GW Instek LCR-800 Series, LCR-817, LCR-827, LCR-829, LCR-826 User Manual

...
LCR-800 Series User Manual
i
CONTENTS PAGE
1. INTRODUCTION……………………………………..……... 1
2. PRECAUTIONS BEFORE OPERATION…….……………. 2-1.Unpacking the instrument………………….……………. 2-2.Checking the Line Voltage…………………..…………… 2-3.Environment……………………………………..……….. 2-4.Equipment Installation, and Operation…………………
2 2 2 3 3
3. PANEL DESCRIPTION…...……………………..………….. 4 OPERATION…………….………………………………...….
4-1.Connects to DUT….………………..………….……..…...
4-2.Start-Up……….....……………………………..………….
4-3.Zeroing……………………………………………………. 4-4.Menu Functions.…..……………………………………… 4-5.Measurement Condition…..……………………………...
8 8 8
8 11 18
4.
5.
SPECIFICATIONS…………………………………………...
35
6. MESSAGE CODE.………………………………………….... 42
7. MAINTENANCE....…………………………………………... 43
8. OPTION 1 (BIN Function for LCR-826/827/829 only)…….. 8-1. BIN Functions for Components Sorting……………..…. 8-2. BIN Setting Conditions:...………………………………..
8-3. BIN Range Sett i ng : Component Sort ing Range.. ....... .....
44 44 46 52
9. OPTION 2 (RS-232, for LCR-816/817/819 only)…………… 61
LCR-800 Series User Manual
ii
SAFETY TERMS AND SYMBOLS
These terms may appear in this manual or on the product:
WARNING : Warning statements ide ntif y condit i on or practice s that could result in injury or loss of life.
CAUTION: Caution statements identify conditions or practices that could result in damage to this product or other property.
The followin g sy mb ols may a p pear in this man ua l or on the product:
DANGER ATTENTION Protective Earth (ground)
High Voltage refer to Manual
Conductor Terminal
Terminal
LCR-800 Series User Manual
iii
FOR UNITED KINGDOM ONLY
NOTE: This lead / applianc e must only be wired by compete nt per sons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the
following c ode:
Green / Yellow: Earth Blue: Neutral Brown: Live (Phase)
As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with the letter E or by the earth symbol
or coloured
Green or Green & Yellow. The wire which is coloured Blue must be connected to the terminal which is
marked with the letter N or coloured Blue or Black. The wire which is coloured Brown must be connected to the terminal
marked with the letter L or P or coloured Brown or Red. If in doubt, consult the instr uctions provided wi th the equipment or contac t
the supplier. This cable/ appliance should be protect ed by a suitably ra ted and approved
HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, cable of 0.75mm
2
should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any moulded mains connector that requires removal /replacement must be destroyed by removal of any fuse & fuse carrier and disposed of immediately, as a plug with bared wires is hazardous if a engaged in live socket. Any re-wiring must be carried out in accordance with the information detailed on this label.
