Each Fluke product is warranted to be free from defects in material and workmanship under normal use and
service. The warranty period is two years and begins on the date of shipment. Parts, product repairs, and services
are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke
authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion,
has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation
or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications
for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software
will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but
have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if
product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased
in one country is submitted for repair in another country.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or
replacement of a defective product which is returned to a Fluke authorized service center within the warranty
period.
To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization
information, then send the product to that service center, with a description of the difficulty, postage and insurance
prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following warranty repair, the product will
be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that failure was caused by
neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including
overvoltage failures caused by use outside the product’s specified rating, or normal wear and tear of mechanical
components, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work.
Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the
repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF
DATA, ARISING FROM ANY CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation
of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every
buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of
competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
The Fluke NORMA 6003/NORMA 6003+/NORMA 6004/NORMA 6004+ Power Analyzers (the Product
or Analyzer) are portable, precision wideband power analyzers. Use the product with the Fluke current
clamp 80i-2010s and High Voltage Differential Probe U1500s for a broad range of electrical loads
including those with electronic switching (inverter type).
The Product has these benefits:
•Up to 4 voltage and 4 current channels (NORMA 6003+, NORMA 6004+).
•Motor Speed and Torque measurements (NORMA 6003+, NORMA 6004+)
The Product is shipped in a container designed to prevent damage during shipment. Inspect the
Product carefully for damage, and immediately report any damage to the shipper.
Figure 1 and Table 1 list the equipment and accessories included with this Product. For specific
information on each part and more accessories, see the printed Getting Started Manual that ships with
Unpack the Product and check for all the standard equipment listed in Table 1 and check the shipping
order for additional items ordered. Report any shortage to the place of purchase or to the nearest
Technical Service Center.
If it is necessary to reship the Product, use the original container. If it is not available, you can order
new packaging from Fluke by identifying the Product model and serial number.
Power Supply Adapter (18V DC, CAT IV 600V
Adapter)
Power line cord (for China): 4894155
Power line cord (for international): 4894137
Mains Adapter set (for international): 4894143
Test lead set
4-channel model:
Test leads, 1.5 m, (4) red, (4) black, (4) blue
Test leads, 0.18 m, (8) blue
Alligator clips, (4) red, (4) black, (4) blue
3-channel model:
Test leads, 1.5 m, (3) red, (3) black, (3) blue
Test leads, 0.18 m, (6) blue
Alligator clips, (3) red, (3) black, (3) blue
11
11
11
Not Shown
4
FRS-485/synchrony adapter (3.81 mm, black, 3 pin)11
GSpeed / torque adapter (3.81 mm, black, 6 pin)None/1None/1
H
USB Flash Drive for PC SW, (contains User
Manual)
Getting Started Manual11
Product Warranty and Certification (China only)11
Calibration Report11
QC certificate (for China only)11
The Analyzer is portable and equipped with a tilt stand that allows the Analyzer to be placed on a flat
surface for users to view the screen from a certain angle, as shown in Figure 2.
Note
Do not use the tilt stand to carry the Product.
Figure 2. Use the tilt stand
Install the supplied hand strap on the left side of the Product. Adjust the handstrap, as shown in
Figure 3.
Figure 3. Install the hand strap
You can hold the Product in one hand when working and use the hand strap to carry the Product when
the Product is not in use.
Before you use the Product for the first time, charge the battery.
When powered by a battery, the battery symbol at the top of the screen indicates the condition of the
battery. indicates low battery power, charge the battery immediately.
To charge the battery and power the instrument, connect the power adapter as shown in Figure 5. To
increase the battery charging speed turn off the Analyzer.
Figure 5. Battery Charging
Caution
To prevent overheating of the battery during charging, do not charge in extremely hot
or cold environments. If the battery is charged under extreme temperatures, the battery
capacity may be reduced. Refer to charging temperature specifications.
Note
Charge the Product for a two-hour minimum at three-month intervals for maximum battery life.
When not in use, the full charged battery will self-discharge in approximately six months.
Batteries stored for long periods will need two to ten charging cycles for full capacity.
This section describes the operation panel of the Product and the location and function of the display.
Please read this section carefully before operating the Product. This manual is applicable to many
models. Because models have different features and accessories, not all of the information in the
manual may apply to your Product.
Table 2 is a list of functional comparison of different models.
The 1 to 5 function keys correspond to the soft keys from top
to bottom on the right next to the screen, the corresponding function
keys are equivalent to the soft keys. The label of the soft key varies
depending on the function and interface.
Back key
Push J to exit the current screen and return to the previous
screen or the previous menu
Select key
When a function on the screen is selected (highlighted in yellow),
push S to select or enter the function.
D
Arrow keys (up, down, left, and right)
Push the keys to go through all selectable functions on the screen
and select one of them, the currently selected function is highlighted
in yellow.
Displays the menu path of the current screen, for example: Configure > Wiring.
Clock
Shows the current date and time of the Analyzer.
D
E
or Integration timer
When integrating, the status bar shows the integral indicator. See Integration for details.
or
Indicates that the communication is connected.
Battery status
Displays battery status.
•d the charging indicator, this shows that the Product is connected to the main
power.
the main power is not connected, and the vertical bar in the icon indicates the
•
battery level.
Main Display Area
The main display area is the main display area of the screen for displaying various setting options,
current measurements, charts and lists.
The main display area displays different content depending on the current interface mode and location,
as detailed in the relevant sections below.
Soft Keys
The soft keys, at the right side of the display, have labels that will vary depending on the functions and
interfaces. The soft keys correspond to the function keys (F1 to F5) on the right side of the screen.
If a soft key has no label, the key has no function.
Except for the main screen, the first menu item for the main measurement screens,
which is used for global configuration. Therefore, the function of the
described again in the menus.
Table 5 shows the connections on the interface panel.
Table 5. Connector Panel
3421
HI
10A FUSED
LO
10A MAX
EXT SE NSOR
10A MAX
EXT SE NSOR
65
ItemDescription
Voltage/current input channel number
Current port protection cover
The protective cover can slide back and forth. When the protective cover slides forward,
the external current sensor input port is exposed. When the protective cover slides back,
the current input port is exposed.
D
E
Voltage input port
Current input port
Input port for external current sensor
Input port for motor speed / torque signal
PinSignal
10A MAX
EXT SE NSOR
1Input for torque analog quantity, 10 V range
2Common port for torque input
F
3
Input for speed analog quantity, 10 V range or input for
speed pulse, A pulse
4Input for speed pulse, pulse B
5Input for speed pulse, pulse Z
6Common port for speed input
Under all view modes (Meter, Scope, Trend, Phasor and Harmonic), the soft key 1 is Configure.
The Configure soft key is global configuration and is applied to all other modes.
Use the Configure menu to set the range for each channel, timing synchronization, wiring, formula,
zero and formula editing.
Enter the Configure Menu:
•From the Main screen of any view mode select
The functions in the Configure menu are as shown in Table 6.
Table 6. Global Configuration
ItemDescription
Sets the voltage/current range for each channel, ratio, filter, and selects
1Range
2Wiring
3Timing Sync
4ZeroEliminates the input offset by zeroing. See Zero.
5Formula
whether to use an external sensor. NORMA 6003+ and NORMA 6004+
can also set the mode and ratio of the motor torque/speed measurement.
See Set Range for details.
Sets the wiring selection and the synchronization source of each wiring
group. For detailed information, see Set the Wiring Selection.
Sets the sync source of each wiring group. For detailed information, see
Set Sync Source and Update Rate.
Defines the efficiency formula and the formula for some numerical
calculation. See Define Calculation Formula for details.
1 (Configure).
Set Range
From the Main screen of any view mode:
1. Select
The Configuration interface of channel 1 is as shown in Table 7.Vol. Range is selected by default.
2. Push S and use the up and down arrow (K L) keys to select the corresponding option. Then
push S again to confirm the option.
3. Use the up and down arrow (K L) keys to move the cursor to Scale. Volt.. Then push S.
An input keyboard appears.
4. Use the arrow (K LM N) keys to move the cursor over the corresponding character on the
keyboard, then push Sto enter the value into the input box at the top. Repeat until all
characters are entered.
is the backspace key, which deletes the character before the cursor.
5. Use the up and down arrow (K L) keys to highlight Ok on the keyboard, then push S to
confirm the input value. To abandon your changes, select Exit on the keyboard and push S or
J.
6. Use the up and down arrow (K L) keys and S to complete the configuration of the remaining
items.
7. Push J to go to the Configure menu.
8. Push 2 to set the range and ratio of CH1 to CH4.
9. Push J repeatedly to go back to the main screen.
Warning
Set the parameters strictly according to the scale of external voltage divider, current
shunt and current clamp. Otherwise, the measurement will not reflect the actual signal
and there is danger of electric shock.
Note
The nameplate of the shunt or probe is usually marked with its parameter settings. Refer to the
relevant information of the external sensor if necessary.
Note
The 10 kHz anti-aliasing filter is in the measurement channel. The anti-aliasing filter is
necessary to analyze Fast Fourier Transform (FFT) data correctly and can be used to ensure
that the measurement bandwidth is much less than 1⁄2 sample rate to avoid signal aliasing.
