Fuji CR Console User Manual

CR Console Operation Manual
FUJI COMPUTED RADIOGRAPHY
CR Console
OPERATION MANUAL
THE USER UTILITY OPERATIONS
12th Edition
2011. 06 897N0520L

END-USER SOFTWARE LICENSE AGREEMENT

FUJIFILM Corporation
The license of the Software (as defined below) granted hereunder is contingent upon the Licensee’s acceptance of all the terms provided herein, and any use of the Software by the Licensee shall be deemed as such acceptance.
This Product contains software ("Software") and related written materials ("Documentation"). The term "Software" shall include any updates, upgrades, modified versions, additions and copies of the Software licensed to you by FUJIFILM Corporation ("FUJI").
1. License. Subject to the terms and conditions of this Agreement, you are granted a limited, non-exclusive, non-transferable and non-
sublicensable license to use the Software and the Documentation only on a single computer until such rights are terminated according to Section 8 below.
2. Ownership. It is acknowledged and agreed that FUJI retains all title and rights to the Software the Documentation and all copies thereof, and no title or rights to the Software or the Documentation is transferred or licensed to you except as expressly set forth herein.
3. Restriction on Use. You may: (i) make and retain one extra copy of the Software for backup purposes only; and (ii) copy the Documentation only as reasonably necessary for user reference and backup purposes. Each copy of the Software and Documentation made by you shall contain the same proprietary notices, labels, and marks as the original thereof without any alteration. You shall not: (i) sell, sublicense, assign, lease, encumber or otherwise transfer the Software or the Documentation or the rights granted hereunder; (ii) reverse engineer, decompile, disassemble or create derivative works of the Software; (iii) remove or alter any proprietary notices, labels or marks on or contained in any part of the Software or the Documentation; (iv) make any copies of the Software or the Documentation except as permitted above; or (v) install the Software on any computer other than as permitted by this Agreement. As a condition of the use of the Software by you, you hereby represent and warrant that you shall not use the Software for any purpose that is unlawful or prohibited by this Agreement. You agree to abide by all applicable laws and regulations in your use of the Software.
4. Export Control. You shall comply with all export and re-export restrictions and regulations imposed by the government of the United States and other relevant countries or regions (“Export Restrictions”). You shall not knowingly transfer, directly or indirectly, any restricted software or technical data received hereunder or the direct product of such data, to any country or region identified as an embargoed destination or country in the Export Restrictions, unless prior written authorization is obtained from FUJI and each appropriate United States or other government agencies. This Section shall survive the expiration or termination of this license agreement.
5. Update and Upgrade. FUJI may, at its sole discretion, make any update or upgrade of the Software available; however, FUJI shall have no obligations to make any update or upgrade of the Software available.
6. Limited Warranty. FUJI hereby warrants that the Software shall operate in substantial conformity with the then current Documentation, when used in strict compliance with such Documentation. This warranty shall be effective only for one year following your receipt of the Software (hereinafter referred to as “Warranty Period”). This warranty is conditioned upon the installation and use of the latest version of the Software released by FUJI in strict compliance with the then current Documentation. If the Software fails to perform as warranted in the Warranty Period, FUJI shall use commercially reasonable efforts to correct such Software. THE FOREGOING STATES THE SOLE AND EXCLUSIVE REMEDIES TO WHICH YOU ARE ENTITLED FOR FUJI'S BREACH OF WARRANTY. FUJI HEREBY DISCLAIMS ANY OTHER WARRANTIES WITH RESPECT TO THE SOFTWARE AND DOCUMENTATION, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION ALL WARRANTIES REGARDING MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE AND NON­INFRINGEMENT.
7. Limitation of Liability. TO THE FULLEST EXTENT PERMITTED BY APPLICABLE LAW, FUJI SHALL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, EXEMPLARY, PUNITIVE, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OF ANY KIND (INCLUDING WITHOUT LIMITATION, LOST PROFITS OR LOST DATA), WHETHER SUCH DAMAGES ARE BASED ON CONTRACT, TORT (INCLUDING NEGLIGENCE), PRODUCT LIABILITY OR OTHERWISE (INCLUDING ANY DAMAGES WHICH ARE IN ANY WAY CONNECTED WITH ANY USE OF THE SOFTWARE OR DOCUMENTATION), EVEN IF IT HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.
8. General Provisions. This Agreement shall be governed by and construed according to the laws of Japan, excluding its conflict of laws rules. You consent to the exclusive jurisdiction of Japanese courts and agree that the venue shall lie exclusively in Japan, in arising out of or relating to this Agreement. The use of the Software or Documentation is not authorized in any jurisdiction that does not give effect to all provisions of the terms and conditions of this Agreement, including without limitation this Section 8.
9. Termination. All the rights granted to you hereunder shall automatically terminate upon your breach of any provision herein, and you must cease the use of the Software and Documentation and destruct all copies of the Software and Documentation in your possession, including any of backup copy, upon such termination. Any provision herein that by its nature survives shall survive the termination of the rights granted to you hereunder.
all disputes
CR Console Operation Manual

CONTENTS

Chapter 1. BEFORE OPERATION
1.1 Functions of the User Utility .................................................................................
1.2 Starting/Ending the User Utility ............................................................................
1.2.1 Starting the User Utility .......................................................................
1.2.2 Ending the User Utility ........................................................................
Chapter 2. MENU SETTING
2.1 Exposure Menu Setting..........................................................................................
2.1.1 Creating Exposure Menus ....................................................................
2.1.2 Editing Exposure Menus ......................................................................
2.1.3 Deleting Exposure Menus ....................................................................
2.2 Study Menu Setting ...............................................................................................
2.2.1 Creating Study Menus ..........................................................................
2.2.2 Editing Study Menus............................................................................
2.2.3 Deleting Study Menus ..........................................................................
2.3 Display Group Setting ...........................................................................................
2-1
1-1
1-2
1-5
1-5
1-7
2-4
2-4
2-22
2-24
2-25
2-26
2-37
2-39
2-40
2.3.1 Relocating Display Groups ..................................................................
2.3.2 Renaming Display Groups ...................................................................
2.3.3 Deleting Display Groups ......................................................................
2.3.4 Setting Non-display to Display Groups................................................
2.4 Display Menu Setting ............................................................................................
2.4.1 Registering Display Menus ..................................................................
2.4.2 Relocating Display Menus ...................................................................
2.4.3 Deleting Display Menus.......................................................................
Chapter 3. TABLE SETTING
3.1 Technologist Information Setting ..........................................................................
3.1.1 Registering Technologist Information..................................................
3.1.2 Editing Technologist Information ........................................................
3.1.3 Relocating Technologist Information...................................................
3.1.4 Deleting Technologist Information ......................................................
3.2 Requesting Department Information Setting .........................................................
2-40
2-40
2-41
2-43
2-44
2-44
2-45
2-46
3-1
3-3
3-4
3-6
3-8
3-8
3-9
3.2.1 Registering Requesting Department Information.................................
3.2.2 Editing Requesting Department Information .......................................
3.2.3 Relocating Requesting Department Information..................................
3.2.4 Deleting Requesting Department Information .....................................
iv 2008. 09 897N0520F
3-10
3-11
3-12
3-13
CR Console Operation Manual
3.3 Film Mark Setting ..................................................................................................
3.3.1 Registering/Editing Film Marks ...........................................................
3.3.2 Relocating Film Marks .........................................................................
3.3.3 Deleting Film Marks .............................................................................
3.4 User Group Management .......................................................................................
3.5 Setting for Comment on Mis-exposure ..................................................................
3.5.1 Registering the Mis-exposure Classification ........................................
3.5.2 Editing the Mis-exposure Classification ...............................................
3.5.3 Relocating the Mis-exposure Classification .........................................
3.5.4 Deleting the Mis-exposure Classification .............................................
3.5.5 Registering Preset Comments ...............................................................
3.5.6 Editing Preset Comments......................................................................
3.5.7 Relocating Preset Comments ................................................................
3.5.8 Deleting Preset Comments....................................................................
Chapter 4. PROPERTY SETTING
4-1
3-14
3-15
3-16
3-17
3-18
3-20
3-21
3-22
3-23
3-24
3-25
3-26
3-27
3-28
Chapter 5. PARAMETER BATCH UPDATING
Chapter 6. BACKUP/RESTORE
6.1 Backup....................................................................................................................
6.2 Restore....................................................................................................................
6-1
Chapter 7. PATIENT'S INFORMATION DB UTILITY
7.1 Patient’s Information Maintenance ........................................................................
7.1.1 Registering Patient’s Information .........................................................
7.1.2 Editing Patient’s Information................................................................
7.1.3 Deleting Patient’s Information..............................................................
7.1.4 Customizing the Patient’s Information List ..........................................
7.2 Backup....................................................................................................................
7.3 Restore....................................................................................................................
7.4 Patient’s Information Import..................................................................................
5-1
7-1
Chapter 8. IMAGE PROCESSING PARAMETERS FILE TRANSFER
6-2
6-5
7-3
7-5
7-6
7-8
7-9
7-11
7-14
7-16
8-1
Chapter 9. EXPOSURE RESULT LOG
9.1 Saving Logs ............................................................................................................
9.2 Clearing Logs .........................................................................................................
2010. 03 897N0520H v
9-1
9-3
9-7
CR Console Operation Manual
Chapter 10. CUSTOMIZING THE SCREEN DISPLAY
10.1 QA Function Shortcut.......................................................................................
10.1.1 Registering QA Function Buttons ........................................................
10.1.2 Relocating QA Function Shortcut Buttons...........................................
10.1.3 Deleting QA Function Shortcut Buttons ..............................................
10.2 Exposure Parameter Shortcut ...........................................................................
10.2.1 Registering Exposure Parameter Shortcut Buttons ..............................
10.2.2 Relocating Exposure Parameter Shortcut Buttons ...............................
10.2.3 Deleting Exposure Parameter Shortcut Buttons...................................
10.3 QA Palette .......................................................................................................
10.3.1 Registering Function Buttons...............................................................
10.3.2 Relocating Function Buttons................................................................
10.3.3 Deleting Function Buttons ...................................................................
10.3.4 Editing Palette Name............................................................................
10.4 QA Image Controller ........................................................................................
10.4.1 Registering QA Image Controller Buttons...........................................
10-1
10-3
10-5
10-6
10-6
10-7
10-9
10-10
10-10
10-11
10-13
10-15
10-15
10-16
10-17
10-18
10.4.2 Relocating QA Image Controller Buttons............................................
10.4.3 Deleting QA Image Controller Buttons................................................
10.5 Patient Information Setting Items.....................................................................
10.5.1 Registering Patient Information Setting Items .....................................
10.5.2 Relocating Patient Information Setting Items ......................................
10.5.3 Deleting Patient Information Setting Items..........................................
10.6 Setting for Screen Display................................................................................
Chapter 11. CALCULATING THE COUNT HOW MANY TIMES THE IP
WAS USED
11.1 Searching by Narrowing Down the Query Conditions.....................................
11.2 Outputting List Files.........................................................................................
10-19
10-19
10-20
10-21
10-22
10-22
10-23
11-1
11-3
11-4
vi 2006. 10 897N0520C
CR Console Operation Manual
FUJI COMPUTED RADIOGRAPHY
Chapter
1
BEFORE OPERATION
2006. 10 897N0520C 1 - 1
CR Console Operation Manual

