Fluke QA-ES III User Manual

PN 4473063 December 2015, Rev. 1
© 2015 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
QA-ES III
Electrosurgery Analyzer
Warranty and Product Support
This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration of instruments is not covered under the warranty.
This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since some jurisdictions do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you. If any provision of this warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
7/07
Notices
All Rights Reserved
Copyright 2015, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language without the written permission of Fluke Biomedical.
Copyright Release
Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke Biomedical.
Unpacking and Inspection
Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking the instrument. Notify the carrier and ask for an agent to be present while the instrument is unpacked. There are no special unpacking instructions, but be careful not to damage the instrument when unpacking it. Inspect the instrument for physical damage such as bent or broken parts, dents, or scratches.
Technical Support
For application support or answers to technical questions, either email techservices@flukebiomedical.com or call 1-800- 850-4608 or 1-440­248-9300. In Europe, email techsupport.emea@flukebiomedical.com or call +31-40-2965314.
Claims
Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all packing materials in their original condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not operate within specifications, or if there are any other problems not caused by shipping damage, please contact Fluke Biomedical or your local sales representative.
Returns and Repairs
Return Procedure
All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factory location. When you return an instrument to Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper packaging or handling.
Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide for repackaging:
Use a double–walled carton of sufficient strength for the weight being shipped.  Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive material around all projecting parts.  Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the instrument.
Returns for partial refund/credit:
Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obtained from our Order Entry Group at 1-440-498-2560.
Repair and calibration:
To find the nearest service center, go to www.flukebiomedical.com/service or In the U.S.A. and Asia: Cleveland Calibration Lab
Tel: 1-800-850-4608 x2564 Email: globalcal@flukebiomedical.com
To ensure the accuracy of the Product is maintained at a high level, Fluke Biomedical recommends the product be calibrated at least once every 12 months. Calibration must be done by qualified personnel. Contact your local Fluke Biomedical representative for calibration.
In Europe, Middle East, and Africa: Eindhoven Calibration Lab
Tel: +31-40-2675300 Email: ServiceDesk@fluke.com
Certification
This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was shipped from the factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibration standards are measured against in-house performance standards using accepted test procedures.
WARNING
Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. Fluke Biomedical will not be responsible for any injuries sustained due to unauthorized equipment modifications.
Restrictions and Liabilities
Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the information in this document will be incorporated in new editions of the publication. No responsibility is assumed by Fluke Biomedical for the use or reliability of software or equipment that is not supplied by Fluke Biomedical, or by its affiliated dealers.
Manufacturing Location
The QA-ES III Electrosurgery Analyzer is manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA, U.S.A.
Table of Contents
Title Page
Introduction .................................................................................................................... 1
Intended Use .................................................................................................................. 1
Safety Information .......................................................................................................... 2
Terminology ................................................................................................................... 5
Unpack the Product ........................................................................................................ 5
Product Familiarization ................................................................................................... 5
Turn On the Product ....................................................................................................... 8
Top Menus ..................................................................................................................... 8
Measurement Connections ............................................................................................ 10
Footswitch Connections ............................................................................................ 10
Neutral Connections .................................................................................................. 10
Setup Communications .................................................................................................. 11
USB Device Port ........................................................................................................ 11
Windows Software Driver ..................................................................................... 11
Device Manager .................................................................................................... 12
i
QA-ES III
Getting Started Manual
Wireless Port............................................................................................................. 13
Product Maintenance ..................................................................................................... 14
Cleaning .................................................................................................................... 15
Troubleshooting ........................................................................................................ 15
Replaceable Parts ..................................................................................................... 15
Accessories .............................................................................................................. 16
General Specifications ................................................................................................... 16
Technical Specifications ................................................................................................ 17
ii
Introduction
The QA-ES III (or the Product) measures the performance of high frequency Electrosurgery Units (ESU) and saves test records that you can transmit to a computer (PC). You can control the Product remotely from the Fluke Ansur software program.
