GW Instek MDO-2000A Series, MDO-2000AG Series, MDO-2202AG, MDO-2302A, MDO-2302AG User Manual

...
Mixed-Domain Oscilloscope
MDO-2000A Series
USER MANUAL
ISO-9001 CERTIFIED MANUFACTURER
June 2019
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan
TABLE OF CONTENTS
Table of Contents
SAFETY INSTRUCTIONS .................................................. 3
GETTING STARTED .......................................................... 8
MDO-2000A Series Overview ....................................... 9
Appearance ................................................................ 12
Set Up ....................................................................... 23
Built-in Help .............................................................. 32
MEASUREMENT ............................................................ 33
Basic Measurement ................................................... 34
Automatic Measurement ............................................ 41
Cursor Measurement ................................................. 55
Math Operation ......................................................... 62
ADVANCED CONFIGURATION ....................................... 72
Acquisition ................................................................ 75
Segmented Memory Acquisition ................................ 80
Display ...................................................................... 92
Horizontal View ......................................................... 97
Vertical View (Channel) .............................................105
Bus Key Configuration ..............................................113
Trigger ......................................................................131
Search.......................................................................162
System Settings and Miscellaneous Settings ............171
ARBITRARY WAVE GENERATOR .................................... 176
Arbitrary Wave Generator Operation .........................177
SPECTRUM ANALYZER .................................................. 210
Spectrum Analyzer operation ....................................211
1
MDO-2000A series User Manual
APPLICATIONS ............................................................. 228
Introduction ............................................................. 229
Go-NoGo application ............................................... 232
DVM application ...................................................... 237
Data Log application ................................................ 239
Digital Filter application .......................................... 241
Mask application ..................................................... 243
FRA application ....................................................... 253
SAVE/RECALL ............................................................... 268
File Format/Utility ................................................... 269
Create/Edit Labels .................................................... 275
Save ......................................................................... 278
Recall ....................................................................... 285
Reference Waveforms ............................................... 291
FILE UTILITIES ............................................................. 293
HARDCOPY KEY ........................................................... 299
REMOTE CONTROL CONFIG ........................................ 303
Interface Configuration ............................................ 304
MAINTENANCE ............................................................ 314
FAQ .............................................................................. 320
APPENDIX .................................................................... 322
Updating the Firmware ............................................ 323
MDO-2000A Series Specifications ............................ 325
Probe Specifications ................................................ 330
Dimensions ............................................................. 332
Declaration of Conformity ........................................ 333
INDEX .......................................................................... 334
2
SAFETY INSTRUCTIONS
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the equipment or to other properties.
DANGER High Voltage
(Note)
Attention required Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to insure your safety and to keep the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the Product name.
3
MDO-2000A series User Manual
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
General Guideline
WARNING
CAUTION
Make sure the BNC input voltage does not
exceed 300Vrms.
Never connect a hazardous live voltage to the
ground side of the BNC connectors. It might lead to fire and electric shock.
Do not place any heavy object on the MDO-
2000A series.
Avoid severe impact or rough handling that
leads to damaging the MDO-2000A series.
Do not discharge static electricity to the MDO-
2000A series.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening. Do not perform measurement at a power source
or building installation (Note below).
Do not disassemble the MDO-2000A series
unless you are qualified.
Note
(Measurement categories) EN 61010-1:2010 specifies the measurement categories and their requirements as follows. The MDO-2000A series falls under category I.
Measurement category IV is for measurements performed at the
source of low-voltage installation.
Measurement category III is for measurements performed in the
building installation.
Measurement category II is for measurements performed on
circuits directly connected to the low voltage installation.
Measurement category I is for measurement performed on
circuits not directly connected to Mains.
Safety Guidelines
4
SAFETY INSTRUCTIONS
Power Supply
WARNING
AC Input voltage: 100 - 240V AC, 50 - 60Hz,
auto selection. Power consumption: 40W for MDO-2000A series.
Connect the protective grounding conductor of
the AC power cord to an earth ground, to avoid electrical shock.
Cleaning the MDO-2000A Series
Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh materials
such as benzene, toluene, xylene and acetone.
