GE VeriDri User Guide

Applications
VeriDri is a dew point temperature or PPMv moisture loop-powered transmit­ter intended for OEM applications such as:
• Desiccant regenerative air or gas dryers
•Drying application for plastics prior to injection/blow molding
•Dryers for medical air
•Glove boxes
Features
•Low-cost
• Thin-film aluminum oxide moisture sensor technology
• Microprocessor-based, all-digital technology for reliable operation
• Compact Size
• NEMA 4X/IP67
• Loop-powered 4 to 20 mA output
• -110° to +40°C dew point tempera­ture/0 to 10,000 PPMv moisture (with constant pressure) overall range capability
• NIST-traceable calibration
For OEM Applications
The GE VeriDri moisture transmitter provides dryer manufacturers and other OEMs with the means to make simple, accurate and reliable moisture measurement of their gas at a low price.
VeriDri is a loop-powered transmitter with a range proportional to 4 to 20 mA. It is offered in a number of factory-configured output ranges cov­ering the overall moisture range of
-110° to +40°C dew point temperature. It can also cover a number of PPMv moisture output ranges from the over­all range of 0 to 10,000 PPMv moisture if it is intended to be operated at a constant pressure. This information must be provided to the factory at the time of order placement.
Improve Product Efficiency and Performance
VeriDri can be used to significantly improve the energy and efficiency of regenerative gas dryer systems. By monitoring the dryer outlet moisture content, desiccant bed regeneration can be initiated on demand, rather
than on a timed cycle. This process has been adopted by many major dryer manufacturers, and has resulted in improved performance and signifi­cant savings in energy costs. In addition, VeriDri can be used to moni­tor overall dryer operations, determine regular maintenance inter­vals and detect faulty operation.
Advanced Moisture Sensor Technology
The VeriDri combines a technological­ly advanced aluminum oxide moisture sensor with state-of-the-art software and electronics for unequaled overall performance.
NIST-Traceable Calibration
Superior sensitivity, speed of response, calibration stability, and wide dynamic range have made GE aluminum oxide moisture probes the standard of per­formance and value in industrial moisture measurement. They are suit­able for laboratory and industrial moisture measurement applications in gases and non-aqueous liquids over a wide range of process conditions.
VeriDri
TM
Moisture Transmitter
© 2004 Panametrics, Inc. All rights reserved. 920-056
GE Panametrics GE Thermometrics GE NovaSensor GE Druck GE Kaye GE General Eastern GE Ruska Now under one name: GE Infrastructure Sensing
GE Infrastructure
Sensing
GE Panametrics has joined other GE high-technology sensing businesses under a new name —
GE Infrastructure
Sensing
All moisture probe calibrations are traceable to the National Institute of Standards and Technology (NIST).
Installation Flexibility for OEM Applications
The rugged, compact VeriDri trans­mitter is designed specifically for OEM installations where space is at a premi­um. It can be installed directly in the process stream or, if necessary, in a sample system. If required, GE can design and build a sample condition­ing system to meet the unique application demands.
The VeriDri transmitter is available with OEM pricing for a minimum quantity of 10 units. Discounts on larger quantities are also available, and blanket orders are accepted.
Specifications
Moisture Ranges
• –110° to 20°C
• –110° to –50°C
• –80° to 20°C
• –80° to –30°C
• –30° to 20°C
• –60° to 40°C
• –150° to 70°F
• –150° to –40°F
• –40° to 70°F
• –100° to 0°F
• –50° to 50°F
•0°to 100°F
•0 to 10 PPMv
•0 to 100 PPMv
•0 to 1,000 PPMv
•0 to 10,000 PPMv
Note: PPMv ranges based on constant pressure, provided at time of order placement.
Operating Temperature
–40° to 60°C (–40° to 140°F)
Storage Temperature
70°C (158°F maximum)
War m-Up Time
Meets specified accuracy in three minutes
Accuracy
• ±2°C from –65° to 40°C dew/frost point
• ±3°C from –80° to –66°C dew/frost point
Repeatability
•±0.5°C from –65° to 40°C dew/frost point
•±1.0°C from –80° to –66°C frost point
Response Time
Less than five seconds for 63% of a step change in moisture content in either wet-up or dry-down cycle
Electrical
Power
•7 to 28 VDC (loop-powered, cus­tomer supplied)
• Output: 4 to 20 mA
• Output Resolution:
0.01 mA
• Max Loop R = 50x(PSV-7) ohms (Where: PSV = Power Supply Voltage)
Example: Given a 24 VDC Power Supply, Max Loop R = 50 x (24 - 7)
= 850
• Cable: 2 m., standard (consult factory for custom lengths)
Mechanical
Sample Connection
3
/
4-in. 16 straight male thread with
O-ring
Operating Pressure
5 µ Hg to 5,000 psig (345 bar)
Enclosure
NEMA 4X/IP67
Dimensions
• Overall: 6.76 × 1.13 in. diameter
• Electronics with cable: 4.08 × 1.13 in. diameter
•Weight: 5 oz. (140 grams)
European Compliance
Complies with EMC Directive 89/336/ EEC and PED 97/23/EC for DN<25
Moisture Sensor
Sensor Type
Thin-film aluminum oxide moisture sensor probe
Calibration
Each sensor is individually computer calibrated against known moisture concentrations, traceable to NIST
Calibration Interval
Sensor recalibration at GE Infrastructure Sensing is recommended every six to 12 months depending on application
Flow Rate
• Gases: Static to 10,000- cm/s linear velocity at 1 atm
• Liquids: Static to 10- cm/s linear velocity at density of 1 g/cc
GE Infrastructure
Sensing
T 800 833 9438 • 978 437 1000 F 978 437 1031 sensing@ge.com www.gesensing.com
All specifications are subject to change for product improvement without notice. Panametrics®is a registered trademark of Panametrics, Inc. GE®is a registered trademark of General Electric Co. Windows
®
is a registered trademark of Microsoft Corporation, which is not affiliated with GE, in the U.S. and other countries. Other company or product names mentioned in this document may be trademarks or registered trademarks of their respective companies, which are not affiliated with GE.
VeriDri transmitter dimensions in inches (cm)
3/4-16 UNF-2A
2.68
(6.81)
1.13
(2.87)
2.58
(6.55)
1.50 (3.81)
Minimal Cable
Bend Allowance
Power and
Measurement
Circuits
+ 7-28VDC
4-20mA _
“Loop-Powered”
Blue
Brown
VeriDri wiring diagram
Loading...