In correspondence concerning this instrument plea se give the model number and
serial number as located on the type plate on the rear of the instrument.
NOTE: The design of this instrument is subject to continuous development and
improvement. Consequently, this instrument may incorporate minor
changes in detail from the information provided in this manual.
Copyright 1997, 1998 Fluke Corporation
All rights reserved. No part of this manual may be reproduced by any means or in
any form without written permission of the copyright owner.
Printed in the Netherlands
Thank you for purchasing this FLUKE oscilloscope. It has been designed and
manufactured to the highest quality standards to give you many yea rs of trouble
free and accurate measurements.
The powerful measuring functions listed below have been combined with an easy
and logical operation to let you use the full power of this instrument each and
every day.
If you have any comments on how this product could be improved, ple ase contact
your local FLUKE organization. FLUKE addresses are listed in the back of the
REFERENCE MANUAL.
III
The REFERENCE MANUAL also contains:
-CHARACTERISTICS AND SPECIFICA TIONS
-PRINCIPLES OF OPERATION
-BRIEF CHECKING PROCEDURE
-PERFORMANCE TEST PROCEDURES
-PREVENTIVE MAINTENANCE PROCEDURES
IV
MAIN FEATURES
There are five models in this family of FLUKE oscilloscopes. Each of these
models is a combination of an analog real-time oscilloscope and a fully featured
digital storage oscilloscope. By pressing a single key , you can switch the
instrument from the analog mode to the digital mode and back. This allows each
of the units to be used in an optimum operating mode for all kinds of signal
conditions. Complex data streams, modulated waveforms, and video signals can
often best be seen in the analog mode of oper ation. The digital mode of operation
is more suited for single events, signals with low repetition freque ncies, and when
automatic measurements need to be performed.
In this family there is a choice of five models. Two models have a ban dwidth of
200 MHz, two have a bandwidth of 100 MHz and one has a bandwidth of 60 MHz.
Beside the 2 channel models with EXT TRIG input, there is a choice of two mode ls
with four fully featured channels, all shown in the following table:
Type NumberBandwidthSample rate Number of
Channels
PM3370B
PM3380B
PM3384B
PM3390B
PM3394B
In the same instrument family, there are two 200-MHz and two 100-MHz analog
oscilloscopes that have specifications similar to the above-mentioned analog/
digital combination oscilloscopes operating in analog mode.
All analog/digital combination oscilloscopes listed above have the following features:
-Autoranging attenuators.
-Realtime clock.
-32K sample acquisition memory in 4 channel versions.
-8K sample acquisition memory, expandable to 32K in 2 channel versions.
-Up to 40 waveforms stored in memory or 204 waveforms with optional
memory extension.
-Autoset function for an instant optimized signal display at the touch of a button.
-Autoranging timebase.
-Cursor measurements with 1% accuracies.
-Extensive set of fully automated voltmeter and time measurement functions.
-Probe operated ’Touch Hold and Measure’ function freezes the display and
instantly displays the signal frequency, amplitude and dc voltage level.
60 MHz
100 MHz
100 MHz
200 MHz
200 MHz
200 MS/s
200 MS/s
200 MS/s
200 MS/s
200 MS/s
2
2
4
2
4
Input
Impedance
1 MΩ
1 MΩ
1 MΩ
1 MΩ/50Ω
1 MΩ/50Ω
-Peak detection for the capture of glitches as narrow as 5 ns.
-Pattern, State and Glitch triggering (2 ns) (2 channel models; 4ns Glitch
triggering only)
-Event delay and pretriggering and posttriggering.
-TV triggering including HDTV and TV line selection.
-Serial interface for printing and plotting.
-Averaging to reduce signal noise and to increase the vertical resolution from
8 to 16 bits.
-Advanced mathematics, including digital low-pass filtering. A Math+ option
adds integration, differentiation, histogramming, and FFT.
-Sine interpolation and magnification which enables true to life four cha nnel
single shot acquisitions with a timebase up to 625 ns/div (32x magnified)
-A delayed timebase with full trigger features.
-An RS-232 (EIA-232-D) interface (standard) and an GPIB/IEEE-488 interface
(optional).
-Autocal for automatic fine tuning of all circuitry to achieve maximu m accuracy
under all user conditions.
-Closed case calibration for efficient maintenance of traceable calibration at
minimum cost.
The following options are available:
-A MATH+ option with more automated measurement functions including
envelope and measurement pass/fail testing. Also in cluded in this option are
Integration, Differentiation, Histogramming, and FFT.
-Memory extension offering 32K acquisition length and the ability to store 156
traces (of 512 samples each) in memory for 2 channel versions.
-IEEE-488.2 interface using the new SCPI (Standard Commands for
Programmable Instruments) industry standard for remote control of test and
measurement equipment.
V
VI
INITIAL INSPECTION
Check the contents of the shipment for completeness and note whether any
damage has occurred during transport. When the conte nts are incomplete or
there is damage, file a claim with the carrier immediately. Then notify the FLUKE
Sales or Service organization to arrange for the repair or replacement of the
instrument or other parts. FLUKE addresses are listed in the back of the
REFERENCE MANUAL.
The following parts should be included in the shipment:
The performance of the instrument can be teste d by us ing the PERFORMANCE
TESTS in the REFERENCE MANUAL.
VII
INSIDE THIS MANUAL
This operating guide contains information on all of the oscilloscope’s features. It
starts with a general introduction, a summary of main capabilities, initial
inspection note and a front and rear view.
Operators safetyChapter 1 should be read before unpacking,
installing, and operating the instrument.
Installation instructionsChapter 2 describes grounding, line cord, fuses,
and backup batteries.
Getting startedChapter 3 provides a 10-minute tutorial intended
for those who are not familiar with Fluke
oscilloscopes.
How to use more advancedChapter 4 provides the more experienced user
functions of the instrumentwith a detailed explanation of the major functions
of the oscilloscope.
Function referenceChapter 5 contains an alphabetized description of
each function. Each description includes an
explanation of local and remote control functions.
CPL protocolChapter 6 provides the CPL commands with an
example of each.
Function indexThe Function Index lists all implemented
functions in alphabetical order.
IndexThe overall index contains all function names
and reference words in alphabetical order. It
includes the relevant chapter and page number
where more detailed information can be found.
Probe Adjust
Squarewave output signal for e.g. probe calibration.
Amplitude is calibrated.
CH1
BNC input socket for vertical channel 1 with probe
indication contact.
CH2
BNC input socket for vertical channel 2 with probe
indication contact.
CH3
BNC input socket for vertical channel 1 with probe
indication contact. (only in 4 channel models)
CH4
BNC input socket for vertical channel 1 with probe
indication contact. (only in 4 channel models)
EXT TRIG
BNC input socket used as an extra external trigger
input with probe indication contact (only in 2 channel
models)
XIII
Ground socket (banana): same potential as safety
ground.
The measuring ground socket and the external
conductor of the BNC sockets are internally
connected to the protective earth conductor of the
three-core mains cable. The measuring ground
socket or the external conductor of the BNC-sockets
must not be used as a protective conductor terminal.
XIV
REAR PANEL CONNECTIONS
Z-MOD
BNC input socket for external intensity-modulation
of the CRT trace.
NC
TXD
RXD
DTR
RS-232 BUS (EIA-232-D)
Input/output socket to connect the oscilloscope to an
RS-232 Interface.
LINE IN
Line input socket. Fuse holder is built in.
RTS
5
9
NC
CTS
ST6065
1
6
DSR
NC=NOT CONNECTED
FUSE
OPTIONAL REAR PANEL CONN ECTIONS
CH1 Y-OUT
BNC output socket with a signal derived from the
Channel 1 input signal.
MAIN TB GATE
BNC output socket with a signal that is "high" when
the Main Timebase i s running and "low" for the other
conditions.
DTB GATE
BNC output socket with a signal that is "high" when
the Delayed Timebase is running and "low" for the
other conditions.
EXT TRIG (only in 4 channel models)
BNC input socket used as an extra external trigger
input for the Main Timebase
DIO4
SHIELD
12
24
LOGIC
GND
SRQ
IFC
ATN
GND11GND
GND
10
NDAC
GND
9
GND8GND
DIO2
DAV
NR
EO1
DIO3
FD
DIO1
1
13
IEEE 488.2 BUS OPTION
If installed you will find here the input/output
socket to connect the oscilloscope to an
IEEE 488 interface.
DIO7
REN
7
DIO5
DIO6
DIO8
6
ST6064
XV
The external conductor of the BNC sockets
and the screening of the interface bus
connectors are internally connected to the
protective earth conductor of the three-core
mains cable. The external conductor of the
BNC sockets and the screening of the
interface bus connectors must not be used as
a protective conductor terminal.
OPERATORS SAFETY1 - 1
1OPERATORS SAFETY
ATTENTION:The instrument is designed for indoor use only.
Read this page carefully before installation and use of the
instrument.
1.1INTRODUCTION
The instrument described in this manual is designed to be used by proper-lytrained personnel only. Adjustment, maintenance and repair of the exposed
equipment shall be carried out only by qualified personnel.
1.2SAFETY PRECAUTIONS
For the correct and safe use of this instrument it is essential that both operating
and service personnel follow generally-accepted safe ty procedures in addition to
the safety precautions specified in this manual. Specific warning and caution
statements, where they apply, will be found throughout the manual. Where
necessary, the warning and caution statements and/or symbols are marked on
the apparatus.
1.3CAUTION AND WARNING STATEMENTS
CAUTION:Is used to indicate correct operating or maintenance
procedures in order to prevent damage to or destruction of the
equipment or other property.
WARNING: Calls attention to a potential danger that requires correct
procedures or practices in order to prevent personal injury.
1 - 2OPERATORS SAFETY
1.4SYMBOLS
Read the safety information in the manual.
Earth.
Conformité Européenne.
Recycling information.
1.5IMPAIRED SAFETY PROTECTION
The use of the instrument in a manner not specified may impair the protection
provided by the equipment. Before use, inspect the instrument and accessories
for mechanical damage!
Whenever it is likely that safety-protection has been impaired, the instrument
must be made inoperative and be secured against any unintended op eration. The
matter should then be referred to qualified technicians. Safe ty prote ction i s likely
to be impaired when, for example, the instrument fails to perform the intended
measurements or shows visible damage.
1.6MEASURING EARTH
The measuring earth socket and the external conductor of the BNC sockets are
internally connected to the protective earth conductor of the three-core mains
cable. The measuring earth socket or the external condu ctor of the BNC-sockets
must not be used to connect a protective conductor.
INSTALLATION INSTRUCTIONS2 - 1
2INSTALLATION INSTRUCTIONS
Attention:You are strongly advised to read this chapter thoroughly before
installing your oscilloscope.
2.1SAFETY INSTRUCTIONS
2.1.1Protective earthing
Before any connection to the input connectors is made, the instrument shall be
connected to a protective earth conductor via the three-core mains cable; the
mains plug shall be inserted only into a socket outlet provided with a protective
earth contact. The protective action shall not be negated by the use of an
extension cord without protective conductor.
WARNING: Any interruption of the protective conductor inside or outside
the instrument is likely to make the instrument dangerous.
Intentional interruption is prohibited.
WARNING: When an instrument is brought from a cold into a warm
environment, condensation may cause a hazardous
condition. Therefore, make sure that the grounding
requirements are strictly adhered to.
2.1.2Mains voltage cord, mains voltage range and fuses
Before inserting the mains plug into the mains socket, make sure that the
instrument is suitable for the local mains voltage.
NOTE: When the mains plug has to be adapted to the local situat io n , such
adaption should be done by a qualified technician only.
WARNING: The instrument shall be disconnected from all voltage
sources when a fuse is to be renewed.
The oscilloscope has a tapless switched-mode power supply that covers most
nominal voltage ranges in use: ac voltages from 100 ... 240 V (r.m.s.). This
obviates the need to adapt to the local mains (line) voltage. The nominal mains
(line) frequency range is 50 Hz ... 400 Hz.
Line fuse rating: 3.15 A T dela yed action, 250 V ( for ordering code see
"INITIAL INSPECTION").
2 - 2INSTALLATION INSTRUCTIONS
The mains (line) fuseholder is located on the rear panel in the mains (line) input
socket. When the mains (line) fuse needs replacing, proceed as follows:
-disconnect the oscilloscope from the mains (line).
-remove the cover of the fuseholder by means of a small screwdriver.
-fit a new fuse of the correct rating and refit the cover of the fuseholder.
WARNING: Make sure that only fuses with the required rated current and
of the specified type are used for replacement. The use of
makeshift fuses and the short-circuiting of fuse holders are
prohibited.
REAR VIEW
Figure 2.1Rear view of the instrument showing the mains input/fuse-holder
and back-up battery compartment.
When the apparatus is connected to its supply, terminals may be live, and the
opening of covers or removal of parts (except those to which access can be
gained by hand) is likely to expose live parts.
