• Voluson i / Voluson e systems with Software version SW 6.x.x (BT06)
• Voluson i / Voluson e systems with Software version SW 7.x.x (BT07)
• Voluson i / Voluson e systems with Software version SW 8.x.x (BT09/11)
• Voluson i / Voluson e systems with Software version SW CC200 (BT14)
(NOVEMBER 8, 2013)
VE8_COVER.FM
Part Number: KTI106052
Revision: 10
0123
GEVOLUSON i / VOLUSON e
WARNING
(EN)
AVERTISSEMENT
(FR)
WARNUNG
(DE)
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
Important Precautions
TRANSLATION POLICY
THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
OR OTHER HAZARDS.
CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.
• SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
• NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.
DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER
SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
• VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.
- i
GEVOLUSON i / VOLUSON e
AVISO
(ES)
ATENÇÃO
(PT-Br)
(PT-pt)
AVISO
AVVERTENZA
(IT)
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.
• SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIÓN.
• NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.
ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL EM
INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
• NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
• O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.
ESTE MANUAL DE ASSISTÊNCIA ESTÁ DISPONÍVEL APENAS EM INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
• NÃO TENTE EFECTUAR REPARAÇÕES NO EQUIPAMENTO SEM TER
CONSULTADO E COMPREENDIDO PREVIAMENTE ESTE MANUAL.
• A INOBSERVÂNCIA DESTE AVISO PODE RESULTAR EM FERIMENTOS NO
TÉCNICO DE ASSISTÊNCIA, OPERADOR OU PACIENTE EM CONSEQUÊNCIA
DE CHOQUE ELÉCTRICO, PERIGOS DE ORIGEM MECÂNICA, BEM COMO DE
OUTROS TIPOS.
IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN
INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.
ii -
GEVOLUSON i / VOLUSON e
HOIATUS
(ET)
VAROITUS
(FI)
ΠΡΟΕΙΔΟΠΟΙΗΣΗ
(EL)
FIGYELMEZTETÉS
(HU)
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES.
• KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST
ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST.
• ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT KÄESOLEVA
TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST.
• KÄESOLEVA HOIATUSE EIRAMINE VÕIB PÕHJUSTADA TEENUSEOSUTAJA,
OPERAATORI VÕI PATSIENDI VIGASTAMIST ELEKTRILÖÖGI, MEHAANILISE
VÕI MUU OHU TAGAJÄRJEL.
TÄMÄ HUOLTO-OHJE ON SAATAVILLA VAIN ENGLANNIKSI.
• JOS ASIAKKAAN PALVELUNTARJOAJA VAATII MUUTA KUIN
ENGLANNINKIELISTÄ MATERIAALIA, TARVITTAVAN KÄÄNNÖKSEN
HANKKIMINEN ON ASIAKKAAN VASTUULLA.
• ÄLÄ YRITÄ KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT
JA YMMÄRTÄNYT TÄMÄN HUOLTO-OHJEEN.
• MIKÄLI TÄTÄ VAROITUSTA EI NOUDATETA, SEURAUKSENA VOI OLLA
PALVELUNTARJOAJAN, LAITTEISTON KÄYTTÄJÄN TAI POTILAAN
VAHINGOITTUMINEN SÄHKÖISKUN, MEKAANISEN VIAN TAI MUUN
VAARATILANTEEN VUOKSI.
ΤΟΠΑΡΟΝΕΓΧΕΙΡΙΔΙΟΣΕΡΒΙΣΔΙΑΤΙΘΕΤΑΙΣΤΑΑΓΓΛΙΚΑΜΟΝΟ.
• ΕΑΝΤΟΑΤΟΜΟΠΑΡΟΧΗΣΣΕΡΒΙΣΕΝΟΣΠΕΛΑΤΗΑΠΑΙΤΕΙΤΟΠΑΡΟΝ
ΕΓΧΕΙΡΙΔΙΟ ΣΕ ΓΛΩΣΣΑ ΕΚΤΟΣ ΤΩΝ ΑΓΓΛΙΚΩΝ, ΑΠΟΤΕΛΕΙ ΕΥΘΥΝΗ ΤΟΥ
ΠΕΛΑΤΗ ΝΑ ΠΑΡΕΧΕΙ ΥΠΗΡΕΣΙΕΣ ΜΕΤΑΦΡΑΣΗΣ.
• ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ
ΕΚΤΟΣ ΕΑΝ ΕΧΕΤΕ ΣΥΜΒΟΥΛΕΥ
ΤΕΙ ΚΑΙ ΕΧΕΤΕ ΚΑΤΑΝΟΗΣΕΙ ΤΟ ΠΑΡΟΝ
ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ.
• ΕΑΝ ΔΕ ΛΑΒΕΤΕ ΥΠΟΨΗ ΤΗΝ ΠΡΟΕΙΔΟΠΟΙΗΣΗ ΑΥΤΗ, ΕΝΔΕΧΕΤΑΙ ΝΑ
ΠΡΟΚΛΗΘΕΙ ΤΡΑΥΜΑΤΙΣΜΟΣ ΣΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ, ΣΤΟ ΧΕΙΡΙΣΤΗ Ή
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝΔΥΝΟΥΣ.
EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ EL.
• HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT,
AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE.
• NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A
KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
• EZEN FIGYELMEZTETÉS FIGYELMEN KÍVÜL HAGYÁSA A SZOLGÁLTATÓ,
MŰKÖDTETŐ VAGY A BETEG ÁRAMÜTÉS, MECHANIKAI VAGY EGYÉB
VESZÉLYHELYZET MIATTI SÉRÜLÉSÉT EREDMÉNYEZHETI.
- iii
GEVOLUSON i / VOLUSON e
VIÐVÖRUN
(IS)
VÝSTRAHA
(CS)
ADVARSEL
(DA)
WAARSCHUWING
(NL)
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU.
• EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN
ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU.
• REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG
SKILIÐ ÞESSA ÞJÓNUSTUHANDBÓK.
• EF EKKI ER FARIÐ AÐ ÞESSARI VIÐVÖRUN GETUR ÞAÐ VALDIÐ MEIÐSLUM
ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS VEGNA RAFLOSTS,
VÉLRÆNNAR EÐA ANNARRAR HÆTTU.
TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE.
•VPŘÍPADĚ, ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTŘEBUJE NÁVOD
V JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA
ÚKOLEM ZÁKAZNÍKA.
• NEPROVÁDĚJTE ÚDRŽBU TOHOTO ZAŘÍZENÍ, ANIŽ BYSTE SI PŘEČETLI
TENTO SERVISNÍ NÁVOD A POCHOPILI JEHO OBSAH.
•VPŘÍPADĚ NEDODRŽOVÁNÍ TÉTO VÝSTRAHY MŮŽE DOJÍT ÚRAZU
ELEKTRICKÁM PROUDEM PRACOVNÍKA POSKYTOVATELE SLUŽEB,
OBSLUŽNÉHO PERSONÁLU NEBO PACIENTŮ VLIVEM ELEKTRICKÉHOP
PROUDU, RESPEKTIVE VLIVEM K RIZIKU MECHANICKÉHO POŠKOZENÍ NEBO
JINÉMU RIZIKU.
DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK.
• HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END
ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
• FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE
DENNE SERVICEMANUAL ER BLEVET LÆST OG FORSTÅET.
• MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFØRE SKADE
PÅ GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR
TEKNIKEREN, OPERATØREN ELLER PATIENTEN.
DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS VERKRIJGBAAR.
• ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS DE
KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN.
• PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE
ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS.
• INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET
ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATIËNT GEWOND
KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK,
MECHANISCHE OF ANDERE GEVAREN.
iv -
GEVOLUSON i / VOLUSON e
BRĪDINĀJUMS
(LV)
ĮSPĖJIMAS
(LT)
ADVARSEL
(NO)
OSTRZEŻENIE
(PL)
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ.
• JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ
VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU.
• NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS
IZLASĪŠANAS UN SAPRAŠANAS.
•ŠĪ BRĪDINĀJUMA NEIEVĒROŠANA VAR RADĪT ELEKTRISKĀS STRĀVAS
TRIECIENA, MEHĀNISKU VAI CITU RISKU IZRAISĪTU TRAUMU APKALPES
SNIEDZĒJAM, OPERATORAM VAI PACIENTAM.
ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA.
• JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ,
VERTIMU PASIRŪPINTI TURI KLIENTAS.
•NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT
VADOVAUTUMĖTĖS ŠIUO EKSPLOATAVIMO VADOVU IR JĮ SUPRASTUMĖTE
• NEPAISANT ŠIO PERSPĖJIMO, PASLAUGŲ TEIKĖJAS, OPERATORIUS AR
PACIENTAS GALI BŪTI SUŽEISTAS DĖL ELEKTROS SMŪGIO, MEC
HANINIŲ AR
KITŲ PAVOJŲ.
DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK.
• HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET
KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
• IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE
SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
• MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT
SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ
GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER.
NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU
ANGIELSKIM.
•JEŚLI FIRMA ŚWIADCZĄCA KLIENTOWI USłUGI SERWISOWE WYMAGA
UDOSTĘPNIENIA PODRĘCZNIKA W JĘZYKU INNYM NIŻ ANGIELSKI,
OBOWIĄZEK ZAPEWNIENIA STOSOWNEGO TłUMACZENIA SPOCZYWA NA
KLIENCIE.
• NIE PRÓBOWAĆ SERWISOWAĆ NINIEJSZEGO SPRZĘTU BEZ UPRZEDNIEGO
ZAPOZNANIA SIĘ Z PODRĘCZNIKIEM SERWISOWYM.
• NIEZASTOSOWANIE SIĘ DO TEGO OSTRZEŻENIA MOżE GROZIĆ
OBRAŻENIAMI CIAłA SERWISANTA, OPERATORA LUB PACJENTA W WYNIKU
PORAŻENIA PRĄDEM, URAZU MECHANICZNEGO LUB INNEGO RODZAJU
ZAGROŻEŃ.
- v
GEVOLUSON i / VOLUSON e
ATE NŢIE
(RO)
ОСТОРОЖНО!
(RU)
(BG)
ПРЕДУПРЕЖДЕНИЕ
UPOZORENJE
(SR)
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ.
• DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ
DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O
TRADUCERE.
• NU ÎNCERCAŢI SĂ REPARAŢI ECHIPAMENTUL DECÂT ULTERIOR
CONSULTĂRII ŞI ÎNŢELEGERII ACESTUI MANUAL DE SERVICE.
• IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA RĂNIREA
DEPANATORULUI, OPERATORULUI SAU PACIENTULUI ÎN URMA
PERICOLELOR DE ELECTROCUTARE, MECANICE SAU DE ALTĂ NATURĂ.
ДАННОЕ РУКОВОДСТВО ПО ОБСЛУЖИВАНИЮ ПРЕДОСТАВЛЯЕТСЯ ТОЛЬКО
НА АНГЛИЙСКОМ ЯЗЫКЕ.
• ЕСЛИ СЕРВИСНОМУ ПЕРСОНАЛУ КЛИЕНТА НЕОБХОДИМО РУКОВОДСТВО
НЕ НА АНГЛИЙСКОМ ЯЗЫКЕ, КЛИЕНТУ СЛЕДУЕТ САМОСТОЯТЕЛЬНО
ОБЕСПЕЧИТЬ ПЕРЕВОД.
• ПЕРЕД ОБСЛУЖИВАНИЕМ ОБОРУДОВАНИЯ ОБЯЗАТЕЛЬНО ОБРАТИТЕСЬ
К ДАНН
ОМУ РУКОВОДСТВУ И ПОЙМИТЕ ИЗЛОЖЕННЫЕ В НЕМ СВЕДЕНИЯ.
• НЕСОБЛЮДЕНИЕ УКАЗАННЫХ ТРЕБОВАНИЙ МОЖЕТ ПРИВЕСТИ К ТОМУ,
ЧТО СПЕЦИАЛИСТ ПО ТЕХОБСЛУЖИВАНИЮ, ОПЕРАТОР ИЛИ ПАЦИЕНТ
ПОЛУЧАТ УДАР ЗЛЕКТРИЧЕСКИМ ТОКОМ, МЕХАНИЧЕСКУЮ ТРАВМУ ИЛИ
ДРУГОЕ ПОВРЕЖДЕНИЕ.
ТОВА СЕРВИЗНО РЪКОВОДСТВО Е НАЛИЧНО САМО НА АНГЛИЙСКИ ЕЗИК.
• АКО ДОСТАВЧИКЪТ НА СЕРВИЗНИ УСЛУГИ НА КЛИЕНТ СЕ НУЖДАЕ ОТ
ЕЗ
ИК, РАЗЛИЧЕН ОТ АНГЛИЙСКИ, ЗАДЪЛЖЕНИЕ НА КЛИЕНТА Е ДА
ПРЕДОСТАВИ ПРЕВОДАЧЕСКА УСЛУГА.
