GE CRx25P CR, Digital Cassettes, CRxFlex CR, DXR250V, DXR250P Brochure

...
Industrial Radiography
Image forming techniques
Issued by GE Inspection Technologies
©
2006 General Electric Company. All Rights Reserved. We reserve the right to technical modifications without prior notice. GEIT-30158EN (06/08)
GE Inspection Technologies
Inspection Technologies
GE Inspection Technologies
www.geinspectiontechnologies.com/en
Industrial Radiography Image forming techniques
1
Digital radiography CR-image of a weld
see acknowledgements*
The first issue of “Industrial Radiography” was published by Agfa in the sixties, for educational and promotional purposes. Some improved editions have been released since, providing information on the latest image forming radiographic techniques. The booklet has been published in a number of languages and has been very much in demand. The latest edition was compiled in the seventies but is now obsolete, because of the large number of computer-aided NDT techniques which have entered the market in recent years.
In 2007 a new edition in the English language was published by GE Inspection Technologies. That edition was compiled by Mr. J.A. de Raad, NDT-expert and consultant, who has a considerable number of publications on the subject of Non-Destructive Testing to his name. Mr. A. Kuiper, an experienced specialist and tutor on industrial radiography, assisted him. Both had been involved in Non-Destructive Testing during their professional careers at Applus RTD NDT & Inspection headquartered in Rotterdam, the Netherlands.
Apart from the developments in conventional radiography, primarily regarding X-ray equipment and films, the 2007 issue describes the now mature methods of digital radiography using radiation-sensitive plate- and panel detectors, including digitisation of traditional film. Several other computer-assisted methods such as the CT technique are also included as well as a separate chapter describing a variety of applications.
In this latest 2008 edition we considerably extended the chapters on digital radiography and special techniques, such as microfocus and X-ray microscopy. In addition, the impact and (non) existence of norms, codes and standards on new NDT-technologies and their applications are addressed.
We trust that this new issue of “Industrial Radiography” will fulfil a need once again.
GE Inspection Technologies, 2008
The author expresses his appreciation to all employed by GE Inspection Technologies and Applus RTD NDT & Inspection who cooperated and provided ample information to update this new edition.
3
Introduction to the overview of “Industrial Radiography”
Image forming techniques
2
Introduction to the overview of “Industrial Radiography” Image forming techniques Preface 13
1. Introduction to industrial radiography 15
2. Basic properties of ionising radiation 19
2.1 Wavelengths of electromagnetic radiation 19
2.2 X-rays 20
2.3 Gamma-rays (γ-rays) 21
2.4 Main properties of X-rays and γ-rays 22
2.5 Radiation energy-hardness 22
2.6 Absorption and scattering 23 Photoelectric effect Compton effect Pair production Total absorption/attenuation
2.7 Penetrating power 25
2.8 Filtering (hardening) 26
2.9 Half-value thickness 26
3. Units and definitions 29
3.1 Units 29
3.2 Definitions 30 Radioactivity Ionisation dose rate Ionisation dose Absorbed energy dose Equivalent dose (man dose)
4. Radiation sources 33
4.1 X-ray tube 33
4.2 The anode 33 Cooling the anode The focal spot Effective focal spot size
4.3 Tube voltage and tube current 35
4.4 Radioactive sources (isotopes) 36 Natural radioactive sources
4
Contents
Artificial radioactive sources
4.5 Advantages and disadvantages of artificial radioactive sources 36
4.6 Properties of radioactive sources 37 Activity (source strength) Specific activity Specific gamma-ray emission (k-factor) Half-life of a radioactive source
5. NDT equipment 39
5.1 X-ray equipment 39 Types of X-ray tubes Bipolar X-ray tubes Unipolar X-ray tubes Special types of X-ray tubes
5.2 High voltage generators 41
5.3 Megavolt equipment 42 The Bètatron The linear accelerator (linac)
5.4 Radioactive sources 45 Average energy level (nominal value)
5.5 Source holders (capsules) 46
5.6 Transport- and exposure containers 46
5.7 Checking for container leakage 49
6. Radiation images, filters and intensifying screens 51
6.1 Radiation images 51
6.2 Radiation filters 53
6.3 Intensifying screens 53 Lead screens Steel and copper screens Fluorescent screens Fluorescent salt screens Fluorometallic screens
7. The X-ray film and its properties 59
7.1 Structure of the X-ray film 59
7.2 Radiographic image 59 Latent image Developing the latent image
7.3 Characteristics of the X-ray film 60 Density (optical) Contrast
7.4 Characteristic curve (density curve) 61
5
7
Stopbath Fixing Final wash Drying in the drying cabinet Roller dryers
10.3 Recommendations for the darkroom 90
10.4 Silver recovery 90
10.5 Automatic film processing 91 NDT-U (universal) film processor NDT-E (economy) film processor
10.6 Checking the developing process and film archiving properties 93 PMC-strips to check the developing process Thiosulphate-test to check the film archival properties
10.7 Storage of exposed films 95
11. Defect discernibility and image quality 97
11.1 Unsharpness 97 Geometric unsharpness Inherent (film) unsharpness Total unsharpness
11.2 Selection source-to-film distance 101
11.3 Other considerations with regard to the source-to-film distance 102 Inverse square law Selection of radiation energy (kV) Selection of gamma source
11.4 Radiation hardness and film contrast 104
11.5 Summary of factors that influence the image quality 105
12. Defect orientation, image distortion and useful film length 107
12.1 Defect detectability and image distortion 107
12.2 Useful film length 108
13. Image quality 111
13.1 Factors influencing image quality 111
13.2 Image quality indicators (IQI) 112 Wire type IQI according to EN 462-1
13.3 List of common IQI’s 114 ASTM IQI’s AFNOR IQI’s Duplex IQI’s
13.4 Position of the IQI 117
13.5 IQI sensitivity values 117
14. Film exposure and handling errors 119
6
Gradient of the density curve Average gradient Effect of developing conditions on the density curve
7.5 Film speed (sensitivity) 65
7.6 Graininess 65
8. Film types and storage of films 67
8.1 The Agfa assortment of film types 67
8.2 Film type selection 70
8.3 Film sizes 70
8.4 Handling and storage of unexposed films 70
9. Exposure chart 73
9.1 Exposure chart parameters 73 Type of X-ray equipment The radioactive source Source-to-film distance Intensifying screens Type of film Density Developing process
9.2 Densitometer 75
9.3 Producing an exposure chart for X-rays 75
9.4 The exposure chart 78
9.5 Use of the exposure chart 78
9.6 Relative exposure factors 80
9.7 Absolute exposure times 80
9.8 Use of the characteristic (density) curve with an exposure chart 81
10. Processing and storage of X-ray films 85
10.1 The darkroom 85 Entrance and colour Darkroom lighting Darkroom layout Tanks
10.2 Chemicals and film-development 86 Making-up processing solutions Developer Fixer Developing time and bath temperatures Film agitation Replenishing
9
16.7 Resolution number of bits 163 Bit depth Lateral resolution
16.8 Comparison of film, CR- and DR methods 164
16.9 Impact and status of CR- and DR standards 165 Development of standards Status of CR standards Status of DR standards Impact of standards Standards for weld inspection Data exchange and tamper proof standard
16.10 Selection of CR- and DR methods 167
16.11 Applications for CR- and DR methods 168 Corrosion detection Weld inspection Dose reduction and controlled area Automated/mechanised inspection Girth weld inspection Useful life of plate and panel
16.12 Work station 171 Hardware and software Versatility of the software Archiving and reliability of images Exchange of data
17. Special radiographic techniques 17 7
17.1 Image magnification techniques 177
17.1.1 Common image magnification technique 177
17.1.2 High resolution X-ray microscopy 178 Magnification factors Microfocus and nanofocus X-ray tubes Tube heads System set-up Effect of focal dimensions Imaging systems for high resolution radiography
17.2 Fluoroscopy, real-time image intensifiers 182 Stationary real-time installations Portable real-time equipment
17.3 Computer Tomography (CT) 185 Unique features Computing capacity and scanning time Reverse engineering CT metrology High resolution and defect sizing
8
15. Film interpretation and reference radiographs 123
15.1 Film interpretation 123
15.2 The film-interpreter 124
15.3 Reference radiographs 124 Weld inspection Casting radiography Examination of assembled objects
16. Digital Radiography (DR) 14 5
16.1 Introduction to DR 145
16.2 Digital image formation 146
16.3 Digitisation of traditional radiographs 146
16.4 Computed Radiography (CR) 148 Two-step digital radiography The CR imaging plate Image development Scanners-Readers CR cassettes Dynamic range-Exposure latitude Exposure time and noise Fading Optimisation Improvements
16.5 Genuine Digital Radiography (DR) 153
One-step digital radiography
16.5.1 Detector types 153 Direct versus indirect detection Linear detectors 2D detectors
16.5.2 Fill Factor 155
16.5.3 Flat panel and flat bed detector systems 156 Amorphous silicon flat panels CMOS detectors and flat bed scanners Limitations
16.6 Image quality and exposure energy 158
16.6.1 Exposure energy 158
16.6.2 Determination of image quality 159
16.6.3 Indicators of image quality- MTF and DQE 160 Factors influencing image quality Image quality definitions Exposure parameters MTF (Modulation Transfer Function) DQE (Defective Quantum Efficiency) Noise, image averaging and DQE
11
20.
