Preface
Information Paragraphs
• Note paragraphs provide information that provides a deeper understanding of
the situation, but is not essential to the proper completion of the instructions.
• Important paragraphs provide information that emphasizes instructions that
are essential to proper setup of the equipment. Failure to follow these
instructions carefully may cause unreliable performance.
• Caution! paragraphs provide information that alerts the operator to a hazardous
situation that can cause damage to property or equipment.
• Warning! paragraphs provide information that alerts the operator to a
hazardous situation that can cause injury to personnel. Cautionary information
is also included, when applicable.
Safety Issues
WARNING! It is the responsibility of the user to make sure all local, county,
state and national codes, regulations, rules and laws related to
safety and safe operating conditions are met for each
installation.
Auxiliary Equipment
Local Safety Standards
The user must make sure that he operates all auxiliary equipment in accordance
with local codes, standards, regulations, or laws applicable to safety.
Working Area
WARNING! Auxiliary equipment may have both manual and automatic
modes of operation. As equipment can move suddenly and
without warning, do not enter the work cell of this equipment
during automatic operation, and do not enter the work
envelope of this equipment during manual operation. If you do,
serious injury can result.
WARNING! Make sure that power to the auxiliary equipment is turned OFF
and locked out before you perform maintenance procedures on
the equipment.
Moisture Target Series 6 Hygrometer User’s Manualiii
Preface
Qualification of Personnel
Make sure that all personnel have manufacturer-approved training applicable to the
auxiliary equipment.
Personal Safety Equipment
Make sure that operators and maintenance personnel have all safety equipment
applicable to the auxiliary equipment. Examples include safety glasses, protective
headgear, safety shoes, etc.
Unauthorized Operation
Make sure that unauthorized personnel cannot gain access to the operation of the
equipment.
Environmental Compliance
Waste Electrical and Electronic Equipment (WEEE) Directive
GE Measurement & Control is an active participant in Europe’s Waste Electrical
and Electronic Equipment (WEEE) take-back initiative, directive
2002/96/EC.
The equipment that you bought has required the extraction and use of natural
resources for its production. It may contain hazardous substances that could impact
health and the environment.
In order to avoid the dissemination of those substances in our environment and to
diminish the pressure on the natural resources, we encourage you to use the
appropriate take-back systems. Those systems will reuse or recycle most of the
materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please
contact your local or regional waste administration.
Visit http://www.ge-mcs.com/en/about-us/environmental-health-and-safety/1741-weee-req.html for take-back instructions and more information
about this initiative.
ivMoisture Target Series 6 Hygrometer User’s Manual
Moisture Target Series 6 Hygrometer User’s Manualvii
Contents
viiiMoisture Target Series 6 Hygrometer User’s Manual
Chapter 1. Features and Capabilities
Chapter 1. Features and Capabilities
1.1 Introduction
The Moisture Target Series 6 (MTS6) is a microprocessor-based,
single-channel hygrometer that measures moisture content in gases. It is
intended for Original Equipment Manufacturer (
suitable for a wide range of process conditions that require real-time
moisture measurement.
The MTS6 accepts any calibration range provided with GE probes (see
Chapter 5, Specifications for more information). It comes equipped with
two standard alarm relays, one fault alarm relay, and a single analog output.
It also has onboard data logging capability using an micro SD card.
1.2 Electronics Unit
The MTS6 displays measurement data on a liquid crystal display (LCD). You
can program your unit and enter probe information using the keys on the
front panel (see Figure 1). The MTS6 accepts line voltages of a universal
power supply from 100 to 240 VAC, or 24 VDC, depending on what is
ordered.
OEM) applications, and is
Figure 1: Front Panel
Moisture Target Series 6 Hygrometer User’s Manual1
Chapter 1. Features and Capabilities
1.3 Probes
The moisture probe is the part of the system that comes in direct contact
with the process. The MTS6 uses any GE M Series (see Figure 2) or a
VeriDri probe (see Figure 3) to measure dew point temperature in °C or °F.
The sensor assembly is secured to the probe mount and is protected with a
sintered stainless steel shield (see Figure 2).
Note: Other types of shields are available upon request.
Figure 2: M Series Probe
Figure 3: VeriDri Probe
2Moisture Target Series 6 Hygrometer User’s Manual
Chapter 2. Installation
Chapter 2. Installation
2.1 Introduction
Installing the MTS6 includes the following steps:
•selecting the recorder output
•mounting the electronics unit
•mounting the sample system
•installing the probe into the sample system
•wiring the input power
•wiring the probe and alarm connections
WARNING! To ensure safe operation, the MTS6 must be panel
mounted and operated as described in this manual. Also,
be sure to follow all applicable local safety codes and
regulations for installing electrical equipment.
