Measuring Low-Value Resistance of up to 200 (Protective
EISO
)75
Conductor and Equipotential Bonding Conductor) ..................77
17.1 Measurement with Constant Test Current..........78
17.2 Protective Conductor Resistance Measurement
with Ramp Sequence – Measurement at PRCDs
with Current-Monitored Protective Conductor using
the PROFITEST PRCD Test Adapter as an
25.6 Internet Addresses for Additional Information ..104
Gossen Metrawatt GmbH3
Page 4
1Safety Instructions
Observe this documentation, in particular all included safety information, in order to protect yourself and others from injury, and to
prevent damage to the instrument.
The operating instructions and the condensed operating instructions should be made available to all users.
General
• Tests/measurements may only be performed by a qualified
electrician, or under the supervision and direction of a qualified electrician. The user must be instructed by a qualified
electrician concerning performance and evaluation of tests
and/or measurements.
• Observe the five safety rules in accordance with
DIN VDE 0105-100:2015-10, VDE 0105-100:2015-10
(EN 50110-1), Operation of electrical installations – Part 100:
General requirements
(1: Shut down entirely. 2: Secure against restart. 3: Assure
absence of voltage at all poles. 4: Ground and short circuit. 5:
Cover neighboring live components, or make them inaccessible).
• Observe and comply with all safety regulations which are
applicable for your work environment.
• Wear suitable and appropriate personal protective equipment
(PPE) whenever working with the instrument.
• The functioning of active medical devices (e.g. pacemakers,
defibrillators) and passive medical devices may be affected by
voltages, currents and electromagnetic fields generated by
the tester and the health of their users may be impaired.
Implement corresponding protective measures in consultation
with the manufacturer of the medical device and your physician. If any potential risk cannot be ruled out, do not use the
instrument.
Accessories
• Use only the specified accessories (included in the scope of
delivery or listed as options) with the instrument.
• Carefully and completely read and adhere to the product documentation for optional accessories. Retain these documents
for future reference.
Handling
• Use the instrument in undamaged condition only.
Inspect the instrument before use. Pay particular attention to
damage, interrupted insulation or kinked cables.
Damaged components must be replaced immediately.
• Accessories and cables may only be used as long as they’re
fully intact.
Inspect accessories and all cables before use. Pay particular
attention to damage, interrupted insulation or kinked cables.
• If the instrument or its accessories don’t function flawlessly,
permanently remove the instrument/accessories from operation and secure them against inadvertent use.
• If the instrument or accessories are damaged during use, for
example if they’re dropped, permanently remove the instrument/accessories from operation and secure them against
inadvertent use.
• The instrument and the accessories may only be used for the
tests/measurements described in the documentation for the
instrument.
• Neither the integrated voltage measuring function nor the
mains connection test may be used to test systems or system
components for the absence of voltage.
Testing for the absence of voltage is only permissible with a
suitable voltage tester or voltage measuring system which fulfills the requirements specified in DIN EN 61243.
Operating Conditions
• Do not use the instrument and its accessories after long periods of storage under unfavorable conditions (e.g. humidity,
dust or extreme temperature).
• Do not use the instrument and its accessories after extraordinary stressing due to transport.
• The instrument must not be exposed to direct sunlight.
• Only use the instrument and its accessories within the limits of
the specified technical data and conditions (ambient conditions, IP protection code, measuring category etc.).
• Do not use the instrument in potentially explosive atmospheres.
Rechargeable Batteries
• When using the charger, only the battery pack (Z502H) may
be inserted in the device.
• Do not use the instrument while charging the battery pack
(Z502H).
• Do not use the test instrument if the battery compartment lid
has been removed.
Touch contact with dangerous voltage is otherwise possible.
• The battery pack (Z502H) may only be charged in undamaged
condition.
Inspect the battery pack (Z502H) before use. Pay particular
attention to leaky and damaged batteries.
Fuses
• The instrument is equipped with fuses. The instrument may
only be used as long as the fuses are in flawless condition.
Defective fuses must be replaced. See detailed operating
instructions.
Measurement Cables and Establishing Contact
• Plugging in the measurement cables must not necessitate any
undue force.
• Never touch conductive ends (e.g. of test probes).
• Fully unroll all measurement cables before starting a test/measurement. Never perform a test/measurement with the measurement cable rolled up.
• Avoid short circuits due to incorrectly connected measurement cables.
• Ensure that alligator clips, test probes or Kelvin probes make
good contact.
Data Security
• Always create a backup copy of your measurement data.
• Observe and comply with the respectively applicable national
data protection regulations. Use the corresponding functions
provided by the test instrument such as access protection, as
well as other appropriate measures.
4Gossen Metrawatt GmbH
Page 5
2Applications
Please read this important information!
2.1Intended Use / Use for Intended Purpose
Measuring and test instruments from the PROFITEST MASTER
series include:
• PROFITEST MBASE+ (M520S)
• PROFITEST MXTRA (M522P)
• PROFITEST MTECH+ (M522R)
• PROFITEST MPRO (M520N)
The test instruments are used to test the effectiveness of protec-
tive measures at stationary electrical systems in accordance with
IEC 60364-6, EN 50110-1 and other country-specific standards.
They can also be used for the testing of electric charging stations
per EN 61851-1 (DIN VDE 0122-1), and for earth measurements.
The test instruments include pre-programmed test sequences for
increased working convenience and user-defined test sequences
can also be programmed as an option.
The test instruments are especially well suited for testing electrical
systems during setup, initial startup, periodic testing and troubleshooting.
The applications range of the test instruments covers all alternating and 3-phase current systems with nominal voltages of 230/
400 V (300/500 V) and nominal frequencies of 16⅔, 50, 60, 200
and 400 Hz.
A system structure is set up in the test instrument and measured
values are assigned to the objects. Completed tests and measured values can be saved and documented in a measurement
and test report.
Safety of the operator, as well as that of the test instrument, is
only assured when it’s used for its intended purpose.
2.2Use for Other than Intended Purpose
Using the test instrument for any purposes other than those
described in these operating instructions, or in the test instrument’s condensed operating instructions, is contrary to use for
intended purpose.
2.3Liability and Guarantee
Gossen Metrawatt GmbH assumes no liability for property damage, personal injury or consequential damage resulting from
improper or incorrect use of the product, in particular due to failure to observe the product documentation. Furthermore, all guarantee claims are rendered null and void in such cases.
Nor does Gossen Metrawatt GmbH accept any liability for data
loss.
2.4Opening the Instrument / Repairs
In order to ensure flawless, safe operation and to assure that the
guarantee isn’t rendered null and void, the test instrument may
only be opened by authorized, trained personnel. Even original
replacement parts may only be installed by authorized, trained
personnel.
Unauthorized modification of the test instrument is prohibited.
If it can be ascertained that the test instrument has been opened
by unauthorized personnel, no guarantee claims can be honored
by the manufacturer with regard to personal safety, measuring
accuracy, compliance with applicable safety measures or any
consequential damages.
If the guarantee seal is damaged or removed, all guarantee claims
are rendered null and void.
2.5Scope of Functions
PROFITEST …
(Article Number)
PRO
M
✓
✓
✓
✓
(M520N)
TECH+
M
✓✓
✓✓
—
—
—
—
✓✓
(M520S)
MBASE+
Testing of Residual Current Devices (RCDs)
UT measurement without tripping the RCD
Tripping time measurement
Measurement of tripping current I
Selective, SRCDs, PRCDs, type G/R
AC/DC sensitive RCDs, types B and B+
Testing of insulation monitoring devices (IMDs)———
Testing of residual current monitoring devices (RCMs)———
Testing for N-PE reversal
Measurement of Loop Impedance Z
Fuse table for systems without RCDs
Without tripping the RCD, fuse table——
15 mA measurement
with 2-pole adapter, probe, earth electrode and current
clamp sensor
Selective earthing resistance R
with probe, earth electrode and
(4-wire measuring method
(3-wire measuring method)
current clamp sensor
via PRO-RE adapter)
E (mains operation)
E (battery operation)
current clamp sensor
via PRO-RE adapter and
)
—
✓✓✓✓
—
Earth loop resistance RELOOP (battery operation)
with 2 clamps (current clamp sensor direct and current
clamp transformer via PRO-RE/2 adapter)
Measurement of equipotential bonding R
Automatic polarity reversal
Insulation resistance R
Variable or rising test voltage (ramp)
Voltage U
Special Measurements
IL, I
Phase sequence
Earth leakage resistance R
Voltage drop (U)
Standing-surface insulation Z
Meter startup (kWh test)
Leakage current with PRO-AB (IL) adapter
Residual voltage test (Ures)
Intelligent ramp (ta + I)
Electric vehicles at charging stations (IEC 61851-1)
Documentation of fault simulations at PRCDs with the
PROFITEST PRCD adapter
Features
Selectable user interface language
Memory (database for up to 50,000 objects)
Automatic test sequence function
/ U
L-N
current measurement with clamp
AMP
L-PE
/ U
INS
N-PE
/ f
E(INS)
ST
LO
2)
RS 232 port for RFID/barcode reader
USB port for data transmission
ETC PC database and report generating software
Measuring category: CAT III 600 V / CAT IV 300 V
DAkkS calibration certificate
1)
The so-called live measurement is only advisable if there’s no bias current within
the system. Only suitable for motor protection switches with small nominal current
values.
2)
Currently available languages D, GB, I, F, E, P, NL, S, N, FIN, CZ, PL
—
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓
———
———
———
——
———
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓
XTRA
(M522R)
M
✓
✓
✓
✓
✓
✓
✓
✓
✓
✓
(M522P)
Gossen Metrawatt GmbH5
Page 6
3Documentation
Note
S
This documentation describes several test instrument.
As a result, features and functions may be described which do
not apply to your instrument. Furthermore, illustrations may differ
from your instrument.
These operating instructions describe a test instrument
with software/firmware version 1.16.20.
List of Abbreviations and their Meanings
RCCBs (residual current circuit breakers / RCDs):
I
I
N
I
F
PRCDPortable residual current device
RCD-Selective RCCB
R
E
SRCDSocket residual current device (permanently installed)
t
a
U
I
U
IN
U
L
Overcurrent protective devices:
I
SC
Z
L-N
Z
L-PE
Earthing:
R
B
R
E
R
ELoop
Low-value resistance at
protective, earthing and bonding conductors:
R
LO+
R
LO–
Insulation:
R
E(INS)
R
INS
R
ST
Z
ST
Current:
I
A
I
L
I
M
I
N
I
P
Voltage:
fLine voltage frequency
Tripping current
Nominal residual current
Rising test current (residual current)
PRCD-S:
with protective conductor detection and monitoring
PRCD-K:
with undervoltage trigger and protective conductor
monitoring
Calculated earthing or earth electrode loop resistance
Time to trip / breaking time
Touch voltage at moment of tripping
Touch voltage
relative to nominal residual current I
N
Touch voltage limit value
Calculated short-circuit current (at nominal voltage)
Supply impedance
Loop impedance
on page 81
– section 20, “Test Sequences (Automatic Test Sequences) –
AUTO Function”, on page 95
Further interesting information: section 21, “Maintenance”, on
page 97.
Testi ng RC Ds
Measurement of Insulation Resistance
”, on page 43
Function)”,
L-N
”, o n pa ge
Measuring Low-Value Resistance of up to 200
SC
”,
6Gossen Metrawatt GmbH
Page 7
5The Instrument
!
XY123
2018-07
D-K
15080-01-01
5.1Scope of Delivery
Standard scope of delivery for PROFITEST MASTER series:
1 Test instrument1 Compact battery pack
(Z502H)
1 Earthing contact plug insert,
country-specific
(PRO-SCHUKO /
GTZ3228000R0001)
1 2-pole measuring adapter
and cable for expansion into
a 3-pole adapter (PRO-A3-II /
Z501O)
2 Alligator clips1 Operating Instructions
1USB cable
1 Neck strap1 ETC software*
1Charger
1 DAkkS calibration certificate
(this document)
(
Z502R)
5.3Meanings of Symbols on the Instrument
Warning concerning a point of danger
(attention, observe documentation!)
Protection category II device
Charging socket for extra-low direct voltage
(for Z502R charger)
The device and its batteries may not be disposed of
with household trash. Further information is included in
the operating instructions.
Indicates EC conformity
If the guarantee seal is damaged or removed, all guarantee claims are rendered null and void.
The special technical knowledge of qualified personnel
is required for electrical installation or repair.
** Download from Internet
5.2Optional Accessories (excerpt)
A complete overview of optional accessories including detailed
information can be found in the data sheet for the test instrument.
The most important accessories are listed here:
• Barcode Profiscanner RS232 (Z502F)
(barcode reader and scanner for RS 232 connection to the
test instrument for identifying systems, electrical circuits and
operating equipment.)
• PRO-HB (Z501V)
Holder for test probes and measuring adapter
• E-SET PROFESSIONAL (Z592Z)
(extensive earth measurement accessories)
CAT III
600 V
CAT IV
300 V
Calibration seal (blue seal):
Consecutive number
Deutsche Akkreditierungsstelle GmbH – calibration lab
Registration number
Date of calibration (year – month)
Measuring category
Gossen Metrawatt GmbH7
Page 8
5.4Instrument Overview
456
8
91011
7
13
31
161715
14
2
192021
22
12
!
RS 232
2
Connector Sockets for Current Clamp Sensor, Probe
and PRO-AB Leakage Current Measuring Adapter
15
16
17
* See section 10.1 on page 36 regarding use of the
probes.
*
*
*
A
Test Instrument and Adapter
Test Instrument and Adapter:
1 Control panel with keys and display screen
2 Eyelets for attaching the neck strap
3 Rotary selector switch
4 Measuring adapter (2-pole)
5 Plug insert (country-specific)
6 Test plug (with retaining ring)
7 Alligator clip (plug-on)
8 Test probes
9 ON/START ▼ key *
10 I
11 Contact surfaces for finger contact
12 Test plug holder
13 Fuses
14 Holders for test probes (8)
8Gossen Metrawatt GmbH
/compens./ZOFFSET key
N
Connections for Current Clamp, Probe, PRO-AB Leakage Current
Measuring Adapter:
15 Current clamp connection 1
16 Current clamp connection 2
17 Probe socket
Interfaces, Charger Connection:
19 USB slave for connection to a PC
20 RS 232 port for connecting barcode or RFID reader
21 Socket for Z502R charger
22 Battery compartment lid (compartment for batteries and
spare fuses)
* Can only be switched on with the key on the instrument
Accessories:
A PRO-HB (Z501V) test probe and measuring adapter holder –
can be purchased separately
Page 9
(1) Control Panel – Display Panel
Attention!
!
Attention!
!
Attention!
!
See section 6.1, “Control Panel”, on page 16.
See section 6.2, “Display”, on page 16.
(2) Eyelets for the Neck Strap
The included neck strap can be attached at the right and left hand
sides of the instrument. You can hang the instrument from your
neck and keep both hands free for measurement.
(3) Rotary Selector Switch
The following basic functions can be selected with the rotary
switch:
SETUP / I
EXTRA / AUTO
N
/ IF / Z
L-P E
/ Z
L-N
/ RE / R
LO
/ R
/ U / SENSOR /
INS
The various basic functions are selected by turning the function
selector switch while the instrument is switched on.
(4) Measuring Adapter
The measuring adapter (2-pole) may only be used together with the test instrument’s test plug.
Use for other purposes is prohibited!
The plug-on measuring adapter (2-pole) with the two test probes
is used for measurements in systems without earthing contact
outlets, e.g. at permanent installations, distribution cabinets and
all three-phase outlets, as well as for insulation resistance and
low-value resistance measurements.
The 2-pole measuring adapter can be expanded to three poles for
phase sequence testing with the included measurement cable
(test probe).
This key has the same function as the
▼ key on the test plug.
(10) IN / I Key (at the control panel)
The following sequences are triggered
by pressing this key, either on the test
plug or at the control panel:
• Starts the tripping test after measurement of touch voltage for
RCCB testing (I
• Measurement of ROFFSET is started in the R
N
).
/ Z
function.
LO
L-N
• Semiautomatic polarity reversal (see section 10.9)
(11) Contact Surfaces
The contact surfaces are located at both sides of the test plug.
When the contact plug is grasped in the hand, contact is automatically made with these surfaces. The contact surfaces are
electrically isolated from the terminals and from the measuring circuit.
In the event a potential difference of greater than 25 V between
protective conductor terminal PE and the contact surface, PE is
displayed. See “LED Indications, Mains Connections and Potential Differences” on page 17.
(12) Test Plug Holder
The test plug with attached plug insert can be reliably secured to
the instrument with the rubberized holder.
(13) Fuses
The two fuses protect the device in case of overload. Phase conductor L and neutral conductor N are fused individually. If a fuse is
defective, and if an attempt is made to perform a measurement
which uses the circuit protected by this fuse, a corresponding
message appears at the display panel.
See section 21.2, “Fuse Replacement”, on page 97.
(5) Plug Insert (country-specific)
The plug insert may only be used together with the test
instrument’s test plug.
Use for other purposes is prohibited!
After the plug insert has been attached, the instrument can be
directly connected to earthing contact outlets. There’s no need to
concern yourself with poling at the plug. The instrument detects
the positions of phase conductor L and neutral conductor N and
automatically reverses polarity if necessary.
The instrument automatically determines whether or not both protective contacts in the earthing contact outlet are connected to
one another, as well as to the system protective conductor, for all
types of protective conductor measurements when the plug insert
is attached to the test plug.
(6) Test Plug
The various country specific plug inserts (e.g. protective contact
plug insert for Germany or SEV plug insert for Switzerland) or the
measuring adapter (2-pole) are attached to the test plug and
secured with a threaded connector.
The controls on the test plug are subject to interference suppression filtering. This may lead to slightly delayed responses as
opposed to controls located directly on the instrument.
(7) Alligator Clip (plug-on)
(8) Test Probes
The test probes comprise the second (permanently attached) and
third (plug-on) poles of the measuring adapter. A coil cable connects them to the plug-on portion of the measuring adapter.
(14) Holders for Test Probes (8)
(15/16) Current Clamp Connector
Only the current clamp transformers offered as accessories may
be connected to these sockets.
(17) Probe Connector Socket
The probe connector socket is required for the measurement of
probe voltage U
tance R
and standing surface insulation resistance.
E
, earth electrode voltage UE, earthing resis-
S-PE
It can be used for the measurement of touch voltage during RCD
testing. The probe is connected with a 4 mm contact-protected
plug.
The instrument determines whether or not the probe has been
properly set and displays results at the display panel.
(19) USB Port
The USB port allows for the exchange of data between the test
instrument and a PC.
(20) RS 232 Port
This port allows for data entry by means of a barcode scanner or
an RFID reader.
(21) Charging Socket
Only the Z502R charger for charging batteries inside the test
instrument may be connected to this socket.
(22) Battery Compartment Lid – Replacement Fuses
Before removing the lid is removed, the instrument must
be disconnected from the measuring circuit at all poles!
(9) ON/Start
The measuring sequence for the function selected in the menu is started by
pressing this key, either on the test
plug or at the control panel. Exception: If the instrument is
▼ Key
The compartment under the lid accommodates the rechargeable
battery pack (Z502H), or commercially available rechargeable batteries or regular batteries.
Two replacement fuses are also located under the battery com-
partment lid.
switched off, it can only be switched on by pressing the key at the
control panel.
Gossen Metrawatt GmbH9
Page 10
5.5Technical Data
BAT
Nominal Ranges of Use
Voltage U
Frequency f
N
N
Overall voltage range
Overall frequency range 15.4 Hz 420 Hz
Line voltageSinusoidal
Temperature range0 C +40C
Battery voltage8 V 12 V
Supply impedance angle Corresponds to cos =1 0.95
Probe resistance< 50 k
120 V(108 V 132 V)
230 V(196 V 253 V)
400 V(340 V 440 V)
16⅔ Hz
(15.4 V 18 Hz)
50 Hz(49.5 V 50.5 Hz)
60 Hz(59.4 V 60.6 Hz)
200 Hz(190 V 210 Hz)
400 Hz(380 V 420 Hz)
65 V 550 V
Reference Conditions
Line voltage230 V 0.1%
Line frequency50 Hz 0.1%
Measured qty. frequency
Measured qty. waveform
45 Hz 65 Hz
Sine (deviation between effective and
rectified value 0.1%)
Supply impedance angle cos =1
Probe resistance 10
Supply voltage12 V 0.5 V
Ambient temperature+ 23 C 2 K
Relative humidity40% 60%
Finger contactFor testing potential difference
to ground potential
Standing surface
insulationPurely ohmic
Power Supply
Batteries8 each AA 1.5 V
We recommend exclusive use of the
included rechargeable battery pack
(2000 mAh, Z502H).
Number of measurements (standard setup with illumination)
– For R
INS
– For R
LO
Battery testSymbolic display of rechargeable
Battery-saving circuitDisplay illumination can be switched off.
Safety shutdownIf supply voltage is too low, the instru-
Recharging socketInserted rechargeable batteries can be
Charging timeZ502R charger: approx. 2 hours *
1 measurement – 25 s pause:
approx. 1100 measurements
Auto polarity reversal / 1
(1 measuring cycle) – 25 s pause:
approx. 1000 (Z502O) measurements
battery voltage
The test instrument is switched off
automatically after the last key operation. The user can select the desired
on-time.
ment is switched off, or cannot be
switched on.
recharged directly by connecting a
charger to the recharging socket:
Z502R charger
Overload Capacity
R
ISO
U
, U
L-P E
L-P E
LO
, Z
E
L-N
L-N
, R
F
RCD, R
Z
R
Protection with
fine-wire fusesFF 3.15 A 10 s,
1200 V continuous
600 V continuous
440 V continuous
550 V (Limits the number of measure-
ments and pause duration. If overload
occurs, the instrument is switched off
by means of a thermostatic switch.)
Electronic protection prevents switching
on if interference voltage is present.
Fuses blow at > 5 A
Electrical Safety
Protection classII
Nominal voltage230/400 V (300/500 V)
Test voltage3.7 kV, 50 Hz
Measuring categoryCAT III 600 V or CAT IV 300 V
Pollution degree2
Fuses
L and N terminals1 G fuse-link ea.
FF 3.15/500G 6.3 × 32 mm
Electromagnetic Compatibility (EMC)
Product standardEN 61326-1
Interference
emission
EN 55022A
Interference
immunity
EN 61000-4-2Contact/atmos. – 4 kV/8 kV
EN 61000-4-310 V/m
EN 61000-4-4Mains connection – 2 kV
EN 61000-4-5Mains connection – 1 kV
EN 61000-4-6Mains connection – 3 V
EN 61000-4-110.5 periods / 100%
The transformation ratio selected at the clamp (1/10/100/1000 mV/A)
×
IN> 300 mA and 5
×
IN > 500 mA and
Key: d = digit(s), rdg. = reading (measured value)
must be set in the “Type” menu with the rotary switch in the “SENSOR”
position.
4
Where R
5
The specified measuring and intrinsic uncertainties already include those
of the respective current clamp.
6
Measuring range of the signal input at the test instrument, UE: 0 … 1.0
V
7
Input impedance
8
DC bias only possible with PROFITEST MXTRA
11
Where Z
TRMS
(0 … 1.4 V
Eselective/REtotal
< 0.6 , ISC > UN/0.5 is displayed
L-PE
< 100
) AC/DC
peak
of the signal input at the test instrument
: 800 k
14Gossen Metrawatt GmbH
Page 15
Characteristic Values, Special Measurements with PROFITEST MPRO and PROFITEST MXTRA
Func-
RE
1
2
3
4
5
6
7
8
9
10
11
Measured
tion
Quantity
RE, 3-pole
RE, 4-pole±(|10% rdg.|+10d) ±(|3% rdg.|+5d)
RE, 4-pole
selective
with clamp meter
BAT
Soil resistivity
(p)
Probe clearance d (p)0.1 … 999 m
RE, 2 clamps
10.0 k … 19.9 k
10.0 k… 49.9 k
1.00 m … 9.99 km
Display Range
0.00 … 9.99
10.0 … 99.9
100 … 999
1.00 k … 9.99 k
10.0 k … 50.0 k
0.00 … 9.99
10.0 … 99.9
100 … 999
1.00 k … 9.99 k
0.0 m … 9.9 m
100 m … 999 m
0.00 … 9.99
10.0 … 99.9
100 … 999
1.00 … 1.99 k
10
11
Reso-
lution
0.01
0.1
1
0.01 k
0.1 k
0.01
0.1
1
0.01 k
0.1 k
0.1 k
0.1 m
1 m
0.01
kWm
0.01
0.1
1
0.01 k
Current /
Signal Freq.