Standards, literature / references, acknowledgements and appendices
215
European norms (EN-standards) Literature and references Acknowledgements Appendices: tables and graphs.
10
17.4 CT for defect detection and sizing 187 Effect of defect orientation 3D CT for sizing of defects in (welded) components
17.5 Neutron radiography (neutrography) 190
17.6 Compton backscatter technique 190
18. Special radiographic applications 193
18.1 Measuring the effective focal spot 193
18.2 Radiographs of objects of varying wall thickness 194
18.3 Radiography of welds in small diameter pipes 195 Elliptical technique Perpendicular technique
18.4 Determining the depth position of a defect 197
18.5 Determining the depth position and diameter of
reinforcement steel in concrete 198
18.6 On-stream inspection - profiling technique 198 Projection technique Tangential technique Selection of source, screens and filters Exposure time
18.7 Flash radiography 201
18.8 Radiography of welds in large diameter pipes 202
19. Radiation hazards, measuring- and recording instruments 207
19.1 The effects of radiation on the human body 179 207
19.2 Responsibilities 207 The client The radiographer
19.3 The effects of exposure to radiation 208
19.4 Protection against radiation 208
19.5 Permissible cumulative dose of radiation 209
19.6 Radiation measurement and recording instruments 210 Radiation measuring instruments Dose rate meters Scintillation counter Personal protection equipment Pendosismeter (PDM) Thermoluminescent dose meter (TLD-badge) Film dose meter (film-badge)
19.7 Dose registration 212
19.8 Radiation shielding 212 Distance Absorbing barrier and distance
Preface
13
To verify the quality of a product, samples are taken for examination or a non-destructive test (NDT) is carried out. In particular with fabricated (welded) assemblies, where a high degree of constructional skill is needed, it is necessary that non-destructive testing is carried out.
Most NDT systems are designed to reveal defects, after which a decision is made as to whether the defect is significant from the point of view of operational safety and/or reliability. Acceptance criteria for weld defects in new constructions have been specified in standards.
However, NDT is also used for purposes such as the checking of assembled parts, the development of manufacturing processes, the detection of corrosion or other forms of deterioration during maintenance inspections of process installations and in research.
There are many methods of NDT, but only a few of these allow the full examination of a component. Most only reveal surface-breaking defects.
One of the longest established and widely used NDT methods for volumetric examination is radiography: the use of X-rays or gamma-rays to produce a radiographic image of an object showing differences in thickness, defects (internal and surface), changes in structure, assembly details etc. Presently, a wide range of industrial radiographic equip­ment, image forming techniques and examination methods are available. Skill and expe­rience are needed to select the most appropriate method for a particular application. The ultimate choice will be based on various factors such as the location of the object to be examined, the size and manoeuvrability of the NDT equipment, the existance of standards and procedures, the image quality required, the time available for inspection and last but not least financial considerations.
This book gives an overview to conventional industrial radiography, as well as digital (computer-aided) techniques and indicates the factors which need to be considered for selection of the most suitable system and procedures to be followed.
At the end of the book a chapter is added describing aspects of radiation safety.
12
1514
1
Introduction to industrial radiography Image forming techniques
In industrial radiography, the usual procedure for producing a radiograph is to have a source of penetrating (ionising) radiation (X-rays or gamma-rays) on one side of the object to be examined and a detector of the radiation (the film) on the other side as shown in figure 1-1. The energy level of the radiation must be well chosen so that sufficient radi­ation is transmitted through the object onto the detector.
The detector is usually a sheet of photographic film, held in a light-tight envelope or cas­sette having a very thin front surface that allows the X-rays to pass through easily. Chemicals are needed to develop the image on film, which is why this process is called the classic or “ wet” process.
Nowadays, different kinds of radiation-sensitive films and detectors not requiring the use of chemicals to produce images, the so-called “dry” process, are used increasingly. These techniques make use of computers, hence the expressions; digital or computer aided radiography (CR) or genuine (true) digital radiography (DR), see chapter 16.
A DR related technique that has been available for many decades is the one in which ima­ges are formed directly with the aid of (once computerless) radiation detectors in combi­nation with monitor screens (visual display units: VDU’s), see chapter 17. This is in fact is an early version of DR. These through transmission scanning techniques (known as fluoroscopy) the storage of images and image enhancement are continually improved by the gradual implementation of computer technology. Nowadays, there is no longer a clear division between conven­tional fluoroscopy with the aid of computers and the entirely computer-aided DR. In time DR will, to some extent, replace conventional fluoroscopy.
Summarising, the image of radiation intensities transmitted through the component can be recorded on:
The conventional X-ray film with chemical development, the “ wet” process, or one of the following “dry” processes:
• A film with memory phosphors and a work station for digital radiography, called computer-assisted radiography or CR.
• Flat panel and flat bed detectors and a computer work station for direct radiography, called DR.
• A phosphorescent or fluorescent screen (or similar radiation sensitive medium) and a closed-circuit television (CCTV) camera as in conventional fluoroscopy, an early version of direct radiography.
Fig. 1-1. Basic set-up for film radiography
screens X-ray film
projection of defect on film
cavity
object
homogeneous
radiation
source
1716
• By means of radiation detectors, e.g.: crystals, photodiodes or semiconductors in a linear array by which in a series of measurements an image is built up of a moving object. This method is applied in systems for luggage checks on airports.
The source of radiation should be physically small (a few millimetres in diameter), and as X-rays travel in straight lines from the source through the specimen to the film, a sharp “image” is formed of the specimen and discontinuities. This geometric image formation is identical to the shadow image with a visible light source. The sharpness of the image depends, in the same way, on the radiation source diameter and its distance away from the surface on which the image is formed.
The “classic” film in its light-tight cassette (plastic or paper) is usually placed close behind the specimen and the X-rays are switched on for an appropriate time (the exposure time) after which the film is taken away and processed photographically, i.e. developed, fixed, washed and dried. In direct radiography (DR), a coherent image is formed directly by means of an computerised developing station. The two methods have a negative image in common. Areas where less material (less absorption) allows more X-rays to be transmitted to the film or detector will cause incre­ased density. Although there is a difference how the images are formed, the interpreta­tion of the images can be done in exactly the same way. As a result, the DR- technique is readily accepted.
The “classic” film can be viewed after photochemical treatment (wet process) on a film viewing screen. Defects or irregularities in the object cause variations in film density (brightness or transparency). The parts of the films which have received more radiation during exposure – the regions under cavities, for example – appear darker, that is, the film density is higher. Digital radiography gives the same shades of black and white images, but viewing and interpretation is done on a computer screen (VDU).
The quality of the image on the film can be assessed by three factors, namely :
1. Contrast
2. Sharpness
3. Graininess
As an example, consider a specimen having a series of grooves of different depths machi­ned in the surface. The density difference between the image of a groove and the back­ground density on the radiograph is called the image contrast. A certain minimum image contrast is required for the groove to become discernible.
With increased contrast:
a. the image of a groove becomes more easily visible b. the image of shallower grooves will gradually also become discernible
Assuming the grooves have sharp-machined edges, the images of the grooves could still be either sharp or blurred; this is the second factor: image blurring, called image unsharpness.
At the limits of image detection it can be shown that contrast and unsharpness are inter­related and detectability depends on both factors.
As an image on a photographic film is made up of grains of silver, it has a grainy appea­rance, dependent on the size and distribution of these silver particles. This granular appe­arance of the image, called film graininess, can also mask fine details in the image.
Similarly, in all other image forming systems these three factors are fundamental para­meters. In electronic image formation, e.g. digital radiography or scanning systems with CCTV and screens, the factors contrast, sharpness and noise are a measure for the image quality; pixel size and noise being the (electronic) equivalent of graininess .
The three factors: contrast, sharpness and graininess or noise are the fundamental para­meters that determine the radiographic image quality. Much of the technique in making a satisfactory radiograph is related to them and they have an effect on the detectability of defects in a specimen.
The ability of a radiograph to show detail in the image is called “radiographic sensitivity”. If very small defects can be shown, the radiographic image is said to have a high (good) sensitivity. Usually this sensitivity is measured with artificial “defects” such as wires or drilled holes. These image quality indicators (IQIs) are described in chapter 13.
