Fluke Hart Scientific 1575 Super-Thermometer, Hart Scientific 1590 Super-Thermometer II User Manual

Page 1
Hart Scientific
1575/1590
Thermometer Readout
User’s Guide
Rev. 5B2802
Page 2
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • Telefax: +1.801.763.1010 E-mail: support@hartscientific.com
www.hartscientific.com
Subject to change without notice. • Copyright © 2005 • Printed in USA
Rev. 5B2802
Page 3

Table of Contents

1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.1 WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.2 CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Authorized Service Centers. . . . . . . . . . . . . . . . . . . . . . 3
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1 General Description . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2.1 Measurement Technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2.2 Performance Issues . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2.2.1 Lead Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2.2.2 Thermoelectric EMF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.2.2.3 Reactance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.2.4 Leakage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.2.5 Self-heating. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.2.6 Component Drift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.2.7 Noise and Resolution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.2.8 Nonlinearity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.2.9 Measurement Speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.2.10 Solid-State Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
3 Specifications and Environmental Conditions . . . . . . . . . 13
3.1 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Environmental Conditions. . . . . . . . . . . . . . . . . . . . . . 15
3.3 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2 Learn About the Features and Components . . . . . . . . . . . . . 17
4.3 2575/2590 Scanner Setup . . . . . . . . . . . . . . . . . . . . . . 17
4.4 Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.5 Probe Connection . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.6 Switch the Power On . . . . . . . . . . . . . . . . . . . . . . . . 18
4.7 Measure Temperature . . . . . . . . . . . . . . . . . . . . . . . . 18
5 Parts and Controls . . . . . . . . . . . . . . . . . . . . . . . . 21
5.1 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.2 Back Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
i
Page 4
5.3 Front Panel Display . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.4 Front Panel Buttons . . . . . . . . . . . . . . . . . . . . . . . . . 25
6 General Operation . . . . . . . . . . . . . . . . . . . . . . . . 29
6.1 Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6.2 Suggestions for Best Results . . . . . . . . . . . . . . . . . . . . 30
6.2.1 Maintaining Accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.2.2 Evaluating Temperature Uncertainty . . . . . . . . . . . . . . . . . . . . . . 32
6.3 Answers to Some Common Questions . . . . . . . . . . . . . . . 33
7 Front Panel Operation . . . . . . . . . . . . . . . . . . . . . . 37
7.1 Channel Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7.1.1 Measure Input 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7.1.2 Measure Input 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.1.3 Alternate Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.2 Sample Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.2.1 Sampling Run/Stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.2.2 Sample N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.2.3 Sample Timing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.2.4 Digital Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
7.2.5 Reset Statistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.3 Memory Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.3.1 Store Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.3.2 Write Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.3.3 View Memory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.3.4 Clear Memory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.3.5 Run/Pause Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.4 Probe Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.4.1 Select Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.4.2 Select Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.4.3 Edit Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.4.3.1 Serial Number. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.4.3.2 Temperature Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.4.3.2.1 ITS-90 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.4.3.2.2 IPTS-68 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
7.4.3.2.3 Callendar-Van Dusen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
7.4.3.2.4 Polynomial . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.4.3.2.5
Thermistor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.4.3.2.6 Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.4.3.2.7 W(T
7.4.3.2.8 Ratio. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.4.3.3 Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.4.3.4 Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.4.3.5 Standby Current. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.4.3.6 Three-wire Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.4.4 Calibrate Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7.4.4.1 Calibrate TPW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
7.4.4.2 Calibrate ITS-90 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
7.4.4.2.1 Calibration Using Only Fixed Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.4.4.2.2 Comparison Calibration With Manual Sequencing . . . . . . . . . . . . . . . . . . . . . 56
7.4.4.2.3 Comparison Calibration With Automatic Sequencing . . . . . . . . . . . . . . . . . . . 56
7.4.4.2.4 Calibration Using SPRT Resistance Comparison. . . . . . . . . . . . . . . . . . . . . . 57
7.4.4.2.5 Calibration With W Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
90
ii
Page 5
7.4.4.2.6 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
7.4.4.2.7 Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
7.4.4.2.8 Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
7.4.4.2.9 Calculate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
7.4.4.2.10 Print Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
7.4.4.3 1.414 x Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7.4.4.4 0.707 x Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7.4.4.5 Conversion Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7.4.5 Probe Disk. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7.4.5.1 Save Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
7.4.5.2 Read Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
7.4.5.3 Save All . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
7.4.5.4 Read All. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.4.5.5 Format Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.5 Display Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.5.1 Data/Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
7.5.2 Select Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
7.5.3 Edit Display - Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
7.5.4 Edit Display - Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
7.5.5 Display Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
7.5.6 Clear Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
7.6 System Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
7.6.1 Unit Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
7.6.2 Parameter Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
7.6.2.1 Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
7.6.2.2 Save Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
7.6.2.3 Load Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
7.6.2.4 Reset Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
7.6.2.5 Screen Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
7.6.3 Disk Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
7.6.3.1 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
7.6.3.2 Record/Stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
7.6.3.3 Record N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
7.6.3.4 Format Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
7.6.3.5 Run Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
7.6.4 Communications Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
7.6.4.1 Serial . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
7.6.4.2 IEEE-488 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
7.6.4.3 Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
7.6.4.4 Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
7.6.4.4.1 Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
7.6.4.4.2 Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
7.6.4.4.3 Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
7.6.4.4.4 Center . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
7.6.4.4.5 Cal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
7.6.5 System Calibration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
7.6.5.1 Set External . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
7.6.5.2 Set Internal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
7.6.5.3 Calibrate Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95
7.6.5.4 Security Lock-out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
7.6.5.5 Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
8 Communications Interface. . . . . . . . . . . . . . . . . . . . 99
8.1 Serial Communications . . . . . . . . . . . . . . . . . . . . . . . 99
8.1.1 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
8.1.2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
8.1.3 Serial Communication Example Program. . . . . . . . . . . . . . . . . . . 100
8.2 IEEE-488 Communications . . . . . . . . . . . . . . . . . . . . 101
iii
Page 6
8.2.1 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
8.2.2 IEEE-488 Interface Functions . . . . . . . . . . . . . . . . . . . . . . . . . 101
8.2.2.1 Local Lockout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
8.2.2.2 Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
8.2.2.3 Serial or Parallel Poll . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
8.2.2.4 Device Clear . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
8.2.3 IEEE-488 Communication Example Program . . . . . . . . . . . . . . . . 102
8.3 Digital Interface Commands . . . . . . . . . . . . . . . . . . . . 103
9 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
9.1 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
9.2 Resistance Ratio Calibration . . . . . . . . . . . . . . . . . . . . 110
9.3 Reference Resistor Calibration. . . . . . . . . . . . . . . . . . . 110
10 2575 Scanner (optional). . . . . . . . . . . . . . . . . . . . . 113
10.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
10.2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
10.3 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
10.3.1 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
10.3.1.1 Resistance Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
10.3.1.2 Control Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
10.3.1.3 Resistance Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
10.3.2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
10.4 Parts and Controls . . . . . . . . . . . . . . . . . . . . . . . . . 116
10.5 Scanner Operation . . . . . . . . . . . . . . . . . . . . . . . . . 119
10.5.1 Soft-key Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
10.5.2 Channel Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
10.5.2.1 Scanner Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
10.5.2.2 Scanner Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
10.5.3 Probe Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
10.5.3.1 Select Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
10.5.4 Digital Interface Commands. . . . . . . . . . . . . . . . . . . . . . . . . . 123
11 2590 Scanner (optional) . . . . . . . . . . . . . . . . . . . . 125
11.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
11.2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
11.2.1 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
11.3 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
11.3.1 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
11.3.1.1 Resistance Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
11.3.1.2 Control Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
11.3.1.3 Resistance Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
11.3.2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
11.4 Parts and Controls . . . . . . . . . . . . . . . . . . . . . . . . . 128
11.5 Scanner Operation . . . . . . . . . . . . . . . . . . . . . . . . . 129
11.5.1 Channel Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
11.5.1.1 Scanner Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
iv
Page 7
11.5.1.2 Scanner Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
11.5.2 Probe Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
11.5.2.1 Select Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
11.5.2.2 Standby Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
11.5.3 Digital Interface Commands. . . . . . . . . . . . . . . . . . . . . . . . . . 133
12 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
13 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . 137
13.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . 137
13.2 CE Comments (1590 Only) . . . . . . . . . . . . . . . . . . . . 140
13.2.1 EMC Directive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
13.2.1.1 Immunity Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
13.2.1.2 Emission Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
13.2.2 Low Voltage Directive (Safety) . . . . . . . . . . . . . . . . . . . . . . . . 141
v
Page 8

Figures

Figure 1 Simplified Schematic Diagram of the Measurement Circuit . . . . . . . 6
Figure 2 Measurement Processing Operations . . . . . . . . . . . . . . . . . . . 7
Figure 3 1575/1590 Super-Thermometer Display . . . . . . . . . . . . . . . . . 24
Figure 4 Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Figure 5 Sample Timing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Figure 6 Smoothing Effect of the Digital Filter . . . . . . . . . . . . . . . . . . 41
Figure 7 Effect of Automatic Reset on Filtered Measurements . . . . . . . . . . 42
Figure 8 Example Calibration Report . . . . . . . . . . . . . . . . . . . . . . . 72
Figure 9 Example Calibration Table Page . . . . . . . . . . . . . . . . . . . . . 73
Figure 10 Data Type Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Figure 11 Graph Type Display . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Figure 12 Example of Data Recorded to Disk or Output Via the Digital Interface . 86
Figure 13 Serial Cable Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Figure 14 Scanner Output Connection. . . . . . . . . . . . . . . . . . . . . . . 114
Figure 15 1575 to 2575 Control Cable Connection . . . . . . . . . . . . . . . . 115
Figure 16 Probe Input Connection . . . . . . . . . . . . . . . . . . . . . . . . . 116
Figure 17 2575 Scanner Front Panel. . . . . . . . . . . . . . . . . . . . . . . . 117
Figure 18 Menu Tree of the 1575 (2575 Scanner Attached) . . . . . . . . . . . 118
Figure 19 2590 to 1590 Front Panel Connections . . . . . . . . . . . . . . . . . 126
Figure 20 2590 To 1590 Control Cable Connection . . . . . . . . . . . . . . . 127
Figure 21 2590 Scanner Front Panel. . . . . . . . . . . . . . . . . . . . . . . . 128
vi
Page 9

Tables

Table1 International Electrical Symbols . . . . . . . . . . . . . . . . . . . . . 1
Table 2 ITS-90 Subranges and Coefficients . . . . . . . . . . . . . . . . . . . 48
Table 3 DIN-43760/IEC-751/ASTM E1137 Callendar-Van Dusen Coefficients. 49
Table 4 Reference Resistor Selection. . . . . . . . . . . . . . . . . . . . . . . 50
Table 5 ITS-90 Calibration Points . . . . . . . . . . . . . . . . . . . . . . . . 54
Table 6 Fixed-point Calibration Setup . . . . . . . . . . . . . . . . . . . . . . 56
Table 7 Manual Comparison Setup . . . . . . . . . . . . . . . . . . . . . . . . 56
Table 8 Automatic Sequencing Calibration Setup . . . . . . . . . . . . . . . . 57
Table 9 SPRT Resistance Comparison Calibration Setup . . . . . . . . . . . . 58
Table 10 Units of Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Table 11 Interface Commands . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Table 12 Interface Commands continued. . . . . . . . . . . . . . . . . . . . . 106
Table 13 Interface Commands continued. . . . . . . . . . . . . . . . . . . . . 107
Table 14 Interface Commands continued. . . . . . . . . . . . . . . . . . . . . 108
Table 15 Standard Resistors Used for Calibration of the Reference Resistors . . 109 Table 16 Standard Resistors Used for Calibration of the Ratio Measurement . . 109
Table 17 2575 Interface Commands . . . . . . . . . . . . . . . . . . . . . . . 123
Table 18 2590 Interface Commands . . . . . . . . . . . . . . . . . . . . . . . 133
vii
Page 10
Page 11
1 Before You Start
1.1 Symbols Used
Table 1 lists the symbols that may be used on the instrument or in this manual and the meaning of each symbol.
Tabl e1 International Electrical Symbols
Symbol Description
AC (Alternating Current)
AC-DC
Battery
Complies with European Union Directives
1 Before You Start
Symbols Used
DC (Direct Current)
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface (Burn Hazard)
Read the User’s Manual (Important Information)
Off
On
1
Page 12
1575/1590 Thermometer Readout
User’s Guide
Symbol Description
Canadian Standards Association
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re fers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Examples include household, office, and laboratory appliances.
C-TIC Australian EMC mark
The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) mark.
1.2 Safety Information
Use this instrument only as specified in this manual. Otherwise, the protection provided by the instrument may be impaired. Refer to the safety information in Sections 1.2.1 and 1.2.2.
The following definitions apply to the terms “Warning” and “Caution”.
“WARNING” identifies conditions and actions that may pose hazards to
the user.
“CAUTION” identifies conditions and actions that may damage the in-
strument being used.
1.2.1 WARNINGS
DO NOT use this unit in environments other than those listed in the User’s Guide.
Follow all safety guidelines listed in the User’s Guide.
Calibration equipment should only be used by trained personnel.
This instrument can measure extreme temperatures. Precautions must be taken to prevent personal injury or damage to objects. Probes may be ex tremely hot or cold. Cautiously handle probes to prevent personal injury. Carefully place probes on a heat/cold resistant surface or rack until they reach room temperature.
DO NOT operate near flammable materials.
Use only a grounded AC mains supply of the appropriate voltage to power the instrument.
DO NOT connect an AC mains supply that does not match the voltage setting on the back of the unit.
DO NOT use this instrument in combination with any probe ( PRT, thermistor, or thermocouple) to measure the temperature or resistance of any device where the probe might come in contact with a conductor that
-
-
2
Page 13
1 Before You Start
Authorized Service Centers
is electrically energized. Severe electric shock, personal injury, or death may occur.
1.2.2 CAUTIONS
If the instrument is dropped, struck, or handled in a way that causes inter
nal or external physical damage, immediately unplug the instrument, dis continue use, and contact a Hart Scientific Authorized Service Center for repair. Do not attempt to disassemble or repair the instrument, batteries, or AC adapter. Refer repairs or replacement components to a Hart Scien tific Authorized Service Center.
The instrument and thermometer probes are sensitive and can be easily
damaged. Always handle these devices with care. DO NOT allow them to be dropped, struck, stressed, or overheated.
Probes are fragile devices which can be damaged by mechanical shock,
overheating, and absorption of moisture or fluids in the wires or hub. Damage may not be visibly apparent but nevertheless can cause drift, in stability, and loss of accuracy. Observe the following precautions:
DO NOT allow probes to be dropped, struck, bent, or stressed.
DO NOT overheat probes beyond their recommended temperature range.
DO NOT allow any part of the probe other than the sheath to be im-
mersed in fluid.
DO NOT allow the probe hub or wires to be exposed to excessive temper-
atures.
Keep the probe wires clean and away from fluids.
-
-
-
-
1.3 Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate service on your Hart product:
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
Customer Support Services
3
Page 14
1575/1590 Thermometer Readout
User’s Guide
Science Park Eindhoven 5108
5692 EC Son
NETHERLANDS
Phone: +31-402-675300
Telefax: +31-402-675321
E-mail: ServiceDesk@fluke.nl
Fluke Int'l Corporation
Service Center - Instrimpex
Room 2301 Sciteck Tower
22 Jianguomenwai Dajie
Chao Yang District
Beijing 100004, PRC
CHINA
Phone: +86-10-6-512-3436
Telefax: +86-10-6-512-3437
E-mail: xingye.han@fluke.com.cn
Fluke South East Asia Pte Ltd.
Fluke ASEAN Regional Office
Service Center
60 Alexandra Terrace #03-16
The Comtech (Lobby D)
118502
SINGAPORE
Phone: +65 6799-5588
Telefax: +65 6799-5588
E-mail: antng@singa.fluke.com
When contacting these Service Centers for support, please have the following information available:
Model Number
Serial Number
Voltage
Complete description of the problem
4
Page 15
2 Introduction
2.1 General Description
The 1575 “Super-Thermometer” and 1590 “Super-Thermometer II” are two of the most accurate resistance and temperature readout instruments available. The 1575 can deliver resistance measurements with up to 4 ppm accuracy while the 1590 can achieve 1 ppm accuracy. In terms of temperature, 1 ppm is equivalent to 0.00025°C with an SPRT at 0°C. When using the 1590 with a stan dards-quality thermistor it’s possible to achieve measurements as accurate as
0.000125°C.
In addition to accuracy, the 1575 or 1590 has a variety of features and functions that make it an indispensable tool in a precision temperature calibration labora
tory. With a wide input range of 0 to 500kΩ it can be used with just about any type of resistance sensor including RTDs, PRTs, SPRTs, HTPRTs, and thermis tors. The driving current is programmable to best suit the sensor. Resistance to temperature calculations are performed automatically using any of a variety of algorithms including the International Temperature Scale of 1990 (ITS-90), IPTS-68, Callendar-Van Dusen, Steinhart-Hart and polynomial equations. The 1575/1590 can even calculate ITS-90 coefficients and generate a report based on resistance and temperature measurements.
The 1575/1590 features an LCD graphics display that shows in a clear and easy-to-read manner a variety of information such as present and past measure­ments, statistical results, and instrument settings. Measurements can be viewed in numeric and graphic plot form. The display is easily configured by the user for various applications. Contrast, brightness, and colors (1590 only) can be ad­justed. Even the tilt of the display on the 1590 can be set as desired. Operation of the 1575/1590 is made simple and convenient with the use of soft-keys. These are buttons having associated labels on the display. The functions of the soft-keys can change between various contexts.
The 1575/1590 features exclusive gold-plated patented DWF connectors (U.S. Patent No. 5,965,625) that are specially designed for this product. Operated with a quick, easy push, the patented DWF connector accepts bare wires, spade terminals, and banana plugs.
The 1575/1590 is able to communicate with computer systems through a vari ety of interfaces. RS-232, IEEE-488, and parallel printer interfaces are all built in. The convenient 31/2" disk drive allows temporary or permanent storage of measurement data, probe coefficients, and instrument settings.
2 Introduction
General Description
-
-
-
-
2.2 Theory of Operation
The 1575 “Super-Thermometer” and 1590 “Super-Thermometer II” require a unique electronic design to achieve the necessary accuracy while meeting size, weight, cost, and speed constraints. This section explains the measurement
5
Page 16
1575/1590 Thermometer Readout
User’s Guide
technique used by these instruments and discusses issues related to performance.
