GE Healthcare DPX-NT Troubleshooting

Page 1
4 1112
Troubleshooting
Chapter 1:Troubleshooting
This chapter contains troubleshooting techniques for diagnostic failures, failing quality assurance tests, mechanical failures and imaging problems.
4.0 Diagnostic Failure Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-119
4.1 Transverse Motion failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-119
4.1.1 Operator Induced - switch closed during scan . . . . . . . . 4-119
4.1.3 Loss of OMI signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-121
4.2 Longitudinal Motion failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-122
4.2.1 Limit Switch Tripped During a Scan . . . . . . . . . . . . . . . . 4-122
4.2.2 Longitudinal Binding . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-123
4.3 Failure of the 28V power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-126
4.4 Emergency Stop Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-127
4.5 Tube Head Thermostat. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-127
4.6 Communication Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-127
4.7 Other Diagnostic Failure Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-127
4.7.1 Reasons For Invalid Diagnostic Failures. . . . . . . . . . . . . 4-128
4.8 Failing Quality Assurance Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-128
4.8.1 Block Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-128
4.8.2 Beam Stop Action . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-128
4.8.3 Mean% Spillover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-128
4.8.4 Reference Counts and Ratio . . . . . . . . . . . . . . . . . . . . . . 4-129
4.8.5 Ratio Fluctuations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-129
4.8.6 Transverse or Longitudinal Mechanics . . . . . . . . . . . . . . 4-129
4.8.7 Tissue Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-130
4.8.8 Bone Mineral of the Standard Chambers . . . . . . . . . . . . 4-130
4.8.9 Symptoms of High and Low KV. . . . . . . . . . . . . . . . . . . . 4-131
4.9 Reference Counts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-131
4.10 Arcing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-134
4.10.1 Limit Switch Tripped During Scan . . . . . . . . . . . . . . . . . 4-135
4.11 Imaging Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-136
4.11.1 White, or Grey in the first or second scan line: . . . . . . . 4-136
4.11.2 Femur Scan Problems . . . . . . . . . . . . . . . . . . . . . . . . . . 4-136
4.11.3 AP-Spine Image Problems: Probable causes . . . . . . . . 4-137
DPX-NT Service Manual (Rev A- 1999) Troubleshooting4117
Page 2
4.11.4 Broken Signal Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-137
4.11.5 Loss of tube head current . . . . . . . . . . . . . . . . . . . . . . . 4-137
4.11.6 X-Ray Relay Failure. . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-137
4.11.7 Unstable AC Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-138
4.11.8 Arcing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-138
4.12 Failing Alignment Test Results . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-138
4.12.1 Image. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-138
4.13 Indicator Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-140
4.13.1 X-ray On LED Blinking. . . . . . . . . . . . . . . . . . . . . . . . . . 4-140
4.13.2 Shutter Open LED Blinking . . . . . . . . . . . . . . . . . . . . . . 4-140
4.13.3 Shutter Not Operating . . . . . . . . . . . . . . . . . . . . . . . . . . 4-141
4.13.4 End of Exposure Alarm During Scan . . . . . . . . . . . . . . . 4-141
4.14 Communications Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-142
4.15 Viewing Quality Assurance Trends . . . . . . . . . . . . . . . . . . . . . . . . 4-142
4.15.1 What to Look for in the QA History . . . . . . . . . . . . . . . . 4-143
4.16 MAX Board Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-143
4.17 FOINK Board Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-144
4.18 OMI Board Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-145
4.19 SBC Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-145
4.20 XORB Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-147
4.21 Detector Motherboard Troubleshooting . . . . . . . . . . . . . . . . . . . . 4-148
4.22 Detector Daughter Board Troubleshooting . . . . . . . . . . . . . . . . . . 4-149
4-118Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 3

4.0 Diagnostic Failure Codes

The following conditions halt the operations of the scanner, and generate a diagnostic failure message which is displayed on the monitor screen:
All of these interrupts are detected by the cSBC
• Transverse Mechanics Failure (4.1)
• Longitudinal Mechanics Failure (4.2)
• Emergency Stop Button activated (4.3)
• X-ray source over temperature (4.4)
• DC power supply failure (4.5)

4.1 Transverse Motion failure

Motion Detection
• cSBC / OMI fail to see transverse motion during a patient scan, an interrupt signal is generated and the cSBC resets and closes the shutter.
• A slotted disk at the end of the arm rotates through an infrared beam (OMI), and pulses are sent to the cSBC board when there is transverse motion.
• The transverse and longitudinal motion detection system (on cSBC) is operational during patient scans and the “find block” test of the daily QA. If the problem occurs during a quality assurance, it may be high voltage arcing or a problem with the OMI.

4.1.1 Operator Induced - switch closed during scan

Cause: When scanning, a defect in the transverse mechanics or a mis­positioned patient may cause one of the transverse limit switches to close when the scanner is acquiring data. Closure of a limit switch prevents further operation of the motor. The shutter will close, the shutter open lamp will go out and the end of exposure alarm will sound. Seconds later the error message will appear on the screen.
Solution: If the patient is not centered on the table top or if the region being scanned is too close to a limit in transverse travel, the limit switch switch may be close while scanning. Re position the patient on the table, further away from the limit.

4.1.2 Mechanical Failures - Unusual noise or irregular motion

DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-119
Page 4
If the Detector / Tube Head motion is irregular, or scraping noises are heard, the Transverse Mechanics may be binding. This symptom will typically be detected as a failure by the daily QA Mechanics Test.
Symptom: If the Detector Array / Tube Head is not moving when scanning check:
• Mechanics are free to move
• Centent
• Stepper Motor
Troubleshooting Binding
Turn off the power to the scanner and move the affected parts by hand. Feel the motion for spots where the carriages are more difficult to move. Listen for unusual noises.
The following are common causes for Transverse Binding
• High Voltage Cable Routing
One of the most common problems is a failure of the Source and Detector to reach the rear Limit Switch due to the Tube Head running into its own high voltage cables. These cables must have a hump formed at the Rear Longitudinal Carriage that allows the lower portion of the Tube Head to pass under the cables.
The High Voltage Cables can also impede transverse motion toward the front Limit Switch. This is caused when the cables have been tied down without enough play for the Tube Head and Detector to move all the way forward. These problems should be investigated by manually tripping both Limit Switches while inspecting for cable conflicts, binding, or tension problems.
• Check to see that the bottom of the Tube Head is not hitting the Transverse Centent
•Wiring
In rare instances, the wires from the Shutter Solenoid and/or Fans can snag on the bolts that protrude through the frame on the foot end of the scanner. This is solved by properly tying down these wires.
• Transverse Belt
The Transverse Belt should not be excessively tightened or this will cause excessive binding in the transverse mechanism. It should be possible to deflect the belt by 4 cm when it is properly tightened. Sometimes the spare belt material near the clamp on the Tube Head Carriage comes into contact with the forward gear and prevents the scanner from going all the way to Home position.
• Drive Wheels
4-120Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 5
The wheels that support the Tube Head and Detector Carriages must be adjusted so that they come into perfect contact with the Transverse Rails. These wheel are best inspected with the table top, front panel and arm covers off so one can sight down the extrusions. However, it is possible to test the wheels by preventing any wheel from turning and seeing if the carriage will still move. By preventing any wheel from turning, you should be able to tell that it slides along the extrusion while the others roll. This indicates that the wheel has not been excessively tightened down. This is least likely to be the source of transverse motion problems, as it is unlikely that the adjustment of the wheels would have become any tighter over time. They would be more likely to loosen over time and fail alignment tests (see Alignment Test, section
4.12). Adjustments can be made by loosening and rotating the eccentric bearings of any of the lower wheels.
• Gear And Pulley Positioning
Check all appropriate gears and pulleys. Verify that the set screws are tightened and the gears and pulleys are not out of position.
• Limit Switch Positioning
If the transverse motion seems to be acceptable, but the number of steps in the Quality Assurance Test is failing, check the position of the Limit Switches.
On Total Body scans, a limit switch out of position could allow the Source/ Detector Carriages to hit the frame or panels before the Limit Switch is actuated.
• Transverse Motor
Check the Transverse Motor for a broken wire in one of the internal coils, or a bad electrical connection to its Centent Motor Controller.
• Transverse Centent
The Centent Motor Controller, if defective, will cause transverse motion problems. Sometimes the controller works well enough to acquire some scans, but it will not provide enough torque to complete every fast scan. The CURRENT SET voltage at terminal 11 on a properly operating controller should be 14 to 16 VDC (transverse motor wired in parallel).
The longitudinal and transverse Centents are identical, and can be exchanged. If the problem remains, the Transverse Motor should be replaced.