The apparatus shall be disconnected from all voltage so urces before it is opened
for any replacement, maintenance or repair.
Capacitors inside the apparatus may still be charged even when the apparatus
has been disconnected from all voltage sources.
Any maintenance and repair of the opened apparatus under voltage shall be
avoided as far as possible and, when inevitable, shall be carried out only by a
skilled person who is aware of the hazard involved.
INSTALLATION INSTRUCTIONS2 - 3
2.2MEMORY BACK-UP BATTERIES
2.2.1General information
Memory backup is provided to store the oscilloscope’ s settings when switched off
so that the instrument returns to the same settings when turned on. T wo AA (LR6)
Alkaline batteries are used.
Note:The batteries are not factory installed and must be installed at the
customer’s site.
Note:This instrument contains batteries. Do not dispose of these batteries with
other solid waste. Used batteries should be disposed of by a qualified
recycler or hazardous materials handler. Contact your authorized Fluke
Service Center for recycling information.
2.2.2Installation of batteries
Proceed as follows:
-Remove all input signals and disconnect the instrument line power.
-Remove the plastic cover of the battery compartment so that the battery
holder becomes accessible.
-Install two penlight batteries (AA) in the battery holder as indicated on the
battery holder.
-Reinstall the cover of the battery compartment.
Note:Frontsettings and autocalibration data disappear after exchange of the
batteries with the instrument disconnected from the line power. After
battery exchange, it is necessery to press the CAL key after the
recommended warming up time.
CAUTION: Never leave the batteries in the oscilloscope at ambient temper-
atures outside the rated range of the battery specifications because of possible damage that may be caused to the
instrument. T o avoid batt ery damage, do not leave the bat teries
in the oscilloscope when it is stored longer than 30 days.
2.3THE FRONT COVER
For ease of removal and reinstallation, the front cover has been designed to snap
on to the front of the instrument.
The front can be removed as follows:
-Fold the carrying handle down so that the oscilloscope occupies a sloping
position (refer to Chapter 2.4 for how to proceed).
-Pull the clamping lip at the top side of the cover slightly outwards.
-Lift the cover off the instrument.
2 - 4INSTALLATION INSTRUCTIONS
2.4HANDLE ADJUSTMENT AND OPERATING
POSITIONS OF THE INSTRUMENT
By pulling both handle ends outwards away from the instrument, the handle can
be rotated to allow the following instrument positions:
-vertical position on its rear feet;
-horizontal position on its bottom feet;
-in three sloping positions on its handle.
The characteristics mentioned in the REFERENCE MANUAL are guaran teed for
the specified positions or when the handle is folded down.
CAUTION:To avoid overheating, ensure that the ventilation holes in the
covers are free of obstruction. Do not position the instrument
in direct sunlight or on any surface that produces or radiates
heat.
In the rear panel of the instrument there is storage space for the mains cable.
There is also a clamping device to fix the end of the mains cable to the re ar panel.
The mains plug then fits in the area where the RS232 co nnector is present. In this
way the instrument can also stand on its rear feet.
MAT4221
Figure 2.2Instrument positions
2.5IEEE 488.2/IEC 625 BUS INTERFACE OPTION
If your oscilloscope is equipped with the IEEE 488.2 interface, it can be used in a
bus system configuration. The protocol used is SCPI (Standard Commands for
Programmable Instruments). For setup information, refer to the function
REMOTE CONTROL IEEE 488.2 in Chapter 5.
The IEEE 488.2 interface is a factory-installed op tio n.
INSTALLATION INSTRUCTIONS2 - 5
2.6RS-232-C SERIAL INTERFACE
Your oscilloscope is equipped with an RS-232-C interface as standard. The
interface can be used in a system for serial communication. The protocol used is
CPL (Compact Programming Language). CPL is a small set of very powerful
commands that can be used for full remote control. Detailed information about this
interface and the CPL protocol is given in Chapter 6 in this manual. For setup
information, refer to the REMOTE CONTROL RS-232 function in Chapter 5
’Function Reference’.
2.7RACK MOUNTING
The rackmount kit (PM 8960/04) allows you to install the oscilloscope in a
standard 19 inch rack.
It is not necessary to open the oscilloscope itself to mount the rackmount kit.
Installation can be done easily by the user.
2.8VERSIONS
The model number of your oscilloscope (e.g. PM33...) is indicated on the text strip
above the CRT. This model number is also represented by the digits 6, 7, 8 and 9
of the 12- digit code on the type plate on the rear panel. The ’A’ or ’B’ series is
indicated by a 1 or 2 on the 5th digit.
The instrument’s serial number is also given on the type plate. This numbe r
consists of a six digit code preceeded by the characters ’DM’.
The instrument version can also be displayed on the CRT after having pressed
menu key UTILITY and then softkey MAINTENANCE.
GETTING STARTED3 - 1
3GETTING STARTED
This chapter provides a 10-minute tutorial intended for those who are not familiar
with Fluke oscilloscopes. Those who are already fami liar can skip this chapter and
continue to Chapter 4.
3.1FRONT-PANEL LAYOUT
This oscilloscope is a combination of an analog oscilloscope and a digital storage
oscilloscope in the same instrument. The basic signal acquisition and display
functions are identical in both operating modes. Differences will be explained in
the text. Switching between the two operating modes is done with the yellow
ANALOG key.
The front panel of the oscilloscope is organized into functional areas. The areas
are discussed in order of typical operation.
Figure 3.1Front panel layout
Note that the front panel shown is that with the most functions. Differences are
explained in Section 4.1. For this getting started procedure, only CH1 and CH2
are used. These are identical for all models.
-More signal detail with the DTB(see Section 3.14)
-Trace storage(see Section 3.15)
3.2SWITCHING ON THE INSTRUMENT
Connect the power cord and set the front panel power switch to ON. For any line
source between 100V to 240V nominal, 50/400 Hz, the instrument auto matically
turns on. After performing the built-in power-up routine, the instrument is
immediately ready for use. The instrument’s settings will be identical to those
when the oscilloscope was switched off (with the batteries installed).
To ensure that you will get the same setup in all cases, press the
and
TEXT OFF
default condition (STANDARD SETUP) and a trace will appear on the screen.
Text is also displayed at the bottom of the screen.
key simultaneously. This will set the instrument in a predefined
STATUS
key
GETTING STARTED3 - 3
3.3SCREEN CONTROLS
The screen controls can be adjusted for optimum trace, text and spot quality by
the controls to the left of the screen.
Figure 3.2Screen control area
The brightness on the screen is adjusted by two controls, one for the trace and
one for the text.
•Turn the
TRACE INTENSITY
control clockwise and verify that only the
brightness of the trace increases.
•Turn the
TEXT INTENSITY
control clockwise and verify that only the
brightness of the text increases.
The sharpness of the trace and
text is optimized by the
When you are making photographs
or are in a dark environment, you
can use the
ILLUMINATION
control
to illuminate the graticule of the
screen.
The trace is adjusted in parallel with
the horizontal graticule lines by the
screwdriver-controlled
ROTATION
control.
TRACE
TRACE
ROTATION
FOCUS
control.
ST5975
9303
3 - 4GETTING STARTED
3.4AUTO SETUP
The best way to start each measurement is by using the AUTOSET key. This
automatically
finds and scales all relevant parameters on all channels.
AUTO SET
1
23
4
ST6659
9303
Figure 3.3Measuring setup
Step 1Connect the probe as shown in figure 3.3.
NOTE:AUTOSET is programmable. Because you have set the instrument in the
"standard setup" before (see Section 3.2), all programmable features are
set to a predefined condition an d the instrument is set in the analog mode .
Programming of AUTOSET is explained in Chapte rs 4 and 5.
Step 2Press the
AUTOSET
key.
The scope flashes the message ’AUTO SETTING...’ on the screen. In
a few seconds the front-panel settings are adjusted for an optimized
display of the applied signal in the analog mode.
Step 3The calibration signal is clearly displayed.
The parameters of the channel and the timebase settings are
displayed at the bottom of the screen.
CH1
200mV
MTB 200µs
CH1
ST6704
GETTING STARTED3 - 5
Step 4To prevent measurement errors, check the pulse response before any
measurement. If the pulse shows overshoot or undershoot, you can
correct this by using the trimmer in the probe’s body. Chapter 4
describes how to adjust the pulse response.
ST5952
In most cases, using AUTO SETUP is sufficient for a good initial display of the
signal(s). After the initial AUTOSET, and to optimize the signal for a more detailed
view, continue with the paragraphs below.
NOTE:If you get lost when adjusting your instrument, just press AUTOSET.
3 - 6GETTING STARTED
3.5MODE SWITCHING BETWEEN ANALOG AND
DIGITAL OPERATING MODES
Y ou can use the yellow ANALOG key to switch from the analog mode to the digital
mode and back at any time. The signal acquisition and display functions of both
operating modes are very similar. However, the nature of the signals you are
using may determine which operating mode you prefer to use. For more
information, refer to the following table:
SIGNAL CRITERIAANALOG MODEDIGITAL MODE
Repetitive signals of UsableUsable
30 Hz and higher
Repetitive signals Causes displayPreferred
below 30 Hzflickering
Single eventsDisplayed for theCan capture and
duration of display for long
the eventterm display
Repetitive signals that arePreferredMay cause alaising
amplitude modulatedUse Peak detect or
Envelope mode
Repetitive signals thatPreferredMay cause aliasing.
are modulated in frequencyUse Envelope mode.
Long serial data streamsPreferred whenWhen using delayed
Delayed sweepsweep to observe
is not used.details, Digital mode
provides better
light output.
Video signalsPreferred whenWhen using delayed
Delayed sweepsweep to observe
is not used.details, Digital mode
provides better
light output.
OTHER CRITERIA
Need to see pretriggerNot possibleUp to full acq uisition
informationlength
GETTING STARTED3 - 7
SIGNAL CRITERIAANALOG MODEDIGITAL MODE
You need to make adjustmentsFastestSlower
to the circuitry and watchdisplaydisplay
the signal changeupdateupdate
Automatic measurementsCan’t useFully implemented
Signal MathAdd, SubtractAll functions
Autorange attenuatorNot availableResults in a displayed
signal with an amplitude of 2 to 6.4 divisions
Autorange timebaseNot availableResults in a signal
display of 2 to 6
waveform periods
ANALOG
1
23
4
RUN/STOP
ST6680
9312
Figure 3.4Analog-Digital switching setup
Step 1Press AUTOSET. The scope performs an AUTOSET in analog mode.
Step 2Press the ANALOG key to change over to the digital mode. Check that
the picture is identical to the one in the analog mode. The text ’DIGIT AL
MODE’ is displayed briefly at the bottom of the screen.
3 - 8GETTING STARTED
Step 3Press AUTOSET again. This time the scope performs the autoset in
digital mode.
Step 4Press the RUN/STOP key and observe that the trace is frozen and
stays on screen even after removing the probe.
Step 5Press the RUN/STOP key to display the actual input signal again.
Reconnect the probe to display the Probe Adjust signal again.
Step 6Press the ANALOG button once again to return to the analog mode. In
the bottom of the screen, the text ’ANALOG MODE‘ is briefly displa yed.
3.6VERTICAL SETUP
This section deals with setting of the input circui ts of the four channels. The ma in
adjustments are AMPLitude, POSition, and the channel input coupling selection
for GND, DC, and AC.
Figure 3.5Vertical setup
Step 1Adjust the absolute ground level by disconnecting the signal and using
the POS control to position the trace in the middle of the screen. A
marker with the channel number (’1-’) at the left of the screen indicates
the ground reference.
POS
1
1
MAT4191
Step 2Reconnect the probe to the Probe Adjust signal for display.
GETTING STARTED3 - 9
CH1
100mV
CH1
500mV
ST6681
Step 3You can change the amplitude of the signal in a 1, 2, 5 sequence by
pressing one of the AMPL keys. Note that the bottom of the screen
shows the AMPL/DIV setting of CH1.
Step 4Press the ON button of CH2 and observe that a second trace is now
visible. The position and amplitude of this channel can be adjusted
similar to the adjustment of CH1. The channel settings are also
displayed in the bottom of the screen.
Press the ON key of CH2 once again to turn this channel off.
Step 5Press the AC DC/GND key of CH1 so that a ’⊥’ sign is displayed in the
bottom text line. This interrupts the input sig nal and connects the inpu t
to the ground. In this case, only the ’base’ line is visible.
Press the AC DC/GND key once again for ac input coupling; the
bottom text line now displays ’~’.
