GW Instek LCR-16, LCR-17 Quick Start Manual

Page 1
DC Bias Test Fixture
LCR-16, LCR-17
QUICK START GUIDE
GW INSTEK PART NO. 82CR-17000M01
ISO-9001 CERTIFIED MANUFACTURER
Page 2
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will Corporation.
The information in this manual was correct at the time of printing. However, Good Will continues to improve its products and therefore reserves the right to change the specifications, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City
236, Taiwan.
Page 3
1
OVERVIEW
The LCR-16 DC Bias Voltage Test Fixture/LCR-17 DC Bias Current Test Fixture should be installed between the LCR meter and the test fixture. The DC Bias Fixtures are used to pass additional DC bias voltage or current to test components.
Specifications
Model
LCR-16
LCR-17
External DC Bias
DC ±45V
DC ±2.5A
Limit (Current/Voltage)
DC ±20mA
DC ±45V
Measurement Frequency
40Hz ~ 1MHz
Internal Capacitance
1mF
N/A
Internal Inductance
N/A
330uH
Dimensions:
113W x 34H x106D mm
Weight
250g
310g
Operating Environment
Temperature
0~50˚C
Relative Humidity
<70%RH
Page 4
2
Fixture Overview
LCR-16 Overview
LCR-17 Overview
Page 5
3
LCR-16/LCR-17 DC Bias Input
INSTALLATION
LCR-16/LCR-17
1. Connect the DC Bias Test Fixture to the LCR meter’s BNC
terminals.
2. Next, connect the test fixture to the DC Bias Test Fixture.
3. Confirm that the output is turned off from the power supply used
for the DC bias. Connect the power supply BNC cable to the input of the DC Bias Test Fixture.
Page 6
4
LCR-16 Installation
LCR-17 Installation
Page 7
5
MEASUREMENT
WARNING
Before measurement, be sure to read the operating instructions to avoid danger.
LCR-16
1. Place the component in the test fixture.
2. Increase the voltage from the power supply gradually from 0V to
the desired voltage value.
3. Perform measurement on the component.
4. Reduce the power supply voltage gradually from the set voltage to
0V.
5. Remove the component from the test fixture.
LCR-17
1. Place the component in the test fixture.
2. Increase the current from the power supply gradually from 0A to
the desired current value.
3. Perform measurement on the component.
4. Reduce the power supply current gradually from the set current to
0A.
5. Remove the component from the test fixture.
Page 8
6
NOTE
Do not remove the BNC cable from the DC Bias input before turning off power to the test component.
Please note that when using the DC bias fixture, the test component may cause electric shock.
Do not perform zeroing or short calibration when the DC Bias input is used.
When the DC Bias input is used, do not turn on the LCR meter’s internal DC Bias settings.