Good Will LCR-821, LCR-826, LCR-816, LCR-829, LCR-827 User Manual

...
LCR-800 Series User Manual
CONTENTS PAGE
1. INTRODUCTION……………………………………..……... 1
LCR-800 Series User Manual
SAFETY TERMS AND SYMBOLS
These terms may appear in this manual or on the product:
2. PRECAUTIONS BEFORE OPERATION…….……………. 2-1.Unpacking the instrument………………….……………. 2-2.Checking the Line Voltage…………………..…………… 2-3.Environment……………………………………..……….. 2-4.Equipment Installation and Operation…………………
3. PANEL DESCRIPTION…...……………………..………….. 4
OPERATION…………….………………………………...….
4. 4-1.Connects to DUT….………………..………….……..…...
4-2.Start-Up……….....……………………………..………….
4-3.Zeroing……………………………………………………. 4-4.Menu Functions.…..……………………………………… 4-5.Measurement Conditions…..……………………………..
SPECIFICATIONS…………………………………………...
5.
6. MESSAGE CODE.………………………………………….... 44
7. MAINTENANCE....…………………………………………...
7-1.Cleaning……………………………………………………
7-2.Battery Replacement……………………………………...
7-3.Recall Calibration………………………………………...
8. OPTION 1 (BIN Function for LCR-826/827/829 only)…….. 8-1. BIN Functions for Components Sorting……………..….
8-2. BIN Setting Conditions.....……………………………….
8-3. BIN Range Setting: Component Sorting Range..............
11 18
35
45
45
45
46
48 48 50 56
2 2 2 3 3
8 8 8 8
The following symbols may appear in this manual or on the product:
WARNING: Warning statements identify condition or practices
that could result in injury or loss of life.
CAUTION: Caution statements identify conditions or practices
that could result in damage to this product or other property.
DANGER ATTENTION Protective Frame or Chassis
High Voltage refer to Manual Conductor Terminal
Terminal
9. OPTION 2……………………………………………………..
9-1.On-line procedure…………………………………………
9-2.RS232 VIEWER Software Operation…………………...
9-3.The Configuration of Cable………………………………
i
65
65
69
74
ii
LCR-800 Series User Manual
FOR UNITED KINGDOM ONLY
NOTE: This lead / appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
Green/ Yellow: Earth
Blue: Neutral
Brown: Live (Phase)
LCR-800 Series User Manual
This cable/appliance should be protected by a suitably rated and approved
HBC mains fuse: refer to the rating information on the equipment and/or
user instructions for details. As a guide, cable of 0.75mm
2
should be
protected by a 3A or 5A fuse. Larger conductors would normally require
13A types, depending on the connection method used.
Any moulded mains connector that requires removal /replacement must be
destroyed by removal of any fuse & fuse carrier and disposed of
immediately, as a plug with bared wires is hazardous if a engaged in live
socket. Any re-wiring must be carried out in accordance with the
information detailed on this label.
As the colours of the wires in main leads may not correspond with the
colours marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth
terminal marked with the letter E or by the earth symbol
or coloured
Green or Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is
marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal
marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact
the supplier.
iii
iv
LCR-800 Series User Manual
LCR-800 Series User Manual
EC Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
No. 69 Lushan Road, Suzhou New District Jiangsu, China.
