Gendex DC User manual

4.3 (3)

1- Opening and Closing Procedures

Master Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Opening the Master Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Local Configuration Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Remote Configuration Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Closing the Master Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Articulated Arm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Opening an End of the Articulated Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Closing an End of the Articulated Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

2- Checks

System Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

3- Adjustments and Maintenance

Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2

Tubehead - Horizontal Drifting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Horizontal Arm - Horizontal Drifting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Articulated Arm - Horizontal Drifting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Articulated Arm - Vertical Drifting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Outer Section. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Inner Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

11

Articulated Arm - Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

i-i

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

4- Configuration Settings

Switch Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

SW2 - Setting for X-Ray Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 SW2 - Setting for Optional 12" (30 cm) Cone . . . . . . . . . . . . . . . . . . . . . . . 3

Jumper Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

JP1 - Setting for 120 V ac . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

5- Problem Resolution

Converter Board (part no. 124-0292) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Converter Board - LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Converter Board - Test Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Converter Board - Connectors and Jumpers . . . . . . . . . . . . . . . . . . . . . . 12

Logic Board (part no. 124-0293) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Logic Board - LEDs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Logic Board - Test Points. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Logic Board - Connectors and Jumpers . . . . . . . . . . . . . . . . . . . . . . . . . 18

Electrical Block Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

6- Replacement Parts

Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2

Articulated Arm and Tubehead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Horizontal Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Master Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Master Control Touch Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

i-ii

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

1- Opening and Closing Procedures

WARNING

Before starting any procedure, review the safety information provided in the Gendex expert® DC User Manual.

1-1

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

Opening the Master Control

Master Control

Opening the Master Control

Tools Required:

No. 2 Phillips Screwdriver

 

 

There are 2 configurations for the Master Control:

Local - Master Control Touch Panel is mounted directly to the Master Control

Remote - Master Control Touch Panel is mounted on a wall plate

1.Turn OFF the power switch.

2.Disconnect the external power source:

FOR A SYSTEM WITH A LINE CORD (120 V ac only), unplug the line cord.

FOR A HARD-WIRED SYSTEM, disconnect the mains supply (external power source) and use a meter to verify the mains supply is disconnected.

3.For Local configurations, continue with Local Configuration Instructions.

For Remote configurations, continue with Remote Configuration Instructions on page 1-6.

Local Configuration Instructions

To open the Master Control when the Master Control Touch Panel is mounted directly, (Local configuration):

1. Remove the Master Control Touch Panel:

(Continued)

1-2

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

Opening the Master Control

(Continued)

a. Pull out at the top of the Master Control Touch Panel.

Figure 1-1 Removing the Master Control Touch Panel

b.Disconnect the control cable and the exposure switch coil-cord or wall switch cord.

Figure 1-2 Disconnecting the Cable and Cord

c.Lay the Master Control Touch Panel aside. This will be re-installed later.

2.Position the horizontal arm so that it is parallel to the wall.

(Continued)

1-3

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

Opening the Master Control

(Continued)

3. Remove the small top cover by pulling up at the corner.

Figure 1-3 Removing the Top Cover

4.Grip the outer cover at the top and at the bottom rear hand-grip cutout and carefully pull it off.

Figure 1-4 Removing the Outer Cover

5.Pull the control cable and the coil-cord back through the cutout for the cables while removing the outer cover.

6.Lay the covers aside for later re-installation.

(Continued)

1-4

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

Opening the Master Control

(Continued)

7.Unscrew the Strain Relief to remove the control cable and coil-cord from the door.

Figure 1-5 Removing the Control Cable and Coil-Cord

8. Remove the two Phillips head screws and open the door carefully.

Figure 1-6 Opening the Door

1-5

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

Opening the Master Control

Remote Configuration Instructions

To open the Master Control in systems with the Master Control Touch Panel mounted on a wall plate (Remote configuration):

1. Pull out at the top of the blank panel.

Figure 1-7 Removing the Blank Touch Panel

2.Lay the blank panel aside. This will be reinstalled later.

3.Position the Horizontal Arm so that it is parallel to the wall.

4.Remove the small top cover by pulling up at the corner, and lay the cover aside; this will be reinstalled later.

Figure 1-8 Removing the Top Cover

(Continued)

1-6

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

Gendex DC User manual

Opening the Master Control

(Continued)

5.Grip the outer cover at the top and at the bottom rear hand-grip cutout and carefully pull it off.

Figure 1-9 Removing the Outer Cover

6.Lay the covers aside to be re-installed later.