The Product makes these measurements and tests:
Generator output: ο Power, RMS ο Current, RMS ο Voltage, peak-to-peak ο Crest Factor
Vessel Sealing Loop Current
HF Leakage Current in various configurations
Contact Quality Monitor (CQM) test
Power Distribution test automatically makes a series
of Generator Output measurements at various loads
Intended Use
The Product is a precision instrument for use in performing tests on high-frequency electrosurgical units in accordance with national and international standards. It is for use by trained service technicians. Tests include automatic power distribution measurement, crest factor measurement, RF leakage measurement, and CQM (contact quality monitor) test. The Product will be used in hospitals, clinical engineering departments, independent service organizations, and at ESU OEMs. The Product will not be used in patient rooms while a patient is present.
1
QA-ES III
Getting Started Manual
Safety Information
A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test.
Table 1. Symbols
Symbol Description
Table 1 is a list of symbols used on the Product or in this manual.
W Risk of Danger. Important information. See Manual. X Hazardous voltage. Risk of electric shock. P Conforms to European Union directives. ) Conforms to relevant North American Safety Standards.
Conforms to relevant Australian EMC standards.
This product complies with the WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With
~
reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste.
2
Electrosurgery Analyzer
Safety Information
XW Warning
To prevent possible electrical shock, fire, or personal injury, follow these guidelines:
Read all safety information before you
use the Product.
Use the Product only as specified, or the
protection supplied by the Product can be compromised.
Limit operation to the specified
measurement category, voltage, or amperage ratings.
Use the correct terminals, function, and
range for measurements.
Carefully read all instructions.
Do not touch voltages >30 V ac rms,
42 V ac peak, or 60 V dc.
Do not use the Product around explosive
gas, vapor, or in damp or wet environments.
Do not use the Product if it operates
incorrectly.
Examine the case before you use the Product. Look for cracks or missing plastic. Carefully look at the insulation around the terminals.
Do not use test leads if they are damaged. Examine the test leads for damaged insulation, exposed metal, or if the wear indicator shows. Check test lead continuity.
Use this Product indoors only.
Use only the mains power cord and
connector approved for the voltage and plug configuration in your country and rated for the Product.
Make sure the ground conductor in the mains power cord is connected to a protective earth ground. Disruption of the protective earth could put voltage on the chassis that could cause death.
Replace the mains power cord if the insulation is damaged or if the insulation shows signs of wear.
3
QA-ES III
Getting Started Manual
Use only current probes, test leads, and adapters supplied with the Product.
Connect the common test lead before the live test lead and remove the live test lead before the common test lead.
Only use probes, test leads, and accessories that have the same measurement category, voltage, and amperage ratings as the Product.
Remove all probes, test leads, and accessories that are not necessary for the measurement.
Do not connect measurement inputs directly to mains.
Do not use test leads if they are damaged. Examine the test leads for damaged insulation and measure a known voltage.
Disable the Product if it is damaged.
Do not use the Product if it is damaged.
Do not touch exposed metal on banana
plugs, they can have voltages that could cause death.
Remove circuit power before you connect the Product in the circuit when you measure current. Connect the Product in series with the circuit.
Connect an approved three-conductor mains power cord to a grounded power outlet.
Do not put the Product where access to the mains power cord is blocked.
Make sure that the Product is grounded before use.
Do not put metal objects into connectors.
Make sure that the space around the Product meets minimum requirements.
Do not use the Analyzer in CAT II, III, or IV environments.
Retractable end of test leads are for use on ESU only.
No probes or accessories supplied with the Analyzer are intended for handheld use. Setup and stand clear when activating the ESU with the footswitch.
4
Electrosurgery Analyzer
Terminology
Terminology
The Product uses the following terminology as described in IEC 60601-2-2:
HF – high frequency surgical signals also called RF (radio frequency).
Neutral Electrode also called Dispersive Electrode.
Contact Quality Monitor (CQM) also called Return
Electrode Monitor (REM trademarked by Covidien) or Return Electrode Current Monitor (RECM).
Unpack the Product
Carefully unpack all items from the box and check that you have the following items:
QA-ES III Electrosurgery Analyzer
Users Manual on CD
Getting Started Manual
Alligator clips, black and red
Dispersive safety lead
CQM safety lead
Jumper safety lead (Shorting Leads)
40 inch stackable/retractable safety leads, black, red
(2), blue, yellow, and green
20 inch stackable/retractable safety leads, black and red
Multi-stacking 4 mm banana plug patch cord, black
RECM Alarm Disabling Lead
Bipolar Activation Lead
USB cable
Ansur Software CD ROM
Product Familiarization
Figure 1 and Table 2 describe the controls and connections on the product.