Note: For the non-enclosure MTS6 package, please refer to Appendix D for
mounting and installation.
Moisture Target Series 6 Hygrometer User’s Manual3
Chapter 2. Installation
2.2 Selecting the Recorder Output
Note: By default, the recorder is set to the current output.
Note: The customer will provide their own cable for connecting the
recorder. Acceptable cables range from 16 to 26AWG.
The MTS6 has one isolated analog recorder output. The recorder output
provides either a current or voltage signal, which is set by switch S1 on the
main PC board.
Complete these steps to check or reset switch
WARNING! Never connect line voltage or any other power input to
S1 (see Figure 8 on page 7).
the recorder output terminals.
1.Make sure the MTS6 is turned off and unplugged.
WARNING! The MTS must be isolated or disconnected from all voltage
sources before changing the recorder output.
2.Remove the screw at the top of the back panel (see Figure 4).
Figure 4: Back Panel
4Moisture Target Series 6 Hygrometer User’s Manual
Chapter 2. Installation
2.2 Selecting the Recorder Output (cont.)
3.Lift the back edge of the cover (see Figure 5), slide the cover back (see
Figure 6), and lift it from the enclosure (see Figure 7 on page 6).
Figure 5: Lifting the Back Edge of the Cover
Figure 6: Sliding the Cover Back
Moisture Target Series 6 Hygrometer User’s Manual5
Chapter 2. Installation
2.2 Selecting the Recorder Output (cont.)
Figure 7: Lifting the Cover
6Moisture Target Series 6 Hygrometer User’s Manual
Chapter 2. Installation
2.2 Selecting the Recorder Output (cont.)
4.Locate switch S1 (see Figure 8, highlighted area).
CAUTION! Use proper ESD grounding prior to changing the
switch.
5.Set switch S1 to the desired position: V for voltage or I for current.
Figure 8: Switch S1 on the Main PC Board
6. After setting the switch, replace the cover and reinsert the rear
enclosure screw.
Moisture Target Series 6 Hygrometer User’s Manual7
Chapter 2. Installation
2.3 Mounting the Electronics Unit
The MTS6 unit can be installed in a panel up to 0.25 in. (6 mm) thick. See
Appendix A, Outline and Installation Drawings, for the required panel
cutout dimensions.
IMPORTANT: For NEMA 4 and IP66 installation, the MTS6 must be mounted
in a rigid, flat panel using the panel gasket and both mounting
brackets provided.
2.3.1Basic Mounting
To mount the MTS6 in a panel with a basic 94 mm (3.69”) x 46 mm (1.81”)
opening, refer to the following figures and complete the following steps:
1.Remove the side panel mount label prior to installation.
Figure 9: Removing Side Panel Mount Label
8Moisture Target Series 6 Hygrometer User’s Manual
Chapter 2. Installation
2.3.1 Basic Mounting (cont.)
2.Slide the small gasket along the MTS6 and place it around the back of
the display (see Figure 10).
Figure 10: Installing the Gasket Behind the Display
3.Slide the MTS6 into the panel cutout (see Figure 11).
Figure 11: Sliding the MTS6 into the Panel Cutout
Moisture Target Series 6 Hygrometer User’s Manual9
Chapter 2. Installation
2.3.1 Basic Mounting (cont.)
4.Behind the panel, insert the mounting brackets into the side holes
provided (see Figure 12).
Figure 12: Installing the Mounting Brackets
5.Hold the chassis and lock each mounting bracket in place by sliding it
toward the rear of the MTS6 (see Figure 13).
Figure 13: Locking the Mounting Brackets in Place
10Moisture Target Series 6 Hygrometer User’s Manual
Chapter 2. Installation
2.3.1 Basic Mounting (cont.)
6.Use a screwdriver to extend the bracket screws to the back of the panel
and secure the MTS6 in the panel cutout (see Figure 14).
Figure 14: Securing the MTS6 to the Panel
7.Using a feeler gauge behind the gasket, check the compression, and
tighten the bracket screws until the gap is 0.028” ±0.002” (see
Figure 15).
Figure 15: Checking the Gasket Compression
Moisture Target Series 6 Hygrometer User’s Manual11
Chapter 2. Installation
2.3.2Adapter Plate Mounting
Some customers may need to retrofit an MTS6 into a cut-out sized for GE’s
previous OEM panel-mount hygrometers: MTS5, MTS4 or PanaDry
models. The previous generations were of a larger size requiring a 5.40”
(137.2 mm) W x 2.65” (67.3 mm) H cut-out. The MTS6 is smaller, requiring
a 3.69” (94 mm) W x 1.81” (46 mm) H cut-out. For those customers who
need to retrofit the larger panel cut-out sizes, GE offers an optional adapter
plate. See Appendix A, Outline and Installation Drawings, for the required
panel cutout.
