Table 1 lists the International Electrical Symbols. Some or all of these
symbols may be used on the instrument or in this manual.
Table1 International Electrical Symbols
SymbolDescription
AC
AC-DC
Battery
Complies with European Union directives
1 Before You Start
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface
Read the User’s Manual
Off
On
Hart ScientificManual Rev. 1806011
1 Before You Start
SymbolDescription
Canadian Standards Association
CAT II
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree
2 per IEC1010-1 revers to the level of Impulse Withstand Volt
age protection provided. Equipment of OVERVOLTAGE CATE
GORY II is energy-consuming equipment to be supplied from
the fixed installation. Examples include household, office, and
laboratory appliances.
1.2Safety Information
Use this instrument only as specified in this manual. Otherwise, the pro
tection provided by the instrument may be impaired. Refer to the safety
information in Warnings and Cautions.
The following definitions apply to the terms “Warning” and “Caution”.
•
“Warning” identifies conditions and actions that may pose hazards to
the user.
•
“Caution” identifies conditions and actions that may damage the instrument being used.
Warnings
To avoid possible electric shock or personal injury, follow these guidelines.
BURN HAZARD – DO NOT touch the IR target surface of the unit.
The temperature of the IR target surface is the same as the actual
temperature shown on the display. If the unit is set at 150°C and the
display reads 150°C, the target surface is at 150°C.
Temperatures above 70°C (158°F) are considered hazardous. Use ex
treme care when working with these temperatures. Observe all warn
ings and cautions given in this manual.
The sheet metal of the instrument may exhibit extreme temperatures
for areas close to the IR target surface.
-
-
-
-
-
DO NOT turn off the unit at temperatures higher than 100°C. This
could create a hazardous situation. Select a set-point less than 100°C
and allow the unit to cool before turning it off.
DO NOT operate this unit without a properly grounded, properly polar
ized power cord.
DO NOT connect this unit to a non-grounded, non-polarized outlet.
2Manual Rev. 1806019133
-
1 Before You Start
HIGH VOLTAGE is used in the operation of this equipment. SEVERE
INJURY OR DEATH may result if personnel fail to observe safety pre
cautions. Before working inside the equipment, turn the power off and
disconnect the power cord.
Always replace the fuse with one of the same rating, voltage, and type.
Allow at least six inches of space between the instrument and nearby
objects. DO NOT place instrument under a flammable structure.
DO NOT use this unit for any application other than calibration work.
DO NOT use this unit in environments other than those listed in the
user’s guide.
DO NOT operate near flammable materials.
-
Use of this instrument at HIGH TEMPERATURES for extended peri
ods of time requires caution.
Completely unattended high temperature operation is not recom
mended for safety reasons.
Before initial use, after transport, and anytime the instrument has not
been energized for more than 10 days, the calibrator must be energized for a dry-out period of 1 to 2 hours before it can be assumed to
meet all of the safety requirements of the IEC1010-1.
This instrument is intended for indoor operation only.
Follow all safety guidelines listed in the user’s manual.
Calibration Equipment should only be used by Trained Personnel.
-
-
Cautions
To avoid possible damage to the instrument, follow these guidelines.
Use the target cover at temperatures below 5°C. If ice forms on the tar
get, the IR probe will not indicate the correct temperature.
DO NOT use fluids to clean out the target surface.
DO NOT change the values of the calibration constants from the fac
tory set values. The correct setting of these parameters is important to
the safety and proper operation of the calibrator.
DO use a ground fault interrupt device.
-
-
1.3Customer Service Information
Hart Scientific can be contacted by writing to:
Hart Scientific, Inc.
799 E. Utah Valley Drive
Hart ScientificManual Rev. 1806013
1 Before You Start
American Fork, UT 84003-9775
Or by calling or faxing:
Telephone: (801) 763-1600
Fax: (801) 763-1010
Our World Wide Web site is: http://www.hartscientific.com
E-mail: support@hartscientific.com
When calling Hart Scientific Customer Service, please have the following
information available:
•
Model Number
•
Serial Number
•
Voltage
4Manual Rev. 1806019133
2Introduction
The Hart Scientific Model 9133 Mid-Range Field IR Calibrator may be
used as a portable instrument or bench top temperature calibrator for cal
ibrating point IR thermometers. The Model 9133 is small enough to use
in the field, and accurate enough to use in the lab. Calibrations may be
done over a range of -30°C to150°C (-22°F to 302°F). Temperature dis
play and setability resolution of the 9133 is 0.1 degrees.