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: ≤80%, 40°C or below; 45%,
41°C ~ 50°C
Altitude: < 2000m Temperature: 0°C to 50°C
Note
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The MDO-2000A series falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
5
MDO-2000A series User Manual
Storage environment
Location: Indoor Temperature: -10°C to 60°C Humidity: Up to 93% RH (non-condensing) /
≤40ºC, up to 65% RH (non-condensing) / 41ºC ~ 60 ºC
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
6
SAFETY INSTRUCTIONS
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the oscilloscope in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code:
As the colours of the wires in main leads may not correspond with the coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, a cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable, any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced in accordance to the above standard.
7
MDO-2000A series User Manual
MDO-2000A Series Overview ....................................... 9
Integrated instruments and series lineup .............................................. 9
Main Features ............................................................................................ 9
Accessories ............................................................................................... 10
Appearance ................................................................ 12
Front panel .............................................................................................. 12
Rear Panel ................................................................................................ 19
MDO-2000A series ................................................................................... 19
MDO-2000AG series ................................................................................ 19
Display ...................................................................................................... 20
Set Up ........................................................................ 23
Tilt Stand .................................................................................................. 23
Power Up ................................................................................................. 23
First Time Use ......................................................................................... 24
How to Use This Manual ...................................................................... 26
Built-in Help .............................................................. 32
GETTING STARTED
This chapter describes the MDO-2000A series in a nutshell, including its main features and front / rear panel. After going through the overview, follow the Set Up section to properly set up the device for first time use. The Set Up section also includes an introduction on how to use this manual effectively.
8
GETTING STARTED
Model name
AWG
(1)
SA
(2)
MDO-2000A
MDO-2000AG
Model name
Frequency bandwidth
Input channels
Max. Real-time Sampling Rate
MDO-2102A/AG
100MHz
2
2GSa/s
MDO-2202A/AG
200MHz
2
2GSa/s
MDO-2302A/AG
300MHz
2
2GSa/s
Features
8 inch, 800 x 480, WVGA TFT display. Available from 100MHz to 300MHz. Real-time sampling rate of 2GSa/s max. Deep memory: 20M points record length. Waveform capture rate of 120,000 waveforms
per second.
Vertical sensitivity: 1mV/div~10V/div.
MDO-2000A Series Overview
Integrated instruments and series lineup
The MDO-2000A series have different hardware implemented instruments:
(1) Arbitrary Wave Generator (2) Spectrum Analyzer
The MDO-2000A series consists of 6 models. Note that throughout the user manual, the term “MDO-2000A series” refers to all models of the series, unless stated otherwise.
Main Features
9
MDO-2000A series User Manual
Segmented Memory: Optimizes the acquisition
memory to selectively capture only the important signal details. Up to 29,000 successive waveform segments can be captured with a time-tag resolution of 4ns.
Waveform Search: Allows the scope to search
for a number of different signal events.
Arbitrary Wave generator: Full-function dual
channel arbitrary waveform generator.
Spectrum Analyzer: A handy tool to perform
signal analysis in the frequency domain.
Powerful embedded applications such as: Data
Logging, Digital Voltmeter, Go-No Go, Mask, Digital filter, FRA etc..
On-screen Help. 32 MB internal flash disk.
Interface
USB host port: front panel, for storage devices. USB device port: rear panel, for remote control
or printing.
Ethernet port as standard. Probe compensation output with selectable
output frequency (1kHz ~ 200kHz).
Calibration output.
Standard Accessories
Part number
Description N/A region dependent
Power cord
GTP-100B-4, for MDO­2102A/AG
Passive probe; 100 MHz
GTP-200B-4, for MDO­2202A/AG
Passive probe; 200 MHz
Accessories
10
GETTING STARTED
GTP-300B-4, for MDO-2302A/AG
Passive probe; 300 MHz
GTL-110
Test lead for AWG.
Standard Apps
Name
Description
Go-NoGo
Go-NoGo testing app.
DataLog
Waveform or image data logging app.
DVM
Digital Voltmeter app.
Digital Filter
High ,low or band pass digital filter for analog inputs.
Mask
Creates shape templates for signal comparison.
Remote Disk
network share drive.
Demo mode
with the GDB-03 demo board.
FRA
(For MDO-2000AG series only)
Optional Accessories
Part number
Description GDB-03
Demo board.