1.1 Functions of the User Utility

Summarized as follows are functions of the user utility. For details of the operation, refer to the pages shown in the parentheses.
Menu Setting
(For details, see [Chapter 2 MENU SETTING].)
Performs various settings related to exposure menus and study menus.
• Exposure Menu Setting
Creates, edits, and deletes exposure menus.
• Study Menu Setting
Creates, edits, and deletes study menus.
• Display Group Setting
Creates, relocates, renames, and deletes display groups of menus for the screen of routine operation mode, and also sets them not to be displayed on the screen.
• Display Menu Setting
Registers, relocates, and deletes menus to be displayed on the routine operation screen.
Table Setting
(For details, see [Chapter 3 TABLE SETTING].)
Sets information tables.
• Technologist Information Setting
Registers, edits, relocates, and deletes technologist information.
• Requesting Department Information Setting
Registers, edits, relocates, and deletes requesting department information.
• Film Mark Setting
Registers, edits, relocates, and deletes film marks.
• User Group Management
Manages user groups.
• Setting for Comment on Mis-exposure
Registers, edits, relocates, and deletes comments on mis-exposures.
Property Setting
Sets the operating environment.
Parameter Batch Updating
Updates exposure parameters for all exposure menus at a time. This function is only for super users, and general users cannot operate it.
(For details, see [Chapter 4 PROPERTY SETTING].)
(For details, see [Chapter 5 PARAMETER BATCH UPDATING].)
1 - 2 2006. 10 897N0520C
CR Console Operation Manual
Backup/Restore
(For details, see [Chapter 6 BACKUP/RESTORE].)
Performs backup/restore of information.
• Backup
Makes a backup of information.
• Restore
Restores information from a backup created.
Patient’s Information DB Utility
(For details, see [Chapter 7 PATIENT'S INFORMATION DB UTILITY].
Performs the maintenance/backup/restore operations of the patient’s information database.
• Patient’s Information Maintenance
Registers, edits and deletes patient’s information.
• Backup
Makes a backup of patient’s information.
• Restore
Restores patient’s information from a backup created.
)
• Patient’s Information Import
Imports patient’s information from external units, and registers them to the database at a time.
Image Processing Parameters File Transfer
PARAMETER FILE TRANSFER].)
(For details, see [Chapter 8 IMAGE PROCESSING
Transfers image processing parameters files recorded in the CR Console to Image Readers.
Exposure Result Log
(For details, see [Chapter 9 EXPOSURE RESULT LOG].)
Clears/saves the exposure result logs for completed studies. However, utilizing this function requires preset by the service engineer.
• Saving Logs
Exports and saves exposure result and mis-exposure logs for completed studies to a medium.
• Clearing Logs
Clears current exposure result and mis-exposure logs.
2006. 10 897N0520C 1 - 3
2010. 03 897N0520H
CR Console Operation Manual
Customizing the Screen Display
DISPLAY].)
(For details, see [Chapter 10 CUSTOMIZING THE SCREEN
Performs customization (a setting for the items to be displayed).
• QA Function Shortcut
Performs register/relocation/delete operations of the QA shortcut buttons to be displayed on the study execution screen.
• Exposure Parameter Shortcut
Performs register/relocation/delete operations of the exposure parameter shortcut buttons to be displayed on the study execution screen.
• QA Palette
Performs register/relocation/delete operations of the QA buttons to be displayed in the palette of the image edit screen.
• QA Image Controller
Performs register/relocation/delete operations of the QA image controller buttons to be displayed in the sub-palette of the image edit screen.
• Patient Information Setting Items
Performs register/relocation/delete operations of the patient information setting items to be displayed in the patient information input screen.
• Setting for Screen Display
Performs display setting for the study execution and study list screens.
Calculating the Count How Many Times the IP Was Used
CALCULATING THE COUNT HOW MANY TIMES THE IP WAS USED].)
Calculates the count how many times the IP was used.
• Searching by Narrowing Down the Query Conditions
Displays on the IP use count list the information of count how many times the IP was used, which is resultant of queries made by narrowing down the search conditions.
• Outputting List Files
Outputs in a “.csv” format file a list of calculated results of the count how many times the IP was used.
(For details, see [Chapter 11
1 - 4 2006. 10 897N0520C

1.2 Starting/Ending the User Utility

1.2.1 Starting the User Utility

CR Console Operation Manual
Starts up the User Utility. For details, refer to [
Chapter 3 TABLE SETTING
]. Before the operation, start up the CR Console to display the Study Reception Screen (or any of the lists).
Select
to display the menu, and select “User Utility”.
A confirmation box is displayed. Select “OK”. The mode for the routine operations is once ended, and the User Utility startup operation is executed.
(“Cancel” Cancels operation.)
2006. 10 897N0520C 1 - 5
2008. 09 897N0520F
CR Console Operation Manual
When the “Enter Password” window is displayed
If a password input is required because of the maintenance setting made at the time of installation, the “Enter Password - User Utility” window is displayed after the startup processing. Input a password and select “OK”.
When the password for general users is already input depending on the maintenance setting, simply select “OK”. If operating the “Parameter Batch Updating” function, start up the User Utility for super users. In this case, input a password for super users and select “OK”.
(“Terminate” → Terminates the User Utility.)
After the completion of starting up the User Utility, the “Main Menu - User Utility” window is displayed. (The “Parameter Batch Updating” menu will not be displayed for the User Utility of general users.)
1 - 6 2006. 10 897N0520C

1.2.2 Ending the User Utility

Ends the User Utility. Before the operation, display the “Main Menu - User Utility” window. Select “Terminate” in the “Main Menu - User Utility” window. The termination processing for User Utility will be executed.
CR Console Operation Manual
After completing the termination of the User Utility, the mode for routine operation is restarted and the initial screen will be displayed.
2006. 10 897N0520C 1 - 7
CR Console Operation Manual
1 - 8 2006. 10 897N0520C
CR Console Operation Manual
FUJI COMPUTED RADIOGRAPHY
MENU SETTING
Chapter
2
2006. 10 897N0520C 2 - 1
CR Console Operation Manual
This Chapter describes how to perform various settings related to exposure menus and study menus. Select “Menu Setting”.
NOTE
• If use of user authentication was set at the time of installation, the “Enter Password” window will appear when you select [Menu Setting]. After you enter the password, select [OK].
(“Cancel”
Cancels operation.)
2 - 2 2008. 09 897N0520F
CR Console Operation Manual
The “Menu Setting - User Utility” window is displayed. Exposure menus and study menus currently registered are listed in the “Exposure/Study Menu” field in the right side of the screen. Creating, editing, and deleting menus are performed in this field. Exposure menus and study menus are listed by group (in the same manner as the display style in a screen for the routine operation mode) in the “Display setting” field at the right side of the screen. Settings related to display of menus and groups are made in this field. For details of creating new menus, see [
Study Menus
For details of display settings, see [
].
2.4.1 Registering Display Menus
2.1.1 Creating Exposure Menus
].
] and [
2.2.1 Creating
Operations are described by the setting, hereafter.
2008. 09 897N0520F 2 - 3
CR Console Operation Manual

2.1 Exposure Menu Setting

Creates, edits, and deletes exposure menus.