• НЕ СЕ ОПИТВАЙТЕ ДА ИЗВЪРШВАТЕ СЕРВИЗНО ОБСЛУЖВАНЕ НА ТОВА
ОБОРУДВАНЕ, ОСВЕН ВСЛУЧАЙ, ЧЕ СЕРВИЗНОТО РЪКОВОДСТВО Е
ПРОЧЕТЕНО И СЕ РАЗБИРА.
• НЕСПАЗВАНЕТО НА ТОВА ПРЕДУПРЕЖДЕНИЕ МОЖЕ ДА ДОВЕДЕ ДО
НАРАНЯВАНЕ НА ДОСТАВЧИКА НА СЕРВИЗНИ УСЛУГИ, НА ОПЕРАТОРА
ИЛИ ПАЦИ
ЕНТА ВСЛЕДСТВИЕНА ТОКОВ УДАР, МЕХАНИЧНИ ИЛИ ДРУГИ
РИСКОВЕ.
OVAJ PRIRUČNIK ZA SERVISIRANJE DOSTUPAN JE SAMO NA ENGLESKOM
JEZIKU.
• AKO KLIJENTOV SERVISER ZAHTEVA JEZIK KOJI NIJE ENGLESKI,
ODGOVORNOST JE NA KLIJENTU DA PRUŽI USLUGE PREVOĐENJA.
• NEMOJTE POKUŠAVATI DA SERVISIRATE OPREMU AKO NISTE PROČITALI I
RAZUMELI PRIRUČNIK ZA SERVISIRANJE.
• AKO NE POŠTUJETE OVO UPOZORENJE, MOŽE DOĆI DO POVREĐIVANJA
SERVISERA, OPERATERA ILI PACIJENTA UZROKOVANOG ELEKTRIČNIM
UDAROM, MEHANIČKIM I DRUGIM OPASNOSTIMA.
vi -
GEVOLUSON i / VOLUSON e
OPOZORILO
(SL)
UPOZORENJE
(HR)
UPOZORNENIE
(SK)
VARNING
(SV)
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
TA SERVISNI PRIROČNIK JE NA VOLJO SAMO V ANGLEŠČINI.
• ČE PONUDNIK SERVISNIH STORITEV ZA STRANKO POTREBUJE NAVODILA V
DRUGEM JEZIKU, JE ZA PREVOD ODGOVORNA STRANKA SAMA.
• NE POSKUŠAJTE SERVISIRATI OPREME, NE DA BI PREJ PREBRALI IN
RAZUMELI SERVISNI PRIROČNIK.
• ČE TEGA OPOZORILA NE UPOŠTEVATE, OBSTAJA NEVARNOST
ELEKTRIČNEGA UDARA, MEHANSKIH ALI DRUGIH NEVARNOSTI IN
POSLEDIČNIH POŠKODB PONUDNIKA SERVISNIH STORITEV, UPORABNIKA
OPREME ALI PACIENTA.
OVAJ SERVISNI PRIRUČNIK DOSTUPAN JE SAMO NA ENGLESKOM JEZIKU.
• AKO KLIJENTOV SERVISER ZAHTIJEVA JEZIK KOJI NIJE ENGLESKI,
ODGOVORNOST KLIJENTA JE PRUŽITI USLUGE PREVOĐENJA.
• NEMOJTE POKUŠAVATI SERVISIRATI OPREMU AKO NISTE PROČITALI I
RAZUMJELI SERVISNI PRIRUČNIK.
• AKO NE POŠTUJETE OVO UPOZORENJE, MOŽE DOĆI DO OZLJEDE
SERVISERA, OPERATERA ILI PACIJENTA PROUZROČENE STRUJNIM
UDAROM, MEHANIČKIM I DRUGIM OPASNOSTIMA.
TÁTO SERVISNÁ PRÍRUČKA JE K DISPOZÍCII LEN V ANGLIČTINE.
• AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO
ANGLIČTINU, POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE
ZODPOVEDNOSŤOU ZÁKAZNÍKA.
• NEPOKÚŠAJTE SA VYKONÁVAŤ SERVIS ZARIADENIA SKÔR, AKO SI
NEPREČÍTATE SERVISNÚ PRÍRUČKU A NEPOROZUMIETE JEJ.
• ZANEDBANIE TOHTO UPOZORNENIA MÔŽE VYÚSTIŤ DO ZRANENIA
POSKYTOVATEĽA SLUŽIEB, OBSLUHUJÚCEJ OSOBY ALEBO PACIENTA
ELEKTRICKÝM PRÚDOM, PRÍPADNE DO MECHANICKÉHO ALEBO INÉHO
NEBEZPEČENSTVA.
DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA.
• OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN
ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA
ÖVERSÄTTNINGSTJÄNSTER.
• FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST
OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN.
• OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I
SKADOR PÅ SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL
FÖLJD AV ELEKTRISKA STÖTAR, MEKANISKA FAROR ELLER ANDRA FAROR.
- vii
GEVOLUSON i / VOLUSON e
DİKKAT
(TR)
(JA)
Traditional
Chinese
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR.
• BU UYARININ GÖZ ARDI EDİLMESİ, ELEKTRİK ÇARPMASI YA DA MEKANİK
VEYA DİĞER TÜRDEN KAZALAR SONUCUNDA TEKNİSYENİN, OPERATÖRÜN
YA DA HASTANIN YARALANMASINA YOL AÇABİLİR.
viii -
GEVOLUSON i / VOLUSON e
(ZH-CN)
(KO)
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
- ix
GEVOLUSON i / VOLUSON e
DANGER
WARNINGWARNING
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY
All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the
equipment shall be performed by licensed electrical contractors. Other connections between pieces of
electrical equipment, calibrations and testing shall be performed by qualified GE personnel. In
performing all electrical work on these products, GE will use its own specially trained field engineers. All
of GE’s electrical work on these products will comply with the requirements of the applicable electrical
codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.
OMISSIONS & ERRORS
If there are any omissions, errors or suggestions for improving this documentation, please contact the
GE Global Documentation Group with specific information listing the system type, manual title, part
number, revision number, page number and suggestion details.
Mail the information to:
GE Healthcare Austria GmbH & Co. OG
Tiefenbach 15
A-4871 Zipf
Austria - Europe
Attn.: “Service Documentation”
GE employees should contact the product's Lead Service Integrator (LSI) to report service
documentation issues. DO NOT use TrackWise for this purpose. The LSI will then use the internal
problem reporting tool to communicate these issues to the writer.
SERVICE SAFETY CONSIDERATIONS
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND
ADJUSTING.
Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury.
For a complete review of all safety requirements, see the Chapter 1, Safety Considerations section in
the Service Manual.
x -
GEVOLUSON i / VOLUSON e
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without
prior written permission of GE.
GE may revise this publication from time to time without written notice.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
This chapter describes important issues related to safely servicing the Voluson i / Voluson e.
The service provider must read and understand all the information presented in this manual before
installing or servicing a unit.
Always observe the instructions given in the manual of the peripheral/auxiliary device.
Under consideration of general maintenance requirements a minimum lifetime of 7 years for the
equipment and 5 years for the probes may be expected. To maintain the safety and performance of the
ultrasound system, a regular check (once per year) by authorized personnel is recommended.
Table 1-1 Contents in Chapter 1
SectionDescriptionPage Number
1-1
1-2
1-3
1-4
1-5
Overview
Important Conventions
Safety Considerations
Electromagnetic Compatibility (EMC)
Customer Assistance
1-1-2Purpose of this Service Manual
This Service Manual is valid for Voluson i / Voluson e ultrasound systems.
!! NOTICE:
For the sake of simplicity, descriptions in this manual will be briefly denot ed
with Voluson i / Voluson e if no special differentiation is required.
This Service Manual provides installation and service information for the Voluson i / Voluson e portable
Ultrasound Scanning System and contains the following chapters:
1.) Chapter 1 - Introduction: Contains a content summary and warnings.
procedures for all changeable Field Replaceable Units (FRU).
9.) Chapter 9 - Renewal Parts: Contains a complete list of field replaceable parts.
10.)Chapter 10 - Care & Maintenance: Provides periodic maintenance procedures.
1-1
1-3
1-8
1-17
1-18
Chapter 1 - Introduction1-1
GEVOLUSON i / VOLUSON e
NOTICE
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
1-1-3Typical Users of the Basic Service Manual
•GE Service Personnel (installation, maintenance, etc.).
•Hospital’s Service Personnel
•Contractors (Some parts of Chapter 2 - Site Preparation)
1-1-4Models Covered by this Manual
Table 1-2 Voluson i / Voluson e Model Designations
Part NumberDescription
H48651KRVoluson i Console (BT06)
H48661DVoluson e Console (BT06)
H48661ADVoluson i Console (BT07)
H48651WNVoluson e Console (BT07)
H48671PWVoluson i Console (BT09/11)
H48671PSVoluson e Console (BT09/11)
H48691LSVoluson i Console CC200 (BT14)
H48691LTVoluson e Console CC200 (BT14)
The Voluson i / Voluson e is a compact, fully portable, phased, linear array ultrasound imaging scanner.
Weighing only 5 kg / 11 lbs. (without peripherals), this laptop-style system offers many advanced
features and is extremely versatile. Depending upon the installed software, the Voluson i / Voluson e
portable ultrasound system can be used for a variety of ultrasound scanning applications.
The Voluson i / Voluson e portable ultrasound systems are based on parallel products in the Voluson
family (Voluson 730, Voluson 730Expert, Voluson 730Pro/ProV and Voluson E8).
Using the same advanced technology, the innovative features and versatility of the system represent
the future of ultrasound scanning.
The Voluson e is a “feature-reduced” version of the Voluson i portable ultrasound system.
!! NOTICE:
That means not all options are available on the Voluson e (marked with * in sections of this Manual).
The Voluson e is only distinguished by the System type, see: Figure 7-1: System Setup - System
Info page (e.g., Voluson e) on page 7-2; it has NO own serial number!
NOTE:Upgrade options may be available with future versions.
1-1-5System History - Hardware and Software Versions
This manual applies to:
•Voluson i / Voluson e systems with Serial Number B00020 - B00742 (BT06)
•Voluson i systems with Serial Number B01001 - B01494 (BT07)
•Voluson e systems with Serial Number B50001 - B50799 (BT07)
•Voluson i systems with Serial Number B02000 - B02779 (BT09)
•Voluson e systems with Serial Number B52000 - B53200 (BT09)
•Voluson i systems with Serial Number B03001 - (BT11)
•Voluson e systems with Serial Number B53261 - (BT11)
•Voluson i systems with Serial Number B05001 - (CC200 / BT14)
•Voluson e systems with Serial Number B57001 - (CC200 / BT14)
1-2Section 1-1 - Overview
GEVOLUSON i / VOLUSON e
DANGER
WARNINGWARNING
CAUTION
NOTICE
BT Version:
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
1-1-6Purpose of Operator Manual(s)
The Operator Manual(s) should be fully read and understood before operating the
Voluson i / Voluson e and also kept near the unit for quick reference.
Section 1-2
Important Conventions
1-2-1Conventions Used in this Manual
Model Designations
This manual covers the Voluson i / Voluson e ultrasound units listed in Table 1-2 on page 1-2.
Icons
Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels and
conventions used on the product and in the service information are described in this chapter.
Safety Precaution Messages
Various levels of safety precaution messages may be found on the equipment and in the service
information. The different levels of concern are identified by a flag word that precedes the precautionary
message. Known or potential hazards are labeled in one of following ways:
INDICATES THE PRESENCE OF A HAZARD THAT WILL CAUSE SEVERE
!! DANGER:
!! WARNING:
!! CAUTION:
!! NOTICE:
BT-Version:
NOTE:Notes provide important information about an item or a procedure.
PERSONAL INJURY OR DEATH IF THE INSTRUCTIONS ARE IGNORED.
INDICATES THE PRESENCE OF A HAZARD THAT CAN CAUSE SEVERE PERSONAL
INJURY AND PROPERTY DAMAGE IF INSTRUCTIONS ARE IGNORED.
Indicates the presence of a hazard that will or can cause minor personal injury and property
damage if instructions are ignored.
Equipment Damage Possible.
Notice is used when a hazard is present that can cause property damage but has absolutely no
personal injury risk.
Example: Disk drive will crash.
Is used when options or features are specific for BT-Software versions.
Information contained in a NOTE can often save you time or effort.
Chapter 1 - Introduction1-3
GEVOLUSON i / VOLUSON e
LASER
LIGHT
Signed
Date
TAG
&
LOCKOUT
EYE
PROTECTION
or
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
1-2-2Standard Hazard Icons
Important information will always be preceded by the exclamation point contained within
a triangle, as seen throughout this chapter. In addition to text, several different graphical
icons (symbols) may be used to make you aware of specific types of hazards that could
cause harm. Even if a symbol isn’t used in this manual, it is included for your reference.
Table 1-3 Standard Hazard Icons
ELECTRICALMECHANICALRADIATION
LASERHEATPINCH
Other hazard icons make you aware of specific procedures that should be followed.
Table 1-4 Standard Icons Indicating a Special Procedure be Used
AVOID STATIC ELECTRICITYTAG AND LOCK OUTWEAR EYE PROTECTION
WEAR HAND PROTECTIONWEAR FOOT PROTECTION
1-4Section 1-2 - Important Conventions
GEVOLUSON i / VOLUSON e
BT Version:
0123
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
1-2-3Product Labels and Icons
The following table describes the purpose and location of safety labels and other important information
provided on the equipment.
