC/AC voltage range
(200mV ~ 600VAC/1000VDC)
Wide AC/DC current range (30mA ~ 300A) Low drift at all ranges Low temperature coefficients
Features
Shunts: 0.001Ω, 0.01Ω, 0.1Ω, 1Ω, 10Ω Current Meter (6 1/2 digits current meter) Voltage Meter (6 1/2 digits voltage meter) Current Monitor Voltage and current can be measured at the
same time.
Interface
USB GPIB for SCPI commands
PCS-1000 Overview
The PCS-1000 uses five high-precision shunt resistors as the basis for accurate current and voltage measurements. The 5 shunt ranges
are 0.001Ω, 0.01Ω, 0.1Ω, 1Ω, 10Ω with a current measurement range
of 300A, 30A, 3A, 300mA and 30mA, respectively.
Main Features
9
Accessories
Standard Accessories
Part number
Description Region dependant
User manual
Region dependant
Power cord
GTL-105A
Alligator clip test leads (3A max): 1x red, 1x black
GTL-207
Banana plug test leads: 1x red, 1x black
GTL-240
USB Cable
PCS-001
Basic Accessory Kit:
Bolt HMS M8*16 x2 Nut hexagon M8*0.75P x2 Spring washer M8
8.4*13.7*1.5T x2 Plain washer M8
8.4*16*1.6T x2
Optional Accessories
Part number
Description GRA-419-J
Rack mount adapter (JIS)
GRA-419-E
Rack mount adapter (EIA)
PCS-1000 User Manual
10
GETTING STARTED
Local AC / DC 300A / 30A 3A Range
Func
AC / DC Range
RMT ACA 30A Auto
mA
A
CURRENT
ACV DCV Auto
mV
V
VOLTAGE
PCS-1000
: Long Push
Auto
DC 1000V AC 600V
MAX
Current Monitor
3A
MAX
30A MAX
DCA 300A ERROR
CAT II 600V
Func Select
Auto
Enter
( No Fuse ) ( Fused )
POWER
INPUT INPUT OUTPUT INPUT
1
6
914
2 3 54
1512 13
7
8 10 11
1. Power Switch
POWER
Turn on or off the main power.
2. AC/DC 30A
Terminal
30A MAX
( No Fuse )
INPUT
Accepts DC/AC. 30A maximum current input.
Warning: The maximum voltage difference between the negative terminal and earth cannot exceed 500Vpeak.
Appearance
Front Panel
11
PCS-1000 User Manual
3. AC/DC 3A
Terminal
3A
MAX
( Fused )
INPUT
Accepts DC/AC. 3A maximum current input. Internally, there is a fuse which protects the instrument from over current: Fuse Rating: T3.5A, 600V
Note: If the fuse is damaged, please contact your dealer or a GW Instek service center to replace the fuse.
Warning: The maximum voltage difference between the negative terminal and earth cannot exceed 500Vpeak.
4. Current
Monitor Sensor
Current Monitor
OUTPUT
Current Monitor Output.
Range 0~300mV (0~full scale of selected input range).
5. AC/DC
Voltage Terminal
DC 1000V AC 600V
MAX
INPUT
Accepts DC 1000V or AC 600V maximum voltage input.
Warning: The maximum voltage difference between the negative terminal and earth cannot exceed 500Vpeak.
12
GETTING STARTED
6. Local
Func
Func
AC / DCLocal
Local: Press to switch to local mode.
Func (long push)
Func: Long push to enter the Function menu. The Function menu is used to configure the instrument.
7. Func
Func
AC / DCLocal
Func
Use the Func arrows keys to scroll through each function when in the Function menu.
8. AC/DC
(Current)
Func
Func
AC / DCLocal
Selects DC or AC current measurement.
9. 300A/30A
Select
Auto
3A Range300A / 30A
Manually select the 300A or 30A measurement range.
10. Select
Auto
3A Range300A / 30A
Select
Use the Select arrow keys to edit parameter values when in the Function menu.
11. 3A Range
Select
Auto
3A Range300A / 30A
3A Range: Manually Select the 30mA, 300mA, or 3A measurement range.
Auto (long push)
Auto: Long push to automatically select 30mA, 300mA or 3A measurement ranges.
13
PCS-1000 User Manual
12. AC/DC
(Voltage)
AC / DC
Selects DC or AC voltage measurement.
13. Range
Enter
Auto
Range
Manually select the voltage measurement range: DC: 200mV, 2V, 20V, 200V, 1000V AC: 200mV, 2V, 20V, 200V, 600V
Enter
Secondary function that confirms selections when in the Function menu.
Auto (long push)
Voltage auto range.
14. Current Meter
RMT ACA 30A Auto
mA
A
CURRENT
ACV DCV Auto
VOLTAGE
DCA 300A ERROR
Displays current measurement.
RMT
The RMT icon will turn on when the instrument is in remote mode.
ACA
AC current measurement mode indicator.
DCA
DC current measurement mode indicator.
300A
300A measurement range indicator. Equivalent to choosing the rear panel 300A terminal.
30A
30A measurement range indicator. Equivalent to choosing the front panel 30A terminal.
14
GETTING STARTED
Auto
Autorange indicator for the 30mA, 300mA and 3A ranges. If the Autorange indicator is off, then that indicates that the range has been manually selected.
mA
Milliamp unit indicator.
A
Ampere unit indicator.
15. Voltage Meter
ACV DCV Auto
mV
V
VOLTAGE
Displays voltage measurement.
ERROR
Indicates an interface error. The SYSTem:ERRor? query can be used to read back error messages. See page 87 and 74 for details.
ACV
AC voltage measurement mode indicator.
DCV
DC voltage measurement mode indicator.
Auto
Autorange indicator. If the Auto indicator is off, then that indicates that the range has been manually selected.
mV
Millivolt unit indicator.
V
Volt unit indicator.
Warning:
For the 3A, 30A and 300A terminals on the front and rear panels, the maximum voltage difference between the negative terminal and earth cannot exceed 500Vpeak.
15
Rear Panel
GPIB
300A MAX.
50 / 60 Hz 35VA MAX.
LINE RATING
FUSE LINE
100VAC 120VAC
220VAC 240VAC
T200mA
SER.NO. LABEL
AND TEST LEADS BEFORE
DISCONNECT POWER CORD
REPLACING FUSE
INPUT
T100mA
( NO FUSE )
17 18
16 20
19
16. Power Cord
Socket
Fuse Socket
100
220
120
240
Accepts the power cord. Input: AC 100/120/220/240V ±10% Line frequency: 50Hz/60Hz Power: 35VA Max Fuse rating: T200mA, 250V for AC 100/120V; T100mA, 250V for AC 220/240V
17. GPIB
Communicati on Port
GPIB used for remote control.
18. USB
Communicati on Port
USB B device port. Used for remote control and firmware update.
PCS-1000 User Manual
16
GETTING STARTED
19. AC/DC 300A
Terminal
300A MAX.
INPUT
( NO FUSE )
Accepts AC/DC. 300A maximum current input.
20. Fan
Temperature controlled fan.
17
PCS-1000 User Manual
Set Up ............................................................................. 19
Power Up ................................................................................................... 19
Rack Mount ............................................................................................... 20
Wire Gauge Considerations ...................................................................... 21
Input Terminals ......................................................................................... 22
Basic Operation .............................................................. 25
Selecting AC/DC Current .......................................................................... 25
Selecting the Current Range ...................................................................... 25
Selecting AC/DC Voltage........................................................................... 26
Selecting the Voltage Range ...................................................................... 27
Voltage Range Conversion Table .............................................................. 28
Crest Factor Table ..................................................................................... 29
Using the Current Monitor Output ........................................................... 30
How to Use the Function Menu ............................................................... 31
View the Software Version ......................................................................... 33
Default Settings ......................................................................................... 34
Setting the USB-UART Baud Rate ............................................................. 35
Setting the GPIB Address .......................................................................... 36
Setting the AD Speed ................................................................................ 37
Setting the Averaging Mode ...................................................................... 38
Setting the Averaging Number for the DCV/ACV/DCA/ACA .................. 39
Setting the Autozero Function .................................................................. 40
Beeper Settings .......................................................................................... 41
OPERATION
18
Set Up
Steps
1. Connect the power cord to the rear panel
socket.
2. Press the power switch.
The unit will perform a calibration data and ROM check and then display the software version momentarily before it is ready to be used.
RMT ACA 30A Auto
mA
A
CURRENT
ACV DCV Auto
mV
V
VOLTAGE
DCA 300A ERROR
Note
In the event the calibration data and ROM check fails, CAL DATA FAIL will be displayed on the screen, as shown below. If the calibration data and ROM check fails, return the unit to an authorized GW Instek service center.
RMT ACA 30A Auto
mA
A
CURRENT
ACV DCV Auto
mV
V
VOLTAGE
DCA 300A ERROR
Note: The CAL DATA FAIL message will remain on the display until it is cleared. Press any key to clear the error message.
Power Up
OPERATION
19
Rack Mount
Background
The PCS-1000 has two types of the racks, GRA­419-E and the GRA-419-J for the EIA and JIS standards, respectively. Both types of the racks are 2U height racks and can fit 1 or 2 units. See the GRA-419 assembly manual for details.
