The information in this manual only applies to CIC Pro Clinical Information Center software version 4.1.
It does not apply to earlier software versions. Due to continuing product innovation, specifications in this
manual are subject to change without notice.
Listed below are GE Medical Systems Information Technologies trademarks used in this document. All other
trademarks contained herein are the property of their respective owners.
APEX and CD TELEMETRY are trademarks of GE Medical Systems Information Technologies registered in
the United States Patent and Trademark Office.
APEXPRO, CD TELEMETRY
®
-LAN, CENTRALSCOPE, CIC PRO, and OCTACOMM are a trademarks of
The CIC Pro Clinical Information Center bears CE mark CE-0459
indicating its conformity with the provisions of the Council Directive 93/
42/EEC concerning medical devices and fulfills the essential
requirements of Annex I of this directive. The product is in radiointerference protection class A in accordance with EN 55011.
The country of manufacture can be found on the equipment labeling.
The product complies with the requirements of standard EN 60601-1-2
“Electromagnetic Compatibility - Medical Electrical Equipment”.
The CE marking prescribed above is only found on the equipment
labeling of the 230V European equipment. Only applicable equipment
will bear the CE marking.
The safety and effectiveness of this device has been verified against
previously distributed devices. Although all standards applicable to
presently marketed devices may not be appropriate for prior devices (i.e.
electromagnetic compatibility standards), this device will not impair the
safe and effective use of those previously distributed devices. See user’s
information.
CE Marking Information Exceptions
EMC: Immunity Performance
Users should be aware of known RF sources, such as radio or TV stations
and hand-held or mobile two-way radios, and consider them when
installing a medical device or system.
Be aware that adding accessories or components, or modifying the
medical device or system may degrade the EMI performance. Consult
with qualified personnel regarding changes to the system configuration.
Revision ACIC Pro Clinical Information CenterCE-1
2001099-183
CE Marking Information
Specified Electromagnetic Environment.
The CIC Pro and ApexPro server is suitable for use in the specified
electromagnetic environment. The customer and/or the user of the CIC
Pro and ApexPro server should assure that it is used in an
electromagnetic environment as described below:
36.202.1 Immunity: ESDDirect - Discharges of 6 KV or greater to the
rear I/O connector area may cause the system
to lock up, thus experiencing loss of data and
loss of functionality. Operator intervention may
be required.
Likelihood of occurrence: Remote
During testing there were 2 occurrences out of
1,920 discharges.
The rear I/O connector area is not considered
to be user accessible during normal operation.
36.202.3.1 Immunity: Fast Transient Transients on the AC power line of +/- 1 KV or
higher may cause momentary network packet
loss (i.e. waveform and/or numeric data), thus
experiencing momentary loss of data at the
time of the surge.
36.202.3.2 Immunity: Fast SurgesSurges on the AC power line of +/- 1 KV or
higher may cause momentary network packet
loss (i.e. waveform and/or numeric data), thus
experiencing momentary loss of data at the
time of the surge.
Floors should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Mains power quality should be that of a typical
commercial and/or hospital environment.
Mains power quality should be that of a typical
commercial and/or hospital environment.
CE-2CIC Pro Clinical Information CenterRevision A
2001099-183
General Information
This manual is an integral part of the product and describes its
The symbol means ATTENTION: Consult accompanying
Information which refers only to certain versions of the product is
The warranty does not cover damages resulting from the use of
GE Medical Systems Information Technologies is responsible for the
CE Marking Information
intended use. It should always be kept close to the equipment.
Observance of the manual is a prerequisite for proper product
performance and correct operation and ensures patient and operator
safety.
documents.
accompanied by the model number(s) of the product(s) concerned.
The model number is given on the nameplate of the product.
accessories and consumables from other manufacturers.
effects on safety, reliability, and performance of the product, only if:
or repairs are carried out by persons authorized by GE Medical
Systems Information Technologies;
the electrical installation of the relevant room complies with the
requirements of the appropriate regulations; and,
the Clinical Information Center is used in accordance with the
instructions for use.
All publications conform with the product specifications and
applicable IEC publications on safety and essential performance of
electromedical equipment as well as with applicable UL and CSA
requirements and AHA recommendations valid at the time of
printing.
The quality management system complies with the international
standards EN ISO 9001, ISO 13485 and EN 46001, and the Council
Directive on Medical Devices 93/42/EEC.