Description of symbols and labels used in combination with this Voluson ultrasound system.
BT-Version:
Product labels and its location depend on BT-version and/or date of manufacture. For description of all
symbols and labels used on the system, refer to Chapter 2 in the Basic User Manual of your system.
Table 1-5 Product Icons
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
Identification and Rating Plate
Manufacturer's name and address
Model and serial numbers
Electrical ratings
Manufacturer’s name and address
Date of manufacture
Serial number
Catalog or model number
on the base of the unit (Main label)
on each probe
Identification and Rating Plate(rear
side of the unit / on plug of each probe)
Identification and Rating Plate(rear
side of the unit / on plug of each probe)
Identification and Rating Plate(rear
side of the unit / on plug of each probe)
Identification and Rating Plate(rear
side of the unit / on plug of each probe)
Device Listing/Certification Labels
Type/Class Label
IP Code (IPX 0)
IP Code (IPX 1)
IP Code (IPX 7)
Laboratory logo or labels denoting conformance with industry
safety standards such as UL or IEC.
CE Conformity mark according to Medical Device Directive
93/42/EEC.
0123: Identification number of notified body TÜV Süd Product
Service.
“Tested and production monitored by TÜV Product Service
NRTL with respect to ELECTRICAL SHOCK, FIRE and
MECHANICAL HAZARDS only in accordance with UL2601-1
and CAN/CSA C22.2 NO.601.1.”
Used to indicate the degree of safety or protection.
degree of protection provided by the enclosure per IEC 60529:
IPX 0 - no protection against ingress of water
IPX 1 - protected against dripping water
IPX 7 - protected against the effects of immersion
Equipment Type BF (man in the box symbol) IEC 878-02-03
indicates B Type equipment having even more electrical
isolation than standard Type B equipment because it is
intended for intimate patient contact.
on the base of the unit (Main label)
on the base of the unit (Main label)
on each probe
Identification and Rating Plate
(rear side of the unit)
various
probe connectors
on the base of the unit (Main label)
Chapter 1 - Introduction1-5
GEVOLUSON i / VOLUSON e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Table 1-5 Product Icons (Continued)
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
“CAUTION
This unit weighs...
Special care must be used to avoid..."
This precaution is intended to prevent injury that may result if
one person attempt to move heavy parts (e.g., the optional
Modo Cart, Dock Cart or Voluson Station) considerable
distances or on an incline due to its weight.
Two people are required whenever a part weighing 16kg
(35 lbs.) or more must be lifted.
Pushing prohibited.
Do not lean on the system. Tipping danger.
Take special care when moving the system
Loading prohibited.
Do not put any items on this shelf. Danger of breaking.
Also items might be crushed when lowering the user interface.
“ATTENTION” - Read and understand all instructions for use”
This symbol advises the reader to consult the accompanying
documents (operator manual or other instructions).
Used in the Service and User Manual
which should be adjacent to equipment
at all times for quick reference.
various
at top cover of the system
rear side of the unit
"DANGER -
Risk of explosion used in..."
The system is not designed for use with flammable anesthetic
gases.
"CAUTION" The equilateral triangle is usually used in
combination with other symbols to advise or warn the user.
ATTENTION - Consult accompanying documents "
This symbol is used to advise the reader to consult the
accompanying documents for important safety-related
information such as warnings and pre-cautions that cannot be
presented on the device itself.
"CAUTION - Dangerous electric voltage" (lightning flash with
arrowhead) is used to indicate electric shock hazards.
Unplug the main plug before opening the system.
CAUTION - Laser radiation Avoid exposure to the beam Class
3B laser product.
CAUTION - Class 3B laser radiation. When open avoid
exposure to the beam.
Indicated in the Service Manual.
various
various
various
various
"OFF/Standby" Indicates the power off/standby position of the
power switch.
CAUTION
This Power Switch DOES NOT ISOLATE Main Supply
1-6Section 1-2 - Important Conventions
adjacent to On-Off/Standby switch
GEVOLUSON i / VOLUSON e
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
Table 1-5 Product Icons (Continued)
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
"Protective Earth" Indicates the protective earth (grounding)
terminal.
"Equipotentiality" Indicates the terminal to be used for
connecting equipotential conductors when interconnecting
(grounding) with other equipment.
Waste Electrical and Electronic Equipment (WEEE) Disposal.
This symbol indicates that waste electrical and electronic
equipment must not be disposed as unsorted municipal waste
and must be collected separately. Please contact an
authorized representative of the manufacturer for information
concerning the decommissioning of your equipment.
This symbol indicates that in the United States of America,
federal law restricts this device to sale by or on the order of a
physician.
GOST-R label (Russia Regulatory Country Clearance)
various
peripherals
isolation transformer on Modo Cart
on the base of the unit (Main label)
on back of the Modo Cart (Main label)
on back of the Dock Cart (Main label)
on back of the Voluson Station
(Main label)
on each probe
Identification and Rating Plate
(rear side of the unit)
Identification and Rating Plate
(rear side of the unit)
These symbols indicate that at least one of the six hazardous
substances of the China RoHS Labelling Standard is above
the RoHS limitation.The number inside the circle is referred to
as the Environmental Friendly Use Period (EFUP). It indicates
the number of years that the product, under normal use, will
remain harmless to health of humans or the environment.
EFUP = 10 for Short Use Products
EFUP = 20 for Medium Use Products
This product consists of devices that may contain mercury,
which must be recycled or disposed of in accordance with
local, state, or country laws.
(Within this system, the backlight lamps in the monitor and the
Touch Panel display, contain mercury.)
Rear side of the unit
on the plug of each probe
Rear side of the unit
on the Display Module Assy (GPD)
Chapter 1 - Introduction1-7
GEVOLUSON i / VOLUSON e
CAUTION
CAUTION
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Section 1-3
Safety Considerations
1-3-1Introduction
The following safety precautions must be observed during all phases of operation, service and repair of
this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this
manual, violates safety standards of design, manufacture and intended use of the equipment.
1-3-2Human Safety
Operating personnel must not remove the system covers. Servicing should be performed by authorized
personnel only. Only personnel who have participated in a Voluson i / Voluson e Training are authorized
to service the equipment.
1-3-3Mechanical Safety
The Voluson i / Voluson e portable ultrasound system weighs 5 kg / 11 lbs. (without peripherals)
!! CAUTION:
!! CAUTION:
when ready for use.
Care must be used when moving it or replacing its parts. Failure to follow the precautions listed
could result in injury, uncontrolled motion and costly damage.
The Cart (incl. Voluson i / Voluson e and peripherals) weighs approximately 30 kg / 66 lbs. when
it is ready for use.
Two people are required when moving on inclines or lifting more than 16 kg (35 lbs).
Move the Cart forward or backward when going up or down inclines. Do not move the Cart
sideways or diagonally.
ALWAYS:
• Be sure the pathway is clear. • Use slow, careful motions.
It is not possible to fix the Voluson i / Voluson e on mounting plate of the Cart.
!! WARNING:
When the Voluson i / Voluson e scanner is used with the Cart, make sure the unit is
firmly placed in correct position since it may become unstable and tip over.
Use extreme caution when elevating the Cart , or if it is moved along any incline. Make
!! WARNING:
sure that the Voluson i / Voluson e scanner and all peripherals are s ecurely place d on
the Cart before attempting to move it.
ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN
!! WARNING:
EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND
PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR
DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN
SERIOUS INJURY AND EQUIPMENT DAMAGE.
NEVER USE A PROBE THAT HAS FALLEN TO THE FLOOR.
!! WARNING:
EVEN IF IT LOOKS OK, IT MAY BE DAMAGED.
WHILE THE SOFTWARE INSTALL PROCEDURE IS DESIGNED TO PRESERVE DATA,
!! WARNING:
YOU SHOULD SAVE ANY PATIENT DATA, IMAGES, SYSTEM SETUPS TO A DVD/CD OR
HARDCOPY.
Always lock the Cart in its parked (locked) position after moving the system. Failure to
!! WARNING:
do so could result in personal injury or equipment damage.
Equipment damage could result if special care is not taken when transporting the
!! WARNING:
1-8Section 1-3 - Safety Considerations
Voluson i / Voluson e system (and Carts if applicable) in a vehicle.
GEVOLUSON i / VOLUSON e
WARNINGWARNING
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
1-3-3Mechanical Safety (cont’d)
Always:
•Eject any DVD/CD from the drive.
•Ensure that the Voluson i / Voluson e system is well prepared and packed in its original packaging
before transporting. Special care must be taken to correctly position the packing material. For
further information, refer to Chapter 3 - Setup Instructions on page 3-1.
•Place the probes in their carrying cases.
•Secure the Cart in an upright position and lock the wheels (brake).
•Ensure that the Voluson i / Voluson e system (and Cart if applicable) is firmly secured while inside
the vehicle.
•Secure the system with straps or as directed otherwise to prevent motion during transport.
•Prevent vibration damage by driving cautiously. Avoid unpaved roads, excessive speeds, and
erratic stops or starts.
1-3-4Electrical Safety
1-3-4-1Safe Practices
The system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety ground. If an extension cord is used with the system, make sure that the total
current rating of the system does not exceed the extension cord rating.
!! WARNING:
1-3-4-2Probes
The power outlet of the AC/DC converter used for this equipment should not be shared with other types
of equipment. Both, the system power cable and the power connector meet international electrical
standards.
CONNECTING A VOLUSON I / VOLUSON E SCANNER TO THE WRONG VOLTAGE
LEVEL WILL MOST LIKELY DESTROY THE SCANNER.
All the probes for the Voluson i / Voluson e portable ultrasound system are designed and manufactured
to provide trouble-free, reliable service. To ensure this, correct handling of probes is important and the
following points should be noted:
•Do not drop a probe or strike it against a hard surface, as this may damage the transducer elements,
acoustic lens, or housing.
•Inspect the probe prior to each use for damage or degradation to the Housing, Cable strain relief,
Lens and Seal.
•Do not use a cracked or damaged probe. In this event, call your field service representative
immediately to obtain a replacement.
•Avoid pulling, pinching or kinking the probe cable, since a damaged cable may compromise the
electrical safety of the probe.
•To avoid the risk of a probe accidentally falling, do not allow the probe cables to become entangled,
or to be caught in the machine’s wheels.
•Never immerse the probe connector or adapter into any liquid.
NOTE:For detailed information on handling probes, refer to the Volu son i / Volu son e Basic User Manual an d
the care card supplied with the probe.
Chapter 1 - Introduction1-9
GEVOLUSON i / VOLUSON e
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
NOTICE
NOTICE
CAUTION
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
1-3-5Auxiliary Devices Safety
Power Supplies for additional equipment MUST comply with IEC 60601-1.
!! WARNING:
DO NOT attempt to use different peripherals and accessories (brand and model;
!! WARNING:
!! WARNING:
connected via the PCMCIA or USB ports) other than approved and provided by
GE!
The ultrasound system is an extremely sensitive and complex medical system.
Any unauthorized peripherals may cause system failure or damage!
The IEC 60601-1-1 standard provides a guideline for safely interconnecting medical devices in systems.
“Equipment connected to the analog or digital interface must comply with the respective IEC/UL
standards (e.g. IEC 60950 / UL 60950 for data processing equipment and IEC 60601-1 / UL 60601-1
for medical equipment). Everybody who connects additional equipment to the signal input portion or
signal output portion configures a medical system, and is therefore responsible that the system
complies with the requirements of the system standard IEC 60601-1-1.
For Ethernet connection, a separation device has to be used.
Special care has to be taken, if the device is connected to computer network (e.g., Ethernet), because
other devices could be connected without any control. There could be a potential difference between
the protective earth and any line of the computer network including the shield.
!! NOTICE:
In this case the only way to operate the system safely is to use an isolated signal link with minimum
4mm creepage distance, 2.5mm air clearance of the isolation device. For computer networks there are
media converters available which convert the electrical to optical signals. Please consider that this
converter has to comply with IEC xxx standards* and is battery operated.
* IEC xxx stands for standards such as:
• IEC 60601 for medical devices
• IEC 60950 for information technology equipment etc.
Internally powered, applied parts type BF according to IEC60601-1 including national deviations
Information for service personnel:
For safety test without AC-mains connection (battery operated): apply Class II limits for electrical safety
test of Voluson i/e (e.g. NC (normal condition) 100µA touch-current acc. IEC60601-1)
For safety test with AC-mains connection (by AC/DC adapter or Voluson Station/Cart): apply Class I
limits for electrical safety test of Voluson i/e (e.g. NC (normal condition) 500µA earth-leakage-current,
100µA enclosure leakage current, 100µA touch-current; SFC (single fault condition, like open PE or
open L/N) 1000µA earth-leakage-current, 500µA enclosure leakage current, 500µA touch-current; acc.
IEC60601-1); Do not perform ground continuity test on accessible conducting parts of the Voluson i/e.
