We thank you very much for purchasing Fuji Electric’s cross-stack laser gas analyzer (Type: ZSS).
First read this instruction manual carefully until an adequate understanding is acquired.
Then proceed to installation, operation and maintenance of the laser gas analyzer. Improper
handling may result in an accident or a failure.
The specifications of the laser gas analyzer may be changed without prior notice for further
product improvement.
Modification of the laser gas analyzer is strictly prohibited unless a written approval is obtained
from the manufacturer. We will not be responsible for any accident attributable to such remod-
eling without permission. If it becomes necessary to modify the laser gas analyzer, contact the
manufacturer in advance.
This instruction manual shall be stored by the person who actually uses the laser gas analyzer.
After reading the manual, be sure to keep it at a place easy to access.
This instruction manual should be delivered to the end user without fail.
This product falls under Category 9 (monitoring and control instruments) set out in Annex I of
the RoHS directive 2011/65/EU, and not for consumer use.
If you return the product to us for repair, provide us with a document that indicates the purpose
of export is repair and a certificate that indicates that the equipment includes no substances re-
stricted by RoHS directive or laws and regulations of the exporting country. We are not liable in
the cases that the re-export from Japan to you is not permitted due to imperfection of the above
documents.
Manufacturer: Fuji Electric Co., Ltd.
Type: Described in nameplate on main frame
Date of manufacture: Described in nameplate on main frame
Product nationality: Japan
Fuji Electric Co., Ltd. 2018
Request
Transcription of a part or the whole of this manual without
permission is prohibited.
The contents of this manual are subject to change without
prior notice.
Issued in August, 2018
Rev. 1st edition September, 2018
INZ-TN8ZSS-E
- i -
Page 3
CONTENTS
PREFACE .......................................................................................................................................................... i
CONTENTS ..................................................................................................................................................... ii
SAFETY INFORMATION ............................................................................................................................ iv
WARRANTY AND MAINTENANCE ......................................................................................................... vii
4.12 How to use the bolt for angle fine adjustment ..................................................................................... 25
4.13 Connecting to control unit .................................................................................................................... 26
5. OPERATION PANEL AND SCREEN ..................................................................................................... 31
5.3 Outline of screen .................................................................................................................................... 33
6. CALIBRATION AND SETTING ............................................................................................................. 42
6.1 Zero calibration ...................................................................................................................................... 42
6.4 Output hold ............................................................................................................................................ 60
6.7 Analog input ........................................................................................................................................... 77
6.8 Analog output ......................................................................................................................................... 83
6.9 Fine adjustment of analog output value ................................................................................................. 84
6.10 Digital Output ...................................................................................................................................... 87
7.1 Maintenance list ..................................................................................................................................... 92
7.3 Zero calibration ...................................................................................................................................... 93
First of all, read this “Safety information” carefully, and then use the analyzer in the correct way.
The following items are important for safe operation and must be fully observed. These safety precautions
are ranked in 2 levels; “DANGER” and “CAUTION.”
If operation is incorrect, a dangerous situation may occur, resulting in death
DANGER
CAUTION
DANGER
or serious injury.
If operation is incorrect, a dangerous situation may occur, resulting in minor
to medium injuries or only physical damage to equipment.
Installation and transportation
(1) When the analyzer (receiver unit and transmitter unit) is installed on in-
cineration facility, make sure the facility has stopped completely.
Installing in an operating facility may cause high temperature gas injection resulting in burn.
(2) This analyzer is not explosion-proof. Do not use it in an atmosphere of
explosive gas. This may result in serious accidents such as explosion,
fire, etc.
(1) The analyzer should be installed in a place conforming with the installa-
CAUTION
tion requirements noted in this instruction manual, and where the weight
of the analyzer can be endured. Otherwise, it may cause a tip-over, drop,
electric shocks, fire or malfunction of the unit.
(2) Ask professional services or your dealer for installation, transportation,
reinstallation, and associated piping and wiring work. Improper installation may result in a falling accident, electric shock, or injury.
(3) Check the installation site once every 6 months to make sure that the in-
stallation surface is free of rattling. If the instrument is used under insecure installation conditions, a falling accident may occur.
(4) During installation, make sure that the inside of the unit is free from ca-
ble chips and other foreign objects. Otherwise, it may cause fire, failure
or malfunction.
(5) For lifting the gas analyzer, be sure to wear protective gloves. Bare hands
may leave you prone to injury. If the temperature in the installation site is
high, it is imperative to wear leather gloves to prevent burn.
(6) The analyzer is heavy. It should be transported carefully by two or more
persons if manually required. Otherwise, bodily harm may ensue.
- iv -
INZ-TN8ZSS-E
Page 6
Wiring
(1) Be sure to connect a ground wire securely to the specified place by per-
CAUTION
DANGER
forming class D grounding work. Otherwise electric shock or malfunction may result.
(2) If the power supply voltage exceeds the rating, electric shock or damage
to the instrument may result. Be sure to use the instrument within the
specified rating range.
(3) Be sure to turn off the power before performing wiring work.
(4) Be sure to use a 600V-IV ground wire 1.6 mm or larger in diameter with
sufficient dielectric strength.
(5) Select input/output wires of materials and diameter that satisfy the rating
of each device. If a wire which cannot endure the rating is used, electric
shock or fire may occur.
(6) Fasten the input/output wires to the floor or wall, and use a wire protec-
tion device.
Operation
(1) When handling the standard gas such as calibration gas, read the instruc-
tion manual of the standard gas carefully, and use the gas correctly.
(2) When toxic fumes, corrosive gas or inert gas is used as calibration gas,
be sure that the position of the air ventilation or exhaust port is suitable.
Otherwise you may inhale exhaust gas. Furthermore, suffocation, brain
disorder, circulatory deficit, or contraction of the breathing system may
occur, resulting in death.
(1) Do not touch the switch with a wet hand. Otherwise it may cause electric
CAUTION
shock.
(2) Do not operate the laser gas analyzer for a long time with its door left
open. Otherwise, dust, foreign matter, etc. may stick on internal walls,
thereby causing faults.
(3) Do not touch the unit terminal block during operation. Otherwise, it may
cause electric shock or injury.
(4) Before leaving unused for a long time or restarting after left at such a
status for an extended length of time, follow the directions of each in-
struction manual because they are different from normal starting or
shutdown. Otherwise, adequate performance will not be provided. Fur-
thermore, an accident or fault may be caused.
(5) Do not allow water to go into the gas analyzer. Otherwise, electric shock
or fire in the instrument may be caused.
(6) Do not smoke nor use a flame near the gas analyzer. Otherwise, it may
result in a fire.
INZ-TN8ZSS-E
- v -
Page 7
Maintenance and inspection
(1) When the analyzer (receiver unit and transmitter unit) is installed on in-
DANGER
CAUTION
cineration facility, make sure the facility has stopped completely. Installing on the operating facility may cause high temperature gas injection, resulting in burn.
(2) If the analyzer is installed or removed from the location higher than op-
erator’s height, set up a fence to keep someone from approaching under
or near the unit. If the analyzer inadvertently falls off and hits someone,
serious injuries may occur, resulting in death.
(1) Be careful not to drop the analyzer on your foot. Otherwise, it may cause
fracture of the bone.
(2) Do not touch the terminal block of each unit of the instrument carelessly
during operation. Otherwise, it may cause electric shock.
(3) Before working, take off a wrist watch, finger ring or the like metallic
accessories. And never touch the instrument with a wet hand to avoid
electric shocks.
(4) If the fuse is blown, eliminate the cause, and then replace it with the one
of the same capacity and type as before. Otherwise, it may cause electric
shock or accident.
(5) Do not wash or splash water on the switch or electrical parts inside the
device. Otherwise it may cause an electric shock, failure, or fire.
(6) Do not use replacement parts other than recommended ones. Otherwise,
adequate performance will not be provided. Furthermore, an accident or
fault may be caused.
(7) Dispose replacement parts such as maintenance parts as incombustibles
in accordance with the local waste disposal requirements.
Handling of laser equipment
When this product is installed in a flue and as long as no one enters in the
CAUTION
CAUTION
flue, this product can be used as Class 1 laser product. However, be sure to
follow the instructions below for safety because the product emits laser beam
when energized.
- Do not remove the transmitter unit, the receiver unit, or any part of them from
the flue without our permission. Otherwise, it may cause a loss of eyesight
and/or skin lesion. Consult us if there is a need for removal.
-
O2 analyzer for high temperature and instrument air purge version O2 analyzer
use the Class 3B laser elements. When anyone enters into the flue, this product is
regarded as a Class 3B product. In this case, safety measures are required; for
example, an entrance detection system, an interlock that stops radiation of laser
beam, etc. Consult us for details.
Others
If the cause of any fault cannot be determined despite reference to the instruction manual, be sure to contact your dealer or Fuji Electric’s technician
in charge of adjustment. If the instrument is disassembled carelessly, you may
get an electric shock or injury.
- vi -
INZ-TN8ZSS-E
Page 8
WARRANTY AND MAINTENANCE
1. Scope of application
To use this equipment, the following conditions must be met:
• the use of the equipment incurs no risk of a serious accident even if a failure or malfunction occurs on the
equipment, and
• in case of product failure or malfunction, safety measures such as redundant design, prevention of mal-
function, fail safe system, foolproof mechanism are provided outside of the equipment.
Be sure to use this instrument under the conditions or environment mentioned in this instruction manual.
Please consult us for specifications for the following applications:
Radiation-related facilities, systems related to charging or settlement, or other usages which
may have large impact on lives, bodies, property, or other rights or interests.
2. Operating conditions and environment
Refer to “SAFETY INFORMATION” and “4. INSTALLATION".
3. Precautions and prohibitions
Refer to “SAFETY INFORMATION”.
4. Warranty
4-1. Period of warranty
1) Warranty period for this product including accessories is one year after delivery.
2) Warranty period for the parts repaired by our service providers is six months after the completion of repair.
4-2. Scope of warranty
1) If any failure or malfunction attributable to Fuji Electric occurs in the period of warranty, we shall pro-
vide the product after repairing or replacing the faulty part for free of charge at the place of purchase or
delivery.
The warranty does not apply to failure or malfunctions resulting from:
a) inappropriate conditions, environment, handling or usage that is not instructed in a catalog, instruction
book or user's manual, or overuse of the product
b) other devices not manufactured by Fuji Electric
c) improper use, or an alteration or repair that is not performed by Fuji Electric
d) inappropriate maintenance or replacement of expendable parts listed in the instruction book or the
catalog
e) damages incurred during transportation or fall after purchase
f) any reason that Fuji Electric is not responsible for, including a disaster or natural disaster such as
earthquake, thunder, storm and flood damage, or inevitable accident such as abnormal voltage.
INZ-TN8ZSS-E
- vii -
Page 9
2) Regardless of the time period of the occurrence, Fuji Electric is not liable for the damage caused by the
factors Fuji Electric is not responsible for, opportunity loss of the purchaser caused by malfunction of Fu-
ji Electric product, passive damages, damage caused due to special situations regardless of whether it was
foreseeable or not, and secondary damage, accident compensation, damage to products that were not
manufactured by Fuji Electric, and compensation towards other operations.
5. Failure diagnosis
Regardless of the time period of the occurrence, if any failure occurs, the purchaser shall perform a primary
failure diagnosis. However, at the purchaser's request, Fuji Electric shall provide the diagnosis service for a
fee. In such a case, the purchaser shall be charged for the service.
6. Service life
This product, excluding limited-life parts and consumable parts, is designed to have a service life of 10 years
when the average annual ambient temperature is 30°C. To ensure the service life, it is important to perform
planned maintenance of the product including limited-life parts and consumable parts.
7. Maintenance plan
Maintenance can be divided into “preventive maintenance” and “corrective maintenance”. Preventive
maintenance can further classified into “daily inspection” and “periodic inspection”. Preventive maintenance
is achieved through systematic implementation of "daily inspection" and "periodic inspection".
(1) Daily inspection
Be sure to perform daily inspection prior to operation to check for any problem in daily operation. For the
specific items of daily inspection, refer to “7. MAINTENANCE”.
(2) Periodic inspection
Periodic inspection is to replace limited-life parts before their service lives are over, thus preventing failure.
Inspection interval: 6 months to 12 months. If you are using the instrument under harsh environment, we
recommend you to shorten the inspection interval. For the specific items of periodic inspection, refer to “7.
MAINTENANCE”.