2
Basic properties of ionising radiation
19
In 1895 the physicist Wilhelm Conrad Röntgen discovered a new kind of radiation, which he called X-rays. The rays were generated when high energy electrons were suddenly stopped by striking a metal target inside a vacuum tube – the X-ray tube. It was subsequently shown that X-rays are an electromagnetic radiation, just like light, heat and radiowaves.
2.1 Wavelengths of electromagnetic radiation
The wavelength lambda (λ) of electromagnetic radiation is expressed in m, cm, mm, micrometer (μm), nanometer (nm) and Ångstrom (1 Å = 0.1 nm).
18
Electromagnetic radiation Wavelength λ m Type
10 km 10
4
1 km 10
3
100 m 10
2
10 m 10
1
1 m 1 10 cm 10
-1
1 cm 10
-2
1 mm 10
-3
100 μm 10
-4
10 μm 10
-5
1 μm 10
-6
100 nm 10
-7
10 nm 10
-8
1 nm 10
-9
0.1 nm 10
-10
0.01 nm 10
-11
1 pm 10
-12
0.1 pm 10
-13
0.01 pm 10
-14
X-ray energy 100 eV 1 keV 10 keV 100 keV 1 MeV 10 MeV 100 MeV
Visible light and Ultraviolet (UV)
X-rays and Gamma-rays (Radiography)
Heat-rays, Infra-red rays, microwaves
Table 1-2. Overview of wavelength, energy and type of electromagnetic radiation
2.2 X-rays
The radiation which is emitted by an X-ray tube is heterogeneous, that is, it contains X-rays of a number of wavelengths, in the form of a continuous spectrum with some superimposed spectrum lines. See fig. 1-2.
The shortest wavelength of the spectrum is given by the Duane-Hunt formula:
In which :
λ = wavelength in nanometers (10
-9
m)
kV = voltage in kilovolts
The average shape of the X-ray spectrum is generally the same however not truely identical for different X-ray sets; it depends chiefly on the energy range of the electrons striking the X-ray tube target and, therefore, on the voltage waveform of the high-voltage generator. A constant potential (CP) X-ray set will not have the same spectrum as a self-rectified set operating at the same nominal kV and current. The spectrum shape also depends on the inherent filtration in the X-ray tube window (glass, aluminium, steel or beryllium).
The energy imparted to an electron having a charge e, accelerated by an electrical poten­tial V is (eV) so the energy of the electrons can be quoted in eV, keV, MeV. These same units are used to denote the energy of an X-ray spectrum line.
The energy of a single wavelength is :
In which: E = the energy in electronVolt (eV) h = Planck’s constant v = frequency
c = the velocity of electromagnetic radiation, such as light (300,000 km/s)
The heterogeneous X-rays emitted by an X-ray tube do not however have a single wavelength, but a spectrum, so it would be misleading to describe the X-rays as (say) 120 keV X-rays. By convention therefore, the ‘e’- in keV- is omitted and the X-rays described as 120 kV, which is the peak value of the spectrum.
21
2.3 Gamma-rays (γ-rays)
Radioactivity is the characteristic of certain elements to emit alpha (α), beta (β) or gamma (γ) rays or a combination thereof. Alpha and beta rays consist of electrically char­ged particles, whereas gamma rays are of an electromagnetic nature.
Gamma rays arise from the disintegration of atomic nuclei within some radioactive sub­stances, also known as isotopes. The energy of gamma-radiation cannot be controlled; it depends upon the nature of the radioactive substance. Nor is it possible to control its intensity, since it is impossible to alter the rate of disintegration of a radioactive substance.
Unlike X-rays, generated to a continuous spectrum, Gamma-rays are emitted in an isola­ted line spectrum, i.e. with one or more discrete energies of different intensities.
Figure 2-2 shows the energy spectrum lines for Selenium75, Cobalt60 and Iridium192. In practical NDT applications, sources (radio active isotopes) are allocated an average nominal energy value for calculation purposes, see section 5.4. Spectrum components with the highest energy levels (keV values) influence radiographic quality the most.
20
λ
min
=
1.234
Ε=h.v λ.v=c
kV
Fig. 1-2. X-ray spectrum – intensity/wavelength distribution The small peaks are the characteristic radiation of the target material
wavelength
intensity
relative intensity
energy (keV)
Fig. 2-2. Energy spectrum (lines) for Se75, Ir192 and Co60
2.4 Main properties of X-rays and γ-rays
X-rays and γ-rays have the following properties in common:
1. invisibility; they cannot be perceived by the senses
2. they travel in straight lines and at the speed of light
3. they cannot be deflected by means of a lens or prism, although their path can be bent (diffracted) by a crystalline grid
4. they can pass through matter and are partly absorbed in transmission
5. they are ionising, that is, they liberate electrons in matter
6. they can impair or destroy living cells
2.5 Radiation energy-hardness
Radiation hardness (beam quality) depends on wavelength. Radiation is called hard when its wavelength is small and soft when its wavelength is long. In industry the quality of the X-ray tube ranges from very soft to ultra hard. The beam quality is related to a tube voltage (kV) range, or keV for isotopes.
The first two columns of table 2-2 below indicate the relationship hardness/tube voltage range applied in NDT. The third column gives the related qualification of the radiation effect, i.e. half-value thickness (HVT), described in detail in section 2.9.
Table 2-2. Comparative values of radiation quality (hardness) against tube voltage.
2.6 Absorption and scattering
The reduction in radiation intensity on penetrating a material is determined by the following reactions :
1. Photoelectric effect
2. Compton effect
3. Pair production
Which of these reactions will predominate depends on the energy of the incident radiation and the material irradiated.
Photoelectric effect
When X-rays of relatively low energy pass through a material and a photon collides with an atom of this material, the total energy of this photon can be used to eject an electron from the inner shells of the atom, as figure 3-2 illustrates. This phenomenon is called the photoelectric effect and occurs in the object, in the film and in any filters used.
Compton effect
With higher X-ray energies (100 keV to 10 MeV), the interaction of photons with free or weakly bonded electrons of the outer atom layers causes part of the energy to be transferred to these electrons which are then
ejected,
as illustrated in figure 4-2.
At the same time the photons will be deflected from the initial angle of incidence and emerge from the collision as radiation of redu­ced energy, scattered in all direc­tions including backward, known as “backscatter”, see section 17.6. In this energy band, the absorp­tion of radiation is mainly due to the Compton effect and less so to the photoelectric effect.
22
Radiation quality Tube voltage Global half-value
Hardness thickness for steel (mm)
Very soft Less than 20 kV Soft 20 – 60 kV Fairly soft 60 – 150 kV 0.5-2 Hard 150 – 300 kV 2-7 Very hard 300 – 3000 kV 7-20 Ultra hard more than 3000 kV > 20
incident
X-ray s
ejected
electron
Fig. 3-2. Photoelectric effect
X-ray
100keV - 10 MeV
ejected
electron
scattered
radiation
Fig. 4-2. Compton effect
23
2.7 Penetrating power
The penetrating power of X-radiation increases with the energy (hardness). The relationship of energy and penetrating power is complex as a result of the various mechanisms that cause radiation absorption. When monochromatic ( homogeneous ­single wave length) radiation with an intensity I
o
passes through matter, the relative
intensity reduction ΔI/I
o
is proportional to the thickness Δt. The total linear absorption
coefficient (μ) consisting of the three components described in section 2.6 is defined by the following formula:
Expressed differently:
In which: Io= intensity at material entry t = thickness I = intensity at material exit e = logarithm: 2.718 μ = total absorption coefficient
Figure 7-2 shows the resulting radiation intensity (logarithmic) as a function of increased material thickness, for soft and hard homogeneous radiation.
When radiation is heterogeneous the graphs are not straight, see figure 7-2, but slightly curved as in figure 8-2.
The slope of the curves becomes gradually shallower (because of selective absorption of the softer radiation) until it reaches the so-called “point-of-homogeneity”. After passing this point the coefficient of absorption remains virtually unchanged, as if the radiation had become homogeneous.
The position of the point of homogeneity varies with the nature of the material irradi­ated. The graph shows that with increasing material thickness, softer radiation is filte­red out, more than hard radiation. This effect is called “hardening”.
2524
Pair production
The formation of ion pairs, see figure 5-2, only occurs at very high energy levels (above 1 MeV). High-energy photons can cause an interaction with the nucleus of the atom involved in the colli­sion. The energy of the photon is here converted tot an electron(e-) and a positron (e+).
Total absorption/attenuation
The total linear absorption or attenuation of X-rays is a combi­nation of the three absorption processes described above, in which the primary X-ray energy changes to a lower form of ener­gy. Secondary X-ray energy arri­ses of a different wavelength and a different direction of travel. Some of this secondary (scatte­red) radiation does not contribu­te to radiographic image forming and may cause loss of image quality through blurring or fog. The contribution of the various causes of X-ray absorption to the total linear absorption coeffi­cient (μ) for steel plotted against radiation energy, are shown in figure 6-2.
