High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases
can be released unexpectedly and with extreme force. High pressure systems should be assembled
and operated only by personnel who have been instructed in proper safety practices.
Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any
form or by any means, electronic or mechanical, for any purpose, without the express written permission of DH Instruments, Inc.
4765 East Beautiful Lane Phoenix AZ 85044-5318 USA.
DH Instruments makes sincere efforts to ensure accuracy and quality of its’ published materials; however, no warranty, expressed
or implied, is provided. DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the
use of the information in this manual or products described in it. Mention of any product or brand does not constitute an
endorsement by DH Instruments of that product or brand. This manual was originally composed in English and was subsequently
translated into other languages. The fidelity of the translation cannot be guaranteed. In case of conflict between the English version
and other language versions, the English version predominates.
Products described in this manual are manufactured under international patents and one or more of the following U.S. patents:
5,142,483, 5,257,640, 5,331,838, 5,445,035. Other U.S. and international patents pending.
DH Instruments, DH, DHI, AutoZ, COMPASS, CalTool, PPC, PPC2+, RPM3 and SDS (Self Defense System) are trademarks,
registered and otherwise, of DH Instruments, Inc.
LabVIEW is a registered trademark of National Instruments Corporation.
Document No. 550094f-01
020909
Printed in the USA.
1.2.3.1 ONE OR TWO RPTS: GROUP 1 (< A1500) .............................................................................................6
1.2.3.2 TWO RPTS: ONE GROUP 1 (> A1500) AND ONE GROUP 2 (> A1500 AND < A10000) OR ONE
1.2.3.3 TWO RPTS: BOTH GROUP 2 (> A1500 AND < A10000) OR LIQUID FILLED GROUP 1.......................8
1.2.3.4 ONE RPT: GROUP 2 (> A1500 AND > A10000) OR GROUP 3 (> A10000) OR LIQUID FILLED
LIQUID FILLED GROUP 1 .........................................................................................................................
GROUP 1 ...................................................................................................................................................
2.3.1 PREPARING FOR OPERATION .................................................................................................................12
2.3.2 FRONT AND REAR PANELS ......................................................................................................................12
2.3.2.1 FRONT PANEL ........................................................................................................................................12
2.3.3 POWER CONNECTION...............................................................................................................................13
2.3.4 TEST PORT CONNECTING ........................................................................................................................13
2.3.4.1 THE ATM AND VENT PORTS .................................................................................................................14
2.4 POWER UP AND VERIFICATION .........................................................................................................15
3.1.1 MAIN RUN SCREEN....................................................................................................................................17
3.1.2 GENERAL OPERATING PRINCIPLES .......................................................................................................18
3.1.2.1 KEYPAD LAYOUT AND PROTOCOL......................................................................................................18
3.2.6.3 DEV (DEVIATION) ...................................................................................................................................32
3.2.9.1 RUNNING AUTOZ IN GAUGE MEASUREMENT MODE ........................................................................40
3.2.9.2 RUNNING AUTOZ IN ABSOLUTE MEASUREMENT MODE..................................................................41
3.3 [SETUP] MENU KEY .............................................................................................................................45
3.3.1 HEAD ...........................................................................................................................................................45
3.4.5.4 RESET - CAL ...........................................................................................................................................59
3.4.6 CAL ..............................................................................................................................................................59
4.3.3 QUERIES AND REPLIES ............................................................................................................................67
4.3.7.1 SCPI AND IEEE-488.2 .............................................................................................................................68
4.3.7.2 PROGRAM TECHNIQUE.........................................................................................................................68
4.4.2.6 SENSE SUBSYSTEM..............................................................................................................................84
4.4.2.7 STATUS SUBSYSTEM ............................................................................................................................86
4.4.2.8 SYSTEM SUBSYSTEM............................................................................................................................90
4.4.2.9 UNIT SUBSYSTEM..................................................................................................................................94
4.5 STATUS SYSTEM .................................................................................................................................96
4.5.1 STATUS REPORTING SYSTEM .................................................................................................................96