2.2.1 Measurement Technique
Fundamentally, the 1575/1590 measures the resistance ratio between two resis tors by comparing their voltages when equal currents are applied. The simpli
­fied schematic in Figure 1 shows the basic components of the measurement circuitry. These include the current source, sensor, reference resistor, relay switch, amplifier, analog-to-digital converter (ADC), and central processing unit (CPU). The reference resistor and sensor are connected in series and the current flows through both simultaneously. The current produces a voltage on each that is proportional to their respective resistances. The voltages are mea sured with the amplifier and ADC. Since only one of the voltages can be mea
-
-
sured at a time, the relay must be used to switch between them.
SENSOR
CURRENT
SOURCE
REFERENCE
Figure 1 Simplified Schematic Diagram of the Measurement Circuit
AMPLIFIERSWITCH
CONVERTER CPU
The voltage on each resistor is measured twice: once with the current in one di rection and again with the current in the opposite direction. Subtracting the two voltage measurements eliminates offset voltages (including those arising from thermoelectric EMF) since they are constant. In summary, one ratio measure
-
ment requires four voltage samples:
1. Sensor, forward current (V
2. Sensor, reverse current (V
3. Reference, forward current (V
4. Reference, reverse current (V
X2
X1
)
)
)
R1
)
R2
The voltage samples are subtracted and divided to produce a ratio of sensor re sistance to reference resistance:
-
-
-
VV
XX
12
r
=
VVRR
RRXR
6
=
12
Page 17
2 Introduction
Theory of Operation
Using this approach, errors from driving current imprecision, voltage offsets, and amplifier and ADC inaccuracies are avoided because these all affect the voltage samples equally.
Each voltage sample requires 0.5s. It takes 0.15s to set the current and relay and allow time for the voltages to settle and 0.35s for the ADC to make a mea surement and send it to the CPU. Since four samples are required, the entire ra tio measurement takes two seconds.
Depending on how the measurement timing is set up, more than one raw ratio sample may be integrated into one raw measurement. Digital filtering is also applied to reduce noise in the measurements. The CPU then calculates the re
­sistance of the sensor by multiplying the measured resistance ratio by the known resistance of the reference resistor. Temperature is calculated from resis tance using one of the built-in conversion algorithms. Finally, statistical values are recalculated to incorporate the latest measurement. Figure 2 below shows this sequence of operations.
-
-
-
Voltage
Samples
Voltage
Samples
Figure 2 Measurement Processing Operations
Raw
Ratio
Raw
Ratio
Integration
SPRTs and PRTs generally require temperature to be calculated according to the International Temperature Scale of 1990 (ITS-90) by applying coefficients unique to the SPRT or PRT. The 1575/1590 does these calculations automati cally. The sensor coefficients are entered into the 1575/1590 by the user and stored in non-volatile memory. Temperature is calculated in any of a variety of units including degrees Celsius (°C), degrees Fahrenheit (°F), or Kelvin (K). Once the temperature reading is obtained it is sent to the display, stored in memory, and, if necessary, written to disk or any of the output ports.
2.2.2 Performance Issues
Measurement of temperature with uncertainty approaching 0.001°C or better can be a significant challenge. Various sources of error inherent in resistance thermometry make it difficult to achieve this level of accuracy. For instance, lead resistance in some cases can cause errors of several tenths of a degree. Problems also arise from sources such as thermoelectric EMF, reactance, and leakage. The accuracy the 1575/1590 achieves is only possible because these effects have been carefully studied and dealt with. The design of this instru ment reduces or eliminates most of the errors that often plague resistance and
Digital
Filter
Resistance Temperature Display
DisplayStatistics
-
-
7
Page 18
1575/1590 Thermometer Readout
User’s Guide
temperature measurements. It also offers other advantages that contribute to the value of this instrument. Consider the following issues.
2.2.2.1 Lead Resistance
Measurements using an electrical sensor can be affected by the resistance in the connecting wires or leads. Resistance also exists in the patented DWF connec tors and the junction between the wires and patented DWF connectors. In com monly used two or three-wire measurement circuits these resistances and their variability causes errors of 0.1 to 1.0°C.
The 1575/1590 uses a four-wire circuit that completely eliminates the effects of lead resistance. In this scheme, often referred to as a Kelvin circuit, the sensor is driven with current from one set of wires and the resulting EMF is sensed with a different set of wires. The signal is passed to an amplifier with a very high input impedance that draws negligible current from the sensor. As a result, no measurable voltage develops along the EMF sensing wires. The 1575/1590 will accurately measure the resistance of sensors even in the presence of lead
resistance that can be as high as 10Ω.
2.2.2.2 Thermoelectric EMF
A resistance sensor such as a PRT contains several junctions between wires of different metals. These act like thermocouples generating small electric poten­tials called thermoelectric EMFs. Unless rejected in some way, the thermoelec­tric EMFs can interfere with the sensor EMF and degrade the accuracy of the measurement. There are three different techniques that can be used to cancel thermoelectric EMF.
Some resistance bridges apply AC driving current and use sensing circuits that detect only the AC signal, rejecting the DC EMFs. This technique is very effec tive at eliminating thermoelectric EMF errors but can lead to other errors. Reactance, leakage, and eddy currents become much more significant with AC.
A different technique sometimes used in DMMs is to periodically switch off current to the sensor and measure the thermoelectric EMF directly. The prob lem with this is it leads to self-heating errors as the sensor warms and cools from the varying current.
The 1575/1590 uses a third technique. Two separate measurements are made and the driving current is simply reversed for the second measurement. Ther moelectric EMF causes errors that are opposite in the two measurements. In es sence, averaging the two measurements cancels the errors. This technique is very effective at eliminating errors from thermoelectric EMF while avoiding the AC related errors and self-heating problems of the other methods. In fact, it’s so effective in the 1575/1590 that the instrument is able to achieve an un certainty of 0.00025°C or better with no observable error caused by thermo electric EMF.
-
-
-
-
-
-
-
-
8
Page 19
2.2.2.3 Reactance
The use of AC driving current can cause errors in resistance thermometry be cause sensors often exhibit significant inductance and capacitance. The 1575/1590 uses DC circuitry that is much less susceptible to these effects. It al lows plenty of time for currents and voltages to settle before it begins a sample. If necessary, the delay time can be increased even further to ensure accuracy. The 1575/1590 is well suited for any type and quality of resistance sensor.
2.2.2.4 Leakage
Resistance sensors can be susceptible to electrical leakage through the insula tion material surrounding the lead wires and sensing element. Leakage is often significant at low temperatures where the insulation absorbs moisture from the air or at high temperatures where the electrical conductivity of the insulating material is relatively high. Leakage and some other effects, such as dielectric absorption and eddy currents, are much more significant with AC than with DC. By operating with DC driving current the 1575/1590 can achieve excellent accuracy with a large variety of sensors and over a wide range of conditions.
2.2.2.5 Self-heating
Another source of trouble in resistance thermometry is self-heating. This re­sults from power being dissipated in the sensor by the driving current. It causes the temperature of the sensor to be higher than it should be. The 1575/1590 is able to achieve full accuracy with small currents that minimize self-heating (1
mA for PRTs and 10 μA for thermistors). The current can be set within a wide range and with excellent resolution. Being able to set the current to precise ar­bitrary values allows self-heating errors to be controlled, measured, and canceled.
2 Introduction
Theory of Operation
-
-
-
2.2.2.6 Component Drift
The accuracy of a typical resistance measuring instrument is seriously limited by the stability, or lack thereof, of its electrical components. The design of the 1575/1590 eliminates sensitivity to variations of the components due to aging or temperature. This is because, in effect, the instrument recalibrates itself dur ing every measurement. Drift of the driving current, amplifier bias current, am plifier offset voltage, amplifier gain, ADC offset, and ADC scale have no effect on the measurement. The accuracy to which the instrument measures resistance is only affected by the drift of one component-the reference resistor. The four resistors built into the 1575/1590 are high-quality, hermetically sealed, low temperature coefficient, metal film resistors that are temperature controlled for excellent stability. Even better stability can be achieved if external standard re sistors are used and they are immersed in a precisely-controlled oil bath.
2.2.2.7 Noise and Resolution
There will always be electrical noise in any measurement circuit; it’s unavoid able. Excessive noise causes measurements to appear to vary randomly over
-
-
-
-
9
Page 20
1575/1590 Thermometer Readout
User’s Guide
time. This makes it impossible to detect small real changes in the parameter be ing measured. In effect, it limits the resolution of the measuring instrument.
Electrical noise in the 1575/1590 comes from a variety of sources. A small amount of noise is generated by the resistors and semiconductor devices in the measurement circuitry. Some noise (quantization noise) results from the limited resolution of the ADC. Electrical interference or EMI from internal or external sources can also be a source of noise. Although it is impossible to completely eliminate all noise, some steps can be and are taken to reduce it. Components were selected for their low-noise qualities. The ADC was chosen, in part, for its excellent resolution (24 bits). Shielding is used to block EMI from reaching the sensitive circuits. To further reduce noise the 1575/1590 uses filtering and EMI suppression devices throughout the circuit. (Since the 1575/1590 operates with DC driving current, it can effectively reject interference coming from the 50/60 Hz. mains supply. AC instruments are more susceptible to this interference.) Fi nally, the CPU applies digital filtering to remove much of the remaining noise. The end result is the capability of making measurements with effective resolu tion as good as 0.25 PPM.
One drawback of digital filtering is that it might make the instrument react more slowly to changes in the resistance or temperature being measured. The 1575/1590 allows the user to adjust the digital filter to achieve the right balance between resolution and response.
2.2.2.8 Nonlinearity
Having taken care of all the other sources of error explained above, all that’s left is nonlinearity. Consider nonlinearity to be curvature in the graph of the re­lationship between the actual resistance ratio and the resistance ratio measured by the 1575/1590. It is a result of imperfections in the analog-to-digital con­verter and also, to a smaller degree, the power supply and amplifier.
To minimize nonlinearity in the 1575/1590 three steps have been taken. First, the best available components have been selected. For instance, the ADC is a dual-slope integrating type that has linearity that’s at least ten times better than other precision integrating or sigma-delta ADCs. Second, the measurement technique that’s used inherently rejects much of the nonlinearity. Because sam ples of opposite polarity are subtracted, zeroth-order errors (offsets), second-or der errors, and all higher even-order components of the nonlinearity are canceled. What’s left are third-order and higher odd-order components that di minish greatly in magnitude the higher the order. The third step is to mathemat ically correct for the third-order nonlinearity. This is the purpose of the “ADC” calibration parameter. This parameter is adjusted during calibration to achieve the best possible linearity.
-
-
-
-
-
-
-
10
2.2.2.9 Measurement Speed
The measurement scheme used by the 1575/1590 offers other advantages in ad dition to accuracy. One of these is speed. The 1575/1590 is capable of complet ing a new measurement in only two seconds. Even if multiple sensors are being measured in turn, the measurement time per sensor is still only two seconds.
-
-
Page 21
Compare this to a typical resistance bridge that takes 30 to 60 seconds to make the first measurement after a sensor is connected. The speed of the 1575/1590 gives it the advantage of allowing greater efficiency as well as better accuracy during a batch calibration process involving a large number of sensors. Inte grating the 1575/1590 with the 2575/2590 multiplexer enhances its capability even more, giving it 10 input channels (or up to 50 for the 1590 with multiple 2590s). The measurement speed of the 1575/1590 makes other applications possible such as tracking fast-changing temperatures, measuring temperature differences, or evaluating thermal response times.
2.2.2.10 Solid-State Design
Unlike a bridge that requires a large, heavy precision ratio transformer and doz ens of relays, the 1575/1590 uses semiconductor circuits and only a few relays to switch between resistors. This gives it better reliability, smaller size, lighter weight, and lower cost. By keeping the size and cost of the measuring circuit small, more space and expense can be dedicated to other important features such as intelligent system control electronics, a convenient user interface, a graphic display, and a built-in disk drive, all contributing to making the 1575/1590 so versatile and useful.
2 Introduction
Theory of Operation
-
-
11
Page 22
Page 23
3 Specifications and Environmental Conditions
3 Specifications and Environmental
Conditions
3.1 Specifications
Specification 1575 1590
Resistance Measurement Range
Resistance Ratio Accuracy
Ω
0to.25
0.25 to 4Ωinput (1Ωrefr, 10 mA)
2.5to40
0to25
25 to 400Ωinput (100Ωrefr, 1 mA)
400 to 1000
0to2.5k
2.5to40kΩinput (10 kΩrefr, 0.01 mA)
40 to 100 k
100 to 500 k
Resistance Accuracy
0to.25
0.25 to 4Ωinput (1Ωrefr, 10 mA)
2.5to40
0to25
25 to 400Ωinput (100Ωrefr, 1 mA)
400 to 1000
0to2.5k
2.5to40kΩinput (10 kΩrefr, 0.01 mA)
40 to 100 k
100 to 500 k
Typical Temperature Accuracy, External Reference
0.25
2.5
25
100
10 k
Typical Temperature Accuracy, Internal Reference
0.25
2.5
25
input (1Ωrefr, 10 mA) 0.00001
Ω
input (10Ωrefr, 3 mA)
Ω
input (100Ωrefr, 1 mA) 0.0001
Ω
input (100Ωrefr, 0.1 mA)
Ω
input (10 kΩrefr, 0.01 mA) 0.025
Ω
input (10 kΩrefr, 0.005 mA)
Ω
input (10 kΩrefr, 0.002 mA)
Ω
input (1Ωrefr, 10 mA) 0.000025
Ω
input 10Ωrefr, 3 mA)
Ω
input (100Ωrefr, 1 mA) 0.0002
Ω
input (100Ωrefr, 0.1 mA)
Ω
input (10 kΩrefr, 0.01 mA) 0.05
Ω
input (10 kΩrefr, 0.005 mA)
Ω
input (10 kΩrefr, 0.002 mA)
Ω
SPRT (0°C, 1Ωrefr, 10 mA)
Ω
SPRT (0°C, 10Ωrefr, 3 mA)
Ω
SPRT (0°C, 100Ωrefr, 1 mA)
Ω
SPRT (0°C, 100Ωrefr, 1 mA) 0.001°C 0.00025°C
Ω
thermistor (25°C, 10 kΩrefr, 0.01 mA)
Ω
SPRT (0°C, 1Ωrefr, 10 mA)
Ω
SPRT (0°C, 10Ωrefr, 3 mA)
Ω
SPRT (0°C, 100Ωrefr, 1 mA)
200 ppm 120 ppm
100 ppm 40 ppm
200 ppm 120 ppm
0.00025°C 0.00013°C
0 to 500 k
Ω
40 ppm 20 ppm
20 ppm 5 ppm
Ω
4 ppm 1 ppm
20 ppm 4 ppm
Ω
10 ppm 5 ppm
50 ppm 25 ppm
Ω
40 ppm 20 ppm
Ω
8 ppm 6 ppm
25 ppm 10 ppm
Ω
20 ppm 10 ppm
60 ppm 30 ppm
0.01°C 0.005°C
0.005°C 0.0013°C
0.001°C 0.00025°C
0.025°C 0.01°C
0.01°C 0.005°C
0.002°C 0.0015°C
Specifications
Ω
0.000005
0.000025
0.012
Ω
0.00001
0.00015
0.025
Ω
Ω
Ω
Ω
Ω
13
Page 24
1575/1590 Thermometer Readout
User’s Guide
Specification 1575 1590
Ω
SPRT (0°C, 100Ωrefr, 1 mA) 0.002°C 0.0015°C
100
Ω
thermistor (25°C, 10 kΩrefr, 0.01 mA)
10 k
Typical RMS Measurement Noise No Filter 20 Second Filter
Ω
input (1Ωrefr, 10 mA)
0.25
Ω
input (10Ωrefr, 3 mA)
2.5
Ω
input (100Ωrefr, 1 mA)
25
Ω
input (100Ωrefr, 1 mA)
100
Ω
input (100Ωrefr, 1 mA)
400
Ω
input (10 kΩrefr, 0.01 mA)
10 k
Minimum Sample Period 2 seconds
Maximum Current
Ω
reference, 1Ωinput
1
Ω
reference, 10Ωinput
10
Ω
reference, 100Ωinput
100
Ω
reference, 10 kΩinput
10 k
Typical Current Accuracy 2% or 0.0002 mA
Power 100–125/200–250 VAC (user-switchable), 50-60
Specified Operating Temperature 20 to 27°C (68 to 81°F)
Operating Temperature Limits 10 to 40°C (50 to 104°F)
Storage Temperature –10 to 50°C (14 to 122°F)
Safety OVER VOLTAGE (Installation) CATEGORY 11,
Size 516 mm x 320 mm x 178 mm
Weight 16 kg (35 lb)
Notes:
Specifications stated in ppm are relative to the reading.
Accuracy specifications are applicable for a one-year calibration interval. In line with normal prudent metrology practices, Hart recommends a short-cycle calibration interval of six months for new units during the first year to ensure that all components are as stable as expected.
Accuracy specifications do not include noise which is stated separately and depends on the digital filter settings.
For measurements above 100kΩ the conversion time should be set to 4 seconds.
The specifications are applicable within the specified operating temperature range. Re sistance accuracy is derated 10 ppm if using the 1 ohm resistor, 5 ppm if using the 10Ω or 10 kΩ resistors, and 2 ppm if using the 100Ω resistor for every degree Celsius out side the specified operating temperature range.
Temperature accuracy specifications do not include sensor errors or external resistor uncertainty.
0.0005°C 0.00025°C
25 ppm 8 ppm
8 ppm 2.5 ppm
2.5 ppm 0.8 ppm
1 ppm 0.3 ppm
2.5 ppm 0.8 ppm
6 ppm 2 ppm
20 mA
15 mA
5 mA
0.1 mA
Hz, 1A max
Pollution Degree 2 per IEC1010-1
(20.3" W x 12.6" D x 7.0" H)
-
-
14
Page 25
3 Specifications and Environmental Conditions
3.2 Environmental Conditions
Environmental Conditions
Although the instrument has been designed for optimum durability and trou ble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty, dirty, or wet environment. Maintenance and cleaning recommentdations can be found in the Maintenacne section of this manual.
