Fluke 9133 Operating Manual

9133
Infrared Thermometer Calibrator
User’s Guide
Rev. 180601
Copyright © 2002 All Rights Reserved
Hart Scientific, Inc. 799 E. Utah Valley Drive American Fork, Utah 84003-9775 Telephone: (801) 763-1600 • Fax: (801) 763-1010 Internet: http://www.hartscientific.com E-mail: support@hartscientific.com
Rev. 180601

Table of Contents

1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Symbols Used. . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Safety Information. . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Customer Service Information . . . . . . . . . . . . . . . . . 3
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Specifications and Environment Conditions . . . . . . . . 7
3.1 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 Environmental Conditions . . . . . . . . . . . . . . . . . . . . 8
3.3 Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4 Safety Guidelines . . . . . . . . . . . . . . . . . . . . . . 11
5 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.2 Set Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.3 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.4 Setting the Temperature . . . . . . . . . . . . . . . . . . . . 14
6 Parts and Controls . . . . . . . . . . . . . . . . . . . . . 15
6.1 Back Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.2 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
7 General Operation . . . . . . . . . . . . . . . . . . . . . 19
7.1 Changing Display Units . . . . . . . . . . . . . . . . . . . . 19
7.2 Switching to 230 V Operation . . . . . . . . . . . . . . . . . 19
7.3 Ice Buildup/Purge . . . . . . . . . . . . . . . . . . . . . . . 19
8 Controller Operation . . . . . . . . . . . . . . . . . . . . 21
8.1 Target Temperature . . . . . . . . . . . . . . . . . . . . . . 21
8.2 Temperature Set-point . . . . . . . . . . . . . . . . . . . . . 21
8.2.1 Programmable Set-points . . . . . . . . . . . . . . . . . . . . . . . . 21
8.2.2 Set-point Value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
8.3 Temperature Scale Units. . . . . . . . . . . . . . . . . . . . 24
8.4 Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
i
8.4.1 Scan Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.4.2 Scan Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
8.5 Set-point Resistance. . . . . . . . . . . . . . . . . . . . . . 25
8.6 Temperature Scale Units. . . . . . . . . . . . . . . . . . . . 25
8.7 Secondary Menu. . . . . . . . . . . . . . . . . . . . . . . . 25
8.8 Heater Power . . . . . . . . . . . . . . . . . . . . . . . . . 25
8.9 Proportional Band . . . . . . . . . . . . . . . . . . . . . . . 26
8.10 Controller Configuration . . . . . . . . . . . . . . . . . . . . 27
8.11 Operating Parameters . . . . . . . . . . . . . . . . . . . . . 28
8.12 Serial Interface Parameters . . . . . . . . . . . . . . . . . . 28
8.12.1 BAUD Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
8.12.2 Sample Period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
8.12.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
8.12.4 Linefeed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
8.13 Calibration Parameters. . . . . . . . . . . . . . . . . . . . . 30
8.13.1 R0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.13.2 ALPHA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.13.3 DELTA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.13.4 BETA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
9 Digital Communication Interface . . . . . . . . . . . . . 33
9.1 Serial Communications . . . . . . . . . . . . . . . . . . . . 33
9.1.1 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
9.1.2 Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
9.1.2.1 BAUD Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
9.1.2.2 Sample Period . . . . . . . . . . . . . . . . . . . . . . . . . 34
9.1.2.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 34
9.1.2.4 Linefeed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9.1.3 Serial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9.2 Interface Commands. . . . . . . . . . . . . . . . . . . . . . 35
10 Calibration Procedure . . . . . . . . . . . . . . . . . . . 39
10.1 Calibration Points . . . . . . . . . . . . . . . . . . . . . . . 39
10.2 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . 39
10.2.1 Compute DELTA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
10.2.2 Compute R0 and ALPHA . . . . . . . . . . . . . . . . . . . . . . . . 40
10.2.3 Compute BETA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
10.2.4 Accuracy and Repeatability . . . . . . . . . . . . . . . . . . . . . . . 41
11 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 43
12 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . 45
12.1 Troubleshooting Problems, Possible Causes, and Solutions . 45
12.2 Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
ii
12.2.1 EMC Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
12.2.2 Low Voltage Directive (Safety) . . . . . . . . . . . . . . . . . . . . . . 47
12.3 Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . . . 48
iii
Figures and Tables
Table1 International Electrical Symbols . . . . . . . . . . . . . . . . . . . 1
Figure 1 Temperature Gradient Between Probe and IR Target
Surface at Ambient of 30°C . . . . . . . . . . . . . . . . . . . . . 8
Figure 2 Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Figure 3 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Figure 4 Controller Operation Flow Chart . . . . . . . . . . . . . . . . . . 22
Figure 5 Serial Cable Wiring . . . . . . . . . . . . . . . . . . . . . . . . . 33
Table 2 Controller Communications Commands . . . . . . . . . . . . . . 36
Table 2 Controller Communications Commands continued . . . . . . . . . 37
Figure 6 Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
iv
1 Before You Start
1.1 Symbols Used
Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual.
Table1 International Electrical Symbols
Symbol Description