4.5.1.1 SCPI STATUS SUBSYSTEM...................................................................................................................96
4.5.1.3 STATUS BYTE REGISTER......................................................................................................................97
4.5.1.4 STANDARD EVENT REGISTER .............................................................................................................98
4.5.1.5 RPT READY STATUS REGISTER ..........................................................................................................99
4.5.2 STATUS SUBSYSTEM..............................................................................................................................100
5.6.1.3 POWER SUPPLY...................................................................................................................................117
5.6.1.6 MAIN BOARD.........................................................................................................................................117
Table 4 One or Two RPTs: Group 1 (> A1500) Specifications................................................................... 6
Table 5 Two RPTs: One Group 1 (> A1500) and One Group 2 (> A1500 and < A10000)
Table 6 Two RPTs: Both Group 2 (> A1500 and < A10000) or Liquid Filled Group 1 Specifications........ 8
Table 7 One RPT: Group 2 (> A1500 and < A10000) or Group 3 (> A10000) or
Table 8 RPM3 Parts List ............................................................................................................................ 11
Table 9. RPM3 Range Identification Summary.......................................................................................... 20
Table 10 Summary of RPM3 Direct Function Key Operation .................................................................... 22
Table 11 PresU - Available Units ............................................................................................................... 46
Figure 1 Rear Panel (Left) One or Two Group 1 RPTs (Right) One or Two Group 1 RPTs with
Figure 2 Schematic (Left) Two Group 1 RPTs (Right) One Group 1 RPT .................................................. 6
Figure 3 Rear Panel (Left) One Group 1 RPT and One Group 2 RPT (Right) One Group 1
Figure 4 Schematic One Group 1 RPT and One Group 2 RPT................................................................... 7
Figure 5 Two Group 2 RPTs (Left) Rear Panel / (Right) Schematic .......................................................... 8
Figure 6 Rear Panel (Left) One Group 2 RPT (Right) One Group 3 RPT ................................................... 9
Figure 7 Schematic (Left) One Group 2 RPT (Right) One Group 3 RPT .................................................... 9
Figure 8. Front Panel .................................................................................................................................12
Figure 9. Test Ports.................................................................................................................................... 13
This manual provides the user with the basic information necessary to operate an RPM3, Reference
Pressure Monitor. It also includes a great deal of additional information provided to help you optimize
RPM3 use and take full advantage of its many features and functions.
Before using the manual, take a moment to familiarize yourself with the Table of Contents structure. All first
time RPM3 users should read Sections 1, 2 and 3. Section 3 provides a comprehensive description of
general RPM3 operating principles. Section 4 is for remote operation from an external computer.
Section 5 provides maintenance and calibration information. Section 6 is a quick troubleshooting guide.
Use it to troubleshoot unexpected RPM3 behavior based on the symptoms of that behavior.
Certain words and expressions have specific meaning as they pertain to RPM3. The Glossary is useful
as a quick reference for exact definition of specific words and expressions as they are used in this
manual.
FOR THOSE OF YOU WHO “DON’T READ MANUALS”, GO DIRECTLY TO SECTION 2.3 TO SET UP YOUR
RPM3 AND THEN SECTION 2.4 FOR POWER UP AND VERIFICATION. THIS WILL GET YOU RUNNING
QUICKLY WITH MINIMAL RISK OF CAUSING DAMAGE TO YOURSELF OR YOUR NEW RPM3. THEN…
WHEN YOU HAVE QUESTIONS OR START TO WONDER ABOUT ALL THE GREAT FEATURES YOU MIGHT
BE MISSING, GET INTO THE MANUAL!
T
T
HHIISS
M
AANNUUAAL
M
L
Manual Conventions
(CAUTION) is used in throughout the manual to identify user warnings and cautions.
(NOTE) is used throughout the manual to identify operating and applications advice and
additional explanations.
[ ] indicates direct function keys (e.g., [RANGE]).
RPM3 is a stand-alone, microprocessor driven, reference pressure monitor intended to accurately
measure gas or liquid pressure in a wide variety of pressure calibration, measurement and testing
applications. It has been designed to provide very high performance and extensive features combined
with maximum versatility and ease of use.
RPM3 uses one or two high accuracy reference pressure transducers (RPTs) and an on-board barometer
to measure pressure. Various internal configurations exist depending on the RPT ranges.
RPM3 can be controlled locally by the operator using its front panel display and keypad or remotely by a
computer using ASCII character command strings over its RS-232 and IEEE-488 interfaces.
RPM3 models are available to measure pressure in ranges from 0.6 psi (4 kPa) to 40 000 psi (275 MPa)
in gauge, compound and absolute measurement modes.
1.2 SPECIFICATIONS
1.2.1 GENERAL SPECIFICATIONS
Specifications below are specific to PG7307. All other specifications are identical to PG7000
specifications found in the PG7000 Operation and Maintenance Manual (P/N 550099).
Power Requirements
Operating Temperature Range
Storage Temperature Range
Vibration
Weight
Dimensions
Microprocessors
Communication Ports
Overall Pressure Ranges
Operating Media
85 to 264 VAC, 47 to 440 Hz, 18 VA max. consumption
15 to 35 °C
-20 to 70 °C
Meets MIL-T-28800D
3.5 kg (7.7 lb) max depending on RPT configuration
8 cm H x 22.5 cm W x 20 cm D (3.1 in. x 8.9 in. x 7.9 in.) approx.
