Gendex DenOptix QST User manual

Operator and Service Manual
Manuale dell’Operatore e Servizio
Manuel de l’Opérateur et de Service
Gebrauschsanweisung
Manual del Usuario y de Servicio
DenOptix
® QST
Phosphor Plat
Operator and Service Manual
English
Inglese
Anglais
Englisch
Inglés
Table of contents
1 Indications for use 5 2 Contraindications 7 3 Safety requirements 9
3.1 Important information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
3.2 Warning and caution statements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
4 Pre-installation information and recommendations 13
4.1 Purpose of the manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
4.2 Abbreviations and explanations of symbols
4.3 Physical description
4.4 Computer requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
4.5 Electrical requirements
4.6 Compliance to standards
4.7 Installation
4.8 Site selection
4.9 Unpacking the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
4.10 Hardware setup and connections
4.11 Network installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Gendex DenOptix® QST
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
5 System operating instructions 27
User and Service Manual
1 Preparing your current X-ray equipment
5.
5.2
Erasing the imaging plate
5.3 Infection control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
5.4 Taking an X-ray
5.5 Scanning the imaging plate
5.6 Preparing for the next patient ning t
ur
T
5.7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
he DenOp
tix QST system on and off
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
. . . . . . . . . . . . . . . . . . . . . . . . . . . .38
Doc # M010-004WWE July 2005
7
Gendex DenOptix® QST
6 Software operating instructions 39 7 Maintenance procedures 41
7.1 Cleaning the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
7.2 Operator maintenance
7.3 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
7.4 Disposal of waste materials and inoperative parts
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
. . . . . . . . . . . . . . . . . . . . . . .50
8 Storage and shipment 51
8.1 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
8.2 Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
9 Warranty statement 53
10 Technical specifications 55
11 Appendix 57
11.1 Appendix A: Storage phosphor technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
11.2 Appendix B: Lighting conditions for handling or erasing imaging plates
11.3 Appendix C: Optional printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
11.4 Appendix D: If you need assistance
©2005 Gendex Dental Systems
Gendex, DenOptix QST, AcuCam and Orthoralix are registered trademarks of Gendex Dental Systems.
ademark of Dentsply International.
CP is a regist
Rinn, X
VixWin and Concept are trademarks of Gendex Dental Systems.
This manual in Eng
ered tr
lish is the original version.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
. . . . . .58
User and Service Manual
2
Doc # M010-004WWE July 2005
Introduction
DenOptix® QST “Quad Speed Technology” System by Gendex®
DenOptix QST Intraoral (FMX) System
DenOptix QST Extraoral (E/O) System
DenOptix QST Intraoral and Extraoral System (Combo) System
Congratulations! Your decision to add the DenOptix QST System to your practice represents a wise investment for the future of your practice.
The DenOptix QST is a revolutionary product designed to completely replace traditional X-ray film and film processors. This system, built on phosphor imaging plate technology, offers the following benefits:
• Diagnostic quality images — every time.
• Lowers the price of dental imaging by eliminating the need for costly film, chemistry and film processors.
User and Service Manual
• Imaging plates are reusable.
• Saves critical time vs. using processing systems. DenOptix QST will process 4 bitewings in under 30 seconds and a FMX will process 20 images in less than 4 seconds per image. Panoramic images can be obtained in less than 80 seconds. A cephalometric image can
ained in less t
be obt
• Significant reduction of X-ray dose as compared to D-speed film.
ks with your current X-ray generating equipment. There is no need to purchase new
or
It w
• equipment.
• Reduces the hassles associated with the waste stream disposal.
han 80 seconds.
3
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
For over 50 years, dental professionals have relied on Gendex to provide X-ray equipment of the highest technology and quality. From our classic AC intraoral units to the software-
®
driven Orthoralix
panoramic, we are dedicated to providing products with the features and high quality you desire. As you enjoy using the new DenOptix QST system, you are taking advantage of our 50 years of manufacturing expertise in the dental industry.
Our motto speaks for itself: Gendex. Imaging Excellence.
Technical support
Call your local Gendex location. For a phone number, please refer to page 62.
Supplies and replacement parts
To order supplies or replacement parts for your DenOptix QST, contact your local Gendex equipment dealer. If your dealer is unable to assist you, call the Gendex location nearest to you.
User and Service Manual
4
Doc # M010-004WWE July 2005
Indications for use
The DenOptix QST is intended for use as a digital dental radiography system using X-ray recording media (phosphor imaging plates) for radiographic diagnostic intraoral and extraoral exposures providing interactive CRT retrieval, viewing and processing of stored computed radiographic images.
The system includes reusable photostimulable phosphor imaging plates, a laser diode scan­ner device and optical reader components, communications electronics and software, and various peripheral accessories.
1
User and Service Manual
5
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
User and Service Manual
6
Doc # M010-004WWE July 2005
Contraindications
There are no contraindications for use.
2
User and Service Manual
7
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
User and Service Manual
8
Doc # M010-004WWE July 2005
Safety requirements
3.1 Important information
It is important that all personnel who will operate the DenOptix QST System read and under­stand this manual before operating the device. All personnel should follow all warnings and cautions as outlined in Section 3.2, for their safety and the safety of others around them.
3.2 Warning and caution statements
In this manual, the following definitions apply for all WARNINGS and CAUTION statements:
Warnings: Any operation, procedure or practice, which, if not strictly observed, may result in injury or long-term health hazards to personnel or patients.
3
User and Service Manual
Cautions: Any operation, procedure or practice, which, if not strictly observed, may result in destruction of equipment or loss of treatment effectiveness.
Warnings
Should be used only by trained professional.
Federal law restricts the sale of this device to physicians, dentists and dental professionals
vice in pr
. Use of t
y
onl in injury.
Do not open the device to service it.
None of the internal parts of the scanner are user serviceable. The only user serviceable parts of the system are outlined in Section 7.2. If there is a service problem, call your qualified Gendex dealer or the Gendex representative nearest you.
his de
ocedures other than those described in this manual may result
9
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Do not place DenOptix QST near:
• X-ray equipment that is constantly energized (e.g., devices such as fluoroscopes which are
energized when not in use)
• Magnetic resonance imaging systems
• Large motor/electric generators
There is potential for electromagnetic interference between devices when using electronic equipment. International standards exist to minimize this potential.
DenOptix QST has been tested against and complies with the international standard EN 60601-1-2, Medical Electrical Equipment - General Requirements for Safety - Collateral Standard; Electromagnetic Compatibility Requirements and Tests, class A.
Use only grounded electrical connections.
Connect the DenOptix QST scanner to a grounded electrical outlet between 100-240 volts AC.
Do not reuse the barrier envelopes.
Once used within the oral cavity, the barrier envelopes are bio-contaminated and should NEVER be reused. Please dispose of in accordance with Section 7.4.
Imaging plates are toxic!
Never place an imaging plate in a patient’s mouth without enclosing it first in a completely sealed barrier envelope. If a patient swallows an imaging plate, contact a physician immedi­ately. The physician must remove the imaging plate. Do not use cracked, bent or chipped imaging plates.
Never open or override the lid lock when the scanner is in use or plugged in.
The scanner lid will lock during operation. In case of power failure, unplug the scanner and open the lid by depressing the lock mechanism as indicated in Section 7.3.
Never use imaging plates with patients that might chew or swallow them.
If the patient bites or chews the plate and damages the protective barrier envelope, rinse the patient’s mouth with a large amount of water. If the patient manages to swallow any of the blue surface of the plate, contact a doctor immediately.