4.1.3 Loss of OMI signal

If the error occurs consistently on the first line of a patient scan or during the find block portion of the daily QA, and the scanner is moving in the transverse direction, then check the following:
Cause: The Interrupt signal is being lost.
DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-121
Page 6
Solution 1: The pulses that normally enter the cSBC board at J14 (the black wire at the center of the connector) may have stopped. These pulses are necessary to keep the cSBC from resetting. These pulses can be seen on a cSBC board LED.
Use the DPX-NT service software (Tools/Diagnostics/Scanner Motion / Motion Commands Tab) to set the joystick speed to 50 steps, enable the joystick and watch the LED. If the OMI is working the LIght will flash when the Transverse Motor is run. If the LED flashes when the mechanics are engaged, but the error still occurs, the interrupt was invalid. Check for arcing in the high voltage system or replace the cSBC.
If the LED D9 (B in figure 4-1) does not flash,
• Verify that the slotted disk at the front of the lower arm rail is in the middle of the slot between the photo diode and photo transistor.
• The slotted disk must be completely flat and remain in the center of the sensor slot during its entire rotation.
• If the slotted disk has been in physical contact with the optical sensor, the sensor may have debris on it, disassemble this mechanism, and clean the sensor and the slots of the disk.
• Check the Cable running from the cSBC to the OMI for a cable break by checking the individual wires for continuity.

4.2 Longitudinal Motion failure

• cSBC/ OMI fail to see longitudinal motion during a patient scan, an interrupt signal is generated by the cSBC which resets and closes the shutter.
• A slotted disk at the foot end of the table on the pulley for the Longitudinal Drive Belt rotates through an infrared beam (OMI), and pulses are sent to the cboard when there is transverse motion.
• The transverse and longitudinal motion detection system (on FOINK) is operational during patient scans and the find block portion of the daily QA. If the problem occurs during a quality assurance, for instance, it may be the high voltage system is arcing or the OMI may be malfunctioning.

4.2.1 Limit Switch Tripped During a Scan

4-122Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 7
Cause: When scanning, a defect in the longitudinal mechanics (binding) or a mis-positioned patient may cause one of the longitudinal limit switches to close when the scanner is acquiring data. Closure of a limit switch prevents further operation of the motor. The shutter will close, the shutter open lamp will go out and the end of exposure alarm will sound. Seconds later the error message will appear on the screen.
Solution: If the patient is not centered on the table top (length wise) or if the region being scanned is too close to a limit in longitudinal travel, the limit switch may be close while scanning. Re position the patient on the table, further away from the limit.

4.2.2 Longitudinal Binding

• If the scan arm fails to move when scanning check the stepper motor, and Centent, if arm motion is irregular, check for binding.
Turn off the power to the scanner and move the affected parts by hand. Feel the motion for spots where the arm is more difficult to move. Listen for unusual noises.
When moving the arm longitudinally, only push on the arm column. Pushing on the forward parts of the arm can ruin Tube Head/Detector alignment.
• Cable Track
The major impediment to longitudinal motion is the plastic Cable Track that runs through the trough at the rear of the scanner.
This track is attached at two points: at the Rear Longitudinal Carriage and to the scanner frame (low, rear and center) each spot by 4 bolts. Also, adequate slack must be left in the cables inside the Cable Track or they will stop the arm from moving fully to the foot end.
Should the Cable Track detach from the scanner frame, it will slide freely in the trough and will eventually cause trouble. This can allow the Cable Track to get in between the Rear Longitudinal Carriage and the scanner frame on the foot end preventing the tripping of the limit switch.
Front Longitudinal Carriage Dragging
Check the distance between the front longitudinal carriage and the longitudinal rail with a go/nogo gauge (See installation Procedure DXAP2000 Chapter 5 appendices). The carriage should not rub the front rail, if necessary insert shims behind the front longitudinal carriage.
• Tube Head Cable Routing
DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-123
Page 8
Problems at the head end of the scanner can be caused by a limit switch being out of position (thus the arm runs into the frame before the Limit Switch) or because the High Voltage Cables are tied down incorrectly where they snake around from the Rear Longitudinal Carriage to the panel under the Tube Head (High Voltage Cable Trough). If the cables are tied down too far toward the head end (at the point where the three tie downs are), they can run into the Longitudinal Motor Assembly and hold the arm away from the head end Limit Switch.
• Slip Clutch
A slip clutch is part of the longitudinal motion system to limit torque. This is a feature to protect the patient should he/she pinch an arm or leg between the back side of the scanner and the Arm Column. If the Slip Clutch is set too loose, it will fail to move the belt and will just "slip" as the motor turns. This may produce the following symptoms:
• failure of Quality Assurance scan Mechanics test
• compressed image in limited areas of the image
• a Longitudinal Mechanics diagnostic failure
• Gear and Pulley Positioning
Check all appropriate gears and pulleys. Verify that the set screws are tightened and the gears and pulleys are not out of position.
• Limit Switch Positioning
If the longitudinal motion seems to be acceptable, but the number of steps in the Quality Assurance Test is failing, check the position of the Limit Switches.
On Total Body scans, a limit switch out of position could allow the Source/ Detector Carriages to hit the frame or panels of the scanner before the Limit Switch.
If the mechanical stop is reached before the Limit Switch is actuated, check carefully to see which part of the Arm Assembly is in contact with the Table Assembly. The front part of the Lower Transverse Extrusion is clamped to the Longitudinal Drive Cable at the front of the scanner. If the Lower Transverse Extrusion is not clamped in such a way that it forms a 90 degree angle with the length of the table, the rollers at the front end of the Lower Transverse Extrusion may strike the end of the scan table before the Limit Switch is actuated.
• Longitudinal Motor
Check the motor for a broken wire in one of the internal coils, or a bad electrical connection to the Centent Motor Controller.
• Longitudinal Centent
4-124Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 9
The longitudinal Centent Motor Controller may be the cause of a failure. The current set voltage should be roughly 9-11 volts at terminal 11 on a properly operating controller. If this voltage is not correct, replace the Centent Controller.
The longitudinal and transverse Centents are identical, and can be exchanged. If the problem remains, the Longitudinal Motor should be replaced.
• Longitudinal Belt
The Longitudinal Belt should not be tightened too much or this will cause the brackets holding the gears to deform at either end of the scanner. When the belt is properly tightened, it should be possible to deflect the upper and lower sides of the belt so that they touch within 8 cm of the gears at either end.
• Drive Wheels
The rollers in front and the wheels in back that support the Arm must be adjusted so that they come into perfect contact with the Longitudinal Rails. Test them by preventing any wheel from turning and see if the carriage will still move. By preventing any wheel from turning, it should be possible to slide the carriage along the rail with one wheel dragging while the others roll. This indicates that the wheel has not been excessively tightened down. Adjustments can be made by loosening and rotating the eccentric bearings of any of the lower wheels.