Motorola 68302
RS-232 (COM1), RS-232 (COM2), IEEE-488
Up to six pressure ranges from 0 to 0.6 psi (4 kPa) to 0 to 40 000 psi
(275 MPa), absolute, gauge and negative gauge, using one or two
reference pressure transducers and a barometer
Any non-corrosive gas or liquid (liquid available only on A0100 or
higher RPTs). RPTs < 10 000 psi (70 MPa) delivered set up for gas
operation, RPTs >
Test port (RPT > 10 000 psi): DH500 (gland and collar type for
coned and left hand threaded
tube, equivalent to AE F250C,
HIP HF4, etc.)
Test (-): 1/8 in. NPT F (with gauge RPT
only)
ATM Ref (On-board barometer): 10-32 UNF
Vent (with SDS only): 10-32 UNF
Maximum working pressure: Range 3 of active RPT+ 4 %
Maximum pressure w/o damage: Standard RPT: 125 % of H3
b Type RPT: 150 % of H3
Maximum pressure applied to
SDS protected TEST port
when SDS is active: 2 000 psi (13 MPa)
Included on all gas operated RPTs of 1 500 psi (10 MPa) or lower.
Isolates RPT from TEST port and vents it to atmosphere when NOT
in use. SDS is NOT included with liquid filled RPTs.
Maximum SDS isolation pressure: 2 000 psi (13 MPa)
Available. Must be specified.
1.2.2 PRESSURE MEASUREMENT SPECIFICATIONS
RPM3 can be configured with one or two reference pressure transducers (RPT). Each
transducer has three ranges (see Section 3.1.2.5). The type and designation of the RPM3’s
RPTs determine its measurement specifications and configuration (see Sections 1.2.2 and
1.2.3). Two types of RPTs are available. Standard RPTs are for the highest measurement
specifications and b Type RPTs for lower measurement performance requirements.
All reference pressure transducers (RPTs) with designation greater than G0030 (200 kPa)
are of the absolute pressure type with an evacuated, permanently sealed reference. Absolute
RPTs can measure both absolute and gauge pressure. Gauge pressures with an absolute
RPT are defined by offsetting atmospheric pressure dynamically with compensation for
atmospheric changes by the on-board barometer. Gauge RPTs cannot measure absolute
pressure.
Precision: Combined linearity, hysteresis, repeatability of measurements made by the reference pressure transducer. When using an absolute RPT
2
Stability: Maximum change in zero and span over specified time period for typical transducer used under typical conditions. As stability can only be
3
Accuracy: Maximum deviation of the reference pressure transducer indication from the TRUE value of the applied pressure including precision,
for gauge mode measurement add ± 2.5 Pa (0.00035 psi) to the precision specification to take into account the resolution and short term
stability of the on-board barometer used for dynamic atmospheric pressure compensation.
predicted and varies from transducer to transducer, stability for a specific RPT should be established from experience.
stability, temperature effect and calibration standard accuracy of ± 0.0035 % of reading.
1.2.2.2 B TYPE REFERENCE PRESSURE TRANSDUCER (RPT)
Table 2 b Type RPT Designations and Ranges
b Type RPT Measurement Specifications (% FS of Active Range)
Transducer Type
Warm Up Time
Resolution
Overpressure Limits
Without effect on calibration
Without permanent damage
Temperature Effect
Micro-machined piezoresistive silicon
None required
10 ppm, user settable by individual range
120 % of Range 3
150 % of Range 3
Fully compensated with active independent temperature
measurement from - 40 to 85 °C
± 0.02 % maximum temperature effect in normal ambient 15 to 35 °C
operating range
Acceleration Affect
Precision
Predicted Stability
1
2
Gauge Mode (w/Autozero)
Absolute Mode (w/Autozero)
Absolute Mode (w/out Autozero)
Measurement Accuracy
3
Gauge Mode (w/Autozero)
Absolute Mode (w/Autozero)
Absolute Mode (w/out Autozero)
1
Precision: Combined linearity, hysteresis, repeatability of measurements made by the reference pressure transducer. When using an absolute RPT
2
Stability: Maximum change in zero and span over specified time period for typical transducer used under typical conditions. As stability can only be
3
Accuracy: Maximum deviation of the reference pressure transducer indication from the TRUE value of the applied pressure including precision,
for gauge mode measurement add ± 2.5 Pa (0.00035 psi) to the precision specification to take into account the resolution and short term
stability of the on-board barometer used for dynamic atmospheric pressure compensation.
predicted and varies from transducer to transducer, stability for a specific RPT should be established from experience.
stability, temperature effect and calibration standard accuracy of ± 0.0035 % of reading.