declare that the below mentioned products:
LCR-817/819/827/829/816/826/821
are herewith confirmed to comply with the requirements set out in the Council Directive on the Approximation of the Law of Member States relating to Electromagnetic Compatibility (89/336/EEC, 92/31/EEC, 93/68/EEC) and Low Voltage Equipment Directive (73/23/EEC, 93/68/EEC). For the evaluation regarding the Electromagnetic Compatibility and Low Voltage Equipment Directive, the following standards were applied:
EN 61326-1: Electrical equipment for measurement, control and laboratory use ––
EMC requirements (1997+A1: 1998) Conducted and Radiated Emissions EN 55011 : 1998 Group I Class A
Current Harmonic EN 61000-3-2: 2000 Voltage Fluctuation EN 61000-3-3: 1995
————————————————
————————————————
————————————————
Electrostatic Discharge IEC 61000-4-2: 1995 Radiated Immunity IEC 61000-4-3: 1995 Electrical Fast Transients IEC 61000-4-4: 1995 Surge Immunity IEC 61000-4-5: 1995 Conducted Susceptibility IEC 61000-4-6: 1996 Voltage Dips/ Interrupts IEC 61000-4-11: 1994
1. INTRODUCTION
The precise LCR meter series are automatic, user programmable instruments
that provide high reliability and great precision for measuring a wide variety of
impedance parameters. The frequency range of LCR-819/829 covers from 12Hz
to 100kHz, LCR-817/827 from 12Hz to 10kHz, LCR-816/826 from 100Hz to
2kHz, and LCR-821 from 12Hz to 200kHz. The basic accuracy is 0.1% for LCR-
827/829/816/826, 0.05% for LCR-817/819/821. The measured results can be
displayed on the high quality LCD monitor with decimal points and units. The
measured result resolution is five full digits for Inductance (L), Capacitance (C),
Resistance (R) and absolute value of impedance (|Z|) [four full digits for
Dissipation factor (D), Quality factor (Q), R with C, R with L, 2 digits after
decimal dot for phase angle (θ)]. In the meantime, the LCD monitor is also shown
the control status and parameters of settings. The keypads are easy for menu
programming. The LCR Meters’ test fixture (option) is a method of convenient,
reliable, guarded 4-terminal/2 wires connection of radial and axial leaded
components to the LCR Meters. Test conditions can be stored and recalled from
internal memory that will reduce the setup time for measurement preparing.
Low Voltage Equipment Directive 73/23/EEC & amended by 93/68/EEC
Safety Requirements IEC/EN 61010-1:2001
v
1
LCR-800 Series User Manual
2. PRECAUTIONS BEFORE OPERATION
2-1. Unpacking the instrument
The product has been fully inspected and tested before shipping from the
factory. Upon receiving the instrument, please unpack and inspect it to
check if there are any damages caused during transportation. If any damage
is found, notify the bearer and/or the dealer immediately.
LCR-800 Series User Manual
2-3. Environment
The normal ambient temperature range of the LCR Meters is from 10° to
50°C. To operate the instrument over this specific temperature range may
cause damage to the circuits.
Do not use the LCR Meters in a place where strong magnetic or electric
field exists as it may disturb the measurement.
2-4. Equipment Installation and Operation
2-2. Checking the Line Voltage
The LCR Meters can be operated with AC power source between 100V
rated voltage and 240V rated voltage at a frequency of 50 to 60Hz, no AC
voltage selector is necessary. Power connection to rear panel is through an
AC inlet module comprised of an AC connector and fuse holder. To change
the fuse proceed as follows:
z Remove the fuse holder by inserting a small flat head screwdriver
behind the small tab to force the holder outward.
z Install the correct fuse (slow-blow, 3A, 250Vac).
z Re-install the fuse holder back into the LCR Meters AC inlet module,
push in and lock.
WARNING: To avoid electrical shock the power cord protective
grounding conductor must be connected to ground.
WARNING: To avoid personal injury, disconnect the power cord
before removing the fuse holder.
Ensure there is proper ventilation for the vents in the LCR meters case. If
this equipment is used not according to the specification, the protection
provided by the equipment may be impaired.
WARNING: This is a Class A product. In a domestic environment
this product may cause radio interference in which case the user
may be required to take adequate measures.
2
3
LCR-800 Series User Manual
LCR-800 Series User Manual
3. PANEL DESCRIPTION
(1). Power Switch
Turns AC Power on or off.
(2). Function key
Soft key functions as indicated on the adjacent LCD monitor.