4.2.3 Loss of OMI Signal

• If the Scan Arm motion is irregular, or scraping noises are heard, the Longitudinal Mechanics may be binding. This symptom will typically be detected as a failure by the daily QA Mechanics Test.
If the error occurs consistently after the first line of a patient scan, and the scanner is moving in the longitudinal direction, then check the following:
• Cause: The Interrupt signal is being lost.
Solution 1: The pulses that normally enter the FOINK board at J (the
black wire at the center of the connector) may have stopped. These pulses are necessary to keep the FOINK board from sending an interrupt to the SBC. These pulses can be seen on a FOINK board LED.
For LED location see figure 4-1. Use the DPX-NT service software (Tools/Diagnostics/Scanner Motion / Motion Commands Tab) to set the joystick speed to 50 steps, enable the joystick and watch the LED. If the OMI / FOINK is working the LIght will flash when the Longitudinal Motor is run. If the LED flashes when the mechanics are engaged, but the error still occurs, the interrupt was invalid. Check for arcing in the high voltage system or replace the FOINK and SBC.
If the LED does not flash,
DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-125
Page 10
• Verify that the slotted disk at the foot end of the scan table is in the middle of the slot between the photo diode and photo transistor.
• The slotted disk must be completely flat and remain in the center of the sensor slot during its entire rotation.
• If the slotted disk has been in physical contact with the optical sensor, the sensor may have debris on it, disassemble this mechanism, and clean the sensor and the slots of the disk.
• Check the Cable running from the FOINK to the OMI for a cable break by checking the individual wires for continuity.

4.3 Failure of the 28V power supply

The 28 VDC power supply is only enabled when the X-ray tube is ramped, use the service software (Tools/Diagnostics/ Scanner X-ray) to attempt to ramp the Tube Head.
When the Supply is ramping the Red and Green LED’s on the MAX board will light.
If the LED’s illuminate and the go out:
• Measure the output of the 28VDC power supply, and verify that it remains constant during the voltage ramping and scanning operations.
• Check the High voltage power supplies, insure they are not arcing (Error Log - see section 3.2) and are ramping.
If the LED’s fail to light:
• This supply is turned on by the X-ray Relay, so verify that the Relay is closing. If not, then either the Relay is bad or it is not receiving the signal from the SBC via the FOINK.
• Check the continuity of the cathode, the filament may have broken, MAX board TP 4, TP 5 and TP 13 should be continuous with the Tube Head control cable connected.
• The Tube Head Thermostat is wired in series with the Relay, so if it has opened, the Relay will not be able to close.
• The 28VDC should also be measured at the Terminal Block. If not, check the continuity of the wiring and refasten all connections. Also, check the wire tie-downs for excess tension they may be putting on the wires.
• It may be necessary to check the wiring from the Terminal Block to the MAX board and to the High Voltage Power Supplies.
4-126Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 11

4.4 Emergency Stop Button

The Emergency Stop button is a normally closed circuit, if it opens the Emergency Stop interrupt will be sent from the FOINK to the SBC.
• Verify that the emergency stop button is out. If it has been pressed in, press it again to release it.
• If 26VDC can be measured on both pins of FOINK connector J17, the Emergency Stop Switch and the wires connecting it to the FOINK board are good and the FOINK board should be replaced.
• If 26 VDC is missing, the Circuit is open between the FOINK and the Switch.
• Insure the switch is functional
• Check the continuity of the wires from the switch to the FOINK board.

4.5 Tube Head Thermostat

There is a thermostat inside the Tube Head. This thermostat is normally closed, but opens when the Tube Head temperature is too high.
• The thermostat will close again automatically after a cool down period of usually less than 30 minutes. If 0 VDC is measured on both pins of FOINK connector J15, the Thermostat and the wires connecting it to the FOINK board are good and the FOINK board should be replaced.

4.6 Communication Error

The computer is not communicating with the SBC board.
Verify communications with the SBC in the service software (Tools/Service Options Comm tab - Test Communications).
If this is unsuccessful, verify that the I/O cable from the computer to the SBC Board is secure. Also, verify that the comm port is configured correctly (see DXPC 2000 Chapter 5 appendices) and that all required drivers are present. If all fails, the SBC or computer serial port is defective.

4.7 Other Diagnostic Failure Codes

DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-127
Page 12
Multiple error codes are possible. These will be displayed one after the other, but the first one displayed is probably the problem. The Error Log will always show the failures in the correct order, the others are generated as a result of the first failure.

4.7.1 Reasons For Invalid Diagnostic Failures

Arcing of the x-ray high voltage system releases a large amount of electromagnetic energy. This energy creates noise in the electronic circuits and may cause the SBC to do unpredictable things, such as give invalid diagnostic errors. Arcing is usually accompanied by white, blue or black lines across patient scans (See Arcing, section 4.10).

4.8 Failing Quality Assurance Test

See chapter 2 section 12, for explanations of the various tests that are performed during the running of the Daily Quality Assurance.
If any of the QA test results fail, none of the results are considered valid. The results will be stored in the Quality Assurance History file, but these values will not be averaged with the other results for calibration purposes. In addition, the software will prevent patient scans until a passing daily QA has been completed. Recent valid QA's are necessary for accurate results.

4.8.1 Block Position

After starting a QA, the operator is prompted to place the QA standard n the table, if the DPX-NT fails to detect the Block in the correct orientation the QA will not progress.
• Check Block position and orientation
The scanner may not be finding the correct "Home" position. The correct "Home" position aligns the center of the x-ray beam with the center of the Brass Piece when the Standard is correctly positioned. The Air Counts must be obtained outside of the QA standard, next to the Brass Piece. If the Reference Counts are obtained with the x-ray beam passing through the Brass Piece, the Quality Assurance Scan will fail.
• Verify that the scanner limit switches are set correctly with the DPX-NT home position jig (see DXAP2000 - DPX-NT Installation Procedure Chapter 5 appendices).

4.8.2 Beam Stop Action

The Beam Stop Action test verifies the ability of the lead shutter to attenuate x-rays from the tube head by verifying that the Background counts when the shutter is closed are lower than counts taken through the brass piece with the shutter open.