Insignificant on all axes
0.03 %
90 day / 1 yr
0.02 % / 0.03 %
0.02 % / 0.03 %
0.03 % / 0.05 %
90 day / 1 yr
0.04 % / 0.05 %
0.04 % / 0.05 %
0.05 % / 0.06 %
1.2.2.3 ON-BOARD BAROMETER
Sensor Technology
Warm Up Time
Resolution
The on-board barometer is NOT used as a source of absolute accuracy. It is
used only to measure changes in atmospheric pressure for dynamic
compensation of the atmospheric pressure offset when using an absolute
reference pressure transducer to make gauge pressure measurements (see
Section 3.4.1).
RPM3 rear panel and internal schematic configurations vary depending on the number (1 or 2)
and pressure ranges of the RPM3’s RPT(s). For configuration purposes, RPTs are divided
into three groups by range designation as in Table 3.
Table 3 RPM3 Configurations
RPT GROUPRPT DESIGNATIONS
1 A1500 or lower 1/8 in. NPT F Yes
2 A2000 to A10000 1/8 in. NPT F No
3 Greater than A10000 DH500 F No
TEST
CONNECTION
SDS
In a single RPT RPM3, the RPT is referred to as the HI RPT.
In a dual RPT RPM3, the higher designation RPT is referred to as the HI RPT and the lower
designation RPT is referred to as the LO RPT.
Identify which RPT or RPTs your RPM3 is equipped with and refer to the corresponding
sections:
• One or two Group 1 (see Section 1.2.3.1)
• One Group 1 and one Group 2 (see Section 1.2.3.2)
• Two Group 2 (see Section 1.2.3.3)
• One Group 2 or one Group 3 (see Section 1.2.3.4)
Liquid filled Group 1 RPTs do NOT have SDS and MUST have their own test port.
The designation of the RPT(s) with which an RPM3 is equipped is given in the label window
just below the front panel display and on the power up introduction screen.
Remove the plastic caps from the RPM3 rear panel pressure connections.
Remove the protective plastic sheet from the front panel display.
Familiarize yourself briefly with the front and rear panels (see Sections 2.3.2.1 and
2.3.2.2).
2.3.2 FRONT AND REAR PANELS
2.3.2.1 FRONT PANEL
The front panel assembly provides a 2 X 20 vacuum fluorescent display of RPM3
operating status, a membrane keypad for local user interface and a soft ON/OFF key.
The rear panel assembly provides pressure connections, communications
interfaces and the power connection module. Pressure fittings are internally
secured to prevent loosening when making and breaking connections.
2
3
1. Pressure Port(s), layout depending on RPT
Configuration (see Section 1.2.3)
2. Electrical Power Connector Configurations
(IEC320-C13)
Figure 9. Test Ports
3. IEEE-488 (GPIB) Connector
4. COM1 Connector
5. COM2 Connector
6. Label, Product (on bottom of case)
4
2.3.3 POWER CONNECTION
• Connect the power cable to the rear panel power module
• Do NOT connect the other end of the power cable to a power source yet
RPM3 is always powered and active when power is supplied through the rear panel power
connector. The front panel ON/OFF key controls a soft ON/OFF (see Section 3.1.2.3).
2.3.4 TEST PORT CONNECTING
Depending on the reference pressure transducer (RPT) configuration of the specific RPM3,
the TEST port layout, TEST port fittings and acceptable test medium differ (see Sections
1.2.3 and 1.2.1).
Using a pressure hose or tube of appropriate pressure rating, connect the appropriate TEST
port to the test system or supply from which pressure is to be measured. The RPM3 TEST
port is either a 1/8 in. NPT F or a DH500 F, high pressure fitting. Always use compatible
male hardware of the same type. Use Teflon tape or another sealer to make 1/8 in. NPT
connections.
If RPM3 is equipped with a gauge RPT (GXXXX), it has a TEST(-) port. The TEST(-) port is
connected to the low side of the gauge RPT. This connection is normally left open to
atmosphere. It can also be connected to the low side of a differential device that is being
calibrated or tested. When measuring very low pressures, this may enhance the results by
helping assure that the RPM3 and the device under test reference ports are at the same
pressure (see Section 1.2.3).
USE THE CORRECT TEST PORT: Some RPM3s have more than one TEST port
corresponding to more than one internal RPT. Before connecting an RPM3 TEST port to a
pressure source, familiarize yourself with the RPTs, their pressure limits and their TEST
port fittings (see Section 1.2.3). In most cases, over-pressuring an RPT by more than
25 % will damage it beyond repair.
USE THE CORRECT PRESSURE CONNECTORS: RPM3 TEST port fittings are either 1/8 in.
NPT F or DH500 F (see Section 1.2.1). Never use fittings other than the corresponding
male fittings in these connectors. Damage to the connectors and dangerous failure
under pressure could result from using incorrect fittings.
