FLUKE ScopeMeter 190C, ScopeMeter 190B, ScopeMeter 120 User Manual

ScopeMeter®190C/190B Series
and ScopeMeter
®
120 Series
Technical Data
ScopeMeter 190C and 190B Series: Speed, performance and analysis power
For demanding applications, the ScopeMeter 190C and 190B Series high-performance oscilloscopes offer specifications usually found on top-end bench instruments. With up to 200 MHz bandwidth,
2.5 GS/s real time sampling and a deep memory of 27,500 points per input they’re ideal for engineers who need the full capabilities of a high-performance scope in a hand­held, battery powered instrument.
Dual input - 200, 100 or 60 MHz bandwidth
Up to 2.5 GS/s real time sampling per input
Connect-and-View™automatic triggering, a full range of man­ual trigger modes plus external triggering
Digital Persistence for analyzing complex, dynamic signals like on an analog oscilloscope (190C)
Fast display update rate for seeing dynamic behavior instantaneously
Automatic capture and replay of 100 screens
27,500 points and more per input record length using ScopeRecord™mode
TrendPlot™paperless recorder for trend analysis up to 22 days
1,000 V CAT II and 600 V CAT III safety certified
Waveform reference for visual comparison and automatic pass/fail testing (190C) of waveforms
Up to 1,000 V independently floating isolated inputs
Four hours rechargeable NiMH battery pack
ScopeMeter 120 Series: Three-in-one simplicity
The compact ScopeMeter 120 Series is the rugged solution for industrial troubleshooting and installation applications. It’s a truly integrated test tool, with oscilloscope, multi­meter and “paperless” recorder in one affordable, easy-to-use instru­ment. Find fast answers to problems in machinery, instrumentation, control and power systems.
A dual input 40 MHz or 20 MHz digital oscilloscope
Two 5,000 counts true-rms digital multimeters
Cursor measurements (Fluke 124)
A dual input TrendPlot recorder
Connect-and-View trigger simplicity for hands-off operation
Shielded test leads for oscillo­scope, resistance, continuity and capacitance measurements
Full bandwidth, VPS40 10:1 40 MHz probe included standard with Fluke 124
Up to seven hours battery operation
600 V CAT III safety certified
Optically isolated RS-232 interface
Rugged, compact case
E
G
M
A
E
N
N
A
T
M
S
Y
T
I
L
A
U

Q
C
E
R
T
I
F
I
E
I
D
T
T
E
O
E
M
Y
S
T
E
M
1
0
0
9
O
S
R
2 Fluke Corporation ScopeMeter 190C/190B Series and ScopeMeter 120 Series
Fluke 199C Fluke 196C Fluke 199B Fluke 196B Fluke 192B
Bandwidth 200 MHz 100 MHz 60 MHz Rise time 1.7 ns 3.5 ns 5.8 ns
Bandwidth limiter: User selectable: 10 kHz, 20 MHz or off Number of inputs: 2 plus external trigger. All inputs isolated from each other and ground.
Input coupling: AC or DC, with ground level indicator Input sensitivity: 2 mV/div to 100 V/div (Fluke 190C
Series); 5 mV/div to 100 V/div (Fluke 190B Series)
Variable attenuator: Variable gain on input channel A Input voltage: 1000 V CAT II, 600 V CAT III rated -
See “general specifications” for further details.
Vertical resolution: 8 bit Accuracy: ±(1.5% of reading + 0.04 x range/div) Input impedance: 1 M± 1% // 15 pF ± 2 pF
Connect-and-View
: Advanced automatic triggering
that recognizes signal patterns, automatically sets up and continuously adjusts triggering, time base and amplitude. Automatically displays stable waveforms of complex and dynamic signals like motor drive and control signals. Video triggering: NTSC, PAL, PAL+, SECAM. Includes field 1, field 2 and line select. Pulse width triggering: Pulse width qualified by time. Allows for triggering <t, >t, =t, t, where t is selec­table in minimal steps of 0.01 div. or 50 nsec. Time delay: One full screen of pre-trigger view or up to 100 screens (1200 divisions) of post-trigger delay
Automatic Capture of 100 Screens
The instrument ALWAYS memorizes last 100 screens (no user interaction or setup required). When an anomaly occurs on screen, the HOLD button can be pressed to review the full screen sequence over and over.