650 Hz low-pass filtering is implemented by a digital filter that helps distinguish the synthesis
frequency and carrier signals of the variable frequency system. Turn on this filter to easily
observe the synthesis frequency of the frequency conversion system.
If measuring a 3-phase system, Fluke recommends that the line filter settings for each channel
in the 3-phase system should be consistent.
•Auto: Autorange. The Analyzer automatically selects the voltage
measurement range with the best resolution.
•10 V
•100 V
•1000 V
Scale of an external voltage sensor (PT). After setting the scale of the
external voltage sensor correctly, the Analyzer automatically calculates the
actual voltage value of the measured signal based on the voltage scale and
the measured value.
The Vol. Ratio represents the voltage ratio of the external sensor.
For example, when connecting the U1500s high-voltage probe with a voltage
ratio of 4mV/V, which represents a voltage ratio of 4:1000, the Vol.
Ratio (V/V) on the Analyzer should be set to 0.004.
Warning
Vol. Line Filter
If the voltage input terminal is not connected to an external
voltage divider in series, set the scale to 1, otherwise the
measured voltage will not be consistent with the actual
voltage, and there is a danger of electric shock.
If the voltage input terminal is connected with an external
voltage divider in series, the actual voltage will be calculated
according to the set scale.
Voltage signal filter options:
•OFF: No filter is used.
•650 Hz: Automatically turn on the line filter, and the cutoff frequency is
650 Hz. Use to observe the synthesis frequency signal of the frequency
conversion system.
•10 kHz: Automatically turn on the line filter, and the cutoff frequency is
10 kHz. Use for anti-aliasing.
Inserting a voltage measurement circuit has a direct impact on the
measurement of voltage and power. When the line filter is turned on, the
measurements do not contain high frequency components. Noise from the
frequency inverter or distorted waveform can be removed.
•Shunt: Auto, 0.1A, 1A, 10A (when "Ext. Sensor" is not selected)
•Ext. Sensor: Auto, 0.1V, 1V, 10V (when "Ext. Sensor" is selected)
Choose whether to use an external sensor.
When the external sensor is selected/deselected, you must reset the current
range.
If an external sensor is not selected, the current signal is directly connected
to the current channel of 10A, and the current is measured via the internal
shunt.
If an external sensor is selected, the current is measured by connecting the
current signal of external sensor output to the BNC interface.
Scale of an external current sensor (CT). After setting the scale of the
external current sensor correctly, the Analyzer automatically calculates the
actual current value of the measured signal based on the current scale and
the measured value.
The ratio settings are, A/A or V/A.
If the internal shunt is selected, the current scale represents the ratio of the
actual current (A) to the input current (A) on the shunt (that is A/A).
If an external sensor (BNC) is selected, the current ratio represents input
voltage (V) on the BNC port to the ratio of the actual current (A) of the
connected sensor (expressed as mV/A).
Cur. Line Filter
18
For example, when connecting a current sensor probe with an output
sensitivity of 1 mV/A, which represents a 1:1000 ratio, the Cur. Ratio (V/A) on
the Analyzer should be set to 0.001.
Current signal filter options:
•Off: No current filter is used.
•650 Hz: Automatically turn on the line filter, and the cutoff frequency is
650 Hz. Use to observe the synthesis frequency signal of the frequency
conversion system.
•10 kHz: Automatically turn on the line filter, and the cutoff frequency is
10 kHz. Use for anti-aliasing.
Inserting a current measurement circuit has a direct impact on current and
power measurements. When the line filter is turned on, the measurements
do not contain high frequency components. Noise from the frequency
inverter or distorted waveform can be removed.
The Configuration interface of Motor Speed / Torque is as shown in Table 8.
The Mode is selected by default (the drop-down menu is highlighted in yellow).
2. Push S and use the up and down arrow (K L) keys to select the corresponding option. Then
push S again to confirm the option.
3. Use the up and down arrow (K L) keys to move the cursor to Ratio. Then push S.
4. Enter the correct values on the keyboard by using the up and down arrow (K L) keys and push
S.
5. Use the up and down arrow (K L) keys to highlight the Ok button on the keyboard, then push
S to confirm the input value.
6. For the Ratio option of the torque, repeat step 3 to 5 to configure the torque ratio in V/Nm.
7. Use the arrow (K LM N) keys to move the cursor to the Apply button and push S. The
settings are not saved to the device if you do not select Apply.
8. Push J repeatedly to go back to the main screen.
Speed
1 (Configure) > 1 (Range) > 5 (Motor).
Table 8. Motor Speed / Torque Channel Configuration
ItemDescription
Mode
Ratio
Speed signal mode options:
•10V (default): Use for speed sensor for analog voltage output. If the
sensor uses current, frequency or other analog output, it must be
converted to voltage before input.
•A: For encoders with single pulse, only the speed value is measured
and recorded, the direction of rotation is not measured and recorded.
•AB: For encoders with AB double-pulse, both speed value and
direction are measured and recorded.
•ABZ: For ABZ incremental encoder, both speed value and direction are
measured and recorded.
Speed sensor ratio.
When Mode is set to 10V, the ratio is in V/rpm, which indicates that how
much voltage (V) it is related to the motor speed (rpm). The formula for the
Analyzer to calculate the actual speed is as follows:
Speed = Voltage/ Ratio
When Mode is set to another option, the unit of the ratio is the number of
pulses per revolution (PPR). The formula for the Analyzer to calculate the
actual speed is as follows:
Table 8. Motor Speed / Torque Channel Configuration (cont.)
ItemDescription
Torque
Mode: The torque input is -10 V to 10 V the supported input signals must be conditioned to operate
within these limits. If the sensor has a current output this must be converted using a shunt resistor
before applying to the input.
Torque sensor ratio. The formula for the Analyzer to calculate the actual
Ratio
torque is:
Torque = Voltage / Ratio
Set the Wiring Selection
To measure the power of various single-phase or three-phase power, different wiring selections are
required, and multiple channel combinations are used. For more information on the connections, see
the Getting Started Manual, shipped with the product or available online.
The Analyzer provides these basic wiring selections:
•1P2W1M, 1-phase 2-wire 1-meter
•3P3W2M, 3-phase 3-wire 2-meter
•3P4W3M, 3-phase 4-wire 3-meter
•3P3W3M, 3-phase 3-wire 3-meter
Note
In the abbreviation of the wiring selection, letter "P" indicates "Phase", letter "W" indicates
"Wire", letter "M" indicates "Meter", that is power meter (includes one set of voltmeter and
ampmeter).
1P2W1M
A
N
Figure 8. 1P2W1M Wiring
For 1P2W1M wiring, connect a voltage measurement channel between the phase line and the zero
line. Connect a current measurement channel in series with the phase line or the zero line (note the
direction) or clamped on the phase line or the zero line with a current sensor.
For 3P4W3M wiring, 3 voltage measurement channels are respectively connected between each
phase line and the neutral line. 3 current measurement channels are respectively connected in series
with each phase line or clamped on each phase line with a current sensor.
For systems without neutral lines, connect 3 voltage measurement channels between the 3 phase lines
and the common ground line. For a 3P5W system, you can use a wiring group of 3P4W3M to measure
the phase and neutral line. You can select one channel and configure it as 1P2W1M to measure the
parameters between the neutral and ground line (NORMA 6004/6004+ only). If the ground and neutral
lines do not exist, connect the common ends of the 3 channels together and suspend, and the total
power measured is still valid.
For this wiring system:
1. The wiring system has the same impedance characteristics for 3 phases, a balanced measurement
system can be established.
2. The wiring system has a terminal for neutral or ground, the common mode component input to the
instrument is small, which reduces the effect of common mode components on the measurement
results.
3. The phase voltage is directly measured, and the line voltage is calculated by Δ. It is more suitable
For 3P3W3M wiring, 3 voltage measurement channels are respectively bridged between 2 phase lines.
3 current measurement channels are respectively connected in series with each phase line or clamped
on each phase line with a current sensor.
For this wiring system:
1. This wiring system has the same impedance characteristics for 3 phases, a balanced
measurement system can be established.
2. When measuring the frequency conversion system with the power frequency input, this wiring
system can appropriately eliminate the power frequency component in the frequency conversion
signal, which is conducive to more accurately capturing the conversion frequency.
3. The line voltage is directly measured, and the phase voltage is calculated by Δ. It is more suitable
for measuring systems of Δ connections.
4. The neutral line current can be measured by summing 3 current channel, but the neutral line
voltage cannot be measured. When the neutral line voltage/neutral line current is present, the total
power measured is also inaccurate.
For 3P3W2M wiring, 1 phase line (C) is used as a common line, 2 voltage measurement channels are
respectively bridged between the other 2 phase lines and this phase line. 2 current measurement
channels are respectively connected in series to the other 2 phase lines, or a current sensor is clamped
on the other 2 phase lines (Also known as two-wattmeter method or Aron method).
For this wiring system:
1. The wiring system can measure 3-phase system with only 2 channels, and 1-2 channels can be
reserved for measuring DC, single-phase AC or 3-phase AC (simultaneous measurement of 2channel 3-phase AC is only available for NORMA 6004/6004+).