X-ray
> 1 MeV
ejected
electron
ejected
positron
Fig. 5-2. Pair production
Fig. 6-2 Absorption coefficient for steel plotted against radiation energy
PE = Photoelectric effect C = Compton effect PP = Pair production
= μ.Δt
I = I
o
.
e
-μt
ΔI
I
o
hard radiation, high tube voltage
hard radiation soft radiation
points of homogeneity
soft radiation, low tube voltage
penetrated material thickness
intensityintesity
penetrated material thickness
Fig. 7-2. Intensity of homogeneous radiation as function of increasing thickness
Fig. 8-2. Intensity of heterogeneous radiation as function of increasing thickness hard radiation
Table 2-2 shows the average HVT-values for steel, table 3-2 shows the values for lead.
For a heterogeneous beam the HVT is not constant; the second HVT is slightly larger than the first. In general, in industry where relatively hard radiation is used, a fixed “average” HVT is applied.
2726
2.8 Filtering (hardening)
All materials, for example a metal layer between the radiation source and the film, cause absorption and filtering. The position of the metal layer plays an important role in the effect it has. A metal layer in front of the object will “harden” the radiation because it fil­ters out the soft radiation. The degree of hardening depends on the type and the thickness of the material. This phenomenon is used to reduce excessive contrast (variation in den­sity) when examining objects of which the thickness varies greatly.
A metal layer between the object and the film filters the soft scattered radiation that occurs in the object, thereby increasing the contrast and consequently the image quality. This method of filtering is for example applied in the use of Cobalt60 in combination with exposure time reducing intensifying screens, which are sensitive to scattered radiation. Lead, copper and steel are suitable filtering materials.
2.9 Half-value thickness
A convenient practical notion (number) of the linear absorption coefficient is the intro­duction of the half-value thickness (HVT). It quantifies the penetrating power of radia­tion for a particular type of material and is defined as the thickness of a particular mate­rial necessary to reduce the intensity of a monochromatic beam of radiation by half, as shown in figure 9-2. This HVT-value depends on the hardness of radiation.
Table 3-2. Half-value thickness for lead
Fig. 9-2. Illustration of half-value thickness
thickness
HVT HVT HVT HVT
intensity
Element/Isotope Symbol Average energy level Half-value thickness
in MeV in mm lead
Ceasium137 Cs137 0.66 8.4 Cobalt60 Co60 1.25 13 Iridium192 Ir192 0.45 2.8 Selenium75 Se75 0.32 2 Ytterbium169 Yb169 0.2 1 Thulium170 Tm170 0.072 0.6
3.1
Units
Until 1978 the “International Commission of Radiation Units and Measurements” (ICRU) used the conventional radiation units of roentgen (R), rad (rd), and curie (Ci). Since 1978 the ICRU has recommended the use of the international system units (SI) with special new units for radiation quantities; the Becquerel, Gray and Sievert.
Table 1-3 shows the relationships of these new units to the older units.
In radiography and radiation safety, units are preceded by prefixes. Table 2-3 shows the ones mostly used.
3
Units and definitions
2928
Table 1-3. Overview of new and old units * disintegrations per second C = Coulomb = A.s J = Joule ** RBE = Relative Biological Effect A = Ampère
Table 2-3. Prefixes
SI –units Formerly used Conversion
Designation of quantity Name Unit Name Unit Old to SI
Designation Designation
Activity (A) Becquerel 1/s* Curie Ci 1 Ci = 37 GBq
(Bq)
Ionisation dose Coulomb (C) C/kg Röntgen R 1 R=2.58 x 10
-4
C/kg
Ionisation dose rate Coulomb (C) C/kg.s R/s
Ampère (A) or A/kg Absorbed energy Gray J/kg Rad Rad 1 Rad = 0.01 Gy dose (D) (Gy) Equivalent dose (H) Sievert J/kg Rem Rem 1 Rem = 0.01 Sv H=D x RBE** (Sv)
Prefix Meaning Value Notation
p pico 0.000000000001 10
-12
n nano 0.000000001 10
-9
μ micro 0.000001 10
-6
m milli 0.001 10
-3
- 11 1
k kilo 1000 10
3
M Mega 1000000 10
6
G Giga 1000000000 10
9
3130
3.2 Definitions
Radioactivity
The activity of a radioactive source of radiation (isotope) is equal to the number of disinte­grations per second. The SI-unit is the Becquerel (Bq) and is equal to 1 disintegration per second. The Becquerel is too small a unit to be used in industrial radiography. Source strengths are, therefore, quoted in Giga Becquerel (GBq). 1 Curie = 37 GBq, see table 2-3.
Ionisation dose rate
The output of an X-ray set or isotope per unit of time is generally quoted at one metre distance from the source, and designated in C/kg, see table 1-3.
Ionisation dose
The ionising effect of radiation in one kilogram of dry air is used to define the ionisation dose. The dose of radiation delivered is equal to the ionisation dose rate multiplied by the amount of time during which radiation takes place. The designation used is C/kg.sec.
The output of an X-ray set, however, is quoted in Sievert per hour, measured at 1 metre distance.
Absorbed energy dose
The radiation energy that is absorbed is expressed in Joules per kilogram (J/kg). The SI-unit is called Gray (Gy) whereby 1 J/kg = 1 Gy.
Equivalent dose (man dose)
The Sievert (Sv) is the SI-unit for the biological effect of ionising radiation upon man. It corresponds with the product of the absorbed energy dose gray (Gy) with a factor that has been experimentally determined and that indicates the relative biological effect (RBE) of the ionising radiation. For X- and
γ-radiation this factor is equal to one, so that the Sievert
is equal to the Gray.
4
Radiation sources
33
4.1 X-Ray tube
The X-ray tube, see figure 1-4, consists of a glass (or ceramic) envelope containing a positive electrode (the anode) and a negative electrode (the cathode) evacuated to an ultra high
vacuum [10
- 9
hPa (hectoPascal)]. The cathode comprises a filament that generates electrons. Under the effect of the electri­cal tension set up between the anode and the cathode (the tube voltage) the electrons from the cathode are attracted to the anode, which accelerates their speed. This stream of electrons is concentrated into a beam by a “cylinder” or “focusing cup”. When the accelerated electrons collide with a target on the anode, part of their energy is converted to X-radiation, know as X-rays.
4.2 The anode
The target is generally made of tungsten. Not only because it has a high atomic number, but also because of its high melting point (approx. 3400˚C). It is essential to use a material with a high melting point because of the substantial amount of heat dissipated as the electron­“bombardment” is concentrated (focused) on a very small surface. Only a part (approx. 0.1 % at 30 keV; 1 % at 200 keV; 40 % at 30 to 40 MeV) of the kinetic energy of the electrons is converted into X-radiation; the remainder is transformed into heat.
Cooling the anode
The heat which accompanies the production of X-radiation is quite considerable, so that the anode has to be cooled. This can be done in a variety of ways :
1. by natural radiation
2. by convection
3. by forced circulation of liquid or gas
4. by conduction
The focal spot
The area of the target which is struck by the electrons, see figure 2-4, is called the focal spot or “the focus”. It is essential that this area is sufficiently large to avoid local overheating, which might damage the anode.
From the radiographic point of view, however, the focus has to be as small as possible in order to achieve maximum sharpness in the radiographic image. This “focal loading” is expressed in Joule/mm
2
. A tungsten target can take a maximum loading of 200 Joule/mm2.
A higher loading might damage the anode.
32
Fig 1-4. Glass envelope X-ray tube
cathode
focusing
cylinder
or cup
filament
electron beam
X-ray beam
anode
glass
target
4.3 Tube voltage and tube current
The voltage across the X-ray tube determines the energy spectrum and so the hardness of the radiation, see figure 3-4. The intensity is proportional to the tube current, see figure 4-4. This graph shows that, contrary to a change in tube voltage, a change in tube current does not shift the spectrum (in other words: the hardness does not change).
The energy spectrum is also influenced by the characteristics of the high voltage applied to the tube. When the spectrum of one X-ray tube on constant voltage is compared with that of another with a current of pulsating voltage, of the same kV value, both spectra will be slightly different. With a current of pulsating voltage there are, during each cycle, moments of relatively low voltage, during which there will be a greater proportion of “soft” X- rays, with their side-effects. This means that a set working on a constant voltage will provide a higher intensity of hard radiation than one on a pulsating voltage; although both working at the same nominal kV value.