2.1.1 Creating Exposure Menus

Creates an exposure menu by copying the information of an existing exposure menu and updating it.
(1) Selecting an exposure menu
Select “Exposure Menu” in the “Exposure/Study Menu” field.
The exposure menus currently registered are listed. Select an exposure menu to be copied. An Exposure menu (button) can be selected in the “Display Setting” field as well. (In that case, the operation above is not necessary.)
NOTE
• Select an exposure menu with the exposure style (anatomical region, positioning) as similar to the exposure menu to be created as possible. Creating an exposure menu for one region by copying that for another (e.g. creating an exposure menu for head by copying that for chest) may lead to an abnormal image.
2 - 4 2008. 09 897N0520F
CR Console Operation Manual
(2) Selecting the processing
Select “New”.
(3) Setting code/name
The “Exposure Menu Setting - User Utility” window is displayed. Set menu codes and a menu name. (Set all items here, otherwise users cannot go to next operation.)
2008. 09 897N0520F 2 - 5
CR Console Operation Manual
Extended Menu Code :
The extended menu code is the code for managing exposure menus. It does not decide EDR (automatic image reading sensitivity setting function) and image processing parameters for images not like the following MPM code. EDR and the image processing parameters are the same as ones for the original exposure menu to be copied. An unused code is set in default. To change the code, enter any code (a maximum of 16 characters, * and @ cannot be used). (Duplication with codes of other exposure menus is not allowed.)
Example> In the case of creating different exposure menus such as “CERVICAL,OBL”/ “CERVICAL,LAT”/“CERVICAL,RAO”/“CERVICAL,LAO” by copying a “CERVICAL SPINE” (MPM code: 0107) menu. (EDR and the image processing parameters are not to be changed.)
Image
Processing
Parameters
Remarks
The exposure parameters (Invert Image, Film Mark, and etc.) can be set differently for each exposure menu. The image processing parameters for the newly created exposure menu are changed when the parameters for the original exposure menu are changed.
Original
exposure
menu
Newly
created
exposure
menu
Menu Name
CERVICAL SPINE 0107 0107-0000
CERVICAL, OBL 0107 0107-0001
CERVICAL, LAT 0107 0107-0002
CERVICAL, RAO 0107 0107-0003
CERVICAL, LAO 0107 0107-0004
MPM Code
Extended
Menu Code
EDR
Same Same
MPM Code :
The MPM code is the code that manages exposure menus and also decides EDR and image processing parameters for images. Users can change the third digit of the code since it does not affect EDR and image processing parameters, but cannot change any other digits. (That is, original exposure menus decide the EDR and image processing parameters for newly created exposure menus.)
Example)
The The code of the original exposure menu to be copied is set in default. Select a code from the dropdown list. If the selected code has already used in other exposure menus, the exposure menu is displayed in the “Related Menu” box.
Example> In the case of creating a new “PELVIS,OBL” menu with changed image processing
parameters and exposure parameters by using a “PELVIS,FRN” (MPM code: 0500) menu. Original exposure menu → PELVIS,FRN (MPM code: 0500) Newly created exposure menu PELVIS,OBL (MPM code: 05x0, “x” is an optional
character to be set.)
Exposure Menu Name :
Enter the exposure menu name (a maximum of 64 characters, * and @ cannot be used).
2 - 6 2008. 09 897N0520F
CR Console Operation Manual
(4) Setting exposure parameter
Select the “Exposure Parameter 1”, “Exposure Parameter 2” or “Exposure Parameter 3” tab to display the folder, and set each exposure parameter.
<Exposure Parameter 1>
Invert Image :
Select an invert display of the image. Normally “Frontal Image” and “Vertical” can be selected. If making “Horizontal” and “Horizontal and Vertical Rotation” selectable, the maintenance setting is required.
Rotate Image :
Rotation display of the image. This item is normally fixed to “None”. If making “Rotate 90 degree” selectable, the maintenance setting is required.
Simplified Image Processing Condition (SS/CS) :
Set CS (contrast shift) and SS (density shift). The value can be changed by selecting the arrow button. It cannot be changed by entering the value through keyboard. CS is the coefficient given to the parameter concerning the tonal curve. It changes the contrast of the image. This value can be set in the range of *(x)0.5 to *2.0. SS is the coefficient given to the parameter concerning the reading sensitivity. It changes the density of the image. This value can be set in the range of /(÷)2.0 to *(x)2.0.
EDR Mode :
Select an EDR (automatic image reading sensitivity setting function) mode. If setting “Fix” mode, enter “S value” (4 to 20047). “S Value” can be changed either by selecting the arrow button or by entering the value through keyboard. If performing the Manual EDR processing, select “Manual”. It is only effective for the Image Readers that can perform the Manual EDR processing.
EIt Value :
Set an EIt value when the EI/DI display is enabled. The EI and the DI (defined in IEC 62494-1:2008) on the “Study Screen” and “Image Edit Screen” are calculated based on the EIt value that has been set in this field.
2010. 10 897N0520J 2 - 7
CR Console Operation Manual
Film Output Format :
Select a film output format.
Image Position :
Select a display position of an image in the image frame.
Film Mark :
Set a film mark (character displayed on the film). This item can be set either by selecting from the dropdown list or by entering characters (a maximum of ten characters) through keyboard. The film marks set separately are listed up in the dropdown list. Refer to [ about the film mark settings.
Distribution Code :
Select a distribution code (a code number to specify the transfer destination) of the image from the dropdown list.
Function :
Select a function (special processing) from the dropdown list.
Routine processing
RT :
Temporal subtraction processing
TS :
Whole spine processing
WS :
Energy subtraction processing
ES :
IVP processing
PS :
Tomography processing
TM :
3.3 Film Mark Setting
]
Usually it is set to the exposure menus for the Image-composition Processing (optional software). If it is set when the user’s unit does not have the image-composing function, the image processing defined to this function is only applied to the exposure menus.
No. of Output Film :
Set the number of films to output. This value can be changed by using arrow buttons or entered by using a keyboard directly. The number of films that can be set varies depending on the setting at the time of installation. If films are not output to external devices other than a printer, do not set this item to “0”.
X-ray Tube/Operational Method Code :
It is used when an X-ray control unit is connected.
2 - 8 2010. 10 897N0520J
<Exposure Parameter 2>
CR Console Operation Manual
Reading Sensitivity Center Shift :
Select a reading sensitivity to the image exposed with high sensitivity (low X-ray dose). Select “high Sensitivity” for images with S value = 800 or higher.
Film Reference Orientation :
Select reference direction to output image on film. This parameter is valid only when using the 8" x 10" 1-image format and the 14" x 14" 2-image format, select “Portrait” for outputting image on film shifted by 90 degrees.
Film Edge :
Select transparent or black for film edge (frame of the image).
Pixel Density :
Select pixel density of the image.
Filing Mode :
Select a mode to be used for filing an image to a filing device, such as ODF. Basically, "OFF" (not to record)/ "ON" (to record) can be selected. "ON*" is not allowed to select. (This parameter is inactivated when no filing device is connected.)
Film Size :
Select a film size. This parameter is used as the exposure actual result information to be sent to the RIS Terminal Units. It does not affect the film size to be used actually. A film size that can be selected will differ depending the settings performed at installation.
Auto Repeat :
Select either of “OFF” (do not perform) / “ON” (perform) for the auto repeat exposures.
2008. 09 897N0520F 2 - 9
CR Console Operation Manual
Number of division exposure :
Set the number of division exposure. This value can be changed by using arrow buttons or entered by using a keyboard directly.
Coding Scheme Designator :
Enter the name of coding scheme designator (a maximum of 16 characters, * and @ cannot be used). Normally, the name of this field does not have to be changed.
<Exposure Parameter 3>
Reading Mode :
Select a desired reading mode used to read images on the Image Reader.
FR : Five lines/mm reading mode SR : Ten lines/mm reading mode
GR : Ten lines/mm GR reading mode “FR: Five lines/mm reading mode” reads images in high-speed mode. Shown below are the IP sizes that can be read in high-speed mode.
• 14”x17”
• 14”x14” Whether FR mode can be selected or not depends on the connected Image Reader. Even if FR mode is selected on Image Readers not available for that mode, images will be read in SR mode instead. There are cases where trimming processing cannot be applied on images read by FR mode.
GPR processing:
When the GPR option is activated, this processing allows OFF (disabled) or ON (enabled) to be selected for GPR (Grid Pattern Removal processing).
Sensitivity level for the VELOCITY :
Select IP erasure mode when [Auto sensitivity change] is specified on the Image Reader FCR VELOCITY U, FCR VELOCITY T or FUJIFILM Unity SpeedSuite. Standard exposure : Erases the IP in high-speed mode. High sensitivity (low X-ray dose) exposure: Erases the IP in standard erasure mode.
2 - 10 2008. 09 897N0520F
(5) Setting display color of the exposure menu
Select “Display Color Setting” in the “Exposure Parameter 1” folder.
The “Display Color Setting - User Utility” window is displayed. Select a color in which exposure menu is to be displayed.
(“Cancel” Cancels operation.)
CR Console Operation Manual
Besides those displayed, an optional color can be selected. Press the “Arbitrarily selectable” button.
The “Color” window is displayed. After selecting a color, select “OK”.
(“Cancel” Cancels operation.)
2008. 09 897N0520F 2 - 11
CR Console Operation Manual
(6) Setting image processing parameter
Select “Image Processing Parameter Setting” in the “Exposure Parameter 1” folder.
The “Image Processing Parameter Setting - User Utility” window is displayed. Set each image processing parameter, and then select “OK”. In case of “DRC” and “TAS” processing, either one or the other is operational at one time. For details of image processing parameters, see [
(“Cancel” Cancels operation.)
-Image processing parameter-
] presented herein.
Specify Image :
Select an image to which the setting of the image processing parameter is to be applied.
GP RP/MFP :