The DenOptix QST scanner is a Class 1 laser device.
Caution: Use of controls or adjustments or performance of procedures other than those spec­ified herein may result in hazardous radiation exposure. A fail-safe switch in the carousel well
ver keeps the laser inactive as long as the cover is open. Only a trained technician from
and co a qualified Gendex dealer should remove the cover from the scanner. Direct eye contact with the output beam from the laser may cause serious damage and possible blindness.
User and Service Manual
10
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Caution statements
Reduce the exposure time on your intraoral X-ray unit.
The DenOptix QST System is designed to produce high quality, diagnostic images at reduced radiation levels. To get the maximum benefit, we recommend that the exposure be reduced as outlined in Section 5.1.
Completely erase the imaging plates.
Before reusing an imaging plate, place it with the blue or white surface up facing a bright light source for at least two (2) minutes as described in Section 5.2.
Mount the imaging plates under low light conditions.
Mount the imaging plates to the carousel under low light conditions as described in Section 5.4. Exposure to direct sunlight or direct indoor lighting will erase the information stored on the imaging plate.
Do not place scanner on or next to a radiator or water source.
Excessive heat or small amounts of water may damage the scanner’s electrical components.
Do not use in the presence of flammable anesthetics mixtures.
Do not autoclave the imaging plates.
Autoclaving will damage the imaging plate. If this happens, discard the imaging plate and replace it with a new one. If an imaging plate becomes contaminated, follow the procedure described in Section 5.6.
Do not leave unscanned, exposed imaging plates in light.
Leaving unscanned, exposed imaging plates in light will cause severe loss of image quality. If left exposed for long time, or to bright light, the image may be erased entirely.
Do not scratch the imaging plates.
When handling t
he imaging plat
e, do no
t touch the active side (blue or white) any sharp
object that might scratch the surface. Do not lay unprotected imaging plates face down.
User and Service Manual
11
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
User and Service Manual
12
Doc # M010-004WWE July 2005
Pre-installation information and recommendations
4.1 Purpose of the manual
The instructions contained in this manual should be carefully followed for safe, trouble-free and effective equipment use.
This manual provides the essential information necessary for the installation, operation and routine care of the DenOptix QST System. The detailed instructions for the imaging software associated with the scanner can be found in the Imaging Software User Manual and Installation Guide. Important instructions for personnel who have been trained in intraoral and panoramic radiography are contained in this manual. This manual is not to be used as a replacement for training in dental radiography.
4
4.2 Abbreviations and explanations of symbols
AC Alternate current
DC Direct current
Hz Hertz; cycles per second
MHz Millions of hertz
LED Light emitting diode
kVp Peak voltage in thousands of volts
es
amic
13
User and Service Manual
IP
CD-ROM Compact disc, read-only
pan
Imaging plat
Panor
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
ceph Cephalometric
I/O Intraoral
Combo Combination system capable of scanning both I/O and panoramic
imaging plates
TIFF Tagged image file format
LP/mm Line pairs per millimeter
PSP PhotoStimulable Phosphor
D.P.I. Dots per inch
CPU Central processing unit (your computer)
IEC International Electrotechnical Commission
| Power on
Power off
Green indicator, ready
Yellow indicator, scanning
Red indicator, error in scanning; refer to operators manual
High voltage
Laser radiation
RAM Random access memory
MB Mega bytes
GB Giga bytes
ersal serial bus
USB U
niv
The CE symbol ensures t provisions of the European Council Directives.
hat t
oduct herein specified meets the applicable
he pr
User and Service Manual
14
Doc # M010-004WWE July 2005
4.3 Physical description
The DenOptix QST system has several configurations. Even if the basic hardware is the same, DenOptix QST is provided with different licenses for scanning different types of plates, according to the chosen model. See the following table for a complete list of components for each model. At the interception of “models” (columns) and “components” (rows) you’ll see the quantity supplied for that item. If no quantity is specified, that item is optional and it can be purchased using the catalog # shown in the same row.
The Intraoral (FMX) and Extraoral (EO) models are upgradeable to the Combo (COMBO) system. Contact your local Gendex representative for details. In addition, the Extraoral and Combo systems can be upgraded to Cephalometric by ordering the catalog number listed on the following table.
Note: Buying some components such as imaging plates or carousels may not allow you to use them without License from Gendex. Upgrade Kits are available for the following: FMX series to upgrade to a Combo system, EO series to Combo system, and EO or Combo series to Ceph.
Gendex DenOptix® QST
Customer Supplied Components
• Bottle of anhydrous isopropyl alcohol for cleaning imaging plates
• Computer system as described in Section 4.4
• Monitor for viewing
User and Service Manual
15
Doc # M010-004WWE July 2005
Table of models and components
Gendex DenOptix® QST
User and Service Manual
16
Doc # M010-004WWE July 2005
DenOptix QST laser scanner
Figure 4-1 Scanner front view
Figure 4-2 Scanner rear view
Gendex DenOptix® QST
1) Carousel well and lid. The lid has a locking mech­anism that engages during scanning and loss of power (see Section 7.3 or handling procedure). The carousel, with imaging plates attached, is inserted into the carousel well for scanning. The carousel lid must be in the closed position prior to scanning. The scanner will not operate with the lid open.
2) LED indicator lights with graphic symbols.
The lights show the current status of the system. Green — Ready to scan Yellow Blinking – Scanning Red — Error
1) DenOptix QST scanner power entry module with
switch and fuses
2) USB 2.0 connector. A USB 2.0 cable (provided)
connects the DenOptix QST scanner to the CPU’s USB 2.0 port.
3) Power cord socket. The power cord (provided) will
be connected from this point to a grounded main outlet.
4) Scanner label
User and Service Manual
17
Doc # M010-004WWE July 2005
Scanner label
Gendex DenOptix® QST
User and Service Manual
Figure 4-3 Intraoral imaging plates
Rear (Black)
1) IP serial number for tracking purposes
2) Date of manufacture
3) Available from Gendex Dental Systems
4) Solid orientation dot
5) IP size indicator. Shows the equivalent film size
6) Orientation circle. Visibly shows on final radiograph for faster orientation
Front (Blue)
18
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Figure 4-4 Panoramic and cephalometric imaging plates
Panoramic Cephalometric
Front (Blue)
Rear (Black)
1) Plastic mounting strips used to hold panoramic imaging plate onto the carousel
2) Active imaging area, white color
3) Patient left ( ) and right ( ) side indicators
4) General information that includes:
• Available from Gendex
• Date of manufacture
• Imaging plate size
• IP serial number
L
R
User and Service Manual
19
Doc # M010-004WWE July 2005
Figure 4-5 Carousels
Gendex DenOptix® QST
Intraoral (QFMX)
Extraoral QEO/Ceph (QCH)
1) Top of the carousel. For the EO/CH carousels, the knob will be positioned based upon whether or not a Pan or Ceph is being scanned. If a Pan or Ceph is loaded, turn the knob towards the mark indicating a Pan or Ceph. If no Pan/Ceph plate is loaded, rotate the knob to the “no Pan or Ceph” position.
2) I/O imaging plate holders. The intraoral carousel can hold up to 39 I/O imaging plates at one time: 20 size 2, 9 size 1, 8 size 0, 2 size 3.
3) Panoramic imaging plate retaining device. The panoramic carousel can hold either the 5x12 in or the 15x30 cm imaging plate.
4) Cephalometric imaging plate retaining device can hold either 8x10 in or 18x24 cm.
5) I/O imaging plate holders on the back of the Extraoral (EO) and Ceph (CH) carousel. The carousel can hold up to four size 2, two size 0 and one size 4 imaging plates.
