GOSSEN PROFITEST MXTRA User Manual

Page 1
Operating Instructions
PROFITEST MASTER Series PROFITEST M
TECH+, MPRO, MXTRA, MBASE+
IEC 60364-6, EN 50110-1
3-447-149-03
1/4.23
Page 2
1 Safety Instructions .............................................................. 4
2 Applications......................................................................... 5
2.1 Intended Use / Use for Intended Purpose ............5
2.2 Use for Other than Intended Purpose...................5
2.3 Liability and Guarantee.........................................5
2.4 Opening the Instrument / Repairs.........................5
2.5 Scope of Functions ..............................................5
3 Documentation.................................................................... 6
4 Getting Started .................................................................... 6
5 The Instrument.................................................................... 7
5.1 Scope of Delivery .................................................7
5.2 Optional Accessories (excerpt) .............................7
5.3 Meanings of Symbols on the Instrument...............7
5.4 Instrument Overview ............................................8
5.5 Technical Data ...................................................10
5.6 Characteristic Values for PROFITEST MTECH+
and PROFITEST MBASE+..................................11
5.7 Characteristic Values for PROFITEST MXTRA
and PROFITEST MPRO .....................................13
6 Operating and Display Elements ....................................... 16
6.1 Control Panel .....................................................16
6.2 Display...............................................................16
6.3 LEDs..................................................................16
6.4 LED Indications, Mains Connections and
Potential Differences ..........................................17
7 Operation........................................................................... 26
7.1 Power Supply ....................................................26
7.1.1 Inserting or Replacing the Battery Pack (Z502H) or Commercially Available Individual
(Rechargeable) Batteries ..............................26
7.1.2 Charging the Battery Pack (Z502H) in the
Tester ..........................................................26
7.2 Switching the Instrument On/Off ........................26
8 Instrument Settings........................................................... 27
9 Database ........................................................................... 31
9.1 Creating Distributor Structures, General .............31
9.2 Transferring Distributor Structures ......................31
9.3 Creating a Distributor Structure in the Test
Instrument..........................................................31
9.3.1 Creating Structures (example for electrical
circuit)..........................................................33
9.3.2 Searching for Structure Elements.................34
9.4 Saving Data and Generating Reports .................34
9.5 Use of Barcode Scanners and RFID Readers.....35
10 General Information on Measurements............................. 36
10.1 Using Cable Sets and Test Probes.....................36
10.2 Test Plug – Changing Inserts..............................36
10.3 Connecting the Instrument .................................36
10.4 Automatic Settings, Monitoring and Shutdown...36
10.5 Measured Value Display and Memory ................36
10.6 Help Function.....................................................37
10.7 Setting Parameters or Limit Values using RCD
Measurement as an Example .............................38
10.8 Freely Selectable Parameter Settings or Limit
Values ............................................................... 39
10.8.1 Changing Existing Parameters .................... 39
10.8.2 Adding New Parameters ............................. 39
10.9 2-Pole Measurement with Rapid or
Semiautomatic Polarity Reversal........................ 40
11 Measuring Voltage and Frequency.................................... 41
11.1 Single-Phase Measurement............................... 41
11.1.1 Voltage Between L and N (U (U
) a
L-PE
nd N and PE
(U
N-PE
),
L and PE
L-N
) with Country-
Specific Plug Insert, e.g. SCHUKO.............. 41
11.1.2 Voltage Between L – PE, N – PE and L – L
with 2-Pole Adapter Connection ................. 41
11.2 3-Phase Measurement (line-to-line voltage) and
Phase Sequence............................................... 41
12
Testing RCDs .......................................................................... 43
12.1 Measuring Touch Voltage (with reference to nomi­nal residual current) with ⅓ Nominal Residual Current and Tripping Test with Nominal Residual
Current.............................................................. 44
12.2
Special Tests for Systems and RCDs.....................46
12.2.1 Testing Systems and RCCBs with Rising Residual Current (AC) for Type AC, A/F, B/B+ and EV/MI RCDs (PROFITEST MTECH+,
PROFITEST MXTRA only)............................ 46
12.2.2 Testing Systems and RCCBs with Rising Residual Current (AC) for Type B/B+ and EV/MI RCDs (PROFITEST MTECH+PROFITEST
MXTRA) ...................................................... 47
12.2.3 Testing RCCBS with 5 × I
....................... 47
N
12.2.4 Testing of RCCBs which are Suitable for
Pulsating DC Residual Current .................... 48
12.3 Testing of Special RCDs.................................... 48
12.3.1 Systems with Type RCD-S Selective
RCCBs ....................................................... 48
12.3.2 PRCDs with Non-Linear Type PRCD-K
Elements..................................................... 49
12.3.3 SRCD, PRCD-S (SCHUKOMAT, SIDOS or
comparable)................................................ 50
12.3.4 Type G or R RCCB ..................................... 50
12.4 Testing Residual Current Circuit Breakers in TN-S
Systems............................................................ 51
12.5 Testing of RCD Protection in IT Systems with High
Cable Capacitance (e.g. In Norway)................... 52
12.6 Testing of 6 mA Residual Current Devices RDC­DD/RCMB (RDC-DD: PROFITEST MXTRA and
PROFITEST MTECH+ only) ............................... 52
13 Testing of Breaking Requirements for Overcurrent Protective
Devices, Measurement of Loop Impedance and Determina­tion of Short-Circuit Current (ZL-PE and I
Functions)..... 54
SC
13.1 Measurements with Suppression of RCD Tripping (PROFITEST MTECH+, PROFITEST
MXTRA only) ..................................................... 54
13.1.1 Measurement with Positive Half-Waves (PROFITEST MTECH+, PROFITEST MXTRA
only)............................................................ 55
13.2 Evaluation of Measured Values.......................... 55
2 Gossen Metrawatt GmbH
Page 3
13.3 Settings for Calculating Short-Circuit Current – Parameter I
14 Measuring Supply Impedance (Z 15 Earthing Resistance Measurement (Function R
.................................................56
SC
Function)...................57
L-N
).............. 59
E
15.1 Earthing Resistance Measurement –
Mains Powered..................................................60
15.2 Earthing Resistance Measurement – Battery Pow­ered, “Battery Mode” (PROFITEST MPRO & PROF-
ITEST MXTRA only)............................................60
15.3
Earthing Resistance, Mains Powered – 2-Pole Mea­surement with 2-Pole Adapter or Country-Specific Plug (Schuko) without Probe
................................61
15.4 Earthing Resistance Measurement. Mains Powered – 3-Pole Measurement: 2-Pole Adapter
with Probe .........................................................62
15.5 Earthing Resistance Measurement, Mains Powered – Measuring Earth Electrode Potential (U
Function) .....................................................63
E
15.6 Earthing Resistance Measurement, Mains Powered – Selective Earthing Resistance Measurement
with Current Clamp Sensor as Accessory ..........64
15.7 Earthing Resistance Measurement, Battery Powered, “Battery Mode” – 3-Pole (PROFITEST
MPRO & PROFITEST MXTRA only) ....................66
15.8
Earthing Resistance Measurement, Battery Powered, “Battery Mode” – 4-Pole ( MPRO
&
PROFITEST MXTRA
PROFITEST
only) ..................... 67
15.9 Earthing Resistance Measurement, Battery Powered, “Battery Mode” – Selective (4-pole) with Current Clamp Sensor and PRO-RE Measuring Adapter as Accessory (PROFITEST MPRO &
PROFITEST MXTRA only)...................................69
15.10Earthing Resistance Measurement, Battery Pow­ered, “Battery Mode” – Ground Loop Measurement (with current clamp sensor and transformer, and pro-re measuring adapter as accessory) (PROFIT-
EST MPRO & PROFITEST MXTRA only).............70
15.11Earthing Resistance Measurement, Battery Powered, “Battery Mode” – Measurement of Soil Resistivity
(PROFITEST MPRO & PROFITEST
E
MXTRAonly) .......................................................71
16
Measurement of Insulation Resistance...................................73
16.1 General..............................................................73
16.2 Special Case: Earth Leakage Resistance (R
17
Measuring Low-Value Resistance of up to 200 (Protective
EISO
)75
Conductor and Equipotential Bonding Conductor) ..................77
17.1 Measurement with Constant Test Current..........78
17.2 Protective Conductor Resistance Measurement with Ramp Sequence – Measurement at PRCDs with Current-Monitored Protective Conductor using the PROFITEST PRCD Test Adapter as an
Accessory ( PROFITEST MXTRAonly).................79
18 Measurement with Accessory Sensors..............................80
18.1 Current Measurement with Current Clamp
Sensor...............................................................80
19 Special Functions – EXTRA Switch Position ......................81
19.1 Voltage Drop Measurement (at Z
LN) –
U Function .......................................................82
19.2 Measuring the Impedance of Insulating Floors and Walls (standing surface insulation impedance) – Z
Function ......................................................83
ST
19.3 Testing Meter Startup with Earthing Contact Plug
– kWh Function .................................................84
19.4 Leakage Current Measurement with PRO-AB Leakage Current Adapter as Accessory – I
Function (PROFITEST MXTRA only) ................85
L
19.5
Testing Insulation Monitoring Devices – IMD Function
(PROFITEST MXTRA only)..................................86
19.6 Residual Voltage Test – U
Function
res
(PROFITEST MXTRA only)..................................88
19.7
Intelligent Ramp – ta+ID (
PROFITEST MXTRA
Function
only).................................... 89
19.8 Testing Residual Current Monitors
– RCM Function ( PROFITEST MXTRA only).......90
19.9 Checking the Operating Statuses of Electric Vehicles at Charging Stations per IEC 61851 ((PROFITEST MTECH+ & PROFITEST MXTRA) ..91
19.10PRCD – Test Sequences for Documenting Fault Simulations at PRCDs with the PROFITEST PRCD
Adapter (PROFITEST MXTRA only).....................92
19.10.1Fault Simulation ...........................................92
20 Test Sequences (Automatic Test Sequences)
– AUTO Function ................................................................95
20.1 General (test sequence layouts)..........................95
20.2 Creation of Test Sequences with ETC ................95
20.3 Using Test Sequences .......................................96
21 Maintenance ......................................................................97
21.1 Test Instrument Firmware/Software....................97
21.1.1 Rechargeable Battery Care ..........................97
21.2 Fuse Replacement .............................................97
21.3 Housing .............................................................97
21.4 Calibration..........................................................97
22 Contact, Support and Service ............................................98
23 CE Declaration ...................................................................98
24 Disposal and Environmental Protection .............................99
25 Appendix ..........................................................................100
25.1 Tables for Determining Maximum and Minimum Display Values in Consideration of the Instrument’s Maximum Measuring and Intrinsic
Uncertainties....................................................100
25.2 At which values should/must an RCD actually be tripped? Requirements for Residual Current
Devices (RCD)..................................................102
25.3 Testing Electrical Machines per DIN EN 60204 –
Applications, Limit Values.................................103
25.4 Periodic Testing per DGUV V 3 (previously BGV A3) – Limit Values for Electrical Systems and Operating
Equipment .......................................................103
25.5 Bibliography.....................................................104
25.6 Internet Addresses for Additional Information ..104
Gossen Metrawatt GmbH 3
Page 4

1 Safety Instructions

Observe this documentation, in particular all included safety infor­mation, in order to protect yourself and others from injury, and to prevent damage to the instrument.
The operating instructions and the condensed operating instruc­tions should be made available to all users.
General
• Tests/measurements may only be performed by a qualified electrician, or under the supervision and direction of a quali­fied electrician. The user must be instructed by a qualified electrician concerning performance and evaluation of tests and/or measurements.
• Observe the five safety rules in accordance with DIN VDE 0105-100:2015-10, VDE 0105-100:2015-10 (EN 50110-1), Operation of electrical installations – Part 100: General requirements (1: Shut down entirely. 2: Secure against restart. 3: Assure absence of voltage at all poles. 4: Ground and short circuit. 5: Cover neighboring live components, or make them inaccessi­ble).
• Observe and comply with all safety regulations which are applicable for your work environment.
• Wear suitable and appropriate personal protective equipment (PPE) whenever working with the instrument.
• The functioning of active medical devices (e.g. pacemakers, defibrillators) and passive medical devices may be affected by voltages, currents and electromagnetic fields generated by the tester and the health of their users may be impaired. Implement corresponding protective measures in consultation with the manufacturer of the medical device and your physi­cian. If any potential risk cannot be ruled out, do not use the instrument.
Accessories
• Use only the specified accessories (included in the scope of delivery or listed as options) with the instrument.
• Carefully and completely read and adhere to the product doc­umentation for optional accessories. Retain these documents for future reference.
Handling
• Use the instrument in undamaged condition only. Inspect the instrument before use. Pay particular attention to damage, interrupted insulation or kinked cables. Damaged components must be replaced immediately.
• Accessories and cables may only be used as long as they’re fully intact. Inspect accessories and all cables before use. Pay particular attention to damage, interrupted insulation or kinked cables.
• If the instrument or its accessories don’t function flawlessly, permanently remove the instrument/accessories from opera­tion and secure them against inadvertent use.
• If the instrument or accessories are damaged during use, for example if they’re dropped, permanently remove the instru­ment/accessories from operation and secure them against inadvertent use.
• The instrument and the accessories may only be used for the tests/measurements described in the documentation for the instrument.
• Neither the integrated voltage measuring function nor the mains connection test may be used to test systems or system components for the absence of voltage. Testing for the absence of voltage is only permissible with a suitable voltage tester or voltage measuring system which ful­fills the requirements specified in DIN EN 61243.
Operating Conditions
• Do not use the instrument and its accessories after long peri­ods of storage under unfavorable conditions (e.g. humidity, dust or extreme temperature).
• Do not use the instrument and its accessories after extraordi­nary stressing due to transport.
• The instrument must not be exposed to direct sunlight.
• Only use the instrument and its accessories within the limits of the specified technical data and conditions (ambient condi­tions, IP protection code, measuring category etc.).
• Do not use the instrument in potentially explosive atmo­spheres.
Rechargeable Batteries
• When using the charger, only the battery pack (Z502H) may be inserted in the device.
• Do not use the instrument while charging the battery pack (Z502H).
• Do not use the test instrument if the battery compartment lid has been removed. Touch contact with dangerous voltage is otherwise possible.
• The battery pack (Z502H) may only be charged in undamaged condition. Inspect the battery pack (Z502H) before use. Pay particular attention to leaky and damaged batteries.
Fuses
• The instrument is equipped with fuses. The instrument may only be used as long as the fuses are in flawless condition. Defective fuses must be replaced. See detailed operating instructions.
Measurement Cables and Establishing Contact
• Plugging in the measurement cables must not necessitate any undue force.
• Never touch conductive ends (e.g. of test probes).
• Fully unroll all measurement cables before starting a test/mea­surement. Never perform a test/measurement with the mea­surement cable rolled up.
• Avoid short circuits due to incorrectly connected measure­ment cables.
• Ensure that alligator clips, test probes or Kelvin probes make good contact.
Data Security
• Always create a backup copy of your measurement data.
• Observe and comply with the respectively applicable national data protection regulations. Use the corresponding functions provided by the test instrument such as access protection, as well as other appropriate measures.
4 Gossen Metrawatt GmbH
Page 5

2 Applications

Please read this important information!

2.1 Intended Use / Use for Intended Purpose

Measuring and test instruments from the PROFITEST MASTER series include:
• PROFITEST MBASE+ (M520S)
• PROFITEST MXTRA (M522P)
• PROFITEST MTECH+ (M522R)
• PROFITEST MPRO (M520N) The test instruments are used to test the effectiveness of protec-
tive measures at stationary electrical systems in accordance with IEC 60364-6, EN 50110-1 and other country-specific standards. They can also be used for the testing of electric charging stations per EN 61851-1 (DIN VDE 0122-1), and for earth measurements. The test instruments include pre-programmed test sequences for increased working convenience and user-defined test sequences can also be programmed as an option.
The test instruments are especially well suited for testing electrical systems during setup, initial startup, periodic testing and trouble­shooting.
The applications range of the test instruments covers all alternat­ing and 3-phase current systems with nominal voltages of 230/ 400 V (300/500 V) and nominal frequencies of 16⅔, 50, 60, 200 and 400 Hz.
A system structure is set up in the test instrument and measured values are assigned to the objects. Completed tests and mea­sured values can be saved and documented in a measurement and test report.
Safety of the operator, as well as that of the test instrument, is only assured when it’s used for its intended purpose.

2.2 Use for Other than Intended Purpose

Using the test instrument for any purposes other than those described in these operating instructions, or in the test instru­ment’s condensed operating instructions, is contrary to use for intended purpose.

2.3 Liability and Guarantee

Gossen Metrawatt GmbH assumes no liability for property dam­age, personal injury or consequential damage resulting from improper or incorrect use of the product, in particular due to fail­ure to observe the product documentation. Furthermore, all guar­antee claims are rendered null and void in such cases.
Nor does Gossen Metrawatt GmbH accept any liability for data loss.

2.4 Opening the Instrument / Repairs

In order to ensure flawless, safe operation and to assure that the guarantee isn’t rendered null and void, the test instrument may only be opened by authorized, trained personnel. Even original replacement parts may only be installed by authorized, trained personnel.
Unauthorized modification of the test instrument is prohibited. If it can be ascertained that the test instrument has been opened
by unauthorized personnel, no guarantee claims can be honored by the manufacturer with regard to personal safety, measuring accuracy, compliance with applicable safety measures or any consequential damages.
If the guarantee seal is damaged or removed, all guarantee claims are rendered null and void.

2.5 Scope of Functions

PROFITEST … (Article Number)
PRO
M
(M520N)
TECH+
M
✓✓
✓✓
✓✓
(M520S)
MBASE+
Testing of Residual Current Devices (RCDs) UT measurement without tripping the RCD Tripping time measurement Measurement of tripping current I Selective, SRCDs, PRCDs, type G/R AC/DC sensitive RCDs, types B and B+ Testing of insulation monitoring devices (IMDs) ——— Testing of residual current monitoring devices (RCMs) ——— Testing for N-PE reversal
Measurement of Loop Impedance Z
Fuse table for systems without RCDs Without tripping the RCD, fuse table —— 15 mA measurement
Earthing resistance RE (mains operation)
I/U measuring method (2/3-wire measuring method via measuring adapter: 2-pole/2-pole + probe)
Earthing resistance RE
1)
(battery operation)
F
/ Z
L-PE
L-N
3 or 4-wire measuring method via PRO-RE
✓✓✓✓ ✓✓✓✓ ✓✓✓✓ ✓✓✓✓
——
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓
adapter Soil resistivity rE (battery operation)
(4-wire measuring method
Selective earthing resistance R
with 2-pole adapter, probe, earth electrode and current clamp sensor
Selective earthing resistance R
with probe, earth electrode and (4-wire measuring method
(3-wire measuring method)
current clamp sensor
via PRO-RE adapter)
E (mains operation)
E (battery operation)
current clamp sensor
via PRO-RE adapter and
)
✓✓✓✓
Earth loop resistance RELOOP (battery opera­tion)
with 2 clamps (current clamp sensor direct and current clamp transformer via PRO-RE/2 adapter)
Measurement of equipotential bonding R
Automatic polarity reversal
Insulation resistance R
Variable or rising test voltage (ramp)
Voltage U Special Measurements
IL, I Phase sequence Earth leakage resistance R Voltage drop (U) Standing-surface insulation Z Meter startup (kWh test) Leakage current with PRO-AB (IL) adapter Residual voltage test (Ures) Intelligent ramp (ta + I) Electric vehicles at charging stations (IEC 61851-1) Documentation of fault simulations at PRCDs with the PROFITEST PRCD adapter
Features Selectable user interface language Memory (database for up to 50,000 objects) Automatic test sequence function
/ U
L-N
current measurement with clamp
AMP
L-PE
/ U
INS
N-PE
/ f
E(INS)
ST
LO
2)
RS 232 port for RFID/barcode reader
USB port for data transmission ETC PC database and report generating software Measuring category: CAT III 600 V / CAT IV 300 V DAkkS calibration certificate
1)
The so-called live measurement is only advisable if there’s no bias current within the system. Only suitable for motor protection switches with small nominal current values.
2)
Currently available languages D, GB, I, F, E, P, NL, S, N, FIN, CZ, PL
✓✓✓✓
✓✓✓✓
✓✓✓✓
✓✓✓✓ ✓✓✓✓ ✓✓✓✓ ✓✓✓✓ ✓✓✓✓ ✓✓✓✓
——— ——— ——— ——
———
✓✓✓✓ ✓✓✓✓ ✓✓✓✓ ✓✓✓✓ ✓✓✓✓ ✓✓✓✓ ✓✓✓✓ ✓✓✓✓
XTRA
(M522R)
M
✓ ✓
✓ ✓ ✓
(M522P)
Gossen Metrawatt GmbH 5
Page 6

3 Documentation

Note
S
This documentation describes several test instrument. As a result, features and functions may be described which do
not apply to your instrument. Furthermore, illustrations may differ from your instrument.
These operating instructions describe a test instrument with software/firmware version 1.16.20.
List of Abbreviations and their Meanings
RCCBs (residual current circuit breakers / RCDs): I
I
N
I
F
PRCD Portable residual current device
RCD- Selective RCCB R
E
SRCD Socket residual current device (permanently installed) t
a
U
I
U
IN
U
L
Overcurrent protective devices: I
SC
Z
L-N
Z
L-PE
Earthing: R
B
R
E
R
ELoop
Low-value resistance at protective, earthing and bonding conductors:
R
LO+
R
LO–
Insulation: R
E(INS)
R
INS
R
ST
Z
ST
Current: I
A
I
L
I
M
I
N
I
P
Voltage: f Line voltage frequency
Tripping current Nominal residual current Rising test current (residual current)
PRCD-S: with protective conductor detection and monitoring PRCD-K: with undervoltage trigger and protective conductor monitoring
Calculated earthing or earth electrode loop resistance
Time to trip / breaking time Touch voltage at moment of tripping Touch voltage
relative to nominal residual current I
N
Touch voltage limit value
Calculated short-circuit current (at nominal voltage) Supply impedance Loop impedance
Operational earth resistance Measured earthing resistance Earth electrode loop resistance
Equipotential bonding conductor resistance (+ pole to PE)
Equipotential bonding conductor resistance (– pole to PE)
Earth leakage resistance (DIN 51953) Insulation resistance Standing surface insulation resistance Standing surface insulation impedance
Breaking current Leakage current (measured with current clamp trans-
former) Measuring current Nominal current Test c u rr e nt
f
N
Nominal voltage rated frequency
U Voltage drop as % U Voltage measured at the test probes during and after
insulation measurement R
U
Batt
U
E
U
INS
(Rechargeable) battery voltage Earth electrode voltage When measuring R
INS
: test voltage for ramp function:
INS
tripping or breakdown voltage U U U U U
L-L
L-N
L-P E N
3~
Voltage between two phase conductors
Voltage between L and N
Voltage between L and PE
Nominal line voltage
Highest measured voltage during determination of
phase sequence U U
S-PE
Y
Voltage between probe and PE
Phase-to-earth voltage

4 Getting Started

1. Read and adhere to the product documentation. In particular
observe all safety information in the documentation, on the instrument and on the packaging. See:
– section 1, “Safety Instructions”, on page 4 – section 2, “Applications”, on page 5 – section 3, “Documentation”, on page 6
2. Familiarize yourself with the test instrument.
See: – section 5, “The Instrument”, on page 7 – section 6, “Operating and Display Elements”, on page 16 – section 7, “Operation”, on page 26
3. Enter the basic settings.
See section 8, “Instrument Settings”, on page 27.
4. Optional but recommended: Create a database in the test
instrument. See section 9, “Database”, on page 31.
5. Read the basic information provided in section 10, “General
Information on Measurements”, on page 36.
6. Perform measurements.
Refer to individual measurements or test sequences (auto­matic sequences):
– section 11, “Measuring Voltage and Frequency”, on page
41 – section 12, “ – section 13, “Testing of Breaking Requirements for Overcur-
rent Protective Devices, Measurement of Loop Impedance
and Determination of Short-Circuit Current (ZL-PE and I
Functions)”, on page 54 – section 14, “Measuring Supply Impedance (Z
on page 57 – section 15, “Earthing Resistance Measurement (Function
R
)”, on page 59
E
– section 16, “
73 – section 17, “
(Protective Conductor and Equipotential Bonding Conductor)
on page 77 – section 18, “Measurement with Accessory Sensors”, on
page 80 – section 19, “Special Functions – EXTRA Switch Position”,
on page 81 – section 20, “Test Sequences (Automatic Test Sequences) –
AUTO Function”, on page 95
Further interesting information: section 21, “Maintenance”, on page 97.
Testi ng RC Ds
Measurement of Insulation Resistance
”, on page 43
Function)”,
L-N
”, o n pa ge
Measuring Low-Value Resistance of up to 200
SC
”,
6 Gossen Metrawatt GmbH
Page 7

5 The Instrument

!
XY123
2018-07
D-K
15080-01-01

5.1 Scope of Delivery

Standard scope of delivery for PROFITEST MASTER series: 1 Test instrument 1 Compact battery pack
(Z502H)
1 Earthing contact plug insert,
country-specific (PRO-SCHUKO / GTZ3228000R0001)
1 2-pole measuring adapter
and cable for expansion into a 3-pole adapter (PRO-A3-II / Z501O)
2 Alligator clips 1 Operating Instructions
1USB cable 1 Neck strap 1 ETC software*
1Charger
1 DAkkS calibration certificate
(this document)
(
Z502R)

5.3 Meanings of Symbols on the Instrument

Warning concerning a point of danger (attention, observe documentation!)
Protection category II device
Charging socket for extra-low direct voltage (for Z502R charger)
The device and its batteries may not be disposed of with household trash. Further information is included in the operating instructions.
Indicates EC conformity
If the guarantee seal is damaged or removed, all guar­antee claims are rendered null and void.
The special technical knowledge of qualified personnel is required for electrical installation or repair.
** Download from Internet

5.2 Optional Accessories (excerpt)

A complete overview of optional accessories including detailed information can be found in the data sheet for the test instrument.
The most important accessories are listed here:
• Barcode Profiscanner RS232 (Z502F) (barcode reader and scanner for RS 232 connection to the test instrument for identifying systems, electrical circuits and operating equipment.)
• PRO-HB (Z501V) Holder for test probes and measuring adapter
• Country-specific plug inserts – PRO-GB/USA (Z503B) – PRO-CH (GTZ3225000R0001)
• Plug inserts for PE and other similar measurements – PRO-RLO-II (Z501P)
(cable length: 10m)
– PRO-RLO 20 (Z505F)
(cable length: 20m)
– PRO-RLO 50 (Z505G)
(cable length: 50m)
• PRO-AB (Z502S) (leakage current measuring adapter for PROFITEST MXTRA)
• PROFITEST PRCD (M512R) (test adapter for testing portable safety switches (types PRCD-K and PRCD-S) with the help of the PROFITEST MXTRA)
• PROFITEST EMOBILITY (M513R) (adapter for standards-compliant testing of single and 3­phase, mode 2 and 3 charging cables with simulation of faults)
• E-SET BASIC (Z593A) (basic earth measurement accessories)
• E-SET PROFESSIONAL (Z592Z) (extensive earth measurement accessories)
CAT III 600 V
CAT IV 300 V
Calibration seal (blue seal): Consecutive number Deutsche Akkreditierungsstelle GmbH – calibration lab Registration number Date of calibration (year – month) Measuring category
Gossen Metrawatt GmbH 7
Page 8

5.4 Instrument Overview

456
8
91011
7
13
31
16 1715
14
2
19 20 21
22
12
!
RS 232
2
Connector Sockets for Current Clamp Sensor, Probe and PRO-AB Leakage Current Measuring Adapter
15
16
17
* See section 10.1 on page 36 regarding use of the
probes.
*
*
*
A
Test Instrument and Adapter
Test Instrument and Adapter:
1 Control panel with keys and display screen 2 Eyelets for attaching the neck strap 3 Rotary selector switch 4 Measuring adapter (2-pole) 5 Plug insert (country-specific) 6 Test plug (with retaining ring) 7 Alligator clip (plug-on) 8 Test probes 9 ON/START ▼ key * 10 I 11 Contact surfaces for finger contact 12 Test plug holder 13 Fuses 14 Holders for test probes (8)
8 Gossen Metrawatt GmbH
/compens./ZOFFSET key
N
Connections for Current Clamp, Probe, PRO-AB Leakage Current Measuring Adapter:
15 Current clamp connection 1 16 Current clamp connection 2 17 Probe socket
Interfaces, Charger Connection:
19 USB slave for connection to a PC 20 RS 232 port for connecting barcode or RFID reader 21 Socket for Z502R charger 22 Battery compartment lid (compartment for batteries and
spare fuses)
* Can only be switched on with the key on the instrument
Accessories:
A PRO-HB (Z501V) test probe and measuring adapter holder –
can be purchased separately
Page 9
(1) Control Panel – Display Panel
Attention!
!
Attention!
!
Attention!
!
See section 6.1, “Control Panel”, on page 16. See section 6.2, “Display”, on page 16.
(2) Eyelets for the Neck Strap
The included neck strap can be attached at the right and left hand sides of the instrument. You can hang the instrument from your neck and keep both hands free for measurement.
(3) Rotary Selector Switch
The following basic functions can be selected with the rotary switch:
SETUP / I EXTRA / AUTO
N
/ IF / Z
L-P E
/ Z
L-N
/ RE / R
LO
/ R
/ U / SENSOR /
INS
The various basic functions are selected by turning the function selector switch while the instrument is switched on.
(4) Measuring Adapter
The measuring adapter (2-pole) may only be used to­gether with the test instrument’s test plug. Use for other purposes is prohibited!
The plug-on measuring adapter (2-pole) with the two test probes is used for measurements in systems without earthing contact outlets, e.g. at permanent installations, distribution cabinets and all three-phase outlets, as well as for insulation resistance and low-value resistance measurements.
The 2-pole measuring adapter can be expanded to three poles for phase sequence testing with the included measurement cable (test probe).
This key has the same function as the
key on the test plug.
(10) IN / I Key (at the control panel)
The following sequences are triggered by pressing this key, either on the test plug or at the control panel:
• Starts the tripping test after measurement of touch voltage for RCCB testing (I
• Measurement of ROFFSET is started in the R
N
).
/ Z
function.
LO
L-N
• Semiautomatic polarity reversal (see section 10.9)
(11) Contact Surfaces
The contact surfaces are located at both sides of the test plug. When the contact plug is grasped in the hand, contact is auto­matically made with these surfaces. The contact surfaces are electrically isolated from the terminals and from the measuring cir­cuit.
In the event a potential difference of greater than 25 V between protective conductor terminal PE and the contact surface, PE is displayed. See “LED Indications, Mains Connections and Poten­tial Differences” on page 17.
(12) Test Plug Holder
The test plug with attached plug insert can be reliably secured to the instrument with the rubberized holder.
(13) Fuses
The two fuses protect the device in case of overload. Phase con­ductor L and neutral conductor N are fused individually. If a fuse is defective, and if an attempt is made to perform a measurement which uses the circuit protected by this fuse, a corresponding message appears at the display panel.
See section 21.2, “Fuse Replacement”, on page 97.
(5) Plug Insert (country-specific)
The plug insert may only be used together with the test instrument’s test plug. Use for other purposes is prohibited!
After the plug insert has been attached, the instrument can be directly connected to earthing contact outlets. There’s no need to concern yourself with poling at the plug. The instrument detects the positions of phase conductor L and neutral conductor N and automatically reverses polarity if necessary. The instrument automatically determines whether or not both pro­tective contacts in the earthing contact outlet are connected to one another, as well as to the system protective conductor, for all types of protective conductor measurements when the plug insert is attached to the test plug.
(6) Test Plug
The various country specific plug inserts (e.g. protective contact plug insert for Germany or SEV plug insert for Switzerland) or the measuring adapter (2-pole) are attached to the test plug and secured with a threaded connector.
The controls on the test plug are subject to interference suppres­sion filtering. This may lead to slightly delayed responses as opposed to controls located directly on the instrument.
(7) Alligator Clip (plug-on)
(8) Test Probes
The test probes comprise the second (permanently attached) and third (plug-on) poles of the measuring adapter. A coil cable con­nects them to the plug-on portion of the measuring adapter.
(14) Holders for Test Probes (8)
(15/16) Current Clamp Connector
Only the current clamp transformers offered as accessories may be connected to these sockets.
(17) Probe Connector Socket
The probe connector socket is required for the measurement of probe voltage U tance R
and standing surface insulation resistance.
E
, earth electrode voltage UE, earthing resis-
S-PE
It can be used for the measurement of touch voltage during RCD testing. The probe is connected with a 4 mm contact-protected plug.
The instrument determines whether or not the probe has been properly set and displays results at the display panel.
(19) USB Port
The USB port allows for the exchange of data between the test instrument and a PC.
(20) RS 232 Port
This port allows for data entry by means of a barcode scanner or an RFID reader.
(21) Charging Socket
Only the Z502R charger for charging batteries inside the test instrument may be connected to this socket.
(22) Battery Compartment Lid – Replacement Fuses
Before removing the lid is removed, the instrument must be disconnected from the measuring circuit at all poles!
(9) ON/Start
The measuring sequence for the func­tion selected in the menu is started by pressing this key, either on the test plug or at the control panel. Exception: If the instrument is
Key
The compartment under the lid accommodates the rechargeable battery pack (Z502H), or commercially available rechargeable bat­teries or regular batteries.
Two replacement fuses are also located under the battery com-
partment lid. switched off, it can only be switched on by pressing the key at the control panel.
Gossen Metrawatt GmbH 9
Page 10