However, even identical X-ray tubes may also show differences in generated energy. The energy generated by one 200 kV X-ray tube will not be true identical to the energy genera­ted by another X-ray tube with the same applied voltage, not even if they are the same type of tube. This behaviour impedes individual calibration in kV of X-ray sets. Another reason why it is hard to calibrate an X-ray tube within a small tolerance band is, that the absolute level and wave characteristics of the supplied high voltage are difficult to measure.
It follows that it is difficult to standardise and calibrate X-ray equipment as far as spectra and kV-settings is concerned, which precludes the exchange of exposure charts, see secti­on 9.1. Each X-ray set therefore requires its own specific exposure chart. Even the exchan­ge of a similar control panel or another (length) of cable between control panel and X-ray tube can influence the level of energy and its spectrum. Usually after exchange of parts or repair the exposure chart for that particular type of X-ray set is normalised (curve-fitting) for the new combination of components. In practice adjusting the zero point of the expo­sure graph is sufficient.
3534
Effective focal spot size
The projections of the focal spot on a surface perpendicular to the axis of the beam of X-rays is termed the “ effective focal spot size” or “ focus size”, see figure 2-4. The effective focus size is one of the parameters in radiography, see section 11-1. The effective focus size, principally determining the sharpness in the radiographic image, has to be as small as pos­sible in order to achieve maximum sharpness. The dimensions of the focus are governed by:
1. The size of the focal spot, and
2. The value of angle α, see figure 2-4.
It should be noted that when in radiography we speak of the “size of the focus” without spe­cifying this more exactly, it is normally the effective focal spot size which is meant. Conventional X-ray tubes have effective focal spot sizes in the range 4 x 4 mm to 1 x 1 mm. There are fine-focus tubes with focal spots from 0.5 x 0.5 mm ad microfocus tubes down to 50 μm diameter or even much less, known as nanofocus tubes.
The effective focal spot size can be determined in accordance with the procedures descri­bed in EN 12543 replacing the old IEC 336 which however is still in use. For more infor­mation on focal spot measurement see section 18.1.
1. Dimension of the electron beam
2. Focal spot
3. Effective focal spot size
4. Anode target
5. True focus size
Fig 3-4. Energy spectra at varying tube voltages and con­stant tube current (here 10mA)
Fig. 4-4. Energy spectra at varying values for tube current and constant high voltage (here 200 kV)
Fig. 2-4. Effective focal spot size
KeV
KeV
relative intensity
relative intensity
4.6 Properties of radioactive sources
Activity (source strength)
The activity of a radioactive substance is given by the number of atoms of the substance which disintegrate per second. This is measured in Becquerels (Bq), 1 Becquerel corresponds to 1 disintegration per second(1 Bq = 1/s).
Specific activity
The specific activity of a radioactive source is the activity of this substance per weight unit, expressed in Bq/g.
Specific gamma-ray emission factor (k-factor)
The k-factor is the generally used unit for radiation output of a source and is defined as the activity measured at a fixed distance. It indicates the specific gamma-emission (gamma constant) measured at 1 metre distance. The higher the k-factor, the smaller the source can be for a particular source strength. A source of small dimensions will improve the sharpness of a radiograph.
Table 1-4 shows the various k-factors and half-life values.
Half-life of a radioactive source
Of an Iridium192 source with an activity of 40 GBq for example 10 GBq will remain after two half-lives (148 days), 5 GBq after three half-lives (222 days) etc.
3736
4.4 Radioactive sources (isotopes)
Natural radioactive sources
The elements from this group which have been used for the purposes of industrial radio­graphy are radium and mesothorium. These give a very hard radiation, making them par­ticularly suitable for examining very thick objects.
A disadvantage of natural sources, next to their high cost, is that it is not possible to make them in dimensions small enough for good quality images and still give sufficient activity.
Artificial radioactive sources
Artificial radioactive sources for NDT are obtained by irradiation in a nuclear reactor. Since 1947, it has been possible to produce radioactive isotopes this way in relatively large quan­tities and in a reasonably pure state and particularly of sufficiently high concentration; the latter being extremely important in NDT because the size of the source has to be as small as possible. Among the many factors deciding a source suitability for non-destructive testing are the wavelength and intensity of its radiation, its half-life and its specific radia­tion. In fact, only a few of the many artificial radio-isotopes available have been found to be suitable for industrial radiography.
4.5 Advantages and disadvantages of artificial radioactive sources
Advantages
1. require no electric power supply; easy to use in the field
2. can be obtained in a range of source diameters, so that if necessary a very short source-to-film distance with a small diameter source can be used, for example, for pipes of small diameter
3. a wide variety of radiation hardnesses
4. higher radiation hardness (more penetration power) than those of conventional X-ray equipment can be selected
Disadvantages
1. cannot be switched off
2. the energy level (radiation hardness) cannot be adjusted
3. the intensity cannot be adjusted
4. limited service life due to source deterioration (half-life)
5. less contrast than X-ray equipment
Isotope Half-life Specific gamma constant
or k-factor
Ytterbium169 31 days 0.05
Iridium192 74 days 0.13
Selenium75 120 days 0.054
Cobalt60 5.3 years 0.35
Caesium137 30 years 0.09
Table 1-4 Various k-factors and half-life values
5
NDT equipment
39
5.1 X-ray equipment
X-ray sets are generally divided in three voltage categories, namely:
1. Up to 320 kV, mainly for use on intermittent, ambulatory work. Tubes are generally of the unipolar alternating current type. Higher voltages are hardly possible with this type of equipment because of insulation problems.
2. Up to 450 kV, mainly for use on continuous, stationary or semi-ambulatory work, because of their dimensions, limited manageability and weight. Tubes are of the bipolar direct current type.
3. Up to 16 MeV, so called Megavolt equipment. Virtually exclusively applied to stationary work.
The first two categories are suitable for radiography on most common objects. Objects of extreme thickness, however, require an energy even higher than 450 kV. In this case Megavolt equipment is used, if alternative sources such as Cobalt60 prove unsuitable. It will normally involve stationary installations of large dimensions and high weight. Lately, portable versions have become available meant for ambulatory use.
Types of X-ray tubes
Depending on the shape of the anode, X-ray tubes produce :
a. a beam of radiation in one direction (directional tube) b. an annular beam (panoramic tube)
X-ray tubes are either unipolar or bipolar.
Bipolar tubes
Figure 1a-5 shows a bipolar tube. The bipolar tube has the advantage that the potential dif­ference with respect to earth on both the anode and the cathode is equal to one-half of the tube voltage, which is a great help from the point-of-view of insulation. The exit window is necessarily situated in the middle of the tube. Bipolar tubes usually operate on direct cur­rent and are generally air, oil or water cooled. They are designed to operate at voltages of 100 to 450 kV and a tube current of up to 20 mA.
Unipolar tubes
In these (shorter) tubes, as shown in figure 1b-5, the anode is held at earth potential and the cathode only has a potential difference to earth. This makes anode cooling a simpler operation. It also means that for low/medium kilo-voltage sets, up to approx. 300 kV as often used in ambulant applications, a single simpler high voltage supply source will suffice. The radiation window is placed asymmetric which can be advantageous in practice.
38
Fig. 1-5. – X-ray tubes A = position of target
Panoramic X-ray beam
Directional X-ray beam
1a-5 Bipolar tube
1b-5 Unipolar tube
Directional X-ray beam
Directional X-ray beam
1c-5 Hollow anode tube giving annular (panoramic) beam
Focus
Focus
41
5.2 High voltage generators
Conventional (trans)portable X-ray equipment for use up to approximately 300 kV are pro­vided with step-up HT transformers, rectifiers and smoothing capacitors. The X-ray tube and the circuitry of this equipment are usually placed in an insulated tank. In most cases these tank type sets use oil for insulation and cooling and weigh approximately 60 kg. Gas is used when weight is important; the set than weighs approximately 30 kg.
Figure 3-5 shows an integrated (all-in-one) tank set for 300 kV with an asymmetric window. At voltages over 300 kV housing eve­rything in one tank becomes very difficult because the high voltage insulation would be inadequate.
Figure 4-5 shows a direct current X-ray tube with a symmetric window. Equipment up to 450 kV operating on direct current is connected to a separate high tension (HT) supply unit by means of HT leads. As a result this equipment is bigger and heavier than “all-in-one” tank sets and mostly meant for stationary or semi-ambulant use.
The 300 kV “all-in-one” tank set and the 450 kV direct current X-ray tube only are of roughly the same dimensions.
Most tank sets are connected to a mains power supply with a frequency of 50 or 60 Hz. At this frequency the supply voltage can be transformed upward. This is followed by rectifying, which occurs in various forms. With some sets the X-ray tube itself functions as rectifier, so called single-phase rectifying. If there is no smoothing applied, considerable changes in voltage per cycle of alternating current will occur. This periodic and greatly varying high voltage affects the intensity and spectrum of the radiation generated, see section 4.3.
