Good Will GPI-826, GPI-825, GPT-815, GPT-805 User Manual

ELECTRICAL SAFETY TESTER
Y
Y
USER MANUAL
WARNING MANUFACTURER ASSUMES NO L IABILIT IF UNIT OPERATED IN AN UNSAFE MANNER.
WARNING
THIS INSTRUMENT GENERATES AND DELIVERS A HAZARDOUSLY HIGH VOLTAGE (5kV). BE EXTREMEL CAREFUL WHEN USING THIS INSTRUMENT. BE SURE TO READ MANUAL SECTION 3 PRECAUTIONS
ELECTRICAL SAFETY TESTER
USER MANUAL
EC Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No.7-1, Jhongsing Road, Tucheng City, Taipei County 236, Taiwan
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
No. 69 Lushan Road, Suzhou New District Jiangsu, China.
declares that the below mentioned product
are herewith confirmed to comply with the requirements set out in the Council Directive on the Approximation of the Law of Member States relating to Electromagnetic Compatibility (89/336/EEC,92/31/EEC) and Low Voltage Equipment Directive (73/23/EEC, 93/68/EEC). For the evaluation regarding the Electromagnetic Compatibility and Low Voltage Equipment Directive, the following standards were applied:
EN 61326-1: Electrical equipment for measurement, control and laboratory use –
EMC requirements (1997+A1: 1998+A2:2001)
Conducted and Radiated Emissions EN 55011 Group I class A: 1998
Current Harmonic EN 61000-3-2: 2000 Voltage Fluctuation EN 61000-3-3: 1995
----------------------------------------------
----------------------------------------------
----------------------------------------------
----------------------------------------------
GPT-805/815, GPI-825/826
Electrostatic Discharge EN 61000-4-2: 1995+A1 :1998 Radiated Immunity EN 61000-4-3: 1996+A1 :1998 Electrical Fast Transients EN 61000-4-4: 1995 Surge Immunity EN 61000-4-5: 1995 Conducted Susceptibility EN 61000-4-6: 1996
Power Frequency Magnetic Field IEC 61000-4-8: 1993 Voltage Dips/ Interrupts EN 61000-4-11: 1994
Low Voltage Equipment Directive 73/23/EEC & amended by93/68/EEC
Safety Requirements IEC/EN 61010-1: 2001
Remark: Also complied with Continuity of Protective Bonding Tester, Insulation
Resistance T est, Voltage Test, and Residual Voltage Test in accordance with the Sub-Clauses 19.2, 19.3, 19.4 and 19.5 of EN 60204-1: 1997
i
ii
ELECTRICAL SAFETY TESTER
USER MANUAL
ELECTRICAL SAFETY TESTER
USER MANUAL
CONTENTS PAGE
1. PRODUCT INTRODUCTION............................................
1-1.Description…………………………..…………………. 1-2.Feature……………………………..……………………
2. SPECIFICATION………………………………………….. 3
3. PRECAUTIONS BEFORE OPERATION……...………... 3-1.Unpacking the instrument……..…………….………... 3-2.Checking the Line Voltage..…………………..……….. 3-3.Environment……………………………………..……..
4. PANEL INTRODUCTION……………………..…………. 4-1.Front Panel……………………………………………... 4-2.Rear Panel……………………………………….……...
5. OPERA T IO N METHO D…… ..…………………………… 5-1.AC Withstanding Test…………………………………. 5-2.DC Withstanding Test…….…………………………… 5-3.Insulation Test…………...…..………………………….
6. MAINTENANCE………..…………………………………. 6-1.Fuse rating and type.…………………………………... 6-2.Cleaning………………..………………………………..
11 12
13
13 14 16
17
17 17
1
1 2
6
6 6 8
9
SAFETY TERMS AND SYMBOLS
These terms may appear in this manual or on the product:
WARNING. Warning statements identify condition or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
The equipment shall not be used for measurements within category II, III and IV.
The following symbols may appear in this manual or on the product:
iii
DANGER ATTENTION Protective Earth (ground) Frame or High Voltage refer to Manual Conductor Terminal Chassis
Terminal Terminal
iv
ELECTRICAL SAFETY TESTER
USER MANUAL
ELECTRICAL SAFETY TESTER
USER MANUAL
FOR UNITED KINGDOM ONLY
NOTE: This lead/appliance must only be wired by competent
persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance
with the following code:
Green/ Yellow: Earth Blue: Neutral Brown: Live (Phase)
As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with the letter E or by the earth
symbol
or coloured Green or Green & Yellow.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red. If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, cable of 0.75mm
2
should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used. Any moulded mains connector that requires removal /replacement must be destroyed by removal of any fuse & fuse carrier and disposed of immediately, as a plug with bared wires is hazardous if a engaged in live socket. Any re-wiring must be carried out in accordance with the information detailed on this label.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
v
vi
ELECTRICAL SAFETY TESTER
USER MANUAL
ELECTRICAL SAFETY TESTER
USER MANUAL
1. PRODUCT INTRODUCTION
1-1. Description
The GPT/GPI-800 Electrical Safety Testers (EST) are designed for AD/DC Withstanding Voltage test and Insulation Resistance test (IR) in order to provide a safe and accurate test environment for the operator. With thoughtful design described in 1-2. Feature insures a safe operation environment of high voltage test to protect user from hazardous impact.
The Electrical Safety Testers comply with the requirement of the electrical equipment & appliance control ordinances and JIS, CSA, UL, BS and other overseas standards as well. The testers can be used for withstanding voltage test of the various types of electrical and equipment and components.
The GPT/GPI-800 series are based on the family of GW withstanding voltage tester including AC Withstanding Voltage test, DC Withstanding Voltage test and Insu lation Resistance test. Please refer to the table as follows:
Function
Model
GPT-805 (500VA) GPT-815 (500VA) GPI-825 (500VA) GPI-826 (100VA)
AC DC IR
ˇ
ˇˇ ˇˇ ˇˇ
1-2. Features
The GPT/GPI-800 series offer several other features :
1) Current and output voltage setting
A safe way to set the current and output voltage without high voltage activated.
2) Easily and quickly setting with front panel control
A user-friendly interface provides user an easy and quick way to set all parameters.
3) Testing time
Use micro processor to control testing time.
4) Adjustable ARC detect level
The ARC detect level can be adjusted with front panel control.
5) Adjustable output voltage during test
The output voltage can be adjusted during testing that add an
operation flexibility.
6) An alert indicator for high voltage
A flashing red LED indicates dangerous situation during high voltage output is activated.
7) Remote I/O controller
In addition to the 9 pin remote I/O controller to control START/RESET, it also provides PASS/FAIL/TEST signal for professional use.