5.5 Technical Data

BAT
Nominal Ranges of Use
Voltage U
Frequency f
N
N
Overall voltage range Overall frequency range 15.4 Hz 420 Hz Line voltage Sinusoidal Temperature range 0 C  +40C Battery voltage 8 V 12 V Supply impedance angle Corresponds to cos =1  0.95 Probe resistance < 50 k
120 V (108 V 132 V) 230 V (196 V 253 V) 400 V (340 V 440 V)
16⅔ Hz
(15.4 V 18 Hz) 50 Hz (49.5 V 50.5 Hz) 60 Hz (59.4 V 60.6 Hz) 200 Hz (190 V 210 Hz) 400 Hz (380 V 420 Hz)
65 V 550 V
Reference Conditions
Line voltage 230 V  0.1% Line frequency 50 Hz  0.1% Measured qty. frequency Measured qty. waveform
45 Hz 65 Hz Sine (deviation between effective and
rectified value  0.1%)
Supply impedance angle cos  =1 Probe resistance  10  Supply voltage 12 V  0.5 V Ambient temperature + 23 C  2 K Relative humidity 40% 60% Finger contact For testing potential difference
to ground potential
Standing surface insulation Purely ohmic
Power Supply
Batteries 8 each AA 1.5 V
We recommend exclusive use of the included rechargeable battery pack (2000 mAh, Z502H).
Number of measurements (standard setup with illumination) – For R
INS
– For R
LO
Battery test Symbolic display of rechargeable
Battery-saving circuit Display illumination can be switched off.
Safety shutdown If supply voltage is too low, the instru-
Recharging socket Inserted rechargeable batteries can be
Charging time Z502R charger: approx. 2 hours *
1 measurement – 25 s pause: approx. 1100 measurements
Auto polarity reversal / 1  (1 measuring cycle) – 25 s pause: approx. 1000 (Z502O) measurements
battery voltage
The test instrument is switched off automatically after the last key opera­tion. The user can select the desired on-time.
ment is switched off, or cannot be switched on.
recharged directly by connecting a charger to the recharging socket: Z502R charger
Overload Capacity
R
ISO
U
, U
L-P E
L-P E
LO
, Z
E
L-N
L-N
, R
F
RCD, R Z
R
Protection with fine-wire fuses FF 3.15 A 10 s,
1200 V continuous 600 V continuous 440 V continuous 550 V (Limits the number of measure-
ments and pause duration. If overload occurs, the instrument is switched off by means of a thermostatic switch.)
Electronic protection prevents switching on if interference voltage is present.
Fuses blow at > 5 A
Electrical Safety
Protection class II Nominal voltage 230/400 V (300/500 V) Test voltage 3.7 kV, 50 Hz Measuring category CAT III 600 V or CAT IV 300 V Pollution degree 2 Fuses
L and N terminals 1 G fuse-link ea.
FF 3.15/500G 6.3 × 32 mm
Electromagnetic Compatibility (EMC)
Product standard EN 61326-1
Interference emission
EN 55022 A
Interference immunity
EN 61000-4-2 Contact/atmos. – 4 kV/8 kV EN 61000-4-3 10 V/m EN 61000-4-4 Mains connection – 2 kV EN 61000-4-5 Mains connection – 1 kV EN 61000-4-6 Mains connection – 3 V EN 61000-4-11 0.5 periods / 100%
Test value F e a t ure
Class
Ambient Conditions
Accuracy 0 … + 40 C Operation –5 … + 50 C Storage –20 … + 60 C (without batteries) Relative humidity
Max. 75%, no condensation allowed
Elevation Max. 2000 m
Mechanical Design
Display Multiple display with dot matrix,
128 × 128 pixels Dimensions W × L × H = 260 × 330 × 90 mm Weight Approx. 2.7 kg with batteries Protection Housing: IP 40, test probe: IP 20
per EN 60529
Data Interfaces
Type USB for PC connection Type RS 232 for barcode and RFID readers
* Maximum charging time with fully depleted batteries.
A timer in the charger limits charging time to no more than 4 hours.
10 Gossen Metrawatt GmbH
Page 11

5.6 Characteristic Values for PROFITEST MTECH+ and PROFITEST MBASE+

Func-
tion
U
Measured
Quantity
U
L-P E
U
N-PE
f
U
3 AC
U
Probe
U
L-N
U
IN
Display Range
0 … 99.9 V 0.1 V
100 … 600 V 1 V ±(|2% rdg.|+1d) ±(|1% rdg.|+1d)
15.0 … 99.9 Hz 100 … 999 Hz
0V … 99.9V
100 V … 600 V
0 … 99.9 V
100 … 600 V
0 … 99.9 V
100 … 600 V
0 … 70.0 V 0.1 V 0.3 × I
10 … 999
1.00 k … 6.51 k 3 … 999
1 k … 2.17 k
R
E
1 … 651 1
0.3 … 99.9 100 … 217
0.2 … 9.9
8
,
8
10 … 130
0 m … 999 m
1.00 … 9.99
0 m… 999 m
1.00 … 9.99
10.0 … 29.9
0 to 9.9 A
10 … 999 A
1.00 … 9.99 kA
10.0 … 50.0 kA
I

IF (IN = 6 mA) 1.8 … 7.8 mA
IF (IN = 10 mA) 3.0 … 13.0 mA 3.0 … 13.0 mA 3.0 … 13.0 mA
I
(IN = 30 mA) 9.0 … 39.0 mA 9.0 … 39.0 mA 9.0 … 39.0 mA
I
F
F
(IN = 100 mA) 30 … 130 mA 1 mA 30 … 130 mA 30 … 130 mA
I
F
(IN = 300 mA) 90 … 390 mA 1 mA 90 … 390 mA 90 … 390 mA
I
F
(IN = 500 mA) 150 … 650 mA 1 mA 150 … 650 mA 150 … 650 mA
I
F
/ UL = 25 V 0 … 25.0 V
U
I
UI / UL = 50 V 0 … 50.0 V 0 … 50.0 V
(IN × 1) 0 … 1000 ms 1ms 6 … 500 mA 0 … 1000 ms
t
A
t
(IN × 2) 0 … 1000 ms 1ms
A
t
(IN × 5) 0 … 40 ms 1 ms
A
Z
()
L-P E
Z
L-N
Z
L-P E
+ DC
I
(Z
SC
Z
Z
L-P E
L-P E
ZL-PE
L-N
+ DC)
Reso­lution
0.1 Hz 1 Hz
0.1 V 1V
0.1 V 1V
0.1 V 1V
1
0.01 k 1
0.01 k
0.1 1
0.1 1
0.1 mA
0.1 V Same as I
1 m
0.01
0.1
0.1 A 1A
10 A
100 A
0.6 … 9.9 0.1 Display range only
Z
(15 mA)
L-P E
ISC (15 mA)
RE (with probe)
[R
(without probe)
E
values same as
Z
R
L-P E
E
RE DC+
U
RE
R
Sel
RE DC+ 80 … 999
Clamp
EXTRA
Z
10.0 … 99.9 100 … 999
100 … 999 mA
0.00 … 9.99 A
10.0 … 99.9 A
0 m … 999 m
1.00 … 9.99
10.0 … 99.9 100 … 999
]
1 k …9.99 k
0 m … 999 m
8
1.00 … 9.99
10.0 … 29.9
E
0  … 999 
E
0 … 253 V 1 V Calculated value
10 k … 199 k
200 k … 999 k
ST
1.00 M … 9.99 M
10.0 M … 30.0 M
 
0.1 1
1mA
0.01 A
0.1 A
1 m
0.01
0.1 1
0.01 k 1 m
0.01
0.1
1 m…
1
1 m…
1
1 k 1 k
0.01 M
0.1 M
Input
Impedance /
Measuring Range
Tes t Cu r rent
0.3 … 600 V
DC
15.4 … 420 Hz
5M
I
= 10 mA × 1.05
N
I
= 30 mA × 1.05
N
I
= 100 mA × 1.05
N
I
= 300 mA × 1.05
N
I
= 500 mA × 1.05
N
1.8 … 7.8 mA 1.8 … 7.8 mA
2 × 6 mA …
2 × 500 mA
5 × 6 mA …
5 × 300 mA
1.3 A AC …
3.7 A AC
0.5A DC,
1.25 A DC
15 mA AC
1.3 … 3.7 A AC
1.3 … 3.7 A AC
1.3 … 3.7 A AC 400 mA AC
40 mA AC
4 mA AC
1.3 … 3.7 A AC
0.5, 1.25 A DC
1.3 … 2.7 A AC
0.5 / 1.25 A DC
2.3 mA at 230 V
N
8
8
8
0.3 V … 600 V
1.0 V … 600 V
1.0 … 600 V
5 V … 70 V
Calculated value
from
R
= U
E
IN / IN
0 … 25.0 V
0 … 1000 ms
0 … 40 ms
0.15 … 0.49
0.50 … 0.99
1.00 … 9.99
0.25 … 0.99
1.00 … 9.99
120 (108 … 132) V 230 (196 … 253) V 400 (340 … 440) V 500 (450 … 550) V
10.0 … 99.9 100 … 999
Calculated value
depending on U
Z
L-P E
I
=
SC
/10 …1000
U
N
0.15 … 0.49
0.50 … 0.99
1.0 …9.99 10 …99.9
100 …999
1 k …9.99 k
0.25 … 0.99
1.00 … 9.99
0.25 … 300 
10 k … 199 k
200 k … 999 k
1.00 M … 9.99 M
10.0 M … 30.0 M
N
:
Nominal Val-
Measuring Un-
ues
1
UN = 120, 230,
400, 500 V
f
= 16.7, 50,
N
60, 200, 400 Hz
1
U 120 V, 230 V,
400 V
fN = 50 Hz,
60 Hz
= 25 V, 50 V
U
L
I
N
6 mA,
10 mA,
30 mA, 100 mA, 300 mA,
500 mA
U
= 120 V,
N
230, 400,
500 V
fN =16.7 Hz, 50 Hz,
60 Hz
UN = 120, 230 V
= 50, 60 Hz
f
N
UN = 120, 230 V
fN = 16.78, 50,
and
60 Hz
UN = 120, 230 V
UN = 400 V
fN = 50, 60 Hz
UN = 120, 230 V
= 50, 60 Hz
f
N
See RE±(|20% rdg.|+20d) ±(|15% rdg.|+20d)
4
UN = 120, 230 V
fN = 50, 60 Hz
= U
U
0
 
±(|2% rdg.|+5d) ±(|1% rdg.|+5d)
±(|0.2% rdg.|+1d) ±(|0.1% rdg.|+1d)
±(|3% rdg.|+5d) ±(|3% rdg.|+1d)
±(|2% rdg.|+5d) ±(|2% rdg.|+1d)
±(|3% rdg.|+5d) ±(|3% rdg.|+1d)
+|10% rdg.|+1d
=
N
2
=
±(|5% rdg.|+1d)
+|10% rdg.|+1d
2
±(|10% rdg.|+30d)
1
±(|10% rdg.|+30d)
±(|5% rdg.|+3d)
±(|18% rdg.|+30d) ±(|10% rdg.|+3d)
±(|10% rdg.|+10d)
±(|8% rdg.|+2d)
Value calculated from Z
±(|10% rdg.|+30d) ±(|10% rdg.|+30d)
±(|5% rdg.|+3d)
1
±(|10% rdg.|+3d) ±(|10% rdg.|+3d) ±(|10% rdg.|+3d)
±(|18% rdg.|+30d)
±(|10% rdg.|+3d)
±(|22% rdg.|+20d) ±(|15% rdg.|+20d)
±(|20% rdg.|+2d) ±(|10% rdg.|+3d)
L-N
±(|10% rdg.|+2d) ±(|5% rdg.|+3d)
certainty
4 ms 3 ms
Value calculated from Z
I
SC
= UN/Z
Uncertainty
±(|2% rdg.|+5d) ±(|2% rdg.|+1d)
±(|1% rdg.|+5d) ±(|1% rdg.|+1d)
±(|2% rdg.|+5d) ±(|2% rdg.|+1d)
+|1% rdg.|–1d …
+|9% rdg.|+1d
±(|3.5% rdg.|+2d)
+|1% rdg.|–1d …
+|9% rdg.|+1d
±(|5% rdg.|+30d) ±(|4% rdg.|+30d)
±(|3% rdg.|+3d)
±(|6% rdg.|+50d)
±(|4% rdg.|+3d)
±(|2% rdg.|+2d) ±(|1% rdg.|+1d)
L-P E
(15 mA)
L-P E
±(|5% rdg.|+30d) ±(|4% rdg.|+30d)
±(|3% rdg.|+3d) ±(|3% rdg.|+3d) ±(|3% rdg.|+3d) ±(|3% rdg.|+3d)
±(|6% rdg.|+50d)
±(|4% rdg.|+3d)
Intrinsic
L-P E
(15 mA):
Connections
Plug
2-Pole
3-Pole
Insert
1
Adapter
Adapter
Probe


option
ally
Z
L-P E


Clamp Meters
WZ12C Z3512A
MFLEX
P300
Gossen Metrawatt GmbH 11
Page 12
PROFITEST MTECH+ and PROFITEST MBASE+
Func-
tion
R
ISO
R
LO
SEN­SOR
6, 7
Measured
Quantity
R
, R
INS
U
R
LO
ROFFSET 0.00 … 9.99 0.01
I
L/Amp
Display Range
1 k … 999 k
1.00 M … 9.99 M
10.0 M … 49.9 M 1 k … 999 k
1.00 M … 9.99 M
10.0 M … 99.9 M 1 k … 999 k
E INS
1.00 M … 9.99 M
10.0 M … 99.9 M
100 M … 200 M
1 k… 999 k
1.00 M… 9.99 M
10.0 M… 99.9 M
100 M … 500 M
10 … 999 V
1.00 … 1.19 kV1V10 V
0.00 … 9.99
10.0 … 99.9 100 … 199
0.0 mA … 99.9 mA 0.1 mA 100 … 999 mA 1 mA
10.0 … 15.0 A 0.1 A
1.00 … 9.99 A 0.01 A
10.0 … 99.9 A 0.1 A
100 A … 150 A 1 A
0.0 … 99.9 mA 0.1 mA 100 … 999 mA 1 mA ±(|7% rdg.|+1d) ±(|5% rdg.|+1d)
0.00 A … 9.99 A 0.01 A 100 mV/A 0.05 A … 10 A
0.00 A … 9.99 A 0.01 A
10.0 … 99.9 A 0.1 A
0.00 A … 9.99 A 0.01 A
10.0 … 99.9 A 0.1 A 100 … 999 A 1 A
0.0 … 99.9 mA 0.1 mA
100 … 999 mA 1 mA
0.00 … 9.99 A
0.00 … 9.99 A 0.01 A
10.0 … 99.9 A 0.1 A
0.00 … 9.99 A 0.01 A
10.0 … 99.9 A 0.1 A ±(|5% rdg.|+2d) ±(|3% rdg.|+2d)
0.00 … 9.99 A 0.01 A
10.0 … 99.9 A 0.1 A ±(|5% rdg.|+7d) ±(|3% rdg.|+7d) 100 … 999 A 1 A ±(|5% rdg.|+2d) ±(|3% rdg.|+2d)
Reso-
lution
1 k
10 k
100 k
1 k
10 k
100 k
1 k
10 k
100 k
1M 1 k
10 k
100 k
1M
0.01
0.1 1
0.01 A
0.01 A
Test Curre nt
I
= 1.5 mA
SC
I 200 mA DC I < 260 mA DC
I 200 mA DC I < 260 mA DC
Tra ns form a-
3
tion ratio
1 V/A 5 A … 15 A
1 mV / A 5 … 150 A
1 V/A 5 … 1000 mA
10 mV/A 0.5 A … 100 A
1 mV / A 5 … 1000 A
1 V/A 30 … 1000 mA
100 mV/A 0.3 … 10 A
10 mV/A 3 … 100 A
10 mV/A 0.5 … 100 A
1 mV / A 5 … 1000 A
Measuring
Range
50 k … 999 k
1.00 M … 49.9 M
50 k … 999 k
1.00 M … 99.9 M
50 k … 999 k
1.00 M … 200 M
50 k … 999 k
1.00 M … 499 M
10 … 1.19 kV
0.10  … 5.99
6.00  … 99.9
0.10 … 5.99
6.00 … 99.9
Nominal
Values
UN = 50 V I
= 1 mA
N
UN = 100 V
= 1 mA
I
N
UN = 250 V
I
= 1 mA
N
UN = 325 V, U
= 500 V,
N
= 1000 V
U
N
= 1 mA
I
N
 
U0 = 4.5 V ±(|4% rdg.|+2d) ±(|2% rdg.|+2d)
fN = 50, 60 Hz
f
=
N
16.7, 50, 60, 200, 400 Hz
f
= 50 Hz,
N
60 Hz
=
f
N
DC, 16.7 Hz,
50 Hz, 60 Hz,
200 Hz
Measuring
Uncertainty
K range
±(|5% rdg.|+10d)
M
±(|5% rdg.|+1d)
±(|3% rdg.|+1d)
55
±(|13% rdg.|+5d) ±(|5% rdg.|+4d)
±(|13% rdg.|+1d) ±(|5% rdg.|+1d)1.00 … 9.99 A 0.01 A
±(|11% rdg.|+4d) ±(|4% rdg.|+3d)
±(|11% rdg.|+1d) ±(|4% rdg.|+1d)
±(|7% rdg.|+2d) ±(|5% rdg.|+2d)
±(|3.4% rdg.|+2d) ±(|3.1% rdg.|+2d) ±(|3.1% rdg.|+1d) ±(|3.1% rdg.|+1d) ±(|3.1% rdg.|+2d) ±(|3.1% rdg.|+1d)
±(|27% rdg.|+100d)
±(|27% rdg.|+11d) ±(|27% rdg.|+12d) ±(|27% rdg.|+11d)
±(|27% rdg.|+100d)
±(|27% rdg.|+11d)
±(|5% rdg.|+12d) ±(|3% rdg.|+12d)
±(|5% rdg.|+50d) ±(|3% rdg.|+50d)
Intrinsic
Uncertainty
k range
±(|3% rdg.|+10d)
M
±(|3% rdg.|+1d)
±(|1.5% rdg.|+1d)
±(|3% rdg.|+2d) ±(|3% rdg.|+2d) ±(|3% rdg.|+1d) ±(|3% rdg.|+1d) ±(|3% rdg.|+2d) ±(|3% rdg.|+1d)
±(|3%
rdg.|+100d)
±(|3% rdg.|+11d) ±(|3% rdg.|+12d) ±(|3% rdg.|+11d)
±(|3%
rdg.|+100d)
±(|3% rdg.|+11d)
Plug
2-Pole
1
Insert
Adapter

Connections
Clamps / Meas. Ranges
3-Pole
WZ12CZ3512AMFLEX
Adapter
I 15A
II
150 A
1A
10 A
100 A
1000
A
P300
3A
30 A
300 A
CP1100
100 A
1000 A
1
U > 230 V with 2 or 3-pole adapter only
2
1×IN> 300 mA and 2 × IN> 300 mA and 5 × IN > 500 mA and I
> 300 mA only up to UN 230 V!
f
5×I
> 300 mA only where UN = 230 V
N
3
The transformation ratio selected at the clamp (1, 10, 100, 1000 mV/A)
Key: d = digit(s), rdg. = reading (measured value)
must be set in the “Type” menu with the rotary switch in the “SENSOR” position.
4
Where R
5
The specified measuring and intrinsic uncertainties already include those of the respective current clamp.
6
Measuring range of the signal input at the test instrument, UE: 0 … 1.0 V
7
Input impedance of the signal input at the test instrument: 800 k
8
DC bias only possible with PROFITEST MTECH+
TRMS
(0 … 1.4 V
Eselective/REtotal
peak
< 100
) AC/DC
12 Gossen Metrawatt GmbH
Page 13

5.7 Characteristic Values for PROFITEST MXTRA and PROFITEST MPRO

Func-
Measured
tion
Quantity
U
L-P E
U
N-PE
f
U
U
I
N
I
F
Z
L-P E
Z
L-N
3 AC
U
Probe
U
L-N
U
IN
R
E
IF (IN = 6 mA) 1.8 … 7.8 mA
(IN = 10 mA) 3.0 … 13.0 mA 3.0 … 13.0 mA 3.0 … 13.0 mA
I
F
(IN = 30 mA) 9.0 … 39.0 mA 9.0 … 39.0 mA 9.0 … 39.0 mA
I
F
I
(IN = 100 mA) 30 … 130 mA 1 mA 30 … 130 mA 30 … 130 mA
F
(IN = 300 mA) 90 … 390 mA 1 mA 90 … 390 mA 90 … 390 mA
I
F
(IN = 500 mA) 150 … 650 mA 1 mA 150 … 650 mA 150 … 650 mA
I
F
/ UL = 25 V 0 … 25.0 V
U
I
/ UL = 50 V 0 … 50.0 V 0 … 50.0 V
I
(IN × 1) 0 … 1000 ms 1 ms 6 … 500 mA 0 … 1000 ms
t
A
t
(IN × 2) 0 … 1000 ms 1 ms
A
t
(IN × 5) 0 … 40 ms 1ms
A
Z
()
L-P E
Z
L-N
Z
L-P E
+ DC
I
(Z
SC
L-P E
ZL-PE +
DC
Z
(15 mA)
L-PE
ISC (15 mA)
Display Range
0 … 99.9 V 0.1 V
100 V … 600 V 1 V ±(|2% rdg.|+1d) ±(|1% rdg.|+1d)
15.0 … 99.9 Hz 100 … 999 Hz
0 … 99.9 V
100 … 600 V
0 … 99.9 V
100 … 600 V
0 … 99.9 V
100 … 600 V
0 … 70.0 V 0.1 V 0.3 × I
10 … 999
1.00 k … 6.51 k 3 … 999
1 k … 2.17 k
1 … 651 1
0.3 … 99.9 100 … 217
0.2 … 9.9 10 … 130
0 m… 999 m
1.00 … 9.99 1 m
0 m … 999 m
1.00 … 9.99
8
10.0 … 29.9
,
0 to 9.9 A
10 … 999 A
1.00 … 9.99 kA
8
10.0 … 50.0 kA
)
0.6 … 99.9 100 … 999
0.10 … 9.99 A
10.0 … 99.9 A
100 … 999 A
0 m … 999 m
1.00 … 9.99
10.0 … 99.9 100 … 999 1 k …9.99 k
0.5 … 99.9 100 … 999
0 m … 999 m
8
1.00 … 9.99
10.0 … 29.9
8
E
0 m … 999 m
1.00 … 9.99
10.0 … 99.9 100 … 999
0 m … 999 m
1.00 … 9.99
8
10.0 … 99.9 100 … 999
10 k … 199 k 1 k
200 k … 999 k
1.00 M … 9.99 M
10.0 M … 30.0 M
0 … 253 V 1 V
R
R
E
probe) +
R
E.sl
RE
(only with probe)
Sel
Clamp
(only with probe)
EXTRA
R
(without
E.sl
probe)
(with probe)
E
R
E (15 mA)
(without/with
probe)
(without
R
E.sl
DC
(with probe)
+ DC
U
R
E.sel
R
E.sel
+ DC
Z
ST
Reso-
lution
0.1 Hz 1 Hz
0.1 V 1V
0.1 V 1V
0.1 V 1V
1
0.01 k 1
0.01 k
0.1 1
0.1 1
0.1 mA
0.1 V Same as I
0.01
0.1
0.1 A 1A
10 A
100 A
0.1 1
0.01 A
0.1 A
11
1A
1 m
0.01
0.1 1
0.01 k
0.1 1
1 m
0.01
0.1
1 m
0.01
0.1 1
1 m
0.01
0.1 1
1 k
0.01 M
0.1 M
Input
Impedance /
Test Curre nt
Measuring
Range
0.3 … 600 V
DC
15.4 … 420 Hz
5M
0.3 … 600 V
1.0 V … 600 V
1.0 … 600 V
5 … 70 V
Calculated value
from
R
= U
E
IN / IN
I
= 10 mA × 1.05
N
I
= 30 mA × 1.05
N
I
= 100 mA × 1.05
N
I
= 300 mA × 1.05
N
I
= 500 mA × 1.05
N
N
1.8 … 7.8 mA 1.8 … 7.8 mA
0 … 25.0 V
2 × 6 mA …
2 × 500 mA
5 × 6
mA
5 × 300 mA
3.7 A AC …
4.7 A AC
3.7 … 4.7 A AC
0.5, 1.25 A DC
15 mA AC
3.7 … 4.7 A AC
3.7 … 4.7 A AC 400 mA AC
40 mA AC
4 mA AC
15 mA AC
3.7 … 4.7 A AC
0.5, 1.25 A DC
3.7 … 4.7 A AC
2.1 A AC
2.1 A AC
400 mA AC
40 mA AC
3.7 … 4.7 A AC
0.5, 1.25 A DC
2.3 mA at 230 V
8
8
8
0 … 1000 ms
0 … 40 ms
0.10 … 0.49
0.50 … 0.99
1.00 … 9.99
0.25 … 0.99
1.00 … 9.99
120 (108 … 132) V 230 (196 … 253) V 400 (340 … 440) V 500 (450 … 550) V
10.0 … 99.9 100 … 999
100 … 12 A
= 120 V)
(U
N
200 mA … 25 A
= 230 V)
(U
N
0.10 … 0.49
0.50 … 0.99
1.0 …9.99 10 …99.9
100 …999 1 k …9.99 k
10 …99.9
100 …999
0.25 … 0.99
1.00 … 9.99
RE = 0.10 …
9.99 W
0.25 … 300 
0.25 … 300 
R
< 10 
E.tot
10 k … 199 k
200 k … 999 k
1.00 M … 9.99 M
10.0 M … 30.0 M
1
1
4
4
 
Nominal
Valu es
UN = 120 V, 230 V, 400 V, 500 V,
f
= 16.7, 50,
N
60, 200, 400 Hz
=
U
N
120 V, 230 V,
2
400 V
fN = 50, 60 Hz
UL = 25, 50 V
I
=
N
6 mA, 10 mA, 30 mA,
100 mA, 300 mA,
2
500 mA
U
= 120, 230,
N
400, 500 V
fN =16.7 Hz, 50 Hz,
60 Hz
UN = 120, 230 V
= 50, 60 Hz
f
N
UN = 120, 230 V fN = 16.7, 50,
60 Hz
U
same as
N
function U
fN = 50, 60 Hz
UN = 120, 230 V
= 50, 60 Hz
f
N
UN = 120, 230 V
= 50, 60 Hz
f
N
UN = 120, 230 V fN = 50, 60 Hz
UN = 120, 230 V
= 50, 60 Hz
f
N
UN = 120, 230 V
= 50, 60 Hz
f
N
= U
U
0
L-N
Measuring
Uncertainty
Intrinsic
Uncertainty
±(|2% rdg.|+5d) ±(|1% rdg.|+5d)
±(|0.2% rdg.|+1d) ±(|0.1% rdg.|+1d)
±(|3% rdg.|+5d) ±(|3% rdg.|+1d) ±(|2% rdg.|+5d) ±(|2% rdg.|+1d) ±(|3% rdg.|+5d) ±(|3% rdg.|+1d)
+|10% rdg.|+1d
±(|2% rdg.|+5d) ±(|2% rdg.|+1d) ±(|1% rdg.|+5d) ±(|1% rdg.|+1d) ±(|2% rdg.|+5d) ±(|2% rdg.|+1d)
+|1% rdg.|–1d …
+|9% rdg.|+1d
±(|5% rdg.|+1d) ±(|3.5% rdg.|+2d)
+|10% rdg.|+1d
+|1% rdg.|–1d …
+|9% rdg.|+1dU
4 ms 3 ms
±(|10% rdg.|+20d)
1
±(|10% rdg.|+20d)
±(|5% rdg.|+3d)
±(|18% rdg.|+30d) ±(|10% rdg.|+3d)
±(|5% rdg.|+20d) ±(|4% rdg.|+20d)
±(|3% rdg.|+3d)
±(|6% rdg.|+50d)
±(|4% rdg.|+3d)
Value calculated from Z
±(|10% rdg.|+10d)
±(|8% rdg.|+2d)
±(|10% rdg.|+20d) ±(|10% rdg.|+20d)
±(|5% rdg.|+3d)
1
±(|10% rdg.|+3d) ±(|10% rdg.|+3d) ±(|10% rdg.|+3d)
±(|10% rdg.|+10d)
±(|8% rdg.|+2d)
±(|18% rdg.|+30d)
±(|10% rdg.|+3d)
Calculated U
Value calculated from
I
SC
= UN/Z
±(|2% rdg.|+2d) ±(|1% rdg.|+1d)
(15 mA)
L-PE
±(|5% rdg.|+20d) ±(|4% rdg.|+20d)
±(|3% rdg.|+3d) ±(|3% rdg.|+3d) ±(|3% rdg.|+3d) ±(|3% rdg.|+3d)
±(|2% rdg.|+2d) ±(|1% rdg.|+1d)
±(|6% rdg.|+50d)
±(|4% rdg.|+3d)
= UN × RE/R
E
±(|20% rdg.|+20d) ±(|15% rdg.|+20d)
±(|22% rdg.|+20d) ±(|15% rdg.|+20d)
±(|20% rdg.|+2d) ±(|10% rdg.|+3d)
±(|10% rdg.|+2d) ±(|5% rdg.|+3d)
L-PE
Plug
2-Pole
1
Insert
Adapter