7.Remove the two Phillips head screws and open the door carefully.

Figure 1-10 Opening the Door

1-7

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

Closing the Master Control

Closing the Master Control

Tools Required:

No. 2 Phillips Screwdriver

 

 

1.Close the door and reinstall the two Phillips screws.

2.Reinstall the outer cover and the top cover.

3.For Local configurations, continue to step 4. For Remote configurations, skip to step 5.

4.For the Master Control Touch Panel that is mounted directly to the Master Control:

a.Put the cables back in the Strain Relief.

b.Pull the control cable and the coil-cord back through the cutout for the cables when reinstalling the outer cover.

c.Connect the control cable and the exposure switch coil-cord to the Master Control Touch Panel before attaching the panel to the outer cover.

d.Skip to step 6.

5.Install the Blank Plate in the outer cover.

6.Make sure all parts and screws are back on.

7.Restore power to the equipment:

FOR A SYSTEM WITH A LINE CORD (120 V ac only), plug in the line cord.

FOR A HARD-WIRED SYSTEM, reconnect the mains supply (external power source) and use a meter to verify the mains supply is connected properly.

8.Turn ON the power switch.

1-8

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

Opening an End of the Articulated Arm

Articulated Arm

Opening an End of the Articulated Arm

1.Turn OFF the power switch.

2.Disconnect the external power source:

FOR A SYSTEM WITH A LINE CORD (120 V ac only), unplug the line cord.

FOR A HARD-WIRED SYSTEM, disconnect the mains supply (external power source) and use a meter to verify the mains supply is disconnected.

CAUTION

There are three pins inserted into posts that hold the two sections of the Trim Cover together. Do not use excessive force when separating the two sections as the pins can be broken. If it is necessary to use a small screwdriver to separate the two sections, position the screwdriver at the location of the pins as shown in Figure 1-11.

Figure 1-11 Locating the Three Pins Inside the Articulated

Arm Trim Covers

(Continued)

1-9

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

Closing an End of the Articulated Arm

(Continued)

3.Separate and remove the plastic Trim Covers from the end of the section of the Articulated Arm. Set the Trim covers aside.

Inner Trim Covers

Outer Trim Covers

Figure 1-12 Articulated Arm Trim Covers

Closing an End of the Articulated Arm

1.Place the two sections at the end of the Articulated Arm, one on each side. Press the two sections together ensuring that the pins of the one section are aligned correctly inside the posts of the other section.

2.Restore power to the equipment:

FOR A SYSTEM WITH A LINE CORD (120 V ac only), plug in the line cord.

FOR A HARD-WIRED SYSTEM, reconnect the mains supply (external power source) and use a meter to verify the mains supply is connected properly.

3.Turn ON the power switch

1-10

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

2- Checks

To prevent unnecessary problems in the future, perform the following set of checks as part of the recommended maintenance as indicated in the User Manual.

WARNING

Failure to perform these checks may result in a system that fails to comply with U.S. Radiation Performance Standards 21 CFR Subchapter J.

CAUTION

To avoid any potential hazard to operators or patients, any unusual operation, problems with mechanical functions, or presence of debris, should be reported to Gendex immediately.

If problems persist, advise the owner NOT TO USE THE SYSTEM.

2-1

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

System Functions

WARNING

Before starting any procedure, review the safety information provided in the Gendex expert® DC User Manual.

If problems are found, refer to the appropriate section in this Service Manual for:

Adjustment procedures for the Horizontal Arm, Articulated Arm, and Tubehead (page 3-1)

Maintenance procedure for the Articulated Arm (page 3-11)

Troubleshooting information (page 5-1)

Electrical block diagrams (page 5-20).

IMPORTANT! Contact GENDEX Technical Support if additional information is required.

____ 1. Tubehead - Check for oil leaks or other evidence that could indicate internal damage. Replace the Tubehead if necessary.

____ 2. Tubehead Rotation - Ensure that the Tubehead maintains its position around the horizontal axis while remaining easy to rotate and position. For horizontal dropping, refer to page 3-2 for the procedure.

Also check for unwanted pivoting of the Tubehead around the vertical axis. Problems with vertical pivot are typically caused when the Horizontal Arm is not properly leveled although the problem can simply be caused when the cabling inside the Tubehead Yoke is too tight.

____ 3. Mounting - Be sure that the wall support is adequate and that the system is properly mounted to the wall.

____ 4. Power Switch - Verify that the switch is working properly and that the Ready Indicator Lamp is illuminated when the power switch is in the ON position.