In most cases, dc input coupling is used to show ac as well as dc components of
the signal. However, in some case s where a small ac sign al is superimp osed on
a large dc voltage, ac input coupling must be used. Then only the ac component
is visible on the screen. The text line shows a ’=’ or ’~’ sign to indicate dc or ac
coupling. Because the calibration signal is a square wave with a low level of 0V
and a high level of +600 mV , the screen sh ows either of the following two displays:
AC INPUT COUPLING
ZERO
1
LEVEL
CH1
200mV
MTB 200µs
CH1
DC INPUT COUPLING
ZERO
1
LEVEL
CH1
200mV
MTB 200µs
CH1
ST6682
3 - 10GETTING STARTED
Step 6Press the ANALOG key to enter the digital mode
Step 7Press the top one (mV) of the AMPL keys, so that the signal has
maximum amplitude.
Press AUTO RANGE and see the signal change to a suitable
attenuator value. When AUTO RANGE is active, the attenuators
automatically adjust when the signal amplitude changes, to keep the
trace on screen.
Step 8Press the key labeled AVERAGE. Noise in the input signal can be
reduced by using the average function. The random noise is reduced
by calculating the average over the last n scans (average factor ca n be
set between 2 and 4096)
NOTE:Refer to Chapter 4 for an explanation of the CH1+CH2, TRIG1, and
INV keys.
GETTING STARTED3 - 11
ST6683
MTB 100µsMTB 500µs
TIME/DIV
3.7TIMEBASE SETUP
The next step is the adjustment of the main timebase controls (
POS
ition, and
MAGNIFY
keys).
MAGNIFY
AUTO
3
2
1
4
RANGE
TIME/DIV
sns
Figure 3.6Timebase setup
Step 1Press the
Step 2Use the
AUTOSET
TIME/DIV
key.
keys on the right hand side of the instrument to
decrease or increase the number of periods of the signal on the
screen.
TIME/DIV, X
VAR
XPOS
ST6435
9312
Step 3Select a timebase of 1 ms/div.
Step 4Press
AUTO RANGE
and see the signal display change to a more
suitable timebase. The AUTO RANGE function au tomatically selects a
timebase that displays 2 to 6 signal periods.
Step 5Press the
Step 6Turn the
ANALOG
X POS
key to switch the scope to the analog mode.
control to shift the signal horizontally (left or right)
across the screen.
3 - 12GETTING STARTED
3.8MAGNIFY (EXPAND)
Step 1You can use the
MAGNIFY
keys to expand the signal on the screen.
The ’MGN’ indication and the corrected timebase settin g are displayed
in the text line.
In the analog mode, magnification is limited to
activate the magnification. The left key will turn off the MAGNIFY
10. The right key will
*
function. On or off is indicated by ’MGN’ in the bottom of the screen.
Step 2Press the
ANALOG
Step 3In the digital mode, pressing the right
signal in
left
operate the
1, *2, *4 ...steps to a maximum of *32 times. Pressing the
*
MAGNIFY
MAGNIFY
key to switch the scope to the digital mode.
MAGNIFY
key compresses the signal to
buttons, or when you turn the X
key expands the
1 again. When you
*
POS
ition
control, a bargraph is displayed showing whic h part of the digital trace
is expanded.
NOTE:The MAGNIFY key and X POS control can also be used after the
oscilloscope is STOPped.
5.00µs
10 DIV
1.25µs
4
*
10 DIV
START OF TIME WINDOW CAN BE VARIED WITH
XPOSOVER THE WHOLE SWEEP RANGE.
1.00ms
40 DIV
ST6684
GETTING STARTED3 - 13
3.9DIRECT TRIGGER SETUP
Now you are ready to set your trigger conditions. Y ou will use one of the channel
selection keys (
control.
LEVEL
TRIG1, TRIG2, TRIG3, TRIG4 or EXT TRIG
Figure 3.7Direct trigger setup
) and the
TRIGGER
Step 1Press the
output is now displayed on channel 1. Turn chann el 2 on to display a
second horizontal trace (channel 2 has no input signal).
Step 2Press the
source. The result is that the signal on channel 1 is no longer triggered
(not stable). The
not triggered. Check also that the right side of the bottom text line
indicates the trigger source (’ch2’).
Step 3Only in 2 channel models
Press
source. Check that the rightside of the bottom textline indicates the
trigger source ’EXT’.
Step 4Press the
The ’ch1’ symbol is displayed in the bottom text line. Triggering
resumes. Turn channel 2 off, by pressing ON again.
Step 5Press the ns key of the MainTB
set to ’2 µs/div’.
AUTOSET
TRIG2
EXT TRIG
TRIG1
key. The square-wave signal of the Probe Adjust
key so that channel 2 is selected as the trigger
LED is on, to indicate that the oscilloscope is
ARM’D
key to select External Trigger input as the trigger
key. Channel 1 is now selected as the trigger source.
TIME/DIV
keys until the timebase is
3 - 14GETTING STARTED
Step 6The same TRIG1 key that was used to select the trigger source is also
used to select the trigger slope. Repeatedly pressing the TRIG1 button
changes the triggering so that it occurs on the leadin g o r tr aili ng edge
of the input signal. Note that the slope is also displayed in the bottom
text line.
TRIG1
MTB 2µs
CH1
200mV
CH1
CH1
200mV
MTB 2µs
CH1
ST6685
Step 7For repetitive signals, you obtain a stable display when each
successive timebase sweep is triggered at the same stable level of the
input signal. You use the TRIGGER LEVEL control to adjust the level.
Turn the control. The precise positio n in relation to the maximum signal
amplitude (between +100 % and -100 %) is displayed on the screen.
SUMMARY
The previous steps covered the basic adjustments. Now you are ready to look
at the special features of the oscilloscope. This includes the use of the cursors,
advanced trigger functions and using the second (delayed) timebase for signal
details.
GETTING STARTED3 - 15
3.10 PRE-TRIGGER VIEW
One of the powerful features in the digital mode is the ability to capture and view
signal contents prior to the actual trigger. The amount of pretrigger information
can be as long as one full acquisition/record. The trigger position is ad justed with
the
TRIGGER POSITION
control.
Step 1Turn the
triggering edge shifts to the center of the screen. A trigger point marker
(s) indicates the trigger point. The part to the left o f the marker is called
pretrigger view. The pretrigger view is indicated in the bottom of the
screen (in divisions)
1
Step 2The
TRIGGER POSITION
delay. Rotate the
readout displays "0". When you continue to turn the TRIGGER
POSITION control clockwise, a positive delay between trigger point
and acquisition is set. The delay is no longer read out in divisions of
pretrigger information, but in seconds, or fractions of seconds to
indicate how much delay is used.
TRIGGER POSITION
TRIGGER
POSITION
control can also be used to adjust time
TRIGGER POSITION
control counter clockwise. Now the
1
DELAY=−5.00dV
PRE-TRIGGER
VIEW
ST6686
control clockwise until the
3 - 16GETTING STARTED
3.11 MORE ADVANCED FEATURES
All basic functions are accessed by dedicated keys for fast and easy operation.
Some of the more advanced features are menu operated. Menu s are called up by
pressing one of the keys identified with
press one of these keys, a menu is displayed on the right side of th e screen. This
menu gives you access to the more advanced functions of the oscilloscope. Use
the blue softkeys to the right of the screen to select the desired functions; the
selected function is indicated by the highlighted text.
3
2
1
4
SOFTKEYS
Figure 3.8Menu keys and softkeys
text on the front panel. After you
blue
TRIGGER WITH
MENUKEY
TRIGGER
ST6433
9303
Step 1Press the key marked
TRIGGER
.
Check that the ’TRIGGER MAINTB’ menu is displayed at the right side
of the screen.
After changing the setting, you can deactivate the menu again to use
the full screen for the signal.
There are two ways to do this:
- Press the
- Press the
TRIGGER
TEXT OFF
key once again.
key.
The TEXT OFF key operates in a 1-2-3 cycle, and allows you to blank
the bottom text line as well.
TRIGGER
MAIN TB
edge tv
logic
ch1
line
TEXT OFF
CH1
MTB 200µs
200mV
CH1
CH1
200mV
MTB 200µs
TEXT OFF
TEXT OFF
CH1
Step 2Use both methods to familiarize yourself with turning the menus and
the bottom text line on and off.
ST6679
GETTING STARTED3 - 17
3.12 CURSOR OPERATION
Cursors are used for accurate amplitude or time measurements of the signal.
1
23
4
TRACK
CURSORS
∆
ST6431
9303
Figure 3.9Cursor setup
Step 1Before you continue, reset the instrument with the STANDARD
SETUP. To do this, press the
STATUS
key and
TEXT OFF
key
simultaneously. Now the instrument is set in the default condition and
operates in analog mode.
Step 2Press the
Step 3Press the
AUTOSET
CURSORS
key.
key to enter the cursors menu.
CURSORS
The menu is now displayed on the screen and the
cursors are turned on.
on off
Step 4Use the second blue softkey from the top to select
one of the three cursor modes:
- Amplitude cursor measurements, indicated by ’=’.
- Time cursor measurements, indicated by ’||’ for
measuring time or frequency.
- Amplitude
’#’. The top text line displays the result of the
time measurements, indicated by
and
READ OUT
measurements (∆V or ∆T).
Step 5Press the second bluesoft key until ’||’ is highlighted.
#
ST6430
Step 6The
TRACK
control moves both cursors at the same time. For
example, to measure the period time of the input signal, set the left
(reference) cursor to a rising edge of the signal.
Step 7 The ∆ contro l moves the right cursor only . Set this cursor the next rising
edge of the signal.
3 - 18GETTING STARTED
Step 8The top text line now shows the pulse repetition time of the signal
(e.g., ch1: ∆T= 500 µs).
200µs
CURSORS
on off
READ OUT
CH1
#
ST6687
ch1:
∆T= 500µs
CH1 200mV
MTB
Step 9Press the second blue softkey until ’=’ is highlighted. Now perform a
peak-to-peak measurement and check that th e amplitude of the signal
(’∆V’) is 600 mV.
NOTE:When you select ’#’, the fifth blue softkey is automatically activated so
that you can choose between using the controls for positioning the
vertical cursors (’||’) or the horizontal cursors (’=’).
The ’READOUT’ submenu is explained in Chapter 4.
Step 10Select the vertical cursors again.
Step 1 1Now switch to the digital mode. Notice the changing readout and the
’X’ indicating where the trace and the cursors intersec t. Since the trace
is digitized, the cursors can be really smart. In the digital mode you can
measure time differences (∆T) and amplitude differences (∆V) at the
same time.
200µs
CURSORS
on off
ch1
-
READ OUT
CH1
ST6688
#
9303
ch1:
∆T= 500µs
CH1 200mV
∆V= 600mV
MTB
Step 12Use the first blue softkey to turn the cursors off. The cursor menu
disappears.
GETTING STARTED3 - 19
3.13 MORE ADVANCED TRIGGER FUNCTIONS
Most of the trigger functions (source, slope, and level) can be controlled with
direct access to the functions (see Section 3.9). A CRT menu is used for more
advanced trigger functions.
TRIGGER
1
23
4
TB MODE
ARM’D
TRIGGER
LEVEL
ST6432
9303
Figure 3.10More advanced trigger setup
Press the menu key
TRIGGER
. This
turns the menu on. An extensive set of
functions is now displayed.
All functions are explained in Chapter 4. For most applications, this menu is not
needed.
3 - 20GETTING STARTED
3.14 MORE SIGNAL DETAIL WITH THE DELAYED
TIMEBASE
When you need to study a part of a signal in more detail, a second (delayed)
timebase is available. This timebase has its own timebase settings and trigger
level adjustment. Additional selections are made in the DELAYED TIMEBASE
menu.
DISPLAY
MAGNIFY
23
1
4
Figure 3.11Delayed timebase setup
Step 1Press the
STATUS
key and
STANDARD SETUP.
Then shift to trace to the upper half of the screen as indicated.
50mV
CH1
MTB 1.00ms
DELAYED TIMEBASE
DTB
sns
TIME/DIV
TEXT OFF
ch1
ST6690
TRIGGER
DELAY
POSITION
TB MODE
key at the same time for
ST6439
9303
GETTING STARTED3 - 21
Step 2Press the DTB key . The DELA YED TIME BASE menu is now displayed
on screen. Turn the delayed time base on with the first softkey.
DELAYED
TIME BASE
DEL’D TB
on off
MAIN TB
on off
starts
trig’d
TRACE
T
SEP
CH1
50mV
MTB1.00ms
DTB 100µs4.882ms
ch1
ST6689
The upper trace is the main timebase trace. This first trace shows an intensified
part. Adjust the TRACE INTENSITY with the control as necessary. The lower
trace is the delayed timebase trace and is an expanded representation of the
intensified part in the upper trace.
Step 3Turn the DELAY knob to shift the intensified part and to select which
part of the main timebase you want to magnify.
Step 4The delayed timebase TIME/DIV keys are used to select the
’magnification factor’. Notice the changing delayed timebase ’TIME/
DIV’ readout at the bottom of the screen.
Step 5The ’T’ symbol at the fourth blue softkey indicates that the cursor
TRACK control can be used to make adjustments. In this menu the
cursor TRACK control is used to change the TRACE SEParation, which
is the distance between the main timebase and the delayed timebase.