5
QA-ES III
Getting Started Manual
1
SWITCHES
LOADS
5kV PEAK 400W MAX
ELECTROSURGERY ANALYZER
QA-ES III
CUT COAG
COMMON CQM
VARIABLE HI VARIABLE LO
FIXED FIXED
2
4
3
F1
F2 F3 F4 F5
BACK
8
USB
7
Figure 1. Product Controls and Connections
5
12
11
9
10
11
6
hwg001.eps
6
Electrosurgery Analyzer
Product Familiarization
Table 2. Controls and Connections
Item Description
Footswitch connections: CUT, COAG, COMMON
CQM connection for the Contact Quality Monitor test
Load connections: VARIABLE HI, VARIABLE LO, FIXED
LCD
Rotary selector knob
USB Device port
Function keys (F1 through F5)
BACK key
Scope out
Mains power connection and On/Off switch
Bumpers to protect back panel
Ground lug
7
QA-ES III
Getting Started Manual
Turn On the Product
Before you turn on the Product, check for damage or wear. Check for adequate ventilation. The Product requires a clear area of 10 cm (4 in) at the rear panel and all vent openings.
Connect the power cord to Mains power and push the power switch. The start-up sequence begins.
During the start-up sequence, a screen shows the firmware version for reference. You can use the start-up sequence to update the firmware.
After the start-up sequence, the Product starts the application. When the Top Menu screen shows, the Product is ready for use.
Top Menus
The Top Menus 1 and 2 have sub-menus to make measurements, set up the instrument, and maintain memory. Use the controls and softkeys to make selections from the menus. Table 3 describes the Top Menu controls.
For complete operating instructions, refer to the Users Manual on the accompanying CD. Manuals are also available at www.flukebiomedical.com.
8
Electrosurgery Analyzer
Top Menus
Table 3. Top Menu Controls
hwg002.eps
Softkey Goes to Softkey Goes to
F1 Generator Output measurement menu F1 Power Distribution measurement menu
F2 Vessel Sealing measurement menu F2 --
F3 HF Leakage measurement menu F3 Memory menu
F4 CQM (Contact Quality Monitor) test menu F4 Clock menu
F5 Top Menu 2 for more functions F5 --
Use the BACK key to return to Top Menu 1.
Note
hwg003.eps
9
QA-ES III
Getting Started Manual
Measurement Connections
Connect the ESU to the Product to make measurements. See the Users Manual for more information.
Footswitch Connections
Some ESUs use a footswitch to turn on the ESU generator. The Product has connections that simulate a footswitch. Use the footswitch feature to automatically turn on the ESU generator at an appropriate time.
Note
As an alternative to a footswitch, some ESUs use a hand switch. For this manual, hand switches and footswitches are both called footswitches.
It is not practical to make a single measurement or a Power Distribution test without the footswitch feature. Without the feature, you must manually turn on the ESU (or use an actual footswitch or hand switch) at the correct time after the delay starts.
The Product has connections for both CUT (yellow) jack and COAG (blue) jack footswitches and a COMMON (gray) connection jack.
The footswitch feature is available during all measurements and CQM tests. In continuous measurement mode, the Product activates the footswitch relays to turn on the ESU. The ESU remains on and the footswitch is engaged until you stop the measurement.
If you are not using the footswitch feature, ignore the footswitch connections.
Neutral Connections
ESUs with CQM alarms have two neutral connections. You must connect the neutral connections together for Generator Output, Vessel Sealing, and HF Leakage measurements.
The Shorting Leads supplied with the Product have a resistance that is sufficiently low to prevent an alarm on the ESU. Some ESUs require the Shorting Lead to have a minimum resistance, as well as, a maximum resistance. A simple jumper may not work. Use the Shorting Leads for all measurements except the CQM test.
10
Electrosurgery Analyzer
Setup Communications
Setup Communications
The Product has a USB Device Port for communication to a computer (PC). Some Products also have wireless functionality. You can use the communications ports to:
Send saved test records to a PC.
Send commands and receive responses to control
the Product remotely.