GRA-419-E
GRA-419-J
PCS-1000 User Manual
20
OPERATION
Background
Before connecting the input terminals to a current/voltage source, the wire gauge of the cables should be considered.
It is essential that the current capacity of the cables is adequate. The rating of the cables must equal or exceed the maximum current input for the selected range.
Recommended wire gauge
Wire Gauge (AWG)
Nominal Cross Section (mm2)
Maximum Current (A) 20
0.5
9 18 1 13 16
1.5
18 14
2.5
24 12 4 34 10 6 45 8
10
64 6
16
88 4
25
120 2
32
145 1
50
190 00
70
240 000
95
290 0000
120
340
WARNING
Withstand voltage wire recommendations
As the PCS-1000 is a CAT II instrument, please ensure that the insulation capacity of the test cables exceed the DUT output voltage when performing current measurement.
Wire Gauge Considerations
21
PCS-1000 User Manual
Background
There are 3 terminals for the 300A, 30A and 3A/300mA/30mA ranges, respectively.
The 300A range uses the rear panel terminals and uses M8 crimped terminal cables.
The 30A range uses the 30A terminal and uses M4 sized crimped terminal cables or banana plugs.
The 3A input terminal uses standard banana plugs (GW Instek part number GTL-105A). The 3A terminal supports 3A, 30mA and 300mA ranges.
WARNING
Ensure any current or voltage sources are disabled before connecting any cables to the PCS-1000.
Steps
1. Turn the power switch off.
POWER
2. Connect the PCS-1000 in series
with the load and source. The current monitor output can be used in conjunction with a voltage meter.
Page 20
WARNING:
Do not short the positive or negative 3A, 30A and 300A terminals.
Input Terminals
22
OPERATION
Current Meter Connection
Load
Power
Source
+ current input
- current input
Voltage Meter Connection
Load
Power
Source
+ voltage input
- voltage input
23
PCS-1000 User Manual
Voltage + Current Meter Connection
Load
Power
Source
+V -V
+A
-A
1
2
3
1 2 3
Measure voltage at load terminal Measure voltage at source terminal Measure current
Current Monitor
DVM
– current monitor+ current monitor
Rear Panel Terminals
24
OPERATION
Background
AC or DC current can be measured when in measurement mode.
Steps
1. Press the AC/DC key under the CURRENT
meter current display to toggle between AC and DC current measurement.
2. The ACA or DCA indicator will be shown on
the display.
RMT ACA 30A Auto
mA
A
CURRENT
ACV DCV Auto
mV
V
VOLTAGE
DCA 300A ERROR
Background
There are 5 selectable current ranges. The range can be manually or automatically selected. Selecting a current range will also select the corresponding the input terminal.
300A/30A
Press the 300/30A key to toggle between the 300A and 30A ranges (as indicated on the display).
The 300A range will select the 300A terminal. The 30A range will select the 30A terminal.
3A
Press the 3A Range key toggle between the 30mA, 300mA and 3A ranges. Selecting the 3A, 30mA or 300mA range will select the 3A terminal.
Basic Operation
Selecting AC/DC Current
Selecting the Current Range
25
PCS-1000 User Manual
Note
The selected range is indicated by the displayed unit (A or mA) and the number of significant digits before the decimal place:
3A: Unit=A; 1 significant digit 30mA: Unit=mA; 2 signicant digits 300mA: Uni=mA; 3 significant digits
Autorange
Long push the Auto (3A Range) key to select autorange.
Auto will be displayed in the CURRENT display when autorange is active.
The autorange function is only applicable for the 3A, 30mA and 300mA ranges. Autorange is not supported for the 30A and 300A ranges.
RMT ACA 30A Auto
mA
A
CURRENT
ACV DCV Auto
mV
V
VOLTAGE
DCA 300A ERROR
Note
Autorange will also be automatically selected when switching from 300A/30A to 3A.
Background
AC or DC voltage can be measured.
Steps
1. Press the AC/DC key under the VOLTAGE
meter display to toggle between AC and DC voltage measurement.
2. The ACV or DCV indicator will be shown on
the display.
RMT ACA 30A Auto
mA
A
CURRENT
ACV DCV Auto
mV
V
VOLTAGE
DCA 300A ERROR
Selecting AC/DC Voltage
26
OPERATION
Background
There are 5 selectable voltage ranges. The range can be manually or automatically selected.
Manual Ranges
Press the Range key to cycle between each voltage range.
ACV: DCV:
200mV, 2V, 20V, 200V, 600V 200mV, 2V, 20V, 200V, 1000V
Note
The selected range is indicated by the displayed unit (V or mV) and the number of significant digits before the decimal place:
200mV: Unit=mV; 3 significant digits 2V: Unit=V; 1 significant digit 20V: Unit=V; 2 significant digits 200V: Unit=V; 3 significant digits AC 600V: Unit=V; 3 significant digits DC 1000V: Unit=V; 4 significant digits
Autorange
Long push the Auto key to select autorange.
Auto will be displayed in the VOLTAGE display when autorange is active.
RMT ACA 30A Auto
mA
A
CURRENT
ACV DCV Auto
mV
V
VOLTAGE
DCA 300A ERROR
Selecting the Voltage Range
27
PCS-1000 User Manual
Waveform
Peak to Peak
AC (True RMS)
DC
Sine
PK-PK
2.828
1.000
0.000 Rectified Sine (full wave)
PK-PK
1.414
0.435
0.900
Rectified Sine (half wave)
PK-PK
2.000
0.771
0.636
Square
PK-PK
2.000
1.000
0.000 Rectified Square
PK-PK
1.414
0.707
0.707
Rectangular Pulse
PK-PK
X
Y
2.000
2K
K=
)2
( DD
D=X/Y
2D
D=X/Y
Triangle Sawtooth
PK-PK
3.464
1.000
0.000
Voltage Range Conversion Table
This table shows the relationship between AC and DC readings in various waveforms.
28
OPERATION
Waveform
Shape
Crest factor
Square wave
1.0
Sine wave
1.414
Triangle sawtooth
1.732
Mixed frequencies
1.414 ~ 2.0
SCR output 100% ~
10%
1.414 ~ 3.0 White noise
3.0 ~ 4.0
AC Coupled pulse train
>3.0
Spike
>9.0
Crest Factor Table
Crest factor is the ratio of the peak signal amplitude to the RMS value of the signal. It determines the accuracy of AC measurement.
If the crest factor is less than 3.0, voltage measurement will not result in error due to dynamic range limitations at full scale.
If the crest factor is more than 3.0, it usually indicates an abnormal waveform as seen from the below table.
29
PCS-1000 User Manual
Background
The current monitor is used to measure the voltage drop across the shunt resistors manually.
The current monitor outputs the full scale current input (for the selected range) as a voltage of 0~300mV.
Shunt Values
Range
Shunt
30 mA
10Ω
300 mA
1Ω
3 A
0.1Ω
30 A
0.01Ω
300 A
0.001Ω
Steps
1. Set the PCS-1000 for normal operation, as
described previously in this chapter, page 25~27.
Make note of the range used and the shunt that is used for that range.
2. Connect the current monitor output to a DVM.
3. Use OHM’s law, V=IR, to determine the
current across the shunt resistor.
For example: If we are using the 3A current range (and thus the 0.1Ω shunt) and the current monitor outputs 150mV, then:
Input current = monitor output / shunt Ω
= 150mV/0.1Ω = 1.5A
Using the Current Monitor Output
30
OPERATION
Background
The function menu allows you to view the software information, set the remote settings, the DCV, ACV, DCA, ACA averaging settings and other settings.
Menu Item
Range/Description
Software Version
Displays the software version on the display.
Factory Default
Load the default settings.
USB to Serial Port Baud Rate
115200, 57600, 38400, 19200, 9600, 4800
GPIB Address
00 ~ 30
AD Speed (measurement resolution)
7_sec (6½ digits), 30_sec (5½ digits), 100_sec (4½ digits)
AVG Mode
SHIFT, TOTAL
DCV AVG
01 ~ 10, 20, 30, 40, 50, 60, 70, 80, 90, 100
ACV AVG
01 ~ 10, 20, 30, 40, 50, 60, 70, 80, 90, 100
DCA AVG
01 ~ 10, 20, 30, 40, 50, 60, 70, 80, 90, 100
ACA AVG
01 ~ 10, 20, 30, 40, 50, 60, 70, 80, 90, 100
Auto Zero
Enable, Disable
Beeper
On, Off
Save Func Set
Saves the settings in the function menus.
Exit Func Set
Exits the function menu.
How to Use the Function Menu
31
PCS-1000 User Manual
Steps
1. Press and long push the Func key.
The software version will be displayed first.
Func
Local
RMT ACA 30A Auto
mA
A
CURRENT
ACV DCV Auto
mV
V
VOLTAGE
DCA 300A ERROR
2. Use the Func keys to
scroll through the menu items.
Func
Func
AC / DCLocal
3. Use the Select keys to
choose the parameter for the selected menu item.
Select
Auto
3A Range300A / 30A
4. Press the Enter key to set the
parameter and go to the next menu item.
Enter
Auto
Range
Save Setup
To save the settings use the Func keys to navigate to SAVE FUNC SET.
Press the Enter key to save all the settings and exit the function menu.
Exit Without Saving
To exit without saving, navigate to the EXIT FUNC SET menu using the Func keys and
press the Enter key to exit without saving any settings.