The delayed timebase can be used in the triggered mode. The trigger ed mode is
selected with the STARTS/TRIG’D softkey . The function of the triggered mode will
be explained in Chapter 4. For this part of "Getting Started", remain in the
STARTS mode.
Step 6Switch the menu off with the TEXT OFF key. Notice that the delayed
timebase is still active and that the most important functions (DELAY
control and TIME/DIV key pair) still allow you to operate the delayed
timebase.
3 - 22GETTING STARTED
3.15 TRACE STORAGE
In the digital mode you not only have the ability to store traces on the screen
(using the
RUN/STOP
key), but also to store traces in memory for later use.
ANALOG
3
2
1
4
SAVE
RECALL
RUN/STOP
ST6691
9312
Figure 3.12Digital memory setup
Store traces on screen:
Step 1Press
AUTOSET
.
Step 2Make sure that the scope is in the digital mode. If not, press the
ANALOG
Step 3With the
key to enter the digital mode.
RUN/STOP
key , new acquisitions are stopped and the display
is frozen. Removing the input signal or pressing a key has no ef fect on
the display . Stopping the acquisition is very useful to do me asurements
on the signal or to make a hard copy.
Step 4Press the
RUN/STOP
key to reactivate the acquisition.
Store traces in memory:
Step 5For this step-by-step introduction you will first clear all memory
locations so that all unnecessary traces are removed.
- press the
- select
- select
CLEAR & PROTECT
CLEAR ALL
- in the confirm menu, select
Sometimes there will be a second confirm menu. Select
SAVE
menu key.
.
.
YES
.
again to
YES
clear protected traces as well.
Press the
TEXT OFF
key to turn the menu off.
GETTING STARTED3 - 23
Here is how traces are stored in memory:
Step 6Use the TRACK control to select an empty memo ry
location such as m1, m2, or m3. Empty locations are
marked with a circle in front of the memory location
number (e.g., m3).
Step 7Press the second blue softkey (’save’). You have
now saved the acquisition signal into memory
’register’ m3. A single register can contain a set of up
to three traces (e.g., CH1, CH2, and EXT (trigger
view)). In this case only one input channel was
SAVE ACQ
TO
MEMORY
m2
m3
m4
save
clear
copy
turned on, so that only one was stored.
Step 8Remove the probe from CH1. Now recall the stored
trace.
Step 9Press the RECALL key.
CLEAR &
PROTECT
ST6705
Step 10Select the previously filled memory register m3 with the TRACK
control. A memory register with trace information is indicated with
Step 11Press the second blue softkey to turn on the display of this register.
Indicated by
Step 12Turn the ∆ control to separate the acquisition (live signal) and the trace
recalled from memory.
NOTE:You are now able to operate nearly all the oscilloscope’s functions in
most routine applications. Please continue with Chapter 4 for a more
detailed description of the oscilloscope’s many advanced features.
Memory indications:
Empty register
Filled register
Displayed register
HOW TO USE THE INSTRUMENT4 - 1
4HOW TO USE MORE ADVANCED
FUNCTIONS OF THE INSTRUMENT
This chapter allows more experienced oscilloscope users to learn more about the
advanced features of this instrument and how to use them. For a complete
description of each function, refer to the next chapter in this manual: "Function
Reference".
This chapter explains the basics of each function and gives examples are given
in a step by step sequence.
Less experienced oscilloscope users should read Chapter 3 before beginning this
chapter.
4.1INTRODUCTION
All of the oscilloscope models in the PM337xB, PM338xB and PM339xB family
combine the features and operation of an analog oscilloscope with that of a fullfeatured Digital Storage Oscilloscope (DSO). Switching between one mode of
operation to the other is done by pressing one (yellow) push button.
Most signal acquisition functions are identical for both modes of operation, even
though the digital mode allows for pretrigger acquisition and display, plus more
powerful logic triggering.
Delayed sweep operation is available in both operating modes.
Cursors operate in both modes.
The digital mode provides access to many powerful calcul ated measurement and
signal analysis functions.
It also has a completely new feature called AUTO RANGE.
AUTO RANGE automatically adjusts the attenuator or the timebase setting when
the amplitude or the frequency of the signal has changed.
This family of oscilloscopes is available with a 60 MHz, 100 MHz or 200 MHz
bandwidth. There are full four-channel instruments as well as 2+2 channel
models.
The economy version have 2 full channels and an external trigger channel.
PM3394B 200 MHz Full Four Channel Oscilloscope.
The PM3394B offers a 200 MHz bandwidth. Four channels provide equal
bandwidth, and sensitivity ranges. Each channel has full AC/DC/GND coupling
capabilities.
In VERTMENU each channel is set to 50Ω or 1 MΩ input impedance.
4 - 2HOW TO USE THE INSTRUMENT
PM3390B 200 MHz 2 Channel Oscilloscope
The PM3390B has the same capabilities as the PM3394B on the channels 1 and
2. The channels 3 and 4 are replaced by an externa l trigger channel. This channel
can only be used as an additional trigger input channel. Signal manipulation as in
the full channels 1 and 2 is not possible. The external trigger signal can be
displayed by using the function TRIG VIEW.
PM3384B 100 MHz Full Four-Channel Oscilloscope
Bandwidth is 100 MHz.
The PM3384B has the same capabilities as the PM3394B. VERT MENU only
offers BW LIMIT selection.
PM3380B 100 MHz 2 Channel Oscilloscope
The PM3380B has the same capabilities as the PM3384B on the channels 1 and
2. The channels 3 and 4 are replaced by an external trigger channel as in the
PM3390B. VERT Menu offers only BWLIMIT selection.
PM3370B 60 MHz 2 Channel Oscilloscope
The PM3370B has a bandwidth of 60 MHz. All other features are equal to those
of the PM3380B.
HOW TO USE THE INSTRUMENT4 - 3
FRONT PANEL LAYOUT
The controls on the front panel are grouped by function. In this chapter, a
description for each group of controls is given in the following sequence:
-Display and probe adjustment(see section 4.2)
-Analog and digital modes(see section 4.3)
-Vertical deflection( see section 4. 4)
-Horizontal deflection and triggering(see section 4.5)
-Digital acquisition and storage(see section 4.6)
-Advanced vertical functions(see section 4.7)
-Advanced horizontal and trigger functions(see section 4.8)
-Memory functions(see section 4.9)
-Cursor functions(see section 4.10)
-Measurement functions(see section 4.11)
-Processing functions(see section 4.12)
-Display functions(see section 4.13)
-Delayed timebase(see section 4.14)
-Hard copy facilities(see section 4.15)
-AUTOSET and other utilities(see section 4.16)
4 - 4HOW TO USE THE INSTRUMENT
Study the front panel of your oscilloscope and observe what functions the different
controls and push buttons (keys) perform. There are three differ ent styles of push
buttons, plus the blue softkeys adjacent to the screen. The push button functions
are as follows:
Direct function key. These keys provide direct access to specific
functions as labeled on the front panel. Examples include
AUTOSET for automatic setup of the oscilloscope and AC DC
(GND) for selection of the instrument’s input coupling.
Menu initialization key, with blue text. Press to produce a menu on
the screen from which you can select functions that are related to
the function name of this key.
Key pair. These pair s serve as up/down controls. They a re used to
select a value from a range.
Softkey. Press to select a function from the menu that has been
initialized by pressing one of the menu initialization keys.
Rotary control. Used for continuously variable contr ol of a function.
HOW TO USE THE INSTRUMENT4 - 5
4.2DISPLAY AND PROBE ADJUSTMENTS
T o help you follow the step-by-step descriptions, each section begi ns by recalling
the standard setting as outlined below.
If you get "lost", you can return to the beginning of each section, because all
functions are set to a predefined state to create a correct start situation.
RECALL STANDARD SETTING
-Simultaneously press the STA TUS and TEXT OFF keys.
Recalling the standard setting always results in a trace on the display, even with
no signal applied to the oscilloscope inputs. Refer for a list of the standard setting
to the section "STA NDARD SETUP/FRONT PANEL RESET" of Chapter 5.
DISPLAY ADJUSTMENT
Before going through the examples that introduce you to
features of this oscilloscope, adjust the display as follows:
-Turn the TRACE INTENSITY control for optimum trace
brightness.
-Turn the TEXT INTENSITY control for optimum
display brightness of the text.
-Adjust trace alignment with the graticule by using the
screwdriver control TRACE ROTATION.
-Turn the FOCUS control until a sharp trace is
obtained.
-You can turn the GRATICULE ILLUMINATION control
to illuminate the graticule lines as desired.
4 - 6HOW TO USE THE INSTRUMENT
DISPLAY LAYOUT
The following illustration shows the layout of th e display with a maximum amou nt
of text.
Most text is active only when specific functions are activated.
HOW TO USE THE INSTRUMENT4 - 7
MENUSTEXT OFF
A menu appears when a key with blue text is pressed.
The menu functions can be selected by pressing the blue softkeys to the right side
of the screen.
-Press the ANALOG key to select the digital mode.
-Press the DISPLAY key to activate the menu.
DISPLAY
Each menu starts with a menu name, which
corresponds with the key that was used to select the
menu. This is sometimes followed by a second name of
the softkey that initialized a submenu.
WINDOWS
on off
VERT
MAGNIFY
off
T
XvsY
The windows function can be switched on and off by
pressing the corresponding blue softkey.
T
indicates that a selection must be made with the
TRACK control.
The arrow ( ) behind TEXT indicates that there is a
TEXT
submenu.
dots
linear
sine
ST7415
9312
The bottom softkey switches (toggles) between the
three functions.
Y ou can turn of f the text by pressing the TEXT OFF key. This can be done to free
up the display area.
-Press the TEXT OFF key three times.
Observe that the text mode follows the following sequence: menu off, settings of f,
both on.
4 - 8HOW TO USE THE INSTRUMENT
CAL SIGNAL AND PROBE ADJUSTMENT
Each measuring probe has been checked and adjusted before delivery. However,
to match the probe to your oscilloscope, you must perform the following
procedure to optimize the pulse response of the combination of oscilloscope input
and probes.
-Connect the probe body to channel 1.
-Connect the probe tip to the Probe Adjust output of the oscilloscope.
-Press the AUTOSET key.
If the display looks like one of the two displays shown on the left, you must adjust
your probe to get the display shown on the right.
The probe output impedance can be adjusted through a hole in the compensation
box of the probe to obtain optimum pulse response. Refer to the following figure.
ST6024
9303
-Adjust the probe until the screen shows the correct compensation.
The probe is now adjusted for optimum pulse response with this oscilloscope. If
you connect the probe to another channel or oscilloscope, it must be adjusted
again to that oscilloscope input.
-Repeat this adjustment for the second probe.
HOW TO USE THE INSTRUMENT4 - 9
4.3ANALOG AND DIGITAL MODES
ANALOG
This instrument is a combination of an analog real-time oscilloscope and
a digital storage oscilloscope, which offers a variety of additional features.
The combination of analog and digital modes in one instrument gives
you the advantages of both modes.
In the
ANALOG MODE
the signal is directly written on the screen. The result is
the "traditional" real-time signal representation. Because of the high update rate
and infinite resolution, this image gives signal details that are vi sible on ly on tr ue
analog oscilloscopes.
In the
DIGITAL MODE
the input signal is sampled. These samples are stored in
memory so that mathematics, calculated measurements, printing, and other
memory functions can be performed on the trace.
You can use the yellow ANALOG push button to switch from the analog mode to
the digital mode and back at any time. The signal acquisition and displa y functions
of both operating modes are very similar. However, the nature of the signals you
are using may determine which operating mode you prefer to use. For more
information, refer to the following table:
SIGNAL CRITERIAANALOG MODEDIGITAL MODE
Repetitive signals of UsableUsable
30 Hz and higher
Repetitive signals Causes displayPreferred
below 30 Hzflickering
Single eventsDisplay forCan capture and
duration ofdisplay for long
the eventterm
Repetitive signals that PreferredMay cause aliasing.
are amplitude modulatedUse Peak detect or
Envelope mode
Repetitive signals thatPreferredMay cause aliasing.
are modulated in frequencyUse Envelope mode.
Long serial data streamsPreferred whenWhen using delayed
Delayed sweepsweep to observe
is not used.details, Digital mode
provides better
light output.
4 - 10HOW TO USE THE INSTRUMENT
SIGNAL CRITERIAANALOG MODEDIGITAL MODE
Video signalsPreferred whenWhen using delayed
Delayed sweepsweep to observe
is not used.details, Digital mode
provides better light
output.
