Fluke T6-600, T6-1000 QuickStart Guide

HOLD
Quick Reference Guide GuÍa de referencia rápida Guia de referência rápida
T6-600/T6-1000
Electrical Tester
with FieldSense Technology
»
�� See “Safety Information”.
Go to www.fluke.com to register your product and find more information, or download this Quick Reference Guide in other languages. Consulte las “Información sobre seguridad”.
Vaya a www.fluke.com para registrar el producto y obtener más información, o descargue esta Guía de referencia rápida en otros idiomas.
Consulte o “Informações de Segurança”. Acesse www.fluke.com.br para registrar seu produto e obter mais informações, ou baixe este Guia de referência rápida em outros idiomas.
Fork Opening 17.8 mm
CAT 600 V
AUTO
SELECT
T6-100 0
ELECTRICAL TESTER
CAT
1000 V
200 A
HOLD
PN 4897598 September 2017 Rev. 1, 10/17
PN 4897598 September 2017
© 2017 Fluke Corporation. All rights reserved.
© 2017 Fluke Corporation. All rights reserved. Specifications are subject to change without notification.
Specifications are subject to change without notification.
1
2
6
3 4
5
7
Overview
Wire Position
1
CAT 600 V
200 A
Finger Guard
2
Display
3
X Hazardous Voltage ≥30 V
4
5
Function Selection/On/Off
6
FieldSense
1000 V
Ground Contact
7
Test Probes/Storage Dock
1 Wire Only
For safe operation, hold the Tester behind the finger guard.
Symbols
WARNING. RISK OF DANGER.
Consult user documentation.
Hazardous voltage ≥30 V.
Battery full charge.
Battery low. Replace.
Back light
Test with fork and FieldSense ground contact.
Test with fork and ground probe.
Test with probes.
FieldSense Measurement: Fluke voltage/current sensing technique.
Good earth ground connection.
No earth ground connection.
Replacement Parts
Test Lead Assembly (T5-RLS) Replace only with Fluke double-insulated leads (T)
TP1 Probe Set, Flat-Blade
Battery Door
Battery Door Screw
Use only specified replacement parts.
Lead Replacement
PN 4462973
PN 4462973
PN 648128
PN 648128
PN 4944370
PN 4944370
PN 1618579
PN 1618579
Specifications
Model T6-1000
Function
FieldSense
Voltage ac
FieldSense
Current ac
FieldSense
Frequency (Hz)
Model T6-600Model T6-1000Model T6-1000
Auto Off
15 min
Voltage ac
Voltage dc
Resistance
[1] Accuracy: ± ( [ % of reading] + [number of least significant digits] ). Accuracy is specified for 1 year after calibration, at 18 °C to 28 °C (64 °F to 82 °F) with relative humidity to 90 %. AC measurements are ac coupled, RMS responding. [2] Add 3 % without an external ground connection. External ground connection required for user wearing insulated gloves, standing on an insulated ladder, or otherwise insulated from earth ground. [3] FieldSense is specified from 16 V to 100 % of range.