1
16 mA/128 Hz
1.6 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
16 mA/128 Hz
16 mA/128 Hz
1.6 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
16 mA/128 Hz
1.6 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
30 V / 128 Hz
100 m … 9.99 km
500 m … 9.99 km
5.00 km … 9.99 km
5.00 km … 9.99 km
5.00 km … 9.99 km
Signal frequency without interference signal
PRO-RE (Z501S) adapter cable for test plug, for connecting earth
probes (E-Set 3/4)
PRO-RE/2 adapter cable for test plug, for connecting the E-CLIP2 generator
clamp
Generator clamp: E-CLIP2 (Z591B)
Clamp meter: Z3512A (Z225A)
Where R
Where R
< 10 or clamp meter current > 500 µA
E.sel/RE
E.H/RE
100 and R
E.E/RE
100
Where d = 20 m
Where d = 2 m
Only where RANGE = 20 k
Only where RANGE = 50 k or AUTO
Te st
Measuring Range
1.00 … 19.9
5.0 … 199
50 … 1.99 k
0.50 k … 19.9 k
0.50 k … 49.9 k
1.00 … 9.99
10.0 … 200
0.10 … 9.99
10.0 … 99.9
Key: d = digit(s), rdg. = reading (measured value)
Measuring
Uncertainty
±(|10% rdg.|+10d
+ 1
±(|15% rdg.|+10d)
±(|20% rdg.|+10d)
6
8
8
±(|20% rdg.|+10d)7±(|12% rdg.|+10d)
9
9
9
±(|10% rdg.|+5d)
±(|20% rdg.|+5d)
±(|3% rdg.|+5d
±(|10% rdg.|+10d)
±(|15% rdg.|+10d)
±(|5% rdg.|+5d)
±(|12% rdg.|+5d)
Intrinsic
Uncertainty
+ 0.5
7
Connections
Adapter for Test Plug
Current Clamps
PRO-RE PRO-RE/2 Z3512A Z591B
2
25
2
354
Gossen Metrawatt GmbH15
Page 16
6Operating and Display Elements
Attention!
!
MEM: Key for memory functions
HELP:
Access context sensitive help
IN: Tripping test
Next step
(semiautomatic measurement)
Start offset measurement
ON/START ▼:
Switch on
Start/stop measurement
ESC
:Return from submenu
SoftkeysFixed Function Keys
• Parameter selection
• Limit value
specification
• Entry functions
LEDs and Connection Symbols
section 6.3
9
10
Battery Display
Measuring Function
Measurement in progress/
Memory Occupancy
Measured
Parameters
PE
Save Value
Battery full
Battery OK
Battery weak
Battery (almost) dead
Battery Display ( section 7.1)
BAT
BAT
BAT
BAT
Memory Occupancy Display
MEM
Memory half full
MEM
Memory full > transfer data to PC
Connection Test – Mains Connection Test ( section 6.4)
NPEL
NPEL
)(
Connection OKL and N reversed
NPELNPEL
x
NPELNPEL
x
x
RUN READY
Connection Test section 6.4
U < 8 V
LPEN
x
LPEN
Quantities
stopped
6.1Control Panel
The display and control panel can be swiveled forward or backward with the detented swivel hinge. The instrument can thus be set to
the optimum reading angle.
6.2Display
The following appear at the display:
• One or two measurement values as three place numeric display with unit of measure and abbreviated measured quantity
• Nominal values for voltage and frequency
• Circuit diagrams
• On-line help
• Messages and instructions
6.3LEDs
MAINS/NETZ LED
This LED is only functional when the instrument is switched on. It
has no function in the voltage ranges U
It lights up green, red or orange, or blinks green or red depending
upon how the instrument has been connected and the selected
function (see also section 6.4, “LED Indications, Mains Connections and Potential Differences”, beginning on page 17).
This LED also lights up if line voltage is present when measuring
R
and RLO.
ISO
UL/RL LED
This LED lights up red if touch voltage is greater than 25 V or 50 V
during RCD testing, as well as after safety shutdown occurs. It
also lights up if R
fallen short of.
RCD • FI LED
This LED lights up red if the RCCB is not tripped within 400 ms
(1000 ms for selective RCDs – type RCD S) during the tripping
test with nominal residual current. It also lights up if the RCCB is
not tripped before nominal residual current has been reached
during measurement with rising residual current.
or RLO limit values have been exceeded or
ISO
L-N
and U
L-P E
.
16Gossen Metrawatt GmbH
The mains connection test may not be used to test systems or system components for the absence of voltage!
Page 17
6.4LED Indications, Mains Connections and Potential Differences
Attention!
!
???
NPEL
NPEL
NPEL
NPELxNPELxNPEL
x
NPEL
NPEL
x
NPEL
LED Signals
Status
Tes t
Plug
Measuring
Function Switch Position Function/Meaning
Adapter
NETZ/
MAINS
NETZ/
MAINS
NETZ/
MAINS
NETZ/
MAINS
NETZ/
MAINS
U
L/RL
FI/RCD
Lights up
green
Blinks
green
Blinks red
Lights up
red
Blinks yel-
low
Lights up
red
Lights up
red
X
X
XX
U, Z
U, Z
U, Z
X
X
XX
Z
L-N
XX
IN / IF , Z
, kWh, IMD, int. ramp, RCM
ST
IN / IF , Z
, kWh, IMD, int. ramp, RCM
ST
IN / IF , Z
, kWh, IMD, int. ramp, RCM
ST
RLO, R
IN / IF , Z
R
, Z
Z
L-N
L-PE
RE, Z
, Z
,
ZST, IMD, kWh, RCM, PRCD,
L-PE
/ Z
L-N
/ Z
L-N
/ Z
L-N
, RE, IL, sensor
ISO
/ Z
L-N
, RLO, RE,
INS
, U, IL, U
, IF, IN,
L-PE
t
+I, RCM
a
e-mobility
IN / IF,
int. ramp
L-PE
L-PE
L-PE
L-PE
, sensor
res
/ RE,
/ RE,
/ RE,
/ R
Correct connection, measurement enabled
N conductor not connected,
measurement enabled
1) No line voltage or
2) PE interrupted
Interference voltage is present at the test probes.
Measurement is disabled.
E
L and N are connected to the phase conductors.
The selected limit value has been violated.
Interference voltage limit value U
has been exceeded.
L
Safety shutdown has occurred.
The test has been manually assessed as “NOT OK”.
The RCCB was not tripped, or was tripped too late
during the tripping test.
Mains Connection Test — Single-Phase System — LCD Connection Pictographs
The mains connection test may not be used to test systems or system components for the absence of voltage!
Status
Tes t
Plug
Measuring
Function Switch
Position
Function/Meaning
Adapter
Is dis-
played
Is dis-
played
Is dis-
played
Is dis-
played
Is dis-
played
Is dis-
played
All except for U
All except for U
All except for U
All except U and RE
RE
All except for U
All except for U
No connection detected
Connection OK
L and N reversed, neutral conductor charged with phase voltage
No mains connection
Standard display without connection messages
Neutral conductor interrupted
Protective conductor PE interrupted,
neutral conductor N and/or phase conductor L charged with phase voltage
Is dis-
played
Is dis-
played
Is dis-
played
Is dis-
played
Gossen Metrawatt GmbH17
All except for U
All except for U
All except for U
All except for U
Phase conductor L interrupted,
neutral conductor N charged with phase voltage
Phase conductor L and protective conductor PE reversed
Phase conductor L and protective conductor PE reversed
Neutral conductor interrupted (with probe only)
L and N are connected to the phase conductors.
Page 18
Mains Connection Test — 3-Phase System — LCD Connection Pictographs
Attention!
!
The mains connection test may not be used to test systems or system components for the absence of voltage!
Status
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
Test PlugMeasuring
Adapter
Function Switch Position Function/Meaning
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
Clockwise rotation
Counter-clockwise rotation
Short between L1 and L2
Short between L1 and L3
Short between L2 and L3
Conductor L1 missing
Conductor L2 missing
Conductor L3 missing
Is displayed
Is displayed
Is displayed
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
Conductor L1 to N
Conductor L2 to N
Conductor L3 to N
18Gossen Metrawatt GmbH
Page 19
Connection Test — Battery Powered Earthing Resistance Measurements, “Battery Mode”
NPEL
PE
PE
Status
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
Tes t
Plug
Measuring
Adapter
PRO-RE
Clamp
meter
PRO-RE
PRO-RE
PRO-RE
Function
Switch Position
R
E
R
E
R
E
R
E
R
E
R
E
Function/Meaning
Standard display without connection messages
Interference voltage at probe S > 3 V
Restricted measuring accuracy
Interference/measuring current ratio > 50 at R
, 1000 at R
E(sel)
Restricted measuring accuracy
at R
Interference current > 0 85 A or interference/measuring current ratio > 100
E(sel):
➭ No measured value, display: RE.Z – – –
Probe H not connected or R
>150k
E.H
➭ No measurement, display: RE – – –
R
> 50 k or
E.H
R
E.H/RE
> 10000
➭ Measured value is displayed, restricted measuring accuracy
Probe S not connected
or R
> 150 k
E.S
or R
× R
E.S
> 25 M²
E.H
➭ No measurement, display: RE – – –
R
> 50 k or
E.S
R
E.S/RE
> 300
➭ Measured value is displayed, restricted measuring accuracy
Probe E not connected or R
> 150 kR
E.E
E.E/RE
> 2000
➭ No measurement, display: RE – – –
R
E.E/RE
>300
➭ Measured value is displayed, restricted measuring accuracy
E(2Z)
PE Test via Finger Contact at the Contact Surfaces on the Test Plug
Status
Tes t
Plug
Measuring
Function Switch
Position
Function/Meaning
Adapter
LCDLEDs
Is displayed
Is displayed
UL/RL
FI/RCD
light up
red
U
L/RL
FI/RCD
light up
red
XX
XX
U
(single-phase
measurement)
U
(single-phase
measurement)
Potential difference 50 V between finger contact and PE (earth contact)
Frequency f 50 Hz
If L is correctly contacted and PE is interrupted (frequency f 50 Hz)
Status bar: Display of Charge Level, Memory Occupancy
Status
Tes t
Plug
Measuring
Function Switch
Position
Function/Meaning
Adapter
Battery
status
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
U,
R
INS
R
, RE,
LO
Z
, Z
L-N
I
, I
F
Setup,
EXTRA,
SENSOR
,
L-PE
N
Battery charge level ≥ 80%
Battery charge level ≥ 50%
,
,
Battery charge level ≥ 30%
Battery charge level ≥ 15%
Battery charge level ≥ 0%
Gossen Metrawatt GmbH19
Page 20
Battery
test
Is displayed
Memory
Status
Is dis-
Is dis-
Is dis-
Is dis-
Is dis-
Is dis-
played
played
played
played
Is dis-
played
Is dis-
played
played
Is dis-
played
played
Z
Error Messages — LCD Connection Pictographs
All
U,
R
,
INS
R
, RE,
LO
, Z
L-N
L-PE
I
, I
F
N
Setup,
EXTRA,
SENSOR
Rechargeable batteries must be recharged or replaced towards the end
of their service life (U < 8 V).
Memory occupancy ≥ 100%
Memory occupancy ≥ 87.5%
Memory occupancy ≥ 75%
Memory occupancy ≥ 62.5%
,
,
Memory occupancy ≥ 50%
Memory occupancy ≥ 37.5%
Memory occupancy ≥ 25%
Memory occupancy ≥ 12.5%
Memory occupancy ≥ 0%
Status
Te st
Plug
Measuring
Function Switch
Position
Adapter
All measurements
XX
XX
XXI
XXZ
XXI
with protective
conductor
I
Z
/ Z
L-N
/ IF
N
L-PE
N
L-PE
/ I
N
Function/Meaning
Potential difference UL between finger contact and PE (earth contact)
(frequency f 50 Hz)
Remedy: Check PE connection
Note: Only if is displayed: Measurement can nevertheless be
started by pressing the start key again.
1) Voltage too high (U > 253 V) for RCD test with direct current
2) U always U > 550 V with 500 mA
3) U > 440 V for I
/ R
E
4) U > 253 V for I
5) U > 253 V for measurement with probe
RCD is tripped too early or is defective.
Remedy: Test circuit for bias current
RCD is tripped too early or is defective.
Remedy: Test with “DC + positive half-wave”.
RCD tripped during touch voltage measurement.
F
Remedy: Check selected nominal test current.
/ IF
N
/ IF with 500 mA
N
R
IF, I∆N,
LO,
EXTRA
ta+I∆
The PRCD has been tripped.
Reason: Poor contact or defective PRCD
Externally accessible fuse is blown.
The voltage ranges remain functional even if fuses have blown.
XXAll except for U
Special case, R
blown fuse.
: Interference voltage during measurement may result in a
LO
Remedy: Replace fuse as described in section 21.2.
20Gossen Metrawatt GmbH
Page 21
XX
Z
L-N
I
/ Z
N
/ IF
L-PE
/ R
Frequency out of permissible range
Remedy: Check the mains connection.
E
All
XX
PRO-RE
X
PRO-RE
PRO-RE/
2
XX
All measurements with
XXR
/ RLO
RINS
RE (bat)
RE (bat)
RE (bat)
probe
ISO
Excessive temperature inside the test instrument
Remedy: Wait for test instrument to cool down
Interference voltage
Remedy: Device under test must be disconnected from all sources of
voltage
Interference voltage > 20 V at the probes:
H to E or S to E
No measurement possible
Probe ES not connected or connected incorrectly
Generator current clamp (E-Clip-2) not connected
Interference voltage at the probe
Overvoltage or overloading of the measuring voltage generator during
measurement of R
INS
Remedy: Ensure absence of voltage at the device under test.
I
/ IF
N
Z
/ Z
L-N
XX
L-PE
ZST, RST, R
E
Meter startup
XXAll
XXR
XR
XEXTRA
XEXTRA
LO
LO
U
U
No mains connection
Remedy: Check the mains connection.
Defective hardware
Remedy:
1) Switch on/off or
2) Briefly remove the batteries.
If error message persists, send instrument to GMC-I Service GmbH.
OFFSET measurement is not sensible.
Remedy: Check system.
OFFSET measurement of RLO+ and RLO– is still possible.
R
OFFSET
> 9.99 :
OFFSET measurement is not sensible.
Remedy: Check system.
Z > 9.99 :
OFFSET measurement is not sensible.
Remedy: Check system.
U
OFFSET
> U:
OFFSET value is greater than the measured value at the consuming system.
OFFSET measurement is not sensible.
Remedy: Check system.
Contact problem or blown fuse
XXR
/ RLO / R
ISO
Remedy: Check test plug or measuring adapter for correct seating in the
E(bat)
test plug, or replace the fuse.
Gossen Metrawatt GmbH21
Page 22
XR
IN/I
F
10 mA30 mA100 mA300 mA500 mA
R
MAX
at I
N
510 170 50 15 9
R
MAX
for I
F
410 140 40 12 7
E
The polarity of the 2-pole adapter must be reversed.
XI
N
/ IF
N and PE are reversed.
1) Mains connection error
Remedy: Check the mains connection.
or
2) Display at the connection pictograph: PE interrupted (x) or
XX
Z
L-N
/ Z
N
/ IF
L-PE
/ R
E
bottom protective conductor bar interrupted with reference to the
keys at the test plug
Cause: Voltage measuring path interrupted
I
Result: Measurement is disabled
Note: Only if is displayed: Measurement can nevertheless be
started by pressing the start key again.
Display at the connection pictograph:
Top protective conductor bar interrupted with reference to the keys at the
XI
N
/ IF
test plug
Cause: Current measuring path interrupted
Result: No measured value display
I
N
RE
/ IF
Probe is not detected, probe not connected
Remedy: Check probe connection.
Clamp is not detected:
– Clamp is not connected or
– Current through clamp is too small (partial earthing resistance too high)
RE
or
– Transformation ratio set incorrectly
Remedy: Check clamp connection and transformation ratio.
Check the batteries in the METRAFLEX P300 and replace if
necessary.
I
Z
RE
RE
All
/ IF
N
L-PE
, R
If you have changed the transformation ratio at the test instrument, a
message appears prompting you to change the setting at the current
clamp sensor as well.
Voltage too high at clamp input or signal distorted
The transformation ratio parameter selected at the test instrument might not
correspond to the transformation ratio at the current clamp sensor.
Remedy: Check transformation ratio or test setup.
Battery voltage is less than or equal to 8 V.
Reliable measurement is no longer possible.
Storage of measured values to memory is disabled.
Remedy: Rechargeable batteries must be recharged or replaced
towards the end of their service life.
Resistance in N-PE path is too high.
Consequence: The required test current cannot be generated and measurement is aborted.
If specified touch voltage UL is exceeded:
and RE: User is prompted to switch to the 15 mA wave.
Z
E
L-PE
RE alternative only:
User is prompted to reduce the measuring range (reduce current.)
Types B, B+ and EV/MI not possible with G/R, SRCD, PRCD
180° not possible for RCD-S, G/R, SRCD, PRCD-S, PRCD-K
DC not possible with G/R, SRCD, PRCD
I
/ IF
N
I
/ IF
N
EXTRA
EXTRA
I
I
R
R
N
N
E
E
RCM
RCM
Half-wave or DC not possible with type AC
DC not possible with type A, F
½ test current not possible with DC
2 × IN / 5 × IN with full-wave only
DC+ with 10 Ω only
No DC bias in the IT network
R
E
EXTRA RCM
I
/ IF
N
EXTRA
RCM
Gossen Metrawatt GmbH23
15 mA only possible in 1 k and 100 ranges!
With RCM: Types AC, F, B+ and EV/MI are not possible.
Measurement with half-wave or DC is not possible in IT systems.
Page 24
All
The parameters you have selected do not make sense in combination
with previously configured parameters. The selected parameter settings
will not be saved.
Remedy: Enter other parameters.
EXTRA ta+I
Messages — LCD Pictographs — Test Sequences
Status
Te st
Plug
Measuring
Function Switch
Position
Adapter
AUTO
AUTO
AUTO
R
E
2-pole measurement via earthing contact plug is not possible in IT systems.
The intelligent ramp is not possible with RCD types RCD-S and G/R.
Function/Meaning
The test sequence includes a measurement which cannot be processed
by the connected test instrument. The corresponding test step must be
skipped. Example: The test sequence includes an RCM measurement
which has been sent to the PROFITEST MTECH+.
The test sequence has been run successfully.
No test sequences have been saved.
Cause: These may have been deleted as a result of any of the following
actions: changing the language, the profile or the DB mode,
or resetting the test instrument to its default settings.
Change to the larger measuring range (test instrument and leakage cur-
L
rent measuring adapter).
Test measurement:
The test has been passed.
L
The leakage current measuring adapter is now ready for use.
Test measurement:
The test has not been passed.
L
The leakage current measuring adapter is defective. Contact our repair
service department.
Test measurement:
L
Check the fuse in the leakage current measuring adapter.
24Gossen Metrawatt GmbH
Page 25
Database and Entry Operations — Pictographs
Attention!
!
Z
EXTRA
EXTRA RCM
I
L-N
N
/ IF
/ Z
L-PE
tA+I
Measured Value Storage with Deviating Electrical Circuit Parameter
The electrical circuit parameter selected by yourself at the test instrument
does not coincide with the parameter entered under object data in the
structure.
Example: Residual operating current is specified as 10 mA in the database, but you have performed measurement with 100 mA. If you want to
perform all future measurements with 100 mA, the value in the database
has to be changed by acknowledging with the key. The measured
value is documented and the new parameter is accepted.
If you want to leave the parameter in the database unchanged, press the
key. The measured value and the changed parameter are only docu-
mented in this case.
All
All
All
All
All
All
Please enter a designation (alphanumeric).
Operation with a Barcode Scanner
Error message when the “EDIT” entry field is opened and rechargeable
battery voltage is less than 8 V. Output voltage is generally switched off
during barcode scanner operation if U is less than 8 V in order to assure
that remaining battery capacity is adequate for entering designations for
devices under test and saving the measurement.
Remedy: Rechargeable batteries must be recharged or replaced towards
the end of their service life.
Operation with a Barcode Scanner
Current flowing through the RS 232 port is too high.
Remedy: The connected device is not suitable for this port.
Operation with a Barcode Scanner
Barcode not recognized, incorrect syntax.
Data cannot be entered at this location within the structure.
Remedy: Observe profile for preselected PC software (see SETUP menu in
section 8).
Measured value cannot be saved at this location within the structure.
Remedy: Make sure that you have selected the right profile for you PC
evaluation program in the SETUP menu (see section 8).
All
All
SETUP
Memory is full.
Remedy: Save your measurement data to a PC and then clear memory at
the test instrument by deleting the database or by importing an empty
database.
Delete measurement or database elements.
This prompt window asks you to confirm deletion (YES).
Data loss after restoring default settings!
Back up your measurement data to a PC before pressing the respective
key.
This prompt window asks you to confirm deletion.
Gossen Metrawatt GmbH25
Page 26
7Operation
Attention!
!
Note
Attention!
!
Attention!
!
Attention!
!
Note
Attention!
!
Attention!
!
Attention!
!
Attention!
!
Attention!
!
Attention!
!
BAT
The protective foil on the two sensor surfaces (finger
contacts) of the test plug must be removed to ensure reliable detection of touch voltages.
7.1Power Supply
The instrument is powered by rechargeable batteries. The
included Master Battery Pack (Z502H) or commercially available
individual rechargeable or regular batteries can be used.
If at all possible use the included battery pack (Z502H)
with sealed cells. This ensures that the complete set of
rechargeable batteries is always replaced at the same
time and that all batteries are inserted with correct polarity, in order to assure that they do not fail.
The included Z502H battery pack has already been inserted
during initial startup (see condensed operating instructions).
7.1.1Inserting or Replacing the Battery Pack (Z502H) or Commercially Available Individual (Rechargeable) Batteries
Before opening the battery compartment, disconnect the
instrument from the measuring circuit (mains) at all poles!
Commercially available, individual rechargeable or regular batteries must comply with the technical data (see
page 10).
➭ Loosen the slotted screw for the rechargeable battery com-
partment lid on the back and remove the lid.
➭ Remove the depleted battery pack or commercially available
rechargeable or regular batteries.
➭ Insert the battery pack or commercially available rechargeable
or regular batteries into the battery compartment.
In the case of commercially available, individual rechargeable or regular batteries: make sure that all of the
batteries are inserted with correct polarity. If just one battery is inserted with reversed polarity, it will not be recognized by the instrument and may result in leakage from
the batteries and damage to the instrument.
➭ Replace the lid and retighten the screw.
Dispose of the battery pack or commercially available,
individual rechargeable or regular batteries in an environmentally sound fashion when their service life has nearly
expired (approx. 80% charging capacity). See section 24,
“Disposal and Environmental Protection”, on page 99.
7.1.2Charging the Battery Pack (Z502H) in the Tester
If commercially available, individual rechargeable batteries are used, they must be charged externally. Do not
use the Z502R charger to charge commercially available
individual batteries. The quality of commercially available,
individual rechargeable batteries cannot be checked and
may result in overheating and thus deformation and explosion when charging them in the instrument.
If commercially available, individual rechargeable batteries are used, they must be charged externally. Do not
use the Z502R charger to charge commercially available
individual batteries.
The quality of commercially available, individual rechargeable batteries cannot be checked and may result
in overheating and thus deformation and explosion when
charging them in the instrument.
Regular batteries may not be charged.
Use only the Z502R charger in order to recharge the
Compact Battery Pack (Z502H) in the test instrument.
The Z502R charger is suitable for mains operation only!
Do not switch the test instrument on during charging.
The charging process may otherwise be impaired.
➭ Verify that the battery pack (Z502H) is inserted, i.e. that com-
mercially available battery packs or batteries are not inserted.
➭ Insert the correct mains plug for your country into the charger
Z502R.
➭ Connect the Z502R charger to the test instrument with the
jack plug, and then to the 230 V mains with the interchangeable plug.
➭ Do not disconnect the charger from the test instrument until
the green LED (charged/ready) lights up.
If the rechargeable batteries or battery pack have not been used
or recharged for a lengthy period of time (> 1 month), thus resulting in excessive depletion:
Observe the charging sequence (indicated by LEDs at the charger) and initiate a second charging sequence if necessary (disconnect the charger from the mains and from the test instrument
to this end, and then reconnect it).
Please note that the system clock stops in this case and must be
set to the correct time after the instrument has been restarted.
7.2Switching the Instrument On/Off
The test instrument is switched on by pressing the ON/START▼
key. The menu which corresponds to the momentary selector
switch position is displayed.
The instrument can be switched off manually by simultaneously
pressing the MEM and HELP keys.
After the period of time selected in the SETUP menu has elapsed,
the instrument is switched off automatically (see “Device Settings”, section 8).
Battery Test
A battery test is performed after switching the instrument on.
If supply voltage has fallen below the permissible
lower limit, the pictograph shown at the right appears.
“Low Batt!!!” is also displayed along with an icon.
The instrument does not function if the batteries have been
depleted excessively, and no display appears.
Ensure adequate power supply by charging the rechargeable battery pack (Z502H/) or by inserting fully charged, commercially
available rechargeable batteries or new batteries. See section 7.1,
“Power Supply”, on page 26.
26Gossen Metrawatt GmbH
Page 27
8Instrument Settings
SETUP
LED and LCD test menu
Rotary switch balancing
Brightness/contrast menu
Calibration date
Display: date/time
Display: auto shutdown
Display: auto shutdown
of display illumination after 15 s.
of the tester after 60 s.