Use Ansur or a terminal emulation program (for example HyperTerminal or Tera Term) to set the COM port options. COM port settings are:
115,200 baud
No parity
8 data bits
1 stop bit
Hardware handshaking is on. (The Product uses
hardware handshaking but does not use XON/XOFF software handshaking.)
Operating system requirements:
Windows Vista
Windows 7
USB Device Port
The USB port on the Product (a Micro Type B connector) connects to a USB controller port on the PC (Type A rectangular connector). Use the supplied USB Type A-to­Micro-B cable to connect the Product to a PC.
The PC sees the USB Port while the cable is connected to the Product (even if the Product is turned off). If the cable is disconnected, the PC must close the port and reopen the connection.
Windows Software Driver
The USB port uses an integrated circuit (IC) to convert USB to RS232. Adapter cables frequently use this IC (FT232R from the company FTDI). When the Product connects to a PC for the first time, the PC registers the Product as a virtual COM port (VCP). The virtual COM port looks like a serial (RS232) device.
The IC is compatible with the USB Version 2.0 Full Speed specification. The USB ID numbers are: VID 0403 and PID 6001.
Windows 8 or later
11
QA-ES III
Getting Started Manual
Device Manager
The Device Manager is a PC program that changes COM port settings. Access the Device Manager according to your operating system, usually one of these methods:
From the Control Panel, select Device Manager.
Select Hardware and Sound and then select
Device Manager (under Devices and Printers).
Table 4 describes the functions of the Device Manager.
Note
While the Device Manager on the PC can set COM port options, the controlling program overwrites the Device Manager settings.
Table 4. COM Port Controls
Function Steps
1. Right-click USB Serial Converter and select
Enable the driver
Properties | Advanced.
2. Select Load VCP and click OK.
3. Click OK. The COM port shows in the Device Driver.
Table 4. COM Port Controls (cont.)
Function Steps
Verify COM port settings
Change the COM port number assigned by Windows:
1. Right-click the COM port.
2. Select Port Settings and
check the settings.
1. Right-click the USB Serial Port (COMx) and select Properties | Port Settings | Advanced.
2. Select the number from the COM Port Number menu and click OK.
3. Scan for hardware changes. The device list shows the new COM port number.
If the COM port number is in use, the number could be assigned to an unused device. Click through the error message to use the number.
12
Electrosurgery Analyzer
Setup Communications
Wireless Port
For Products with wireless functionality, the wireless port communicates with a PC that has an 802.15 (Bluetooth) wireless interface. For PCs without the interface, use a commercially available USB adapter. The PC starts the interface when you connect the adapter. (Additional software is not necessary.)
The PC sees the wireless port while the Product is on. When the Product is turned off, the PC closes the port. When the wireless device is assigned to a COM port, the COM port reopens when the Product is turned on again.
Note
The wireless port on the Product is a Classic Bluetooth port not a Low Energy Bluetooth port.
To install a wireless device:
1. Right-click the Bluetooth Devices icon and select
Add a Device, or select Show Bluetooth Devices | Add a Device.
The Product shows in the window. The serial number of the Product is part of the name.
Note
It is okay if the icon is a headset, or if the name is Bluetooth headset. These are defaults and the name will change to the Product.
2. Select the Product and click Next.
The system prompts you to compare the codes. Ignore the message and continue with the next step.
3. Make sure Yes is selected and click Next.
4. Select Driver Software Installation.
The system installs two standard serial-over­Bluetooth-link COM ports. The Bluetooth Peripheral Device will fail. Ignore the message and close the window. The Add a device window shows the device successfully added to the computer.
5. Close the Add a device window.
6. Right-click the Bluetooth icon and select Show Bluetooth Devices.
The Product name (including serial number) shows. Ignore the message about the missing driver for the Bluetooth Peripheral Device.
7. Right-click the Product and select Properties. The Hardware section shows a COM port for a
Standard Serial-over-Bluetooth link. Use this Outgoing COM port for the interface.
The Bluetooth Settings COM ports show that the Product has 2 COM ports: Outgoing (initiated by the PC) and Incoming (initiated by the Product). The system uses the Outgoing port only.
13
QA-ES III
Getting Started Manual
Wireless Settings—For Products with wireless functionality, Table 5 shows the settings. You do not need to change the default settings. Choose a method to open Bluetooth settings:
Right-click the Bluetooth icon and select Open Settings.