2. When measuring the frequency conversion system with the power frequency input, this wiring
system can appropriately eliminate the power frequency component in the frequency conversion
signal, which is conducive to more accurately capturing the conversion frequency.
3. The system impedance for the 3 phases is different. The phase voltage and part of the line voltage
are calculated by Δ. In addition, this wiring system cannot measure the voltage and current of the
neutral line. When the neutral line voltage/neutral line current is present, the total power measured
is also inaccurate.
Number of channels occupied332
Balanced test impedanceYesYesNo
The power frequency components of frequency
conversion systems can be eliminated
Phase voltage
Line voltage
Common mode interference voltage inputtedLowModerateHigh
Neutral line current measurementYesYesNo
Note
Where the current is measured may have some effect on the measurement results. To mitigate
this effect, consider where the voltage and current are measured. For a signal with high
voltage and low current, the voltage channel should be connected to the side that is closest to
the source, the current channel should be connected in series (or the sensor should be placed)
to the side that is closest to the load (as shown in the wiring diagrams). For a signal with low
voltage and high current, connect the voltage channel to the side that is closest to the load,
and connect the current channel in series (or the sensor should be placed) to the side that is
closest to the source. Please refer to the appendix Measurement Error Caused by the Channel
Resistance.
NoYesYes
Direct
measurement
Indirect
calculation
Indirect
calculation
Direct
measurement
Indirect
calculation
Partial direct
measurement
Wiring Group
When you select a specific wiring selection, use a combination of 2 or more channels, which is called a
wiring group.
For example, when the wiring selection of channel 1 is set to 3P3W2M, channel 1 and channel 2 form
a wiring group.
2. The Configuration interface of wiring selection is as shown in Figure 12.
1 (Configure) > 2 (Wiring).
Figure 12. Wiring
Note
The optional wiring selection is related to the number of channels equipped with the Analyzer,
and the interface of wiring selection will be different. Proper selection of wiring is conducive to
ensuring more accurate measurements.
The wiring Configuration interface is generally divided into these parts:
•Channel selection: The upper right part shows the wiring group configuration of all 3/4 channels.
Different wiring groups can be selected by using the left and right arrow (M N) keys. The name of the
selected wiring group is highlighted in red.
•Wiring selection: The lower right part shows the wiring options supported by the currently selected
channel. Use the up and down arrow (K L) keys to select the wiring mode of the corresponding
channel.
•Wiring diagram: The upper left part shows the wiring diagram of the currently selected wiring
mode in the “Wiring Selection”.
•Formula: The lower left part shows the parameter calculation method of the currently selected
In the Wiring screen (Figure 12), use the left and right arrow (M N) keys to select a channel or wiring
group in Channel Configuration. The wiring selection of the selected channel/wiring group is
highlighted.
2. Use the up and down arrow (K L) keys to select the appropriate wiring mode in Wiring Selection.
3. Push S.
4. Use the arrow keys (M N K L) and S to complete the configuration of the remaining items.
5. Push J to go back to the previous menu. Push the button continuously to go back to the main
screen.
Set Sync Source and Update Rate
The sync source is a reference input signal, and the time interval of the measurement is determined
according to the frequency of the selected sync source signal, and the sample values in this interval
are used to generate measurements such as voltage, current and power, thereby ensuring calculation
with the signal of the entire cycle to avoid measurement errors due to signal truncation.
1. From the Main screen of any view mode select
The setup interface of Timing/Sync is as shown in Table 10.
Table 10. Timing and Sync
ItemDescription
The time interval at which the DSP calculates the data values.
•100 ms
Update rate
Source
The wiring groups that can be configured in the interface depends on the wiring selection. See
Set the Wiring Selection for details.
•200 ms
•500 ms
•1 s
The sync source of each wiring group.
All channels in each wiring group can only share the same sync source.
The available sync sources include voltage and current signals for all
channels in the wiring group.
1 (Configure) > 3 (Timing Sync).
Note
2. Use the up and down arrow (K L) keys to move the cursor to Update Rate.
3. Push S and use the up and down arrow (K L) keys to select the corresponding option. Then
push S again to confirm the option.
4. Use the up and down arrow (K L) keys to move the cursor to the first wiring group in Source.
5. Push S and use the up and down arrow (K L) keys to select the corresponding option. Then
push S again to confirm the option.
6. Use the left and right arrow (M N) keys and S to select each of the remaining wiring groups one
by one.
7. Push J to go back to the previous menu. Push Jcontinuously to go back to the main
screen.
Set Sync Source
When the Analyzer measures AC signals, most of the measurements are calculated using the full-cycle
sampling method to reduce the truncation error. In this case, a channel needs to be selected as the
sync source to determine the length of the entire cycle.
Figure 13. Set Sync Source
For each wiring group, you can select a sync source, and the voltage or current of any channel in the
wiring group can be selected as the sync source. In general, the voltage of the power frequency system
is more stable than the current. Fluke recommends using the voltage as the sync source. The voltage
of the frequency conversion system has a large high-frequency component, while the high-frequency
component of the current is relatively small. Therefore, when the current is relatively stable, the current
can be selected as the sync source. However, if the current value is small or it is not stable, then the
voltage should be selected as the sync source.
Zero
The function of zeroing is to reset the current value of the specified channel to calibrate the zero
position of the Analyzer itself and the external sensor. The Analyzer supports zeroing each channel.
1. From the Main screen of any view mode select 1 (Configure) > 4 (Zero).
2. For each channel you want to zero:
a. Use the arrow (K LM N) keys to move the cursor to the option you want to zero,
b. Push S, the selection box of the current item turns yellow and a black “×” symbol shows,
indicating that it has been selected. If you push S again, the selection box will be restored
to blank, indicating that it is not selected.
3. Use the arrow (K LM N) keys to move the cursor to the Zero button and push S.
The Analyzer allows the user to define efficiency and some numerical calculation formulas.
Define Efficiency Calculation Formula
The Analyzer allows the user to edit the efficiency calculation formula as needed.
1. From the Main screen of any view mode select 1 (Configure) > 5 (Formula) > 1 (ƞ).
The interface of calculation formula definition is as shown in Table 11.
2. Use the up and down arrow (K L) keys to move the cursor to the numerator of the efficiency (ƞ)
formula. When the numerator portion turns yellow, push S.
A keyboard shows on the screen.
The keyboard is divided into 3 parts: the top area shows all the measured parameters of the main
unit, the middle “Remote” area shows the measured parameters of the remote slave unit in the
dual-unit interconnection mode (in single unit mode, this area is gray and not selectable). The
bottom area contains operators and buttons.
Table 11. Define Calculation Formula
ItemDescription
Operators and buttons
Measured parameters from main unit
This part of the keyboard lists all measurable power symbols for the
Analyzer's current wiring selection. For the specific meaning of each
parameter, refer to section Measurement Calculation Method in the
appendix.
Measured parameters from the slave
•The measured parameters of the remote slave unit in the dual-unit
interconnection mode.
•In single unit mode, this part is gray and cannot be selected.
•For the specific meaning of each parameter, refer to section
3. Use the arrow (K LM N) keys to move the cursor over the corresponding character on the
keyboard, then push S to enter the value into the input box at the top.
4. Continue to use the arrow (K LM N) keys and S to enter an expression that conforms to the
efficiency calculation rule in the input box.
Note
During input process, if the input does not conform to the rules, the Analyzer will indicate the
input is incorrect. Please double check that the input conforms to the efficiency calculation
expression rules.
5. Use the arrow (K LM N) keys to move the cursor to OK and push S.
6. For the denominator of the efficiency (ƞ) formula, continue to use the arrow (K LM N) keys and
S to enter a formula.
7. Push Jto go back to the main screen.
Define hvf/hcv Calculation
hvf/hcv is a numerical measurement item used for measuring rotating motors according to IEC60034. It
is necessary to configure the type of motor and rated voltage/current according to the actual test
requirements.
1. From the Main screen of any view mode select 1 (Configure) > 5 (Formula) > 2 (hvf hcf).
The interface of calculation formula definition is as shown in Figure 14.
In the hvf/hcv calculation page, the upper part is used to calculate hvf, and the lower part is used to
calculate hcf.
Figure 14. hvf/hcf Calculation Formula
2. Use the arrow (K LM N) keys to move the cursor to Motor Type, select Non 3-phase or 3-phase,
k≠3n and then push S.
3. Use the arrow (K LM N) keys to move the cursor to “U
Specify and then push S.
oIf Use Measured is selected, the measured voltage fundamental amplitude is used as the rated
voltage.
oIf Specify is selected, you can specify the rated voltage. Move the cursor to the corresponding
input box, then push S, and input the corresponding value using the on-screen keyboard,
then move the cursor to OK key on the keyboard, and push S.
4. Use the arrow (K LM N) keys and Sto enter the hcf settings.
For I
,
N
oIf Use Measured is selected, the measured current fundamental amplitude is used as the rated
current.
oIf Specify is selected, the rated current can be specified.
5. Use the arrow (K LM N) keys to move the cursor to the Apply button and push S. The
settings are not saved to the device if you do not select Apply.