(3). Function key
Soft key functions as indicated on the adjacent LCD monitor.
(4). Function key
Soft key functions as indicated on the adjacent LCD monitor.
(5). Function key
Soft key functions as indicated on the adjacent LCD monitor.
(6). MENU key
Enters menu display mode or exits sub menu back to main menu.
(7). Compound key
z For making numerical entries as labeled. z Turns the “RANGE HOLD” mode on or off.
(8). Compound key
z For making numerical entries as labeled. z Turns the “CONSTANT VOLTAGE” mode on or off.
(9). Compound key
z For making numerical entries as labeled. z Measures the unit of Dissipation and Quality Factor in PPM.
(Note: The models of LCR-816 and 826 do not have this function.)
F1
F2
F3
F4
(10). Compound key
z For making numerical entries as labeled. z Selects the “INTERNAL BIAS” mode or “EXTERNAL BIAS” mode. (If
this key function is switched to “INTERNAL BIAS” mode, the bottom
of LCD monitor will display the “INT.B” message. If the external DC
bias is selected, the bottom of LCD monitor will display the “EXT.B”
message.)
(11). Compound key
z For making numerical entries as labeled. z Turns the INTERNAL BIAS mode or EXTERNAL BIAS mode “ON” or
“OFF”.
(12). Compound key
z For making numerical entries as ““ (the negative sign). z Inputs the “TEST FREQUENCY”.
(13). START (Compound key)
z Starts measurement sequence. Normally used in the “MANU” (Triggered)
mode.
z Selects “AUTO” or “MANU” mode by pressing this key for 3 seconds at
least.
z The LCR Meters will process the measurement automatically, if the
“AUTO” mode is selected.
key (ENTER)
(14).
This key enables programming of all special functions, test frequency, test
voltage, averaging, delay, and nominal value etc.
(15). Symbol key
Inputs the decimal point
4
5
LCR-800 Series User Manual
LCR-800 Series User Manual
(16). Numeral key—“2”
(17). Numeral key—“3”
(18). Numeral key—“5”
(19). Numeral key—“6”
(20). Numeral key—“9”
(21). Primary Display
This line can display the measured Inductance, Capacitance, or Resistance (LCR-821 has Ζ).
(22). Secondary Display
This line can display the measured Quality Factor or Dissipation or ESR or EPR (LCR-821 has Θ).
(23). Instrument status or indicates measurement results based on entered test
limits.
(24). Test conditions
(25). Input terminals
BNC connectors, connects to device under test (DUT).
FRONT PANEL
Connectors of the LCR Meters
Lforce (current, low)
Lsense (potential low)
Hsense (potential high)
Hforce (current, high)
BIAS
6
7
LCR-800 Series User Manual
LCR-800 Series User Manual
4. OPERATION
4-1. Connects to DUT
The LCR Meters utilize the structure of four wires measurement which allows
accurate, easy, and stable measurements and avoids mutual inductance and
interference from measurement signals, noise and other factors inherent with
other types of connections. For the accuracy of measurement, GOODWILL
produces the cable set and test fixture (option) for connection directly to the
front panel BNC connectors.
4-2. Start-Up
Connects the power cord of the LCR Meters to the mains socket-outlet. Presses
POWER button of front panel to apply the AC power to the LCR Meters.
the
4-3. Zeroing
In order to eliminate strayed capacitance and impedance of test cable during the
measurement, the LCR Meters should be zeroed to correct for test cable and/or
test fixture errors before taking measurements. The corrections are calculated
and stored in memory of the LCR Meters during the zeroing process. Open and
short circuit zeroing should be done for test cable and/or test fixture. For the best
accuracy, the test cable and/or test fixture should be zeroed once per day at least
and each time test cable or test fixture is changed in general.