Instrument can be set up for triggering on glitches or intermittent anomalies and will operate in “baby-sit” mode and will capture 100 events.
Alternatively, the 199C and 196C can be set up in waveform compare mode to store only matching (“Pass”) or only non-matching (“Fail”) acquired wave­forms in the replay memory bank for further analysis. Replay: Manual or continuous replay. Displays the captured 100 screens as a “live” animation. Each screen is labelled with date and timestamp. The contents can also be viewed by manually scrolling backwards and forwards “screen by screen.” Replay storage: Up to 2 sets of 100 screens each can be saved for later recall and analysis
Waveform Compare and Pass/Fail Testing
Waveform compare: Provides storage and display of a reference waveform for visual comparison with newly acquired waveforms. Reference is derived from an acquired waveform and can be modified in the ScopeMeter or externally using FlukeView®Software. Pass/Fail Testing (199C, 196C): In waveform compare mode, the Color Scopemeter can be set up to store only matching (“Pass”) or only non-matching (“Fail”) acquired waveforms in the replay memory bank for further analysis.
Automatic Scope Measurements
Vdc, Vac rms, Vac+dc, Vpeak max, Vpeak min, Vpeak to peak, Aac, Adc, Aac+dc, frequency (Hz), risetime, fall­time, power factor, Watts, VA, VA reactive, phase, pulse width (pos/neg), dutycycle (pos/neg), temperature °C, temperature °F, dBV, dBm into 50 and 600
Vpwm ac, Vpwm ac+dc for measurement on pulse width modulated motordrives and frequency inverters
Cursor Measurements
Source: Input A, input B or the Mathematical Result trace Dual horizontal lines: Voltage at cursor 1 and 2, voltage between cursors Dual vertical lines: Time between cursors, 1/T between cursors (in Hz), voltage between markers, risetime with markers, falltime with markers Single vertical line: Min-Max and Average voltage at cursor position Zoom: Up to 8x horizontal zoom
Technical Specifications 190C and 190B Series
Oscilloscope Mode
Vertical Deflection
Display and Acquisition
Fluke 199C Fluke 196C Fluke 199B Fluke 196B Fluke 192B
Maximum 2.5 GS/s 1 GS/s 500 MS/s real-time sample rate
Number of 2 2 2 digitizers
Time base 5 ns/div 10 ns/div range to 5 s/div to 5 s/div
Horizontal
Fluke 199C Fluke 199B Fluke 196C
Fluke 196B Fluke 192B
Display 144 mm (5.67˝) full 144 mm (5.67˝)
color LCD monochrome LCD
Display modes Input A, input B, dual, Input A, input B, dual,
average, invert, replay, average, invert,
Digital Persistence mode replay, persistence
(short/medium/ (on/off)
long/infinite)
Maximum record length: 1,200 points per input in Scope mode; 27,500 points per input in ScopeRecord™roll mode (5 ms/div to 2 min/div)
Accuracy: ±(0.01 % of reading + 1 pixel) Glitch capture: 50 nsec (at 5 µsec/div to 1 min/div);
250 nsec (at 2 min/div)
Visible screen width: 12 divisions in scope mode Waveform mathematics: A + B, A - B, A * B, all
with user selectable scaling of resultant; A versus B (X - Y-mode) Acquisition modes: Normal, auto, single shot, ScopeRecord™roll, glitch capture, waveform compare, waveform compare with automatic “Pass / Fail testing” (in 199C and 196C only)
Trigger and Delay
Source: Input A, input B, external trigger input. All input references isolated from each other and from ground Modes: Automatic Connect-and-View,™free run, single shot, edge, delay, video, video line, selectable pulsewidth
Recorder Mode
ScopeRecord-Roll Mode
Dual input waveform storage mode
Source and display: Input A, Input B, Dual Memory depth: 27,500 points per input or more.