(3) Corrective maintenance
Corrective maintenance is a measure to be taken after a trouble has occurred. Refer to “7. MAINTENANCE”
and “8 TROUBLESHOOTING”. If the measures mentioned in this instruction manual do not solve the
problem, please contact our sales office or service office.
- viii -
INZ-TN8ZSS-E
Page 10
8. Limited-life parts and consumable parts
This product contains the following limited-life parts and consumable parts which may affect the service life
of the product itself.
(1) Aluminum electrolytic capacitors
Design life: Design life: 10 years under general working conditions (annual average of ambient tem-
perature: 30°C)
Symptoms when a capacitor loses its capacity: deterioration of power quality, malfunction
Factors which affect battery life: temperature. The life is shortened by half when the temperature rises
by 10°C. (Arrhenius' law)
Replacement: Estimate the lifetime of capacitor according to your operating environment, and have
the capacitor replaced or overhauled at appropriate time, at least once in 10 years. Do not use capaci-
tors beyond its lifetime. Otherwise, electrolyte leakage or depletion may cause odor, smoke, or fire.
Please contact Fuji Electric or its service providers when an overhaul is required.
(2) LCD
Design life: approximately three years for continuous use
Symptoms when depleted: the display may have some kind of problem, or the backlight may not work.
Factors which affect battery life: temperature
The life is shortened by half when the temperature rises by 10°C. (Arrhenius' law)
Replacement: Estimate the lifetime according to your operating environment, and replace it at appro-
priate time.
(3) Backup battery for clock (CR1220)
The clock does not use the battery during the primary power supply.
Replacement: when the total time that no primary power is supplied has exceeded five years.
9. Spare parts and accessories
Refer to “2. CHECKING DELIVERED ITEMS” or “7. MAINTENANCE” for details.
10. Period for repair and provision of spare parts after product discontinuation (maintenance period)
The discontinued models (products) can be repaired for five years from the date of discontinuation. Also,
most spare parts used for repair are provided for five years from the date of discontinuation. However, some
electric parts may not be obtained due to their short life cycle. In this case, repair or provision of spare parts
may be difficult even in the above period.
Please contact Fuji Electric or its service providers for further information.
INZ-TN8ZSS-E
- ix -
Page 11
1. DESCRIPTION
1.1 Introduction
Cross-stack laser gas analyzer (ZSS) provides continuous measurement of HCl in flue gas incineration, NH3
concentration of denitration equipment and heat treat furnace, and O
within a short response time. The cross-stack configuration eliminates the need for transfer of the preparation
measurement gas to the analyzer for proper measurement. Dust resistant construction enables installation
upstream of bug filter units and the application for which injection volume of calcium hydroxide is controlled while measuring HCl concentration. The analyzer adopts near-infrared laser as light source. The analyzer targets only one spectrum line among a large number of absorption spectrum lines, and carry out a
measurement while controlling the temperature and the driving current of the laser. Since the range of wavelengths to be measured is as narrow as a few nanometers, the analyzer receives minimum interference by
other crossovers. For the concentration detection, the modulated intensity of signal amplitude is employed
instead of the amount of change of light.
1.2 Compliance
, CO, and CO2 for combustion control
2
First read this instruction manual carefully, and then make a plan for periodic inspection to perform appropriate maintenance management.
This analyzer uses the invisible infrared laser (excluding O
analyzer). Do not watch the laser beam directly
2
or scattering light.
Laser class: Class 1 (IEC/EN 60825-1)
IP rating : IP65
Measurement category : CAT II
Pollution degree : 2
Altitude : ≤ 2000 m
EC DIRECTIVE COMPLIANCE
LVD (2014/35/EU)
EN 61010-1
EN 62311
EN 60825-1
EMC (2014/30/EU)
EN 61326-1 (Table 2)
EN 55011 (Group 1 Class A)
EN 61000-3-2 (Class A)
EN 61000-3-3
EN 61326-2-3
RoHS (2011/65/EU)
EN 50581
INZ-TN8ZSS-E
- 1 -
Page 12
2. CHECKING DELIVERED ITEMS
Upon receiving the recorder unit, check if the correct quantity of the accessories are supplied.
Separately supplied document are given first priority. When you have purchased or want to purchase spare
parts for 1-year operation or a list of calibration/installation fixtures, refer to “APPENDIX 2”at the end of
this Manual.
Table 2-1 Products Table 2-2 Standard accessories
No. Product name Quantity No. Product name Quantity
1. Control unit 1 1. Bolt (*1) 8 (16)
2. Receiver box 1 2. Nut (*1) 8 (16)
3. Transmitter box 1 3. Spring washer (*1) 8 (16)
4. Angle adjustment unit 2 4. Flat washer (*1) 8 (16)
Cable between receiver unit and
5.
control unit 1 5.
Cable between receiver unit and
6.
transmitter unit
7. Power supply fuse (250 V AC/T1A) 2
8. Instruction manual 1
1 6. Bolt for angle fine adjustment (*2) 6
9.
10.
11.
*1: When the 9th code is “B”, 16 pieces are provided. For other cases, 8 pieces are provided. The length of
the bolts are 55 mm when the 9th code is “A”, and 70 mm when the 9th code is “B”, “C”, or “D”.
Inch-sized bolts are not provided.
*2: The bolts may be delivered being attached to the angle adjustment unit.
Companion flange packing or flange
packing specified for use in high
temperature
Bolt for connecting the receiving unit
and the transmitter unit
Ferrite core (for power cable, outside
the transmitter case)
Ferrite core (for power cable, inside
the transmitter case)
2
12
1
1
- 2 -
INZ-TN8ZSS-E
Page 13
INZ-TN8ZSS-E
- 3 -
Page 14
3. NAME AND EXPLANATION OF EACH PART
3.1 Overall composition
The analyzer consists of 3 units; “Transmitter unit” to transmit the laser, “Receiver unit” to receive light, and
“Control unit” to display and output signals.
- 4 -
INZ-TN8ZSS-E
Page 15
Name Description
(1)
Receiver box Accommodates a photodiode that receives the laser light, PCB, and
others.
(2)
Transmitter box Accommodates the laser element, a Peltier device that controls the
laser temperature, PCB, and others.
(3)
Probe for adjusting the received light intensity
A probe (BNC socket) used to check the received light intensity, in a
form of voltage value, based on which the optical axis is adjusted.
Both the transmitter unit and the receiver unit have a probe.
(4)
Angle adjustment unit Used to adjust the light axis of the laser emitted from the transmitter
unit. The maximum adjustable angle is 5 degree.
(5)
Bolt for angle fine adjustment The longer the optical path length is, the more the light intensity is
affected by angle adjustment. If the stack diameter is large, therefore, use the bolt for angle fine adjustment to set the optical axis.
(6)
Air purge inlet
(7)
Companion flange
(8)
Cable between the receiver
unit and the transmitter unit
(9)
Cable between the receiver
unit and the control unit
(10)
Mounting bracket for the
control unit
(11)
Display
(12)
Cable entry
Inlet for instrument air or N2 gas used for purging. The purging is required
for preventing water condensation and dust contamination on the lens.
*The instrument air used for purging shall not include oil or mist.
*For purging line, use the pipe connectable to a 10/8 mm fitting.
Flanges for connecting the stack and the transmitter unit, and the stack and
the receiver unit.
*The companion flange shall be prepared by customer.
Cable that transmits the electricity to the transmitter unit, and the signals
between the receiver unit and the transmitter unit. You can select the cable
length up to 25 meters.
Cable that transmits the electricity to the receiver unit, and the signals between the receiver unit and the control unit. You can select the cable length
up to 100 meters.
A bracket for attaching the control unit on the wall or the like. It has two
12-mm diameter holes, and a 12-mm width oval hole.
Displays the measurement value and alarm.
10-mm diameter entries for the power cable, signal cable from the receiver
unit, AI, AO, DI, and DO.
INZ-TN8ZSS-E
- 5 -
Page 16
Input / output terminal
3.2 Wiring diagram
- 6 -
INZ-TN8ZSS-E
Page 17
INZ-TN8ZSS-E
- 7 -
Page 18
4. INSTALLATION
This analyzer is not explosion-proof. Do not use it in an atmosphere of ex-
DANGER
CAUTION
plosive gas. Otherwise, it can result in serious accidents such as explosion,
fire, etc.
The analyzer should be installed in a place conforming with the installa-
tion requirements noted in this instruction manual. Otherwise, it may
cause toppling, dropping, electric shocks, fire or malfunction of the unit.
Request assistance from the professionals or the vendors when mount-
ing, moving, re-mounting and carrying out piping and wiring works associated with these activities. A poor installation may cause accidental
tip over, electric shock, injury, etc.
During installation, make sure that the inside of the unit is free from ca-
ble chips and other foreign objects. Otherwise, it may cause fire accident
or malfunction.
For lifting the analyzer, be sure to wear protective gloves. Bare hands
may leave you prone to an injury. If the temperature in the installation
location is high, be sure to wear leather gloves. Otherwise, you may suffer a burn.
The analyzer is heavy. It should be transported carefully by two or more
persons if manually required. Otherwise, bodily harm may ensue.
- 8 -
INZ-TN8ZSS-E
Page 19
4.1 Requirements
4.1.1 Installation conditions of receiver unit and transmitter unit
4.1.1
Select a location that meets the following conditions.
(1) Ambient temperature : A place where the temperature is within -20 to 55°C and there is no
(2) Ambient humidity : A place where the humidity is 90% RH or lower not subjected to
(3) Measured gas temperature : Refer to “1-1 (4) Measurable component and range” in APPENDIX
(4) Measured gas pressure : ±10kPa
(5) Measured gas moisture : 50vol% or less (no condensation)
(6) Measured gas velocity : 25m/s or less (However, consultation is necessary for the environ-
(7) Dust : 15 g/Nm
(8) Companion flange : Prepare the flange diameter selected by the 9th digit of the code
(9) Air purge : Prepare the instrumentation air containing no oil or water. If there is
(10) Air purge flow rate : 20L/min or more (depending on measured gas temperature, velocity,
(11) Path lengths (diameter) : 0.5 to 10 m (0.5 to 5 m for CO+O
(12) Vibration : 0.5G or less (0.2G or less for frequency 20 to 40Hz)
(13) A place with less corrosive gases
(14) A place accessible for maintenance and check
(15) A place with less electrically induced disturbances such as high electric currents or sparks in the sur-
(16) Light axis fluctuation: ≤ 0.3°
Notes:
Receiver unit and transmitter unit
sudden temperature change.
condensation.
1. A place where the stack is not distorted or laser light axis is not
deflected by sudden temperature change.
ment where dust (1g/Nm
ists.) (Prevention of dust deposition or dew condensation due to increase of air purge flow rate is required)
components, device specification, optical light path, particle diameter, and so on)
symbols. When the purchase specification is provided, prepare the
flange diameter described in the specification.
a possibility of containing oil or water, install an oil filter or a mist
filter. When the instrumentation air cannot be supplied, install a
compressor. Prepare N
cluding air purge version (4th digit of code symbol is T or V).
pressure, moisture or dust) (One-side air purge flow rate (L/min) ≥
Gas velocity (m/s) × 10)
(When the optical path length is 1m or less)
(Refer to “4.2 Mounting dimensions”.)
rounding.
• When there is a possibility of deflecting the laser light axis or reducing the light intensity due to the
large vibration acceleration, contact the manufacturer before installation.
• When “Box Temperature Warning” is occurred under the influence of gas temperature, mount the reducer to keep the receiver box and the transmitter box away from the stack.
• When you measure the high temperature gas of 500°C or more, install a pressure sensor on the purge
line to monitor the purging status. Measuring the high temperature gas without air purge may cause
damage to the analyzer.
• Be sure to keep purging the analyzer once you installed it. Purging is required even you suspend the
operation. Leaving the analyzer without purging causes permanent contamination which may result in
a malfunction or error.
3
or less (depending on the conditions such as measurable
3
or more) or water (25vol% or more) ex-
for O2 analyzer and CO+O2 analyzer ex-
2
analyzer)
2
INZ-TN8ZSS-E
- 9 -
Page 20
4.1.2 Control unit
Select a location that meets the following conditions.
(1) Ambient temperature : –5 to 45°C
(2) Ambient humidity : 90 % RH or less
(3) Power supply : Rated voltage : 100V to 240V AC
Rated frequency : 50Hz/60Hz
(4) A place where the instrument receives no vibration
(5) A place where the ambient air is clean
(6) A place where has enough space for maintenance work or inspection
(7) A place where induced electrical noise, such as large electrical current or spark, is low
(8) Install a breaker that meets the requirements of IEC60947-1 and IEC60947-3 in the facility in which
the analyzer is placed.