Select GP (Tonal Conversion Processing), RP (Spatial Frequency Processing) or MFP (Multi­objective Frequency Processing). MFP can be set when the MFP option is activated.
DRC :
Select ON (enabled) or OFF (disabled) of DRC (Dynamic Range Control processing).
TAS :
Select ON (enabled) or OFF (disabled) of TAS (Tomographic Artifacts Suppression processing).
2 - 12 2008. 09 897N0520F
CR Console Operation Manual
FNC :
Select ON (enabled) or OFF (disabled) of FNC (Frequency Noise Control processing).
Return to default :
If this button is selected, parameter value of the Fuji Standard is set. If an exposure menu of the Fuji Standard is not displayed in the “Related Menu” box. This button is disabled.
NOTES
• To change image processing parameters, consult a physician who is going to read images, and then verify changed parameters on multiple images. Moreover, if image processing parameters thus changed affect image reading adversely, adjust those parameters appropriately.
• There is apprehension that depending on the exposure conditions applied, subject to be exposed or X-ray generator used for exposure, image processing task becomes burdensome, which causes then artifacts to occur, resulting consequently in undesirable effects in image reading.
2008. 09 897N0520F 2 - 13
CR Console Operation Manual
(7) Setting IP logical size
Select the “IP Logical Size” tab to display the folder and set size for IP logical reading.
FCR Cassette Type :
Select an IP logical reading size for the cassette type Image Reader from the dropdown list.
FCR Upright Built-in Type :
Select an IP logical reading size for the built-in type Image Reader for standing figures from the dropdown list and select “IP Reading Reference Position”.
FCR Bed Built-in Type :
Select an IP logical reading size for the built-in type Image Reader for laying figures from the dropdown list.
Velocity U / Velocity Ufp:
Select an IP logical reading size for the FCR VELOCITY U or Velocity Ufp Image Reader from the dropdown list. Select then “IP Reading Reference Position”.
Velocity T / Velocity Tfp / Unity :
Select an IP logical reading size for the FCR VELOCITY T, Velocity Tfp or FUJIFILM Unity SpeedSuite Image Reader from the dropdown list.
2 - 14 2008. 09 897N0520F
(8) Trimming Setting
Select the “Trimming” tab to display the folder and set trimming processing. (The “Trimming” tab will be displayed only when the trimming option has been activated.)
CR Console Operation Manual
Trimming Size :
Specify one trimming size for the selected exposure menu.
Direction of Image Extraction :
Specify whether to extracting an image to be trimmed in the specified size vertically or horizontally.
Trimming Reference Position :
Select a reference position when subjecting an image to trimming processing.
NOTE
• Trimming processing is not available in the cases a) to c) below, even if the size of a read image is the same as that of the image before it has been subjected to trimming processing as specified in “Trimming Size:” above. a) Images read from the FCR 5000 series at standard resolution b) Composed images or source images for composition c) Energy subtraction images d) Some of the images read in FR mode
• With regard to an image that has been subjected to trimming processing to 10”x14” or 14”x10” size, do not change image magnification ratio or image rotation direction on the image display unit side. Because the trimming frame position information varies as the image magnification ratio or image rotation direction changes, 10”x14” or 14”x10” sized images cannot be output correctly even if they are reprinted via the CR Console.
• Modifying an exposure menu or replacing an image after image reading processing will retain the trimming setting determined before such a change, resulting thus in disabling retrimming processing at initial menu setting after such a change.
• Specified trimming processing will not be reflected on images extracted in a size such as 10”x12” or 8”x10” and subjected to filing and soft copy at standard resolution.
2008. 09 897N0520F 2 - 15
CR Console Operation Manual
(9) Setting the allowable S value and L value ranges
Select “Setting the Allowable S/L Value Range”. The “Setting the Allowable S/L Value Range ­User Utility” window opens.
On this window, you can set up allowable S and L value ranges for exposure menus to be created and edited.
S value :
Maximum and minimum S values are determined for each exposure menu.
The range including both the maximum and minimum values will be taken as the allowable range for the S value.
L value :
Maximum and minimum L values are determined for each exposure menu.
The range including both the maximum and minimum values will be taken as the allowable range for the L value.
Select “OK”.
(“Cancel” Cancels operation.)
An attempt to have an image, whose S and L values exceed the maximum and minimum values determined here, be read on the CR Console will display a message like that shown below.
Select “OK”.
2 - 16 2008. 09 897N0520F
CR Console Operation Manual
The target image is then read on the study screen. When it is read, the S value and L value displayed in the study menu blinks for the purpose of warning.
If you select
, a message like that shown below appears.
If you select “Finish”, the study ends here. The S value displayed for the purpose of warning then stops blinking.
2008. 09 897N0520F 2 - 17
CR Console Operation Manual
Furthermore, when this processing ends as the “Finish” button is pressed, a warning mark is displayed to the related study on the “Local WL” list, as shown on the window below.
To use the S value/L value warning function, perform Image Reader calibration regularly. When doing so, use the Image Reader QC tool so that the calibration is performed unfailingly. If images are read on an Image Reader not undergone due calibration, the warning can be displayed illegally when the S value/L value warning function is used.
(10) Completing creation
Select “OK”.
(“Cancel” Cancels operation.)
The created exposure menu is not displayed on the routine operation screen until registering the display of the menu. To continue to register the display of the exposure menu, perform the operation in [
2.4.1 Registering Display Menus
].
2 - 18 2008. 09 897N0520F
Image processing parameter −
The following shows characteristics of each parameter.
GP (Tonal Conversion Processing) Parameter
Parameter Characteristic Input Range
CR Console Operation Manual
GA
GT
GC
GS
Inclination of the tonal curve. Adjusts the contrast of images.
Shape of the tonal curve. Attains the same effect as changing γ curve of various X-ray films.
Central density of inclination of the tonal curve. Changes “GA” centering on the density determined here.
Amount of parallel translation of the tonal curve. Changes display density of the entire image.
–4.1 to –0.1
0.1 to 4.0 (the first decimal place excluding 0.0)
A to Z (26 types)
0.3 to 2.64 (the second decimal place)
–1.44 to 1.44 (the second decimal place)
RP (Spatial Frequency Processing) Parameter
Parameter Characteristic Input Range
0 to 9 (10 ranks)
F, P to X (10 types)
RN
RT
Enhanced frequency band in frequency processing. The frequency band determined here becomes the subject of enhancement processing.
Shape of enhancement curve according to the density of images. By changing RT, the level of enhancement in the specific density area can be adjusted to the area with a specific density.
RE
Level of enhancement of frequency processing. Changes the strength of processing according to the value.
0.0 to 9.9 (the first decimal place) 10 to 16 (no decimal place)
2008. 09 897N0520F 2 - 19
CR Console Operation Manual
DRC (Dynamic Range Control Processing) Parameter
Parameter Characteristic Input Range
0 to 9 (10 sizes)
A to T (20 types)
0.0 to 2.0 (the first decimal place)
DRN
DRT
DRE
Mask size of dynamic range control. Sets the area for processing along with “DRT”.
Type of dynamic range control filter. Sets the density area for processing.
Level of enhancement of dynamic range control filter processing. Changes the strength of processing according to the value.
TAS (Tomographic Artifacts Suppression Processing) Parameter
Parameter Characteristic Input Range
0 to 9 (10 sizes)
0 : The orbit is at right angle to the green
line of the cassette.
1 : The orbit is parallel to the green line
of the cassette.
0.0 to 9.9 (the first decimal place) 10 to 16 (no decimal place)
ORN
ORD
ORE
Mask size of tomographic artifacts suppression. Sets the most effective size according to the tomographic angle.
Processing direction of tomographic artifacts suppression. Sets according to the reading direction and the orbital direction of tomography.
Level of enhancement of tomographic artifacts suppression. Changes the strength of processing according to the value.
FNC (Frequency Noise Control Processing) Parameter
Parameter Characteristic Input Range
Depending on the irradiated X-ray dose, determines
FFC
FNB
FNT
FNE
the degree of FNC to be applied. A: For general-purpose exposures, B: For linac graphy.
Determines the degree of FNC to be applied with regard to the frequency components. Low-frequency control will be enhanced gradually depending on how it is close to A.
Determines the degree of FNC to be applied with regard to the image density.
Determines the degree of FNC that is to be applied. The degree of FNC applied will be changed depending on the entered value.
A to Z (26 types)
A to F (6 types)
A to C (3 types)
0.0 to 1.0 (first decimal point)
2 - 20 2008. 09 897N0520F
MFP (Multi-objective Frequency Processing) Parameter (option)
Parameter Characteristic Input Range
CR Console Operation Manual
A to Z (26 types)
MRB
Enhanced frequency band in frequency processing. The frequency band determined here is targeted for enhancement processing.
F, P to Z (12 types)
MRT
Shape of enhancement curve according to the density of images. By changing MRT, the level of enhancement in the specific density area can be adjusted.
MRE
MDB
Level of enhancement of frequency processing. Changes the strength of processing according to the specified value.
Frequency type applied to the DR compression filter.
0.0 to 9.9 (the first decimal place) 10 to 16 (no decimal place)
A to Z (26 types)
Sets processing area, in conjunction with “MDT”.
A to T (20 types)
MDT
Type applied to the DR compression filter. Sets a density area where processing is to be performed.
MDE
Level of enhancement of DR compression filter processing. Changes the strength of processing according to the
0.0 to 2.0 (the first decimal place)
specified value.
Parameters that correspond to GP, such as GA, GT, GC and GS, are gradation parameters used for MFP processing.
For the characteristics, see descriptions in the “GP” (Tonal Conversion Processing) Parameter.
2008. 09 897N0520F 2 - 21
CR Console Operation Manual