40
Special types of X-ray tubes
Unipolar X-ray tubes with a long hollow anode, as shown in fig. 1c-5, are generally known as “rod anode tube” and can be inserted into pipes or vessels. These tubes produce an annular (panoramic) beam over 360º, so allowing a complete circumferential weld to be radiographed in one exposure.
Figure 2-5 shows the conical anode of a (360º) panoramic tube, which allows a circumferential weld to be radiographed centrally, hence uni­formly, from within. With this anode the axis of the electron beam must strike the top of the cone in such a way that the centre of the generated X-ray beam is perpendicular to the longitudinal axis of the tube.
Note: Anodes shaped so that the cen­tre of the generated X-ray beam is not perpendicular (oblique) to the centre line of the tube (which was acceptable in the past), are no longer allowed when work is to be perfor­med to official standards.Tubes that produce a real perpendicular beam are known as "true panoramics"
There are also panoramic tubes in which the electron beam is focused over an extended length by means of a magnetic lens or an electrostatic lens (Wehnelt-cylinder) to produce a very small focal spot size. These sets are called microfocus rod anode tubes with which a very small focal spot size, of less than 10 micrometers, can be achieved. Since the anode can be damaged relatively easy through overheating the anode is usually interchangeable. This requires a separate vacuum unit in order to restore the vacuum after replacement. The advantage of this construction is that with different types of anodes, different radiation patterns can be obtained for special applications. The maximum energy level is usually below 150 kV.
However, there are 150 kV microfocus tubes with a fixed anode for enlarging or scanning purposes, see section 17.1. With these tubes the tube current has to be kept low, because of heat dissipation limitations of the non-interchangeable anode.
Some X-ray tubes used in the radiography of plastics and aluminium are equipped with a beryllium window to allow the softer radiation generated at the lower tube voltages of 5 to 45 kV, to pass.
Fig. 2-5. Anode configuration for an annular panoramic tube
Fig. 4-5. Direct current X-ray tube for 450 kV with a symmetric window
X-ray
electron beam
cathode
X-ray
anode
filament
Window
Window
Fig. 3-5. “All-in-one“ 300 kV tank set with an asymmetric window
43
The linear accelerator (linac)
The energy levels mostly used for linacs (linear accelerators) are 4 MeV and 8 MeV. Linear accelerators can be constructed for one or two energy levels.
In the travelling-wave linac, the acceleration of electrons from a heated filament to very high energies results from the electrons “riding” a high-frequency (3-10 MHz) electromag­netic wave travelling in a straight line down an acceleration tube (the hollow guide). The electrons are bunched into pulses at a frequency of a few hundred pulses per second. The target, which the electrons strike to generate X-radiation, is at the opposite end of the main wave guide of the filament assembly. This is a transmission type target from which the radiation beam passes in a straight line.
The X-ray output from a linear accelerator is many times higher than from a Bètatron of the same energy. An 8 MeV linac with a 2 mm diameter focal spot can deliver a radiation dose rate of 30 Sv/minute at 1 metre distance from the focus. Small light-weight portable linacs of 3 MeV capacity can have outputs of 1.5 Sv/minute at 1 metre distance.
42
The intensity of radiation is increased by double-phased rectifying and varying degrees of smoothing. At very low voltage ripple these sets are considered constant potential (CP) equipment.
In the latest types of tank sets the mains frequency is first converted to a high frequency alternating current and only then transformed upward, which makes it easier still to smooth the ripple. At very high frequencies, up to 50 kHz, smoothing is hardly necessary anymore and such X-ray sets can be called CP systems. Additional features may be built in, for example an automatic warm-up facility, see note below. This type of circuitry with advanced electronics leads to a higher degree of reliability and significant space and weight reduction compared with earlier power supply systems. As a result of the various improve­ments that have gradually been implemented, present day (high frequency) AC X-ray sets perform as well as true CP sets.
Note: Because of the high vacuum prevailing inside the X-ray tube, it carefully has to be warmed-up after a period of rest. During rest the vacuum deteriorates. This warm-up procedure has to be done in accordance with the supplier’s instructions, to prevent high voltage short-circuiting which might damage the tube or render it useless.
5.3 Megavolt equipment
The equipment described in sections 5.1 and 5.2 is used to generate X-radiation up to approximately 450 kV. However, sometimes higher energy levels are needed. Several types of equipment have been built to operate in the 1 MeV to 16 MeV range. In industrial radiography almost exclusively Bètatrons or linear accelerators (linacs) are used. Operating high-energy X-ray installations requires (costly) safety precautions.
The Bètatron
The Bètatron is an electron accelerator, which can produce X-radiation in the 2-30 MeV energy range. The electrons are emitted into a round-sectioned donut shaped glass vacuum tube, as shown in figure 5-5. After several millions of revolutions the electrons reach maximum energy and are deflected towards the target. On the target, part of the electron energy is converted into a tangentially directed beam of X-radiation.
To obtain a reasonably high radiation intensity, most Bètatrons have been designed to ope­rate in the 10-30 MeV energy range, as these voltages achieve maximum conversion rate of electron energy into radiation. Even so the output of Bètatrons is usually small compared to linacs. Transportable low energy Bètatrons (2-6 MeV) have been built, but these gene­rally have a low radiation output, which limits their application.
One advantage of Bètatrons is that they can be built with very small (micromillimeter) focal spots. A disadvantage is that with these very high energy levels the X-ray beam is usu­ally narrow, and the coverage of larger film sizes is only possible by using increased sour­ce-to-film distances. The extended exposure times required can be a practical problem.
Fig. 5-5. Bètatron
X-rays
1. Ring-shaped accelerator tube
2. Anode
3. Cathode filament (emitting electrons)
4.&5 magnetic fields
6. Coils
7. Auxiliary winding
1
2
3
6
5
7
7
4
5.4 Radioactive sources
Table 1-5 shows various radioactive sources for industrial NDT. The most commonly used ones are Cobalt, Iridium and increasingly Selenium. Selenium is very attractive while it permits lighter containers than Iridium. Due to its average energy level it often is a good alternative for an X-ray tube, also attractive while no electricity is needed.
Average energy level (nominal value)
The spectrum of a source has one or more energy lines, as shown in figure 2-2. For sources with multiple energy lines an average energy level is assumed, the so-called nominal value.
On the basis of these spectra data it is clear that Co60, Cs137 and Ir192 sources produce high-energy radiation and are the­refore well suited to irradiate thick materials. Yb169, on the other hand, is a source that produces relatively soft radiation and is of a very small size (0.5 mm), which makes it particularly suitable for radiographic examination of circum­ferential welds in pipes of a small diameter and thin wall thick­ness, with the source centrally positioned so that the weld can be exposed uniformly in one exposure, as shown in figure 8-5.
4544
The main properties of a linear accelerator are:
1. very high output of radiation
2. very small focal spot dimensions (<2 mm)
3. considerable weight (approx. 1200 kg for an 8 MeV stationary installation)
Figure 7-5 shows an 8 MeV linac in a radiation bunker examining a pump housing.
Fig.7-5. Linac and pump house in a radiation bunker
Table 1-5. Radioactive sources used in industrial radiography, in sequence of nominal (average) energy level
Table 2-5. Radiation spectra and nominal values
Element Symbol Mass Specific gamma Average energy
Number constant level
k-factor in MeV
Cobalt60 Co 60 0.35 1.25 Caesium137 Cs 137 0.09 0.66 Iridium192 Ir 192 0.13 0.45 Selenium75 Se 75 0.054 0.32 Ytterbium169 Yb 169 0.05 0.2 Thulium170 Tm 170 0.001 0.072
Source Number of Main energy Nominal value
spectrum lines levels in MeV in MeV
Cobalt60 2 1.17 and 1.34 MeV 1.25 MeV. Caesium137 1 0.66 MeV 0.66 MeV Iridium192 >10 0.3; 0.31; 0.32; 0.47 en 0.6 MeV 0.45 MeV. Selenium75 >4 120, 140 and 400 keV 320 keV. Ytterbium169 >6 0.06 and 0.2 MeV 200 keV. Thulium170 2 52 and 84 keV 72 keV.
source
film
Fig. 8-5. Ytterbium169 source in central position for exposure of circumferential weld
Fig. 6-5. Linear electron accelerator (linac)
X-rays
target
wave guide
electron gun
vacuum pump
magnetron
focus coils
Also greatly depleted uranium (with the highest radiation absorption) is used for shielding, resulting in very compact exposure containers. A disadvantage of this material, however, is that it has a certain minimal radioactivity, which is reason that in some countries the use of depleted uranium is not allowed. Regardless of the shielding material used, all containers have a considerable weight in common.
There are several solutions to the problem of safely storing a source on the one hand, and of putting it in a simple but absolutely safe manner in its radiation position on the other hand. Two regularly used constructions for this purpose are: source S is situated in a rota­ting cylinder, as shown in figure 11-5, or in an S-channel container as shown in figure 12-5.