Figure 4-6 Disposable barrier envelopes
User and Service Manual
Front (Blue)
Back (Clear)
1) The front of the barrier envelope is blue. This is the side that should face the X-ray tube.
2) Barrier envelope adhesive strip. When this is removed, the envelope can be sealed and pro­vides protection from cross-contamination
20
Doc # M010-004WWE July 2005
4.4 Computer requirements
Minimum Computer Requirements Recommended
CPU Speed 800 MHz Pentium 4 2.4 GHz Operating system Microsoft® Windows® 2000 /XP Pro Microsoft® Windows® XP Pro
SP2 SP2
RAM 128 MB 256+ MB Hard Drive 6 GB 40+ GB USB port USB 2.0 USB 2.0 Monitor S-VGA with 0.25mm/0.26mm S-VGA with 0.25mm/0.26mm
dot pitch dot pitch
Video Display Adapter 4 MB RAM, 800 x 600 display, 8+ MB RAM, 1024x768
true color true color
Keyboard, Mouse Standard Standard Backup Device Highly recommended Printer Recommended (see appendix C)
1
To be evaluated based on the number of images taken, and the image file size used by the imaging software.
4.5 Electrical requirements
Gendex DenOptix® QST
1
Voltage 100-240 V AC 50/60 Hz power supply auto-senses the input voltage. Power 110 watts maximum Power Cord Standard line cord provided (Medical Grade for 115V)
4.6 Compliance to standards
The DenOptix QST System conforms to the following standards:
Standard Title
UL 60601-1 Medical Electrical Equipment, Part 1: General requirements for safety
1 CFR Chap
2 Subchapter J
MDD 93/42/ECC CAN/CSA-C22.2 Medical Electrical Equipment, Part 1: General requirements for safety
No. 601.1 IEC 6060 IEC 6060
IEC 60825-1 Safety of Laser Products ISO 1 ISO 14971 Risk Analysis
ter I Performance Standard for Light-Emitting Products
opean Medical De
Eur
1-1 1-1-2
485 Medical devices- Quality management systems
3
Medical Electrical Eq Medical Electrical Eq
ety- Collateral standard; Electromagnetic Compatibility
saf
vice Directive (CE Mark)
uipment, Par uipment, Par
t 1: General requirements for safety t 1: Gener
uirements for
al req
User and Service Manual
21
Doc # M010-004WWE July 2005
4.7 Installation
The DenOptix QST System is designed to be installed by a qualified equipment professional from your local dental dealer. Ask your local dental dealer for more information.
We do not recommend that an untrained person try to install and configure the system.
4.8 Site selection
The DenOptix QST scanner can be located almost anywhere in the dental office. The site you pick should have:
• Subdued lighting conditions. You should have the ability to turn down/off the lights and block sunlight with blinds. The area to be used for mounting imaging plates should not exceed 20 lux as measured on a light meter. This will give you about one minute to mount all of your imaging plates without encountering excess signal fading. If you cannot meas­ure the light in the room where you wish to mount the imaging plates, you should turn off the room lights, close all blinds and leave the door open just enough so that you can see to mount the imaging plates. This should make the light level about 10-20 lux. If you need more than one minute, refer to appendix B.
Gendex DenOptix® QST
User and Service Manual
• A stable, flat countertop large enough to hold the scanner plus provide a working area. We suggest a minimum of 189x369 in. (46x92 cm). The computer does not need to be on the countertop, but must be within the length of the USB 2.0 cable (9ft or 3m provided). The countertop must be able to hold the weight of the unit or a minimum of 60 lbs (27 kg).
• Access to a standard grounded electric outlet.
• Enough room to allow the operator to mount imaging plates and use the computer effectively.
• Do not position the DenOptix QST scanner in a dusty environment. Excessive dust levels can result in additional scanner service, beyond normal service intervals.
ou choose to put the system in your darkroom, please prepare the site by:
If y
• Removing all old film processing equipment and plumbing.
• Remove the safelight filter and adjust the intensity of the light by lowering the bulb wattage until the intensity of the light around the scanner is 20 lux or less.
• We recommend that you keep the light off and the door open when mounting imaging plates to the carousel. The light from the open door will allow you to see enough to mount the imaging plates but will not adversely effect the image.
22
Doc # M010-004WWE July 2005
4.9 Unpacking the unit
The DenOptix QST System is shipped in one box. Completely unpack the box and save it in a safe, dry location. You may need to repack the unit for shipping if you ever encounter a main­tenance outage, or relocate your office.
Inventory the contents of the box according to the “Table of Models and Components” (Section 4.3) and ensure that you have all the components for your DenOptix QST system as they’re listed in the Table. If any items are damaged or missing, contact your dental dealer immediately.
Gendex DenOptix® QST
Figure 4-7
Scanner, 1 each
Figure 4-10
Disposable Barrier Envelopes 500 Size 2 (FMX and Combo units), 200 Size 2 (EO units)
Figure 4-12
15x30 cm or 5x12 in Imaging Plate, 1 each (EO and Combo units)
Figure 4-8
Intraoral Carousel 1 each (FMX and Combo units)
Figure 4-11a
Size 0 Imaging Plates, 2 each (EO and Combo units)
3
e 4-1
igur
F
4 cm or 8x1
8x2
1 Plate, 1 each (Included with
hase of Ceph kit)
purc
0 in Imaging
Figure 4-9
EO/CH Carousel 1 each (EO, Combo units, and Ceph upgrade)
Figure 4-11b
Size 2 Imaging Plates, 20 each (FMX and Combo units), 8 each (EO units)
User and Service Manual
23
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Figure 4-14
Power Cord, 1 each
Figure 4-15
User Manual and Installation Guide (Software and Hardware), 1 each
4.10 Hardware setup and connections
NOTE: Before you start the hardware setup, ensure that you have an acceptable com­puter as outlined in Section 4.4. The computer must have a USB 2.0 port available.
Step 1: Pick a location using the guidelines from Section 4.8.
Step 2: Set up the computer and monitor per the manufacturer’s recommendations. Use an
ergonomic setup to minimize repetitive motion injuries.
Step 3: Turn the power on to the monitor and computer.
Step 4: Connect the device end of the USB 2.0
cable to the USB 2.0 connector on the back of the DenOptix QST scanner. Connect the scanner to a grounded power outlet. Make sure the other end of the USB 2.0 cable is connected to the USB 2.0 port on the computer.
Step 5: Connect the computer to a network
if desired. Refer to your Imaging
are User Manual and Installation
tw
Sof Guide for more information on how to configure the Imaging Software in a
twork.
ne
tep 6: Turn the power on the scanner. The green light on the scanner should be lit at this
S
he trouble shooting guidelines in Section 7.3.
w t
time. If it is no
Step 7: After the Windows operating system is loaded, a message indicating that a “New
Hardware” has been found will appear. Follow the on-screen instructions to final­ize the installation.
t, follo
Figure 4-16
USB 2.0 cable (6 ft, 2 m) 1 each
User and Service Manual
24
Doc # M010-004WWE July 2005
4.11 Network installation
Refer to your network/computer professional for network installation, configuration and maintenance. Refer to your Imaging Software User Manual and Installation Guide for more information on how to configure the Imaging Software in a network.