LCR-800 Series User Manual
iv
EC Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No. 95-11, Pao-Chung Rd., Hsin-Tien City, Taipei Hsien, Taiwan
declares that the below mentioned product
LCR-817/819/827/829/816/826
are herewith confirmed to comply with the requirements set out in the Council Directive on the Approximation of the Law of Member States relating to Electromagnetic Compatibility (89/366/EEC, 92/31/EEC, 93/68/EEC) and Low Voltage Equipment Directive (73/23/EEC). For the evaluation regarding the Electromagnetic Compatibility and Low Voltage Equipment Directive, the following standards were applied:
EN 61326-1: Electrical equipment for measurement, control and laboratory use ––
EMC requireme nt s (19 97+A1: 1998)
Conducted and Radiated Emissions EN 5 5011 Gr oup I cla ss A: 1998
Electrostatic Discharge
EN 61000-4-2: 1995 Current Harmonic EN 61000-3-2: 1995
Radiated Immunity
EN 61000-4-3: 1996 Voltage Fluctuation EN 61000-3-3: 1995
Electrical Fast Transients
EN 61000-4-4: 1995
————————————————
Surge Immunity
EN 61000-4-5: 1995
————————————————
Conducted Susceptibility
EN 61000-4-6: 1996
————————————————
Power Frequency Magnetic field
EN 61000-4-8: 1993
————————————————
Voltage Dips/ Interrupts
EN 61000-4-11: 1994
————————————————
Oscillatory Waves
EN 61000-4-12: 1995
Low Voltage Equipment Directive 73/23/EEC & amended by 93/68/EEC
Safety Requirements EN 61010-1: 1993+A2: 1995; IEC 61010-1: 1990+A1: 1992+A2: 1995
LCR-800 Series User Manual
1
1. INTRODUCTION
The pr eci se LCR m eter ser ies ar e au t omat ic, u ser pr ogra mma ble in str umen ts th at pr ovide h igh r elia bilit y and gr eat preci sion for mea sur ing a wi de var iet y of impedance parameters. The frequency range of LCR-819/829 covers from 12Hz to 100kHz, LCR-817/827 from 12Hz to 10kHz, and LCR-816/826 from 100Hz to 2kHz. The basic accuracy is 0.1% for LCR827/829, 0.05% for LCR-817/819, and 0.2% for LCR-816/826. The measured results can be displayed on the high quality LCD monitor with decimal points and units. The measured result resolution is five full digits for Inductance (
L), Capa citan ce (C), an d Resista nce
(
R) (four full d igit s for D issi pation (D), Quality factor (Q) or C with R). In the
meantime, the LCD monitor is also shown the control status and parameters of settings. The keypads are easy for menu programming. The LCR Meters’ test fixture (option) is a method of convenient, reliable, guarded 4-terminal/2 wires connection of radial and axial leaded components to the LCR Meters. Test conditions can be stored and recalled from internal memory that will reduce the setup time for measurement preparing.
LCR-800 Series User Manual
2
2. PRECAUTIONS BEFORE OPERATION
2-1. Unpacking the instrument
The product has been fully inspected and tested before shipping from the factory. Upon receiving the instrument, please unpack and inspect it to check if there ar e any damag es ca used durin g tr an spor tat i on. If an y dama g e is found, not ify th e bearer an d /or the dealer im mediately.
2-2. Checking the Line Voltage
The LCR Meters can be operated with AC power source between 100V rated voltage and 240V rated voltage at a frequency of 50 to 60Hz, no AC voltage selector is necessary. Power connection to rear panel is through an AC inl et modul e com p rised of an AC conn ect or and fuse hold er. To change the fus e p roceed a s fol lows:
! Remove the fuse holder by inserting a small flat head screwdriver
behind the small tab to force the holder outward.
! Install the correct fuse (slow-blow, 3A, 250Vac). ! Re-install the fuse holder back into the LCR Meters AC inlet module,
push in and lock.
WARNING. To avoid electrical shock the power cord protective grounding conductor must be connected to ground.
WARNING. To avoid personal injury, disconnect the power cord before removing the fuse holder.
LCR-800 Series User Manual
3
2-3. Environment
The norm al ambien t temperat ure range of th e LCR Meters is fr om 10° to 50°C. To operate th e instrument over this specific temp erat ure r ange may
cause damage to the circuits.
Do not use th e LCR M eters in a place wh ere str ong ma gnetic or el ectric field exists as it may disturb the measurement.
2-4. Equipment Installation, and Operation
Ensure there is proper ventilation for the vents in the LCR meters case. If this equipment is used not according to the specification, the protection provided by the equipment may be impaired.
WARNING : This is a Class A product. In a domestic environment this prduct may cause radio interference in which case the user may be required to take adequate measures.
LCR-800 Series User Manual
4
3. PANEL DESCRIP TI ON
(1). Power S witc h
Turns AC Power on or off.
