Fluke ScopeMeter 190B Series, ScopeMeter 190C Series User Manual

4822 872 00741 Nov. 2002, Rev.1 3/04
© 2002 , 2003 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies.
Fluke 190B/C
Medical Functions
Users Manual Supplement
Chapter Title Page
1 General.......................................................................................................................... 1-1
About this Manual .......................................................................................................... 1-1
Safety Information.......................................................................................................... 1-1
Limited Warranty, Limitation of Liability......................................................................... 1-1
About the Medical Version ............................................................................................. 1-2
MA 190 Accessory Kit.................................................................................................... 1-2
2 Using the Medical Functions....................................................................................... 2-1
mAs Measurements ....................................................................................................... 2-1
mVs measurements ................................................................................................. 2-3
mWs measurements ................................................................................................. 2-3
Extended Offset ............................................................................................................. 2-4
Oscilloscopes and dynamic range............................................................................. 2-5
Triggering .................................................................................................................. 2-6
Tip for video signal measurements............................................................................ 2-6
Specifications ............................................................................................................ 2-6
Smart Average ............................................................................................................... 2-7
Smart average vs. normal average ........................................................................... 2-8
Non Interlaced Video Triggering..................................................................................... 2-9
Chapter 1
General
About this Manual
This manual supplement is an addition to the Users Manual that is included with the ScopeMeter 190B/C series test tool kit. It provides user information about the extended test tool functionality that is available in the Fluke190B/C /M versions.
Safety Information
Read and comply with the safety instructions that you find in the ScopeMeter 190B/C series Getting Started Manual and Users Manual.
The MA 190 kit accessories are intended for low voltage (42 Vpk) use only, and do not comply with the voltage rating on most other ScopeMeter accessories. Carefully read the warning below which is an explicit addition to the general accessory and ScopeMeter safety statements.
Warning
To avoid electrical shock or fire, do NOT connect the following accessories to voltages more than 42 Vpk (30 Vrms) or circuits of more than 4800 VA :
CS 190-1R, 1 Ohm shunt resistor
TR 190-50R, 50 Ohm terminator
TA 190-50R, 50 Ohm terminator / 10:1
attenuator
Limited Warranty, Limitation of Liability
Read the Warranty and Limitation of Liability statements that you find in the ScopeMeter 190B/C Users Manual that is included in the test tool kit shipment (on CD).
1-1
Fluke 190B/C Medical
ST8723
User Manual
About the Medical Version
The /M Medical version is a standard Fluke 190B/C Series ScopeMeter test tool with the following additional functionality:
mAs measurements
non interlaced video triggering
smart averaging
extended offset
MA 190 Accessory Kit
The MA 190 is an optional accessory kit that provides optimal connectivity to the ScopeMeter test tool in medical and video applications.
Before use carefully read the WARNING on page 1-1.
The kit contains:
TR 190-50R, 50 Ohm BNC feed-through terminator for video and HF measurements (Max. 5 V) . See Fig. 1-1.
TA 190-50R, 50 Ohm 10:1 attenuator feed-through terminator for additional dynamic range and offset. See Fig. 1-1.
CS 190-1R, 1 Ohm current shunt for sensitive current and mAs measurements. (Max. 500 mA)
. See Fig. 2-1.
Safety designed 50 Ohm BNC cable with isolated plastic connectors. (1.5 m)
Female BNC to male 4 mm banana plug adapter
Two female to female 4 mm banana plug adapters
Note 1
CS 190-1R, the TR 190-50R, and the
The TA 190-50R are not available as individual accessory, but only as part of the MA 190 kit.
VIDEO SIGNAL SOURCE
50 BNC CABLE
TR190-50R
or
TA190-50R
Figure 1-1. 50 Ohm cable termination using the
TR190-50R or TA190-50R
1-2
mAs Measurements
The mAs function allows you to make mAs (milliampere x seconds) cursor measurements on input A or input B. You can do this on live waveforms, frozen waveforms (HOLD), recorded waveforms, and on saved waveforms.
Chapter 2
Using the Medical Functions
50 BNC CABLE
To perform mAs measurements on input A, do the following:
1
Connect the current shunt CS190-1R to input A using the BNC cable (see Figure 2-1)
2 Connect the current shunt resistor CS 190-1R to
the current source using the BNC to banana adapter (see Figure 2-1)
CS190-1R
I in
BLACK
I out
RED
Figure 2-1. mAs measurement
ST8722
2-1
Fluke 190B/C Medical
User Manual
3
4
5
6
Display the SCOPE key labels
Open the Reading 1 menu
Select on A, then accept
Select mAs, then accept. This opens the Current Probe menu.
7
Select for example 1 V/A, then accept.
After closing the information banner the screen shows (see figure 2-2) :
Top left reading, mA (or A, µA, etc.): the difference between the current amplitude measured on the left cursor and measured on the right cursor:
Top right reading, mAs (or As, µAs, etc.): the integrated current over the time between the cursors.
Figure 2-2. mAs measurement screen
Note: the B versions F3 key label does not show RMS as
these versions do not provide RMS measurements on the trace section between the cursors.
8 Highlight the left or right cursor.
9 Move the highlighted cursor to the
desired position on the waveform.
2-2
Using the Medical Functions
mAs Measurements
2
From step 2, you can also proceed as follows to do mAs measurements on input A:
3
4 Open the Probe on A menu.
5
6
7
In Scope mode display the INPUT
A key labels.