1.Fit the larger gasket around the adapter plate (see Figure 16).
Figure 16: Installing the Adapter Plate Gasket
12Moisture Target Series 6 Hygrometer User’s Manual
Chapter 2. Installation
2.3.2 Adapter Plate Mounting (cont.)
2.Fit the adapter plate into the panel cutout (see Figure 17).
Figure 17: Inserting the Adapter Plate
3.Behind the panel, place the metal backing plate over the four adapter
plate mounting screws (see Figure 18).
Figure 18: Applying the Backing Plate
Moisture Target Series 6 Hygrometer User’s Manual13
Chapter 2. Installation
2.3.2 Adapter Plate Mounting (cont.)
4. Apply nuts to the four screws and secure the assembly to the panel (see
Figure 19 and Figure 20). Use a feeler gauge behind the gasket, check
the compression, and tighten the nuts until the gap is 0.032” ±0.002”.
Figure 19: Securing the Assembly to the Panel
Figure 20: Plate Assembly Mounting Complete
14Moisture Target Series 6 Hygrometer User’s Manual
Chapter 2. Installation
2.3.2 Adapter Plate Mounting (cont.)
Now mount the MTS6 using steps 1-6 in Basic Mounting on page 8.
Re-verify the adapter plate gap after the MTS6 is mounted to the adapter
plate. The installation should appear similar to Figure 21 and Figure 22.
Figure 21: MTS6 Installation with Adapter Plate - Rear
Figure 22: MTS6 Installation with Adapter Plate - Front
Moisture Target Series 6 Hygrometer User’s Manual15
Chapter 2. Installation
2.4 Mounting the Sample System
Figure 23: Typical Sample System
16Moisture Target Series 6 Hygrometer User’s Manual
Chapter 2. Installation
2.4 Mounting the Sample System (cont.)
The sample system is normally fastened to a flat metal plate that has four
mounting holes. Upon request, GE can also provide the sample system in an
enclosure.
Complete the following steps to mount the sample system:
1.Fasten the sample system plate or enclosure to a vertical wall or panel
with a bolt in each of the four corners.
2.Connect the sample system inlet to the process and the outlet to the
return, using appropriate stainless steel fittings and tubing.
CAUTION! Do not start the process flow through the system
until the probe has been properly installed (see the
following section).
2.5 Installing the Probe
GE probes are usually installed in a sample system to protect the probe from
any damaging elements in the process. The probe is mounted in a
cylindrical container called the
your sample system.
sample cell, which is included as part of
Standard M2 probes and VeriDri probes are mounted into the sample system
or process line with 3/4-16 straight threads that are sealed with an o-ring.
Other fittings are available for special applications.
CAUTION! If the probe is to be mounted directly in the
process line, without a sample system, consult GE for proper
installation instructions and precautions.
Refer to Figure 24 on page 18, and complete these steps to install the probe
into the sample cell:
1.Insert the probe into the sample cell and thread the probe into the
sample cell fitting. Make sure you do not to cross the threads.
2.Tighten the probe securely.
Moisture Target Series 6 Hygrometer User’s Manual17
Chapter 2. Installation
2.5 Installing the Probe (cont.)
3.Identify the sample cell inlet port as the connection that is perpendicular
to the installed probe.
CAUTION! For maximum protection of the aluminum oxide
sensor, the probe shield should always be left in place.
Figure 24: Probe/Sample Cell Assembly
2.6 Wiring the System
Wiring the MTS6 system includes the following steps:
•connecting the probe
•connecting the recorder output
•connecting the alarms
•installing the power cable
WARNING! To ensure safe operation, the MTS6 must be installed and
operated as described in this manual. Also, be sure to
follow all applicable local safety codes and regulations for
installing electrical equipment.
18Moisture Target Series 6 Hygrometer User’s Manual
2.6 Wiring the System (cont.)
The CAUTION! symbol is a reminder that MTS6 components can
be damaged if electrical connections are not correctly made.
Figure 25: MTS6 Rear Panel Connections - AC Version
Chapter 2. Installation
Figure 26: MTS6 Rear Panel Connections - DC Version
Moisture Target Series 6 Hygrometer User’s Manual19
Chapter 2. Installation
2.6 Wiring the System (cont.)
Figure 27: MTS6 Wiring Diagram
20Moisture Target Series 6 Hygrometer User’s Manual