2.1.2 Editing Exposure Menus

Edits information of exposure menus.
NOTE
• To edit FUJI Standard exposure menu information, be sure to subject to editing an exposure menu copied using the same procedure as that of creating a new exposure menu. If original FUJI standard exposure menus are subjected to editing, there may be cases where the CR Console does not operate normally.
(1) Selecting the exposure menu
Select “Exposure Menu” in the “Exposure/Study Menu” field.
The exposure menus currently registered are listed. Select an exposure menu to be edited. Exposure menu (button) can be selected in the “Display Setting” field as well. (In that case, the operation above is not necessary.)
(2) Selecting the processing
Select “Edit”.
2 - 22 2008. 09 897N0520F
(3) Various settings
The “Exposure Menu Setting - User Utility” window is displayed. To edit exposure menus, use the procedures in [
(3) Setting code/name through (9) Setting the
allowable S value and L value ranges under 2.1.1 Creating Exposure Menus
CR Console Operation Manual
].
(4) Completing editing
Select “OK”.
(“Cancel” Cancels operation.)
2008. 09 897N0520F 2 - 23
CR Console Operation Manual

2.1.3 Deleting Exposure Menus

Deletes exposure menus.
(1) Selecting the exposure menu
Select “Exposure Menu” in the “Exposure/Study Menu” field.
The exposure menus currently registered are listed. Select an exposure menu to be deleted.
NOTE
• Once the Fuji Standard exposure menu is deleted, it cannot be restored. (To restore exposure menus created by the user, they must be re-created.)
• Exposure menu which has been registered to be displayed cannot be deleted. To delete such exposure menus, perform in advance the operation described in [
2.4.3 Deleting Display Menus
Exposure menus registered in study menus cannot be deleted, either.
(2) Selecting the processing Completing deletion
Select “Delete”.
].
2 - 24 2008. 09 897N0520F

2.2 Study Menu Setting

p
y
Creates, edits, and deletes study menus. The study menu is the menu to which multiple exposure menus for study are registered together. For example, once the study menu named “THORA.SPINE 2R” to which “THORA.SPINE,FRN” and “THORA.SPINE,LAT” are registered is created, only by selecting the study menu, the necessary exposure menus can be ready for the study. As it is not necessary to select exposure menus every time performing the same study, this increases a work efficiency.
<Operation Flow>
In the case of study menus
Select “THORA.SPINE 2R”.
Menus for “THORA.SPINE,FRN” and “THORA.SPINE,LAT” are
anded.
ex
In the case of exposure menus
CR Console Operation Manual
Select “THORA.SPINE,FRN”.
Process “THORA.SPINE,FRN”.
Process “THORA.SPINE,FRN”.
When the processing is completed, “THORA.SPINE,LAT” is selected automaticall
Process “THORA.SPINE,LAT”.
All processing is completed.
.
Select “THORA.SPINE,LAT”.
Process “THORA.SPINE,LAT”.
All processing is completed.
2008. 09 897N0520F 2 - 25
CR Console Operation Manual

2.2.1 Creating Study Menus

Creates study menus.
(1) Selecting the processing
Select “Study Menu” in the “Exposure/Study Menu” field.
Select “New”.
2 - 26 2008. 09 897N0520F
CR Console Operation Manual
(2) Setting code/name
The “Study Menu Setting - User Utility” window is displayed. Set a menu code and a menu name. (Set both items here, otherwise users cannot go to next operation step.)
Setting menu code and menu name
Study Menu Code :
The study menu code is the code to manage study menus. Enter the desired code (a maximum of 16 characters, * and @ cannot be used). (Duplication with codes of other study menus is not allowed.)
Study Menu Name :
Enter the study menu name (a maximum of 64 characters, * and @ cannot be used).
2008. 09 897N0520F 2 - 27
CR Console Operation Manual
(3) Selecting the exposure menu
Select an exposure menu to be registered in the “Exposure Menu List”, and then select
To select more than one (a maximum of 48 menus) exposure menu, repeat this operation.
.
The selected exposure menus are listed in the “Selected Menu”, and an exposure order reflects the order of the selection. To cancel the selected exposure menu, select that exposure menu , and then select
To change the order of a selected exposure menu, select the exposure menu and then select
upwards)/
(moves downwards).
.
(moves
2 - 28 2010. 10 897N0520J
(4) Setting film output format
Select a film output format.
CR Console Operation Manual
Default :
2-on-1 :
4-on-1 :
Outputs the image on a film in the format set for each exposure menu. Outputs the image on a film in 2-on-1 format. Outputs the image on a film in 4-on-1 format.
NOTE
• When 2-on-1 or 4-on-1 format is set, the film output format cannot be changed in the routine operation.
2008. 09 897N0520F 2 - 29
CR Console Operation Manual
(5) Setting film output position
When “2-on-1” or “4-on-1” is set for the film output format, set an output position of image on film. Select “Film Output Position Setting”.
The “Film Output Position Setting - User Utility” window is displayed. Select “OK” after completing the following operation.
(“Cancel” Cancels operation.)
2 - 30 2008. 09 897N0520F
CR Console Operation Manual
Relocating images :
Relocates images to desired positions. Display a page that has an image whose position is to be changed (
/ : page turn button),
select the image, and then move the image to the objective position. The selected image will be replaced to an existing image in the transfer destination. Passing on the page turn button on the way to the blank button makes page turn accordingly.
(“Set Order of Exposure” Relocates images in exposure order.)
2008. 09 897N0520F 2 - 31
CR Console Operation Manual
Re-arranging after release of all image disposition :
Releases a current image disposition and re-arranges image positions. Select “Cancel All”. The current entire image disposition will be released.
Display the desired page for which an image will be located (
/ : page turn button), select an
exposure menu for the image from the “Registered Exposure Menu” field, and then move it to the objective position. The image will be located. Passing on the page turn button on the way to the blank button makes page turn accordingly.
(“Set Order of Exposure” Relocates images in exposure order.)
Repeat this operation to arrange all images. Confirm that “Registered” is displayed for all exposure menus at the “Registered Exposure Menu” field after the operation.
2 - 32 2008. 09 897N0520F
Deleting images :
Deletes arranged images. Display a desired page that has an image to be deleted (
image, and then move it to
. The image will be deleted.
CR Console Operation Manual
/ : page turn button), select the
After deletion, arrange all images by performing either operation of the “Relocating images” or “Re-arranging after release of all image disposition”. For how to arrange all images, see [
Relocating images: in 2.2.1 Creating Study Menus]
[Re-arranging after release of all image disposition: in 2.2.1 Creating Study Menus
].
and
Confirm that “Registered” is displayed for all exposure menus at the “Registered Exposure Menu” field after the operation.
2008. 09 897N0520F 2 - 33
CR Console Operation Manual
(6) Setting display color
Select “Display Color Setting”.
The “Display Color Setting - User Utility” window is displayed. Select a color for displaying study menu name.
(“Cancel” Cancels operation.)
Besides those displayed, an optional color can be selected. Press the “Arbitrarily selectable” button.
The “Color” window is displayed. After selecting a color, select “OK”.
(“Cancel” Cancels operation.)
2 - 34 2008. 09 897N0520F
CR Console Operation Manual
(7) Setting trimming processing
Select “Trimming”.
The “Setting the Trimming for Image Extraction - User Utility” dialog then opens. (The “Trimming” button can be selected only when the trimming option has been activated.)
Trimming Size :
Specify one trimming size for the selected study menu.
Direction of Image Extraction :
Specify whether to extracting an image to be trimmed in the specified size vertically or horizontally. Select “OK” after you have specified a trimming size and a direction of image extraction.
(“Cancel” Cancels operation.)
2008. 09 897N0520F 2 - 35
CR Console Operation Manual
(8) Completing creation
Select “OK”.
(“Cancel” Cancels operation.)
The created study menu is not displayed on the Routine Operation Screen until registering the display of the menu. To continue to register the display of the exposure menu, perform the operation in [
2.4.1 Registering Display Menus
].
2 - 36 2008. 09 897N0520F

2.2.2 Editing Study Menus

Edits information of study menus.
(1) Selecting the study menu
Select “Study Menu” in the “Exposure/Study Menu” field.
The study menus currently registered are listed. Select a study menu to be edited. Study menu (button) can be selected in the “Display Setting” field as well. (In that case, the operation above is not necessary.)
CR Console Operation Manual
(2) Selecting the processing
Select “Edit”.
2008. 09 897N0520F 2 - 37
CR Console Operation Manual
(3) Various settings
The “Study Menu Setting - User Utility” window is displayed. To edit study menus, use the procedures in [
processing under 2.2.1 Creating Study Menus
(2) Setting code/name through (7) Setting trimming
].
(4) Completing editing
Select “OK”.
(“Cancel” Cancels operation.)
2 - 38 2008. 09 897N0520F

2.2.3 Deleting Study Menus

Deletes study menus.
(1) Selecting the study menu
Select “Study Menu” in the “Exposure/Study Menu” field.
The exposure menus currently registered are listed. Select a study menu to be deleted.
CR Console Operation Manual
NOTE
• A study menu which was registered to be displayed cannot be deleted. To delete such a study menu, perform in advance the operation described in [
2.4.3 Deleting Display Menus
(2) Selecting the processing Completing deletion
Select “Delete”.
].
2008. 09 897N0520F 2 - 39
CR Console Operation Manual

2.3 Display Group Setting

Creates, relocates, renames, and deletes display groups of menus for the screen of routine operation mode, and also sets them not to be displayed on the screen.

2.3.1 Relocating Display Groups

Relocates display groups. Select the tab of a display group to be relocated, and then move it to left or right.

2.3.2 Renaming Display Groups

Renames display groups.
(1) Selecting the display group
Select the tab of a display group to be renamed.
(2) Setting name → completing rename
Select “Edit Display Group”.
2 - 40 2008. 09 897N0520F
The “Display Group Setting - User Utility” window is displayed. Enter the display group name (a maximum of 64 characters, * and @ cannot be used), and then select “OK”.
(“Cancel” Cancels operation.)

