Fluke Ti100-FC, TiR110, TiR125 Data sheet

TiR125, TiR110,
TiR105 Building
Diagnostic and the
TiR110
Ti100 General Use
Thermal Imagers
The lightest, most-rugged, easiest–to-use professional thermal imagers around.
Key features
Made in
the U.S.A.
of U.S. and
non U.S.
parts
enabled
TiR125
Ti100
Fluke CNX
Building problems, defects and general maintenance
Energy audit, building inspection, and weatherization
TiR105
•Exclusive IR-OptiFlex in good focus from 1.2 m (4 ft) and beyond for optimum image clarity and scanning convenience. For shorter distances change to manual mode with the touch of a finger (TiR110 and TiR125).
•Always have references handy—IR-PhotoNotes system—Quickly identify and keep track of inspection locations by adding digital images of important information and surrounding areas (TiR110 and TiR125).
•Find problems faster and easier with Fluke IR-Fusion (TiR125, TiR110, TiR105). Accurately identify potential issues by combining digital and IR images.
•Capture additional measurements fast and conveniently from up to five wireless modules at once using your Fluke thermal imager and CNX
•Get further clarification with AutoBlend of a digital and partially transparent IR image into a single information-filled image (TiR125 only).
•Multi-mode video recording—focus-free video in visible light and infrared with full IR-Fusion (TiR110 and TiR125 only).
•Easily communicate the location of problems with the Electronic 8-Point Cardinal Compass (TiR125 and TiR110 only).
Technical Data
focus system—ensures that images are
annotation
wireless modules.
mode - the blending
®
technology
Restoration, water damage, and roofing
Patented Fluke IR-Fusion® Technology
Enjoy the industry’s only point-and-shoot IR-Fusion camera. Fluke patented technology provides the user with both a digital and an infrared image in one to precisely document problem areas.
Spend less time in the field and more time growing your business with these new Fluke thermal imagers.
IR-OptiFlex focus system
Scan for issues significantly faster than before with Fluke’s revolutionary, ultra-rugged focus system. The IR-OptiFlex focus system gives you optimum focus by combining focus-free, ease-of-use with the flexibility of manual focus on the same camera.
Detailed specifications
TiR125 TiR110 TiR105 Ti100
Building diagnostics General use
IR resolution (FPA size) 160 x 120 FPA Uncooled Microbolometer Spectral band 7.5 μm to 14 μm (long wave) Capture or refresh rate 9 Hz
NETD (Thermal sensitivity) ≤ 0.08 °C at 30 °C target temp (80 mK)
FOV (Field of view) 22.5 °H x 31 °V IFOV (Spatial resolution) 3.39 mRad Temperature measurement range
(not calibrated below -10 °C)
-20 °C to +150 °C (-4 °F to +302 °F)
Temperature measurement accuracy ± 2 °C or 2 % (at 25 °C nominal, whichever is greater) Focus mechanism IR-OptiFlex
IR-Fusion
CNX
®
technology
Wireless enabled Yes
PIP, FULL IR, FULL VISIBLE,
AutoBlend
focus system Focus-free 1.2 m (4 ft) and beyond
PIP (1.2 m (4 ft)
PIP, FULL IR, FULL VISIBLE
to 4.6 m (15 ft)), FULL IR,
FULL VISIBLE
High termperature, low
Color alarms
temperature (dewpoint), and
Low temperature (dewpoint)
isotherm
Standard palettes
Blue-Red, Grayscale, Inverted Grayscale, High-contrast,
Hot Metal, Ironbow, Amber, Inverted Amber
Blue-Red, Ironbow, Grayscale,
Blue-Red, Grayscale, Inverted
Ultra Contrast
palettes
Grayscale, High-contrast,
Hot Metal, Ironbow, Amber,
Blue-Red, Grayscale, Ironbow
Inverted Amber
Hot/cold markers Yes User definable spot markers Three on camera and in SmartView
®
in SmartView® only
Centerpoint Yes Centerbox (MIN/AVG/MAX) Yes Level and span control Manual and auto Minimum span in auto mode 2.5 °C 5 °C Minimum span in manual mode 2.0 °C 2.5 °C Minimum IR focus distance 15.25 cm (6 in) 122 cm (48 in) Weight 0.726 kg (1.6 lb) Size 28.4 x 8.6 x 13.5 cm (11.2 x 3.4 x 5.3 in) LCD display 3.5 inch diagonal (portrait format) Visible camera 2 megapixel industrial-grade N/A Minimum parallax ~45.7 – 55.9 cm (~ 18 in – 22 in) ~122 cm (48 in) N/A IR-PhotoNotes
annotation system Yes (3 images)
Laser pointer Yes Torch Yes Electronic (cardinal) compass Yes Emissivity correction Yes Transmission correction Yes Background (reflected) compensation Yes Voice annotation (audio) Yes (60 seconds) per image Multi-mode video output Streaming USB video output Multi-mode video recording
(standard AVI with MPEG encoding)
Multi-mode video recording (radiometric .is3)
Yes, radiometric .is3 for approx. 2.5 to 5 minutes de­pending upon thermal scene
Yes (AVI with MPEG encoding)
Memory review thumbnail review Battery (field-replaceable,
rechargeable)
Two One
Battery life 4+ hours (each)* External battery charging base Included Optional (accessory) Charging power supply Ye s
*Assumes 50 % brightness of LCD
≤ 0.10 °C at 30 °C target temp
(100 mK)
-20 °C to + 250 °C (-4 °F to 482 °F)
No, full IR only
Amber
2 Fluke Corporation TiR125, TiR110, TiR105 Building Diagnostic and Ti100 General Use Thermal Imagers
Loading...
+ 1 hidden pages