Each point consists of Min-Max pair Min-Max values: Min-Max values are measured at high sample rates ensuring capture and display of glitches
Time base range 5 ms/div to 1 min/div 2 min/div
Recorded timespan 6 sec to 24 hr 48 hr Glitch capture 50 ns 250 ns Sample rate 20 MS/s 4 MS/s Resolution 200 µsec to 2 sec 4.8 sec
Recording modes: Single sweep, continuous roll, Start-on-Trigger, Stop-on-Trigger Stop-on-Trigger (through external): ScopeRecord mode can be stopped by an individual trigger event, or by an interruption of a repetitive trigger signal
Horizontal scale: Time from start, time of day Zoom: Up to 100x Memory: Up to 2 dual input ScopeRecord waveforms
can be saved for later recall and analysis
TrendPlot™recording
Single or dual input electronic paperless chart recorder. Plots, displays and stores meter and scope measurements.
Source and display: Input A, Input B or DMM input Memory depth: 18,000 points recording. Per record
point a minimum, a maximum and an average value, plus a date and timestamp are recorded. Ranges: 5 s/div to 30 min/div in normal view mode; 5 min/div to 48 hr/div in view all mode, giving overview of total record Recorded timespan: Up to 22 days with a resolution of 1 minute Recording mode: Continuous roll for the duration of the full recordable timespan
Measurement speed: 5 measurements per second Horizontal scale: Time from start, time of day Zoom: Up to 64x zoom Memory: Up to 2 TrendPlot recordings can be saved
for later recall and analysis
Cursor measurements - all recorder modes
Source: Input A, B or DMM input Dual vertical lines: Min-Max or Average
voltage. Time between cursors Single vertical line: Min-Max or Average voltage. Absolute date and time or time from start
Meter Mode
Via 4 mm banana inputs. Fully isolated from scope inputs and scope ground. The specified accuracy is valid over the temperature range 18 °C to 28 °C (65 °F to 82 °F). Add 10 % of specified accuracy for each degree C below 18 °C or above 28 °C.
Maximum Resolution: 5,000 counts Voltmeter Ranges: 500 mV, 5 V, 50 V, 500 V, 1,000 V Accuracy:
Vdc ± (0.5 % + 5 counts) Vac true rms
15 Hz...60 Hz: ± (1 % + 10 counts) 60 Hz...1 kHz: ± (2.5 % + 15 counts)
Vac+dc true rms
dc...60 Hz: ± (1 % + 10 counts) 60 Hz...1 kHz: ± (2.5 % + 15 counts)
Ohms: Ranges: 500 , 5 k, 50 k, 500 k, 5 M, 30 M Accuracy: ± (0.6 % + 5 counts)
Other Meter Functions
Continuity: Beeper on < 50 (± 30 Ω) Diode test: Up to 2.8 V Amps: Adc, Aac, Aac+dc using an optional current
clamp or shunt
Scaling factors: 0.1 mV/A ... 100 V/A Temperature (°C, °F): With optional accessories. Scale
factors 1 mV/°C or 1 mV/°F
Input impedance: 1 M± 1% // 10 pF ± 2 pF Advanced meter functions: Auto/manual ranging,
relative measurements (Zero reference), TrendPlot recording
ScopeMeter 190C/190B Series and ScopeMeter 120 Series Fluke Corporation 3
4 Fluke Corporation ScopeMeter 190C/190B Series and ScopeMeter 120 Series
General Specifications
Case
Design: Rugged, shock proof with integrated protective holster
Drip and dust proof: IP51 according to IEC529 Shock and Vibration: Shock 30 g, Vibration
(sinusoidal and random) 3 g according to MIL-PRF-28800F Class 2
Display size: 115 . 2 x 86.4 mm (4.54 x 3.4 in.) Resolution: 320 x 240 pixels Contrast and brightness: User adjustable,
temperature compensated
Memory Save and Recall
Scope memories: 10 memory locations that each can contain two waveforms plus corresponding setup Recorder memories: 2 memory locations that each can contain 100 captured dual input scope screens, or a dual input ScopeRecord (27,500 Min-Max pairs per input), or a dual input Trendplot (18,000 min-max pairs + average values)
Real-Time Clock
Time and date stamp for ScopeRecord, 100 captured screens and TrendPlots
Power
Line power: Country specific line voltage adapter/battery charger included
Battery power: Rechargeable NiMH (installed) Battery operating time: 4 hours Battery charging time: 4 hours Battery power saving functions: Auto power down
with adjustable power down time. On screen battery power indicator
Mechanical Data
Size: 256 x 169 x 64 mm (10.1 x 6.6 x 2.5 in) Weight: 2 kg (4.4 lbs)