The instrument features:
•
Rapid heating and cooling
•
RS-232 interface capability
Built in programmable features include:
•
Temperature scan rate control
•
Eight set-point memory
•
Adjustable readout in °C or °F
The temperature is accurately controlled by Hart’s digital controller. The
controller uses a precision platinum RTD as a sensor and controls the
well temperature with a FET driven thermal electric device.
2 Introduction
-
-
For improved uncertainties, a 0.125" diameter external reference thermometer may be used in the calibration probe hole at the top of the instrument. When using an external reference thermometer, the instrument
accuracy, stability, and ambient temperature effects on the surface can
be minimized.
The LED front panel continuously shows the current temperature. The
temperature may be easily set with the control buttons to any desired
temperature within the specified range. The instrument’s multiple fault
protection devices insure user and instrument safety and protection.
The Model 9133 calibrator was designed for portability, low cost, and
ease of operation. Through proper use, the instrument will provide con
tinued accurate calibration of temperature sensors and devices. The
user should be familiar with the safety guidelines and operating proce
-
dures of the calibrator as described in this user guide.
Hart ScientificManual Rev. 1806015
-
3 Specifications and Environment Conditions
3Specifications and Environment
Conditions
3.1Specifications
Temperature
†
Range
Accuracy
Stability±0.1°C (±0.18°F)
Target Size2.25" (57 mm)
Target Emissivity 0.95
Resolution0.1
Heating Time15 minutes (25°C to 150°C)
Cooling Time15 minutes (25°C to -20°C)
Power115 VAC (±10%), 3 A, or 230 VAC (±10%), 1.6 A, switch-
Size6" H x 11.25" W x 10.5" H (152 x 286 x 267 mm)
Weight10 lb. (4.6 kg)
†
When using a purge, the high and low range are reduced. This amount de-
pends on the flow and temperature of the purge.
‡
Temperature difference between the IR target surface and the reference probe
are calculated as shown in the graph in Figure 1 on page 8. Note: As the tem
perature of the IR target surface increases, the temperature difference de
creases.
‡
–30°C to 150°C (-22°F to 302°F)
±0.4°C (±0.72°F) [using 1/8" PRT]
able, 50/60 Hz
-
-
Hart ScientificManual Rev. 1806017
3 Specifications and Environment Conditions
0
C
0.15
0.1
0.05
0
-0.05
delta T °
-0.1
-0.15
-0.2
-30-10103050709011013015
Figure 1 Temperature Gradient Between Probe and IR Target Surface at Ambient of 30°C
Reference Probe °C
3.2Environmental Conditions
Although the instrument has been designed for optimum durability and
trouble-free operation, it must be handled with care. The instrument
should not be operated in an excessively dusty or dirty environment.
Maintenance and cleaning recommendations can be found in the Maintenance section of this manual.
The instrument operates safely under the following conditions:
•
temperature range: 5–50°C (41–122°F)
•
ambient relative humidity: 15–60%
•
pressure: 75kPa–106kPa
•
mains voltage within ±10% of nominal
•
vibrations in the calibration environment should be minimized
•
altitudes less than 2,000 meters
3.3Warranty
Hart Scientific, Inc. (Hart) warrants this product to be free from defects in
material and workmanship under normal use and service for a period as
stated in our current product catalog from the date of shipment. This war
ranty extends only to the original purchaser and shall not apply to any
-
8Manual Rev. 1806019133
3 Specifications and Environment Conditions
product which, in Hart’s sole opinion, has been subject to misuse, alter
-
ation, abuse or abnormal conditions of operation or handling.
Software is warranted to operate in accordance with its programmed in
structions on appropriate Hart products. It is not warranted to be error
free.
Hart’s obligation under this warranty is limited to repair or replacement of
a product which is returned to Hart within the warranty period and is de
termined, upon examination by Hart, to be defective. If Hart determines
that the defect or malfunction has been caused by misuse, alteration,
abuse or abnormal conditions or operation or handling, Hart will repair
the product and bill the purchaser for the reasonable cost of repair.
To exercise this warranty, the purchaser must forward the product after
calling or writing Hart for authorization. Hart assumes NO risk for in-tran
sit damage.
For service or assistance, please contact the manufacturer.
Hart Scientific, Inc.
799 East Utah Valley Drive
American Fork, UT 84003-9775
Phone: (801) 763-1600 – Fax: (801) 763-1010
E-mail: support@hartscientific.com
THE FOREGOING WARRANTY IS PURCHASER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY
IMPLIED WARRANTY OR MERCHANTABILITY, OR FITNESS FOR
ANY PARTICULAR PURPOSE OR USE. HART SHALL NOT BE LIABLE
FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES OR LOSS WHETHER IN CONTRACT, TORT, OR OTHER
-
WISE.
-
Hart ScientificManual Rev. 1806019
Loading...
+ 32 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.