(2). Function keyF1
Soft key functions as indicated on the adjacent LCD monitor.
(3). Function keyF2
Soft key functions as indicated on the adjacent LCD monitor.
(4). Function keyF3
Soft key functions as indicated on the adjacent LCD monitor.
(5). Function keyF4
Soft key functions as indicated on the adjacent LCD monitor.
(6). MENU key
Enters menu display mode or exits sub menu back to main menu.
(7). Compound key
! For making numerical entries as labeled. ! Turns the “RANGE HOLD” mode on or off.
(8). Compound key
! For making numerical entries as labeled. ! Turns the “CONSTANT VOLTAGE” mode on or off.
(9). Compound key
! For making numerical entries as labeled. ! Measures the unit of Dissipation and Quality Factor in PPM.
(Note: The models of LCR-816 and 826 do not have this function.)
LCR-800 Series User Manual
5
(10). Compound key
! For making numerical entries as labeled.
! Selects th e “INTERNAL BIAS” mode or “EXTE RNAL BIAS” mode. ( If
this key function is switched to “INTERNAL BIAS” mode, the bottom of LCD monitor will display the “INT.B” message. If the external DC bias is selected, the bottom of LCD monitor will display the “EXT.B” message.)
(11). Compound key
! For making numerical entries as labeled. ! Turns the INTERNAL BIAS mode or EXTERNAL BIAS mode “ON” or
“OFF”.
(12). Compound key
! For making numerical entries as ““ (the negative sign). ! Input s th e “TEST FREQUE N C Y”.
(13). START (Compound k ey)
! Starts measurement sequence. Normally used in the “MANU”
(Triggered) mode.
! Selects “AUTO” or “MANU” mode by pressing this key for 3 seconds at
least.
! The LCR Meters will process the measurement automatically, if the
“AUTO” mode is selected .
(14).
↵↵
key (ENTER)
Thi s key ena bles pr ogrammin g of all speci al fun ction s, t est fr equ enc y, t est volt a ge, averaging, delay, and nom in al val u e etc.
(15). Symbol key
In puts the d ecimal p oint
(16). Numeral key—“2” (17). Numeral key—“3” (18). Numeral key—“5” (19). Numeral key—“6” (20). Numeral key—“9” (21). Primary Displa y
This line can display the measured Inductance, Capacitance, or Resistance.
(22). Secondary Display
LCR-800 Series User Manual
6
This line can display the measured Quality Factor or Dissipation or ESR or EPR.
(23). Instrument status or indicates measurement results based on entered test
limits.
(24). Test conditions (25). Input terminals
BNC connect ors, connects to device u nder tes t (DUT) .
Conn ectors of t he LCR Met ers
BIAS
Lforce (current, low)
Lsens e ( p otential low)
Hsen se (potential high )
Hforce (current, high)
LCR-800 Series User Manual
7
FRONT PANEL
LCR-800 Series User Manual
8
4. OPERATION
4-1. Connects to DUT
The LCR Meter s utilize th e stru cture of four wires measur ement which all ows accurate, easy, and stable measurements and avoids mutual inductance and interference from measurement signals, noise and other factors inherent with other types of connections. For the accuracy of measurement, GOODWILL produces the cable set and test fixture (option) for connection directly to the fron t pan el BNC connect ors.
4-2. Start-Up
Conn ects th e power cor d of th e LCR M eter s to th e main s socket- outlet . Presses the
POWER button of front panel to apply the AC power to the LCR Meters.