RMT ACA 30A Auto
mA
A
CURRENT
ACV DCV Auto
mV
V
VOLTAGE
DCA 300A ERROR
Note
If the settings in the function menu are not saved, then the settings will only apply until the unit is reset.
32
OPERATION
Note
The display uses a 7 segment LED display. The appendix has an ASCII Table if you have trouble understanding the characters on the LED display character set. See page 88.
Background
The display will show the software version.
Display
RMT ACA 30A Auto
mA
A
CURRENT
ACV DCV Auto
mV
V
VOLTAGE
DCA 300A ERROR
Steps
Long push the Func key.
The software version is displayed on the screen (it is the first item in the function menu).
Exit
To exit, use the Func keys to change the menu to the EXIT FUNC SET menu item. Press the Enter key to exit.
View the Software Version
33
PCS-1000 User Manual
Background
The Factory Default function will restore the default settings.
Steps
1. Long push the Func key.
The function menu will appear.
1. Use the Func keys to navigate to the
FACTORY DEFAULT menu.
2. Press the Enter key to set the mode.
See page 88 for a list of the default settings.
Default Settings
34
OPERATION
Background
The baud rate settings are used for remote control via the USB B port. The USB B connection uses a virtual COM port to simulate a serial port (UART) connection. The baud rate can be set to 115200, 57600, 38400, 19200, 9600,
4800.
See the Communication Interface chapter on page 42 chapter for details on remote control.
Note
The USB driver needs to be installed for the baud rate settings to be applicable. See page 48 for details.
Steps
1. Long push the Func key.
The function menu will appear.
2. Use the Func keys to navigate to the
BAUDRATE settings.
3. Use the Select keys to select a baud rate.
Press the Enter key to set the baud rate.
4. Use the Func keys to change the menu to
the SAVE FUNC SET menu item. Press the Enter key to save.
Note
To exit without saving, navigate to the
EXIT FUNC
SET
menu using the
Func
keys and press the
Enter
key to exit without saving any settings.
Setting the USB-UART Baud Rate
35
PCS-1000 User Manual
Background
The GPIB port is used for remote control. The GPIB address can be set between 00 ~ 30.
See the Communication Interface chapter on page 42 chapter for details on remote control.
Steps
1. Long push the Func key.
The function menu will appear.
2. Use the Func keys to navigate to the
ADDRESS settings.
3. Use the Select keys to select the GPIB
address. Press the Enter key to set the address.
4. Use the Func keys to navigate to the
SAVE FUNC SET menu item. Press the Enter key to save.
Note
To exit without saving, navigate to the
EXIT FUNC
SET
menu using the
Func
keys and press the
Enter
key to exit without saving any settings.
Setting the GPIB Address
36
OPERATION
Background
The ADC IC speed has a number of settings. The higher the setting, the lower the accuracy and resolution of the meter.
Range:
Seconds (resolution): 7 (6½ digits), 30 (5½ digits), 100 (4½ digits)
Steps
1. Long push the Func key.
The function menu will appear.
2. Use the Func keys to navigate to the AD
SPEED menu.
3. Use the Select keys to select the AD speed.
Press the Enter key to set the speed.
By default the AD Speed is set to 7 (6½ digits).
4. Use the Func keys to navigate to the
SAVE FUNC SET menu item. Press the Enter key to save.
Note
To exit without saving, navigate to the
EXIT FUNC
SET
menu using the
Func
keys and press the
Enter
key to exit without saving any settings.
Setting the AD Speed
37
PCS-1000 User Manual
Background
There are two different types of averaging modes, SHIFT or TOTAL.
SHIFT is a box car averaging mode while TOTAL will average all the collected samples to get the average value.
Range
SHIFT, TOTAL
Steps
1. Long push the Func key.
The function menu will appear.
2. Use the Func keys to navigate to the AVG
MODE menu.
3. Use the Select keys to select the Averaging
Mode. Press the Enter key to set the mode.
By default the average mode is set to SHIFT.
4. Use the Func keys to navigate to the
SAVE FUNC SET menu item. Press the Enter key to save.
Note
To exit without saving, navigate to the
EXIT FUNC
SET
menu using the
Func
keys and press the
Enter
key to exit without saving any settings.
Setting the Averaging Mode
38
OPERATION
Background
Each of the different measurement modes (DCV, ACV, DCA, ACA) can have the number of averages set individually.
Range
01 ~ 10, 20, 30, 40, 50, 60, 70, 80, 90, 100
Steps
1. Long push the Func key.
The function menu will appear.
2. Use the Func keys to navigate to the DCV
AVG, ACV AVG, DCA AVG or ACA AVG
menu.
3. Use the Select keys to select the number of
averages for the selected mode. Press the Enter key to set the mode.
By default the number of averages is 10.
4. Use the Func keys to navigate to the
SAVE FUNC SET menu item. Press the Enter key to save.
Note
To exit without saving, navigate to the
EXIT FUNC
SET
menu using the
Func
keys and press the
Enter
key to exit without saving any settings.
Setting the Averaging Number for the DCV/ACV/DCA/ACA
39
PCS-1000 User Manual
Background
The Autozero function will automatically perform a zero calibration when the unit is turned on.
Range
Enable, Disable
Steps
1. Long push the Func key.
The function menu will appear.
2. Use the Func keys to navigate to the
AUTOZERO menu.
3. Use the Select keys to enable autozero.
Press the Enter key to set the mode.
By default the Autozero is already enabled.
4. Use the Func keys to navigate to the
SAVE FUNC SET menu item. Press the Enter key to save.
Note
To exit without saving, navigate to the
EXIT FUNC
SET
menu using the
Func
keys and press the
Enter
key to exit without saving any settings.
Setting the Autozero Function
40
OPERATION
Background
The beeper sound that is used for key presses and other system sounds can be turned on or off using this menu.
Range
On, Off
Steps
1. Long push the Func key.
The function menu will appear.
2. Use the Func keys to navigate to the
BEEPER menu.
3. Use the Select keys to the beeper on or off.
Press the Enter key to set the mode.
By default the beeper sound is turned on.
4. Use the Func keys to navigate to the
SAVE FUNC SET menu item. Press the Enter key to save.
Note
To exit without saving, navigate to the
EXIT FUNC
SET
menu using the
Func
keys and press the
Enter
key to exit without saving any settings.
Beeper Settings
41
PCS-1000 User Manual
Interface Configuration ................................................... 44
Configure GPIB Interface ........................................................................... 44
GPIB Function Check ................................................................................. 45
USB Driver Installation ............................................................................... 48
USB Interface Settings ............................................................................... 50
USB Function Check ................................................................................... 53
Return to Local Operation .......................................................................... 55
Command Syntax ............................................................ 56
Command List ................................................................. 59
Configure Commands ................................................................................ 61
CONFigure .................................................................................................. 61
CONFigure:CURRent ................................................................................. 62
CONFigure:CURRent[:DC] ......................................................................... 62
CONFigure:CURRent:AC ............................................................................ 63
CONFigure:VOLTage ................................................................................. 63
CONFigure:VOLTage[:DC] ......................................................................... 64
CONFigure:VOLTage:AC ............................................................................ 65
CONFigure:AVERage:MODE ..................................................................... 65
Measure Commands .................................................................................. 66
MEASure ..................................................................................................... 66
MEASure:CURRent[:DC] ............................................................................. 66
MEASure:CURRent:AC ............................................................................... 67
MEASure:VOLTage[:DC] ............................................................................. 67
MEASure:VOLTage:AC ............................................................................... 67
READ ........................................................................................................... 67
COMMUNICATION
INTERFACE
This chapter describes basic configuration of IEEE488.2 based remote control.
42
COMMUNICATION INTERFACE
Sense Commands ...................................................................................... 69
[SENSe:]CURRent:RANGe ......................................................................... 69
[SENSe:]CURRent:DC:AVERage:COUNt ................................................... 70
[SENSe:]CURRent:AC:AVERage:COUNt ................................................... 70
[SENSe:]VOLTage:RANGe ......................................................................... 70
[SENSe:]VOLTage:DC:AVERage:COUNt ................................................... 71
[SENSe:]VOLTage:AC:AVERage:COUNt ................................................... 72
System Commands ................................................................................... 73
SYSTem:BEEPer:STATe ............................................................................. 73
SYSTem:ERRor........................................................................................... 74
SYSTem:LOCal........................................................................................... 74
SYSTem:REMote ........................................................................................ 75
SYSTem:RWLock ....................................................................................... 75
SYSTem:VERSion ....................................................................................... 75
SYSTem:OUTPut:FORMat ........................................................................ 75
Status Commands ..................................................................................... 77
STATus:OPERation:CONDition ................................................................ 77
STATus:OPERation:ENABle ...................................................................... 78
STATus:OPERation[:EVENt] ...................................................................... 78
STATus:PRESet .......................................................................................... 79
STATus:QUEStionable:CONDition ........................................................... 79
STATus:QUEStionable:ENABle ................................................................. 80
STATus:QUEStionable[:EVENt] ................................................................. 80
Common Commands ................................................................................ 81
*IDN? ......................................................................................................... 81
*ESE ........................................................................................................... 81
*ESR? ......................................................................................................... 82
*SRE ........................................................................................................... 82
*STB? ......................................................................................................... 83
*PSC ........................................................................................................... 84
*OPC .......................................................................................................... 84
*TST? ......................................................................................................... 85
*CLS ........................................................................................................... 85
*RST ........................................................................................................... 85
*WAI ........................................................................................................... 85
Status Registers .............................................................. 86
Error Messages ............................................................... 87
43
PCS-1000 User Manual
Configure GPIB
1. Connect the GPIB cable from the GPIB
controller to the PCS-1000.