Time, language
1
2
3
and battery test menu
0b
0a
0
Return to main menu
MAINS LED: test green
MAINS LED: test red
UL/RL LED: test red
RCD-FI LED: test red
Cell test
Inverse cell test
Hide all pixels
Show all pixels
Acoustic signal test
1
Return to main menu
Brightness/contrast submenu
Set time
Default settings
Language for
user interface
3
3a
3b
3c
3e
Set date
Duty cycle
for display illumination / tester
0b
Return to submenu
0a
Display Illumination On-time
Time, On-Time and Default Settings
Menu Selection for Operating Parameters
LED testsLCD and acoustic signal tests
Test Instrument On-Time
3f
4
No automatic shutdown,
continuously on
DB MODE submenu
3g
Current inspector
Firmware/software info page 97
Brightness and Contrast Settings
Create
or select
inspector
(deletion via ETC only)
Distributor structures
3d
Gossen Metrawatt GmbH27
Page 28
LED and LCD test menu
Rotary switch balancing
Brightness/contrast menu
Calibration date
Display: date/time
Display: auto shutdown
Display: auto shutdown
of display illumination after 15 s.
of the tester after 60 s.Time, language
1
2
3
and battery test menu
0b
0a
0
Return to main menu
Brightness/contrast submenu
Set time
Default settings
Language for
user interface
3
3a
3b
3c
3e
Set date
Duty cycle
for display illumination / tester
Set time
Menu Selection for Operating Parameters
Set Time, Language, Acoustic Signal
Set date
Select time
Increase hours
Increase minutes
Decrease hours
Apply settings
Decrease minutes
3a
Increase seconds
Decrease seconds
Return to submenu
Select date
Decrease
Decrease
day
Apply settings
month
3b
Decrease
year
Return to submenu
Increase
Increase
day
month
Increase
year
3f
4
Current inspector
DB MODE submenu
3g
Firm/software info page 97
Brightness and Contrast Settings
Create or select inspector
(deletion via ETC only)
Distributor structures
3d
28Gossen Metrawatt GmbH
Page 29
Significance of Individual Parameters
Note
Note
Attention!
!
Attention!
!
Attention!
!
0a
0b
2
2
2
2
3c
3d
3e
Test Instrument On-Time
The period of time after which the test instrument is automatically
shut off can be selected here. This selection has a considerable
influence on the service life and the charging status of the batteries.
LCD Illumination On-Time
The period of time after which LCD illumination is automatically
shut off can be selected here. This selection has a considerable
influence on the service life and the charging status of the batteries.
Submenu: Rotary Switch Balancing
Proceed as follows in order to precision adjust the rotary switch:
1 Press the TESTS Rotary Switch / Battery Test softkey in order
to access the rotary switch balancing menu.
2 Then press the softkey with the rotary switch icon.
SETUP
3 Make sure that the rotary switch is set to
The level mark to the left of the number should be aligned to the
black function field with the respective rotary switch designation. The
value of the number can be displayed within a range of -1 to 101 and
should be between 45 and 55. In the case of -1 or 101, the rotary
switch position does not match the measuring function shown at the
display.
If the displayed value is not within this range, readjust the position by
pressing the
edges readjustment.
readjust
softkey . A brief acoustic signal acknowl-
.
User Interface Language (CULTURE)
➭ Select the desired country setup with the appropriate country
code.
All structures, data and sequences are deleted when the
language is changed!
Back up your structures,
measurement data and sequences to a PC before
pressing the key.
The prompt window shown at
the right asks you to confirm
deletion.
Profiles for Distributor Structures (PROFILES)
The profiles are laid out in
a tree structure. The tree
structure for the utilized
PC evaluation program
may differ from that of the
PROFITEST MASTER. For
this reason, the PROFITEST MASTER provides
the user with the opportunity of adapting this structure.
Selecting a suitable profile
determines which object
combinations are made
possible. For example,
this makes it possible to
create a distributor which is subordinate to another or to save a
measurement to a given building.
➭
Select the PC evaluation program you intend to use.
If labeling in the LCD image of the rotary switch does not
correspond with its actual position, a continuous acoustic signal is generated as a warning when the readjust
softkey is pressed.
4 Acknowledge by pressing the softkey with the rotary switch icon.
The display is then switched to the next measuring function.
5 Turn the rotary switch clockwise to the next measuring function (after
SETUP
comes
IN).
6 Repeat steps 3 through 5 until all rotary switch functions have
been tested, and if necessary readjusted.
7 Press ESC, in order to return to the main menu.
Submenu: Battery Level Query
If battery voltage has dropped to 8.0 V or less, the UL/RL LED lights
up red and an acoustic signal is generated as well.
Measuring Procedure
If battery voltage drops to below 8.0 V
during the course of a measuring
sequence, this is only indicated by means
of a pop-up window. Measured values
are invalid. The measurement results cannot be saved to
memory.
All structures, data and sequences are deleted when the
profile is changed!
Back up your structures,
measurement data and sequences to a PC before
pressing the key.
The prompt window shown at
the right asks you to confirm
deletion.
If you haven’t selected a suitable PC evaluation
program and, for example, if measured value
storage to the selected location within the structure is not possible, the pop-up window shown
at the right appears.
Default Settings (GOME SETTING)
The test instrument is returned to its original default settings when
this key is activated.
All structures, data and sequences are deleted!
Back up your structures, measurement data and sequences to
a PC before resetting.
➭ Press ESC in order to return to the main menu.
Gossen Metrawatt GmbH29
Page 30
Adjusting Brightness and Contrast
Note
Note
3f
Jump back to
Increase brightness
Decrease brightness
Increase contrast
Decrease contrast
previous menu
3g
4
Add a new inspector
Accept letter/character
Select letter/character
Select letter/character
Delete letter/character
Toggle: upper/lowercase,
vowel mutations & special chars.
Accept name
Select inspector
Select inspector
Accept inspector
DB-MODE – Display of
the Database in the Text or
ID Mode
Selecting, Adding or Deleting an Inspector
See also section 10.8 on page 39 regarding the entry of a text.
Creating Structures in the TXT MODE
The database in the test instrument is set to the text mode as a
default feature and “TXT” appears in the header. You can create
structure elements in the test instrument and label them in plain
text, e.g. Customer XY, Distributor XY and Circuit XY.
Creating Structures in the ID MODE
You can work in the ID MODE as an alternative, in which case “ID”
appears in the header. You can create structure elements in the
test instrument and label them with any desired ID numbers.
Structures can be created in the test instrument in either
the text mode or the ID mode.
In contrast to this, designations and ID numbers are
always assigned in the report generating program.
If no texts or ID numbers have been entered to the test instrument
when creating structures, the report generating program creates
the missing entries automatically. These can then be edited in the
report generating program and transferred back to the test instrument if required.
(delete inspector via ETC only)
The inspector cannot be changed. If an inspector’s name
is incorrect, it can be deleted and a new inspector can be
created with the correct name.
Changes are not retroactive. Deleted inspectors are
retained for tests which have already been performed.
30Gossen Metrawatt GmbH
Page 31
9Database
Note
9.1Creating Distributor Structures, General
A complete distributor structure with data for electrical circuits
and RCDs can be created in the test instrument.
This structure makes it possible to assign measurements to the
electrical circuits of various distributors, buildings and customers.
There are two possible procedures:
• On location or at the
construction site:
create a distributor
structure in the test
instrument.
A distributor structure with up to
50,000 structure elements can be created in the test
instrument, which is
saved to the instrument’s flash memory.
or
• Create and save an image of an existing distributor structure
at a PC with the help of ETC report generating software (Electric Testing Center) – see Help > Getting Started (F1). The distributor structure is then transferred to the test instrument.
The test instrument and the PC must be connected with a USB
cable in order to transfer structures and data.
The rotary selector switch may not be set to the “U” position during data transmission.
The following image
appears at the display
during transfer of structures and data.
9.3Creating a Distributor Structure in the Test Instrument
Overview of the Meanings of Icons used to Create Structures
IconsMeaning
Main
Sublevel
level
Memory Menu, Page 1 of 3
Cursor UP: scroll up
Notes regarding ETC
The following steps must be completed before using the software:
• Install the USB device driver
(required for operation of the test instrument at a PC):
GMC-I Driver Control software for installing the USB device
driver can be downloaded from our website:
https://www.gmc-instruments.de/services/download-center/
• Install ETC report generating software:
The most up-to-date version of ETC can be downloaded free
of charge from the mygmc page of our website as a ZIP file, if
you have registered your test instrument:
https://www.gmc-instruments.de/services/mygmc/
Cursor DOWN: scroll down
ENTER: Acknowledge selection.
+ – change to sub-level
(expand directory) or
– + change to main level
(close directory)
Display the complete structure designation (max.
63 characters) or ID number (25 characters) in a
zoom window.
Temporarily switch back and forth between structure designation and ID number.
These keys don’t have any effect on the main setting in the setup menu (see “DB Mode” on
page 30).
Hide the zoom window
Change display to menu selection
Memory Menu, Page 2 of 3
Add a structure element
9.2Transferring Distributor Structures
The following data transfer operations are possible:
• Transfer a distributor structure from the PC to the test instrument.
• Transfer a distributor structure including measured values
from the test instrument to the PC.
Gossen Metrawatt GmbH31
Page 32
IconsMeaning
Distributor
A check mark to the right of a structure element means that all measurements
within the respective hierarchy have been passed.
x: at least one measurement has not been passed.
No symbol: measurement has not yet been performed.
Building
Customer
RCD
Circuit
Equipment
Same type of element as in the Windows Explorer:
+: sub-objects available, display by pressing .
–: sub-objects are displayed, hide by pressing .
Equipment
Meanings of icons from top to bottom:
Customer, building, distributor, RCD, electrical cir-
cuit, operating equipment, machine and earth
electrode (display of the icons depends on the
selected structure element).
Selection: UP/DOWN scroll keys and
In order to add a designation to the selected
structure element, refer to the edit menu in following column.
Delete the selected structure element.
Show measurement data, if a measurement has
been performed for this structure element.
Edit the selected structure element.
Memory Menu, Page 3 of 3
Search for ID number.
> Enter complete ID number.
Search for text.
> Enter full text (complete word).
Search for ID number or text.
Distributor Structure Symbology / Tree Structure
Continue searching.
Edit menu
Cursor LEFT:
Select an alphanumeric character.
Cursor RIGHT:
Select an alphanumeric character.
ENTER: accept an individual character.
Acknowledge entry
Scroll left
Scroll right
Delete character
Switching amongst different types of alphanumeric characters:
AUpper case letters
aLower case letters
0Numbers
@Special characters
32Gossen Metrawatt GmbH
Page 33
9.3.1Creating Structures (example for electrical circuit)
Note
Note
Note
Scroll up
Scroll down
Comfirm selection /
Display object
Nextpage
change level
or ID number
Create object
Delete object
VA: Show measurement data
Edit designation
Scroll up
Scroll down
Comfirm selection
Select character
Select character
Accept character
Delete character
Character selection:
✓Save object designation
A, a, 0, @
Select parameter
Parameter settings list
Acknowledge parameter selection
Acknowledge parameter setting
and return to page 1/3
Select parameter setting
in the database menu.
After selection with the MEM key, all setting options for the creation of a tree structure are made available on three menu pages
(1/3, 2/3 and 3/3). The tree structure consists of structure elements, referred to below as objects.
Selecting the Position at which a New Object will be Added
Use the keys in order to select structure elements.
Change to the sub-level with the key.
Go to the next page with the >> key
Entering a Designation
Enter a designation and then acknowledge it with ✓.
Acknowledge your entry with ✓and , because the entry
will otherwise not be accepted.
Entering a Comment
Enter a comment and then acknowledge it with ✓.
Creating a New Object
Press the key in order to create a new object.
Select a new object from a list.
Acknowledge your entry with ✓ and , because the entry
will otherwise not be accepted.
Setting Electrical Circuit Parameters
For example, nominal current values must be entered here for the
selected electrical circuit. Measuring parameters which have been
accepted and saved in this way are subsequently accepted by
the current measuring menu automatically when the display is
switched from the structure view to measurement.
Electrical circuit parameters changed during structure
creation are also retained for individual measurements
(measurement without saving data).
If you change the electrical circuit parameters specified by the
structure in the test instrument, a warning is displayed when
the change is saved (see error message on page 25).
Select the desired object from the list with the keys and
acknowledge with the key.
Depending upon the profile selected in the test instrument’s
SETUP menu (see section 8), the number of object types may be
limited, and the hierarchy may be laid out differently.
Gossen Metrawatt GmbH33
Page 34
9.3.2Searching for Structure Elements
Scroll up
Scroll down
Comfirm selection /
Display object
Menu selection page 3/3
change level
or ID number
Search for ID number
Search for text
Search for ID number or text.
Select character
Select character
Accept character
Delete character
Character selection:
✓ Save object designation
Continue searching
End search
Regardless of the currently selected object, the search is started
at database.
Go to page 3/3 in the database menu.
After selecting text search...
If no further matches are found, the message shown above is displayed.
9.4Saving Data and Generating Reports
Preparing and Executing a Measurement
Measurements can be performed and stored to memory for each
structure element. Proceed as follows, adhering to the prescribed
sequence:
➭ Select the desired measurement with the rotary knob.
➭ Start the measurement by pressing the ON/START ▼ or I
Upon completion of measurement, the Floppy Disk softkey is
displayed.
➭ Briefly press the Save Value key.
The display is switched to the memory menu or the
structure view.
➭ Navigate to the desired memory location, i.e. to the desired
structure element / object, for which the measurement data
will be saved.
➭ If you would like to save a comment along with the
measurement, press the key shown at the right and
enter a designation via the “EDIT” menu as described in
section 9.3.1.
➭ Complete data storage by pressing the “STORE” key.
N
key.
... and entering the desired text (only full matches are found – no
wild cards, case sensitive) ...
... the first match is displayed.
Further matches can be found by selecting
the icon shown at the right.
Saving Error Messages (pop-ups)
If a measurement is ended without a measured value due to an
error, the measurement can be saved along with the pop-up by
pressing the “Save Value” key. The corresponding text is read out
in ETC instead of the pop-up symbol. This only applies to a limited
number of pop-ups (see below). Neither a symbol nor a text can
be accessed in the test instrument’s database itself.
Alternative Storage Procedure
➭ The measured value can be saved to the last se-
lected object in the structure diagram by pressing and holding the Save Value key, without switching the
display to the memory menu.
34Gossen Metrawatt GmbH
Page 35
Note
If you change the parameters in the measurement view,
Note
Note
Note
they’re not saved for the structure element. A measurement with changed parameters can nevertheless be
saved to the structure element, and any changed parameters are documented in the report for each measurement.
Retrieving Saved Measured Values
➭ Switch the display to the distributor structure by pressing the
MEM key and select the desired electrical circuit with the scroll
keys.
➭ Switch to page 2
by pressing the key shown here:
➭ Display the measurement data
by pressing the key shown here:
One measurement with date
and time, as well as any comment you might have entered, is
displayed in each screen.
Example:
RCD Measurement
Data Evaluation and Report Generation with the Report Generating
Program
All data, including the distributor structure, can be transferred to
the PC and evaluated with the help of the report generating program. Additional information can be entered here subsequently
for the individual measurements. After pressing the appropriate
key, a report including all measurements within a given distributor
structure is generated, or the data are exported to an Excel
spreadsheet.
The database is exited when the rotary selector switch is
turned. Previously selected parameters in the database
are not used for the measurement.
9.5Use of Barcode Scanners and RFID Readers
Search for an Already Scanned Barcode
The search can be started from any switch setting and menu.
➭ Scan the object’s barcode.
The recognized barcode is displayed inversely.
➭ This value is accepted after pressing the ENTER key.
A previously selected object is not taken into consideration by the search.
A check mark in the header means that the respective
measurement has been passed.
An X means that the measurement has not been passed.
➭ Scrolling amongst measurements
is possible with the keys shown here:
➭ The measurement can be deleted with the key
shown here:
A prompt window asks you to confirm
deletion.
With the help of the key shown at the right (MW: measured value / PA: parameter), the setting parameters
can be displayed for this measurement.
Continued Searching in General
Regardless of whether or not an object has been found,
searching can be continued by pressing the key shown at
the right:
– Object found: Searching is continued below the previously
selected object.
– Further object found: The entire database is searched at all
levels.
Reading In a Barcode for Editing
If the menu for alphanumeric entry is active, any value scanned by
means of a barcode or RFID reader is accepted directly.
Using a Barcode Printer (accessory)
The following functions are made possible with the help of a barcode printer:
• Read-out of ID numbers as barcodes – for quick and convenient acquisition for periodic testing
• Print out repeatedly occurring designations such as test
object types encrypted as barcodes in a list, allowing them to
be read in as required for comments
➭ Scrolling amongst measurements
is possible with the keys shown here:
Gossen Metrawatt GmbH35
Page 36
10General Information on Measurements
Attention!
!
Note
10.1Using Cable Sets and Test Probes
• Scope of delivery: 2-pole measuring adapter and cable for
expansion into a 3-pole adapter (PRO-A3-II/)
•Optional accessory: PRO-RLO II (Z501P) 2-pole measuring
adapter with 10 m cable
• Optional accessory: KS24 cable set (GTZ3201000R0001)
Measurements per DIN EN 61010-031 may only be performed in
environments in accordance with measuring categories III and IV
with the safety cap attached to the test probe at the end of the
measurement cable.
In order to establish contact inside 4 mm jacks, the safety caps
have to be removed by prying open the snap fastener with a
pointed object (e.g. the other test probe).
Grip and hold the test plug and test probes securely
when they have been inserted, for example, into a
socket. Danger of injury exists if tugging at the coil cord
occurs, which may cause the test plug or test probes to
snap back.
10.2Test Plug – Changing Inserts
The test plug can be fitted with various inserts (e.g. two-pole
measuring adapter or country-specific plug insert).
In order to change inserts, unscrew the retaining ring until you can
pull out the currently used insert. Then mount the desired insert
and retighten the retaining ring.
(See overview in section 5.4 on page 8.)
10.3Connecting the Instrument
For systems with earthing contact sockets, connect the instrument to the mains with the test plug to which the appropriate,
country-specific plug insert is attached. Voltage between phase
conductor L and protective conductor PE may not exceed 253 V!
Poling at the socket need not be taken into consideration. The
instrument detects the positions of phase conductor L and neutral conductor N, and automatically reverses polarity if necessary.
This does not apply to the following measurements:
– Voltage measurement in switch position U
– Insulation resistance measurement
– Low-resistance measurement
The positions of phase conductor L and neutral conductor N are
identified on the plug insert.
If measurement is to be performed at three-phase outlets, at distribution cabinets or at permanent connections, the measuring
adapter must be attached to the test plug. Connection is established with the test probes: one at PE or N and the other at L.
The 2-pole measuring adapter must be expanded to 3 poles with
the included measurement cable for the performance of phase
sequence testing.
Touch voltage (during RCCB testing) and earthing resistance can
be, and earth-electrode potential, standing surface insulation
resistance and probe voltage must be measured with a probe.
The probe is connected to the probe connector socket with a
4 mm contact-protected plug.
10.4Automatic Settings, Monitoring and Shutdown
The test instrument automatically selects all operating conditions
which it’s capable of determining itself. It tests line voltage and
frequency. If these lie within their valid nominal ranges, they
appear at the display panel. If they are not within nominal ranges,
prevailing voltage (U) and frequency (f) are displayed instead of UN
and f
.
N
Touch voltage which is induced by test current is monitored for
each measuring sequence. If touch voltage exceeds the limit
value of > 25 V or > 50 V, measurement is immediately interrupted. The U
If battery voltage falls below the permissible limit value the instrument cannot be switched on, or it is immediately switched off.
LED lights up red.
L/RL
The measurement is interrupted automatically, or the measuring
sequence is blocked (except for voltage measuring ranges and
phase sequence testing) in the event of:
• Impermissible line voltages (< 60 V, > 253 V / > 330 V /
> 440 V or > 550 V) for measurements which require line voltage
• Interference voltage during insulation resistance or low resistance measurements
• Overheating at the instrument
As a rule, excessive temperatures only occur after approximately 50 measurement sequences at intervals of 5 seconds,
or Z
when the rotary selector switch is set to the Z
position.
L-P E
L-N
If an attempt is made to start a measuring sequence, an
appropriate message appears at the display panel.
The instrument only switches itself off automatically after completion of an automatic measuring sequence, and after the predetermined on-time has expired (see section 7.2). On-time is reset to
its original value as defined in the setup menu, as soon as any key
or the rotary selector switch is activated.
The instrument remains on for approximately 75 s in addition to
the preset on-time for measurements with rising residual current
in systems with selective RCDs.
The instrument always shuts itself off automatically!
10.5Measured Value Display and Memory
The following items appear at the display panel:
• Measured values with abbreviations and units of measure
• Selected function
• Nominal voltage
• Nominal frequency
• Error messages
Measured values for automatic measuring sequences are stored
and displayed as digital values until the next measurement
sequence is started, or until automatic shutdown occurs.
If the upper range limit is exceeded, the upper limit value is displayed and is preceded by the “>” symbol (greater than), which
indicates measurement value overrun.
The depiction of LEDs in these operating instructions
may vary from the LEDs on the actual instrument due to
product improvements.
Testing Earthing Contact Sockets for Correct Connection
The testing of earthing contact sockets for correct connection
prior to protective measures testing is simplified by means of the
instrument’s error detection system.
The instrument indicates improper connection as follows:
• Impermissible line voltage (< 60 V or > 253 V):
The MAINS/NETZ LED blinks red and the measuring
sequence is disabled.
• Protective conductor not connected or potential to earth 50 V at 50 Hz (switch position U – single-phase measurement):
If the contact surfaces are touched (finger contact*) while PE is
being contacted (via the country-specific plug insert, e.g.
SCHUKO, as well as via the PE test probe at the 2-pole
adapter) PE appears (only after a test sequence has been
started). The U
* For reliable detection of touch voltages, both sensor surfaces on the
test plug must be touched with unprotected fingers/palm, i.e. with direct skin contact (see also section 7).
• Neutral conductor N not connected
(during mains dependent measurements):
The MAINS/NETZ LED blinks green.
• One of the two protective contacts is not connected:
This is checked automatically during testing for touch current
U
. Poor contact resistance at one of the contacts leads to
IN
and RCD/FI LEDs light up red as well.
L/RL
36Gossen Metrawatt GmbH
Page 37
one of the following displays depending upon poling of the
Note
Attention!
!
plug:
– Display at the connection pictograph:
PE interrupted (x), or underlying protective conductor bar interrupted with reference to the keys at the test plug
Cause:
Voltage measuring path interrupted
Consequence: measurement is disabled
– Display at the connection pictograph:
Top protective conductor bar interrupted
with reference to the keys at the test plug
Cause: current measuring path interrupted
Result: no measured value display
See “LED Indications, Mains Connections and Potential
Differences” on page 17.
Reversal of N and PE in a system without RCCBs cannot
be detected and is not indicated by the instrument.
In a system including an RCCB, the RCCB is tripped
during touch voltage measurement without RCCB tripping (automatic Z
are reversed.
measurement), insofar as N and PE
L-N
10.6Help Function
The following information can be displayed for each switch position and basic function after it has been selected with the rotary
selector switch:
• Wiring diagram
• Measuring range
• Nominal range of use as well as measuring and intrinsic
uncertainties
• Nominal value
➭ Press the HELP key in order to query online help.
➭ If several pages of help are available for the respective mea-
suring function, the HELP key must be pressed repeatedly.
➭ Press the ESC key in order to exit online help.
Gossen Metrawatt GmbH37
Page 38
10.7Setting Parameters or Limit Values using RCD Measurement as an Example
1
2
2
3
4
4
5
6
2
4
3
5
6
1 Access the submenu for setting the desired parameter.
2 Select a parameter using the or scroll key.
3 Switch to the setting menu for the selected parameter with the
scroll key.
4 Select a setting value using the or scroll key.
5 Acknowledge the setting value with the key. This value is
transferred to the settings menu.
6 The setting value is not permanently accepted for the respec-
tive measurement until
returned to the main menu. You can return to the main menu
by pressing ESC instead of ✓, without accepting the newly
selected value.
✓ is pressed, after which the display is
Parameter Lock (plausibility check)
Individually selected parameter settings are checked for plausibility before transfer to the measurement window.
If you select a parameter setting which doesn’t make sense in
combination with other parameter settings which have already
been entered, it’s not accepted. The previously selected parameter setting remains unchanged.
Remedy: select another parameter setting.
38Gossen Metrawatt GmbH
Page 39
10.8Freely Selectable Parameter Settings or Limit Values
Note
Note
Select the EDIT menu.
Select value/U/M.
Select value/U/M.
Accept value/U/M.
Delete character
✓ Save value (to list)
Select the
EDIT+ menu.
Select value/U/M.
Select value/U/M.
Accept value/U/M.
Delete character
✓ Save value (to list)
10.8.1 Changing Existing Parameters
Individual parameters can be changed for certain measuring functions, i.e. adjusted within predetermined limits.
The EDIT menu doesn’t appear until after switching to the
right-hand column and selecting the editable parameter .
Example for RLO Measuring Function – Parameter: LIMIT RLO
1 Open the submenu for setting the desired parameter (no fig-
ure, see section 10.7).
2 Select the editable parameter – identified with the icon –
with the or scroll key.
3 Select the edit menu by pressing the key.
10.8.2 Adding New Parameters
For certain measuring functions, additional values within predefined limits can be added in addition to the fixed values.
The EDIT+ menu doesn’t appear until after switching to the
right-hand column.
Example for IN Measuring Function – Parameter: I
1 Open the submenu for setting the desired parameter
(no figure, see section 10.7).
Select the edit menu by pressing the key.
N
4 Select the respective characters with the left or right cursor
key. The character is accepted by pressing the key. The
value is acknowledged by selecting
key.
Observe the predefined limits for the new setting value.
Enter any places to the right of the decimal point as well.
✓ and then pressing the
2 Select the respective characters with the LEFT or RIGHT cursor
key. The character is accepted by pressing the key. The
value is acknowledged by selecting
key. The new parameter is added to the list.
Observe the predefined limits for the new setting value.
Enter any places to the right of the decimal point as well.