Z

E.sl

L-P E
Connections
3-Pole
Adapter
Probe
option-
ally
Clamp Meters
WZ12C Z3512A
MFLEX
P300
Gossen Metrawatt GmbH 13
Page 14
Characteristic Values for PROFITEST MXTRA and PROFITEST MPRO
Func-
Measured
tion
Quantity
IMD test
EXTRA
R
, R
INS
R
ISO
R
SEN-
SOR
6, 7
R
LO
ROFFSET 0.00 … 9.99 0.01
I
L/Amp
1.00M … 9.99 M
10.0M … 49.9 M
1.00 M … 9.99 M
10.0 M … 99.9 M
E INS
1.00 M … 9.99 M
10.0 M … 99.9 M 100 M … 200 M
U
LO
Display Range
20 k… 648 k
2.51 M
1k … 999 k
1 k … 999 k
1 k … 999 k
1 … 999 k
1.00 … 9.99 M
10.0 … 99.9 M 100 … 500 M
10 … 999 V DC
1.00 … 1.19 kV1V10 V
0.00 … 9.99
10.0 … 99.9 100 … 199
0.0 … 99.9 mA 0.1 mA 100 … 999 mA 1 mA
10.0 … 15.0 A 0.1 A
1.00 … 9.99 A 0.01 A
10.0 … 99.9 A 0.1 A 100 … 150 A 1 A
0.0 … 99.9 mA 0.1 mA 100 … 999 mA 1 mA ±(|7% rdg.|+1d) ±(|5% rdg.|+1d)
0.00 … 9.99 A 0.01 A 100 mV/A 0.05 … 10 A ±(|3.4% rdg.|+2d) ±(|3% rdg.|+2d)
0.00 … 9.99 A 0.01 A
10.0 … 99.9 A 0.1 A ±(|3.1% rdg.|+1d) ±(|3% rdg.|+1d)
0.00 … 9.99 A 0.01 A
10.0 … 99.9 A 0.1 A ±(|3.1% rdg.|+2d) ±(|3% rdg.|+2d) 100 … 999 A 1 A ±(|3.1% rdg.|+1d) ±(|3% rdg.|+1d)
0.0 … 99.9 mA 0.1 mA 100 … 999 mA 1 mA
0.00 … 9.99 A
0.00 … 9.99 A 0.01 A
10.0 … 99.9 A 0.1 A
0.00 … 9.99 A 0.01 A
10.0 … 99.9 A 0.1 A ±(|5% rdg.|+2d)
0.00 … 9.99 A 0.01 A
10.0 … 99.9 A 0.1 A ±(|5% rdg.|+7d) 100 … 999 A 1 A ±(|5% rdg.|+2d)
Reso-
Test Current Measuring Range
lution
IT line voltage
1 k
UN = 90 …
0.01 M
1 k
10 k
100 k
1 k
10 k
100 k
1 k
10 k
100 k
1M
1 k
10 k
100 k
1M
0.01
0.1 1
550 V
I
= 1.5 mA
SC
I 200 mA DC I < 260 mA DC
I 200 mA DC I < 260 mA DC
Tra ns fo rma -
tion
ratio 3
1 V/A 5 … 15 A
1 mV / A 5 … 150 A
1 V/A 5 … 1000 mA
10 mV/A 0.5 … 100 A
1 mV / A 5 … 1000 A
1 V/A
0.01 A
0.01 A
100 mV/A 0.3 … 10 A
10 mV/A 3 … 100 A
10 mV/A 0.5 … 100 A
1 mV / A 5 … 1000 A
20 k… 199 k 200 k … 648 k
2.51 M
50 k … 999 k
1.00 M … 49.9 M
50 k … 999 k
1.00 M … 99.9 M
50 k … 999 k
1.00 M … 200 M
50 k … 999 k
1.00 M … 499 M
10 … 1.19 kV ±(|3% rdg.|+1d)
0.10 … 5.99
6.00 … 99.9
0.10 … 5.99
6.00 … 99.9
30 …
1000 mA
Nominal
Values
IT system nomi-
nal voltages
U
=
N
120 V, 230 V,
400 V, 500 V fN = 50, 60 Hz
U
= 50 V
N
= 1 mA
I
N
UN = 100 V
= 1 mA
I
N
UN = 250 V
I
= 1 mA
N
UN = 325 V
= 500 V
U
N
= 1000 V
U
N
I
= 1 mA
N
= 4.5 V ±(|4% rdg.|+2d) ±(|2% rdg.|+2d)
U
0
fN = 50, 60 Hz
=
f
N
16.7, 50, 60, 200, 400 Hz
fN = 50, 60 Hz
f
=
N
DC, 16.7 Hz,
50 Hz, 60 Hz,
200 Hz
Measuring
Uncertainty
±7%
±12%
±3%
K range
±(|5% rdg.|+10d)
M
±(|5% rdg.|+1d)
55
±(|13% rdg.|+5d) ±(|5% rdg.|+4d)
±(|13% rdg.|+1d) ±(|5% rdg.|+1d)1.00 … 9.99 A 0.01 A
±(|11% rdg.|+4d) ±(|4% rdg.|+3d)
±(|11% rdg.|+1d) ±(|4% rdg.|+1d)
±(|7% rdg.|+2d) ±(|5% rdg.|+2d)
±(|3.1% rdg.|+2d) ±(|3% rdg.|+2d)
±(|3.1% rdg.|+1d) ±(|3% rdg.|+1d)
±(|27% rdg.|+100d) ±(|3% rdg.|+100d)
±(|27% rdg.|+11d) ±(|27% rdg.|+12d) ±(|27% rdg.|+11d)
±(|27% rdg.|+100d) ±(|3% rdg.|+100d)
±(|27% rdg.|+11d)
±(|5% rdg.|+12d)
±(|5% rdg.|+50d) ±(|3% rdg.|+50d)
Intrinsic
Uncertainty
±5%
±10%
±2%
k range
±(|3% rdg.|+10d)
M
±(|3% rdg.|+1d)
±(|1.5% rdg.|+1d)
±(|3% rdg.|+11d) ±(|3% rdg.|+12d) ±(|3% rdg.|+11d)
±(|3% rdg.|+11d)
±(|3% rdg.|+12d)
±(|3% rdg.|+2d)
±(|3% rdg.|+7d) ±(|3% rdg.|+2d)
Connections
Plug
Insert
1
2-Pole
Adapter
Clamps / Meas. Ranges
3-Pole
Adapter
WZ12C Z3512A


I 15A
II
150 A
1A
10 A
100 A
1000
A
MFLEX
P300
3A
30 A
300 A
CP1100
100 A
1000 A
1
U > 230 V, with 2 or 3-pole adapter only
2
1
×
IN> 300 mA and 2
I
> 300 mA only up to UN 230 V!
f
3
The transformation ratio selected at the clamp (1/10/100/1000 mV/A)
×
IN> 300 mA and 5
×
IN > 500 mA and
Key: d = digit(s), rdg. = reading (measured value)
must be set in the “Type” menu with the rotary switch in the “SENSOR” position.
4
Where R
5
The specified measuring and intrinsic uncertainties already include those of the respective current clamp.
6
Measuring range of the signal input at the test instrument, UE: 0 … 1.0 V
7
Input impedance
8
DC bias only possible with PROFITEST MXTRA
11
Where Z
TRMS
(0 … 1.4 V
Eselective/REtotal
< 0.6 , ISC > UN/0.5 is displayed
L-PE
< 100
) AC/DC
peak
of the signal input at the test instrument
: 800 k
14 Gossen Metrawatt GmbH
Page 15
Characteristic Values, Special Measurements with PROFITEST MPRO and PROFITEST MXTRA
Func-
RE
1 2
3
4 5 6 7 8 9 10 11
Measured
tion
Quantity
RE, 3-pole
RE, 4-pole ±(|10% rdg.|+10d) ±(|3% rdg.|+5d)
RE, 4-pole
selective
with clamp meter
BAT
Soil resistivity
(p)
Probe clearance d (p) 0.1 … 999 m
RE, 2 clamps
10.0 k … 19.9 k
10.0 k… 49.9 k
1.00 m … 9.99 km
Display Range
0.00 … 9.99
10.0 … 99.9 100 … 999
1.00 k … 9.99 k
10.0 k … 50.0 k
0.00 … 9.99
10.0 … 99.9 100 … 999
1.00 k … 9.99 k
0.0 m … 9.9 m
100 m … 999 m
0.00 … 9.99
10.0 … 99.9 100 … 999
1.00 … 1.99 k
10 11
Reso-
lution
0.01
0.1 1
0.01 k
0.1 k
0.01
0.1 1
0.01 k
0.1 k
0.1 k
0.1 m 1 m
0.01
kWm
0.01
0.1
1
0.01 k
Current /
Signal Freq.
1
16 mA/128 Hz
1.6 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz 16 mA/128 Hz
16 mA/128 Hz
1.6 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz 16 mA/128 Hz
1.6 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
0.16 mA/128 Hz
30 V / 128 Hz
100 m … 9.99 km 500 m … 9.99 km
5.00 km … 9.99 km
5.00 km … 9.99 km
5.00 km … 9.99 km
Signal frequency without interference signal PRO-RE (Z501S) adapter cable for test plug, for connecting earth probes (E-Set 3/4) PRO-RE/2 adapter cable for test plug, for connecting the E-CLIP2 generator clamp Generator clamp: E-CLIP2 (Z591B) Clamp meter: Z3512A (Z225A) Where R Where R
< 10 or clamp meter current > 500 µA
E.sel/RE E.H/RE
100 and R
E.E/RE
100
Where d = 20 m Where d = 2 m Only where RANGE = 20 k Only where RANGE = 50 k or AUTO
Te st
Measuring Range
1.00 … 19.9
5.0 … 199
50 … 1.99 k
0.50 k … 19.9 k
0.50 k … 49.9 k
1.00 … 9.99
10.0 … 200
0.10 … 9.99
10.0 … 99.9
Key: d = digit(s), rdg. = reading (measured value)
Measuring
Uncertainty
±(|10% rdg.|+10d
+ 1 
±(|15% rdg.|+10d) ±(|20% rdg.|+10d)
6
8 8
±(|20% rdg.|+10d)7±(|12% rdg.|+10d)
9 9 9
±(|10% rdg.|+5d) ±(|20% rdg.|+5d)
±(|3% rdg.|+5d
±(|10% rdg.|+10d) ±(|15% rdg.|+10d)
±(|5% rdg.|+5d)
±(|12% rdg.|+5d)
Intrinsic
Uncertainty
+ 0.5 
7
Connections
Adapter for Test Plug
Current Clamps
PRO-RE PRO-RE/2 Z3512A Z591B
2
25
2
354
Gossen Metrawatt GmbH 15
Page 16

6 Operating and Display Elements

Attention!
!
MEM: Key for memory functions
HELP:
Access context sensitive help
IN: Tripping test
Next step (semiautomatic measurement)
Start offset measurement
ON/START ▼:
Switch on
Start/stop measurement
ESC
: Return from submenu
SoftkeysFixed Function Keys
• Parameter selection
• Limit value specification
• Entry functions
LEDs and Connection Symbols 
section 6.3
9
10
Battery Display
Measuring Function
Measurement in progress/
Memory Occupancy
Measured
Parameters
PE
Save Value
Battery full
Battery OK
Battery weak
Battery (almost) dead
Battery Display ( section 7.1)
BAT BAT
BAT BAT
Memory Occupancy Display
MEM
Memory half full
MEM
Memory full > transfer data to PC
Connection Test – Mains Connection Test ( section 6.4)
NPEL
NPEL
)(
Connection OK L and N reversed
NPEL NPEL
x
NPEL NPEL
x
x
RUN READY
Connection Test section 6.4
U < 8 V
LPEN
x
LPEN
Quantities
stopped

6.1 Control Panel

The display and control panel can be swiveled forward or backward with the detented swivel hinge. The instrument can thus be set to the optimum reading angle.

6.2 Display

The following appear at the display:
• One or two measurement values as three place numeric dis­play with unit of measure and abbreviated measured quantity
• Nominal values for voltage and frequency
• Circuit diagrams
• On-line help
• Messages and instructions

6.3 LEDs

MAINS/NETZ LED
This LED is only functional when the instrument is switched on. It has no function in the voltage ranges U It lights up green, red or orange, or blinks green or red depending upon how the instrument has been connected and the selected function (see also section 6.4, “LED Indications, Mains Connec­tions and Potential Differences”, beginning on page 17). This LED also lights up if line voltage is present when measuring R
and RLO.
ISO
UL/RL LED
This LED lights up red if touch voltage is greater than 25 V or 50 V during RCD testing, as well as after safety shutdown occurs. It also lights up if R fallen short of.
RCD • FI LED
This LED lights up red if the RCCB is not tripped within 400 ms (1000 ms for selective RCDs – type RCD S) during the tripping test with nominal residual current. It also lights up if the RCCB is not tripped before nominal residual current has been reached during measurement with rising residual current.
or RLO limit values have been exceeded or
ISO
L-N
and U
L-P E
.
16 Gossen Metrawatt GmbH
The mains connection test may not be used to test sys­tems or system components for the absence of voltage!
Page 17

6.4 LED Indications, Mains Connections and Potential Differences

Attention!
!
???
NPEL
NPEL
NPEL
NPELxNPELxNPEL
x
NPEL
NPEL
x NPEL
LED Signals
Status
Tes t Plug
Mea­suring
Function Switch Position Function/Meaning
Adapter
NETZ/
MAINS
NETZ/
MAINS
NETZ/
MAINS
NETZ/
MAINS
NETZ/
MAINS
U
L/RL
FI/RCD
Lights up
green
Blinks
green
Blinks red
Lights up
red
Blinks yel-
low
Lights up
red
Lights up
red
X
X
XX
U, Z
U, Z
U, Z
X
X
XX
Z
L-N
XX
IN / IF , Z
, kWh, IMD, int. ramp, RCM
ST
IN / IF , Z
, kWh, IMD, int. ramp, RCM
ST
IN / IF , Z
, kWh, IMD, int. ramp, RCM
ST
RLO, R
IN / IF , Z
R
, Z
Z
L-N
L-PE
RE, Z
, Z
,
ZST, IMD, kWh, RCM, PRCD,
L-PE
/ Z
L-N
/ Z
L-N
/ Z
L-N
, RE, IL, sensor
ISO
/ Z
L-N
, RLO, RE,
INS
, U, IL, U
, IF, IN,
L-PE
t
+I, RCM
a
e-mobility
IN / IF,
int. ramp
L-PE
L-PE
L-PE
L-PE
, sensor
res
/ RE,
/ RE,
/ RE,
/ R
Correct connection, measurement enabled
N conductor not connected, measurement enabled
1) No line voltage or
2) PE interrupted Interference voltage is present at the test probes.
Measurement is disabled.
E
L and N are connected to the phase conductors.
The selected limit value has been violated.
Interference voltage limit value U
has been exceeded.
L
Safety shutdown has occurred. The test has been manually assessed as “NOT OK”.
The RCCB was not tripped, or was tripped too late during the tripping test.
Mains Connection Test — Single-Phase System — LCD Connection Pictographs
The mains connection test may not be used to test systems or system components for the absence of voltage!
Status
Tes t Plug
Mea­suring
Function Switch Position
Function/Meaning
Adapter
Is dis-
played
Is dis-
played
Is dis-
played
Is dis-
played
Is dis-
played
Is dis-
played
All except for U
All except for U
All except for U
All except U and RE
RE
All except for U
All except for U
No connection detected
Connection OK
L and N reversed, neutral conductor charged with phase voltage
No mains connection
Standard display without connection messages
Neutral conductor interrupted
Protective conductor PE interrupted, neutral conductor N and/or phase conductor L charged with phase volt­age
Is dis-
played
Is dis-
played
Is dis-
played
Is dis-
played
Gossen Metrawatt GmbH 17
All except for U
All except for U
All except for U
All except for U
Phase conductor L interrupted, neutral conductor N charged with phase voltage
Phase conductor L and protective conductor PE reversed
Phase conductor L and protective conductor PE reversed Neutral conductor interrupted (with probe only)
L and N are connected to the phase conductors.
Page 18
Mains Connection Test — 3-Phase System — LCD Connection Pictographs
Attention!
!
The mains connection test may not be used to test systems or system components for the absence of voltage!
Status
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
Test Plug Measuring
Adapter
Function Switch Position Function/Meaning
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
Clockwise rotation
Counter-clockwise rotation
Short between L1 and L2
Short between L1 and L3
Short between L2 and L3
Conductor L1 missing
Conductor L2 missing
Conductor L3 missing
Is displayed
Is displayed
Is displayed
U
(3-phase measurement)
U
(3-phase measurement)
U
(3-phase measurement)
Conductor L1 to N
Conductor L2 to N
Conductor L3 to N
18 Gossen Metrawatt GmbH
Page 19
Connection Test — Battery Powered Earthing Resistance Measurements, “Battery Mode”
NPEL
PE
PE
Status
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
Is displayed
Tes t Plug
Mea­suring Adapter
PRO-RE
Clamp
meter
PRO-RE
PRO-RE
PRO-RE
Function Switch Position
R
E
R
E
R
E
R
E
R
E
R
E
Function/Meaning
Standard display without connection messages
Interference voltage at probe S > 3 V Restricted measuring accuracy
Interference/measuring current ratio > 50 at R
, 1000 at R
E(sel)
Restricted measuring accuracy at R
Interference current > 0 85 A or interference/measuring current ratio > 100
E(sel):
No measured value, display: RE.Z – – –
Probe H not connected or R
>150k
E.H
No measurement, display: RE – – –
R
> 50 k or
E.H
R
E.H/RE
> 10000
Measured value is displayed, restricted measuring accuracy
Probe S not connected or R
> 150 k
E.S
or R
× R
E.S
> 25 M²
E.H
No measurement, display: RE – – –
R
> 50 k or
E.S
R
E.S/RE
> 300
Measured value is displayed, restricted measuring accuracy
Probe E not connected or R
> 150 kR
E.E
E.E/RE
> 2000
No measurement, display: RE – – –
R
E.E/RE
>300
Measured value is displayed, restricted measuring accuracy
E(2Z)
PE Test via Finger Contact at the Contact Surfaces on the Test Plug
Status
Tes t Plug
Mea­suring
Function Switch Position
Function/Meaning
Adapter
LCD LEDs
Is displayed
Is displayed
UL/RL
FI/RCD
light up
red
U
L/RL
FI/RCD
light up
red
XX
XX
U
(single-phase
measurement)
U
(single-phase
measurement)
Potential difference  50 V between finger contact and PE (earth contact) Frequency f 50 Hz
If L is correctly contacted and PE is interrupted (frequency f 50 Hz)
Status bar: Display of Charge Level, Memory Occupancy
Status
Tes t Plug
Mea­suring
Function Switch Position
Function/Meaning
Adapter
Battery
status
Is dis­played
Is dis­played
Is dis­played
Is dis­played
Is dis­played
U,
R
INS
R
, RE,
LO
Z
, Z
L-N
I
, I
F
Setup,
EXTRA,
SENSOR
,
L-PE
N
Battery charge level ≥ 80%
Battery charge level ≥ 50%
,
,
Battery charge level ≥ 30%
Battery charge level ≥ 15%
Battery charge level ≥ 0%
Gossen Metrawatt GmbH 19
Page 20
Battery
test
Is dis­played
Memory Status
Is dis-
Is dis-
Is dis-
Is dis-
Is dis-
Is dis-
played
played
played
played Is dis-
played Is dis-
played
played Is dis-
played
played
Z
Error Messages — LCD Connection Pictographs
All
U,
R
,
INS
R
, RE,
LO
, Z
L-N
L-PE
I
, I
F
N
Setup,
EXTRA,
SENSOR
Rechargeable batteries must be recharged or replaced towards the end of their service life (U < 8 V).
Memory occupancy ≥ 100%
Memory occupancy ≥ 87.5%
Memory occupancy ≥ 75%
Memory occupancy ≥ 62.5%
,
,
Memory occupancy ≥ 50%
Memory occupancy ≥ 37.5%
Memory occupancy ≥ 25%
Memory occupancy ≥ 12.5%
Memory occupancy ≥ 0%
Status
Te st Plug
Mea­suring
Function Switch Position
Adapter
All measurements
XX
XX
XX I
XX Z
XX I
with protective
conductor
I
Z
/ Z
L-N
/ IF
N
L-PE
N
L-PE
/ I
N
Function/Meaning
Potential difference  UL between finger contact and PE (earth contact) (frequency f 50 Hz) Remedy: Check PE connection Note: Only if is displayed: Measurement can nevertheless be
started by pressing the start key again.
1) Voltage too high (U > 253 V) for RCD test with direct current
2) U always U > 550 V with 500 mA
3) U > 440 V for I
/ R
E
4) U > 253 V for I
5) U > 253 V for measurement with probe
RCD is tripped too early or is defective. Remedy: Test circuit for bias current
RCD is tripped too early or is defective. Remedy: Test with “DC + positive half-wave”.
RCD tripped during touch voltage measurement.
F
Remedy: Check selected nominal test current.
/ IF
N
/ IF with 500 mA
N
R
IF , I∆N,
LO,
EXTRA
ta+I∆
The PRCD has been tripped. Reason: Poor contact or defective PRCD
Externally accessible fuse is blown. The voltage ranges remain functional even if fuses have blown.
X X All except for U
Special case, R
blown fuse.
: Interference voltage during measurement may result in a
LO
Remedy: Replace fuse as described in section 21.2.
20 Gossen Metrawatt GmbH
Page 21
XX
Z
L-N
I
/ Z
N
/ IF
L-PE
/ R
Frequency out of permissible range Remedy: Check the mains connection.
E
All
XX
PRO-RE
X
PRO-RE
PRO-RE/
2
XX
All measurements with
XX R
/ RLO
RINS
RE (bat)
RE (bat)
RE (bat)
probe
ISO
Excessive temperature inside the test instrument Remedy: Wait for test instrument to cool down
Interference voltage Remedy: Device under test must be disconnected from all sources of
voltage Interference voltage > 20 V at the probes: H to E or S to E No measurement possible
Probe ES not connected or connected incorrectly
Generator current clamp (E-Clip-2) not connected
Interference voltage at the probe
Overvoltage or overloading of the measuring voltage generator during measurement of R
INS
Remedy: Ensure absence of voltage at the device under test.
I
/ IF
N
Z
/ Z
L-N
XX
L-PE
ZST, RST, R
E
Meter startup
XX All
XX R
XR
XEXTRA
XEXTRA
LO
LO
 
 
U
U
No mains connection Remedy: Check the mains connection.
Defective hardware Remedy:
1) Switch on/off or
2) Briefly remove the batteries. If error message persists, send instrument to GMC-I Service GmbH.
OFFSET measurement is not sensible. Remedy: Check system. OFFSET measurement of RLO+ and RLO– is still possible.
R
OFFSET
> 9.99 :
OFFSET measurement is not sensible. Remedy: Check system.
Z > 9.99 : OFFSET measurement is not sensible. Remedy: Check system.
U
OFFSET
> U:
OFFSET value is greater than the measured value at the consuming sys­tem.
OFFSET measurement is not sensible. Remedy: Check system.
Contact problem or blown fuse
XXR
/ RLO / R
ISO
Remedy: Check test plug or measuring adapter for correct seating in the
E(bat)
test plug, or replace the fuse.
Gossen Metrawatt GmbH 21
Page 22
XR
IN/I
F
10 mA 30 mA 100 mA 300 mA 500 mA
R
MAX
at I
N
510 170 50 15 9
R
MAX
for I
F
410 140 40 12 7
E
The polarity of the 2-pole adapter must be reversed.
XI
N
/ IF
N and PE are reversed.
1) Mains connection error Remedy: Check the mains connection. or
2) Display at the connection pictograph: PE interrupted (x) or
XX
Z
L-N
/ Z
N
/ IF
L-PE
/ R
E
bottom protective conductor bar interrupted with reference to the keys at the test plug Cause: Voltage measuring path interrupted
I
Result: Measurement is disabled
Note: Only if is displayed: Measurement can nevertheless be started by pressing the start key again.
Display at the connection pictograph: Top protective conductor bar interrupted with reference to the keys at the
XI
N
/ IF
test plug Cause: Current measuring path interrupted Result: No measured value display
I
N
RE
/ IF
Probe is not detected, probe not connected Remedy: Check probe connection.
Clamp is not detected: – Clamp is not connected or
– Current through clamp is too small (partial earthing resistance too high)
RE
or – Transformation ratio set incorrectly
Remedy: Check clamp connection and transformation ratio.
Check the batteries in the METRAFLEX P300 and replace if necessary.
I
Z
RE
RE
All
/ IF
N
L-PE
, R
If you have changed the transformation ratio at the test instrument, a message appears prompting you to change the setting at the current clamp sensor as well.
Voltage too high at clamp input or signal distorted The transformation ratio parameter selected at the test instrument might not
correspond to the transformation ratio at the current clamp sensor. Remedy: Check transformation ratio or test setup.
Battery voltage is less than or equal to 8 V. Reliable measurement is no longer possible. Storage of measured values to memory is disabled. Remedy: Rechargeable batteries must be recharged or replaced
towards the end of their service life. Resistance in N-PE path is too high.
Consequence: The required test current cannot be generated and mea­surement is aborted.
If specified touch voltage UL is exceeded:
and RE: User is prompted to switch to the 15 mA wave.
Z
E
L-PE
RE alternative only: User is prompted to reduce the measuring range (reduce current.)
22 Gossen Metrawatt GmbH
Page 23
Entry Plausibility Check – Parameters Combination Checking — LCD Pictographs
Status
Tes t Plug
Mea­suring Adapter
Function Switch Position
I
N
I
/ IF
N
EXTRA

ta + I∆
>I
N
I
/ IF
N
Function/Meaning
Parameter out of range
5 × 500 mA is not possible
Types B, B+ and EV/MI not possible with G/R, SRCD, PRCD
180° not possible for RCD-S, G/R, SRCD, PRCD-S, PRCD-K
DC not possible with G/R, SRCD, PRCD
I
/ IF
N
I
/ IF
N
EXTRA
EXTRA
I
I
R
R
N
N
E
E
RCM
RCM
Half-wave or DC not possible with type AC
DC not possible with type A, F
½ test current not possible with DC
2 × IN / 5 × IN with full-wave only
DC+ with 10 Ω only
No DC bias in the IT network
R
E
EXTRA  RCM
I
/ IF
N
EXTRA
RCM
Gossen Metrawatt GmbH 23
15 mA only possible in 1 k and 100 ranges!
With RCM: Types AC, F, B+ and EV/MI are not possible.
Measurement with half-wave or DC is not possible in IT systems.
Page 24
All
The parameters you have selected do not make sense in combination with previously configured parameters. The selected parameter settings will not be saved.
Remedy: Enter other parameters.
EXTRA  ta+I
Messages — LCD Pictographs — Test Sequences
Status
Te st Plug
Mea­suring
Function Switch Position
Adapter
AUTO
AUTO
AUTO
R
E
2-pole measurement via earthing contact plug is not possible in IT sys­tems.
The intelligent ramp is not possible with RCD types RCD-S and G/R.
Function/Meaning
The test sequence includes a measurement which cannot be processed by the connected test instrument. The corresponding test step must be skipped. Example: The test sequence includes an RCM measurement which has been sent to the PROFITEST MTECH+.
The test sequence has been run successfully.
No test sequences have been saved. Cause: These may have been deleted as a result of any of the following
actions: changing the language, the profile or the DB mode, or resetting the test instrument to its default settings.
Error Messages — LCD Pictographs — PRO-AB Leakage Current Measuring Adapter
Status
Te st Plug
Mea­suring
Function Switch Position
Function/Meaning
Adapter
Measuring range exceeded
EXTRA  I
EXTRA  I
EXTRA  I
EXTRA  I
Change to the larger measuring range (test instrument and leakage cur-
L
rent measuring adapter).
Test measurement: The test has been passed.
L
The leakage current measuring adapter is now ready for use.
Test measurement: The test has not been passed.
L
The leakage current measuring adapter is defective. Contact our repair service department.
Test measurement:
L
Check the fuse in the leakage current measuring adapter.
24 Gossen Metrawatt GmbH
Page 25
Database and Entry Operations — Pictographs
Attention!
!
Z
EXTRA EXTRA  RCM
I
L-N
N
/ IF
/ Z
L-PE
tA+I
Measured Value Storage with Deviating Electrical Circuit Parameter
The electrical circuit parameter selected by yourself at the test instrument does not coincide with the parameter entered under object data in the structure.
Example: Residual operating current is specified as 10 mA in the data­base, but you have performed measurement with 100 mA. If you want to perform all future measurements with 100 mA, the value in the database
has to be changed by acknowledging with the key. The measured value is documented and the new parameter is accepted.
If you want to leave the parameter in the database unchanged, press the
key. The measured value and the changed parameter are only docu-
mented in this case.
All
All
All
All
All
All
Please enter a designation (alphanumeric).
Operation with a Barcode Scanner Error message when the “EDIT” entry field is opened and rechargeable
battery voltage is less than 8 V. Output voltage is generally switched off during barcode scanner operation if U is less than 8 V in order to assure that remaining battery capacity is adequate for entering designations for devices under test and saving the measurement.
Remedy: Rechargeable batteries must be recharged or replaced towards the end of their service life.
Operation with a Barcode Scanner Current flowing through the RS 232 port is too high. Remedy: The connected device is not suitable for this port.
Operation with a Barcode Scanner Barcode not recognized, incorrect syntax.
Data cannot be entered at this location within the structure. Remedy: Observe profile for preselected PC software (see SETUP menu in
section 8).
Measured value cannot be saved at this location within the structure. Remedy: Make sure that you have selected the right profile for you PC
evaluation program in the SETUP menu (see section 8).
All
All
SETUP
Memory is full.
Remedy: Save your measurement data to a PC and then clear memory at the test instrument by deleting the database or by importing an empty database.
Delete measurement or database elements.
This prompt window asks you to confirm deletion (YES).
Data loss after restoring default settings!
Back up your measurement data to a PC before pressing the respective key. This prompt window asks you to confirm deletion.
Gossen Metrawatt GmbH 25
Page 26

7 Operation

Attention!
!
Note
Attention!
!
Attention!
!
Attention!
!
Note
Attention!
!
Attention!
!
Attention!
!
Attention!
!
Attention!
!
Attention!
!
BAT
The protective foil on the two sensor surfaces (finger contacts) of the test plug must be removed to ensure re­liable detection of touch voltages.

7.1 Power Supply

The instrument is powered by rechargeable batteries. The included Master Battery Pack (Z502H) or commercially available individual rechargeable or regular batteries can be used.
If at all possible use the included battery pack (Z502H) with sealed cells. This ensures that the complete set of rechargeable batteries is always replaced at the same time and that all batteries are inserted with correct polar­ity, in order to assure that they do not fail.
The included Z502H battery pack has already been inserted during initial startup (see condensed operating instructions).
7.1.1 Inserting or Replacing the Battery Pack (Z502H) or Com­mercially Available Individual (Rechargeable) Batteries
Before opening the battery compartment, disconnect the instrument from the measuring circuit (mains) at all poles!
Commercially available, individual rechargeable or regu­lar batteries must comply with the technical data (see page 10).
Loosen the slotted screw for the rechargeable battery com-
partment lid on the back and remove the lid.
Remove the depleted battery pack or commercially available
rechargeable or regular batteries.
Insert the battery pack or commercially available rechargeable
or regular batteries into the battery compartment.
In the case of commercially available, individual re­chargeable or regular batteries: make sure that all of the batteries are inserted with correct polarity. If just one bat­tery is inserted with reversed polarity, it will not be recog­nized by the instrument and may result in leakage from the batteries and damage to the instrument.
Replace the lid and retighten the screw.
Dispose of the battery pack or commercially available, individual rechargeable or regular batteries in an environ­mentally sound fashion when their service life has nearly expired (approx. 80% charging capacity). See section 24, “Disposal and Environmental Protection”, on page 99.

7.1.2 Charging the Battery Pack (Z502H) in the Tester

If commercially available, individual rechargeable batter­ies are used, they must be charged externally. Do not use the Z502R charger to charge commercially available individual batteries. The quality of commercially available, individual rechargeable batteries cannot be checked and may result in overheating and thus deformation and ex­plosion when charging them in the instrument.
If commercially available, individual rechargeable batter­ies are used, they must be charged externally. Do not use the Z502R charger to charge commercially available individual batteries. The quality of commercially available, individual re­chargeable batteries cannot be checked and may result in overheating and thus deformation and explosion when charging them in the instrument.
Regular batteries may not be charged.
Use only the Z502R charger in order to recharge the Compact Battery Pack (Z502H) in the test instrument.
The Z502R charger is suitable for mains operation only!
Do not switch the test instrument on during charging. The charging process may otherwise be impaired.
Verify that the battery pack (Z502H) is inserted, i.e. that com-
mercially available battery packs or batteries are not inserted.
Insert the correct mains plug for your country into the charger
Z502R.
Connect the Z502R charger to the test instrument with the
jack plug, and then to the 230 V mains with the interchange­able plug.
Do not disconnect the charger from the test instrument until
the green LED (charged/ready) lights up.
If the rechargeable batteries or battery pack have not been used or recharged for a lengthy period of time (> 1 month), thus result­ing in excessive depletion:
Observe the charging sequence (indicated by LEDs at the char­ger) and initiate a second charging sequence if necessary (dis­connect the charger from the mains and from the test instrument to this end, and then reconnect it).
Please note that the system clock stops in this case and must be set to the correct time after the instrument has been restarted.