2. Turn the PCS-1000 on.
3. Long push Func key to enter the
function menu.
Page 31
4. Use the Func keys to go to the ADDRESS
function.
5. Select the address using the Select keys.
GPIB Address
00~30
6. Press the Enter key to confirm the selection.
Note
RMT will be displayed on the screen when the unit is remote mode.
GPIB constraints
Maximum 14 devices altogether, 20m cable
length, 2m between each device
Unique address assigned to each device At least 2/3 of the devices turned On No loop or parallel connection
Interface Configuration
Configure GPIB Interface
To use GPIB the GPIB address must first be set.
44
COMMUNICATION INTERFACE
Background
To test the GPIB functionality, National Instruments Measurement and Automation Explorer can be used. This program is available on the NI website, www.ni.com, via a search for the VISA Run-time Engine page, or
“downloads” at the following URL,
http://www.ni.com/visa/
Requirements
Operating System: Windows XP, 7, 8
Functionality check
1. Start the NI Measurement and Automation
Explorer (MAX) program. Using Windows, press:
Start>All Programs>National Instruments>Measurement & Automation
GPIB Function Check
45
PCS-1000 User Manual
1. From the Configuration panel access;
My System>Devices and Interfaces>GPIBX (where X is the GPIB card number that is connected to the PCS-1000).
2. Click Scan for Instruments.
3. Double click on the Instrument 0 icon.
1
2
3
46
COMMUNICATION INTERFACE
4. Click on Communicate with Instrument.
5. In the communicator window that appears,
ensure *IDN? is written in the Send test box.
6. Click on the Query button to send the *IDN?
query to the instrument.
7. The following string should be returned:
GWInstek, PCS-1000, xxxxxxxxx, Vx.xx
(Manufacturer, model, serial, software version)
209.1711 mm.
4
5
6
7
47
PCS-1000 User Manual
Background
The USB driver is actually a virtual COM port driver that simulates a serial port (UART) connection.
Note: The USB driver should not need to be manually installed if your operating system has been fully updated. In most cases, the PCS-1000 driver should be automatically installed when connected to the PC.
If the driver is not automatically detected, or if your operating system is not fully updated, it may be necessary to install the USB driver, as shown below.
Requirements
Operating System: Windows XP, Vista, 7, 8, 8.1
Note
The following installation instructions only apply if the USB driver does not get automatically installed.
Steps
1. Connect the PCS-1000 to a PC using the USB
Type A-Type B cable (GTL-240).
2. The Windows Found New Hardware wizard
should pop up asking you to install the device driver.
3. Select Locate and install driver software.
4. You will now be asked to insert a disk that
contains the USB driver.
Insert the User Manual CD. Windows will automatically install the USB driver.
Note: If the Windows Security pop-up appears,
USB Driver Installation
48
COMMUNICATION INTERFACE
choose Install this driver software anyway.
5. The PCS-1000 will now become available in the
device tree under PORTS (COM & LPT) in the Windows Device Manager.
Alternate Installation
If the Found New Hardware wizard does not appear or you wish to install the driver from another location, the driver can be also installed from the Windows Device Manager.
1. Open the Windows Device Manager. Using
Windows 7, press:
Start>Control Panel>Hardware and Sound>Device Manager
2. From the device tree go to: Other devices>USB
Serial Port
The yellow error sign indicates that a driver has not been installed.
49
PCS-1000 User Manual
3. Right-click USB Serial Port and select Update
Driver Software.
Select Browse my computer for driver software when prompted.
Select the directory with the USB drivers from the User Manual CD when prompted.
Note: If the Windows Security pop-up appears, choose Install this driver software anyway.
4. The PCS-1000 will now become available in the
device tree under PORTS (COM & LPT).
Note
If required, the USB drivers can be downloaded from http://www.ftdichip.com/Drivers/VCP.htm.
If the drivers are downloaded, they can be installed using the Alternate Installation method described on the previous page.
Baud Rate Settings
1. Connect the USB cable from the PC to the rear
panel USB-B port on the PCS-1000.
2. Turn the PCS-1000 on.
3. Long push Func key to enter the
function menu.
Page 31
4. Use the Func keys to go to the
BAUDRATE function.
5. Select the baud rate using the Select keys.
Baud Rate
4800, 9600(default), 19200, 38400, 57600, 115200
USB Interface Settings
50
COMMUNICATION INTERFACE
6. Press the Enter key to confirm the selection.
7. Use the Func keys to go to the SAVE
FUNC SET function.
8. Press the Enter key to save the baud rate
settings.
Edit UART Settings
1. Connect the PCS-1000 to the PC using the GTL-
240 USB cable.
2. Open the Windows Device Manager, using
Windows 7, click:
Start>Control Panel>Hardware and Sound>Device Manager:
3. In the device tree go to: PORTS (COM &
LPT)>PCS-1000 (COM XX)
4. Right-click PCS-1000 and select Properties.
5. Go to the Port Settings tab and from there you
can set any other UART settings such as data bits, parity, number of stop bits and the flow control.
51
PCS-1000 User Manual
52
COMMUNICATION INTERFACE
Background
To test the USB functionality, National Instruments Measurement and Automation Explorer can be used. This program is available on the NI website, www.ni.com, via a search for the VISA Run-time Engine page, or
“downloads” at the following URL,
http://www.ni.com/visa/
Requirements
Operating System: Windows XP, 7, 8, 8.1
Functionality check
1. Open the Windows Device Manager to see
which COM port the PCS has been assigned. Using Windows 7, press:
Start>Control Panel>Hardware and Sound>Device Manager
The COM port number will be shown in the device tree under: PORTS (COM & LPT)>PCS- 1000 (COM XX)
2. Start the NI Measurement and Automation
Explorer (MAX) program. Using Windows, press:
Start>All Programs>National Instruments>Measurement & Automation
USB Function Check
53
PCS-1000 User Manual
3. From the Configuration panel access;
My System>Devices and Interfaces>Serial & Parallel>COMX (where X is the COM port
number assigned to the PCS-1000).
4. Click on the Port Settings tab at the bottom.
5. Make sure the Baud rate settings are correct
(PCS-1000 default = 9600 baud).
6. Click on Open Visa Test Panel.
3
4
5
6
54
COMMUNICATION INTERFACE
7. Click on Input/Output.
8. In the Select or Enter Command drop down list,
ensure *IDN?\n is selected.
9. Click on the Query button to send the *IDN?
query to the instrument.
10. The following string should be returned:
GWInstek, PCS-1000, xxxxxxxxx, Vx.xx
(Manufacturer, model, serial, software version)
7
9
8
10
Steps
1. Press the Local key to return to local operation.
2. The RMT icon will turn off when you have
returned to local mode.
Return to Local Operation
55
PCS-1000 User Manual
Compatible Standard
IEEE488.2
Partial compatibility
SCPI, 1999
Partial compatibility
Command Structure
SCPI commands follow a tree-like structure, organized into nodes. Each level of the command tree is a node. Each keyword in a SCPI command represents each node in the command tree. Each keyword (node) of a SCPI command is separated by a colon (:).
For example, the diagram below shows an SCPI sub-structure and a command example.
MEASure
CURRent
MEASure:CURRent:DC?
AC
DC
Command types
There are a number of different instrument commands and queries. A command sends instructions or data to the unit and a query receives data or status information from the unit.
Command types
Simple
A single command with/without a parameter
Example
*IDN?
Command Syntax
56
COMMUNICATION INTERFACE
Query
A query is a simple or compound command followed by a question mark (?). A parameter (data) is returned.
Example
meas:curr:dc?
Compound
Two or more commands on the same command line. Compound commands are separated with either a semi­colon (;) or a semi-colon and a colon (;:).
A semi-colon is used to join two related commands, with the caveat that the last command must begin at the last node of the first command.
A semi-colon and colon are used to combine two commands from different nodes.
Example
conf:curr?;:meas:volt:dc?
57
PCS-1000 User Manual
Command Forms
Commands and queries have two different forms, long and short. The command syntax is written with the short form of the command in capitals and the remainder (long form) in lower case.
The commands can be written in capitals or lower-case, just so long as the short or long forms are complete. An incomplete command will not be recognized.
Below are examples of correctly written commands.
Long form
CONFigure:VOLTage? CONFIGURE:VOLTAGE? configure:voltage?
Short form
CONF:VOLT? conf:volt?
Square Brackets
Commands that contain square brackets indicate that the contents are optional. The function of the command is the same with or without the square bracketed items, as shown below.
For “MEASure:CURRent[:DC]?”, both MEASure:CURRent:DC?” and “MEASure:CURRent?” are both valid forms.