NEVER connect a pressure source to the TEST(-) port. The pressure applied to this port
should be maintained at standard atmospheric pressure ± 3 psi (20 kPa). Exceeding
these limits may damage the RPT.
USE THE PROPER MEDIUM: b Type RPTs are intended for use only with dry (instrument
grade, non-condensing), non-corrosive gases. Contaminated gases may cause
measurement shifts and offsets. Standard RPTs are delivered for set up for gas
operation (A6000 and lower) or liquid operation (A10000 and higher). Contaminating
gas operated RPTs with liquids may cause aberrant measurements.
SDS Self Defense System: RPTs designated A1500 or lower include the SDS Self Defense
System. SDS, operated properly, allows a Group 1 RPT TEST port (see Section 1.2.3) to
be left connected to a pressure up to 2 000 psi (13 MPa) without damage to the RPT.
Do NOT attempt to use SDS in this manner without first becoming thoroughly familiar
with its operations and limitations (see Sections 3.1.2.6, 3.2.8, and 3.4.2).
2.3.4.1 THE ATM AND VENT PORTS
The ATM pass through is connected to the on-board barometer. This connection
assures that the on-board barometer actually measures ambient atmospheric
pressure rather than the pressure inside the RPM3 case that may vary slightly
from ambient pressure. The ATM port should be left open and unobstructed.
The VENT port is only present on RPM3s equipped with SDS Self Defense
System (Group 1 RPTs only) (see Sections 1.2.3 and 3.1.2.6). The VENT port is
connected to the SDS vent valve to assure that any gases vented through SDS
escape outside the RPM3 case. A connection may be made to the VENT port to
direct these gases if desired but the port must NOT be obstructed. Obstructing
the VENT port may interfere with SDS operation.
NEVER plug or obstruct the ATM pass through as this may adversely affect
gauge mode operation and autozeroing on an absolute transducer.
NEVER plug or obstruct the VENT pass through as this may interfere with
SDS operation and RPT autozeroing (see Section 3.1.2.6).
2.4 POWER UP AND VERIFICATION
2.4.1 APPLY POWER
Connect the RPM3 power cable to an electric supply of 85 to 264 VAC (47 to 440 Hz).
Observe the front panel display as RPM3 initializes, error checks and goes to the main run
screen (see Section 3.1.1).
RPM3 is always powered and active when power is supplied through the rear panel power
connector. The front panel ON/OFF key controls a soft ON/OFF (see Section 3.1.2.3).
If the RPM3 fails to reach the main run screen, service is required. Record the sequence of
operation and displays observed.
Any SDS present in RPM3 is active at power up. This causes SDS to flash over the
measured pressure (see Section 3.1.2.6).
The active range on power up is the same as the range that was active at the last power
If the RPM3 has an absolute RPT (designated AXXXX), check that it operates properly in
absolute mode.
Make sure that the TEST port of the RPT is open to atmosphere.
Use [RANGE] to change ranges if necessary (see Section 3.2.1). Select a range of the
absolute RPT.
Press [MODE] and select absolute mode (available on AXXXX RPTs only). Change the
pressure unit if desired (see Section 3.2.2).
If SDS is ON (SDS flashes over the pressure indication on the top display line), turn SDS
OFF. Press [SDS] and select <2 Yes> to defeat SDS (see Section 3.2).
Do NOT defeat SDS with a pressure higher than the active range maximum applied to the
Observe the current value of atmospheric pressure. Check that the value agrees with the
local value of atmospheric pressure within measurement tolerances. Repeat this process for
all the ranges on both RPTs if the RPM3 has two absolute RPTs. Check that the values of
atmospheric pressure measured by the different ranges agree with each other within RPM3
measurement tolerances (see Section 1.2.2). If they do NOT agree within tolerances, RPM3
may need calibration or repair.
2.4.2.2 CHECKING GAUGE MODE PRESSURE MEASUREMENT
Make sure that the TEST port(s) of the RPT(s) is/are open to atmosphere.
Use [RANGE] to change ranges if necessary (see Section 3.2.1). Select a range
of the RPT.
Press [MODE] and select gauge mode. Change the pressure unit if desired
(see Section 3.2.2).
If SDS is ON (SDS flashes over the pressure indication on the top display line),
turn SDS OFF. Press [SDS] and select <2Yes> to defeat SDS (see Section
3.2.8).
Do NOT defeat SDS with a pressure higher than the RPT maximum applied to
the TEST port. Damage to the RPT may result.
The value indicated should be near zero (± 6 psi, 35 kPa). Press [AutoZ]. This
runs AutoZ to zero the range (see Section 3.2.9). Upon return to the main run
screen, observe that the indication of measured pressure has zeroed.