Safety
Compliance:
EN61010-1 (1993) Pollution degree 2 UL 3111-1 (1994) CAN/CSA C22.2 No.1010.1 (1992) ANSI/ISA S82.01 (1994)
Input Voltage Ratings
Maximum probe voltage: 1,000 V CAT II/600 V CAT III
(Maximum voltage between 10:1 probe tip (VPS200) and reference lead)
Floating voltage: 1,000 V CAT II/600 V CAT III
(Maximum voltage between earth ground and any terminal (signal input or shielding))
Independently isolated inputs:
1,000 V CAT II/600 V CAT III
(Maximum voltage between any terminal of one input or probe (VPS200) and any other terminal of another input or probe (VPS200))
Maximum voltage on BNC input directly (input A or B): 300 V CAT III Maximum voltage on meter input:
1,000 V CAT II/600 V CAT III
Environmental
Operating temperature: 0 °C to +50 °C Storage temperature: -20 °C to +60 °C Humidity:
10 °C to 30 °C: 95% RH non condensing 30 °C to 40 °C: 75% RH non condensing 40 °C to 50 °C: 45% RH non condensing
Maximum operating altitude: 3,000 m (10,000 feet) Maximum storage altitude: 12 km (40,000 feet) Electro-Magnetic-Compatibility (EMC): EN 61326-1
for emission and immunity
Optically Isolated PC / Printer Interface
PC communication: Transfer instrument settings, screen images, waveform data and waveform references, compatible with FlukeView®software for Windows®via optional PM9080. To printer: Supports HP Laserjet®, DeskJet, Epson FX/LQ and Postscript printers via optional PAC 91
Warranty
3 years, parts and labor on mainframe instrument 1 year on accessories
Fluke 199C, 196C Fluke 199B, 196B, 192B
Display Bright full-color LCD Bright monochrome LCD
with CCFL backlight, with CCFL backlight,
80 (30) Cd/m
2
125 (75) Cd/m2with
with (without) (without) power adapter
power adapter
Accessories
Standard Accessories Fluke 199C, Fluke 196C, Fluke 199B, Fluke 196B, Fluke 192B
Rechargeable battery pack (installed) BP190 Line voltage adapter / Battery charger BC190 Voltage probes and accessories 10:1 voltage probe (VPS200, 1 red + 1 grey) including hook clip, ground lead with mini alligator clip, ground lead with
hook clip, 4 mm add-on probe tip, ground lead to 4 mm banana plug Multimeter test leads TL75 Hard Point test lead set (1red, 1 black) User manual 9 language versions on CD-ROM, “Getting Started” booklet included with instrument
ScopeMeter 190C/190B Series and ScopeMeter 120 Series Fluke Corporation 5
Technical Specifications
ScopeMeter 120 Series
Cursor Measurements (124 only)
Sources: Input A, Input B Modes: Single or dual vertical cursor, dual horizontal
cursor, rise- or falltime
Measurements:
Single vertical line: Average, min value, max value, time from start of recording in roll mode Dual vertical lines: V at markers, time between cursors, 1/T between cursors (in Hz) Dual horizontal lines: High, low or V - readout, rise- and falltime: transition time, 0 %-level, 100 %­level, with markers at 10 % and 90 % Accuracy: As oscilloscope
Dual Input Meter
The specified accuracy is valid over the temperature range 18 °C to 28 °C (64 °F to 82 °F). Add 10 % of specified accuracy for each °C below 18 °C or above 28 °C (64 °F to 82 °F) Max. meter bandwidth: 40 MHz (Fluke 124), 20 MHz (Fluke 123)
Voltage measurements
Measurement selection: Vdc, Vac rms, Vac+dc rms, Vpeak max, Vpeak min, V
pk-pk
Ranges: 500 mV, 5 V, 50 V, 500 V, 1250 V Full scale reading: 5,000 counts
Accuracy
Vdc: ± (0.5 % + 5 counts) Vac rms:
1 Hz to 60 Hz: ± (1 % + 10 counts) 60 Hz to 1 kHz: ± (2.5 % + 15 counts) 20 kHz to 1 MHz ± (5 % + 20 counts)
Vac+dc true-rms:
DC to 60 Hz: ± (1 % + 10 counts) 60 Hz to 1 kHz: ± (2.5 % + 15 counts) 20 kHz to 1 MHz (5 % + 20 counts)
Vpeak:
Max peak or Min peak: 5% of full scale Peak-to-peak: 10% of full scale
Ohms
Ranges: 500 , 5 k, 50 k, 500 k, 5 M, 30 M Max. resolution: 5,000 counts Accuracy: ± (0.6 % of reading + 5 counts)
Capacitance
Ranges: 50 nF to 500 µF Max. resolution: 5,000 counts Accuracy: ± (2 % of reading + 10 counts)
Other meter functions
Frequency: Up to 70 MHz (Fluke 124) or up to 40 MHz (Fluke 123)
Continuity: Beeper on < 30 Diode test: Up to 2.8 V Amps: Amp dc, Amp ac, Amp ac+dc using an optional
current clamp or shunt.