6. Push J to go back to the previous menu.
” (rated voltage) select Use Measured or
N
Meter
When the Analyzer is turned on, it defaults to the Meter mode, and the current measurements are
displayed in a list on the main screen.
You can also push l to enter the Meter mode.
Table 12 lists the items displayed on the screen.
Table 12. Meter Screen
ItemDescription
Current range information for each module
•The representation with (Auto) indicates auto range, which
Range configuration
information
Wiring configuration
information
Update rate
Current state
automatically switches the range according to the signal.
•The representation without (Auto) indicates that the current range is
fixed, the range of the value displayed and the range will not change
with the signal.
Current wiring selection for each wiring group.
The current update rate of the Analyzer. See Set Sync Source and Update
Rate.
•Preview: Currently in Preview mode.
•User Item: Currently in User Item mode.
•HOLD: Currently in data hold state. The value displayed in the value
list area does not change with the actual signal.
The measurement items are displayed in a list according to the user's
settings. See Preview and User Item View for details.
www.GlobalTestSupply.com
Table 12. Meter Screen (cont.)
ItemDescription
1 (Configure): Global configuration, see Global Configuration.
Power Analyzers
Operation
Soft keys
Bottom BarDisplays the current page number.
2 (Preview): Enter Preview mode, see Preview.
3 (User Item): Enter User-defined Mode, see User Item View.
4 (Integral): Integration Operation, see Integration
Preview
In Preview mode, all currently available measurements are shown in a certain order on the screen.
Enter Preview Mode:
•From the Main screen of Meter select 2 (Preview).
The Preview screen is as shown in Table 13 and the state label shows Preview.
Table 13. Meter Screen - Preview
ItemDescription
1Sort by Function
2Sort by Channel
Push 1 (Sort by Func.) to sort the measurements by functions.
The label of Sort by Func. is highlighted in dark blue.
Push 2 (Sort by CHANNEL) to sort the measurements by
channels.
Displays the number of items
3Display
4
5
The number of measurements shown on each screen up to
20 measurements:
•4
•8
•12
•16
•20
Page up key
When the number of measurements available exceeds the number
of items displayed on each screen, the and keys can be used
to scroll forward or backward to display more measurements.
The page number of the current screen is displayed in the prompt
bar at the bottom. “1/40” means there are 40 pages in total, and the
current screen is page 1.
Page down key
Page down key for scrolling backward to display next page.
The number of measurements can be very large depending on different wiring selection and the
number of channels. Even if the maximum number of measurements is displayed on each screen,
there could be dozens of screen pages.
The Analyzer features a User Item View function, which allows users to select up to 20 most interesting
measurements and display them on a single screen, thus avoiding frequent use of the and keys.
Enter User Item View:
•From the main screen of Meter select 3 (User Item).
The User Item view is as shown in Table 14 and the data source label shows User Item”
Table 14. Meter Screen - User Item View
ItemDescription
1Edit
2Add
3Del
4
5
When you enter the User Item view for the first time, the default items are U1rms... U4rms,
I1rms... I4rms dependent on models. You can add up to 20 measurements to the list.
Edit the current item
Change the item selected at the current location.
Add measurements
Add measurements at the end of the list. Up to 20 measurements can be
added.
Delete the current item
Delete the item at the current location.
Move forward the currently selected item
Moves the currently selected measurement forward by one position in the list. If
the currently selected measurement is already at the top of the list, no action is
taken when the button is pushed.
Move backward the currently selected item
Moves the currently selected measurement forward by one position in the list. If
the currently selected measurement is already at the end of the list, no action is
taken when the button is pushed.
Note
The Analyzer can remember the User Item views. Even if the Analyzer is turned off, the list
defined last time will be automatically displayed when you enter the User Item view next time.
Add Measurement Items
1. From the main screen of Meter, select 3 (User Item) > 2 (Add).
All available measurement items are listed by categories in a tree structure. See Measurement
Calculation Method for details of measurement items.
2. For each measurement to add (up to 20 measurements):
a. Use 1 through 4to select a category.
b. Use the arrow (K LM N) keys to move the cursor over the corresponding item.
In most cases, there are 2 dialog boxes displayed in the measurement tree menu, the one on
the left lists measurement items, and the one on the right lists the supported channels or wiring
groups. The yellow block represents the cursor position, and the yellow box indicates the
corresponding branch in the current tree structure.
Note
Under Dual-Unit Interconnection mode, "Remote" option appears in this menu. See Dual-Unit
Interconnection for details.
c. Push S to add the current option to the end of the list.
If the left and right dialogs are displayed, first select the name of the measurement item in the
left dialog box, then select the channel or wiring group to be added in the right dialog box, and
finally push S to confirm.
Operation
3. Push J repeatedly to go back to the main screen of Meter.
Delete Measurement Items
1. From the main screen of Meter, select 3 (User Item).
2. For each item to delete:
a. Use the arrow (K LM N) keys to move the cursor to the measurement item you want to delete.
b. Push 3 (Del).
3. Push J to go back to the main screen of Meter.
Edit the Current Measurement Item
1. From the mains screen of Meter, select 3 (User Item) > 1 (Edit).
2. Use the arrow (K LM N) keys to move the cursor to the measurement item you want to edit.
3. Select the measurement item by following the operation similar to section Add Measurement Items.
Unlike the section Add Measurement Items, the measurement items added will directly replace the
measurement items at the cursor position, instead of adding the newly added measurement to the
end of the list.
Change the Position of the Measurement Item
You can sort the measurement items displayed in the User Defined view:
1. From the main screen of Meter select 3 (User Item).
2. To move each item:
a. Use the arrow (K LM N) keys to move the cursor to the measurement you want to move.
A settling time is required during the automatic range switching, and the sampled value during
this time is not accurate, which will cause integral error. Therefore, in order to obtain accurate
integration results, Fluke recommends setting the channel used for integration to manual
range, and the range should cover the instantaneous voltage and current.
Harmonic
In HARMONIC mode, the user can analyze the signals in the form of a table/bar/FFT.
Enter Harmonic Mode
•From the main screen of current measurement screen select k.
Table 15 lists the items displayed on the screen.
Table 15. Harmonic Screen
ItemDescription
Operation
Items
The name of the currently selected items depends on the options in Item.
See Select the Harmonics to Display for details.
Basic FrequencyThe fundamental frequency (Hz) of the reference signal (sync source).
Total Harmonic
Distortion
Fundamental Content
Harmonic Content
Total harmonic distortion of the voltage or current of the currently item, U
or I
. The unit is %.
thd
Fundamental content of the voltage or current of the currently selected
Item, U
or Ifc. The unit is %.
fc
Harmonic content of the voltage or current of the currently selected Item,
U
or Ihc. The unit is%.
hc
Harmonic
Voltage/Current
Harmonic voltage or current factor of the currently Item, hvf or hcf.
Factors
Table, Bar and FFTDisplays table, bar or FFT depending on the selected display mode.
1 (Configure): Global configuration, see Global Configuration.
2 (Item): Select the harmonic items displayed on the screen see Select
the Harmonics to Display.
3 (Mode): Harmonic display mode. You can push this button to cycle
through harmonic table, bar and FFT.
Soft Keys
4( /Axis): Page Up key or Axis Setup key. It is the Page Up key
under the table display mode, and it is the Axis Setup key under the bar
and the FFT mode. See Linear and Logarithmic Axis.
thd
5 (): Page Down, is used to scroll down the page to display more
content.
The harmonic bar is as shown in Figure 16 where the horizontal axis represents the harmonic order
and the vertical axis represents the harmonic amplitude in V or A.
The harmonic bar of each screen shows 50 harmonics. The pages can be turned forward or backward
through 5 (Next Page) button to display more harmonic bar.
The Phasor mode displays the magnitude and phase relationship of the voltage and current of each
phase in the selected wiring group in tabular and graphical form.
Enter Phasor Mode:
•From the Current measurement screen push p.
The phasor screen shown in Table 17 is divided into left and right parts, the left part is a phasor
table, and the right part is a phasor graph. The content displayed in the table on the left will be
displayed synchronously in the graph on the right.
Table 17. Phasor Screen
ItemDescription
Table
Voltage phasor information
The voltage phasor information of each phase or line in the selected wiring group,
including fundamental amplitude and phase angle relative to sync source.
See Set the Wiring Selection for information on wiring group.
Voltage unbalance
The 3-phase voltage unbalance of the currently selected wiring group.
40
Current phasor information
The current phasor information of each phase in the selected wiring group, including
fundamental amplitude and phase angle relative to sync source. See Select the Phasor
Data to Display for details. See Set the Wiring Selection for information on wiring group.
The 3-phase current unbalance of the currently selected wiring group.
Phase information
The relative phase angle between the voltage and current of each phasor in the
currently selected wiring group, the unit is degrees (°).
Frequency
The frequency of the sync source signal in the currently selected wiring group. See
Set Sync Source and Update Rate for details.
Wiring Selection
The wiring selection of the selected wiring group. The current interface data shows
the corresponding wiring group, and to switch the wiring group use 2 (Wiring
Group).
Voltage phasor
The voltage phasor of the selected wiring group. The right side of the horizontal axis
is the reference and the phase angle of the reference signal is 0°.
