Fluke 2063513, MULTIMETER WITH TEMPERATURE 16, 16 Instruction Sheet

Page 1
16
®
Instruction Sheet
P
To ensure that the meter is used safely, follow these instructions:
Do not use the meter if the meter or test leads appear damaged, or if
Disconnect the live test lead before disconnecting the common test
When using the probes, keep your fingers behind the finger guards on
Do not use the VCheck
Read First: Safety Information
you suspect that the meter is not operating properly.
lead.
the probes.
mode to measure voltages in circuits that
could be damaged by this mode's low input impedance (2 k).
Turn off power to the circuit under test before cutting, desoldering, or
breaking the circuit. Small amounts of current can be dangerous.
Do not apply more than 600V rms between a meter terminal and earth
ground.
Use caution when working with voltages above 60V dc or 30V ac rms.
Such voltages pose a shock hazard.
PN 2063513 December 2002 Rev.1, 9/04
2002,2004 Fluke Corporation. All rights reserved. Printed in China
Page 2
Symbols
Press button.
Press button to switch between modes.
Double insulation.
Manual ranging mode. ip14i.eps
Automatic Range Selection
The meter defaults to autoranging when you turn it on. The 4000 mV range can be entered only with manual range selection.
Manual Range Selection
MULTIMETER
16
To return to autoranging, press for
OFF
CAT
600V
COM
+
2 seconds, or change the measurement mode.
ip15i.eps
Battery Saver
If the meter is ON but inactive and not connected to voltage for more than 45 minutes, the display goes blank to preserve battery life. To resume operation, press any button.
Battery Saver is disabled in MIN/MAX record mode.
Page 3
AC and DC Voltage
Also refer to V•Check. Volts AC Volts DC
Input Impedance ≅5 MΩ Input Impedance ≅10 MΩ 50 Hz to 400 Hz
MULTIMETER
16
DCAC
4000 mV 4V 40V 400V 600V
OFF
V
AC / DC
CAT
600V
COM
+
Resistance and Continuity
eR
OFF
Turn off circuit power before testing. Also refer to V•Check.
MULTIMETER
16
V
AC / DC
CAT
600V
COM
+
ip01i.eps
Resistance
MULTIMETER
16
Continuity
Mk
OFF
COM
Detects shorts and opens 250 µS.
V¥Check
A
CAT
TEMPERATURE
600V
+
Low Impedance
V¥Check
MULTIMETER
16
OFF
COM
V¥Check
A
CAT
TEMPERATURE
600V
+
OFF
COM
Short Open
<25
MULTIMETER
16
CAT
600V
TEMPERATURE
+
V¥Check
A
ip02i.eps
Page 4
W
V•Check
If a dc or ac voltage greater than about 4.5V is present across the inputs when the meter is set to
R, G, or e, the meter switches automatically
to dc or ac voltage mode (VCheck mode).
W Warning
Repetitive transients on a dc bus will cause VCheck to select ac volts, even though a hazardous dc voltage may be present. To avoid a misleading display and possible electric shock, manually select the proper volts function for measurements on these circuits.
When VCheck is activated, the meter has low input impedance (LoZ) ≅2 k. This load can alter the voltages in electronic control circuits. Do
not use VCheck to measure voltage in circuits that could be damaged by a 2 k load.
Hint: VCheck can be effectively used to eliminate “ghost” voltages.
16
MULTIMETER
If AC detected If DC detected
Disable and re-enable V•Check
For , and , Disables V•Check and
Meter puts 2 k load on circuit.
OFF
COM
V•Check
A
CAT
TEMPERATURE
600V
+
or
locks meter in selected mode.
2 sec
Re-enables
or
or
V•Check
ip09i.eps
Page 5
Temperature
16
MULTIMETER
C F
OFF
COM
V•Check
A
CAT
TEMPERATURE
600V
+
Note correct connector polarity.
80BK
Integrated
Temperature
Probe (Type K)
Vent
or
Pipe
ip19f.eps
To meet stated accuracy, the 80BK temperature adapter must be at the same temperature as the meter.
To avoid possible electric shock, DO NOT apply thermocouple tip to any conductor that is greater than 30V AC, 42.4V pk, or 60V DC to earth.
W Warning
Page 6
Capacitance
E
Turn off circuit power; then disconnect and discharge the capacitor before measuring capacitance.
MULTIMETER
16
If the capacitor requires more discharging, diSC is displayed while the capacitor discharges.
OFF
COM
V•Check
A
CAT
TEMPERATURE
600V
+
1 µF 10 µF 100 µF 10,000 µF
+
Note correct probe polarity for polarized capacitors.
ip05i.eps
Page 7
Microamps
µA
Range 0 to 200 µA
MULTIMETER
16
µA AC µA DC
OFF
COM
Flame
sensor
V•Check
A
CAT
TEMPERATURE
600V
+
Control
module
probe
Example: Flame rectification circuit.