Notes:
• The breaker should be installed close enough to the analyzer so that an operator can use it without difficulty.
• The breaker should be identified as the one for this product.
• The breaker must meet the rating of the analyzer.
- 10 -
INZ-TN8ZSS-E
Page 21
4.2 Mounting dimensions
Keep purging the analyzer once you install it on a stack.
Otherwise, the contamination of optical surface may result in the analyzer failure.
Fig. 4-1
*1: When the flue gas contains a large amount of dust, ensure the minimum length of 50 mm.
*2: When the flue gas temperature is high (400°C or more), ensure the minimum distance of 400 mm.
Note that in that case you have to be even more careful in adjusting the angle of the companion flange
because the angle range
within which the companion flange can be adjusted is narrow.
*3: Make sure to install the analyzer in the place where it is easy to operate, and where there is enough
clearance.
*4: Scaffold is required for both the receiver unit side and the transmitter unit side.
*5: Use a flanged valve where there is a risk of gas blowout or where the operator may be exposed to a
dangerous situation.
*6: Tightening torque for the companion flange shall be 118±14 N·m. When tightening the bolts of the
flange or the angle adjustment unit, apply grease to the bolts. (Recommended grease is the one contains
molybdenum.)
INZ-TN8ZSS-E
- 11 -
Page 22
4.3 Mounting range of companion flange
Mount the companion flange so that it satisfies the conditions of the following figure (Fig. 4-2). If the conditions below are not satisfied, light cannot be received even if the light axis adjustment is performed by the
angle adjustment unit. In such a case, mount the companion flange again.
When θ (angle determined by laser light source and flange diameter) is bigger than 5°, mount the companion
flange within the angle γ (≤ 5°) inside the circle with radius A.
When the distance between laser light source and stack is long, or θ angle is less than 5° because flange diameter is small, both radius A and angle γ will become smaller, and the mounting conditions will be strict.
Fig. 4-2
4.4 Items required for adjustment test
It is recommended to prepare the following items for the adjustment test before installation.
(1) Cable between receiver unit and control unit (for calibration)
(2) Cable between receiver unit and transmitter unit (for calibration)
(3) Calibration gas cell
(4) Power supply drum (or power supply extension cable)
(5) Digital multimeter
(6) BNC for light axis adjustment (ZZP*ZSSTQ505298)
(7) Flow meter (about 2L/min)
(8) N
(9) Gas cylinder corresponding to the span (80 to 100% of span)
(10) Regulator (for gas cylinder) (For HCl/NH
(11) 10/8 mm PTFE tube, 10 m (we recommend to prepare the tube with some extra length)
(12) Rc1/4 × 10/8 joint, two or more
(13) Tools (2 spanners, measure, cutter, Phillips screwdriver, flat-blade driver, hexagonal wrench, tube
(14) Plastic sheet (used to prevent the drop of parts and/or tools)
Gas cylinder
2
meter, prepare a stainless steel regulator. If using a brass
3
regulator, use it only once and then discard it. A brass regulator may get rusty inside, and the rust
absorbs gas to disturb accurate measurement.)
cutter)
- 12 -
INZ-TN8ZSS-E
Page 23
4.5 Setup procedure
Page
(1)
Installation site check
Check that the transmitter unit, the receiving unit, and the control unit are installed in a
location that meets the requirements. (“4. INSTALLATION”)
↓
(2)
Installation dimension check
Check that the transmitter unit and the receiver unit are installed in a manner that meets
the dimension requirements. (“4.2 Mounting dimensions”)
↓
(3)
Flange position check
Check that the companion flange for the transmitter unit and for the receiver unit are
installed in positions that meet the requirements. (“4.3 Mounting range of companion
flange”)
↓
(4)
Purging line check
Check that one 10/8 mm tube for purging, equipped with a flowmeter or a flow regulator, is prepared for each of the transmitter unit and the receiver unit.
↓
(5)
Power supply and RS-485 wiring
Check that the power supply of rated voltage 100 to 240V AC ±10%, and rated frequency 50/60Hz is prepared at the installation location for the control unit. Check the
wiring for RS-485 communication.
↓
(6)
Zero calibration
*If the analyzer has been turned off for a long time, warm-up the transmitter for about
90 minutes before calibration. (“6.1 Zero calibration”).
↓
(7)
Light intensity check
Take a memo of the output value from the probe for adjusting the received light intensity
during zero calibration. (“4.6 Received light ”)
↓
(8)
Span calibration
*If you feed any corrosive gas such as hydrogen chloride and carbon oxide, use the ex-
haust gas tube long enough and at an adequate position so that no one breathe in the
gas.
*Flow the zero gas after span calibration. (“6.2 Span calibration”)
↓
(9)
Angle adjustment
Attach the angle adjustment unit on each companion flange. Adjust the angle of the
transmitter unit and the receiver unit. (“4.7.3 Adjustment procedure”)
↓
(10)
Installation of purging equipment
(“4.8 Piping system diagram”)
↓
(11)
Installation of the transmitter box and the receiver box
(“4.9 Installation of the transmitter box and the receiver box”)
↓
P. 8
P. 11
P. 12
P. 19
P.26, 28
P.42 to 45
P. 15
P.46, 48
P.16, 18
P. 19
P. 20
INZ-TN8ZSS-E
- 13 -
Page 24
(12)
Cable connection
Connect the cable between the transmitter unit and the receiver unit, and the cable between the receiver unit and the control unit. (“4.10 Cable connection”)
↓
(13)
Turning on the control unit
↓
(14)
Light intensity adjustment
Adjust the light intensity to as close to the value on nameplate as possible.
*This may be difficult when you adjust the angle while the furnace is in operation.
↓
(15)
Optical path length setting
Enter the optical path length on the control panel. (“6.5 Parameter setting”)
↓
(16)
Analog signal wiring
↓
(17)
Output item setting
Select the items to be transmitted. (“6.8 Analog output”)
↓
(18)
Analog signal adjustment
Make a analog output setting in reference to “6.9 Fine adjustment of analog output value”.
↓
(19)
Analog input setting
If you use external sensors for gas pressure, gas temperature, gas flow rate, O
H
O, make necessary settings for them. For the items with no external sensor, set the
2
parameter to fixed value. Gas temperature, O
, and HCl have to be properly set for
2
, and /or
2
measurement.
If there is a difference between the gas temperature during furnace suspension and the
fixed gas temperature, measurement error may occur. (“6.7 Analog input”)
↓
(20)
Other parameters setting
↓
(21)
Digital signal wiring
↓
(22)
Analog signal adjustment (as needed)
↓
(23)
Alarm output check (as needed)
P.21, 22
P.63 to 67
P. 28
P. 63
P.63 to 67
P.63 to 67
P. 83
P. 28
P. 84
P. 87
- 14 -
INZ-TN8ZSS-E
Page 25
4.6 Received light intensity check
(1) Remove the cap from the probe (BNC socket) of the receiver unit, and connect the probe and the digital
voltmeter using the BNC cable for light axis adjustment.
(2) Read the DC voltage value. The receiver unit emits a voltage about 3 V DC when the light transmit-
tance is 100%.
(3) The voltage value during zero gas is supplied with the calibration cell connected becomes the reference
light intensity for light axis adjustment which is performed after the equipment is attached to a companion flange (Factory-set voltage is described on a nameplate on the covers of the receiver unit,
transmitter unit, and control unit).
Note 1) CO + O
Note 2) Before connecting or removing the BNC cable, make sure that no static electricity has built up on the
cable. If any, discharge static electricity.
• Single laser version
analyzer has BNC sockets for each of CO and O2.
2
• Dual laser version (CO + O
analyzer)
2
INZ-TN8ZSS-E
Fig. 4-3
- 15 -
Page 26
4.7 Angle adjustment
Fig. 4-4
4.7.1 How to use the angle adjustment unit 1
(1) When you tighten the flange B
Turn the fixing bolt and the anchor nut clockwise at the same time.
If it is difficult to turn them, slightly loosen the anchor nut first.
(approx. one-tenth rotation)
(2) When you loosen the flange B
Turn the fixing bolt and the anchor nut counter-clockwise at the
same time. If it is difficult to turn them, slightly loosen the anchor
nut first.
(approx. one-tenth rotation)
(3) When you fix the flange B
When angle is determined after adjustment in 1) or 2), turn them in
the direction opposite to each other to fasten them. Be careful because if the fixing bolt and the anchor nut are too far away from the
flange B, the angle will slip.
Note) Pay attention to the insertion angle of the tool. Take utmost
care not to turn at the angle which may let the bellows contact
with the head of the spanner. Otherwise, the spanner may crush
the bellows.
Fig. 4-5
Fig. 4-6
Fig. 4-7
- 16 -
INZ-TN8ZSS-E
Page 27
4.7.2 How to use the angle adjustment unit 2
This section describes movement of the optical axis by operating the angle adjustment unit.
Example) When you adjust the fixing bolt and the anchor nut shown in the figure below, encircled with a
solid line
(1) If the flange A and the flange B are parallel, the laser beam points horizontal direction.
(2) If you screw up the bolt shown in the below figure, the place where the bolt is screwed up moves to-
wards the flange A, and the laser beam faces downward.
(3) If you loosen the fixing bolt as the figure below, the place where the bolt is loosen aparts from the
flange A, and the laser beam point upward.
In the same way, if you screw up the bolt and nut colored gray in the below figure, the light axis moves as
・
follows.
INZ-TN8ZSS-E
- 17 -
Page 28
4.7.3 Adjustment procedure
4.7.3.1 When an optical axis adjusting tool (laser pointer) is used
Do not watch the laser pointer beam directly with the optical measuring
CAUTION
(1) When the flanges A and B shown in Fig. 4-4 are extremely tilted, adjust them in a flat place before
mounting them to the stack so that they are parallel to each other. Refer to “4.7.1 How to use the angle
adjustment unit 1”.
(2) Mount the angle adjustment unit to the companion flange on the stack.
- Make sure the fixing bolts come to the positions shown in Fig. 4-8.
- Mount the angle adjustment unit so that the companion flange and the flanges of the angle adjustment
unit are as concentric as possible.
- Do not mount the transmitter box and the receiver box yet.
(3) With the hexagon socket head bolts, attach the laser pointer to the angle adjustment unit of the trans-
mitter side, and the laser scope on the angle adjustment unit of the receiver side. (The laser pointer has
two kinds of holes at different distance from the center. Fix it with three bolts.)
(4) Emit light from the laser pointer, and adjust the angle adjustment unit referring to “4.7.2 How to use the
angle adjustment unit 2” so that the pointer’s light comes the center of the target.
(5) Using the attached hexagonal socket screws, fix the laser pointer to the angle adjustment unit on the
receiver unit, and the laser scope to the angle adjustment unit on the transmitter unit.
(6) Adjust the optical axis in the same way as the step 4.
(7) When you have finished adjustment, retighten the fixing bolts and nuts lightly. Take care not to move
the light axis. Remove the laser pointer and the target.
4.7.3.2 When an optional optical axis adjusting tool (laser pointer) is not used
(1) When the flanges A and B shown in Fig. 4-4 are extremely tilted, adjust them in a flat place before
mounting them to the stack so that they are parallel to each other. Refer to “4.7.1 How to use the angle
adjustment unit 1”.
(2) Attach the angle adjustment unit to the companion flange in a manner that:
- Make sure the fixing bolts come to the positions shown in Fig. 4-8.
- Mount the angle adjustment unit so that the companion flange and the flanges of the angle adjustment
unit are as concentric as possible.
device. Otherwise, it may cause serious damage to your eyes.
Fig. 4-8
(3) Mount the transmitter box and the receiver box in reference to 4.9 Installation of the transmitter box
and the receiver box.
(4) Adjust the light intensity, referring to “4.11.2 When an optional optical axis adjusting tool (laser point-
er) is not used”.
- 18 -
INZ-TN8ZSS-E
Page 29
4.8 Piping system diagram
Fig. 4-9
INZ-TN8ZSS-E
- 19 -
Page 30
4.9 Installation of the transmitter box and the receiver box
(1) Prepare “Receiver box or transmitter box” and “Hexagonal socket bolt” as shown in the following fig-
ure (Fig. 4-10). Check that the O-ring is mounted on near the lens (refer to Figure below) of the receiver unit or the transmitter unit
(2) Mount “Receiver box or transmitter box” on “Angle adjustment unit” so that the socket is positioned
bottom. (Be careful not to touch the lenses of the transmitter box and the receiver box during installation.)