7.2 Switching the Instrument On/Off

The test instrument is switched on by pressing the ON/START key. The menu which corresponds to the momentary selector switch position is displayed.
The instrument can be switched off manually by simultaneously pressing the MEM and HELP keys.
After the period of time selected in the SETUP menu has elapsed, the instrument is switched off automatically (see “Device Set­tings”, section 8).
Battery Test
A battery test is performed after switching the instrument on. If supply voltage has fallen below the permissible
lower limit, the pictograph shown at the right appears. “Low Batt!!!” is also displayed along with an icon.
The instrument does not function if the batteries have been depleted excessively, and no display appears.
Ensure adequate power supply by charging the rechargeable bat­tery pack (Z502H/) or by inserting fully charged, commercially available rechargeable batteries or new batteries. See section 7.1, “Power Supply”, on page 26.
26 Gossen Metrawatt GmbH
Page 27

8 Instrument Settings

SETUP
LED and LCD test menu
Rotary switch balancing
Brightness/contrast menu
Calibration date
Display: date/time
Display: auto shutdown
Display: auto shutdown
of display illumination after 15 s.
of the tester after 60 s.
Time, language
1
2
3
and battery test menu
0b
0a
0
Return to main menu
MAINS LED: test green
MAINS LED: test red
UL/RL LED: test red
RCD-FI LED: test red
Cell test
Inverse cell test
Hide all pixels
Show all pixels
Acoustic signal test
1
Return to main menu
Brightness/contrast submenu
Set time
Default settings
Language for user interface
3
3a 3b
3c
3e
Set date
Duty cycle for display illumination / tester
0b
Return to submenu
0a
Display Illumination On-time
Time, On-Time and Default Settings
Menu Selection for Operating Parameters
LED tests LCD and acoustic signal tests
Test Instrument On-Time
3f
4
No automatic shutdown, continuously on
DB MODE submenu
3g
Current inspector
Firmware/software info page 97
Brightness and Contrast Settings
Create
or select
inspector
(deletion via ETC only)
Distributor structures
3d
Gossen Metrawatt GmbH 27
Page 28
LED and LCD test menu
Rotary switch balancing
Brightness/contrast menu
Calibration date
Display: date/time
Display: auto shutdown
Display: auto shutdown
of display illumination after 15 s.
of the tester after 60 s. Time, language
1
2
3
and battery test menu
0b
0a
0
Return to main menu
Brightness/contrast submenu
Set time
Default settings
Language for user interface
3
3a 3b
3c
3e
Set date
Duty cycle for display illumination / tester
Set time
Menu Selection for Operating Parameters
Set Time, Language, Acoustic Signal
Set date
Select time
Increase hours
Increase minutes
Decrease hours
Apply settings
Decrease minutes
3a
Increase seconds Decrease seconds
Return to submenu
Select date
Decrease
Decrease
day
Apply settings
month
3b
Decrease year
Return to submenu
Increase
Increase
day
month
Increase
year
3f
4
Current inspector
DB MODE submenu
3g
Firm/software info page 97
Brightness and Contrast Settings
Create or select inspector
(deletion via ETC only)
Distributor structures
3d
28 Gossen Metrawatt GmbH
Page 29
Significance of Individual Parameters
Note
Note
Attention!
!
Attention!
!
Attention!
!
0a
0b
2
2
2
2
3c
3d
3e
Test Instrument On-Time
The period of time after which the test instrument is automatically shut off can be selected here. This selection has a considerable influence on the service life and the charging status of the batter­ies.
LCD Illumination On-Time
The period of time after which LCD illumination is automatically shut off can be selected here. This selection has a considerable influence on the service life and the charging status of the batter­ies.
Submenu: Rotary Switch Balancing
Proceed as follows in order to precision adjust the rotary switch:
1 Press the TESTS Rotary Switch / Battery Test softkey in order
to access the rotary switch balancing menu.
2 Then press the softkey with the rotary switch icon.
SETUP
3 Make sure that the rotary switch is set to
The level mark to the left of the number should be aligned to the black function field with the respective rotary switch designation. The value of the number can be displayed within a range of -1 to 101 and should be between 45 and 55. In the case of -1 or 101, the rotary switch position does not match the measuring function shown at the display. If the displayed value is not within this range, readjust the position by pressing the edges readjustment.
readjust
softkey . A brief acoustic signal acknowl-
.
User Interface Language (CULTURE)
Select the desired country setup with the appropriate country
code.
All structures, data and se­quences are deleted when the language is changed!
Back up your structures, measurement data and se­quences to a PC before pressing the key. The prompt window shown at the right asks you to confirm deletion.
Profiles for Distributor Structures (PROFILES)
The profiles are laid out in a tree structure. The tree structure for the utilized PC evaluation program may differ from that of the PROFITEST MASTER. For this reason, the PROFIT­EST MASTER provides the user with the opportu­nity of adapting this struc­ture.
Selecting a suitable profile determines which object combinations are made possible. For example, this makes it possible to create a distributor which is subordinate to another or to save a measurement to a given building.
Select the PC evaluation program you intend to use.
If labeling in the LCD image of the rotary switch does not correspond with its actual position, a continuous acous­tic signal is generated as a warning when the readjust softkey is pressed.
4 Acknowledge by pressing the softkey with the rotary switch icon.
The display is then switched to the next measuring function.
5 Turn the rotary switch clockwise to the next measuring function (after
SETUP
comes
IN).
6 Repeat steps 3 through 5 until all rotary switch functions have
been tested, and if necessary readjusted.
7 Press ESC, in order to return to the main menu.
Submenu: Battery Level Query
If battery voltage has dropped to 8.0 V or less, the UL/RL LED lights up red and an acoustic signal is generated as well.
Measuring Procedure
If battery voltage drops to below 8.0 V during the course of a measuring sequence, this is only indicated by means of a pop-up window. Measured values are invalid. The measurement results cannot be saved to memory.
All structures, data and se­quences are deleted when the profile is changed!
Back up your structures, measurement data and se­quences to a PC before pressing the key. The prompt window shown at the right asks you to confirm deletion.
If you haven’t selected a suitable PC evaluation program and, for example, if measured value storage to the selected location within the struc­ture is not possible, the pop-up window shown at the right appears.
Default Settings (GOME SETTING)
The test instrument is returned to its original default settings when this key is activated.
All structures, data and sequences are deleted! Back up your structures, measurement data and sequences to a PC before resetting.
Press ESC in order to return to the main menu.
Gossen Metrawatt GmbH 29
Page 30
Adjusting Brightness and Contrast
Note
Note
3f
Jump back to
Increase brightness
Decrease brightness
Increase contrast
Decrease contrast
previous menu
3g
4
Add a new inspector
Accept letter/character
Select letter/character
Select letter/character
Delete letter/character Toggle: upper/lowercase,
vowel mutations & special chars.
Accept name
Select inspector
Select inspector
Accept inspector
DB-MODE – Display of the Database in the Text or ID Mode
Selecting, Adding or Deleting an Inspector
See also section 10.8 on page 39 regarding the entry of a text.
Creating Structures in the TXT MODE
The database in the test instrument is set to the text mode as a default feature and “TXT” appears in the header. You can create structure elements in the test instrument and label them in plain text, e.g. Customer XY, Distributor XY and Circuit XY.
Creating Structures in the ID MODE
You can work in the ID MODE as an alternative, in which case “ID” appears in the header. You can create structure elements in the test instrument and label them with any desired ID numbers.
Structures can be created in the test instrument in either the text mode or the ID mode. In contrast to this, designations and ID numbers are always assigned in the report generating program.
If no texts or ID numbers have been entered to the test instrument when creating structures, the report generating program creates the missing entries automatically. These can then be edited in the report generating program and transferred back to the test instru­ment if required.
(delete inspector via ETC only)
The inspector cannot be changed. If an inspector’s name is incorrect, it can be deleted and a new inspector can be created with the correct name. Changes are not retroactive. Deleted inspectors are retained for tests which have already been performed.
30 Gossen Metrawatt GmbH
Page 31

9 Database

Note

9.1 Creating Distributor Structures, General

A complete distributor structure with data for electrical circuits and RCDs can be created in the test instrument. This structure makes it possible to assign measurements to the electrical circuits of various distributors, buildings and customers.
There are two possible procedures:
• On location or at the construction site: create a distributor structure in the test instrument. A distributor struc­ture with up to 50,000 structure ele­ments can be cre­ated in the test instrument, which is saved to the instru­ment’s flash mem­ory.
or
• Create and save an image of an existing distributor structure at a PC with the help of ETC report generating software (Elec­tric Testing Center) – see Help > Getting Started (F1). The dis­tributor structure is then transferred to the test instrument.
The test instrument and the PC must be connected with a USB cable in order to transfer structures and data.
The rotary selector switch may not be set to the “U” posi­tion during data transmission.
The following image appears at the display during transfer of struc­tures and data.

9.3 Creating a Distributor Structure in the Test Instrument Overview of the Meanings of Icons used to Create Structures

Icons Meaning
Main
Sublevel
level
Memory Menu, Page 1 of 3
Cursor UP: scroll up
Notes regarding ETC
The following steps must be completed before using the soft­ware:
• Install the USB device driver (required for operation of the test instrument at a PC): GMC-I Driver Control software for installing the USB device driver can be downloaded from our website: https://www.gmc-instruments.de/services/download-center/
• Install ETC report generating software: The most up-to-date version of ETC can be downloaded free of charge from the mygmc page of our website as a ZIP file, if you have registered your test instrument: https://www.gmc-instruments.de/services/mygmc/
Cursor DOWN: scroll down
ENTER: Acknowledge selection. + – change to sub-level
(expand directory) or
+ change to main level
(close directory)
Display the complete structure designation (max. 63 characters) or ID number (25 characters) in a zoom window.
Temporarily switch back and forth between struc­ture designation and ID number.
These keys don’t have any effect on the main set­ting in the setup menu (see “DB Mode” on page 30).
Hide the zoom window
Change display to menu selection
Memory Menu, Page 2 of 3
Add a structure element

9.2 Transferring Distributor Structures

The following data transfer operations are possible:
• Transfer a distributor structure from the PC to the test instru­ment.
• Transfer a distributor structure including measured values from the test instrument to the PC.
Gossen Metrawatt GmbH 31
Page 32
Icons Meaning
Distributor
A check mark to the right of a structure element means that all measurements within the respective hierarchy have been passed. x: at least one measurement has not been passed. No symbol: measurement has not yet been performed.
Building
Customer
RCD
Circuit
Equipment
Same type of element as in the Windows Explorer: +: sub-objects available, display by pressing . –: sub-objects are displayed, hide by pressing .
Equipment
Meanings of icons from top to bottom: Customer, building, distributor, RCD, electrical cir-
cuit, operating equipment, machine and earth electrode (display of the icons depends on the selected structure element).
Selection: UP/DOWN scroll keys and In order to add a designation to the selected
structure element, refer to the edit menu in follow­ing column.
Delete the selected structure element.
Show measurement data, if a measurement has been performed for this structure element.
Edit the selected structure element.
Memory Menu, Page 3 of 3
Search for ID number. > Enter complete ID number.
Search for text. > Enter full text (complete word).
Search for ID number or text.
Distributor Structure Symbology / Tree Structure
Continue searching.
Edit menu
Cursor LEFT: Select an alphanumeric character.
Cursor RIGHT: Select an alphanumeric character.
ENTER: accept an individual character.
Acknowledge entry
Scroll left
Scroll right
Delete character
Switching amongst different types of alphanu­meric characters:
A Upper case letters
a Lower case letters
0 Numbers
@ Special characters
32 Gossen Metrawatt GmbH
Page 33

9.3.1 Creating Structures (example for electrical circuit)

Note
Note
Note
Scroll up
Scroll down
Comfirm selection /
Display object
Nextpage
change level
or ID number
Create object
Delete object
VA: Show measurement data
Edit designation
Scroll up
Scroll down
Comfirm selection
Select character
Select character
Accept character
Delete character
Character selection:
Save object designation
A, a, 0, @
Select parameter
Parameter settings list
Acknowledge parameter selection
Acknowledge parameter setting
and return to page 1/3
Select parameter setting
in the database menu.
After selection with the MEM key, all setting options for the cre­ation of a tree structure are made available on three menu pages (1/3, 2/3 and 3/3). The tree structure consists of structure ele­ments, referred to below as objects.
Selecting the Position at which a New Object will be Added
Use the  keys in order to select structure elements. Change to the sub-level with the key. Go to the next page with the >> key
Entering a Designation
Enter a designation and then acknowledge it with ✓.
Acknowledge your entry with and , because the entry will otherwise not be accepted.
Entering a Comment
Enter a comment and then acknowledge it with ✓.
Creating a New Object
Press the key in order to create a new object.
Select a new object from a list.
Acknowledge your entry with ✓ and , because the entry will otherwise not be accepted.
Setting Electrical Circuit Parameters
For example, nominal current values must be entered here for the selected electrical circuit. Measuring parameters which have been accepted and saved in this way are subsequently accepted by the current measuring menu automatically when the display is switched from the structure view to measurement.
Electrical circuit parameters changed during structure creation are also retained for individual measurements (measurement without saving data).
If you change the electrical circuit parameters specified by the structure in the test instrument, a warning is displayed when the change is saved (see error message on page 25).
Select the desired object from the list with the keys and acknowledge with the key.
Depending upon the profile selected in the test instrument’s SETUP menu (see section 8), the number of object types may be limited, and the hierarchy may be laid out differently.
Gossen Metrawatt GmbH 33
Page 34

9.3.2 Searching for Structure Elements

Scroll up
Scroll down
Comfirm selection /
Display object
Menu selection page 3/3
change level
or ID number
Search for ID number
Search for text
Search for ID number or text.
Select character
Select character
Accept character
Delete character
Character selection:
Save object designation
Continue searching
End search
Regardless of the currently selected object, the search is started at database.
Go to page 3/3 in the database menu.
After selecting text search...
If no further matches are found, the message shown above is dis­played.

9.4 Saving Data and Generating Reports

Preparing and Executing a Measurement Measurements can be performed and stored to memory for each
structure element. Proceed as follows, adhering to the prescribed sequence:
Select the desired measurement with the rotary knob.Start the measurement by pressing the ON/START ▼ or I
Upon completion of measurement, the Floppy Disk softkey is displayed.
Briefly press the Save Value key.
The display is switched to the memory menu or the structure view.
Navigate to the desired memory location, i.e. to the desired
structure element / object, for which the measurement data will be saved.
If you would like to save a comment along with the
measurement, press the key shown at the right and enter a designation via the “EDIT” menu as described in section 9.3.1.
Complete data storage by pressing the “STORE” key.
N
key.
... and entering the desired text (only full matches are found – no wild cards, case sensitive) ...
... the first match is displayed. Further matches can be found by selecting
the icon shown at the right.
Saving Error Messages (pop-ups)
If a measurement is ended without a measured value due to an error, the measurement can be saved along with the pop-up by pressing the “Save Value” key. The corresponding text is read out in ETC instead of the pop-up symbol. This only applies to a limited number of pop-ups (see below). Neither a symbol nor a text can be accessed in the test instrument’s database itself.
Alternative Storage Procedure
The measured value can be saved to the last se-
lected object in the structure diagram by pressing and holding the Save Value key, without switching the display to the memory menu.
34 Gossen Metrawatt GmbH
Page 35
Note
If you change the parameters in the measurement view,
Note
Note
Note
they’re not saved for the structure element. A measure­ment with changed parameters can nevertheless be saved to the structure element, and any changed param­eters are documented in the report for each measure­ment.
Retrieving Saved Measured Values
Switch the display to the distributor structure by pressing the
MEM key and select the desired electrical circuit with the scroll keys.
Switch to page 2
by pressing the key shown here:
Display the measurement data
by pressing the key shown here:
One measurement with date and time, as well as any com­ment you might have entered, is displayed in each screen. Example: RCD Measurement
Data Evaluation and Report Generation with the Report Generating Program
All data, including the distributor structure, can be transferred to the PC and evaluated with the help of the report generating pro­gram. Additional information can be entered here subsequently for the individual measurements. After pressing the appropriate key, a report including all measurements within a given distributor structure is generated, or the data are exported to an Excel spreadsheet.
The database is exited when the rotary selector switch is turned. Previously selected parameters in the database are not used for the measurement.

9.5 Use of Barcode Scanners and RFID Readers Search for an Already Scanned Barcode

The search can be started from any switch setting and menu. Scan the object’s barcode. The recognized barcode is displayed inversely. This value is accepted after pressing the ENTER key.
A previously selected object is not taken into consider­ation by the search.
A check mark in the header means that the respective measurement has been passed. An X means that the measurement has not been passed.
Scrolling amongst measurements
is possible with the keys shown here:
The measurement can be deleted with the key
shown here:
A prompt window asks you to confirm deletion.
With the help of the key shown at the right (MW: mea­sured value / PA: parameter), the setting parameters can be displayed for this measurement.
Continued Searching in General
Regardless of whether or not an object has been found, searching can be continued by pressing the key shown at the right:
– Object found: Searching is continued below the previously
selected object.
– Further object found: The entire database is searched at all
levels.
Reading In a Barcode for Editing
If the menu for alphanumeric entry is active, any value scanned by means of a barcode or RFID reader is accepted directly.
Using a Barcode Printer (accessory)
The following functions are made possible with the help of a bar­code printer:
• Read-out of ID numbers as barcodes – for quick and conve­nient acquisition for periodic testing
• Print out repeatedly occurring designations such as test object types encrypted as barcodes in a list, allowing them to be read in as required for comments
Scrolling amongst measurements
is possible with the keys shown here:
Gossen Metrawatt GmbH 35
Page 36

10 General Information on Measurements

Attention!
!
Note

10.1 Using Cable Sets and Test Probes

• Scope of delivery: 2-pole measuring adapter and cable for expansion into a 3-pole adapter (PRO-A3-II/)
Optional accessory: PRO-RLO II (Z501P) 2-pole measuring adapter with 10 m cable
• Optional accessory: KS24 cable set (GTZ3201000R0001)
Measurements per DIN EN 61010-031 may only be performed in environments in accordance with measuring categories III and IV with the safety cap attached to the test probe at the end of the measurement cable.
In order to establish contact inside 4 mm jacks, the safety caps have to be removed by prying open the snap fastener with a pointed object (e.g. the other test probe).
Grip and hold the test plug and test probes securely when they have been inserted, for example, into a socket. Danger of injury exists if tugging at the coil cord occurs, which may cause the test plug or test probes to snap back.

10.2 Test Plug – Changing Inserts

The test plug can be fitted with various inserts (e.g. two-pole measuring adapter or country-specific plug insert).
In order to change inserts, unscrew the retaining ring until you can pull out the currently used insert. Then mount the desired insert and retighten the retaining ring.
(See overview in section 5.4 on page 8.)

10.3 Connecting the Instrument

For systems with earthing contact sockets, connect the instru­ment to the mains with the test plug to which the appropriate, country-specific plug insert is attached. Voltage between phase conductor L and protective conductor PE may not exceed 253 V! Poling at the socket need not be taken into consideration. The instrument detects the positions of phase conductor L and neu­tral conductor N, and automatically reverses polarity if necessary. This does not apply to the following measurements:
– Voltage measurement in switch position U – Insulation resistance measurement – Low-resistance measurement
The positions of phase conductor L and neutral conductor N are identified on the plug insert.
If measurement is to be performed at three-phase outlets, at dis­tribution cabinets or at permanent connections, the measuring adapter must be attached to the test plug. Connection is estab­lished with the test probes: one at PE or N and the other at L.
The 2-pole measuring adapter must be expanded to 3 poles with the included measurement cable for the performance of phase sequence testing.
Touch voltage (during RCCB testing) and earthing resistance can be, and earth-electrode potential, standing surface insulation resistance and probe voltage must be measured with a probe. The probe is connected to the probe connector socket with a 4 mm contact-protected plug.

10.4 Automatic Settings, Monitoring and Shutdown

The test instrument automatically selects all operating conditions which it’s capable of determining itself. It tests line voltage and frequency. If these lie within their valid nominal ranges, they appear at the display panel. If they are not within nominal ranges, prevailing voltage (U) and frequency (f) are displayed instead of UN and f
.
N
Touch voltage which is induced by test current is monitored for each measuring sequence. If touch voltage exceeds the limit value of > 25 V or > 50 V, measurement is immediately inter­rupted. The U
If battery voltage falls below the permissible limit value the instru­ment cannot be switched on, or it is immediately switched off.
LED lights up red.
L/RL
The measurement is interrupted automatically, or the measuring sequence is blocked (except for voltage measuring ranges and phase sequence testing) in the event of:
• Impermissible line voltages (< 60 V, > 253 V / > 330 V / > 440 V or > 550 V) for measurements which require line volt­age
• Interference voltage during insulation resistance or low resis­tance measurements
• Overheating at the instrument As a rule, excessive temperatures only occur after approxi­mately 50 measurement sequences at intervals of 5 seconds,
or Z
when the rotary selector switch is set to the Z position.
L-P E
L-N
If an attempt is made to start a measuring sequence, an appropriate message appears at the display panel.
The instrument only switches itself off automatically after comple­tion of an automatic measuring sequence, and after the predeter­mined on-time has expired (see section 7.2). On-time is reset to its original value as defined in the setup menu, as soon as any key or the rotary selector switch is activated.
The instrument remains on for approximately 75 s in addition to the preset on-time for measurements with rising residual current in systems with selective RCDs.
The instrument always shuts itself off automatically!

10.5 Measured Value Display and Memory

The following items appear at the display panel:
• Measured values with abbreviations and units of measure
• Selected function
• Nominal voltage
• Nominal frequency
• Error messages
Measured values for automatic measuring sequences are stored and displayed as digital values until the next measurement sequence is started, or until automatic shutdown occurs. If the upper range limit is exceeded, the upper limit value is dis­played and is preceded by the “>” symbol (greater than), which indicates measurement value overrun.
The depiction of LEDs in these operating instructions may vary from the LEDs on the actual instrument due to product improvements.
Testing Earthing Contact Sockets for Correct Connection The testing of earthing contact sockets for correct connection
prior to protective measures testing is simplified by means of the instrument’s error detection system.
The instrument indicates improper connection as follows:
Impermissible line voltage (< 60 V or > 253 V): The MAINS/NETZ LED blinks red and the measuring sequence is disabled.
Protective conductor not connected or potential to earth 50 V at 50 Hz (switch position U – single-phase measurement): If the contact surfaces are touched (finger contact*) while PE is being contacted (via the country-specific plug insert, e.g. SCHUKO, as well as via the PE test probe at the 2-pole adapter) PE appears (only after a test sequence has been started). The U
* For reliable detection of touch voltages, both sensor surfaces on the
test plug must be touched with unprotected fingers/palm, i.e. with di­rect skin contact (see also section 7).
Neutral conductor N not connected (during mains dependent measurements): The MAINS/NETZ LED blinks green.
One of the two protective contacts is not connected: This is checked automatically during testing for touch current U
. Poor contact resistance at one of the contacts leads to
IN
and RCD/FI LEDs light up red as well.
L/RL
36 Gossen Metrawatt GmbH
Page 37
one of the following displays depending upon poling of the
Note
Attention!
!
plug: – Display at the connection pictograph:
PE interrupted (x), or underlying protec­tive conductor bar interrupted with refer­ence to the keys at the test plug Cause:
Voltage measuring path interrupted
Consequence: measurement is disabled
Display at the connection pictograph:
Top protective conductor bar interrupted with reference to the keys at the test plug
Cause: current measuring path interrupted Result: no measured value display
See “LED Indications, Mains Connections and Potential Differences” on page 17.
Reversal of N and PE in a system without RCCBs cannot be detected and is not indicated by the instrument. In a system including an RCCB, the RCCB is tripped during touch voltage measurement without RCCB trip­ping (automatic Z are reversed.
measurement), insofar as N and PE
L-N

10.6 Help Function

The following information can be displayed for each switch posi­tion and basic function after it has been selected with the rotary selector switch:
• Wiring diagram
• Measuring range
• Nominal range of use as well as measuring and intrinsic uncertainties
• Nominal value
Press the HELP key in order to query online help.If several pages of help are available for the respective mea-
suring function, the HELP key must be pressed repeatedly.
Press the ESC key in order to exit online help.
Gossen Metrawatt GmbH 37
Page 38

10.7 Setting Parameters or Limit Values using RCD Measurement as an Example

1
2
2
3
4
4
5
6
2
4
3
5
6
1 Access the submenu for setting the desired parameter. 2 Select a parameter using the or scroll key. 3 Switch to the setting menu for the selected parameter with the
scroll key. 4 Select a setting value using the or scroll key. 5 Acknowledge the setting value with the key. This value is
transferred to the settings menu. 6 The setting value is not permanently accepted for the respec-
tive measurement until
returned to the main menu. You can return to the main menu
by pressing ESC instead of , without accepting the newly
selected value.
is pressed, after which the display is
Parameter Lock (plausibility check)
Individually selected parameter settings are checked for plausibil­ity before transfer to the measurement window.
If you select a parameter setting which doesn’t make sense in combination with other parameter settings which have already been entered, it’s not accepted. The previously selected parame­ter setting remains unchanged.
Remedy: select another parameter setting.
38 Gossen Metrawatt GmbH
Page 39

10.8 Freely Selectable Parameter Settings or Limit Values

Note
Note
Select the EDIT menu.
Select value/U/M.
Select value/U/M.
Accept value/U/M.
Delete character
Save value (to list)
Select the EDIT+ menu.
Select value/U/M.
Select value/U/M.
Accept value/U/M.
Delete character
Save value (to list)

10.8.1 Changing Existing Parameters

Individual parameters can be changed for certain measuring func­tions, i.e. adjusted within predetermined limits.
The EDIT menu doesn’t appear until after switching to the right-hand column and selecting the editable parameter .
Example for RLO Measuring Function – Parameter: LIMIT RLO
1 Open the submenu for setting the desired parameter (no fig-
ure, see section 10.7).
2 Select the editable parameter – identified with the icon –
with the or  scroll key.
3 Select the edit menu by pressing the key.

10.8.2 Adding New Parameters

For certain measuring functions, additional values within pre­defined limits can be added in addition to the fixed values.
The EDIT+ menu doesn’t appear until after switching to the right-hand column.
Example for IN Measuring Function – Parameter: I
1 Open the submenu for setting the desired parameter
(no figure, see section 10.7).
Select the edit menu by pressing the key.
N
4 Select the respective characters with the left or right cursor
key. The character is accepted by pressing the key. The value is acknowledged by selecting key.
Observe the predefined limits for the new setting value. Enter any places to the right of the decimal point as well.
and then pressing the
2 Select the respective characters with the LEFT or RIGHT cursor
key. The character is accepted by pressing the key. The value is acknowledged by selecting key. The new parameter is added to the list.
Observe the predefined limits for the new setting value. Enter any places to the right of the decimal point as well.
and then pressing the
Gossen Metrawatt GmbH 39
Page 40
10.9 2-Pole Measurement with Rapid or Semiautomatic Polarity
L1-N L2-N
L3-N L1-L2 L2-L3 L1-L3
N-PE L1-PE L2-PE L3-PE
L+N-PE
L1-N L2-N
L3-N L1-L2 L2-L3 L1-L3
Z
L-PE
Z
L-N
L1-PE L2-PE L3-PE
R
iso
N-PE L1-PE L2-PE L3-PE
L1-N L2-N
L3-N L1-L2 L2-L3 L1-L3
U
L1-N L2-N
L3-N L1-L2 L2-L3 L1-L3
N-PE L1-PE L2-PE L3-PE
L+N-PE
L1-N L2-N
L3-N L1-L2 L2-L3 L1-L3
Z
L-PE
Z
L-N
L1-PE L2-PE L3-PE
R
iso
L1-PE L2-PE L3-PE
N-PE
L1-N L2-N
L3-N L1-L2 L2-L3 L1-L3
U
Reversal
Rapid, semiautomatic polarity reversal is possible for the following measurements:
• Voltage U
• Loop impedance Z
• Internal line resistance Z
• Insulation resistance R
LP-E
L-N
INS
Rapid Polarity Reversal at the Test Plug
The polarity parameter is set to AUTO. Fast and convenient switching amongst all polarity variants, or
switching to the parameter settings submenu, is possible by pressing the I
key at the instrument or the test plug.
N
Semiautomatic Polarity Reversal in Memory Mode
The polarity parameter is set to AUTO. If testing is to be conducted with all polarity variants, automatic
polarity changing takes place after each measurement after sav- ing.
Polarity variants can be skipped by pressing the I instrument or the test plug.
40 Gossen Metrawatt GmbH
key at the
N
Page 41

11 Measuring Voltage and Frequency

Note
U
2
1
Select the Measuring Function
Switch Back and Forth Between Single and 3-Phase Measure­ment
Press the softkey shown at the left in order to switch back and forth between single and 3-phase mea­surement. The selected phase measurement is dis­played inversely (white on black).