The zeroing process of open and short circuits are following:
Open Circuit
z The test cable or test fixture should be open with no component connected. z Press MENU key. z Press F1 key to select “OFFSET” menu. z Press F1 key to select open circuit zeroing (the “CAP OFFSET” is indicated
on the adjacent LCD monitor.).
z After the BAR at the bottom of LCD monitor is filled to the full, the zeroing
process is done.
z If the zeroing process is successful, a message of “OK” will appear on the
LCD monitor. If failed, a message of “FAIL” will appear on the LCD
monitor.
Short Circuit
z The test cable should be connected or test fixture shorted (using a clean
copper wire, as short as possible).
z Press MENU key. z Press F1 key to select “OFFSET” menu. z Press F2 key to select short circuit zeroing (the “R/L OFFSET” is indicated
on the adjacent LCD monitor.).
z After the BAR at the bottom of LCD monitor is filled fully, the zeroing
process is done.
z If the zeroing process is successful, a message of “OK” will appear on the
LCD monitor. If failed, a message of “FAIL” will appear on the LCD
monitor.
Test Condition:
Test voltage=According to the actual test voltage.
Test speed = According to the actual test speed.
R.H = OFF
C.V = According to the actual test status.
For the summary, the zeroing menu can be chosen through menu selection as
shown in Figure 4-1 above.
NOTE: The “Open Circuit” and “Short Circuit” have to pass the test,
otherwise, the accuracy of the LCR Meters will become worse.
8
9
LCR-800 Series User Manual
LCR-800 Series User Manual
L 1.2345H Q 0.6789
TESTING
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
CAP. R/L OFFSET
SET SORT
SET PARAMETER
CALIBRATION
R.H OFF C.V OFF
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
OFFSET
SORT
SETTING
CALBRAT
EXIT
Press menu
key
Press F1 key
to the zeroing
menu
4-4. Menu Functions
All the LCR Meters’ programmable functions are controlled by the easy to use
menu displays. User can enter the menu mode by selecting the
MENU key that
calls up four top level menus, OFFSET, SORT, SETTING and CALBRAT.
Each one of these is comprised of a sub menu list whose functions are described
in detail below. User can enter one of four functions by pressing the
corresponding function key (just adjacent LCD monitor, see figure 4-2).
Press F1 key
to the zeroing
CAP. R/L OFFSET
SET SORT
SET PARAMETER
CALIBRATION
OFFSET
SORT
SETTING
CALBRAT
EXIT
menu
Press F2 key
to the bin sort
menu
Press F3 key
to the utility
menu
Press F4 key
to the
calibration
menu
OPEN TEST
SHORT TEST
Figure 4-1: Summary of zeroing menu
CAP
OFFSET
R/L
OFFSET
EXIT
10
Press F1 KEY for
open test
Press F2 KEY for
short test
Press MENU key
to exit menu screen
Figure 4-2: four main menu screen
4-4-1. Primary & Secondary Display
For the LCR Meters, four combinations of two parameters can be measured and
displayed simultaneously. One referred to the “Primary Display” (displayed
first) and the other to the “Secondary Display”. Depending on the component
type the primary and secondary display could be L & Q, C & D, C & R or R and
Q, and add L & R and Z & θ to LCR-821. The parameter can be chosen by
pressing
F3 key as shown in Figure 4-3.
11
LCR-800 Series User Manual
LCR-800 Series User Manual
Primary
Display
L 1.2345H Q 0.6789
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
Figure 4-3. Primary & Secondary display
User can select R/Q for resistor measurement; select L/Q or L/R for inductor
measurement; select Z/θ for impedance measurement, select either C/D or C/R
for capacitor measurement.
Note: Only LCR-821 can select Z/θ and L/R measurement modes.
4-4-2. Series & Parallel Equivalent Circuit
Impedance that is neither a pure resistance nor a pure reactance can be
represented at any specific frequency by either a series or a parallel combination
of resistance and reactance. Such representation is called “equivalent circuit”.
The component value of the “Primary Display” depends on which equivalent
circuit (series or parallel) is chosen. In normal, the component manufacturer
shall specify how a component is to be measured (usually series) and at what
frequency. Please refer to Figure 4-4.