Gendex DenOptix® QST
User and Service Manual
25
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
User and Service Manual
26
Doc # M010-004WWE July 2005
System operating instructions
5.1 Preparing your current X-ray equipment
The DenOptix QST System produces X-ray images of high quality and low noise yet offers a dose reduction in intraoral surveys up to 80% of the dose required for D-speed intraoral film. More details on how this technology works can be found in appendix A. Almost any intrao­ral or panoramic X-ray unit can be used with the DenOptix QST system. There is no need to purchase a new one.
The imaging plates produce excellent diagnostic images over a wide range of exposures. With the DenOptix QST system, it is possible to get diagnostic images at the same exposure time you use for X-ray film or as much as 80% reduction of the dose required for D-speed film. This means that it is more difficult to over- or under-expose an image. We recommend, how­ever, that you start with the exposure times outlined in the table below.
5
User and Service Manual
Intraoral equipment
Exposure times indicated hereunder are estimates that must be verified and adjusted depending on specific and actual local conditions, e.g. supply voltage, X-ray tube yield, timer accuracy, beam filtration, etc.
27
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Intraoral equipment suggested exposure times, in seconds and impulses
DC SUPPLY 60kV 65 kV 70 kV AC SUPPLY 50 kV 60 kV 65 kV 70 kV 75 kV
Lower incisor/cuspid 0.32 sec. 0.25 sec. 0.16 sec. 0.16 sec. 0.12 sec.
19 impulses 15 impulses120 imp. 10 imp. 7 imp.
Lower premolar 0.40 sec. 0.32 sec. 0.16 sec. 0.16 sec. 0.12 sec.
24 imp. 19 imp. 25 imp. 10 imp. 7 imp.
Lower molar 0.50 sec. 0.40 sec. 0.20 sec. 0.20 sec. 0.16 sec.
30 imp. 24 imp. 32 imp. 12 imp. 10 imp.
Upper incisor/cuspid 0.40 sec. 0.32 sec. 0.16 sec. 0.16 sec. 0.12 sec.
24 imp. 19 imp. 10 imp. 10 imp. 7 imp.
Upper premolar 0.50 sec. 0.40 sec. 0.25 sec. 0.20 sec. 0.16 sec.
30 imp. 24 imp. 15 imp. 12 imp. 10 imp.
Upper molar 0.64 sec. 0.50 sec. 0.32 sec. 0.25 sec. 0.20 sec.
38 imp. 30 imp. 19 imp. 15 imp. 12 imp.
Reference conditions:
• Adult patient
• Anodic current 7 mA.
• Source to Detector Distance SSD = 250 mm (10 inches)
• Total (inherent) filtration 2 mm Al equip.
User and Service Manual
Unless the image is grossly over- or under-exposed, image density can be corrected by sub­sequent software processing (contrast/brightness — see Imaging Software User Manual and Installation Guide).
1
One impulse is equal to .017 sec. at 60 Hz. The recommended number of impulses can be adjusted as necessary for proper exposure, resulting in adequate image density and contrast and low noise gratuity.
Panoramic/cephalometric equipment
The DenOptix QST system requires no adjustments to the panoramic/cephalometric exposure settings (kV, mA or time). The DenOptix QST system can be used with almost any existing panoramic/cephalometric unit, regardless of age or manufacturer.
The use of a cassette without an intensifying screen is highly recommended. If you choose to use a cassette with an intensifying screen, you must at a minimum remove the front
t a low quality image due
screen. If t
ensifying screens are no
he int
t remo
ou will g
ed, y
v to the shielding effect from a decreased dose reaching the imaging plate. To remove the intensifying screens, follow the directions below.
28
e
Doc # M010-004WWE July 2005
• If you have a soft vinyl cassette, simply remove both intensifying screens from the cassette.
• If you have a hard sided cassette, remove intensifying screens (these are usually glued).
• Cassettes without intensifying screens are also available from Gendex.
Contact your local dental dealer or the panoramic/cephalometric equipment manufacturer for additional assistance.
5.2 Erasing the imaging plate
Imaging plates should be erased just prior to use. Scanning an imaging plate does not erase all the image information. To completely erase the imaging plate, expose the front surface (blue or white in color) to direct, intense light for 2 minutes. Fluorescent light is highly recommended.
An effective way to erase the imaging plate is to use a lightbox such as the RINN® Universal Viewer. We recommend fixing the lightbox under a cabinet. Make sure to lay the imaging plate down with the ACTIVE SIDE UP (blue or white face) facing the lightbox shining down. Expose the plates to light for 2 minutes.
Gendex DenOptix® QST
Alternatively, place the imaging plates within 8 inches of a lamp equipped with a lit 100-watt bulb for two minutes, minimum. The imaging plates are now completely erased and are ready for reuse.
To avoid scratching the imaging plates, DO NOT lay the imaging plate active (“blue or white”) face down on a lighbox.
Note: Erasing time will vary depending on the quality and intensity of the erasing light. For more detailed information, see appendix B.
User and Service Manual
29
Doc # M010-004WWE July 2005
5.3 Infection control
Prior to use, the DenOptix QST I/O imaging plates must be placed in barrier envelopes to minimize the possibility of cross-contamination. With proper application and use of the barriers, it will not be necessary to routinely cold sterilize the imaging plates. After erasing the imaging plates, follow these directions:
Gendex DenOptix® QST
Figure 5-1
Insert the imaging plate into the barrier envelope, ensuring that the black side of the imaging plate can be seen through the clear side of the barrier envelope. Pay careful attention to the location of the orientation dot.
Figure 5-2
With the imaging plate completely inserted, seal the envelope by removing the adhesive strip and pressing the envelope closed. The imaging plate is now protected and can be used intraorally.
User and Service Manual
30
Doc # M010-004WWE July 2005
5.4 Taking an X-ray
Intraoral imaging
Gendex DenOptix® QST
Figure 5-3
Ensure the computer, monitor and DenOptix QST scanner are switched on and properly connected. The green LED indicates that the system is ready to scan.
Figure 5-4
Launch your imaging software. Refer to your Imaging Software User Manual for details (VixWin shown).
Figure 5-5
Place an erased I/O imaging plate in the sealed barrier envelope and position in the patient’s mouth. Make sure that the blue side (front of the barrier envelope and, therefore, front of the imaging plate) is toward the X-ray source. Wear appropriate gloves and protective attire.
We recommend the use of a positioning device. The orientation dot should be posi-
tioned toward the occlusal surface for peri­apical projections. Fold back the barrier prior to inserting it into the positioning device. This will ensure that the imaging plate is firmly held in place. Ensure that the imaging plate (not the barrier) is in the center of the aiming ring. Expose the X-ray in the usual manner. Exposure settings should be in accordance with Section 5.1.
User and Service Manual
31
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Figure 5-6
Wearing gloves, remove the imaging plate in the barrier envelope from the patient’s mouth. Wipe off any excess saliva with a paper towel. Disinfect the barrier envelope by dipping it into a cold sterilant solution, if desired.
After each exposure, open the barrier envelope at the “V” cut in the center of the packet, with a steady pull, and let the plate fall into the transfer container (Dark Box provided with the system).
At the end of the intraoral examination, remove and dispose of gloves and move to a semi­darkened room or area (see Section 4.8). Remove lid and begin the process of loading the imaging plates.
Do not touch the transfer container with contaminated gloves.
Figure 5-7
Remove your gloves and wash any powder from your hands. Mount the I/O imaging plates onto the carousel by sliding the right side of the imag­ing plate into the correct size I/O mounting hole in the carousel. Ensure that the blue side faces out.
User and Service Manual
32
Repeat until all I/O imaging plates are mounted.
Figure 5-8
Open the carousel well cover on the DenOptix QST scanner and insert the carousel. Close the lid. You are now ready to scan your I/O imaging plates, as outlined in Section 5.5.