11.1 Single-Phase Measurement Connection

11.1.2 Voltage Between L – PE, N – PE and L – L with 2-Pole Adapter Connection

Press the softkey shown at the left in order to switch back and forth between the country-specific plug insert, e.g. SCHUKO, and the 2-pole adapter. The selected connection type is displayed inversely (white on black).
See section 10.9 concerning 2-pole measurement with rapid or semiautomatic polarity reversal.
A probe must be used in order to measure probe voltage U
S-PE
11.1.1 Voltage Between L and N (U and PE
(U
) with Country-Specific Plug Insert, e.g.
N-PE
SCHUKO
Press the softkey shown at the left in order to switch back and forth between the country-specific plug insert, e.g. SCHUKO, and the 2-pole adapter. The selected connection type is displayed inversely (white on black).
L-N
.
),
L and PE
(U
L-PE
) a
nd N

11.2 3-Phase Measurement (line-to-line voltage) and Phase Sequence

Connection
The measuring adapter (2-pole) is required in order to connect the instrument, and is expanded to a 3-pole measuring adapter with the included measure­ment cable.
If you view the country-specific plug insert, e.g. SCHUKO, from the front, you’ll see two embossed let­ters, namely L and N. Automatic polarity reversal does not take place during voltage measurement. You can thus specify the terminal to which the phase is connected in the socket. If (mains) voltage is displayed for UL-PE, then the phase is located where L appears on the con­nector. If (mains) voltage is displayed for N-PE, then the phase is located where N appears on the connector.
Gossen Metrawatt GmbH 41
Page 42
Press softkey U3~.
Note
Clockwise Rotation
Counterclockwise Rotation
A clockwise phase sequence is required at all 3-phase electrical outlets.
• Measurement instru­ment connection is usually problematic with CEE outlets due to contact problems. Measurements can be executed quickly and reliably without contact problems with the help of the Z500A variable plug adapter set available from GMC.
• Connection for 3-wire measurement: L1-L2-L3 at plug in clockwise direction as of PE socket
Direction of rotation is indicated by means of the following dis­plays:
See section 6.4 regarding all indications for the mains connection test.
Voltage Polarity
If the installation of single-pole switches to the neutral conductor is prohibited by the standards, voltage polarity must be tested in order to assure that all existing single-pole switches are installed to the phase conductors.
42 Gossen Metrawatt GmbH
Page 43

12 Testing RCDs

Attention!
!
Note
I
N
3
-------
I
N
(measurement up to 1000 ms)
t
a
I
a
t
S
The testing of residual current devices (RCDs) includes:
• Visual inspection
•Testing
• Measurement Use the test instrument for testing and measurement.
When testing systems with RCCBs, they may switch off. This may occur even though it’s not normally provided for by the test. Leakage currents may be present which, in combination with the test current of the test instru­ment, exceed the shutdown threshold value of the RCCB. PCs which are operated in proximity to such RCCB systems may switch off as a consequence. This may result in inadvertent loss of data. Before conducting the test, precautions should therefore be taken to ensure that all data and programs are suitably backed up and the computer should be switched off, if necessary. The manufacturer of the test instrument assumes no liability for any direct or indirect damage to equipment, comput­ers, peripheral equipment or data bases when perform­ing the tests.
Measuring Method
The following must be substantiated by generating a fault current downstream from the RCD:
• That the RCD is tripped no later than upon reaching its nominal fault current value
• That the continuously permissible touch voltage value UL agreed upon for the respective system is not exceeded
This is achieved by means of:
• Touch voltage measurement 10 measurements with full-waves and extrapolation of I
• Substantiation of tripping within 400 ms or 200 ms with IN
• Substantiation of tripping current with rising residual current This value must be between 50% and 100% of IN (usually about 70%).
N
RCD/FI Table Waveform
Differential Current
Suddenly occurring
Alternating current
Slowly rising
Suddenly occurring
Pulsating direct current
Slowly rising
Direct current
Direct current up to 6 mA
* only PROFITEST MTECH+, PROFITEST MXTRA
Correct RCD/RCCB Function
Typ e AC Type A/F Type B */
B+*
Type EV/
MI*
✔✔✔✔
✔✔✔
✔✔
Test Standard
The following must be substantiated per IEC 60364-6: – Touch voltage occurring at nominal residual current may not
exceed the maximum permissible value for the system.
– Tripping of the RCCB must occur within 400 ms (1000 ms for
selective RCDs) at nominal residual current.
Important Notes
• The test instrument permits simple measurements at all types of RCDs. Select RCD, SRCD, PRCD etc.
• Measurement must be executed at one point only per RCD (RCCB) within the connected electrical circuits. Low-resis­tance continuity must be substantiated for the protective con­ductor at all other connections within the electrical circuit (R or U
).
B
• The measuring instruments often display 0.1 V touch voltage in TN systems due to low protective conductor resistance.
• Be aware of any bias currents within the system. These may cause tripping of the RCDs during measurement of touch volt­age UB, or may result in erroneous displays for measurements with rising current: Display = I
- I
F
bias current
• Selective RCDs identified with an can be used as the sole means of protection for automatic shutdown if they adhere to the same shutdown conditions as non-selective RCDs (i.e. t
< 400 ms). This can be verified by measuring breaking time.
a
• Type B RCDs may not be connected in series with type A or F RCDs.
LO
• No premature tripping with the test instrument, because test­ing is begun with 30% residual current (if no bias current occurs within the system).
Bias Magnetization Only AC measurements can be performed with the 2­pole adapter. Suppression of RCD tripping by means of bias magnetization with direct current is only possible via a country-specific plug insert, e.g. SCHUKO, or the 3­pole adapter.
Gossen Metrawatt GmbH 43
Page 44
Measurement With or Without Probe
Attention!
!
I
N
Nom. res. current: 10 … 500 mA
Type 1: RCD, SRCD, PRCD …
Nominal current: 6 … 125 A
Type 2*: AC , A , F
, B , B+
, EV, MI
k
* Types B, B+, EV, MI =
AC/DC sensitive
Phase displacement: 0°/180°
X times tripping current:
Negative/positive half-wave
Negative/positive direct current
1, 2, 5 (I
N
max. 300 mA)
Waveform:
Connection:
Without/with probe
System type:
TN/TT, IT
Touch voltage:
Time to trip:
< 25 V, < 50 V, < 65 V
Measurements can be performed with or without a probe. Measurements with probe require that the probe and reference
earth are of like potential. This means that the probe must be positioned outside of the potential gradient area of the earth elec­trode (R
The distance between the earth electrode and the probe should be at least 20 m.
The probe is connected with a 4 mm contact-protected plug. In most cases this measurement is performed without probe.
) in the RCD safety circuit.
E
The probe is part of the measuring circuit and may carry a current of up to 3.5 mA in accordance with IEC 61557 / EN 61557.
12.1 Measuring Touch Voltage (with reference to nominal resid­ual current) with ⅓ Nominal Residual Current and Tripping Test with Nominal Residual Current
Select Measuring Function
Connection
Testing for the absence of voltage at the probe can be performed with the U
function (see also section 11.1 on page 41).
PROBE
Set Parameters for I
N
44 Gossen Metrawatt GmbH
Page 45
1) Measuring Touch Current Without Tripping the RCD
Attention!
!
Note
Note
Attention!
!
Measuring Method
The instrument uses a measuring current of only 1/3 nominal residual current for the determination of touch voltage U occurs at nominal residual current. This prevents tripping of the
IN
which
RCCB. This measuring method is especially advantageous, because
touch voltage can be measured quickly and easily at any electrical outlet without tripping the RCCB.
The usual, complex measuring method involving testing for the proper functioning of the RCD at a given point, and subsequent substantiation that all other systems components requiring pro­tection are reliably connected at low resistance values to the selected measuring point via the PE conductor, is made unneces­sary.
N-PE Reversal Test
Additional testing is conducted in order to determine whether or not N and PE are reversed. The pop-up window shown at the right appears in the event of reversal.
2) Tripping Test after the Measurement of Touch Voltage
Press the
The tripping test need only be performed at one measuring point for each RCCB.
If the RCCB is not tripped at nominal residual current, the MAINS/NETZ LED blinks red (line voltage disconnected) and, amongst other values, time to trip t appear at the display panel.
If the RCCB is not tripped at nominal residual current, the RCD/FI LED lights up red.
I
key.
N
and earthing resistance RE
a
In order to prevent the loss of data in data processing systems, perform a data backup before starting the measurement and switch off all consumers.
Start Measurement
Amongst other values, touch voltage U resistance R
appear at the display panel.
E
The measured earthing resistance value RE is acquired with very little current. More accurate results can be obtained with the selector switch in the R The DC + function can be selected here for sys­tems with RCCBs.
and calculated earthing
IN
position.
E
Touch Voltage Too High
If touch voltage U residual current I U
LED lights up red.
L/RL
If the limit value for touch voltage is exceeded during the mea­surement process, U for Germany (65 V applies for Austria – standard: ÖVE/ÖNORM E 8001-1 section 5.3).
Safety shutdown: At up to 70 V, a safety shutdown is tripped within 3 s in accordance with IEC 61010.
Touch voltages of up to 70 V are displayed. If the value is greater than 70 V, U
IN
, which has been measured with 1/3 nominal
IN
and extrapolated to IN, is > 50 V (> 25 V), the
N
> 50 V (> 25 V), safety shutdown occurs
IN
> 70 V is displayed.
Limit Values for Permissible, Continuous Touch Voltage
The limit for permissible, continuous touch voltage is equal to U
= 50 V for alternating voltages (international agreement).
L
Lower values have been established for special applications (e.g. medical applications: U
If touch voltage is too high, or if the RCCB is not tripped, the system must be repaired (e.g. earthing resistance is too high, defective RCCB etc.)!
=25V).
L
Unintentional Tripping of the RCD due to Bias Current in the System Any bias current which might occur can be ascertained as described in section 18.1 on page 80 with the help of a current
clamp transformer. The RCCB may be tripped during the testing of touch voltage if extremely large bias currents are present within the system, or if a test current was selected which is too great for
3-Phase Connections
For proper RCD testing at three-phase connections, the tripping test must be conducted for one of the three phase conductors (L1, L2 or L3).
the RCCB. After touch voltage has been measured, testing can be performed
to determine whether or not the RCCB is tripped within the selected time limit values at nominal residual current.
Unintentional Tripping of the RCD due to Leakage Current in the Measur­ing Circuit
Measurement of touch voltage with 30% nominal residual current does not normally trip an RCCB. However, the trip limit may be exceeded as a result of leakage current in the measuring circuit, e.g. due to interconnected consumers with EMC circuit, e.g. fre-
Inductive Power Consumers
Voltage peaks may occur within the measuring circuit if inductive consumers are shut down during an RCCB trip test. If this is the case, the test instrument might not display any measured value (– – – ). If this message appears, switch all consumers off before performing the trip test. In extreme cases, one of the fuses in the test instrument may blow, and/or the test instrument may be damaged.
quency converters or PCs.
Gossen Metrawatt GmbH 45
Page 46

12.2 Special Tests for Systems and RCDs

Note
Attention!
!
I
F
Nom. res. current: 10 … 500 mA
Type 1: RCD, SRCD, PRCD …
Nominal current: 6 … 125 A
Type 2*: AC , A , F
, B , B+
, EV, MI
k
* Types B, B+, EV, MI =
AC/DC sensitive
Sinusoidal
Negative/positive half-wave
Waveform:
Connection:
Without/with probe
System type:
TN/TT, IT
Negative/positive direct current
Touch voltage:
Tripping limit values:

12.2.1 Testing Systems and RCCBs with Rising Residual Current (AC) for Type AC, A/F, B/B+ and EV/MI RCDs (PROFITEST MTECH+, PROFITEST MXTRA only)

Measuring Method
The instrument generates a continuously rising residual current of (0.3 … 1.3) × I The instrument stores the touch voltage and tripping current val­ues which were measured at the moment tripping of the RCCB occurred, and displays them.
One of the touch voltage limit values, U can be selected for measurement with rising residual current.
within the system for the testing of RCDs.
N
=25V or UL=50/65V,
L
Select Measuring Function
Connection
Start Measurement
Set Parameters for I
Measuring Procedure
After the measuring sequence has been started, the test current generated by the instrument is continuously increased starting at
0.3 times nominal residual current, until the RCCB is tripped. This can be observed by viewing gradual filling of the triangle at I.
If touch voltage reaches the selected limit value (U
F
or 25 V) before the RCCB is tripped, safety shutdown occurs. The UL/RL LED lights up red.
Safety shutdown: At up to 70 V, safety shutdown is trig­gered within 3 s in accordance with IEC 61010.
If the RCCB is not tripped before rising current reaches nominal residual current I
If bias current is present within the system during mea­surement, it’s superimposed onto the residual current which is generated by the instrument and influences measured values for touch voltage and tripping current. See also section 12.1.
, the RCD/FI LED lights up red.
N
=65V, 50V
L
Evaluation
According to IEC 60364-6, rising residual current must, however, be used for measurements in the evaluation of RCDs, and touch voltage at nominal residual current IN must be calculated from the measured values. The faster, more simple measuring method should thus be taken advantage of (see section 12.1).
46 Gossen Metrawatt GmbH
Page 47
12.2.2 Testing Systems and RCCBs with Rising Residual Current
Note
Note
S
I
N
Negative direct current
Positive direct current
Waveform:
180°: Start with neg. half-wave
0°: Start with pos. half-wave
5 times tripping current
X times tripping current:
(AC) for Type B/B+ and EV/MI RCDs (PROFITEST MTECH+PROFITEST MXTRA)
In accordance with IEC 61557 / EN 61557, it must be substanti­ated that, with smooth direct current, residual operating current is no more than twice the value of rated residual current IN. A con­tinuously rising direct current, beginning with 0.2 times rated residual current IN, must be applied to this end. If current rise is linear, rising current may not exceed twice the value of I period of 5 seconds.
Testing with smoothed direct current must be possible in both test current directions.
within a
N
12.2.3 Testing RCCBS with 5 × IN
Measurement of time to trip is performed here with 5 times nomi­nal residual current.
Measurements performed with 5 times nominal fault cur­rent are required for testing type and G RCCBs in the manufacturing process. They’re used for personal safety as well.
Measurement can be started with the positive half-wave at “0°” or with the negative half-wave at “180°”.
Both measurements must nevertheless be performed. The longer of the two tripping times is decisive regarding the condition of the tested RCCB. Both values must be less than 40 ms.
Start Measurement
Select Measuring Function
Set Parameter – Start with Positive or Negative Half-Wave
Set Parameter – 5 Times Nominal Current
Gossen Metrawatt GmbH 47
The following restrictions apply to the selection of tripping current multiples relative to nominal current: 500 mA: 1 × IN, 2 × I
N
Page 48
12.2.4 Testing of RCCBs
Note
Note
Note
I
N
Neg. half-wave
Pos. half-wave
Negative direct current
Positive direct current
Waveform:
X times tripping current:
50% IN*
* No-trip test
with 50% I
N
S
I
N
I
F
or
Type 1:
which are Suitable for Pulsating DC Residual Current
In this case, RCCBs can be tested with either positive or negative half-waves. The standard calls for tripping at 1.4 times nominal current.
Select Measuring Function
Set Parameter – Positive or Negative Half-Wave
Set Parameter – Test With and Without “No-Trip Test”

12.3 Testing of Special RCDs

12.3.1 Systems with Type RCD-S Selective RCCBs

Selective RCCBs are used in systems which include two series connected RCCBs which are not tripped simultaneously in the event of a fault. These selective RCCBs demonstrate delayed response characteristics and are identified with the symbol.
Measuring Method
The same measuring method is used as for standard RCCBs (see sections 12.1 on page 44 and 12.2.1 on page 46).
If selective RCDs are used, earthing resistance may not exceed half of the value for standard RCCBs.
For this reason, the instrument displays twice the measured value for touch voltage.
Select Measuring Function
Set Parameter – Selective
Start Measurement
No-Trip Test
If, during the no-trip test which lasts for 1 sec­ond, the RCD trips too early at 50% IN, i.e. before the actual tripping test starts, the pop-up window shown at the right appears.
The following restrictions apply to the selection of tripping current multiples relative to nominal current: 500 mA: double and five-fold nominal current is not pos­sible in this case.
Tripping Test
Press the IN key. The RCCB is tripped. Blinking bars appear
According to DIN EN 50178 (VDE 160), only type B RCCBs (AC-DC sensitive) can be used for equipment with > 4 kVA, which is capable of generating smooth DC residual current (e.g. frequency converters). Tests with pulsating DC fault current only are not suitable for these RCCBs. Testing must also be conducted with smooth DC residual current in this case.
Measurement is performed with positive and negative half-waves for testing RCCBs during manufacturing. If a circuit is charged with pulsating direct current, the func­tion of the RCCB can be executed with this test in order to assure that the RCCB is not saturated by the pulsating direct current so that it no longer trips.
48 Gossen Metrawatt GmbH
at the display panel, after which time to trip t sistance R
The tripping test need only be performed at one measuring point for each RCCB.
are displayed.
E
and earthing re-
A
Page 49
Note
Selective RCDs demonstrate delayed response charac-
I
N
I
F
or
Type 1:
teristics. Tripping performance is briefly influenced (up to 30 s) due to pre-loading during measurement of touch voltage. In order to eliminate pre-charging caused by the measurement of touch voltage, a waiting period must be observed prior to the tripping test. After the measuring sequence has been started (tripping test), blinking bars are displayed for approximately 30 seconds. Tripping times of up to 1000 ms are permissible. The tripping test is executed immediately after once again pressing the I key.
N
Set Parameter – PRCD with Non-Linear Elements

12.3.2 PRCDs with Non-Linear Type PRCD-K Elements

The PRCD-K is a portable RCD with electronic residual current evaluation laid out as an inline device which switches all poles (L, N and PE). Undervoltage tripping and protective conductor moni­toring are additionally integrated into the PRCD-K.
The PRCD-K is equipped with undervoltage tripping, for which reason it has to be operated with line voltage, and measurements may only be performed in the on state (PRCD-K switches all poles).
Terminology (from DIN VDE 0661)
Portable protective devices are circuit breakers which can be con­nected between power consuming devices and permanently installed electrical outlets by means of standardized plug-and­socket devices. A reusable, portable protective device is a protective device which is designed such that it can be connected to movable cables.
Please be aware that a non-linear element is usually integrated into PRCDs, which leads to immediate exceeding of the greatest permissible touch voltage during U than 50 V).
PRCDs which do not include a non-linear element in the protec­tive conductor must be tested in accordance with section 12.3.3 on page 50.
measurements (UI greater
I
Objective (from DIN VDE 0661)
Portable residual current devices (PRCDs) serve to protect per­sons and property. They allow for the attainment of increased lev­els of protection as provided by protective measures utilized in electrical systems for the prevention of electrical shock as defined in DIN VDE 0100-410. They are to be designed such that they can be installed by means of a plug attached directly to the pro­tective device, or by means of a plug with a short cable.
Measuring Method
The following can be measured, depending upon the measuring method:
•Time to trip tA: tripping test with nominal residual current I (the PRCD-K must be tripped at 50% nominal current)
Tripping current I for testing with rising residual current I
N
F
Start Measurement
Select Measuring Function
Connection
Gossen Metrawatt GmbH 49
Page 50

12.3.3 SRCD, PRCD-S (SCHUKOMAT, SIDOS or comparable)

Note
I
N
I
F
or
Type 1:
I
N
Type 1:
Waveform:
Negative direct current
Positive direct current
180°: Start with neg. half-wave
0°: Start with pos. half-wave
RCCBs from the SCHUKOMAT SIDOS series, as well as others which are of identical electrical design, must be tested after selecting the corresponding parameter.
Monitoring of the PE conductor is performed for RCDs of this type. The PE conductor is monitored by the summation current transformer. If residual current flows from L to PE, tripping current is cut in half, i.e. the RCCB must be tripped at 50% nominal resid­ual current I
Whether or not PRCDs and selective RCDs are of like design can be tested by means of touch voltage U touch voltage U of an otherwise error-free system, the PRCD more than likely con­tains a non-linear element.
.
N
measurement. If a
of greater than 70 V is measured at the PRCD
IN
IN
PRCD-S
The PRCD-S (portable residual current device – safety) is a spe­cial, portable, protective device with protective conductor detec­tion or protective conductor monitoring. The device serves to pro­tect persons from electrical accidents in the low-voltage range (130 to 1000 V). The PRCD-S must be suitable for commercial use, and is installed like an extension cable between an electrical consumer – as a rule an electrical tool – and the electrical outlet.
Select Measuring Function

12.3.4 Type G or R RCCB

In addition to standard RCCBs and selective RCDs, the special characteristics of the type G RCCB can also be tested with the test instrument.
The type G RCCB is an Austrian specialty which complies with device standard ÖVE/ÖNORM E 8601. Erroneous tripping is min­imized thanks to its greater current carrying capacity and short­term delay.
Select Measuring Function
Set Parameter – Type G/R (VSK)
Set Parameter – SRCD / PRCD
Start Measurement
Touch voltage and time to trip can be measured in the G/R-RCD switch position.
It must be observed that time to trip for type G RCCBs may be as long as 1000 ms when measurement is made at nominal residual current. Set the limit value corre­spondingly.
Then select 5 × I
for the G/R setting) and repeat the tripping test beginning with the positive half-wave at 0° and the negative half-wave at 180°. The longer of the two tripping times is decisive regard­ing the condition of the tested RCCB.
in the menu (this is selected automatically
N
Set Parameter – Start with Positive or Negative Half-Wave
50 Gossen Metrawatt GmbH
Page 51
Set Parameter – 5 Times Nominal Current
Note
Note
5 times tripping current
S
U
IN
= RE×IN = 1 × 30 mA = 30 mV = 0.03 V
The following restrictions apply to the selection of tripping current multiples relative to nominal current: 500 mA: 1 ×, 2 × IN
Start Measurement

12.4 Testing Residual Current Circuit Breakers in TN-S Systems Connection

RCCBs can only be used in TN-S systems. An RCCB would not work in a TN-C system because PE is directly connected to the neutral conductor in the outlet (it does not bypass the RCCB). This means that residual current would be returned via the RCCB and would not generate any differential current, which is required in order to trip the RCCB.
As a rule, the display for touch voltage is also 0.1 V, because the nominal residual current of 30 mA together with minimal loop resistance result in a very small voltage value:
In both cases tripping time must be between 10 ms (minimum delay time for type G RCCBs!) and 40 ms.
Type G RCCBs with other nominal residual current values must be tested with the corresponding parameter setting under menu item I
. In this case as well, the limit value must be appropriately
N
adjusted.
The RCD parameter setting for selective RCCBs is not suitable for type G RCCBs.
Gossen Metrawatt GmbH 51
Page 52
12.5 Testing of RCD Protection in IT Systems with High Cable
Note
System type:
I
N
I
F
or
Time to trip
6mA 60 mA 200 mA
10.0 s
0.3 s
0.1 s
Capacitance (e.g. In Norway)
The desired system type (TN/TT or IT) can be selected for RCD test type U
A probe is absolutely essential for measurement in IT systems, because touch voltage U not otherwise be measured.
After selecting the IT system setting, connection with probe is selected automatically.
(IN, ta), and for earthing measurement (RE).
IN
which occurs in these systems can-
IN
Set Parameter – Select System Type
Start Measurement

12.6 Testing of 6 mA Residual Current Devices RDC-DD/RCMB (RDC-DD: PROFITEST MXTRA and PROFITEST MTECH+ only)

DIN VDE 0100-722 (Requirements for special installations or locations – Supplies for electric vehicles) specifies that all outlets for charging electric vehicles must be protected by a separate residual current device (RCD). Furthermore, additional protection is required for multiphase charging with smooth DC fault current. Either a type B RCD, an RDC-DD (residual direct current detecting device) or an RCMB (residual current monitoring module) can be used to this end.
Select Measuring Function
Set Parameter – Type RDC
Set Parameter – Time to Trip
The RDC-DD is tested with nominal residual currents of 6 to 200 mA.
Start Measurement
52 Gossen Metrawatt GmbH
Page 53
Set Parameter – Type RCMB
Note
Time to trip
300 mA 6mA 60 mA
0.04 s
10.0 s
0.3 s
Set Parameter – Time to Trip
The RCMB is tested with nominal residual currents of 6 to 300 mA.
Start Measurement
Gossen Metrawatt GmbH 53
Page 54
13 Testing of Breaking Requirements for Overcurrent Protective Devices, Measurement of Loop Impedance
Note
Note
Note
Note
Z
L-PE
Start
t1 t3
Measure
t2
Operation
RCD Disabled!
t
I
F
/mA
Suppression of RCCB tripping for RCCBs which are sensitive to pulsed current
and Determination of Short-Circuit Current (ZL-PE and ISC Functions)
Testing of overcurrent protective devices includes visual inspec­tion and measurement.
Measuring Method
Loop impedance Z is ascertained in order to determine if the breaking requirements for protective devices have been fulfilled.
Loop impedance is the resistance within the current loop (utility station – phase conductor – protective conductor) when a short­circuit to an exposed conductive part occurs (conductive connec­tion between phase conductor and protective conductor). Short­circuit current magnitude is determined by the loop impedance value. Short-circuit current I value set forth by IEC 60364, so that reliable breaking of the pro­tective device (fuse, automatic circuit breaker) is assured.
The measured loop impedance value must therefore be less than the maximum permissible value.
Tables containing permissible display values for loop impedance and minimum short-circuit current display values for ampere rat­ings for various fuses and circuit breakers can be found in the help texts and in section 25 from page 100. Maximum device error in accordance with IEC 61557 / EN 61557 has been taken into consideration in these tables. See also section 13.2.
In order to measure loop impedance Z test current of 3.7 to 7 A (60 to 550 V) depending on line voltage and line frequency. At 16 Hz, The test has a duration of no more than 1200 ms.
is measured and short-circuit current ISC
L-P E
may not fall below a predetermined
SC
, the instrument uses a
L-P E
If the limit value for touch voltage is exceeded during this mea­surement process (> 50 V), safety shutdown occurs for Germany (65 V applies for Austria – standard: ÖVE/ÖNORM E 8001-1 section 5.3). The shutdown value can be adjusted within a range of 25 to 65 V (see section 10.8).
The test instrument calculates short-circuit current ISC based on measured loop impedance current calculation is made with reference to nominal line voltage for line voltages which lie within the nominal ranges for 120, 230 and 400 V systems. This also applies between phases L-L at 500 V. If line voltage does not lie within these nominal ranges, the instrument calculates short-circuit current I vailing line voltage and measured loop impedance Z
Z
and line voltage. Short-circuit
L-P E
based upon pre-
SC
L-P E
.
Select Measuring Function
Connection Schuko / 3-Pole Adapter
Connection 2-Pole Adapter
Loop impedance should be measured for each electrical circuit at the farthest point, in order to ascertain maxi­mum loop impedance for the system.
Observe national regulations, e.g. the necessity of con­ducting measurements without regard for RCCBs in Austria.
3-Phase Connections
Measurement of loop impedance to earth must be performed at all three phase conductors (L1, L2, and L3) for the testing of over­current protective devices at three phase outlets.

13.1 Measurements with Suppression of RCD Tripping (PROFITEST MTECH+, PROFITEST MXTRA only)

The test instruments make it possible to measure loop impedance in TN systems with type A , F and AC 100, 300, 500 mA nominal residual current).
The test instrument generates a direct current to this end, which saturates the RCCB’s magnetic circuit. The test instrument then superimposes a measuring cur­rent which only demonstrates half­waves of like polar­ity. The RCCB is no longer capable of detecting this mea­suring current and is consequently not tripped during measurement.
A four conductor measuring cable is used between the instru­ment and the test plug. Cable and measuring adapter resistance is automatically compensated during measurement and does not affect measurement results.
RCCBs (10, 30,
Loop impedance measurement in accordance with the procedure for the suppression of RCCB tripping is only possible with type A and F RCDs.
Bias Magnetization
54 Gossen Metrawatt GmbH
Only AC measurements can be performed with the 2­pole adapter. Suppression of RCD tripping by means of bias magnetization with direct current is only possible via a country-specific plug insert, e.g. SCHUKO, or the 3­pole adapter (neutral conductor N required).
Page 55
13.1.1 Measurement with Positive Half-Waves (PROFITEST
Z
L-PE
Nom. current: 2 … 160 A, 9999 A
Tripping characteristics:
Diameter*: 1.5 … 70 mm
Cable type*: NY…, H07…
Number of wires*: 2 … 10-wire
A, B/L, C/G, D, E, H, K, GL/GG & factor
Sinusoidal
15 mA sinusoidal
Waveform:
DC-L and positive half-wave
Touch voltage:
DC-H and positive half-wave
2-pole
Measurement with country-specific
plug insert (e.g. Schuko)
Note
Selecting the test probe, as well as Lx-PE reference or AUTO, is only relevant with regard to report generation.
Semiautomatic measurement
See also section 10.9 regard­ing the AUTO parameter.
Polarity Selection
measurement
MTECH+, PROFITEST MXTRA only)
Measurement by means of half-waves plus direct current makes it possible to measure loop impedance in systems which are equipped with RCCBs.
In the case of DC measurement with half-waves, selection can be made between two variants:
DC-L: Reduced bias current
but faster measurement as a result
DC-H: Higher bias current providing more reliability with regard
to non-tripping of the RCD
Select Measuring Function
Set Parameters
Start Measurement
* Parameters used for report generation and do not influence the mea-
surement
Sinusoidal (full-wave) Setting for circuit without RCD 15 mA sinusoidal Setting for motor protection switch only
DC + half-wave Setting for circuit with RCD
Gossen Metrawatt GmbH 55
with small nominal current
Semiautomatic measurement (conductor reference change)

13.2 Evaluation of Measured Values

The maximum permissi-
ble loop impedance Z
which may be dis-
PE
played after allowance has been made for the instrument’s maximum measuring and intrinsic uncertainties (under nor­mal measuring condi­tions) can be determined with the help of Table 1 on page 100. Intermedi­ate values can be inter­polated.
The maximum permissi­ble nominal current for the protective device (fuse or circuit breaker) for a line voltage of 230 V after allowance has been made for maximum measuring error can be determined with the help of Table 6 on page 101 based on measured short-circuit current (corresponds to IEC 60364-6).
L-
Page 56
Special Case: Suppressing Display of the Limit Value
Limit Value:
ISC < Limit Value
UL R
L
The limit value cannot be ascertained. The inspec­tor is prompted to evalu­ate the measured val­ues himself, and to acknowledge or reject them with the help of the softkeys.
Measurement passed:
. key
Measurement failed: key
The measured value can only be saved after it has been evaluated.
13.3 Settings for Calculating Short-Circuit Current – Parameter I
Short-circuit current ISC is used to test shutdown by means of an overcurrent protective device. In order for an overcurrent protec­tive device to be tripped on time, short-circuit current ISC must be greater than tripping current Ia (see table 6 in section 25.1). The variants which can be selected with the “Limits” key have the fol­lowing meanings:
I
:IaThe measured value displayed for ISC is used
SC
I
:Ia+% The measured value displayed for Z
SC
I
: 2/3 Z In order to calculate ISC, the measured value dis-
SC
I
:3/4 ZZ
SC
SC
without any correction to calculate Z
rected by an amount equal to the test instrument’s
L-P E
L-P E
.
is cor-
measuring and intrinsic uncertainties in order to calculate I
played for Z sponding to all possible deviations (these are
.
SC
is corrected by an amount corre-
L-PE
defined in detail by IEC 60364-6 as Z
2/3 × U0/Ia).
s(m)
3/4 × U0/I
s(m)
a
Z Loop impedance I
Short-circuit current
SC
U Momentary voltage at the test probes, “UN” is displayed if
U
deviates from nominal voltage by 10%
max.
f Frequency of the applied voltage,
“fN” is displayed if frequency f max. deviates from nominal frequency by 1%
Tripping current
I
a
(see data sheet for circuit breakers / fuses)
% Test instrument intrinsic error
56 Gossen Metrawatt GmbH
Page 57
14 Measuring Supply Impedance (Z
Z
L-N
Diameters: 1.5 … 70 mm
Cable type: NY…, H07…
Number of wires: 2 … 10-wire
Tripping characteristics:
A, B/L, C/G, D, E, H, K, GL/GG & factor
Semiautomatic Measurement
See also section 10.9 regard­ing the AUTO parameter.
L-PE relationships are not possible here.
Polarity Selection
Limit Value:
ISC < Limit Value
UL R
L
I
SC
Function)
L-N
Measuring Method (internal line resistance measurement)
Supply impedance Z method used for loop impedance Z
54). However, the current loop is completed via neutral conductor N rather than protective conductor PE as is the case with loop impedance measurement.
is measured by means of the same
L-N
(see section 13 on page
L-P E
Select Measuring Function
Connection Schuko
Connection 2-Pole Adapter
Set Parameters
Settings for Calculating Short-Circuit Current – Parameter I
SC
Short-circuit current ISC is used to test shutdown by means of an overcurrent protective device. In order for an overcurrent protec­tive device to be tripped on time, short-circuit current I greater than tripping current I variants which can be selected with the “Limits” key have the fol-
(see table 6 in section 25.1). The
a
must be
SC
lowing meanings: I
:IaThe measured value displayed for ISC is used
SC
I
:Ia+% The measured value displayed for Z
SC
without any correction to calculate Z
rected by an amount equal to the test instrument’s
L-N
L-N
.
is cor-
measuring and intrinsic uncertainties in order to calculate I
I
: 2/3 Z In order to calculate ISC, the measured value dis-
SC
played for Z sponding to all possible deviations (these are
.
SC
is corrected by an amount corre-
L-N
defined in detail by IEC 60364-6 as Z
2/3 × U0/Ia).
I
:3/4 ZZ
SC
s(m)
3/4 × U0/I
s(m)
a
Gossen Metrawatt GmbH 57
back and forth between the country-specific plug insert, e.g. SCHUKO, and the 2-pole adapter. The selected connection type is displayed inversely (white on black).
Press the softkey shown at the left in order to switch
Z Loop impedance I
Short-circuit current
SC
U Momentary voltage at the test probes, “UN” is displayed if
U
deviates from nominal voltage by 10%
max.
f Frequency of the applied voltage,
“fN” is displayed if frequency f
deviates from nominal
max.
frequency by 1% Tripping current
I
a
(see data sheet for circuit breakers / fuses)
% Test instrument intrinsic error
Page 58
Start Measurement
Display of U
(UN / fN)
L-N
If the measured voltage value lies within a range of 10% of the respective nominal line voltage of 120, 230 or 400 V, the respec­tively corresponding nominal line voltage is displayed. In the case of measured values outside of the 10% tolerance, the actual measured value is displayed.
Displaying the Fuse Table
After measurement has been performed, permissible fuse types can be displayed by pressing the HELP key.
The table shows maximum permissible nominal current depen­dent on fuse type and breaking requirements.
Key: I
= breaking current, ISC = short-circuit current, IN = nominal
a
current, t
58 Gossen Metrawatt GmbH
= time to trip
a
Page 59