Current phasor
The current phasor of the selected wiring group. The right side of the horizontal axis
is the reference and the phase angle of the reference signal is 0°.
J
K
Voltage and current scale
For ease of reading, only the outermost circle scale is marked in the phasor graph.
1 (Configure): Global configuration, see Global Configuration.
2 (Wiring Group): Select a wiring group for measurement, see Set the Wiring
Selection.
3 (Y-Δ): Switch the circuit connection (Y/Δ). represents a star connection.
represents a triangle connection. See Select Circuit Type.
Only phasor data of one wiring group can show on the phasor screen at the same time. The user can
select a group from the currently active wiring groups to display its phasor data on the phasor screen.
Enter Wiring Group Menu:
•From the main screen of Phasor, select 2 (Wiring Group).
Select the Phasor Data to Display:
1. In Wiring Group, use the left and right arrow (M N) keys to select a channel or wiring group in Channel Configuration. The wiring selection of the selected channel/wiring group is highlighted.
As you move the cursor, you can observe that the data displayed in the
instantly according to the current options.
2. Push J to go back to the main screen of Phasor.
Note
The wiring group currently available for display depends on the wiring selection. See Set the
Wiring Selection for details.
Phasor screen is updated
Select Circuit Type
From the main screen of Phasor:
1. Select 3 (Y-Δ).
2. Push 3 (Y-Δ) to switch between the star (Y) and the delta (Δ) connection. The current
connection type is displayed on the tab page.
Note
The Y-Δ option is only valid when a 3-phase wiring group is selected. When Y is selected, the
phasor table displays the voltage of each phase, and the phasor diagram is plotted in the form
of Y. When Δ is selected, the phasor table displays the voltage of each line, and the phasor
diagram is plotted in the form of Δ.
Zoom Phasor
In the Phasor screen, the Analyzer automatically determines the magnitude of each phasor in the
phasor diagram based on the current measurements. However, the user can also manually zoom the
phasor diagram of the voltage and current for ease of viewing.
Enter the Zoom Menu:
1. From the main screen of Phasor select 5 (Zoom).
The options are as shown in Table 18.
2. Push 1 (
3. Push 3 () or 4()to zoom in or out the currently selected phasor.
4. To restore to default Auto Zoom, push 5 (
5. Push J to go back to the main screen of Phasor.
Select the Phasor to Zoom
Push 1 (ZOOM) to cycle through the currently selected phasor, namely U
ZOOM
U/I
or I.
The currently selected phasor shows under the “ZOOM” softkey, such as ,
which means the currently selected phasor is voltage.
Zoom in/out phasor
Power Analyzers
Operation
3
4
5Auto
Each time you push 3 (zoom in) or 4(zoom out), the currently selected
phasor is zoomed in or out by a certain step.
The steps to zoom in/out the coefficient are: 1, 2, 5, 10.
Auto zoom
Push 5 to enter AUTO mode, and the system automatically determines the
most appropriate display scale. All manual zooming settings are ignored.
Auto zoom features:
•The largest phasor does not exceed the outermost circle in polar
coordinates
•Clearly display each phasor direction point in polar coordinates
•Clearly display the name of each phasor
Scope
Use Scope mode to view the waveform of the signal under measurement. The supported signals
include voltage and current for example. Up to 4 signals can be selected simultaneously for waveform
display.
In addition, the Product provides cursor measurement function that can be used for various
measurements or calculations of waveforms. See Measure with The Cursor for details.
Enter Scope Mode:
•From the Current measurement screen, push q.
The Scope screen is shown in Table 19.
•The difference between the 2 cursors is displayed when the double
horizontal cursors are activated.
11
Difference
between cursors /
average at cursor
Value at cursor
/time difference
•The average value of the cursor position is displayed when the single
vertical line cursor is activated.
•The difference between the 2 cross points of the 2 cursors and
waveform is displayed when the double vertical cursors are activated.
•If the cursor is off, no values are displayed in this area.
See Measure with The Cursor for details.
•The values of the 2 cursors show separately when the double
horizontal cursors are activated. HIGH: Value of the upper cursor.
LOW: Value of the lower cursor.
•The maximum and minimum values of the cursors show when the
single vertical cursor is activated. MAX: Maximum value. MIN:
Minimum value.
•The time difference (t) between 2 cursors and the frequency value
(1/t) calculated accordingly are shown when the double vertical
cursors are activated.
•If the cursor is off, no values are shown in this area.
Up to 4 waveforms can be displayed simultaneously on the screen, and the user can select 1 to 4 items
from all available items for display.
Enter the Signal Menu:
•From the main screen of Scope select 2 (Signal).
The Signal menu lists all currently available parameters, including voltage and current for
example.The specific parameters depend on the current system configuration. See Set the Wiring
Selection.
Select the Waveform to Display:
1. In the Signal menu, select up to 4 parameters:
a. Use the arrow (K L M N) keys to move the cursor over the corresponding item.
b. Push S, the selection box of the current item turns yellow and a black “×” symbol is
appears, indicating the selection. To remove a selection, push S again, the selection box
will be cleared.
When a parameter is selected or cleared, the state is reflected in the waveform screen.
2. Push J to go back to the main screen of Scope.
Note
If 4 signals are selected, all other signals are grayed out. When you push the arrow (K L M N)
keys, the cursor can only move among the selected 4 signals. To select a different signal, you
must clear 1 of the 4 signals and then add a new one.
Trigger Setup
Use Trigger to determine when the Analyzer should start displaying waveforms on the screen so that
the waveforms displayed by each refresh can be aligned. Once the set trigger condition is met, the
screen is triggered, the waveform displayed on the screen is aligned at that point.
The Analyzer offers a variety of trigger methods to support multiple trigger sources and trigger modes.
Enter the Trigger Setup Menu:
•From the main screen of Scope select, 3 (Trigger).
The options are as shown in Table 20.
Table 20. Trigger Setup
ItemDescription
Push 1 (Source) to cycle through the current trigger sources.
The currently selected trigger source is shown under the “Source” label
1Source
with a color block of the same color as the waveform.
The selectable trigger sources are the selected signal in the “Signal”
menu, which is the signal displayed on the “Vertical Axis Scale” of the
screen, see Select the Waveform to Display.
•Method: options are Auto, Normal and Single Shot. The default is
2Trigger Option
4
5
If the trigger level is not set properly, under the “Auto” mode, the starting point of the waveform
displayed (signal level at the left side of the screen) may be unstable, it will jitter left and right.
The waveform may not be displayed at all under “Normal” mode.
Auto mode.
•Slope: options are Positive, Negative and Dual. The default
is Positive.
Trigger Level
The up and down arrow keys increase or decrease the trigger level. Each
time the key is pushed, the trigger mark on the screen moves in the
up/down direction.
Note
Operation
Trigger Setup:
1. From the main screen of Scope select 3 (Trigger).
2. Push 1 (Source) until the block with the same color as the waveform of the trigger signal is
shown.
3. Push 2 (Method).
4. Use the up and down arrow (K L) keys to move the cursor to Type and push S.
5. Use the up and down arrow (K L) keys to select the appropriate option and push S.
6. Use the up and down arrow (K L) keys to move the cursor to Slope and push S.
7. Use the up and down arrow (K L) keys to select the appropriate option and push S.
Use the cursor to measure the waveform digitally.
Enter the Cursor Setup Menu:
•From the main screen of Scope select 4(Cursor).
The options are as shown in Table 21.
Table 21. Cursor Setup
ItemDescription
Select the measured signal
Push 1 (Select) to cycle through the current measured signal.
1Select
2Type
3Move
The currently selected signal is shown under the Select label with a color block
of the same color as the signal waveform.
The selectable signals are the selected signal in the Select menu, which is the
signal shown on the Vertical Axis Scale of the screen, see Select the
Waveform to Display.
Measurement type
Push 2 to cycle through and select the cursor measurement types. Cursor
Measurement types include single vertical line, double vertical lines and double
horizontal lines.
• (single vertical line)
•(double vertical lines)
•(double horizontal lines)
Move the cursor
Each time you push 3, the currently selected cursor will be cycled through.
The selected cursor is indicated in yellow and is displayed below the Move
label.
After selecting a cursor, use the arrow (K L M N) keys to move the cursor to the
corresponding position on the waveform.
4Cursor
Use the Cursor
The following content uses the double vertical lines as an example to introduce the use of the cursor,
which is similar to the use of the cursor of a single vertical lines and double horizontal lines.
1. From the main screen of Scope select 4(Cursor).
2. Push 4(Cursor) until “ON” is shows on the label.
Turn the cursor on/off
Push 4to turn the cursor function on or off.
www.GlobalTestSupply.com
Power Analyzers
Operation
4. Push 1 (Select) to select the signal to be measured.
The difference between the crossing points of the two cursors and the signal under measurement is
shown in the upper left corner of the screen, such as “204.7 V”. The time difference (426 ms)
between the two cursors and the frequency (2.347 Hz) calculated are displayed in the upper part of
the screen.
5. Push 3 (Move) until shows on the label, indicating that the left cursor is selected and a
diamond mark is displayed on the left cursor.