(Continued)

2-2

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

(Continued)

____ 5. Master Controls - With the power switch in the ON position, verify that a Time Selection value indicates on the Master Control Touch Panel display.

Also check the function of the selector switches for the Anatomical Time Selection, Imaging Type Selection and Patient Selection. Pressing the selection buttons should cause indicator lamps to indicate the selected item.

____ 6. Quickset Tubehead Control - Verify that the indicator lamps are consistent with the Master Control selections. Verify that the selector button works properly.

____ 7. Push-button exposure switch - Verify that the push-button switch in the face of the operator controls is functioning properly.

____ 8. Coil-cord exposure switch - If the coil-cord switch is used, inspect the switch housing and coiled cord for damage or wear. Replace if there is evidence of damage present.

____ 9. Exposure Indicators - Make several exposures and verify that the Radiation Indicator Lamp illuminates and the audible signal is heard.

____ 10. Premature Termination - Select an exposure of 2 seconds using the manual adjustment buttons. Initiate an exposure, but release the exposure switch after a brief period of time before the timer terminates the exposure. Verify that exposure terminates immediately upon release of the exposure switch.

____ 11. X-ray Beam Size - Position the cone of the Tubehead directly onto a panoramic film or several occlusal films taped together to form a large rectangle. Process the film and verify that the resulting image is containable in a 6 cm circle.

____ 12. Check that the arms are evenly balanced and that all movements are smooth and quiet. Verify that the Horizontal and Articulated Arms do not drift.

2-3

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

This page intentionally left blank

2-4

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

3- Adjustments and Maintenance

Procedures include adjustments for the:

Tubehead (page 3-2)

Horizontal Arm (page 3-2)

Articulated Arm (page 3-5)

Procedures include maintenance for the Articulated Arm (page 3-11).

Preliminary Information

The balance of the Articulated Arm is initially set at the factory. Other friction and drift adjustments are set during the installation process.

Adjustments provided in this document are to accommodate customer preferences or changes in the leveling of the system due to age.

Note! The mechanical adjustments should not be used to compensate for a system that is not properly leveled on the wall.

Before making any adjustments, first verify that the unit is properly leveled on the wall; then make adjustments as necessary.

WARNING

Before starting any procedure, review the safety information provided in the Gendex expert® DC User Manual.

3-1

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

Tubehead - Horizontal Drifting

Adjustments

Tubehead - Horizontal Drifting

The Tubehead will drift from the correct horizontal position if the four hex locking nuts are not adjusted properly. This can happen when the supplied 8" (20 cm) cone is replaced with the longer, heavier 12" (30 cm) cone.

Note! The adjustment of the four hex locking nuts does not lock the Tubehead in place but provides proper friction to allow the Tubehead to be rotated and then to remain in the position to which it has been rotated.

Tools Required:

2.5 mm Allen Wrench

4.5mm Hex Wrench

1.Turn OFF the power switch.

2.Disconnect the external power source:

FOR A SYSTEM WITH A LINE CORD (120 V ac only), unplug the line cord.

FOR A HARD-WIRED SYSTEM, disconnect the mains supply (external power source) and use a meter to verify the mains supply is disconnected.

3.Unscrew the two screws on the Tubehead Yoke Cover Plate, using a

2.5mm Allen wrench and remove the cover plate from the underside of the Tubehead yoke. (It may be necessary to position a thin screwdriver at the base and gently pry off the cover.)

Figure 3-1 Removing the Tubehead Yoke Cover Plate

(Continued)

3-2

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

Tubehead - Horizontal Drifting

(Continued)

4.Tighten the locking nuts:

a.Rotate the Tubehead to its maximum position (until it stops).

b.Locate the first hex locking nut, and tighten it with the 4.5 mm hex wrench.

Figure 3-2 Two of the Four Hex Locking Nuts

c.Rotate the Tubehead until you see the next hex locking nut. Tighten the nut.

d.Repeat step c. for the remaining two hex nuts.

Note! Tighten all four hex nuts equally, to apply adequate friction to prevent drifting.

5.Reposition the cover plate and snap it into place. Screw in the two screws that secure the cover plate.

6.Restore power to the equipment:

FOR A SYSTEM WITH A LINE CORD (120 V ac only), plug in the line cord.

FOR A HARD-WIRED SYSTEM, reconnect the mains supply (external power source) and use a meter to verify the mains supply is connected properly.

7.Turn ON the power switch.

3-3

032-0207-EN Rev 0

Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER

Loading...
+ 43 hidden pages