4.8.3 Mean% Spillover

4-128Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 13
An increase in the Spillover percentage over time is an indication that the detector is losing resolution, even if the Spillover test does not fail. A large change in the Spillover can be explained in some cases by standard positioning (or a small change in the home position of the scanner). If the beam is not fully eclipsed by the brass piece during the Spillover test, the Spillover value increases dramatically.
Spillover stability is a test of detector bias drift. if the detector bias is drifting, the mean spillover value will also drift.

4.8.4 Reference Counts and Ratio

During this test the baseline is established which will later be used for comparing with the values obtained during the scan of the standard. Care should be taken that the x-ray beam is not missing the standard, nor being obstructed by the brass piece on the standard during the High and Low Air Count test.
The Ratio value should remain fairly constant over time.

4.8.5 Ratio Fluctuations

The ratio of High to Low Channel Air Counts is the way the bone density measurement is calculated by the changes in the ratio of High to Low Channel counts. The Ratio should remain constant as long as the x-ray beam quality and the resolution of the detector remain constant.
It is possible to view the Ratio trends in the Quality Assurance History. Check both the 3mA Ratio and the 150 Ratio to determine if the either Ratio has changed significantly.
The problems which could cause a change in air ratio are:
· deterioration of the resolution of the detector
· current leakage through the transorbs on the XORB board
· a faulty High Voltage Power Supply (kV unstable over time)
· changes to the x-ray tube insert
These are very difficult to diagnose by a method other than substitution of new components until the Reference Counts Ratio returns to normal.
Detector deterioration or unstable High Voltage Power Supply, will usually cause the Alignment Test Scan to fail.
The specification for XORB Board transorbs is that they must allow less than
0.5 micro amperes reverse bias current. At LUNAR, each transorb is measured by applying a 5 Volt reverse bias to the transorb and a 100 kW 1% resistor connected in series. The voltage measured across the 100 kW resistor must then be less than 50 mV.

4.8.6 Transverse or Longitudinal Mechanics

Failure of this test is caused by one or more of the following:
DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-129
Page 14
• Incorrect Limit Switch positioning
• Defective Limit Switch
• A mechanical constraint
• A defective Motor
• A defective Centent Motor Controller
The values recorded for these tests should remain fairly constant over time. Variations between QA's of under 25 steps should not be a cause for concern as 1 transverse step = 0.05 mm and 1 longitudinal step = 0.1 mm, so the actual variation is only a few millimeters.
If the number of steps continually increases from QA to QA, this could indicate an impediment to the scanner's motion and should be rectified (see 4.1 (Transverse) or 4.2 (Longitudinal)).

4.8.7 Tissue Value

The tissue value should not be the sole failing parameter. The failure of this test is usually accompanied by problems with Bone Mineral values (see below) or Reference Counts Ratio trends (see above). If this is failing alone it is probably a problem with the counting system or high or low kV (4.8.9).

4.8.8 Bone Mineral of the Standard Chambers

The software contains values which it expects the scanner to measure for the scan of each chamber. If the mean of the BM measurements made for the chamber does not fall within the predefined software limits the QA will fail. If the percent coefficient of variance is over 3%, the QA will fail.
These failures will always occur if the Reference Counts or Reference Ratio test have deviated severely from normal results. However if the Reference Count results look normal, and the values are approximately equal to the numbers obtained during the scanner installation, then the counts may be unstable.
Another cause of failing on the bone chamber measuremnts can be arcing during the QA. This can be detected by examining the QA History graphically and looking for variation of the Large BM values. Note the few data points that vary. These are the early signs of arcing.
This will also be apparent (but not obvious) on the QA Results printout. The arc occurred in the third standard scan and elevated the BM values. If a customer reports a failing QA because of a bone chmaber measurement being too high, be aware that this could be an early warning of arcing. Obtain from the site the QA history file and error log for analysis. Look for variation of the Large BM values.
It is very important to notice these early warning signs of arcing so that the system can be re-greased before any damage is done to the high voltage cable connectors or the tube head.
4-130Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 15

4.8.9 Symptoms of High and Low KV

It has been noted in a small number of cases that defective high voltage power supplies, or an arcing tube head may produce a voltage other than 76 kV. The DPX-NT monitors kV once per scan sweep.
Note: :It would be prudent in both of these cases to take a positive
and negative power supply, AND a tube head. Also, the XORB test points may look normal (indicating 76 KV) in each case as well. This usually occurs where the power supplies are defective and so the monitor voltages returned to the XORB may not show the true voltage. When the tube head arcs causing the voltage to go too high, then the XORB test points may indeed show the actual voltages.

4.9 Reference Counts

4.9.1 Unstable Counts

To examine the scanner's ability to count detector pulses consistently over longer periods of time, run the Alignment Test option from the service software program.
• If the "End of Exposure Alarm" rings during the time the Alignment Test scan is running, see "Alarm Pings During Scan" in section 4.13.4. If the Shutter Open or the X-ray On lamps on the front panel blink, during the test, see section 4.13 also.
• While the Alignment Test scan is running, measure the voltage at test points 1, 2, 5 and 6 of the XORB board. They should have approximately the following values respectively: -0.150VDC, 3.8VDC, 0.150VDC, and
3.8VDC.
• A pure DC voltage on an oscilloscope at test point 1 and 5 of the XORB will verify that there is constant current through the filament of the X-ray Insert.
• If there is AC ripple in excess of 0.2 Vpp, the power supplies should be replaced.
• Start and stop the x-rays several times while observing the voltage at TP3 and TP7 of the XORB board. This is the programming voltage from the SBC, and although this voltage is dependent on the feedback information returned to the SBC, the voltages at the XORB board test points should be approximately the same each time the x-rays are produced. The AC ripple on this signal must be less than 0.2 Vpp. Replace the SBC board if the proper control signal is not present.
DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-131
Page 16
• When the x-rays are off between scans, verify that the shutter aperture lines up with the aperture to the Tube Head beneath. Take some alignment pictures and verify that the image is a bright, rectangular image and the entire alignment box is visible. If the alignment box is not visible, the shutter Tube Head and detector apertures are not aligned.
• Open and close the shutter to make sure that it returns to the same position each time. Turn the x-rays back on, and make sure that you obtain nearly the same count rate each time the shutter is opened.