6. Use the left and right arrow (M N) keys to move the left cursor to the appropriate position.
7. Push 3 (Move) until shows on the label, indicating that the right cursor is selected and a
diamond mark is displayed on the right cursor.
8. Use the left and right arrow (M N) keys to move the right cursor to the appropriate position.
9. Read the difference, time difference and frequency value at the top of the screen.
Zoom and Shift the Waveform
The Scope supports horizontal and vertical zooming and shifting. You can select a singe waveform
independently for vertical zooming and shifting and the waveforms are not affected by each other.
Enter the Zoom Menu:
•From the main screen of Scope select 5 (Zoom/Shift).
The options are as shown in Table 22.
Table 22. Zoom In/Out the Waveform
ItemDescription
Push 1 (Select) to cycle through the current selected signal.
The currently selected signal is displayed under the “Select” label with a
1Select
2DIR
color block of the same color as the signal waveform.
The selectable signals are the selected signal in the “Select” menu, which
is the signal displayed on the “Vertical Axis Scale” of the screen, see
Select the Waveform to Display.
Zoom direction
Push 2 to cycle through and select the zoom direction. Horizontal and
vertical zooming of waveforms are supported. The currently selected zoom
direction is displayed below the “DIR” softkey label, indicated by a
horizontal or vertical two-way arrow.
• (horizontal): Zooms the time axis of all signal waveforms in step of 1,
2, 5 and 10.
• (Vertical): Zooms the currently selected signal vertically.
Each time you push 3 (zoom in) or 4(zoom out), the waveform is
zoomed in or out in a certain step in the horizontal or vertical direction.
(Vertical zooming will be applied to the selected waveform. Horizontal
zooming will be applied to all waveforms currently displayed.)
Auto zoom
Push 5 to enter the Auto Zoom mode, the system automatically zooms
all displayed waveforms to the appropriate scale. All manual zooming
settings are ignored.
Shift waveform
In the menu, use the up and down arrow (K L) keys to move the currently
selected waveform up and down on the screen.
Trend
The Trend mode displays and records the trend of numerical measurements. Any numerical
measurement item can be displayed as a trend, up to 4 trend curves can be displayed simultaneously
on the screen.
When the trend function is activated, the Analyzer continuously records the digital reading of the
measurements and displays it as a graph. Like a paper chart recorder, the trend graph can be scrolled
from right to left.
The bottom of the screen shows the recording time since it starts.
Enter Trend Mode:
•From the Current measurement screen, push t.
The Trend screen is as shown in Table 23.
You can select 1 to 4 items from all available items to show on the screen.
Enter the Item menu:
•From the main screen of Trend select 2 (Item).
The Item menu is as shown in Table 24.
Table 24. Select the Trend to Display
ItemDescription
1Display
2AddAdd items to plot trend curves and up to 4 collection items can be added.
3DelDelete the trend item that is currently added.
Display/hide Trend Curve:
1. From the main screen of Trend select 2 (Item) > 1 (Display).
2. In Display dialog, use the up and down arrow (K L) buttons to move the cursor over the trend
curve you want to display/hide.
3. Push S to select or deselect the current item depending on its previous state.
When a trend curve is selected or deselected, it will immediately be reflected in the trend screen to
display or hide the trend curve.
4. Push J to go back to the main screen of Trend.
The Display menu is equivalent to a buffer that can hold up to 4 items. Use 2 (Add) and
3 (Del) to add the 4 most commonly used items to the buffer. Then select the trend data to
show/hide the trend curve through the Display menu. This avoids frequently using the
cumbersome 2 (Add) and 3 (Del) keys to switch the displayed trend curve.
Set the trend curve currently displayed. You can choose from up to 4 trend
curves. All 4 trend curves are displayed by default.
Note
Add Trend Item:
1. From the main screen of Trend select 2 (ITEM) > 2 (Add).
The Add menu is shown. All available items are listed by categories in a tree structure. The
available trend items contain the most measurement items under Meter function.
2. Use 1 to 4button to enter the category.
3. For each item to add (up to 4):
a. Use the navigation arrow (K L) keys to move the cursor over the corresponding item.
In most cases, there are 2 dialog boxes displayed in the measurement tree menu, the one on
the left lists the names of items, and the one on the right lists the supported channels or wiring
groups. The yellow block represents the cursor position, and the yellow box indicates the
corresponding branch in the current tree structure.
b. Push S to enter its sub-menu (if any).
c. Use the arrow (K L) keys to move the cursor over the corresponding item.
d. Push S to add the current item. The item will be highlighted in yellow.
After the item is added successful, you will be straight back to the top-level parameter menu.
4. Push J repeatedly to go back to the main screen of Trend.
Note
If there are already 4 trend items in the “Display” menu, you cannot add a 5th item regardless
of whether they are all displayed on the trend screen. You must delete one or more trend items
and then add other items.
Delete Trend Item:
1. From the main screen of Trend select 2 (ITEM) > 3 (Del).
2. In “Del” dialog, use the up and down arrow (K L) buttons to move the cursor over the item to be
delete.
3. Push S to delete the current item from the Display menu.
4. Go back directly to ITEM screen.
Note
When deleting trend data, remove an item from the Display menu by using the Del menu even
if its trend curve is not displayed on the screen.
Set Measurement Period
The measurement period is the time interval used to plot the trends and provides two modes of Update
Rate and Interval. In the Update Rate mode, the interval of the trends is the update rate of the
Analyzer, which can achieve accurate evenly spaced acquisition, which is suitable for analyzing the
process of rapid change. In the Interval mode, the interval of the trends is determined by the trend
function itself, which is not as accurate as the Update Rate mode, but it can record the trends for a long
time.
Enter the Measurement Period Setup Menu:
•From the main screen of Trend select 3 (Interval).
The measurement period setup dialog is shown in Table 25.
Table 25. Measurement Period Setup
ItemDescription
By update rate
The measurement period is the same as the update rate of the Analyzer by
Update rate
default.
The update rate is a kind of global configuration of the Analyzer. See Set
Sync Source and Update Rate for information on update rate.
Customize sampling interval
You can customize sampling interval as measurement period. The
minimum interval is 1 second and the maximum is 24 hours
Interval
Customize the Interval
1. From the main screen of Trend select 3 (Interval).
2. If Update Rate is selected, use the up and down arrow (K L) keys to move the cursor to Interval.
Push S. The Interval option is selected.
•H: Customize the Hour value of the interval.
•M: Customize the Minute value of the interval.
•S: Customize the Second value of the interval.
3. Use the up and down arrow (K L) keys to move the cursor to
4. Use the left and right arrow (M N) keys to adjust the Hour value.
5. Push S.
6. Use the left and right arrow (M N) keys and S to set Minutes and Seconds.
7. Push J to go back to the main screen of Trend.
H (Hour) input box.
Zoom and Shift the Trend
The Trend supports vertical and horizontal shifting as well as vertical zooming. Each trend graph can
be shifted and zoomed vertically independently, and they are not affected by each other.
Enter the Zoom Menu:
•From the main screen of Trend select 5 (Zoom/Shift).
The options are as shown in Table 26.
Table 26. Zoom the Trend
ItemDescription
Select the trend
Push 1 (Select) to cycle through the currently selected trend.
The currently selected trend is displayed under Select with a color
block of the same color as the trend curve.
The available trend is the selected items in the ITEM menu, which is
the items displayed on the Vertical Axis Scale of the screen, see
Select the Trend to Display.
Zoom in/out waveform
Each time you push 3 (zoom in) or 4(zoom out) button, the
selected waveform will be zoomed in or out vertically by a certain
step.
www.GlobalTestSupply.com
Table 26. Zoom the Trend (cont.)
ItemDescription
Restore auto zoom
5Restore
When this button is pushed, it will restore to the display scale
automatically determined by the system.
Shift waveform
Power Analyzers
Operation
K LM N
In the interface of this menu, use the navigation arrow (K L M N) keys
to shift the currently selected trend up and down on the screen, and
the left and right movement will be applied to all displayed trends.
Data Management
The Analyzer can save the measurement data and the current configuration of the Analyzer in the
internal memory, or log the data displayed on the current screen in Meter mode for a long time. The
data can be downloaded to a PC by using the Fluke Power Analyzer software. For more information,
please refer to the instructions for using the PC software.
Use mto manage the data stored in the internal memory.
Save Data
On any measurement screen, push s and the Save dialog box will show on the screen.
•Push 1 (Data) to save all valid measurements at the current time to the internal memory. When
it is saved successfully, a prompt message containing the name of the file will show on the screen.
•Push 2 (Settings) to save the current system settings and the configuration of all measurement
functions. When the save is successful, a message containing the name of the file shows on the
screen.
Note
The Analyzer automatically creates the file name based on the category of the saved data and
system time. To modify the file name, after it is saved, push m and follow the prompt.
(See Memory)
Logger
Enter the Logger Menu:
•In Meter mode, push and hold sfor more than 2 seconds, to show the Logger menu on the
screen.
The functions in the Logger menu are as shown in Table 27
1. Use the up and down arrow (L K) keys to move the cursor to Interval. Push
option is selected.