OTHER CRITERIA
Need to see pretriggerNot possibleUp to one screen
information
You need to make adjustmentsFastestSlower display
to the circuitry and watchdisplay updateupdate
the signal change
Automatic measurementsCan’t useFully implemented
Signal MathAdd, Subtract onlyAll functions
Add, Subtract, Multiply
Signal AnalysisNot availableFull analysis, with
Integration,Math Plus option
Differentiation, FFTinstalled
Autorange timebaseNot AvailableResults in a signal
display of 2 to 6
waveform periods
HOW TO USE THE INSTRUMENT4 - 11
STANDARD SETTING
-Simultaneously press the STA TUS and TEXT OFF keys.
-Connect the Probe Adjust signal to channel 1.
The Probe Adjust signal, now supplied to the input, is a square wave with a lower
level of 0V and a top level of 600 mV . The osci lloscope always operates in analog
mode after a recall of the standard setting.
-Press the AUTOSET key for optimum signal display.
ANALOG TO DIGITAL MODE SWITCHING
Switching from analog to digital mode and vice-versa can be done by pressing the
yellow
ANALOG
switch between the modes with no changes in the display.
-Press the ANALOG key again.
The oscilloscope is automatically set in the digital mode. This is indicated by a
message ’DIGITAL MODE’ that appears briefly on the screen.
key. If no Digita l mode-only features were activated, you can
4 - 12HOW TO USE THE INSTRUMENT
TRACE STORAGERUN/STOP
The digital mode offers a set of powerful features. One of the most importa nt
advantages is that you can store one or more traces in memory or on screen.
-Press the RUN/STOP key to stop the signal acquisition and freeze the
display on the screen.
The trace is frozen and stays on the screen. Observe that the signal stays stored
even when you remove the probe.
Now the instrument has been STOPped, and most keys have been disabled. The
only keys that continue to function are those directly relate d to di spla y functio ns.
This includes trace shift using the
operable.
A frozen trace can be used for comparisons with other traces, mathematics,
cursor measurements, automatic measurements and more.
The
RUN/STOP
key is used to end the STOP mode and start the aquisition a gain.
-Press the RUN/STOP key.
This starts the acquisition again so that the actual input signal is displayed. You
can use the RUN/STOP key at any time.
control. All measurement functions are still
POS
RUN/STOP
RUNNING
FROZEN
ST6482
NOTE: The following section explains the basic functions regardless of
operating mode. The oscilloscope will react almost identically in either
mode. Where necessary, different behavior will be explained.
9312
HOW TO USE THE INSTRUMENT4 - 13
4.4VERTICAL DEFLECTION
1
AMPL
POS
RANGE
mV
VAR
V
AUTO
ON
VERT MENU
TRIG1
CH1+CH2
AC DC
GND
2
AMPL
POS
RANGE
mV
VAR
V
AUTO
ON
AVERAGE
TRIG2
INV
AC DC
GND
ST6437
9312
The section shown on the left contains all
direct vertical deflection controlls for the input
channels 1 and 2.
Refer to Section 4.1 for the differences
between model numbers.
To start this section with the settings in a
predefined state, you must recall the
standard setting.
STANDARD SETTING
-Simultaneously press the STA TUS and TEXT OFF keys.
-Connect the Probe Adjust signal to channel 1.
The Probe Adjust signal, now supplied to the input, is a square wave with a lower
level of 0V and a top level of 600 mV.
-Press the AUTOSET key.
4 - 14HOW TO USE THE INSTRUMENT
VERTICAL COUPLINGAC, DC, GND
The input coupling after AUTOSET is ac. Since the Probe Adjust signal is a pulse
type signal with a 50% duty cycle, its mean value is at the signal’s 50% amplitude
level. When the input is ac coupled, the mean value will be displayed at the
ground level of the oscilloscope. As a result of this, the displayed waveform is
centered on the screen.
AC coupling can be used to examine small ac components that are su perimposed
on large dc voltages.
-Press the AC DC GND key once for DC input coupling.
Since the Probe Adjust signal is a pulse-type signal with a lower level of 0V, and
a higher level of 600 mV, and since the oscilloscope adjusts the display position
of the ground level to coincide with the screen center , switchin g from ac coupling
to dc coupling results in an upward shift of the display position of the signal.
The coupling sign in the lower left hand corner of the scr een changes from ∼(ac)
to = (dc).
The ground level for each of the channels is indica ted by a dash after the ch annel
identifier, i.e., ’1-’ for the ground level of channel 1.
0.5ms
600mV
0V
AC DC
GND
1-
200mV
CH1
GND Coupled
ST6708
AUTO SET
AC DC
GND
Input Signal
AC DC
1-
200mV
CH1
AC Coupled
GND
1-
200mV
CH1
DC Coupled
-Press the AC DC GND key to obtain ground coupling.
A straight line is now displayed. This is the 0V (ground) level of the input. This
level serves as the 0 volt reference for amplitude measurements.
The coupling sign ’⊥’ now indicates ground coupling.
HOW TO USE THE INSTRUMENT4 - 15
VERTICALPOSITION
POS
Use the POS control to adjust the ground level to any desired vertical
position on the screen.
ST6158
9303
1
POS
1
MAT4171
-Use the position control to position the line in the middle of the screen.
Observe that the channel identifier ’1-’ shifts with the trace.
-Press the AC DC GND key again to obtain ac input coupling.
Since the Probe Adjust signal is a pulse type signal with a 50% duty cycle, its
mean value is at the 50% amplitude level of the input signal. When the input is ac
coupled, the mean value will be displayed at the ground level of the oscilloscope.
As a result of this, the displayed waveform is centered on the screen.
4 - 16HOW TO USE THE INSTRUMENT
VERTICALAMPLITUDE
-Press the upper key of the AMPL keys
AMPL
Pressing the upper key increases the amplitude of the displayed
waveform. As the amplitude of the displayed waveform changes, the
screen readout of the input sensitivity changes as well.
If you adjust the displayed amplitude to 6 divisions, you will notice that
the readout in the lower left hand corner of the screen reads 100 mV/div.
ST6159
The amplitude of the signal is then: 100 mV/div times 6 div = 600 mV.
AMPL
100mV
CH1
AMPL/divAMPL/div
500mV
CH1
MAT4165
Try other sensitivity settings as follows :
-Use the AMPL keys to step through the attenuator range.
Observe that the sensitivity readout changes in steps following a 1, 2, 5
sequence. The AMPL keys allow you to step up and down through the
sensitivities from 5 V/div to 2 mV/div and vice-versa. This sequence enables a
quick selection between the oscilloscope’s sensitivity positions and are such that
almost every input signal can be made visible with sufficient amplitude.
-
Adjust AMPL to 100 mV
.
A waveform with an amplitude of six divisions is displayed.
HOW TO USE THE INSTRUMENT4 - 17
VERTICALAUTO RANGE
AUTO
RANGE
The AUTO RANGE function results in an amplitude display of 2 to 6
divisions.
-Press the AUTO RANGE key.
Observe that the amplitude of the signal changes from 6 divisions to 3 divisions.
In the upper right corner of the display ATT 1 is displayed. This indicates that the
AUTO RANGE function is active on input channel 1.
-Press the upper AMPL key once.
The AUTO RANGE function is switched off and the ATT 1 indication disappears
The attenuator is now back in manual control.
VERTICALVARIABLE AMPLITUDE
AMPL
-Enter the VAR mode by simultaneously pressing both AMPL keys.
-Adjust amplitude with either AMPL key.
Press both AMPL buttons simultaneously to select the VARiable
mV
mode. This mode is used to make fine adjustments of the input
VAR
amplitude settings between the 1-2-5 steps.
When the V ARiable mode is turned of f, the oscilloscope selects the
V
nearest ’1-2-5’ value.
ST5971
Observe that the sensitivity steps that can be selected are much finer than before
and that the displayed amplitude is continuously adjustable.
Note:The VAR values are calibrated amplitude settings as well. This enables
you to make accurate measurements and readouts even when
intermediate settings are used.
-Adjust AMPL to 220 mV.
-
Press both AMPL keys to turn the VARiable mode off
.
4 - 18HOW TO USE THE INSTRUMENT
VERTICALCH1+CH2
-Using two probes, connect the Probe Adjust signal to Ch1 and Ch2.
-Press AUTOSET.
-Both Ch1 and Ch2 are now displayed.
-Adjust POS and AMPL to get a display as illustrated on the left below.
-Press the CH1+CH2 key.
In the analog mode, a third trace will appear on the screen. This trace has twice
the amplitude of the Probe Adjust signal. The position of the third trace is affe cted
by the position controls of both Ch1 and Ch2.
The display mode is indicated in the screen as ’1+2’.
1-
1-
CH1+CH2
2-
500mV
CH1
CH2
500mV
MTB
CHP
200µs
ch1
2-
500mV
CH1
1+2
MTB
CHP
500mV
CH2
200µs
ch1
ST5970
-Press the ANALOG key to switch to the Digital mode.
Observe that switching to the digital mode turns off the third trace. The ’1+2’
indication disappears.
-Press the CH1+CH2 key once again.
Observe that in the digital mode the (CH1+CH2) trace can be used as well.
Simultaneous display of both channels together with the sum of both chann els is
not possible in the digital mode.
1-
CH1+CH2
1+2
500mV
CH1
CH2
500mV
MTB
200µs
ch1
2-
500mV
CH1
CH2
500mV
MTB
ALT
200µs
ch1
ST6720
-Press the ANALOG key to return to the analog mode.
HOW TO USE THE INSTRUMENT4 - 19
VERTICALINVERT
The INVERT function in Channel 2 can be used to make it easier to do out-ofphase signal comparisons. The most common use of the INVERT function is to
obtain the display or make the acquisition of the voltage difference between two
channels. This is done by displaying or capturing the sum of Ch1 and Ch2 as
follows. (This is referred to as the differential mode.)
-Using two probes, connect the Probe Adjust signal to Ch1 and Ch2.
-Press AUTOSET.
-Both Ch1 and Ch2 are now displayed.
-Adjust POS and AMPL to get a display as illustrated on the left below.
-Press the CH1+CH2 key.
-Press the INV key of channel 2.
Since channel 2 is inverted before it is added to channel 1, the result will be CH1CH2. This is indicated as ’1-2’ on the screen. The signals on both inputs are the
same, resulting in a straight line at ground level.
(If the line is not straight, this may be an indication that the pr obes are not correctly
adjusted. When one of the probes is not properly adjusted, the input signals at the
input connectors of the oscilloscope will be unequal. The difference of the two
input signals shows up in the differential mode).
CH1+CH2
1-
INV
2-
200mV
CH1
1+2
MTB
CHP
200mV
CH2
200µs
ch1
CH1
CH2↓
1-
2-
200mV
1-2
MTB
200µs
200mV
CHP
ch1
ST5969
4 - 20HOW TO USE THE INSTRUMENT
VERTICAL MENUBANDWIDTH LIMITER
The Bandwidth Limiter reduces the bandwidth of the vertical channels to 20 MHz.
This is done by activating a filter in the vertical channels. This fea ture can be used
both in analog mode and in digital mode to suppress high frequency noise. For
repetitive signals and when in the digital mode,
method to reduce noise without limiting the bandwidth.
-Press the ’BW LIMIT’ softkey to turn it on.
Observe that the displayed line becomes "thinner" as an indication that the
amount of noise is reduced.
The text ’BWL’ appear s in the bottom of the screen to indica te that the functio n is
active.
-Press the ’BW LIMIT’ softkey to turn it off again.
averaging
is the preferred
HOW TO USE THE INSTRUMENT4 - 21
VERTICAL MENU50
The 200 MHz models offer the selection of an i nput impedance of 50Ω on all input
channels. The EXTTRIG input has no 50Ω possibility. The impedance selection
is a subfunction in VERT MENU.
The 50Ω input impedance is used to ob tain a corre ct im pedance to match signal
sources of the same impedance. For interconnection, a coaxial 50Ω cable must
be used. The 50Ω position is indicated on the display with the Lz sign (Low Z =
Low Impedance).
-Disconnect all signals from the inputs.
-Using a 50Ω coaxial cable, connect a 4 Vpp, 2 kHz signal from a
generator with 50Ω output impedance to channel 1.
-Press the AUTOSET key.
-Press VERT MENU.
-Press the 50Ω CH1 softkey.
VERT MENU
50Ω CH.
ON
OFF
1V
CH1
CH1
Observe that the displayed signal amplitude changes to half of the original
amplitude. This is caused by the change of the input impedance to 50Ω, which
results in a voltage division between the 50Ω output of the generator and the 50Ω
input of the oscilloscope.
L
1V
Z
ST5968
9312
Ω
-Remove the input signal from CH1.
VERTICALPROBE RANGE INDICATOR
-Connect the probe again.
CHx
detector
multiply the displayed amplitude by 10 or 100 when you use a Fluke probe with
range indication.
Since the 10:1 probe delivered with the instrument is a high
impedance probe, the 50Ω termination is automatically turned off.