This operator manual has been prepared by the Technical Writing staff
of GE Medical Systems Information Technologies. It provides operating
instructions for the CIC Pro with GE Medical Systems Information Technologies patient monitors and interface devices.
The CIC Pro can be used as a central station for viewing bedside
monitored patients, telemetry patients, or both. All of the basic operating
instructions for the CIC Pro itself are in this manual. Information
presented in this manual applies when the type of monitoring (using
either a bedside patient monitor or telemetry) is not a factor in CIC Pro
operation. For information about monitoring functions that are specific
to the patient monitor or telemetry system, refer to the operator’s
manual for that product.
Intended Audience
The Basics: About This Manual
This manual is geared for clinical professionals. Clinical professionals
are expected to have working knowledge of medical procedures,
practices, and terminology as required for monitoring of critically ill
patients.
Revision ACIC Pro Clinical Information Center1-3
2001099-183
Definitions
The Basics: About This Manual
The following formats are used in this manual to highlight various web
viewer features and functions.
Black textIndicates keys on the keyboard, text to be entered, or hardware items such as buttons or
switches on the equipment.
Italicized textIndicates software terms that identify menu items, buttons, or options in various windows.
Ctrl+EscIndicates a keyboard operation. A (+) sign between the names of two keys indicates that
you must press and hold the first key while pressing the second key once.
For example, “Press Ctrl+Esc” means to press and hold down the Ctrl key while pressing
the Esc key.
<Space>Indicates you must press the spacebar. When instructions are given for typing a precise text
string with one or more spaces, the point where the spacebar must be pressed is indicated
as: <Space>. The purpose of the < > brackets is to ensure you press the spacebar when
required.
EnterIndicates you must press the “Enter” or “Return” key on the keyboard. Do not type “enter”.
Illustrations and Names
All illustrations in this manual are provided as examples only. They may
not necessarily reflect your monitoring setup or data displayed on your
monitor.
In this manual, all names appearing in examples and illustrations are
fictitious. The use of any real person’s name is purely coincidental.
1-4CIC Pro Clinical Information CenterRevision A
2001099-183
Revision History
The Basics: About This Manual
This manual has a revision letter, located at the bottom of each page.
This revision letter changes whenever the manual is updated. Revision A
is the initial release of the document.
RevisionDateComments
A22 April 04Manual released with this document number to
correspond with CIC Pro software version 4.1
Revision ACIC Pro Clinical Information Center1-5
2001099-183
The Basics: About the Clinical Information Center
About the Clinical Information Center
This manual addresses the operation of the Clinical Information Center.
The Clinical Information Center is referred to as the CIC Pro throughout
the rest of this document.
The CIC Pro displays data from up to 16 bedside devices (patient
monitors, Unity Network
transmitters, etc.) at one time. This data is collected for each individual
monitor and sent to the CIC Pro via the Unity Network. Included are the
bed number, the primary ECG waveform, heart rate, alarm messages,
etc.
The vital sign numerics displayed are determined at the GE Medical
Systems Information Technologies bedside monitor or the telemetry
receiver cabinet. GE bedside monitors send numerics for all actively
monitored parameters as well.
Control is provided through the mouse or optional touchscreen display
and keyboard.
ID connectivity devices, telemetry
Telemetry Compatibility
The following page shows an illustration of the CIC Pro with the various
components, indicators, and controls labeled.
Refer to the service manual for installation instructions, including
mounting and support requirements.
The following issues apply with regard to compatibility with CIC Pro
software version 4.1
Due to changes in Telemetry Unit Defaults and Telemetry Alarm
Control defaults, Centralscope
TM
central stations and previous
software versions of CIC are not in-unit compatible with CIC Pro
software version 4.1.
The CD Telemetry-LAN software is compatible with all versions of
CIC Pro software.
The ApexPro telemetry system software 3.x is only compatible with
CIC Pro software version 4.x.
The ApexPro telemetry system software 1.7 and 2.x is compatible
with CIC software version 3.x.
The ApexPro telemetry system software 1.x is compatible with CIC
software version 2.x.
The ApexPro telemetry system is not compatible with versions of CIC
software earlier than 2.x.
1-6CIC Pro Clinical Information CenterRevision A
2001099-183
Front View
The Basics: About the Clinical Information Center
001B
Clinical Information Center
The components shown in the photograph above are representative of a
typical CIC Pro. They may not appear identical to the components of
your system. The CIC Pro processing unit is not shown.
