Fluke Victoreen 451P, Victoreen 451P-DE-SI Operator's Manual

Victoree
451P & 451P-DE-SI
Ion Chamber Survey Meter
Manual No. 1020101000 Rev. 2 ©2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies
February 2005
Operators Manual
Fluke Biomedical
Radiation Management Services
6045 Cochran Road Cleveland, Ohio 44139
440.498.2564
Table Of Contents
Section 1: General Information
1.1 Introduction .................................................................................................. 1-1
1.2 Features....................................................................................................... 1-1
1.3 Receiving Inspection.................................................................................... 1-2
1.4 Specifications............................................................................................... 1-2
Section 2
2.1 External Controls ......................................................................................... 2-1
2.2 Installation.................................................................................................... 2-2
Section 3: Theory Of Operation
3.1 Introduction .................................................................................................. 3-1
3.2 Firmware...................................................................................................... 3-2
Section 4: Maintenance
4.1 Introduction .................................................................................................. 4-1
4.2 Routine Cleaning ......................................................................................... 4-1
4.3 Storage ........................................................................................................ 4-1
4.4 Battery Replacement ................................................................................... 4-2
4.5 High Voltage Battery Board Replacement ................................................... 4-3
4.6 Replacement Parts Information ................................................................... 4-3
4.7 Recalibration and Service Information ......................................................... 4-4
:
Operation
..................................................................................... 2-1
................................................................................. 4-1
...................................................................... 1-1
...................................................................... 3-1
i
General Information
Introduction
1
Section 1
GENERAL INFORMATION
1.1 Introduction
The Model 451P Ion Chamber Survey Meter is a hand-held, pressurized, battery operated unit designed to measure gamma and x-ray radiation above 25 keV, and beta radiation above 1 MeV, using the latest CMOS and LCD technology. The 451P case is constructed of high strength ABS plastic. A gasket seals moisture out of the unit and provides a cushion for the internal components. The Model 451P readout consists of a 2 1/2 digit liquid crystal display and a 100 segment analog bargraph. The bargraph contains a zero segment and twenty groups of five segments each. A permanent scale is located on the display screen. The major divisions of the scale indicate the units corresponding to the range that the meter is measuring. Units of measurement are displayed next to the 2 1/2 digit display. LOW BAT and FREEZE will appear on the display when the instrument is operating in these modes. External controls consist of an ON/OFF button and a MODE button. The unit is auto-ranging and auto-zeroing and has an auto ON backlight. The backlight automatically comes on when the ambient light conditions fall below twilight levels. An internal factory set alarm will blink the display if the rate exceeds 5 R/hr. Two 9-volt batteries, located in the rear of the instrument, provide over 200 hours of continuous operation.
Model 451P & 451P-DE-SI Ion Chamber Survey Meter
1.2 Features
1.2.1 Overrange Rate
If the instantaneous radiation rate measured by the Model 451P exceeds 5 R/h, the 'R' in 'mR/h' or the 'Sv' in 'Sv/h' displayed will blink to notify the user that a potential error in the integrated radiation value exists. The blinking will stop when the integrated value is cleared. This, however, may also clear the log data RAM.
1-1
Victoreen 451P & 451P-DE-SI
Operators Manual
1.2.2 Low Battery Indicator
There are about six hours of operation remaining when the LOW BATTERY indicator first becomes visible. When the LOW BATTERY indicator blinks, there is less than one hour of operation remaining. These times are for two batteries installed and from the first occurrences of these indications. If the instrument is turned off during a low battery condition, the batteries will recover somewhat, but time of operation remaining will be less. The LCD low battery indicator is disabled during communication mode. Be sure to use fresh batteries when calibrating this instrument.
WARNING
If chamber bias is LO, the instrument can’t measure high radiation rates accurately. Service to the instrument is required.
Verify the chamber bias is low. Turn the unit off and replace with new 9-volt batteries. Power up the unit. After the self-test, if blinking low battery reoccurs the bias is low. Refer to Section 4 Maintenance.