Command Format
AUTO
1 2 3
CUR R :RAN G
1. Command
header
2. Space
3. Parameter 1
Parameters
Type
Description
Example
<Boolean>
Boolean logic
0, 1
<NR1>
integers
0, 1, 2, 3
58
COMMUNICATION INTERFACE
<NR2>
decimal numbers
0.1, 3.14, 8.5
<NR3>
floating point
4.5e-1, 8.25e+1
<NRf>
any of NR1, 2, 3
1, 1.5, 4.5e-1
<block data>
Definitive length arbitrary block data. A single decimal digit followed by data. The decimal digit specifies how many 8-bit data bytes follow.
Message Terminator
LF
Line feed code
Configure Commands
CONFigure .................................................................. 61
CONFigure:CURRent .................................................. 62
CONFigure:CURRent[:DC] ......................................... 62
CONFigure:CURRent:AC ............................................ 63
CONFigure:VOLTage .................................................. 63
CONFigure:VOLTage[:DC] ......................................... 64
CONFigure:VOLTage:AC ............................................ 65
CONFigure:AVERage:MODE ..................................... 65
Measure Commands
MEASure ..................................................................... 66
MEASure:CURRent[:DC] ............................................. 66
MEASure:CURRent:AC ............................................... 67
MEASure:VOLTage[:DC] ............................................. 67
MEASure:VOLTage:AC ............................................... 67
READ ........................................................................... 67
Sense Commands
[SENSe:]CURRent:RANGe .......................................... 69
[SENSe:]CURRent:DC:AVERage:COUNt .................... 70
[SENSe:]CURRent:AC:AVERage:COUNt .................... 70
[SENSe:]VOLTage:RANGe .......................................... 70
[SENSe:]VOLTage:DC:AVERage:COUNt .................... 71
[SENSe:]VOLTage:AC:AVERage:COUNt .................... 72
Command List
59
PCS-1000 User Manual
System Commands
SYSTem:BEEPer:STATe ............................................... 73
SYSTem:ERRor ............................................................ 74
SYSTem:LOCal ............................................................ 74
SYSTem:REMote .......................................................... 75
SYSTem:RWLock ......................................................... 75
SYSTem:VERSion ......................................................... 75
SYSTem:OUTPut:FORMat .......................................... 75
Status Commands
STATus:OPERation:CONDition .................................. 77
STATus:OPERation:ENABle ........................................ 78
STATus:OPERation[:EVENt] ........................................ 78
STATus:PRESet ............................................................ 79
STATus:QUEStionable:CONDition ............................. 79
STATus:QUEStionable:ENABle ................................... 80
STATus:QUEStionable[:EVENt] ................................... 80
Common Commands
*IDN? ........................................................................... 81
*ESE ............................................................................. 81
*ESR? ........................................................................... 82
*SRE ............................................................................. 82
*STB? ........................................................................... 83
*PSC ............................................................................. 84
*OPC ............................................................................ 84
*TST? ........................................................................... 85
*CLS ............................................................................. 85
*RST ............................................................................. 85
*WAI............................................................................. 85
60
COMMUNICATION INTERFACE
CONFigure .................................................................. 61
CONFigure:CURRent .................................................. 62
CONFigure:CURRent[:DC] ......................................... 62
CONFigure:CURRent:AC ............................................ 63
CONFigure:VOLTage.................................................. 63
CONFigure:VOLTage[:DC] ......................................... 64
CONFigure:VOLTage:AC ............................................ 65
CONFigure:AVERage:MODE ..................................... 65
CONFigure
Query
Description
The CONFigure query will return both the current and voltage configuration as a string.
Query Syntax
CONFigure?
Return Parameter
<string>
Current mode, range unit, voltage mode, range unit.
Query Example
CONF?
>CURR:DC 0.01,VOLT:DC 0.1
Note
The range that is returned is the base unit. See the table below:
Unit
Voltage Range
Current Range
1000
1000VDC
N/A
600
600ACV
N/A
100
200V
300A
10
20V
30A
1
2V
3A
0.1
200mV
300mA
0.01
N/A
30mA
Configure Commands
61
PCS-1000 User Manual
CONFigure:CURRent
Query
Description
The CONFigure:CURRent query will return the current range unit.
Query Syntax
CONFigure:CURRent?
Return Parameter
<string>
Returns the current mode and range unit.
Query Example
CONF:CURR?
> DC 0.01
Note
The range that is returned is the base unit. See the table below:
Unit
Current Range
100
300A
10
30A
1
3A
0.1
300mA
0.01
30mA
CONFigure:CURRent[:DC]
Set
Description
This command will set the current mode to DC and set the range. If the range is not specified, then it will not change.
Syntax
CONFigure:CURRent[:DC] [<Range> | AUTO]
Parameter
<Range>
Current range <NRf>: 0.00000001~305 The unit will automatically be set to the closest range.
AUTO
Autorange; Only applicable for the ≤3A ranges.
Autorange is not supported for the 30A and 300A ranges.
62
COMMUNICATION INTERFACE
Example
CONF:CURR 20
Sets the current mode to DC and the range to 30A
Example
CONF:CURR
Sets the current mode to DC. The range is not changed.
CONFigure:CURRent:AC
Set
Description
This command will set current mode to AC and set the range. If the range is not specified, then it will not change.
Syntax
CONFigure:CURRent:AC [<Range> | AUTO]
Parameter
<Range>
Current range <NRf>: 0.00000001~305 Current range. The unit will automatically be set to the closest range.
AUTO
Autorange; Only applicable for the ≤3A
ranges.
Autorange is not supported for the 30A and 300A ranges.
Example
CONF:CURR:AC 100
Sets the current mode to AC and the range to 300A.
Example
CONF:CURR:AC
Sets the current mode to AC. The range is not changed.
CONFigure:VOLTage
Query
Description
The CONFigure:VOLTage query will return the voltage mode and the voltage range unit.
Query Syntax
CONFigure:VOLTage?
Return Parameter
<string>
Returns the voltage mode and range unit.
63
PCS-1000 User Manual
Query Example
CONF:VOLT?
>DC 0.1
The mode is DCV and the range is 200mV.
Note
The range that is returned is the base voltage unit. See the table below:
Unit
Voltage Range
1000
1000VDC
600
600ACV
100
200V
10
20V 1 2V
0.1
200mV
CONFigure:VOLTage[:DC]
Set
Description
This command will set the voltage mode to DC and set the DCV range. If the range is not specified then it will not be changed.
Syntax
CONFigure:VOLTage[:DC] [<Range> | AUTO]
Parameter
<Range>
Voltage range <NRf>: 0.0000001 ~ 1050 The unit will automatically be set to the closest range.
AUTO
Autoset
Example
CONF:VOLT:DC 20
Sets the voltage mode to DC and the DCV range to 20V.
Example
CONF:VOLT:DC
Sets the voltage mode to DC. The range stays the same.
64
COMMUNICATION INTERFACE
CONFigure:VOLTage:AC
Set
Description
This command will set the voltage mode to AC and set the ACV range. If the range is not specified then it will not be changed.
Syntax
CONFigure:VOLTage:AC [<Range> | AUTO]
Parameter
<Range>
Voltage range <NRf>: 0.0000001~630 The unit will automatically be set to the closest range.
AUTO
Autoset
Example
CONF:VOLT:AC 20
Sets the voltage mode to AC and the ACV range to 20V.
Example
CONF:VOLT:AC
Sets the voltage mode to AC. The range stays the same.
CONFigure:AVERage:MODE
Set
Query
Description
This command will set or query the average mode.
Syntax
Query Syntax
CONFigure:AVERage:MODE {0|1,TOTAL|SHIFT}
CONFigure:AVERage:MODE?
Parameter
0, TOTAL
Total mode
1, SHIFT
Shift mode
Return Parameter
Total
Total mode
Shift
Shift mode
Example
CONF:AVER:MODE 0
Sets the average mode to Total mode.
65
PCS-1000 User Manual
MEASure ...................................................................... 66
MEASure:CURRent[:DC] .............................................. 66
MEASure:CURRent:AC ................................................ 67
MEASure:VOLTage[:DC] .............................................. 67
MEASure:VOLTage:AC ................................................ 67
READ ............................................................................ 67
MEASure
Query
Description
This query will return all the measurements.
Query Syntax
MEASure?
Return Parameter
<NRf>
Returns the current measurement voltage measurement:
<current>,<voltage>
Query Example
MEAS?
> 9.9768E-1, 3.21E-1
Returns the current measurement (0.99A) and voltage (0.321V) measurement.
MEASure:CURRent[:DC]
Query
Description
This query will return the DC current.
Query Syntax
Measure:CURRent[:DC]?
Return Parameter
<NRf>
Return the DC current.
Query Example
MEAS:CURR:DC?
>+9.9067E-1
Returns DC current measurement (0.99A).
Measure Commands
66
COMMUNICATION INTERFACE
MEASure:CURRent:AC
Query
Description
This query will return the AC current.
Query Syntax
MEASure:CURRent:AC?
Return Parameter
<NRf>
Returns the AC current.
Query Example
MEAS:CURR:AC?
>+9.9067E-1
Returns the AC current measurement (0.9A).
MEASure:VOLTage[:DC]
Query
Description
This query will return the DC voltage.
Query Syntax
MEASure:VOLTage[:DC]?
Return Parameter
<NRf>
Returns the DC voltage
Query Example
MEAS:VOLT:DC?
>+1.5E+1
Returns the DC voltage measurement (15.0 V).
MEASure:VOLTage:AC
Query
Description
This query will return the AC voltage.