For full accuracy, operate the instrument in ambient temperatures between 20–27°C (68-81°F). Do not operate the instrument in an environment colder than 10°C (50°F) or warmer than 40°C(104°F).
The instrument operates safely under the following conditions:
Temperature range: 10 to 40°C
Ambient relative humidity: 15 to 50%
Pressure: 75 to 106 kPa
Vibrations should be minimized
Altitude less than 2,000 meters
3.3 Warranty
Fluke Corporation, Hart Scientific Division (Hart) warrants this product to be free from defects in material and workmanship under normal use and service for a period as stated in our current product catalog from the date of shipment. This warranty extends only to the original purchaser and shall not apply to any product which, in Hart’s sole opinion, has been subject to misuse, alteration, abuse or abnormal conditions of operation or handling.
Software is warranted to operate in accordance with its programmed instruc tions on appropriate Hart products. It is not warranted to be error free.
Hart’s obligation under this warranty is limited to repair or replacement of a product which is returned to Hart within the warranty period and is determined, upon examination by Hart, to be defective. If Hart determines that the defect or malfunction has been caused by misuse, alteration, abuse or abnormal condi tions or operation or handling, Hart will repair the product and bill the pur chaser for the reasonable cost of repair.
To exercise this warranty, the purchaser must forward the product after calling or writing to an Authorized Service Center (see Section 1.3 on page 3) for au thorization. The Service Centers assume NO risk for in-transit damage.
THE FOREGOING WARRANTY IS PURCHASER’S SOLE AND EXCLU SIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EX PRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OR MERCHANTABILITY, OR FITNESS FOR ANY PARTICULAR PURPOSE OR USE. HART SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAM AGES OR LOSS WHETHER IN CONTRACT, TORT, OR OTHERWISE.
-
-
-
-
-
-
-
-
15
Page 26
Page 27
4 Quick Start
4 Quick Start
This section briefly explains the basics of setting up and operating your 1575 or 1590 thermometer readout.
4.1 Unpacking
Carefully unpack the 1575 or 1590 thermometer readout and inspect it to make sure all components are present and in satisfactory condition. Verify that the following items are present:
1575 or 1590 Thermometer Readout
Serial Cable
Manual
Report of Calibration
Calibration Label
If all items are not present, call a Hart Scientific Authorized Service Center.
4.2 Learn About the Features and Components
Unpacking
Familiarize yourself with the features and components of the 1575 or 1590 by reading Section 5.
4.3 2575/2590 Scanner Setup
If a 2575 or 2590 scanner was purchased, refer to Section 10, 2575 Scanner or Section 11, 2590 Scanner concerning specifics to each scanner.
4.4 Power Source
The 1575/1590 is normally configured to use 100–125 VAC, 50-60Hz. It may optionally be switched to use 200–250 VAC, 50-60Hz. The switch on the back of the instrument is used for switching the voltage. Make sure the voltage switch is always set correctly!
4.5 Probe Connection
Your SPRT or thermistor probe connects to the 1575/1590 using the patented DWF connector posts of Input 1. The input accepts a four-wire probe. The C1 and C2 patented DWF connectors source current to the probe and P1 and P2 measure the voltage on the probe. A resistor used for a reference should be connected to Input 2 in the same fashion.
17
Page 28
1575/1590 Thermometer Readout
User’s Guide
There are also ‘Guard’ patented DWF connectors for each channel and a ‘Chas sis’ patented DWF connector that can be used for attaching guard wires and for grounding. Probes that have shielded cables will be less susceptible to electro magnetic interference (EMI). Best results may be obtained if the guard wire is attached to the ‘Guard’ patented DWF connector, the ‘Chassis’ patented DWF connector, or both. This depends on environmental conditions and placement of the probe. If the probe has no shield or guard try simply connecting a wire be tween the ‘Guard’ and ‘Chassis’ patented DWF connectors. Experiment with each scheme to find what works best in your application.
The probe patented DWF connectors are operated by simply pressing on them. Pressing on a patented DWF connector opens it allowing you to insert wires with or without spade terminals. When the patented DWF connector is released it will clamp down on the wire. Banana plugs can also be inserted into the hole in the top of the patented DWF connectors.
4.6 Switch the Power On
Power is switched on and off with the power button located on the front panel, lower left corner. The instrument will take one or two minutes to power up, ini­tialize itself, and begin normal operation. During initialization the display re­mains dark.
4.7 Measure Temperature
To make temperature measurements using your probe you must set up the following:
1.
Select the channel to which your probe is connected. To do this enter the CHANNEL MENU and press the INPUT 1 soft-key if the probe is connected to Input 1 or press the INPUT 2 soft-key if it is connected to the Input 2 (Section 7.1). If you are using the 2575 or 2590 Scanner, press the appropriate channel button on the 2575/2590 front panel or use the SELECT CHANNEL soft-key to select the channel as explained in the scanner’s manual. Use the RETURN soft-key to exit the menu.
2.
Set up the reference for the probe. To do this enter the PROBE MENU, select EDIT PROBE (Section 7.4.3), and press Enter to select the probe number. In the EDIT PROBE window, move the cursor down
to the REFERENCE selection using the U and D keys. Use the L and R keys and Enter to select the appropriate reference for your probe (Section 7.4.3.3). For example, a 25Ω SPRT may use the 100Ω refer
ence. Be sure to always press Enter after making any selection.
3.
Set the current for the probe. This is also done in the EDIT PROBE function. Move the cursor to CURRENT using the U and D keys. Use
the numeric keys and Enter to enter the appropriate current for your par ticular probe. Your probe specifications should state the required current.
For example, a 25Ω SPRT may use 1 mA.
-
-
-
-
-
18
Page 29
4 Quick Start
Measure Temperature
4.
Set the temperature conversion type. Move the cursor to the CON
-
VERSION selection. Use the L and R keys and Enter to select ITS-90 or whatever conversion is appropriate for your probe (Section
7.4.3.2). Refer to the calibration report or specifications for your probe. Most SPRTs and RTDs use the ITS-90.
5.
Enter the characterization coefficients for your probe. Use the nu meric keys and Enter to enter each coefficient. The coefficients are deter
-
­mined when the probe is calibrated. Refer to the calibration report or specifications for your probe. When finished exit out to the main menu using the RETURN soft-key or Exit.
6.
Begin sampling. If the 1575/1590 is not now sampling press the RUN soft-key in the SAMPLE MENU to begin sampling (Section 7.2.1).
The 1575/1590 should now be able to measure temperature accurately. If de sired, you can modify the filter settings in the SAMPLE MENU to minimize noise (Section 7.2.4). You may change the units from °C to any other desired units such as °F or K using the UNIT soft-key in the SYSTEM menu (Section
7.6.1). You can set up the display to show statistical data or graph measure­ments in the DISPLAY MENU (Section 7.5).
-
19
Page 30
Page 31
5 Parts and Controls
This section describes the exterior features of the 1575/1590.
5.1 Front Panel
The following are found on the front of the 1575/1590.
Display
The display is a 320 x 240 pixel color graphics LCD device with a bright CCFT backlight. The contrast can be adjusted using the front panel buttons. The angle of the display can be adjusted for best viewing by pushing it in at the top (1590 only). The display is used to show measurements, status information, operating parameters, and soft-key functions. Details of the display are explained in Sec tion 5.3.
Soft-keys
The soft-keys are the six buttons to the immediate right of the display. The functions of the soft-keys are indicated on the display to the left of the buttons. They may change depending on the menu or function that’s selected.
5 Parts and Controls
Front Panel
-
Buttons
The front panel buttons allow you to enter numeric data for operating parame­ters, move the cursor on the display, and adjust the brightness and contrast of the display. The functions of the buttons are further explained in Section 5.4.
Patented DWF Connector Posts
The input patented DWF connector posts are located on the right side of the front panel. Thermometer probes can be connected to the patented DWF con nectors with spade lugs, banana plugs, or bare wires. There are two channel in puts. These are labeled Input 1 and Input 2 on the front panel. Input 1 is usually used when measuring with only one resistance thermometer. Input 2 can be used with a second probe for comparative temperature measurements. It can also be used to connect an external reference resistor (see Section 7.6.5.1 on page 94) for use when measuring with Input 1. Each input has five patented DWF connectors for four-wire (Kelvin) resistance measurements. The C1 and C2 patented DWF connectors supply current to the probe. The P1 and P2 pat ented DWF connectors are used to sense the voltage potential on the probe. The guard or shield of the probe cable can be connected to the “Guard” patented DWF connector. This connects to the ADC circuit common which is isolated from chassis ground. The “Chassis” patented DWF connector connects directly to the chassis and is grounded to earth through the power cable. This can be used as a convenient grounding point if needed.
-
-
-
21
Page 32
1575/1590 Thermometer Readout
User’s Guide
Disk Drive
The 3.5" floppy disk drive is located at the left side of the front panel. The disk drive can be used to store probe data and operating parameters and to log mea surement data. The drive is MS-DOS compatible allowing data to be trans ferred between the 1590 and PC-compatible computers. The drive can read, write, and format with either 720 kb or 1.44 Mb disks. The disk drive has an LED indicator at the top that illuminates when the drive is in use, and an eject button at the bottom for removing disks. Disks are inserted into the drive with the top side to the right.
-
-
CAUTION: Be careful not to accidentally push the power switch when at
tempting to remove disks from the drive.
Power Switch
The power switch is located at the bottom left of the front panel. It is illumi nated when power is on.
Rack Mounting Screws
The 1575/1590 can be removed from the outer cabinet for installation into a rack system. It is removed by first removing the four front panel screws and the four back panel screws. (It may be easier to take out if the top of the box is removed.)
5.2 Back Panel
The following are found on the back of the 1575/1590.
Power Cord
The power supply cord attaches to the rear panel next to the fan. Plug the cord into an AC mains supply appropriate for the voltage switch setting. Above the power connector is an auxiliary power output that can supply up to 1A of current.
Voltage Switch
The 1590 can be used with either 115 or 230V AC nominal voltages. The volt age is selected with the red slide switch located directly above the power connector.
-
-
-
22
WARNING: Make sure the voltage switch is always set correctly. Setting
the switch incorrectly may damage the instrument.
Page 33
5 Parts and Controls
Back Panel
Fan
The fan is necessary to keep the internal components cool. Always make sure air can flow freely behind and around the instrument.
Serial Connector
The serial connector is a 9-pin subminiature D type located on the back panel. The serial (RS-232) interface can be used to transmit measurements and control the operation of the 1575/1590.
Printer Connector
The 25-pin subminiature D parallel connector can be used to connect the 1575/1590 to a printer to record measurements or print calibration reports.
IEEE-488 Connector
The 1575/1590 can be connected to an IEEE-488 bus (GPIB) through this con nector on the back panel. The IEEE-488 bus can be used to read measurements and control the operation of the instrument.
Analog Output
The two binding posts on the back panel are used for the analog output. This provides a DC potential between +5V and -5V that indicates the value of measurements.
-
Scanner Connector
The 15-pin connector on the back panel can be used to interface with the Hart 2575 or 2590 Scanner. It provides power and digital control signals for the scanner. Operation of the 1575/1590 with the 2575 or 2590 Scanner is ex
-
plained in Section 10, 2575 Scanner or Section 11, 2590 Scanner respectively.
23
Page 34
1575/1590 Thermometer Readout
User’s Guide
5.3 Front Panel Display
The front panel display is shown in detail in Figure 3 and its features are de
-
scribed below.
Figure 3 1575/1590 Super-Thermometer Display
The display is a 320 x 240 pixel graphics LCD device with a bright CCFT backlight. The contrast can be adjusted using the front panel buttons marked
C. The tilt of the display can be adjusted for best viewing by pressing the dis play in at the top (1590 only). The 1575/1590 has a screen saver feature which can be used to extend the lifetime of the display (see Section 7.6.2.5). The dis play is used to show measurements, status information, operating parameters, and soft-key functions. The types of data shown on the screen are described below.
-
-
24
Measurement Window
The most recent measurement is shown in large digits in the box at the top of
the screen. Out-of-range measurements are displayed as “..........”.The
unit of measurement is shown to the right of the measurement (“C” for Cel
-
sius, “F” for Fahrenheit, “K” for Kelvin, “Ω” for resistance in ohms, “W” for ITS-90 W(T), and “r” for ratio of input resistance to reference resistance).
Channel Indicator
To the left of the measurement is a number that indicates the channel that was measured.
Page 35
5 Parts and Controls
Front Panel Buttons
Warning Messages
From time-to-time important warning messages may appear immediately below the measurement. “FILTER RESET” indicates the filter has been reset (see Sec tion 7.2.4). “WARNING: SUBRANGE EXCEEDED” indicates the measure
-
ment is outside the ITS-90 subrange (see Section 7.4.3.2.1).
Sampling Status
Below the measurement window is a box where the sampling status and mes sages are displayed. The status window indicates which channel is being mea
-
­sured or else shows that sampling is suspended. The sliding bar in the status window shows the progress of the sampling process.
Data Fields
Each of the five data fields can be set up to display any of a variety of values such as previous measurements, maximum and minimum measurements and deviations from a reference value (see Section 7.5.4). Alternately, this area may be used to show measurements graphically.
Oven Indicator
This indicator shows whether the oven is on or off. If it was recently turned on, the time remaining until the oven is warmed up and stabilized is shown. The oven regulates the temperature of the internal reference resistors to minimize drift (see Section 7.6.5.2).
-
Date and Time
The date is shown at the bottom of the screen in the order of month, day, and year. The time of day is shown at the bottom of the display in either 12-hour mode (12:00 a.m. to 11:59 p.m.) or 24-hour mode (0:00 to 23:59).
Soft-key Functions
The six boxes on the right side of the display indicate the functions of the soft-keys. These functions change with each menu.
Editing Windows
While setting up and operating the 1575/1590, you are often required to enter or select parameters. Editing windows appear on the screen when necessary to show the values of parameters and allow you to change them.
5.4 Front Panel Buttons
The functions of the front panel buttons are described below.
25
Page 36
1575/1590 Thermometer Readout
User’s Guide
Soft-keys
The six soft-keys to the right of the display are used to select menus or menu functions. The functions of the soft-keys are indicated in the small boxes on the display just to the left of the soft-keys. The functions of the soft-keys change depending on the selected menu. The bottom soft-key is usually RETURN. Pressing this key exits from a sub-menu or window and returns to the previous menu. (For an outline of the menu structure, see Section 7.)
Numeric Keys
The ten digit keys, the decimal point, +/-, and EXP keys are used to type nu meric data or make numbered selections.
Enter
The Enter key is used to enter a new parameter value or option. When the value of any parameter is changed Enter must be pressed to accept the new value. If
Exit, Home, U,orD, are pressed before Enter, the cursor will leave the pa rameter and any changes made to it will be canceled. Within a window with a list of parameters, pressing Enter will also move the cursor down to the next parameter. If the cursor is at the bottom of the list, pressing Enter without changing the parameter will exit the window. The Enter button may also be used during some operations to affirm or continue with an action or choice.
-
-
26
NOTE: The Enter button must always be pressed after changing a param-
eter to accept the new value or option.
Del
When entering or editing a numeric parameter the Del (delete) key is used to delete a digit that is underscored by the cursor.
Exit
The Exit key is used to cancel an operation, exit a window, or return from a lower menu to a higher menu. In any window, pressing Exit will immediately exit the window and skip to the next window or return to the menu. If a param eter is entered or changed and Exit is pressed before Enter, the change will be canceled. During some operations the Exit button may be used to cancel the operation.
Home
The Home key is used to return from any sub-menu, window or entry to the main menu. If a parameter is entered or changed and Home is pressed before Enter, the change will be canceled. Home may also be used to refresh the screen if it happens to become corrupted for some reason.
-
Page 37
5 Parts and Controls
Front Panel Buttons
The up and down arrow keys are used to move the cursor through a list of pa
­rameters in a window. Note that parameter entry will not be accepted if the up or down button was pressed before Enter. Thus these keys can also be used to intentionally cancel a change to a parameter. If the list is longer than can be dis
played in a window, U and D can be used to scroll the list.
The left and right arrow keys have two functions. When entering or editing a numeric parameter these can be used to move from digit to digit. When setting some parameters these are used to change the option. Remember, you must al
-
ways press Enter to save a new value or option.
-
These are the display contrast adjustment buttons. They can be used at any time to increase or decrease the contrast of the display.
-
27
Page 38
Page 39
6 General Operation
This section provides applications, suggestions for best results, and answers to some common questions concerning operating the 1575/1590 Super-Thermometer.
6.1 Applications
Listed here are some of the applications for which the 1575/1590 Thermometer is well suited.
High-precision temperature measurement using SPRTs. The excellent
accuracy of the 1575/1590 makes it ideal for high-precision temperature measurement using SPRTs. The 1575/1590 accepts ITS-90 characteriza tion coefficients (see Section 7.4.3.2) for direct temperature readout. The 1575/1590 can measure using a wide range of SPRTs - capsule SPRTs, 25
and 100( midrange SPRTs, and 2.5Ω and 0.25Ω high-temperature SPRTs (see Section 7.4.3.3 and 7.4.3.4). You have the option of using the high-precision internal resistors or your own standard resistor as a refer­ence (see Section 7.4.3.3). The 1575/1590 makes it convenient to use your SPRT in conjunction with a triple point of water cell to eliminate er­ror due to SPRT and reference resistor drift (see Section 7.4.4.1).
High-resolution temperature measurement using thermistors. The 1575/1590 can also read sensors with resistances of 10 kΩ and greater
(see Section 7.4.3.3 and 7.4.3.4). This makes the 1575/1590 an excellent readout for thermistor temperature sensors. The superb accuracy and reso­lution of the 1575/1590 along with the extra sensitivity of thermistors gives you the capability of measuring temperature with remarkable reso­lution and accuracy. The 1575/1590 will accept Steinhart-Hart thermistor characterization coefficients for direct readout of temperature (see Section
7.4.3.2.5).
Resistance measurement and calibration. The 1575/1590 can be used to measure and compare fixed resistances with great precision. Since it is based on DC measurement technology it can be used with both AC and DC resistors (see Section 7.4.3.2.6, Section 7.6.1, and Section 9).
Calibration of PRTs. The 1575/1590 is able to calculate ITS-90 charac terization coefficients. The CAL ITS-90 function directs the user through a step-by-step procedure for measuring temperatures and corresponding resistances and calculating the coefficients (see Section 7.4.4.2). The 1575/1590 will even create a calibration report and table that can be printed or written to disk.