To measure flame rectification circuits:
1. Turn function switch to far right position.
ip20i.eps
2. Push select button 3 times to select µA.
3. Connect meter between flame sensor probe and control module.
4. Turn heating unit on and record µA measurement.
Page 8
Diode Test
G
Turn off circuit power before testing. For best results diodes should be measured out of circuit. Also refer to V•Check.
Good Diode
16
MULTIMETER
Typical voltage
MULTIMETER
16
for silicon diode
Low Impedance
V•Check
OFF
COM
V•Check
A
CAT
TEMPERATURE
600V
+
OFF
COM
V•Check
A
CAT
TEMPERATURE
600V
+
Single beep
ip03i.eps
Forward Bias Reverse Bias
Bad Diode
MULTIMETER
16
MULTIMETER
16
OFF
CAT
600V
COM
+
or
Shorted Open
V•Check
A
TEMPERATURE
OFF
COM
V•Check
A
CAT
TEMPERATURE
600V
+
ip04i.eps
Page 9
MIN MAX
µF
(Records the lowest and highest measurements)
VCheck, autoranging, and Battery Saver are disabled. Put the meter
in the proper range before entering MIN MAX.
When the reading changes more than about 50 digits, the meter gives a short beep. When a new minimum or maximum is recorded, the meter gives a long beep.
Enter
MIN MAX
Present Reading with
Input Change Alert
TM
Maximum Reading
Minimum Reading
2 sec
To Exit
ip10i.eps
Page 10
MIN MAX with Elapsed Time
µF
Records the hours and minutes between when MIN MAX was entered and the last high and low was recorded. OL is displayed for times longer than 99:59.
To enable the MIN MAX timer, hold down switch from OFF to either measurement mode.
Enter MIN MAX. View
Present Reading
Maximum Reading Maximum Reading
M while turning the rotary
Elapsed Time
Minimum Reading
Elapsed Time
2 sec
Minimum Reading Exit MIN MAX
ip11i.eps
Page 11
Disabling the Beeper
To disable the beeper for all modes, hold down r for 2 seconds while turning the meter on.
Continuity Capture
To set up the meter to capture intermittent shorts and opens, turn the switch to
GR, connect the leads to the circuit; then press M.
Captures transitions longer than 250 µs (1/4000th of a second).
Transitions after the first transition cause the meter to beep, but the display does not change.
To reset the display to the current condition, press
To exit, press M for 2 seconds, or change the measurement mode.
M.
ip12i.eps
Page 12
Maintenance
Clean the case with a damp cloth and detergent. Do not use abrasives or solvents.
Battery Replacement
Remove the test leads before disassembling the case.
C
O
M
6
0
0
V
+
+
Replacement Parts
ip13i.eps
Fluke TL-75 (Double-insulated leads) PN 855705 80BK (Integrated Temperature Probe) PN 1273124
Service and Parts
This meter should be serviced only by a qualified service technician. To locate an authorized service center, call:
USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500
Or, visit Fluke's Web site at www.fluke.com
.
Page 13
Specifications
Accuracy is specified for a period of one year after calibration, at 18°C to 28°C (64°F to 82°F) with relative humidity to 90%. AC conversions are ac-coupled, average responding, and calibrated to the rms value of a sine wave input. Accuracy specifications are given as follows:
±([% of reading] + [number of least significant digits])
Function Range Resolution Accuracy
-10°C to 400°C
Temperature
(Type K Thermocouple) -40°C to -10°C
Error does not include Type K Thermocouple errors.
Function Range Resolution Accuracy
4000 mV1
f
(50 to 400 Hz)
E
4.000V
40.00V
400.0V 600V
4000 mV1
4.000V
40.00V
400.0V 600V
14°F to 752°F
-40°F to 14°F
1 mV
0.001V
00.01V
000.1V 1V
1 mV
0.001V
00.01V
000.1V 1V
0.1°C or
0.1°C or
0.2°F
0.2°F
±(1.0% + 0.8°C)
typical
±(1.0% + 1.5°F) typical
±(5.0% + 1.5°C)
typical
±(5.0% + 3.3°F) typical
±(1.9% + 3) ±(1.9% + 3) ±(1.9% + 3) ±(1.9% + 3) ±(1.9% + 3)
±(0.9% + 2) ±(0.9% + 2) ±(0.9% + 1) ±(0.9% + 1) ±(0.9% + 1)
400.0
4.000 k
|
40.00 k
400.0 k
4.000 M
40.00 M
1.000 µF
10.00 µF
100.0 µF 10000 µF
ML 2.000V 0.001V ±(1.9% + 2)