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Panoramic/cephalometric imaging
Figure 5-9
Place the erased panoramic or cephalomet-
ric imaging plate into the cassette with the
white side pointed toward the tubehead.
Insert the cassette into the panoramic or
cephalometric unit and expose in the usual
manner.
Figure 5-10
With the EO/CH carousel you can scan 4 #2 size imaging plates, 2 #0 size and 1 #4 size imaging plate. Follow the steps previously outlined for intraoral imaging but mount the I/O imaging plates on the panoramic carousel.
When scanning a Pan or Ceph plate, rotate the knob in the direction of Pan or Ceph. When no pan or ceph is loaded, rotate the knob in the direction of “NO PAN or CEPH”.
User and Service Manual
33
Figure 5-11
In a semi-darkened room (see Section 4.8), open the cassettes. With the white surface pointed away from the carousel, insert one side of the imaging plate under the edge of the panoramic or cephalometric imaging plate clip. Push with your thumbs until you hear a “click.” NOTE: If the IP does not click into place, slide the IP to one side then the other until it clicks into place.
Panoramic shown.
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Figure 5-12
Wrap the imaging plate around the carousel and push the free end under the edge of the other clip. Push until you hear a “click.”
Panoramic shown.
Figure 5-13
Open the carousel well cover on the DenOptix QST scanner and insert the carousel. Close the cover. You are now ready to scan your panoramic or cephalometric imaging plate.
User and Service Manual
34
Doc # M010-004WWE July 2005
5.5 Scanning the imaging plate
plates by simply clicking the appropriate plate. The plates in black will be skipped (not acquired).
Gendex DenOptix® QST
Figure 5-14
Once the scanner has been loaded and the patient file has been opened, click on the scanner icon on the imaging software toolbar. The software will automatically sense whether an FMX, EO or Ceph carousel is loaded.
Figure 5-15
When you scan a carousel, you can select from several options (subject to change, depending on the imaging software used).
It is possible to select how many plates to scan, and what size.
Plates can be arranged and saved to cus­tomized templates. You can select the
User and Service Manual
You can choose between 150, 300, and 600 dpi. Scan time will be longer as you chose high­er dpi. Make the appropriate choice by clicking on one of the combinations and then click­ing on “scan”.
NOTE: The more specific you are, the faster the scan time will be. For example, scanning the top row takes less than 30 seconds at 300 dpi. Scanning 20 imaging plates at 300 dpi. takes approximately 75 seconds.
Figure 5-16
emplates scanned will be
ou can also sa
6. Y
Doc # M010-004WWE July 2005
ve templates
35
The last four t
ed for quick retrieval by simply clicking
v
sa the “T” in the lower right as shown in
igure 5-1
F by selecting “templates” and save. After you name the template, you can retrieve this template by selecting “Templates” and “Load”. A window box will appear request­ing the name of the template to load.
Gendex DenOptix® QST
Figure 5-17
The procedure for scanning a panoramic/cephalometric plate is the same as for the intraorals.
It is possible to select how many plates to scan, and what size. As with intraorals, you can choose between 150, 300, and 600 DPI. Scan time will be longer as you choose higher DPI. To switch between panoramic and cephalo­metric, click on the “P” or “C” symbol in the lower right hand corner. To change between inch and metric, click “IN” or “CM”.
Choose your imaging plates by click­ing on each of the imaging plate boxes you want to acquire and then clicking on “scan”.
The plates in black will be skipped (not acquired).
To save the template, click on the template menu choice and save.
When the scan is complete, the I/O images will appear on the screen.
Note: if an error occurs during the scan, the scanner will stop. Follow the troubleshooting guidelines to start scanning again. If the resulting image is not diagnostic, erase the imag­ing plate and start again.
User and Service Manual
36
Doc # M010-004WWE July 2005
5.6 Preparing for the next patient
Once scanning is complete, it is time to get ready for the next patient.
Cross-contamination
There is no reason to routinely sterilize imaging plates unless you believe they have been con­taminated. If an imaging plate has touched a contaminated surface, wipe it gently with lint­free gauze dampened with a cold sterilant as recommended by the manufacturer. DO NOT SOAK. Two percent (2%) Gluteraldehyde solutions will not damage the imaging plate if used as described. IMAGING PLATES CANNOT BE AUTOCLAVED.
If a carousel has touched a contaminated surface, the intraoral imaging plate holders should be removed (see Section 7.1) and wiped with a cold sterilant solution. The remainder of the carousel can be sprayed with an aerosol disinfectant or soaked in a cold sterilant solution.
If the outside of the DenOptix QST scanner has been touched by a contaminated surface, it must be cleaned. BEFORE PROCEEDING, TURN THE SCANNER OFF AND DISCONNECT FROM THE POWER OUTLET. The outside of the scanner can now be wiped with a towel dampened with cold sterilizing solution and then allowed to air dry. DO NOT SPRAY OR SOAK THE SCAN­NER. If necessary, wipe the inside of the carousel well with a cloth very lightly dampened with cold sterilizing solution. NEVER SPRAY THE INSIDE OF THE CAROUSEL WELL. Be careful not to allow solvent into the DenOptix QST scanner, which could damage the electronics inside. Allow to air dry before connecting the power cord or turning the unit on.
Gendex DenOptix® QST
User and Service Manual
It is the operator’s ultimate responsibility to ensure that correct infection control procedures, consistent with those recommended by the Board of Dental Examiners, are followed and that effective cold sterilant and disinfecting sprays are used.
Follow the steps in Sections 5.2 and 5.3 to erase the imaging plate and protect it from cross­contamination. You are now ready to take an X-ray on the next patient.
37
Doc # M010-004WWE July 2005
5.7 Turning the DenOptix QST system on and off
The DenOptix QST System is designed to be left on continuously. The laser is only operational while the unit is actually scanning. If you wish to turn the system off, please remove the carousel and use the following procedure.
1. Wait until any scan in progress is complete and the green light is on.
2. Save any changes you have made and close down the imaging software (for example VixWin Pro software). To protect your data, frequently back-up your files.
3. Turn the power switch in the back of the scanner to the “Off” position.
4. Turn off all computer components following any instructions from the manufacturer.
If you live in an area of frequent thunderstorms or variable line voltages, you may want to turn the unit off on a more regular basis. An uninterruptible power supply (UPS) is recom­mended for areas with power quality issues.
Gendex DenOptix® QST
User and Service Manual
38
Doc # M010-004WWE July 2005
Software operating instructions
The DenOptix QST imaging systems can operate with different imaging software. For example, VixWin imaging software from Gendex optimizes digital X-ray images from the DenOptix QST systems. Please refer to your Imaging Software User Manual and Installation Guide for detailed instructions.
6
User and Service Manual
39
Doc # M010-004WWE July 2005
User and Service Manual
40
Doc # M010-004WWE July 2005
Maintenance procedures
The DenOptix QST System is designed for many years of trouble-free operation. It is manufac­tured from the highest quality components to ensure excellent performance. Maintenance that can be performed by the operator is minimal.
7.1 Cleaning the system
If the scanner, imaging plates or carousel becomes contaminated, clean in accordance with the directions given in Section 5.6. If, however, a component merely becomes soiled, clean as follows:
Cleaning the scanner
Turn off the DenOptix QST scanner, as described in Section 5.7, before cleaning. Wipe the outside surfaces with a paper towel dampened with a cold sterilant solution or household, non-abrasive cleaner (window cleaner works well). DO NOT SPRAY OR SOAK THE SCANNER. Wipe the inside of the carousel well with a very slightly damp cloth. Be careful not to allow running or dripping solvents into the DenOptix QST scanner. This could cause damage to the electronics inside. Allow to air dry before plugging in or turning back on.