Scaling factors: 0.1 mV/Amp to 100 V/Amp Temperature (°C, °F): With optional accessories. Scale
factors 1 mV/°C or 1 mV/°F
Number of inputs: 2 Input impedance: 1 M±1 % // 10 pF ± 2 pF Advanced meter functions: Auto/manual ranging,
TouchHold®, Relative measurements (zero reference), TrendPlot recording
Oscilloscope Mode
Vertical deflection
Bandwidth and Risetime Fluke 124 Fluke 123
Bandwidth (risetime)
with VPS40 probes 40 MHz 20 MHz
input A and B directly 40 MHz 20 MHz
with STL120 Shielded Test Leads 12.5 MHz 12.5 MHz
Instrument risetime (input directly) 8.75 ns 17.5 ns
Number of inputs: 2 Input coupling: AC, DC with ground level indicator Input sensitivity: 5 mV to 500 V/div (with the
included VPS40 (Fluke 124) and STL120 shielded test leads measure up to 600 Vrms, CAT III) Input voltage: 600 V CAT III. See General Specifica­tions for more detailed information
Vertical resolution: 8 bit Accuracy: ± (2 % of reading + 0.05 x range/div) Input impedance: 1 M±1 % // 225 pF with STL120
shielded test leads; 1 M±1 % // 20 pF ± 3 pF with BB120; 5 MΩ ±1 % // 15.5 pF with VPS40, 10:1 Volt­age probe
Horizontal
Max. sample rate (both channels simultaneously):
Fluke 124: 2.5 GS/s for repetitive signals; 25 MS/s for single shot Fluke 123: 1.25 GS/s for repetitive signals; 25 MS/s for single shot
Number of digitizers: 2 Time base range: 10 ns/div to 1 min/div (Fluke 124);
20 ns/div to 1 min/div (Fluke 123) Maximum record length: 512 Min-Max points per input
Accuracy: ± (0.1 % of reading + 1 pixel) Glitch detect: 40 ns
Display and acquisition
Display modes: Input A, input A and B, envelope, smooth Acquisition modes: Normal (including glitch capture), single shot, roll
Trigger and delay
Source: Input A, input B, external via optional ITP120 Modes: Automatic Connect-and-View, Free Run, Edge,
Single Shot, Video, Video Line Connect-and-View: Advanced automatic triggering that recognizes signal patterns and automatically sets up and continuously adjusts triggering, time base and amplitude. Automatically displays stable pictures of complex and dynamic signals like motor drive and control signals Video triggering: NTSC, PAL, PAL+, SECAM. Includes line select Time delay: Up to 10 divisions pre-trigger view
Measurements
Vdc, Vac, Vac+dc, Vpeak max, Vpeak min, Vpeak to peak, frequency (Hz), positive pulse width, negative pulse width, positive duty cycle, negative duty cycle, Amp ac, Amp dc, Amp ac+dc, Phase, Temperature °C, Temperature °F, dBV, dBm into 50 and 600 . (Amps, °C or °F with optional probes)
6 Fluke Corporation ScopeMeter 190C/190B Series and ScopeMeter 120 Series
Mechanical data
Size: 50 x 115 x 232 mm (2 x 4.5 x 9.1 in) Weight: 1.2 kg (2.64 lbs)
Safety
Compliance: EN61010.1 (1993) Pollution degree 2, UL3111-1 (1994), CAN/CSA-C22.2 No. 1010.1 (1992), ANSI/ISA S82.01 (1994)
Input voltage ratings
Maximum input voltage: 600 V CAT III
(Maximum voltage between input and reference lead)
Maximum input voltage using VPS40 Probe: 600 V CAT III, 1000 V CAT II (Maximum voltage between
probe tip input and reference lead)
Floating voltage: 600 V CAT III
(Maximum voltage between earth ground and any terminal (signal input or reference lead))
Maximum voltage between reference leads:
Instrument has common grounds connected via self recovering fault protection. For different ground potential measurements between inputs use DP120 differential voltage probe
Environmental
Operating temperature: 0 °C to +50 °C Storage temperature: -20 °C to +60 °C Humidity:
10 °C to 30 °C, 95% RH non condensing; 30 °C to 40 °C, 75% RH non condensing; 40 °C to 50 °C, 45% RH non condensing Maximum operating altitude: 2,000 m (6,500 feet); 3,000 m (10,000 feet) voltages 400 V
Maximum storage altitude: 12 km (40,000 feet) Electro-Magnetic Compatibility:
Emission EN50081-1 (EN55022 and EN60555-2) Immunity EN50082-2 (IEC1000–4-2, -3, -4, -5)
Optically isolated PC/Printer interface
To printer: Supports HP Laserjet,®Deskjet,®Epson FX/LQ and postscript printers via optional PAC91 To P C: Transfer instrument settings, screen images and data, compatible with FlukeView®software for Windows®via optional PM9080