Temperature comparison. The 1575/1590 allows you to connect and measure two probes simultaneously. You can compare two temperatures and display the difference automatically (see Section 7.1.3). With the 2575 or 2590 Scanner added to the system you can automatically measure up to 50 probes (10 with the 2575).
6 General Operation
Applications
-
-
29
Page 40
1575/1590 Thermometer Readout
User’s Guide
Self-heating tests. The 1575/1590 offers very flexible control of the exci
tation current used in measuring sensors and resistors. The current can be finely adjusted to provide the optimum resolution and accuracy with mini mum self-heating error (see Section 7.4.3.4). Easy-to-use controls enable you to double or halve the power to the sensor for quick self-heating tests (see Section 7.4.4.3 and 7.4.4.4).
Observing the melt-freeze cycle of a metal freeze-point cell. The
1575/1590 allows on-screen graphing of measurements (see Section
7.5.1). One convenient application for this is to visually follow the
melt-freeze cycle of a metal freeze-point cell.
Testing baths, dry-wells, furnaces, and ovens. The excellent resolution
and accuracy of the 1575/1590 makes it very useful as a temperature ref erence for constant-temperature baths, dry-wells, furnaces, and ovens. The graphing feature (see Section 7.5.1) or statistical standard deviation calculation (see Section 7.5.3) can be used to evaluate the temperature sta bility of these devices.
Recording temperature measurements on disk. The internal disk drive
gives you the capability of conveniently transferring data from the 1575/1590 to your computer (see Section 7.6.3). The 1575/1590 can re­cord on disk a series of measurements taken over a long period of time. You can use this to observe long-term fluctuations of temperature. The data can be downloaded to a computer for analysis, graphing, and print­ing.
Recording temperature with a strip-chart recorder. The 1575/1590 an-
alog output allows you to connect the 1575/1590 to a strip-chart recorder and graph measurements directly on paper (see Section 7.6.4.4. You can easily program the analog output for any desired relationship between the measured value and the output voltage.
Using other types resistance sensors. The 1575/1590 can calculate tem perature using an arbitrary polynomial instead of the standard algorithms (see Section 7.4.3.2.4). This makes it possible to use the 1575/1590 with practically any type of resistance sensor.
Remote communications. The 1575/1590 is equipped with a variety of communications interfaces. With the parallel printer interface you can connect the Super-Thermometer directly to a printer and record measure ments or print calibration reports (see Section 7.6.4.3). The serial RS-232 interface and IEEE-488 GPIB interface allow you to connect the Super-Thermometer with a computer in an automated system capable of monitoring or controlling processes (see Section 7.6.4.1 and Section
7.6.4.2).
-
-
-
-
-
-
30
6.2 Suggestions for Best Results
This section provides some suggestions to help you get the best results from the 1575/1590 Super-Thermometer.
Page 41
6.2.1 Maintaining Accuracy
6 General Operation
Suggestions for Best Results
There are many potential problems that can lead to inaccurate results when us ing the 1575/1590. Some of these are: improper calibration of the 1575/1590, incorrect setup of the 1575/1590 parameters, wrong probe characterization co efficients, a damaged probe or sensor, poor connection of the probe wires, or misuse of the 1575/1590. Always be alert to anything that may be wrong and always try to verify your results. When possible, perform regular checks to ver ify the accuracy of your measurements. Consider the following suggestions:
1.
Maintain calibration of the 1575/1590. It is recommended that the 1575/1590 be calibrated at intervals no greater than one year. For new units the calibration interval should be six months for the first year. On a more frequent basis, you can verify the resistance accuracy using stan
­dard calibrated resistors. This is most important when the 1575/1590 is used in applications where accuracy is especially critical. (Section 9 ex
-
plains how to calibrate or verify the accuracy of the 1575/1590.)
2.
Double-check the reference resistor and current. Proper selection of the reference resistor and driving current is very important to achieve full accuracy. You should always verify the reference resistor and current se­lection before making measurements. Consider displaying the reference resistance and current in the programmable data fields so that they can be readily checked (see Section 7.5.3).
3.
Set the digital filter correctly. The digital filter helps to improve resolu­tion and accuracy. If it not set up properly the filter can be ineffective. A common mistake is to set the filter threshold too low. This can cause the filter to reset too often allowing excessive noise to pass through (see Sec­tion 7.2.4). The message “FILTER RESET” helps alert you if this is a problem.
4.
Check the probe coefficients carefully. Using the wrong probe coeffi cients can cause significant error with temperature measurements. Al
-
­ways verify that the correct characterization coefficients are being used. It is good practice to display the probe serial number in one of the pro
­grammable data fields to help you verify that the correct parameters are associated with the probe being measured (see Section 7.5.3). You can use the CONV TEST function to test calculations of temperature with the probe coefficients (see Section 7.4.4.5).
5.
Use the parameter lock-out feature to protect critical parameters.
Set up the parameter safety lock-out to prevent access to critical parame
­ters such as the reference resistor calibration values (see Section 7.6.5.4). This will help protect these parameters from being inadvertently changed.
6.
Compare measurements of your working SPRT to a backup SPRT.
Resistance thermometers, especially SPRTs are very fragile and can eas
­ily be damaged or lose accuracy if mishandled. They are susceptible to thermal shock and contamination. Periodically check your working
-
-
-
31
Page 42
1575/1590 Thermometer Readout
User’s Guide
SPRT to make sure it is accurate. One way to do this is to compare its measurements to those of a second SPRT.
7.
Check system accuracy with a fixed-point cell. An excellent way to verify the accuracy of the entire system, including the 1575/1590, refer ence resistors, and SPRT, is to check measurements using a fixed-point cell. The most accurate, useful, and economical fixed point is the triple point of water. Other fixed points such as the gallium melting point, mer cury triple point, and tin freezing point can also be used for verification.
8.
Check accuracy after calibrating a PRT. After calibrating a PRT com pare measurements using the PRT and its new characterization coeffi cients with your reference SPRT at each calibration point.
9.
Be aware of other problems that can affect accuracy. Be careful of other potential problems that can affect accuracy. When using baths make sure probes are immersed adequately to prevent stem conduction error. When comparing or calibrating one probe against another it is im portant that the bath have good temperature stability and uniformity. Make sure probe wires have good low-resistance electrical connections. Avoid excessive electromagnetic interference. Keep leads as short as possible and use shielding if necessary. Prevent abrupt or extreme changes in ambient temperature around the 1575/1590 and wire connec­tions. Operate the 1575/1590 in a stable environment free from signifi­cant drafts.
-
-
-
-
-
32
6.2.2 Evaluating Temperature Uncertainty
The 1575/1590 Thermometer Readout is often used for calibrating PRTs and in other applications where precision temperature measurements are critical. It is important to know how much uncertainty the 1575/1590 contributes to total system uncertainty. The specification section of this manual (Section 3.1) pro vides typical temperature uncertainty figures for some particular conditions. The actual uncertainty of your measurements may differ. A significant factor that affects uncertainty is the sensitivity of the probe. The greater the sensitiv ity, the lower the uncertainty. For SPRTs sensitivity generally decreases as tem perature increases. As a result, uncertainty will be greater at higher temperatures. To determine the exact uncertainty at a given temperature take the uncertainty of the resistance measurement (ohms) and divide by the slope of the resistance-temperature curve (ohms per degree) at that temperature. As
an example, suppose the 1590 and a 25.5Ω SPRT are used to measure a tem perature of 419°C. The SPRT’s resistance at this temperature is about 65.45Ω and its slope is 0.08916 Ω/°C. The resistance measurement uncertainty is 6 ppm x 65.45Ω = 0.00039Ω so the temperature uncertainty of the 1590 is
0.00039Ω / 0.08916 Ω/°C = 0.0044°C.
Realize that this is only the contribution of the 1590 to total uncertainty. The SPRT or sensor contributes additional uncertainty. External reference resistors, if used, must also be considered. With comparison calibrations using baths or dry-wells you must take into account uncertainty caused by temperature varia
-
-
-
-
-
Page 43
Answers to Some Common Questions
6 General Operation
tions and uniformity. Carefully consider all sources of uncertainty when evalu ating total system uncertainty.
6.3 Answers to Some Common Questions
Following are answers to some often-asked questions.
Can the 1575/1590 be used with thermocouples?
No. The 1575/1590 makes precision resistance ratio measurements and is not capable of making voltage measurements which are required with thermocouples.
Is the 1575/1590 an AC or DC device?
The 1575/1590 is considered a DC measuring device in that the driving current remains constant for the duration of each sample. The current is reversed in-be tween samples so that offsets can be canceled. This occurs at a very low rate (with a period of one second or greater) and plenty of time is allowed for set­tling before sampling the voltage. With AC bridges the current alternates at a rate of 25 Hz or more in a sinusoidal fashion. They require specially-certified AC standard resistors. The 1575/1590 operates just as well with AC or DC standards.
What is the uncertainty of temperature measurements using the 1575/1590?
The specification section of this manual provides typical temperature uncer­tainty figures for some particular conditions. The actual uncertainty of your measurements may vary depending on the resistance of your sensor. Uncer­tainty also depends significantly on the quality of the SPRT or sensor and its calibration. Section 6.2.2 provides some suggestions for evaluating uncertainty.
-
-
How can I verify the accuracy of the measurements?
There are many methods you can use to verify that measurements you make with the 1575/1590 are accurate. Some of these methods are explained in Sec
­tion 6.2.1. It is important that the resistance accuracy of the 1575/1590 be regu larly calibrated and verified. This can be done easily by the user with one or more standard resistors (see Section 9). Temperature measurement can also be verified using fixed points as explained in Section 6.2.1.
Can I calibrate the 1575/1590 myself?
Yes, if you have the necessary standard resistors. These resistors must have ad equate certifiable uncertainty. The calibration procedure is explained in Section
9.
-
-
33
Page 44
1575/1590 Thermometer Readout
User’s Guide
If I only use the internal 100Ω reference resistor, do I need to maintain cali bration of the other reference resistors?
If you only use certain internal reference resistors it is not necessary to cali
-
brate the unused resistors. For example, if you only use the 1575/1590 for mea suring 10 kΩ thermistors there may be no need to calibrate the 1Ω,10Ω,and 100Ω internal reference resistors. Likewise, if the 1575/1590 will only be used
with external reference resistors you do not need to calibrate any of the internal reference resistors. Nevertheless, you should still periodically verify the resis
­tance ratio accuracy of the 1575/1590. If you choose to maintain only a limited calibration take necessary precautions to ensure that the instrument will not be used outside its calibrated ranges.
If two probesare connected tothe 1575/1590, is current supplied to both at the same time?
During a measurement, the 1575/1590 drives only the probe and reference re
­sistor with current. If two probes are connected and measured alternately while using one of the internal reference resistors, the current will alternate between the two probes. It is possible that this could cause slight error in the measure­ment if self-heating is significant. Self-heating error can be evaluated with sim­ple tests (see Section 7.4.4.3). When using the 2590 Scanner with the 1590 this problem can be avoided since the 2590 incorporates an independent standby current source for each channel.
Can the 1575/1590 calculate ITS-90 coefficients?
Yes. The ITS-90 calibration feature allows you either measure resistance data or enter previously obtained values to calculate ITS-90 coefficients (see Section
7.4.4.2.8).
-
-
34
Can the 1575/1590 generate Callendar-Van Dusen, IPTS-68, or thermistor coefficients?
The 1575/1590 is not able to calculate coefficients for characterizations other than ITS-90. The 1575/1590 can be used to obtain resistance and temperature measurements for calibration, but calculation of coefficients for Callendar-Van Dusen, IPTS-68, or thermistor sensors must be done with external software.
Can the 1575/1590 convert IPTS-68 or Callendar-Van Dusen coefficients to ITS-90 coefficients?
The 1575/1590 can be used to calculate ITS-90 coefficients from IPTS-68 or Callendar-Van Dusen coefficients using the following procedure: First, you must set up a probe record with your IPTS-68 or Callendar-Van Dusen coeffi
­cients (use EDIT PROBE, see Section 7.4.3.2.2). Second, use the CONV TEST function to enter resistances and obtain temperatures near each of the required ITS-90 calibration points for Range 4 and Range 7 or 8 (Section 7.4.4.2 and
7.4.4.5). Carefully note the resistances and corresponding temperatures. Third, use EDIT PROBE to select ITS-90 and the appropriate subranges (Section
7.4.3.2.1) for one probe record (there’s no need to enter values for A[4], B[4],
Page 45
Answers to Some Common Questions
6 General Operation
etc.). Fourth, select the probe for ITS-90 calibration using SETUP in CAL ITS-90 (Section 7.4.4.2.6). Fifth, enter the resistance and temperatures using DATA (Section 7.4.4.2.8). Sixth, calculate the ITS-90 coefficients using CAL CULATE (Section 7.4.4.2.9).
-
Can the 1575/1590 directly control abath or otherdevice using digitalcommu nications?
No. The 1575/1590 is intended to be used only as a measuring device. The 1575/1590 has no means of sending controlling commands to other equipment. A separate computer must be used for this purpose.
Can I connect any multi-channel scanner to the 1575/1590?
The Hart 2590 Scanner is specially designed to be used with the 1590 and like wise, the 2575 with the 1575. The inputs of the 1575/1590 can be connected to third-party scanners but the 1575/1590 will not be able to control the scanner directly. You would have to select channels manually or use a computer to con trol the scanner.
What has been done to verify and validate the accuracy of the 1575/1590 and its internal software?
Hart Scientific makes every effort to ensure that your 1575/1590 Super-Ther­mometer operates reliably and accurately. It is put through an extensive valida­tion procedure. Each function and calculation is tested and verified. Each of the temperature conversion algorithms are tested over the full temperature range and the results are compared to established data. Each unit that is shipped out is tested to make sure that it meets accuracy specifications and that each compo­nent functions properly.
-
-
-
35
Page 46
Page 47
7 Front Panel Operation
This chapter explains operation of the 1575/1590 thermometer in detail using the front panel buttons and soft-keys. Aside from the display contrast adjust ment which is controlled by dedicated buttons on the front panel, all instrument functions are controlled using the soft-key menu system. This menu system is outlined in the Figure 4 on page 38. In each sub-menu the RETURN soft-key (not shown) returns to the next higher menu.
7.1 Channel Menu
The CHANNEL menu contains functions that control the input channels. You can select either channel or you can have the Super-Thermometer alternate be tween them. Within the channel menu there are three soft-keys: INPUT 1, IN PUT 2, and ALT N. If the 2575 or 2590 Scanner is connected to the thermometer, two additional soft-key functions, SCANNER and SCANNER SEQ, appear in this menu (refer to the 2575 and 2590 Scanner Sections).
7 Front Panel Operation
Channel Menu
-
-
-
7.1.1 Measure Input 1
Press the INPUT 1 soft-key to select Input 1 for measurement. Measurements made with this channel will use the probe setup parameters specified for Input 1 (refer to Section 7.4, “Probe Menu”).
37
Page 48
1575/1590 Thermometer Readout
User’s Guide
MAIN MENU
CHANNEL MENU *
INPUT 1 INPUT 2 ALT N
SAMPLE MENU
RUN/STOP SAMPLE N SAMPLE TIMING DIGITAL FILTER RESET STATS
MEMORY MENU
STORE SAMPLE WRITE VALUE VIEW MEMORY CLEAR MEMORY RUN/PAUSE
PROBE MENU
INP 1/INP 2 * SELECT PROBE EDIT PROBE CAL PROBE
PROBE DISK
DISPLAY MENU
DATA/GRAPH SELECT DISPLAY EDIT DISPLAY DISPLAY RESOL Set display resolution CLEAR GRAPH
SYSTEM MENU
UNITS PARAM MENU
DISK MENU
COMMUNICATIONS MENU
SYSTEM CAL
*These menus and soft-keys change when the 2575 or 2590 scanner is connected.
1590 only
Figure 4 Menu Tree
Select Input 1 for measurement Select Input 2 for measurement Alternate between Input 1 and Input 2
Start or stop continuous sampling Start sampling for N measurements Set conversion time, sample interval, and integration period Select digital filter, response time, and reset threshold
Reset maximum, minimum, average, spread, standard deviation
Store present measurement into a memory register Enter a value into a memory register View memory register contents Clear all memory registers Pause sampling temporarily or resume sampling
Select channel for probe editing Select probe setup for channel Edit parameters for selected probe
CAL TPW CAL ITS-90 Calibrate probe ITS-90 coefficients
1.414xCURRENT
0.707xCURRENT CONV TEST Test probe characterization coefficients
SAVE PROBE READ PROBE SAVE ALL READ ALL FORMAT DISK
TIME SAVE PARAMS LOAD PARAMS RESET PARAMS SCREEN PARAMS Set up screen saver, brightness, color
SETUP RECORD/STOP RECORD N FORMAT DISK Format disk RUN APP
SERIAL IEEE488 PRINTER ANALOG OUTPUT Set analog output parameters
SET EXT SET INT CAL REFER SECURITY INFO Show system information
Calibrate probe water triple point resistance
Multiply current by 1.414 Multiply current by 0.707
Save probe setup to disk Read probe setup from disk Save all probes to disk Read all probes from disk Format disk
Select type of display Select display setup Edit display setup
Clear data from graph (graph type display only)
Select units - C, K, F, ohms, ratio, W
Set date and time Save all parameters to disk Read all parameters from disk Reset all parameters to factory defaults
Setup disk sampling parameters Start or stop data logging to disk Record a number of measurements to disk
Run application program on disk
Set serial setup and sampling parameters Set IEEE-488 setup and sampling parameters Set parallel interface sampling parameters
Set external reference resistance Set internal reference resistances Calibrate reference resistor Set security lockout for parameters
38
Page 49
7.1.2 Measure Input 2
Press the INPUT 2 soft-key to select Input 2 for measurement. Measurements made with this channel will use the probe setup parameters specified for Input
2.
7.1.3 Alternate Channels
Press the ALT N soft-key to select the alternating measurement mode. With this mode the channels are automatically switched after a certain number of mea surements are made on each. For example, if n is set to 5 then five measure ments will be taken with Input 1, the next five will be taken with Input 2, and the sequence repeats.
The alternate mode works with either of the sampling modes described below in Section 5.2. If continuous sampling is selected, the channels will alternate indefinitely while measurements are made. If SAMPLE N is selected, the chan nels will be switched alternately until the total number of measurements reaches the number specified with SAMPLE N.