These parts are either isolated (floating) or connected to functional earth but not protective earth, i.e.
the connection is not designed to withstand 25A ground continuity test and can damage the Voluson i/e.
Always observe national applicable regulations which might require different limits
The system integrator (any person connecting the medical device to other devices) is responsible
that the connections are safe.
If in doubt, consult the technical service department or your local representative.
The leakage current of the entire system including any / all auxiliary equipment must not exceed
!! CAUTION:
the limit values as per EN 60601-1-1:1990 (IEC 60601-1-1) respectively other valid national or
international standards. All equipment must comply with UL, CSA and IEC requirements.
1-10Section 1-3 - Safety Considerations
GEVOLUSON i / VOLUSON e
CAUTION
CAUTION
CAUTION
WARNINGWARNING
NOTICE
BT Version:
Main Label
Quality Control label
Assembly label
Windows XP
License key label
Warning Label
Voluson i / Voluson e
Hg Vermont label
(for US only)
(Windows XP embedded
with Software 8.x.x)
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
1-3-5Auxiliary Devices Safety (cont’d)
Please observe that some printers may not be medical devices! If the Bluetooth Printer and/or
!! CAUTION:
!! CAUTION:
!! CAUTION:
!! WARNING:
Line Printers are no medical devices, they have to be located outside of the patient environment
(according to IEC 60601-1 / UL 60601-1).
Auxiliary equipment must only be connected to the main console with the special main outlet
provided for the electrical safety of the system.
Auxiliary equipment with direct main connection requires galvanic separation of the signal and/
or control leads.
For hardware installation procedures see: Chapter 3 - Connection of Auxiliary Devices on page 3-23.
After each installation, the leakage currents have to be measured according to
IEC 60601-1 respectively UL 60601-1 (see: Section 10-7-2 on page 10-13).
!! NOTICE:
1-3-6Label Location
BT-Version:
When the Voluson i / Voluson e scanner is used with the Cart:
• make sure the voltage jumper on the isolation transformer of the Cart is in the correct position.
• all peripherals mounted on the Cart should be firmly secured in position using straps.
• use only the on board power outlets from transformer and USB ports of Voluson i / Voluson e
to connect devices.
The Voluson i / Voluson e ultrasound system comes equipped with product labels and icons.
These labels and icons represent pertinent information regarding the operation of the unit.
Product labels and its location depend on BT-version and/or date of manufacture.
Refer to Section 1-2-3 on page 1-5 and Chapter 2 in the Basic User Manual of your system.
Figure 1-1 Labeling (on base of the Voluson i / Voluson e system)
Chapter 1 - Introduction1-11
GEVOLUSON i / VOLUSON e
BT Version:
1
3
4
5
6
7
8
9
2
4
1
2
3
4
5
6
7
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
1-3-6-1Main Label on Voluson i / Voluson e system
The Main Label is located on the base of the Voluson i / Voluson e system.
The layout of the Identification and Rating Plate depe nds on BT-version and/or date of manufacture.
Figure 1-2 Main Label (located on the base of the system, e.g., Voluson e)
1 Manufacturing date4 Model type and Serial number7 CE marking
2 TÜV NRTL Classified5 WEEE Disposal Icon8 FDA Guidance document
3 Manufacturer6 Safety type: Type BF9 GHOST R label
1-3-6-2Main Label on optional Modo Cart
The Main Label is located on the back of the Modo Cart.
GE Medical Systems
Kretztechnik GmbH & Co OHG
Tiefenbach 15
4871 Zipf Austria
VOLUSON Cart
Input:
230(220-240)V~ 50/60Hz
Output:
max. 300VA
Figure 1-3 Main Label (located on back of the Modo Cart)
1 Model Type4 Manufacturer
115(100-130)V/
330W
100-130V~/220-240V~
SN: BCxxxx
!
Protection Class I
7S126_2
XXXXX
2 Input/Output Specification 5 Serial number
3 WEEE Disposal Icon6 Manufacturing date
7 Protection Class
1-12Section 1-3 - Safety Considerations
GEVOLUSON i / VOLUSON e
1
2
3
4
5
1
2
3
4
5
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
1-3-6-3Main Label on optional Dock Cart
The Main Label is located on the back of the Dock Cart.
Figure 1-4 Main Label (located on back of the Dock Cart)
1 Model Type4 Manufacturer
2 Input Specification 5 Manufacturing date
3 WEEE Disposal Icon
1-3-6-4Main Label on optional Voluson Station
The Main Label is located on the back of the Voluson Station.
Figure 1-5 Main Label (located on back of the Voluson Station)
1 Model Type4 Manufacturer
2 Input Specification 5 Manufacturing date
3 WEEE Disposal Icon
Chapter 1 - Introduction1-13
GEVOLUSON i / VOLUSON e
DANGER
WARNINGWARNING
WARNINGWARNING
WARNINGWARNING
CAUTION
WARNINGWARNING
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
1-3-7Dangerous Procedure Warnings
Warnings, such as the examples below, precede potentially dangerous procedures throughout this
manual. Instructions contained in the warnings must be followed.
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT
!! DANGER:
IN THIS EQUIPMENT.
USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
EXPLOSION WARNING
!! WARNING:
!! WARNING:
!! WARNING:
DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE.
OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT
CONSTITUTES A DEFINITE SAFETY HAZARD.
DO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT
BECAUSE OF THE DANGER OF INTRODUCING ADDITIONAL HAZARDS, DO NOT
INSTALL SUBSTITUTE PARTS OR PERFORM ANY UNAUTHORIZED MODIFICATION
OF THE EQUIPMENT.
Use only batteries approved by GE as suitable for the Voluson i / Voluson e portable
ultrasound system. In addition to Section 1-3-8 "Voluson i / Voluson e Battery Safety"
on page 1-14, READ all safety precautions that come from the battery and battery
charger manufacturers.
1-3-8Voluson i / Voluson e Battery Safety
NOTE:The Voluson i / Voluson e system is supplied with a lithium ion battery in the battery bay, as standard.
The lithium ion battery provides power when an AC power source is not available. Lithium ion batteries
last longer than conventional batteries and do not require replacement as often. You can expect one
hour of battery life with a single, fully-charged battery.
The lithium ion technology used in the system’s battery is significantly less hazardous to the
environment than the lithium metal technology used in some other batteries (such a watch batteries).
Used batteries should not be placed with common household waste products. Contact local authorities
for the location of a chemical waste collection program nearest you.
NOTE:Regulations vary for different countries. Dispose of a used battery in accordance with local regulations.
Use only batteries approved by GE as suitable for use with the Voluson i / Voluson e portable
!! CAUTION:
!! WARNING:
ultrasound system.
The Voluson i / Voluson e battery has a safety device.
DO NOT ATTEMPT TO DIS-ASSEMBLE OR ALTER THE BATTERY!
Always observe following precautions:
1-14Section 1-3 - Safety Considerations
GEVOLUSON i / VOLUSON e
WARNINGWARNING
CAUTION
NOTICE
NOTICE
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
1-3-8Voluson i / Voluson e Battery Safety (cont’d)
•Charge batteries only when the ambient temperature is between 0° and 65° C (32° and 149° F).
and discharge the batteries between -10° and 55° C (14° and 131° F).
•Do not short-circuit the battery by directly connecting the negative terminals with metal objects.
•Do not heat the battery or discard it in a fire.
•Do not expose the battery to temperatures over 60° C (140° F). Keep the battery away from fire and
other heat sources.
•Do not charge the battery near a heat source, such as, a fire or heater.
•Do not leave the battery in direct sunlight.
•Do not pierce the battery with a sharp object, hit it, or step on it.
•Do not use a damaged battery.
•Do not apply solder to a battery.
•Do not connect the battery to an electrical power outlet.
In the event that the Voluson i / Voluson e system will not be used for a long period, it
!! WARNING:
!! CAUTION:
is necessary to remove the battery from the system while not in use.
To prevent the battery bursting, igniting, or fumes from the battery causing equipment damage,
always observe the following precautions:
•Do not immerse the battery in water or allow it to get wet.
•Do not place the battery into a microwave oven or pressurized container.
•If the battery leaks or emits an odor, remove it from all possible flammable sources.
•If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal during
use, recharging or storage, immediately remove it and stop using it.
•If you have any questions about the battery, consult your local GE representative.
•Storage of the battery pack:
Short term (less than 1 month): between 0° C (32° F) and 50° C (122° F).
Long-term (more than 3 months): between 10° C (50° F) and 35° C (95° F).
For further details refer to Chapter 3 - Internal Lithium Ion Battery on page 3-73.
When storing a battery pack for more than 6 months, it is necessary to charge the battery pack at least
once every 6 months in order to prevent leakage and deterioration in performance (as a result of selfdischarging).
When charging the battery for the first time after long-term storage, recover the battery pack to original
!! NOTICE:
performance through repeating several cycles of full charging and discharging.
NOTE:When shipped, the Voluson i / Voluson e battery is in a state of being approximately 30% charged.
Before use, it is necessary to fully charge and discharge the battery pack up to 3 times, in order to utilize
Li-lon smart packs.
Chapter 1 - Introduction1-15
GEVOLUSON i / VOLUSON e
NOTICE
Signed
Date
TAG
&
LOCKOUT
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
1-3-9Lockout/Tagout Requirements (For USA Only)
Follow OSHA Lockout/Tagout requirements to protect service personnel from injuries caused by
unexpected energizing or start-up of equipment during service, repair, or maintenance.
Energy Control and Power Lockout for Voluson i / Voluson e.
!! NOTICE:
When servicing parts of the system where there is exposure to voltage greater than 30 Volts:
Unplug the system
Maintain control of the system power plug
Remove Battery
There are no test points to verify isolation, you must wait for at least 20 seconds for capacitors to
discharge.
Beware that the Power Supply, Front End Processor and Back End Processor may be energized even
if the power is turned off when the cord is still plugged into the AC Outlet.
UT
1-3-10Returning/Shipping Probes and Repair Parts
Equipment being returned must be clean and free of blood and other infectious substances.
GEMS policy states that body fluids must be properly removed from any part or equipment prior to
shipment. GEMS employees, as well as customers, are responsible for ensuring that parts/equipment
have been properly decontaminated prior to shipment. Under no circumstance should a part or
equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or
an ultrasound probe).
The purpose of the regulation is to protect employees in the transportation industry, as well as the
people who will receive or open this package.
NOTE:The US Department of Transportation (DOT) h as ruled that “items what were saturated and/or dripping
with human blood that are now caked with dried blood; or which were used or intended for use in patient
care” are “regulated medical waste” for transportation p urpose and must be transported as a hazardous
material.
1-16Section 1-3 - Safety Considerations
GEVOLUSON i / VOLUSON e
WARNINGWARNING
WARNINGWARNING
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
Section 1-4
Electromagnetic Compatibility (EMC)
1-4-1What is EMC?
Electromagnetic compatibility describes a level of performance of a device within its electromagnetic
environment. This environment consists of the device itself and its surroundings including other
equipment, power sources and persons with which the device must interface. Inadequate compatibility
results when a susceptible device fails to perform as intended due interference from its environment or
when the device produces unacceptable levels of emission to its environment. This interference is often
referred to as radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through
space or conducted over interconnecting power of signal cables. In addition to electromagnetic energy,
EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and
disturbances in the electrical power supply.
For applicable standards refer to Chapter 2 in the Basic User Manual.
1-4-2Compliance
The Voluson i / Voluson e unit conforms to all applicable conducted and radiated emission limits and to
immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line
transient requirements as mentioned in IEC 60601-1-2.
NOTE:For CE Complia nce, it is critical that all covers, sc rews, shielding, gaskets, mesh, cla mps, are in
good condition, installed tightly without skew or stress. Proper installation following all
comments noted in this service manual is required in order to achieve full EMC performance.
1-4-3Electrostatic Discharge (ESD) Prevention
Do not touch any boards with integrated circuits prior to taking the necessary ESD
!! WARNING:
!! WARNING:
precautions:
1.) When installing boards, ESD may cause damage to a board. ALWAYS connect
yourself, via an arm-wrist strap, to any exposed metal part of the system.
2.) Follow general guidelines for handling of electrostatic sensitive equipment.
Risk of electrical shock: System must be turned off and Battery has to be removed.
Avoid all contact with electrical contacts, conductors and components.
Always use non-conductive handles designed for the removal and rep lacement of ESD
sensitive parts. All parts that have the potential for storing energy must be discharged
or isolated before making contact.
Chapter 1 - Introduction1-17
GEVOLUSON i / VOLUSON e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Section 1-5
Customer Assistance
1-5-1Contact Information
If this equipment does not work as indicated in this Service Manual or in the Basic User Manual, or if
you require additional assistance, please contact the local distributor or appropriate support resource,
as listed below.