Use [RANGE] to change ranges, repeat the zeroing process for each range.
If a range fails to zero properly, RPM3 may need repair.
It is normal for RPM3 to indicate a value other than zero when vented when
gauge mode is first entered or ranges are changed, especially if AutoZ is OFF
and/or RPM3 has been OFF for some time or its location has changed.
2.5 SHORT TERM STORAGE
The following is recommended for short term storage of RPM3:
Vent the RPM3 TEST port.
Disconnect the power supply.
When RPM3 will NOT be used for some time, it may be left powered, but use the soft ON/OFF key to turn
OFF the display.
RPM3 is designed to offer the optimum balance between simple, straight forward operation and the
availability of a wide variety of functions with a high level of operator discretion if desired. The local
operator interface is through a 2 x 20 character alpha-numeric display and a 4 x 4 multi-function keypad.
3.1.1 MAIN RUN SCREEN
The RPM3 main run screen is its home display that is reached on power up and from which
other functions and menus are accessed. It is the top level of all menu structures.
The main run screen is where the RPM3 is left in normal operation. It displays the current
measured pressure as well as a variety of additional information if desired.
1. <k> Ready/NOT ready indication, <k> when ready, <-> or <¯> indicating direction of measured pressure evolution when
NOT ready (see Section 3.1.2).
2. <PRESSURE1> Numerical value and sign of pressure
measured by active RPT and range. Shows result of last
average in average display mode (see Section 3.2.6.1).
<**SDS**>: To indicate when SDS is ON (see Section
3.1.2.6) for the active RPT, **SDS** is displayed alternating
with PRESSURE1.
3. <UNIT> Current unit of measure (see Section 3.2.2).
4. <M> Pressure measurement mode: <g> for gauge, <a> for absolute (see Sections 3.2.3 and 3.2.2).
5. <h> Indicates whether a head correction is applied. <h> if applied, blank if NOT (see Section 3.2.7).
6. <z> Indicates whether the autozero function is ON or OFF. <z> if ON; blank if OFF (see Section 3.4.1).
7. <RR> Indicates active RPT (<H> = high, <L> = low) and range (<1> = low, <2> = mid, <3> = hi) (see Section 3.2.1).
8. <DISPLAY FUNCTION> Information displayed depends on current display function.
9. <D> Indication of what is being displayed on the bottom line of the display as set by the DISPLAY function
(see Section 3.2.6). Choices include:
• <s> Current DISPLAY mode is average (see Section 3.2.6.1).
• <R> Current DISPLAY mode is rate (see Section 3.2.6.2); or if <n avg> is in the bottom right hand corner
of the display, current DISPLAY mode is average and this is the instantaneous reading average screen
(see Section 3.2.6.1).
• <H> Current DISPLAY mode is hi/lo (see Section 3.2.6.5).
• <D> Current DISPLAY mode is deviation (see Section 3.2.6.3).
• <S> <-> or <¯> Current DISPLAY mode is RPT (see Section 3.2.6.4).
• <F> Current DISPLAY mode is freeze (see Section 3.2.6.6).
• Blank, NO character Current DISPLAY mode is clean (see Section 3.2.6.7).
* PRESSURE1 UNITM hZRR
D DISPLAY FUNCTION
RPM3 has a screen saver function that causes the display to dim if NO key is pressed
for 10 minutes. Pressing a key restores full power to the display. The screen saver
activation time can be changed or screen saving can be completely suppressed (see
Section 3.4.7.1).
The RPM3 has a 4 x 4 keypad for local operator access to direct functions,
function menus and for data entry.
The [ENTER] key generally causes execution or forward movement in the menu tree.
The [ESCAPE] key moves back in the menu tree and/or causes execution to cease or suspend
without changes being implemented. Pressing [ESCAPE] repeatedly eventually returns to the
main run screen. From the main run screen, pressing [ESCAPE] allows momentary viewing of
the RPM3 introduction screen.
The [+/-] key changes a numerical sign when editing. It also toggles through multiple screens
when available.
The [
They are also used to scroll through choices.
] and []keys allow reverse and forward cursor movement when editing data entry.
Figure 10 Keypad
1. The Function/Data keys allow very commonly
used functions to be accessed directly from the
main run screen by a single keystroke. The
name of the function is on the bottom half of the
key (see Section 3.1.2.7). These keys enter
numerical values when editing.
2. The Editing and Execution keys are for
execution, suspending execution, backing up in
menus and editing entries.
3. The Menu/Data keys provide access to function
menus from the main run screen. The menu
name is on the bottom half of the key. The
SETUP menu is for more frequently used
functions (see Section 3.3). The SPECIAL menu
is for functions that are NOT generally used as a
part of day to day operation (see Section 3.4).