Right-click Bluetooth from the Start menu.
If you installed an adapter, select Control Panel |
Devices and Printers, then right-click the adapter and select Bluetooth Settings.
Table 5. Bluetooth Settings
Option Recommended Setting
Allow Bluetooth devices to connect to this computer.
Alert me when a new Bluetooth device wants to connect.
Show the Bluetooth icon in the notification area.
Allow Bluetooth devices to find this computer.
Selected (Required)
Selected
Selected
Not selected (The PC uses the Outgoing COM port to find the Product.)
Product Maintenance
XW Warning
To prevent possible electrical shock, fire, or personal injury, follow these guidelines:
Do not use an extension cord or adapter plug.
Do not operate the Product with covers removed or the case open. Hazardous voltage exposure is possible.
Disconnect the mains power cord and input cables before you remove the Product covers.
Remove the input signals before you clean the Product.
Use only specified replacement parts.
Have an approved technician repair the
Product.
After troubleshooting or maintenance, restart the Product and ensure that it starts without errors.
For radio frequency certification, see
www.flukebiomedical.com
14
Electrosurgery Analyzer
Product Maintenance
Cleaning
WCaution
For safe operation and maintenance of the Product:
Do not spray cleaning solution or water directly on the Product.
Do not pour or drip liquid onto the Product.
The Product needs little maintenance or special care. To clean, wipe with a damp cloth.
Troubleshooting
Table 6 shows possible problems and solutions.
Table 6. Troubleshooting
Symptom Resolution
The Product does not show the Top Menu.
The Product fails during the initial self-test.
Time stamp on saved records is not accurate.
Ensure the Product is connected to power.
Contact Fluke Biomedical Technical support.
Set the clock and then turn on the Product and keep it on for a minimum of 1 hour.
Replaceable Parts
Table 7 lists the replaceable parts for the Product.
Table 7. Replaceable Parts
Item
AC285 Large Alligator Clips, Black, Red
ESU-Dispersive Safety Lead 4635167 ESU-CQM Safety Lead 4635171 ESU-Jumper Safety Lead 2772209
Micro USB cable (2 m) 4114833
40 Inch Safety Leads, Black, Red (Stackable/Retractable)
40 Inch Blue Safety Lead (Stackable/Retractable)
40 Inch Yellow Safety Lead (Stackable/Retractable)
40 Inch Green Safety Lead (Stackable/Retractable)
Fluke Biomedical
Part Number
1610159
2772159
4635180
4635198
4635209
15
QA-ES III
Getting Started Manual
Table 7. Replaceable Parts (cont.)
Item
20 Inch Black Safety Lead (Stackable/Retractable)
20 Inch Red Safety Lead (Stackable/Retractable)
Fluke Biomedical
Part Number
4635211
4635227
40 Inch Black Safety Lead (Stackable) 4635230 Multi-stacking 4 mm banana plug
patch cord
4605232
RECM Alarm Disabling Lead 4635253 Bipolar Activation Lead 4635266
Accessories
Table 8 is a list of optional accessories.
Table 8. List of Optional Accessories
Item
International Dispersive Lead (1/4 in Phono Plug)
Test Probe Set, 0.080 Brass Tip 1909216
Fluke Biomedical
Part Number
4635248
General Specifications
Physical
Housing ............................... Metal case
Size (HxWxL) ...................... 14.5 cm x 35 cm x 47 cm
(5.75 in x 13.75 in x 18.5 in)
Weight ................................. 7.5 kg (16.5 lbs)
Power
Power Requirements .......... 100 V ac, 115 V ac, 230 V ac,
50 Hz / 60 Hz, universal input 100 V/115 V: 20 VA 230 V: 30 VA
User Interface
LCD ..................................... Monochrome 240 pixels x
64 pixels, 8 lines x 40 characters, white LED backlight
Keys .................................... 6 (1 fixed, 5 soft-defined) and
Rotary selector knob
Environmental Specifications
Temperature
Operating ........................ 10 °C to 40 °C (50 °F to 104 °F)
Storage ........................... -20 °C to 60 °C (-4 °F to 140 °F)
Humidity .............................. 10 % to 90 % non-condensing
Altitude ................................ 2000 m maximum
IP Rating ............................. IEC60529:IP20
Ansur QA-ES MK III Plug-In License 4704312
16
Electrosurgery Analyzer
Technical Specifications
Electromagnetic Compatibility (EMC)
IEC 61326-1: Basic Emissions
Classification ...................... IEC CISPR11: Group 1, Class A.