NOTE:Prepare vital system information (see: Section 7-2 on page 7-2) befor e you call:
•System Type
•System Serial number (also visible on label on back of the system)
•Application Software version
•Backup version
•additional information about installed software
Table 1-6 Phone numbers for Customer Assistance
LOCATIONPHONE NUMBER
USA
GE Medical Systems
Ultrasound Service Engineering
9900 Innovation Drive (RP-2123)
Wauwatosa, WI 53226, USA
Canada1-800-668-0732
Latin America
Europe
GE Ultraschall Deutschland GmbH
Beethovenstraße 239
Postfach 11 05 60, D-42655 Solingen
Germany
GE Healthcare Austria GmbH & Co OG
Tiefenbach 15
A-4871 Zipf
Austria
+43 (0) 7682-3800-0+43 (0) 7682-3800-47
1-18Section 1-5 - Customer Assistance
GEVOLUSON i / VOLUSON e
CAUTION
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
Chapter 2
Site-Preparation
Section 2-1
Overview
2-1-1Purpose of Chapter 2
This chapter provides the information required to plan and prepare for the installation of a
Voluson i / Voluson e ultrasound unit.
Included are descriptions of the facility and electrical needs to be met by the purchaser.
Table 2-1 Contents in Chapter 2
SectionDescriptionPage Number
2-1
2-2
2-3
Overview
General Console Requirements
Facility Needs
Section 2-2
General Console Requirements
2-2-1Environmental Requirements
Table 2-2 Environmental Requirements
!! CAUTION:
2-2-1-1Cooling
The cooling requirement for the Voluson i / Voluson e is 300 BTU/hr. This figure does not include cooling
needed for lights, people, or other equipment in the room.
Operating
Temperature
10 to 30oC
(50 to 86
o
F)
If the system has been in storage or has been transported, please see the acclimation requirements
before powering
Operating
Humidity
30 to 80% rH
non-condensing
ON and/or using the system. See ”Installation Warn ings” on page 3-2
Heat Dissipation
300 BTU pr hour
Storage
Temperature
o
-10 to 40
(14 to 104
C
o
F)
Storage
Humidity
< 90% rH
non-condensing
2-1
2-1
2-5
Air Pressure
700 - 1060 hPa
NOTE:Each person in the room places an additional 300 BTU/hr demand on the cooling system.
2-2-1-2Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
Chapter 2 - Site-Preparation2-1
GEVOLUSON i / VOLUSON e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
2-2-2Electrical Requirements
GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound
equipment. This dedicated power shall originate at the last distribution panel before the system.
Sites with a main power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
Sites with a main power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
NOTE:Please note that image artifacts can occur, if at any time within the facility, the Ground from the main
facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-2-1Voluson i / Voluson e Power Requirements
The Voluson i / Voluson e is powered either by its internal battery or by a separate power supply adapter
unit connected to a separate outlet for any range of 100 - 250 VAC, 47-63 Hz.
Table 2-3 Electrical Specifications for Voluson i / Voluson e
AC DC converter100V AC to 250V AC20V±10%6A47-63 Hz
2-2-2-2Inrush Current
Inrush current is not a factor for consideration, due to the inrush current limiting properties of the power
supplies.
2-2-2-3Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without extension cords. Other adequate
outlets for the external peripherals, medical and test equipment needed to support this unit must also
be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state, and
national electrical codes.
2-2-2-4Main Power Plug
The Voluson i / Voluson e portable ultrasound system is supplied with a main power plug, as standard.
In the event that the unit arrives without a power plug, or with the wrong plug, contact your GE dealer.
When necessary, the installation engineer will supply the appropriate power plug to meet the applicable
local regulations.
2-2-2-5Power Stability Requirements
•Voltage drop-out
Max 10 ms.
•Power Transients
(All applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient, including
line frequency, synchronous, asynchronous, or aperiodic transients.
2-2Section 2-2 - General Console Requirements
GEVOLUSON i / VOLUSON e
NOTICE
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
2-2-3EMI Limitations
Ultrasound systems are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. Ultrasound machines also generate EMI. The
Voluson i / Voluson e complies with limits as stated on the EMC label. However, there is no guarantee
that interference will not occur in a particular installation.
Possible EMI sources should be identified before the unit is installed, and should not be on the same
!! NOTICE:
line as the ultrasound system. A dedicated line should be used for the ultrasound system.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect.
Sources of EMI include the following:
•medical lasers
•scanners
•cauterizing guns
•computers
•monitors
•fans
•gel warmers
•microwave ovens
•portable phones
•broadcast stations and mobile broadcasting machines
Table 2-4 EMI Prevention/ Abatement
EMI RuleDetails
Be aware of RF sources.
Ground the unit.
Replace and/or reassemble
all screws, RF gaskets,
covers and cores.
Replace broken RF gaskets.
Do not place labels where
RF gaskets touch metal.
Use GE-specified harnesses
and peripherals.
Take care with cellular
phones.
Keep the unit at least 5m (16.4 ft) away from other EMI sources. Special shielding may be
required to eliminate interference problems caused by high frequency, high powered radio
or video broadcast signals.
Poor grounding is the most likely reason a unit will have noisy images.
Check grounding of the power cord and power outlet.
After you finish repairing or updating the system, replace all covers and tighten all screws.
Any cable with an external connection requires a magnet wrap at each end.
Install the shield over the front of the card cage. Loose or missing covers or RF gaskets
allow radio frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket.
Do not turn ON the unit until any loose metallic part is removed and replaced, if required.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit
RF leakage. In case a label has been found in such a location, move the label to a different,
appropriate location.
The interconnect cables are grounded and require ferrite beads and other shielding.
Cable length, material, and routing are all important; do not make any changes that do not
meet all specifications.
Cellular phones may transmit a 5 V/m signal that causes image artifacts.
Properly address peripheral
cables.
Do not allow cables to lie across the top of the system.
Loop any peripheral cable excess length into one bundle.
Chapter 2 - Site-Preparation2-3
GEVOLUSON i / VOLUSON e
NOTICE
CAUTION
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
2-2-4Probe Environmental Requirements
Operation: Ambient temperature 18° to 30° C
Storage: -10° to 50° C
NOTE:Temperature in degrees C. Conversion to degrees F = °C (9/5) + 32).
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
!! NOTICE:
-10 TO + 50 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
2-2-5Time and Manpower Requirements
Site preparation takes time. Begin Site preparation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
Only one person is required to unpack the Voluson i / Voluson e ultrasound unit.
!! CAUTION:
At least two people must be available to unpack the Cart.
Attempts to move the Cart considerable distances (or on an incline) by one person alone, could
result in personal injury, and/or damage to the system.
2-2-6System Specifications
2-2-6-1Physical Dimensions of Voluson i / Voluson e
The physical dimensions of the Voluson i / Voluson e unit are summarized in Tab le 2- 5.
Table 2-5 Physical Dimensions of Voluson i / Voluson e
HeightWidthDepth
59 cm / 2.3 inches358 mm / 14.2 inches313 mm / 12.4 inches5 kg / 11 lbs.
2-2-6-2Acoustic Noise Output
max. 45 dB(A)
2-2-6-3Electrical Specifications
Please refer to Section 2-2-2-1 "Voluson i / Voluson e Power Requirements" on page 2-2.
Weight
(without peripherals)
2-4Section 2-2 - General Console Requirements
GEVOLUSON i / VOLUSON e
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
Section 2-3
Facility Needs
2-3-1Purchaser Responsibilities
The work and materials needed to prepare the site is the responsibility of the purchaser. Delay,
confusion, and waste of manpower can be avoided by completing pre installation work before delivery.
Use the Pre-installation checklist (provided in Ta bl e 2 -6 ) to verify that all needed steps have been taken.
Table 2-6 Voluson i / Voluson e Pre-Installation Check List
ActionYesNo
Schedule at least 3 hours for installation of the system.
Notify installation team of the existence of any variances from the basic installation.
Make sure system and probes have been subject to acclimation period.
Environmental cooling is sufficient.
Lighting is adjustable to adapt to varying operational conditions of the scanner.
Electrical facilities meet system requirements.
EMI precautions have been taken and all possible sources of interference have been
removed.
Mandatory site requirements have been met.
If a network is used, be sure to have all applicable Network information (IP addresses,
AE Title, Port Number, etc.) and a dedicated network outlet is available.
Purchaser responsibility includes:
•Procuring the materials required.
•Completing the preparations before delivery of the ultrasound system.
•Paying the costs for any alterations and modifications not specifically provided in the sales contract.
NOTE:All electrical installations that are preliminary to the positioning of the equipment at the site prepared for
the equipment must be performed by licensed electrical contractors. Other connections between pieces
of electrical equipment, calibrations, and testing must also be performed by qualified personnel. The
products involved (and the accompanying electrical installations) are highly sophisticated and special
engineering competence is required. All electrical work on these products must comply with the
requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified
personnel to perform electrical servicing on the equipment.
The desire to use a non–listed or customer provided product or to place an approved product further
from the system than the interface kit allows presents challenges to the installation team. To avoid
delays during installation, such variances should be made known to the individuals or group performing
the installation at the earliest possible date (preferably prior to the purchase).
The ultrasound suite must be clean prior to delivery of the machine. Carpet is not recommended
because it collects dust and creates static. Potential sources of EMI (electromagnetic interference)
should also be investigated before delivery. Dirt, static, and EMI can negatively impact system reliability.
Chapter 2 - Site-Preparation2-5
GEVOLUSON i / VOLUSON e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
2-3-2Mandatory Site Requirements
The following are mandatory site requirements. Additional (optional) recommendations, as well as a
recommended ultrasound room layout, are provided in Section 2-3-3 "Site Recommendations" .
•A dedicated single branch power outlet of adequate amperage (see Table 2-3 on page 2-2) that
meets all local and national codes and is located less than 2.5 m (8.2 ft) from the unit’s proposed
location. Refer to: Section 2-2-2 "Electrical Requirements" on page 2-2.
•A door opening of at least 76 cm (2.5 ft) in width (if using the Cart).
•The proposed location for the unit is at least 0.2 m (0.67 ft) from the walls, to enable cooling.
•Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m of the unit to connect cables.
•Power outlets for other medical equipment and gel warmer.
•Power outlets for test equipment within 1 m (3.3 ft) of the ultrasound unit.
•Clean and protected space for storage of probes (either in their case or on a rack).
•Material to safely clean probes (performed using a plastic container, never metal).
•In the case of a network option:
-An active network outlet in the vicinity of the ultrasound unit.
-A network cable of appropriate length (regular Pin-to-Pin network cable).
-An IT administrator who will assist in configuring the unit to work with your local network. A
fixed IP address is required. Refer to the form provided in Figure 3-141 on page 3-120 for
network details that are required.
NOTE:All relevant preliminary network port installations at the prepared site must be performed by a uthorized
contractors. The purchaser of GE equipment must utilize only qualified personnel to perform servicing
of the equipment.
2-3-3Site Recommendations
The following are (optional) site recommendations. Mandatory site requirements are provided in the
Mandatory Site Requirements section, above.
•Door opening of 92 cm (3 ft) in width (if using the Cart).
•Accessible circuit breaker for a dedicated power outlet.
•Sink with hot and cold running water.
•Receptacle for bio–hazardous waste, for example, used probe sheaths.
•Emergency oxygen supply.
•Storage area for linens and equipment.
•Nearby waiting room, lavatory, and dressing room.
•Dual level lighting (bright and dim).
•Lockable cabinet for software and manuals.
2-6Section 2-3 - Facility Needs
GEVOLUSON i / VOLUSON e
36 IN.
(92 CM)
Dedicated Power Outlets
Hospital Network
Cabinet for
Software and Manuals
(optional)
VOLUSON i
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
2-3-3-1Recommended Ultrasound Room Layout
Figure 2-1 shows a floor plan illustrating the recommended layout of the Ultrasound Room and
depicting the minimal room layout requirements.
Figure 2-1 Recommended Floor Plan 4.3m x 5.2m (14ft x 17ft)
Chapter 2 - Site-Preparation2-7
GEVOLUSON i / VOLUSON e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
2-3-4Networking Setup Requirements
2-3-4-1Stand-alone Unit (without Network Connection)
None
2-3-4-2Unit Connected to Hospital’s Network
Supported networks:
•Ethernet
•Wireless LAN
2-3-4-3Purpose of the DICOM® Network Function
DICOM® (Digital Imaging and Communications in Medicine) services provide the operator with
clinically useful features for moving images and patient information over a hospital network.
Examples of DICOM® services include the transfer of images to workstations for viewing or transferring
images to remote printers. As an added benefit, transferring images in this manner frees up the onboard monitor and peripherals, enabling viewing to be done while scanning continues.
With DICOM®, images can be archived, stored, and retrieved faster, easier, and at a lower cost.
To configure the Voluson i / Voluson e ultrasound unit to work with other network connections, the
network administrator must provide some necessary information.
Use the Connectivity Setup Worksheet on page 3-119 to record required information that must include:
• Voluson i / Voluson e Details:DICOM® network details for the Voluson i / Voluson e unit, incl.
the host name, local port, IP address, AE title and net mask.
• Routing Information:IP addresses for the default gateway and other routers in use at
the site.
• DICOM® Application
Information:
Installation see: Section 3-12 "Network IP Address Configuration" on page 3-116.
Details of DICOM® devices in use at the site, including the
DICOM® host name, AE title, DICOM® port number and IP
addresses.