✓ and then pressing the
Gossen Metrawatt GmbH39
Page 40
10.92-Pole Measurement with Rapid or Semiautomatic Polarity
L1-N
L2-N
L3-N
L1-L2
L2-L3
L1-L3
N-PE
L1-PE
L2-PE
L3-PE
L+N-PE
L1-N
L2-N
L3-N
L1-L2
L2-L3
L1-L3
Z
L-PE
Z
L-N
L1-PE
L2-PE
L3-PE
R
iso
N-PE
L1-PE
L2-PE
L3-PE
L1-N
L2-N
L3-N
L1-L2
L2-L3
L1-L3
U
L1-N
L2-N
L3-N
L1-L2
L2-L3
L1-L3
N-PE
L1-PE
L2-PE
L3-PE
L+N-PE
L1-N
L2-N
L3-N
L1-L2
L2-L3
L1-L3
Z
L-PE
Z
L-N
L1-PE
L2-PE
L3-PE
R
iso
L1-PE
L2-PE
L3-PE
N-PE
L1-N
L2-N
L3-N
L1-L2
L2-L3
L1-L3
U
Reversal
Rapid, semiautomatic polarity reversal is possible for the following
measurements:
• Voltage U
• Loop impedance Z
• Internal line resistance Z
• Insulation resistance R
LP-E
L-N
INS
Rapid Polarity Reversal at the Test Plug
The polarity parameter is set to AUTO.
Fast and convenient switching amongst all polarity variants, or
switching to the parameter settings submenu, is possible by
pressing the I
key at the instrument or the test plug.
N
Semiautomatic Polarity Reversal in Memory Mode
The polarity parameter is set to AUTO.
If testing is to be conducted with all polarity variants, automatic
polarity changing takes place after each measurement after sav-ing.
Polarity variants can be skipped by pressing the I
instrument or the test plug.
40Gossen Metrawatt GmbH
key at the
N
Page 41
11Measuring Voltage and Frequency
Note
U
2
1
Select the Measuring Function
Switch Back and Forth Between Single and 3-Phase Measurement
Press the softkey shown at the left in order to switch
back and forth between single and 3-phase measurement. The selected phase measurement is displayed inversely (white on black).
11.1Single-Phase Measurement
Connection
11.1.2 Voltage Between L – PE, N – PE and L – L
with 2-Pole Adapter Connection
Press the softkey shown at the left in order to switch
back and forth between the country-specific plug
insert, e.g. SCHUKO, and the 2-pole adapter. The
selected connection type is displayed inversely
(white on black).
See section 10.9 concerning 2-pole measurement with rapid or
semiautomatic polarity reversal.
A probe must be used in
order to measure probe voltage U
S-PE
11.1.1 Voltage Between L and N (U
and PE
(U
) with Country-Specific Plug Insert, e.g.
N-PE
SCHUKO
Press the softkey shown at the left in order to switch
back and forth between the country-specific plug
insert, e.g. SCHUKO, and the 2-pole adapter. The
selected connection type is displayed inversely
(white on black).
L-N
.
),
L and PE
(U
L-PE
) a
nd N
11.23-Phase Measurement (line-to-line voltage) and Phase
Sequence
Connection
The measuring adapter
(2-pole) is required in
order to connect the
instrument, and is
expanded to a 3-pole
measuring adapter with
the included measurement cable.
If you view the country-specific plug insert, e.g.
SCHUKO, from the front, you’ll see two embossed letters, namely L and N. Automatic polarity reversal does
not take place during voltage measurement. You can
thus specify the terminal to which the phase is connected
in the socket. If (mains) voltage is displayed for UL-PE,
then the phase is located where L appears on the connector. If (mains) voltage is displayed for N-PE, then the
phase is located where N appears on the connector.
Gossen Metrawatt GmbH41
Page 42
➭ Press softkey U3~.
Note
Clockwise Rotation
Counterclockwise Rotation
A clockwise phase
sequence is required at all
3-phase electrical outlets.
• Measurement instrument connection is
usually problematic
with CEE outlets
due to contact problems.
Measurements can be executed quickly and reliably without
contact problems with the help of the Z500A variable plug
adapter set available from GMC.
• Connection for 3-wire measurement: L1-L2-L3 at plug in
clockwise direction as of PE socket
Direction of rotation is indicated by means of the following displays:
See section 6.4 regarding all indications for the mains
connection test.
Voltage Polarity
If the installation of single-pole switches to the neutral conductor
is prohibited by the standards, voltage polarity must be tested in
order to assure that all existing single-pole switches are installed
to the phase conductors.
42Gossen Metrawatt GmbH
Page 43
12Testing RCDs
Attention!
!
Note
I
N
3
-------
I
N
(measurement up to 1000 ms)
t
a
I
a
t
S
The testing of residual current devices (RCDs) includes:
• Visual inspection
•Testing
• Measurement
Use the test instrument for testing and measurement.
When testing systems with RCCBs, they may switch off.
This may occur even though it’s not normally provided
for by the test. Leakage currents may be present which,
in combination with the test current of the test instrument, exceed the shutdown threshold value of the
RCCB. PCs which are operated in proximity to such
RCCB systems may switch off as a consequence. This
may result in inadvertent loss of data. Before conducting
the test, precautions should therefore be taken to ensure
that all data and programs are suitably backed up and
the computer should be switched off, if necessary. The
manufacturer of the test instrument assumes no liability
for any direct or indirect damage to equipment, computers, peripheral equipment or data bases when performing the tests.
Measuring Method
The following must be substantiated by generating a fault current
downstream from the RCD:
• That the RCD is tripped no later than upon reaching
its nominal fault current value
• That the continuously permissible touch voltage value
UL agreed upon for the respective system is not exceeded
This is achieved by means of:
• Touch voltage measurement
10 measurements with full-waves and extrapolation of I
• Substantiation of tripping within 400 ms or 200 ms with IN
• Substantiation of tripping current with rising residual current
This value must be between 50% and 100% of IN (usually
about 70%).
N
RCD/FI TableWaveform
Differential Current
Suddenly occurring
Alternating
current
Slowly rising
Suddenly occurring
Pulsating
direct current
Slowly rising
Direct current
Direct current
up to 6 mA
* only PROFITEST MTECH+, PROFITEST MXTRA
Correct RCD/RCCB Function
Typ e AC Type A/F Type B */
B+*
Type EV/
MI*
✔✔✔✔
✔✔✔
✔✔
✔
Test Standard
The following must be substantiated per IEC 60364-6:
–Touch voltage occurring at nominal residual current may not
exceed the maximum permissible value for the system.
– Tripping of the RCCB must occur within 400 ms (1000 ms for
selective RCDs) at nominal residual current.
Important Notes
• The test instrument permits simple measurements at all types
of RCDs. Select RCD, SRCD, PRCD etc.
• Measurement must be executed at one point only per RCD
(RCCB) within the connected electrical circuits. Low-resistance continuity must be substantiated for the protective conductor at all other connections within the electrical circuit (R
or U
).
B
• The measuring instruments often display 0.1 V touch voltage
in TN systems due to low protective conductor resistance.
• Be aware of any bias currents within the system. These may
cause tripping of the RCDs during measurement of touch voltage UB, or may result in erroneous displays for measurements
with rising current:
Display = I
- I
F
bias current
• Selective RCDs identified with an can be used as the sole
means of protection for automatic shutdown if they adhere to
the same shutdown conditions as non-selective RCDs (i.e.
t
< 400 ms). This can be verified by measuring breaking time.
a
• Type B RCDs may not be connected in series with type A or F
RCDs.
LO
• No premature tripping with the test instrument, because testing is begun with 30% residual current (if no bias current
occurs within the system).
Bias Magnetization
Only AC measurements can be performed with the 2pole adapter. Suppression of RCD tripping by means of
bias magnetization with direct current is only possible via
a country-specific plug insert, e.g. SCHUKO, or the 3pole adapter.
Gossen Metrawatt GmbH43
Page 44
Measurement With or Without Probe
Attention!
!
I
N
Nom. res. current: 10 … 500 mA
Type 1: RCD, SRCD, PRCD …
Nominal current: 6 … 125 A
Type 2*: AC , A , F
, B, B+
, EV, MI
k
* Types B, B+, EV, MI =
AC/DC sensitive
Phase displacement: 0°/180°
X times tripping current:
Negative/positive half-wave
Negative/positive direct current
1, 2, 5 (I
N
max. 300 mA)
Waveform:
Connection:
Without/with probe
System type:
TN/TT, IT
Touch voltage:
Time to trip:
< 25 V, < 50 V, < 65 V
Measurements can be performed with or without a probe.
Measurements with probe require that the probe and reference
earth are of like potential. This means that the probe must be
positioned outside of the potential gradient area of the earth electrode (R
The distance between the earth electrode and the probe should
be at least 20 m.
The probe is connected with a 4 mm contact-protected plug.
In most cases this measurement is performed without probe.
) in the RCD safety circuit.
E
The probe is part of the measuring circuit and may carry
a current of up to 3.5 mA in accordance with IEC 61557
/ EN 61557.
12.1Measuring Touch Voltage (with reference to nominal residual current) with ⅓ Nominal Residual Current and Tripping
Test with Nominal Residual Current
Select Measuring Function
Connection
Testing for the absence of voltage at the probe can be performed
with the U
function (see also section 11.1 on page 41).
PROBE
Set Parameters for I
N
44Gossen Metrawatt GmbH
Page 45
1) Measuring Touch Current Without Tripping the RCD
Attention!
!
Note
Note
Attention!
!
Measuring Method
The instrument uses a measuring current of only 1/3 nominal
residual current for the determination of touch voltage U
occurs at nominal residual current. This prevents tripping of the
IN
which
RCCB.
This measuring method is especially advantageous, because
touch voltage can be measured quickly and easily at any electrical
outlet without tripping the RCCB.
The usual, complex measuring method involving testing for the
proper functioning of the RCD at a given point, and subsequent
substantiation that all other systems components requiring protection are reliably connected at low resistance values to the
selected measuring point via the PE conductor, is made unnecessary.
N-PE Reversal Test
Additional testing is conducted in order to
determine whether or not N and PE are
reversed. The pop-up window shown at
the right appears in the event of reversal.
2) Tripping Test after the Measurement of Touch Voltage
➭
Press the
The tripping test need
only be performed at
one measuring point for
each RCCB.
If the RCCB is not tripped at nominal residual current,
the MAINS/NETZ LED blinks red (line voltage disconnected) and,
amongst other values, time to trip t
appear at the display panel.
If the RCCB is not tripped at nominal residual current,
the RCD/FI LED lights up red.
I
key.
N
and earthing resistance RE
a
In order to prevent the loss of data in data processing
systems, perform a data backup before starting the
measurement and switch off all consumers.
Start Measurement
Amongst other values, touch voltage U
resistance R
appear at the display panel.
E
The measured earthing resistance value RE is acquired
with very little current. More accurate results can be
obtained with the selector switch in the R
The DC + function can be selected here for systems with RCCBs.
and calculated earthing
IN
position.
E
Touch Voltage Too High
If touch voltage U
residual current I
U
LED lights up red.
L/RL
If the limit value for touch voltage is exceeded during the measurement process, U
for Germany (65 V applies for Austria – standard: ÖVE/ÖNORM E
8001-1 section 5.3).
Safety shutdown: At up to 70 V, a safety shutdown is
tripped within 3 s in accordance with IEC 61010.
Touch voltages of up to 70 V are displayed. If the value is greater
than 70 V, U
IN
, which has been measured with 1/3 nominal
IN
and extrapolated to IN, is > 50 V (> 25 V), the
N
> 50 V (> 25 V), safety shutdown occurs
IN
> 70 V is displayed.
Limit Values for Permissible, Continuous Touch Voltage
The limit for permissible, continuous touch voltage is equal to
U
= 50 V for alternating voltages (international agreement).
L
Lower values have been established for special applications (e.g.
medical applications: U
If touch voltage is too high, or if the RCCB is not tripped,
the system must be repaired (e.g. earthing resistance is
too high, defective RCCB etc.)!
=25V).
L
Unintentional Tripping of the RCD due to Bias Current in the System
Any bias current which might occur can be ascertained as
described in section 18.1 on page 80 with the help of a current
clamp transformer. The RCCB may be tripped during the testing
of touch voltage if extremely large bias currents are present within
the system, or if a test current was selected which is too great for
3-Phase Connections
For proper RCD testing at three-phase connections, the tripping
test must be conducted for one of the three phase conductors
(L1, L2 or L3).
the RCCB.
After touch voltage has been measured, testing can be performed
to determine whether or not the RCCB is tripped within the
selected time limit values at nominal residual current.
Unintentional Tripping of the RCD due to Leakage Current in the Measuring Circuit
Measurement of touch voltage with 30% nominal residual current
does not normally trip an RCCB. However, the trip limit may be
exceeded as a result of leakage current in the measuring circuit,
e.g. due to interconnected consumers with EMC circuit, e.g. fre-
Inductive Power Consumers
Voltage peaks may occur within the measuring circuit if inductive
consumers are shut down during an RCCB trip test. If this is the
case, the test instrument might not display any measured value
(– – – ). If this message appears, switch all consumers off before
performing the trip test. In extreme cases, one of the fuses in the
test instrument may blow, and/or the test instrument may be
damaged.
quency converters or PCs.
Gossen Metrawatt GmbH45
Page 46
12.2Special Tests for Systems and RCDs
Note
Attention!
!
I
F
Nom. res. current: 10 … 500 mA
Type 1: RCD, SRCD, PRCD …
Nominal current: 6 … 125 A
Type 2*: AC , A , F
, B, B+
, EV, MI
k
* Types B, B+, EV, MI =
AC/DC sensitive
Sinusoidal
Negative/positive half-wave
Waveform:
Connection:
Without/with probe
System type:
TN/TT, IT
Negative/positive direct current
Touch voltage:
Tripping limit values:
12.2.1 Testing Systems and RCCBs with Rising Residual Current
(AC) for Type AC, A/F, B/B+ and EV/MI RCDs
(PROFITEST MTECH+, PROFITEST MXTRA only)
Measuring Method
The instrument generates a continuously rising residual current of
(0.3 … 1.3) × I
The instrument stores the touch voltage and tripping current values which were measured at the moment tripping of the RCCB
occurred, and displays them.
One of the touch voltage limit values, U
can be selected for measurement with rising residual current.
within the system for the testing of RCDs.
N
=25V or UL=50/65V,
L
Select Measuring Function
Connection
Start Measurement
Set Parameters for I
Measuring Procedure
After the measuring sequence has been started, the test current
generated by the instrument is continuously increased starting at
0.3 times nominal residual current, until the RCCB is tripped. This
can be observed by viewing gradual filling of the triangle at I.
If touch voltage reaches the selected limit value (U
F
or 25 V) before the RCCB is tripped, safety shutdown occurs. The
UL/RL LED lights up red.
Safety shutdown: At up to 70 V, safety shutdown is triggered within 3 s in accordance with IEC 61010.
If the RCCB is not tripped before rising current reaches nominal
residual current I
If bias current is present within the system during measurement, it’s superimposed onto the residual current
which is generated by the instrument and influences
measured values for touch voltage and tripping current.
See also section 12.1.
, the RCD/FI LED lights up red.
N
=65V, 50V
L
Evaluation
According to IEC 60364-6, rising residual current must, however,
be used for measurements in the evaluation of RCDs, and touch
voltage at nominal residual current IN must be calculated from
the measured values.
The faster, more simple measuring method should thus be taken
advantage of (see section 12.1).
46Gossen Metrawatt GmbH
Page 47
12.2.2 Testing Systems and RCCBs with Rising Residual Current
Note
Note
S
I
N
Negative direct current
Positive direct current
Waveform:
180°: Start with neg. half-wave
0°: Start with pos. half-wave
5 times tripping current
X times tripping current:
(AC) for Type B/B+ and EV/MI RCDs (PROFITEST
MTECH+PROFITEST MXTRA)
In accordance with IEC 61557 / EN 61557, it must be substantiated that, with smooth direct current, residual operating current is
no more than twice the value of rated residual current IN. A continuously rising direct current, beginning with 0.2 times rated
residual current IN, must be applied to this end. If current rise is
linear, rising current may not exceed twice the value of I
period of 5 seconds.
Testing with smoothed direct current must be possible in both
test current directions.
within a
N
12.2.3 Testing RCCBS with 5 × IN
Measurement of time to trip is performed here with 5 times nominal residual current.
Measurements performed with 5 times nominal fault current are required for testing type and G RCCBs in the
manufacturing process. They’re used for personal safety
as well.
Measurement can be started with the positive half-wave at “0°” or
with the negative half-wave at “180°”.
Both measurements must nevertheless be performed. The longer
of the two tripping times is decisive regarding the condition of the
tested RCCB. Both values must be less than 40 ms.
Start Measurement
Select Measuring Function
Set Parameter – Start with Positive or Negative Half-Wave
Set Parameter – 5 Times Nominal Current
Gossen Metrawatt GmbH47
The following restrictions apply to the selection of
tripping current multiples relative to nominal current:
500 mA: 1 × IN, 2 × I
N
Page 48
12.2.4 Testing of RCCBs
Note
Note
Note
I
N
Neg. half-wave
Pos. half-wave
Negative direct current
Positive direct current
Waveform:
X times tripping current:
50% IN*
* No-trip test
with 50% I
N
S
I
N
I
F
or
Type 1:
which are Suitable for Pulsating DC Residual Current
In this case, RCCBs can be tested with either positive or negative
half-waves. The standard calls for tripping at 1.4 times nominal
current.
Select Measuring Function
Set Parameter – Positive or Negative Half-Wave
Set Parameter – Test With and Without “No-Trip Test”
12.3Testing of Special RCDs
12.3.1 Systems with Type RCD-S Selective RCCBs
Selective RCCBs are used in systems which include two series
connected RCCBs which are not tripped simultaneously in the
event of a fault. These selective RCCBs demonstrate delayed
response characteristics and are identified with the symbol.
Measuring Method
The same measuring method is used as for standard RCCBs (see
sections 12.1 on page 44 and 12.2.1 on page 46).
If selective RCDs are used, earthing resistance may not exceed
half of the value for standard RCCBs.
For this reason, the instrument displays twice the measured value
for touch voltage.
Select Measuring Function
Set Parameter – Selective
Start Measurement
No-Trip Test
If, during the no-trip test which lasts for 1 second, the RCD trips too early at 50% IN, i.e.
before the actual tripping test starts, the pop-up
window shown at the right appears.
The following restrictions apply to the selection of tripping
current multiples relative to nominal current:
500 mA: double and five-fold nominal current is not possible in this case.
Tripping Test
➭ Press the IN key. The RCCB is tripped. Blinking bars appear
According to DIN EN 50178 (VDE 160), only type B
RCCBs (AC-DC sensitive) can be used for equipment
with > 4 kVA, which is capable of generating smooth DC
residual current (e.g. frequency converters).
Tests with pulsating DC fault current only are not suitable
for these RCCBs. Testing must also be conducted with
smooth DC residual current in this case.
Measurement is performed with positive and negative
half-waves for testing RCCBs during manufacturing. If a
circuit is charged with pulsating direct current, the function of the RCCB can be executed with this test in order
to assure that the RCCB is not saturated by the pulsating
direct current so that it no longer trips.
48Gossen Metrawatt GmbH
at the display panel, after which time to trip t
sistance R
The tripping test need
only be performed at
one measuring point for
each RCCB.
teristics. Tripping performance is briefly influenced (up to
30 s) due to pre-loading during measurement of touch
voltage. In order to eliminate pre-charging caused by the
measurement of touch voltage, a waiting period must be
observed prior to the tripping test. After the measuring
sequence has been started (tripping test), blinking bars
are displayed for approximately 30 seconds. Tripping
times of up to 1000 ms are permissible. The tripping test
is executed immediately after once again pressing the I
key.
N
Set Parameter – PRCD with Non-Linear Elements
12.3.2 PRCDs with Non-Linear Type PRCD-K Elements
The PRCD-K is a portable RCD with electronic residual current
evaluation laid out as an inline device which switches all poles (L,
N and PE). Undervoltage tripping and protective conductor monitoring are additionally integrated into the PRCD-K.
The PRCD-K is equipped with undervoltage tripping, for which
reason it has to be operated with line voltage, and measurements
may only be performed in the on state (PRCD-K switches all
poles).
Terminology (from DIN VDE 0661)
Portable protective devices are circuit breakers which can be connected between power consuming devices and permanently
installed electrical outlets by means of standardized plug-andsocket devices.
A reusable, portable protective device is a protective device which
is designed such that it can be connected to movable cables.
Please be aware that a non-linear element is usually integrated
into PRCDs, which leads to immediate exceeding of the greatest
permissible touch voltage during U
than 50 V).
PRCDs which do not include a non-linear element in the protective conductor must be tested in accordance with section 12.3.3
on page 50.
measurements (UI greater
I
Objective (from DIN VDE 0661)
Portable residual current devices (PRCDs) serve to protect persons and property. They allow for the attainment of increased levels of protection as provided by protective measures utilized in
electrical systems for the prevention of electrical shock as defined
in DIN VDE 0100-410. They are to be designed such that they
can be installed by means of a plug attached directly to the protective device, or by means of a plug with a short cable.
Measuring Method
The following can be measured, depending upon the measuring
method:
•Time to trip tA: tripping test with nominal residual current I
(the PRCD-K must be tripped at 50% nominal current)
•Tripping current I for testing with rising residual current I
N
F
Start Measurement
Select Measuring Function
Connection
Gossen Metrawatt GmbH49
Page 50
12.3.3 SRCD, PRCD-S (SCHUKOMAT, SIDOS or comparable)
Note
I
N
I
F
or
Type 1:
I
N
Type 1:
Waveform:
Negative direct current
Positive direct current
180°: Start with neg. half-wave
0°: Start with pos. half-wave
RCCBs from the SCHUKOMAT SIDOS series, as well as others
which are of identical electrical design, must be tested after
selecting the corresponding parameter.
Monitoring of the PE conductor is performed for RCDs of this
type. The PE conductor is monitored by the summation current
transformer. If residual current flows from L to PE, tripping current
is cut in half, i.e. the RCCB must be tripped at 50% nominal residual current I
Whether or not PRCDs and selective RCDs are of like design can
be tested by means of touch voltage U
touch voltage U
of an otherwise error-free system, the PRCD more than likely contains a non-linear element.
.
N
measurement. If a
of greater than 70 V is measured at the PRCD
IN
IN
PRCD-S
The PRCD-S (portable residual current device – safety) is a special, portable, protective device with protective conductor detection or protective conductor monitoring. The device serves to protect persons from electrical accidents in the low-voltage range
(130 to 1000 V). The PRCD-S must be suitable for commercial
use, and is installed like an extension cable between an electrical
consumer – as a rule an electrical tool – and the electrical outlet.
Select Measuring Function
12.3.4 Type G or R RCCB
In addition to standard RCCBs and selective RCDs, the special
characteristics of the type G RCCB can also be tested with the
test instrument.
The type G RCCB is an Austrian specialty which complies with
device standard ÖVE/ÖNORM E 8601. Erroneous tripping is minimized thanks to its greater current carrying capacity and shortterm delay.
Select Measuring Function
Set Parameter – Type G/R (VSK)
Set Parameter – SRCD / PRCD
Start Measurement
Touch voltage and time to trip can be measured in the G/R-RCD
switch position.
It must be observed that time to trip for type G RCCBs
may be as long as 1000 ms when measurement is made
at nominal residual current. Set the limit value correspondingly.
➭ Then select 5 × I
for the G/R setting) and repeat the tripping test beginning with
the positive half-wave at 0° and the negative half-wave at
180°. The longer of the two tripping times is decisive regarding the condition of the tested RCCB.
in the menu (this is selected automatically
N
Set Parameter – Start with Positive or Negative Half-Wave
50Gossen Metrawatt GmbH
Page 51
Set Parameter – 5 Times Nominal Current
Note
Note
5 times tripping current
S
U
IN
= RE×IN = 1 × 30 mA = 30 mV = 0.03 V
The following restrictions apply to the selection of
tripping current multiples relative to nominal current:
500 mA: 1 ×, 2 × IN
Start Measurement
12.4Testing Residual Current Circuit Breakers in TN-S Systems
Connection
RCCBs can only be used in TN-S systems. An RCCB would not
work in a TN-C system because PE is directly connected to the
neutral conductor in the outlet (it does not bypass the RCCB).
This means that residual current would be returned via the RCCB
and would not generate any differential current, which is required
in order to trip the RCCB.
As a rule, the display for touch voltage is also 0.1 V, because the
nominal residual current of 30 mA together with minimal loop
resistance result in a very small voltage value:
In both cases tripping
time must be between
10 ms (minimum delay
time for type G RCCBs!)
and 40 ms.
Type G RCCBs with other nominal residual current values must
be tested with the corresponding parameter setting under menu
item I
. In this case as well, the limit value must be appropriately
N
adjusted.
The RCD parameter setting for selective RCCBs is not
suitable for type G RCCBs.
Gossen Metrawatt GmbH51
Page 52
12.5Testing of RCD Protection in IT Systems with High Cable
Note
System type:
I
N
I
F
or
Time to trip
6mA
60 mA
200 mA
10.0 s
0.3 s
0.1 s
Capacitance (e.g. In Norway)
The desired system type (TN/TT or IT) can be selected for RCD
test type U
A probe is absolutely essential for measurement in IT systems,
because touch voltage U
not otherwise be measured.
After selecting the IT system setting, connection with probe is
selected automatically.
(IN, ta), and for earthing measurement (RE).
IN
which occurs in these systems can-
IN
Set Parameter – Select System Type
Start Measurement
12.6Testing of 6 mA Residual Current Devices RDC-DD/RCMB
(RDC-DD: PROFITEST MXTRA and PROFITEST MTECH+
only)
DIN VDE 0100-722 (Requirements for special installations or
locations – Supplies for electric vehicles) specifies that all outlets
for charging electric vehicles must be protected by a separate
residual current device (RCD).