Drivers, others
USB driver
LabVIEW driver
11
Appearance
LCD
Variable knob and Select key
Autoset, Run/Stop, Single and Default settings
CH1~CH2, EXT TRIG
Trigger controls
Function keys
USB Host port, Probe Compensation terminals
Hardcopy key
Math, Reference and Bus keys
Bottom menu keys
Vertical controls
Option key
Menu key
Side menu keys
Power button
Horizontal controls
EXT TRIG
LCD Display
8” WVGA TFT color LCD. 800 x 480 resolution,
wide angle view display.
Menu Off Key
Use the Menu Off key to hide the onscreen menu system.
Option Key
The Option key is used to access installed options.
Front panel
MDO-2000A series User Manual
12
GETTING STARTED
Menu Keys
The side menu and bottom menu keys are used to make selections from the soft-menus on the LCD user interface.
To choose menu items, use the 7 Bottom menu keys located on the bottom of the display panel.
To select a variable or option from a menu, use the side menu keys on the side of the panel. See page 26 for details.
VARIABLE
POSITION
HORIZONTAL
TRIGGER
LEVEL
PUSH TO
ZERO
PUSH TO
ZERO
SCALE
Measure
Display
Cursor
Help
Test
Save/Recall
Acquire
Utility
Autoset
Run/Stop
Single
Default
Select
Search
Set/Clear
Zoom
Menu
50 %
Force-Trig
POSITION
POSITION
VERTICAL
MATH
REF
BUS
SCALESCALE
PUSH TO
ZERO
PUSH TO
ZERO
M
R
B
Digital Storage Oscilloscope
GDS-2202E
200 MHz 1 GS/s
Visual Persistence Oscilloscope
Side menu keys
Bottom menu keys
Hardcopy Key
The Hardcopy key is a quick-save or quick-print key, depending on its configuration. For more information see pages 300 (save) or 300 (print).
Variable Knob and Select Key
Select
VARIABLE
The Variable knob is used to increase/decrease values or to move between parameters.
The Select key, when lit up, is used to make selections.
13
MDO-2000A series User Manual
Function Keys
The Function keys are used to enter and configure different functions on the MDO-2000A series.
Measure
Measure
Configures and runs automatic measurements.
Cursor
Cursor
Configures and runs cursor measurements.
APP
APP
Configures and runs GW Instek applications.
Acquire
Acquire
Configures the acquisition mode, including Segmented Memory acquisition.
Display
Display
Configures the display settings.
Help
Help
Shows the Help menu.
Save/Recall
Save/Recall
Used to save and recall waveforms, images, panel settings.
Utility
Utility
Configures the Hardcopy key, display time, language, probe compensation and calibration. It also accesses the file utilities menu.
Autoset
Autoset
Press the Autoset key to automatically set the trigger, horizontal scale and vertical scale.
14
GETTING STARTED
Run/Stop Key
Run/Stop
Press to Freeze (Stop) or continue (Run) signal acquisition (page 37). The run stop key is also used to run or stop Segmented Memory acquisition (page 83).
Single
Single
Sets the acquisition mode to single triggering mode.
Default Setup
Default
Resets the oscilloscope to the default settings.
Horizontal Controls
The horizontal controls are used to change the position of the cursor, set the time base settings, zoom into the waveforms/traces and search for events.
Horizontal Position
POSITION
PUSH TO
ZERO
The Position knob is used to position the waveforms/traces horizontally on the display screen. Pressing the knob will reset the position to zero.
SCALE
SCALE
The Scale knob is used to change the horizontal scale (TIME/DIV).
Zoom
Zoom
Press Zoom in combination with the horizontal Position knob.
Play/Pause
The Play/Pause key allows you to view each search event in succession – to effectively “play” through each search event. It is also used to play through a waveform/trace in zoom mode.
15
MDO-2000A series User Manual
Search
Search
The Search key accesses the search function menu to set the search type, source and threshold.
Search Arrows
Use the arrow keys to navigate the search events.
Set/Clear
Set/Clear
Use the Set/Clear key to set or clear points of interest when using the search function.
Trigger Controls
The trigger controls are used to control the trigger level and options.
Level Knob
LEVEL
Used to set the trigger level. Pressing the knob will reset the level to zero.
Trigger Menu Key
Menu
Used to bring up the trigger menu.