The S-channel container is usually provided with a means to move the source out from a distance (after all, distance is the safest protection from radiation). This may be done by means of a flexible cable in a hose (Teleflex design) as shown in figures 13-5 and 14-5. With this construction it is possible to extend the flexible hose in such a way that the source can safely be moved several metres out of the container to the most favourable exposure position.
4746
5.5 Source holders (capsules)
All gamma-ray sources for radiography are supplied in hermetically sealed, corrosion resistant source holders (capsules), made out of monel, vanadium or titanium. The Atomic Energy Authority in the country of origin encapsulates the radioactive materi­al. The supplier will supply the source with a certificate which indicates the type of source, its serial number, the activity at a certain date, and a disintegration graph.
The radiation material proper, also called the source or pellet, ranges in size from 1 to 4 mm. The size is dictated by the specific radiation activity of the source material. The outside dimensions of the cylindrical capsule are approximately 5.5 x 15 mm, as shown in figure 9-5.
5.6 Transport- and exposure containers
Transportation and handling of sealed sources are subject to strict international safety regulations, as a source is continuously emitting radiation in all directions, in contrast to an X-ray tube which can be switched off. During transportation and use the source must be surrounded by a volume of radiation absorbing material, which in turn is encapsulated in a container. The level of radioactivity at the outside surface of the container shall not exceed the legally established maximum limit.
Like the transport container, the exposure container needs to be robust and must function safely at all times. The exposure container, also called camera, must be fail-safe and water­and dirt proof. It must also not be effected by impact. Moreover, if the radiation-absorbing material, for example lead, melts (in a fire) the radia­tion absorbing qualities must not be lost. This requires a casing made of a material with a high melting point, for example steel. Besides lead, increasingly a new sintered material with very high tungsten content (97%) is used as shielding material. This material is easily worked and finished and not prone to melting.
Fig. 9-5. Cross-section of a capsule for a radioactive source
15 m m
5.5 O
/
Fig. 10-5. Sealed capsule
handling / operation side
flexible connection
storage position of the source
casing/container
exposure position
of the source
shielding
Fig 11-5. Exposure container with source S in a rotating inner cylinder
Fig. 12-5. S-channel container with source S in storage position
Fig. 13-5. Exposure container with S-channel and flexible operating hose ad cable
open
SS
S
closed
Figure 14-5 shows an S-chan­nel container with a flexible (metal) hose and cable in rol­led up (transport) position.
Figure 15-5 shows a more recent (2006) S-channel Selenium75 container with operation hoses and pigtail. Selenium75 radio-isotope is becoming popular since new production (enrichment) methods resulted in a much better k-factor. Thus for a cer­tain activity (source strength) a much smaller source size (focus) is achieved. This results in a better/sharper image qua­lity than could be achieved with the old Selenium75 production method.
Due to its average energy level of 320 kV, Selenium75 increasingly replaces X-ray equipment for a thickness range from 5 mm to 30 mm of steel. This eliminates the need for electric power, very attractive in the field for reasons of electrical safety and more convenient at remote- or work locations with difficult access (high, deep, offshore, refineries, etc). Last but not least, a Selenium container is of much lower weigth than needed for an Iridium192 container with the same source strength.
To enable radiography on work sites with (many) people in the vicinity, for example on offshore installations or in assembly halls, containers with rotating cylinders and collimators were developed so that only the beam of radiation required for the radiograph is emitted. The remainder of radiation is absorbed by the collimator material which allows people to work safely at a distance of a few metres while radiography is in progress. Such containers with collimators are known by the name of “CARE” (Confined Area Radiation Equipment) or “LORA” (Low Radiation) equipment.
4948
Fig. 14-5. S-channel container with the flexible cable and deployment mechanism.
Fig. 15-5. S-channel container for Selenium75 with pigtail (at right) and operating hoses (at left)
Fig. 16a-5. Gamma container with collimator on a circumferential weld in a pipe
handle
tungsten container
rotating cylinder source
Collimator
base pipe and weld
film lead shielding
boundaries of the beam of radiation
Fig. 16b-5. Cross-section of CARE/LORA container on the pipe
Without collimating the minimum safety distance is considerably more than 10 metres (in all directions!).
Such containers with collimators are particularly suitable for frequent and identical repetitive NDT work, for example radiographic testing of welds in pipes of < 300 mm diameter. Figure 16a-5 shows such a special container with collimator set up for a double wall radiograph. The cross-section drawing of figure 16b-5 shows the boundaries of the beam of radiation. For bigger focus-to-film distances, longer collimators are used to restrict the beam of radiation.
This type of container is suitable for Iridium sources up to 1000 GBq and weighs “only” approx. 20 kg. A similar system (Saferad) with a weight of up to 15 kg exists, using Selenium75, which almost eliminates the usual disruption to construction, maintenance and process operations in the vicinity of the exposure.
5.7 Checking for container leakage
A sealed radioactive source (capsule) might start to leak and become an open source as a result of corrosion, mechanical damage, chemical reactions, fire, explosion etc. Regular mandatory “wipe-tests” by specialists serve to detect leakage at an early stage.
measuring tape
weld
lead
51
50
6
Radiation images, filters and intensifying screens
To influence the effects of radiation on an image, filters and intensifying screens are used to :
filter / harden the radiation to influence contrast and/or
to intensify the effect of radiation to improve contrast
6.1 Radiation images
The intensity of a beam of X-rays or gamma-rays undergoes local attenuation as it passes through an object, due to absorption and scattering of the radiation. On a uniform object attenuation of the primary beam will also be uniform and the film evenly exposed. If the object contains defects or is of variable thickness, the surface of the film will be unevenly exposed resulting in a shadow image of the object and the defects in it. When the film is processed the variations in radiation intensity show up as varying film densities; higher radiation intensity producing higher film density resulting in a negative X-ray image as shown in figure 1-6.
When the primary beam is partly absorbed in the object, some radiation, as shown in figure 2-6, will be scattered and reach the film as secondary radiation by an indirect path. The quality of the radiograph is reduced by this scattered radiation, and it is important to keep its effects to a minimum.
At any point P on the film, therefore, the total radiation reaching that point is made up of some transmitted primary radiation forming the image of cavity (N), the “image forming”­or direct radiation intensity Ip, and some secondary “ non-image forming” , scattered radi­ation, intensity Is. Hence, the total radiation intensity at P is (I
p
+ Is).
The ratio Is/Ipis called the “scattered radiation factor” and can be as high as 10 for great wall thicknesses, which means that the scattered radiation is ten times higher than the image-forming radiation. The ratio (I
p+Is
)/Ip = 1+Is/Ipis called the “build-up factor” and
is of considerable importance for the detectability of defects. It usually has a value between 2 and 20, depending on radiation energy and object thickness.
It must also be appreciated that any object in the neighbourhood of the object being exam­ined (table, walls, ground and so on) which is struck by the gamma- or X-rays will partial­ly reflect these rays in the form of “backscatter” which is liable to fog the film.
Backscatter coming from the object under examination is less hard than the primary radi­ation that has caused it and can be intercepted by a metal filter between object and film. Radiation scattered by objects nearby the film can be intercepted by means of a protective sheet of lead at the rear face of the film cassette.
defect
object
object
film
A
A
C
N
D
E
B
F
P
source (S)
primary radiation
Fig. 1-6. The negative X-ray image
film
negative (shadow) image on the film
radiation intensity
after passing
through the object
Fig. 2-6. Image forming and non-image forming radiation
Only the radiation from source (S) that reaches the film in straight lines via beam section DE, produces an image of cavity N at P. The remainder, not reaching P directly, is scattered radiation, no defect image forming thus reducing the image quality.
53
6.2 Radiation filters
When a metal plate, usually lead or copper, is placed between the tube window and the object, radiation “hardening” occurs leading to a lower image contrast. This may be counter-balanced by a metal filter placed immediately behind the object (i.e. between object and film). This filter will cause the (softer) scattered radiation passing through the object to be absorbed by the filter to a greater extent than the primary (har­der) radiation. This also improves the image quality.
If the edges of an object being radiographed are not close to the film (as in the case of a cylindrical body in figure 3-6) considerable scatter of the primary radiation can occur, leading to fogging. This scatter can be prevented by positioning sheets of lead foil between the object and the film as illustrated in this figure.
Reducing the contrast by filtration is also desirable when a radiographic image of an object of widely varying thicknesses has to be obtained on a single film see section 18.2.
Typical filter thicknesses are :
0.1 – 0.25 mm lead for 300 kV X-rays
0.25 – 1.0 mm lead for 400 kV X-rays
6.3 Intensifying screens
The radiographic image is formed by only approximately 1 % of the amount of radiation energy exposed at the film. The rest passes through the film and is consequently not used. To utilise more of the available radiation energy, the film is sandwiched between two inten­sifying screens. Different types of material are being used for this purpose.