Warranty
3 years, parts and labor on mainframe instrument 1 year on accessories
Recorder Mode
Trendplot recording
Dual input electronic paperless chart recorder. Plots and displays the actual, minimum, maximum and average of any measurement.
Source and display: Input A, Input A and B Range: 15 s/div to 2 days per division (automatic) Recorded timespan: Up to 16 days with a resolution
of 1.5 hours Recording mode: Continuous with automatic vertical scaling and horizontal time compression Measurement speed: 2.5 measurements per second maximum Horizontal scale: Time from start
General Specifications
Case
Design: Rugged, shock proof with integrated protective holster
Drip and dust proof: IP51 according to IEC529 Shock and vibration: Shock 30 g Vibration 3 g (sinu-
soidal) according to MIL-PRF-2800F Class 2
Display
Bright LCD with CCFL backlight, 60 (35) cd/m
2
with (without) power adapter
Size: 72 x 72 mm (2.8 x 2.8 inch) Resolution: 240 x 240 pixels Contrast and brightness: User adjustable,
temperature compensated
Memory Save and Recall
20 (10 in Fluke 123) instrument screens with user set-ups and user text
Real-time clock
Time and date stamp TrendPlot recording
Power
Line power: Country specific line voltage adapter/battery charger included Battery power: Rechargeable Ni-MH BP130 (installed in Fluke 124) or rechargeable NiCd BP120 (installed in Fluke 123) Battery operating time: Up to 7 hours using BP130, up to 5 hours using BP120 Battery charging time: 5 hours (Fluke 123), 7 hours (Fluke 124) Battery power saving functions: Auto power down with adjustable power down time. On-screen battery power indicator
Accessories
Standard Accessories Fluke 123, Fluke 124
Rechargeable battery pack (installed) BP120 (Fluke 123), BP130 (Fluke 124) Line voltage adapter / Battery charger PM8907 Voltage probes and accessories STL120 Shielded Test lead set; VPS40 high impedance 10:1 probe, 40 MHz (1 black, included with Fluke 124 only);
HC120 hook clips; ground leads with mini alligator clips; AC120 alligator clips; BB120 BNC-to-Shielded banana adapter Multimeter test leads TL75 Hard Point test lead (1 black) User manual 14 language versions on CD-ROM, “Getting Started” booklet included with instrument
Add user text to instrument setups and send these to the instrument for operator reference and instructions
Capture complete Replay sequence into the PC for further analysis and documentation
English, French and German versions included on a single CD-ROM
Note: Some functionality may be available with specific ScopeMeter models only
System requirements
Pentium 90 or better
CD-ROM drive
Windows®95 / 98 / Me / NT 4.0 / 2000 / XP
One free RS 232 port
PM9080 Optically isolated RS 232 adapter/cable, available separately or included in SCC190/SCC120 kit and in ScopeMeter ‘S’ versions
Supported Instruments
Full support for Fluke 199C, 199B, 199, 196C, 196B, 196, 192B, 192, 124, 123
Overvoltage Category Summary description
CAT IV Three phase at utility connection, any outdoors
conductors (under 1,000 V)
• Outside and service entrance
• Service drop from pole to building
• Run between meter and panel
• Overhead line to detached building
• Underground line to well pump
CAT III Three-phase distribution (under 1,000 V),
including single phase commercial lighting and distribution panels
• Feeders and short branch circuits
• Distribution panel devices
• Heavy appliance outlets with “short” connections to service entrance
CAT II Single-phase receptable connected loads
• Outlets and long branch circuits
• All outlets at more than 10 m (30 ft) from Category III source
• All outlets at more than 20 m (60 ft) from Category IV source
CAT I Electronic
• Electronic equipment
• Low energy equipment with transient limiting protection
ScopeMeter 190C/190B Series and ScopeMeter 120 Series Fluke Corporation 7
To protect your instrument and – more importantly – yourself, choose a test tool that can withstand the electrical hazards present in the environment in which you plan to use it.