7.2 Sample Menu
The SAMPLE MENU contains functions that allow you to control the measure­ment process including timing, digital filter, and statistics. The soft-keys within the sampling menu are RUN/STOP, SAMPLE N, SAMPLE TIMING, DIGI­TAL FILTER, and RESET STATS.
7 Front Panel Operation
Sample Menu
-
-
-
39
Page 50
1575/1590 Thermometer Readout
User’s Guide
7.2.1 Sampling Run/Stop
The RUN/STOP soft-key is used to start and stop sampling. The soft-key label will indicate the present status by graying RUN or STOP.
The run/stop sampling function overrides any of the or disk sampling functions (see Section 7.6.3 and 7.6.4). As long as sampling is inhibited with the STOP soft-key, no new measurements will be transmitted via the communications in terfaces or written to disk.
7.2.2 Sample N
The SAMPLE N soft-key programs the Super-Thermometer to make a certain number of measurements then stop. You will be prompted to enter the desired number of measurements. Press Enter after setting the number.
The 1575/1590 will begin making measurements and then stop after the speci fied number of measurements has been obtained. The SAMPLE N soft-key will indicate the number of samples remaining. The RUN/STOP soft-key will indi cate (by graying) RUN while the samples are being taken and STOP when fin­ished. Channels may be switching manually or automatically without affecting the total number of measurements made.
7.2.3 Sample Timing
The SAMPLE TIMING soft-key allows the user to set any of the sample timing parameters which include conversion time, sample interval, and integration pe­riod. (Access to the sample timing parameters can be locked out as explained in Section 7.6.5.4.) Figure 5 below shows how these parameters affect the mea­surement process.
-
-
-
40
new measurement new measurement
Figure 5 Sample Timing
conversion time
sample interval
samples
averaged
integration
period
The conversion time is the time in seconds required for the thermometer to make a single unprocessed (raw) measurement. Normally this is set to two sec onds. It can be increased if more settling time is needed after the current is re versed (see Section ). In some cases a PRT may have excessive reactance, leakage, or dielectric absorption and a longer than normal delay may be neces
time
-
-
-
Page 51
7 Front Panel Operation
Sample Menu
sary. Four seconds is recommended when measuring resistance above 100kΩ. To determine if there is error in the measurement due to inadequate delay, com pare measurements using the normal conversion time with measurements using twice the normal conversion time. The range of the conversion time is 2 to 10 seconds.
The sample interval is the time in seconds required to process and display a new measurement on the screen. For example, if you are interested in making a measurement once every hour, you can set the sample interval to 3600s. Nor mally the sample interval will be set to two seconds so measurements are dis
-
­played as often as possible. The range of the sample interval is 2 to 32000 seconds but must be equal to or greater than the conversion time. The sample interval affects the time scale of the graph display (see Section 7.5.1).
The integration period is the length of time in seconds over which one or more raw samples are averaged to obtain one measurement for display. For example, suppose you want to make a measurement once every hour. If you want the measurement to be the average of only the last one minute you would set the integration time to 60 seconds. If you want the measurement to be the average of all raw samples during the hour you would set the integration period to 3600 seconds. The range of the integration period is 2 to 32000 but must be between the conversion time and sample interval inclusive.
-
7.2.4 Digital Filter
The digital filter helps eliminate noise in the measurements and improves reso­lution. It does this by averaging, in effect, the present measurement with past measurements. The result of the digital filtering on a series of measurements is illustrated in Figure 6 below.
Without Filter With Filter
Figure 6 Smoothing Effect of the Digital Filter
The DIGITAL FILTER soft-key configures the digital filter, selecting the type of filter, response time, and reset threshold. (Access to the digital filter parame ters can be locked out as explained in Section 7.6.5.4.)
The filter type is selected with FILTER TYPE. Two types of digital are avail able: moving average (AVERAGE) and exponential (EXP). The moving aver age filter, also called a sliding average filter, averages all of the past
-
-
-
41
Page 52
1575/1590 Thermometer Readout
User’s Guide
measurements taken within the time specified by the response time. The expo
­nential filter simulates a 1-pole analog filter where more recent measurements are given more weight than earlier measurements. The filter can also be dis
-
abled by selecting NONE.
The response time of the filter is adjusted with RESPONSE TIME. For the moving average filter the response time is the length of time in seconds over which measurements are averaged. For the exponential filter the response time is the time required for the output to reach 63% of its final value with a step change in the input. There is a tradeoff between response time and smoothing effectiveness. Increasing the response time smooths the measurements and im
­proves the effective resolution. However, if the response time is too long you will not be able to accurately measure changing temperatures. The response time of the filter should be set to the smallest value that achieves the desired resolution and accuracy.
To help allow the filter to track changing temperatures while still removing noise, the digital filter has an automatic reset feature. The filter will reset if the difference between the raw measurement and the previous filtered measurement exceeds a specified reset threshold. This limits the amount of error the filter can cause. When the filter resets its memory of past measurements is erased. The effect of automatic reset is illustrated in Figure 7. Notice that with reset the measurement follows more closely the actual temperature during a sudden large change in temperature. If the input changes continuously the output may appear to stair-step as the filter attempts to track the input.
42
reset threshold
filter with reset
without filter
filter without reset
Figure 7 Effect of Automatic Reset on Filtered Measurements
The reset threshold can be adjusted within the DIGITAL FILTER window. The smaller the threshold the closer the measurements will be to the actual tempera
-
Page 53
7 Front Panel Operation
Memory Menu
ture during sudden large temperature changes. However, the threshold should not be too small or the reset may be triggered by ordinary noise making the fil ter ineffective. The message “FILTER RESET” appears below the temperature display whenever the filter resets to help alert you if there is a problem. The threshold should be set so it is just larger than the maximum deviation caused by noise. For many applications a threshold value of 0.01 to 0.1 is ideal. In cases where the temperature variations are greater, the threshold value should be increased. For temperature units (°C, K, °F) the threshold value is the devia tion in temperature that will cause reset. For resistance measurements the threshold is given as a percentage of the measured resistance or reference resis tance, whichever is greater. For ratio measurements the threshold is given as a percentage of the measured ratio or one, whichever is greater.
NOTE: The filter threshold value must be greater than the noise in the
measurement or the filter will be ineffective.
-
-
-
The digital filter can also be reset manually at any time by pressing the DIGI TAL FILTER soft-key and then pressing Exit. The filter resets whenever the fil­ter window is exited. The filter will also be reset if the system units are changed unless the units are only changed from one type of temperature units to another, e.g., °C to °F.
7.2.5 Reset Statistics
Pressing the RESET STATS soft-key causes the statistical registers to be cleared. Any subsequent statistical calculations comprise only measurements obtained from that time forward. Separate statistical registers and calculations are used for each channel. The reset function resets the statistics of both chan­nels simultaneously.
A variety of statistical functions can be applied to the measurements and the re sults displayed in various fields on the display. The available statistical func tions are maximum, minimum, spread, average, and standard deviation. (For more information on setting up the display to show statistical calculations, see Section 7.5).
7.3 Memory Menu
The MEMORY menu allows you to store measurements or arbitrary values in memory. The numbers can then be displayed on the screen or used as refer ences for difference measurements (refer to Section 7.5.3). There are 100 avail able memory locations.
-
-
-
-
-
43
Page 54
1575/1590 Thermometer Readout
User’s Guide
The memory menu includes the soft-keys STORE SAMPLE, WRITE VALUE, VIEW MEMORY, CLEAR MEMORY, and RUN/PAUSE.
44
7.3.1 Store Sample
The STORE SAMPLE soft-key places the most recent measurement in a mem­ory location you specify. Since the measurement may be updated while you are trying to store it in memory, you may first want to pause sampling temporarily using the RUN/PAUSE soft-key. When you press the STORE SAMPLE soft-key, a small window appears requesting you to enter a location number. Enter a number between 1 and 100. The measurement displayed at the top of the screen will be stored in memory along with its units.
7.3.2 Write Value
With WRITE VALUE you can enter any number into any memory location. This may be useful if you want to display the differences between measure ments and a particular value (see Section 7.5.3). When you press this key a small window appears that allows you to enter the memory location you want, the value, and the units. The value will stay in memory until you overwrite it or clear it.
7.3.3 View Memory
The VIEW MEMORY function allows you to scan the memory to view the data it contains. This is useful for finding specific data or unused locations. When you press the key a small window appears requesting you to enter the begin ning location. A larger window then appears showing the contents of that loca
-
-
-
Page 55
tion and the next four. You can use the or scroll buttons to scroll the list. Press Exit when you are finished viewing.
7.3.4 Clear Memory
The CLEAR MEMORY soft-key clears the contents of memory, setting all the values to 0.0°C. When you press the key you will be prompted to confirm. Press Enter to clear the memory or press Exit to cancel.
7.3.5 Run/Pause Sampling
When the RUN/PAUSE soft-key is pressed, the instrument will temporarily stop taking measurements. This is useful for holding the last measurement until it can be stored in memory as explained in Section 7.3.1. Measuring will re sume when the key is pressed again or the memory menu is exited. The word RUN or PAUSE is grayed out to indicate the current status.
7.4 Probe Menu
The PROBE menu includes functions for specifying probe characterization co­efficients and measuring parameters. It also includes functions for calibrating PRTs according to ITS-90. Access to the probe menu can be locked out as ex­plained in Section 7.6.5.4.
7 Front Panel Operation
Probe Menu
-
Probe parameters are defined and stored in memory together as a set. Up to 50 probe definitions can be stored in memory. These are referred to by number, e.g., “PROBE 1". Any input channel can use any of the probe definitions.
45
Page 56
1575/1590 Thermometer Readout
User’s Guide
The probe menu includes INP 1/INP 2, SELECT PROBE, EDIT PROBE, CAL PROBE, and PROBE DISK soft-keys. If the 2590 Scanner is connected to the 1575/1590, the first soft-key will appear as SELECT CHAN instead. To change the probe definition for a particular channel, first use INP 1/INP 2 to select the channel to edit then use SELECT PROBE to assign a different probe number or EDIT PROBE to change the parameters.
7.4.1 Select Channel
The INP 1/INP 2 soft-key allows you to select the channel for which you want to edit the probe parameters. This function does not select a channel to mea sure. When you first enter the probe menu, the channel that is currently being measured is selected. An asterisk (*) appears next to INP 1 or INP 2 to indicate the selected channel. Press this soft-key to select the other channel for editing. Note that this soft-key does not actually change the channel being measured.
7.4.2 Select Probe
The SELECT PROBE soft-key allows you to choose which probe definition to use with the selected channel. When you press this soft-key a list of probe defi-
nitions appears with the serial numbers of the probes. Use the U and D but­tons to move to the desired probe number and press Enter. The probe window will be updated to show the parameters for the selected probe. One restriction is you can not select a probe for INPUT 2 that has EXTERNAL selected as the reference.
-
46
7.4.3 Edit Probe
The EDIT PROBE soft-key is used to set the parameters included in a probe definition. You will be prompted to enter the probe number to edit. Enter two digits for the probe number (enter 01 for probe number 1, 20 for probe number
20). Then press the Enter button twice. After a probe is selected a new window will appear showing the probe parameters. Use the U and D buttons to move
up and down the list. The window will scroll if needed. You can edit any or all of the parameters in any order. When you are finished, press Exit to return to the menu.
NOTE: Be sure to press Enter after changing a parameter to store the
new value.
These parameters, described in the following sections, can be specified for the probe:
Serial number
Conversion type (ITS-90, Callendar-Van Dusen, etc.)
Reference resistor
Current
Page 57
Standby current (1590 with 2590 only)
Three-wire offset
Sub-range (ITS-90 only)
Characterization coefficients applicable to conversion type and selected
subrange
7.4.3.1 Serial Number
For SERIAL NO. you can enter up to ten characters that consist of digits from 0 to 9, upper case letters from “A” to “Z”, “.” and “-”. To enter a letter, repeat edly press the Exp button until the desired letter appears.
7.4.3.2 Temperature Conversion
CONVERSION specifies what algorithm is used to calculate temperature from the measured resistance of the probe. Options include ITS-90, IPTS-68, Callendar-Van Dusen (CVD), POLYNOMIAL, and thermistor. You can also
choose to display resistance R(Ω), ITS-90 ratio W(T overriding the system units (see Section 7.6.1). Each of the conversion options is explained in more detail below. The probe characterization parameters that appear below CONVERSION in the edit list will depend on the type of conver­sion you select.
7 Front Panel Operation
), or resistance ratio,
90
Probe Menu
-
NOTE: The unit in which a measurement is displayed can be affected by both the conversion type and the System Units settings. Ratio or ohms set­tings will override temperature units.
7.4.3.2.1 ITS-90
Most PRTs and SPRTs are characterized according to the International Temper ature Scale of 1990 (ITS-90). If you select ITS-90 for CONVERSION, you must then select one or two sub-ranges then enter the coefficients for the sub-ranges. You may select both a low temperature sub-range and a high tem perature sub-range. At temperatures where the high and low ranges overlap, the low range takes precedence. If the probe is calibrated for only one range, set the unused high or low range to NONE. If both sub-ranges are set to NONE the temperature will be calculated using the ITS-90 reference function.
If the temperature falls outside the selected sub-range or if no sub-ranges are selected a warning message will appear: “WARNING: SUBRANGE EX CEEDED”. This indicates that the calibration of the probe does not apply at this temperature and it may not be accurate. If the 1575/1590 is unable to cal culate temperature because the measured resistance is too low or too high or
there is an error with the coefficients “..........” will be displayed indicating the
measurement is out-of-range.
-
-
-
-
47
Page 58
1575/1590 Thermometer Readout
User’s Guide
Table 2 shows the various ITS-90 subranges with the temperature range and co efficients. All subranges also include R(TPW) (the resistance at 273.16K) as a coefficient. For details about the ITS-90 see NIST Technical Note 1265, Guide
lines for Realizing the International Temperature Scale of 1990.
Tabl e 2 ITS-90 Subranges and Coefficients
Subrange Number Temperature Range Coefficients
Low Range
1 13.8033 to 273.16K a
2 24.5561 to 273.16K a2,b2,c1-c
3 54.3584 to 273.16K a3,b3,c
4 83.8058 to 273.16K a4,b
5 234.3156 to 302.9146K a5,b
High Range
6 273.15 to 1234.93K a
7 273.15 to 933.473K a7,b7,c
8 273.15 to 692.677K a8,b
9 273.15 to 505.078K a9,b
10 273.15 to 429.7485K a
11 273.15 to 302.9146K a
1,b1,c1-c5
3
1
4
5
6,b6,c6,d6
7
8
9
10
11
-
-
48
7.4.3.2.2 IPTS-68
The IPTS-68 option is for probes calibrated to the International Practical Tem perature Scale of 1968. IPTS-68 probes are often only calibrated over a limited temperature range and some of the coefficients may not be available (e.g. a b
). Enter 0.0 for unused coefficients. You can select the ITS-90 scale conver
3
sion by setting the SCALE parameter to T90. This causes the temperature to be adjusted by an amount equal to the difference between the IPTS-68 and ITS-90 temperatures. (The accuracy of the scale conversion is ±0.001°C between 0 and 630°C, ±0.002°C between –200 and 0°C, and ±0.003°C between –259 and –200°C.)
7.4.3.2.3 Callendar-Van Dusen
The Callendar-Van Dusen (CVD) conversion can be used with some industrial probes such as those that conform to the DIN-43760, IEC-751, or ASTM E1137 standards. The Callendar-Van Dusen equation is shown below. The stan dard coefficients for DIN-43760, IEC-751, or ASTM E1137 are shown in Table
3.
-
,
3
-
-
Page 59
rt C
([ ])°=
Tabl e 3 DIN-43760/IEC-751/ASTM E1137 Callendar-Van Dusen Coefficients
Coefficients
R
0
α δ β
7.4.3.2.4 Polynomial
The POLYNOMIAL conversion can be used with non-standard temperature sensors that you fit to a polynomial equation. The equation has the form,
tR C A AR AR AR A R()[ ]°=+ + + +
Rt
⎪ ⎪ ⎨
R
⎪ ⎩
+−
1
0
100 100
⎩ ⎧
1
+−
αδ βt
0
Coefficient Values
100.0
0.00385055
1.4998
0.109
01 233
100 100
1
⎛ ⎜ ⎝
tt t t
⎛ ⎜ ⎝
10αδ
1
−−
3
K
10
1001100
10
⎞ ⎟ ⎠
t
7 Front Panel Operation
t
3
t 0
Probe Menu
<
If the desired polynomial equation is less than order ten, set the unnecessary higher order coefficients to 0.
7.4.3.2.5 Thermistor
The thermistor conversions use the Steinhart-Hart equation or its inverse as shown below. The 1575/1590 can use coefficients for either of the two equa tions. If your probe has coefficients for temperature as a function of resistance, select THRM-T(R). If the coefficients are for resistance as a function of tem perature equation, select THRM-R(T). If there is no coefficient for the second order term set A
T R K A A InR A InR A InR()[ ] [ ( ) ( )]=+ + +
RTK B BT BT BT( [ ]) exp[ ]= +++
7.4.3.2.6 Resistance
The R(Ω) option causes the measurement to be displayed as a resistance in ohms rather than temperature. The resistance is displayed for the probe regard less of the selected system units. There are no coefficients for this conversion option.
7.4.3.2.7 W(T90)
The W(T90) option causes the measurement to be displayed as an ITS-90 resis tance ratio rather than temperature. The ratio is displayed for the probe regard
or B2to 0.
2
01 2
−−−
0112
-
-
2
2
31
3
3
3
-
-
-
49
Page 60
1575/1590 Thermometer Readout
User’s Guide
less of the selected system units. For the ratio to be displayed properly, the R(TPW) value must be accurate. You may use the CAL TPW soft-key to cali brate this parameter (see Section 7.4.4.1).
7.4.3.2.8 Ratio
The RATIO option causes the measurement to be displayed as resistance ratio rather than temperature. The ratio is the measured resistance divided by the ref erence resistance. There are no coefficients for this conversion option.
7.4.3.3 Reference
REFERENCE specifies which reference resistor to use with the probe. You can select among the four internal reference resistors or you can select an external resistor. You should choose the reference resistor closest to the resistance to be measured (see Table 4). Using the wrong reference resistor can result in exces sive noise in the measurement and poor accuracy. For 25 ohm and 100 ohm PRTs and SPRTs the 100 ohm internal reference resistor or a 100 ohm external reference resistor is recommended.