4.9.2 No Counts

When the table to the right of the peak graph on the Quality Assurance Results printout is entirely filled with zeros, use the Signal monitor program in the Diagnostics (see section 5.1) to create x-rays at 76 kV and 150 uA and open the Shutter for sampling. Then check the following items:
I.Is the amber X-ray On LED illuminated?
A.If it is, skip to part II.
B.If not, the voltage or current ramping has probably failed. Are both the red and green LED's on the MAX Board illuminated?
1.Voltage or Current Ramping Failures
If not, is the red LED illuminated?
a.X-ray Relay or FOINK
If not, measure the output of the 28 Volt Power Supply. Check the operation of the X-ray Relay or the FOINK Board which controls its operation. If all of the above are working, the red LED may be defective.
b.MAX Board Fuse
If so, the fuse is blown on the MAX Board (see section 4.16).
2.If the red and green MAX Board LED's are illuminated, verify the following test point voltages:
a.Current Ramping Failure
• XORB TP1 is approximately 0.150 VDC. If incorrect, go to step d. If this test point is correct, verify that XORB TP5 is also approximately 0.150 VDC. If this voltage is incorrect, substitute a new Positive High Voltage Power Supply.
b.Voltage Ramping Failure
• XORB TP2 and XORB TP6 are approximately 3.8 VDC. If these voltages are incorrect, verify that the voltages on XORB TP3 and XORB TP7 are approximately 3.8 VDC. If TP3 and TP7 are not equal the XORB jumper at J26 is set in the wrong position. If they are equal but incorrect, test the cable from the SBC to XORB, or substitute a new SBC Board.
4-132Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 17
c.X-ray On LED
Test the X-Ray On LED by inserting it into the Power On receptacle. Replace if defective.
d.High Voltage System
• Current ramping has failed. If TP1 and TP5 are zero and do not change, check the polarity of the High Voltage Cables. The X-ray Insert is essentially a diode, and will not conduct current from the anode to the cathode.
• If either of the test points is at 1.0 VDC, the High Voltage Power Supply is delivering as much current as it possibly can, and has automatically limited the voltage.
• Check TP2 and TP6. If the voltage is approximately 3.8 VDC, the voltage has been set properly.
Feel the heat sinks on the back of the High Voltage Power Supplies. If one of the power supplies is cold, this is usually the defective one. If one is warm and the other is hot, replace the hot one. The best troubleshooting technique may be to substitute power supplies.
• If either TP2 or TP6 are incorrect, the possibility exists that one of the High Voltage Cables or the Tube Head is shorted. The short may be possible to find with an ohm meter, but often it takes several kV to break down the defective component. It will be difficult to troubleshoot this problem by any method other than part substitution.
Do not attempt to ramp the power supply above 50 kv without a cable connected, or arcing will occur near the connector.
• Ramping between 40 and 50 kV is a good way to determine whether the power supplies are able to increase voltage from 0 to 50 kV.

4.9.3 Decreasing Reference Counts

A long term, continuous decrease in the Reference Counts unaccompanied by a change in Reference Ratio is a serious problem. A count rate of at least 500,000 Low Channel Reference Counts is necessary to maintain precision of results when scanning thick patients. Reference Count values should change by less than 10% from the day of installation. The possible causes are:
I.Lead Filings - Shutter Wear
DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-133
Page 18
A rough edge around the shutter aperture may be wearing away on the tube head below. The lead filings fall off the shutter aperture and gather on the filter below, causing a reduction in x-rays over time. Any wear should be easily visible on the underside of the shutter paddle. If there is visible wear, the lead dust must be removed from the top of the filter. The Collimator assembly must be removed, the filter cleaned, and then the Collimator assembly reinstalled and realigned.
II.A Change in Current Control Settings
Use the Signal monitor option of the service software program to produce x­rays (Tools/Diagnostics? Scanner X-Ray). Select 76 kV and 750 µA operation.
• Verify that the SBC is properly controlling the current. TP11 of the MAX board should be approximately 0.75 volts. This voltage is dependent on feedback information from the power supplies, which makes it difficult to troubleshoot by any means other than SBC substitution.
• Check the current through the X-ray Insert. The absolute value of the voltage at test points 1 and 5 on the XORB board is proportional to the current through the X-ray Insert. 1 millivolt is equal to 1 micro-ampere of current. A current setting of 750 µA should give a reading of 0.750 VDC at test points 1 and 5. If either of these voltages vary from the expected by more than 10 millivolts, the MAX board could be at fault. If TP1 and TP5 are more than 15 millivolts apart, substitute new high voltage power supplies.

4.10 Arcing

The X-ray Tube Head Insert is an evacuated glass enclosure. An AC current is applied to the filament inside the insert. It glows like the filament in a light bulb, and electrons are boiled off into the evacuated space. A high voltage is applied between the anode and the cathode causing electrons to rush toward the anode, striking it and creating x-rays. As long as the insert is properly evacuated, there can be no internal arc. However, no insert can be totally evacuated and impurities can be ionized creating a lightning like effect; arcing. During the arc the resistance of the insert is dramatically decreased and a large amount of current flows.
• The significant amounts of electromagnetic energy released inside the scanner can cause problems with the electronics of the scanner system and may result in abnormal operations of the scanner. An arcing scanner will have one or more of the following symptoms:
• complete lock up of the system
• a diagnostic error code message which is undefined or inappropriate
• a vertical stripe or artifact in the image of the patient's scan (effects all 16 detectors at once)
4-134Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 19
• Transverse Motion Failure message.
To specifically locate the source of an arc, it is necessary to find out which high voltage power supply provides the excessive current. This may be done by recording the power supply current monitors at TP1 and TP5 of the XORB board with a storage oscilloscope. An arc typically draws 1.5 mA from the power supply for about 1 ms. The trigger voltage should therefore be set at
1.5V for TP1 and for TP5.
• Arcing can also occur inside the high voltage connectors. This will usually result in a plainly visible black or brown carbon track through the grease on the connector. In any case, once the high voltage connectors have been removed from the Tube Head and the power supplies, they should not be re-connected without first being cleaned and regreased (see procedure DXSE0002 in the chapter 5 appendices).
• After cleaning the old grease off of the connectors, they should be carefully inspected for carbon tracks. Look for these tracks on both the rubber cable connectors and on the phenolic sockets of the Tube Head. If such tracks are found after cleaning, the following are the options for returning the scanner to service:
• If carbon tracks are found on the rubber cable connectors, they can be removed by excising the damaged section with a sharp blade. Severe tracks can burn quite deep into the rubber, so care must taken to remove all the damaged rubber. After the carbon tracks have been removed, additional grease must be used when repacking the connection to fill in the volume of the removed rubber.
• If the above procedure requires too much rubber to be removed, or the carbon track looks very severe, another option is to replace the high voltage cable.
• Arc tracks will be impossible to see on the sockets in the Tube Head. Therefore, if arc tracks are seen on the rubber cable connectors, the sockets should be sanded with emory cloth as a precaution. After sanding the socket, remember to flush the socket with cleaning solution to remove any particles.
• If the tracking inside the socket is too severe or the carbon track cannot be removed, another option is to replace the Tube Head.
Note: IT IS VERY IMPORTANT TO REMOVE ANY CARBON
TRACKS ON THESE SURFACES! Carbon tracks that are simply covered by grease or not completely removed will still provide a path for arcs to follow.

4.10.1 Limit Switch Tripped During Scan

If a Limit Switch is tripped during an install test, verify the Limit Switches' location with the Service Software (Tools/Diagnostics/Scanner Motion). If this problem occurs any time after install, see the problem description below.
DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-135
Page 20
If the transverse mechanics of the scanner become imprecise, it may cause a Limit Switch to be tripped during a scan. This will usually only happen during a total body scan or during an Alignment Test test since both of these scans involve full-width scanning. Once the switch is closed, the motors stop, the End-of-Exposure Alarm sounds, the amber SHUTTER OPEN lamp turns off and after a few seconds a Diagnostic Failure appears on the screen.
• During full-width scanning, the Detector Carriage comes very close to the Limit Switches, so any imprecision will cause a Limit Switch to be tripped and the scan will be aborted with a Diagnostic Failure message for Transverse Motion Failure.
A cause of this problem is a loosening of the first drive Reduction Belt which connects the Transverse Motor to the first Reduction Pulley. This loosening causes the belt to "walk" on the pulleys causing enough imprecision in the motion to trip a switch. To tighten the belt, first remove the Pulley Shroud and loosen all four nuts that hold the motor in place. Then, while holding the motor such that the belt is pulled taught, tighten the nuts to secure the motor in place. Replace the shroud and test the scanner (an Alignment Test test works well).
If the Tube Head cables come into contact with the cable bundle entering the cable track, it may be impossible for the scanner to complete all of the necessary transverse steps away from the operator. Consequently, on each scan line the detector will move closer to the front transverse limit switch, and the switch may be eventually closed. Form the cable bundle exiting from the cable track into an arch such that the Tube Head cables will move under the arch rather than running into the bundle. This arch must not be too high. If it is, there will not be sufficient slack in the Tube Head cable bundle to allow the Tube Head to move to the front transverse limit switch. Also, if the arch is too small, it may cause too much slack when the Tube Head is at the forward side of the table. This causes the Tube Head cable bundle to rub against the inside of the front panel causing a scraping noise to be heard. If the above does not solve the problem then check the other mechanical components. Electrical components that could cause such a failure are the Centent Motor Controller, the Transverse Motor, the FOINK board or the SBC board.