4-3. Zeroing
In ord er to elimina te stra yed capa citan ce and impedan ce of test ca ble du rin g the meas urement , the LCR Meter s shoul d be zeroed to corr ect for t est cabl e and/ or test fixture errors before taking measurements. The corrections are calculated and stored in memor y of the LCR Met ers dur ing the z eroin g pr ocess. Open and short circu it zeroing should be d one for test cable and/or te s t fixture. For the be s t accur a cy, th e test ca ble an d/ or test fix tur e sh oul d be zer oed on ce p er day a t l east and each time tes t cabl e or test fi xture is changed in general. The z eroing pr ocess of open an d sh ort cir cu its are following:
Open Circuit
! The t es t cable or tes t fixture should be open with n o com p onent connect ed . ! Press MENU key. ! Press F1 key to select “OFFSET” menu. ! Press F1 key to sele ct open cir cu it zeroin g (the “CAP OFFSE T” i s indicated
on the adjacent LCD mon itor.).
LCR-800 Series User Manual
9
! After the BAR at the bottom of LCD monitor is filled to the full, the zeroing
process is done.
! If th e zeroing pr ocess is su ccessful, a messag e of “OK” will appear on the
LCD monitor. If failed, a message of “FAIL” will appear on the LCD monitor.
Short Circuit
! The test cable should be connected or test fixture shorted (using a clean
copper wire, as short as possible).
! Press MENU key. ! Press F1 key to select “OFFSET” menu. ! Press F2 key to select short circuit zeroing (the “R/L OFFSET” is indicated
on the adjacent LCD mon itor.).
! After the BAR at the bottom of LCD monitor is filled fully, the zeroing
process is done.
! If th e zeroing pr ocess is su ccessful, a messag e of “OK” will appear on the
LCD monitor. If failed, a message of “FAIL” will appear on the LCD monitor.
Test Condition:
Test voltage=1V Test speed = SLOW R.H = OFF C.V = OFF
For the summary, the zeroing menu can be chosen through menu selection as shown i n Fi gure 4-1 above.
NOTE: The “Open Circuit” and “Short Circuit” have to pass the test,
otherwise, the accuracy of the LCR Meters will become worse.
LCR-800 Series User Manual
10
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
L 1.2345H Q 0.6789
TESTING
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
R.H OFF C.V OFF
OFFSET
SORT
SETTING
CALBRAT
EXIT
CAP. R/L OFFSET
SET SORT
SET PARAMETER
CALIBRATION
Press menu
key
Press F1 key
to the zeroing
menu
EXIT
OPEN TEST
SHORT TEST
CAP
OFFSET
R/L
OFFSET
Press F1 KEY for
open test
Press
F2
KEY for
short test
Figure 4-1: Summary of zeroing menu
LCR-800 Series User Manual
11
4-4. Menu Functions
All the LCR Meters’ programmable functions are controlled by the easy to use menu displays. User can enter the menu mode by selecting the
MENU key that
calls up four top level menus, OFFSET, SORT, SETTING and CALBRAT. Each one of these is comprised of a sub menu list whose functions are described in detail below. User can enter one of four functions by pressing the corresponding function key (just adjacent LCD monitor, see figure 4-2).
4-4-1. Primary & Secondary Display
For the LCR Meters, four combinations of two parameters can be measured and displayed simultaneously. One referred to the “Primary Display” (displayed fir st) and the oth er t o the “Secondary Display”. Depen ding on the com ponen t type t he primar y and secondar y displ ay could be L & Q, C & D, C & R or R and Q. The parameter can be chosen by the user through
F3 key as shown in
Figure 4-3.
OFFSET
SORT
SETTING
CALBRAT
EXIT
CAP. R/L OFFSET
SET SORT
SET PARAMETER
CALIBRATION
Press F1 key
to the zeroing
menu
Press F2 key
to the bin sort
menu
Press F3 key
to the utility
menu
Press F4 key
to the
calibration
menu
Press
MENU
key
to exit menu screen
Figure 4-2: four main menu screen
LCR-800 Series User Manual
12
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
L 1.2345H Q 0.6789
TESTING
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
R.H OFF C.V OFF
Press F3 key successively in order to select the different measurement mode (L/Q, C/D, C/Q, R/Q)
Primary
Display
Secondary
Display
Figure 4-3. Primary & Secondary display
User can select R/Q for resistor measurement; select L/Q for inductor meas uremen t; select either C/D or C/R for capaci tor measuremen t.