At the same time, the sensitivity readout is adjusted automatically
ST6021
when a 10:1 or 100:1 probe is used. This way you don’t have to
4 - 22HOW TO USE THE INSTRUMENT
4.5HORIZONTAL DEFLECTION AND TRIGGERING
Before starting with the horizontal deflection functions, you must set the
instrument to a predefined state to create a correct start situation.
STANDARD SETTING
-Simultaneously press the STATUS and TEXT OFF keys.
-Connect the Probe Adjust signal to channel 1.
The Probe Adjust signal, now supplied to the input, is a square wave with a lower
level of 0V and a top level of 600 mV.
-Press AUTOSET.
TIMEBASETIME/DIV
TIME/DIV
4 periods/cycles of the square wave are displayed on the screen.
ST6839
-Press the ’s’ (left) key of the MainTB TIME/DIV key pair a few times.
The more you press the left (’s’) key, the slower the timebase will run. This results
in the number of displayed periods/cycles of the input signal to increase.
-Press the ’ns’ (right) key of the MainTB TIME/DIV key pair a few times.
The number of displayed periods/cycles decreases as the timebase speed
increases.
Observe that the timebase speeds are adjustable in steps following a 1-2-5
sequence.
TIME/DIV
MTB 100µsMTB 500µs
ST6709
-
Set the timebase to 200 µs
.
The signal is displayed with four periods on the screen.
HOW TO USE THE INSTRUMENT4 - 23
TIMEBASEAUTO RANGE
AUTO
RANGE
The AUTO RANGE function continuously adjusts the timebase to a
display of 2 to 6 waveform periods.
-Press the AUTO RANGE key.
Observe that the timebase of the signal display changes to display 2 to 6 signal
periods. AUTO TB in the upper right corner of the display indicates that the AUT O
RANGE timebase function is active.
-Press one of the TIME/DIV keys.
The AUTO RANGE function is switched off and the AUTO TB indication
disappears.
TIMEBASEMAIN TB VAR
TIME/DIV
-Activate the VAR function by simultaneously pressing the TIME/DIV keys.
-Press one key of the MainTB VAR keys.
Observe that the timebase indication is now changing continuously instead of
following the 1-2-5 step sequence.
The V AR timebase readout values are calibrated values. This enables accurate
timing measurements using VAR timebase settings.
Fine adjustment of timebase speeds between the 1-2-5 steps
can be made by simultaneously pressing the MainTB TIME/DIV
keys.
ST6710
-
Adjust MainTB VAR to 850 µs/div.
-Turn off the VAR function by pressing the MainTB TIME/DIV key pair
simultaneously.
Observe that the timebase is set to the nearest step value (1 ms/div).
In the Digital mode, the timebase speeds are determined by an XTal oscillator.
4 - 24HOW TO USE THE INSTRUMENT
TIMEBASEMAGNIFY
MAGNIFY
The displayed signal can be expanded horizontally so that more
signal detail becomes visible. In the analog mode, a
magnification is possible and the entire sweep length can be made
ST6711
visible by turning the X POS control.
10
*
-Press the right MAGNIFY key.
The text ’MGN’ appears in the bottom of the scree n to indicate that the function i s
active. Observe that the timebase indication is changed from 1.00 ms/div to
100 µs/div.
MAGNIFY
CH1
ch1
200mV
CH1
100µs
MTB
200mV
MTB
100µs
MGN
ch1
ST6504
-Press the ANALOG key to set the scope in the digital mode.
-Press the right MAGNIFY key.
Observe that the signal expands in
This magnification factor is displayed briefly on the screen.
-Set the MAGNIFY to
4.
*
1, *2, *4 ... steps to a maximum of 32 times.
*
HOW TO USE THE INSTRUMENT4 - 25
XPOS
TIMEBASEX POS
With X POS the displayed signal is shifted horizontally across the
display.
-Turn the X POS control clockwise.
1
XPOS
1
MAT4199
Observe that a bar graph () is displayed.The block on the bar graph
shows which part of the digital trace is displayed as expa nded. This bl ock on the
bar graph moves as you use the XPOS control.
-Press the RUN/STOP key.
Observe that the MAGNIFY key and the X POS control can still b e used after the
acquisition has STOPped. This allows you to look at signal details eve n after the
signal has been acquired.
-Press the RUN/STOP key.
TRIGGERING
-Simultaneously press the STA TUS and TEXT OFF keys.
-Press AUTOSET.
4 - 26HOW TO USE THE INSTRUMENT
TRIGGERSOURCE
-Press the TRIG 2 key in the CH2 section of the front panel to select
channel 2 as trigger source.
The indication in the lower right hand readout area of the screen now displays ’ch2’.
Observe that the signal is not triggered. The ARM’D LED is on.
-Press the TRIG 1 key in the CH1 section of the front panel to select
channel 1 as trigger source.
Observe that the indication has now changed into ’ch1’ and the signal is triggered
again.
TRIG2
CH1
ch2
TRIG1
CH1
ch1
ST5955
TRIGGERSLOPE
TRIG1
The trigger source selection key is also used to select the trigger slope
between the positive-and negative-going edge of the triggering signal.
-Press the TRIG 1 key in the CH1 section of the front panel a second time.
Observe that the signal now starts with the trailing edge instead of a leading edge.
Also the trigger slope symbol in the lower right hand corner of the readout area
indicates a trailing edge.
TRIG1
ch1
ch1
ST6025
HOW TO USE THE INSTRUMENT4 - 27
7
TRIGGERTRIGGER LEVEL
TRIGGER
LEVEL
After each AUTOSET, the trigger level is always clamped within the
signal amplitude range to assure stable triggering on most signa ls. The
trigger level is adjustable, but it is limited between the minimum
(-100%) and the maximum (+100%) amplitude levels of the signal.
-Turn the TRIGGER LEVEL control.
As you turn the trigger level control, the
trigger level readout in the bottom of the
screen shows the trigger level relative
to the signal amplitude. For example, it
may appear as ’Level=+ 14%’.
TRIGGER
1-
LEVEL=+17%
LEVEL
ST654
4 - 28HOW TO USE THE INSTRUMENT
TRIGGER
POSITION
TRIGGERTRIGGER POSITION
-Press the ANALOG key to select the digital mode.
-Turn the TRIGGER POSITION control counterclockwise.
One of the outstanding capabilities of a Digital Storage Oscilloscope is the ability
to capture and display signal details before the trigger moment.
When you turn the TRIGGER POSITION control counterclockwise, the trigger
point ( ) is shifted to the right. This allows you to look at the signal before the
trigger point. The pretrigger view is displayed in the bottom of the screen, and
calibrated in divisions. The trigger point is indicated on the screen with an arrow.
Pretrigger recording can be adjusted and is limited to one full screen, or -10
divisions.
-Turn the TRIGGER POSITION control clockwise.
When the control is rotated clockwise, delay is added. The trigger point shifts to
the left. The delay can be adjusted to a maximum of1000 divisions, so that it is
outside the screen. Positive trigger delay is indicated in seco nds or in fractions of
seconds.
-Set the delay to 0 with TRIGGER POSITION
HOW TO USE THE INSTRUMENT4 - 29
TRIGGERSINGLE SHOT
-Make sure the instrument is in the Digital Mode.
-Press AUTOSET.
This sets up the scope with the proper amplitude and timebase settings.
-Remove the probe tip from the Probe Adjust output.
-Turn the TRIGGER POSITION to mid-screen (counterclockwise).
-Press the SINGLE key.
This
automatically
prepares the scope for a single acquisition.
Observe that the screen has been cleared, and that the ARM’D LED is on. This
indicates that the scope is armed and waits for a trigger. The trigger level is
automatically set to .5 divisions (indicated by ’T-’).
-Look at the screen and touch the Probe Adjust output with the probe tip.
Observe that the oscilloscope ’runs’ once and that a signal appea rs on the screen.
The ARM’D LED is turned off. At the moment you touched the Probe Adjust output
with the probe, the scope was triggered. The trigger moment is displayed in the
center of the screen. The left side of the screen displays the part of the signal
before triggering (pretrigger information).
T−
Probe Adjust
T−
ST6505
9303
In SINGLE SHOT also ’Dual Slope’ triggering is possible. See chapter 5
TRIGGER MAIN TB function.
4 - 30HOW TO USE THE INSTRUMENT
4.6DIGITAL ACQUISITION AND STORAGE
This section gives you a short introduction to digital acquisition and storag e in
order to provide the basic knowledge and terms. This informa tion is necessary for
you to understand all digital statements in the following sections of the manual.
ANALOG:
DIGITAL:
In the
In the
INPUT
STAGE
INPUT
STAGE
mode the input signals are directly displayed on the screen.
analog
mode the channels are applied to the input stage of the digital circuit:
digital
signal
ADC
data
ACQUISITION
MEMORY
REGISTER
MEMORY
DISPLAY
DISPLAY
ST6721
the analog-to-digital convertors (ADC’s). The AD C’s co nvert the analo g signal(s)
into digital data. Depending on the selected timebase speed, this sa2pling and
conversion is done at a rate of up to 200 million samp les per second ( 200MS/s).
This high sampling rate allows you to observe fast signal variations.
Digital information from the adc is then stored in a Acquisition Me mory. The data
acquisition in this memory can be stopped to freeze the trace on the screen. You
can also store (save) complete registers or separate si gnals into a "background "
memory, called Register Memory. Traces can be stored as long as you like and
can be recalled at any time.
The CombiScopes in this range have a standard memory of 8K. The record length
for each trace is the maximum number of samples divided by the number of
traces. With the standard scope, the memory depth of 8K is available for two
channels. The memory is partitioned so that 1x8K, 2x4K or 4x2K records can be
captured. For maximum update rate, records can be made as short as 512 poin ts.
If the memory expansion option is installed, the record length can be adjusted
between 32K and 512 points.
HOW TO USE THE INSTRUMENT4 - 31
4.7ADVANCED VERTICAL FUNCTIONS
All basic functions of the oscilloscope are accessible via direct action front panel
keys.
More advanced functions are easily accessible via the menus behind the menu
initialization keys (keys with their function name in blue text).
STANDARD SETTING
Before continuing with the advanced vertical functions, you must set the
instrument to a predefined state to ensure a correct start situation.
-Simultaneously press the STA TUS and TEXT OFF keys.
-Connect the Probe Adjust signal to channel 1.
The Probe Adjust signal, now supplied to the input, is a square wave with a lower
level of 0V and a top level of 600 mV.
-Press AUTOSET.
4 - 32HOW TO USE THE INSTRUMENT
ACQUIREAVERAGE
The
average
function averages the input data over a number of successive
acquisitions. The average function is used to reduce the influence of random
noise in the input signal. There is no loss of bandwidth when the averag e function
is activated, but the signal must be repetitive. The number of sample s over which
the average is calculated can be selected by the user.
-Use the probe to connect Ch1 to the Probe Adjust output.
-Press AUTOSET.
-Connect the second probe to channel 2, but do not connect it to a signal!
-Press the CH1+CH2 key to add Ch1 and Ch2 together.
-Press the mV key of the CH2 AMPL keys so that the noise picke d up by
the probe tip of the second probe appears on the CH1+CH2 trace.
-Press the AVERAGE key.
For maximum ease of use, the scope offers direct access to the important
average feature.
The text ’Average = 8’ appears in the bottom of the screen to indicate that the
function is active. The default value is 8. The indication changes into ’AVG’.
Observe that the noise on the trace is reduced.
-To c hange the Average constant, press the ACQUIRE key.
-Turn the TRACK control clockwise.
The Average factor increases when the TRACK control is turned clockwise.
Observe that this reduces noise on the trace even more.
TRACK
Average=2
AVG
-
Turn the TRAC K control counterclockwise so that the Average factor is
8 again.
Average=256
AVG
ST6481
HOW TO USE THE INSTRUMENT4 - 33
on
ENVELOPE
ST6480
ENV
ACQUIREENVELOPE
The
ENVELOPE
mode records the minimum and the maximum of th e signal over
a number of acquisitions.
-Press the ACQUIRE key.
-Press the ’ENVELOPE’ softkey to turn it on.
The A VERAGE mode and the ENVELOPE modes are mutually exclusive.
Observe that AVERAGE is automatically switched off. The text ’ENV’ appears in
the bottom of the screen to indicate that the function is active.
The minimum and maximum of the input signal is stored and becom es visible on
the screen. This process continues until it is turned off by the user.
-Turn the POS control of Ch1.
Observe that every time you change the settings of the scope (like trace POS) the
ENVELOPE process automatically starts again.
-Press the ’ENVELOPE’ softkey to turn off the ENVELOPE mode.
ACQUIREPEAK DETECTION
PEAK DET ection automatically catches peak values (up to glitch es of 10 ns width)
of the input signal during a single acquisition.
-Press the ’PEAK DET’ softkey to turn it on.
The text ’PKD’ appears in the bottom of the screen to indicate that the function is
active. Observe that any glitch on the input signal becomes visi ble on the scree n.