(3) Fix it with the 6 “hexagonal socket bolts”.
Fig. 4-10
- 20 -
INZ-TN8ZSS-E
Page 31
4.10 Cable connection
4.10.1 Between transmitter unit and receiver unit
The receiver unit and the transmitter unit are connected with the “Cable between receiver unit and transmitter unit”.
Both ends of it are fitted with a female 16-pin connector (waterproof type). The connector has no polarity.
Fix the Cable between receiver unit and transmitter unit to the stack, etc. to prevent the light axis from deflecting by its own weight.
Fig. 4-11
Fig. 4-12
INZ-TN8ZSS-E
- 21 -
Page 32
4.10.2 Between receiver unit and control unit
The receiver unit and the control unit are connected with the “Cable between receiver unit and control unit”.
Both ends of it are fitted with a male 10-pin connector (waterproof type). The connector has no polarity.
Perform wiring in the way that the cable between receiver unit and control unit is not pulled.
Fig. 4-13
- 22 -
INZ-TN8ZSS-E
Page 33
4.11 Light intensity adjustment
Note that if you loosen multiple bolts and nuts at a time, it takes considerable time to re-adjust the angle.
4.11.1 When an optional optical axis adjusting tool (laser
pointer) is used
(1) Connect the digital voltmeter using the BNC cable, referring to “4.6 Received light ”.
(2) The reference light intensity is based on the output in zero calibration or voltage value of light intensity
which is described in a nameplate.
If the output is not as much as that measured at the time of calibration before the adjustment, find a position where the received light intensity increases, moving the transmitter box from side to side and up
and down slowly. (*1)(*2)(*3)
e.g.) In the case that the light intensity increases when the
whole transmitter box is moved upward.
As the light intensity increases when the light axis turns
→
downward, tighten the fixing bolt in reference to “4.7.2
How to use the angle adjustment unit 2”.
(3) If the output voltage is less than the value obtained during a zero calibration despite your adjustment of
the transmitter box to make the output maximum, fix the transmitter box and then move the receiver
box to adjust the optical axis.
Do not loosen more than one bolt or loosen fixing bolt and nut widely in both cases of adjustment.
Light axis may deflect again.
(4) Retighten all volts to fix them (If the light intensity decreases at retightening, adjust again).
*1) Received light intensity might not increase to its maximum that was confirmed by “4.6 Received light ”
due to the influences from dust and moisture in the stack.
*2) When using the high-speed/AGC version (i.e. the 22th digit of code symbols is "H"), adjust the light axis
under the status that the level of AGC is "01" (refer to "5.3 Outline of screen ").
*3) When using CO + O
be maximum.
Carry out the adjustment of the O
reverse way.
analyzer, check the received light intensity at two points, and then adjust them to
2
analyzer first and the CO analyzer next because it is easier than the
2
INZ-TN8ZSS-E
- 23 -
Page 34
4.11.2 When an optional optical axis adjusting tool (laser
pointer) is not used
(1) Connect the digital voltmeter using the BNC cable, referring to “4.6 Received light ”.
(2) The reference light intensity is based on the output in zero calibration or voltage value of light intensity
which is described in a nameplate.
When the light intensity is completely zero V, move the transmitter box from side to side and up and
down slowly to find the position where the received light intensity increases. If you cannot still confirm
the light intensity, loosen the fixing bolts and nuts further, and check reaction of the digital voltmeter
while moving the box back and forth slowly.
(3) If you find reaction of light intensity increases, fasten the fixing bolts and nuts temporally, and perform
the operation in 2). (*4)
(4) If output is not as much as that measured at the time of calibration even if output is at its maximum on
the transmitter side, fix the box at the maximum output angle temporality and adjust the light axis,
moving the box on the receiver side similarly. When you can confirm the output at the time of adjusting
any of the boxes, do not loosen the fixing bolts and nuts too much, or the light axis can deflect further.
Do not loosen more than one bolt or loosen fixing bolt and nut widely in both cases of adjustment.
Light axis may deflect again.
(5) Move the transmitter unit and the receiver unit from side to side and up and down to adjust until the
voltage value of light intensity of digital multimeter becomes smaller in all directions.
(6) Retighten all volts to fix them (If the light intensity decreases at retightening, adjust again).
*4) Received light intensity might not increase to its maximum that was confirmed by “4.6 Received light ”
due to the influences from dust and moisture in the stack.
- 24 -
INZ-TN8ZSS-E
Page 35
4.12 How to use the bolt for angle fine adjustment
(standard accessory)
This bolt is used for adjustment when the light path length in the stack is long and when the light intensity
misses too much at the time of retightening in the procedure in “4.11 Light intensity adjustment”.
When each retightening is finished, perform fine adjustment by pressing with this bolt.
When you use the fine adjustment screw, set it to the flange A before mounting the receiver and the transmitter boxes on the angle adjustment unit. The fine adjustment screw cannot be set after mounting these boxes.
Fig. 4-14
INZ-TN8ZSS-E
- 25 -
Page 36
4.13 Connecting to control unit
Do not supply the power until all the wiring work is completed.
To ensure the watertight property of the cable glands, use the cables that have an outer diameter from 6 mm
to 10 mm.
4.13.1 AC power connection
AC power connecting terminal is positioned at the lower left of the control unit (see Fig. 4-15.). Use the
AC cable that has a flame resistance of 600 V, a cross-sectional area of 1.25 mm
2
(AWG16) or more, and an
outer diameter between 6 mm through 10 mm. Connect the grounding wire to the protective earth terminal.
When connecting the AC cable to the input terminal, the ground cable should be longer than the L, N line.
The cable length should be adjusted so that when the AC cable is pulled from the external, the L, N line will
be removed from the terminal block first, and the ground cable will be removed last.
Add a ferrite core both inside and outside of the case. If you install the control unit outside, we recommend
taking a measure to protect the ferrite core attached outside the case from deterioration; for example, cover it
with plastic tape.
Fig. 4-15
- 26 -
INZ-TN8ZSS-E
Page 37
When noise source is in the vicinity
Avoid installing this analyzer near an electrical apparatus which produces power source noise. (Such
as high frequency furnace, electric welder, etc.) If
the analyzer must be used near such equipment, a
separate power line should be used for avoiding
noise.
In case noise may enter from a relay, solenoid
valve, etc. through power supply, connect a varistor
(e.g. ENA211 OKAYA) to the noise source as
shown in Fig. 4-16. If the varistor is located away
from the noise source, no effect is obtainable.
Fig. 4-16
CAUTION
• For power line, use the cables that have a cross-sectional area of 1.25 mm2 or more.
• Connect the power cable to the socket with the ground slot (class D grounding). For grounding line, use
the cable that has a cross-sectional area of 2 mm
2
or more.
• For signal lines, use a shielded wire in order to suppress the influences by noise.
• For the digital contacts and the digital input lines, use cables that have a cross-sectional area of 0.12
2
mm
(AWG26) or more.
• After the AC power cable is connected, put a power terminal cover on.
• Do not install the instrument near objects which considerably disturb power waveforms. Do not share
their power supplies either. Otherwise, it may cause a display error.
• Power supply and output signal lines should be separated from each other.
• Use M4 solderless terminals for terminal connection.
• For the power line and the grounding line, use solderless terminals of the type that core wires and
shield lines are crimped separately.
INZ-TN8ZSS-E
- 27 -
Page 38
4.13.2 Connecting analog input/output
Connect the analog input and output properly, referring to “3.2 Wiring diagram”. According to your order, 4–
20 mA DC or 1–5 V DC analog signal is emitted. For connection of the signal cable, use an insulated converter (WS2DC Series, manufactured by Fuji Electric Technica Co., Ltd. or equivalent) to supress the influences by noise.
Analog input signal shall be 4 to 20mA DC.
Note) If you connect the analog input to the analog output terminal, the printed circuit board may be dam-
aged. Check the terminal before connecting them.
4.13.3 Connecting digital input/output
Connect the contact input and output properly, referring to “3.2 Wiring diagram”. If it is provided with the
separately submitted approved drawing, connect the contact input/output as shown in the drawing. For wiring, use shielded wires of which cross-sectional area are 0.12 mm
2
(AWG26) or more.
4.13.4 RS-485 connection
(1) Open the cover of the control unit. Remove the two screws shown in the below figure.
Fig. 4-17
- 28 -
INZ-TN8ZSS-E
Page 39
(2) Open the control panel, and you can see the terminal box on the right side of the printed board.
Fig. 4-18
INZ-TN8ZSS-E
- 29 -
Page 40
(3) Connect the cables as shown in the figure 4-19.
Cable cross-sectional area shall be between 0.2 mm
2
(AWG24) and 0.12 mm2 (AWG26).
Terminate both ends of a cable (between ② and ③) by using 100Ω (≥ 1/2W) resistors.
Put the terminal cover back on after you finished wiring.
Do not use an unassigned terminal as a relay terminal.
Fig. 4-19
Terminal No. Signal
1 Ground
2 RTxD+
3 RTxD-
4 NC
- 30 -
INZ-TN8ZSS-E
Page 41
5. OPERATION PANEL AND SCREEN
5.1 Operation panel
Name Functions Name Functions
(1) Mode key Used to display the menu
mode.
(2) Escape key Used to return to a previous
screen or cancel the setting in
midway.
(3) Up key Used to move the cursor,
change the selected item and
increase numeral value.
(4) Down key Used to move the cursor,
change the selected item and
decrease numeral value.
(5) Side key Used to move the cursor and
change numeral digit.
(6) Entry key Used for confirmation of se-
lected items or values, and for
execution of calibration.
INZ-TN8ZSS-E
- 31 -
Page 42
5.2 Screen configuration
- 32 -
INZ-TN8ZSS-E
Page 43
5.3 Outline of screen
5.3.1 “Measurement” screen
(appears when the power is turned ON)
On the measurement screen, the display shows the components being measured, alarm, reference-O
DOWN or ENT to move up, down, or across the screen.
INZ-TN8ZSS-E
-corrected concentration, and analog input. If there are more than four items to be displayed, use UP,
2
- 33 -
Page 44
5.3.1.1 Name (functions)
(1) Measured component ······ Displays the gas component to be measured in molecular formula.
(2) Concentration value ······ Displays the measured concentration. The value is reversed during the
value is held and during automatic calibration of laser waveform is in
operation.
(3) Unit ······ Displays the unit of concentration such as ppm, vol%, etc.
(4) Instantaneous ······
value / Average value
Indicates whether the displayed concentration value is instantaneous
value or average value. (The reversed value is displayed.)
(5) Wet / Dry ······ Indicates whether the displayed concentration is wet base or dry base.
(The reversed value is displayed.)
H
O is fixed to “Wet”.
2
(6) Range display ······ Displays the current full scale range.
(7) Alarm ······ Displays all the alarm. When more than one alarm has occurred, dis-
play is switched by 3 seconds.
(8) O
conversion value ······ If components of the HCl meter is provided with O2 conversion output
2
when ordering, it displays O
conversion. “Conversion**” is displayed
2
as “Conversion HCl” on the display area. For changeover of instanta-
neous value/average value or wet/dry, display and analog output can
be performed independent of the original conversion on “Display /
Output” screen. For the contents of O
conversion, refer to “5.3.1.2 O2
2
conversion concentration value”.
(9) Analog input ······ Displays the analog input value which was set at the “Analog Input”
screen. Analog input to be displayed are “Temperature”, “Pressure”,
“Velocity”, “O
” and “H2O”. They are not displayed when selecting
2
fixed value is selected.
(10) Analog input range ······ Displays the setting range of 4 to 20mA DC which was set at the “An-
alog Input” screen.
(11) Analog input value ······ Displays the analog input value corresponding to the setting range of 4
to 20mA DC which was set at the “Analog Input” screen.
(12) AGC level ······ Displays current level of AGC. (when using High-speed AGC version
i.e. the 22th digit of type code is "H", or CO + O
meter i.e. the 4th
2
digit is "V", "U", or "S")
(13) Light transmission ······ displays the percentage of the light transmitted through gas.
5.3.1.2 O2 conversion concentration value
O2 conversion concentration value is calculated from the following equation of measured component (Cs),
instantaneous concentration of O
Conversion output =
● On: Oxygen conversion reference value (%)
● Os: Oxygen concentration (%)
● Cs: Gas concentration for target component
If you want to change the oxygen conversion reference value and limit value to
other than default value, give instructions before delivery or contact our technical
service representative.
and O2 correction reference value.