Furthermore, additional protection is required for multiphase
charging with smooth DC fault current. Either a type B RCD, an
RDC-DD (residual direct current detecting device) or an RCMB
(residual current monitoring module) can be used to this end.
Select Measuring Function
Set Parameter – Type RDC
Set Parameter – Time to Trip
The RDC-DD is tested with nominal residual currents of 6
to 200 mA.
Start Measurement
52Gossen Metrawatt GmbH
Page 53
Set Parameter – Type RCMB
Note
Time to trip
300 mA
6mA
60 mA
0.04 s
10.0 s
0.3 s
Set Parameter – Time to Trip
The RCMB is tested with nominal residual currents of 6 to
300 mA.
Start Measurement
Gossen Metrawatt GmbH53
Page 54
13Testing of Breaking Requirements for Overcurrent Protective Devices, Measurement of Loop Impedance
Note
Note
Note
Note
Z
L-PE
Start
t1t3
Measure
t2
Operation
RCD Disabled!
t
I
F
/mA
Suppression of RCCB tripping for
RCCBs which are sensitive to pulsed
current
and Determination of Short-Circuit Current (ZL-PE and ISC Functions)
Testing of overcurrent protective devices includes visual inspection and measurement.
Measuring Method
Loop impedance Z
is ascertained in order to determine if the breaking requirements
for protective devices have been fulfilled.
Loop impedance is the resistance within the current loop (utility
station – phase conductor – protective conductor) when a shortcircuit to an exposed conductive part occurs (conductive connection between phase conductor and protective conductor). Shortcircuit current magnitude is determined by the loop impedance
value. Short-circuit current I
value set forth by IEC 60364, so that reliable breaking of the protective device (fuse, automatic circuit breaker) is assured.
The measured loop impedance value must therefore be less than
the maximum permissible value.
Tables containing permissible display values for loop impedance
and minimum short-circuit current display values for ampere ratings for various fuses and circuit breakers can be found in the
help texts and in section 25 from page 100. Maximum device
error in accordance with IEC 61557 / EN 61557has been taken
into consideration in these tables. See also section 13.2.
In order to measure loop impedance Z
test current of 3.7 to 7 A (60 to 550 V) depending on line voltage
and line frequency. At 16 Hz, The test has a duration of no more
than 1200 ms.
is measured and short-circuit current ISC
L-P E
may not fall below a predetermined
SC
, the instrument uses a
L-P E
If the limit value for touch voltage is exceeded during this measurement process (> 50 V), safety shutdown occurs for Germany
(65 V applies for Austria – standard: ÖVE/ÖNORM E 8001-1
section 5.3).
The shutdown value can be adjusted within a range of 25 to 65 V
(see section 10.8).
The test instrument calculates short-circuit current ISC based on
measured loop impedance
current calculation is made with reference to nominal line voltage
for line voltages which lie within the nominal ranges for 120, 230
and 400 V systems. This also applies between phases L-L at
500 V. If line voltage does not lie within these nominal ranges, the
instrument calculates short-circuit current I
vailing line voltage and measured loop impedance Z
Z
and line voltage. Short-circuit
L-P E
based upon pre-
SC
L-P E
.
Select Measuring Function
Connection
Schuko / 3-Pole Adapter
Connection
2-Pole Adapter
Loop impedance should be measured for each electrical
circuit at the farthest point, in order to ascertain maximum loop impedance for the system.
Observe national regulations, e.g. the necessity of conducting measurements without regard for RCCBs in
Austria.
3-Phase Connections
Measurement of loop impedance to earth must be performed at
all three phase conductors (L1, L2, and L3) for the testing of overcurrent protective devices at three phase outlets.
13.1Measurements with Suppression of RCD Tripping
(PROFITEST MTECH+, PROFITEST MXTRA only)
The test instruments make it possible to measure loop impedance
in TN systems with type A, F and AC
100, 300, 500 mA nominal residual current).
The test instrument
generates a direct
current to this end,
which saturates the
RCCB’s magnetic
circuit.
The test instrument
then superimposes
a measuring current which only
demonstrates halfwaves of like polarity. The RCCB is no
longer capable of
detecting this measuring current and is
consequently not
tripped during measurement.
A four conductor measuring cable is used between the instrument and the test plug. Cable and measuring adapter resistance
is automatically compensated during measurement and does not
affect measurement results.
RCCBs(10, 30,
Loop impedance measurement in accordance with the
procedure for the suppression of RCCB tripping is only
possible with type A and F RCDs.
Bias Magnetization
54Gossen Metrawatt GmbH
Only AC measurements can be performed with the 2pole adapter. Suppression of RCD tripping by means of
bias magnetization with direct current is only possible via
a country-specific plug insert, e.g. SCHUKO, or the 3pole adapter (neutral conductor N required).
Page 55
13.1.1 Measurement with Positive Half-Waves (PROFITEST
Z
L-PE
Nom. current: 2 … 160 A, … 9999 A
Tripping characteristics:
Diameter*: 1.5 … 70 mm
Cable type*: NY…, H07…
Number of wires*: 2 … 10-wire
A, B/L, C/G, D, E, H, K, GL/GG & factor
Sinusoidal
15 mA sinusoidal
Waveform:
DC-L and positive half-wave
Touch voltage:
DC-H and positive half-wave
2-pole
Measurement with country-specific
plug insert (e.g. Schuko)
Note
Selecting the test probe, as well
as Lx-PE reference or AUTO, is
only relevant with regard to
report generation.
Semiautomatic measurement
See also section 10.9 regarding the AUTO parameter.
Polarity Selection
measurement
MTECH+, PROFITEST MXTRA only)
Measurement by means of half-waves plus direct current makes it
possible to measure loop impedance in systems which are
equipped with RCCBs.
In the case of DC measurement with half-waves, selection can be
made between two variants:
DC-L: Reduced bias current
but faster measurement as a result
DC-H: Higher bias current providing more reliability with regard
to non-tripping of the RCD
Select Measuring Function
Set Parameters
Start Measurement
* Parameters used for report generation and do not influence the mea-
surement
Sinusoidal (full-wave)Setting for circuit without RCD
15 mA sinusoidalSetting for motor protection switch only
played after allowance
has been made for the
instrument’s maximum
measuring and intrinsic
uncertainties (under normal measuring conditions) can be determined
with the help of Table 1
on page 100. Intermediate values can be interpolated.
The maximum permissible nominal current for
the protective device (fuse or circuit breaker) for a line voltage of
230 V after allowance has been made for maximum measuring
error can be determined with the help of Table 6 on page 101
based on measured short-circuit current (corresponds to IEC
60364-6).
L-
Page 56
Special Case: Suppressing Display of the Limit Value
Limit Value:
ISC < Limit Value
UL R
L
The limit value cannot be
ascertained. The inspector is prompted to evaluate the measured values himself, and to
acknowledge or reject
them with the help of the
softkeys.
Measurement passed:
. key
Measurement failed:
key
The measured value can
only be saved after it has
been evaluated.
13.3Settings for Calculating Short-Circuit Current
– Parameter I
Short-circuit current ISC is used to test shutdown by means of an
overcurrent protective device. In order for an overcurrent protective device to be tripped on time, short-circuit current ISC must be
greater than tripping current Ia (see table 6 in section 25.1). The
variants which can be selected with the “Limits” key have the following meanings:
I
:IaThe measured value displayed for ISC is used
SC
I
:Ia+% The measured value displayed for Z
SC
I
:2/3 ZIn order to calculate ISC, the measured value dis-
SC
I
:3/4 ZZ
SC
SC
without any correction to calculate Z
rected by an amount equal to the test instrument’s
L-P E
L-P E
.
is cor-
measuring and intrinsic uncertainties in order to
calculate I
played for Z
sponding to all possible deviations (these are
.
SC
is corrected by an amount corre-
L-PE
defined in detail by IEC 60364-6 as
Z
2/3 × U0/Ia).
s(m)
3/4 × U0/I
s(m)
a
ZLoop impedance
I
Short-circuit current
SC
UMomentary voltage at the test probes, “UN” is displayed if
U
deviates from nominal voltage by 10%
max.
fFrequency of the applied voltage,
“fN” is displayed if frequency f max. deviates from nominal
frequency by 1%
Tripping current
I
a
(see data sheet for circuit breakers / fuses)
% Test instrument intrinsic error
56Gossen Metrawatt GmbH
Page 57
14Measuring Supply Impedance (Z
Z
L-N
Diameters: 1.5 … 70 mm
Cable type: NY…, H07…
Number of wires: 2 … 10-wire
Tripping characteristics:
A, B/L, C/G, D, E, H, K, GL/GG & factor
Semiautomatic Measurement
See also section 10.9 regarding the AUTO parameter.
L-PE relationships are not
possible here.
Polarity Selection
Limit Value:
ISC < Limit Value
UL R
L
I
SC
Function)
L-N
Measuring Method (internal line resistance measurement)
Supply impedance Z
method used for loop impedance Z
54). However, the current loop is completed via neutral conductor
N rather than protective conductor PE as is the case with loop
impedance measurement.
is measured by means of the same
L-N
(see section 13 on page
L-P E
Select Measuring Function
Connection
Schuko
Connection
2-Pole Adapter
Set Parameters
Settings for Calculating Short-Circuit Current – Parameter I
SC
Short-circuit current ISC is used to test shutdown by means of an
overcurrent protective device. In order for an overcurrent protective device to be tripped on time, short-circuit current I
greater than tripping current I
variants which can be selected with the “Limits” key have the fol-
(see table 6 in section 25.1). The
a
must be
SC
lowing meanings:
I
:IaThe measured value displayed for ISC is used
SC
I
:Ia+% The measured value displayed for Z
SC
without any correction to calculate Z
rected by an amount equal to the test instrument’s
L-N
L-N
.
is cor-
measuring and intrinsic uncertainties in order to
calculate I
I
:2/3 ZIn order to calculate ISC, the measured value dis-
SC
played for Z
sponding to all possible deviations (these are
.
SC
is corrected by an amount corre-
L-N
defined in detail by IEC 60364-6 as
Z
2/3 × U0/Ia).
I
:3/4 ZZ
SC
s(m)
3/4 × U0/I
s(m)
a
Gossen Metrawatt GmbH57
back and forth between the country-specific plug
insert, e.g. SCHUKO, and the 2-pole adapter. The
selected connection type is displayed inversely
(white on black).
Press the softkey shown at the left in order to switch
ZLoop impedance
I
Short-circuit current
SC
UMomentary voltage at the test probes, “UN” is displayed if
U
deviates from nominal voltage by 10%
max.
fFrequency of the applied voltage,
“fN” is displayed if frequency f
deviates from nominal
max.
frequency by 1%
Tripping current
I
a
(see data sheet for circuit breakers / fuses)
%Test instrument intrinsic error
Page 58
Start Measurement
Display of U
(UN / fN)
L-N
If the measured voltage value lies within a range of 10% of the
respective nominal line voltage of 120, 230 or 400 V, the respectively corresponding nominal line voltage is displayed. In the case
of measured values outside of the 10% tolerance, the actual
measured value is displayed.
Displaying the Fuse Table
After measurement has been performed, permissible fuse types
can be displayed by pressing the HELP key.
The table shows maximum permissible nominal current dependent on fuse type and breaking requirements.
Key: I
= breaking current, ISC = short-circuit current, IN = nominal
a
current, t
58Gossen Metrawatt GmbH
= time to trip
a
Page 59
15Earthing Resistance Measurement (Function RE)
Note
Attention!
!
Note
Start
t1
t3
Measure
t2
Operation
RCD Disabled!
t
I
F
/mA
Suppression of RCCB tripping for
RCCBs which are sensitive to pulsed
current
Earthing resistance RE is important for automatic shutdown in
system segments. It must have a low value in order to assure that
high short-circuit current flows and the system is shut down reliably by the RCCB in the event of a fault.
Test Se tu p
Earthing resistance (RE) is the sum of the earth electrode’s dissipation resistance and earth conductor resistance. Earthing resistance is measured by applying an alternating current via the earth
conductor, the earth electrode and dissipation resistance. This
current, as well as voltage between the earth electrode and a
probe, are measured.
The probe is connected to the probe connector socket (17) with a
4 mm contact protected plug.
Direct measurement with probe (mains powered earthing measurement)
Direct measurement of earthing resistance RE is only possible
within a measuring circuit which includes a probe. However, this
means that the probe and reference earth must be of like potential, i.e. that they are positioned outside of the potential gradient
area. The distance between the earth electrode and the probe
should be at least 20 m.
Measurement without probe (mains powered earthing measurement)
In many cases, especially in extremely built-up areas, it’s difficult,
or even impossible, to set a measuring probe. In such cases,
earthing resistance can be measured without a probe. In this
case, however, the resistance values for the operational earth
electrode RB and phase conductor L are also included in the
measurement results.
The instrument measures earthing resistance RE by means of the
ammeter-voltmeter test.
Resistance RE is calculated from the quotient of voltage UE and
current I
The test current which is applied to earthing resistance is controlled by the instrument (see section 5.5, “Technical Data”, on
page 10 for pertinent values).
A voltage drop is generated which is proportional to earthing
resistance.
Gossen Metrawatt GmbH59
where UE is between the earth electrode and the probe.
E
Measurement cable and measuring adapter resistance
are compensated automatically during measurement and
have no effect on measurement results.
If dangerous touch voltages occur during measurement
(> 50 V), the measurement is interrupted and safety shutdown occurs.
Probe resistance does not affect measurement results
and may be as high as 50 k.
The probe is part of the measuring circuit and may carry
a current of up to 3.5 mA in accordance with IEC 61557
/ EN 61557.
Measurement with or without earth electrode voltage depending
upon entered parameters and the selected type of connection:
RANGEConnectionMeasuring
Functions
xx / xx k
10 / U
xx / xx k *
* This parameter results in automatic selection of probe connection.
No probe measurement
*
E
No U
Probe measurement
activated
U
E
Probe measurement
activated
No U
Clamp measurement
activated
No U
measurement
E
is measured
measurement
E
measurement
E
Measuring method with suppression of RCD tripping
(mains powered earthing measurement) (PROFITEST MTECH+,
PROFITEST MXTRA only)
The test instrument makes it possible to measure earthing resistance in TN systems with type A, F and AC
RCCBs
(10, 30, 100, 300, 500 mA nominal residual current).
The test instrument
generates a direct
current to this end,
which saturates the
RCCB’s magnetic
circuit.
The test instrument
then superimposes
a measuring current
which only demonstrates half-waves
of like polarity. The
RCCB is no longer
capable of detecting this measuring
current, and is consequently not
tripped during measurement.
A four conductor measuring cable is used between the instrument and the test plug. Cable and measuring adapter resistance
is automatically compensated during measurement and does not
affect measurement results.
Bias Magnetization
Only AC measurements can be performed with the 2pole adapter. Suppression of RCD tripping by means of
bias magnetization with direct current is only possible via
a country-specific plug insert, e.g. SCHUKO, or the 3pole adapter (neutral conductor N required).
Limit Values
Earthing resistance (earth coupling resistance) is determined primarily by the electrode’s contact surface and the conductivity of
the surrounding earth.
The specified limit value depends on the type of electrical system
and its shutdown conditions in consideration of maximum touch
voltage.
Evaluating Measured Values
The maximum permissible displayed resistance values which
assure that the required earthing resistance is not exceeded, and
for which maximum device operating error has already been
taken into consideration (at nominal conditions of use), can be
determined with the help of Table 2 on page 100. Intermediate
values can be interpolated.
The 5 following types of measurement and connection are possible:
• 3-wire measurement via PRO-RE adapter
• 4-wire measurement via PRO-RE adapter
• Selective measurement: 2-pole measurement
with probe
and current clamp sensor
At left in figure:
2-pole measuring
adapter for contacting PE and L
measuring points
At right in figure:
The PRO-Schuko
measuring
adapter can be
used as an alternative.
Select Measuring Function
Select Operating Mode
• Selective measurement with clamp meter (4pole)
via PRO-RE adapter
• 2-clamp measurement via PRO-RE/2 adapter
• Measurement of soil resistivity
via PRO-RE adapter
Figure at right:
PRO-RE adapter for connect-
ing earth electrode, auxiliary
earth electrode, probe and
auxiliary probe to the test
instrument for
3/4-pole measurement,
selective measurement and
measurement of soil resistivity
Figure at right:
PRO-RE/2 measuring adapter as
accessory for connecting the EClip 2 generator clamp for 2-clamp
measurement and earth loop resistance measurement.
E
The selected operating mode is displayed inversely:
mains~ in white against a black background.
Special Case: Manual Measuring Range Selection (test current
selection)
When the measuring range is selected manually, accuracy values are only valid starting at 5% of the upper limit range value
(except for the 10 W range; separate display for small values).
)
E
Set Parameters
❏ Measuring range: AUTO
10 k (4 mA), 1 k (40 mA), 100 (0.4 A), 10 (> 3.7 A)
In systems with RCCBs, resistance or test current must be selected such that it is less than tripping current (½ I
❏ Touch voltage: UL < 25 V, < 50 V, < 65 V, see section 10.8 re-
garding freely selectable voltage
❏ Transformer ratio: depends on utilized current clamp sensor
❏ Connection: 2-pole adapter, 2-pole adapter + probe,
2-pole adapter + clamp meter
❏ System type: TN or TT
❏ Test current waveform
See section 15.4 through section 15.6 regarding advisable
parameters for the respective measurement and connection
types.
N
).
Performing Measurements
See section 15.4 through section 15.6.
Select Measuring Function
Select Operating Mode
The selected operating mode is displayed inversely:
white battery icon against black background.
Set Parameters
❏ Measuring range: AUTO, 50 k, 20 kW, 2 kW, 200 W, 20 W
❏ Current clamp sensor transformer ratio:
In the event that it’s impossible to set a probe, earthing resistance
can be estimated by means of an “earth loop resistance measurement” without probe.
The measurement is performed exactly as described in section
15.4, “Earthing Resistance Measurement. Mains Powered – 3-Pole
Measurement: 2-Pole Adapter with Probe”, on page 62. However,
no probe is connected to the probe connector socket (17).
The resistance value R
method also includes operational earth electrode resistance R
obtained with this measuring
ELoop
and resistance at phase conductor L. These values must be subtracted from the measured value in order to determine earthing
resistance.
If conductors of equal cross section are assumed (phase conductor L and neutral conductor N), phase conductor resistance is half
as great as supply impedance Z
conductor). Supply impedance can be measured as described in
section 14 from page 57. In accordance with IEC 60364, the
operational earth electrode R
2 ”.
10 k (4 mA), 1 k (40 mA), 100 (0.4 A), 10 (3.7 … 7 A)
In systems with RCCBs, resistance or test current must be selected such that it is less than tripping current (½ I
This measurement is only possible with a probe (see section
15.4). Earth electrode potential U
the earth electrode between the earth electrode terminal and reference earth if a short-circuit occurs between the phase conductor and the earth electrode. The measurement of earth electrode
potential is required by Swiss standard NIV/NIN SEV 1000.
is the voltage which occurs at
E
Measuring Method
In order to determine earth electrode potential the instrument first
measures earth electrode loop resistance R
ately thereafter earthing resistance R
values and then calculates earth electrode potential with the following equation:
The calculated value is displayed at the display panel.
. The instrument stores both
E
, and immedi-
ELoop
Select Measuring Function
Select Operating ModeSelect Measuring Range
Set Parameters
❏ Measuring range: 10 / U
❏ Connection: 2-pole adapter + probe
❏ Touch voltage: UL < 25 V, < 50 V, < 65 V, see section 10.8
regarding freely selectable voltage
❏ Test current waveshape: sinusoidal only in this case (full-wave)!
❏ System type: TN/TT, IT
❏ Transformer ratio: irrelevant in this case
E
Start Measurement
Connection
The following diagram appears if the
2-pole adapter is connected incorrectly.
To be connected: 2-pole adapter and probe
Gossen Metrawatt GmbH63
Page 64
15.6Earthing Resistance Measurement, Mains Powered – Selective Earthing Resistance Measurement with Current Clamp Sensor
T
E
S
T
I
N
S
T
R
U
M
E
N
T
W
a
t
e
r
P
i
p
e
SE
2
E
1
B
U
Probe
I
Clamp
-----------------
R
E
as Accessory
As an alternative to the conventional measuring method, measurement can also be performed with a current clamp sensor.
1 k (40 mA), 100 (0.4 A), 10 (3.7 … 7 A)
In the case of systems with RCCBs, the DC offset and positive
half-wave (DC + ) functions can be selected (only in the
10 range and only with the METRAFLEX P300).
❏ Connection: 2-pole adapter + clamp meter
After parameter selection: automatic setting to 10 measuring
range and 1 V/A or 1000 mV/A transformer ratio
❏ Touch voltage: UL < 25 V, < 50 V, < 65 V, for information regard-
ing freely selectable voltage see section 10.8
❏ Test current waveshape:
Sinusoidal (full-wave), DC offset and
positive half-wave (DC + )
E2
❏ System type: TN/TT, IT
❏ Current clamp sensor transformation ratio: see table below
Set Parameters at Current Clamp Sensor
❏ Current clamp sensor measuring range: see table below
Select Measuring Range at the Current Clamp Sensor
Test InstrumentMETRAFLEX P300 ClampTest Instrument
Parameters
Transformation Ratio
1:1
1 V / A
01:10
100 mV / A
1:100
10 mV / A
SwitchesMeasuring
Range
3 A (1 V/A)3A0.5 … 100 mA
30 A (100 mV/A)30 A5 … 999 mA
300 A (10 mV/A)300 A0.05 … 10 A
Measuring Range
Important Instructions for Use of the Current Clamp Sensor
• Use only the METRAFLEX P300 or the Z3512A current clamp
Connection
To be connected: 2-pole adapter, clamp and probe
64Gossen Metrawatt GmbH
sensor for this measurement.
• Read and adhere to the operating instructions for
the METRAFLEX P300 current clamp sensor, as well as the
safety precautions included therein.
• Observe direction of current flow (see arrow on the current
clamp sensor).
• Use the clamp in the permanently connected state. The sensor may not be moved during measurement.
• The current clamp sensor may only be used at an adequate
distance from powerful extraneous fields.
• Before use, always inspect the electronics housing, the connector cable and the current sensor for damage.
• In order to prevent electric shock, keep the surface of the
METRAFLEX clean and free of contamination.
• Before use, make sure that the flexible current sensor, the connector cable and the electronics housing are dry.
Page 65
Start Measurement
Note
Note
In the event that you have changed the transformation ratio at the
test instrument, a pop-up window appears indicating that this
new setting also has to be entered to the connected current
clamp sensor.
Note regarding
currently
selected transformation
at the tester
ratio
RE
RE
: Selective earthing resistance measured via clamp
Clamp
: Total earthing resistance measured via probe,
Probe
comparative value
The following diagram appears if the
2-pole adapter is connected incorrectly.
The selected operating mode is displayed inversely:
white battery icon against black background.
Set Parameters
❏ Measuring range: AUTO, 50 k, 20 k, 2 k, 200 , 20
❏ Connection: 3-pin
❏ Transformer ratio: irrelevant in this case
Earthing Resistance Measurement According to the 3-Wire
Method
❏ Distance d (for measuring
Start Measurement
): irrelevant in this case
E
Connection
➭ Position the spikes for the probe and the auxiliary electrode at
least 20, respectively 40 meters from the electrode (see figure
above).
➭ Make sure that no excessively high contact resistances occur
between the probe and the ground.
➭ Attach the PRO-RE adapter (Z501S) to the test plug.
➭ Connect the probe, the auxiliary electrode and the electrode
via the 4 mm banana plug sockets at the PRO-RE adapter.
In doing so, observe labeling on the banana plug sockets.
Terminal ES/P1 is not connected.
The resistance of the measurement cable to the earth electrode is
incorporated directly into the measurement results.
In order to keep error caused by measurement cable resistance
as small as possible, a short connector cable with large crosssection should be used between the earth electrode and terminal
E for this measuring method.
The measurement cables must be well insulated in order
to prevent shunting. In order to keep the influence of possible coupling to a minimum, the measurement cables
should not cross each other or run parallel to each other
over any considerable distance.
Earthing Resistance Measurement
According to the 4-Wire Method
R
E
4-Wire Method
Select Measuring Function
Select Operating Mode
Set Parameters
❏ Measuring range: AUTO, 50 k, 20 k, 2 k, 200 , 20
❏ Connection: 4-pin
❏ Transformer ratio: irrelevant in this case
❏ Distance d (for measuring
PROFITEST MPRO
The selected operating mode is displayed inversely:
white battery icon against black background.
&
PROFITEST MXTRA
): irrelevant in this case
E
only)
The 4-wire method is used in the case of high cable resistance
between the earth electrode and the instrument terminal.
The resistance of the cable between the earth electrode and the
“E” terminal at the instrument is measured in this case.
Connection
➭ Position the spikes
for the probe and the auxiliary electrode at least 20, respectively 40 meters from the electrode (see figure above).
➭ Make sure that no excessively high contact resistances occur
between the probe and the ground.
➭ Attach the PRO-RE adapter (Z501S) to the test plug.
➭ Connect the probes, the auxiliary electrode and the electrode
via the 4 mm banana plug sockets at the PRO-RE adapter.
In doing so, observe labeling on the banana plug sockets.