Note:Regardless of the timebase speed selected, the scope is set to the
highest sample rate, so that glitches up to the sample fr equency can be
caught (5 ns at 200 MS/s).
-Press the ’PEAK DET’ to turn it off again.
4 - 34HOW TO USE THE INSTRUMENT
4.8ADVANCED HORIZONTAL AND TRIGGER
FUNCTIONS
All basic timebase and trigger functions of the oscilloscope are accessible via
direct action front panel keys. More advanced functions are easily accessible via
the menus behind the menu initialization keys (keys with the function in blue text).
STANDARD SETTING
Before continuing with the advanced functions, you must set the instrument to a
predefined state to ensure a correct start situation.
-Simultaneously press ST ATUS and TEXT OFF.
-Connect the Probe Adjust signal to channel 1.
-Press the AUTOSET key.
TRIGGERMODE
-Press the TRIGGER key.
On the screen the ’TRIGGER MAIN TB’ menu is displayed. The menu functions
can be selected by pressing the blue softkeys to the right of the screen. The top
softkey selects the trigger mode (’edge’ or ’tv’). The lower part of the menu is
optimized for the selected trigger modes.
-Press the first softkey to select ’tv’ trigger mode.
Observe that the lower part of the menu is optimized for tv applications.
-Press the first softkey to return to ’edge’.
Note :In the Digital mode, an additional selection is available. This is called
"logic" or "glitch". A full description of the Logic and Glitch triggering
functions can be found in Chapter 5.
HOW TO USE THE INSTRUMENT4 - 35
TRIGGERLEVEL
-Press the ’level-pp’ softkey to turn it off.
TRIGGER
LEVEL
The automatic level detection circuitry is turned off. The trigger level is
no longer clamped within the peak-peak range of the signal. You must
adjust the proper trigger level. While you turn the TRIGGER LEVEL
control, the actual trigger level is displayed in (m)V.
An indicator ’T-’ is displayed next to the channel identifi er when the combination
of trigger coupling and channel input coupling is useful (both ac coupled or both
dc coupled).
When the ’T-’ is within the signal range, a stable display of the signal is ob taine d.
When proper triggering takes place, the ARM’D LED is off.
Use the TRIGGER LEVEL control to move the indicator (T-) vertically.
TRIGGER
T-
LEVEL
T-
ST6738
When the trigger level indicator (T-) is outside the signal range, tr iggering is lo st.
Because of the nature of the (Probe Adjust) input signal, it appears that only two
lines are drawn. The timebase is not properly triggered, as is indicated by the
ARM’D LED being turned on.
Note:’T-’ is an indication for reference only . Its position can differ slightly from
the actual trigger level.
4 - 36HOW TO USE THE INSTRUMENT
TRIGGER MODETRIGGERED
-Press the TB MODE key.
TB MODE
-Press the softkey next to the ’auto trig single’ function to select ’trig’.
In the triggered mode, a signal is visible only when proper triggering occurs.
-Move the indicator (T-) in and out of the signal range.
In the screen, the timebase mode (TB MODE) menu appears. ’auto’ is
intensified. With the "auto" mode turned on, the timebase will run free as
soon as no triggering signal is detected, in order to provide a base line.
The ’Auto’ mode works well with most signal that have frequencies of
30 Hz or higher.
T−
ST6737
In the
T−
Analog
TRIGGER
LEVEL
mode, the result is a stable and triggered signal when the indicator
is inside the signal range, and no trace appears on the screen when the indicator
is outside the signal range.
Disabling the ’auto’ function by selecting ’trig’ can be useful for displ aying very low
frequency signals. The scope waits for a real trigger before the sweep starts even
if the signal frequency is very low.
In the
mode, the same criteria exist for the use of the ’auto’ function and
Digital
the ’trig’ function, but there is a difference. When the scope is in the ’trig’ mode
and triggering is lost, the acquisition will stop, but the display will reflect the last
information before the triggering stopped. This may lead you into thinking that the
scope is properly triggered because the display appears to be stable, while in fact,
you are displaying stored information only. Watch the ARM’D LED !
HOW TO USE THE INSTRUMENT4 - 37
TRIGGER MODEROLL
-Press the ANALOG key to select the digital mode.
-Press the TB Mode key to enter the TB mode menu
The TB MODE menu is extended with extensive timebase modes. The
differences are as follows:
•A Roll mode
•Selection of ’Real-time only’
•Delay by events
•Selection of Acquisition length
-Press the softkey next to the ’ROLL on off’ function to select ’on’.
The ’ROLL on’ function is now intensified.
Observe that the timebase is automatically set to 200 ms. The trace moves from
the right to the left.
-Press the ’STOP ON TRIGGER’ softkey to select ’yes’.
The trace stops when the trigger (in this case the start of the first period) reaches
the left of the screen.
Using the roll mode, you can monitor signals like temperature changes and
chemical processes at low timebase speeds. In this mode the scope operates l ike
a four-pen plotter. Over 36 hours of events can be recorded in memory and
plotted later.
Unlike a paper and pen plotter, the scope can record glitches as narrow as 5 ns,
and the record can be stopped when a trigger condition is met.
4 - 38HOW TO USE THE INSTRUMENT
TRIGGERLINE
-Connect a sine-wave signal of 4 V/300 Hz to channel 1.
-Simultaneously press the STATUS and TEXT OFF keys.
-Press AUTOSET.
-Select a timebase speed of 1 ms/div.
-Press the TRIGGER key.
-Press the softkey next to the function ’ch1 line’
When ’line’ is selected, this function is intensified. The line frequency is used as
the trigger source.
-Slowly vary the frequency of the sine-wave input signal.
Observe that when the input frequency is close to an integer multiple of the line
frequency , the run ning sine wave slows down or even stands still. Line trigger ing
can be used to display signals or signal components that are related to the line
frequency (e.g., hum or power supply ripple).
to select ’line
’.
HOW TO USE THE INSTRUMENT4 - 39
4.9MEMORY FUNCTIONS
The next section deals with storing and recalling traces in memory for later use.
Functions related to trace storage and recall are easily accessible via the m enus
selection keys labeled ’SAVE" and ’RECALL’.
STANDARD SETTING
Before continuing with the memory functions, you must first set the instrument to
the default setting to ensure a correct start situation.
-Simultaneously press the STA TUS and TEXT OFF keys.
-Connect the Probe Adjust signal to channel 1.
-Press AUTOSET.
The Probe Adjust signal, now supplied to the input, is a square wave with a lower
level of 0V and a top level of 600 mV.
TRACE STORAGECLEAR & PROTECT
SAVE
To be sure that all unnecessary stored traces are removed, you first will clear all
memory locations.
-Press the ’CLEAR & PROTECT’ softkey.
-Press the ’clear all’ softkey.
The scope now displays the confirmation submenu to prevent you from removi ng
traces you did
-Press the ’yes’ confirm softkey.
If any memory locations were protected (’PROTECT ON’ in CLEARS PROTECT
MEMORY submenu), a second confirmation must be given.
-Press the ’yes’ confirm softkey again.
-
Press the ANALOG key to select the digital mode.
-
Press the SA VE key to enter the ’SA VE ACQ TO MEMOR Y’ menu.
want to remove.
not
4 - 40HOW TO USE THE INSTRUMENT
TRACE STORAGESAVE
-Turn the TRACK control t o select the memory location in which to store
the trace data.
Observe that eight memory locations ’m1 to m8’ (or 50 memory locations when
extended memory is installed) scroll through the menu. All locations ar e marked
with a circle before the memory number.
-Select memory location ’m1’.
-Press the ’save’ sof tkey.
The signal is saved in memory location m1.
Observe that the circle before ’m1’ is solid now to indicate that this memory
location is occupied.
-Set the MainTB to 500 µs.
-Select location ’m2’.
-Press the ’save’ sof tkey.
-Set the MainTB to 200 µs/div.
-Select location ’m3’.
-Press the ’save’ sof tkey.
Observe that memory locations ’m2’ and ’m3’ are now solid bullets as well.
TRACE STORAGECLEAR
You can also clear a memory location.
-Select location ’m2’.
-Press the ’clear’ softkey.
This clears memory location 2. This is indicated by the open circle.
HOW TO USE THE INSTRUMENT4 - 41
TRACE STORAGECOPY
A trace can be copied from one memory location to another.
-Press the ’COPY’ softkey.
The ’COPY MEMORY’ is now displayed. The source (’FROM’) and destination
(’TO’) memory locations can be selected in this submenu.
-Turn the TRACK control to select ’m3’ as source from.
-Turn the ∆ control to select ’m2’ as destination to.
-Press the ’copy’ softkey.
The contents of ’m3’ have now been copied into ’m2’ without d estroying the data
in ’m3’.
4 - 42HOW TO USE THE INSTRUMENT
TRACE STORAGERECALL
RECALL
If you completed all of the previous steps, three memory locations are
occupied by trace information. You can display these stored traces at
any time and in any combination.
-Press the RECALL key.
-Turn the TRACK control to select ’m1’.
-Press the ’DISPLAY on off’ softkey to display m1.
Observe that the stored trace on memory location ’m1’ is displayed and that the
circle before m1 is solid.
An open circle indicates an empty memory location.
A circle with a dot indicates that there is a trace stored, but it is not displayed.
A solid circle indicates trace data in memory are displayed on the screen.
Empty memory locations cannot be displayed.
You can shift traces separately over the screen by means of the ∆ control.
-Turn the POSition control.
The live trace shifts.
-Turn the TRACK
control to select ’m2’.
-Press the ’DISPLAY on
POS
RECALL
REGISTER
MEMORIES
acq
m1
m2
DISPLAY
on off
CLEAR
DISPLAY
Y-POS
−2.30D
T
∆
off’ softkey to on.
-
Turn the ∆ control.
∆
trace
register
ST6722
9303
Observe that the second stored trace shifts.
HOW TO USE THE INSTRUMENT4 - 43
TRACE STORAGESAVING MULTIPLE TRACES
Each memory location can store two channels.
-Connect the probe to channel 1.
-Press AUTOSET.
-Switch on channel 2.
-Turn the channel 2 POS control counterclockwise.
The upper trace displays the signal of channel 1, and the lower line represents
channel 2.
-Press the SAVE key.
-Turn the TRACK control to select memory location ’m1’.
-Press the ’save’ softkey to store two traces in ’m1’.
Observe that m1 was already filled, so the confirmation submenu is displayed.
-Press the softkey next to ’yes’.
This overwrites memory ’m1’ with the new traces.
TRACE STORAGERECALLING MULTIPLE TRACES
-Press the RECALL menu key.
The memory location is already preset to the last saved (m1).
-Press the ’DISPLAY on off’ softkey so that ’m1’ is displayed.
-
Turn the ∆ control counterclockwise.
Observe that both traces are displayed and shifted to the lower screen.
-Press the softkey next to ’trace register’ to select ’trace’.
Observe that the menu displays both traces of register 1 (m1.1 and m1.2). Both are
filled: m1.1 with the signal of channel 1 and m1.2 with the signal from channel 2.
-Turn the TRACK control to select m1.1.
-Press the ’DISPLAY on off’ softkey to off.
Observe that the stored trace of channel 1 is not displayed any more.
-Press the ’CLEAR DISPLAY’ softkey to clear all stored traces before
continuing to the next section.
4 - 44HOW TO USE THE INSTRUMENT
4.10 CURSORS FUNCTIONS
The
TRACK
analog
CURSORS
∆
Cursors are provided to make fast and accurate amplitude
and time measurements. These can be done in digital
ST6171
9303
mode as well as in analog mode.
mode of the oscilloscope is often used to accurately display complex
waveforms, such as in AM, FM, and Video.
The
mode will often be the preferred operating mode for single accuracies
digital
which have to be studied or analyzed.
When operating in
mode, the cursors are always set to the optimum place
digital
for that particular measurement and the readout is directly displayed on the
screen.
This is the fastest and easiest way for most of the measurements .
But for measurements on very complex signals with different waveforms in a
trace, the
cursor measurements gives you a very helpful tool to find your
analog
measurement.
STANDARD SETTING
Start with the standard setting. This ensures you have the corre ct start co ndi tion.
-Press the STATUS and TEXT OFF keys simultaneously.
-Connect the Probe Adjust signal to channel 1.
The Probe Adjust signal now supplied to the input is a square wave with a lower
level of 0V and a top level of 600 mV.
-Press the AUTOSET key.
HOW TO USE THE INSTRUMENT4 - 45
CURSORSON/OFF
There are two sets of cursors : amplitude cursors and time cursors. Amplitude
cursors are two horizontal lines, and Time cursors are two vertical lines. The
dashed lines are referred to as reference cursors and the dotted lines are referred
to as delta (∆) cursors.
Cursor measurements can be made in both operating modes of the instrument.
Amplitude and time cursors can be displayed at the same time.
-Press the ANALOG key to set the scope in mode.
-Press the CURSORS key to enter the cursors menu.