Group 1 have intentionally generated and/or use conductively coupled radio-frequency energy which is necessary for the internal functioning of the equipment itself. Class A equipment is suitable for use in non­domestic locations and/or directly connected to a low­voltage power supply network
USA (FCC) ......................... Intentional Radiators
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. (15.19)
Changes or modifications not expressly approved by Fluke could void the user's authority to operate the equipment. (15.21)
Safety
IEC 61010-1:Overvoltage category II, pollution degree 2 IEC 61010-2-030: Measurement 5,000 V
Wireless Module Listing
FCC (United States) compliant
(Class A)............................. FCC ID: X3ZBTMOD3
IC (Industry Canada)
compliant ............................ IC: 8828A-M OD3
CE (European) certified ...... CE0051
Technical Specifications
Specifications apply for a period of one year from date of the most recent calibration.
Measures ................................ Cut and coag waveforms
Power and current
measurements ....................... true RMS
Bandwidth ............................. 30 Hz to 5 MHz at -3 dB including
Delay Time for single
measurements ....................... 0.2 seconds to 4.0 seconds from
Duty Cycle
Variable Load ..................... 10 seconds on, 30 seconds off, at
Fixed 200 Load .............. 10 seconds on, 30 seconds off, at
Generator Output Measurements
Load Resistance
Variable ......................... 0 , 10 , 20 , 25 to 2500
Accuracy ........................ ±2.5 %
Monopolar and bipolar outputs
loads
Foot Switch activation to start of measurement
100 W, all loads
400 W
(by 25 ), 2500 to 5200 (by 100 )
17
QA-ES III
Getting Started Manual
Power
Ranges ............................ 0.0 W to 99.9 W
100 W to 500 W
Accuracy ...................... < 10 W: ±5 % + 1 W
10 W: ±5 %
Maximum
At 25% duty cycle (10 seconds on,
30 seconds off) ........... 10 : 300 W,
20 to 2900 : 400 W, 3000 to 5200 : 200 W
At 10% duty cycle (5 seconds on,
45 seconds off) ........... 10 : 300 W,
20 to 2400 : 500 W, 2425 to 2900 : 400 W, 3000 to 5200 : 200 W
Current
RMS ........................... 0 mA to 5,500 mA
Accuracy ..................... ±(2.5 % of reading + 1 mA)
Voltage
Peak .......................... 10 kV Peak to Peak
Accuracy ..................... ±(10 % of reading + 50 V)
Crest Factor ........................ 1.4 to 16.0
Defined as the ratio of Peak voltage to RMS voltage (Vpk / Vrms), using the larger of the 2 peaks (positive or negative)
Vessel Sealing Measurement
Loop Current, RMS ............. 0 mA to 5500 mA
Accuracy ............................ ±(2.5 % of reading + 1 mA)
HF Leakage Current
Fixed Load .......................... 200
Load Accuracy .................... ±2.5 %
Power rating ........................ 400 W
Additional Fixed Load ......... 200
Current, RMS ...................... 0 mA to 5500 mA
Accuracy ............................. ±(2.5 % of reading + 1 mA)
CQM Test (Contact Quality Monitor)
Resistances ........................ 0 to 475 (by 1 )
Accuracy ............................. 0 to 10 ±0.5 Ω,
11 and above ±5 %
Power rating ....................... 0.5 W
Auto Time interval ............... 1 second to 5 seconds
Oscilloscope Output
1 V per ampere of input current, typical.
Footswitch Simulations
Cut and Coag
Communications
USB Device Port ................. Micro B connector
Full speed
Wireless Port ...................... 802.15, Speed:115,200 baud
Memory
Test Records ...................... 5,000
Non-volatile ........................ retained through power cycling
Calibration
Traceable to the International System of Units (SI) through the appropriate National Metrology Institutes such as NIST or through intrinsic standards.
18
Loading...