EN61010 establishes international safety require­ments for electrical measurement equipment. It separates the various electrical environments into installation categories based on the danger from high
International Safety Standards
voltage-energy transients. To choose the right tool, the voltage rating alone does not determine the safety. It is the combination of voltage rating and installation category that determines maximum tran­sient withstand capability of the tool. CAT III rated instruments are recommended for measurement on industrial power distribution systems.
FlukeView software adds PC power to your Fluke ScopeMeter Test Tools.
FlukeView ScopeMeter software helps you get more out of your ScopeMeter:
Store instrument’s screen copies on the PC, in color (with Fluke 190C Series) or in black and white (Fluke 190B and 120 Series)
Copy color screen images into your reports and documentation (color screen images with Fluke 190C Series only)
Capture and store waveform data from your ScopeMeter on your PC
Create and archive waveform references for auto­matic (Fluke 190C Series) or visual (Fluke 190B and 190C Series) comparison
Use cursors for parameter measurement
Includes waveform analysis, e.g., FFT spectrum analysis
Copy waveform data into your spreadsheet for detailed analysis
Extended recording of up to four user-selected measurements help you monitor and analyze slow moving signals and related events
Logging of other readings directly into other appli­cation programs, eg., spreadsheet
FlukeView®ScopeMeter®Software for Windows
®
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries (425) 446-5500 or Fax (425) 446-5116 Web access: http://www.fluke.com/
©2002 Fluke Corporation. All rights reserved. Trademarks are the property of their respective owners. Printed in U.S.A. 11/2002 1629083 D-ENG-N Rev C
Fluke. Keeping your world
up and running.
Fluke-199C/S Color ScopeMeter (200 MHz / 2.5 GS/s)
with SCC190 kit Fluke-199C Color ScopeMeter (200 MHz / 2.5 GS /s) Fluke-196C/S Color ScopeMeter (100 MHz / 1 GS/s)
with SCC190 kit Fluke-196C Color ScopeMeter (100 MHz / 1 GS/s) Fluke-199B/S ScopeMeter (200 MHz / 2.5 GS/s) with
SCC190 kit Fluke-199B ScopeMeter (200 MHz / 2.5 GS /s) Fluke-196B/S ScopeMeter (100 MHz / 1 GS/s) with
SCC190 kit Fluke-196B ScopeMeter (100 MHz / 1 GS/s) Fluke-192B/S ScopeMeter (60 MHz / 500 MS /s) with
SCC190 kit Fluke-192B ScopeMeter (60 MHz / 500 MS/s) Fluke-124/S Industrial ScopeMeter, 40 MHz, with
SCC120 kit Fluke-124 Industrial ScopeMeter, 40 M Hz Fluke-123/S Industrial ScopeMeter, 20 MHz, with
SCC120 kit Fluke-123 Industrial ScopeMeter, 20 M Hz BP120 Rechargeable NiCd Battery for use with
Fluke 120 Series
BP130 Rechargeable NiMH Battery for use with
Fluke 120 Series
BP190 Rechargeable NiMH Battery for use with
Fluke 190 Series ScopeMeters
SW90W FlukeView ScopeMeter Software for
Windows PM9080 Optically isolated RS232 adapter/cable SCC190 Software - Cable - Case kit for Fluke
19 0 Series SCC120 Software - Cable - Case kit for Fluke
120 Series C195 Durable, universal soft carrying case for
ScopeMeters and accessories C190 Hard shell protective carrying case for
Fluke 190 Series ScopeMeters C125 Durable, protective soft carrying case for