2-8Section 2-3 - Facility Needs
GEVOLUSON i / VOLUSON e
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
Chapter 3
Setup Instructions
Section 3-1
Overview
3-1-1The Purpose of Chapter 3
This chapter contains information needed to setup the Voluson i / Voluson e ultrasound unit.
Included are procedures to receive, unpack and configure the equipment.
NOTE:Depending on the customer’s specific requirements, the Vo luson i / Voluson e may have been supplied
with or without the Voluson Modo Cart, Dock Cart or Voluson Station (optional).
A worksheet is provided (see: page 3 - 119 to page 3 - 120) to help ensure that all the required
information is available, prior to setup the system.
Table 3-1 Contents in Chapter 3
SectionDescriptionPage Number
3-1
3-2
3-3
3-4
3-5
3-6
3-7
3-8
3-9
3-10
3-11
3-12
3-13
3-14
Overview
Set Up Reminders
Receiving and Unpacking the Equipment
Preparing for Set Up
Connection of Auxiliary Devices
Completing the Set Up
Printer Installation
System Configuration
Available Probes
Software/Option Configuration
Connectivity Setup
Network IP Address Configuration
Connectivity Setup Worksheet
Paperwork
3-1
3-2
3-4
3-9
3-23
3-68
3-80
3-98
3-106
3-106
3-107
3-116
3-119
3-121
Chapter 3 - Setup Instructions3-1
GEVOLUSON i / VOLUSON e
CAUTION
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Section 3-2
Set Up Reminders
3-2-1Average Installation Time
The Voluson i / Voluson e installation and functional checkout will take approximately one hour;
Voluson i / Voluson e consoles with optional equipment may take slightly longer.
Once the site has been prepared, the average installation time required is shown in Ta bl e 3 -2 below.
Table 3-2 Average Installation Time
Average
Description
Unpacking the scanner0.5 hours
Installing the scanner / options / printers0.5 - 1.5 hoursdepends on required configuration
® Option (connectivity)
DICOM
Installation TimeComments
0.5 - 1.5 hoursdepends on configuration amount
3-2-2Installation Warnings
1.) The Voluson i / Voluson e portable ultrasound system weighs only 5 kg (11 lbs.), without
peripherals. However since the Carts (incl. Voluson i / Voluson e and peripherals) weighs
approximately 33 kg (72 lbs.), two persons are required to unpack it.
2.) There are no operator serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service
personnel should carry out servicing and troubleshooting.
NOTE:For information on shipping carton labels, refer to Figure 3-1 on page 3-3.
3-2-2-1System Acclimation Time
After being transported the Voluson i / Voluson e system may be very cold, or hot. It requires one hour
for each 2.5°C increment if it's temperature is below 10°C or above 40°C.
Equipment damage possibility. Turning the system on without acclimation after arriving at site
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
!! DANGER:
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT
TOUCH THE UNIT!
Two people are required to unpack the optional Cart because of its weight.
!! CAUTION:
!! CAUTION:
!! CAUTION:
!! CAUTION:
!! CAUTION:
Two people are required whenever a part weighing 16kg (35 lbs.) or more must be lifted.
If the unit is very cold or hot, do NOT turn on its power until it has had sufficient time to
acclimate to its operating environment.
To prevent electrical shock, connect the unit to a properly grounded power outlet.
Do NOT use a three to two prong adapter. This defeats safety grounding.
Do NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is
present.
Do not operate this unit unless all board covers and frame panels are securely in place, to
ensure optimal system performance and cooling.
(When covers are removed, EMI may be present).
!! CAUTION:
!! CAUTION:
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the
Voluson i / Voluson e.
Keep manuals near the unit for reference.
ACOUSTIC OUTPUT HAZARD
Although the ultrasound energy transmitted from the Voluson i / Voluson e portable
ultrasound system is within FDA limitations, avoid unnecessary exposure.
Ultrasound energy can produce heat and mechanical damage.
Figure 3-1 Environmental Labels
Chapter 3 - Setup Instructions3-3
GEVOLUSON i / VOLUSON e
CAUTION
CAUTION
NOTICE
NOTICE
Check whether delivery is complete (according to
packing list) and check visual damage!
The envelope with delivery address, packing list and
invoice is located on the front panel of the carton.
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Section 3-3
Receiving and Unpacking the Equipment
Please read this section carefully before unpacking the Voluson i / Voluson e portable
!! CAUTION:
!! CAUTION:
ultrasound system and its (optional) peripherals.
The Voluson i / Voluson e ultrasound system, together with peripherals, probes, cables and accessories
are shipped from the factory in a single durable shipping crate which is mounted on a raised wooden
platform base - see Figure 3-3 on page 3-5.
If the optional Modo Cart, Dock Cart or Voluson Station has been ordered, this is shipped in a separate
shipping crate.
Transport only with forklift or stacker truck.
During transportation pay attention to the point of gravity.
Only one person is required to unpack the Voluson i / Voluson e ultrasound unit.
At least two people must be available to unpack the Modo Cart, the Dock Cart or the Voluson
Station. Attempts to move the Cart considerable distances (or on an incline) by one person
alone, could result in personal injury, and/or damage to the system.
Move the Cart forward or backward when going up or down inclines. Do not move the Cart
sideways or diagonally.
Table 3-4 Shipping Cartons - Dimensions and Weights
DescriptionHeightWidthDepth
Voluson i / Voluson e
incl. peripherals and accessories
Voluson - Modo Cart only106 cm / 41.7 inch75 cm / 29.5 inch70 cm / 28 inch58.5 kg / 129 lbs
Voluson - Dock Cart only100 cm / 39,3 inch 63 cm / 24,8 inch60 cm / 24 inch59 kg / 130 lbs
Voluson Station only91 cm / 35,8 inch73 cm / 28,7 inch 69 cm / 27,1 inch61 kg / 134 lbs
* Weight is approximate and will vary depending upon the supplied peripherals
75 cm / 30 inch93 cm / 37 inch70 cm / 28 inch50 kg / 110 lbs
NOTE:Before unpacking the unit, inspect the crate for visible damage.
The device must only be transported in the original packaging!
!! NOTICE:
Each shipping crate is sealed with cross-head screws. A Phillips 2 screwdriver is needed to open the
crate. It is recommended to keep and store the shipping crate and all other packing materials (including
the support foams, anti-static plastic cover, etc.), in case the unit (and/or Cart) has to be moved to a
different location.
Unpack the devices such a way that packaging can be reused.
For warranty purposes, storage of the above is required for one year from date of purchase.
If the shipping crate is damaged, please inform the GE sales representative immediately.
!! NOTICE:
Weight*
Figure 3-2 envelope at front panel of the crate
3-4Section 3-3 - Receiving and Unpacking the Equipment
GEVOLUSON i / VOLUSON e
CAUTION
Wooden Base
top cover
of crate
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
Section 3-3 Receiving and Unpacking the Equipment (cont’d)
3-3-1Unpacking Procedure
1.) Loosen the screws and then remove the top cover from the crate.
!! CAUTION:
Figure 3-3 Shipping crate - Voluson i / Voluson e with Peripherals
When using sharp tools to open packing materials, take care to avoid cutting or damaging any
of the contents.
2.) Remove the packing slip (see: Figure 3-2) detailing the contents of the shipping crate and keep
close to hand, ready to check whether delivery is complete (according to packing list).
3.) Remove the surface packing material (bubble packs, foam sections, silica gel, etc.).
Continue to remove additional packing material as applicable during the following procedure steps.
4.) Carefully remove the boxes containing the Voluson i / Voluson e scanner, peripherals, probes, etc.
as well as the package containing further accessories (e.g. cables and software CD).
Figure 3-4 Shipping crate with Top open
5.) Before commencing the installation procedure, check visual damage!
Chapter 3 - Setup Instructions3-5
GEVOLUSON i / VOLUSON e
CAUTION
CAUTION
Wooden Base
Securing straps
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-3-2Unpacking the Voluson i / Voluson e Modo Cart
1.) Cut the straps securing the shipping carton to the wooden base.
When using sharp tools to open packing materials, take care to avoid cutting or damaging any
!! CAUTION:
of the contents.
Figure 3-5 Shipping carton - Voluson Modo Cart
2.) Carefully cut the adhesive tape used to close the shipping carton and open the carton.
3.) Remove the packing slip (shipping consignment note) detailing the contents of the shipping carton
and keep close to hand, ready to check whether delivery is complete (according to packing list).
4.) Remove the packing material in front of and surrounding the Modo Cart. Continue to remove
additional packaging material as applicable during the following procedure steps.
5.) Using two people, carefully roll the Modo Cart down the unloading ramp to remove it completely
from the shipping carton.
!! CAUTION:
Have two people available to move the Modo Cart
Attempts to move the unit considerable distances or on an incline by one person could
result in injury or damage, or both.
Make sure the Modo Cart is standing firmly on the floor before applying the locking brake.
6.) Before commencing the installation procedure, check visual damage!
3-6Section 3-3 - Receiving and Unpacking the Equipment
GEVOLUSON i / VOLUSON e
CAUTION
CAUTION
Wooden Base
Securing straps
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
3-3-3Unpacking the Voluson i / Voluson e Dock Cart
1.) Cut the straps securing the shipping carton to the wooden base.
When using sharp tools to open packing materials, take care to avoid cutting or damaging any
!! CAUTION:
of the contents.
!! CAUTION:
Figure 3-6 Shipping carton - Voluson Dock Cart
2.) Carefully cut the adhesive tape used to close the shipping carton and open the carton.
3.) Remove the packing slip (shipping consignment note) detailing the contents of the shipping carton
and keep close to hand, ready to check whether delivery is complete (according to packing list).
4.) Remove the packing material in front of and surrounding the Dock Cart. Continue to remove
additional packaging material as applicable during the following procedure steps.
5.) Using two people, carefully roll the Dock Cart down the unloading ramp to remove it completely from
the shipping carton.
Have two people available to move the Dock Cart
Attempts to move the unit considerable distances or on an incline by one person could
result in injury or damage, or both.
Make sure the Dock Cart is standing firmly on the floor before applying the locking brake.
6.) Before commencing the installation procedure, check visual damage!
Chapter 3 - Setup Instructions3-7
GEVOLUSON i / VOLUSON e
CAUTION
CAUTION
Wooden Base
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-3-4Unpacking the Voluson i / Voluson e Voluson Station
1.) Cut the straps securing the shipping carton to the wooden base.
When using sharp tools to open packing materials, take care to avoid cutting or damaging any
!! CAUTION:
of the contents.
!! CAUTION:
Figure 3-7 Shipping carton - Voluson Station
2.) Carefully cut the adhesive tape used to close the shipping carton and open the carton.
3.) Remove the packing slip (shipping consignment note) detailing the contents of the shipping carton
and keep close to hand, ready to check whether delivery is complete (according to packing list).
4.) Remove the packing material in front of and surrounding the Voluson Station. Continue to remove
additional packaging material as applicable during the following procedure steps.
5.) Using two people, carefully roll the Voluson Station down the unloading ramp to remove it
completely from the shipping carton.
Have two people available to move the Voluson Station
Attempts to move the unit considerable distances or on an incline by one person could
result in injury or damage, or both.
Make sure the Voluson Station is standing firmly on the floor before applying the locking brake.
6.) Before commencing the installation procedure, check visual damage!
3-8Section 3-3 - Receiving and Unpacking the Equipment
GEVOLUSON i / VOLUSON e
NOTICE
b
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
Section 3-4
Preparing for Set Up
3-4-1Verify Customer Order and Inspect Components
1.) After unpacking the equipment, it is important to verify that all items ordered by the customer have
!! NOTICE:
been received. Compare all items listed on the packing slip (delivery note) with those received.
It is recommended to keep and store the shipping carton and all other packing materials (including the
support foams, anti-static plastic cover, etc.), in case the unit has to be moved to a different location.
Unpack the devices such a way that packaging can be reused.
For warranty purposes, storage of the above is required for one year from date of purchase.
2.) Visually inspect the system components using the following checklist.
Table 3-5 Damage Inspection Checklist - Voluson i / Voluson e System
StepItemRecommended Procedure
1
Main labelEnter Serial Number: __________________________ (printed on main label, see: Figure 1-2 on page 1-12)
2
ConsoleVerify that the system is switched OFF and unplugged. Clean the console and control panel.
3
Control Console
4
Probes
5
LCD Display
6
FansVerify that the system’s cooling fans and peripheral fans are operating.
7
Rear Panel
8
Covers
9
Peripherals
10
AC DC SystemCheck the AC DC unit is not damaged or cracked board and verify that the output cable is properly secured.
11
Power Cord
Physically inspect the control console for missing or damaged items.
After switching on the system, verify the proper illumination of all the control panel buttons.
Check all probes for wear and tear on the lens, cable, and connector. Look for bent or damaged pins on the
connector and in the connector socket on the unit. Verify that the EMI fingers around the probe connector
socket housing are intact. Check the probe locking mechanism and probe switch.
Clean the LCD display by gently wiping with a dry, soft, lint-free non-abrasive folded cloth. Inspect the monitor
for scratches and raster burn.