15 Earthing Resistance Measurement (Function RE)

Note
Attention!
!
Note
Start
t1
t3
Measure
t2
Operation
RCD Disabled!
t
I
F
/mA
Suppression of RCCB tripping for RCCBs which are sensitive to pulsed current
Earthing resistance RE is important for automatic shutdown in system segments. It must have a low value in order to assure that high short-circuit current flows and the system is shut down reli­ably by the RCCB in the event of a fault.
Test Se tu p
Earthing resistance (RE) is the sum of the earth electrode’s dissi­pation resistance and earth conductor resistance. Earthing resis­tance is measured by applying an alternating current via the earth conductor, the earth electrode and dissipation resistance. This current, as well as voltage between the earth electrode and a probe, are measured.
The probe is connected to the probe connector socket (17) with a 4 mm contact protected plug.
Direct measurement with probe (mains powered earthing mea­surement)
Direct measurement of earthing resistance RE is only possible within a measuring circuit which includes a probe. However, this means that the probe and reference earth must be of like poten­tial, i.e. that they are positioned outside of the potential gradient area. The distance between the earth electrode and the probe should be at least 20 m.
Measurement without probe (mains powered earthing measure­ment)
In many cases, especially in extremely built-up areas, it’s difficult, or even impossible, to set a measuring probe. In such cases, earthing resistance can be measured without a probe. In this case, however, the resistance values for the operational earth electrode RB and phase conductor L are also included in the measurement results.
Measuring method (with probe) (mains powered earthing mea­surement)
The instrument measures earthing resistance RE by means of the ammeter-voltmeter test. Resistance RE is calculated from the quotient of voltage UE and current I The test current which is applied to earthing resistance is con­trolled by the instrument (see section 5.5, “Technical Data”, on page 10 for pertinent values).
A voltage drop is generated which is proportional to earthing resistance.
Gossen Metrawatt GmbH 59
where UE is between the earth electrode and the probe.
E
Measurement cable and measuring adapter resistance are compensated automatically during measurement and have no effect on measurement results.
If dangerous touch voltages occur during measurement (> 50 V), the measurement is interrupted and safety shut­down occurs.
Probe resistance does not affect measurement results and may be as high as 50 k.
The probe is part of the measuring circuit and may carry a current of up to 3.5 mA in accordance with IEC 61557 / EN 61557.
Measurement with or without earth electrode voltage depending upon entered parameters and the selected type of connection:
RANGE Connection Measuring
Functions
xx / xx k
10 / U
xx / xx k *
* This parameter results in automatic selection of probe connection.
No probe measurement
*
E
No U Probe measurement
activated U
E
Probe measurement activated No U
Clamp measurement activated No U
measurement
E
is measured
measurement
E
measurement
E
Measuring method with suppression of RCD tripping (mains powered earthing measurement) (PROFITEST MTECH+, PROFITEST MXTRA only)
The test instrument makes it possible to measure earthing resis­tance in TN systems with type A , F and AC
RCCBs
(10, 30, 100, 300, 500 mA nominal residual current). The test instrument
generates a direct current to this end, which saturates the RCCB’s magnetic circuit. The test instrument then superimposes a measuring current which only demon­strates half-waves of like polarity. The RCCB is no longer capable of detect­ing this measuring current, and is con­sequently not tripped during measurement.
A four conductor measuring cable is used between the instru­ment and the test plug. Cable and measuring adapter resistance is automatically compensated during measurement and does not affect measurement results.
Bias Magnetization Only AC measurements can be performed with the 2­pole adapter. Suppression of RCD tripping by means of bias magnetization with direct current is only possible via a country-specific plug insert, e.g. SCHUKO, or the 3­pole adapter (neutral conductor N required).
Limit Values
Earthing resistance (earth coupling resistance) is determined pri­marily by the electrode’s contact surface and the conductivity of the surrounding earth.
The specified limit value depends on the type of electrical system and its shutdown conditions in consideration of maximum touch voltage.
Evaluating Measured Values
The maximum permissible displayed resistance values which assure that the required earthing resistance is not exceeded, and for which maximum device operating error has already been taken into consideration (at nominal conditions of use), can be determined with the help of Table 2 on page 100. Intermediate values can be interpolated.
Page 60

15.1 Earthing Resistance Measurement – Mains Powered

Note
RER
E
The following types of measurement and connection are possible:
2-wire measurement via 2-pole adapter
2-pole measurement via earthing contact plug (not possible in IT systems)
3-wire measurement via 2-pole adapter and
probe

15.2 Earthing Resistance Measurement – Battery Powered, “Battery Mode” (PROFITEST MPRO & PROFITEST MXTRA only)

The 5 following types of measurement and connection are possi­ble:
3-wire measurement via PRO-RE adapter
4-wire measurement via PRO-RE adapter
Selective measurement: 2-pole measurement with probe
and current clamp sensor At left in figure: 2-pole measuring
adapter for con­tacting PE and L measuring points
At right in figure: The PRO-Schuko
measuring adapter can be used as an alter­native.
Select Measuring Function
Select Operating Mode
Selective measurement with clamp meter (4­pole)
via PRO-RE adapter
2-clamp measurement via PRO-RE/2 adapter
Measurement of soil resistivity
via PRO-RE adapter
Figure at right: PRO-RE adapter for connect-
ing earth electrode, auxiliary earth electrode, probe and auxiliary probe to the test instrument for 3/4-pole measurement, selective measurement and measurement of soil resistivity
Figure at right: PRO-RE/2 measuring adapter as
accessory for connecting the E­Clip 2 generator clamp for 2-clamp measurement and earth loop resis­tance measurement.
E
The selected operating mode is displayed inversely: mains~ in white against a black background.
Special Case: Manual Measuring Range Selection (test current selection)
(R AUTO, R = 10 k (4 mA), 1 k (40 mA), 100 (0,4 A), 10 (3.7 … 7 A), 10 /U
When the measuring range is selected manually, accuracy val­ues are only valid starting at 5% of the upper limit range value (except for the 10 W range; separate display for small values).
)
E
Set Parameters
Measuring range: AUTO
10 k (4 mA), 1 k (40 mA), 100 (0.4 A), 10 (> 3.7 A) In systems with RCCBs, resistance or test current must be se­lected such that it is less than tripping current (½ I
Touch voltage: UL < 25 V, < 50 V, < 65 V, see section 10.8 re-
garding freely selectable voltage
Transformer ratio: depends on utilized current clamp sensorConnection: 2-pole adapter, 2-pole adapter + probe,
2-pole adapter + clamp meter
System type: TN or TTTest current waveform
See section 15.4 through section 15.6 regarding advisable parameters for the respective measurement and connection types.
N
).
Performing Measurements
See section 15.4 through section 15.6.
Select Measuring Function
Select Operating Mode
The selected operating mode is displayed inversely: white battery icon against black background.
Set Parameters
Measuring range: AUTO, 50 k, 20 kW, 2 kW, 200 W, 20 W Current clamp sensor transformer ratio:
1:1 (1 V/A,) 1:10 (100 mV/A), 1:100 (10 mV/A), 1:1000 (1 mV/A)
Connection: 3-pole, 4-pole, selective, 2-clamp, Distance d (for measuring
See section 15.7 through section 15.11 regarding advisable parameters for the respective measurement and connection types.
E
): xx m
(Rho)
E
Performing Measurements
See section 15.7 through section 15.11.
60 Gossen Metrawatt GmbH
Page 61
15.3
T
E
S
T
I
N
S
T
R
U
M
E
N
T
Ri
W
a
t
e
r
P
i
p
e
E
2
E
1
B
R
E
Limit Value:
RE > Limit Value
UL R
L

Earthing Resistance, Mains Powered – 2-Pole Measurement with 2-Pole Adapter or Country-Specific Plug (Schuko) without Probe

Key
R
Operational earth electrode
B
R
Earthing resistance
E
R
Internal resistance
i
R
Earthing resistance through equipotential bonding systems
X
R
Probe resistance
S
PAS Equipotential bonding busbar RE Overall earthing resistance (R
In the event that it’s impossible to set a probe, earthing resistance can be estimated by means of an “earth loop resistance measure­ment” without probe.
The measurement is performed exactly as described in section
15.4, “Earthing Resistance Measurement. Mains Powered – 3-Pole Measurement: 2-Pole Adapter with Probe”, on page 62. However, no probe is connected to the probe connector socket (17).
The resistance value R method also includes operational earth electrode resistance R
obtained with this measuring
ELoop
and resistance at phase conductor L. These values must be sub­tracted from the measured value in order to determine earthing resistance.
If conductors of equal cross section are assumed (phase conduc­tor L and neutral conductor N), phase conductor resistance is half as great as supply impedance Z conductor). Supply impedance can be measured as described in section 14 from page 57. In accordance with IEC 60364, the operational earth electrode R 2 ”.
must lie within a range of “0 to
B
//RE2//water pipe)
E1
(phase conductor + neutral
L-N
B
Set Parameters
Measuring range: AUTO, 10 k (4 mA), 1 k (40 mA), 100
(0.4 A), 10 (3.7 … 7 A). In systems with RCCBs, resistance or test current must be selected such that it is less than trip­ping current (½ I
N
).
Connection: 2-Pole adapterTouch voltage: UL < 25 V, < 50 V, < 65 VTest current waveshape: Sinusoidal (full-wave), 15 mA sinusoidal
(full-wave), DC offset and positive half-wave
System type: TN/TT, ITTransformer ratio: irrelevant in this case
Start Measurement
1) Measurement: Z
2) Measurement: Z
3) Calculation:
amounts to Ri = 2 × R
LN
amounts to R
L-PE
RE1 amounts to Z
L-PE
ELoop
– ½ × Z
L
, where RB = 0
L-N
The value for operational earth conductor resistance RB should be ignored in the calculation of earthing resistance, because it’s gen­erally unknown.
The calculated earthing resistance thus includes operational earth conductor resistance as a safety factor.
If the parameter is selected, steps 1 through 3 are executed automatically by the test instrument.
Select Measuring Function
Select Operating Mode
Gossen Metrawatt GmbH 61
Page 62

15.4 Earthing Resistance Measurement. Mains Powered – 3-Pole Measurement: 2-Pole Adapter with Probe

Note
T
E
S
T
I
N
S
T
R
U
M
E
N
T
W
a
t
e
r
P
i
p
e
SE
2
E
1
B
R
E1
U
Probe
I
----------------
=


R
E
Limit Value:
RE > Limit Value
UL R
L
Key
R
Operational earth electrode
B
R
Earthing resistance
E
R
Earthing resistance through equipotential bonding sys-
X
tems
R
Probe resistance
S
PAS Equipotential bonding busbar RE Overall earthing resistance (R
Measurement of R
E
//RE2//water pipe)
E1
Select Measuring Function
Select Operating Mode
Set Parameters
Measuring range: AUTO
10 k (4 mA), 1 k (40 mA), 100 (0.4 A), 10 (3.7 … 7 A) In systems with RCCBs, resistance or test current must be se­lected such that it is less than tripping current (½ I
N
).
Connection: 2-pole adapter + probeTouch voltage: UL < 25 V, < 50 V, < 65 V, see section 10.8 re-
garding freely selectable voltage
Test current waveshape:
Sinusoidal (full-wave), 15 mA sinusoidal (full-wave), DC offset and positive half-wave
System type: TN/TT, ITTransformer ratio: irrelevant in this case
Start Measurement
Connection
To be connected: 2-pole adapter and probe
62 Gossen Metrawatt GmbH
The following diagram appears if the 2-pole adapter is connected incor­rectly.
Page 63

15.5 Earthing Resistance Measurement, Mains Powered – Measuring Earth Electrode Potential (UE Function)

Note
T
E
S
T
I
N
S
T
R
U
M
E
N
T
Ri
W
a
t
e
r
P
i
p
e
SE
2
E
1
B
U
E
UNRE
R
ELoop
--------------------=
R
E
Limit Value:
RE > Limit Value
UL R
L
This measurement is only possible with a probe (see section
15.4). Earth electrode potential U the earth electrode between the earth electrode terminal and ref­erence earth if a short-circuit occurs between the phase conduc­tor and the earth electrode. The measurement of earth electrode potential is required by Swiss standard NIV/NIN SEV 1000.
is the voltage which occurs at
E
Measuring Method
In order to determine earth electrode potential the instrument first measures earth electrode loop resistance R ately thereafter earthing resistance R values and then calculates earth electrode potential with the fol­lowing equation:
The calculated value is displayed at the display panel.
. The instrument stores both
E
, and immedi-
ELoop
Select Measuring Function
Select Operating Mode Select Measuring Range
Set Parameters
Measuring range: 10 / UConnection: 2-pole adapter + probeTouch voltage: UL < 25 V, < 50 V, < 65 V, see section 10.8
regarding freely selectable voltage
Test current waveshape: sinusoidal only in this case (full-wave)!System type: TN/TT, ITTransformer ratio: irrelevant in this case
E
Start Measurement
Connection
The following diagram appears if the 2-pole adapter is connected incor­rectly.
To be connected: 2-pole adapter and probe
Gossen Metrawatt GmbH 63
Page 64
15.6 Earthing Resistance Measurement, Mains Powered – Selective Earthing Resistance Measurement with Current Clamp Sensor
T
E
S
T
I
N
S
T
R
U
M
E
N
T
W
a
t
e
r
P
i
p
e
SE
2
E
1
B
U
Probe
I
Clamp
-----------------



R
E
as Accessory
As an alternative to the conventional measuring method, measurement can also be performed with a current clamp sensor.
Key
R
Operational earth
B
R
Earthing resistance
E
R
Cable resistance
L
R
Earthing resistance through equipotential bonding
X
systems
R
Probe resistance
S
PAS Equipotential bonding busbar RE Overall earthing resistance (R
// RE2 // water pipe)
E1
Measurement without clamp: RE = RE1 // R
Measurement with clamp: RE = RE2 =
Select Measuring Function
Select Operating Mode
Set Parameters at Tester
Measuring range (test current selection):
1 k (40 mA), 100 (0.4 A), 10  (3.7 … 7 A) In the case of systems with RCCBs, the DC offset and positive half-wave (DC + ) functions can be selected (only in the 10 range and only with the METRAFLEX P300).
Connection: 2-pole adapter + clamp meter
After parameter selection: automatic setting to 10 measuring range and 1 V/A or 1000 mV/A transformer ratio
Touch voltage: UL < 25 V, < 50 V, < 65 V, for information regard-
ing freely selectable voltage see section 10.8
Test current waveshape:
Sinusoidal (full-wave), DC offset and
positive half-wave (DC + )
E2
System type: TN/TT, ITCurrent clamp sensor transformation ratio: see table below
Set Parameters at Current Clamp Sensor
Current clamp sensor measuring range: see table below
Select Measuring Range at the Current Clamp Sensor
Test Instrument METRAFLEX P300 Clamp Test Instrument
Parameters
Transformation Ratio
1:1
1 V / A
01:10
100 mV / A
1:100
10 mV / A
Switches Measuring
Range
3 A (1 V/A) 3A 0.5 … 100 mA
30 A (100 mV/A) 30 A 5 … 999 mA
300 A (10 mV/A) 300 A 0.05 … 10 A
Measuring Range
Important Instructions for Use of the Current Clamp Sensor
• Use only the METRAFLEX P300 or the Z3512A current clamp
Connection
To be connected: 2-pole adapter, clamp and probe
64 Gossen Metrawatt GmbH
sensor for this measurement.
• Read and adhere to the operating instructions for the METRAFLEX P300 current clamp sensor, as well as the safety precautions included therein.
• Observe direction of current flow (see arrow on the current clamp sensor).
• Use the clamp in the permanently connected state. The sen­sor may not be moved during measurement.
• The current clamp sensor may only be used at an adequate distance from powerful extraneous fields.
• Before use, always inspect the electronics housing, the con­nector cable and the current sensor for damage.
• In order to prevent electric shock, keep the surface of the METRAFLEX clean and free of contamination.
• Before use, make sure that the flexible current sensor, the con­nector cable and the electronics housing are dry.
Page 65
Start Measurement
Note
Note
In the event that you have changed the transformation ratio at the test instrument, a pop-up window appears indicating that this new setting also has to be entered to the connected current clamp sensor.
Note regarding currently selected trans­formation at the tester
ratio
RE RE
: Selective earthing resistance measured via clamp
Clamp
: Total earthing resistance measured via probe,
Probe
comparative value
The following diagram appears if the 2-pole adapter is connected incor­rectly.
Gossen Metrawatt GmbH 65
Page 66

15.7 Earthing Resistance Measurement, Battery Powered, “Battery Mode” – 3-Pole (PROFITEST MPRO & PROFITEST MXTRA only)

Note
PROFITEST MPRO / PROFITEST MXTRA
E
SH
20 m 20 m
S
HESE
R
E
3-Wire Method
Select Measuring Function
Select Operating Mode
The selected operating mode is displayed inversely: white battery icon against black background.
Set Parameters
Measuring range: AUTO, 50 k, 20 k, 2 k, 200 , 20 Connection: 3-pinTransformer ratio: irrelevant in this case
Earthing Resistance Measurement According to the 3-Wire Method
Distance d (for measuring
Start Measurement
): irrelevant in this case
E
Connection
Position the spikes for the probe and the auxiliary electrode at
least 20, respectively 40 meters from the electrode (see figure above).
Make sure that no excessively high contact resistances occur
between the probe and the ground.
Attach the PRO-RE adapter (Z501S) to the test plug.Connect the probe, the auxiliary electrode and the electrode
via the 4 mm banana plug sockets at the PRO-RE adapter. In doing so, observe labeling on the banana plug sockets. Terminal ES/P1 is not connected.
The resistance of the measurement cable to the earth electrode is incorporated directly into the measurement results.
In order to keep error caused by measurement cable resistance as small as possible, a short connector cable with large cross­section should be used between the earth electrode and terminal E for this measuring method.
The measurement cables must be well insulated in order to prevent shunting. In order to keep the influence of pos­sible coupling to a minimum, the measurement cables should not cross each other or run parallel to each other over any considerable distance.
66 Gossen Metrawatt GmbH
Page 67
15.8 Earthing Resistance Measurement, Battery Powered, “Battery Mode” – 4-Pole (
Note
Note
PROFITEST MPRO / PROFITEST MXTRA
S
HESE
E
SH
20 m 20 m
Earthing Resistance Measurement According to the 4-Wire Method
R
E
4-Wire Method
Select Measuring Function
Select Operating Mode
Set Parameters
Measuring range: AUTO, 50 k, 20 k, 2 k, 200 , 20 Connection: 4-pinTransformer ratio: irrelevant in this caseDistance d (for measuring
PROFITEST MPRO
The selected operating mode is displayed inversely: white battery icon against black background.
&
PROFITEST MXTRA
): irrelevant in this case
E
only)
The 4-wire method is used in the case of high cable resistance between the earth electrode and the instrument terminal.
The resistance of the cable between the earth electrode and the “E” terminal at the instrument is measured in this case.
Connection
Position the spikes
for the probe and the auxiliary electrode at least 20, respec­tively 40 meters from the electrode (see figure above).
Make sure that no excessively high contact resistances occur
between the probe and the ground.
Attach the PRO-RE adapter (Z501S) to the test plug.Connect the probes, the auxiliary electrode and the electrode
via the 4 mm banana plug sockets at the PRO-RE adapter. In doing so, observe labeling on the banana plug sockets.
In the case of the 4-wire method, the earth electrode is connected to the “E” and “ES” terminals with two separate measurement cables, the probe is connected to the “S” terminal and the auxiliary earth electrode is connected to the “H” terminal.
The measurement cables must be well insulated in order to prevent shunting. In order to keep the influence of pos­sible coupling to a minimum, the measurement cables should not cross each other or run parallel to each other over any considerable distance.
Start Measurement
Potential Gradient Area
Information regarding suitable positioning of the probe and the auxiliary earth electrode can be obtained by observing voltage characteristics ordissipation resistance in the ground.
The measuring current from the earth tester which flows via the earth electrode and the auxiliary earth electrode causes a given potential distribution in the form of a potential gradient area, (cf. figure “Voltage Curve in Homogeneous Earth between Earth Elec­trode E and Auxiliary Earth Electrode H” on page 68.) Resistance distribution is analogous to potential distribution.
Dissipation resistance of the earth electrode and the auxiliary earth electrode differs as a rule. The potential gradient area and the resistance gradient area are thus not symmetrical.
Dissipation Resistance of Small Scope Earth Electrodes
The arrangement of the probe and the auxiliary earth electrode are very important for correct determination of the dissipation resistance of earth electrodes. The probe must be positioned between the earth electrode and the auxiliary earth electrode within the so-called neutral zone (ref­erence earth) (cf. figure “Probe Distance S Outside of the Overlap­ping Potential Gradient Areas on the Perpendicular Bisector of Earth Electrode E and Auxiliary Earth Electrode H” on page 68.) The voltage or resistance curve is thus nearly horizontal within the neutral zone.
Proceed as follows in order to select suitable probe and auxiliary earth electrode resistances:
Drive the auxiliary earth electrode into the ground at a dis-
tance of roughly 40 meters from the earth electrode.
Position the probe halfway between the earth electrode and
the auxiliary earth electrode and determine earthing resis­tance.
Reposition the probe 2 3 m closer to the earth electrode,
and then 2 3 m closer to the auxiliary earth electrode and measure earthing resistance in each position.
If all 3 measurements result in the same measured value, this is the correct earthing resistance. The probe is in the neutral zone.
Gossen Metrawatt GmbH 67
Page 68
However, if the three measured values for earthing resistance dif-
d = distance from electrode to aux. electrode E = earth electrode H = auxiliary earth electrode I = measuring current K = neutral zone (reference earth) UE= earth potential RE= UE / I = earthing resistance
= potential
I
I
d
E
H
U
E
K
Voltage Curve in Homogeneous Earth between Earth Electrode E and Auxiliary Earth Electrode H
E = electrode location H = aux. electrode location S = probe location
S
HE
Probe Distance S Out­side of the Overlapping Potential Gradient Areas on the Perpendicular Bisector of Earth Elec­trode E and Auxiliary Earth Electrode H
Curve I (KI) Curve II (KII)
mWmW
5 10 15 20 25 30 40 60 80
100
0.9
1.28
1.62
1.82
1.99
2.12
2.36
2.84
3.68 200
10 20 40 60
80 100 120 140 160 200
0.8
0.98
1.60
1.82
2.00
2.05
2.13
2.44
2.80 100
S1, S2 = inflection points KI = curve I KII = curve II
S1, S2 = inflection points KI = curve I KII = curve II
S
1
S
2
K I
K II
4
3
2
1
0
10 20 30 40 50 60 70 80 90 100 m KI 20 40 60 80 100 120 140 160 180 200 m KII
5
R
A/H
R
A/E
0 0
S
HESE
Earthing Resistance Measurement for a Large Scope Earthing System
fer from each other, either the probe is not located in the neutral zone, or the voltage or resistance curve is not horizontal at the point at which the probe has been inserted.
Correct measurements can be obtained in such cases by either increasing distance between the earth electrode and the auxiliary earth electrode, or by moving the probe to the perpendicular bisector between the earth electrode and the auxiliary earth elec­trode (see figure below). When the probe is moved to the perpen­dicular bisector, its location is removed from the sphere of influ­ence of the two potential gradient areas caused by the earth elec­trode and the auxiliary earth electrode.
Measured resistance values are displayed as a table, and then
plotted graphically as depicted in “Earthing Resistance Mea­surement for a Large Scope Earthing System” on page 68. (Curve I).
If a line parallel to the abscissa is drawn through inflection point S1, this line divides the resistance curve into two parts. Measured at the ordinate, the bottom part results in sought dissi­pation resistance of the earth electrode R dissipation resistance of the auxiliary earth electrode R With a measurement setup of this type, dissipation resistance of
, and the top value is
A/E
A/H
.
the auxiliary earth electrode should be less than 100 times the dissipation resistance of the earth electrode.
In the case of resistance curves without a well-defined horizontal area, measurement should be double checked after repositioning the auxiliary earth electrode. This additional resistance curve must be entered to the first diagram with a modified abscissa scale such that the two auxiliary earth electrode locations are superim­posed. The initially ascertained dissipation resistance value can be checked with inflection point S2.
Notes Regarding Measurement in Difficult Terrain
In extremely unfavorable terrain (e.g. sandy soil after a lengthy period without rain), auxiliary earth electrode and probe resistance can be reduced to permissible values by watering the ground around the auxiliary earth electrode and the probe with soda water or salt water. If this does not suffice, several earth spikes can be parallel con­nected to the auxiliary earth electrode.
In mountainous terrain or in the case of very rocky subsoil where earth spikes cannot be driven into the ground, wire grates with a mesh size of 1 cm and a surface area of about 2 square meters can be used. These grates are laid flat onto the ground, are wet­ted with soda water or salt water and may also be weighted down with sacks full of moist earth.
Dissipation Resistance of Large Scope Earthing Systems
Significantly large distances to the probe and the auxiliary earth electrode are required for measuring large scope earthing sys­tems. Calculations are based on 2½ or 5 times the value of the earthing system’s largest diagonal. Large scope earthing systems of this sort often demonstrate dis­sipation resistances of only a few ohms, which makes it especially important to position the measuring probe within the neutral zone. The probe and the auxiliary earth electrode should be positioned at a right angle to the direction of the earthing system’s largest lin­ear expansion. Dissipation resistance must be kept small. If nec­essary, several earth spikes must be used at a distance of 1 to 2 m from each other and connected to this end.
However, in actual practice large measuring distances are fre­quently not possible to due difficult terrain. If this is the case, proceed as shown in figure “Earthing Resis­tance Measurement for a Large Scope Earthing System” on page 68.
Auxiliary earth electrode H is positioned as far as possible
from the earthing system.
The area between the earth electrode and the auxiliary earth
electrode is sampled with the probe in equal steps of 5 meters each.
68 Gossen Metrawatt GmbH
Page 69
15.9 Earthing Resistance Measurement, Battery Powered, “Battery Mode” – Selective (4-pole)
Note
PROFITEST MPRO /PROFITEST MXTRA
R
E
with Current Clamp Sensor and PRO-RE Measuring Adapter as Accessory (PROFITEST MPRO & PROFITEST MXTRA only)
General
Set Parameters at Tester
Measuring range: 200
After switching to selective measurement, the AUTO mea- suring range is activated automatically if a measuring range of greater than 200 had been selected.
Connection type: selectiveCurrent clamp sensor transformer ratio:
1:1 (1 V/A,) 1:10 (100 mV/A), 1:100 (10 mV/A)
Distance d (for measuring
): irrelevant in this case
E
Set Parameters at Current Clamp Sensor
Current clamp sensor measuring range: see table below
Select Measuring Range at the Current Clamp Sensor
When measuring earthing resistance in systems with several par­allel connected earth electrodes, total resistance of the earthing system is measured.
Two earth spikes (auxiliary earth electrode and probe) are set for this measurement. Measuring current is fed between the earth electrode and the auxiliary earth electrode and voltage drop is measured between the earth electrode and the probe.
The current clamp is positioned around the earth electrode to be measured, and thus only that portion of the measuring current which flows through the earth electrode is measured.
Test Instrument Z3512A Clamp
Parameters
Transformation Ratio
1:1
1 V / A
01:10
100 mV / A
1:100
10 mV / A
Switches Measuring Range
1 A / × 1 1A
10 A / × 10 10 A
100 A / × 100 100 A
Important Instructions for Use of the Current Clamp Sensor
Connection
Position the spikes
for the probe and the auxiliary electrode at least 20, respec­tively 40 meters from the electrode (see figure above).
Make sure that no excessively high contact resistances occur
between the probe and the ground.
Attach the PRO-RE adapter (Z501S) to the test plug.Connect the probes, the auxiliary electrode and the electrode
via the 4 mm banana plug sockets at the PRO-RE adapter. In doing so, observe labeling on the banana plug sockets.
Connect the Z3512A current clamp sensor to jacks (15) and
(16) at the test instrument.
Attach the current clamp sensor to the earth electrode.
• Use only the Z3512A current clamp sensor for this measure­ment.
• Use the clamp in the permanently connected state. The sen­sor may not be moved during measurement.
• The current clamp sensor may only be used at an adequate distance from powerful extraneous fields.
• Make sure that the current clamp sensor’s connector cable is laid separate from the probe cables to the greatest possible extent.
Start Measurement
Select Measuring Function
Select Operating Mode
The selected operating mode is displayed inversely: white battery icon against black background.
Gossen Metrawatt GmbH 69
Page 70
15.10 Earthing Resistance Measurement, Battery Powered, “Battery Mode” – Ground Loop Measurement
Note
PROFITEST MPRO / PROFITEST MXTRA
R
E
(with current clamp sensor and transformer, and pro-re measuring adapter as accessory) (PROFITEST MPRO & PROFITEST MXTRA only)
2-Clamp Measuring Method
Select Measuring Function
Select Operating Mode
The selected operating mode is displayed inversely: white battery icon against black background.
Set Parameters at Tester
Measuring range: in this case always AUTO
After selecting 2-clamp measurement, switching to the AUTO range takes place automatically. It is then no longer
In the case of earthing sys­tems which consist of sev­eral earth electrodes (R1 … Rx) which are connected to each other, earthing resis­tance of a single electrode (Rx) can be ascertained with the help of 2 current clamps without disconnecting Rx or using spikes.
This measuring method is especially well suited for buildings or systems for which probes and auxiliary earth electrodes cannot be used, or where it’s impermissible to disconnect earth electrodes.
Furthermore, this “spike-free” measurement is performed as one of three measurements for lightning protection systems, in order to determine whether or not current can be dissipated.
Figure at right: PRO-RE/2 measuring adapter as accessory for connecting the E­Clip 2 generator current clamp
Connection
Connection: 2 clampsCurrent clamp sensor transformer ratio:
Distance d (for measuring
Set Parameters at Current Clamp Sensor
Current clamp sensor measuring range: see table below
Select Measuring Range at the Current Clamp Sensor
Important Instructions for Use of the Current Clamp Sensor
• Use only the Z3512A current clamp sensor for this measure-
• Use the clamp in the permanently connected state. The sen-
• The current clamp sensor may only be used at an adequate
• Make sure that the connector cables from the two clamps are
possible to change the range!
1:1 (1 V/A), 1:10 (100 mV/A), 1:100 (10 mV/A)
): irrelevant in this case
E
Test Instrument Z3512A Clamp
Parameters
Transformation Ratio
1:1
1 V / A
01:10
100 mV / A
1:100
10 mV / A
ment.
sor may not be moved during measurement.
distance from powerful extraneous fields.
laid separate from each other to the greatest possible extent.
Switches Measuring Range
1 A / × 1 1A
10 A / × 10 10 A
100 A / × 100 100 A
Start Measurement
No probes or auxil-
The earth electrode is not disconnected.Attach the PRO-RE/2 adapter (Z502T) to the test plug.Connect the E-Clip 2 generator clamp (current clamp trans-
Connect the Z3512A current clamp sensor to jacks 15 and 16
Attach the 2 clamps to an earth electrode (earth spike) at dif-
70 Gossen Metrawatt GmbH
iary earth electrodes are required.
former) via the 4 mm safety plugs at the PRO-RE/2 adapter.
at the test instrument.
ferent heights with a clearance of at least 30 cm.
Page 71
15.11 Earthing Resistance Measurement, Battery Powered, “Battery Mode” – Measurement of Soil Resistivity E
Note
ES
ESH
dd d
Measurement of Soil Resistivity
R
E
+E (%)
10
20
30
-10
-20
-30
Jan. March May July Sept. Nov.
Soil Resistivity E Relative to Season Without the Effects of Precipitation (earth electrode depth < 1.5 m)
(PROFITEST MPRO & PROFITEST MXTRA only)
General
Select Measuring Function
Select Operating Mode
The selected operating mode is displayed inversely: white battery icon against black background.
Set Parameters
The determination of soil resistivity is necessary for the planning of earthing systems. Reliable values need to be ascertained which take even the worst possible conditions into account (see „Geo­logic Evaluation“ on page 71).
Soil resistivity is decisive with regard to the magnitude of an earth electrode’s dissipation resistance. Soil resistivity can be measured with the test instrument using the method according to Wenner.
Four earth spikes of greatest possible length are driven into the ground in a straight line at distance d from one another, and are connected to the earth tester (see figure above). The earth spikes usually have a length of 30 to 50 cm. Longer earth spikes can be used for soil which demonstrates poor con­ductivity (sandy soil etc.). The depth to which the earth spikes are driven into the ground may not exceed one twentieth of distance d.
Measuring range: AUTO, 50 k, 20 k, 2 k, 200 , 20 Connection:
(Rho)
E
Transformer ratio: irrelevant in this caseDistance d for measurement of
: adjustable from 0.1 to 999
E
Start Measurement
Erroneous measurement may result if piping, cables or other underground metal conduits run parallel to the measuring setup.
Soil resistivity is calculated as follows:
E=2 d R Where: = 3.1416 d = distance in m between two earth spikes R = ascertained resistance value in (this value corresponds to R
determined with the 4-wire method)
E
as
Connection
Position the spikes
for the probe and the auxiliary electrode at equal distances (see figure above).
Make sure that no excessively high contact resistances occur
between the probe and the ground.
Attach the PRO-RE adapter (Z501S) to the test plug.Connect the probes, the auxiliary electrode and the electrode
via the 4 mm banana plug sockets at the PRO-RE adapter. In doing so, observe labeling on the banana plug sockets.
Geologic Evaluation
Except in extreme cases, the ground is measured down to a depth which is roughly equal to probe distance d. And thus it’s possible to arrive at conclusions regarding the ground’s stratification by varying probe distance. Layers which are highly conductive (water table) into which earth electrodes should be installed, can thus be discovered within a region which is otherwise not very conductive.
Soil resistivity is subject to considerable fluctuation which may be due to various causes such as porosity, moisture penetration, concentration of dissolved salts in the ground water and climatic fluctuation.
Characteristic values for and the soil’s negative temperature coefficient) can be approxi­mated quite closely by means of a sinusoidal curve.
relative to season (soil temperature
E
Gossen Metrawatt GmbH 71
Page 72
A number of typical soil resistivity values for various types of
R
A
2
E
I
----------- -
=
R
A
E
I
-----
=
R
A
2
E
3D
----------- -
=
D 1,13 F
2
=
R
A
2
E
2D
----------- -
=
D 1,13 F
2
=
R
A
2
E
4,5 a
------------
=
R
A
E
D
--------- -
=
D 1,57 J
3
=
ground are summarized in the following table.
Type of Soil Soil Resistivity
Marshy ground 8 60 Arable soil, loamy and clayey soil, moist gravel 20 300 Moist sandy soil 200 600 Dry sandy soil, dry gravel 200 2000 Rocky ground 300 8000 Rock 10
[m]
E
4
10
10
Calculating Dissipation Resistance
Formulas for calculating dissipation resistance for common types of earth electrodes are included in this table. These rules of thumb are entirely adequate for actual practice.
Number Earth Electrode Rule of Thumb Subsidiary Variable
1
2
3 Ring earth electrode
4 Mesh earth electrode
5 Ground plate
6
Earth strip
(star type earth electrode)
Earth rod
(buried earth electrode)
Hemispherical earth
electrode
RA= dissipation resistance () 
= soil resistivity (m)
E
I = length of the earth electrode (m) D = diameter of a ring earth electrode, diameter of the equivalent surface
area of a mesh earth electrode or diameter of a hemispherical earth electrode (m)
F = surface area (sq. meters) of the enclosed surface or a ring or mesh
earth electrode
a = Edge length (m) of a square ground plate; a is replaced with the fol-
lowing for rectangular plates:
bxc, where b and c are the two
sides of the rectangle.
J = volume (cubic meters) of an individual foundation footing
72 Gossen Metrawatt GmbH
Page 73

16 Measurement of Insulation Resistance

Attention!
!
Note
Note
Note
R
ISO
Voltage type: constant
Test voltag e: 15, 50, 100, 250, 325, 500, 1000 V
Voltage type: rising/ramp
Earth leakage resistance:
xxx V*
2-pole measurement (selection relevant for report genera- tion only): measurements between Lx-PE / N-PE / L+N-PE / Lx-N / Lx-Ly / AUTO*
where x, y = 1, 2, 3
Limit Value:
I > I
Limit
U
INS
(U
INS
)
STOP
Lower limit value:
U
INS
(U
INS
)
Enterable range:
> 40 V … < 999 V
Upper limit value:
Limit Value:
R
INS
< Limit Value
UL R
L
U
INS
(U
INS
)
Insulation resistance may only be measured at voltage­free devices.