Query Syntax
MEASure:VOLTage:AC?
Return Parameter
<NRf>
Returns the AC voltage.
Query Example
MEAS:VOLT:AC?
>+2.5E+1
Returns the AC voltage measurement (25V).
READ
Query
Description
The read command will return current and voltage reading.
67
PCS-1000 User Manual
Query Syntax
READ?
Return Parameter
<NRf>
Returns the current and voltage readings, respectively <current>,<voltage>
Query Example
READ?
> +9.9067E-1,+2.5E+1
Returns the current and voltage readings.
68
COMMUNICATION INTERFACE
[SENSe:]CURRent:RANGe .......................................... 69
[SENSe:]CURRent:DC:AVERage:COUNt .................... 70
[SENSe:]CURRent:AC:AVERage:COUNt .................... 70
[SENSe:]VOLTage:RANGe .......................................... 70
[SENSe:]VOLTage:DC:AVERage:COUNt .................... 71
[SENSe:]VOLTage:AC:AVERage:COUNt .................... 72
[SENSe:]CURRent:RANGe
Set
Query
Description
Sets or queries the current range.
Syntax
Query Syntax
[SENSe:]CURRent:RANGe {<Range>|AUTO}
[SENSe:]CURRent:RANGe?
Parameter /
Return Parameter
<Range>
Current range <NRf>: 0.00000001305 Sets the current range in amps. The unit will automatically choose the closest range that is programmed.
AUTO
Sets the range to AUTO; Only applicable for the ≤3A ranges.
Autorange is not supported for the 30A and 300A ranges.
Example
CURR:RANG AUTO
Sets the current range to AUTO.
Note
The range that is returned is the base unit. See the table below:
Unit
Current Range
100
300A
10
30A
1
3A
.1
300mA
.01
30mA
Sense Commands
69
PCS-1000 User Manual
[SENSe:]CURRent:DC:AVERage:COUNt
Set
Query
Description
This query will set or return average count setting for DC current.
Syntax
Query Syntax
[SENSe:]CURRent:DC:AVERage:COUNt (NR1)
[SENSe:]CURRent:DC:AVERage:COUNt?
Parameter /
Return Parameter
<NR1>
The average count setting for DC current.
1~10, 20, 30, 40, 50, 60, 70, 80, 90, 100
Query Example
CURR:DC:AVER:COUN?
>10
The average count setting for DC current is 10.
[SENSe:]CURRent:AC:AVERage:COUNt
Set
Query
Description
This query will set or return average count setting for AC current.
Syntax
Query Syntax
[SENSe:]CURRent:AC:AVERage:COUNt (NR1)
[SENSe:]CURRent:AC:AVERage:COUNt?
Return Parameter
<NR1>
The average count setting for AC current.
1~10, 20, 30, 40, 50, 60, 70, 80, 90, 100
Query Example
CURR:AC:AVER:COUN?
>10
The average count setting for AC current is 10.
[SENSe:]VOLTage:RANGe
Set
Query
Description
Sets or queries the voltage range.
Syntax
Query Syntax
[SENSe:]VOLTage:RANGe {<Range>|AUTO}
[SENSe:]VOLTage:RANGe?
70
COMMUNICATION INTERFACE
Parameter /
Return Parameter
<Range>
Sets the voltage range in volts. The unit will automatically choose the closest range that is programmed.
DC Range <NRf>: 0.0000001 ~ 1050
AC Range <NRf>: 0.0000001 ~ 600
AUTO
Sets the range to AUTO.
Example
VOLT:RANG AUTO
Sets the voltage range to auto.
Note
The range that is returned is the base voltage unit. See the table below:
Unit
Voltage Range
1000
1000VDC
600
600ACV
100
200V
10
20V
1
2V
0.1
200mV
[SENSe:]VOLTage:DC:AVERage:COUNt
Set
Query
Description
This command will set or return the average count setting for DC voltage.
Syntax
Query Syntax
[SENSe:]VOLTage:DC:AVERage:COUNt <NR1>
[SENSe:]VOLTage:DC:AVERage:COUNt?
Parameter /
Return Parameter
<NR1>
The average count setting for DC voltage.
1~10, 20, 30, 40, 50, 60, 70, 80, 90, 100
Query Example
VOLT:DC:AVER:COUN?
>10
The average count setting for DC voltage is 10.
71
PCS-1000 User Manual
[SENSe:]VOLTage:AC:AVERage:COUNt
Set
Query
Description
This query will set or return the average count setting for AC current.
Syntax
Query Syntax
[SENSe:]VOLTage:AC:AVERage:COUNt <NR1>
[SENSe:]VOLTage:AC:AVERage:COUNt?
Return Parameter
<NR1>
The average count setting for AC voltage.
1~10, 20, 30, 40, 50, 60, 70, 80, 90, 100
Query Example
VOLT:AC:AVER:COUN?
>10
The average count setting for AC voltage is 10.
72
COMMUNICATION INTERFACE
SYSTem:BEEPer:STATe .............................................. 73
SYSTem:ERRor ............................................................ 74
SYSTem:LOCal ............................................................ 74
SYSTem:REMote ......................................................... 75
SYSTem:RWLock ........................................................ 75
SYSTem:VERSion ........................................................ 75
SYSTem:OUTPut:FORMat ......................................... 75
SYSTem:BEEPer:STATe
Set
Query
Description
Sets or queries the beeper status.
Syntax
Query Syntax
SYSTem:BEEPer:STATe {0|1}
SYSTem:BEEPer:STATe?
Parameter/ Return Parameter
1
Beeper on
0
Beeper off
Query Example
SYST:BEEP:STAT?
>1
The beeper is on.
System Commands
73
PCS-1000 User Manual
SYSTem:ERRor
Query
Description
Queries the error queue. Error messages are stored in FIFO order. Up to 20 error messages are stored in the error queue. The first error message that is stored is the first message that is returned. Each time a message is returned it is also cleared from the queue. When the error queue is queried and there are no error messages, 0, "No error" will be returned. If the error queue is full (20 messages) and an error occurs, the last-stored error message will be overwritten with the
-350,"Error queue overflow" message. This message will remain, and no additional messages will be stored until it is cleared.
See page 87 for a list of the error messages.
Query Syntax
SYSTem:ERRor?
Return Parameter
<string>
Returns the next error message in the error queue.
Query Example
SYST:ERR?
> 0, "No error."
Returns no error in the error queue.
SYSTem:LOCal
Set
Description
Returns the unit back to local mode. This command will enable all panel keys that may have been locked.
Syntax
SYSTem:LOCal
74
COMMUNICATION INTERFACE
SYSTem:REMote
Set
Description
Sets the PCS-1000 operation to remote mode. All panel keys except the Local key are locked.
Syntax
SYSTem:REMote
SYSTem:RWLock
Set
Description
Sets the PCS-1000 operation to remote mode. All panel keys are locked, including the Local key.
Syntax
SYSTem:RWLock
SYSTem:VERSion
Query
Description
Queries the SCPI version number.
Query Syntax
SYSTem:VERSion?
Return Parameter
<string>
Returns the SCPI version as a string.
Query Example
SYST:VERS?
>1999.0
Returns the SCPI version number as 1999.0.
SYSTem:OUTPut:FORMat
Set
Query
Description
Sets or queries the output formatting. There are 4 types of output formatting: 0, 1, 2, 3.
Format “0” is the default format.
The following table will show how each format will differ from each other when the MEASure? query is used.
Format
Description
Example
0
Returns the output in NR3 format.
+0.0E+0,-4.0E-7
75
PCS-1000 User Manual
1
Returns the output in NR3 format + unit.
+0.0E+0 ADC,-5.0E­7 VDC
2
Returns the output in NR2 format.
+0.00000000,-
0.0000004
3
Returns the output in NR2 format + unit.
+0.00000000 ADC,-
0.0000004 VDC
Syntax
Query Syntax
SYSTem:OUTPut:FORMat (03)
SYSTem:OUTPut:FORMat?
Parameter /
Return Parameter
<NR1>
0~3
Example
SYST:OUTP:FORM?
>3
Returns the format as NR2 + unit.
76
COMMUNICATION INTERFACE
STATus:OPERation:CONDition ................................. 77
STATus:OPERation:ENABle ....................................... 78
STATus:OPERation[:EVENt] ....................................... 78
STATus:PRESet ........................................................... 79
STATus:QUEStionable:CONDition ............................ 79
STATus:QUEStionable:ENABle .................................. 80
STATus:QUEStionable[:EVENt] .................................. 80
STATus:OPERation:CONDition
Query
Description
Returns the contents of the Standard Operation Condition Register.
Bit
Bit weight
Description
0 1 Calibrating
1~3 ~ Not used
4
16
Measuring
5~7 ~ Not used
8
256
Config Change
9~15
~
Not used
Query Syntax
STATus:OPERation:CONDition?
Return Parameter
<NR1>
0~65535: Returns the bit weight of the Standard Operation Condition Register.
Query Example
STAT:OPER:COND?
> 256
Indicates that the configuration has been changed.
Status Commands
77
PCS-1000 User Manual
STATus:OPERation:ENABle
Set
Query
Description
Returns or sets the contents of the Standard Operation Enable Register.