Check the rear panel connectors for bent pins, loose connections and loose or missing hardware. Screw all
the cable connectors tightly to the connector sockets on the panel. Verify that the labeling is in good condition.
Check that all screws are tightly secured in place, that there are no dents or scratches and that no internal
parts are exposed.
Check and clean the peripherals in accordance with the manufacturer’s directions.
To prevent EMI or system overheating, dress the peripheral cables inside the peripheral cover.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any deterioration.
Verify continuity. Replace the power cord, as required.
NOTE:Report any items that are missing, back-ordered, or damaged, to your GE sales representative. The
contact address is shown in Contact Information on page 1-18.
Chapter 3 - Setup Instructions3-9
GEVOLUSON i / VOLUSON e
b
b
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-4-1Verify Customer Order and Inspect Components (cont’d)
•Optional: Visually inspect the Modo Cart components using the following checklist.
Table 3-6 Damage Inspection Checklist - Modo Cart (optional)
StepItemRecommended Procedure
1
Main labelEnter Serial Number: __________________________ (printed on main label)
2
Probe HoldersClean the gel wells with warm water and a damp cloth to remove all traces of gel.
3
CoversCheck that all screws are in place, all chassis and internal covers are installed.
4
Transformer
5
Power Cord
6
Front Castors
7
Rear CastorsCheck that the rear castors can roll and swivel.
Modo Cart
8
Voltage
Check the isolation transformer connectors and the associated cabling for good connection and proper
insulation. Japan only: Exchange pull out protection of main power cord.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any deterioration.
Verify continuity.
Check that the front castors can swivel, and can be placed in the locked position by pressing the foot
brake lever n each castor.
Verify that the voltage of the optional Modo Cart is correct.
see: Section 3-4-3 "Modo Cart Voltage Configuration" on page 3-20
NOTE:Report any items that are missing, back-ordered, or damaged, to your GE sales representative. The
contact address is shown in Contact Information on page 1-18.
•Optional: Visually inspect the Dock Cart components using the following checklist.
Table 3-7 Damage Inspection Checklist - Dock Cart (optional)
StepItemRecommended Procedure
1
Main labelEnter Serial Number: __________________________ (printed on main label)
2
Probe HoldersClean the gel wells with warm water and a damp cloth to remove all traces of gel.
3
CoversCheck that all screws are in place, all chassis and internal covers are installed.
4
Transformer
5
Power Cord
6
Front Castors
7
Rear CastorsCheck that the rear castors can roll and swivel.
Dock Cart
8
Voltage
Check the power supply connectors and the associated cabling for good connection and proper
insulation. Japan only: Exchange pull out protection of main power cord.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any deterioration.
Verify continuity.
Check that the front castors can swivel, and can be placed in the locked position by pressing the foot
brake lever n each castor.
Verify that the voltage of the optional Dock Cart is correct.
see: Section 3-4-4 "Dock Cart / Dock Cart light / Voluson Station Voltage Configuration" on page 3-21
NOTE:Report any items that are missing, back-ordered, or damaged, to your GE sales representative. The
contact address is shown in Contact Information on page 1-18.
3-10Section 3-4 - Preparing for Set Up
GEVOLUSON i / VOLUSON e
b
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
3-4-1Verify Customer Order and Inspect Components (cont’d)
•Optional: Visually inspect the Voluson Station components using the following checklist.
Table 3-8 Damage Inspection Checklist - Voluson Station (optional)
StepItemRecommended Procedure
1
Main labelEnter Serial Number: __________________________ (printed on main label)
2
Probe HoldersClean the gel wells with warm water and a damp cloth to remove all traces of gel.
3
CoversCheck that all screws are in place, all chassis and internal covers are installed.
4
Transformer
5
Power Cord
6
Front Castors
7
Rear CastorsCheck that the rear castors can roll and swivel.
Voluson Station
8
Voltage
Check the power supply connectors and the associated cabling for good connection and proper
insulation. Japan only: Exchange pull out protection of main power cord.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any deterioration.
Verify continuity.
Check that the front castors can swivel, and can be placed in the locked position by pressing the foot
brake lever n each castor.
Verify that the voltage of the optional Voluson Station is correct.
see: Section 3-4-4 "Dock Cart / Dock Cart light / Voluson Station Voltage Configuration" on page 3-21
NOTE:Report any items that are missing, back-ordered, or damaged, to your GE sales representative. The
contact address is shown in Contact Information on page 1-18.
Chapter 3 - Setup Instructions3-11
GEVOLUSON i / VOLUSON e
6
6
5
7
8
9
2
2
1
10
10
4
3
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-4-1-1Front and Side View of the Voluson i / Voluson e System
Figure 3-8 below shows the Voluson i / Voluson e portable ultrasound system components that are
visible from the front and side of the ultrasound unit.
Figure 3-8 Front and Side View of the Voluson i / Voluson e
keyboard and different buttons used to operate the Voluson i / Voluson e
4Alphanumeric Keyboard (upper row of keys are assigned to ultrasound-
specific functions)
5Trackball and Trackball buttons
6Air Vents
7Probe Port: one active probe connector
8Probe Locking Lever
9ON/OFF Switch
10Release Latches
3-12Section 3-4 - Preparing for Set Up
GEVOLUSON i / VOLUSON e
1
2
3
4
5
6
7
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
3-4-1-2Rear View of the Voluson i / Voluson e System
Figure 3-9 shows the view of the Voluson i / Voluson e ultrasound unit rear panel and external
peripheral/accessory connectors:
Figure 3-9 View of the Voluson i / Voluson e Rear Panel
1Two interchangeable USB ports (Printers, DVD+R/RW Drive and other
peripherals)
2Docking connector (connector for Voluson Station Docking Adapter)
3Port for DC IN (AC Adapter)
4SVGA Output (VCR option or CRT monitor option)
5LAN 10/100 Base-TX Ethernet network connector
6PCMCIA port for PC card (Wireless LAN)
(only available with GPC1-3)
7Ejection lever for PCMCIA device
Chapter 3 - Setup Instructions3-13
GEVOLUSON i / VOLUSON e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-4-1-3Front View of the Modo Cart
Figure 3-10 Modo Cart - Front View
1Changeable probe/gel holders
2Moving handle
3Footrest
4Upper peripheral storage shelf (e.g., for DVD+R/RW Drive)
5Middle peripheral storage shelf (e.g., for Color printer, DVD recorder, ...)
6Lower peripheral storage shelf (e.g., for B/W printer)
7Front Caster wheels
8Brakes
3-14Section 3-4 - Preparing for Set Up
GEVOLUSON i / VOLUSON e
1
2
3
4
5
6
7
8
9
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
3-4-1-4Rear View of the Modo Cart
Figure 3-11 Modo Cart - Rear View
1Cable hooks
2Power outlets for Peripherals
3Isolation Transformer Fuses for changing between 115V and 230V
4Mains Power Switch
5Power Cord + pull out protection mounted
6Screws for changing the height of the mounting platform
7Cable Management clips
8Voltage Output Selector switch (115V and 230V)
9Place for AC/DC Adapter of Voluson i / Voluson e
Chapter 3 - Setup Instructions3-15
GEVOLUSON i / VOLUSON e
33
1
23
4
6
5
7
9
8
9
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-4-1-5Front View of the Dock Cart / Dock Cart light
Figure 3-12 Dock Cart/Dock Cart light - Front View
1Tray Recorder
2USB Hub (not available on Dock Cart light)
3Changeable probe/gel holders
4Moving handle
5Upper peripheral storage shelf (e.g., for Color printer)
6Lower peripheral storage shelf (e.g., for B/W printer)
7Footrest
8Brakes
9Front Caster Wheels
3-16Section 3-4 - Preparing for Set Up
GEVOLUSON i / VOLUSON e
1
2
3
4
5
6
7
8
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
3-4-1-6Rear View of the Dock Cart / Dock Cart light
Figure 3-13 Dock Cart/Dock Cart light - Rear View
1Lan Isolation Box (not available on Dock Cart light)
2Moving Handles
3Release Lever (not available on Dock Cart light)
4Cable Guide
5Probe Mux Box (not available on Dock Cart light)
6Mains Power Switch
7Power Supply Fuses for changing between 115V and 230V
8Voltage Output Selector switch (115V and 230V)
Chapter 3 - Setup Instructions3-17
GEVOLUSON i / VOLUSON e
1
2
3
4
5
6
7
7
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-4-1-7Front View of the Voluson Station
Figure 3-14 Voluson Station - Front View
1Patient Monitor
2Changeable probe/gel holders
3optional Probe Mux Box
4Cover BW Printer and Console Bottom Shelf
5Footrest
6Brakes
7Front Caster Wheels
3-18Section 3-4 - Preparing for Set Up
GEVOLUSON i / VOLUSON e
1
2
3
4
5
6
7
8
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
3-4-1-8Rear View of the Voluson Station
Figure 3-15 Voluson Station - Rear View
1Release Lever
2Changeable probe/gel holders
3Additional Shelf Peripherals (e.g., for Color printer)
4Storage Tray Back
5Mains Power Switch
6Power Supply Fuses for changing between 115V and 230V
7Voltage Output Selector switch (115V and 230V)
8Back Caster Wheels
Chapter 3 - Setup Instructions3-19
GEVOLUSON i / VOLUSON e
CAUTION
CAUTION
Isolation Transformer
Voltage
Fuses
Output
Selector
Fuse holder
(e.g. 230V)
Fuse
Fuse
Fuse holder
(back)
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-4-2System Voltage Settings
NOTE:Not necessary at the Voluson i / Voluson e portable ultrasound system:
For details see: Table 2-3 "Electrical Specifications for Voluson i / Voluson e" on page 2-2.
3-4-3Modo Cart Voltage Configuration
NOTE:Only required when using the Voluson i / Voluson e with the optional Modo Cart.
Check Fuses and Voltage Output Selector setting located on the isolation transformer of the Modo Cart.
Before starting the system or peripherals connected to the Modo Cart, always make sure:
!! CAUTION:
• Rating of isolation transformer fuses is compatible. • Output Selector is in correct position.
The wrong fuses and position of the voltage output selector may cause major damage on
Voluson i / Voluson e and connected peripherals.
Rating of fuses depend on Voltage (refer to description on the isolation transformer box).
!! CAUTION:
If required switch Output Selector to correct position and exchange fuses of the isolation transformer.
a.) Open the fuse protection on the rear of the isolation transformer.
b.) Pull out the fuse holder and exchange the 2 fuses.
Figure 3-16 Change fuses of isolation transformer
c.) Push the fuse holder back into position (correct voltage should be shown in upper line,
e.g, 230V, see: Figure 3-16, above) and then close protection.
3-20Section 3-4 - Preparing for Set Up
GEVOLUSON i / VOLUSON e
CAUTION
CAUTION
Power Supply
Voltage
Fuses
Output
Selector
Fuse holder
(e.g. 230V)
Fuses
Fuse Holder
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
3-4-4Dock Cart / Dock Cart light / Voluson Station Voltage Configuration
NOTE:Only required when using the Voluson i / Voluson e with the optional Dock Cart, Dock Cart light or
Voluson Station.
Check Fuses and Voltage Output Selector setting located on the power supply of the Cart.
!! CAUTION:
!! CAUTION:
Before starting the system or peripherals connected to the Cart, always make sure:
• Rating of power supply fuses is compatible. • Output Selector is in correct position.
The wrong fuses and position of the voltage output selector may cause major damage on
Voluson i / Voluson e and connected peripherals.
Rating of fuses depend on Voltage (refer to description on the power supply).
If required switch Output Selector to correct position and exchange fuses of the power supply.
a.) Open the fuse protection on the rear of the power supply.
b.) Pull out the fuse holder and exchange the 2 fuses.
Figure 3-17 Change Main Power Voltage / Change fuses of power supply
c.) Push the fuse holder back into position (correct voltage should be shown in upper line,
e.g, 230V) and then close protection.
Chapter 3 - Setup Instructions3-21
GEVOLUSON i / VOLUSON e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-4-5EMI Protection
The Voluson i / Voluson e Ultrasound Unit has been designed to minimize the effects of Electo-Magnetic
Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system
from image artifacts caused by this interference. For this reason, it is imperative that all covers and
hardware are installed and secured before the unit is put into operation.
Ensure that the system is protected from electromagnetic interference (EMI), as follows:
•Operate system at least 15 feet away from equipment that emits strong electromagnetic radiation.
•Operate the system in an area enclosed by walls, floors and ceilings comprised of wood, plaster or
concrete, which help prevent EMI.
•Shield the system when operating it in the vicinity of radio broadcast equipment, if necessary.
•Do not operate mobile phones or other EMI emitting devices in the ultrasound room.
•Verify that all EMI rules listed in the following table are followed:
The Voluson i / Voluson e
environmentally qualified facilities, in terms of the prevention of radio wave interference. Operation of
the ultrasound unit
television sets situated near the medical equipment.