50% Key
50 %
Sets the trigger level to the half way point (50%).
Force - Trig
Force-Trig
Press to force an immediate trigger of the waveform.
Vertical POSITION
POSITION
PUSH TO
ZERO
Sets the vertical position of the waveform/trace. Push the knob to reset the vertical position to zero.
Channel Menu Key
CH1
Press the CH1~4 key to set and configure the corresponding channel.
16
GETTING STARTED
(Vertical)SCALE Knob
SCALE
Sets the vertical scale of the channel (TIME/DIV).
External Trigger Input
EXT TRIG
Accepts external trigger signals (page 131). Only on 2 channel models.
Input impedance: 1MΩ Voltage input: ±15V (peak), EXT trigger capacitance: 16pF.
Math Key
MATH
M
Use the Math key to set and configure math functions.
Reference Key
REF
R
Press the Reference key to set or remove reference waveforms.
BUS Key
BUS
B
The Bus key is used for serial bus (UART, I2C, CAN, LIN) configuration.
Channel Inputs
CH1
Accepts input signals. Input impedance: 1MΩ. Capacitance: 16pF CAT I
USB Host Port
2V
Type A, 1.1/2.0 compatible. Used for data transfer.
17
MDO-2000A series User Manual
Ground Terminal
2V
Accepts the DUT ground lead for common ground.
Probe Compensation Output
2V
The probe compensation output is used for probe compensation. It also has an adjustable output frequency.
By default this port outputs a 2Vpp, square wave signal at 1kHz for probe compensation.
Please see page 173 for details.
Power Switch
POWER
Used to turn the power on/off.
: ON
: OFF
18
GETTING STARTED
Calibration output
Key lock Power input socket
Fan
Go/No Go output
LAN portUSB Device port
Calibration output
Key lock Power input socket
Fan
Go/No Go output
LAN portUSB Device port
AWG channels output
Calibration Output
Outputs the signal for vertical scale accuracy calibration (page 315).
Rear Panel
MDO-2000A series
MDO-2000AG series
19
MDO-2000A series User Manual
USB Device Port
The USB Device port is used for remote control.
LAN (Ethernet) Port
The LAN port is used for remote control over a network or when combined with the Remote Disk app, allows the scope to be mounted to a share disk.
Power Input Socket
Power cord socket accepts AC mains, 100 ~ 240V, 50/60Hz.
For power up sequence, see page
23.
Security Slot
Kensington security slot compatible.
Go-No Go Output
Outputs Go-No Go test results (page 232) as a 500us pulse signal.
AWG Output
GEN 1
Output the GEN1 or GEN2 signal from the Arbitrary Wave Generator function. (see page 178).
Display
Below is a general description of the main display. As the display changes while activating the different functions of the MDO­2000A/AG, please refer to each function sub-chapters of this user manual for more details.
20
GETTING STARTED
Memory bar
Reference
waveform
Analog
Waveform
Bus
decoding
Channel status Horizontal status
Trigger configuration
Waveform
frequency
Date and time
Trigger position
Trigger status
Acquisition mode
Trigger level
Channel Indicators
Memory length
and sample rate
Analog Waveforms
Shows the analog input signal waveforms.
Channel 1: Yellow
Channel 2: Blue
Bus decoding
Shows serial bus data decoding. The values are displayed in hex or binary.
Reference waveform
Reference waveform(s) can be displayed for reference, comparison or other operations.
Channel Indicators
The channel indicators for each activated channels are located at the zero volt level of each signals. Any active channel is shown with a solid color.
Example:
Bus indicator(B)
Reference waveform indicator
Math indicator
Trigger Position
Shows the position of the trigger.
Horizontal Status
Shows the horizontal scale and position.
Date and Time
Current date and time (page 173).
Trigger Level
Shows the trigger level on the graticule.
Memory Bar
21
MDO-2000A series User Manual
The ratio and the position of the displayed waveform compared to the internal memory (page
97).
Trigger Status
Trig’d
Triggered.
PrTrig
Pre-trigger.
Trig?
Not triggered, display not updated.
Stop
Trigger stopped. Also appears in Run/Stop (page 37).
Roll
Roll mode.
Auto
Auto trigger mode.
For trigger details, see page 131.