4.11 Imaging Problems

4.11.1 White, or Grey in the first or second scan line:

The software is not perfect in its ability to determine the correct grey level of the entire scan based on the first line of data. Sometimes the grey levels will be set incorrectly during acquisition. During analysis the grey level is easily adjusted to give a good image, and the results are not affected by this imaging problem. This problem is most often found on very thin patients. Adding extra tissue equivalent material (rice bag or saline solution) to a very thin patient may be all that is necessary to cure the symptom.

4.11.2 Femur Scan Problems

4-136Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 21
Most image discontinuity problems occur in femur images performed at 3 mA on thin or osteoporotic subjects. These lines usually occur in the trochanter area where the x-ray beam is least attenuated by tissue.
These lines are caused by Automatic Gain Stabilizer (AGS) trying to adjust the input signal which is at an excessively high count rate. However, the AGS is not at fault, and the correct action is for the customer to attenuate the x-ray beam. Lunar normally recommends placing a rice bag (on its side to create 12 cm of patient thickness in the x-ray beam) along the side of the patient's thigh for thin or osteoporotic patients.

4.11.3 AP-Spine Image Problems: Probable causes

• the shutter closes (causes the shutter open lamp to go out, and the End of Exposure Alarm to sound).
• increase or reduction of high voltage to the detector (causes horizontal artifacts in the image).
• loss of the signal from the detector (this effects all 16 detectors at once)
• loss of current to the x-ray insert filament (28 volt power supply error message, the orange lamp to go out, and the End of Exposure Alarm to sound).
• bad x-ray relay contacts (providing power to tube head power supplies)
• reduction in AC Line voltage
• loss of proper high voltage on the x-ray tube
• faulty operation of the AGS system (white lines are short and found only where scan lines move from air into tissue).
• arcing (may cause vertical artifacts in the image)

4.11.4 Broken Signal Cable

The cables which carry the voltage to the detector and return the signal from the detector array are part of the upper and lower cable bundle assemblies. Any break or significant pinch of the cable can cause reduction in counts in either or both channels. This could result in horizontal artifacts which effect all 16 detectors at once.

4.11.5 Loss of tube head current

The tube head control cable contains the wires which provide power to the filament transformer. If the wire to the transformer center tap breaks, the x­rays will stop, the orange lamp will go off, and the end-of-exposure alarm will sound. If either of the other two wires break, the MAX board fuse will blow. Always ask the operator if a "ping" sound (the end of exposure alarm) is heard, and whether the Shutter Open or X-Ray On lamp turns off. In the case of a tube head current problem, the X-Ray On light will turn off.

4.11.6 X-Ray Relay Failure

DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-137
Page 22
See section 4.16.

4.11.7 Unstable AC Line

If the customer's AC line conditions are unstable, voltages 11% below the scaners AC input configuration will begin to drop the high voltage on the x-ray tube. The counts will drop significantly with only a change of a few kV. However, this normally causes black or dark grey lines in the scan image.

4.11.8 Arcing

If the tube head is arcing, the arcs will be more likely to occur at the higher current settings. Arcing generally also causes other strange symptoms such as:
• Scanner stops in the middle of a patient or QA scan with no error message or with transverse mechanics fail
• The QA scan fails intermittently on bone chamber measurements
• The scanner periodically leaves the message on the screen "Starting X­rays Please Wait" for a longer than normal period of time (more than 10 seconds)
If you see arcing symptoms, try the following service action:
1.View the Quality Assurance History (see section 4.15). In particular note the trend in BM Values as discussed in the Failing QA's section of this chapter.
If there is evidence of failures in the mean BM value:
2.Repack the high voltage connections as described in the chapter 5 appendices. Note if an arc track is visible in the insulation compound.
3.Replacement of the tube head if no arc tracks are discovered, or if the repacking of the connectors does not eliminate all arcing symptoms.
Summary
Except for the x-ray relay, the cause of the problem will be difficult to find. Try to obtain some additional information. If the customer is hearing the end-of­exposure alarm, or seeing the yellow x-ray on lamp flickering, you can deduce the tube head control cable is bad. Likewise, it is possible to decide on the shutter/collimator/fan cable. In the case of no alarm or lamp indications, or in the case where large deviations occur in the peak, the upper and lower cable bundles should be replaced, as problems with the coaxial cable will be too expensive to troubleshoot.

4.12 Failing Alignment Test Results

4.12.1 Image

4-138Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 23
The Image is a graphic illustration of the counts recorded during the Alignment Test. A good image has a black field with evenly distributed grey dots and perhaps a few randomly distributed dots of various colors (green or blue usually).
Figure 4-6. Failing Ailigment Test
A poor Deviation Image will have lines, streaks or spots (indicating counts higher or lower than expected). When the image is analized, click first on points and then back to results, this will typically point type failing points (they will have a yellow box around them).
• The information tab (see figure 4-2) will show the test outcome (pass / fail).
• Lines or streaks going longitudinally in the image indicate an alignment problem. Recheck the alignment of the scanner including the levelness of the scanner. Also, check the scanner frame for any bending that may have occurred at the site or during shipment.
DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-139
Page 24
• Lines or streaks going transversely across the Alignment Test image indicate a problem with the AGS, on the Detector Daughter Board in the detector array. Again, replace parts until a passing Alignment Test is obtained.
• Troubleshooting tip: if the Alignment test can be run with the Alignment Aperture on, this will make the test more sensitive as the beam will be further collimated to the exact size of the detector. Any failures (beam straying off the detector) will be exaggerated.