Bit
Bit weight
Description
0 1 Calibrating
1~3 ~ Not used
4
16
Measuring
5~7 ~ Not used
8
256
Config Change
9~15
~
Not used
Syntax
STATus:OPERation:ENABle (065535)
Query Syntax
STATus:OPERation:ENABle?
Parameter / Return Parameter
<NR1>
0~65535: Indicates the bit weight of the Standard Operation Enable Register.
Query Example
STAT:OPER:ENAB 273
Enables bit 0, 4 and 8 of the Standard Operation Enable Register.
STATus:OPERation[:EVENt]
Query
Description
Returns the contents of the Standard Operation Event Register.
Bit
Bit weight
Description
0 1 Calibrating
1~3 ~ Not used
4
16
Measuring
5~7 ~ Not used
8
256
Config Change
9~15
~
Not used
Query Syntax
STATus:OPERation[:EVENt]?
Return Parameter
<NR1>
0~65535: Returns the bit weight of the Standard Operation Event Register.
78
COMMUNICATION INTERFACE
Example
SYST:OPER?
>256
Indicates that bit 8 has been latched.
STATus:PRESet
Set
Description
Resets the Standard Event Enable Register, the Questionable Data Enable Register and the Standard Operation Enable Register to their default state.
Syntax
STATus:PRESet
STATus:QUEStionable:CONDition
Query
Description
Returns the contents of the Questionable Data Condition Register.
Bit
Bit weight
Description
0 1 Volt Overload
1 2 Current Overload
2~15
~
Not used
Query Syntax
STATus:QUEStionable:CONDition?
Return Parameter
<NR1>
0~65535: Returns the bit weight of the Questionable Data Condition Register.
Query Example
STAT:QUES:COND?
> 1
Indicates there was a voltage overload.
79
PCS-1000 User Manual
STATus:QUEStionable:ENABle
Set
Query
Description
Returns or sets the contents of the Questionable Data Enable Register.
Bit
Bit weight
Description
0 1 Volt Overload
1 2 Current Overload
2~15
~
Not used
Syntax
STATus:QUEStionable:ENABle (065535)
Query Syntax
STATus:QUEStionable:ENABle?
Parameter / Return Parameter
<NR1>
0~65535: Indicates the bit weight of the Questionable Data Enable Register.
Query Example
STAT:QUES:ENAB 3
Enables bit 1and 2 of the Questionable Data Enable Register.
STATus:QUEStionable[:EVENt]
Query
Description
Returns the contents of the Questionable Data Event Register.
Bit
Bit weight
Description
0 1 Volt Overload
1 2 Current Overload
2~15
~
Not used
Query Syntax
STATus:QUEStionable[:EVENt]?
Return Parameter
<NR1>
0~65535: Returns the bit weight of the Questionable Data Event Register.
Example
SYST:QUES?
>0
Indicates that no events have been latched.
80
COMMUNICATION INTERFACE
*IDN? .......................................................................... 81
*ESE ............................................................................ 81
*ESR? .......................................................................... 82
*SRE ............................................................................ 82
*STB? .......................................................................... 83
*PSC ............................................................................ 84
*OPC ........................................................................... 84
*TST?........................................................................... 85
*CLS ............................................................................ 85
*RST ............................................................................ 85
*WAI ............................................................................ 85
*IDN?
Query
Description
Returns the manufacturer, model number, serial number and software version number.
Query Syntax
*IDN?
Query Example
*IDN?
>GWInstek,PCS-1000,xxxxxxxxx,Vx.xx
*ESE
Set
Query
Description
Returns or sets the contents of the Standard Event Enable Register.
Bit
Bit weight
Description
0 1 Operation Complete
1 2 Not used
2 4 Query Error
3 8 Device Error
4
16
Execution Error
5
32
Command Error
6
64
Not used
7
128
Power On
Common Commands
81
PCS-1000 User Manual
Syntax
*ESE (0255)
Query Syntax
*ESE?
Parameter / Return Parameter
<NR1>
0~255: Indicates the bit weight of the Standard Event Enable Register.
Query Example
*ESE 189
Enables all bits except for bit 1 and 6.
*ESR?
Query
Description
Queries the Standard Event Register.
Bit
Bit weight
Description
0 1 Operation Complete
1 2 Not used
2 4 Query Error
3 8 Device Error
4
16
Execution Error
5
32
Command Error
6
64
Not used
7
128
Power On
Query Syntax
*ESR?
Parameter
<NR1>
0~255: Indicates the bit weight of the Standard Event Register.
Query Example
*ESR?
>32
Indicates a command error was encountered.
*SRE
Set
Query
Description
Returns or sets the contents of the Service Request Enable Register.
Bit
Bit weight
Description
0 1 Not used
1 2 Not used
82
COMMUNICATION INTERFACE
2 4 ERR: Error queue
3 8 QUES: Questionable Data Register summary bit
4
16
MAV: Message available bit
5
32
ESB: Event summary bit
6 ~ ~
7
128
OPER: Standard Operation Register summary bit
Syntax
*SRE (0255)
Query Syntax
*SRE?
Parameter / Return Parameter
<NR1>
0~255: Indicates the bit weight of the Service Request Enable Register.
Query Example
*SRE? >188
Indicates that bits 2, 3, 4, 5 and 7 are enabled.
*STB?
Query
Description
Queries the Status Byte Register.
Bit Summary
Bit
Bit weight
Description
0 1 Not used
1 2 Not used
2 4 ERR: Error queue
3 8 QUES: Questionable Data Register summary bit
4
16
MAV: Message available bit
5
32
ESB: Event summary bit
6
64
MSS: Master summary bit of the Service Request Register and the Status Byte Register.
7
128
OPER: Operation status register summary bit
Query Syntax
*STB?
Parameter
<NR1>
0~255: Indicates the bit weight of the Status Byte Register.
83
PCS-1000 User Manual
Query Example
*STB?
>4
Indicates that there is a message in the error queue.
*PSC
Set
Query
Description
The Power on Status Clear command enables the unit to clear the Service Request Enable, the Standard Event Enable and other event enable registers at power up.
Syntax
*PSC (0|1)
Query Syntax
*PSC?
Parameter / Return Parameter
0
Disabled
1
Enabled
Query Example
*PSC 0 Disables the clearing of the event registers at power up.
*OPC
Set
Query
Description
The Operation Complete command will set bit 0 of the Standard Event Register when all pending operations are complete. The OPC? query will return 1 when all pending operations are complete.
Syntax
*OPC
Query Syntax
*OPC?
Return Parameter
1
Enabled
Query Example
*OPC?
>1
Indicates that all pending operations are complete.
84
COMMUNICATION INTERFACE
*TST?
Query
Description
Self-test query. This query will initiate a self-test and return the result.
Query Syntax
*TST?
Parameter
0
All tests have passed.
1
One of more tests have failed.
Query Example
*TST?
>0
Indicates that all tests have passed.
*CLS
Set
Description
The Clear Status command will clear the Status Byte Register by clearing the error queue, and register groups that connect to the Status Byte Register with a summary bit.
Syntax
*CLS
*RST
Set
Description
The Reset command will reset the unit to factory default settings.
Syntax
*RST
*WAI
Set
Description
The Wait command will make the unit wait until all pending operations are complete.
Syntax
*WAI
85
Summary Bit (RQS)
Serial Poll
*STB?
*SRE
*ESE*ESR?
Condition
Calibrating 0
Event Enable
Not Used 1
Not Used 2
Not Used 3
Measuring 4
Not Used 5
Not Used 6
Not Used 7
Config Change 8
Not Used 9
Not Used 10
Not Used 11
Not Used 12
Not Used 13
Not Used 14
Not Used 15
0 1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
0 1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
OPC (Operation Complete)
Not Used
QUE (Query Error)
Device Error
EXE (Execution Error)
CME (Command Error)
Not Used
PON (Power On)
0
1
2
3
4
5
6
7
0 1 2 3 4 5 6 7
0 1 2 3 4 5 6 7
ERR
QUES
MAV ESB MSS
OPER
0 1 2 3 4 5 6 7
Output
Buffer
: : :
Error Que
: : :
EnableEvent
Status
Byte
Register
Service
Request
Enable
Register
Standard Operation Register
Standard Event Register
Event Enable
0 1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
0 1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
Questionable Data Register
Condition
Voltage Overload 0
Current Overload 1
Not Used 2 Not Used 3 Not Used 4 Not Used 5 Not Used 6 Not Used 7 Not Used 8 Not Used 9 Not Used 10 Not Used 11 Not Used 12 Not Used 13 Not Used 14 Not Used 15
STAT:QUES:COND?
STAT:QUES:EVEN?
STAT:QUES:ENAB
STAT:QUES:ENAB?
*ESE?
STAT:OPER:COND?
STAT:OPER:EVEN?
STAT:OPER:ENAB
STAT:OPER:ENAB?
*SRE?