Fork Opening 17.8 mm
Calibration 1-year calibration cycle
Temperature Operating -10 °C to +50 °C (+14 °F to +122 °F) Storage -30 °C to +60 °C (-22 °F to +140 °F)
Altitude
Operating Storage
Relative Humidity 0 % to 95 %, 5 °C to 30 °C (41 °F to 86 °F) 0 % to 75 %, 30 °C to 40 °C (86 °F to 104 °F) 0 % to 45 %, 40 °C to 50 °C (104 °F to 122 °F)
Battery Type and Life 2x AA (IEC LR6) 360 hr continuous, typical is 200 hr using the FieldSense function
Temperature coefficient 0.1 x (specified accuracy) / °C for <18 °C or >28 °C (<64.4 °F or >82.4 °F)
Requires
Test Leads
No
No
No
Yes
Yes
Yes
Yes
Yes
2000 m 10 000 m
Range
1000 V
200.0 A
45 Hz to 66 Hz
1000 V
1000 V
2000 Ω 2000 Ω
20.00 kΩ
100.0 kΩ
Model T6-600
Range
600 V
200.0 A
600 V
600 V
Resolution
1 V
0.1 A
1 Hz
1 V
1 V
1 Ω
0.01 kΩ
0.1 kΩ
[1]
Accuracy
±(3 % + 3 counts) 45 Hz to 66 Hz
±(3 % + 3 counts) 45 Hz to 66 Hz
±(1 % + 2 counts)
±(1.5 % + 2 counts) 45 Hz to 66 Hz
±(1 % + 2 counts)
±(1 % + 2 counts)
[2][3]
[3]
Accessories
● HT6 Belt Holster
● TPAK Hanger
● PRV240FS
● AC285 SureGrip™ Alligator Clips
● AC220 SureGrip™ Alligator Clips
Contact Fluke
USA: 1-800-44-FLUKE (1-800-443-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-6714-3114 Singapore: +65-6799-5566-5655 China: +86-400-921-08365 Anywhere in the world: +1-425-446-5500
A
CAT
HOLD
HOLD
T6 - 1000
Electrical Tester
HOLD
Hz
FieldSense
T6 - 1000
Electrical Tester
HOLD
Hz
FieldSense
200 A
CAT
1000 V
CAT 600 V
How to Make MeasurementsHow to Make Measurements
1
2A
2B
/
HOLD
45 Hz - 66 Hz T6-600 max: 600 V rms CAT III
(AC)
T6-1000 max: 1000 V rms CAT III 600 V rms CAT IV
T6-600 max: 600 V CAT III T6-1000 max: 1000 V CAT III
(DC)
600 V CAT IV
Model T6-1000
1
HOLD
2
Use the Ground Probe
3
Hz
/ / /
1
HOLD
Hz
2
Insert Test Probes into the storage dock
3
Hold hand firmly on FieldSense Ground Contact
Hz
Model T6-600
1
HOLD
AUTO
SELECT
2
Use the Ground Probe
3
/ /
1
HOLD
AUTO
SELECT
2
Insert Test Probes
into the storage dock
3
Hold hand firmly on FieldSense Ground Contact
1
2A
2B
HOLD
Tips: FieldSense Technology
Use the black probe to make the ground connection if:
● wearing gloves or insulated footwear
● standing on an insulated ladder
● additional ground contact is required for the application
W Do not hold or touch wire during a measurement.
FieldSense measurement looks like this when properly grounded:
If you see: Solution:
● Display color turns green
● Display shows a valid measurement >16 V rather than dashes or 0.0
Y shows for measurement ≥30 V
● shows on the
display
● Display color is gray
● VAC measurement does not show on display
● Display color is gray
● VAC measurement does not show on display
● Make sure that your bare hand is firmly touching the FieldSense Ground Contact on the battery door.
● Check that the black probe is properly stored in the dock.
● Check the wire position in the
fork opening. See the section about wire position.
● Measurement is <16 V
● Check the wire position in the fork opening. See the section about wire position.
HOLD
4
4
/
/
200 A
CAT 600 V
CAT
1000 V
HOLD
5
HOLD
Hz
=
Display turns green for valid measurement. See Tips.
200 A
CAT 6 00 V
CAT
1000 V
200 A
CAT 600 V
CAT
1000 V
Display turns green for valid measurement. See Tips.
5
CAT 6 00 V
CAT
200 A
1000 V
Hz
=
4
CAT 600 V
200 A
Display turns green for valid measurement. See Tips.
5
CAT 600 V
200 A
=
4
CAT 600 V
200 A
CAT 1000 V
200
600 V
Display turns green for valid measurement. See Tips.
5
CAT 600 V
200 A
HOLD
1
200 A
CAT 600 V
CAT
1000 V
Low Battery Indicator
2
Do not overtighten screw
2 AA
=
Loading...