These keys enter numerical values when editing.
Key press confirmation is provided by both tactile
and audible feedback. A single tone confirms a
valid entry, a descending two note tone signals
an invalid entry. The audible valid entry feedback
can be suppressed or modified by pressing
[SPECIAL] and selecting <7Intern>, <2sound>
(see Section 3.4.7.2).
Some screens go beyond the two lines provided by the display. This is
indicated by a flashing arrow in the second line of the display. Use []
and [
] to move the cursor to access the lines that are NOT visible or
directly enter the number of the hidden menu choice if you know it.
3.1.2.2 SOUNDS
RPM3 is equipped with a variable frequency tone device to provide audible
feedback and alarms. The beeper is used for the following indications.
Valid key press - Brief beep, choice between three frequencies or NO sound is
available (see Section 3.4.7.2)
Invalid key press - Descending two tone blurp
Leak check routine completed - Three two second beeps (see Section 3.3.5)
UL (upper limit) exceeded - Intermittent one second beeps (see Section 3.2.4)
Pmax! (overpressure limit) exceeded - Eight second high frequency beep
RPM3 is equipped with a soft [ON/OFF] key and indicator LED on the bottom left
hand corner of the front panel. The purpose of the soft ON/OFF key is to put
RPM3 into a dormant mode in which the display is turned OFF but power is still
supplied and SDS and overpressure functions are still active. When RPM3 is
ON, the ON/OFF indicator is ON continuously. When RPM3 is soft OFF, the
ON/OFF indicator blinks every five seconds.
The soft [ON/OFF] key can also be used to reset from an overpressure condition
(see Section 3.2.4.1).
When RPM3 is soft OFF, receiving a remote command turns it ON.
Turning RPM3 soft OFF does NOT turn SDS ON. Do NOT assume that SDS
is ON when an RPM3 is soft OFF. Disconnecting power completely turns
SDS ON (see Section 3.1.2.6).
3.1.2.4 PRESSURE READY <*>/NOT READY (<↑> OR <↓>)
The character to the left of the measured pressure on the main run screen
provides a pressure Ready (<*>)/Not Ready (<↑> or <↓>) indication.
This indication is intended to provide the user with a clear and objective
indication of when a stable pressure has been achieved. Ready <*> is indicated
when the current stability (rate of change) of pressure is less than the stability
limit.
The user can set the stability limit (see Section 3.3.4). The ready indication is
often used when comparing the RPM3 and a test device to indicate when a valid
reading can be made.
Ready (<*>)/Not Ready (<↑> or <↓>) character indications are:
<k > Pressure ready (stable)
<↑> Pressure NOT ready (unstable) and decreasing
<↓> Pressure NOT ready (unstable) and increasing
3.1.2.5 MULTIPLE PRESSURE RANGES
RPM3 has one or two reference pressure transducers (RPT) each of which has
three ranges for a total of three or six pressure ranges. This multi-ranging
feature allows accuracy to be optimized for the range of pressure in which you
are working. Generally, the best range to select (see Section 3.2.1) is that
whose full scale is closest to, but greater than, the maximum pressure of the
device or system under test.
RPM3 handles all of the data operations needed to make range changes occur
transparently to the user when the RANGE function is used for range selection.
For a range change to be executed, the current pressure applied to the RPT on
which the range is being selected must be lower than the current upper limit (UL)
of that range (see Section 3.2.4).
When ranges are changed, the upper limit is automatically changed to the default
for that range or to the last upper limit set for that range. In addition, most other
functions and settings are range specific (see Section 3.2.1).
Each RPM3 has three or six ranges. In general, settings and operational
adjustments are specific to the range currently in use, as if you had six
instruments rather than one. The DISPLAY function, HEAD functions and
AUTO READRT function are NOT range specific. In remote mode, most
settings are RPT specific rather than
RPM3 internal pressure schematics and rear panel TEST port configurations
change depending on the number of RPTs and their ranges. See Section 1.2.3
for details on possible RPM3 configurations.
Ranges and Identification
The currently active RPT and range is continuously indicated in the upper right
hand corner of the main run screen and most other screens (see Section 1.2.2
for complete listing of the RPTs available and their ranges).
Hi RPT The RPT in a single RPT RPM3 or the RPT with the highest
maximum range in a dual RPT RPM3 is referred to as the Hi RPT.
Lo RPT The RPT with the lower maximum range in a dual RPT RPM3
is referred to as the Lo RPT.
Range 1, 2 or 3 The three ranges of an RPT are referred to as 1 = lo range,
2 = mid range, 3 = hi range.