2
21 – On
21 – Os
× Cs
(Value that is set according to application: default value 12%)
(O
analog input value or fixed value that is set at the “Analog Input” screen.
2
In the case O
input exceeds the limit value, calculate from the limit value.
2
(Default value of the limit value: 20%.))
- 34 -
INZ-TN8ZSS-E
Page 45
5.3.2 “Menu” screen
Press the key while the “Measurement” screen is displayed, and the “Menu” screen appears. (It does
MODE
not appear when pressing the key while other screen is displayed. In that case, Press the key to
display the “Measurement” screen, and then press the key.)
MODE
MODE
-5+
5.3.2.1 Name (functions)
(1) Zero Calibration ············ Used for performing zero calibration.
(2) Span Calibration ············ Used for performing span calibration.
(3) Alarm Log ··················· Displays the alarm occurred in the past.
(4) Alarm Setting ··············· Used for setting range of upper/lower limit alarm or analog output range
of the measurement value.
(5) Output Hold ················· Used for holding analog output.
(6) Parameter Setting ·········· Used for setting each parameter.
(7) Display / Output ············ Used for setting the measurement value to be displayed on the “Meas-
urement” screen such as switching “Instantaneous value / Average val-
ue” or “Wet / Dry”.
(8) Input Signal ················· Used when analog input range or fixed value are set for concentration
correction or air purge alarm.
(9) AO Select ··················· Used to determine what to output to the analog output terminal.
(10) Analyzer Information ······ Display the calculation information of concentration.
(It is user for trouble analysis)
INZ-TN8ZSS-E
- 35 -
Page 46
5.3.3 “Alarm Record” screen
The screen displays the alarm record occurred in the past. The ten newest errors are logged. The oldest error
will be deleted one by one every time a new alarm occurs. New errors are displayed from the top on the
screen.
It displays the date, time and component when an alarm occurred, alarm contents, recovery date, and recovery time from the left to right.
If you turn off the power in a state that the alarm is generated, the highlighted time and date when the power
was turned off are displayed for “Alarm recovery time” and “Alarm recovery date”.
All the alarms except “Connection Error” which indicates communication failure between the receiver unit
and the control unit will be activated after at least five minutes from the power off. The alarm of “Low Light
Transmission” is activated after 1 minute continuation. Though it takes 6 minutes from just after the power
on.
5.3.3.1 Name (functions)
(1) Alarm occurrence date ············ Displays the date when device failure, high gas temperature or
alarm occurred.
(2) Alarm occurrence time ············ Displays the time when alarm occurred.
(3) Alarm occurrence component ···· Displays the component and analog input for which alarm oc-
curred.
(4) Alarm contents ····················· Displays the contents of alarm.
(5) Alarm recovery date ··············· Displays the date when alarm is recovered. Nothing is displayed
for the alarm which is not recovered.
(6) Alarm recovery time ·············· Displays the time when alarm is recovered. Nothing is displayed
for the alarm which is not recovered.
(7) Select Delete key ·················· Deletes the selected alarm.
(8) All Delete key ······················ Deletes all alarms.
- 36 -
INZ-TN8ZSS-E
Page 47
r
r
r
5.3.3.2 Basic Operation
Moving the cursor
Alarm Record
2018-04-01
12:00:00
Record page change
Date Day
04
01 11:23
/
03
29 23:14
/
03
25 15:23
/
03
20 04:41
/
02
18
00:08
/
01
12
/
11:22
Select Delete
HC
HC
Press.
Air
Tem p.
HC
Alarm
Over H - Limit
ℓ
LD Temp. Error
ℓ
AI Unde
Low Air Purge
High Gas Temp.
Low Light Trans.
ℓ
Alarm Record
Date Day
04
/
03
/
03
/
03
/
02
/
01
/
01 11:23
29 23:14
25 15:23
20 04:41
18
00:08
12
11:22
Select Delete
HC
HC
Press.
Air
Tem p.
HC
Alarm Record
Alarm
Over H - Limit
ℓ
LD Temp. Error
ℓ
AI Unde
Low Air Purge
High Gas Temp.
Low Light Trans.
ℓ
All Delete
All Delete
Day Repair
31 03:23
28 13:28
20
09:44
27
08:11
2018-04-01
12:00:00
Day Repair
31 03:23
28 13:28
20
09:44
27
08:11
2018-04-01
12:00:00
The cursor is moved.
Date Day
01 11:23
04
/
03
29 23:14
/
03
25 15:23
/
03
20 04:41
/
02
18
00:08
/
01
12
/
11:22
Select Delete
HC
HC
Press.
Air
Tem p.
HC
Alarm
Over H - Limit
ℓ
LD Temp. Error
ℓ
AI Unde
Low Air Purge
High Gas Temp.
Low Light Trans.
ℓ
All Delete
Day Repair
31 03:23
28 13:28
20
09:44
27
08:11
Move the cursor to the bottom and press the
Alarm Record
Date Day
21 17:56
12
/
HC
Alarm
Low Light Trans.
ℓ
2018-04-01
12:00:00
Day Repair
22 11:49
key, and the record page is changed.
Select Delete
All Delete
INZ-TN8ZSS-E
- 37 -
Page 48
r
r
Selected alarm deletion
Alarm Record
2018-04-01
12:00:00
Date Day
04
/
03
/
03
/
03
/
02
/
01
/
ENT
01
11:23
HC
23:14
29
25
20
18
12
15:23
04:41
00:08
11:22
Select Delete
HC
Press.
Air
Tem p.
HC
Alarm
Over H -Limit
ℓ
LD Temp. Error
ℓ
AI Unde
Low Air Purge
High Gas Temp.
Low Light Trans.
ℓ
Alarm Record
Date Day
01
04
11:23
/
29
03
23:14
/
03
15:23
25
/
20
03
04:41
/
02
18
00:08
/
01
12
/
11:22
Select Delete
HC
HC
Press.
Air
Tem p.
HC
ENT
Alarm
Over H -Limit
ℓ
LD Temp. Error
ℓ
AI Unde
Low Air Purge
High Gas Temp.
Low Light Trans.
ℓ
Alarm Record
All Delete
All Delete
Day Repair
31 03:23
28 13:28
20 09:44
08:11
27
-5+
2018-04-01
12:00:00
Day Repair
31 03:23
28 13:28
20 09:44
27
08:11
2018-04-01
12:00:00
Move the cursor to the alarm to be deleted
by pressing the or the key.
Press the key, and the “Select Delete”
ENT
is reversed.
Date Day
01
04
11:23
/
29
03
23:14
/
20
03
04:41
/
18
02
00:08
/
12
01
11:22
/
12
21
/
17:56
Select Delete
HC
HC
Air.
HC
HC
Alarm
Over H -Limit
ℓ
LD Temp. Error
ℓ
Low Air Purge
High Gas Temp.Tem p.
Low Light Trans.
ℓ
Low Light Trans.
ℓ
All Delete
Day Repair
31 03:23
20 09:44
08:11
27
22 11:49
The alarm aligned with the cursor is deleted
by pressing the key, and old alarm is
ENT
shifted up.
- 38 -
INZ-TN8ZSS-E
Page 49
r
r
r
r
r
r
All alarms deletion
Alarm Record
2018-04-01
12:00:00
DateDay
04/01 11:23
03/29 23:14
03/25 15:23
03/20 04:41
02
/
01
/
ENT
HC
HC
Press.
18
12
Tem p.
00:08
11:22
HC
Select Delete
Alarm
Over H -Limit
ℓ
LD Temp. Error
ℓ
AI Unde
Low Air PurgeAi
High Gas Temp.
Low Light Trans.
ℓ
Alarm Record
DateDay
04/01 11:23
03/29 23:14
03/25 15:23
03/20 04:41
18
02
00:08
/
01
12
/
11:22
Select Delete
HC
HC
Press.
Tem p.
HC
Alarm
Over H -Limit
ℓ
LD Temp. Error
ℓ
AI Unde
Low Air PurgeAi
High Gas Temp.
Low Light Trans.
ℓ
Alarm Record
All Delete
All Delete
Day Repair
31 03:23
28 13:28
20 09:44
27 08:11
-5+
2018-04-01
12:00:00
Day Repair
31 03:23
28 13:28
20 09:44
27 08:11
2018-04-01
12:00:00
Press the key at the desired place.
ENT
“Select Delete” is reversed.
Press the key and the “All Delete” is
reversed.
DateDay
04/01 11:23
03/29 23:14
03/25 15:23
03/20 04:41
18
02
00:08
/
01
12
/
11:22
Select Delete
HC
HC
Press.
Tem p.
HC
Alarm
Over H -Limit
ℓ
LD Temp. Error
ℓ
AI Unde
Low Air PurgeAi
High Gas Temp.
Low Light Trans.
ℓ
All Delete
Day Repair
31 03:23
28 13:28
20 09:44
27 08:11
Press the key.
ENT
ENT
All alarms are deleted.
INZ-TN8ZSS-E
- 39 -
Page 50
5.3.3.3 Alarm types
All the alarms except for “Connection Error” which indicates communication failure between the receiver
unit and the control unit will be activated after at least five minutes from the power on. The alarm of “Low
Light Transmission” is is triggered when the low light transmission status is continued for one minute, which
means that it takes six minutes after start-up of the analyzer.
Alarm display Alarm contents Probable causes
LD Temp. Error Peltier which is cooling the laser
cannot control the set temperature.
Laser temperature
control failure
The analyzer cannot control the
laser temperature within the required time.
Low Light Trans. Light intensity required for the
measurement cannot be obtained.
High Light Trans. Light quantity is too much for
measurement.
Overrange Signal input is too much.
Connection Error Communication between receiver
unit and control unit does not occur
properly.
LD communication
error
Communication between the trans-
mitter unit and the receiver unit
failed.
High Gas Temp. It is reported when exceeded gas
temperature is detected.
Out of Range It is reported when exceeded gas
pressure is detected.
Peltier failure
•
Thermistor failure
•
The receiver unit and the transmitter
•
unit are used at an installation location
exceeding the set range.
The unit is used in an environment
where the temperature of transmitter
unit is higher than 55°C
The unit is used in an environment
•
where gas temperature is beyond the
specification.
Light intensity is insufficient in an en-
•
vironment with high dust.
Light intensity is insufficient in an en-
•
vironment with high water vapor.
Contamination of window and conden-
•
sation are caused by insufficient air
purge.
Optical path is blocked due to dust.
•
Optical axis is deflected due to vibra-
•
tion.
Optical axis is deflected due to distor-
•
tion of a stack.
Optical axis is deflected due to external
•
faults.
Optical path is too short
•
Failure of PD analog board
•
Failure of PD analog board
•
Break of wiring
•
Influence of high frequency noise
•
Poor contacts of the connector unit
•
CPU board failure
•
PD digital board failure
•
The actual gas temperature is more than
•
450°C.
The value set to analog input on the
•
“Analog Input” screen is not correct.
Actual gas pressure is outside the speci-
•
fication range.
The value of the input range set on the
•
“Analog Input” screen is not correct.
- 40 -
INZ-TN8ZSS-E
Page 51
Alarm display Alarm contents Probable causes
AI Under The analog input that the control
unit received is out of the range
which is set in the “Analog Input”
screen.
AI terminal is not connected to the ex-
•
ternal input device when channel setting
is set on the “Analog Input” screen.
AI terminal and setting channel do not
•
match.
Input is not 4 to 20mA DC.
•
The input value is ≤ -10% of the analog
•
input range.
The input value is ≥ 110% of the analog
•
input range
Box Temp. Warning Temperature in the receiver unit
and the transmitter unit exceeds the
temperature for normal operation.
The unit is used at an installation loca-
•
tion exceeding the set range.
The unit is used in an environment
•
where gas temperature is beyond the
specification.
Insufficient air purge causes rise in
•
temperature.
The distance between the Receiver /
•
Transmitter unit and the stack is not
maintained sufficiently.
Low Air Purge Purge pressure is lower than the
value set for alarms on the “Analog
Input” screen.
Air purge pressure is lower than alarm
•
setting or the analyzer is not purged.
AI terminal is not connected to the ex-
•
ternal input device when the air purge
pressure is set to channel setting on the
“Analog Input” screen.
Over H-Limit “Analog Output / Alarm Record” is
set to “Over H-Limit” or “Over H/L
Limit” on the “Alarm Setting”
screen, and the measured value exceeds the higher limit.
Concentration beyond the Range limit is
•
measured.