Secondary
Display
TESTING
R.H OFF C.V OFF
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
Press F3 key successively in order to select the different measurement mode (R/Q, C/D, C/R, L/Q, Z/θ, L/R)
z Measurement Function
Measurement Mode Series Equivalent Circuit Parallel Equivalent
Circuit R / Q C / D C / R
Z / θ (degree)
L / R
Suggested Test Conditions:
Inductors less than 10µH: Series, 100kHz.
Inductors from 10 µH to 1mH: Series, 10kHz.
Inductors from 1mH to 1H: Series, 1kHz.
Inductors greater than 1H: Series, 0.1kHz.
Capacitors less than 10pF: Parallel, 100kHz.
Capacitors from 10 to 400pF: Series or Parallel, 10kHz.
Capacitors from 400 to 1µF: Series, 1kHz.
Capacitors greater than 1µF: Series, 0.1 or 0.12kHz.
Resistor less than 1k: Series, 1kHz.
Resistor from 1k to 10MΩ: Parallel, 0.25kHz.
Resistor greater than 10 MΩ: Parallel, 0.03kHz
Unless for special reason, always select “Series” for capacitors and inductors.
This has traditionally been standard practice. For very small capacitance or
inductance, select a higher test frequency for better accuracy. For very large
capacitance or inductance, select a lower test frequency for better accuracy. For
dc resistance, select a lower test frequency to minimize ac effects.
Because the reactive component most likely to be represented in a low resistance
resistor is series inductance, the “Series” is selected for a resistor below about
1k. If a resistor large than 10M, select “Parallel” that because the reactive
component most likely to be represent in a high resistance resistor is shunt
capacitance. If the Q is less than 0.1, the measured Rp is probably very close to
the dc resistance.
99 99 99 9 99
12
13
LCR-800 Series User Manual
LCR-800 Series User Manual
The total loss of a capacitor can be expressed in several ways, including D and
“ESR” (Equivalent Series Resistance). “ESR” is typically much larger than
actual “ohmic” series resistance of the wire leads and foils that are in series with
the heart of a capacitor physically, because ESR includes also the effect of
dielectric loss. ESR is related to D by the formula: ESR = Rs = D/ωCs. Where ω
represents “omega” = 2 pi time frequency.
Although it is traditional to measure series inductance of inductors, there are
situations in which the parallel equivalent circuit better represents the physical
component. For small “air-core” inductors, the significant loss mechanism is
usually “ohmic” or “copper loss” in the wire, therefore the series circuit is
appropriate. Nevertheless, for an “iron core”, the significant loss mechanism can
be “core loss”, therefore, the parallel equivalent circuit is appropriate which
being a better model of the inductor.
SPEED
L 1.2345H Q 0.6789
TESTING
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
Figure 4-4. Selections of Series & Parallel Circuit
R.H OFF C.V OFF
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
Press F4 key to select the Series or Parallel circuit
4-4-3. Measurement Displays
The measured results of the LCR Meters can be shown on the LCD monitor in
three ways: VALUE
appropriate item for measurement. Please refer to Figure 4-5.
VALUE
The LCD monitor will display the measured value of both the primary and
secondary parameter, shown with decimal and units. The resolution of primary
display (L, C, R or Z) is five digits. The resolution of secondary display (D, Q, R
with C or R with L) is four digits. The resolution of secondary display (θ) is at 2
digits after decimal point. The message “TESTING” is displayed when a test is
in process.
DELTA%
The “DELTA%” shows the percent deviation of the measured L, C, R or Z value
from a stored NOMINAL VALUE. The sign of deviation is indicated.
DELTA
The LCR difference is similar to the DELTA% except that the deviation is shown
in suitable units (ohms, henries, etc).