TIP: Program the data window to show the reference resistor so you can
easily see that it’s set correctly (see Section 7.5.3).
-
-
-
Tabl e 4 Reference Resistor Selection
Input Resistance
0to1 1
1to10 10
10 to 1000 100
1kto500k 10k
If you use an external standard resistor as the reference, be sure you have en tered its correct resistance with the SET EXT function in the SYSTEM - REF ERENCE menu (see Section 7.6.5.1). An external reference resistor must be connected to Input 2 (see Section 5.1).
NOTE: When using an external reference resistor, you must enter its pre cise value with the SET EXT function in order for the current and the mea sured resistance to be accurate.
7.4.3.4 Current
The driving current for the probe should be optimized to achieve the most accu rate measurement possible. If the current is too high there may be too much er ror in the measurement because of self-heating. If the current is too low noise, drift in the measurements may be excessive and accuracy will be poorer. The
Internal Reference
-
-
-
-
-
-
50
Page 61
optimum current depends on the probe. Most 25Ω SPRTs and 100Ω PRTs and RTDs work well with 1.0 mA. High-temperature 0.25Ω SPRTs may require 10
mA. Thermistor probes use a much smaller current, typically 0.01 mA. Consult the specifications for your probe for the recommended driving current.
TIP: Program the data window to show the current so you can easily see
that it’s set correctly (see Section 7.5.3).
The range and resolution with which the current can be set will depend on the reference resistance and the input resistance. The actual current may differ slightly from the setting because of limited resolution and accuracy of the cur rent source. Current may also be limited if the resistance is large.
7.4.3.5 Standby Current
This option is only enabled when the 1590 is used with the 2590 Scanner. It sets the standby current applied to the probe while another is being measured. Refer to Section 11, 2590 Scanner, for instructions on using the standby cur­rent feature of the scanner.
7.4.3.6 Three-wire Offset
7 Front Panel Operation
Probe Menu
-
The three-wire offset (3W OFFSET) parameter is used when measuring tem­perature using two-wire or three-wire RTDs. It adjusts the measured resistance to compensate for lead resistance. This is not necessary with four-wire probes since the lead resistance is canceled automatically.
CAUTION: When using four-wire probes the 3W OFFSET value must be
set to zero or the measurement may not be accurate.
When using three-wire or two-wire probes, the 1575/1590 is unable to directly cancel the lead resistance. The measured resistance will be too high by the amount of resistance of one lead if it’s three-wire or two leads if it’s two-wire. The true resistance can only be calculated by subtracting the lead resistance er ror from the measurement. The exact amount of resistance to subtract must be entered for the 3W OFFSET parameter.
The lead resistance error of a three-wire probe can be determined by measuring the resistance between the two common leads and dividing by two. Connect the probe to the input with one of the matched wires connect to C1 and P1 and the other connected to C2 and P2. Set up the measurement to measure resistance
(you may use the 100Ω reference and 1 mA of current).
NOTE: Be sure to divide the measured lead resistance by two to obtain
the correct 3W OFFSET value for a three-wire probe.
-
51
Page 62
1575/1590 Thermometer Readout
User’s Guide
7.4.4 Calibrate Probe
The probe calibration (CAL PROBE) sub-menu contains functions for calibrat ing probes with the triple point of water, calibrating ITS-90 coefficients, mea suring self-heating error, and testing the temperature conversion. These are CAL TPW, CAL ITS-90, 1.414 x CURRENT, 0.707 x CURRENT, and CONV TEST.
7.4.4.1 Calibrate TPW
The CAL TPW calibration function allows you to easily calibrate a probe at the triple point of water. It adjusts the ITS-90 R(TPW) value. The calibration will be performed on the probe being measured. The input channel must be set on one channel (not alternating or scanning). The probe must be using the ITS-90 conversion.
Obtaining a precise measurement is very important when calibrating the probe. It is recommended that you use filtering with a time constant of 20 to 60 sec onds. Be sure the filter threshold is high enough.
To begin the calibration, place the probe in the triple point of water cell and al­low the temperature to equilibrate. Press the CAL TPW soft-key. A window ap­pears asking for the immersion depth in millimeters, of the probe in the triple point of water cell. If you enter the value for the depth, the thermometer will compensate for the error in temperature caused by the hydrostatic pressure of the water. This error amounts to approximately 0.0002°C at a depth of 250 mm. If you do not care to correct for this small error, enter 0.
-
-
-
52
The Super-Thermometer will begin to measure the resistance of the probe. You may graph the measurements by pressing the GRAPH soft-key. When you are
Page 63
7 Front Panel Operation
Probe Menu
satisfied that the measurement is stable and accurate, press the STOP soft-key (or Enter).
When STOP is pressed, the final value of the measurement will be shown. Press the STORE soft-key (or Enter) to save the new value. The measured re­sistance will replace the previous value for the water triple point resistance R(TPW) for the given probe.
53
Page 64
1575/1590 Thermometer Readout
User’s Guide
7.4.4.2 Calibrate ITS-90
The CAL ITS-90 function facilitates the calibration of probes to the ITS-90. It aids you in collecting the necessary resistance and temperature data and calcu lating the ITS-90 deviation function coefficients. Calibration can be performed over any of the ITS-90 subranges between Range 4 and Range 11 inclusive. The resistance-temperature data can be measured or entered using the front panel keys. Once the calibration coefficients are calculated, the 1575/1590 can print the results with a list of the derived coefficients and a resistance versus temperature table.
Calibration of an SPRT to the ITS-90 requires measuring the resistance of the probe at various fixed points for primary calibrations, or at known temperatures near the fixed-point temperatures for comparison calibrations. The fixed points required for the various subranges are listed in Table 5. Also shown is the rec ommended range for each point for comparison calibrations. For details on the fixed points and the ITS-90 in general, refer to NIST Technical Note 1265,
Guidelines for Realizing the International Temperature Scale of 1990.
Tabl e 5 ITS-90 Calibration Points
-
-
Subrange Number Temperature Range Fixed Points
Low Range
4 83.8058 to 273.16K Argon (Ar), 83.8058K
Mercury (Hg), 234.3156K Water (TPW), 273.16K
5 234.3156 to 302.9146K Mercury (Hg), 234.3156K
Water (TPW), 273.16K Gallium (Ga), 302.9146K
High Range
6 273.15 to 1234.93K Water (TPW), 273.16K
Tin (Sn), 505.078K Zinc (Zn), 692.677K Aluminum (Al), 933.473K Silver (Ag), 1234.93K
7 273.15 to 933.473K Water (TPW), 273.16K
Tin (Sn), 505.078K Zinc (Zn), 692.677K Aluminum (Al), 933.473K
8 273.15 to 692.677K Water (TPW), 273.16K
Tin (Sn), 505.078K Zinc (Zn), 692.677K
9 273.15 to 505.078K Water (TPW), 273.16K
Indium (In), 429.7485K Tin (Sn), 505.078K
Comparison Temperature Range
65 to 210K 225 to 250K 263 to 283K
225 to 250K 263 to 283K 290 to 325K
263 to 283K 470 to 550K 600 to 800K 850 to 1050K 1100 to 1400K
263 to 283K 470 to 550K 600 to 800K 850 to 1050K
263 to 283K 470 to 550K 600 to 800K
263 to 283K 350 to 450K 470 to 550K
54
Page 65
7 Front Panel Operation
Probe Menu
Subrange Number Temperature Range Fixed Points
10 273.15 to 429.7485K Water (TPW), 273.16K
Indium (In), 429.7485K
11 273.15 to 302.9146K Water (TPW), 273.16K
Gallium (Ga), 302.9146K
Comparison Temperature Range
263 to 283K 350 to 450K
263 to 283K 290 to 325K
The CAL ITS-90 sub-menu has individual soft-keys for each of the various steps for calibrating an SPRT. The soft-keys are SETUP, MEASURE, DATA, CALCULATE,andPRINTREPORT.ThestepstoanITS-90calibrationare summarized here.
1. Set up the options for the probes to be calibrated using the SETUP soft-key. This includes stating which probes and channels to use for the probes to be calibrated and the reference SPRT. You also specify the se
­quencing parameters. The probe specific parameters such as the serial number, subrange, current, and reference resistor must be set up previ
-
ously using EDIT PROBE in the PROBE MENU.
2. Measure the resistance of the probes at each calibration temperature us­ing the MEASURE soft-key. The temperature may be measured using a reference SPRT or specified as a fixed-point temperature.
3. After obtaining the resistance and temperature measurements you may examine the data using the DATA soft-key.
4. The ITS-90 coefficients are calculated from the resistance-temperature data using the CALCULATE soft-key.
5. A report is generated for the probe using the PRINT REPORT soft-key. The report shows the new ITS-90 coefficients along with applicable data. The report may optionally include a table of resistances for various temperatures.
The ITS-90 calibration program is designed to be flexible enough to allow for a variety of calibration methods. The calibration setup options will depend on the method to be used. The next sections will describe five possible methods. Fol lowing that the calibration steps and their corresponding soft-keys will be ex
-
-
plained. The five calibration methods to be described are as follows:
1. Calibration using only fixed points
2. Comparison calibration with manual sequencing
3. Comparison calibration with automatic sequencing
4. Calibration using SPRT resistance comparison
5. Calibration with W values using fixed points
7.4.4.2.1 Calibration Using Only Fixed Points
In this method of calibration, the resistances of the probe or probes being cali brated are measured at known fixed-point temperatures. The temperatures are
-
55
Page 66
1575/1590 Thermometer Readout
User’s Guide
entered, not measured. Normally, the fixed points that are used will be the stan dard specified fixed points for the ITS-90 such as the triple point of water, the freezing point of tin, etc.
Tabl e 6 Fixed-point Calibration Setup
REFR: FIXED POINTS
TEMP: ONCE
MEAS: PROMPT
The setup options for the fixed-point calibration are shown in Table 5. Options not indicated are irrelevant. The calibration may done for just one probe, for multiple probes using just one channel, or multiple probes using multiple channels.
7.4.4.2.2 Comparison Calibration With Manual Sequencing
With this method of calibration, the probes are calibrated in baths or other sta­ble temperature sources and the temperatures are measured with a reference SPRT. Fixed points may be used for one or more of the calibration points. User input will be required to initiate the resistance measurement at each point. The calibration may be done either for just one probe, for multiple probes using just one channel, or multiple probes using multiple channels.
The setup parameters for the comparison calibration method are shown in Table
7. Options not indicated are irrelevant. In the table, PROBE 2 is shown set up as the reference SPRT on INPUT 2 as an example. A different probe and chan­nel may be used instead. For best accuracy, the temperature measurement should be repeated between each probe’s resistance measurement (TEMP set to REPEAT).
-
56
Tabl e 7 Manual Comparison Setup
REFR: PROBE 2
CHANNEL: INPUT 2
TEMP: REPEAT
MEAS: PROMPT
RREF: NORM
7.4.4.2.3 Comparison Calibration With Automatic Sequencing
With this method of calibration, the probes are calibrated in a bath or other sta ble temperature source and the temperatures are measured with a reference SPRT. Fixed points may be used for one or more calibration points. All the measurements at each calibration point are made automatically with a specified amount of delay between each. This method is useful when using the 1575/1590 in conjunction with the 2590 Scanner for calibrating a large number of probes simultaneously.
-
Page 67
The setup parameters for the automatic sequencing calibration method are shown in Table 8. Options not indicated are irrelevant. In the table, PROBE 2 is shown set up as the reference SPRT on INPUT 2 as an example. A different probe and channel may be used if desired. For best accuracy, the temperature measurement should be repeated between each resistance measurement (TEMP set to REPEAT).
Tabl e 8 Automatic Sequencing Calibration Setup
REFR: PROBE 2
CHANNEL: INPUT 2
TEMP: REPEAT
MEAS: TIMED
DELAY: 4.0
RREF: NORM
7.4.4.2.4 Calibration Using SPRT Resistance Comparison
With this method of comparison calibration, the resistances of the probes are measured using the reference SPRT as the reference resistor. This can result in improved accuracy when the temperature of the bath is drifting. The calibration proceeds as follows: First, the resistance of the reference SPRT is measured and its temperature is calculated. Then, the resistance of the probe to be calibrated (UUT) is measured using the SPRT as the reference resistor. The resistance of the UUT is calculated by multiplying the measured ratio by the previously measured SPRT resistance. If the bath is drifting, the resistances of the probe and SPRT will change in the same proportion so the ratio, and thus the mea­sured resistance, will remain constant. Likewise, any error in measuring the bath temperature has little effect on the results since the recorded SPRT resis tancewillreflectthesameerror.Onefurther advantage of this method is that if the reference SPRT and UUT have similar nominal resistances the resistance measurements at each calibration point will be made with a resistance ratio close to 1:1. The 1575/1590 is most accurate at ratios near 1:1.
The setup parameters for the SPRT resistance comparison method are shown in Table 8. Options not indicated are irrelevant. PROBE 2 is shown set up as the reference SPRT but any other probe may be used instead. The channel of the reference SPRT must be INPUT 2 since it will also be used as the reference re
7 Front Panel Operation
Probe Menu
-
-
57
Page 68
1575/1590 Thermometer Readout
User’s Guide
sistor for the UUT. The temperature measurement need not be repeated be tween resistance measurements so TEMP is set to ONCE.
Tabl e 9 SPRT Resistance Comparison Calibration Setup
REFR: PROBE 2
CHANNEL: INPUT 2
TEMP: ONCE
MEAS: PROMPT
RREF: SPRT
7.4.4.2.5 Calibration With W Values
Calibration of an SPRT can also be performed using fixed points by obtaining the W value at each point. The W value is obtained by measuring the resistance of the SPRT at the fixed point and at the triple point of water at each point and then computing the ratio. This process is more time consuming but may give greater precision. It helps to reduce error caused by the SPRT drifting during the calibration process. The effect of self-heating can also be observed during calibration and zero-power coefficients produced.
For this calibration process, the W values must be obtained outside of the CAL ITS-90 function instead of using the MEASURE soft-key. The steps are as follows:
1. At each calibration point, you measure the resistance of the probe at the fixed point and again at the triple point of water. W is calculated by di­viding the fixed-point resistance by the triple point of water resistance,
RT
WT
()
()
=
RTPW
()
-
58
The self-heating can be determined while measuring resistance by in
­creasing the current by a factor of 1.414 and noting the increase in the measurement (see Section 7.4.4.3). Subtract the difference from the mea sured value (with normal current) to obtain the zero-power resistance.
2. You must now calculate resistances again to provide the necessary data for the 1575/1590 to calculate the coefficients. The resistances are calcu lated using the formula,
RT RTPWW T() ( ) ()=
The R(TPW) value used for these calculations is not to critical as long as the same value is used for each point. You may use the most recent water triple point resistance measurement.
3. Set up the probe definition in preparation for calculation of coefficients. Select the ITS-90 conversion and select the ranges you want to calibrate (see Section 7.4.3.2.1).
-
-
Page 69
4. Enter the CAL ITS-90 function and use SETUP. You must specify that
5. Enter the DATA function in CAL ITS-90. Select your probe number. En
6. Now you are ready to calculate coefficients. Enter the CALCULATE
7.4.4.2.6 Setup
The SETUP function is used to select options for the calibration process. Be fore beginning the setup, you must properly set up the probe definitions for each probe to be calibrated as well as the reference SPRT. You should set the serial number, conversion (ITS-90), reference resistor, current, and subranges. For the probe that will be used as the reference SPRT, you must also enter the correct characterization coefficients. Use the EDIT PROBE function in the PROBE MENU to set these parameters. Also, make sure that the sample timing and digital filter parameters are set appropriately in the SAMPLE MENU. To remind you that the probe parameters must be set before beginning the calibra­tion, a message appears when you begin SETUP. Press the NEXT soft-key (also Enter or Exit) to continue.
The setup parameters are grouped in four screens. The first screen sets up the temperature reference. You must specify the type of reference. If only fixed points will be used, select FIXED POINTS. If an SPRT will be used to measure the temperature at any or all of the points, select the probe number of the refer ence SPRT.
If an SPRT was selected as the temperature reference, specify the channel to use with the SPRT. It is often most convenient to use Input 2 unless the SPRT or other probes will be using an external reference resistor. Note: For an SPRT resistance comparison type calibration, you must select Input 2 for the refer ence SPRT. Press the NEXT soft-key (or Exit) to continue.
The next screen selects the calibration probes and their channels. For each probe to be calibrated, select the channel to use. If the 2590 scanner is con nected to the 1575/1590 Thermometer, the channels SCANNER 1 through SCANNER 10 will also be available along with INPUT 1 and INPUT 2. Do not choose INPUT 2 for any probe set up to use an external reference resistor. Se lect NO CAL for all probe numbers that will not be used. You will not be al lowed to set the channel for a probe number that is designated as the reference SPRT or one that does not have the conversion set to ITS-90 with a valid subrange. Press the NEXT soft-key (or Exit) to continue.
7 Front Panel Operation
Probe Menu
you want to calibrate the probe by selecting a channel for it (see Section
7.4.4.2.6). Since you will not actually be measuring the data, you may select any arbitrary channel as long as the probe is not set to NO CAL. You may ignore the other setup parameters since they pertain only to the measurement process.
ter the resistances (from Step 2) and temperatures for each calibration point (see Section 7.4.4.2.8). For the water triple point, enter the same resistance that was used to calculate the resistances in Step 2.
function. Select your probe number and press CALCULATE (see Sec
-
tion 7.4.4.2.9).
-
-
-
-
-
-
-
59
Page 70
1575/1590 Thermometer Readout
User’s Guide
The next screen sets the parameters relating to the measurement sequence. The user can specify whether the temperature is to be measured only once (ONCE) at the beginning for each calibration point or measured just prior to every resis tance measurement (REPEAT). Repeating the temperature measurement for each probe may give better accuracy if the temperature is drifting.
You can select the measurement mode as TIMED or PROMPT. The TIMED mode is useful when calibrating multiple probes simultaneously using the 2575 or 2590 Scanner. If TIMED is selected then the user must specify the DELAY time in minutes between measurements. After switching to the next probe, the 1575/1590 will measure for this amount of time before automatically recording the reading. The delay time should be at least five times the digital filter time constant to allow enough time for transients to settle. With the PROMPT mode, the thermometer will wait for you to press the STOP and STORE soft-keys to record the next measurement. After setting these parameters, press the NEXT soft-key (or Exit) to continue.