In the case of the 4-wire method, the earth electrode is
connected to the “E” and “ES” terminals with two separate
measurement cables, the probe is connected to the “S”
terminal and the auxiliary earth electrode is connected to
the “H” terminal.
The measurement cables must be well insulated in order
to prevent shunting. In order to keep the influence of possible coupling to a minimum, the measurement cables
should not cross each other or run parallel to each other
over any considerable distance.
Start Measurement
Potential Gradient Area
Information regarding
suitable positioning of
the probe and the auxiliary earth electrode can be obtained by
observing voltage characteristics ordissipation resistance in the
ground.
The measuring current from the earth tester which flows via the
earth electrode and the auxiliary earth electrode causes a given
potential distribution in the form of a potential gradient area, (cf.
figure “Voltage Curve in Homogeneous Earth between Earth Electrode E and Auxiliary Earth Electrode H” on page 68.) Resistance
distribution is analogous to potential distribution.
Dissipation resistance of the earth electrode and the auxiliary
earth electrode differs as a rule. The potential gradient area and
the resistance gradient area are thus not symmetrical.
Dissipation Resistance of Small Scope Earth Electrodes
The arrangement of the probe and the auxiliary earth electrode
are very important for correct determination of the dissipation
resistance of earth electrodes.
The probe must be positioned between the earth electrode and
the auxiliary earth electrode within the so-called neutral zone (reference earth) (cf. figure “Probe Distance S Outside of the Overlapping Potential Gradient Areas on the Perpendicular Bisector of
Earth Electrode E and Auxiliary Earth Electrode H” on page 68.)
The voltage or resistance curve is thus nearly horizontal within the
neutral zone.
Proceed as follows in order to select suitable probe and auxiliary
earth electrode resistances:
➭ Drive the auxiliary earth electrode into the ground at a dis-
tance of roughly 40 meters from the earth electrode.
➭ Position the probe halfway between the earth electrode and
the auxiliary earth electrode and determine earthing resistance.
➭ Reposition the probe 2 3 m closer to the earth electrode,
and then 2 3 m closer to the auxiliary earth electrode and
measure earthing resistance in each position.
If all 3 measurements result in the same measured value, this is
the correct earthing resistance. The probe is in the neutral zone.
Gossen Metrawatt GmbH67
Page 68
However, if the three measured values for earthing resistance dif-
d = distance from electrode to aux. electrode
E = earth electrode
H = auxiliary earth electrode
I = measuring current
K = neutral zone (reference earth)
UE= earth potential
RE= UE / I = earthing resistance
= potential
I
I
d
E
H
U
E
K
Voltage Curve in Homogeneous Earth between
Earth Electrode E and Auxiliary Earth Electrode H
E = electrode location
H = aux. electrode location
S = probe location
S
HE
Probe Distance S Outside of the Overlapping
Potential Gradient Areas
on the Perpendicular
Bisector of Earth Electrode E and Auxiliary
Earth Electrode H
Curve I (KI)Curve II (KII)
mWmW
5
10
15
20
25
30
40
60
80
100
0.9
1.28
1.62
1.82
1.99
2.12
2.36
2.84
3.68
200
10
20
40
60
80
100
120
140
160
200
0.8
0.98
1.60
1.82
2.00
2.05
2.13
2.44
2.80
100
S1, S2 = inflection points
KI= curve I
KII= curve II
S1, S2 = inflection points
KI= curve I
KII= curve II
S
1
S
2
K I
K II
4
3
2
1
0
10 20 30 40 50 60 7080 90 100 m KI
20 40 60 80 100 120 140 160 180 200 m KII
5
R
A/H
R
A/E
0
0
S
HESE
Earthing Resistance Measurement for a Large Scope
Earthing System
fer from each other, either the probe is not located in the neutral
zone, or the voltage or resistance curve is not horizontal at the
point at which the probe has been inserted.
Correct measurements can be obtained in such cases by either
increasing distance between the earth electrode and the auxiliary
earth electrode, or by moving the probe to the perpendicular
bisector between the earth electrode and the auxiliary earth electrode (see figure below). When the probe is moved to the perpendicular bisector, its location is removed from the sphere of influence of the two potential gradient areas caused by the earth electrode and the auxiliary earth electrode.
➭ Measured resistance values are displayed as a table, and then
plotted graphically as depicted in “Earthing Resistance Measurement for a Large Scope Earthing System” on page 68.
(Curve I).
If a line parallel to the abscissa is drawn through inflection point
S1, this line divides the resistance curve into two parts.
Measured at the ordinate, the bottom part results in sought dissipation resistance of the earth electrode R
dissipation resistance of the auxiliary earth electrode R
With a measurement setup of this type, dissipation resistance of
, and the top value is
A/E
A/H
.
the auxiliary earth electrode should be less than 100 times the
dissipation resistance of the earth electrode.
In the case of resistance curves without a well-defined horizontal
area, measurement should be double checked after repositioning
the auxiliary earth electrode. This additional resistance curve must
be entered to the first diagram with a modified abscissa scale
such that the two auxiliary earth electrode locations are superimposed. The initially ascertained dissipation resistance value can
be checked with inflection point S2.
Notes Regarding Measurement in Difficult Terrain
In extremely unfavorable terrain (e.g. sandy soil after a lengthy
period without rain), auxiliary earth electrode and probe resistance
can be reduced to permissible values by watering the ground
around the auxiliary earth electrode and the probe with soda
water or salt water.
If this does not suffice, several earth spikes can be parallel connected to the auxiliary earth electrode.
In mountainous terrain or in the case of very rocky subsoil where
earth spikes cannot be driven into the ground, wire grates with a
mesh size of 1 cm and a surface area of about 2 square meters
can be used. These grates are laid flat onto the ground, are wetted with soda water or salt water and may also be weighted down
with sacks full of moist earth.
Dissipation Resistance of Large Scope Earthing Systems
Significantly large distances to the probe and the auxiliary earth
electrode are required for measuring large scope earthing systems. Calculations are based on 2½ or 5 times the value of the
earthing system’s largest diagonal.
Large scope earthing systems of this sort often demonstrate dissipation resistances of only a few ohms, which makes it especially
important to position the measuring probe within the neutral zone.
The probe and the auxiliary earth electrode should be positioned
at a right angle to the direction of the earthing system’s largest linear expansion. Dissipation resistance must be kept small. If necessary, several earth spikes must be used at a distance of 1 to
2 m from each other and connected to this end.
However, in actual practice large measuring distances are frequently not possible to due difficult terrain.
If this is the case, proceed as shown in figure “Earthing Resistance Measurement for a Large Scope Earthing System” on
page 68.
➭ Auxiliary earth electrode H is positioned as far as possible
from the earthing system.
➭ The area between the earth electrode and the auxiliary earth
electrode is sampled with the probe in equal steps of 5 meters
each.
with Current Clamp Sensor and PRO-RE Measuring Adapter as Accessory (PROFITEST MPRO & PROFITEST MXTRA only)
General
Set Parameters at Tester
❏ Measuring range: 200
After switching to selective measurement, the AUTO mea-
suring range is activated automatically if a measuring
range of greater than 200 had been selected.
❏ Connection type: selective
❏ Current clamp sensor transformer ratio:
1:1 (1 V/A,) 1:10 (100 mV/A), 1:100 (10 mV/A)
❏ Distance d (for measuring
): irrelevant in this case
E
Set Parameters at Current Clamp Sensor
❏ Current clamp sensor measuring range: see table below
Select Measuring Range at the Current Clamp Sensor
When measuring earthing resistance in systems with several parallel connected earth electrodes, total resistance of the earthing
system is measured.
Two earth spikes (auxiliary earth electrode and probe) are set for
this measurement. Measuring current is fed between the earth
electrode and the auxiliary earth electrode and voltage drop is
measured between the earth electrode and the probe.
The current clamp is positioned around the earth electrode to be
measured, and thus only that portion of the measuring current
which flows through the earth electrode is measured.
Test InstrumentZ3512A Clamp
Parameters
Transformation Ratio
1:1
1 V / A
01:10
100 mV / A
1:100
10 mV / A
SwitchesMeasuring Range
1 A / × 11A
10 A / × 1010 A
100 A / × 100100 A
Important Instructions for Use of the Current Clamp Sensor
Connection
➭ Position the spikes
for the probe and the auxiliary electrode at least 20, respectively 40 meters from the electrode (see figure above).
➭ Make sure that no excessively high contact resistances occur
between the probe and the ground.
➭ Attach the PRO-RE adapter (Z501S) to the test plug.
➭ Connect the probes, the auxiliary electrode and the electrode
via the 4 mm banana plug sockets at the PRO-RE adapter.
In doing so, observe labeling on the banana plug sockets.
➭ Connect the Z3512A current clamp sensor to jacks (15) and
(16) at the test instrument.
➭ Attach the current clamp sensor to the earth electrode.
• Use only the Z3512A current clamp sensor for this measurement.
• Use the clamp in the permanently connected state. The sensor may not be moved during measurement.
• The current clamp sensor may only be used at an adequate
distance from powerful extraneous fields.
• Make sure that the current clamp sensor’s connector cable is
laid separate from the probe cables to the greatest possible
extent.
Start Measurement
Select Measuring Function
Select Operating Mode
The selected operating mode is displayed inversely:
white battery icon against black background.
(with current clamp sensor and transformer, and pro-re measuring adapter as accessory) (PROFITEST MPRO & PROFITEST
MXTRA only)
2-Clamp Measuring Method
Select Measuring Function
Select Operating Mode
The selected operating mode is displayed inversely:
white battery icon against black background.
Set Parameters at Tester
❏ Measuring range: in this case always AUTO
After selecting 2-clamp measurement, switching to the
AUTO range takes place automatically. It is then no longer
In the case of earthing systems which consist of several earth electrodes (R1 …
Rx) which are connected to
each other, earthing resistance of a single electrode
(Rx) can be ascertained with
the help of 2 current clamps
without disconnecting Rx or
using spikes.
This measuring method is
especially well suited for
buildings or systems for
which probes and auxiliary
earth electrodes cannot be used, or where it’s impermissible to
disconnect earth electrodes.
Furthermore, this “spike-free” measurement is performed as one
of three measurements for lightning protection systems, in order
to determine whether or not current can be dissipated.
Figure at right:
PRO-RE/2 measuring adapter as
accessory for connecting the EClip 2 generator current clamp
Connection
❏ Connection: 2 clamps
❏ Current clamp sensor transformer ratio:
❏ Distance d (for measuring
Set Parameters at Current Clamp Sensor
❏ Current clamp sensor measuring range: see table below
Select Measuring Range at the Current Clamp Sensor
Important Instructions for Use of the Current Clamp Sensor
• Use only the Z3512A current clamp sensor for this measure-
• Use the clamp in the permanently connected state. The sen-
• The current clamp sensor may only be used at an adequate
• Make sure that the connector cables from the two clamps are
possible to change the range!
1:1 (1 V/A), 1:10 (100 mV/A), 1:100 (10 mV/A)
): irrelevant in this case
E
Test InstrumentZ3512A Clamp
Parameters
Transformation Ratio
1:1
1 V / A
01:10
100 mV / A
1:100
10 mV / A
ment.
sor may not be moved during measurement.
distance from powerful extraneous fields.
laid separate from each other to the greatest possible extent.
SwitchesMeasuring Range
1 A / × 11A
10 A / × 1010 A
100 A / × 100100 A
Start Measurement
➭ No probes or auxil-
➭ The earth electrode is not disconnected.
➭ Attach the PRO-RE/2 adapter (Z502T) to the test plug.
➭ Connect the E-Clip 2 generator clamp (current clamp trans-
➭ Connect the Z3512A current clamp sensor to jacks 15 and 16
➭ Attach the 2 clamps to an earth electrode (earth spike) at dif-
70Gossen Metrawatt GmbH
iary earth electrodes are required.
former) via the 4 mm safety plugs at the PRO-RE/2 adapter.
at the test instrument.
ferent heights with a clearance of at least 30 cm.
Soil Resistivity E Relative to Season Without the Effects
of Precipitation (earth electrode depth < 1.5 m)
(PROFITEST MPRO & PROFITEST MXTRA only)
General
Select Measuring Function
Select Operating Mode
The selected operating mode is displayed inversely:
white battery icon against black background.
Set Parameters
The determination of soil resistivity is necessary for the planning of
earthing systems. Reliable values need to be ascertained which
take even the worst possible conditions into account (see „Geologic Evaluation“ on page 71).
Soil resistivity is decisive with regard to the magnitude of an earth
electrode’s dissipation resistance. Soil resistivity can be measured
with the test instrument using the method according to Wenner.
Four earth spikes of greatest possible length are driven into the
ground in a straight line at distance d from one another, and are
connected to the earth tester (see figure above).
The earth spikes usually have a length of 30 to 50 cm. Longer
earth spikes can be used for soil which demonstrates poor conductivity (sandy soil etc.). The depth to which the earth spikes are
driven into the ground may not exceed one twentieth of distance
d.
❏ Transformer ratio: irrelevant in this case
❏ Distance d for measurement of
: adjustable from 0.1 to 999
E
Start Measurement
Erroneous measurement may result if piping, cables or
other underground metal conduits run parallel to the
measuring setup.
Soil resistivity is calculated as follows:
E=2 d R
Where:
= 3.1416
d = distance in m between two earth spikes
R = ascertained resistance value in (this value corresponds to R
determined with the 4-wire method)
E
as
Connection
➭ Position the spikes
for the probe and the
auxiliary electrode at equal distances (see figure above).
➭ Make sure that no excessively high contact resistances occur
between the probe and the ground.
➭ Attach the PRO-RE adapter (Z501S) to the test plug.
➭ Connect the probes, the auxiliary electrode and the electrode
via the 4 mm banana plug sockets at the PRO-RE adapter.
In doing so, observe labeling on the banana plug sockets.
Geologic Evaluation
Except in extreme cases, the ground is measured down to a
depth which is roughly equal to probe distance d.
And thus it’s possible to arrive at conclusions regarding the
ground’s stratification by varying probe distance. Layers which
are highly conductive (water table) into which earth electrodes
should be installed, can thus be discovered within a region which
is otherwise not very conductive.
Soil resistivity is subject to considerable fluctuation which may be
due to various causes such as porosity, moisture penetration,
concentration of dissolved salts in the ground water and climatic
fluctuation.
Characteristic values for
and the soil’s negative temperature coefficient) can be approximated quite closely by means of a sinusoidal curve.
relative to season (soil temperature
E
Gossen Metrawatt GmbH71
Page 72
A number of typical soil resistivity values for various types of
Formulas for calculating dissipation resistance for common types
of earth electrodes are included in this table.
These rules of thumb are entirely adequate for actual practice.
NumberEarth ElectrodeRule of ThumbSubsidiary Variable
1
2
3Ring earth electrode
4Mesh earth electrode
5Ground plate—
6
Earth strip
(star type earth electrode)
Earth rod
(buried earth electrode)
Hemispherical earth
electrode
—
—
RA= dissipation resistance ()
= soil resistivity (m)
E
I = length of the earth electrode (m)
D = diameter of a ring earth electrode, diameter of the equivalent surface
area of a mesh earth electrode or diameter of a hemispherical earth
electrode (m)
F = surface area (sq. meters) of the enclosed surface or a ring or mesh
earth electrode
a = Edge length (m) of a square ground plate; a is replaced with the fol-
lowing for rectangular plates:
bxc, where b and c are the two
sides of the rectangle.
J = volume (cubic meters) of an individual foundation footing
Insulation resistance may only be measured at voltagefree devices.
16.1General
Select Measuring Function
Connection
2 pole adapter or test
plug
Breakdown Current for Ramp Function
In order to suppress the influence of parallel capacitances
on the device under test when measurement is started,
shutdown at the respective breakdown current I
not occur until a minimum voltage of 5 V is exceeded.
Limit Values for Breakdown Voltage
lim
does
Set Parameters
* Freely adjustable voltage (see section 10.8)
Polarity Selection
* AUTO parameter (see section 10.9)
Gossen Metrawatt GmbH73
The test instrument always measures insulation between
the L and PE terminals.
N and PE must be interrupted for systems without RCD.
Checking Measurement Cables Before Measurements
Before performing insulation measurement, the test
probes on the measurement cables should be short-circuited in order to assure that the instrument displays a
value of less than 1 k. In this way, incorrect connection
can be avoided and interrupted measurement cables can
be detected.
Limit Values for Constant Test Voltage
❏ Test Voltage
A test voltage which deviates from nominal voltage, and is usually
lower, can be selected for measurements at sensitive components, as well as systems with voltage limiting devices.
❏ Voltage Type
The “U
detect weak points in the insulation, as well as to determine
response voltage for voltage limiting components. After pressing
the ON/START ▼ key, test voltage is continuously increased until
the specified nominal voltage U
which is measured at the test probes during and after testing.
After measurement, this voltage drops to a value of less than 10 V
(see section entitled “Discharging the Device Under Test”).
Insulation measurement with rising test voltage is ended:
• As soon as specified maximum test voltage U
the measured value is stable
or
• As soon as specified maximum test voltage is reached
(e.g. after sparkover occurs at breakdown voltage).
Specified maximum test voltage U
down voltage which occurs is displayed for U
The constant test voltage function offers two options:
” rising test voltage function (ramp function) is used to
INS
is reached. U is the voltage
N
is reached and
N
or any triggering or break-
N
INS
.
Page 74
• After briefly pressing the ON/START ▼ key, specified test volt-
Note
Note
age U
is read out and insulation resistance R
N
As soon as the measured value is stable (settling time may be
is measured.
INS
several seconds in the case of high cable capacitance values),
measurement is ended and the last measured values for R
and U
the test probes during and after testing. After measurement,
are displayed. U is the voltage which is measured at
INS
INS
this voltage drops to a value of less than 10 V (see section
entitled “Discharging the Device Under Test”).
or
• As long as you press and hold the ON/START ▼ key, test voltage
U
is applied and insulation resistance R
N
not release the key until the measured value has settled in
is measured. Do
INS
(settling time may be several seconds in the case of high cable
capacitance values). Voltage U, which is measured during
testing, corresponds to voltage U
START ▼ key, measurement is ended and the last measured
values for R
less than 10 V after measurement (see the section entitled
INS
and U
are displayed. U drops to a value of
INS
. After releasing the ON/
INS
“Discharging the Device Under Test”.
❏ Pole Selection Report Entry
The poles between which testing takes place can only be entered
here for reporting purposes. The entry itself has no influence on
the actual polarity of the test probes or the pole selection.
❏ Limits – Setting the Limit Value
The limit value for insulation resistance can be set as desired. If
measurement values occur which are below this limit value, the
red U
0.5 M to 10 M is available. The limit value is displayed above
LED lights up. A selection of limit values ranging from
L/RL
the measured value.
Start Measurement – Rising Test Voltage (ramp function)
Press briefly:
• The selected voltage limit is reached
• The selected current limit is reached
or
• Sparkover occurs (spark gaps)
Differentiation is made amongst the following three procedures for
insulation measurement with ramp function:
Testing overvoltage limiters or varistors and determining their
tripping voltage:
– Select maximum voltage such that the anticipated breakdown
voltage of the device under test is roughly one third of this
value (observe manufacturer’s data sheet if applicable).
– Select the current limit value in accordance with actual
requirements or the manufacturer’s data sheet (characteristic
curve of the device under test).
Determining tripping voltage for spark gaps:
– Select maximum voltage such that the anticipated breakdown
voltage of the device under test is roughly one third of this
value (observe manufacturer’s data sheet if applicable).
– Select the current limit value in accordance with actual
requirements within a range of 5 to 10 A (response characteristics are too unstable with larger current limit values, which
may result in faulty measurement results).
Detecting weak points in the insulation:
– Select maximum voltage such that it does not exceed the test
object’s permissible insulation voltage; it can be assumed that
an insulation fault will occur even with a significantly lower
voltage if an accordingly lower maximum voltage value is
selected (nevertheless at least greater than anticipated breakdown voltage) – the ramp is less steep as a result (increased
measuring accuracy).
– Select the current limit value in accordance with actual
requirements within a range of 5 to 10 A (see also settings
for spark gaps).
Start Measurement – Constant Test Voltage
Quick polarity reversal if
parameter is set to AUTO:
01/10 … 10/10: L1-PE … L1-L3
If semiautomatic polarity reversal is selected (see section
10.9), the corresponding icon is displayed instead of the
ramp.
General Notes Regarding Insulation Measurements with Ramp
Function
Insulation measurement with ramp function serves the following
purposes:
• Detect weak points in the test object’s insulation
• Determine tripping voltage of voltage limiting components and
test them for correct functioning These components may
include, for example, varistors, overvoltage limiters (e.g.
DEHNguard® from Dehn+Söhne) and spark gaps.
The test instrument uses continuously rising test voltage for this
measuring function, up to the maximum selected voltage limit.
The measuring procedure is started by pressing the ON/START ▼
key and runs automatically until one of the following events
occurs:
Press and hold for longterm measurement:
Quick polarity reversal if parameter is set to AUTO: 01/10 … 10/10: L1-PE
… L1-L3
The instrument’s batteries are rapidly depleted during the
insulation resistance measurement. When using the con-stant test voltage function, only press and hold the start
key ▼ until the display has become stable (if long-term
measurement is required).
74Gossen Metrawatt GmbH
Page 75
Special Condition for Insulation Resistance Measurement
Attention!
!
Attention!
!
Attention!
!
Attention!
!
R
ISO
Limit Value:
RE(INS) > Limit Value
UL R
L
R
EINS
Voltage type: constant
Test vol t ag e :
50, 100, 250, 325, 500, 1000 V*
Voltage type: rising/ramp
Earth leakage
* Freely adjustable voltage (see section 10.8)
resistance:
Insulation resistance can only be measured at voltagefree objects.
If measured insulation resistance is less than the selected limit
value, the UL/RL LED lights up.
If an interference voltage of
insulation resistance is not measured. The MAINS/NETZ LED lights
up and the interference voltage pop-up message appears.
All conductors (L1, L2, L3 and N) must be tested against PE!
Do not touch the instrument’s terminal contacts during
insulation resistance measurements!
If nothing has been connected to the terminal contacts, or if a
resistive load component has been connected for measurement,
your body would be exposed to a current of approximately 1 mA
at a voltage of 1000 V. The resultant perceptible shock may lead
to injury (e.g. resulting from a startled reaction etc.).
25 V is present within the system,
Discharging the Device Under Test
16.2Special Case: Earth Leakage Resistance (R
This measurement is performed in order to determine electrostatic discharge capacity for floor coverings in accordance with
EN 1081.
EISO
)
Select Measuring Function
Set Parameters
If measurement is performed at a capacitive object such
as a long cable, it becomes charged with up to approx.
1000 V! Touching such objects is life endangering!
When an insulation resistance measurement has been performed
on a capacitive object it’s automatically discharged by the instrument after measurement has been completed. Contact with the
device under test must be maintained to this end. The falling voltage value can be observed at the U display.
Do not disconnect the DUT until less than 10 V is displayed for
U!
Evaluating Measured Values
Instrument measuring error must be taken into consideration in
order to assure that the limit values set forth in DIN VDE regulations are not fallen short of. The required minimum display values
for insulation resistance can be determined with the help of Table
3 on page 100. These values take maximum device error into
consideration (under nominal conditions of use). Intermediate values can be interpolated.
Connection and Test Setup
Gossen Metrawatt GmbH75
➭ Rub the floor covering at the point at which measurement is to
be performed with a dry cloth.
➭ Place the 1081 floor probe onto the point of measurement
and load it with a weight of at least 300 N (30 kg).
➭ Establish a conductive connection between the measuring
electrode and the test probe and connect the measuring
adapter (2-pole) to an earth contact, e.g. the earthing contact
at a mains outlet or a central heating radiator (prerequisite: reliable ground connection).
Page 76
Start Measurement
The limit value for earth
leakage resistance from the relevant regulations applies.
76Gossen Metrawatt GmbH
Page 77
17
Attention!
!
Note
Note
R
LO
ROFFSET: ON OFF
Polarity: +/– to PE
Polarity: +/– to PE
with ramp sequence
Limit Value:
RLO > Limit Value
UL R
L
Measuring Low-Value Resistance of up to 200 (Protective Conductor and Equipotential Bonding Conductor)
According to the regulations, the measurement of low-value resistance at protective conductors, earth conductors or bonding conductors must be performed with (automatic) polarity reversal of
the test voltage, or with current flow in one (+ pole to PE) and then
the other direction (– pole to PE).
Low-resistance may only be measured at voltage-free
objects.
Select Measuring Function
Connection
via 2-pole adapter only!
❏ ROFFSETON/OFF
– Compensation for Measurement Cables up to 10
If measurement cables or extension cables are used, their resistance can be automatically subtracted from the measurement
results. Proceed as follows:
➭ Switch R
footer.
➭
Select a polarity option or automatic polarity reversal.
➭ Short-circuit the end of the measurement extension cable with
the second test probe at the instrument.
➭ Start measurement of offset resistance with IN.
First of all, an intermittent acoustic warning is
generated and a blinking message appears, in
order to prevent inadvertent deletion of a previously saved offset value.
➭ Start offset measurement by pressing
the triggering key once again or abort
offset measurement by pressing the ON/
START ▼
OFFSET from OFF to ON. Roffset = 0.00 appears in the
key
(in this case =
If offset measurement is stopped upon appearance of a
pop-up error window indicating Roffset > 10 or a differ-ence between RLO+ and RLO– of greater than 10%, the last
measured offset value is retained. Inadvertent deletion of
a previously ascertained offset value is thus practically
ruled out. The respectively smaller value is otherwise
stored to memory as an offset value. The maximum offset
value is 10.0 . Negative resistances may result due to
the offset value.