2.3.3 Deleting Display Groups

Deletes display groups.
CR Console Operation Manual
(1) Selecting the display group
Select the tab of a display group to be deleted.
2008. 09 897N0520F 2 - 41
CR Console Operation Manual
(2) Setting the processing completing deletion
Select “Edit Display Group”.
The “Display Group Setting - User Utility” window is displayed. Select “Delete”.
A confirmation box is displayed. Select “OK”.
(“Cancel” Cancels operation.)
2 - 42 2008. 09 897N0520F

2.3.4 Setting Non-display to Display Groups

Sets display groups not to be displayed.
(1) Selecting the display group
Select the tab of a display group not to be displayed.
(2) Selecting the processing completing setting
CR Console Operation Manual
Select “Change Display”. The folder of the selected display group is inactivated (displayed in dark color). To release non-display, select “Change Display” again.
2008. 09 897N0520F 2 - 43
CR Console Operation Manual

2.4 Display Menu Setting

Registers, relocates, and deletes menus to be displayed on the routine operation screen.

2.4.1 Registering Display Menus

Registers display menus.
(1) Selecting the display group
Select the tab of a display group to which the menu is to be registered, and then display a page which has a blank button (
/ : page turn button).
(2) Executing registration
Select the menu to be registered from the list in the “Exposure/Study Menu” field, and then move it onto the desired blank button. Passing on the page turn button on the way to the blank button makes page turn accordingly.
Page is turned when passing here.
Moves to other page’s buttons.
Moves to current page’s buttons.
2 - 44 2008. 09 897N0520F

2.4.2 Relocating Display Menus

Relocates display menus.
(1) Selecting the display group
Select the tab of a display group to which the menu to be relocated is registered, and then display the page to which the menu is registered (
CR Console Operation Manual
/ : page turn button).
(2) Executing relocation
Select the menu to be located, and then move it onto the desired blank button. If the menu is moved onto a menu on which a menu is already displayed, the moved menu and the existing menu switch their locations. Passing on the page turn button on the way to the desired button makes page turn accordingly.
Page is turned when passing here.
Moves to other page’s buttons.
Moves to current page’s buttons.
2008. 09 897N0520F 2 - 45
CR Console Operation Manual

2.4.3 Deleting Display Menus

Deletes display menus.
(1) Selecting the display group
Select the tab of a display group to be deleted is registered, and then display the page to which the menu is registered (
/ : page turn button).
(2) Executing deletion
Select the menu to be delet1ed, and then move it onto
.
2 - 46 2008. 09 897N0520F
CR Console Operation Manual
FUJI COMPUTED RADIOGRAPHY
Chapter
3
TABLE SETTING
2006. 10 897N0520C 3 - 1
CR Console Operation Manual
This Chapter describes settings for information tables.
Select “Table Setting”.
The “Table Setting Menu - User Utility” window is displayed. Perform the operation of each menu, and then select “OK”.
Operations are described by the menu, hereafter.
3 - 2 2006. 10 897N0520C

3.1 Technologist Information Setting

Registers, edits, relocates, and deletes technologist information.
Select “Technologist Information Setting”.
CR Console Operation Manual
The “Technologist Information Setting - User Utility” window is displayed. Perform the operation described as follows, and then select “OK”.
(“Cancel” Cancels operation.)
2006. 10 897N0520C 3 - 3
CR Console Operation Manual
NOTE
• When the user authentication has been specified to be used according the Maintenance Utility settings, the “Enter Password - User Utility” dialog will open if you select “Technologist Information Setting”.
Enter a user password registered as the “Administrator” and then select “OK”. The “Technologist
Information Setting - User Utility” window will be displayed.
(“Cancel” → Cancels operation.)

3.1.1 Registering Technologist Information

Registers technologist information.
(1) Selecting the processing
Select “Register”.
(2) Setting code/name → Completing registration
The “Technologist Information Input - User Utility” window is displayed. Set necessary items accordingly, and then select “OK”.
(“Cancel” Cancels operation.)
3 - 4 2006. 10 897N0520C
CR Console Operation Manual
Technologist Code :
Enter the technologist code (a maximum of two characters, * and @ cannot be used). (Duplication with codes of other technologists is not allowed.)
Technologist Name :
Enter the technologist name (a maximum of 64 characters, and * cannot be used). (Duplication with names of other technologists is not allowed.)
User Name :
Enter a user name to be used when logging into the system (a maximum of 64 characters, and * cannot be used). (Duplication with codes of other technologists is not allowed.) Note that a box where a user name is to be set up will appear when the user authentication has been specified to be used according to the Maintenance Utility settings.
Password :
Enter a password to be used when logging into the system (a maximum of 20 characters). (Duplication with codes of other technologists is not allowed.) Note that a box where a password is to be set up will appear when the user authentication has been specified to be used according to the Maintenance Utility settings.
User Group :
Select from the dropdown list a user group to be assigned to the technologist. Available capabilities are limited by the function allowed to the user group. For details of this function, see [
3.4 Use Group Management
]. Note that a box where a user group is to be set up will appear when the user authentication has been specified to be used according to the Maintenance Utility settings.
NOTE
If more than 150 technologist names are registered, the display of the Retake Analysis function may slow down. In this case, delete unused technologist information after confirmation. For details, refer to [Chapter 3 TABLE SETTING].
2008. 03 897N0520E 3 - 5
CR Console Operation Manual

3.1.2 Editing Technologist Information

Edits technologist information.
(1) Selecting the technologist code
Select the code of the technologist information to be edited.
(2) Selecting the processing
Select “Edit”.
(3) Setting code/name → Completing editing
The “Technologist Information Input - User Utility” window is displayed. Set necessary items accordingly, and then select “OK”.
(“Cancel” Cancels operation.)
3 - 6 2006. 10 897N0520C
CR Console Operation Manual
Technologist Code :
Enter the technologist code (a maximum of two characters, * and @ cannot be used). (Duplication with codes of other technologists is not allowed.)
Technologist Name :
Enter the technologist name (a maximum of 64 characters, and * cannot be used). (Duplication with names of other technologists is not allowed.)
User Name :
Enter a user name to be used when logging into the system (a maximum of 64 characters, and * cannot be used). (Duplication with codes of other technologists is not allowed.) Note that a box where a user name is to be set up will appear when the user authentication has been specified to be used according to the Maintenance Utility settings.
Password :
Enter a password to be used when logging into the system (a maximum of 20 characters). (Duplication with codes of other technologists is not allowed.) Note that a box where a password is to be set up will appear when the user authentication has been specified to be used according to the Maintenance Utility settings.
User Group :
Select from the dropdown list a user group to be assigned to the technologist. Available capabilities are limited by the function allowed to the user group. For details of this function, see [
3.4 Use Group Management
]. Note that a box where a user group is to be set up will appear when the user authentication has been specified to be used according to the Maintenance Utility settings.
2006. 10 897N0520C 3 - 7
CR Console Operation Manual

3.1.3 Relocating Technologist Information

Relocates technologist information (technologist codes).
(1) Selecting the technologist code
Select the code of the technologist information to be relocated.
(2) Performing relocation
Select
(moves upwards)/ (moves downwards) to relocate the information.

3.1.4 Deleting Technologist Information

Deletes technologist information.
(1) Selecting the technologist code
Select the code of the technologist information to be deleted.
(2) Selecting the processing Completing deletion
Select “Delete”.
3 - 8 2006. 10 897N0520C

3.2 Requesting Department Information Setting

Registers, edits, relocates, and deletes requesting department information.
Select “Requesting Department Information Setting”.
CR Console Operation Manual
The “Requesting Department Information Setting - User Utility” window is displayed. Perform the operation described as follows, and then select “OK”.
(“Cancel” Cancels operation.)
2006. 10 897N0520C 3 - 9
CR Console Operation Manual

3.2.1 Registering Requesting Department Information

Registers requesting department information.
(1) Selecting the processing
Select “Register”.
(2) Setting code/name → Completing registration
The “Requesting Department Information Input - User Utility” window is displayed. Set the requesting department code and requesting department name, and then select “OK”.
(“Cancel” Cancels operation.)
Requesting Department Code :
Enter the requesting department information code (a maximum of 16 characters, * and @ cannot be used). (Duplication with codes of other requesting department is not allowed.)
Requesting Department Name :
Enter the requesting department name (a maximum of 64 characters, * and @ cannot be used). (Duplication with names of other requesting department is not allowed.)
3 - 10 2006. 10 897N0520C
3.2.2 Editing Requesting Department Information
Edits requesting department information.
(1) Selecting the requesting department code
Select the code of the requesting department information to be edited.
(2) Selecting the processing
CR Console Operation Manual
Select “Edit”.
(3) Setting code/name → Completing editing
The “Requesting Department Information Input - User Utility” window is displayed. Set requesting department code and requesting department name, and then select “OK”.
(“Cancel” Cancels operation.)
2006. 10 897N0520C 3 - 11
CR Console Operation Manual
Requesting Department Code :
Enter the requesting department information code (a maximum of 16 characters, * and @ cannot be used). (Duplication with codes of other requesting departments is not allowed.)
Requesting Department Name :
Enter the requesting department name (a maximum of 64 characters, * and @ cannot be used). (Duplication with names of other requesting departments is not allowed.)