2. Use the up and down arrow (L K) keys to move the cursor to H (Hour) input box.
3. Use the left and right arrow (N M) keys to adjust the Hour value.
4. Continue using the arrow keys to adjust the Minute and Second value.
5. Push 1 (Confirm) to take the adjustment into effect and back to main screen of Logger menu.
6. Push 2 (Cancel) to go back to the main screen of Logger menu without any change.
Table 28. Logger Interval
ItemDescription
By update rate (default)
The logger interval is the update rate of the Analyzer.
Update rate
The update rate is a kind of global configuration of the Analyzer. See Set
SyncSet Sync Source and Update Rate for information on update rate.
, the Interval
Customize acquisition interval
Acquisition interval can be customized, the minimum interval is 1
second and the maximum is 24 hours
Interval
During logging, the user cannot switch screens and perform other measurement functions. If
the user tries to perform other operations, the Analyzer will prompt you to stop the logger.
The logger function is available in the Meter mode only. In other modes, when you push and
hold s, the Analyzer will show a message that indicates the Logger is only available in
Meter mode.
Even in Meter mode, the logger function is not supported if the user freezes the display with
h. You must exit the Hold mode first before you can use the Logger function.
Memory
1. Push m to enter the
The available file types are shown in Table 29.
1DataView and manage saved data files.
2Logger SessionsView and manage Logger items.
3Screen CapturesView and manage saved screenshot files.
4SettingsView and manage saved settings files.
The available file actions are shown in Table 30
3. Select the action. To delete one file, rename, or playback, select the file and then select the action.
Table 30. Memory File Operation
ItemDescription
1DelDelete a file
2Del AllDelete all files of the current type.
The Analyzer uses an automatic naming method when saving files. With this
3Rename
function key, the user can rename each file to a name that is easy to
remember.
Page up key
4
5
K L M N
When you select “Delete”, the Analyzer deletes the selected files or deletes all files of the
current category without any prompts. The files deleted by these operations cannot be restored
locally. Please operate with caution.
When the number of files available exceeds the number of items displayed
on each screen, the and keys can be used to scroll forward or
backward to display more files.
Page down key
Page down key for scrolling backward to display next page.
Playback
Use the navigation arrow (K L M N) keys to playback the selected file.
The Analyzer supports Dual-Unit Interconnection, that is, 2 Analyzers work synchronously in host/slave
mode to realize 6-channel or 8-channel configuration.
2 Analyzers are connected in a point-to-point manner by using the RS-485/sync adapter provided with
the Analyzer to connect the RS-485 communication ports on the right side of the 2 Analyzers, which
are 3-pin connector sockets, as shown in the Figure 19.
For the 2 Analyzers, one works as a host and the other works as a slave.
Figure 19. Dual-Unit Interconnection
Note
The host can only add user defined measurement items that have been added in the slave.
See User Item View for more information on user defined measurement items.
Dual-Unit Interconnection Measurement require specific RS-485 communications parameters. These
are:
•Baud Rate = 921600,
•Parity = None,
•EOL = CRLF,
•Data Bit = 8,
•Stop bits = 1.
After the 2 Analyzers are connected through the RS-485 port, follow these steps to make
measurements:
1. Set RS-485 communication parameters of the slave.
a. From the main screen in any operating mode of the slave, select r > 2 (Comm.), and
then select the RS-485 connection method. Follow the instructions in Set Communication
Parameters.
b. Push J repeatedly to go back to the main screen of the Meter.
2. Set RS-485 communication parameters of the host. On the host:
a. Push r> 2 (Comm.) and then select the RS-485 connection method. Follow the
instructions in Set Communication Parameters.
b. Push r> 3 (Remote) to enter the Remote menu.
The host’s screen prompts that the remote device is being searched, and the host will
automatically scan the connected device.
After the Analyzer as a slave is searched by the host the connected device shows in the
Available devices list, such as Address 2.
c. Push 1 (Connect) to enter the Remote menu. If the connection succeeds, the label for 2
changes to Disconnected.
d. Push J repeatedly to go back to the main screen of the System.
e. Follow the instructions in section User Item View to enter the interface of Add Measurement
Items. Notice that 5 (Remote) is added to the category of Add measurement items.
f.Push 5 (Remote).
All user defined measurement items selected on the slave will be listed on the host’s screen.
g. Use the arrow (K L M N) keys to move the cursor to the corresponding measurement item and
push S.
The corresponding measurement items show in the list of user defined measurement items on
the host (the measurement item name has the suffix "*").
3. To exit Dual-Unit interconnection mode:
On the host, push r > 3 (Remote) > 2 (
Or
On the slave, push 1 (Local).
If the measurement item is not listed in the Remote interface on the host, exit the Dual-Unit
interconnection mode and add the measurement items on the slave. See User Item View for
details.
Note
Under the Dual-Unit interconnection mode, the slave will be locked for remote operation and
the buttons on the control panel will be locked (except for (1 (Local)) and no other
operations can be performed.
Connect the Computer
The Fluke Power Analyzer software is a software that communicates between the Analyzer and PCs
through RS-485 or mini-USB interfaces. With this software, data can be controlled and summarized
simply and quickly on the PC.
Figure 20. Connect the Computer
With the Fluke Power Analyze software:
•The PC screen shows the measurements, waveforms, phasors, harmonics and trends from the
Analyzer Select from all measurements such as voltage, current, power and harmonics.
•Instrument settings: settings of the connected Analyzer can be changed in the PC software.
•Files in the Analyzer memory can be downloaded to a PC and analyzed.
•Operating system required: Windows 10 / Windows 8 / Windows 7 (32bit/64bit).
For more information about using the Analyzer with the PC software, refer to the instructions for using
the PC software.
System Settings
The System Settings function provides the system level setup options, including general settings,
communication parameter settings, local/remote control, as well as the test of the Analyzer itself.
Enter the System Settings Function:
•From the main screen in any operating mode push r.
All information listed in the work area of the System screen is reference information and cannot be
changed. This information is useful for troubleshooting and maintenance.
Note
Table 31. System Information
ItemDescription
DescriptionProduct name, such as Power Analyzer.
Model NameProduct model, for example Norma 6004+
Serial No.
Calibration Date
Firmware VersionThe version number of the internal firmware of the Product.
DSP Version
FPGA VersionThe version number of the FPGA in the Product.
Soft Keys
The unique serial number of the product, which may be requested when
Last calibration date of the Product. See Calibration for information on
calibration.
The version numbers of the Digital Signal Processing (DSP) device in the
Product.
1 (Settings): Sets general options for the Analyzer, see General
Settings.
2 (Comm.): Sets the serial communication parameters of the Analyzer,
see Set Communication Parameters.
3 (Remote): Connects remote analyzer, see Remote.
4(Maint.): Diagnoses and tests the LCD and buttons/keys of the
The Settings screen contains some general instrument settings such as interface language, time and
date, date format, log storage settings, as well as firmware upgrades and factory resets.
Enter the Settings screen:
•From the Main screen in any operating mode select r > 1 (Settings).
The options listed on the Settings screen include:
•Language
•Time date
•Date format
•Auto-off
•Firmware upgrade
•Factory default
•Memory
See Firmware Upgrade for the operation of firmware upgrade.
Language
Users can select the language supported by this unit.
Change UI Language:
1. In the Settings screen, use the up and down arrow (K L) keys to move the cursor to the
Language option (the drop-down menu is highlighted in yellow).
2. Push S and use the up and down arrow (K L) keys to select the corresponding option. Then
push S again to confirm the option.
3. Push J repeatedly to go back to the main screen of the Settings.
Set Time and Date
In the Settings screen:
1. Use the up and down arrow (K L) keys to move the cursor to the Date Time option (the drop-
down menu is highlighted in yellow).
2. In the Time Date dialog box, use the left and right arrow (M N) keys to select the year, month, day,
hour and minute, and use the up and down arrow (K L) keys to adjust the value.
3. Push J repeatedly to go back to the main screen of the Settings.
Auto-Off
If set to ON, the power turns off automatically after 30 minutes without operation, communication,
integral, or logging.
The user can select the operation mode of the Analyzer when the memory is full, which can be set to
stop the logging by the system, to manually clear log files saved in the Product by the user, or to
automatically overwrite the old files by the system.
1. In the Settings screen, use the up and down arrow (K L) keys to move the cursor to the Logger Memory option (the drop-down menu is highlighted in yellow).
2. Push S and use the up and down arrow (K L) keys to select the corresponding option. Then
push S again to confirm the option.
3. Push J repeatedly to go back to the main screen of the Settings.
Set Communication Parameters
The Analyzer supports communication via USB virtual serial port and RS-485. Host/slave
communication is realized by RS-485 communication in the dual-unit interconnection mode. For
details, see Dual-Unit Interconnection.
In order to ensure normal communication, the communication port of the Product needs to be set up
correctly.
Enter the Comm. Setup Interface:
•From the Main screen in any operating mode select r > 2 (Comm.).
The "Comm." settings screen is shown, the items that can be set for USB virtual serial port and
RS-485 port are as shown in Table 30.