ESC
).
Set Parameters
Measuring ROFFSET
The ROFFSET x.xx message now appears in the footer at the display, where x.xx can be
a value between 0.00 and 10.0. This value is subtracted from
the actual measuring results for all subsequent R
ments, if the R
R
OFFSET must be redetermined in the following cases:
• After switching to a different polarity option
• After switching from ON to OFF and back again
The offset value can be deliberately deleted by switching ROFFSET
from OFF to ON.
OFFSET ON/OFF key has been set to ON.
Only use this function when performing measurements
with extension cables.
When different extension cables are used, the above
described procedure must always be repeated.
measure-
LO
❏ Type / Polarit y
Gossen Metrawatt GmbH77
The direction in which current flows can be selected here.
❏ Limits – Setting the Limit Value
The limit value for resistance can be set as desired. If measured
values occur which are above this limit value, the red UL/RL LED
lights up. Limit values can be selected within a range of 0.10 to
10.0 (editable). The limit value is displayed above the measured
value.
Page 78
17.1Measurement with Constant Test Current
Attention!
!
Note
Start Measurement
Press and hold for longterm measurement
Measuring Low-Value Resistance
Measurement cable and 2-pole measuring adapter resistance is compensated automatically thanks to the 4-wire
method and thus doesn’t effect measurement results.
However, if an extension cord is used its resistance must
be measured and deducted from the measurement
results.
Resistances which do not demonstrate a stable value
until after a “settling in period” should not be measured
with automatic polarity reversal, but rather one after the
other with positive and negative polarity.
Examples of resistances whose values may change
during measurement include:
– Incandescent lamp resistance, whose values change
due to warming caused by test current
– Resistances with a large conductive component
– Contact resistance
The test probes should always be in contact with the device
under test before the Start key
If the DUT is energized, measurement is disable as soon as it’s
contacted with the test probes.
If the start key
the test probes afterwards, the fuse blows.
Which of the two fuses has blown is indicated in the pop-up
window with the error message by means of an arrow.
In the case of single-pole measurement, the respective value is
saved to the database as R
Polarity SelectionDisplayCondition
+ pole to PERLO+None
– pole to PERLO–None
Pole to PE
▼ is pressed first and the DUT is contacted with
LO
RLOWhere RLO 10%
RLO+
RLO–
▼ is activated.
.
Where RLO > 10%
Automatic Polarity Reversal
After the measuring sequence has been started, the instrument
performs the measurement with automatic polarity reversal, first
with current flow in one direction, and then in the other. In the
case of long-term measurement (press and hold ON/START ▼ key),
polarity is switched once per second.
If the difference between RLO+ and RLO– is greater than 10%
with automatic polarity reversal, RLO+ and RLO– values are displayed instead of RLO. The respectively larger value, RLO+ or
RLO–, appears at the top and is saved to the database as the
RLO value.
Evaluating Measured Values
See Table 4 on page 100.
Calculation of Cable Lengths for Common Copper Conductors
If the HELP key is activated after performance of resistance measurement, the cable lengths corresponding to common conductor cross sections are displayed.
If results vary for the two different current flow directions, cable
length is not displayed. In this case, capacitive or inductive components are apparently present which would distort the calculation.
This table only applies to cables made with commercially available
copper conductors and cannot be used for other materials (e.g.
aluminum).
Evaluating Measurement Results
Differing results for measurements in both directions indicate voltage at the DUT (e.g. thermovoltages or unit voltages).
Measurement results can be distorted by parallel connected
impedances in load current circuits and by equalizing current,
especially in systems which make use of overcurrent protection
devices (previous neutralization) without an isolated protective
conductor. Resistances which change during measurement (e.g.
inductance), or a defective contact, can also cause distorted
measurements (double display).
In order to assure unambiguous measurement results, causes of
error must be located and eliminated.
In order to find the cause of the measuring error, measure resistance in both current flow directions.
The instrument’s batteries are exposed to excessive stress during
insulation resistance measurement. For measurement with current flow in one direction, only press and hold the ON/START ▼ key
as long as necessary for the measurement.
78Gossen Metrawatt GmbH
Page 79
17.2Protective Conductor Resistance Measurement with Ramp Sequence
Note
Measuring PhaseDemagnetizing
and Waiting Time
Result
Time [s]
Rising
Phase
Test C ur re n t [ A]
0136
0.25
Before Polarity ReversalorRestart
Representation of Measuring and Waiting Times for
Protective Conductor Resistance Measurement at
PRCDs with the PROFITEST MXTRA
– Measurement at PRCDs with Current-Monitored Protective Conductor using the PROFITEST PRCD Test Adapter as an
Accessory ( PROFITEST MXTRA only)
Application
Protective conductor current is monitored for certain types of
PRCDs. Direct activation or deactivation of the test current
required for protective conductor resistance measurements of at
least 200 mA results in tripping of the PRCD and thus to interruption of the protective conductor connection. Protective conductor
measurement is no longer possible in this case.
A special ramp sequence for test current activation and deactivation in combination with the PROFITEST PRCD test adapter permits protective conductor resistance measurement without tripping the PRCD.
Ramp Function Time Sequence
Due to the physical characteristics of the PRCD, measuring times
for this ramp function amount to several seconds.
If test current polarity is revered, additional waiting time is also
required during polarity reversal.
This is programmed into the test sequence in the “automatic
polarity reversal” operating mode.
Reverse polarity manually, e.g. from
“+pole with ramp” to “-pole
with ramp” . The test instrument then detects the reversal of current
flow direction, stops measurement for the
required waiting time and simultaneously
displays a corresponding message (see
figure at right).
Connection
➭ Read the operating instructions for the PROFITEST PRCD
adapter, in particular section 4.1. It includes connection instructions for offset measurement and for protective conductor resistance measurement.
Selecting the Polarity Parameter
➭ Select the desired polarity parameter with ramp.
Measuring ROFFSET
➭ Perform offset measurement as described on page 77, in
order to assure that the test adapter’s connector contacts are
not included in the measurement results.
The offset is only retained in memory until the polarity
parameter is changed. If measurement is performed with
manual polarity reversal (+pole or -pole), the offset measurement has to be repeated in both polarities before
each measurement.
Measuring Protective Conductor Resistance
➭ Determine whether or not the PRCD is activated. If not, acti-
vate it.
➭ Perform protective conductor measurement as described in
section 17.1 above. Start the test sequence by briefly pressing the ON/START ▼ key. The predetermined duration of the
measuring phase can be extended by pressing and holding
the ON/START ▼ key.
Start Measurement
PRCD Tripping due to Poor Contacting
Good contact must be assured between the test probes at the 2pole adapter and the device under test or the sockets at the
PROFITEST PRCD test adapter during measurement. Interruptions can result in considerable test current fluctuation which
causes the PRCD to trip under unfavorable conditions.
If this is the case, tripping of the PRCD is
automatically detected by the test instrument and indicated by a corresponding
error message (see figure at the right). In
this case as well, the test instrument
automatically takes subsequently
required waiting time into account before
you can reactivate the PRCD and start the measurement over
again.
Gossen Metrawatt GmbH79
The symbol shown at the right appears during the magnetization phase (rising curve) and the subsequent measuring phase (constant current).
If measurement is aborted already during the rising phase, no
measurement results can be ascertained or displayed.
After measurement, the demagnetization phase (falling
curve) and subsequent waiting time are indicated by the
inverted symbol shown at the right.
No new measurements can be started during this time.
Measurement results cannot be read and measurement
with the same or another polarity cannot be started until
the symbol at the right appears.
Page 80
18Measurement with Accessory Sensors
Attention!
!
Attention!
!
Attention!
!
SENSOR
Clamp
output range
Limit Value:
I < and I > Limit Value
UL R
L
18.1Current Measurement with Current Clamp Sensor
Bias, leakage and circulating current up to 1 A, as well as leakage
current up to 1000 A can be measured with the help of special
current clamp sensors, which are connected to sockets 15 and
16.
Danger: High-Voltage!
Use only current clamp sensors which are specifically offered as accessories by Gossen Metrawatt GmbH.
Other current clamp sensors might not be terminated
with an output load at the secondary side. Dangerously
high voltage may endanger the user and the device in
such cases.
Maximum input voltage at the test instrument!
Do not measure any currents which are greater than
specified for the measuring range of the respective
clamp. Input voltage for clamp connector sockets 15
and 16 at the test instrument may not exceed 1 V!
Set Parameters
The transformation ratio parameter must be correspondingly set
at the test instrument depending upon the respectively selected
measuring range at the current clamp sensor.
Be sure to read and adhere to the operating instructions
for current clamp sensors and the safety precautions included therein, especially those regarding the approved
measuring category.
Select Measuring Function
Select Measuring Range at the Current Clamp Sensor
Tes t
Instrument
Transforma-
tion Ratio
Parameter
1:1
1 V / A
01:10
100 mV / A
1:100
10 mV / A
1:1000
1 mV / A
WZ12C
Switch
1 mV / mA
—x 100 [mV/A]—0 … 10 A
—x 10 [mV/A]—0 … 100 A
1 mV / Ax 1 [mV/A]
Clamp MetersTest
Z3512A
Switch
x 1000
[mV/A]
WZ12C
Measuring
Range
1 mA …
15 A
1 A …
150 A
Z3512A
Measuring
Range
0 … 1 A
0 … 1000 A
Instrument
Measuring
Range
5 …
999 mA
0.05 …
10 A
0.5 …
100 A
5 …
150A/999A
Specifying limit values results in automatic evaluation at the end of
the measurement.
Connection
Start Measurement
Test InstrumentClampTest
Transformation Ratio
Parameter
1:1
1 V / A
01:10
100 mV / A
1:100
10 mV / A
80Gossen Metrawatt GmbH
METRAFLEX P300
Switch
3 A (1 V/A)3A5 … 999 mA
30 A (100 mV/A)30 A0.05 … 10 A
300 A (10 mV/A)300 A0.5 … 100 A
METRAFLEX P300
Measuring Range
Instrument
Measuring
Range
Press the OK key to stop
the measurement.
Page 81
19Special Functions – EXTRA Switch Position
EXTRA
Select the EXTRA Switch Position
Overview of Special Functions
SoftkeyMeaning / Special
Function
Voltage drop
measurement
U function
Standing surface
insulation
impedance
Z
function
ST
Meter startup
test
kWh function
Leakage current
measurement
function
I
L
Insulation monitor
test
IMD function
Residual voltage
test
Ures function
Intelligent ramp
ta + Ifunction
Section /
Page
PROFITEST MBASE+
PROFITEST MTECH+
PROFITEST MPRO
PROFITEST MXTRA
section
✓
✓
✓
✓
19.1 on
page 82
section
19.2 on
page 83
section
19.3 on
page 84
section
19.4 on
page 85
section
19.5 on
page 86
section
19.6 on
page 88
section
19.7 on
page 89
✓✓✓✓
✓✓✓✓
✓✓✓✓
———
———
———
———
Selecting Special Functions
The list of special functions is accessed by pressing the uppermost softkey. Select the desired function with the appropriate
icon.
✓
section
19.8 on
page 90
RCM residual
current monitor
———
RCM function
✓
section
19.9 on
page 91
Testing of electric
vehicle operating
statuses at charging
—
—
✓
stations per
IEC 61851-1
✓
section
19.10 on
page 92
Documentation of
fault simulations at
PRCDs with the PRO-
———
FITEST PRCD adapter
Gossen Metrawatt GmbH81
Page 82
19.1Voltage Drop Measurement (at ZLN) – U Function
Note
1
2
Nominal current: 2 … 160 A
Polarity selection: Lx-N
Diameters: 1.5 … 70 mm
Cable type: NY…, H03… - H07…
Number of wires: 2 … 10-wire
Tripping characteristics: B, L
Limit Value:
U % > Limit Value
UL R
L
U
Red
1
2
Significance and Display of U (per IEC 60364-6)
Voltage drop from the intersection of the distribution network and
the consumer system to the point of connection of an electrical
power consumer
should not exceed 4% of nominal line voltage.
Calculating voltage drop (without offset):
U = Z
L-N
Calculating voltage drop (with offset):
U = (Z
L-N
U in % = 100 × U / U
See also section 14 regarding the measurement procedure and
connection.
(electrical outlet or device connector terminals)
× nominal current of the fuse
– Z
) × nominal current of the fuse
OFFSET
L-N
Connection and Test Setup
Set Parameters
Measurement Without OFFSET
Proceed as follows:
➭ Switch OFFSET from ON to OFF.
Determine OFFSET (in %).
Proceed as follows:
➭ Switch OFFSET from OFF to ON. U
➭ Connect the test probe to the point of common coupling
(measuring device / meter).
➭ Start measurement of offset with I
First of all, an intermittent acoustic warning is
generated and a blinking message appears, in
order to prevent inadvertent deletion of a previously saved offset value.
➭ Start offset measurement by pressing
the triggering key once again, or abort
offset measurement by pressing the
ON/START ▼ key (in this case ESC).
OFFSET = 0.00% is displayed.
.
N
If nominal current IN is changed by U
value is automatically adjusted.
OFFSET
, the offset
Set Limit Values
U
OFFSET x.xx % is displayed and x.xx can be a value within a
range of 0.00 to 99.9%.
An error message appears in a pop-up window
if Z exceeds 9.99
Start measurement with OFFSET.
TABLimit value per German technical connection conditions
for connection to low-voltage mains between the distribution network and the measuring device
DINLimit value per DIN 18015-1: U < 3%
VDELimit value per DIN VDE 0100-520: U < 4%
NLLimit value per NIV: U < 5%
82Gossen Metrawatt GmbH
between the measuring device and the consuming device
between the distribution network and the consuming
device (adjustable up to 10% in this case)
Page 83
19.2Measuring the Impedance of Insulating Floors and Walls
Note
Attention!
!
OK
NOT OK
(standing surface insulation impedance) – Z
Function
ST
Measuring Method
The instrument measures the impedance between a weighted
metal plate and earth. Line voltage available at the measuring site
is used as an alternating voltage source. The ZST equivalent circuit is considered a parallel circuit.
Connection and Test Setup
Start Measurement
Evaluate Measured Value
The measured value has to be evaluated after measurement has
been completed:
Use the measuring setup described in section 16.2 (triangular probe) or the one outlined below:
➭ Cover the floor or the wall at unfavorable locations, e.g. at
joints or abutments, with a damp cloth measuring approx.
270 × 270 mm.
➭ Place the 1081 Probe on top of the damp cloth and load the
probe with a weight of 750 N (75 kg, i.e. one person) for
floors, or 250 N (25 kg) for walls, e.g. press against the wall
with one hand which is insulated with a glove).
➭ Establish a conductive connection to the 1081 Probe, and
connect it to the probe connector socket at the instrument.
➭ Connect the instrument to a mains outlet with the test plug.
Do not touch the metal plate or the damp cloth with your
bare hands.
No more than 50% line voltage may be applied to these
parts! Current with a value of up to 3.5 mA may flow!
The measured value would be distorted as well.
Resistance values must be measured at several points in order to
provide for adequate evaluation. Measured resistance may not be
less than 50 k at any given point. If the measured value is
greater than 30 M, Z
play panel.
In the event that “NOT OK” is selected, an error is indicated by the
UL/RL LED which lights up red.
See also Table 5 on page 101 with regard to evaluating measured
values.
The measured value cannot be saved to memory and included in
the test report until it has been evaluated.
> 30.0M always appears at the dis-
ST
Save measured value
Gossen Metrawatt GmbH83
Page 84
19.3Testing Meter Startup with Earthing Contact Plug
Note
Note
OK
NOT OK
– kWh Function
Energy consumption meters can be tested for correct startup with
this function.
Connection L – N
Earthing Contact Plug
Start Measurement
The measured value cannot be saved to memory and included in
the test report until it has been evaluated.
Save measured value
Special Case
Startup of energy consumption meters which are connected
between L and L or L and N can be tested with this function.
Connection L – L
2-Pole Adapter
The meter is tested with
the help of an internal load resistor and a test current of approximately 250 mA. After pressing the Start key, test power is displayed and the meter can be tested for proper startup within a
period of 5 seconds. The “RUN” pictograph is displayed.
TN systems: All 3 phase conductors must be tested against N,
one after the other.
In other types of systems, all phase conductors (active conductors) must be tested against one another.
If minimum power is not reached, the test is either not
started or aborted.
Evaluate Measured Value
The measured value has to be evaluated after measurement has
been completed:
If an earthing contact outlet is not available, you can use
the 2-pole adapter. N must be contacted with the PE test
probe (L2), and then measurement must be started.
If PE is contacted with the PE test probe (L2) during the
meter startup test, approximately 250 mA flow through
the protective conductor and any upstream RCD is
tripped.
In the event that “NOT OK” is selected, an error is indicated by the
UL/RL LED which lights up red.
84Gossen Metrawatt GmbH
Page 85
19.4Leakage Current Measurement with PRO-AB Leakage
Attention!
!
Note
Current Adapter as Accessory – I
(PROFITEST MXTRA only)
Function
L
Application
Measurement of touch voltage in accordance with
DIN VDE 0107, part 10, as well as continuous leakage and
patient auxiliary current per IEC 62353 (VDE 0750-1) / IEC 601-1
/ EN 60601-1, is possible using the PRO-AB leakage current
measuring adapter as an accessory with the test instrument.
As specified in the standards listed above, current values of up to
10 mA can be measured with this measuring adapter. In order to
be able to fully cover this measuring range using the measurement input provided on the test instrument (2-pole current clamp
input), the measuring instrument is equipped with range switching
including transformation ratios of 10:1 and 1:1. In the 10:1 range,
voltage dividing takes place at the same ratio.
Connection and Test Setup
In order to perform the leakage current measurement, the
adapter’s measurement outputs must be plugged into the measurement inputs at the left-hand side of the test instrument (2pole current clamp input and probe input).
Either of the leakage current measuring adapter’s inputs is connected to reference earth (e.g. safe earth electrode / equipotential
bonding) via a measurement cable. The metallic housing (accessible part) of the device under test is contacted with a test probe or
alligator clip which is connected to the other input by means of a
second measurement cable.
Measuring Procedure
Refer to the operating instructions for the PRO-AB leakage current measuring adapter regarding performance of the measurement.
The test plug should be located in the storage slot during
leakage current measurement. Under no circumstances
may the test plug be connected with any system components, including PE / ground potential (measured values
might otherwise be distorted).
Testing the PRO-AB Adapter
The adapter should be tested before use and at regular intervals
(see adapter operating instructions).
The measurement can be started or stopped by pressing the ON/START ▼ key. Leakage current measurement is a long-term mea-
surement, i.e. is continues until it’s stopped by the user. The
momentary measured value is displayed continuously during
measurement.
The self-test must be deactivated in the menu (set
“TEST ON/OFF” function key to “OFF”) in order to perform a
measurement.
Always start with the large measuring range (10:1), unless there’s
no doubt that small measured values can be expected, in which
case the small measuring range can be used (1:1). The measuring
range must be selected at the measuring adapter, as well as in
the menu using the corresponding function key (RANGE). It must
be assured that the range settings at the adapter and at the test
instrument are always identical, in order to prevent any distortion
of measurement results.
Depending on the magnitude of the measured values, the range
setting can or must (in the case of overranging) be manually corrected at the measuring adapter and the test instrument.
Individual limit values can be adjusted after pressing the Limits
function key. Exceeded limit values are indicated by the red limit
value LED at the test instrument.
Gossen Metrawatt GmbH85
Page 86
19.5
1
3
2
Conductor
Relationship:
Testing Insulation Monitoring Devices – IMD Function
(PROFITEST MXTRA only)
Application
Insulation monitoring devices (IMDs) or earth fault detection systems (EDSs) are used in IT systems in order to monitor adherence
to a minimum insulation resistance value as specified by
DIN VDE 0100-410.
They’re used in power supplies for which a single-pole earth fault
may not result in failure of the power supply, for example in operating rooms or photovoltaic systems.
Insulation monitors can be tested with the help of this special
function. After pressing the ON/START ▼ key, an adjustable insula-
tion resistance is activated between one of the two phases of the
IT system to be monitored and ground to this end. This resistance
can be changed in the MAN± manual sequence mode with the
help of the + or – softkey, or varied automatically from R
in the AUTO operating mode. Testing is ended by once again
R
min
pressing the ON/START ▼ key.
Time during which the momentary resistance value prevails since
changing the value at the system is displayed. The IMD’s display
and response characteristics can be subsequently evaluated and
documented with the help of the OK or NOT OK softkey.
max
to
Connection L – N
Resistance RSTART (3)
Numerous parameters are available for setting resistance RSTART,
with which measurement is begun.
Conductor Relationship / Resistance Range (2)
– Conductor relationship: The corresponding conductor relation-
ship can be selected for documentation of the measuring
point.
– Resistance range: A range of values can be selected for testing
the display of resistance at the IMD.
The parameter is set as a percentage with reference to the resistance momentarily introduced by the test instrument.
Upper and lower limit values are displayed in the measuring view.
When selecting test resistance, don’t forget that an excessively
high test current could damage sensitive system components.
Set Parameters
Measuring Procedure (1)
There are two ways to conduct the test:
– MAN: Resistance is changed manually by tapping the respec-
tive softkeys.
– AUTO: Resistance is changed automatically every 2 seconds
beginning with R
START.
Measuring Procedure
➭ Set the parameters.
➭ Start: Press the ON/START ▼ key.
➭ A resistance is introduced between the phase and protective
conductors and time measurement is started.
➭ Manual testMAN + –: press the or key to increase
or reduce test resistance R
➭ Automatic test AUTO: the resistance value is changed automati-
cally.
➭ Time to trip ta is restarted each time resistance is changed.
➭ Press I
➭ In order to end measurement, press the ON/START ▼ key as
soon as the IMD indicates that insulation resistance has been
fallen short of.
➭ Display of measured values
➭ Evaluation query: Measurement OK?
➭ If evaluation is NOT OK: the UL/RL LED lights up red.
➭ Save: by pressing the soft key.
in order to change the conductor relationship.
N
L-PE.
86Gossen Metrawatt GmbH
Page 87
Measurement can be
OK
NOT OK
aborted by pressing the ON/START ▼ or ESC key.
The following measured values are displayed:
–
RL-PE: Active test resistance with upper and lower limit values
– ta: Response time (during which momentary resistance is
applied until the measurement is ended)
- R
– R
min
with reference to the number of possible resistances
– U
L1PE
conductor L1 and protective conductor PE
– U
L2PE
conductor L2 and protective conductor PE
– U
L1L2
conductors L1 and L2
– I
LPE
: Status display indicating momentary resistance
max
: Momentary voltage at the test probes between phase
: Momentary voltage at the test probes between phase
: Momentary voltage at the test probes between phase
: Test current flowing through the active resistance
– f: Frequency of the applied voltage
Evaluation
In order to evaluate the measurement, it must be stopped. This
applies to manual as well as automatic measurement. Press the
ON/START ▼ or ESC key to this end. The stopwatch is stopped and
the evaluation window appears.
Retrieving Saved Measured Values
The measured value cannot be saved to memory and included in
the test report until it has been evaluated (see also section 9.4).
With the help of the key
shown at the right
(MW: measured value / PA: parameter), the setting
parameters can be displayed for this measurement.
Gossen Metrawatt GmbH87
Page 88
19.6Residual Voltage Test – U
Note
Limit Value:
U % > Limit Value
UL R
L
U
only)
Function (PROFITEST MXTRA
res
Application
The EN 60204 standard specifies that after switching supply
power off, residual voltage must drop to a value of 60 V or less
within 5 seconds at all accessible, active components of a
machine to which a voltage of greater than 60 V is applied during
operation.
Testing for the absence of voltage is performed as follows with the
test instrument by means of a voltage measurement which
involves the measurement of discharge time:
In the case of voltage dips of greater than 5% of momentary line
voltage (within 0.7 seconds), the stopwatch is started and
momentary undervoltage is displayed as U
indicated by the red UL/RL LED.
The function is ended after 30 seconds after which Ures and tu
data can be deleted and the function can thus be restarted by
pressing the ESC key.
after 5 seconds and
res
Connection
Measuring Sequence – Long-Term Measurement
Testing is selected as a
continuous measurement because residual
voltage testing is triggered automatically and
voltage measurement is
always active for safety
reasons.
If, for example, conductors are exposed when a machine
is switched off – e.g. if plug connectors are disengaged –
which are not protected against direct contact, maximum
permissible discharge time is 1 second!
Limit Values
Set Limit Values
88Gossen Metrawatt GmbH
Page 89
19.7
Nom. res. current: 10 … 500 mA
Type 1: RCD, SRCD, PRCD …
Nominal current: 6 … 125 A
Type 2: AC , A/F , B
* Type B = AC/DC sensitive
Touch voltage:
< 25 V, < 50 V, < 65 V
Intelligent Ramp – ta+I
(
PROFITEST MXTRA
only)
Function
Application
The advantage of this measuring function in contrast to individual
measurement of I
breaking time and breaking current by means of a test current
which is increased in steps, during which the RCD is tripped only
once.
The intelligent ramp is
subdivided into time
segments of 300 ms
each between the initial
current value (35% I
and the final current
value (130% I
results in a gradation for
which each step corresponds to a constant
test current which is
applied for no longer
than 300 ms, assuming
that tripping does not
occur.