4.13 Indicator Failures

4.13.1 X-ray On LED Blinking

The amber X-RAY ON LED should glow steadily once it is illuminated, until the x-rays are turned off. If the X-RAY ON LED blinks while the shutter is open for a patient scan, the End of Exposure Alarm should sound.
The amber LED is controlled by a comparator on the FOINK Board. This LED illuminates when approximately 40µA (or greater) of current is passing through the X-ray Insert. On Spellman system scanners the amber light may appear to fade out slowly when x-rays are ramped down. This due to the Spellman power supplies bleeding off current and is not a problem.
If the voltage ramping fails, the High Voltage Power Supplies must be shut completely off before a second ramping attempt is made. If the amber X-RAY ON LED turns on momentarily, then turns off for a few seconds before coming on steadily, the system is having difficulty ramping the high voltage. One of the High Voltage Power Supplies may be defective.
The SBC does not monitor the Insert current once the proper level has been reached. Therefore, if the current to the Insert is interrupted, the SBC will not recognize the problem and will not alert the computer, or terminate the exposure.
If the X-RAY ON LED blinks once the x-ray high voltage has been set, there must be a problem in one of the following areas:
• Tube Head Control Cable-The most likely cause of an intermittent problem in this circuit is a broken wire in the Tube Head Control Cable. The Red wire is most likely the broken one, as a broken blue or black wire should blow the MAX Board Fuse. Turn the scanner off and test for continuity between MAX Board test points TP4, TP5, and TP13.
• MAX Board-The MAX Board may be operating intermittently. Substitute a replacement MAX Board.

4.13.2 Shutter Open LED Blinking

4-140Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 25
The amber Shutter Open LED is controlled by the Limit Switch on the Shutter/ Collimator Assembly. If the Limit Switch is defective or improperly adjusted, the Shutter Open LED may be switched on and off. This will usually be accompanied by the sound of the End of Exposure Alarm if an exposure is underway.
If the Shutter is actually opening and closing intermittently and uncontrollably, the problem is either on the FOINK, or a broken wire between the FOINK and the Shutter/Collimator Assembly. Remove connector J11 from the FOINK Board and measure the resistance between pin 5 and pin 6. This will be a measurement of the resistance of the Shutter Solenoid and the cable through the Cable Track. If the problem occurs on every scan, run a scan while making this continuity check. If the cable and solenoid seem to be working properly, substitute a new FOINK Board.

4.13.3 Shutter Not Operating

If the Shutter Solenoid will not open, the problem may be caused by one of the following:
• The Shutter Open LED-If the Shutter Open LED fails, a properly operating FOINK board will prevent operation of the Shutter Solenoid. Check the LED for continuity, replace it, or substitute a working amber LED temporarily.
• Solenoid Cable-Remove J11 from the FOINK Board and check continuity between pin 5 and 6 (see Shutter Open LED Blinking above).
• FOINK Board-Replace the FOINK Board.

4.13.4 End of Exposure Alarm During Scan

If the sharp "Ping" sound of the End of Exposure Alarm is heard during a scan, look at the computer display to see if a Diagnostic Failure Code is being reported. If so, see section 4.1.
If no Diagnostic Failure Code is reported, the scanner may still be in motion continuing with the scan. Abort the scan and remove the patient from the table. In any case, note the status of the SHUTTER OPEN light and the X­RAY ON light.
• Both the SHUTTER OPEN light and the X-RAY ON light are on and steady. This would indicate a faulty FOINK board. Replace it
The SHUTTER OPEN light is off. This could indicate the following:
• The LED becomes defective during the scan and since the shutter solenoid and this light are wired in series, the Shutter closed and the Alarm sounded.
• The cable to the amber Shutter Open light broke during the scan with the same result as above.
• The shutter solenoid failed and the Shutter closed followed by the Alarm sounding.
DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-141
Page 26
• The FOINK is faulty.
• The SBC is faulty.
The X-RAY ON light is off. This could indicate the following:
• X-ray production has halted. This turns off the X-ray On light and sounds the Alarm.

4.14 Communications Failures

Should the scanner, host computer and or Detector Mother Board lose communications with each other, there are a number of things to check:
A good troubleshooting tool is the COMM test in Tools/Service Options ­COMM tab.
Check
• the I/O cable connections at the serial port on the back of the Host computer and at the SBC. Be sure both connections are tight and that the thumb screws are used to hold the connectors firmly together.
• inspect the connector on the SBC. This connector is very fragile and may have been damaged. Be very careful when connecting the I/O cables to the SBC.
• Check and inspect the Signal Cable input at the SBC
• Check the Connection at the Detector Mother Board and the Bulkhead, where the two halves of the signal cable meet.
• Verify the power LED’s on the SBC are lit (See section 4.19)
• Verify that the Power LED’s and PLD programming LED’s are lit on the DMB (see section 4.21)
If the ports are properly configured and all power is present, and communication still fails, then the probable causes are:
• A faulty I/O cable.
• A faulty SBC.
• A faulty serial port / serial port card.
Replacement of these parts is the best troubleshooting method.

4.15 Viewing Quality Assurance Trends

4-142Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 27
DPX-NT software includes the capability to view, print or graph data contained in the Quality Assurance database. View and print display selected results of many of the tests run during the Daily Quality Assurance in tabular form. The graph feature will graph the results of only one of the Quality Assurance parameters. All features allow for the user to select the time period to be examined.

4.15.1 What to Look for in the QA History

All categories should remain steady over time. Check the PASSED column for failed QA’s. If a failing QA is found, try to determine the cause, including operator error.
The limit for % Spillover is 13%.
If any sudden jumps are noticed in the values of any of the categories, determine first if these are due to service work such as a Tube Head replacement. If service work causes a shift in the QA values, then compare present values to those obtained only after the service.

4.16 MAX Board Troubleshooting

The MAX board controls the current through the X-ray Insert in the Tube Head. DPX-NT scanners use MAX version 02B currently.
The current at a given kV across the insert is a function of insert filament temperature. The MAX board adjusts the current by regulating the filament temperature to provide the proper current as set by the SBC. The relationship is Insert µA X 1000 = SBC input (in volts DC), so for an SBC input of 5VDC, Insert µA = 5000.
The Positive High Voltage Power Supply has a current monitor output (mA MONITOR); the voltage at this output is proportional to the current passing through the X-ray Insert. This monitor voltage is connected to the MAX board as feedback (mA FEEDBACK).
What to look for
• Make sure that all connectors are fully seated.
• Check Lemo connectors for loose parts.
• Check for loose wires on the insulation displacement connectors.
• Check the position of the shorting jumpers on pin headers JB1, JB2, and JB3 for: Be sure pins 1 and 2 are connected as indicated by silk screen.
When the X-ray Relay is on, both the red and green LED's on the MAX board should be illuminated. If both are out, no power is getting to the MAX board. If only the red LED is lit, then the MAX board fuse is burned out or missing.
Electrical tests
DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-143
Page 28
When the board is powered up, and x-rays are being generated, the following measurements may be performed:
With a VOM set on 200 VDC, you should measure about 18 to 21 VDC on TP's 4, 13 and 5 referenced to TP3 (GND).
Note: If an oscilloscope is available, you should see about 18 to 21
VDC on TP13. However, TP 4 & 5 should have a 36 to 42 volt peak square wave. A VOM will show the average value of the square wave, i.e. 18 to 21 VDC.