7
User and Service Manual
Cleaning imaging plates
Imaging plates should be handled carefully. For the best images, take care not to scratch them and k
eep them dust-free. Use the following procedure to clean them:
on gauze (not cotton balls). Gently wipe the cotton gauze over the
tt
. Use lint-free, 1
1
dry imaging plate surface. Wipe back and forth and then in a circular motion.
2. To clean any remaining stains, dampen the gauze in anhydrous (water-free) isopropyl alcohol and wipe using t
3. Completely dry the surface by wiping with another piece of cotton gauze. Ensure that the imaging plate is completely dry before use.
00% co
he same motion as above.
41
Doc # M010-004WWE July 2005
Cleaning the carousel
Gendex DenOptix® QST
Figure 7-1
Remove the intraoral imaging plate holder from the carousel by removing the three screws along the vertical plastic retention piece. After removing the three screws, the intraoral plate holder will come free from the carousel.
Figure 7-2
Spray the soiled surfaces of the imaging plate holder and carousel with a general household non-abrasive cleaner or soap and water. Do not use strong alkaline or ammonia based cleaners. Wipe clean and allow to air dry.
User and Service Manual
42
Figure 7-3
Insert the dry imaging plate holder under the vertical plastic retention piece on the carousel. Slightly tighten the screws to hold the first side in place. Take the other side of the intraoral plate holder and place under the plastic reten­tion piece and tighten the screws.
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Cleaning the EO/CH (panoramic and cephalometric) carousel
Figure 7-4
To clean the EO/CH carousel, you will follow similar steps to the intraoral carousel. Remove the carousel imaging plate holder by removing the three screws of the plastic retention piece on both sides of the carousel. The imaging plate holder will easily come free.
Spray the carousel with a general household non­abrasive cleaning solution. Do not use strong alkaline or ammonia based cleaners.
Wipe clean and allow to air dry. If the intraoral imag­ing plate holder is soiled, remove it by removing both vertical plastic retention pieces by removing the three screws on each side and following the same instruc-
tions as for the intraoral image plate holder cleaning.
Clean with the general household non-abrasive spray or soap and water. Allow to dry com­pletely before re-assembling. Re-assemble by placing one side of the imaging plate holder under the plastic retention device and tighten screws. Repeat step on the other side of the imaging plate holder, replace plastic retention device and tighten the three screws.
User and Service Manual
43
Doc # M010-004WWE July 2005
7.2 Operator maintenance
Scanner maintenance
The only part of the DenOptix QST scanner that may require operator maintenance is the fuse. If you experience any other service-related problem, contact your local Gendex dealer. Follow these directions to change the fuse.
Gendex DenOptix® QST
Figure 7-5
Turn the DenOptix QST scanner off as per Section 5.7. Unplug the power cord and USB 2.0 cable.
WARNING: To avoid possible electrical shock, ensure that the power cord is unplugged.
Figure 7-6
Insert a small, flat tipped screwdriver into the top of the fuse cover on the back of the scanner. Pry the top of the connector open.
igure 7-7
F
Remove the fuse holder. Change both fuses. Ensure that new fuses meet the following spec­ifications: 250 V, 2 amp, Time Lag, High Breaking Capacity. Insert the fuses, REF 5101-0005, into the fuse holder.
User and Service Manual
44
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Figure 7-8
Insert the fuse holder and close the cover. Plug the power cord back in and turn the system on as outlined in Section 5.7.
Carousel maintenance
The carousel from the DenOptix QST scanner can be maintained by the operator. If the intra­oral imaging plate holder on an intraoral, panoramic or cephalometric carousel becomes damaged, it can be replaced by the operator. Order a replacement holder from your local Gendex dealer. The damaged holder can be removed and the new holder mounted as described in Section 7.1.
Imaging plate maintenance
The DenOptix QST imaging plates have no operator serviceable parts.
Lubrication
The DenOptix QST Digital System might occasionally require lubrication. If the scanner should start to squeak during scanning, please contact your local Gendex dealer. Do not position the DenOptix QST scanner in a dusty environment. Excessive dust levels can result in addi­tional scanner service, beyond normal service intervals.
User and Service Manual
45
Doc # M010-004WWE July 2005
7.3 Troubleshooting
Gendex DenOptix® QST
Trouble
No power No green light
Lid is Locked
Probable cause
Scanner not plugged in
Blown fuse
Main power switch or power supply is bad
Outlet does not have power
System is scanning
Power Failure or lack of power.
Corrective action
Check power cord at outlet and at the connection on the back of the scanner. Turn on the scan­ner, then turn on the computer and monitor.
See Section 7.2
Call your service representative.
Ensure that the outlet is grounded and has power.
Wait for scanning to end before attempting to open the lid. DO NOT OVERIDE THE LOCKING MECHANISM DURING SCAN­NING.
Unplug the scanner from the power source. Slide a thin tool or credit card into the gap between the lid and the scanner housing and simply depress the lock. You can then remove the carousel
User and Service Manual
Red, Green and/or Yellow LED(s) do not work; sys­tem operates normally
Defective LED Call your service representative
46
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Trouble
The Scanner does not ini­tialize when the software is opened
Probable cause
The scanner has not been turned on
The lid on the carousel well is open
The USB 2.0 cable between the scanner and computer is loose or defective.
The computer does not recognize that the scan­ner is connected or the scanner has not been added to the Device Manager
There is a hardware problem with the DenOptix QST scanner.
Corrective action
If the green power indicator light is off, switch on the DenOptix QST scanner per Section 5.7.
Close the lid.
Reconnect the cable. Check for tightness. Swap with a known good cable if possible.
Turn off the scanner and re-install the DenOptix QST systems software driver.
Turn the unit on.
Call your service representative
User and Service Manual
After scanning, no image appears on the monitor
There are no imaging plates on the carousel.
The imaging plate is mounted backwards on the carousel.
The imaging plate was erased prior to scan­ning.
Hardware failure.
X-ray source failed.
47
Ensure that the exposed imag­ing plate is properly mounted for scanning
The imaging plates are to be mounted with the white or blue surface visible. Turn incorrectly mounted plates around and rescan
Ensure that imaging plates are mounted under low light condi­tions as outlined in Section 4.8
Call your service representative.
Call your service representative.
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Trouble
Image is too dark
Image contains ghost images or shadows
Probable cause
Imaging plate has been overexposed
Imaging plate was not completely erased.
Imaging plate was exposed with the back of the IP facing the tubehead You may notice the writing from the back of the plate on the image.
Corrective action
Adjust brightness with software. If this is not possible, retake image at lower exposure. See exposure guidelines Section 5.1.
Increase the amount of time that is used to erase plates. Alternatively, increase the inten­sity of the light source. To test if an imaging plate is completely erased, simply mount an erased plate on a carousel and scan it. If no image is obtained, your procedure is effective at com­pletely erasing imaging plates.
Insure the imaging plates are inserted properly into the barri­er envelope and the proper ori­entation to the X-ray source is maintained
User and Service Manual
Imaging plates have been stored in barrier envelopes for too long a period.
Partial erasure of the image due to exposure to light during handling of the imaging plates.
48
Do not store imaging plates in barrier envelopes for more than one week.
