Fluke T6 Quick Start Guide

HOLD
Quick Reference Guide
True
RMS
T6-600/T6-1000
Electrical Tester
with FieldSense Technology
» See
Go to www.󰘱uke.com to register your product and 󰘰nd more information, or
download this Quick Reference Guide in other languages.
Safety Information
Fork Opening 17.8 mm
CAT 600 V
200 A
1000 V
.
1
2
6
3
HOLD
4
5
AUTO
SELECT
T6-1000
ELECTRICAL TESTER
PN 4897598 September 2017 Rev. 4, 7/18
© 2017-2018 Fluke Corporation. All rights reserved.
Speci󰘰cations are subject to change without noti󰘰cation.
7
Overview
Wire Position
1
Wire must contact the bottom of the
CAT 600 V
200 A
Finger Guard
2
Display
3
X Hazardous Voltage ≥30 V
4
Function Selection/On/O󰀨
5
6
FieldSense Ground Contact
7
Test Probes/Storage Dock
CAT
fork opening.
1000 V
Model T6-1000
1 Wire Only
For safe operation, hold the
Tester behind the 󰘰nger guard.
Model T6-600
Auto O󰀨
15 min
Symbols
W
»
WARNING. RISK OF DANGER. Test with fork and FieldSense ground contact.
Consult user documentation. Test with fork and ground probe.
Hazardous voltage ≥30 V. Test with probes.
Y
Battery full charge.
Battery low. Replace.
Back light
J
J
FieldSense Measurement: Fluke voltage/current
sensing technique.
Good earth ground connection.
No earth ground connection.
Specications
Function
FieldSense
Voltage ac true-rms
FieldSense
Current ac true-rms
FieldSense
frequency (Hz)
Voltage ac true-rms Yes 1000 V 600 V 1 V
Voltage dc Yes 1000 V 600 V 1 V ±(1 % + 2 counts)
Resistance
[1] Accuracy: ±( [ % of reading] + [number of least signi󰘰cant digits] ). Accuracy is speci󰘰ed for 1 year after calibration, at
18 °C to 28 °C (64 °F to 82 °F) with relative humidity to 90 %. AC measurements are accoupled, RMS responding.
[2] Add 3 % typical without an external ground connection. External ground connection required for user wearing insulated
gloves, standing on an insulated ladder, or otherwise insulated from earth ground.
[3] FieldSense is speci󰘰ed from 16 V to 100 % of range.
Fork opening 17.8 mm
Temperature
Operating
Storage
Altitude
Operating
Storage
Relative Humidity 0 % to 90 %, 5 °C to 30 °C (41 °F to 86 °F)
Battery Type and Life 2x AA (IEC LR6) 360 hr continuous, typical is 200 hr using the FieldSense function
Temperature coe󰘲cient 0.1 x (speci󰘰ed accuracy) / °C for <18°C or >28 °C (<64.4 °F or >82.4 °F)
Requires
Test Leads
No 1000 V 600 V 1 V
No 200.0 A 200.0 A 0.1 A
No 45 Hz to 66 Hz 1 Hz ±(1 % + 2 counts)
Yes 2000 Ω 2000 Ω 1 Ω
Yes 100.0 kΩ 0.1 kΩ
-10 °C to +50 °C (+14 °F to +122 °F)
-30 °C to +60 °C (-22 °F to +140 °F)
2000 m
10 000 m
0 % to 75 %, 30°C to 40 °C (86 °F to 104 °F)
0 % to 45 %, 40°C to 50 °C (104 °F to 122 °F)
Model T6-1000 Model T6-600
Range Range Resolution Accuracy
±(3 % + 3 counts) 45 Hz to 66 Hz
±(3 % + 3 counts) 45 Hz to 66 Hz
±(1.5 % + 2 counts) 45 Hz to 66 Hz
±(1 % + 2 counts)Yes 20.00 kΩ 0.01 kΩ
Replacement Parts
Test Lead Assembly (T5-RLS) Replace only with Fluke double-insulated leads ()
TP1 Single Probe, Flat-Tip, Red PN 648128
TP1 Single Probe, Flat-Tip, Black PN 648102
TP38 Single Probe, Round-Tip, Red PN 1276841
TP38 Single Probe, Round-Tip, Black PN 1276852
Battery Door PN 4944370
Battery Door Screw PN 1618578
Use only specied replacement parts
.Lead Replacement
[1]
[2] [3]
[3]
Accessories● HT6 ●
● HT6 Belt Holster
● TPAK Hanger
● PRV240FS Proving Unit
● AC285 SureGrip™ Alligator Clips
● AC220 SureGrip™ Alligator Clips
● C60 softcase
Contact FlukeUSA:
USA: 1-800-44-FLUKE (1-800-443-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-6714-3114
Singapore: +65-6799-5566-5655
China: +86-400-921-08365
Anywhere in the world: +1-425-446-5500
PN 4462973
0
C
C
HOLD HOLD
How to Make Measurements
1
2A
2B
/
HOLD
45 Hz - 66 Hz T6-600 max: 600 V rms CAT III
(AC)
T6-1000 max: 1000 V rms CAT III 600 V rms CAT IV
45 Hz - 66 Hz T6-600 max: 600 V CAT III
(DC)
T6-1000 max: 1000 V CAT III 600 V CAT IV
Model T6-1000
1
HOLD
2
Use the Ground Probe
3
Hz
// /
1
HOLD
Hz
2
Insert Test Probes into the storage dock
3
Hold hand 󰘰rmly on
FieldSense Ground Contact
Hz
Model T6-600
1
HOLD
AUTO
SELECT
2
Use the Ground Probe
3
//
1
HOLD
AUTO SELECT
2
Insert Test Probes into the storage dock
3
Hold hand rmly on
FieldSense Ground Contact
1
2A
2B
HOLD
STips: FieldSense Technology
Use the black probe to make the ground connection if:
● wearing gloves or insulated footwear
● standing on an insulated ladder
● additional ground contact is required for the application
W Do not hold or touch wire under test during a measurement. This will modify
the operator's voltage pontential with respect to earth and give invalid measurements.
FieldSense measurement looks like this when properly grounded:
● Display color turns green
● Display shows a valid measurement >16 V
rather than dashes or 0.0
Y shows for measurement ≥30 V
If you see: Solution:
● Make sure that your bare hand
show on the
display
● Display color is gray
● VAC measurement
does not show on display
● Display color is gray
● VAC measurement
does not show on display
is 󰘰rmly touching the FieldSense Ground Contact on the battery door.
● Check that the black probe is properly stored in the dock.
● Make sure the wire position in
the fork opening is correct. See the
Overview
position.
● Measurement is <16 V
● Make sure the wire position in
the fork opening is correct. See the
Overview
position.
section about wire
section about wire
HOLD
HOLD
4
4
/
/
200 A
CAT 600 V
CAT
1000 V
Display turns green for valid measurement. See Tips.
5
CAT 600 V
CAT
200 A
1000 V
HOLD
Hz
=
200 A
CAT 600 V
CAT
1000 V
Display turns green for valid measurement. See Tips.
5
CAT 600 V
CAT
200 A
1000 V
Hz
=
4
CAT 600 V
200 A
Display turns green for valid measurement. See Tips.
5
CAT 600 V
200 A
=
4
CAT 600 V
200 A
20
600 V
Display turns green for valid measurement. See Tips.
5
CAT 600 V
200 A
1000 V
HOLD
1
200 A
CAT 600 V
CAT
1000 V
Low Battery Indicator
2
Do not overtighten screw
2 AA
=