Revision ACIC Pro Clinical Information Center1-7
2001099-183
Power
Turning Power On
The Basics: Power
WARNING
Check the service manual for AC power requirements
before installation.
When you properly connect all cables, press the power switches on the
back of the processing unit and the monitor. (Refer to the service manual
for information on system connections.) A green indicator light means
the power is on. (Refer to the figure on the previous page.)
CAUTION
Do NOT load any software other than that specified by
GE Medical Systems Information Technologies onto the
CIC Pro. Installation of software not specified by GE
Medical Systems Information Technologies may cause
damage to the server or loss or corruption of data.
1-8CIC Pro Clinical Information CenterRevision A
2001099-183
Controls
Mouse
The Basics: Controls
Operating the CIC Pro is done with the mouse, the keyboard or an
optional touchscreen display.
The mouse is the primary means of user interaction with the CIC Pro.
Left Mouse ButtonRight Mouse Button
Mouse Pointer Shapes
027A
Depending on your action, or the mode of operation, the mouse pointer on
the CIC Pro screen shapes differently.
Arrow
The mouse pointer is arrow-shaped when you operate in user
mode. When the mouse pointer is arrow-shaped, you can choose
028A
commands, highlight options, and move from window to
window.
I-Beam
The mouse pointer is shaped like an I-beam when you are in a
030A
text entry field.
Cross
The mouse pointer is cross-shaped when you are operating in
service mode. For more information on the CIC Pro’s service
029A
mode, refer to the service manual.
Revision ACIC Pro Clinical Information Center1-9
2001099-183
Left Mouse Button
Right Mouse Button
The Basics: Controls
Use the left mouse button for making patient-oriented selections and
initiating actions on the screen of the CIC Pro.
Use the right mouse button for making system-oriented selections. When
you press the right mouse button, the right-click menu opens.
NOTE
The right mouse button only opens the right-click menu. It does not
highlight selections or initiate actions other than the items you
select.
Right-Click Menu—Example
Right-click menu items available within the user mode are:
Select Care Unit then Bed Number
Select Waveform #1 color
Select Waveform #2
Select Waveform #3
Select Waveform #4
023B
1-10CIC Pro Clinical Information CenterRevision A
2001099-183
The Basics: Controls
“Clicking” the Mouse
For purposes of this manual, the term “clicking” refers to positioning the
mouse pointer on a selection and pressing the left mouse button.
NOTE
In situations where the right mouse button is pressed, this is
specifically called out. In all other cases, assume you should press the
LEFT mouse button.
Optional Touchscreen Display
The touchscreen display is an optional component of the CIC Pro. A
touchscreen display (or simply touchscreen) is a screen with areas that
are sensitive to touch. Essentially, any area that can be selected using
the mouse pointer, can be selected using the touchscreen.
Guidelines for using the touchscreen display:
The touchscreen feature does not function properly if tape or paper is
stuck to the screen’s surface.
Do not use pencils, pens, and other sharp, pointed objects to activate
the touchscreen.
The right mouse click function cannot be used with a touchscreen
display.
NOTE
Any time the manual says to “click with the mouse” you can
optionally tap the display control with a finger.
Revision ACIC Pro Clinical Information Center1-11
2001099-183
Keyboard
Use the keyboard to enter patient information when you admit a patient.
To enter patient data, position the mouse pointer in the text entry field.
When the mouse pointer is positioned correctly, its shape changes from
an arrow to an I-beam. The I-beam mouse pointer indicates that you can
type, select text, or reposition the insertion point.
Silence Alarms Keyboard Key
Press the Silence Alarms key located on the keyboard to silence the
audible alarm tones.
This sends a silence alarms command to those beds displayed in the
multiple patient viewer or single patient viewer who have queued or are
sounding an audio alarm on the CIC Pro.
The Basics: Controls
107A
1-12CIC Pro Clinical Information CenterRevision A
2001099-183
Display Formats
The Basics: Display Formats
NOTE
Display format is subject to licensing restrictions. Refer to the service
manual for licensing information.
You can configure the CIC Pro to display patient information for one to
16 patients. Set up the display format in the CIC Setup window. For
more information on configuring the display, refer to Chapter 4, “CIC
Setup”.
Revision ACIC Pro Clinical Information Center1-13
2001099-183
Terminology
Bedside Monitoring
Telemetry Monitoring
The Basics: Terminology
For the purposes of this document, the following terms apply:
A bedside monitor is a stationary monitor (user-configured or factory
configured).