4.17 FOINK Board Troubleshooting

The FOINK is a multipurpose interface board designed for use with the DPX­NT. Operation is both analog and digital.
Optical isolation is used to isolate the +26VDC supply from the +5VDC and +12VDC (logic) supplies, and eliminate interference between the motor circuits and the event (detector) signals.
What to look for
Figure 4-7. FOINK Diagnostic LED’s
FOINK LED’s:
A - (D10) Longitudinal OMI, when this LED flashes, the Longitudinal OMI infrared beam is
B - (D9) Transverse OMI, when this LED flashes, the Longitudinal OMI infrared beam is
4-144Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 29
C- (D8) X-ray on (28 VDC supply), when this LED is lit when the +28 VDC power supply is on
D - (D7) Mechanics engage, when LED is out the motor(s) are enabled
E - (D6) - Errors Clear, must be lit - out when FOINK detects an error (failsafe)
F - (D5) Reset when this LED is lit the SBC is in Reset

4.18 OMI Board Troubleshooting

The OMI board, being a rather uncomplicated device, has little in the way of things to watch for in installation and use. There are three such boards, one on the Front Longitudinal Carriage (transverse), at the foot of the scanner on the longitudinal idler shaft (longitudinal), and on the collimator (shutter open detect).
What to look for
• Check the connector to each board for proper seating.
• Examine the area where the wires attach to the board itself; it is possible the may fray or break.
• Insure that the interrupter disk does not rub against the plastic detector housing. If it does, adjust by loosening the two machine screws that attach the plastic detector, reposition the detector and re-tighten the screws.
Electrical tests
The signal from the OMI board is nominally rectangular in shape. The signal is generated when the Interrupter Disk alternately passes and occludes a light beam between the two posts on the photo-transducer. With the beam occluded by the disk, the output signal should be about 5 VDC (close counts). When the beam is allowed to pass through a slot in the disk, the signal output is near ground potential (see FOINK troubleshooting 4.17).
Symptoms of possible OMI failure
• Scanner shuts down at the beginning of a patient scan.
• Scanner shuts down unexpectedly during a scan.
• Either condition above should be accompanied by a Diagnostic Failure Transverse or Longitudinal (see section 4.0).

4.19 SBC Troubleshooting

DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-145
Page 30
The SBC consists of a single circuit board that uses the 8032 microprocessor with a RS-422 interface to the Host PC and a RS-422 interface to the Detector Motherboard (DMB)
The SBC is located on the electronics pan in the front center.
Figure 4-8. DPX-NT SBC highlight indicates the location of the diagnostic
LED’s
Figure 4-9. DPX-NT SBC Diagnostic LED’s
What to look for
• Green LED (C in figure 4-5) is on then you have +5VDC to the board.
• Red LED (A in figure 4-5) is on then you have +12VDC to the board.
4-146Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 31
• Amber LED (B in figure 4-5) is on then you have -12VDC to the board.
• Red LED (E in figure 4-5) is the DTR reset line, when this LED is lit, the Detector, FOINK, or SBC are detecting an error.
When the DTR Reset Line is lit:
• Check the LED’s on the DMB (see section 4.21)
• Check the FOINK LED’s
• Check the Signal Cable
The Service Software can be used to verify communications between the Host / SBC and the DMB.
Go to Tools - Service Options - Communications Tab - test communications
• Red LED (D in figure 4-5) the scanner Reset, when this LED is lit, the communications with the Host PC are in Reset.
• Verify port settings on the Host PC (See DXPC 2000, chapter 5 Appendices)
• Verify communications cable
• Green LEDs (G and H in figure 4-5) flash when data is being received by the SBC
• Green LED (F in figure 4-5) is not assigned a function at this time.
The reset button near the communications cable connector causes a reset signal at the SBC microprocessor. Also, disconnecting the serial cable will cause a reset state until the cable is re-attached and communications reinstated.

4.20 XORB Troubleshooting

The XORB board routes a number of important signals to the rest of the system and protects the system from transients by shunting them to ground with transorbs.
What to look for
• All connectors firmly seated.
• Check to see that all wires and cables are firmly affixed to their respective connectors. Pay particular attention to Lemo connectors, since the jam nuts on the connectors are prone to be loose.
DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-147
Page 32
• Check and tighten, if necessary, all mounting screws for the XORB board.
Electrical Tests
• Measure impedance between XORB ground plane and the electrical pan. It should be less than an ohm.
• There should be no low impedance (less than 10 ohms) reading between any signal or signal ground and chassis ground on the XORB when the as-associated signal connectors are disconnected.
The following test points should have the following readings:
Table 4-12. XORB Test Points
Test Point kV mA Limits
TP 1 76 .750 0.74 - 0.76
TP 5 76 .750 0.74 - 0.76
TP 1 76 3.00 2.97 - 3.03
TP 5 76 3.00 2.97 - 3.03
TP 2 76 .750 3.76 - 3.84
TP 6 76 .750 3.76 - 3.84

4.21 Detector Motherboard Troubleshooting

Figure 4-10. DMB Diagnostic LED’s
4-148Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 33
Diagnostic LED’s in figure 4-6
• A - (Green) Power +12 VDC Present
• B - (Green) Power - 12 VDC Present
• If the 12 VDC power LED’s are not illuminated power down the scanner, wait 30 seconds and power back up (see section Chapter
2.1.2)
• Check +5, +/-12 VDC Power Supply 1
• C - (Green) DMB PLD Programmed and ready
• If the DMB PLD LED is not lit verify FIRMWARE version with the Service Software (Tools/Set Download Parameters / Query), if necessary download to the DMB.
• D - (Green) DDB 1 PLD Programmed
• Verify the DDB card is seated in its socket
• E - (Green) DDB 2 PLD Programmed
• F - (Green) DDB 3 PLD Programmed
• G - (Green) DDB 4 PLD Programmed
• H - (Red) Data Sent to SBC (should flash when acquiring data)
• I - (Red) Data Sent to SBC (should flash when acquiring data)
• J - (Green) Detector Reset
• Verify Communications with the SBC
• Verify All DC voltages are Present on the DMB
• K - (Green) +5 VDC present
• Check +5, +/-12 VDC Power Supply 1
• L - Detector Mother Board Reset Button

4.22 Detector Daughter Board Troubleshooting

DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-149
Page 34
Figure 4-11. Test Point Board
The Test point board gives the service engineer access to all of the signals for the individual Detector Daughter Boards
The Test Point Board is broken into 2 areas:
Analog Signals
DMB Controlled Programming Information (same data to all DDB PLD’s
FIRMWARE from DMB Flash RAM)
• DCA REF - DCA Reference Levels - High and Low Energy Reference Levels for event counting windows
LEL = 1.19 VDC
LEH = 1.66 VDC
HEL = 1.96 VDC
HEH = 2.79 VDC
• AGS REF AGS Reference Levels - High and low Energy
CENTER = 2.35 VDC
LOWER = 2.00 VDC
• Detector Bias Setting (BIAS =.450VDC)
• +5 VDC Ref (+5.00 VDC)
• - 5 VDC Ref (-5.00 VDC)
4-150Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Page 35
Individual Detector Signals (Stored in DMB Flash RAM)
• GAIN - combination of Peak and AGS signal gain - setting for Shaping Amplifier - visible only with Oscilloscope
• AGS - AGS amplify or Attenuate signal - visible only with Oscilloscope
• BiPolar - Bipolar signal out of the shaping Amplifier
Digital Signals
• AGS ROLL - logic signal for AGS rollover - this signal is not currently utilized by the software
• LE - Low Energy Event detected and counted
• HE - High Energy Event detected and counted
• SAMPLE - Sample clock signal
• AGS ENABLE - Operate Calibrate signal - locks out AGS for Peaking operation
DPX-NT Service Manual (Rev A- 1999) Troubleshooting 4-151
Page 36
4-152Troubleshooting DPX-NT Service Manual (Rev A- 1999)
Loading...