3.2.3 Relocating Requesting Department Information

Relocates requesting department information (requesting department codes).
(1) Selecting the requesting department code
Select the code of the requesting department to be relocated.
(2) Performing relocation
Select
(move upwards)/ (move downwards) to relocate the information.
3 - 12 2006. 10 897N0520C

3.2.4 Deleting Requesting Department Information

Deletes requesting department information.
(1) Selecting the requesting department code
Select the code of the requesting department information to be deleted.
(2) Selecting the processing Completing deletion
CR Console Operation Manual
Select “Delete”.
2006. 10 897N0520C 3 - 13
CR Console Operation Manual

3.3 Film Mark Setting

Registers, edits, relocates, and deletes film marks.
Select “Film Mark Setting”.
The “Film Mark Setting - User Utility” window is displayed. Perform the operation described as follows, and then select “OK”.
3 - 14 2006. 10 897N0520C
3.3.1 Registering/Editing Film Marks
Registers/edits film marks.
(1) Selecting the button
When newly registering a film mark, select a blank button. When editing the film mark, display a page which has the desired film mark (
CR Console Operation Manual
/ : page turn button), and then select the button.
(2) Completing registration/editing
Enter the film mark (a maximum of ten characters, * and @ cannot be used), and then select “Register”.
Input field for the film mark
2006. 10 897N0520C 3 - 15
CR Console Operation Manual

3.3.2 Relocating Film Marks

Relocates film marks.
Display the page to which the film mark to be relocated is registered (
and select the film mark, and then move it onto the desired button. If the film mark is moved onto a button on which a film mark is already displayed, the moved film mark and the film mark to be relocated switch their locations. Passing on the page turn button on the way to the desired button makes page turn accordingly.
Page is turned when passing here.
/ : page turn button)
Moves to current page’s buttons.
Moves to other page’s buttons.
3 - 16 2006. 10 897N0520C

3.3.3 Deleting Film Marks

Deletes film marks.
(1) Selecting the film mark
CR Console Operation Manual
Display the page to which the film mark to be deleted (
the film mark.
(2) Completing deletion
Select “Delete”.
/ : page turn button), and then select
2006. 10 897N0520C 3 - 17
CR Console Operation Manual

3.4 User Group Management

Manages user groups. This function is available when the user authentication has been specified to be used according the Maintenance Utility settings.
Select “User group management”.
The “Enter Password - User Utility” window will open. Enter a user password registered as the “Administrator” and then select “OK”.
(“Cancel” Cancels operation.)
The “User group management - User Utility” window will open. Perform operations described herein and then select “OK”.
(“Cancel” Cancels operation.)
3 - 18 2006. 10 897N0520C
(1) Selecting the user group
Select a user group you wish to edit.
(2) Selecting the processing
Select “Edit”.
CR Console Operation Manual
(3) Setting the permission function Completing editing
The “User group setting - User Utility” window will open. Select and check a box for the function to be permitted to the user group. Unchecked functions are not permitted to the user group. Note that the “User name” and “User group name” cannot be changed.
Select then “OK”.
2006. 10 897N0520C 3 - 19
CR Console Operation Manual

3.5 Setting for Comment on Mis-exposure

Registers, edits, relocates, and deletes comments on mis-exposures. This function is available when a reason for the mis-exposure has been specified to be input according to the Maintenance Utility settings.
Select “Setting for comment on mis-exposure”.
The “Setting for comment on mis-exposure - User Utility” window will open. After registering a classification of the mis-exposure, you can register preset comments for each mis­exposure classification. Perform operations described herein and then select “OK”.
(“Cancel” Cancels operation.)
3 - 20 2006. 10 897N0520C

3.5.1 Registering the Mis-exposure Classification

Registers the mis-exposure classification.
(1) Selecting the processing
Select “Register” from the “Classification of mis-exposure” item.
(2) Completing the registration
The “Input classification of mis-exposures - User Utility” window will open. Input a classification for the mis-exposure and then select “OK”.
(“Cancel” Cancels operation.)
CR Console Operation Manual
2006. 10 897N0520C 3 - 21
CR Console Operation Manual

3.5.2 Editing the Mis-exposure Classification

Edits a mis-exposure classification.
(1) Selecting the mis-exposure classification
Select a mis-exposure classification you wish to edit.
(2) Selecting the processing
Select “Edit” from the “Classification of mis-exposure” item.
(3) Completing editing
The “Input classification of mis-exposures - User Utility” window will open. Input a mis-exposure classification and then select “OK”.
(“Cancel” Cancels operation.)
3 - 22 2006. 10 897N0520C

3.5.3 Relocating the Mis-exposure Classification

Relocates a mis-exposure classification.
(1) Selecting a mis-exposure classification
Select a mis-exposure classification you wish to relocate.
(2) Executing relocation
CR Console Operation Manual
Select
(moves upward)/ (moves downward) to relocate the failed image category accordingly.
2006. 10 897N0520C 3 - 23
CR Console Operation Manual

3.5.4 Deleting the Mis-exposure Classification

Deletes a mis-exposure classification. Using this function will fully delete also preset comments registered to the mis-exposure classification to be deleted.
(1) Selecting the mis-exposure classification
Select a mis-exposure classification you wish to delete.
(2) Selecting the processing Completing the deletion
Select “Cancellation” from the “Classification of mis-exposures” item.
3 - 24 2006. 10 897N0520C

3.5.5 Registering Preset Comments

Registers preset comments.
(1) Selecting the mis-exposure classification
Select a mis-exposure classification for which you wish to register preset comments.
(2) Selecting the processing
Select “Register” from the “Preset comments” item.
CR Console Operation Manual
(3) Completing the registration
The “Input of comment on mis-exposure - User Utility” window will open. Input preset comments and then select “OK”.
(“Cancel” Cancels operation.)
2006. 10 897N0520C 3 - 25
CR Console Operation Manual

3.5.6 Editing Preset Comments

Edits preset comments.
(1) Selecting the mis-exposure classification Selecting preset comments
Select a mis-exposure classification and then preset comments you wish to edit.
(2) Selecting the processing
Select “Edit” from the “Preset comments” item.
(3) Completing editing
The “Input of comment on mis-exposure - User Utility” window will open. Input preset comments and then select “OK”.
(“Cancel” Cancels operation.)
3 - 26 2006. 10 897N0520C

3.5.7 Relocating Preset Comments

Relocates preset comments.
(1) Selecting the mis-exposure classification Selecting preset comments
Select a mis-exposure classification and then preset comment you wish to relocate.
(2) Selecting the processing Completing deletion
CR Console Operation Manual
Select
(moves upward)/ (moves downward) to relocate the preset comments accordingly.
2006. 10 897N0520C 3 - 27
CR Console Operation Manual