The last setup screen is used to set up SPRT resistance comparison calibrations. This screen only appears if a reference SPRT was specified in the first screen and its channel was set to INPUT 2. You specify whether to use the normal ref­erence resistor (determined by REFERENCE in EDIT PROBE) or the SPRT connected to Input 2 when measuring the resistances of the calibration probes. Select NORM if the resistances are to be measured in ratio to the normal refer­ence resistor (e.g. 100( internal resistor). Select SPRT if you desire to measure the resistances in direct ratio with the SPRT. This mode may help improve the accuracy when the temperature of the bath is drifting. If a fixed point is used for a calibration point, such as the triple point of water, the resistance measure­ments for that point will use the normal reference resistor rather than the SPRT. After setting this parameter press the RETURN soft-key (or Exit) to go back to the ITS-90 calibration menu. You are now ready to begin measuring using the MEASURE function.
-
60
7.4.4.2.7 Measure
The MEASURE function is used to collect the temperature and resistance mea surements for the calibration. This requires that the setup step be performed previously. The measurement process involves measuring the temperature and then the probe resistances at each of the calibration points. You may measure the calibration points in any order. The calibration point selection screen lists the points required for the subranges you selected and allows you to select
-
Page 71
7 Front Panel Operation
Probe Menu
which one you want to measure next. Press the SELECT soft-key (or Enter) to begin measuring at the indicated point.
If a reference SPRT was specified in the setup, the next screen will prompt to begin the temperature measurement. Press the START soft-key (or Enter) to be­gin measuring with the SPRT or press FIXED POINT if a fixed point is to be used or the temperature is already known. This screen will not appear if the TIMED mode was selected in the setup.
61
Page 72
1575/1590 Thermometer Readout
User’s Guide
When START is pressed the Super-Thermometer will begin measuring the tem perature. The temperature (in Kelvin) will appear in the window. You may graph the measurements by pressing the GRAPH soft-key. After allowing ade quate time for the measurement to stabilize, press the STOP soft-key (or Enter) to stop measuring. If the TIMED measurement mode was selected in the setup, measuring will stop automatically and the value stored after the specified length of time has elapsed.
-
-
62
If the thermometer detects that the noise of the measurements exceeds the filter threshold, a warning message will appear. To fix this, return, go into the FIL
­TER parameter window in the SAMPLING MENU, and increase the filter threshold value. Start over with the calibration.
When STOP is pressed the final value of the measurement will be shown. Press the STORE soft-key (or Enter) to save the value. The temperature measurement will be stored with the calibration data for the given probe or for all probes de pending on the TEMP setting in the setup. If the SPRT is on INPUT 2 and the
-
Page 73
7 Front Panel Operation
Probe Menu
SPRT was selected as the reference resistor, the resistance of the SPRT will be stored internally for use during the calibration probe resistance measurement.
If you want to use a fixed-point cell for the calibration point, press FIXED POINT instead of START. You will be requested to store the fixed temperature value. You may change the temperature if necessary using the numeric keys and Enter. Press the STORE soft-key (or Enter) to accept the value.
63
Page 74
1575/1590 Thermometer Readout
User’s Guide
If the temperature, whether measured or specified, is outside the recommended range, a message will appear to warn you. The range for each calibration point is listed in Table 4. Press the NEXT soft-key (or Enter) to accept the tempera ture or REDO to go back and measure or enter the temperature again.
The next step is to measure the resistance of the calibration probe or probes. The screen will show the probe number, serial number and channel for the first calibration probe. Make sure the probe is connected properly. Press the START soft-key (or Enter) to begin the measurement. This screen will not appear if the TIMED mode was selected in the setup.
-
64
The thermometer will begin measuring the resistance of the probe. If the RREF mode was set to SPRT, the resistance will be obtained by measuring the ratio of the probe resistance to the SPRT on INPUT 2 and multiplying by the SPRT re
­sistance measured in the temperature step. If the reference type was specified as NORM or if a fixed point is being used, the resistance will be measured with the reference resistor specified with REFERENCE in EDIT PROBE. You may graph the measurements by pressing the GRAPH soft-key. Press the STOP
Page 75
7 Front Panel Operation
Probe Menu
soft-key (or Enter) to stop measuring, or if TIMED mode was selected, wait for the measurement to be recorded automatically.
When STOP is pressed, the final value of the measurement will be shown. Press the STORE soft-key (or Enter) to save the value. The measurement will be stored with the calibration data as the resistance value of the given probe.
65
Page 76
1575/1590 Thermometer Readout
User’s Guide
If more than one probe was selected for calibration, the others will also be mea sured in the same way. If the REPEAT measurement mode was selected, the temperature measurement step will be repeated prior to the resistance measure ment for each probe.
Once all the probes are measured at this calibration point, the procedure will re turn to the calibration point selection screen. You may select the next point and continue with the measurements.
When all the measurements have been made at each calibration point, press the RETURN soft-key (or Exit) to return to the menu. At this point, you may ex amine the measurement data with the DATA function and calculate the ITS-90 coefficients using the CALCULATE function.
7.4.4.2.8 Data
The DATA soft-key can be used to examine the calibration data obtained in the measurement step. You may also enter resistance-temperature data obtained previously and use it to calculate coefficients. When you press this soft-key, a window appears showing the data for the first probe.
-
-
-
-
66
You may select any of the calibration probes using the LR buttons. Use the UD buttons to scroll down through the list. Data may be entered using the
numeric buttons. To exit the window, press the RETURN soft-key (or Exit).
Only data for those probes assigned a channel in SETUP can be viewed here. Also, only the points included in the selected ranges of the probes will be shown. If you intend to calculate coefficients using existing data, you must first select a probe number and set the probe parameters to specify ITS-90 conver
-
sion and select the ranges you want to calibrate (use EDIT PROBE in the
Page 77
PROBE MENU). You must also specify that you want to calibrate the probe in the SETUP function by selecting a channel. Since you will not actually be mea suring the data, you may select any arbitrary channel as long as the probe is not set to NO CAL. You may ignore the other setup parameters since they pertain to the measurement step.
7.4.4.2.9 Calculate
The CALCULATE soft-key allows you to calculate ITS-90 coefficients for a probe using the measured or entered resistance-temperature data. When you press the soft-key, a window appears to warn you that this step will change the coefficients of the calibration probes. This is a precaution against inadvertently changing the coefficients. Press NEXT (or Enter) to continue. Press RETURN (or Exit) to cancel and return to the menu.
Next, a window will show the first probe in the set and ask if you want to pro ceed with calculating the coefficients. You can select any probe in the set using
the LR buttons. Only those probes assigned a channel in the setup can be selected. Press the CALCULATE soft-key (or Enter) to continue with the calcu lation. If the calculations are successful, the old probe coefficients will be re­placed with new ones. To view the new coefficients, use EDIT PROBE in the PROBE MENU.
7 Front Panel Operation
Probe Menu
-
-
-
If there is an error with the data that causes the calculation to fail, or if the probe does not conform to the ITS-90 specifications for W(302.9146K), W(234.3156K), or W(1234.93K), a warning message will appear, “WARN
-
ING: THIS PROBE DOES NOT MEET ITS-90 SPECIFICATIONS”. When finished, press the RETURN soft-key (or Exit) to return to the menu. If desired
67
Page 78
1575/1590 Thermometer Readout
User’s Guide
you may print a report for each of the probes calibrated using the PRINT RE PORT function.
It should be noted that the particular values for coefficients the 1575/1590 cal culates are very sensitive to slight variations in the temperatures and resistances used. You may obtain different values for the coefficients each time you cali brate a probe even if the calibration is correct and accurate. The 1575/1590 may also produce coefficients that differ slightly from values calculated using other software because of differences in mathematical precision, matrix solu tion approaches, and recursion methods. You may test the validity of the calcu lated coefficients using the CONV TEST function (see Section 7.4.4.5). You should also compare measurements with the probe to those of a known accurate SPRT to verify that it’s properly calibrated.
7.4.4.2.10 Print Report
The PRINT REPORT soft-key can be used to print a calibration report for any of the probes that were calibrated. The report will show the probe serial num ber, the date, the current that was used, and the coefficients that were produced. The table can also include the measured resistance and temperature data and a resistance-temperature table. The report can be printed from the printer port, serial interface, IEEE-488 interface, or written to disk.
The first window that appears allows you to select the probe and output port for the report. You may use the LR buttons to select the probe. Only the probes
assigned channels in SETUP can be selected. Press Enter after changing the probe number.
For the output port, you may select among the printer parallel port, serial port, IEEE-488 port, or disk. Printing to disk may be useful if you want to customize the report using a word processor. When printing to disk, the filename of the re­port will be PROBEn.RPT where n is the probe number. If you print the report from the serial or IEEE-488 port, you should first check that the interface pa rameters (baud rate, address, etc.) are set correctly using the SERIAL or IEEE488 soft-keys in the SYSTEM - COMM MENU. Press Enter after chang ing the port.
If the parameters do not need to be changed, you may simply press the NEXT soft-key to go to the next screen. You may also press the RETURN soft-key from this screen to return to the menu after printing the reports.
The next window that appears asks whether or not you want to include the mea sured data with the report and allows you to select the units for this data. If the parameters do not need to be changed, you may simply press the NEXT soft-key to go to the next screen.
The next window asks whether or not you want to include the resistance-tem perature table with the report. You can also specify the units to use in the table, the minimum and maximum temperatures, and the temperature interval. If the parameters do not need to be changed, you may simply press the NEXT soft-key to go to the next screen.
-
-
-
-
-
-
-
-
-
-
68
Page 79
At this point, all the options are set and you may now print the report. Press the PRINT soft-key (or Enter) to print the report. Figures 8 and 9 show an example of a printed report and a page of the resistance versus temperature table. After the report is printed, the first screen will reappear allowing you to select an other probe to print.
7.4.4.3 1.414 x Current
This function in the CAL PROBE menu is useful for measuring the self-heating error with a probe. When the 1.414 x CURRENT soft-key is pressed the driv ing current will be increased by a factor of pated in the sensor by a factor of two and doubles the self-heating error. By observing the change in measured temperature with twice the power, you can determine the amount of self-heating error present with the original driving current. When increasing the current, be careful to observe the limits of the cur rent. It is useful to program one of the display fields to show the actual current (see Section 7.5.3).
7.4.4.4 0.707 x Current
The 0.707 x CURRENT soft-key can be used to restore the original current af­ter checking the self-heating. It reduces the driving current by a factor of
7 Front Panel Operation
. This increases the power dissi
2
Probe Menu
-
-
-
-
.
2
7.4.4.5 Conversion Test
The CONV TEST soft-key can be used to test the accuracy of the resis­tance-to-temperature conversion calculations and probe characterization coeffi­cients. It allows you to enter a resistance value and then it will display the corresponding temperature values in (C, K, and if ITS-90 or IPTS-68 is used, W values. You can compare the resistance and temperature to values given in a calibration report table for the probe to verify that the conversion parameters are set up properly. The CONV TEST function tests the probe parameters of the channel currently selected for measurement.
7.4.5 Probe Disk
The PROBE DISK sub-menu contains functions for writing and reading probe configuration data to and from the floppy disk. You can also format a new disk. The disk can be used to exchange information between the 1575/1590 ther mometer and PC compatible computers or it can be used to backup operating parameters for later use. The functions in the PROBE DISK menu are de scribed below.
7.4.5.1 Save Probe
The SAVE PROBE soft-key allows you to save a set of probe parameters to disk. When the key is pressed you will be prompted to enter the probe number (1 to 16). After you press Enter the parameters will be saved to a file with the name “PROBEn.CFG” where n is the probe number. If a file with this name al ready exists on the disk, you will be asked if you want to overwrite it. The
-
-
-
69
Page 80
1575/1590 Thermometer Readout
User’s Guide
probe parameter files are in ASCII text format so they can be easily read or written by other software applications.
7.4.5.2 Read Probe
The READ PROBE soft-key allows you to read a set of probe parameters from a disk. When the key is pressed, the disk will be read and a window will appear showing the probe files found on the disk. Use the buttons to select a file and press Enter.
70
A new window appears prompting you to confirm to replace the current param eters with those in the file. You may set a different probe with the parameters in the file by changing the probe number using the (( buttons. Press Enter to pro ceed or Exit to abort.
CAUTION: Using the READ PROBE or READ ALL functions will over
write current probe parameters.
7.4.5.3 Save All
You can use the SAVE ALL soft-key to save the parameters for all of the probes at once. Each probe will be given a different file with the name “PROBEn.CFG” where n is the probe number. Any probe parameter files saved previously will be overwritten.
-
-
-
Page 81
7.4.5.4 Read All
You can use the READ ALL soft-key to read all available probe parameter files from disk replacing the current parameters. The parameters will be assigned to the probes according to the filenames. Be careful with the READ ALL func tion; you will lose all of the current probe parameters and characterization coefficients.
7.4.5.5 Format Disk
Before a floppy disk can be used to store information, it must be formatted. You can format the disk on a PC-compatible computer or you can use the 1575/1590 FORMAT DISK function. The 1575/1590 can use 3.5" disks of ei ther 720 kb (double density) or 1.44 Mb (high density) format. Generally, disks of one type cannot be formatted to the opposite density. When you begin the
format operation, you must specify the density by using the LR buttons and pressing Enter.
The 1575/1590 will try to format the disk at the density you select. If the disk cannot be formatted at this density, a message will appear to tell you so. If this happens, select the other density and try again. Be careful with the format com-
mand - any data previously stored on the disk will be erased.
7 Front Panel Operation
Display Menu
-
-
7.5 Display Menu
Within the DISPLAY menu you can set up the display to show data in which you are interested. You can have the display show past measurements, memory data, differences between measurements, or the results of any of a variety of
71
Page 82
1575/1590 Thermometer Readout
User’s Guide
REPORT OF CALIBRATION
International Temperature Scale of 1990
Platinum Resistance Thermometer Serial Number: 2
20 March 1995
Current: 1.0 mA
Temperature(K) W(T90)
83.8071 0.21586101
234.3141 0.84415349
273.16 1.00000000
505.0759 1.89270529
692.6744 2.56876956
Coefficients
a4: -1.26508267E-04 b4: -8.61659096E-05 a8: -1.03200171E-04 b8: 9.448039801E-06
Resistance at 273.16 K: 25.57249 ohms
Figure 8 Example Calibration Report
72
Page 83
PROBE CALIBRATION TABLE Serial No: 2
T(C) W(T90) T(C) W(T90)
-190.0 0.21300745 -140.0 0.42767855
-189.0 0.21735026 -139.0 0.43189838
-188.0 0.22169381 -138.0 0.43611544
-187.0 0.22603772 -137.0 0.44032975
-186.0 0.23038161 -136.0 0.44454135
-185.0 0.23472518 -135.0 0.44875026
-184.0 0.23906813 -134.0 0.45295651
-183.0 0.24341017 -133.0 0.45716013
-182.0 0.24775107 -132.0 0.46136114
-181.0 0.25209058 -131.0 0.46555959
-180.0 0.25642851 -130.0 0.46975548
-179.0 0.26076466 -129.0 0.47394886
-178.0 0.26509886 -128.0 0.47813975
-177.0 0.26943095 -127.0 0.48232817
-176.0 0.27376080 -126.0 0.48651416
-175.0 0.27808826 -125.0 0.49069773
-174.0 0.28241324 -124.0 0.49487893
-173.0 0.28673562 -123.0 0.49905776
-172.0 0.29105531 -122.0 0.50323426
-171.0 0.29537223 -121.0 0.50740846
-170.0 0.29968632 -120.0 0.51158037
-169.0 0.30399751 -119.0 0.51590002
-168.0 0.30830574 -118.0 0.51991744
-167.0 0.31261096 -117.0 0.52408265
-166.0 0.31691315 -116.0 0.52824567
-165.0 0.32121225 -115.0 0.53240653
-164.0 0.32550826 -114.0 0.53656525
-163.0 0.32980114 -113.0 0.54072184
-162.0 0.33409087 -112.0 0.54487634
-161.0 0.33837745 -111.0 0.54902876
-160.0 0.34266087 -110.0 0.55317913
-159.0 0.34694113 -109.0 0.55732746
-158.0 0.35121822 -108.0 0.56147377
-157.0 0.35549214 -107.0 0.56561809
-156.0 0.35976290 -106.0 0.56976043
-155.0 0.36403052 -105.0 0.57390081
-154.0 0.36829500 -104.0 0.57803925
-153.0 0.37255635 -103.0 0.58217576
-152.0 0.37681460 -102.0 0.58631037
-151.0 0.38106975 -101.0 0.59044309
-150.0 0.38532183 -100.0 0.59457394
-149.0 0.38957085 -99.0 0.59870294
-148.0 0.39381684 -98.0 0.60283009
-147.0 0.39805983 -97.0 0.60695542
-146.0 0.40229983 -96.0 0.61107895
-145.0 0.40653686 -95.0 0.61520067
-144.0 0.41077097 -94.0 0.61932062
-143.0 0.41500216 -93.0 0.62343881
-142.0 0.41923047 -92.0 0.62755524
-141.0 0.42345593 -91.0 0.63166993
7 Front Panel Operation
Display Menu
Figure 9 Example Calibration Table Page
73
Page 84
1575/1590 Thermometer Readout
User’s Guide
statistical calculations. You can also set up the display to graph measurements over time. Sixteen different display setups can be stored in memory.
The display menu includes the soft-keys DATA/GRAPH, SELECT DISPLAY, EDIT DISPLAY, DISPLAY RESOL and CLEAR GRAPH (graph display only).
7.5.1 Data/Graph
The DATA/GRAPH soft-key allows you to select between two types of dis plays: data type and graph type. The data type display shows five lines of nu
-
-
74
Page 85
7 Front Panel Operation
Display Menu
meric information (see Figure 10). Each line can be programmed to show a specific type of information such as a statistical calculation or probe parameter.
Figure 10 Data Type Display
The graph type display shows a plot of past measurements (see Figure 11). You can set the upper and lower bounds of the graph. The time scale of the graph is
75
Page 86
1575/1590 Thermometer Readout
User’s Guide
determined by the sample interval set with the SAMPLE TIMING soft-key in the SAMPLE MENU.