Do not leave exposed imaging plates in well-lit areas. Even in the barrier envelope, some light penetrates and can partially
he imaging plates.
ase t
er Transfer imaging plates from their protective barriers to the DenOptix QST scanner as quickly as possible using the Dark Box (within one hour of exposure).
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Trouble
Image shows artifacts
The barrier envelope did not seal properly
Imaging plates fall out into the DenOptix QST scanner carousel well
Probable cause
The imaging plate sur­face is not clean and has dust, powder or stains on it. The surface may be scratched.
The imaging plate was removed from the panoramic cassette too quickly, which resulted in a static discharge
After the protective strip covering the adhesive on the barrier envelope was removed, the adhesive was touched or picked up dirt.
The imaging plates have not been loaded properly. See Section 5.4 for more information.
Corrective action
Clean the imaging plate as out­lined in Section 7.1. If the plate is scratched or stained, do not use the plate again.
Clean the imaging plate with anhydrous isopropyl alcohol AND remove from cassette more slowly.
Immediately seal the barrier envelopes after removing the adhesive strip.
Practice loading erased imaging plates in full daylight until you become familiar with the load­ing procedure. You will hear a “click” when the panoramic screen is loaded properly.
User and Service Manual
Extraoral plates don’t stay on the carousel during scanning
The red light stays on and the scanner stops
Broken clips Replace broken clips
Hardware failure
49
Call your service representative
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
7.4 Disposal of waste materials and inoperative parts
Disposal regulations vary from country to country. Therefore, it is difficult to give specific instructions on the disposal of DenOptix QST waste materials and inoperative parts. In gen­eral, we believe the following guidelines to be true.
Material Items Recycle? Comments
Plastic Barrier Envelopes No Dispose of with other
Carousel Latches No non-recyclable plastic items. Holders for I/O Plates No
Aluminum Housing, Optic Module Yes Remove all non-aluminum
Carousels Yes parts before recycling. Scan Engine Frame Yes Carriage, Optic Module Ye s
Storage Phosphors I/O Imaging Plates No Contain barium, which may be
Panoramic Imaging Plates No regulated. Contact your local Ceph. Imaging Plates No Gendex for more information
on what to do with damaged imaging plates.
Other Materials All Others No All other materials should be
sent to a landfill.
User and Service Manual
50
Doc # M010-004WWE July 2005
Storage and shipment
8.1 Storage
The DenOptix QST System has been designed for long-term operation in a normal office environ­ment. The system must be protected from adverse conditions such as excessive moisture, cold or heat. If the system is to be stored for a long period of time, pack the unit in its original carton. While in storage, keep at moderate temperatures and protect against moisture and humidity.
When you are ready to use the system again, unpack it and allow the scanner to come back to room temperature. Then, follow the guidelines outlined in Section 5.
8.2 Shipment
8
User and Service Manual
The DenOptix QST System is designed to be shipped in its original shipping container by normal commercial carriers. Ensure that the carousel has been removed from the scanner and that the carousel well is empty.
If the system is to be moved long distances or under adverse conditions, it should be protected. This can be accomplished by wrapping the DenOptix QST scanner in a plastic bag before putting it in the original packing materials.
If you are shipping the scanner, it is necessary to park the optics module in the center of travel prior to shipment. Contact your local Gendex rep for details.
51
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
User and Service Manual
52
Doc # M010-004WWE July 2005
Warranty statement
DenOptix QST scanner
The DenOptix QST Scanner and Carousels are designed expressly for use in a dental office environment and this warranty is not applicable to other uses. The scanner and carousels are warranted against defects arising from faulty materials or workmanship for two (2) years from date of purchase. Parts will be repaired or replaced at our option. The DenOptix QST Scanner and Carousels must be installed and operated in accordance with the Gendex written instruc­tions furnished with the unit.
DenOptix QST imaging plates
The DenOptix QST Imaging Plates are designed expressly for use with the DenOptix QST scan­ner. DenOptix QST Imaging Plates will be replaced if defective in manufacturing or packaging only. With proper handling, DenOptix QST Imaging Plates are designed for years of effective use. They can, however, be damaged if folded, creased, scratched or dented, and, therefore no additional warranty is provided.
9
User and Service Manual
53
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
User and Service Manual
54
Doc # M010-004WWE July 2005
Technical specifications
DenOptix QST scanner
Height 39.4 cm 15.5 in Width 49.3 cm 19.4 in Depth 27.4 cm 10.8 in Weight (empty) 16 kg 35 lbs Interface Cables USB 2.0 cables Voltage 100-240 V AC Frequency 50/60 Hz Power 110 VA max. Laser Classification Compliance per DHHS Radiation Performance Standards 21 CFR,
Ch I, Subch. J+EN60825 Class 1 Laser Device Operating Conditions 15° to 35°C (59° to 95° F); 5%-95% RH NC Storage and Shipping -40° to 70°C (-40° to 158°F) Temperatures Class I Grounded For Indoor Use Only Suitable for continuous operation
10
User and Service Manual
DenOptix QST intraoral imaging plates
Size 0 Size 1 Size 2 Size 3 Size 4 I/O dimensions 22x35 mm 24x40 mm 31x41 mm 27x54 mm 57x76 mm Recommended Dose Significant reduction in relation to typical D-speed film depending
upon type of X
orage Store in their clear plastic shipping box, at room temperature
t
S Operating and 18° to 35°C (64° to 95° F); 5%-80% RH NC
e Conditions
ag
or
t
S
-ray device.
55
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
DenOptix QST panoramic/cephalometric imaging plates
Dimensions Pan: 5x12 in; 15x30 cm Ceph: 8x10 in; 18x24 cm Recommended Dose Equal to standard panoramic systems equipped with rare-earth
intensifying screens (400 speed) Storage Store in their clear plastic shipping envelope Operating and 18° to 35°C (64° to 95° F); 5%-80% RH NC Storage Conditions
User and Service Manual
56
Doc # M010-004WWE July 2005
Appendix
11.1 Appendix A: Storage phosphor technology
The DenOptix QST System produces high quality X-ray images at very low doses as compared to intraoral X-ray film. The imaging plates in the DenOptix QST Systems are much more effi­cient at capturing X-ray energy than film. This means that the timer on your intraoral X-ray source can be turned down to a very low level without affecting the image quality.
11
User and Service Manual
How storage phosphor imaging plates work
Each DenOptix QST Imaging Plate is made up of a very thin layer of tiny storage phosphor crystals that are bonded together and coated on a flexible sheet of plastic. These storage phosphor crystals have the ability to capture the energy of X-rays and store the pattern as a
ent image (see diagram).
lat
57
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
In essence, they act as an “energy trap,” storing X-ray energy. The amount of energy stored is directly proportional to the amount of X-ray energy the crystal was exposed to.
The DenOptix QST scanner further excites the phosphor crystals to an unstable state by exposing them to a red laser. The phosphor crystals then release a blue light and return to their stable ground state. The DenOptix QST scanner reads this blue light and, with the imag­ing software, produces an image.
This process does not completely erase the imaging plate. Some crystals remain as “energy traps.” This information can be erased by exposing the imaging plate to light. A few seconds of low levels of light will usually not effect the image quality. Once erased, the imaging plate can be re-exposed and the process can begin again. With proper handling, DenOptix QST Imaging Plates can be continually reused.
11.2 Appendix B: Lighting conditions for handling or erasing imaging plates
Without light, storage phosphor imaging would not work. The red light from the laser scanner excites the exposed imaging plates, causing them to emit blue light. After your computer interprets the blue light and changes it into a digital image, light is used to erase the residual information from the imaging plate.