For further details and EMI Prevention/Abatement refer to Section 2-2-3 "EMI Limitations" on page 2-3.
in an inappropriate environment can cause electronic interference to radios and
ultrasound unit is approved for use in hospitals, clinics and other
3-22Section 3-4 - Preparing for Set Up
GEVOLUSON i / VOLUSON e
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
Section 3-5
Connection of Auxiliary Devices
Table 3-9 below outlines the Voluson i/e hardware installation procedures described in the sub-sections.
Table 3-9 Connection Procedures
Sub-sectionDescriptionPage Number
Connection and Usage of the optional Modo Cart3-25
3-5-1
3-5-2
3-5-3
3-5-4Connecting the Wireless Network Adapter/Interface3-43
3-5-5General Remarks and Hints when using external USB-Devices3-44
3-5-6Connecting the USB Hub3-45
3-5-7Connecting the DVD+R/RW Drive3-46
3-5-8Connecting the USB Flash Memory Stick3-49
3-5-9Connecting the external USB Hard disk (Handydrive)3-50
3-5-10Connecting the Black & White Printer3-51
3-5-11Connecting the Color Printer3-52
3-5-12Connecting the DeskJet Color Printer3-53
3-5-13Connecting the VCR (VideoCassette Recorder)3-55
3-5-14Connecting the DVD Recorder3-60
Attaching the Voluson i / Voluson e to the Modo Cart3-25
Mounting small Peripherals (e.g., Black & White Printer)3-27
Mounting large Peripherals (e.g., Color Printer)3-28
Connection and Usage of the optional Dock Cart3-29
Attaching the Voluson i / Voluson e to the Dock Cart3-29
Installation of the Probe Mux Box3-31
Installation of the Probe Shelf3-32
Installation of the Top Probe Holder3-32
Mounting small Peripherals (e.g., Black & White Printer)3-33
Mounting large Peripherals (e.g., Color Printer)3-34
Connection and Usage of the optional Voluson Station3-35
Attaching the Voluson i / Voluson e to the Voluson Station3-35
Installation Gel Condom Holder / Probe Holder for Endocavity Probe3-37
Installation of the Additional Shelf Peripherals3-38
Installation of the Console Bottom Shelf3-38
Installation of the Cover BW Printer Shelf3-38
Installation of the Patient Monitor3-39
Installation of the Probe Mux Box Voluson Station3-40
Mounting small Peripherals (e.g., Black & White Printer)3-41
Mounting large Peripherals (e.g., VCR)3-42
Chapter 3 - Setup Instructions3-23
GEVOLUSON i / VOLUSON e
CAUTION
WARNINGWARNING
WARNINGWARNING
CAUTION
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Power Supplies for additional equipment MUST comply with IEC 60601-1. Auxiliary equipment
!! CAUTION:
with direct main connection requires galvanic separation of the signal and/or control leads.
DO NOT attempt to use different peripherals and accessories other than approved and
!! WARNING:
provided by GE!
Any unauthorized peripherals may cause system failure or damage!
After each installation, the leakage currents have to be measured according to
!! WARNING:
!! CAUTION:
NOTE:For more detailed Safety Considerations when connecting auxiliary devices to the Voluson i / Voluson
IEC 60601-1 respectively UL 60601-1 (see: Section 10-7-2 on page 10-13).
Please observe that some printers may not be medical devices! If they are not medical devices,
they have to be located outside of the patient environment (acc. to IEC 60601-1 / UL 60601-1).
e system, please review: Chapter 1 - Auxiliary Devices Safety on page 1-10.
3-24Section 3-5 - Connection of Auxiliary Devices
GEVOLUSON i / VOLUSON e
WARNINGWARNING
WARNINGWARNING
CAUTION
NOTICE
CAUTION
front sliding guides
DVD Drive
(upper storage shelf)
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
3-5-1Connection and Usage of the optional Modo Cart
DO NOT CONNECT any devices to any AC power other than the on-board outlets from
!! WARNING:
!! WARNING:
!! CAUTION:
3-5-1-1Attaching the Voluson i / Voluson e to the Modo Cart
!! NOTICE:
the transformer, as this might increase leakage currents and sacrifice patient safety.
Before starting the Voluson i / Voluson e or Peripherals that are connected to the Modo
Cart, ALWAYS make sure that rating of the isolation transformer fuses is compatible
and the voltage output selector is in the correct position.
The wrong fuses and position of the voltage output selector may cause major damage
on the Voluson i / Voluson e system and connected peripherals.
Make sure that all peripherals are firmly secured in position using Velcro tapes or straps.
Place heavy and large devices on the middle storage shelf.
A battery or an empty battery shell should always be attached to the Voluson i / Voluson e system
before attaching it to the Modo Cart.
1.) Hold the Voluson i / Voluson e slightly tilted towards you and place it over the front sliding guides.
!! CAUTION:
Figure 3-18 Voluson i / Voluson e attached to the Modo Cart
It is not possible to fix the Voluson i / Voluson e on the mounting plate of the Modo Cart!
2.) Take a power extender cable and plug it into the AC main adapter of the Voluson i / Voluson e.
Figure 3-19 AC (main) Adapter connection
Chapter 3 - Setup Instructions3-25
GEVOLUSON i / VOLUSON e
protection plate
L-shaped circular
plug of power cable
connected to DC IN
Upper storage shelf
(for DVD+/R/RW Drive)
Middle storage shelf
(e.g., Color printer)
Lower storage shelf
(e.g., B&W printer)
Bottom shelf
Isolation Transformer,
AC/DC Adapter
on-board outlets for peripheals
main power cable + pull out protection
AC Adapter
Circuit breaker on Modo Cart
(power extender cables connected)
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-5-1-1Attaching the Voluson i / Voluson e to the Modo Cart (cont’d)
3.) Connect the L-shaped circular power cable plug of the AC Adapter to the power inlet plug DC IN on
the rear of the Voluson i / Voluson e unit (see: Figure 3-20 below).
Figure 3-20 main power cable connected to Voluson i / Voluson e´s power inlet (DC IN) connector
4.) Plug the other end of the power extender cable to an outlet of the transformer on the Modo Cart.
5.) Put the AC adapter beside the transformer into the shelf and fasten it (see: Figure 3-21 below).
6.) Place the DVD Drive on the upper peripheral shelf of the Modo Cart and fasten it.
7.) Take a power extender cable and plug it into the adapter of the DVD Drive; the other end connect
to an on-board outlet of the isolation transformer on the Modo Cart.
8.) Plug the power cable and the USB-cable into the DVD-drive; the other end of the USB cable plug
to an USB port on the rear of the Voluson i / Voluson e.
9.) Mount and connect other peripherals as described in Section 3-5-1-2 and Section 3-5-1-3.
10.)Connect the main power cable to the isolation transformer of the Modo Cart; the other end connect
to a hospital-grade main power outlet with the proper rated voltage.
Figure 3-21 Isolation Transformer of Modo Cart / Peripherals mounted and secured
11.)Press the Circuit breaker on the Modo Cart. To start up the system, press the On/Off button
(see: Figure 3-87) on the control panel of the Voluson i / Voluson e system (for ~3 seconds).
3-26Section 3-5 - Connection of Auxiliary Devices
GEVOLUSON i / VOLUSON e
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
3-5-1-2Mounting small Peripherals (e.g., Black & White Printer)
1.) Place the B&W Printer on the lower peripheral shelf of the Modo Cart.
2.) Take a Velcro tape and pull it under the shelf (left image at Figure 3-22 below).
Figure 3-22 pull Velcro tape under shelf and printer
3.) Pull up the Velcro tape on the left side. The other end of the tape place on the printer
(right image at Figure 3-22, above).
4.) Place the end of the Velcro tape on the top of the other end (left image at Figure 3-23 below).
Figure 3-23 fasten Velcro tape
NOTE:Make sure that the Velcro tape is firmly fastened. For reference see: Figure 3-21 on page 3-26.
5.) Connect the peripheral device and mount the pull out protection of the power cable.
Links to connection schemes can be found in Table 3-9, “Connection Procedures,” on page 3-23.
NOTE:A power extender cable from the power outlets of the M odo Carts isolation transform er should be used
to connect peripheral devices. Please connect the power extender cable to the peripherals first,
then connect the other end to the isolation transformer of the Modo Cart.
(Cables can be routed inside the cable management clips, see: Figure 3-11 on page 3-15.)
Connect B/W printer
or Color printer
Connect isolation
General using cable
transformer
Figure 3-24 Connect peripheral device to the Modo Cart (isolation transformer)
Chapter 3 - Setup Instructions3-27
GEVOLUSON i / VOLUSON e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-5-1-3Mounting large Peripherals (e.g., Color Printer)
1.) Place the Color Printer, VCR or DVD recorder on the middle peripheral shelf of the Modo Cart.
2.) Take a Velcro tape and pull it under the shelf (left image at Figure 3-25 below).
Figure 3-25 pull Velcro tape under shelf and printer
3.) Pull up the Velcro tape on the front side. The other end of the tape place on the printer
(right image at Figure 3-25, above).
4.) Place the end of the Velcro tape on the top of the other end (left image at Figure 3-26 below).
Figure 3-26 fasten Velcro tape
NOTE:Make sure that the Velcro tape is firmly fastened. For reference see: Figure 3-21 on page 3-26.
5.) Connect the peripheral device and mount the pull out protection of the power cable.
Links to connection schemes can be found in Table 3-9, “Connection Procedures,” on page 3-23.
NOTE:A power extender cable from the power outlets of the Modo Car ts isolation transformer sh ould be used
to connect peripheral devices. Please connect the power extender cable to the peripherals first,
then connect the other end to the isolation transformer of the Modo Cart.
(Cables can be routed inside the cable management clips, see: Figure 3-11 on page 3-15.)
Connect B/W printer
or Color printer
Connect isolation
General using cable
transformer
Figure 3-27 Connect peripheral device to the Modo Cart (isolation transformer)
3-28Section 3-5 - Connection of Auxiliary Devices
GEVOLUSON i / VOLUSON e
WARNINGWARNING
WARNINGWARNING
CAUTION
NOTICE
probe connector
locking lever
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
D
3-5-2Connection and Usage of the optional Dock Cart
DO NOT CONNECT any devices to any AC power other than the on-board outlets from
!! WARNING:
!! WARNING:
!! CAUTION:
3-5-2-1Attaching the Voluson i / Voluson e to the Dock Cart
!! NOTICE:
the transformer, as this might increase leakage currents and sacrifice patient safety.
Before starting the Voluson i / Voluson e or Peripherals that are connected to the Dock
Cart, ALWAYS make sure that rating of the power supply fuses is compatible and the
voltage output selector is in the correct position.
The wrong fuses and position of the voltage output selector may cause major damage
on the Voluson i / Voluson e system and connected peripherals.
Make sure that all peripherals are firmly secured in position using Velcro tapes or straps.
Place heavy and large devices on the lower storage shelf.
A battery or an empty battery shell should always be attached to the Voluson i / Voluson e system
before attaching it to the Dock Cart.
1.) Hold the Voluson i / Voluson e slightly tilted towards you and place it on the top shelf next to
the arm rest.
2.) Now push the Voluson i / Voluson e away from the armrest until it locks in place.
Figure 3-28 Voluson i / Voluson e attached to the Dock Cart
3.) Push the probe connector into the probe socket of the Voluson i / Voluson e and lift the
locking lever upwards to lock the probe connector.The Probe Box enables the connection of up to
3 probes to the Voluson i / Voluson e.
4.) Connect the cables to the rear of the Voluson i / Voluson e. The cables to connect are listed
below. The order of connection is not important:
•L-Shaped circular plug of the power cable
•USB Cable - Connect the USB plug to the lower USB connector on the rear-panel of the
Voluson i / Voluson e. To connect USB Devices to the system, plug them into the Dock Cart’s
USB ports.
•LAN cable - Connect the ethernet cable, routing from the LAN isolation box, into the network port
of the Voluson i / Voluson e.
5.) Place the DVD Drive on the Tray recorder (see Figure 3-10 on page 3-14) and connect it.
Chapter 3 - Setup Instructions3-29
GEVOLUSON i / VOLUSON e
Tray recorder for
DVD+/R/RW Drive
USB port
Upper storage shelf
(e.g., B&W printer)
Lower storage shelf
(e.g., Color printer)
Power Supply
Circuit breaker on Dock Cart
main power cable
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
3-5-2-1Attaching the Voluson i / Voluson e to the Dock Cart (cont’d)
6.) Plug the power cable and the USB-cable into the DVD-drive; the other end of the USB cable plug
to an USB port on the Voluson i / Voluson e.
7.) Mount and connect other peripherals as described in Section 3-5-2-5 and Section 3-5-2-6.
8.) Connect the main power cable to the Power Supply of the Dock Cart; the other end connect to a
hospital-grade main power outlet with the proper rated voltage.
Figure 3-29 Peripherals mounted and secured / Power Supply of Dock Cart / USB-port on Dock Cart
9.) Press the Circuit breaker on the Dock Cart. To start up the system, press the On/Off button
(see: Figure 3-87) on the control panel of the Voluson i / Voluson e system (for ~3 seconds).
3-30Section 3-5 - Connection of Auxiliary Devices
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.