Figure 11 Graph Type Display
76
Page 87
7.5.2 Select Display
The SELECT DISPLAY soft-key allows you to choose which display configu rationtouse.
7 Front Panel Operation
Display Menu
-
There are 16 display configurations each for both the data and graph type dis­play. The display configuration determines the type of data shown for the data display or the scale limits for the graph display. You select the display using the
LR buttons. After you press Enter, the display will change to the configura­tion you selected.
7.5.3 Edit Display - Data
Press the EDIT DISPLAY soft-key to edit a display configuration. For the data type display, you can program the five data fields to show any of a variety of
77
Page 88
1575/1590 Thermometer Readout
User’s Guide
parameters. Use the U and D buttons to move the highlight to any field. Press Enter to change the parameter for the field.
A new window will appear showing the available parameters. Move the cursor to the parameter you want. The window will scroll if you move past the top or bottom of the screen.
78
Page 89
7 Front Panel Operation
Display Menu
Press Enter to select the desired parameter for the specified field. For some pa rameters, such as history, you may have to specify additional information, like the location number, in another window that appears. The data display parame ter options are described below.
(1) HISTORY
This option will show past measurements. You can select which history mea
­surement to display. HISTORY(1) refers to the last measurement, HISTORY(2) refers to the measurement before that and so on. Up to 100 past measurements are available for viewing. As new measurements are made, past measurements will be shifted down in the history buffer and the history field will be updated accordingly.
(2) MEMORY
This option will show data that is stored in a specified memory location. For storing data in memory, refer back to Section 7.3.1. When you select this op
­tion, you will be prompted to enter a memory number. Up to 100 memory loca tions are available.
(3) T-MEMORY
This option displays the difference between the present measurement and the value stored in a specific memory location. This is useful for comparing mea­surements to a reference temperature. For storing data in memory, refer back to Section 7.3.1. When you select this option, you will be prompted to enter a memory number. There are 100 memory locations. If there is a conflict be-
tween the units of the T value and the MEMORY value, “..........” will be
displayed.
-
-
-
(4) T-HISTORY
This option displays the difference between the present measurement and a pre vious measurement in the history buffer. This can be used to calculate the rate at which the temperature is changing. When you select this option, you will be prompted to enter the history number. HISTORY(1) refers to the last measure ment, HISTORY(2) refers to the measurement before that and so on. Up to 100 past measurements are available for viewing. If there is a conflict between the
unitsof the T value and the HISTORY value, “..........” will be displayed.
(5) T(1)-T(2)
This option displays the difference between Input 1 and Input 2. This is useful when measuring the accuracy of one probe against another. It will use the most recent measurements made on the two channels regardless of how long ago they were made. To keep the measurements accurate, you should sample both channels periodically. To do this automatically, use the alternating channel mode as explained in Section 7.1.3.
(6) T(2)-T(1)
This option displays the difference between Input 2 and Input 1. This is just the reverse of the above option.
-
-
79
Page 90
1575/1590 Thermometer Readout
User’s Guide
(7) MAXIMUM
This displays the maximum of all measurements made on the channel since the statistics were reset (see Section 7.2.5).
(8) MINIMUM
This displays the minimum of all measurements made on the channel since the statistics were reset.
(9) SPREAD
This displays the difference between the maximum and minimum of all mea surements made on the channel since the statistics were reset.
(10) AV E R A G E
This displays the average of all measurements made on the channel since the statistics were reset.
(11) STD DEV
This displays the sample standard deviation of all measurements made on the channel since the statistics were reset. This function can be used to measure temperature stability or noise in the measurement system.
(12) T(C)
This displays the most recent temperature measurement in degrees Celsius.
-
80
(13) T(K)
This displays the most recent temperature measurement in Kelvin.
(14) T(F)
This displays the most recent temperature measurement in degrees Fahrenheit.
(15) T(
)
This displays the resistance of the most recent measurement in ohms.
(16) T(r)
This displays the most recent measurement as a resistance ratio.
(17) T(W)
This displays the most recent measurement as an ITS-90 or IPTS-68 W ratio.
(18) CURRENT
This option displays the actual current. The actual current may differ from the programmed value depending on the resolution and voltage limitation with the given input resistance.
(19) CONVERSION
This displays the type of temperature conversion algorithm such as ITS-90.
Page 91
(20) REFERENCE
This displays the resistance of the reference resistor used to make the measure ment. It also indicates whether the reference was one of the internal resistors or an external resistor connected to Input 2.
(21) PROBE SER#
This displays the serial number of the probe used to make the measurement.
7.5.4 Edit Display - Graph
For the graph type display, you can program the upper and lower limits of the graph with the EDIT DISPLAY. Selecting AUTO causes the graph to be scaled automatically. If you select SET and press Enter, a you will be prompted to en ter the limit.
The horizontal time scale of the graph cannot be set directly. It depends on the sample interval specified with the SAMPLE TIMING soft-key in the SAMPLE MENU. The scale of the horizontal axis is the length of time required to make 216 measurements.
This function also allows you to select the graph style between POINTS or LINES . POINTS causes temperature measurements to be represented as sepa­rate points on the display. LINES causes lines to be drawn that connect the measurement points.
7 Front Panel Operation
Display Menu
-
-
7.5.5 Display Resolution
The DISPLAY RESOL soft-key allows you to select the resolution with which measurements are displayed on the screen. There are settings for the different units. You can select resolutions of between 1 and 0.00000001. The specified resolution is used for displaying the present measurement as well as any values shown in the data fields. Data written to disk or transmitted via the communica
-
81
Page 92
1575/1590 Thermometer Readout
User’s Guide
tions ports will also use this resolution. Use the (( buttons and then press Enter to set the resolution.
82
7.5.6 Clear Graph
The CLEAR GRAPH soft-key is only available when the graph type display is selected. It clears all of the past data from the graph. If the limits are set to AUTO, the graph will automatically rescale to the new measurements.
Page 93
7.6 System Menu
7 Front Panel Operation
System Menu
The SYSTEM menu contains various functions that control settings such as the temperature units, time, reference resistor calibration values, and communica­tions parameters. The functions and sub-menus in the system menu are UNIT, PARAM MENU, DISK MENU, COMM MENU, and SYSTEM CAL menu.
7.6.1 Unit Menu
The UNIT soft-key allows you to select the units with which measurements are displayed, saved to disk, and transmitted over the communications channels. You can choose any the units shown in Table 10. Access to the UNIT function can be locked out as explained in Section 7.6.5.4.
Tabl e 1 0 Units of Measure
Units Description
C Temperature in degrees Celsius
F Temperature in degrees Fahrenheit
K Temperature in degrees Kelvin
Ω
r Ratio of resistance to reference resistance
W Ratio of resistance to characteristic resistance (ITS-90 and
Resistance in ohms
IPTS-68 only)
83
Page 94
1575/1590 Thermometer Readout
User’s Guide
W values are only meaningful with ITS-90 and IPTS-68 conversions. For ITS-90, W refers to the ratio of measured resistance to the resistance at the tri ple point of water. For IPTS-68, W refers to the ratio of measured resistance to the resistance at 0°C.
Measurement with a particular probe can be forced to be displayed in Ω,W,or ratio using CONVERSION in EDIT PROBE (see Section 7.4.3.2).
7.6.2 Parameter Menu
The PARAM MENU contains functions for saving or loading all of the operat ing parameters to or from the disk and for setting the date and time. The param eter sub-menu includes the functions TIME, SAVE PARAMS, LOAD PARAMS, RESET PARAMS and SCREEN PARAMS.
7.6.2.1 Time
The current time is always shown at the bottom of the data type screen. The time can also be recorded with measurements that are transmitted over the com munications channels or saved to disk. Access to the TIME function can be locked out as explained in Section 7.6.5.4.
The TIME soft-key lets you set the time and date. A window appears showing the hour, minute, month and so on. You can move to any item and change it. You can set the clock to one of two modes: 12-hour mode (12:00 a.m. to 11:59 p.m.) and 24-hour mode (0:00 to 23:59). If you will be transmitting time-stamped measurements to a computer it may be more convenient to use the 24-hour format. Press Enter after you make any changes to a parameter.
-
-
-
-
84
NOTE: You may need to set the date again after January 1, 2000.
7.6.2.2 Save Parameters
With the SAVE PARAMS soft-key you can save all of the operating parameters to disk. It is a good idea to keep a backup of the parameters in case they are lost or inadvertently changed. You can also keep different sets of parameters on disk for different applications. All of the instrument’s parameters are saved to one file on the disk. These include all sampling variables, probe data, display configurations, communications parameters, and reference resistor calibration values. The file is in a special binary format for efficient storage. As a result it cannot be easily edited on a computer.
When you use the SAVE PARAMS function, a window will appear requesting you to enter a file number between 0 and 9. Data will be saved to a file named “SETUPn.CFG” where n is the number you specify. Press Enter twice to save the parameters. If a file with this name already exists on the disk, you will be asked if you want to overwrite it.
Page 95
7.6.2.3 Load Parameters
7 Front Panel Operation
System Menu
With the LOAD PARAMS soft-key you can load all of the operating parame ters from disk. This is useful for restoring previous parameters if they have been lost or changed. To save parameters to disk, see Section 7.6.2.2 above. Access to the LOAD PARAMS function may be locked out if any parameters are protected as explained in Section 7.6.5.4.
After you select this function, a window appears asking you to select a file. Se lect a number and then press Enter twice. If you load a parameter file, all of the current operating parameters will be overwritten. You will be asked to confirm by pressing Enter.
CAUTION: Using the LOAD PARAMS or RESET PARAMS functions af
fects instrument settings which can affect accuracy.
7.6.2.4 Reset Parameters
The RESET PARAMS function causes all of the operating parameters to be re­set to the factory default values. Use this function only when necessary. Access to the RESET PARAMS function may be locked out if any parameters are pro­tected as explained in Section 7.6.5.4.
7.6.2.5 Screen Parameters
The SCREEN PARAMS function lets you set some of the display options which are COLORS (1590 only), BRIGHTNESS, and SCREEN SAVER.
COLORS lets you select one of various color schemes. BRIGHTNESS lets you adjust the brightness of the display backlight. SCREEN SAVER lets you en­able, disable, or set the DELAY TIME for the screen saver. The screen saver can help extend the lifetime of the display. The screen saver shuts off the dis play when the instrument is left unattended for a long period of time. After the display shuts off, pressing any button will reactivate it.
-
-
-
-
7.6.3 Disk Menu
The disk sub-menu contains functions for setting up disk data logging and for formatting new disks. Data logging allows you to record measurements in a file on a disk. Figure 20 shows an example of recorded data. This example is with PERIOD set to AUTO and TIME and DATE ON (see Section 7.6.3.1 below.)
85
Page 96
1575/1590 Thermometer Readout
User’s Guide
The channel mode is set to ALT N with N=1 (see Section 7.1.3.) Note that each measurement is preceded by the channel number.
1: 0.0103 C 13:06:47 3-22-95
2: 232.3503 C 13:06:49 3-22-95
1: 0.0104 C 13:06:51 3-22-95
2: 232.3502 C 13:06:54 3-22-95
1: 0.0104 C 13:06:56 3-22-95
2: 232.3503 C 13:06:58 3-22-95
Figure 12 Example of Data Recorded to Disk or Output Via the Digital Interface
7.6.3.1 Setup
With the SETUP soft-key you can program the disk sampling parameters. You can specify the file number, the sample period (in seconds), and whether to re­cord the time or date with the measurements. When you press the key, a win­dow appears allowing you to set these parameters.
86
PERIOD is the interval, in seconds, between recorded measurements. You may set it to a specific value or you can set it to AUTO. If you set the period to AUTO, each new measurement will be recorded as soon as it is available. The period will be the same as the sample interval set in SAMPLE TIMING (see Section 7.2.3). Using AUTO will insure that you record every measurement and will record samples with a more precise interval.
Page 97
The TIME and DATE parameters determine whether the current time and date are recorded along with measurements. It may be more convenient to use the 24-hour time format (Section 7.6.2.1).
The FILENAME is the name of the file to which the data will be recorded. You can select from “SAMPLE0.DAT” to “SAMPLE9.DAT” using the numeric keys 0 through 9. If the file does not yet exist on the disk, a new file will be created. If the file already exists, it will overwrite the data in the file. A mes sage will warn you if data will be overwritten.
7.6.3.2 Record/Stop
The RECORD/STOP soft-key is used to enable or disable disk sampling. The word RECORD or STOP is grayed out to indicate the current status. Before starting disk sampling, make sure the disk sampling parameters are set cor rectly with the SETUP soft-key.
7.6.3.3 Record N
The RECORD N soft-key allows you to record a specified number of measure­ments to disk. When this soft-key is pressed, a window will appear prompting you to enter a number. Set the number of measurements you want and then press Enter.
The 1575/1590 will begin recording measurements to disk and then stop after the specified number of measurements has been obtained. The RECORD N soft-key will be highlighted while recording and it will indicate the number of samples remaining. The RECORD/ STOP soft-key will indicate RECORD while the samples are being recorded and then STOP when finished. Before starting disk sampling, make sure the disk sampling parameters are set cor­rectly with the SETUP soft-key.
7 Front Panel Operation
System Menu
-
-
7.6.3.4 Format Disk
Before a floppy disk can be used to store information, it must be formatted. You can format the disk on a PC- compatible computer or you can use the 1575/1590 FORMAT DISK function. The 1575/1590 can use 3.5" disks of ei ther 720 kb (double density) or 1.44 Mb (high density) format. Generally, disks of one type cannot be formatted to the other. When you begin the format opera tion, you must specify the density by using the (( buttons and pressing Enter.
The 1575/1590 will try to format the disk at the density you select. If the disk cannot be formatted at this density, a message will appear to tell you so. If this happens, select the other density and try again. Be careful with the format com mand; any data previously stored on the disk will be erased.
7.6.3.5 Run Application
The RUN APP soft-key is for running software on disk such as specialized ap plications or upgrade installation software.
-
-
-
-
87
Page 98
1575/1590 Thermometer Readout
User’s Guide
7.6.4 Communications Menu
The communications sub-menu, accessed with the COMM MENU soft-key, contains functions for setting up the serial, IEEE-488, and printer interfaces and the analog output. The next sections describe the soft-key functions for set ting the communications parameters. (For specifics on operating the 1575/1590 through the digital communications channels, see Section 8).
7.6.4.1 Serial
The serial RS-232 interface is useful for communications between the 1575/1590 and a remote device such as a computer. It allows the remote device to receive measurement data from the thermometer or send commands to con trol the operation of the thermometer. (For a list of commands and other details related to the serial communications, see Section 8). Parameters that control the action of the serial interface are set with the SERIAL soft-key. These include the baud rate, sample period, duplex mode, linefeed, and time and date stamp. Pressing this soft-key brings up a window as shown below.
-
-
88
BAUDRATE determines the rate at which data is transferred. You must set the baud rate of the thermometer to be the same as that of the equipment it will be communicating with. The options are: 110, 150, 300, 600, 1200, 2400, 4800, 9600, and 19200. Use the arrow buttons to change the setting and then press Enter to save the setting.
By setting PERIOD, you can have the thermometer automatically transmit mea surements over the serial channel. The period is the interval, in seconds, be
­tween transmitted measurements. If you set the period to AUTO, each new measurement will be transmitted as soon as it is available. The period will be
-
Page 99
7 Front Panel Operation
System Menu
the same as the sample interval set in SAMPLE TIMING (see Section 7.2.3. Using AUTO will insure that you read every measurement and will read sam
­ples with a more precise interval. If the period is set to SET, the user must enter a value, in seconds, that will be used to transmit measurements. If the period is set to OFF, measurements will not be transmitted (unless requested using the “tem” command, see Section 8.3). Data sampled to the serial port is formatted the same as data recorded to disk. Figure 12 shows an example of how the data might look. The example is with the PERIOD set to AUTO and TIME and DATE ON (see below). The channel mode is set to ALT N with N=1 (see Sec
-
tion 7.1.3). Each measurement value is preceded by the channel number.
DUPLEX determines whether commands received by the 1575/1590 will be echoed back. If duplex is FULL, commands will be echoed back. If it is HALF they will not.
LINEFEED determines whether a linefeed character will be transmitted at the end of a line of data. Data transmitted from the serial port is always followed by a carriage return. If you set the linefeed ON, a linefeed will also be transmit ted immediately after the carriage return.
The TIME and DATE parameters determine whether the current time and date are transmitted along with measurements. The time can be set to 12-hour or 24-hour time format (Section 7.6.2.1).
The serial interface is configured for eight data bits, one stop bit, and no parity. The PC or instrument connected to the serial port must have these same settings.
-
7.6.4.2 IEEE-488
The IEEE-488 (GPIB) interface works very much like the serial communica­tions. One advantage of the IEEE-488 bus over serial is that many devices can be connected to a single cable. It is also faster than serial communications. With IEEE-488, as with serial, the remote device can request measurement data from the thermometer and can send commands to control its operation. (For a list of commands and other details related to IEEE-488 communications, see Section 8.) Parameters that control the action of the interface are set with the
89
Page 100
1575/1590 Thermometer Readout
User’s Guide
IEEE488 soft-key. These include the address, termination character, and time and date stamp. Pressing this soft-key brings up a window as shown below.
DEVICE ADDRESS sets the IEEE-488 address of the 1575/1590 Super-Ther­mometer. To be used as a device on an IEEE-488 bus the thermometer must be given a unique address. It can be any number from 0 to 30. It must be different from any address used by any other devices also connected to the bus.
TERM determines the character or characters that are sent after each line of data. You can select linefeed, carriage return, or both as the termination.
The TIME and DATE parameters determine whether the current time and date are transmitted along with measurements (in response to the “tem” command, see Section 8.3). The time can be set in 12-hour or 24-hour time format (Sec tion 7.6.2.1).
7.6.4.3 Printer
With the printer (parallel) interface, you can connect a printer to the thermome ter and have it record measurements on paper. Data sampled to the printer port is formatted the same as data recorded to disk. Figure 20 shows an example of how the data might look. The example is with the PERIOD set to AUTO and TIME and DATE ON (see below). The channel mode is set to ALT N with N=1 (see Section 7.2.3). Each measurement value is preceded by the channel number.
To transmit data to a printer, you must set the period. You can also set the line
-
-
-
90
Loading...