4-4-2. Series & Parallel Equivalent Circuit
Impedance that is neither a pure resistance nor a pure reactance can be repr esen t ed at an y speci fic fr eq uen c y by either a s eri es or a p aral lel combi n at ion of resistance and reactance. Such repr esentation is called “equivalent circuit”. The component value of the “Primary Display” depends on which equivalent circuit (series or parallel) is chosen. In normal, the component manufacturer shall specify how a component is to be measured (usually ser ies) and at what frequency.
Suggested Test Conditions:
Inductors le s s t han 10µH: Series, 100kHz.
Inductors from 1 0 µH to 1mH: Series, 10kHz.
Inductors from 1mH to 1H: Se ries, 1kHz .
Inductors greater than 1H: S eri es, 0.1k Hz.
Capacitors less than 10pF: Parallel, 100kHz.
Capacitors from 10 to 400pF: Series or Parallel, 10kHz.
Capacitors from 400 to 1µF: Series, 1 kHz.
Capacitors greater than 1µF: Series, 0.1 or 0.12kHz.
Resistor less than 1kΩ: Series, 1kHz.
Resistor from 1kΩ to 10MΩ: Parallel, 0.25kHz.
Resistor greater th an 10 MΩ: Parallel, 0.03kHz
LCR-800 Series User Manual
13
Unless for special reason, always select “Series” for capacitors and inductors. This has traditionally been standard practice. For very small capacitance or inductance, select a higher test frequency for better accuracy. For very large capa citance or ind uctance, select a lower test fr equency for better accur acy. For dc r esistance, select a l ower test frequency to minimize ac effects . Becau se th e r eactive component most lik ely to be represented in a low resi s tance resi stor is series ind uctance, th e “Seri es” is sel ected for a resist or below about
1k. If a resistor large than 10MΩ, select “Parallel” that because the reactive component most likely to be represent in a high resistance resistor is shunt capa citan ce. If th e Q is less th an 0.1 , the mea sured Rp is proba bly ver y cl ose to the d c r es istan ce. The total loss of a capacitor can be expressed in several ways, including D and “ESR” (Equivalent Series Resistance). “ESR” is typically much larger than actu al “ohm ic” s er ies r esi stan ce of th e wir e l ead s and foil s th at ar e in ser ies wit h the heart of a capacitor physically, because ESR includes also the effect of
dielectric loss. ESR is related to D by the formula: ESR = Rs = D/ωCs. Wher e ω represents “omeg a” = 2 p i time freq u ency. Although it is traditional to measure series inductance of inductors, there are situation s in which the parallel equivalent circuit better represents the physical component. For small “air-core” inductors, the significant loss mechanism is usually “ohmic” or “copper loss” in the wire, therefore the series circuit is appr opr iat e. Never th el ess, for an “ir on cor e” , th e sign i fican t loss mech an is m can be “core loss”, therefore, the parallel equivalent circuit is appropriate which being a better model of the inductor.
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
L 1.2345H Q 0.6789
TESTING
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
R.H OFF C.V OFF
Press F4 key to select the
Series
or
Parallel
circuit
Figure 4 -4. Sel ecti ons of S erie s & Parallel Circui t
4-4-3. Measur em ent Displ ays
LCR-800 Series User Manual
14
The measured results of the LCR Meters can be shown on the LCD monitor in thr e e wa ys: VAL UE
, DELTA%, or DELTA . User can pres s F2 key to select the
appropriate it em for mea s uremen t. VALUE
The LCD monitor will display the measured value of both the primary and secondary parameter, shown with decimal and units. The resolution of primary display (L, C, or R) is five digits. The resolution of secondary display (D, Q or R with C) is four digits. The message “TESTING” is displayed when a test is in process.