Table 9. RPM3 Range Identification Summary
REFERENCE PRESSURE
TRANSDUCER AND RANGE
Lo RPT, Lo range Lo, 1 L1
Lo RPT, Mid range Lo, 2 L2
Lo RPT, Hi range Lo, 3 L3
Hi RPT, Lo range Hi, 1 H1
Hi RPT, Mid range Hi, 2 H2
Hi RPT, Hi range Hi, 3 H3
* The display symbol is included in the upper, right hand corner of most RPM3 menu displays as a convenient
indicator of active range.
DESIGNATION DISPLAY SYMBOL *
3.1.2.6 SDS SELF DEFENSE SYSTEM
RPTs designated A1500 or lower that are NOT liquid filled are equipped with the
SDS self defense system. The SDS system includes hardware and embedded
software logic to protect RPTs from overpressure. When SDS is ON for an RPT,
that RPT is isolated from the RPM3 TEST port and opened to the VENT port
(See Figure 11 and Sections 1.2.3 and 5.6.1.1).
In dual RPTs with a single TEST port, SDS is used to protect the Lo RPT from
overpressure when the high RPT is in use.
RPM3 internal logic uses SDS to protect against accidental overpressure.
Whenever RPM3 is ON (or in soft OFF mode) it monitors the pressure read by its
RPTs. If the pressure reaches the maximum pressure limit for a range, SDS will
activate automatically to isolate the RPT from the TEST port and vent it to
atmosphere.
SDS can also be used to isolate an RPM3 from a test system which may
sometimes be subjected to pressures higher than the RPM3s maximum
pressure. When configuring a multi-range system, this can eliminate the need
for external valving or connecting and disconnecting RPM3s. When used in this
manner, SDS should be activated prior to applying pressure. Automatic SDS
activation should only be used in an emergency overpressure situation.
SDS ON - ISOLATION Closed / VENT Open
SDS OFF - (Defeated) ISOLATION Open / VENT Closed
Figure 11 SDS Schematic
The maximum pressure that should be applied to an RPM3 TEST port when
SDS is ON is 2 000 psi (13 MPa).
Though the SDS self defense system includes features to automatically
protect RPTs against accidental overpressure, SDS should NOT be
considered a fail-safe overpressure protection system. SDS cannot
guarantee that overpressure damage will NOT occur. SDS automatic
activation may protect an RPT in case of accidental overpressure but
conventional measures for protecting against overpressure should also
always be followed. Damage to RPTs due to overpressure is NOT covered by
the RPM3 product warranty even when SDS is present.
3.1.2.7 DIRECT FUNCTION KEYS SUMMARY
Local operation of RPM3 is through the 4 x 4 pressure sensitive keypad. To
minimize the use of multi-layered menu structure, the 4 x 4 keypad numerical
keys also provide direct access to the most commonly used functions. The
function accessed is labeled on the bottom half of the keys. Direct function keys
are active whenever RPM3 is in its main run screen. Table 10 summarizes the
operation of the direct function keys.
Table 10 provides a brief summary of direct function key operation. It may be
useful to keep a copy of this summary near the RPM3, especially when first
becoming acquainted with its operation.
Table 10 Summary of RPM3 Direct Function Key Operation
Direct function keys are active from the main run screen.
See corresponding manual sections for full detail.
DIRECT FUNCTION KEYS ARE ACTIVE FROM ANY RUN SCREEN
See corresponding manual sections for full detail.
Menu of commonly used setup
features including run LEAK
CHECK.
(main, ambient, system).
View/adjust upper limit.
Menu of commonly used setup
features including run LEAK
CHECK.
Menu of less frequently used
internal functions and settings.
Adjust height of fluid head
correction. Set to zero to defeat
head correction.
Turn SDS (Self Defense System)
ON/OFF (if present). SDS ON is
indicated by flashing <***SDS***>.
SDS must be turned OFF to
measure pressures applied to
TEST port. Use with caution as
turning SDS OFF with
overpressure on test port could
damage RPT.
Runs AutoZ to rezero active
range. Should be used in gauge
mode whenever vented.
3.2 DIRECT FUNCTION KEYS
Adjust display resolution of
measured pressure and other
indications and settings.
Select [DISPLAY] function for
second line of RPM3 display.
Choices include <average>,
To view and/or change the active pressure measurement range.
PRINCIPLE
Each RPM3 has three or six ranges (see Section 3.1.2.5).
The [RANGE] key allows the range values to be viewed and a range selection to be made.
Most RPM3 settings such as pressure unit of measure (UNIT) and measurement mode
(MODE), are range specific. Changes made while in one range apply to that range only and
do NOT affect the other ranges (see Section 3.2.1).
OPERATION
Pressing [RANGE] activates the range viewing and selecting function. Pressing [RANGE]
key again or [+/-] while in the RANGE function steps through displays of available ranges:
Lo to Hi.