The actual path length is longer than the
•
measured path length set at the “Parameter Setting” screen.
The actual temperature is lower than the
•
temperature (fixed value) set at the
“Analog Input” screen.
Under L-Limit “Analog Output / Alarm Record” is
set to “Under L-Limit” or “Over
H/L Limit” on the “Alarm Setting”
screen, and the measured value is
less than the lower limit.
Concentration beyond the Range limit is
•
measured.
The actual path length is longer than the
•
measured path length set at the “Parameter Setting” screen.
The actual temperature is lower than the
•
temperature (fixed value) set at the
“Analog Input” screen.
INZ-TN8ZSS-E
- 41 -
Page 52
6. CALIBRATION AND SETTING
6.1 Zero calibration
Select “Zero Calibration” from the “Menu” screen and press the key.
ENT
6.1.1 Preparation
Do not detach the transmitter unit during calibration while the pow-
CAUTION
Parts name Quantity Remarks
Calibration gas cell 1 To be ordered separately
Cable between receiver unit and
transmitter unit
Cable between receiver unit and
control unit (for calibration)
Zero gas (N2) 1 To be ordered separately
Pressure regulator 1 To be ordered separately
Pipe (PTFE tube, etc) Several m 10/8 mm or larger
Joint (Rc1/4) 2 For connecting the zero gas line and the
Flow meter 1 2L/min or more
Thermometer 1 4 to 20mA output, for temperature correc-
Pressure gauge 1 4 to 20mA output, for pressure correction
Others (joint, etc)
(1) Turn OFF the power.
(2) Remove the receiver box and the transmitter box using the hexagon wrench. Never remove the boxes
while the system into which the analyzer is incorporated is in operation. Otherwise, hot temperature gas
may blow out.
er is supplied. If you turn the transmitter unit towards a person, the
laser beam can damage his/her cornea.
The laser beam is the invisible infrared light. Do not watch the laser
beam directly or scattering light.
Do not watch the laser beam directly with the optical measuring
device. Otherwise, it may cause serious damage to your eyes.
1 To be ordered separately
1 To be ordered separately
exhaust line to the calibration gas cell
tion (required for range HCl and NH
(required for low range HCl and NH
)
3
)
3
- 42 -
INZ-TN8ZSS-E
Page 53
(3) Connect the transmitter box and the receiver box to the calibration gas cell by using the hexagon socket
head bolts.
*If the cable between the receiver unit and the transmitter unit or/and the cable between the receiver
unit and the control unit cannot be used for calibration because they are fixed, prepare the substitute
cables for calibration.
(4) Connect the cables as shown in the following figure.
INZ-TN8ZSS-E
- 43 -
Page 54
(5) Attach two Rc1/4″ pipe fittings to the inlet and outlet of the calibration gas cell. Connect the tube from
the standard gas (N
) cylinder to the inlet side.
2
(6) Connect the exhaust gas tube to the outlet side. For exhaust gas line, use a tube as large and short as
possible.
*If you use thermometer and/or pressure gauge, attach them on the exhaust gas line, and connect their
4–20mA output to the analog input terminal of the control unit. Make settings for the analog input in
reference to 6.7.1 “Analog input setting: sensor input”.
(7) After all the pipes and cables are connected, flow the calibration gas with a flow rate of 1.5 to 2.0
L/min.
- 44 -
INZ-TN8ZSS-E
Page 55
ppm
ppm
6.1.2 Zero calibration
(1) If the power is OFF, turn it ON.
(2) Check if the flow of N
(3) Display the “Zero Calibration” screen.
(4) Point the to “Component”, and press the key.
(5) Press the key or the key to select the measured gas component to be zero-calibrated.
When there is only one component, it is not necessary to select. Note that, gas component and flowing
gas are not equivalent.
(6) Point the to “Cell Length”, and press the key.
gas is approximately 1.5 to 2.0 L/min.
2
Zero Calibration
Component
Cell Length
Cal.Gas Value
Meas.
HC
ℓ
1000 mm
0000
00.12
2018-04-01
12:00:00
ZeroCal.Star
LastCal.Day
2018/01/01
Cal.Finish
(7) Enter the length of calibration gas cell. Standard cell length is 1000mm. (When the range is low con-
centration, the length of calibration gas cell can be either 500mm or 200mm.)
(8) See the “Meas.” and make sure the indication value is stable.
(9) Point the to “ZeroCal.Start”, and press the key to start the zero calibration.
ENT
(10) “ZeroCal.Start” blinks for about 30 seconds.
(11) When the calibration is completed, the cursor moves to “Cal.Finish”, and the current date is displayed
at “LastCal.Day”. When the date is updated, you can move the cursor. If there are two gas components
to be calibrated, repeat the steps from 5 to 11.
INZ-TN8ZSS-E
- 45 -
Page 56
6.2 Span calibration
Select the “Span Calibration” from the “Menu” screen and press the key.
Note that the span calibration described in this subsection is not available for H
analyzer has no span calibration gas cylinder.
ENT
O analyzer because the H2O
2
6.2.1 Preparation
If toxic fume, corrosive gas or inert gas is used as calibration gas, be
DANGER
CAUTION
Preparation is basically same as that of zero calibration.
(1) Stop the flow of zero gas, and switch it to span calibration gas.
(2) If you use toxic gas for span calibration, fix the exhaust gas outlet at such a position that no one inhale
the toxic gas.
(3) When pipe connection and exhaust are completed, allow the gas to flow with a flow rate between 1.5 to
2.0 L/min.
Notes:
• Use a new regulator when possible. Do not use the regulator that was used for alkaline gas like ammonia for acid gas like HCl. Otherwise, the acid gas is absorbed by the alkaline gas inside the regulator and the indication value may become unstable.
• For HCl, when you bring a gas cylinder which has been unused for long period into use or a new gas
cylinder, it takes time until the indication value is stabilized. Supply gas for a while until it is stabilized.
• When the indication value does not change, even if you supply gas for several tens of minutes, the inside of the regulator may be in rust. Replace it with the new one.
sure that the position of air ventilation or exhaust port is suitable.
Otherwise you may inhale exhaust gas. Suffocation, brain disorder,
circulatory deficit, or contraction of the breathing system may occur,
resulting in death.
Do not detach the transmitter unit during calibration while the pow-
er is supplied. If you turn the transmitter unit towards a person, the
laser beam can damage his/her cornea.
The laser beam is the invisible infrared light. Do not watch the laser
beam directly or scattering light.
Do not watch the laser beam directly with the optical measuring
device. Otherwise, it may cause serious damage to your eyes.
- 46 -
INZ-TN8ZSS-E
Page 57
• About Span Calibration
INZ-TN8ZSS-E
- 47 -
Page 58
ppm
ppm
h
6.2.2 Span calibration
(1) If the power is OFF, turn it ON.
(2) Check if the flow of span gas is approximately 1.5 to 2.0 L/min.
(3) Display the “Span Calibration” screen.
(4) Point the to “Component”, and press the key.
(5) Press the key or the key to select the measured gas component to be span-calibrated.
When there is only one component, it is not necessary to select.
Span Calibration
2018-04-01
12:00:00
Component
Cell Lengt
Cal.Gas Value
Meas.
Gas Temp.
Gas Press.
LastCal.Day
HC
1000 mm
0010.02
00.12
0025
- 00.50
2018/01/01
ℓ
℃
kPa
Span Cal.Start
Cal.Finish
(6) Point the to “Cell Length”, and press the key.
(7) Enter the length of calibration gas cell. Standard cell length is 1000mm. (When the range is low con-
centration, the length of calibration gas cell can be either 500mm or 200mm.)
(8) Point the to “Cal.Gas Value”, and press the key.
(9) Enter the concentration displayed on the gas cylinder.
(10) Point the to “Gas Temp.”, and press the key.
(11) Connect the thermometer to the pipe. When output signal (4 to 20mA) is entered in the AI terminal of
the control unit, read the temperature value on the “Measurement” screen and enter the value. When the
range is low concentration, the value may be affected by gas temperature.
(12) Enter the value in “Gas Press.” in the same manner of “Gas Temp”.
(13) See the value of “Meas.” At the “Span Calibration” screen, and make sure that the value is not com-
pletely different from that of “Cal.Gas Value”, and the indication value is stable. For easily-absorbed
gases such as HCl, it requires a certain length of time to be stabilized. It takes 5 minutes until the gas is
stabilized at the earliest, and it may not be stabilized even an hour, depending on the diameter, length of
exhaust tube, the gas cylinder and regulator.
(14) Point the to “Span Cal.Start” and press the key to start span calibration.
ENT
(15) “Span Cal.Start” blinks for about 30 seconds.
(16) When the calibration is completed, the cursor moves to “Cal.Finish”, and the current date is displayed
at “LastCal.Day”. When the date is updated, you can move the cursor. If there are two components to
be calibrated, repeat the steps from 5 to 16.
- 48 -
INZ-TN8ZSS-E
Page 59
6.2.3 H2O calibration by matching with values obtained by
manual analysis
This method does not use the calibration cell. Calibration is carried out without uninstalling the receiver unit
and the transmitter unit.
6.2.3.1 Required data
(1) H2O concentration measured by manual analysis
(2) H
6.2.3.2 Preparation for span calibration
Calibration gas concentration should be calculated by following procedure prior to span calibration.
(1) Check the time took for manual analysis and the H
(2) Average the H
(3) Plug the values obtained in the above steps 1 and 2 into the following equation to calculate the span
The following procedures should be conducted just before starting the span calibration.
(4) Check on the H
(5) Plug the mean value obtained in the step 4 and the span calibration coefficient in the step 3 into the fol-
O concentration measured by the laser gas analyzer at the same time as the manual analysis
2
O concentration measured by the manual analysis.
2
O concentration values measured by the laser gas analyzer over the same time as manual
2
analysis.
For example, if you carried out the manual analysis from 9:00 to 9:30, average the H
O concentration
2
measured by the laser gas analyzer from 9:00 to 9:30.
calibration coefficient.
Coefficient of span calibration =
e.g.: When manual value in the step 1 is 15[vol%] and average value of H
Manual analysis value in the step 1 [vol%]
Average H
O concentration in the step 2 [vol%]
2
O concentration is 10 [vol%]
2
in the step 2.
Coefficient of span calibration=
O span calibration screen that the fluctuation of the measured values is within ±1 vol%.
2
15 [vol%]
10 [vol%]
= 1.5
Calculate the mean value of appropriately 10 seconds.
lowing equation to obtain the calibration gas concentration.
Calibration gas concentration = Mean concentration × Span calibration coefficient
For example, when the mean value is 5 vol% and the span calibration coefficient is 1.5, the calibration
gas concentration is:
5 [vol%] × 1.5 = 7.5 [vol%]
INZ-TN8ZSS-E
- 49 -
Page 60
6.2.3.3 Procedure of span calibration
In [Menu] screen, point the cursor to [span
calibration] with and then
press .
ENT
Point the cursor to [Component] with
and then press the key to reverse
out the cell.
Select the [H
press .
O] with and then
2
ENT
Point the cursor to [Cell Length] with
and then press the key to reverse
out the cell. Input the length of measurement
cell with and then press .
ENT
- 50 -
INZ-TN8ZSS-E
Page 61
Point the cursor to [Cal Gas Value] with
and then press the key to
reverse out the cell.
By using the keys, enter the cal-
ibration gas concentration calculated in
6.2.3.2, and then click the key.
ENT
Point the cursor to [Gas Temp] with
and then press the key to reverse
out the cell.
Input the 25 [°C] with and then
press .
ENT
Point the cursor to [Gas Press] with
and then press the key to reverse
out the cell. Input the 0.00[kPa] with
and then press .
ENT
INZ-TN8ZSS-E
- 51 -
Page 62
Point the cursor to [Span Cal Start] with
and then press .
ENT
Calibration will be completed in 30 seconds
and automatically cursor will move to [Cal
Finish].
- 52 -
INZ-TN8ZSS-E
Page 63
6.3 Alarm setting
Select “Alarm Setting” from the “Menu” screen, and press the key. On the “Alarm Setting Gas” screen,
ENT
select a component by moving the with the key and the key.
Press the key. “Alarm Setting” screen is displayed.
ENT
6.3.1 Alarm value ON/OFF
ON validates the High/Low limit alarm output, alarm display, alarm record for the measured concentration.
Select OFF to invalidate.
Note
When OFF is selected, “High/Low limit setting”, “Analog Output / Alarm Record” cannot be set.