Table 32. Communication Parameters
ItemOption
Connect
Parity
Baud Rate
USB (virtual serial) port
RS-485
O: Odd parity
E: Even parity
N: No parity
9600
19200
38400
57600
112500
230400
460800
576000
921600
To ensure proper communication between the host/slave Analyzer, you must configure both
Analyzers correctly. Please refer to Dual-Unit Interconnection.
LF
CRLF
7 Bits
8 Bits
1 Bits
1.5 Bits
2 Bits
Note
Remote
The Analyzer supports Dual-Unit interconnection. This menu allows 2 Analyzers to be configured in
host/slave mode for 6 or 8 channels configurations. See Dual-Unit Interconnection for details.
Instrument Maintenance
The Analyzer provides a self-diagnosis function that allows the user to test the button/key and display
screen of the Product.
Enter Diagnosis:
•From the Main screen in any operating mode select r > 4(Maint.)
Table 33. Diagnostics
ItemDescription
Push K, the screen will show alternately full-screen of solid colors of blue,
1LCD Test
2Keys Test
black, white, red and green.
Push J to go back to the test menu.
Test if each key is valid. All the keys on the control panel are shown on the
screen. The original function of all the keys on the control panel will be
disabled.
Each time a physical key on the control panel is pushed, the corresponding
key on the screen changes from white to yellow.
Pushing J will not take you back to the main menu. The screen will
automatically go back to the Maint. menu only after all physical
buttons/keys on the control panel have been tested.
The only required maintenance is to replace the battery regularly and to replace the fuse, if necessary.
Cleaning
Clean the case with a damp cloth and a weak soap solution. Do not use abrasives, isopropyl alcohol, or
solvents to clean the case.
Warning
Before cleaning the Analyzer, please be sure to unplug the power adapter/charger from
the power jack and disconnect all test leads.
Change the Battery
The battery is charged in the Analyzer, the battery can be changed if a freshly charged battery is
required for extended battery power operation or the battery is no longer keeping charge. Charge the
Analyzer every three months if the Analyzer is not in use.
Warning
To prevent electric shock, remove all probes, test leads and accessories before the
battery compartment cover is opened.
Do not use any battery which is not designed or recommended by Fluke for use with
the Product.
1. Turn off the Product, unplug the power adapter/charger from the power jack, and disconnect all test
leads.
2. Place the operating panel face down on a flat surface and fold the tilt stand or remove it from the
Product.
3. Use a Phillips screwdriver to remove the 2 screws from the fuse compartment cover and remove
the cover.
4. Insert a flat screwdriver into the gap to the right of the fuse holder and gently pry it until it is possible
to hold the fuse by one hand and take it out.
5. Install new fuses and reinstall and fix the fuse cover and stand in the reverse order of the above.
6. Check the circuit before reconnecting to ensure the current does not blow the fuse again.
3. During the upgrade, make sure that the battery is in good contact and can power the unit normally,
and the USB connection is reliable.
4. After the instrument is powered on, select r > 1 (Settings).
5. Use the up and down arrow (K L) keys to move the cursor to Firmware Upgrade (the drop-down
menu is highlighted in yellow).
6. Push S.
Note
If the USB memory is not inserted, the Product will prompt to insert the USB. Follow the steps
above to insert the USB memory. Or check if the USB memory is connected securely.
The upgrade will take a few minutes. Wait for the upgrade to complete. The Analyzer automatically
restarts after the upgrade.
Note
To ensure the success of the upgrade, the battery must be installed during the firmware
upgrade and the battery level must be at least 50 %. The power adapter must be connected.
Specifications
General Specifications
Size (HxWxL).............................................. 298 mm x 215 mm x 96 mm
Weight.........................................................3.5 kg
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is
necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a
low-voltage power supply network that supplies buildings used for domestic purposes. There may be potential
difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated
disturbances.
Caution: This equipment is not intended for use in residential environments and may not provide adequate
protection to radio reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test
object.
Korea (KCC).......................................................Class A Equipment
(Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user
should take notice of it. This equipment is intended for use in business environments and not to be used in homes.
USA (FCC) .........................................................47 CFR 15 subpart B.
This product is considered an exempt device per clause 15.103.
Power Analyzers
Specifications
Electrical Specifications
Modules
NORMA 6003..................................................3 Voltage + 3 Current
NORMA6003+ ................................................3 Voltage + 3 Current + 1 Motor
NORMA 6004 .................................................4 Voltage + 4 Current
NORMA 6004+ ...............................................4 Voltage + 4 Current + 1 Motor
Measurement Error Caused by the Channel Resistance
By wiring a circuit to match the load, you can minimize the effects of power loss on measurement
accuracy. The wiring of the source and load is discussed in this section.
I Rin=0.025Ω
iv+iL
iviL
+
Source
U Rin=2MΩ
Load
Figure A1. Connect the Voltage Measurement near the load
Connect the voltage measurement circuit near the load (as shown in Figure A1). The current
measurement circuit measures the sum of i
Connect the Voltage Measurement near the loadL
current) is the current flowing through the load of the circuit under measurement, and i
and iV. iL (desired
(error current)
V
is the current flowing through the voltage measurement circuit. Because the current flowing through the
circuit under measurement is i
voltage measurement circuit of the Analyzer is approximately 2 MΩ. If the input voltage is 1000 V, i
approximately 0.5 mA (1000 V/2 MΩ). Only when the load current i
, only iV reduces measurement accuracy. The input resistance of the
L
is 0.5 A or more, the effect of iV on
L
V
is
the measurement accuracy is 0.1% or less.
I Rin=0.025Ω
e
i
Load
e
L
Source
+
eLe
+
i
U Rin=2MΩ
Figure A2. Connect the Voltage Measurement near the source
Connect the voltage measurement circuit near the source (as shown in Figure A2). The voltage
measurement circuit measures the sum of e
and eI. eL (desired voltage) is the load voltage, and eI
L
(error voltage) is the voltage drop across the current measurement circuit. The input resistance of the
current measurement circuit of the Analyzer, I Rin is approximately 0.025 Ω. If the input current is 10 A,
the effect of e
on the measurement accuracy is approximately 0.25 V (10 A×0.025 Ω). Only when the
is 250 V or more, the effect on the measurement accuracy is less than 0.1 %. In
L
summary, when measuring low voltage and large current, it is recommended to use the wiring method
of the figure above; when measuring high voltage and small current, it is recommended to use the
wiring method of the figure below.
Effects of Leakage Capacitance
Each measurement channel of the Analyzer is isolated from each other. However, these isolated
channels still have a leakage capacitance C
capacitance of the instrument itself, the isolation capacitance of the power adapter, and the antiinterference capacitance inside the instrument. In general, the test object of the Analyzer has a higher
voltage relative to the earth, and we call this voltage common mode voltage V
voltage can generate a current i
through the leakage capacitance Cs. This current ics will have a
cs
certain impact on the measurement results of the Analyzer.
As shown in Figure A3, during voltage measurement, the current i
circuit through R
measurement and R
, Rs- and Rin. Rs+ and Rs- are internal resistances of the power supply under
s+
is the test internal resistance of the power analyzer. The effect on the
in
measurement voltage is: . Because R
the power supply under measurement R
terminal (black terminal) of the measurement circuit.
relative to the earth, which consists of the distributed
s
. The common mode
com
is coupled to the measurement
cs
is larger, internal resistance of
in
has the main impact. Rs- is connected to the common
s-
RinRs+
2M Ohm
U
Rs-
Ics
Vcom
Cs
∆U= Ics[Rs-//(Rs+ + Rin)]
Figure A3. Effects of Leakage Current on Voltage Measurement
During current measurement, the current i
flows through the test resistance R
cs
, and the internal
shunt
resistance of the test resistance and the common terminal of the measurement circuit R
voltage, that is coupled into the measurement circuit. The effect on the measurement
Figure A4. Effects of Leakage Current on Current Measurement
Methods to Reduce the Effect of Leakage Capacitance
1. High frequency components more easily generate current on the leakage capacitance than low
frequency components. When measuring the small current with the built-in current channel, if you
do not pay attention to the high-frequency components of the current, it is recommended to turn on
the 650Hz line filter to effectively reduce the effect of leakage capacitance.
2. Connecting the built-in current channel to terminal N or ground can effectively reduce the effect of
leakage capacitance.
3. Disconnecting the power adapter and using battery only can reduce the effect of leakage
capacitance.
4. When measuring voltage, if the 2 terminals of the voltage source under measurement do not have
the same impedance to the ground, it is recommended to connect the black terminal to the terminal
with lower impedance (to avoid leakage current flowing through the internal resistance of the
voltage source under measurement).
Broadband Measurement
The Analyzer has a bandwidth of up to 500 kHz and a sample rate of 200 kHz. When the measurement
bandwidth is greater than half of the sample rate, aliasing may be caused, the observed waveform is
not the real waveform, and the measured frequency is not the real frequency.
To measure a time-based digital quantity, such as RMS, AC and DC components, the results of the
measurement after aliasing are still accurate. However, note that if the signal frequency is close to or
equal to the sample frequency and its multiples, the measurement results are probably invalid.
If the signal under measurement has a frequency component higher than 100 kHz, and it is not desired
to be measured under aliasing, the line filter of the corresponding channel can be turned on according
to the required bandwidth.