Select Probe Type: Current, then accept.
Select for example 1 V/A, then accept.
Display the cursor key labels (B­and C-versions are different).
8
9
10
11
Press to highlight . Observe that two vertical cursors are displayed.
Press to highlight mAs.
Highlight the left or right cursor.
Move the highlighted cursor to the desired position on the waveform.
mVs measurements
When you select Voltage in the Probe on A (B) menu, the F3 key label shows mVs.
mWs measurements
When you are in the Math AxB mode, and you select in the Probe on A (B) menu for one channel Voltage and for the other channel Current, the F3 key label will show
mWs.
2-3
Fluke 190B/C Medical
User Manual
Extended Offset
The extended offset function enlarges the vertical trace positioning range for input A and input B to 16 divisions (maximum). This allows you to zoom-in vertically to study small details of the signal.
To choose the extended offset range for input A do the following:
1
2 Open the INPUT A menu.
3
4
Display the INPUT A key labels.
Jump to Offset range.
(2x)
Select Extended and accept the extended offset range.
To move the trace:
5
When moving down, the following banner will show up:
Figure 2-3. Extended offset banner
When moving up a similar banner shows up.
Move the trace up or down until the ground marker ( - ) hits the top - or bottom grid line.
2-4
Using the Medical Functions
Extended Offset
2
Oscilloscopes and dynamic range
To prevent waveforms from being distorted, in particular signals with large very steep edges, the signal should stay within the dynamic range, just as with any Oscilloscope.
The dynamic range limits are the same as the trace offset limits given in the section Specifications below.
In practice you must take care when using high time base speeds when looking at signals that have large very steep edges. They can be distorted when they jump from outside the dynamic range directly to the visible area of the screen.
Figure 2-4 shows a pulse wave within the dynamic range (left), and a pulse wave that is partially outside the dynamic range (right). The example, particularly the 16 div trace offset, applies to one of the 5 most sensitive ranges.
+8
-
16
Avoid this for
steep edges
Trace offset limits (= dynamic range) Ground marker position example
Figure 2-4. Max. trace offset lowest ranges
2-5
Fluke 190B/C Medical
User Manual
Triggering
The extended offset range allows you to shift the signal up and down. Be aware that only the “visible” part is sampled by the ADC, and that the trigger level has to stay in the visible area too. Depending on what you are looking for, it might be required to switch to Edge Trigger or to time-qualified Pulse Trigger mode.
Tip for video signal measurements
To gain full profit of the Extended Offset for video signal measurements:
use the 50 10:1 attenuator feed-through terminator
select Probe on A attenuation 10:1
select range 500 mV/d
Now you can move the trace down for 16 divisions.
Specifications
Maximum trace offset with respect to center screen:
Vertical ranges 2 mV to 50 mV/div (direct 1:1):
positive going signals ...................................... -16 div
negative going signals ...................................+10 div
Other vertical ranges:
positive going signals ........................................ -8 div
negative going signals .....................................+8 div
2-6
Using the Medical Functions
Smart Average
2
Smart Average
Most users do not like to use averaging as the standard default mode. Occasional deviations in a waveform just distort the averaged waveshape, and do not show up on screen clearly. When a signal really changes, for instance when you probe around, it takes quite some time before the new waveshape is stable. With smart averaging you can quickly probe around, and incidental waveform changes like a line flyback in video show up on screen instantly without distorting the main signal.
To smooth the waveform, do the following:
1
2
Display the SCOPE key labels.
Open the Waveform Options menu.
3
4
5
6
Jump to the Average: field.
Select On… to open the Average menu.
Jump to the Average: field.
Select Smart and accept.
2-7
Fluke 190B/C Medical
User Manual
Smart average vs. normal average
Normal average smoothes the waveform of repetitive signals. It gives you the averaged waveform over successive acquisitions. A single waveform that deviates strongly will affect the smoothed waveform. This happens for example when looking at an interlaced video system. The fly-back line can distort the waveform, and you will not see it flash by as a separate waveform (See Figure 2-5).
Another characteristic of normal average is that when the signal really changes, for example if you are probing around, it takes some time before the new signal becomes stable. The displayed waveform slowly changes from one waveform to another waveform.
Smart average gives you the primary waveform, but displays an incidental trace of different waveshape directly and without affecting the primary waveform . You can now for example look at the averaged waveform on an interlaced video system, and still see the fly-back line flash by (See Figure 2-6).
When the signal changes are of more permanent nature, the averaging algorithm automatically restarts and displays the new waveform instantly, without displaying the distorted intermediate waveshapes.
Figure 2-5. Normal Average
2-8
Figure 2-6. Smart Average
Using the Medical Functions
Non Interlaced Video Triggering
2
Non Interlaced Video Triggering
To select triggering on a non interlaced video signal, do the following:
Apply a video signal to the red input A.
1
2
3
4
5
Display the TRIGGER key labels.
Open the Trigger Options menu.
Select Video on A to open the
Trigger on Video menu.
Select positive signal polarity for video signals with negative going sync pulses.
6
7
Select Non interlaced to open the Scan rate menu.
Select the required Scan rate and confirm the selection.
Trigger problems
The non-interlaced trigger circuits expect synchronization pulses that comply with standard video, or that are of similar amplitude but with shorter pulsewidth. The circuit however covers most other “standards” too. In case of incidental problems you always can define the trigger conditions in more detail using pulse width triggering.
2-9
Loading...