Point the to “H-Limit” or “L-Limit” by the
key and the key, and press the
key.
( )
INZ-TN8ZSS-E
- 53 -
Page 64
Select “ON” or “OFF” by the key or the
key, and press the key.
( )
ENT
When you press the key, the cursor returns
to the initial position.
ENT
6.3.2 Setting of alarm value
Make a setting of the High/Low limit alarm for the measured concentration. To change the alarm setting, set
the Alarm ON/OFF setting to ON, and then change the numeric value.
Point the to “H-Limit” to change the setting of H-Limit value, and point “L-Limit” to change the set-
ting of the L-Limit value.
Note
Point the to “H-Limit” to set the H-limit val-
ue, and point “L-Limit” to set the L-Limit value,
and then press the key.
( )
- 54 -
INZ-TN8ZSS-E
Page 65
Select “ON” by the key or the key,
( )
ENT
( )
and press the key.
Press the key.
ENT
Change the numeric value by the key or
the key, and press the key.
ENT
( )
ENT
Click the key, and the returns to the
initial position.
-5+
Set the value so that H-Limit is larger than L-Limit, and that (H-Limit – L-Limit) is larger than hysteresis
width.
Note
INZ-TN8ZSS-E
- 55 -
Page 66
6.3.3 Analog output / alarm record
Set the recording range of the external alarm output such as analog output or the alarm record.
Setting range can be selected from “H-Limit”, “L-Limit” and “H/L Limit”.
When “OFF” is set to the ON/OFF setting of the alarm value, “Analog Output / Alarm Record” cannot be
set. Select “ON” again. The alarm output which is “H-Limit”, “L-Limit” or “H/L Limit” cannot be per-
formed for 5 minutes after turning on the power.
Note
Point the to “Analog Output / Alarm Record”
by the key and the key, and press
the key.
( )
( )
ENT
Select output range by the key and the
key, and press the key.
ENT
Range H-Limit : Alarm output is provided only when a value exceeds the Range H-Limit.
Range L-Limit : Alarm output is provided only when a value is lower than Range L-Limit.
Range H/L-Limit : Alarm output is provided when a value exceeds the Range H-Limit or it is lower than
Explanation of alarm range
Range L-Limit.
- 56 -
INZ-TN8ZSS-E
Page 67
6.3.4 Hysteresis setting
Set the hysteresis to prevent possible chattering of the alarm output near the alarm setting value.
Point the to “Hysteresis” by the key or
the key, and press the key.
( )
Change the numeric value by the key or
the key, and move the digit by the
key.
Press the key to make the hysteresis valid.
ENT
( )
ENT
Setting Range
0 to 20% FS
%FS: Indicates the rate for which the range width of each component is regarded as 100%.
INZ-TN8ZSS-E
- 57 -
Page 68
Hysteresis mode (in case of H-Limit)
Alarm output is turned ON when the value exceeds the H-Limit. The alarm goes off when the measured value has gone below the hysteresis.
- 58 -
INZ-TN8ZSS-E
Page 69
6.3.5 Gas temperature alarm
When the gas temperature is out of the range between the higher limit and the lower limit, the analyzer emits
the “High Gas Temp” alarm.
Set the limit values so that the upper limit is larger than the lower limit.
Note
( )
ENT
By using the keys, move the cursor
to the “Gas Temp upper limit alarm” or the “Gas
Temp lower limit alarm”, and press the
key.
ENT
By using the keys, change the values
and then press the key.
ENT
( )
ENT
INZ-TN8ZSS-E
- 59 -
Page 70
N
6.4 Output hold
Before a maintenance work, be sure to carry out the setting explained in this subsection to hold the analog
output.
Select “Output Hold” from the “Menu” screen, and press the key.
ENT
6.4.1 Output hold
Set output hold to “Last Meas.” or “Pre-set” to hold analog output. (Indication value on the “Measurement”
screen is not held. However, it is highlighted during the output hold time.)
Output Hold
HCℓ
o
2018-04-01
020
12:00:00
%FS
Point the to the measurable component to
hold output by the key or the key,
and press the key.
( )
Output Hold
HCℓ
Pre-set
2018-04-01
020
12:00:00
%FS
Select either “Last Meas.” or “Pre-set” by the
key or the key. Press the key
to validate the setting.
ENT
( )
ENT
- 60 -
INZ-TN8ZSS-E
Page 71
t
Output Hold
HCℓ
Output Hold
HCℓ
2018-04-01
12:00:00
020
%FS
ENT
2018-04-01
020
Pre-se
( )
Pre-set
12:00:00
%FS
When “Last Meas.” is selected, cursor returns to
the starting position.
When “Pre-set” is selected, numeric value is
highlighted. Change the numeric value by the
key or the key, and move the digit to
the right by the key.
After the numeric value is changed, press the
ENT
key.
“Last Meas.” : Holds the value for which “Last Meas.” is determined by the key.
“Pre-set” : Holds the %FS value for which range is currently validated.
Setting Range
ENT
Example) When range is from 0 to 10ppm, and the set value is 20%FS, the value corre-
sponding to 2ppm is output regardless of the measurement value.
O2 Conversion Hold
“Last Meas.” : Holds the measurement value for which “Last Meas.” is determined by the key, and
the value calculated by O
value. When the O2 value is fixed, it holds the value calculated
2
by the fixed value.
“Pre-set” : Even if O
determined by the key.
analog input is entered, it holds the value calculated by O2 fixed value which is
2
ENT
Last Meas. of average value
ENT
If you select the “Last Meas” while using the average value output, the analyzer hold the value averaged
before you press the key.
ENT
INZ-TN8ZSS-E
- 61 -
Page 72
6.4.2 Remote hold (DI3 terminal)
You can remotely hold the output by using the external contact output (DI3 terminal, option).
The value to be hold is “Last Meas” only.
By applying a voltage (pulse width 2.0 seconds or more) to the remote hold input terminal, you can hold the
analog output. The hold is cancelled when you stop applying the voltage.
- 62 -
INZ-TN8ZSS-E
Page 73
6.5 Parameter setting
On the “Parameter Setting” screen, “Path length” related to measurement value and “Average Period” related
to average value output are set. Items to be set are as follows.
Path length : Enter the optical path lengths.
Purge length : Enter the total length of the purge length A and the purge length B shown in the figure
Avg. Period : This parameter is for the average value output. Set the time (in minutes) used for cal-
cv O
Reset Avg.Gas : This parameter is for the average value output. This allows you to reset the average
Key Lock : No key operation except for the key lock OFF can be performed.
Disp.OFF : Set the on or off of the screen timeout function. If you turn on the screen timeout, you
Date/Time : Set the current year/month/date, hour: minute: second.
Setting items
in 6.5.1. Set whether you use the blast purge or not. The purge length setting is required
only for the O
culation of the moving average.
: Set the reference O2 concentration used for converting the measured values.
2
value.
also have to set the time (in minutes) until the screen automatically goes off after the
last operation.
measurement for combustion control.
2
Select the “Parameter Setting” from the “Menu” screen, and press the key.
ENT
6.5.1 Path length
Enter the diameter of the stack where the receiver unit and the transmitter unit are attached. It does not include the lengths of companion flange. This value has a direct effect on the measurement value, so be sure to
enter correctly. Otherwise, measurement value will not be properly displayed or output. Enter the value in the
millimeter. (Input range is from 100 to 15000mm.)
Note
In default setting, the analyzer calculates the gas concentration assuming the optical path length is one
meter. Be sure to set the correct path length because the volume of the light absorbed by a gas changes
with the optical path length.
INZ-TN8ZSS-E
- 63 -
Page 74
Point the to “Path lengths” by the key
( )
( )
ENT
and the key, and press the key.
Change the numeric value by the key or
the key, and move the digits by the
key. Press the key to validate the set input
value.
ENT
- 64 -
INZ-TN8ZSS-E
Page 75
6.5.2 Purge length
The “purge length” setting is required only for the O2 analyzer for combustion control. In other cases, you do
not have to edit this parameter.
Setting range: 100–15000 mm
If you ordered the “blast purge” option, you have to set the “blast purge” column as well.
Move the cursor to “Purge Length” by using the
keys, and press the key.
( )
Edit the value by using th keys to
( )
( )
ENT
ENT
change the value, and the key to move to
the next digit. Press the ENT key to set the value.
(The cursor will move to “Blast Purge”.) The
“Purge length” setting is completed.
If you ordered the “blast purge” option, move on
to the next step.
Select “Blast Purge” or “None” by using the
keys, and press the key.
Press the key twice, and the cursor returns
to the initial position.
ENT
ENT
INZ-TN8ZSS-E
- 65 -
Page 76
- 66 -
INZ-TN8ZSS-E
Page 77
6.5.3 Moving average time
This parameter is required if you set the indication and the output to the average in “6.6.1 Instantaneous/average value”. Setting range is from 1 to 60 minutes. Up to the time you set, the analyzer emit the average value at the time.
The average value is reset when you press the ENT key after changing the setpoint.
When the moving average time is set to 60 minutes
Averaging operation
Reset
Output
0
Sampling rate: twice per second
Mode 1:
Until the time reaches 60 minutes, the analyzer calculates the average by dividing the available values
by the time passed.
*Available values: the values obtained after you reset the average to the each moment.
The quation is:
Mode 2:
Until the time reaches 60 minutes, the analyzer calculates the average by dividing the available values
by 60 minutes.
The equation is:
In both modes, it is after 60 minutes that you can obtain the first moving average value for 60 minutes.
60min
Σ
Σ
60
Time
INZ-TN8ZSS-E
- 67 -
Page 78
Move the cursor to the “Avg. Period” by using
the keys, and then press the
key.
( )
Change the value by using the keys,
and move the digit by using the key. Press
( )
( )
ENT
ENT
ENT
the key to set the value, and the cursor will
move to the next column, “Mode 1”.
Select either of “Mode 1” or “Mode 2” by using
the keys, and press the key.
If you press the key twice, the cursor will
return to the initial position.
ENT
ENT
- 68 -
INZ-TN8ZSS-E
Page 79
6.5.4 O2 conversion
This parameter allows you to set the reference O2 concentration value used for converting the measured values. For the detail of the O
conversion, refer to “5.3.1.2 O2 conversion concentration value”.
2
Move the cursor to the “cv O2” by using the
keys, and then press the key.
( )
Change the value by using the keys,
and move the digit by using the key. Press
ENT
the key to set the value.
( )
ENT
INZ-TN8ZSS-E
- 69 -
Page 80
6.5.5 Average value reset
When the indication value and output value are set to the average value in the “6.6.1 Instantaneous/average
value”, resetting the average value works on both the measured value and the O
indication value and the output value are set to the instantaneous value, nothing changes even if the average
value is reset.
Point the to “Reset Avg.Gas” by the
key and or key, and press the key.
( )
converted value. When the
2
Select the measurable component to be reset by
the or the key. When there is only
one component, the component is fixed.
Press the key to reset the average value.
ENT
( )
ENT
- 70 -
INZ-TN8ZSS-E
Page 81
Remote average value reset (DI1 terminal)
Apply voltage to the input terminal of remote average value reset (DI1 terminal, option) to reset the aver-
age value.
12 to 24 VDC/
5 to 20 mA
OFF
Remote resetting signal
Starts averaging
Reset is kept during short-circuit.
When switching the action from short-circuiting to opening, average value action starts.
6.5.6 Key lock
Key lock disables any setting change except for turning off the key lock.
On the “Menu” screen, you cannot enter any
screen other than the “Parameter Setting”.
INZ-TN8ZSS-E
On the “Parameter Setting” screen, you cannot
edit any items other than the “Key Lock”.
- 71 -
Page 82
6.5.7 Backlight timeout
This parameter allows you to set the time until the backlight automatically turns off after the display returns
to the measurement screen. The setting range is from 1 to 60 minutes. The maximum time that the backlight
can last is approximately 58000 hours.
Point the to “Disp.OFF” by the or the
key, and press the key.
( )
Turn it to “ON” by the or the key,
and press the key.
ENT
( )
ENT
Change the numeric value by the key or
the key, and move the digits by the
key. Press the key to validate the set input
value.
ENT
- 72 -
INZ-TN8ZSS-E
Page 83
6.5.8 Time and date
Set the time and date every three months because the time error occurs depending on the ambient temperature.
Move the cursor to the “Date/Time” by using the
keys, and then press the key.
( )
Change the value by using the keys,
and move the digit by using the key. Press
ENT
the key to set the value.