These monitors are connected directly to the patient. They are set up
with a unit name as well as a bed name (e.g., IMC-BED4). For a userconfigured monitor, data is processed by an acquisition module. For a
factory-configured monitor, data is processed within the monitor itself.
Telemetry monitoring occurs when ECG data is transmitted by a
telemetry transmitter to a telemetry receiver cabinet over an established
antenna system and viewed at a designated location. The display
monitor identifies a telemetry bed by placing an asterisk next to the bed
name (e.g., IMC-BED4*). ECG data is processed by the telemetry
receiver cabinet.
Switching Transmitters
If you wish to switch to an ApexPro transmitter from a CDT-LAN
transmitter (e.g. Apex, Apex S) or vice versa, while a patient is admitted,
you must:
Discharge the patient (losing stored data).
Switch transmitters.
Re-admit the patient.
Monitoring stops if you switch transmitters while a patient is admitted.
1-14CIC Pro Clinical Information CenterRevision A
2001099-183
Locked or Unlocked Beds
The clinical information center can be configured with the bed names
either in locked or unlocked mode. When locked, the bed names are
permanently assigned to specific windows. For more information, refer to
“Locked/Unlocked Beds” on page 1-18.
NOTE
It is possible to admit a patient to a window with a bed name that is
locked to NONE. To avoid duplication of patient waveforms, a
window locked as NONE should not be used to admit a patient.
The Basics: Terminology
Revision ACIC Pro Clinical Information Center1-15
2001099-183
The Basics: Multiple Patient Viewer
Multiple Patient Viewer
The multiple patient viewer of the CIC Pro displays information for up to
16 patients along with the Main Menu buttons. (For more information on
the Main Menu buttons, refer to “Main Menu Buttons” on page 1-23.)
Each patient slot contains an ECG parameter window and ECG
waveform. Any default and user-selected leads appear in the waveform
window. Depending on the selected display format, additional monitored
parameters display for patients monitored on a GE Medical Systems
Information Technologies bedside monitor. Refer to Chapter 4, “CIC
Setup”, for information on modifying display formats.
1-16CIC Pro Clinical Information CenterRevision A
2001099-183
The Basics: Multiple Patient Viewer
Q
P
O
N
M
L
K
A
B
C
D
E
F
G
H
J
I
031C
Clinical Information Center Multiple Patient Viewer
A Browse buttonJ Alarm Condition/Quick Access Buttons
B Click Here to Open This Patient’s Single Patient ViewerK Message Area
C Colored Border Alarm IndicatorL Patient Name
D Alarm MessageM Bed Name / Unit Name
E Empty Patient WindowN Additional Parameter Information
F Admit ButtonO Patient’s Waveform Window
G Click Here to Print this Patient’s WaveformsP Patient’s ECG Parameter Window
H Click Here to Open Graph All Patients MenuQ Header bar: Displays time, maker of software, product name,
I Main Menu Buttons
and server name.
Revision ACIC Pro Clinical Information Center1-17
2001099-183
The Basics: Multiple Patient Viewer
Telemetry and Network Bed Names
Telemetry bed names are designated with an asterisk (*) to distinguish
them from bedside monitored patients.
Network beds (such as Unity Network ID) are designated with a plus
sign (+).
The same bed may be used for patients connected to a bedside monitor,
an Unity Network ID connection, and the telemetry monitoring system.
Duplicate bed numbers are not allowed, so the system adds these
designators to accommodate this feature.
Monitor SetupBed Name Designation
ECG Source = Bedside MonitorUnit Name | Bed Name
ECG Source = TelemetryUnit Name | Bed Name*
ECG Source = Bedside Monitor with Unity Network ID connectionUnit Name | Bed Name+
ECG Source = Telemetry with Unity Network ID connectionUnit Name | Bed Name*+
Table 1. Bed Name Designations
Locked/Unlocked Beds
Throughout this document, references are made to “locked” beds. If a bed
is locked, it means that a particular bed is allocated permanently to a
particular slot on the CIC Pro and users are unable to move the bed to
another slot on the CIC Pro.
Within the service mode, this option is
enabled. You can choose between
LOCK and UNLOCK.
034B
Lock/Unlock Beds Right-Click Menu
1-18CIC Pro Clinical Information CenterRevision A
2001099-183
Loading...
+ 192 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.