1. The 4000 mV range can be entered only in manual range mode. Use the 4000 mV range with accessories.
2. The beeper is guaranteed to come on at <25 and turn off at >250. The meter detects opens or shorts ≥250 µs.
0.1
0.001 k
0.01 k
0.1 k
0.001 M
0.01 M
0.001 µF
0.01 µF
0.1 µF 1 µF
±(0.9% + 2) ±(0.9% + 1) ±(0.9% + 1) ±(0.9% + 1) ±(0.9% + 1) ±(1.5% + 3)
±(1.9% + 2) ±(1.9% + 2) ±(1.9% + 2)
1000 µF ±(1.9% + 2)
>1000 µF ±(10% + 90) typical
2
Page 14
Function Range Resolution Accuracy Burden Voltage
N
(50 Hz to 400 Hz)
M
0 to 200 µA0.1 µA ±(2% + 3 counts) <5 mV/µA
0 to 200 µA0.1 µA ±(1% + 2 counts) <5 mV/µA
Function
f
Overload
Protection
600V rms
1
>5 M <100 pF
Input Impedance (Nominal)
VCheck and LoZ = >2 kΩ <200 pF (ac coupled)
E
f 600V rms >60 dB at dc 50 or 60 Hz
E
J
L 600V rms 2.4-3.0V dc 2.400V dc
600V rms
600V rms >100 dB at dc, 50 or 60 Hz >50 dB at 50 Hz or 60 Hz
600V rms <1.5V dc <450 mV dc <1.5V dc
>10M <100 pF
VCheck and LoZ = >2 kΩ <200 pF
Common Mode Rejection
Ratio (1 k Unbalanced)
Open Circuit Test Voltage
Short Circuit Current
Normal Mode Rejection
Full Scale Voltage
To 4.0 M 40 M
2
2
J 600V rms <500 µA
L
1. 3 x 106 V Hz maximum
2. 2 k input impedance up to 50V. Impedance increases with input voltage to >300 k at 600V.
600V rms
0.95 mA (typical)
MIN MAX Recording Accuracy and Response Time
Specified accuracy of the measurement function ±12 digits in dc for changes >200 ms in duration (±40 digits in ac). Typical 100 ms response to 80%.
Example 1: This would mean ±1.2° when recording temperature. Example 2: This would mean ±1.2 µA when recording µA or ±12A if used with a
dc amp probe (with a mV input).
Page 15
MIN MAX Recording with Elapsed Time
Elapsed Time Resolution Accuracy
0 to 100 hours (99:59) 1 minute 0.3% typical
Maximum Voltage Between any Terminal and Earth Ground:
Display: 3 3/4-digits, 4000 counts, updates 4/sec Operating Temperature: -10°C to 50°C (14°F to 122°F) Storage Temperature:
Temperature Coefficient:
Relative Humidity: 0% to 90% (-10°C to 35°C; 14°F to 95°F)
Battery Type: 9V, NEDA 1604 or IEC 6F22 Battery Life: 650 continuous hours with alkaline
600V rms
-30°C to 60°C (-22°F to 140°F) indefinitely (to -40°C (-40°F) for 100 hrs)
(.1 x specified accuracy)/°C (<18°C or >28°C)
0% to 70% (35°C to 50°C; 95°F to 122°F)
450 continuous hours with carbon-zinc
Shock: 1 meter drop 6 sides. Vibration: Class II vibration MIL-PRF-28800F Size (H x W x L): 3.46 cm x 7.05 cm x 14.23 cm
(1.35 in x 2.75 in x 5.55 in)
Weight: 286g (10 oz) Safety:
EMI Regulations:
Certifications:
Designed to Protection Class II requirement of UL3111, ANSI/ISA-S82, CSA C22.2 No 231, and VDE 0411, and IEC 1010 overvoltage Category III (CAT III, 600 Volts).
Complies with FCC Part 15, Class B, and VDE 0871B. Trademark of TÜV Product Services. Complies with EN 61010-1:
1993.
P $
TUV, UL and VDE
Page 16
LIMITED WARRANTY & LIMITATION OF LIABILITY
Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is three years and begins on the date of shipment. Parts, product repairs and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries or to any product which, in Fluke's opinion, has been misused, altered, neglected or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center or send the product, with a description of the difficulty, postage and insurance prepaid (FOB Destination), to the nearest Fluke authorized service center. Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that the failure was caused by misuse, alteration, accident or abnormal condition of operation or handling, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
Fluke Corporation Fluke Europe B.V. P.O. Box 9090 P.O. Box 1186 Everett WA 5602 B.D. Eindhoven 98206-9090 The Netherlands
5/94