( )
ENT
INZ-TN8ZSS-E
- 73 -
Page 84
6.6 Display/AO setting
Make a setting of “Inst./Avg.” or “Wet/Dry” of the measurement value displayed on the “Measurement”
screen.
“Inst./Avg.” : Select “instantaneous value” or “average value” for each measurable com-
“Wet/Dry” : Select “Wet” or “Dry” for each measurable component. When “Dry” is set,
Select “Display/Output” from the “Menu” screen, and press the key.
Setting items
ponent.
H
O setting of the “Analog Input” is required.
2
ENT
6.6.1 Instantaneous/average value
Make a setting of “Instantaneous value” or “Average value” for each measurable component. The setting is
reflected on the display on the “Measurement” screen.
● You can set the instantaneous value and the average value for the O
of measured value.
● If you select the average value, the analyzer provides moving averages obtained based on the time you set
in the “Moving average time” of the parameter setting screen.
Remote switching between the instantaneous value and the average value (DI2 terminal)
Applying a voltage to the optional DI2 terminal allows you to switch between instantaneous value and the
average value. If the measured value has been set to the instantaneous value and the O
average value, applying a voltage change the measured value to the average value and the O
the instantaneous value.
conversion regardless of the setting
2
conversion to the
2
conversion to
2
- 74 -
INZ-TN8ZSS-E
Page 85
6.6.2 Wet/dry
Make a setting of “Wet” or “Dry” for each measurable component. The setting is reflected on the display on
the “Measurement” screen.
● You can set the O
ue.
● This analyzer performs the measurement in wet environment. If you want to obtain dry values, you need
to set the values for H
• If you use an external moisture meter: Connect the 4–20 mA DC output from the moisture meter to the
analog input terminal of the ZSS control unit, and set the range with reference to “6.7.1 Analog input
setting: sensor input”.
• No moisture meter: On the Display/AO setting screen, set the H
calculate the dry values based on the fixed H
• If you use the laser gas analyzer that can measure H
by the laser gas analyzer or the H
Wet can be converted to Dry using the following expression.
Measurement value (dry) = Measurement value (wet) × 100/ (100 – moisture content (%))
conversion value to whichever of wet or dry, regardless of the setting of measured val-
2
O, in any way among the followings:
2
O value to “fixed”. The analyzer will
2
O concentration.
2
O: you can select either the H2O values measured
2
O values provided by an external moisture meter.
2
( )
( )
ENT
ENT
Select the component to be set by the key
or the key. Press the key to move
the cursor to the “Inst./Avg.” setting.
ENT
Select “Inst.” or “Avg.” by the key or the
key. Press the key to move the cur-
sor to “Wet/Dry” setting.
ENT
INZ-TN8ZSS-E
- 75 -
Page 86
Select the “Inst” or “Avg” by the key or
( )
( )
ENT
ENT
the key. Press the key to move the
ENT
cursor to “Range Control” setting.
Press the key to move the cursor to the set
component.
ENT
-5+
The key can move the cursor backward in
the middle of the setting. The setting fixed by the
ENT
key does not return to the previous setting,
even if the cursor is moved by the key.
-5+
- 76 -
INZ-TN8ZSS-E
Page 87
6.7 Analog input
By providing the signals from external sensors to the control unit, you can reflect the changes in gas conditions into the measurement result. The analyzer can accept two 4–20 mA DC inputs from a thermometer,
pressure gauge, flowmeter, oxygen analyzer, and/or moisture meter. Wire between each sensor and the analog input terminal of the control unit in reference to “3.2 Wiring diagram”.
Set the parameters of an item with no sensor input to “fixed” and enter a value for each. For items whose
change is negligible, you can set the item to a fixed value.
Be sure to make settings for all the items. Otherwise, the analyzer cannot deliver accurate measurement.
Select “Input Signal” from the “Menu” screen, and press the key.
ENT
6.7.1 Analog input setting: sensor input
The description here is an example when using a thermometer.
Connect the 4-20 mA DC signal from a thermometer to Ch1 or Ch2 of AI terminal.
If you make a setting but there is no signal input, the analyzer emits an “analog input error” alarm.
Select the “Gas Temp.” by the key or the
key. Press the key to move the cur-
sor to “Channel” setting.
( )
INZ-TN8ZSS-E
- 77 -
Page 88
Select the channel (CH1 or CH2) connected to AI
terminal by the key or the key, and
( )
( )
ENT
ENT
press the key.
ENT
Enter the temperature (°C) corresponding to 4mA
DC output of the output signal of the thermome-
ter. Change the numeric value by the key
or the key, and move the digits by the
key. Press the key to validate the set
ENT
input value.
Enter the temperature (°C) corresponding to
20mA DC output of the output signal of the
thermometer. Change the numeric value by the
key or the key, and move the digits
by the key. Press the key to validate
ENT
the set input value.
( )
ENT
Implement the same procedure as for the sensor
setting of pressure gauge, flow meter, oxygen
analyzer (O
) or the moisture meter (H2O).
2
- 78 -
INZ-TN8ZSS-E
Page 89
conversion
O
2
The analyzer converts the measured concentration into the value in reference to the O
con-
2
centration provided by an external oxygen analyzer.
C
21
21
Where:
C : converted concentration
Cs : measured gas concentration (%)
Os : Measured O
On : Reference O
concentration (%)
2
concentration (%) *Default value is 12%
2
INZ-TN8ZSS-E
- 79 -
Page 90
6.7.2 Analog input setting: fixed value
The description here is an example of the H2O input.
In such cases that the gas conditions do not change, that no external sensor is available, or that no analog input terminals are left, set a fixed value. Note that if there is a large difference between the setpoint and the
actual value, the analyzer may not be able to deliver accurate measurement.
Select “H2O” by the key or the key.
Press the key to move the cursor to
“Channel” setting.
( )
( )
( )
ENT
ENT
Select “Fixed” by the key or the
key, and press the key.
ENT
Enter the fixed value.
Change the numeric value by the key or
the key, and move the digits by the
key. Press the key to validate the set input
value.
ENT
- 80 -
INZ-TN8ZSS-E
Page 91
The same procedure can be implemented as for
pressure gauge, thermometer or oxygen analyzer
(O
).
2
6.7.3 Air purge pressure
If a pressure gauge is installed on air purge pipe, and connected to the control unit, the analyzer can emit an
alarm when the purge pressure is below the setpoint.
● Be sure to install a pressure gauge if the purge pressure is unstable.
● If you measure high-temperature gas without purging the analyzer, the analyzer may get damaged.
When there is no sensor to be connected, set the air purge pressure to “No”. If you select “CH1” or “CH2”
while there is no signal input, the analyzer emits an “analog signal error” alarm.
( )
( )
ENT
Select “Air Purge” by the key or the
key. Press the key to move the cursor to
“Channel” setting.
Select the channel (CH1 or CH2) connected to AI
terminal by the key or the key, and
press the key.
ENT
INZ-TN8ZSS-E
- 81 -
Page 92
Enter the purge pressure value to output alarm.
Change the numeric value by the key or
the key, and move the digits by the
key. Press the key to validate the set input
value.
ENT
( )
ENT
Enter the pressure value corresponding to 4,
20mA DC output in the same manner.
- 82 -
INZ-TN8ZSS-E
Page 93
6.8 Analog output
There are two analog outputs (4 to 20mA DC output) as a standard specification. This allows you to emit the
measured value and the O
the instantaneous value separately. In addition to the measured value and the O
mittance is available as the analog output.
Default value is set to “No”, so be sure to set it after installation.
Select “AO Select” from the “Menu” screen, and press the key.
conversion value individually. For each of them you can set the average value or
2
conversion value, the trans-
2
ENT
When the number of the components to be measured is one, “Component” or “No” can be selected. When
there is O
conversion output, “O2 Conversion” is also selectable.
2
Select the analog terminal to be output by the
or the key. Press the key to
move the cursor to “Output Component” setting.
( )
INZ-TN8ZSS-E
Select the component to be output by the
or the key. Press the key to validate
the set input value.
- 83 -
ENT
Page 94
6.9 Fine adjustment of analog output value
The analog output values may slightly deviate from appropriate values due to load resistance. In such a case,
adjust the analog output by the following method.
Point the cursor to "Password" in the bottom left
of the “Menu” screen by the key or the
key.
Press the key to change the display to
“00000”.
ENT
Enter “02404” to the Password. Select the digits
by the key, and press the key or the
key to move the cursor to “Password”.
Press the key to display the “Factory
ENT
Menu” screen.
Point the cursor to “Analog Output Tuning” by
the key or the key, and press the
ENT
key.
- 84 -
INZ-TN8ZSS-E
Page 95
This device has the simulation output function.
Point the cursor to “Output Fine Tuning” at the
bottom and press the key. Select an item
from “Meas Output”, “Zero Output”, “Half Out-
put”, or “Span Output” by the key or the
ENT
key, and press the key.
ENT
Selected analog output adjustment is performed
on all outputs.
Point the cursor to the analog output to be ad-
justed while "Zero output" is selected for Analog
fine adjustment, and press the key.
ENT
The cursor then moves to 4-digit numeric value
in the “Zero” column, press the key or the
key to adjust the output.
Press the key, and the “output fine tuning”
will change to “Span Output” and the cursor will
move to numeric values on the first row of the
span.
Edit the value by using the keys.
INZ-TN8ZSS-E
- 85 -
Page 96
Press the key twice, and the cursor will
return to the initial position and the “output fine
tuning” will change back to “Meas Output”.
- 86 -
INZ-TN8ZSS-E
Page 97
6.10 Digital Output
Follow the instructions to check if the digital output relays properly work, and/or if the wiring has been done
appropriately.
6.10.1 Checking the digital output by using the alarm
The terminal number and the initial setting of each relay are shown in parenthesis.
(1) Light intensity low (DO1, SPST-NO)
Remove the receiver box or the transmitter box to generate a “light intensity low” alarm. Do not look
into the transmitter box. Otherwise, it may cause serious damage to your retinae or cornea. The “light
intensity low” alarm starts after about a minute during operation and about six minutes right after the
start-up. After you check the alarm output, attach the receiver box or the transmitter box to the angle
adjustment unit where they were.
(2) Device failure (DO2, SPST-NO)
Turn off the power, and disconnect the cable between the receiver unit and the control unit. Turn on the
power again, and the “connection error” appears on the screen and a “device error” alarm is emitted.
After you check the alarm output, turn off the power, connect the cable, and turn on the power again.
(3) During hold / during calibration (DO3, SPST-NO)
Hold the output in reference to “6.4 Output hold”. Check the output during hold / during calibration.
After checking it, release the hold.
(4) Overrange/underrange (DO4, SPST-NO)
Set the range limit so that the current measured value is beyond the upper or lower limit, in reference to
“6.3 Alarm setting”. After you check that the overrange/underrange alarm is emitted, set the alarm to
OFF or change the setting to the appropriate one.
(5) Environmental error (DO5)
Edit the analog input setting to channel 1 or channel 2 while no sensor is connected to the AI terminal.
“AI Under”, which means the analog input error, appears on the screen, and an environmental error
alarm is emitted.
(6) Power interruption (DO6, closed during power interruption)
De-energize the analyzer and check the output. In the default setting, the contact is closed when no
power is supplied, and opened when the power is supplied.
INZ-TN8ZSS-E
- 87 -
Page 98
6.10.2 Changing the relay contact
Open the cover of the control unit. Change the position of the shorting plug for the switch.
The check cell is intended to be used to check a deviation of the span point without removing the device
from the stack. However, if the gas concentration inside the stack is not stable, the check is unavailable.
You cannot check the zero point as far as the measured gas exists in the stack.
INZ-TN8ZSS-E
- 89 -
Page 100
6.11.2 Concentration of the gas which is fed to the check cell
and gas concentration fluctuation
Based on the Lambert-Beer Law, absorption intensity is proportional to gas concentration and the length
where the gas exists (optical path lengths or stack length).
Since the check cell length is 0.25m, the gas concentration to be fed is calculated from the following equation.
Required gas cylinder concentration
= (Measuring range - Measured gas concentration inside the stack (should be stabilized)) × 4
For example, when the measurement range is from 0 to 50 ppm, and the gas concentration inside the stack is
10 ppm,
Required gas cylinder concentration = (50 - 10) × 4
= 160ppm
Feeding the gas of 160ppm to the check cell is similar to feeding the gas at the span point.
Deviation of the span point can be checked by studying difference between the concentration output at this
time and 50ppm.
- 90 -
INZ-TN8ZSS-E
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.