1.2.3 Warm-up Time
The pressured ion chamber collection potential is 105 Vdc and is derived from the 5 lithium cells. The warm-up time for an instrument that has been off 12 or more hours is about four minutes for readings less than 20 μR/h in a 10 μR/h or less background.
1.3 Receiving Inspection
Upon receipt of the unit:
1. Check the shipping cartons(s) and their contents for in-shipment damage. If damage is evident, file a claim with the carrier and contact Fluke Biomedical, Radiation Management Services Repair Coordinator at 440.498.2564 or 800.850.4606 immediately.
2. Check that all items listed on the packing slip are present and in good condition. If any items are missing or damaged, contact Fluke Biomedical, Radiation Management Services Repair Coordinator at 440.498.2564 or 800.850.4606 immediately.
1.4 Specifications
Radiation Detected Beta above 1 MeV & gamma above 25 KeV
Operating Ranges 0 to 500 μR/hr (0 to 5 μSv/h), 0 to 5 mR/h (0 to 50 μSv/h)
0 to 50 mR/h (0 to 500 μSv/h), 0 to 500 mR/h (0 to 5 mSv/h) 0 to 5 R/h (0 to 50 mSv/h)
Accuracy ± 10% of reading between 10% and 100% of full-scale indication on any range,
exclusive of energy response (calibration source is
137
Cs)
Detector 230 cc volume air ionization chamber, pressurized to 6 atmospheres
Plastic chamber wall 200 mg/cm2 thick
1-2
General Information
Specifications
Warm-Up Time Less than one minute for initial operation when the instrument is in temperature
equilibrium with the surrounding area and about 4 minutes for readings less than 20 μR/h in a 10 μR/h or less background
Drift After seven minutes operation, 0.04 mR/h equivalent, or less
Response Time Time measured from 10% to 90% of final value for a step increase/decrease in
radiation rate such that a range change does not occur:
0 to 500 μR/h (0 to 5 μSv/h) range: 5 seconds 0 to 5 mR/hr (0 to 50 μSv/h) range: 2 seconds 0 to 50 mR/h (0 to 500 μSv/h) range: 1.8 seconds 0 to 500 mR/h (0 to 5 mSv/h) range: 1.8 seconds 0 to 5 R/h (0 to 50 mSv/h) range: 1.8 seconds
NOTE: In pulsating field, the instantaneous rate should not exceed 5 R/h for
proper integration; instantaneous exposure rate is still limited to 5 R/h
Precision Within 5% reading
Readout Liquid Crystal Display: contains an analog bar graph with a permanent scale on
the display and a 2 1/2 digit display
Analog Display: the bargraph consists of 100 segments, 2 ½ inches long; the
scale has five major divisions; the appropriate value for the operating range of the instrument will appear below the scale
Digital Display: the digital display is 2 1/2 digits followed by a significant zero
digit depending on the operating range of the instrument. The leading 1/2 digit is blank, or a “1” or a “0” for clarity. Units of measure appear to the right of the digital display. Appropriate multipliers also appear on the display
Units: as indicated under Range, programmable in R/h or Sv/h. Appropriate
multipliers also appear on the display
Auto-On Backlight: turns on when ambient light is less then twilight conditions
1
External Controls ON/OFF button, MODE button
Automatic Features Ranging and zeroing are fully automatic.
Environmental Operating Temperature Range: - 4º to + 122ºF (- 20º to + 50ºC)
Relative Humidity Range: 0 to 100%, non-condensing
Geotropism: less than 1%
Dimensions 8.5 (l) x 4.5 (w) x 8.6 in (h) (21 x 11.4 x 21.3 cm)
Weight Approximately 2.6 lb (1.2 kg)
Batteries Two 9-volt batteries provide over 200 hours continuous operation
* Specifications are subject to change without notice.
1-3
Loading...
+ 15 hidden pages