Fluke 120 Series ScopeMeters C120 Hard shell protective carrying case for
Fluke 120 Series ScopeMeters DP120 Differential Voltage Probe for use with
Fluke 120 Series VPS40 40 MHz, 10:1 Voltage probe set for use
with Fluke 120 Series
Ordering Information
Selection Guide
19 0C Color ScopeMeter Series 190B ScopeMeter Series Fluke 120 Series
Fluke 199C Fluke 196C Fluke 199B Fluke 196B Fluke 192B Fluke 124 Fluke 123 Bandwidth 200 MHz 100 MHz 200 MHz 100 MHz 60 MHz 40 MHz 20 MHz Max real time sample rate 2.5 GS/s 1 GS/s 2.5 G S/s 1 GS/s 500 MS/s 25 M S/s
Max equivalent time
2.5 GS/s 1.25 GS/s
sample rate Display 144 mm full color LCD 144 mm monochrome LCD 102 mm monochrome LCD Digital persistence Yes, gives analog oscilloscope like
——
waveform decay (user selectable) Envelope mode Yes Yes Yes Waveform compare Visual reference and automatic Visual reference
‘Pass / Fail’ testing
Max record length
in Scope Mode: 1200 points per input channel 512 min/max points per input
in ScopeRecord mode: 27,500 points per input or more (5 ms/div to 2 min/div) — Number of inputs 2 plus external / DMM input, all isolated from each other and from ground 2 Number of digitizers 2 2 Independently floating Up to 1000 V between inputs, references and ground
isolated inputs Input sensitivity 2 mV/div to 100 V/div 5 mV/div. to 100 V/div 5 mV/div to 500 V/div Glitch capture
Up to 3 ns using pulse width triggering; 50 ns peak detect at 5 ms/div to 1 min/div
40 ns
Timebase range in Scope mode
5 ns/div to 2 min/div 10 ns/div to 10 ns/div to 20 ns/div to
2 min/div 1 min/div 1 min/div
Trigger types Connect-and-View™, Free Run, Single Shot, Edge, Delay, Video Frame, Connect-and-View™, Free Run,
Video Line, Selectable Pulse Width and External Single Shot, Edge, Video
Scope measurements
7 cursor measurements, 30 automatic measurements cursors + 26 26 automatic
automatic measurements
measurements Waveform mathematics A + B, A - B, A x B, A versus B (X-Y-mode, giving Lissajous diagrams) — ScopeRecord trigger modes Start on trigger, stop on trigger — Capture last 100 screens Automatic, with replay capability — Dual input TrendPlot Yes, with cursors and zoom Yes Memory for screens and 10 screens with set-up;
set-ups 5 more memories are made available upon registration of the ScopeMeter
20 10
Memory for recordings Two, each can store 100 scope screens, a ScopeRecord or a TrendPlot True-rms multimeter 5000 counts, Volts, Amps, Ohms, Continuity, Diode, Temp Safety certified (EN61010-1) 1000 V CAT II / 600 V CAT III (instrument and included accessories) 600 V CAT III (instrument and
included accessories) Battery (installed) 4 hr Ni-MH BP190 7 hr NiMH 5 hr NiCd Line Power Adapter / battery-charger included Size 25 x 16.9 x 6.4 cm (10.1 x 6.7 x 2.5 in) 232 x 115 x 50 mm
(9.2 x 4.5 x 2 in) Weight 2 kg (4.4 lb) 1.2 kg (2.64 lbs) PC and Printer Interface Using optional optically isolated RS-232 adapter/cable Warranty 3 years on main instrument, 1 year on the standard accessories
Detailed technical specifications and optional accessories can be found on the Fluke web site. There, you can also download a fully-functional instrument which is based on the real instrument firmware. Check it out at: www.fluke.com/scopemeter.