3.5.8 Deleting Preset Comments

Deletes preset comments.
(1) Selecting the mis-exposure classification Selecting preset comments
Select a mis-exposure classification and then preset comments you wish to delete.
(2) Selecting the processing
Select “Cancellation” from the “Preset comments” item.
3 - 28 2006. 10 897N0520C
CR Console Operation Manual
FUJI COMPUTED RADIOGRAPHY
PROPERTY SETTING
Chapter
4
2006. 10 897N0520C 4 - 1
CR Console Operation Manual
This Chapter describes settings for the operating environment. Select “Property Setting”.
The “Property Setting - User Utility” window is displayed. Set each item, and then select “OK”. The description of each item is displayed in the “Description” field at the time of setting.
(“Cancel” Cancels operation.)
Time Setting :
Set time of the unit. Operation after the button is selected is the same operation as performed on the Windows Control Panel.
4 - 2 2006. 10 897N0520C
CR Console Operation Manual
<Study setting 1>
Set study mode for the first time after the startup :
Select the first study screen mode from which the study is begun after the CR Console starts up.
One-image display :
Six-image display :
Only one image is displayed in the display field.
A maximum of six images are displayed simultaneously in the
display field.
Set the Existing Exposure List When Entering Patient Information :
Select an action to be taken to the list of exposure menus already selected when receiving new studies.
Store Exposure List :
A new study can be received while leaving exposure menus already
selected for the previous study.
Clear Exposure List :
A new study can be received clearing off exposure menus already
selected for the previous study.
The method at the time pushed “Study Finish” button (There is no unregistration menus.) :
Select an action to be taken to the corresponding study after
is selected on the Study
Execution Screen. The action set here is applied only when image reading has been completed to all exposure menus. (When there remain in the study exposure menus whose images are unread, the study will be stacked in the “Local WL” with either of the following settings.)
Finish exposure only :
The study is stacked in the “QA” list as the study to which only
exposures have been performed. It cannot be selected when “Automatically distribute images to all attributes whenever an image is read.” or “Distribute only print attribute automatically at each reading time.” is set for “Auto-distribution Setting” in “Distribution setting” folder.
Finish exposure and QA :
The study is stacked in the “Queue” list as the study to which
exposures and image editing (QA) have been performed.
The method at the time pushed “Study Finish” button (The unregistration menu exists.) :
Select an action to be taken to the corresponding study after
is selected on the Study
Execution Study. The action set here is applied only when exposure menu(s) without IP registration remain in the study.
The unregistration menu is deleted, and study is finished :
The exposure menu to which IP registration has not been performed is deleted automatically, and the study is finished.
The unregistration menu is not deleted, and study is suspended :
The exposure menu without IP registration is not deleted but suspended, and the study is stacked in “Local WL”.
The dialog of the confirmation is indicated :
A confirmation box is displayed to select either of the above two actions.
2006. 10 897N0520C 4 - 3
CR Console Operation Manual
Default Menu Code :
Select the exposure menu that is set automatically when the image without exposure menu setting is read. Select “Setup” to display the exposure menu selection box, select the exposure menu to be set, and then select “OK”.
(“Cancel” Cancels operation.)
Default Requesting Department :
Select the requesting department to be set initially when receiving new studies.
Default Sex :
Select the sex to be set initially when receiving new studies.
Trimming Frame Color :
Select “Select Display Color” to display “ Display Color Setting - User Utility” and select a desired color used for the trimming frame.
(“Cancel” Cancels operation.)
4 - 4 2006. 10 897N0520C
CR Console Operation Manual
Besides those displayed, an optional color can be selected. Press the “Arbitrarily selectable” button.
The “Color” window is displayed. After selecting a color, select “OK”.
(“Cancel” Cancels operation.)
2006. 10 897N0520C 4 - 5
CR Console Operation Manual
<Study setting 2>
Setup for automatically issuing a patient ID# :
Select whether or not to issue a patient ID automatically when a new study is started. You can also select to use the examination order issuance function to automatically issue a patient ID.
Auto-issuance (use the order issuance function for screening) :
A patient ID is issued automatically by the examination order issuance function when a new study is started. With this function enabled, the Examination Order Issuance Screen designed for group examination is displayed instead of the Study Reception Screen that is normally displayed. Issuance format is determined by the Maintenance Utility setting. You can make settings in the ‘‘Patient ID’’ and ‘‘Facility Identification Code’’ fields on the Examination Order Issuance Screen.
Automatically issuing a patient ID# :
A patient ID is issued automatically without using the examination order issuance function when a new study is started. ID issuance format is determined by the Maintenance Utility setting. If the determined ID issuance format includes "User-specified character string" or "Auto increment", you can set up those values on the study screen. Note that when an optional card reader is used, the patient name registered in the card will prevail.
Not automatically issuing an patient ID# :
No patient ID is issued automatically when a new study is started.
NOTE
When ‘‘Automatically issuing a patient ID #.’’ is selected, a patient ID is issued automatically even for a study for which an exposure has not yet been performed. In cases such as when a study on the ‘‘Local WL’’ list is resumed or finished, note that the issued patient IDs may not be sequential.
The examination order issuance function can only be used at the following resolutions:
1024x768
1280x1024
4 - 6 2010. 03 897N0520H
CR Console Operation Manual
Setup for starting the next study using the study end button :
After
was selected on the study screen, select processing to be performed for the next
study.
Start the study :
The system starts the next study without returning to the Study Reception Screen. At the start of the next study, patient information items except “Technologist” will be canceled. Note, however, that the default setting remains selected for the selection of Sex. Whether or not to retain the exposure menu complies with what is selected in “Set the Existing Exposure List When Entering Patient Information” in <Study setting 1>.
Not to start the study :
The system returns to the Study Reception Screen without starting the next study.
Menu name font size settings :
Select from among the following the size of a font to be displayed in the exposure menu on the Exposure Menu Selection Screen. This setting will be enabled only when the menu name is too long to display on the screen.
Normal size : Medium size : Small size :
Sets a font size predetermined on the CR Console. Sets a font size smaller than the normal size. Sets a font size smaller than the medium size.
2009. 09 897N0520G 4 - 7
CR Console Operation Manual
<Distribution setting 1>
Auto-distribution Setting :
Select an auto-distribution action to be taken to study images. When the Scroll View Option is activated, setting for the auto-distribution action is fixed to “After an image has been read, only Print attribute will be output automatically when a specific time elapses. On the “Delay time” setting unit specify a time required for outputting the Print attribute.”, and others are not displayed.
Distributes automatically after QA completion. :
When image editing (QA) is completed after an image has been read, it will be auto-distributed (including film output).
Automatically distribute images to all attributes whenever an image is read. :
When an image is read, it will be auto-distributed (including film output).
Distributes only print attribute automatically at each reading time. :
When an image is read, it will be output on film automatically. In such an instance, other delivery methods excepting film output will be executed after the exposure terminating operation is done on the Study Execution Screen.
Distributes all attributes individually. :
Usually, the same action for “Distributes automatically after QA completion.” is taken. However, because the individual delivery icon is displayed on the Study Execution Screen, you can deliver a read image (including film output) individually in advance. Note that studies whose images have been delivered individually with this setting cannot be stacked in the “QA” list after completing only exposures.
Distributes Print attribute only individually. :
Usually, the same action for “Distributes automatically after QA completion.” is taken. However, because the individual delivery icon is displayed on the Study Execution Screen, you can output a read image individually on film in advance. Note that studies whose images have been delivered with this setting can be stacked in the “QA” list after completing only exposures.
4 - 8 2009. 09 897N0520G
CR Console Operation Manual
After an image has been read, only Print attribute will be output automatically when a
specific time elapses :
When a certain period of time has passed after an image was read, it will be
output automatically on film. In this case, image distribution other than film output is executed after the exposure operation is finished at the Study Execution Screen.
Select a time period (in minutes) ranging from image reading to film output from
dropdown list.
When setting this operation, “Finish Exposure only” cannot be selected for the
study finish method. When either one of the following items is set to “Finish Exposure only”, a confirmation box is displayed. Select “YES”. The study finish method will be changed to “Finish Exposure and QA”.
• Item of “The method at the time pushed “Study Finish” button (There is no unregistration menu.)” in the “Study setting 1” folder.
• Item of “Setup of an automatic study end” in the “Other setting” folder.
• Item of “The method at the time pushed “Study Finish” button on local worklist in the “Other setting” folder.
(“Cancel” Cancels the operation. A setting performed in the “Auto-distribution Setting” item
will be restored.)
Output stitched image :
A stitched long-view image to be printed on film or transferred to external devices is determined.
Film output :
Select images to be used for film output from among the following.
Source images for stitching :
Source images used for creation of a stitched long-view image are output.
Source images for stitching and stitched image (reduced to 14”x17” size) :
Source images used for stitching are output, and a stitched long-view image is reduced and output on 14”x17” film.
Stitched image (reduced to 14”x17” size) :
A stitched long-view image is reduced and output on 14”x17” film.
Stitched image (in life size) :
A stitched long-view image is output in its life size.
Stitched image (in life size) and stitched image (reduced to 14”x17” size) :
A stitched long-view image is output in its life size and reduced and output on 14”x17” film.
2009. 09 897N0520G 4 - 9
CR Console Operation Manual
Network transfer :
Select images to be used for network transfer from among the following.
Source images for stitching :
Source images for stitching and stitched image (reduced to 14”x17” size) :
Stitched image (reduced to 14”x17” size) :
Stitched image (in life size) :
Stitched image (in life size) and stitched image (reduced to 14”x17” size) :
NOTE
A stitched long-view image output in its life size will not reflect results of Automatic Shuttering
Processing and Manual Shuttering Processing.
For shared study information, do not use the User Utility function to change the output type that
has been set up as the default for the “Output stitched image”.
Source images used for creation of a stitched long-view image are output.
Source images used for stitching are output, and a stitched long-view image is reduced and output on 14”x17” film.
A stitched long-view image is reduced and output on 14”x17” film.
A stitched long-view image is output in its life size.
A stitched long-view image is output in its life size and reduced and output on 14”x17” film.
Setting the reduction ratio for output of an image on 8”x 10” film. :
Select processing for output of reduced images on 8”x10”-sized film.
Reduced image output is not allowed. :
The reduced-image output on 8”x10”-sized film is not performed.
Reduced image output is allowed. :
Outputs reduced images with the one-image format on 8”x10”-sized film. (Excepting the images to which the 100%-size output for 14”x17” and 14”x14”-sized films is set.)
PRINT Attribute :
Select a study image printer. If the first printer is not available, the output will be yielded to the second printer. Printouts will be yielded to a printer where an appropriate film size is available, in accordance with the specified “Image size” and “Film format”. Note that depending on the printer function, a study film may be output to any other printer.
FILE Attribute :
Select a study image filing device.
Energy Subtraction Exposure Output Image :
Select the type of an output image derived from energy subtraction processing.
4 - 10 2009. 09 897N0520G
CR Console Operation Manual
<Distribution setting 2>
Setting the Position for Layout of Film Annotation When the Medical Examination Format is
Applied :
This setting is enabled when the 12on1 format option has been activated. Select layout of character strings (Top End or Bottom End) that represent film’s additional information, when outputting 12 image frames of multiple studies on one film.
Top End :
Assigns additional information at top end when applying a medical examination
format.
Bottom End :
Assigns additional information at bottom end when applying a medical
examination format.
Setting the Image Annotation Layout Position When the Medical Examination Format is
Applied :
This setting is enabled when the 12on1 format option has been activated. Select layout of character strings (Top Left, Top Right, Bottom Left or Bottom Right) that represents image’s additional information, when outputting 12 image frames of multiple studies on one film.
Top Left :
Assigns additional information at top left when applying a medical examination
format.
Bottom Left :
Assigns additional information at bottom left when applying a medical
examination format.
Top Right :
Assigns additional information at top right when applying a medical
examination format.
Bottom Right :
Assigns additional information at bottom right when applying a medical
examination format.
2009. 09 897N0520G 4 - 11
CR Console Operation Manual
Setting the Output of Medical Examination Format Images from the Media :
This setting is enabled when the 12on1 format option and the media storage option have been activated appropriately. Select whether or not to output 12 image frames of multiple studies derived from the media on one film.
Images are output in the medical examination format :
Images are not output in the medical examination format :
Shown below are the study requirements applicable to image output in the medical examination format (a film format for output of 12 image frames on one film).
• “Study” and “QA” should have been shut down on the CR Console.
• Images to be output should not be long-view, energy subtraction or mis-exposure images.
You can output images in the medical examination format from the “Today” and “All” lists available on the CR Console. You can also output images from the media list if the media option has been activated appropriately.
12 image frames of multiple studies are output on one film.
Images of multiple studies are output according to the pre-determined format. If “Images are not output in the medical examination format” is selected, images are output in the ordinary format.
4 - 12 2009. 09 897N0520G
Loading...