16.1 General Select Measuring Function

Connection
2 pole adapter or test plug
Breakdown Current for Ramp Function
In order to suppress the influence of parallel capacitances on the device under test when measurement is started, shutdown at the respective breakdown current I not occur until a minimum voltage of 5 V is exceeded.
Limit Values for Breakdown Voltage
lim
does
Set Parameters
* Freely adjustable voltage (see section 10.8)
Polarity Selection
* AUTO parameter (see section 10.9)
Gossen Metrawatt GmbH 73
The test instrument always measures insulation between the L and PE terminals. N and PE must be interrupted for systems without RCD.
Checking Measurement Cables Before Measurements
Before performing insulation measurement, the test probes on the measurement cables should be short-cir­cuited in order to assure that the instrument displays a value of less than 1 k. In this way, incorrect connection can be avoided and interrupted measurement cables can be detected.
Limit Values for Constant Test Voltage
Test Voltage
A test voltage which deviates from nominal voltage, and is usually lower, can be selected for measurements at sensitive compo­nents, as well as systems with voltage limiting devices.
Voltage Type
The “U detect weak points in the insulation, as well as to determine response voltage for voltage limiting components. After pressing the ON/START ▼ key, test voltage is continuously increased until the specified nominal voltage U which is measured at the test probes during and after testing. After measurement, this voltage drops to a value of less than 10 V (see section entitled “Discharging the Device Under Test”).
Insulation measurement with rising test voltage is ended:
• As soon as specified maximum test voltage U the measured value is stable
or
• As soon as specified maximum test voltage is reached (e.g. after sparkover occurs at breakdown voltage).
Specified maximum test voltage U down voltage which occurs is displayed for U
The constant test voltage function offers two options:
rising test voltage function (ramp function) is used to
INS
is reached. U is the voltage
N
is reached and
N
or any triggering or break-
N
INS
.
Page 74
• After briefly pressing the ON/START ▼ key, specified test volt-
Note
Note
age U
is read out and insulation resistance R
N
As soon as the measured value is stable (settling time may be
is measured.
INS
several seconds in the case of high cable capacitance values), measurement is ended and the last measured values for R and U the test probes during and after testing. After measurement,
are displayed. U is the voltage which is measured at
INS
INS
this voltage drops to a value of less than 10 V (see section entitled “Discharging the Device Under Test”).
or
As long as you press and hold the ON/START ▼ key, test voltage U
is applied and insulation resistance R
N
not release the key until the measured value has settled in
is measured. Do
INS
(settling time may be several seconds in the case of high cable capacitance values). Voltage U, which is measured during testing, corresponds to voltage U START ▼ key, measurement is ended and the last measured values for R less than 10 V after measurement (see the section entitled
INS
and U
are displayed. U drops to a value of
INS
. After releasing the ON/
INS
“Discharging the Device Under Test”.
Pole Selection Report Entry
The poles between which testing takes place can only be entered here for reporting purposes. The entry itself has no influence on the actual polarity of the test probes or the pole selection.
Limits – Setting the Limit Value
The limit value for insulation resistance can be set as desired. If measurement values occur which are below this limit value, the red U
0.5 M to 10 M is available. The limit value is displayed above
LED lights up. A selection of limit values ranging from
L/RL
the measured value.
Start Measurement – Rising Test Voltage (ramp function)
Press briefly:
• The selected voltage limit is reached
• The selected current limit is reached or
• Sparkover occurs (spark gaps)
Differentiation is made amongst the following three procedures for insulation measurement with ramp function:
Testing overvoltage limiters or varistors and determining their tripping voltage:
– Select maximum voltage such that the anticipated breakdown
voltage of the device under test is roughly one third of this value (observe manufacturer’s data sheet if applicable).
– Select the current limit value in accordance with actual
requirements or the manufacturer’s data sheet (characteristic curve of the device under test).
Determining tripping voltage for spark gaps:
– Select maximum voltage such that the anticipated breakdown
voltage of the device under test is roughly one third of this value (observe manufacturer’s data sheet if applicable).
– Select the current limit value in accordance with actual
requirements within a range of 5 to 10 A (response charac­teristics are too unstable with larger current limit values, which may result in faulty measurement results).
Detecting weak points in the insulation:
– Select maximum voltage such that it does not exceed the test
object’s permissible insulation voltage; it can be assumed that an insulation fault will occur even with a significantly lower voltage if an accordingly lower maximum voltage value is selected (nevertheless at least greater than anticipated break­down voltage) – the ramp is less steep as a result (increased measuring accuracy).
– Select the current limit value in accordance with actual
requirements within a range of 5 to 10 A (see also settings for spark gaps).
Start Measurement – Constant Test Voltage
Quick polarity reversal if parameter is set to AUTO: 01/10 … 10/10: L1-PE … L1-L3
If semiautomatic polarity reversal is selected (see section
10.9), the corresponding icon is displayed instead of the ramp.
General Notes Regarding Insulation Measurements with Ramp Function
Insulation measurement with ramp function serves the following purposes:
• Detect weak points in the test object’s insulation
• Determine tripping voltage of voltage limiting components and test them for correct functioning These components may include, for example, varistors, overvoltage limiters (e.g. DEHNguard® from Dehn+Söhne) and spark gaps.
The test instrument uses continuously rising test voltage for this measuring function, up to the maximum selected voltage limit. The measuring procedure is started by pressing the ON/START key and runs automatically until one of the following events occurs:
Press and hold for long­term measurement:
Quick polarity reversal if parameter is set to AUTO: 01/10 … 10/10: L1-PE … L1-L3
The instrument’s batteries are rapidly depleted during the insulation resistance measurement. When using the con- stant test voltage function, only press and hold the start key until the display has become stable (if long-term measurement is required).
74 Gossen Metrawatt GmbH
Page 75
Special Condition for Insulation Resistance Measurement
Attention!
!
Attention!
!
Attention!
!
Attention!
!
R
ISO
Limit Value:
RE(INS) > Limit Value
UL R
L
R
EINS
Voltage type: constant
Test vol t ag e :
50, 100, 250, 325, 500, 1000 V*
Voltage type: rising/ramp
Earth leakage
* Freely adjustable voltage (see section 10.8)
resistance:
Insulation resistance can only be measured at voltage­free objects.
If measured insulation resistance is less than the selected limit value, the UL/RL LED lights up.
If an interference voltage of insulation resistance is not measured. The MAINS/NETZ LED lights up and the interference voltage pop-up message appears.
All conductors (L1, L2, L3 and N) must be tested against PE!
Do not touch the instrument’s terminal contacts during insulation resistance measurements!
If nothing has been connected to the terminal contacts, or if a resistive load component has been connected for measurement, your body would be exposed to a current of approximately 1 mA at a voltage of 1000 V. The resultant perceptible shock may lead to injury (e.g. resulting from a startled reaction etc.).
25 V is present within the system,
Discharging the Device Under Test
16.2 Special Case: Earth Leakage Resistance (R
This measurement is performed in order to determine electro­static discharge capacity for floor coverings in accordance with EN 1081.
EISO
)
Select Measuring Function
Set Parameters
If measurement is performed at a capacitive object such as a long cable, it becomes charged with up to approx. 1000 V! Touching such objects is life endangering!
When an insulation resistance measurement has been performed on a capacitive object it’s automatically discharged by the instru­ment after measurement has been completed. Contact with the device under test must be maintained to this end. The falling volt­age value can be observed at the U display.
Do not disconnect the DUT until less than 10 V is displayed for U!
Evaluating Measured Values
Instrument measuring error must be taken into consideration in order to assure that the limit values set forth in DIN VDE regula­tions are not fallen short of. The required minimum display values for insulation resistance can be determined with the help of Table 3 on page 100. These values take maximum device error into consideration (under nominal conditions of use). Intermediate val­ues can be interpolated.
Connection and Test Setup
Gossen Metrawatt GmbH 75
Rub the floor covering at the point at which measurement is to
be performed with a dry cloth.
Place the 1081 floor probe onto the point of measurement
and load it with a weight of at least 300 N (30 kg).
Establish a conductive connection between the measuring
electrode and the test probe and connect the measuring adapter (2-pole) to an earth contact, e.g. the earthing contact at a mains outlet or a central heating radiator (prerequisite: re­liable ground connection).
Page 76
Start Measurement
The limit value for earth leakage resistance from the relevant regulations applies.
76 Gossen Metrawatt GmbH
Page 77
17
Attention!
!
Note
Note
R
LO
ROFFSET: ON OFF
Polarity: +/– to PE
Polarity: +/– to PE
with ramp sequence
Limit Value:
RLO > Limit Value
UL R
L
Measuring Low-Value Resistance of up to 200 (Protective Conductor and Equipotential Bonding Conductor)
According to the regulations, the measurement of low-value resis­tance at protective conductors, earth conductors or bonding con­ductors must be performed with (automatic) polarity reversal of the test voltage, or with current flow in one (+ pole to PE) and then the other direction (– pole to PE).
Low-resistance may only be measured at voltage-free objects.
Select Measuring Function
Connection
via 2-pole adapter only!
❏ ROFFSET ON/OFF
– Compensation for Measurement Cables up to 10
If measurement cables or extension cables are used, their resis­tance can be automatically subtracted from the measurement results. Proceed as follows:
Switch R
footer.
Select a polarity option or automatic polarity reversal.
Short-circuit the end of the measurement extension cable with
the second test probe at the instrument.
Start measurement of offset resistance with IN.
First of all, an intermittent acoustic warning is generated and a blinking message appears, in order to prevent inadvertent deletion of a previ­ously saved offset value.
Start offset measurement by pressing
the triggering key once again or abort offset measurement by pressing the ON/
START
OFFSET from OFF to ON. Roffset = 0.00 appears in the
key
(in this case =
If offset measurement is stopped upon appearance of a pop-up error window indicating Roffset > 10 or a differ- ence between RLO+ and RLO– of greater than 10%, the last measured offset value is retained. Inadvertent deletion of a previously ascertained offset value is thus practically ruled out. The respectively smaller value is otherwise stored to memory as an offset value. The maximum offset value is 10.0 . Negative resistances may result due to the offset value.
ESC
).
Set Parameters
Measuring ROFFSET
The ROFFSET x.xx mes­sage now appears in the footer at the display, where x.xx can be a value between 0.00 and 10.0. This value is subtracted from the actual measuring results for all subsequent R ments, if the R
R
OFFSET must be redetermined in the following cases:
• After switching to a different polarity option
• After switching from ON to OFF and back again The offset value can be deliberately deleted by switching ROFFSET
from OFF to ON.
OFFSET ON/OFF key has been set to ON.
Only use this function when performing measurements with extension cables. When different extension cables are used, the above described procedure must always be repeated.
measure-
LO
Type / Polarit y
Gossen Metrawatt GmbH 77
The direction in which current flows can be selected here.
Limits – Setting the Limit Value
The limit value for resistance can be set as desired. If measured values occur which are above this limit value, the red UL/RL LED lights up. Limit values can be selected within a range of 0.10 to
10.0 (editable). The limit value is displayed above the measured value.
Page 78

17.1 Measurement with Constant Test Current

Attention!
!
Note
Start Measurement
Press and hold for long­term measurement
Measuring Low-Value Resistance Measurement cable and 2-pole measuring adapter resis­tance is compensated automatically thanks to the 4-wire method and thus doesn’t effect measurement results. However, if an extension cord is used its resistance must be measured and deducted from the measurement results.
Resistances which do not demonstrate a stable value until after a “settling in period” should not be measured with automatic polarity reversal, but rather one after the other with positive and negative polarity. Examples of resistances whose values may change during measurement include: – Incandescent lamp resistance, whose values change
due to warming caused by test current – Resistances with a large conductive component – Contact resistance
The test probes should always be in contact with the device under test before the Start key If the DUT is energized, measurement is disable as soon as it’s contacted with the test probes. If the start key the test probes afterwards, the fuse blows. Which of the two fuses has blown is indicated in the pop-up window with the error message by means of an arrow.
In the case of single-pole measurement, the respective value is saved to the database as R
Polarity Selection Display Condition
+ pole to PE RLO+ None – pole to PE RLO None
Pole to PE
is pressed first and the DUT is contacted with
LO
RLO Where RLO 10% RLO+
RLO
is activated.
.
Where RLO > 10%
Automatic Polarity Reversal
After the measuring sequence has been started, the instrument performs the measurement with automatic polarity reversal, first with current flow in one direction, and then in the other. In the case of long-term measurement (press and hold ON/START ▼ key), polarity is switched once per second.
If the difference between RLO+ and RLO– is greater than 10% with automatic polarity reversal, RLO+ and RLO– values are dis­played instead of RLO. The respectively larger value, RLO+ or RLO–, appears at the top and is saved to the database as the RLO value.
Evaluating Measured Values
See Table 4 on page 100.
Calculation of Cable Lengths for Common Copper Conductors
If the HELP key is activated after performance of resistance mea­surement, the cable lengths corresponding to common conduc­tor cross sections are displayed.
If results vary for the two different current flow directions, cable length is not displayed. In this case, capacitive or inductive com­ponents are apparently present which would distort the calcula­tion.
This table only applies to cables made with commercially available copper conductors and cannot be used for other materials (e.g. aluminum).
Evaluating Measurement Results
Differing results for measurements in both directions indicate volt­age at the DUT (e.g. thermovoltages or unit voltages).
Measurement results can be distorted by parallel connected impedances in load current circuits and by equalizing current, especially in systems which make use of overcurrent protection devices (previous neutralization) without an isolated protective conductor. Resistances which change during measurement (e.g. inductance), or a defective contact, can also cause distorted measurements (double display).
In order to assure unambiguous measurement results, causes of error must be located and eliminated.
In order to find the cause of the measuring error, measure resis­tance in both current flow directions.
The instrument’s batteries are exposed to excessive stress during insulation resistance measurement. For measurement with cur­rent flow in one direction, only press and hold the ON/START ▼ key as long as necessary for the measurement.
78 Gossen Metrawatt GmbH
Page 79
17.2 Protective Conductor Resistance Measurement with Ramp Sequence
Note
Measuring Phase Demagnetizing
and Waiting Time
Result
Time [s]
Rising Phase
Test C ur re n t [ A]
01 3 6
0.25
Before Polarity ReversalorRestart
Representation of Measuring and Waiting Times for Protective Conductor Resistance Measurement at PRCDs with the PROFITEST MXTRA
– Measurement at PRCDs with Current-Monitored Protective Conductor using the PROFITEST PRCD Test Adapter as an Accessory ( PROFITEST MXTRA only)
Application
Protective conductor current is monitored for certain types of PRCDs. Direct activation or deactivation of the test current required for protective conductor resistance measurements of at least 200 mA results in tripping of the PRCD and thus to interrup­tion of the protective conductor connection. Protective conductor measurement is no longer possible in this case.
A special ramp sequence for test current activation and deactiva­tion in combination with the PROFITEST PRCD test adapter per­mits protective conductor resistance measurement without trip­ping the PRCD.
Ramp Function Time Sequence
Due to the physical characteristics of the PRCD, measuring times for this ramp function amount to several seconds.
If test current polarity is revered, additional waiting time is also required during polarity reversal. This is programmed into the test sequence in the “automatic polarity reversal” operating mode.
Reverse polarity manually, e.g. from “+pole with ramp” to “-pole with ramp” . The test instru­ment then detects the reversal of current flow direction, stops measurement for the required waiting time and simultaneously displays a corresponding message (see figure at right).
Connection
Read the operating instructions for the PROFITEST PRCD
adapter, in particular section 4.1. It includes connection in­structions for offset measurement and for protective conduc­tor resistance measurement.
Selecting the Polarity Parameter
Select the desired polarity parameter with ramp.
Measuring ROFFSET
Perform offset measurement as described on page 77, in
order to assure that the test adapter’s connector contacts are not included in the measurement results.
The offset is only retained in memory until the polarity parameter is changed. If measurement is performed with manual polarity reversal (+pole or -pole), the offset mea­surement has to be repeated in both polarities before each measurement.
Measuring Protective Conductor Resistance
Determine whether or not the PRCD is activated. If not, acti-
vate it.
Perform protective conductor measurement as described in
section 17.1 above. Start the test sequence by briefly press­ing the ON/START ▼ key. The predetermined duration of the measuring phase can be extended by pressing and holding the ON/START ▼ key.
Start Measurement
PRCD Tripping due to Poor Contacting
Good contact must be assured between the test probes at the 2­pole adapter and the device under test or the sockets at the PROFITEST PRCD test adapter during measurement. Interrup­tions can result in considerable test current fluctuation which causes the PRCD to trip under unfavorable conditions.
If this is the case, tripping of the PRCD is automatically detected by the test instru­ment and indicated by a corresponding error message (see figure at the right). In this case as well, the test instrument automatically takes subsequently required waiting time into account before you can reactivate the PRCD and start the measurement over again.
Gossen Metrawatt GmbH 79
The symbol shown at the right appears during the mag­netization phase (rising curve) and the subsequent mea­suring phase (constant current).
If measurement is aborted already during the rising phase, no measurement results can be ascertained or displayed.
After measurement, the demagnetization phase (falling curve) and subsequent waiting time are indicated by the inverted symbol shown at the right. No new measurements can be started during this time.
Measurement results cannot be read and measurement with the same or another polarity cannot be started until the symbol at the right appears.
Page 80

18 Measurement with Accessory Sensors

Attention!
!
Attention!
!
Attention!
!
SENSOR
Clamp
output range
Limit Value:
I < and I > Limit Value
UL R
L

18.1 Current Measurement with Current Clamp Sensor

Bias, leakage and circulating current up to 1 A, as well as leakage current up to 1000 A can be measured with the help of special current clamp sensors, which are connected to sockets 15 and
16.
Danger: High-Voltage!
Use only current clamp sensors which are specifically of­fered as accessories by Gossen Metrawatt GmbH. Other current clamp sensors might not be terminated with an output load at the secondary side. Dangerously high voltage may endanger the user and the device in such cases.
Maximum input voltage at the test instrument!
Do not measure any currents which are greater than specified for the measuring range of the respective clamp. Input voltage for clamp connector sockets 15 and 16 at the test instrument may not exceed 1 V!
Set Parameters
The transformation ratio parameter must be correspondingly set at the test instrument depending upon the respectively selected measuring range at the current clamp sensor.
Be sure to read and adhere to the operating instructions for current clamp sensors and the safety precautions in­cluded therein, especially those regarding the approved measuring category.
Select Measuring Function
Select Measuring Range at the Current Clamp Sensor
Tes t
Instrument
Transforma-
tion Ratio
Parameter
1:1
1 V / A
01:10
100 mV / A
1:100
10 mV / A
1:1000
1 mV / A
WZ12C Switch
1 mV / mA
x 100 [mV/A] 0 … 10 A
x 10 [mV/A] 0 … 100 A
1 mV / A x 1 [mV/A]
Clamp Meters Test
Z3512A
Switch
x 1000
[mV/A]
WZ12C
Measuring
Range
1 mA …
15 A
1 A …
150 A
Z3512A
Measuring
Range
0 … 1 A
0 … 1000 A
Instrument
Measuring
Range
5 …
999 mA
0.05 … 10 A
0.5 … 100 A
5 …
150A/999A
Specifying limit values results in automatic evaluation at the end of the measurement.
Connection
Start Measurement
Test Instrument Clamp Test
Transformation Ratio
Parameter
1:1
1 V / A
01:10
100 mV / A
1:100
10 mV / A
80 Gossen Metrawatt GmbH
METRAFLEX P300
Switch
3 A (1 V/A) 3A 5 … 999 mA
30 A (100 mV/A) 30 A 0.05 … 10 A
300 A (10 mV/A) 300 A 0.5 … 100 A
METRAFLEX P300
Measuring Range
Instrument
Measuring
Range
Press the OK key to stop the measurement.
Page 81

19 Special Functions – EXTRA Switch Position

EXTRA
Select the EXTRA Switch Position
Overview of Special Functions
Softkey Meaning / Special
Function
Voltage drop measurement
U function Standing surface
insulation impedance
Z
function
ST
Meter startup test
kWh function Leakage current
measurement
function
I
L
Insulation monitor test
IMD function Residual voltage
test
Ures function Intelligent ramp
ta + Ifunction
Section / Page
PROFITEST MBASE+
PROFITEST MTECH+
PROFITEST MPRO
PROFITEST MXTRA
section
19.1 on page 82
section
19.2 on page 83
section
19.3 on page 84
section
19.4 on page 85
section
19.5 on page 86
section
19.6 on page 88
section
19.7 on page 89
✓✓✓✓
✓✓✓✓
✓✓✓✓
———
———
———
———
Selecting Special Functions
The list of special functions is accessed by pressing the upper­most softkey. Select the desired function with the appropriate icon.
section
19.8 on page 90
RCM residual current monitor
———
RCM function
section
19.9 on page 91
Testing of electric vehicle operating statuses at charging
stations per IEC 61851-1
section
19.10 on page 92
Documentation of fault simulations at PRCDs with the PRO-
———
FITEST PRCD adapter
Gossen Metrawatt GmbH 81
Page 82
19.1 Voltage Drop Measurement (at ZLN) – U Function
Note
1
2
Nominal current: 2 … 160 A
Polarity selection: Lx-N
Diameters: 1.5 … 70 mm
Cable type: NY…, H03… - H07…
Number of wires: 2 … 10-wire
Tripping characteristics: B, L
Limit Value:
U % > Limit Value
UL R
L
U
Red
1
2
Significance and Display of U (per IEC 60364-6)
Voltage drop from the intersection of the distribution network and the consumer system to the point of connection of an electrical power consumer should not exceed 4% of nominal line voltage.
Calculating voltage drop (without offset): U = Z
L-N
Calculating voltage drop (with offset): U = (Z
L-N
U in % = 100 × U / U See also section 14 regarding the measurement procedure and
connection.
(electrical outlet or device connector terminals)
× nominal current of the fuse
– Z
) × nominal current of the fuse
OFFSET
L-N
Connection and Test Setup
Set Parameters
Measurement Without OFFSET
Proceed as follows:
Switch OFFSET from ON to OFF.
Determine OFFSET (in %).
Proceed as follows:
Switch OFFSET from OFF to ON. UConnect the test probe to the point of common coupling
(measuring device / meter).
Start measurement of offset with I
First of all, an intermittent acoustic warning is generated and a blinking message appears, in order to prevent inadvertent deletion of a pre­viously saved offset value.
Start offset measurement by pressing
the triggering key once again, or abort offset measurement by pressing the
ON/START ▼ key (in this case ESC).
OFFSET = 0.00% is displayed.
.
N
If nominal current IN is changed by U value is automatically adjusted.
OFFSET
, the offset
Set Limit Values
U
OFFSET x.xx % is displayed and x.xx can be a value within a
range of 0.00 to 99.9%. An error message appears in a pop-up window
if Z exceeds 9.99
Start measurement with OFFSET.
TAB Limit value per German technical connection conditions
for connection to low-voltage mains between the distribu­tion network and the measuring device
DIN Limit value per DIN 18015-1: U < 3%
VDE Limit value per DIN VDE 0100-520: U < 4%
NL Limit value per NIV: U < 5%
82 Gossen Metrawatt GmbH
between the measuring device and the consuming device
between the distribution network and the consuming device (adjustable up to 10% in this case)

Page 83
19.2 Measuring the Impedance of Insulating Floors and Walls
Note
Attention!
!
OK
NOT OK
(standing surface insulation impedance) – Z
Function
ST
Measuring Method
The instrument measures the impedance between a weighted metal plate and earth. Line voltage available at the measuring site is used as an alternating voltage source. The ZST equivalent cir­cuit is considered a parallel circuit.
Connection and Test Setup
Start Measurement
Evaluate Measured Value
The measured value has to be evaluated after measurement has been completed:
Use the measuring setup described in section 16.2 (trian­gular probe) or the one outlined below:
Cover the floor or the wall at unfavorable locations, e.g. at
joints or abutments, with a damp cloth measuring approx. 270 × 270 mm.
Place the 1081 Probe on top of the damp cloth and load the
probe with a weight of 750 N (75 kg, i.e. one person) for floors, or 250 N (25 kg) for walls, e.g. press against the wall with one hand which is insulated with a glove).
Establish a conductive connection to the 1081 Probe, and
connect it to the probe connector socket at the instrument.
Connect the instrument to a mains outlet with the test plug.
Do not touch the metal plate or the damp cloth with your bare hands. No more than 50% line voltage may be applied to these parts! Current with a value of up to 3.5 mA may flow! The measured value would be distorted as well.
Resistance values must be measured at several points in order to provide for adequate evaluation. Measured resistance may not be less than 50 k at any given point. If the measured value is greater than 30 M, Z play panel.
In the event that “NOT OK” is selected, an error is indicated by the UL/RL LED which lights up red.
See also Table 5 on page 101 with regard to evaluating measured values.
The measured value cannot be saved to memory and included in the test report until it has been evaluated.
> 30.0M always appears at the dis-
ST
Save measured value
Gossen Metrawatt GmbH 83
Page 84
19.3 Testing Meter Startup with Earthing Contact Plug
Note
Note
OK
NOT OK
– kWh Function
Energy consumption meters can be tested for correct startup with this function.
Connection L – N Earthing Contact Plug
Start Measurement
The measured value cannot be saved to memory and included in the test report until it has been evaluated.
Save measured value
Special Case
Startup of energy consumption meters which are connected between L and L or L and N can be tested with this function.
Connection L – L 2-Pole Adapter
The meter is tested with the help of an internal load resistor and a test current of approxi­mately 250 mA. After pressing the Start key, test power is dis­played and the meter can be tested for proper startup within a period of 5 seconds. The “RUN” pictograph is displayed.
TN systems: All 3 phase conductors must be tested against N, one after the other. In other types of systems, all phase conductors (active conduc­tors) must be tested against one another.
If minimum power is not reached, the test is either not started or aborted.
Evaluate Measured Value
The measured value has to be evaluated after measurement has been completed:
If an earthing contact outlet is not available, you can use the 2-pole adapter. N must be contacted with the PE test probe (L2), and then measurement must be started. If PE is contacted with the PE test probe (L2) during the meter startup test, approximately 250 mA flow through the protective conductor and any upstream RCD is tripped.
In the event that “NOT OK” is selected, an error is indicated by the UL/RL LED which lights up red.
84 Gossen Metrawatt GmbH
Page 85
19.4 Leakage Current Measurement with PRO-AB Leakage
Attention!
!
Note
Current Adapter as Accessory – I (PROFITEST MXTRA only)
Function
L
Application
Measurement of touch voltage in accordance with DIN VDE 0107, part 10, as well as continuous leakage and patient auxiliary current per IEC 62353 (VDE 0750-1) / IEC 601-1 / EN 60601-1, is possible using the PRO-AB leakage current measuring adapter as an accessory with the test instrument.
As specified in the standards listed above, current values of up to 10 mA can be measured with this measuring adapter. In order to be able to fully cover this measuring range using the measure­ment input provided on the test instrument (2-pole current clamp input), the measuring instrument is equipped with range switching including transformation ratios of 10:1 and 1:1. In the 10:1 range, voltage dividing takes place at the same ratio.
Connection and Test Setup
In order to perform the leakage current measurement, the adapter’s measurement outputs must be plugged into the mea­surement inputs at the left-hand side of the test instrument (2­pole current clamp input and probe input).
Either of the leakage current measuring adapter’s inputs is con­nected to reference earth (e.g. safe earth electrode / equipotential bonding) via a measurement cable. The metallic housing (accessi­ble part) of the device under test is contacted with a test probe or alligator clip which is connected to the other input by means of a second measurement cable.
Measuring Procedure
Refer to the operating instructions for the PRO-AB leakage cur­rent measuring adapter regarding performance of the measure­ment.
The test plug should be located in the storage slot during leakage current measurement. Under no circumstances may the test plug be connected with any system compo­nents, including PE / ground potential (measured values might otherwise be distorted).
Testing the PRO-AB Adapter
The adapter should be tested before use and at regular intervals (see adapter operating instructions).
The measurement can be started or stopped by pressing the ON/ START ▼ key. Leakage current measurement is a long-term mea- surement, i.e. is continues until it’s stopped by the user. The momentary measured value is displayed continuously during measurement.
The self-test must be deactivated in the menu (set “TEST ON/OFF” function key to “OFF”) in order to perform a measurement.
Always start with the large measuring range (10:1), unless there’s no doubt that small measured values can be expected, in which case the small measuring range can be used (1:1). The measuring range must be selected at the measuring adapter, as well as in the menu using the corresponding function key (RANGE). It must be assured that the range settings at the adapter and at the test instrument are always identical, in order to prevent any distortion of measurement results.
Depending on the magnitude of the measured values, the range setting can or must (in the case of overranging) be manually cor­rected at the measuring adapter and the test instrument.
Individual limit values can be adjusted after pressing the Limits function key. Exceeded limit values are indicated by the red limit value LED at the test instrument.
Gossen Metrawatt GmbH 85
Page 86
19.5
1
3
2
Conductor
Relationship:

Testing Insulation Monitoring Devices – IMD Function (PROFITEST MXTRA only)

Application
Insulation monitoring devices (IMDs) or earth fault detection sys­tems (EDSs) are used in IT systems in order to monitor adherence to a minimum insulation resistance value as specified by DIN VDE 0100-410.
They’re used in power supplies for which a single-pole earth fault may not result in failure of the power supply, for example in oper­ating rooms or photovoltaic systems.
Insulation monitors can be tested with the help of this special function. After pressing the ON/START ▼ key, an adjustable insula- tion resistance is activated between one of the two phases of the IT system to be monitored and ground to this end. This resistance can be changed in the MAN± manual sequence mode with the help of the + or – softkey, or varied automatically from R
in the AUTO operating mode. Testing is ended by once again
R
min
pressing the ON/START ▼ key. Time during which the momentary resistance value prevails since
changing the value at the system is displayed. The IMD’s display and response characteristics can be subsequently evaluated and documented with the help of the OK or NOT OK softkey.
max
to
Connection L – N
Resistance RSTART (3)
Numerous parameters are available for setting resistance RSTART, with which measurement is begun.
Conductor Relationship / Resistance Range (2)
Conductor relationship: The corresponding conductor relation-
ship can be selected for documentation of the measuring point.
Resistance range: A range of values can be selected for testing
the display of resistance at the IMD.
The parameter is set as a percentage with reference to the resis­tance momentarily introduced by the test instrument.
Upper and lower limit values are displayed in the measuring view.
When selecting test resistance, don’t forget that an excessively high test current could damage sensitive system components.
Set Parameters
Measuring Procedure (1)
There are two ways to conduct the test: – MAN: Resistance is changed manually by tapping the respec-
tive softkeys.
AUTO: Resistance is changed automatically every 2 seconds
beginning with R
START.
Measuring Procedure
Set the parameters.Start: Press the ON/START ▼ key.A resistance is introduced between the phase and protective
conductors and time measurement is started.
Manual test MAN + –: press the or key to increase
or reduce test resistance R
Automatic test AUTO: the resistance value is changed automati-
cally.
Time to trip ta is restarted each time resistance is changed.Press IIn order to end measurement, press the ON/START ▼ key as
soon as the IMD indicates that insulation resistance has been fallen short of.
Display of measured valuesEvaluation query: Measurement OK?If evaluation is NOT OK: the UL/RL LED lights up red.Save: by pressing the soft key.
in order to change the conductor relationship.
N
L-PE.
86 Gossen Metrawatt GmbH
Page 87
Measurement can be
OK
NOT OK
aborted by pressing the ON/START or ESC key.
The following measured values are displayed:
RL-PE: Active test resistance with upper and lower limit values
ta: Response time (during which momentary resistance is
applied until the measurement is ended)
- R
R
min
with reference to the number of possible resistances
U
L1PE
conductor L1 and protective conductor PE
U
L2PE
conductor L2 and protective conductor PE
U
L1L2
conductors L1 and L2
I
LPE
: Status display indicating momentary resistance
max
: Momentary voltage at the test probes between phase
: Momentary voltage at the test probes between phase
: Momentary voltage at the test probes between phase
: Test current flowing through the active resistance
f: Frequency of the applied voltage
Evaluation
In order to evaluate the measurement, it must be stopped. This applies to manual as well as automatic measurement. Press the ON/START ▼ or ESC key to this end. The stopwatch is stopped and the evaluation window appears.
Retrieving Saved Measured Values
The measured value cannot be saved to memory and included in the test report until it has been evaluated (see also section 9.4).
With the help of the key shown at the right (MW: measured value / PA: parameter), the setting parameters can be displayed for this measurement.
Gossen Metrawatt GmbH 87
Page 88
19.6 Residual Voltage Test – U
Note
Limit Value:
U % > Limit Value
UL R
L
U
only)
Function (PROFITEST MXTRA
res
Application
The EN 60204 standard specifies that after switching supply power off, residual voltage must drop to a value of 60 V or less within 5 seconds at all accessible, active components of a machine to which a voltage of greater than 60 V is applied during operation.
Testing for the absence of voltage is performed as follows with the test instrument by means of a voltage measurement which involves the measurement of discharge time: In the case of voltage dips of greater than 5% of momentary line voltage (within 0.7 seconds), the stopwatch is started and momentary undervoltage is displayed as U indicated by the red UL/RL LED.
The function is ended after 30 seconds after which Ures and tu data can be deleted and the function can thus be restarted by pressing the ESC key.
after 5 seconds and
res
Connection
Measuring Sequence – Long-Term Measurement
Testing is selected as a continuous measure­ment because residual voltage testing is trig­gered automatically and voltage measurement is always active for safety reasons.
If, for example, conductors are exposed when a machine is switched off – e.g. if plug connectors are disengaged – which are not protected against direct contact, maximum permissible discharge time is 1 second!
Limit Values
Set Limit Values
88 Gossen Metrawatt GmbH
Page 89
19.7
Nom. res. current: 10 … 500 mA
Type 1: RCD, SRCD, PRCD …
Nominal current: 6 … 125 A
Type 2: AC , A/F , B
* Type B = AC/DC sensitive
Touch voltage:
< 25 V, < 50 V, < 65 V
Intelligent Ramp – ta+I (
PROFITEST MXTRA

only)
Function
Application
The advantage of this measuring function in contrast to individual measurement of I breaking time and breaking current by means of a test current which is increased in steps, during which the RCD is tripped only once.
The intelligent ramp is subdivided into time segments of 300 ms each between the initial current value (35% I and the final current value (130% I results in a gradation for which each step corre­sponds to a constant test current which is applied for no longer than 300 ms, assuming that tripping does not occur.
And thus both tripping current and tripping time are measured and displayed. Measured quantities are acquired with reduced accuracy.
and tA is the simultaneous measurement of
N
)
N
). This
N
Start Touch Voltage Measurement
Start Tripping Test
Connection
Set Parameters
The measurement sequence can be broke off prematurely at any time by pressing the ON/START ▼ key.
Measurement Results
Gossen Metrawatt GmbH 89
Page 90
19.8 Testing Residual Current Monitors
Nom. res. current: 10 … 500 mA
Waveform:
Nominal current: 6 … 125 A
Typ e: A , B *
* Type B = AC/DC sensitive
X times tripping current:
Connection: with/without probe
System type: TN/TT, IT
Touch voltage:
< 25 V, < 50 V, < 65 V
– RCM Function ( PROFITEST MXTRA only)
General
Residual current monitors (RCMs) monitor residual current in elec­trical systems and display it continuously. As is also the case with residual current devices, external switching devices can be con­trolled in order to shut down supply power in the event that a specified residual current value is exceeded.
However, the advan­tage of an RCM is that the user is informed of fault current within the system before shutdown takes place.
As opposed to individual measurement of I
, measurement results
t
A
must be evaluated man­ually in this case.
If an RCM is used in combination with an external switching device, the combination must be tested as if it were an RCD.
N
and
Connection
Measure Touch Voltage
No-Trip Test with ½ × I
and 10 s
N
Set Parameters for I
After 10 seconds have elapsed, no residual current may be indi­cated. The measurement must then be evaluated. In the event that “NOT OK” is selected (in case of false alarm), an error is indi­cated by the UL/RL LED which lights up red.
The measured value cannot be saved to memory and included in the test report until it has been evaluated.
F
Tripping test with 1 × I
N
– Measurement of Signal Response Time (Stopwatch Function) with Residual Current Generated by the Test Instrument
Measurement must be stopped manually by pressing the ON/ START or I
in order to document tripping time.
90 Gossen Metrawatt GmbH
In the event that NOT OK is selected, an error is indicated by the UL/ RL LED which lights up red.
The measured value cannot be saved to memory and included in the test report until it has been evaluated.
key immediately after indication of residual current,
N
Page 91

19.9 Checking the Operating Statuses of Electric Vehicles at Charging Stations per IEC 61851 ((PROFITEST MTECH+ & PROFITEST MXTRA)

A charging station is a facility designed to charge electric vehicles in accordance with IEC 61851-1, and is equipped with essential elements including a plug connector, conductor protection, an RCD, a circuit breaker and a safety communication device (PWM). Depending on where it’s used, other function modules may be added, for example for mains connection and metering.
Selecting the Adapter (test box)
Simulation of Operating Statuses per IEC 61851-1 with the MEN­NEKES Test Box
(Statuses A through E) The MENNEKES test box is used exclusively to simulate the vari-
ous operating statuses of a fictitious electric vehicle connected to a charging station. Settings for the simulated operating statuses can be found in the operating instructions for the test box.
The simulated operating statuses can be saved to as a visual inspection and documented in the report generating program.
Select the respective status to be checked with the
SECLECT STATUS key at the test instrument.
Status C – non-gassing vehicle detected
• Vehicle is ready for charging / power is connected
• Voltage between PE and CP: +6 V / -12 V
Status D – gassing vehicle detected
• Vehicle is ready for charging / power is connected
• Voltage between PE and CP: +3 V / -12 V
Status A – charging cable connected to charging point only
• CP signal is activated.
• Voltage between PE and CP is 12 V.
Status B – charging cable connected to charging point and vehicle
• Charging cable is locked into place at the charging point and
the vehicle.
• Vehicle is not yet ready for charging.
• Voltage between PE and CP: +9 V / -12 V
Status E – cable is damaged
• Short-circuit between PE and CP
• Charging cable is unlocked at the charging point.
• Voltage between PE and CP is +0 V.
Semi-Automatic Changing of Operating Statuses
As an alternative to manual sta­tus changing via the parame­ters menu for the SECLECT STATUS softkey at the test instrument, quick and con­venient switching amongst the statuses is also possible. The AUTO status parameter has to be selected to this end. After responding to the visual inspection prompt and saving the results, automatic switching to the next status ensues – the 01/05 key display corresponds to A/E (01 = A, 02 = B, 03 = C, 04 = D, 05 = E).
Status variants can be skipped by pressing the I instrument or the test plug.
key at the test
N
Gossen Metrawatt GmbH 91
Page 92
19.10 PRCD – Test Sequences for Documenting Fault Simula-
Attention!
!
tions at PRCDs with the PROFITEST PRCD Adapter (PROFIT­EST MXTRA only)
The PROFITEST PRCD test adapter can be used in combination with the test instrument.
Read the respective operating instructions before using the PROFITEST PRCD.
Measurements with the PROFITEST PRCD connected to the test instrument:
• Measurement of the PRCD’s insulation resistance using the test instrument’s R
• Measurement of the PRCD’s protective conductor resistance using the test instrument’s RLO function. Please note that the protective conductor measurement is a modified RLO mea­surement with ramp sequence for PRCDs (seesection 17).
• Tripping test with nominal residual current using the test instrument’s I
• Measurement of time to trip using the test instrument’s I function (see section 12.3).
• Varistor test for PRCD-K: measurement via ISO ramp (see section 16).
F
function (see section 16).
INS
function (see section 12.3).
N

19.10.1 Fault Simulation

The procedure for the PROFITEST PRCD, including the proce­dure with the device under test, is described in the operating instructions for the PROFITEST PRCD. This section describes the procedure for the test instrument.
Procedure
Prepare error simulation at the PROFITEST PRCD. Refer to
the operating instructions for the PROFITEST PRCD.
Select the test sequence at the test instrument.Execute each of the test sequence steps at the PROFITEST
PRCD and document evaluation and assessment at the test instrument.
Select the PRCD to be Tested
Testing performed by simulating faults is carried out without con­nection to the test instrument, but it’s accompanied and docu­mented by the test instrument. The test sequence is opened in the test instrument to this end and the specified steps are exe­cuted at the PROFITEST PRCD. Afterwards, evaluation and assessment of each test step (OK or not OK) is performed at the test instrument for later documentation.
There are three preset test sequences: – PRCD-S (single-phase / 3-pole): 11 test steps – PRCD-K (single-phase / 3-pole): 4 test steps – PRCD-S (3-phase / 5-pole): 18 test steps
Interaction Between PROFITEST PRCD and Test Instrument
Switch Position at
Display at Test Instrument
Test Step Icon
Meaning
PROFIT­EST PRCD
ON 1~ON
ON 3~ON
BREAK Lx
Lx <-> PE
Lx <-> N
PE-U
EXT
Uext -> PE
PROBE
PRCD-Ip
—AUTOAUTO
Single-phase PRCD activated
3-phase PRCD activated
Interrupted phase
Wires reversed between phase conductor and PE or neutral conductor
PE to phase
Contact ON key on PRCD with probe
Protective conductor current measurement with current clamp transformer
Semi-automatic change of fault simulations
The test steps are displayed at the test instrument. Their mean­ings and the associated switch positions at the PROFITEST PRCD are listed in the above table.
92 Gossen Metrawatt GmbH
Page 93
Overview of Test Sequences and their Test Steps
PRCD-S, single-phase: 11 test steps
Selection Examples, PRCD-S Test Sequence (single-phase) – 11 Test Steps
Simulation of Interruption (steps 1 to 6)
PRCD-S, 3-phase: 18 test steps
PRCD-K, single-phase: 5 test steps
Reversed Conductor Simulation (step 7)
Simulation of PE to Phase (step 8)
Contact ON Key at PRCD with Probe (step 10)
Measurement of Protective Conductor Current with a Current Clamp Transformer (step 11)
Gossen Metrawatt GmbH 93
Page 94
Selection Examples, PRCD-S Test Sequence (3-phase) – 18 Test Steps
Simulation of Interruption (steps 1 to 10)
Reversed Conductor Simulation (steps 11 to 16)
Semi-Automatic Changing of Fault Simulations (Statuses)
As an alternative to manual sta­tus changing via the parameters menu for the respective PRCD selection at the test instrument (PRCD-S 1~, PRCD-K 1~ or PRCD- S3~), quick and convenient switching amongst the fault simulations is also possible. The AUTO status parameter has to be selected to this end. After responding to the visual inspec­tion prompt and saving the results, automatic switching to the next fault simulation ensues.
Skipping Test Steps
Test steps can be skipped during fault simulation by pressing the
key at the test instrument or the test plug.
I
N
Simulation of PE to Phase (step 17)
Measurement of Protective Conductor Current with a Current Clamp Transformer (step 18)
94 Gossen Metrawatt GmbH
Page 95
20 Test Sequences (Automatic Test Sequences)
AUTO
1
2 3 4
5 6 7
8
109
11 12
Select the utilized test instrument.
!
– AUTO Function
Select AUTO Switch Position at the Test Instrument
With the rotary switch in the AUTO position, all of the test sequences in the device are displayed.
If there aren’t any test sequences in the instrument, NO DATA appears.

20.1 General (test sequence layouts)

If the same sequence of tests will be run frequently (one after the other with subsequent report generation), for example as speci­fied in the standards, it’s advisable to make use of test sequences.
Automated test sequences can be compiled from manually cre­ated individual measurements with the help of the test sequence function.
A test sequence consists of up to 200 individual steps, which are executed one after the other. Fundamentally, differentiation is made amongst three types of individual steps:
•Note (Visual Inspection test step) Test sequences are inter- rupted when a pop-up message is displayed for the inspector. The test sequences is not resumed until the message has been acknowledged. Sample Message Before Insulation Resistance Measurement “Disconnect the device from the mains!”
• Visual inspection, testing and report generation: The test sequence is interrupted when a passed/failed evaluation is displayed. The comment and the results of the evaluation are saved to the database.
• Measurement (“User-Evaluated Measurement” test step): same as individual measurements with instruments with storage and parameters configuration

20.2 Creation of Test Sequences with ETC

The test sequences are created at the PC with the help of ETC software, and are then transferred to the test instrument.
Measurement parameters are also configured at the PC. How­ever, parameters can be changed at the test instrument during the test sequence before the respective measurement is started.
When the test step is started once more, the parameter settings specified in ETC are loaded again.
ETC does not subject the parameters to a plausibility check. As a result, the newly created test sequence should be checked at the test instrument before it’s permanently added to the database.
Limit values are not currently set in ETC, and have to be adjusted during the automatic test. Accessing the Menu for Editing Test Sequences
In order to be able to edit existing test sequences (e.g. add test steps or change parameter settings), they first have to be loaded to ETC.
There are two ways to do this:
•ETC: Extras Test Sequences Load Test Sequences
or
• ETC: Device Test Sequences Receive Test Sequences
(from the file “test_sequence_xyz.seq”)
(from the connected test instrument)
Operating Overview: Creating Test Sequences at a PC
1 Create a new test sequence – enter a designation. 2 Change the designation of the selected test sequence. 3 Duplicate the selected test sequence,
(Copy) is added to the name of the duplicated sequence. 4 Delete the selected test sequence. 5 Create or add a new test step for the selected test sequence.
– Select the test step type from the list to this end and either accept or edit
its designation. 6 Duplicate the selected test step. 7 Delete the selected test step. 8 Change position of the selected test step within the sequence. 9 Select test parameters for the selected test step type from the list.
10 Select a setting for the measuring parameter from the list. 11 Accept change to the measuring parameter. 12 Exit the test sequences menu.
Saving Test Sequences in ETC to the PC
We recommend saving default test sequences, as well as edited and new test sequences, to the PC or to other data storage media using the desired filename (test_se­quence_xyz.seq) with the help of the following menu command: Extras Test Sequences Save Test Sequences. Data loss resulting from certain administrative operations is prevented in this way (see following notes).
Due to the fact that only up to 10 test sequences, can be trans­ferred to the test instrument, no more than 10 test sequences should be saved to any given file.
Test sequences which have been saved to a file can be reloaded to ETC at any time by clicking “Extras Test Sequences Load Test Sequences”. Sequences can be further edited by clicking “Extras Test Sequences Edit Test Sequences”.
Transferring Test Sequences from the PC to the Test Instrument
After executing the following ETC command, all previously cre­ated test sequences (up to 10) are transferred to the connected test instrument: “Device Test Sequences Send Test Sequences”.
Gossen Metrawatt GmbH 95
Page 96
Attention!
!
Test sequences which have been loaded to the test in­strument are deleted when: – New test sequences are received from the PC – Selection lists are received from the PC – Backup data is restored to the test instrument – The user interface language is changed – The test instrument’s entire database is deleted – The test instrument is reset to its default settings – The firmware is updated
For as long as the test sequences are being transferred, a progress bar is displayed at the PC and the illustration shown to the right appears at the test instrument’s display.
After data transmission has been completed, the display is switched to the database memory menu.
The display is returned to the measuring menu for the respective switch position by clicking ESC.
The selected test sequence (SEQU.1 in this case) is started with the ON/START ▼ key.
When a test step of the measurement type is executed, the same screen layout appears as is also the case for individual measure­ments. The current test step number appears in the header instead of the memory and battery icons. After pressing the Save key twice, the next test step is displayed.
Setting Parameters and Limit Values
Parameters and limit values can also be changed while a test sequence is running or before the respective measurement is started. The respective change only affects the active test sequence and is not saved.
Skipping Test Steps
There are two ways to skip test steps or individual measurements:
• Select the test sequence, change to the test step column at the right with the help of the cursor, select the x and press the ON/START ▼ key.
• The navigation menu can be opened within the test sequence by pressing the navigation key (cursor left-right). Jumping to the next or the previous test step is possible using the separate scroll keys which then appear. The navigation menu can be exited again and the current test step can be displayed by pressing the ESC key.
th
test step

20.3 Using Test Sequences Test Sequence Commands

Acknowledge message
Discard event
Confirm event
To prev io u s / to next step
Save measurement results
Configuring Test Sequence Parameters
Measurement parameters are also configured at the PC. How­ever, parameters can be changed at the test instrument during the test sequence before the respective measurement is started.
Selecting and Starting a Test Sequence at the Test Instrument
Aborting or Ending a Test Sequence
An active sequence can be aborted by pressing the ESC key and then acknowledging.
Sequence Ended appears after the last test step is completed. The initial menu, List of Test Sequences, is once again displayed after acknowledging the prompt.
96 Gossen Metrawatt GmbH
Page 97

21 Maintenance

Note
Note
Attention!
!
Note

21.1 Test Instrument Firmware/Software

The layout of the test instruments makes it possible to adapt device software to the latest standards and regulations. Beyond this, suggestions from customers result in continuous improve­ment of the test instrument software, as well as new functions.
Query Current Status
Turn the rotary switch to the SETUP position.Press the SW-Info CALIBRATION key.
Press any key in order to return to the main menu.
Update
Internal test instrument firmware/software can be updated via the USB port with the help of a PC and an interface cable.
The firmware/software with the required version is transferred to the test instrument with the help of the MASTER Updater Soft­ware-Tool. Currently installed test instrument firmware/software is overwritten.
The MASTER Updater can be downloaded free of charge from www.gossenmetrawatt.com. Registration with myGMC is required to this end. Operating instructions for the Firmware Update Tool are available here as well.
Prerequisite for transfer: The rotary selector switch is not set to the U position.
Severe damage to the instrument may occur if incorrect fuses are used. Only original fuses from Gossen Metrawatt GmbH may be used (order no. 3-578-285-01 / SIBA 7012540.3.15 SI-EINSATZ FF 3.15/500 6.3X32). Only original fuses assure required protection by means of suitable blowing characteristics. Short-circuiting of fuse terminals or the repair of fuses is prohibited, and is life endangering! The instrument may be damaged if fuses with incorrect ampere ratings, breaking capacities or blowing charac­teristics are used!
Remove the blown fuse and insert a new one.Insert the fuse compartment lid after the fuse has been re-
placed and secure it by turning clockwise.

21.3 Housing

No special maintenance is required for the housing. Keep outside surfaces clean. Use a slightly dampened cloth for
cleaning. In particular for the protective rubber surfaces, we rec­ommend a moist, lint-free microfiber cloth. Avoid the use of cleansers, abrasives or solvents.

21.4 Calibration

Use of your instrument and resultant stressing influence the instrument and lead to deviation from warranted accuracy values.
In the case of strict measuring accuracy requirements, as well as in the event of severe stressing (e.g. severe climatic or mechanical stress), we recommend a relatively short calibration interval of once per year. If this is not the case, a calibration interval of 2 to 3 years is usually adequate.
Please contact GMC-I Service GmbH for calibration services (see section 22, “Contact, Support and Service”, on page 98).
A sticker with an instrument-specific guideline value for the cali­bration interval and information regarding the service provider is included on the instrument as an aid.
Establish a USB connection between the PC and the test in-
strument.
Switch the PC and the test instrument on.Follow the instructions displayed by the MASTER Updater and
the associated operating instructions.

21.1.1 Rechargeable Battery Care

Check to make sure that no leakage has occurred at the rechargeable batteries at short, regular intervals, or after the instrument has been in storage for a lengthy period of time.
Remove rechargeable batteries during lengthy periods of non-use (e.g. vacation). This prevents excessive deple­tion or leakage, which may result in damage to the test instrument.

21.2 Fuse Replacement

If a fuse has blown due to overloading, a corresponding message error appears at the display panel. The instrument’s voltage mea­suring ranges are nevertheless still functional.
Disconnect the device from the measuring circuit at all poles!Loosen the slotted screws at the fuse compartment lid next to
the mains power cable with a screwdriver. The fuses are now accessible.
Replacement fuses can be accessed after opening the battery
compartment lid.
Date on Calibration Certificate / Calibration Interval Begins Upon Receipt
Your instrument is furnished with a calibration certificate on which a date appears. This date may be further in the past if your instrument has been stored for some time prior to sale. The instruments are stored in accordance with the spec­ified conditions. Drift is thus negligible for a duration of 1 year. Longer storage periods are highly unusual. Consequently, the instrument’s characteristic values lie within the specifications and the first calibration interval can be determined as of the date of receipt.
Gossen Metrawatt GmbH 97
Page 98

22 Contact, Support and Service

23 CE Declaration

Gossen Metrawatt GmbH can be reached directly and simply – we have a single number for everything! Whether you require sup­port or training, or have an individual inquiry, we can answer all of your questions here:
+49-911-8602-0 Monday to Thursday:
Friday:
8 a.m. to 4 p.m. 8 a.m. to 2 p.m.
Or contact us by e-mail at: info@gossenmetrawatt.com
Do you prefer support by e-mail?
Measuring and Test Technology: support@gossenmetrawatt.com
Industrial Measuring Technology: support.industrie@gossenmetrawatt.com
Enquiries concerning training and seminars can also be submitted by e-mail and online:
training@gossenmetrawatt.com https://www.gossenmetrawatt.com/training
The instrument fulfills all requirements of applicable EU directives and national regulations. We confirm this with the CE mark. The CE declaration is available upon request.
A calibration certificate is included with the instrument.
Please contact GMC-I Service GmbH for repairs, replacement parts and calibration
1)
:
+49-911-817718-0 Beuthener Str. 41 service@gossenmetrawatt.com www.gmci-service.com
90471 Nürnberg Germany
1)
DAkkS calibration laboratory per DIN EN ISO/IEC 17025 – accredited by the Deutsche Akkreditierungsstelle GmbH under refer­ence number D-K-15080-01-01.
98 Gossen Metrawatt GmbH
Page 99

24 Disposal and Environmental Protection

Attention!
!
Attention!
!
Proper disposal makes an important contribution to the protection of our environment and the conservation of natural resources.
Environmental Damage Improper disposal results in environmental damage. Follow the instructions concerning return and disposal included in this section.
The following comments refer specifically to the legal situation in the Federal Republic of Germany. Owners or end users who are subject to other national requirements are required to comply with the respectively applicable national requirements and to imple­ment them correctly on site. Relevant information can be obtained, for example, from the responsible national authorities or national distributors.
Waste Electrical Equipment, Electrical or Electronic Accessories and Waste Batteries (including rechargeable batteries)
Electrical equipment and batteries (including rechargeable batter­ies) contain valuable raw materials that can be recycled, as well as hazardous substances which can cause serious harm to human health and the environment, and they must be recycled and dis­posed of correctly.
The symbol on the left depicting a crossed-out garbage can on wheels refers to the legal obligation of the owner or end user (German electrical and electronic equipment act ElektroG and German battery act BattG) not to dis­pose of used electrical equipment and batteries with unsorted municipal waste (“household trash”). Waste batteries must be removed from the old device (where possible) without destroying them and the old device and the waste batteries must be dis­posed of separately. The battery type and its chemical composi­tion are indicated on the battery’s labelling. If the abbreviations “Pb” for lead, “Cd” for cadmium or “Hg” for mercury are included, the battery exceeds the limit for the respective metal.
Please observe the owner’s or end user’s responsibility with regard to deleting personal data, as well as any other sensitive data, from old devices before disposal.
Old devices, electrical or electronic accessories and waste batter­ies (including rechargeable batteries) used in Germany can be returned free of charge to Gossen Metrawatt GmbH or the service provider responsible for their disposal in compliance with applica­ble regulations, in particular laws concerning packaging and haz­ardous goods. Further information regarding returns can be found on our website.
Packaging Materials
We recommend retaining the respective packaging materials for the case that you might require servicing or calibration in the future.
Danger of Asphyxiation Resulting from Foils and Other Packaging Materials Children and other vulnerable persons may suffocate if they wrap themselves in packaging materials, or their components or foils, or if they pull them over their heads or swallow them. Keep packaging materials, as well as their components and foils, out of the reach of babies, children and other vulnerable persons.
In accordance with German packaging law (VerpackG), the user is obligated to correctly dispose of packaging and its components separately, and not together with unsorted municipal waste (“household trash”).
Private end consumers can dispose of packaging free of charge at the responsible collection point. Packaging which is not subject to so-called system participation is returned to the appointed ser­vice provider. Further information regarding returns can be found on our website.
Gossen Metrawatt GmbH 99
Page 100

25 Appendix

25.1 Tables for Determining Maximum and Minimum Display Values in Consideration of the Instrument’s Maximum Measuring and Intrinsic Uncertainties

Table 1
Z
(full-wave) / Z
L-P E.
()
Limit Valu e
Display Value
0.10 0.07 0.10 0.05
0.15 0.11 0.15 0.10
0.20 0.16 0.20 0.14
0.25 0.20 0.25 0.18
0.30 0.25 0.30 0.22
0.35 0.30 0.35 0.27
0.40 0.34 0.40 0.31
0.45 0.39 0.45 0.35
0.50 0.43 0.50 0.39
0.60 0.51 0.60 0.48
0.70 0.60 0.70 0.56
0.80 0.70 0.80 0.65
0.90 0.79 0.90 0.73
1.00 0.88 1.00 0.82
1.50 1.40 1.50 1.33
2.00 1.87 2.00 1.79
2.50 2.35 2.50 2.24
3.00 2.82 3.00 2.70
3.50 3.30 3.50 3.15
4.00 3.78 4.00 3.60
4.50 4.25 4.50 4.06
5.00 4.73 5.00 4.51
6.00 5.68 6.00 5.42
7.00 6.63 7.00 6.33
8.00 7.59 8.00 7.24
9.00 8.54 9.00 8.15
9.99 9.48 9.99 9.05
Max.
Z
L-N
(+/- half-wave) ()
L-P E.
Limit
Value
Max.
Display Value
Table 3
R
M
Limit
Valu e
Display Value
0.10 0.12 10.0 10.7
0.15 0.17 15.0 15.9
0.20 0.23 20.0 21.2
0.25 0.28 25.0 26.5
0.30 0.33 30.0 31.7
0.35 0.38 35.0 37.0
0.40 0.44 40.0 42.3
0.45 0.49 45.0 47.5
0.50 0.54 50.0 52.8
0.55 0.59 60.0 63.3
0.60 0.65 70.0 73.8
0.70 0.75 80.0 84.4
0.80 0.86 90.0 94.9
0.90 0.96 100 106
1.00 1.07 150 158
1.50 1.59 200 211
2.00 2.12 250 264
2.50 2.65 300 316
3.00 3.17
3.50 3.70
4.00 4.23
4.50 4.75
5.00 5.28
6.00 6.33
7.00 7.38
8.00 8.44
9.00 9.49
Min.
INS
Limit
Value
Min.
Display Value
Table 2
R
/ R
()
E
Limit
Value
0.10 0.07 10.0 9.49 1.00 k 906
0.15 0.11 15.0 13.6 1.50 k 1.36 k
0.20 0.16 20.0 18.1 2.00 k 1.81 k
0.25 0.20 25.0 22.7 2.50 k 2.27 k
0.30 0.25 30.0 27.2 3.00 k 2.72 k
0.35 0.30 35.0 31.7 3.50 k 3.17 k
0.40 0.34 40.0 36.3 4.00 k 3.63 k
0.45 0.39 45.0 40.8 4.50 k 4.08 k
0.50 0.43 50.0 45.4 5.00 k 4.54 k
0.60 0.51 60.0 54.5 6.00 k 5.45 k
0.70 0.60 70.0 63.6 7.00 k 6.36 k
0.80 0.70 80.0 72.7 8.00 k 7.27 k
0.90 0.79 90.0 81.7 9.00 k 8.17 k
1.00 0.88 100 90.8 9.99 k 9.08 k
1.50 1.40 150 133
2.00 1.87 200 179
2.50 2.35 250 224
3.00 2.82 300 270
3.50 3.30 350 315
4.00 3.78 400 360
4.50 4.25 450 406
5.00 4.73 500 451
6.00 5.68 600 542
7.00 6.63 700 633
8.00 7.59 800 724
9.00 8.54 900 815
Max.
Display Value
Limit
Valu e
ELoop
Max.
Display Value
Limit Value
Max.
Display Value
Table 4
R
Limit
Valu e
Display Value
0.10 0.07 10.0 9.59
0.15 0.12 15.0 14.4
0.20 0.17 20.0 19.2
0.25 0.22 25.0 24.0
0.30 0.26 30.0 28.8
0.35 0.31 35.0 33.6
0.40 0.36 40.0 38.4
0.45 0.41 45.0 43.2
0.50 0.46 50.0 48.0
0.60 0.55 60.0 57.6
0.70 0.65 70.0 67.2
0.80 0.75 80.0 76.9
0.90 0.84 90.0 86.5
1.00 0.94 99.9 96.0
1.50 1.42
2.00 1.90
2.50 2.38
3.00 2.86
3.50 3.34
4.00 3.82
4.50 4.30
5.00 4.78
6.00 5.75
7.00 6.71
8.00 7.67
9.00 8.63
Max.
LO
Limit
Value
Max.
Display Value
100 Gossen Metrawatt GmbH
Loading...