L 1.2345H Q 0.6789
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
, DELTA%, or DELTA. User can press F2 key to select the
SPEED
SLOW
TESTING
R.H OFF C.V OFF
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
Press F2 key to select the three different measuring display ("VALUE", "DELTA%", or "DELTA").
14
Figure 4-5. Types of measurement display
15
LCR-800 Series User Manual
4-4-4. Nominal Value
LCR-800 Series User Manual
Allows entry of a “Nominal Value” for the primary parameter which is the basis
for the measurement result in “DELTA” or “DELTA %”. Accepts numerical entry
up to five digits with decimal. Units are depended on which measurement
displays selected.
Steps of “Nominal Value” input (Figure 4-6):
z Press MENU key. z Press F2 key to select “SORT” menu. z Press F1 key to select “Nominal Value” (the “NOM.VAL” is indicated on
the adjacent LCD monitor).
z Input the nominal value via the numeral keys (5 digits with decimal
maximum).
z Press key z After the BAR at the bottom of LCD monitor is filled fully, the “Nominal
Value” input is done.
4-4-5. Selection of Measurement Speed
One of three measurement speeds SLOW, MEDIUM, or FA ST could be
selected (Figure 4-7). The continuous mode speeds are about 1, 5, and 12
measurement per second respectively. The trade-off is accuracy vs. speed. LCR-
817/819 will take a more accurate measurement at a slower rate. The trade-off is
as follows
L 1.2345H Q 0.6789
TESTING
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
CAP. R/L OFFSET
SET SORT
SET PARAMETER
CALIBRATION
R.H OFF C.V OFF
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
OFFSET
SORT
SETTING
CALBRAT
EXIT
Press menu
key
Press F2 key
to the sort
menu
SLOW speed : More than 1 measurement per second, at 0.05% accuracy
(or better)
MEDIUM speed : More than 3 measurements per second, at 0.1% accuracy
(or better)
FAST speed : More than 7 measurements per second, at 0.24% accuracy
(or better)
* For the details of accuracy, please refer to the specifications.
* Regarding the models LCR-826/827/829, please refer to 4-5-10 Handler
Interface Timing.
16
NOM.VAL = 77.000pF
OPTION 1
NOM.VAL
24.870
HANDLER
EXIT
Figure 4-6. Steps of “Nominal Value” input.
17
Press F1 key to input the nominal value
LCR-800 Series User Manual
Press F1 key to select the three different measurement speed. (SLOW, MEDIUM, or FAST)
SPEED
L 1.2345H Q 0.6789
TESTING
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
Figure 4-7. Selection of measurement speed
4-5. Measurement Conditions
4-5-1. Bias Voltage
There are two available bias voltage modes: “Internal” and “External”.
R.H OFF C.V OFF
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
LCR-800 Series User Manual
Press the numerical 7 key to select either internal or external bias voltage
7
BIAS
4
PPM
1
C.V
0
R.H
8
ON/OFF
5 6
2 3
.
Press the numerical 8 key to turn either internal or external bias voltage on or off.
9
S T A R T
_
FREQ
Internal:
An internal DC 2 volts bias voltage will apply to the device under test.
External:
An external DC bias voltage between 0 and 30 volts can be applied to the device
under test. The external bias connection is located on the rear panel. The
maximum current is 200mA. The supply of bias voltage has to be floating, don’t
connect either side to ground. It’s better to wait approximately 1 second for
taking a reading after initiating a testing process, therefore, the device under test
will stabilize after bias voltage applied. The DC bias voltage should be applied
only to capacitors in general. If the DC bias voltage is applied to device of low
impedance, the unreliable testing results will occur. When the external bias is
applied, please join the function of C.V.ON (Constant voltage On) in use.
18
L 1.2345H Q 0.6789
TESTING
F : 1.000 kHz V : 1.000 V
AUTO MANU INT.B OFF
Indication of internal or
external bias voltage
Figure 4-8. Selection of “BIAS” voltage
R.H OFF C.V OFF
19
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
Loading...
+ 28 hidden pages