Lead screens
Under the impact of X-rays and gamma-rays, lead screens emit electrons to which the film is sensitive. In industrial radiography this effect is made use of: the film is placed between two layers of lead to achieve the intensifying effect and intensity improvement of approxi­mately factor 4 can be realised. This method of intensification is used within the energy range of 80 keV to 420 keV, and applies equally to X-ray or gamma-radiation, such as pro­duced by Iridium192.
Intensifying screens are made up of two homogeneous sheets of lead foil (stuck on to a thin base such as a sheet of paper or cardboard) between which the film is placed: the so called front and back screens.
The thickness of the front screen (source side) must match the hardness of the radiation being used, so that it will pass the primary radiation while stopping as much as possible of the secondary radiation (which has a longer wavelength and is consequently less penetrating).
52
Scattered radiation also occurs in radiographic examination of cylindrical objects, as shown in figure 3-6.
The effects of scattered radiation can be further reduced by :
• limiting the size of the radiation beam to a minimum with a diaphragm in front of the tube window
• using a cone to localise the beam, a so called collimator
• the use of masks: lead strips around the edges of the object.
1
S
X-ray s
Fig. 3-6. Scattered radiation in radiogra­phy of cylindrical objects. Scattered radiation from object 1 causes a spurious band at B, object 2 at A etc, unless lead strips are used as shown in the lower part of this figure
film
film
lead strips
A B C D
23
123
55
Steel and copper screens
For high-energy radiation, lead is not the best material for intensifying screens. With Cobalt60 gamma-rays, copper or steel have been shown to produce better quality radio­graphs than lead screens. With megavoltage X-rays in the energy range 5-8 MeV (linac) thick copper screens produce better radiographs than lead screens of any thickness.
Fluorescent screens
The term fluorescence (often mistaken for phosphorescence) is used to indicate the cha­racteristic of a substance to instantly instantly emit light under the influence of electro­magnetic radiation. The moment radiation stops, so does the lighting effect. This pheno­menon is made good use of in film based radiography. Certain substances emit so much light when subjected to ionising radiation, that they have considerably more effect on the light sensitive film than the direct ionising radiation itself.
The term phosphorescence is used to describe the same luminescent phenomenon, but once the electromagnetic radiation ceases, light fades slowly (so called after-glow).
NDT additionally uses the “memory effect” of some phosphorous compounds to store a latent radiographic image in order to develop it later into a visible image with the aid of laser stimulation, see section 16.4.
Fluorescent salt screens
Fluorescent screens consist of a thin, flexible base coated with a fluorescent layer made up from micro-crystals of a suitable metallic salt (rare earth; usually calcium tungstate) which fluoresce when subjected to radiation. The radiation makes the screen light up. The light intensity is in direct proportion to the radiation intensity. With these screens a very high intensification factor of 50 can be achieved, which means a significant reduction in expo­sure time. The image quality, however, is poor because of increased image unsharpness. Fluorescent screens are only used in industrial radiography when a drastic reduction of exposure time, in combination with the detection of large defects, is required.
Fluorometallic screens
Apart from fluorescent and lead intensifying screens, there are fluorometallic screens which to a certain extent combine the advantages of both. These screens are provided with a lead foil between the film base and the fluorescent layer. This type of screen is intended to be used in combination with so-called RCF-film (Rapid Cycle Film) of the types F6 or F8, see section 8.1.
The degree of intensification achieved largely depends on the spectral sensitivity of the X-ray film for the light emitted by the screens. Due to the considerable exposure time reduc­tion the application is attractive for work on lay barges and in refineries.
54
The lead foil of the front screen is usually 0.02 to 0.15 mm thick. The front screen acts not only as an intensifier of the primary radiation, but also as an absorbing filter of the softer scatter, which enters in part at an oblique angle, see figure 2-6. The thickness of the back screen is not critical and is usually approx. 0.25 mm.
The surface of lead screens is polished to allow as close a contact as possible with the sur­face of the film. Flaws such as scratches or cracks on the surface of the metal will be visible in the radiograph and must, therefore, be avoided. There are also X-ray film cassettes on the market with built-in lead screens and vacuum packed to ensure perfect contact bet­ween emulsion and lead foil surface.
Figure 4a-6 and figure 4b-6 clearly show the positive effect of the use of lead screens.
Summarizing, the effects of the use of lead screens are :
• improvement in contrast and image detail as a result of reduced scatter
• decrease in exposure time
Fig. 4a-6. Radiograph of a casting without lead intensifying screens
Fig. 4b-6. Radiograph of a casting with lead intensi­fying screens
On balance for on-stream inspection, the relative time saving is much smaller; usually no more than a factor 2 for an F6-film (at Ir192 and Co60) instead of 10 in the D7 lead screen technique. See the bold figures (2.5 and 1.7) in table 2-6.
Figure 6-6 gives an overview of graphs from which the relative exposure times can be dedu­ced when using different films and screens at 200 kV, (for film-density 2). The graph shows that an F8-film with RCF screen (point C) is approximately 8 times faster than a D8-film with lead (point B) and approximately 15 times faster than a D7-film with lead (point A). Since on-stream examination as well as examination of concrete, and also flash radiogra­phy (see section 18.7) allow concessions to image quality, a special fluorometallic screen (NDT1200) has been developed with extremely high light emission. In combination with an F8-film it may result in a reduction in exposure time at a factor 100 at 200 kV, against a D7-film with lead (point D as opposed to point A in figure 6-6), or even a factor 140 to 165, depending on source selection, see table 2-6. The intensification factor of the NDT1200 screens increases significantly at lower temperatures.
Table 2-6 shows the effect of radiation hardness on relative exposure times for the various film/screen combinations compared with D7 film with lead screen. Noticeably, for the NDT1200 screen and F-8 film the factor increases with the increase in energy, but for the F6 film the factor decreases at energy levels exceeding 300 keV.
It is clear from the above tables and graphs that there are many ways to reduce the expo­sure time or radiation dose needed. The required image quality is decisive (a higher expo­sure rate automatically means reduced image quality), and next the economic factors, for example the cost of the screens against time saved need to be weighed.
5756
To achieve satisfactory radiographs with fluorometallic screens, they should be used in combination with the appropriate F-film type.
When used correctly and under favourable conditions, exposure time can be reduced by a factor 5 to 10, compared with D7 film in combination with lead screens. This is not a con­stant factor because the energy level applied (radiation hardness) and ambient temperatu­re also affects the extent of fluorescence. For example, at 200 kV a factor 10 can be achie­ved, but with Iridium192 (nominal value 450 kV) it will only be a factor 5 compared to D7 film. Table 1-6 shows the relative exposure factors for the RCF-technique.
A total processing cycle of a few minutes is possible with the use of an automatic film pro­cessor which makes it a very attractive system to deploy offshore (on lay barges) where weld examination has to be done at a very fast rate and few concessions are made towards image quality. Fig. 5-6 shows that a time saving at 10
(3.7-2.8)
or 10
0.9
works out at approxi-
mately a factor 8. The actual time saving is often closer to factor 10.
These RCF screens are also used for “on-stream” examination - also known as profile radio­graphy- (see section 18.6), whereby long exposure times and mostly hard (gamma) radia­tion are applied because of the penetrating power required. However, the relatively long exposure time (causing reciprocity) and hard radiation (Cobalt60) together considerably reduce the light emission effect, as tables 1-6 and 2-6 show.
Film system Relative exposure time
200 kV Ir192 (450 kV)
F6 + RCF screens 0.1 0.2
D7 + lead screens 1.0 1.0
Table 1-6. Relative exposure factors for RCF technique
Table 2-6. Relative exposure times for NDT1200, RCF and lead screens.
Relative exposure times
Energy level Screen type Film F8 Factor Film F6 Factor Film D7
100 kV NDT1200 0.01 100 0.05 20
RCF 0.03 33 0.17 6
none 1
300 kV NDT1200+Pb 0.008 125 0.04 25
RCF 0.02 50 0.13 8
Lead 1
Ir192 NDT1200+Pb 0.007 140 0.06 17
450 keV RCF 0.035 30 0.4 2.5
Lead 1
Co60 NDT1200 0.006 165 0.1 10
1.25 MeV RCF 0.04 25 0.6 1.7 Lead 1
F6+
RCF screen
F8 +
NDT 1200
F8+RCF
D8+lead
D7+lead
density
density
log. rel.bel. log. rel.bel.
D7+lead
Fig. 5-6. Relative exposure time RCF and lead intensifying screen, for 300kV
Fig. 6-6. Speed comparison F8 film +NDT1200 and RCF versus D7 and D8 +lead, for 200kV
Loading...
+ 84 hidden pages