PCS-1000 User Manual
Status Registers
86
COMMUNICATION INTERFACE
Command Errors
0,"No error"
-101,"Invalid character"
-102,"Syntax error"
-103,"Invalid separator"
-108,"Parameter not allowed"
-109,"Missing parameter"
-113,"Undefined header"
-121,"Invalid character in number"
-123,"Numeric overflow"
-131,"Invalid suffix"
-148,"Character data not allowed"
-151,"Invalid string data"
Execution Errors
-222,"Data out of range"
-224,"Illegal parameter value"
Device Specific Errors
-300,"Device-specific error"
-330,"Self-test failed"
-350,"Error queue overflow"
Query Errors
-410,"Query INTERRUPTED"
-420,"Query UNTERMINATED"
-521,"Input buffer overflow"
-522,"Output buffer overflow"
Error Messages
87
PCS-1000 User Manual
Initial Settings
Default Setting
Current Meter
DCA
Voltage Meter
DCV
Current Range
Auto (Auto range only for 30mA, 300mA, 3A)
Voltage Range
Auto
Baud rate
9600
GPIB address
08
AD Speed
7 seconds ( 6½ digits)
AVG Mode
Shift
DCV AVG
10 (samples)
ACV AVG
10 (samples)
DCA AVG
10 (samples)
ACA AVG
10 (samples)
Autozero
Enable
Beeper
On
0 1 2 3 4 5 6 7 8 9 A B C
D
E F G H I J K L M N O P Q
R
S T U V W X Y Z ( ) + - ,
APPENDIX
PCS Default Settings
The following default settings are the factory configuration settings when the unit first ships. See page 34 to restore the factory default settings.
LED ASCII Table Character Set
Use the following table to read the LED display messages.
88
APPENDIX
Power Supply
100 V / 120 V / 220 V / 240 V ±10%
Power Line Frequency
50/60Hz
Operating Environment
Full accuracy for 0˚C to 50˚C, Full accuracy to 80% R.H. at 40˚C
Storage Environment
-40˚C to 70˚C
Power Consumption
Max 35VA
Dimensions
210mm (W) * 80mm (H) * 390mm (D)
Weight
Approximately 6 kg
DC Voltage
Range
1 Year 23˚C ± 5˚C
Temperature Coefficient/˚C
200.0000 mV
0.0050 + 0.0035
0.0005 + 0.0005
2.000000 V
0.0050 + 0.0010
0.0005 + 0.0001
20.00000 V
0.0050 + 0.0010
0.0005 + 0.0001
200.0000 V
0.0050 + 0.0010
0.0005 + 0.0001
1000.000 V
0.0050 + 0.0020
0.0005 + 0.0001
Accuracy specification : ± (% of reading + % of range) Voltage input Resistance: 10M for all DC Voltage ranges
DC Current Range
Burden Voltage
1 Year 23˚C ± 5˚C
Temperature Coefficient/˚C
30.00000 mA
<0.4 V
0.01 + 0.005
0.001 + 0.002
300.0000 mA
<0.5 V
0.01 + 0.005
0.001 + 0.002
3.000000 A
<0.8 V
0.01 + 0.005
0.001 + 0.002
30.00000 A
<0.8 V
0.01 + 0.005
0.001 + 0.002
300.0000 A
<0.8 V
0.02 + 0.005
0.001 + 0.002
Accuracy specification : ± (% of reading + % of range)
PCS-1000 Specifications
The specifications apply when the PCS is powered on for at least 30 minutes.
General
DC Characteristics
89
PCS-1000 User Manual
True RMS AC Voltage
Range
Frequency
1 Year 23˚C ± 5˚C
Temperature Coefficient/˚C
200.0000 mV
45 Hz - 2 kHz 2 kHz - 10 kHz 10 kHz - 20 kHz
0.5 + 0.05
1.0 + 0.05
2.0 + 0.10
0.005 + 0.005
2.000000 V
0.005 + 0.005
20.00000 V
0.005 + 0.005
200.0000 V
0.005 + 0.005
600.000 V
0.005 + 0.005
Accuracy specification : ± (% of reading + % of range)
True RMS AC Current
Range
Frequency
1 Year 23˚C ± 5˚C
Temperature Coefficient/˚C
(AC+DC)
30.00000 mA
45 Hz - 2 kHz 2 kHz - 10 kHz
0.5 + 0.05
1.0 + 0.05
0.03 + 0.006
300.0000 mA
0.03 + 0.006
3.000000 A
0.03 + 0.006
30.00000 A
45 Hz - 400 Hz
0.5 + 0.05
0.03 + 0.006
300.0000 A
0.03 + 0.006
Accuracy specification : ± (% of reading + % of range)
Range
Shunt Value
DC Accuracy
AC Accuracy 400 Hz
Max Input DC/AC rms
30.00000 mA
10
0.01%
0.1%
30 mA
300.0000 mA
1
0.01%
0.1%
300 mA
3.000000 A
0.1
0.01%
0.1%
3 A
30.00000 A
0.01
0.01%
0.1%
30 A
300.0000 A
0.001
0.02%
0.1%
300 A
AC Characteristics
Current Monitor Accuracy
90
APPENDIX
GPIB
300A MAX.
50 / 60 Hz 35VA MAX.
LINE RATING
FUSE LINE
100VAC 120VAC
220VAC 240VAC
T200mA
SER.NO. LABEL
AND TEST LEADS BEFORE
DISCONNECT POWER CORD
REPLACING FUSE
INPUT
T100mA
( NO FUSE )
18
7.8
51
25.6
250
351.3
386
25
Precision Current Shunt
32
213
150
15.4
12.4
80
48
255
GPIB
300A MAX.
50 / 60 Hz 35VA MAX.
LINE RATING
FUSE LINE
100VAC
120VAC
220VAC
240VAC
T200mA
SER.NO. LABEL
AND TEST LEADS BEFORE DISCONNECT POWER CORD
REPLACING FUSE
INPUT
T100mA
( NO FUSE )
Local AC / DC 300A / 30A 3A Range
Func
AC / DC Range
RMT ACA 30A Auto
mA
A
CURRENT
ACV DCV Auto
mV
V
VOLTAGE
Precision Current Shunt
PCS-1000
: Long Push
Auto
DC 1000V AC 600V MAX
Current Monitor
3A MAX
30A MAX
DCA 300A ERROR
CAT II 600V
Func Select
Auto
Enter
( No Fuse ) ( Fused )
POWER
INPUT INPUT OUTPUT INPUT
PCS Dimensions
scale = mm.
91
PCS-1000 User Manual
Type of Product: Digital Current and Voltage Meter
Model Number: PCS-1000
EMC
EN 61326-1: EN 61326-2-1:
Electrical equipment for measurement, control and
laboratory use –– EMC requirements (2013)
Conducted & Radiated Emission EN 55011: 2009+A1:2010
Electrostatic Discharge EN 61000-4-2: 2009
Current Harmonics EN 61000-3-2: 2006+A1: 2009+A2: 2009
Radiated Immunity EN 61000-4-3: 2006+A1:2008+A2:2010
Voltage Fluctuations EN 61000-3-3: 2008
Electrical Fast Transients IEC 61000-4-4: 2012
-------------------------
Surge Immunity EN 61000-4-5: 2006
-------------------------
Conducted Susceptibility EN 61000-4-6: 2009
-------------------------
Power Frequency Magnetic Field EN 61000-4-8: 2010
-------------------------
Voltage Dip/ Interruption EN 61000-4-11: 2004
Low Voltage Equipment Directive 2006/95/EC
Safety Requirements
EN 61010-1: 2010 EN 61010-2-030: 2010
Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No. 7-1, Jhongsing Rd, Tucheng Dist., New Taipei City 236, Taiwan
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
No. 69 Lushan Road, Suzhou New District Jiangsu, China. declare that the below mentioned product
are herewith confirmed to comply with the requirements set out in the Council Directive on the Approximation of the Law of Member States relating to Electromagnetic Compatibility (2004/108/EC) and Low Voltage Directive (2006/95/EC). For the evaluation regarding the Electromagnetic Compatibility and Low Voltage Directive, the following standards were applied:
92
INDEX
INDEX
AC/DC Current ......................... 25
AC/DC Voltage ......................... 26
Accessories ................................. 10
AD speed .................................... 37
Autozero function ...................... 40
Average mode ............................ 38
Average number settings .......... 39
Baudrate ...................................... 35
Beeper settings ........................... 41
Caution symbol ............................ 4
Cleaning the instrument ............. 5
Connections ................................ 22
Current monitor ......................... 30
Current range selection ............. 25
Declaration of conformity ......... 92
Default settings .................... 34, 88
Dimensions ................................. 91
Display format ........................... 88
Disposal instructions ................... 6
EN61010
measurement category .................. 5
pollution degree.............................. 6
Environment
safety instruction ............................ 6
Error messages ........................... 87
Front panel diagram .................. 11
Function menu ........................... 31
GPIB address .............................. 36
GPIB function check .................. 45
Ground
symbol .............................................. 4
LCD conversion .......................... 88
List of features .............................. 9
Overview ....................................... 9
Power on/off
safety instruction ............................ 5
Power up ..................................... 19
Rack mount ................................. 20
Rear panel ................................... 16
Remote control ........................... 42
Command list ................................ 59
Command syntax ......................... 56
GPIB configuration....................... 44
return to local mode ..................... 55
USB configuration ........................ 50
Restore default settings ............. 34
Service operation
about disassembly .......................... 5
Software version ........................ 33
Specifications .............................. 89
Status Registers .......................... 86
UK power cord ............................. 7
USB driver installation .............. 48
USB function check .................... 53
Voltage range selection ............. 27
Warning symbol ........................... 4
Waveform type crest factor
comparison ............................. 29
Waveform type voltage
comparison ............................. 28
Wire gauge chart ........................ 21
93
Loading...