The menu is now displayed. Time cursors (’||’) are selected by default. In the
digital mode, two ’x’s mark the actual measuring point wh ere the cursors intersect
with the signal. In the analog mode these ’x’ indicators are not available.
CURSORSTIME
Using the TRACK and ∆ controls, you can adjust the position of the cursors. The
time and amplitude difference between the cursors is read in the top of the screen.
Cursor readout parameters can be
changed and are selected via the
’READ OUT’ softkey.
-Turn the TRACK cont rol and
observe that both cursors
move.
-
Turn the Delta (∆) control
and observe that only the
delta cursor moves
.
TRACK
ch1: ∆T= 354µs ∆V= 584mV
MTB 200us
200mV
CH1
∆
ch1
CURSORS
on off
#
-
ch1
-
READOUT
ST6730
9303
4 - 46HOW TO USE THE INSTRUMENT
CURSORSVOLT
-Use the softkeys, to select the amplitude cursors (’=’ intensified).
Two amplitude cursors will appear in the display.
The menu permits the channel to be selected for which the amplitude cursors
apply.
When channel 1 is the only channel displayed, ’ch1’ is the only selection.
CURSORS
on off
-
Turn the TRACK
control to move both
cursors, and turn the
Delta (∆) control to
move only the ∆
.
cursor
∆
TRACK
ch1: ∆T=
CH1
200mV
s ∆V=360mV
MTB 200us
ch1
-
ch1
-
READOUT
#
ST6731
9303
CURSORSAMPLITUDE & TIME
Amplitude and Time cursors can be displayed at the same time. When both
amplitude and time cursors are displayed, you must select which cursors are
affected by the TRACK and ∆ controls. This selection is made by toggling the
softkey ’CONTROL’.
Channel selection applies only to the voltage cursors. The time cursor always
applies to all channels.
To use both types of cursors at the same time, do the following:
-Press the softkey next to ’= || #’ so that ’#’ is intensified.
-Press the CONTROL softkey so that ’=’ is intensified.
-Turn the TRACK control and observe that only the amplitude cursors
move.
-Press the CONTROL softkey so that ’||’ is intensified.
-Turn the TRACK control and observe that only the time cursors move.
HOW TO USE THE INSTRUMENT4 - 47
CURSORSREADOUT
-Press the softkey next to ’READOUT’.
CURSORS
READOUT
The ’CURSOR READOUT’
selection menu for horizontal and
vertical measurements is now
displayed. Y ou can make the
following selections from this menu:
ch1: ∆T= 460s ∆V= 600mV
∆T, 1/∆T, ∆T-ratio, T-trg and phase
for time measurements.
∆V, V1, V2, ∆V-ratio for voltage
measurements.
∆
is the time difference between the time cursors. The readout is in se conds (s)
T
or fractions of seconds.
results in the inverted value of the time difference, and is displ ayed in Hz.
1/∆T
When the time cursors are set exactly one period apart this represents the
frequency of the signal.
∆
allows for a relative measurement of two cursor readings. It is displayed
T-ratio
as a percentage.
ase is used to measure a phase difference between two signa ls, or to
Ph
determine the phase of a detail within one signal.
gives the timedifference between the cursors and the trigger poin t.
T-trg
∆T1/∆T
∆T-ratio
T-trg
ph
∆V
V1&V2
∆V-ratio
RETURN
ST6732
9312
∆
is the voltage difference between the voltage cursors (readout: ∆ V=...V).
V
displays the absolute voltages for each voltage cursor in relation to ground
V1, V2
(readout: V1=Vreference, V2=Vdelta).
∆
allows for a relative measurement of two cursor readings. It is displayed
V-ratio
as a percentage.
4 - 48HOW TO USE THE INSTRUMENT
CURSORS READOUT∆T-RATIO/PHASE
First, the reference distance between the two cursors is set. This then is set to
100 % (360°) by pressing the ∆T=100 % (360°) softkey. Changing the distance
between the cursors now results in a % (°) reading of the reference.
-
Press the softkey next to ∆T until ∆T-ratio (phase) is intensified.
-Set the time cursors to be exactly one signal period apart.
-
Press the softkey next to ∆T=100% (360°).
In the cursor readout line of the screen, the value for ∆T is now ’∆T=100%’
(∆T=360°)
Observe that the 100% (360°) reference does not change when the TRACK
control is turned.
ch1: ∆T=100%
TRACK
100%
∆V=600mV
∆
∆T=100%
50%
ch1: ∆T=50%
TRACK
-Adjust the cursors to a distance of half a signal period.
Now the ∆T in the readout area will read ’∆T= 50%’ (∆T=180°)
This is the way to use the cursors for duty cycle measurements.
∆V=600mV
∆
ST6733
HOW TO USE THE INSTRUMENT4 - 49
4.11 MEASUREMENT FUNCTIONS
When operating in the
amplitude and time data.
When operating in the
automated
measurements to be performed simultaneously. These measurements are
updated automatically, so that when signals change, the measured values
automatically changed at the same time. The read- outs are displayed in the
upper section of the screen.
amplitude and time measurement functions. You can select two
mode, you can use the cursors to measure
analog
mode, the scope has an extensive set of
digital
fully
STANDARD SETTING
Start with the standard setting.This ensures a correct start.
-Simultaneously press the STA TUS and TEXT OFF keys.
-Connect the Probe Adjust signal to channel 1.
The Probe Adjust signal now supplied to the input is a square wave with a lower
level of 0V and a top level of 600 mV.
-Press AUTOSET.
4 - 50HOW TO USE THE INSTRUMENT
MEASUREMEAS1-PKPK
MEASURE
-Press the ANALOG key to select the digital mode.
-Press the MEASURE key.
The displayed menu gives access to the two measurements MEAS 1 and
MEAS 2. Each measurement can be independently turned on and off. In this
menu, you can select the measurement in MEAS 1 and MEAS 2 function.
Observe that the measurements are linked to the active channel. The two default
measurements are ’pkpk’ and ’freq’.
-Press the second softkey to turn MEAS1 to ’on’.
The screen automatically displays the result of the peak-to-peak measurement:
’ch1: pkpk= ....mV’.
-Press the softkey next to ’MEAS1’.
The menu displays the ’SELECT MEAS 1" submenu. You can choose from three
main measurements: volt, time, or delay. The lower part of the menu optimizes to
the selected main measurement.
The actual type of measurement is selected by using the ’TRACK’ control. The
following voltage measurements are available : dc, roo t- mean squar e, minimum
peak, maximum peak, peak-to-peak, low level, high level.
The following time measurements are available : frequency, period, pulse width,
rise time, fall time, duty cycle.
The ’delay’ measurements include the time delay between leading or trailing
edges between any two channels or traces. (Channel or trace
must
be on the
display)
MEASUREMEAS1-RMS
-Select ’volt’ measurements.
-Turn the TRACK control to select ’rms’.
Observe that the screen displays the result of the rms measurement: ’ch1 rms = ....V’.
HOW TO USE THE INSTRUMENT4 - 51
MEASUREMEAS 2-FREQ
-Press the ’RETURN’ softkey to return to the ’MEASURE’ menu.
-Turn on MEAS 2.
MEAS 2 performs a frequency measurement on the same signal. The screen
displays this result as : ’ch1 freq= ....kHz’.
-Remove the Probe Adjust signal from channel 1
Because of the absence of an input signal, no frequency can be measured. This
results in the display ’ch1 freq= ----Hz’.
MEASUREDELAY
-Connect the Probe Adjust signal to channels 1 and 2.
-Turn on channel 2.
-Shift the signal on channel 1 to the upper half of the screen and the
signal on channel 2 to the lower half of the screen.
-Press the softkey next to MEAS1.
The SELECT MEAS1 submenu is displayed.
-Press the first softkey to select ’delay’
The delay measurement is
performed automatically.
Observe that the first
measurement is linked to the
channel 1 signal (on positive
SELECT
MEAS 1
volt
time
delay
-
ch1
ch2
T
slope) and the second is listed to
the channel 2 signal (on
negative slope).
The delay is displayed in the top
corner of the screen
’ch1: del = .... µs’.
1−
2−
DELAY
ch1
ch2
-
RETURN
∆
ST6723
9303
-Press the RETURN softkey to return to the MEASURE menu.
This measurement is very useful when you must compensate unequal cable delays.
4 - 52HOW TO USE THE INSTRUMENT
MEASURECURSOR LIMIT
With cursor limited measurements it is possible to perform measurements on a
part of the waveform.
-Press the softkey next to CURSOR LIMIT& STATIST.
-Set CURSOR LIMITED to ’yes’.
-Turn the ∆ control to reduce the area between the cursors.
Notice that the measurement result shows dashes when the area between the
cursors gets too small.
-Press the RETURN softkey to return to the MEASURE menu.
-Turn off the measurements MEAS1 and MEAS2.
-Disconnect the probe from channel 2 before continuing to the next
section.
HOW TO USE THE INSTRUMENT4 - 53
MEASURETOUCH, HOLD & MEASURE
™
The probes delivered with the oscilloscope offer a unique and innovative way to
perform a number of functions directly from a push button mounted on th e side of
the probe. One of these functions is called ’TOUCH HOLD and MEASURE ™’.
The TOUCH HOLD & MEASURE ™ function is a very quick way to instantly
display the four main measurements. You need only one single push button. This
push button is located nearest to the place of measurement on the probe;
therefore, this button is referred to as COMMAND switch.
3
2
1
4
Before using the COMMAND
switch on the probe, you must
first adapt the instrument’
reaction to this function. This is
COMMAND
ST6837
done in the ’UTILITY >>
PROBE’ menu.
9303
The procedure to set up for TOUCH, HOLD and MEASURE ™ is:
-Press the UTILITY key.
-Press the softkey labeled ’PROBE’.
-Select ’q.meas’.
-Press the COMMAND push button on the probe.
Observe that the screen displays the signal frequency and amplitude dc voltage
level (dc, pkpk, rms, and freq) measurements.
The probe
COMMAND
switch gives you the fastest and easiest method of
trouble-shooting. You can keep your eyes on the circuit to be measured without
having to move your eyes to press a push button on the scope.
NOTE: The probe COMMAND switch can also be used for other functions.
Programming of other functions is done in the UTILITY menu
(see section 5)
4 - 54HOW TO USE THE INSTRUMENT
4.12 PROCESSING FUNCTIONS
MATH
In addition to those traditional oscilloscope functions, this range of fers capabilities
to mathematically change the contents of each memory location. To do so, the
functions under ’MATHEMATICS’ allow you to calculate new waveforms from
existing other traces serving as input data.
If your oscilloscope is equipped with the MATH + option, more functions are
available than the ones described in this section. Additional information on the
additional functions is described in a separate manual.
The section below describes the mathematical functions implemented in the
standard oscilloscopes versions.
Two mathematical functions MATH 1 and MATH 2 can be used as two
independent processes. They can be used in series or in parallel.
The result of MATH 1 is always stored in memory location ’m1’.
The result of MATH 2 is always stored in memory location ’m2’.
When used for mathematical functions, existing memory locations are
overwritten. (Use the ’Copy’ function to save important traces in another memo ry
location, before using the mathematical functions.)
Most oscilloscopes, including most Digital Storage Oscilloscopes, limit
their capabilities to the display of amplitude (in volts), versus time.
STANDARD SETTING
Start with the standard setting to ensure that you have the correct start condition.
-Simultaneously press the STATUS and TEXT OFF keys.
-Connect the Probe Adjust signal to channel 1.
-Press AUTOSET.
The Probe Adjust signal now supplied to the input is a square wave with a lower
level of 0V and a top level of 600 mV.
HOW TO USE THE INSTRUMENT4 - 55
MATHEMATICSFILTER
-Press the ANALOG key to select the digital mode.
-Press the MATH menu key.
The MA TH m enu i s displayed on the scr een. Observe that MATH 1 is the default.
-Press the softkey next to MATH 1 to enter the MATH 1 submenu.
This menu is used to select one of the four mathematic functions: ’add, sub, m ul,
filter’.
Observe that the default process for MA TH 1 is the ’filter’ function. T ypical use of
the digital low-pass filter is to suppress noise even after a (si ngle shot) acquisition
has to be captured.
You can select between the acquired signal or a stored signal to be the source.
The trace information in the source trace is mathematically filtered by a low-p ass
filter.
The resulting trace is automatically stored in memory location ’m1 an d displayed
on the screen.
-Turn the TRACK control to select ’acq’.
-Press the ’ENTER’ softkey.
-Press the key next to ’on off’ to turn MA TH 1 on.
1-
3dB=4.25kHz
200mV
CH1
MTB
200µs
ch1
TRACK
1-
3dB=11.2kHz
200mV
CH1
MTB
200µs
-Turn the TRACK control to change the cut-off frequency.
-Switch off MATH 1 again before continuing to the next section.
ch1
ST6745
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.