And thus both tripping
current and tripping time are measured and displayed. Measured
quantities are acquired with reduced accuracy.
and tA is the simultaneous measurement of
N
)
N
). This
N
Start Touch Voltage Measurement
Start Tripping Test
Connection
Set Parameters
The measurement sequence can be broke off prematurely at any
time by pressing the ON/START ▼ key.
Measurement Results
Gossen Metrawatt GmbH89
Page 90
19.8Testing Residual Current Monitors
Nom. res. current: 10 … 500 mA
Waveform:
Nominal current: 6 … 125 A
Typ e: A , B *
* Type B = AC/DC sensitive
X times tripping current:
Connection: with/without probe
System type: TN/TT, IT
Touch voltage:
< 25 V, < 50 V, < 65 V
– RCM Function ( PROFITEST MXTRA only)
General
Residual current monitors (RCMs) monitor residual current in electrical systems and display it continuously. As is also the case with
residual current devices, external switching devices can be controlled in order to shut down supply power in the event that a
specified residual current value is exceeded.
However, the advantage of an RCM is that
the user is informed of
fault current within the
system before shutdown
takes place.
As opposed to individual
measurement of I
, measurement results
t
A
must be evaluated manually in this case.
If an RCM is used in
combination with an
external switching
device, the combination
must be tested as if it
were an RCD.
N
and
Connection
Measure Touch Voltage
No-Trip Test with ½ × I
and 10 s
N
Set Parameters for I
After 10 seconds have elapsed, no residual current may be indicated. The measurement must then be evaluated. In the event
that “NOT OK” is selected (in case of false alarm), an error is indicated by the UL/RL LED which lights up red.
The measured value cannot be saved to memory and included in
the test report until it has been evaluated.
F
Tripping test with 1 × I
N
– Measurement of Signal Response Time (Stopwatch Function)
with Residual Current Generated by the Test Instrument
Measurement must be stopped manually by pressing the ON/
START ▼ or I
in order to document tripping time.
90Gossen Metrawatt GmbH
In the event that NOT OK is selected, an error is indicated by the UL/RL LED which lights up red.
The measured value cannot be saved to memory and included in
the test report until it has been evaluated.
key immediately after indication of residual current,
N
Page 91
19.9Checking the Operating Statuses of Electric Vehicles at
Charging Stations per IEC 61851 ((PROFITEST MTECH+ &
PROFITEST MXTRA)
A charging station is a facility designed to charge electric vehicles
in accordance with IEC 61851-1, and is equipped with essential
elements including a plug connector, conductor protection, an
RCD, a circuit breaker and a safety communication device (PWM).
Depending on where it’s used, other function modules may be
added, for example for mains connection and metering.
Selecting the Adapter (test box)
Simulation of Operating Statuses per IEC 61851-1 with the MENNEKES Test Box
(Statuses A through E)
The MENNEKES test box is used exclusively to simulate the vari-
ous operating statuses of a fictitious electric vehicle connected to
a charging station. Settings for the simulated operating statuses
can be found in the operating instructions for the test box.
The simulated operating statuses can be saved to as a visual
inspection and documented in the report generating program.
Select the respective status to be checked with the
SECLECT STATUS key at the test instrument.
Status C – non-gassing vehicle detected
• Vehicle is ready for charging / power is connected
• Voltage between PE and CP: +6 V / -12 V
Status D – gassing vehicle detected
• Vehicle is ready for charging / power is connected
• Voltage between PE and CP: +3 V / -12 V
Status A – charging cable connected to charging point only
• CP signal is activated.
• Voltage between PE and CP is 12 V.
Status B – charging cable connected to charging point and vehicle
• Charging cable is locked into place at the charging point and
the vehicle.
• Vehicle is not yet ready for charging.
• Voltage between PE and CP: +9 V / -12 V
Status E – cable is damaged
• Short-circuit between PE and CP
• Charging cable is unlocked at the charging point.
• Voltage between PE and CP is +0 V.
Semi-Automatic Changing of Operating Statuses
As an alternative to manual status changing via the parameters menu for the
SECLECT STATUS softkey at the
test instrument, quick and convenient switching amongst the
statuses is also possible. The
AUTO status parameter has to
be selected to this end. After
responding to the visual
inspection prompt and saving
the results, automatic switching
to the next status ensues – the
01/05 key display corresponds
to A/E (01 = A, 02 = B, 03 = C, 04 = D, 05 = E).
Status variants can be skipped by pressing the I
instrument or the test plug.
key at the test
N
Gossen Metrawatt GmbH91
Page 92
19.10 PRCD – Test Sequences for Documenting Fault Simula-
Attention!
!
tions at PRCDs with the PROFITEST PRCD Adapter (PROFITEST MXTRA only)
The PROFITEST PRCD test adapter can be used in combination
with the test instrument.
Read the respective operating instructions before using
the PROFITEST PRCD.
Measurements with the PROFITEST PRCD connected to the test
instrument:
• Measurement of the PRCD’s insulation resistance using the
test instrument’s R
• Measurement of the PRCD’s protective conductor resistance
using the test instrument’s RLO function. Please note that the
protective conductor measurement is a modified RLO measurement with ramp sequence for PRCDs (seesection 17).
• Tripping test with nominal residual current using the test
instrument’s I
• Measurement of time to trip using the test instrument’s I
function (see section 12.3).
• Varistor test for PRCD-K: measurement via ISO ramp (see
section 16).
F
function (see section 16).
INS
function (see section 12.3).
N
19.10.1 Fault Simulation
The procedure for the PROFITEST PRCD, including the procedure with the device under test, is described in the operating
instructions for the PROFITEST PRCD. This section describes the
procedure for the test instrument.
Procedure
➭ Prepare error simulation at the PROFITEST PRCD. Refer to
the operating instructions for the PROFITEST PRCD.
➭ Select the test sequence at the test instrument.
➭ Execute each of the test sequence steps at the PROFITEST
PRCD and document evaluation and assessment at the test
instrument.
Select the PRCD to be Tested
Testing performed by simulating faults is carried out without connection to the test instrument, but it’s accompanied and documented by the test instrument. The test sequence is opened in
the test instrument to this end and the specified steps are executed at the PROFITEST PRCD. Afterwards, evaluation and
assessment of each test step (OK or not OK) is performed at the
test instrument for later documentation.
There are three preset test sequences:
– PRCD-S (single-phase / 3-pole): 11 test steps
– PRCD-K (single-phase / 3-pole): 4 test steps
– PRCD-S (3-phase / 5-pole): 18 test steps
Interaction Between PROFITEST PRCD and Test Instrument
Switch
Position
at
Display at
Test Instrument
Test StepIcon
Meaning
PROFITEST PRCD
ON1~ON
ON3~ON
BREAK Lx
Lx <-> PE
Lx <-> N
PE-U
EXT
Uext -> PE
PROBE
PRCD-Ip
—AUTOAUTO
Single-phase PRCD activated
3-phase PRCD activated
Interrupted phase
Wires reversed between phase
conductor and PE or neutral
conductor
PE to phase
Contact ON key on PRCD with
probe
Protective conductor current
measurement
with current clamp transformer
Semi-automatic change of
fault simulations
The test steps are displayed at the test instrument. Their meanings and the associated switch positions at the PROFITEST
PRCD are listed in the above table.
92Gossen Metrawatt GmbH
Page 93
Overview of Test Sequences and their Test Steps
PRCD-S, single-phase: 11 test steps
Selection Examples, PRCD-S Test Sequence (single-phase) –
11 Test Steps
Simulation of Interruption (steps 1 to 6)
PRCD-S, 3-phase: 18 test steps
PRCD-K, single-phase: 5 test steps
Reversed Conductor Simulation (step 7)
Simulation of PE to Phase (step 8)
Contact ON Key at PRCD with Probe (step 10)
Measurement of Protective Conductor Current with a Current
Clamp Transformer (step 11)
Gossen Metrawatt GmbH93
Page 94
Selection Examples, PRCD-S Test Sequence (3-phase) –
18 Test Steps
Simulation of Interruption (steps 1 to 10)
Reversed Conductor Simulation (steps 11 to 16)
Semi-Automatic Changing of Fault Simulations (Statuses)
As an alternative to manual status changing via the parameters
menu for the respective PRCD
selection at the test instrument
(PRCD-S 1~, PRCD-K 1~ or PRCD-S3~), quick and convenient
switching amongst the fault
simulations is also possible. The
AUTO status parameter has to
be selected to this end. After
responding to the visual inspection prompt and saving the
results, automatic switching to
the next fault simulation ensues.
Skipping Test Steps
Test steps can be skipped during fault simulation by pressing the
key at the test instrument or the test plug.
I
N
Simulation of PE to Phase (step 17)
Measurement of Protective Conductor Current with a Current
Clamp Transformer (step 18)
94Gossen Metrawatt GmbH
Page 95
20Test Sequences (Automatic Test Sequences)
AUTO
1
2 3 4
5 6 7
8
109
1112
Select the utilized test
instrument.
!
– AUTO Function
Select AUTO Switch Position at the Test Instrument
With the rotary switch in the AUTO position, all of the test
sequences in the device are displayed.
If there aren’t any test sequences in the instrument, NO DATA
appears.
20.1General (test sequence layouts)
If the same sequence of tests will be run frequently (one after the
other with subsequent report generation), for example as specified in the standards, it’s advisable to make use of test
sequences.
Automated test sequences can be compiled from manually created individual measurements with the help of the test sequence
function.
A test sequence consists of up to 200 individual steps, which are
executed one after the other.
Fundamentally, differentiation is made amongst three types of
individual steps:
•Note (Visual Inspection test step) Test sequences are inter-
rupted when a pop-up message is displayed for the inspector.
The test sequences is not resumed until the message has
been acknowledged.
Sample Message Before Insulation Resistance Measurement
“Disconnect the device from the mains!”
• Visual inspection, testing and report generation: The test
sequence is interrupted when a passed/failed evaluation is
displayed. The comment and the results of the evaluation are
saved to the database.
• Measurement (“User-Evaluated Measurement” test step): same
as individual measurements with instruments with storage and
parameters configuration
20.2Creation of Test Sequences with ETC
The test sequences are created at the PC with the help of ETC
software, and are then transferred to the test instrument.
Measurement parameters are also configured at the PC. However, parameters can be changed at the test instrument during
the test sequence before the respective measurement is started.
When the test step is started once more, the parameter settings
specified in ETC are loaded again.
ETC does not subject the parameters to a plausibility check. As a
result, the newly created test sequence should be checked at the
test instrument before it’s permanently added to the database.
Limit values are not currently set in ETC, and have to be adjusted
during the automatic test.
Accessing the Menu for Editing Test Sequences
In order to be able to edit existing test sequences (e.g. add test
steps or change parameter settings), they first have to be loaded
to ETC.
There are two ways to do this:
•ETC: Extras Test Sequences Load Test Sequences
or
• ETC: Device Test Sequences Receive Test Sequences
(from the file “test_sequence_xyz.seq”)
(from the connected test instrument)
Operating Overview: Creating Test Sequences at a PC
1 Create a new test sequence – enter a designation.
2 Change the designation of the selected test sequence.
3 Duplicate the selected test sequence,
(Copy) is added to the name of the duplicated sequence.
4 Delete the selected test sequence.
5 Create or add a new test step for the selected test sequence.
– Select the test step type from the list to this end and either accept or edit
its designation.
6 Duplicate the selected test step.
7 Delete the selected test step.
8 Change position of the selected test step within the sequence.
9 Select test parameters for the selected test step type from the list.
10 Select a setting for the measuring parameter from the list.
11 Accept change to the measuring parameter.
12 Exit the test sequences menu.
Saving Test Sequences in ETC to the PC
We recommend saving default test sequences, as well as edited
and new test sequences, to the PC or to other data
storage media using the desired filename (test_sequence_xyz.seq) with the help of the following menu command:
Extras Test Sequences Save Test Sequences. Data loss
resulting from certain administrative operations is prevented in this
way (see following notes).
Due to the fact that only up to 10 test sequences, can be transferred to the test instrument, no more than 10 test sequences
should be saved to any given file.
Test sequences which have been saved to a file can be reloaded
to ETC at any time by clicking “Extras Test Sequences Load
Test Sequences”.
Sequences can be further edited by clicking
“Extras Test Sequences Edit Test Sequences”.
Transferring Test Sequences from the PC to the Test Instrument
After executing the following ETC command, all previously created test sequences (up to 10) are transferred to the connected
test instrument: “Device Test Sequences Send Test
Sequences”.
Gossen Metrawatt GmbH95
Page 96
Attention!
!
Test sequences which have been loaded to the test instrument are deleted when:
– New test sequences are received from the PC
– Selection lists are received from the PC
– Backup data is restored to the test instrument
– The user interface language is changed
– The test instrument’s entire database is deleted
– The test instrument is reset to its default settings
– The firmware is updated
For as long as the test
sequences are being
transferred, a progress
bar is displayed at the
PC and the illustration
shown to the right
appears at the test
instrument’s display.
After data transmission
has been completed,
the display is switched
to the database memory
menu.
The display is returned
to the measuring menu
for the respective switch
position by clicking ESC.
The selected test sequence (SEQU.1 in this case) is started with the
ON/START ▼ key.
When a test step of the measurement type is executed, the same
screen layout appears as is also the case for individual measurements. The current test step number appears in the header
instead of the memory and battery icons. After pressing the Save
key twice, the next test step is displayed.
Setting Parameters and Limit Values
Parameters and limit values can also be changed while a test
sequence is running or before the respective measurement is
started. The respective change only affects the active test
sequence and is not saved.
Skipping Test Steps
There are two ways to skip test steps or individual measurements:
• Select the test sequence, change to the test step column at
the right with the help of the cursor, select the x
and press the ON/START ▼ key.
• The navigation menu can be
opened within the test sequence
by pressing the navigation key
(cursor left-right). Jumping to the
next or the previous test step is
possible using the separate scroll
keys which then appear.
The navigation menu can be exited again and the current test
step can be displayed by pressing the ESC key.
th
test step
20.3Using Test Sequences
Test Sequence Commands
Acknowledge message
Discard event
Confirm event
To prev io u s /
to next step
Save measurement
results
Configuring Test Sequence Parameters
Measurement parameters are also configured at the PC. However, parameters can be changed at the test instrument during
the test sequence before the respective measurement is started.
Selecting and Starting a Test Sequence at the Test Instrument
Aborting or Ending a Test Sequence
An active sequence can be aborted by pressing the ESC key and
then acknowledging.
Sequence Ended appears after the last test step is completed. The
initial menu, List of Test Sequences, is once again displayed after
acknowledging the prompt.
96Gossen Metrawatt GmbH
Page 97
21Maintenance
Note
Note
Attention!
!
Note
21.1Test Instrument Firmware/Software
The layout of the test instruments makes it possible to adapt
device software to the latest standards and regulations. Beyond
this, suggestions from customers result in continuous improvement of the test instrument software, as well as new functions.
Query Current Status
➭ Turn the rotary switch to the SETUP position.
➭ Press the SW-Info CALIBRATION key.
➭ Press any key in order to return to the main menu.
Update
Internal test instrument firmware/software can be updated via the
USB port with the help of a PC and an interface cable.
The firmware/software with the required version is transferred to
the test instrument with the help of the MASTER Updater Software-Tool. Currently installed test instrument firmware/software is
overwritten.
The MASTER Updater can be downloaded free of charge from
www.gossenmetrawatt.com. Registration with myGMC is
required to this end. Operating instructions for the Firmware
Update Tool are available here as well.
Prerequisite for transfer:
The rotary selector switch is not set to the U position.
Severe damage to the instrument may occur if incorrect
fuses are used.
Only original fuses from Gossen Metrawatt GmbH may
be used (order no. 3-578-285-01 / SIBA 7012540.3.15
SI-EINSATZ FF 3.15/500 6.3X32).
Only original fuses assure required protection by means
of suitable blowing characteristics. Short-circuiting of
fuse terminals or the repair of fuses is prohibited, and is
life endangering!
The instrument may be damaged if fuses with incorrect
ampere ratings, breaking capacities or blowing characteristics are used!
➭ Remove the blown fuse and insert a new one.
➭ Insert the fuse compartment lid after the fuse has been re-
placed and secure it by turning clockwise.
21.3Housing
No special maintenance is required for the housing.
Keep outside surfaces clean. Use a slightly dampened cloth for
cleaning. In particular for the protective rubber surfaces, we recommend a moist, lint-free microfiber cloth. Avoid the use of
cleansers, abrasives or solvents.
21.4Calibration
Use of your instrument and resultant stressing influence the
instrument and lead to deviation from warranted accuracy values.
In the case of strict measuring accuracy requirements, as well as
in the event of severe stressing (e.g. severe climatic or mechanical
stress), we recommend a relatively short calibration interval of
once per year. If this is not the case, a calibration interval of 2 to 3
years is usually adequate.
Please contact GMC-I Service GmbH for calibration services (see
section 22, “Contact, Support and Service”, on page 98).
A sticker with an instrument-specific guideline value for the calibration interval and information regarding the service provider is
included on the instrument as an aid.
➭ Establish a USB connection between the PC and the test in-
strument.
➭ Switch the PC and the test instrument on.
➭ Follow the instructions displayed by the MASTER Updater and
the associated operating instructions.
21.1.1 Rechargeable Battery Care
Check to make sure that no leakage has occurred at the
rechargeable batteries at short, regular intervals, or after the
instrument has been in storage for a lengthy period of time.
Remove rechargeable batteries during lengthy periods of
non-use (e.g. vacation). This prevents excessive depletion or leakage, which may result in damage to the test
instrument.
21.2Fuse Replacement
If a fuse has blown due to overloading, a corresponding message
error appears at the display panel. The instrument’s voltage measuring ranges are nevertheless still functional.
➭ Disconnect the device from the measuring circuit at all poles!
➭ Loosen the slotted screws at the fuse compartment lid next to
the mains power cable with a screwdriver. The fuses are now
accessible.
➭ Replacement fuses can be accessed after opening the battery
compartment lid.
Date on Calibration Certificate / Calibration Interval Begins Upon
Receipt
Your instrument is furnished with a calibration certificate
on which a date appears. This date may be further in the
past if your instrument has been stored for some time
prior to sale.
The instruments are stored in accordance with the specified conditions. Drift is thus negligible for a duration of
1 year. Longer storage periods are highly unusual.
Consequently, the instrument’s characteristic values lie
within the specifications and the first calibration interval
can be determined as of the date of receipt.
Gossen Metrawatt GmbH97
Page 98
22Contact, Support and Service
23CE Declaration
Gossen Metrawatt GmbH can be reached directly and simply –
we have a single number for everything! Whether you require support or training, or have an individual inquiry, we can answer all of
your questions here:
+49-911-8602-0
Monday to Thursday:
Friday:
8 a.m. to 4 p.m.
8 a.m. to 2 p.m.
Or contact us by e-mail at:
info@gossenmetrawatt.com
Do you prefer support by e-mail?
Measuring and Test Technology:
support@gossenmetrawatt.com
The instrument fulfills all requirements of applicable EU directives
and national regulations. We confirm this with the CE mark. The
CE declaration is available upon request.
A calibration certificate is included with the instrument.
Please contact GMC-I Service GmbH for repairs, replacement
parts and calibration
1)
:
+49-911-817718-0Beuthener Str. 41
service@gossenmetrawatt.com
www.gmci-service.com
90471 Nürnberg
Germany
1)
DAkkS calibration laboratory per DIN EN ISO/IEC 17025 –
accredited by the Deutsche Akkreditierungsstelle GmbH under reference number D-K-15080-01-01.
98Gossen Metrawatt GmbH
Page 99
24Disposal and Environmental Protection
Attention!
!
Attention!
!
Proper disposal makes an important contribution to the protection
of our environment and the conservation of natural resources.
Environmental Damage
Improper disposal results in environmental damage.
Follow the instructions concerning return and disposal
included in this section.
The following comments refer specifically to the legal situation in
the Federal Republic of Germany. Owners or end users who are
subject to other national requirements are required to comply with
the respectively applicable national requirements and to implement them correctly on site. Relevant information can be
obtained, for example, from the responsible national authorities or
national distributors.
Waste Electrical Equipment, Electrical or Electronic Accessories
and Waste Batteries (including rechargeable batteries)
Electrical equipment and batteries (including rechargeable batteries) contain valuable raw materials that can be recycled, as well as
hazardous substances which can cause serious harm to human
health and the environment, and they must be recycled and disposed of correctly.
The symbol on the left depicting a crossed-out garbage
can on wheels refers to the legal obligation of the owner
or end user (German electrical and electronic equipment
act ElektroG and German battery act BattG) not to dispose of used electrical equipment and batteries with unsorted
municipal waste (“household trash”). Waste batteries must be
removed from the old device (where possible) without destroying
them and the old device and the waste batteries must be disposed of separately. The battery type and its chemical composition are indicated on the battery’s labelling. If the abbreviations
“Pb” for lead, “Cd” for cadmium or “Hg” for mercury are included,
the battery exceeds the limit for the respective metal.
Please observe the owner’s or end user’s responsibility with
regard to deleting personal data, as well as any other sensitive
data, from old devices before disposal.
Old devices, electrical or electronic accessories and waste batteries (including rechargeable batteries) used in Germany can be
returned free of charge to Gossen Metrawatt GmbH or the service
provider responsible for their disposal in compliance with applicable regulations, in particular laws concerning packaging and hazardous goods. Further information regarding returns can be found
on our website.
Packaging Materials
We recommend retaining the respective packaging materials for
the case that you might require servicing or calibration in the
future.
Danger of Asphyxiation Resulting from Foils and Other
Packaging Materials
Children and other vulnerable persons may suffocate if
they wrap themselves in packaging materials, or their
components or foils, or if they pull them over their heads
or swallow them.
Keep packaging materials, as well as their components
and foils, out of the reach of babies, children and other
vulnerable persons.
In accordance with German packaging law (VerpackG), the user
is obligated to correctly dispose of packaging and its components
separately, and not together with unsorted municipal waste
(“household trash”).
Private end consumers can dispose of packaging free of charge
at the responsible collection point. Packaging which is not subject
to so-called system participation is returned to the appointed service provider. Further information regarding returns can be found
on our website.
Gossen Metrawatt GmbH99
Page 100
25Appendix
25.1Tables for Determining Maximum and Minimum Display Values in Consideration of the Instrument’s Maximum Measuring and
Intrinsic Uncertainties
Table 1
Z
(full-wave) / Z
L-P E.
()
Limit
Valu e
Display Value
0.100.070.100.05
0.150.110.150.10
0.200.160.200.14
0.250.200.250.18
0.300.250.300.22
0.350.300.350.27
0.400.340.400.31
0.450.390.450.35
0.500.430.500.39
0.600.510.600.48
0.700.600.700.56
0.800.700.800.65
0.900.790.900.73
1.000.881.000.82
1.501.401.501.33
2.001.872.001.79
2.502.352.502.24
3.002.823.002.70
3.503.303.503.15
4.003.784.003.60
4.504.254.504.06
5.004.735.004.51
6.005.686.005.42
7.006.637.006.33
8.007.598.007.24
9.008.549.008.15
9.999.489.999.05
Max.
Z
L-N
(+/- half-wave) ()
L-P E.
Limit
Value
Max.
Display Value
Table 3
R
M
Limit
Valu e
Display Value
0.100.1210.010.7
0.150.1715.015.9
0.200.2320.021.2
0.250.2825.026.5
0.300.3330.031.7
0.350.3835.037.0
0.400.4440.042.3
0.450.4945.047.5
0.500.5450.052.8
0.550.5960.063.3
0.600.6570.073.8
0.700.7580.084.4
0.800.8690.094.9
0.900.96100106
1.001.07150158
1.501.59200211
2.002.12250264
2.502.65300316
3.003.17
3.503.70
4.004.23
4.504.75
5.005.28
6.006.33
7.007.38
8.008.44
9.009.49
Min.
INS
Limit
Value
Min.
Display Value
Table 2
R
/ R
()
E
Limit
Value
0.100.0710.09.491.00 k906
0.150.1115.013.61.50 k1.36 k
0.200.1620.018.12.00 k1.81 k
0.250.2025.022.72.50 k2.27 k
0.300.2530.027.23.00 k2.72 k
0.350.3035.031.73.50 k3.17 k
0.400.3440.036.34.00 k3.63 k
0.450.3945.040.84.50 k4.08 k
0.500.4350.045.45.00 k4.54 k
0.600.5160.054.56.00 k5.45 k
0.700.6070.063.67.00 k6.36 k
0.800.7080.072.78.00 k7.27 k
0.900.7990.081.79.00 k8.17 k
1.000.8810090.89.99 k9.08 k
1.501.40150133
2.001.87200179
2.502.35250224
3.002.82300270
3.503.30350315
4.003.78400360
4.504.25450406
5.004.73500451
6.005.68600542
7.006.63700633
8.007.59800724
9.008.54900815
Max.
Display Value
Limit
Valu e
ELoop
Max.
Display Value
Limit
Value
Max.
Display Value
Table 4
R
Limit
Valu e
Display Value
0.100.0710.09.59
0.150.1215.014.4
0.200.1720.019.2
0.250.2225.024.0
0.300.2630.028.8
0.350.3135.033.6
0.400.3640.038.4
0.450.4145.043.2
0.500.4650.048.0
0.600.5560.057.6
0.700.6570.067.2
0.800.7580.076.9
0.900.8490.086.5
1.000.9499.996.0
1.501.42
2.001.90
2.502.38
3.002.86
3.503.34
4.003.82
4.504.30
5.004.78
6.005.75
7.006.71
8.007.67
9.008.63
Max.
LO
Limit
Value
Max.
Display Value
100Gossen Metrawatt GmbH
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.