LCR-800 Series User Manual
15
DELTA% The “DE LTA%” shows th e percen t deviation of the measur ed L, C, or R va lue from a stored N O MINAL VALUE. The sign of devi at ion is indic at e d.
DELTA The LCR differ en ce is similar to th e DELTA% except tha t the d evia t ion is shown in suitable units (ohms, henries, etc).
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
L 1.2345H Q 0.6789
TESTING
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
R.H OFF C.V OFF
Press F2 key to select the three different measuring display ("VALUE", "DELTA%", or "DELTA").
Figure 4-5. Types of measurement display
4-4-4. Nominal Value
Allows entry of a “Nominal Value” for the primary parameter which is the basis for th e measur ement result in “DELTA” or “DELTA %”. A ccept s n umer ica l entry up to five digits with decimal. Units are depended on which measurement disp lays selected.
LCR-800 Series User Manual
16
Steps of “Nom in al Value” in p ut (Figur e 4- 6 ):
! Press MENU key. ! Press F2 key to select “SORT” menu. ! Press F1 key to select “Nominal Value” (the “NOM.VAL” is indicated on
the adjacent LCD m on itor).
! Input the nominal value via the numeral keys (5 digits with decimal
maximum).
! Press
↵↵
key
! After the BAR at the bottom of LCD monitor is filled fully, the “Nominal
Value” input is done.
4-4-5. Selection of Measurement Speed
One of three measurement speeds SLOW, MEDIUM, or FAST could be selected (Figure 4-7). The continuous mode speeds are about 1, 5, and 12 meas urement per second resp ectivel y. Th e trade-off is a ccuracy vs. speed. LC R­817/819 will take a more accurate measurement at a slower rate. The trade-off is as follows section
SLOW
speed : More than 1 mea surement per second, at 0.05% accur acy ( or
better)
MEDIUM
speed : More than 3 measurements per second, at 0.1% accuracy
(or better)
FAST
speed :
More than 7 measurements per second, at 0.24% accuracy (or better)
* For th e d etails of a ccu ra cy, please refer to the specifica tions. * Regarding the models LCR-827 & 829, please refer to 4-6-2 Handler Interface
Timing.
LCR-800 Series User Manual
17
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
L 1.2345H Q 0.6789
TESTING
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
R.H OFF C.V OFF
OFFSET
SORT
SETTING
CALBRAT
EXIT
CAP. R/L OFFSET
SET SORT
SET PARAMETER
CALIBRATION
Press menu
key
Press F2 key
to the sort
menu
NOM.VAL
24.870
HANDLER
EXIT
NOM.VAL = 77.000pF
OPTION 1
Press F1 key to input the nominal value
Figure 4-6. Steps of “Nominal Value” input.
LCR-800 Series User Manual
18
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
L 1.2345H Q 0.6789
TESTING
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
R.H OFF C.V OFF
Press
F1
key to select the three different measurement
speed. (SLOW, MEDIUM, or FAST)
Figure 4-7. Selection of measurement speed
4-5. Measurement Conditions
4-5-1. Bias Vo ltage
There are two ava ila bl e bi as volt age modes: “Internal” and “External”.
Internal:
An internal DC 2 volts bias voltage will apply to the device under test.
External:
An ex t er n al DC bi a s vol t a g e bet we en 0 a nd 30 volt s can be a ppl i ed t o t h e d evi ce under test. The external bias connection is located on the rear panel. The maximum current is 200mA. The supply of bias voltage has to be floating, don’t connect either side to ground. It’s better to wait approximately 1 second for takin g a read ing aft er init iatin g a test ing pr ocess, th erefor e, th e device u nder t est will stabilize after bias voltage applied. The DC bias voltage should be applied only to capacitors in general. If the DC bias voltage is applied to device of low impedance, the unreliable testing results will occur.
Loading...
+ 50 hidden pages