User and Service Manual
Between X-ray exposure and scanning, however, one must be careful handling the imaging plates. An imaging plate should never be removed from its protective barrier and exposed to a strong source of light. This act would erase the latent image from the plate. Therefore, you must load imaging plates onto the carousel under controlled lighting conditions. The recom­mendations below have been developed to ensure that you obtain excellent image quality.
Recommendations on lighting conditions for loading carousels
Fluorescent lights
If loading can be accomplished in less than 60 seconds, the fluorescent lighting in the area should not exceed 20 lux. If loading will take longer than 60 seconds, the room conditions
o be dar
need t
Incandescent lighting
Incandescent lighting is not as efficient as fluorescent lighting at fading imaging plates. We recommend lighting conditions of around 20 lux. Under incandescent lighting, however, you have about 2 minutes to complete the loading process. If it takes longer than 2 minutes, decrease t
ker. At 10 lux, the operator has about 5 minutes to accomplish loading.
0 lux.
he amount of light t
o 1
58
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
Sunlight
Sunlight is efficient light for erasing imaging plates but it is not recommended that IPs are left exposed to sunlight for extreme periods of time. Additionally, it is very difficult to judge the intensity of sunlight. Loading carousels under sunlight must be avoided.
If you cannot darken your mounting area to 20 lux or less, contact your local Gendex representative for suggested mounting times.
Recommendations on lighting conditions for erasing imaging plates
In order to completely erase imaging plates, 99.5% of the image information must be removed. Scanning only removes part of the information. To completely erase imaging plates, follow the guidelines given below. It is important that you measure the light intensity with a light meter at the point where the imaging plate will be positioned. The intensity of light changes dramatically as you move away from the source. Erasing the imaging plates for periods of time longer than suggested does not harm them. A good place to store your imaging plates is under the erasing light source.
Fluorescent lights
We recommend using fluorescent lights for erasing imaging plates. A lightbox typically gives off between 1000 and 5000 lux. Measure your lightbox and use the lowest value to deter­mine the erasing time. At 1000 lux, an imaging plate will be erased in 1 minute. At 2000 lux or more, 30 seconds is sufficient.
User and Service Manual
Incandescent lighting
At 1000 lux, an imaging plate will be erased in about 2 minutes. At 2000 lux or more, erase for 1 minute.
Sunlight
We do not recommend using it.
If you are unable to directly measure the light output from your erasing light source, or do not have an erasing light source greater than 1000 lux, use the practical test, below, to
ermine your erasing time.
t
de
Practical test
1. Completely erase five size 2 imaging plates by subjecting them to direct, intense light for about 20 minutes.
Scan the imaging plates. No image should appear on the monitor. If an image does appear,
2. repeat st
3. Place the imaging plates in barrier envelopes as described in Section 5.3.
ep 1 wit
h a more intense light source.
59
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
4. Using the exposure conditions outlined in Section 5.1, expose each imaging plate. If pos­sible, place an object, such as an extracted tooth, in the path of the X-ray beam.
5. Turn on your erasing light source. We recommend a light box or a minimum l00-watt incandescent light bulb (see Section 5.2).
6. Remove one imaging plate from its barrier envelope and place it with the blue side toward the light for 2 minutes.
7. Scan the imaging plate. If an image appears, the plate was not completely erased. Repeat step 6, adding a minute to the exposure time. Continue to repeat until no image appears on the monitor. Record the erasing time that resulted in no image. This is the time required to erase an unscanned imaging plate.
8. Repeat steps 1-4. This time, under low light conditions, remove the imaging plates from the barrier envelopes and mount them on a carousel. Scan the imaging plates.
9. Remove the scanned imaging plates and place them, in the dark, into a light-protected area (such as a light-tight drawer).
10. Now repeat steps 6 and 7, exposing one plate (blue side toward light) to the erasing light
for 2 minutes, the next plate for 3 minutes, etc., until you find the point where no image appears on the monitor. This is the time required to erase a scanned imaging plate.
The time required to erase an imaging plate will vary depending on the quality and intensity of the erasing light source. It should be fairly easy to find a light source that will erase imaging plates in 2-3 minutes.
User and Service Manual
60
Doc # M010-004WWE July 2005
11.3 Appendix C: Optional printer
Because the DenOptix QST system can be interfaced with different applications software, and printer technology is constantly evolving, we can only suggest a list of suitable technical requirements for printers.
To interface with the PC, the printer must have a digital input and not a video (analog) input.
Using the standard Windows® drivers allows the use of the same printer for all other Windows applications (word processing, etc.).
Inkjet printers
Black and white or color hard copy. Printers should have a minimum of 600 dpi resolution and pseudo-randomic dithering.
We recommend using the photographic quality glossy paper and printing on the highest resolution setting available.
Thermal printers
Thermal printers print in shades of grey with at least 64 levels. Thermal printers use a special type of paper which guarantees a print life of about 3 years.
Gendex DenOptix® QST
User and Service Manual
Dye sublimation printers
Suitable for high printing quality whenever the DenOptix QST System is used in combination with an intraoral camera.
Laser printers
If a Laser Printer is adopted, contact the Printer manufacturer or a representative to make sure that it can achieve at least 256 gray shades. Non medical grade Laser Printers may not be able t
o achieve a good image quality.
Application software
Please refer to the instructions in the application software for special information about
s.
er
print
61
Doc # M010-004WWE July 2005
11.4 Appendix D: If you need assistance
The DenOptix QST System is designed to provide years of trouble-free service, and carries a limited warranty. If you need assistance, first contact your local Gendex dealer. They have been trained to handle most technical service issues. If you cannot get the information you need there, simply call the Gendex office nearest to you.
Gendex DenOptix® QST
North America
USA - Canada
KaVo Dental - Gendex Dental Systems 340 E. Main Street Lake Zurich, IL 60047 USA
Customer Service:
Tel. +1.888.275.5286 Fax +1.847.640.4891
Technical Service:
Tel. +1.800.769.2909 Fax +1.847.640.5310
www.gendex.com
Europe
Italia-European Headquarters
KaVo Dental - Gendex Dental Systems Via A. Manzoni 44 - 20095 Cusano Milanino - MI Tel. +39.02.618008.1 Fax +39.02.618008.09 www.gendex-dental.com
Deutschland
KaVo Dental - Gendex Dental Systems GmbH Albert-Einstein-Ring 15 - 22761 Hamburg Tel. +49.40.899688.0 Fax +49.40.899688.19 www.gendex.de
France
KaVo Dental - Division Imagerie Gendex ZAC Paris Nord 2 Parc des Reflets - BP 46044 95912 Roissy CDG Cedex Tel. +01 56 48 72 00 Fax + 01 56 48 72 25 www.gendex-dental.com
España
KaVo Dental S.L. División Radológica C/ Joaquín María López, 41- 28015 - Madrid Tel. +34 915.493.700 Fax. +34 915.437.054 www.gendex-dental.com
User and Service Manual
62
Other
Asia - Centr
KaVo Dental - Gendex Dental Systems Via A. Manzoni 44 - 20095 Cusano Milanino - MI
y
al
It Tel. +39.02.618008.1 Fax +39.02.618008.09 www.gendex-dental.com
al & Sout
h Amer
Doc # M010-004WWE July 2005
ica
Gendex DenOptix® QST
User and Service Manual
63
Doc # M010-004WWE July 2005
Gendex DenOptix® QST
User and Service Manual
64
Doc # M010-004WWE July 2005
User and Service Manual
65
Doc # M010-004WWE July 2005
© 2005 Gendex Dental Systems – M010-004WWE
Loading...