Acquisition Mode
Normal mode
Peak detect mode
Average mode
For acquisition details, see page 75.
Signal Frequency
Shows the trigger source frequency.
Indicates the frequency is less than 2Hz (lower frequency limit).
Trigger Configuration
Trigger source, slope, voltage and coupling.
Horizontal Status
Horizontal scale, horizontal position.
For trigger details, see page 131.
Channel Status
Channel 1, DC coupling, 2V/Div.
For channel details, see page 105.
22
Set Up
Tilt
To tilt, pull the legs forward, as shown below.
Stand
To stand the scope upright, push the legs back under the casing as shown below.
Requirements
The MDO-2000A series accepts line voltages of 100 ~ 240V at 50 or 60Hz.
Step
1. Connect the power cord to
the rear panel socket.
Tilt Stand
GETTING STARTED
Power Up
23
2. Press the POWER key. The
display becomes active in ~ 30 seconds.
: ON
: OFF
POWER
Note
The MDO-2000A series recovers the state right before the power is turned OFF. The default settings can be recovered by pressing the Default key on the front panel. For details, see page 285.
First Time Use
Background
This section describes how to connect, adjust the scale and compensate the probe. Before operating the MDO-2000A series in a new environment, run these steps to make sure the instrument performs at its full potential.
1. Power On
Follow the procedures on the previous page.
2. Firmware
Update to the latest firmware.
Page 323
3. Set the Date and Time
Set the date and time.
Page 173.
4. Reset System
Reset the system by recalling the factory settings. Press the Default key on the front panel. For details, see page 285.
Default
5. Connect the probe
Connect the probe that you will use for measurements to the Channel 1 input and to the probe compensation output. This output provides by default a 2V peak to peak, 1kHz square wave for signal compensation.
Set the probe attenuation to x10 if the probe has adjustable attenuation.
MDO-2000A series User Manual
24
GETTING STARTED
EXT TRIG
x1
x10
X 10
X 1
CH1
2V
6. Capture Signal (Autoset)
Press the Autoset key. A square waveform appears on the center of the screen. For Autoset details, see page 35.
Autoset
7. Select Vector Waveform
Press the Display key, and set the display to Vector on the bottom menu.
Display
25
MDO-2000A series User Manual
8. Compensate the probe
Turn the adjustment point on the probe to make the square waveform as flat as possible.
9. Start operations
Continue with the other operations.
Measurement: page 33
Configuration: page 72
Using the Spectrum Analyzer: page 210.
Applications: page 228 Save/Recall: page 268
File Utilities: page 293
Hardcopy key: page 299
Remote Control: page 303
Maintenance: page 314
Background
This section describes the conventions used in this manual to operate the MDO-2000A series.
Throughout the manual any reference to pressing a menu key refers to the keys directly below or beside any menu icons or parameters.
Under
Compensation
Normal
Over
Compensation
How to Use This Manual
26
GETTING STARTED
When the user manual says to “toggle” a value or
parameter, press the corresponding menu item. Pressing the item will toggle the value or parameter.
Active parameters are highlighted for each menu item. For example in the example below, Coupling is currently set to DC.
If a menu item can be toggled from one value or parameter to another, the available options will be visible, with the current option highlighted. In the example below the slope can be toggled from a rising slope to a falling slope or either slope.
Menu item
Parameter
Menu item
Active
parameter
Optional
parameters
Menu item
Selecting a Menu Item, Parameter or Variable
When the user manual says to “select” a value
from one of the side menu parameters, first press the corresponding menu key and use the Variable knob to either scroll through a parameter list or to increase or decrease a variable.
Example 1
1
2
3
27
MDO-2000A series User Manual
3. Press a bottom menu key to access
the side menu.
4. Press a side menu key to either set
a parameter or to access a sub menu.
5. If accessing a sub menu or setting
a variable parameter, use the Variable knob to scroll through menu items or variables. Use the Select key to confirm and exit. The Select key is lit-up when such selection can be made.
Select
VARIABLE
6. Press the same bottom menu key
again to reduce the side menu.
Example 2
For some variables, a circular arrow icon indicates that the variable for that menu key can be edited with the Variable knob.
1
7. Press the desired menu key to select it. The
circular arrow will become highlighted.
28
Loading...
+ 309 hidden pages