AC
AC-DC
Battery
Complies with European Union directives
1 Before You Start
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface
Read the User’s Manual
Off
On
Hart Scientific Manual Rev. 180601 1
1 Before You Start
Symbol Description
Canadian Standards Association
CAT II
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 revers to the level of Impulse Withstand Volt age protection provided. Equipment of OVERVOLTAGE CATE GORY II is energy-consuming equipment to be supplied from the fixed installation. Examples include household, office, and laboratory appliances.
1.2 Safety Information
Use this instrument only as specified in this manual. Otherwise, the pro tection provided by the instrument may be impaired. Refer to the safety information in Warnings and Cautions.
The following definitions apply to the terms “Warning” and “Caution”.
“Warning” identifies conditions and actions that may pose hazards to the user.
“Caution” identifies conditions and actions that may damage the in­strument being used.
Warnings
To avoid possible electric shock or personal injury, follow these guide­lines.
BURN HAZARD DO NOT touch the IR target surface of the unit.
The temperature of the IR target surface is the same as the actual temperature shown on the display. If the unit is set at 150°C and the display reads 150°C, the target surface is at 150°C.
Temperatures above 70°C (158°F) are considered hazardous. Use ex treme care when working with these temperatures. Observe all warn ings and cautions given in this manual.
The sheet metal of the instrument may exhibit extreme temperatures for areas close to the IR target surface.
-
-
-
-
-
DO NOT turn off the unit at temperatures higher than 100°C. This could create a hazardous situation. Select a set-point less than 100°C and allow the unit to cool before turning it off.
DO NOT operate this unit without a properly grounded, properly polar ized power cord.
DO NOT connect this unit to a non-grounded, non-polarized outlet.
2 Manual Rev. 180601 9133
-
1 Before You Start
HIGH VOLTAGE is used in the operation of this equipment. SEVERE INJURY OR DEATH may result if personnel fail to observe safety pre
cautions. Before working inside the equipment, turn the power off and disconnect the power cord.
Always replace the fuse with one of the same rating, voltage, and type.
Allow at least six inches of space between the instrument and nearby objects. DO NOT place instrument under a flammable structure.
DO NOT use this unit for any application other than calibration work.
DO NOT use this unit in environments other than those listed in the
user’s guide.
DO NOT operate near flammable materials.
-
Use of this instrument at HIGH TEMPERATURES for extended peri ods of time requires caution.
Completely unattended high temperature operation is not recom mended for safety reasons.
Before initial use, after transport, and anytime the instrument has not been energized for more than 10 days, the calibrator must be ener­gized for a dry-out period of 1 to 2 hours before it can be assumed to meet all of the safety requirements of the IEC1010-1.
This instrument is intended for indoor operation only.
Follow all safety guidelines listed in the user’s manual.
Calibration Equipment should only be used by Trained Personnel.
-
-
Cautions
To avoid possible damage to the instrument, follow these guidelines.
Use the target cover at temperatures below 5°C. If ice forms on the tar get, the IR probe will not indicate the correct temperature.
DO NOT use fluids to clean out the target surface.
DO NOT change the values of the calibration constants from the fac
tory set values. The correct setting of these parameters is important to the safety and proper operation of the calibrator.
DO use a ground fault interrupt device.
-
-
1.3 Customer Service Information
Hart Scientific can be contacted by writing to:
Hart Scientific, Inc.
799 E. Utah Valley Drive
Hart Scientific Manual Rev. 180601 3
1 Before You Start
American Fork, UT 84003-9775
Or by calling or faxing:
Telephone: (801) 763-1600
Fax: (801) 763-1010
Our World Wide Web site is: http://www.hartscientific.com
E-mail: support@hartscientific.com
When calling Hart Scientific Customer Service, please have the following information available:
Model Number
Serial Number
Voltage
4 Manual Rev. 180601 9133
2 Introduction
The Hart Scientific Model 9133 Mid-Range Field IR Calibrator may be used as a portable instrument or bench top temperature calibrator for cal ibrating point IR thermometers. The Model 9133 is small enough to use in the field, and accurate enough to use in the lab. Calibrations may be done over a range of -30°C to150°C (-22°F to 302°F). Temperature dis play and setability resolution of the 9133 is 0.1 degrees.
The instrument features:
Rapid heating and cooling
RS-232 interface capability
Built in programmable features include:
Temperature scan rate control
Eight set-point memory
Adjustable readout in °C or °F
The temperature is accurately controlled by Hart’s digital controller. The controller uses a precision platinum RTD as a sensor and controls the well temperature with a FET driven thermal electric device.
2 Introduction
-
-
For improved uncertainties, a 0.125" diameter external reference ther­mometer may be used in the calibration probe hole at the top of the in­strument. When using an external reference thermometer, the instrument accuracy, stability, and ambient temperature effects on the surface can be minimized.
The LED front panel continuously shows the current temperature. The temperature may be easily set with the control buttons to any desired temperature within the specified range. The instrument’s multiple fault protection devices insure user and instrument safety and protection.
The Model 9133 calibrator was designed for portability, low cost, and ease of operation. Through proper use, the instrument will provide con tinued accurate calibration of temperature sensors and devices. The user should be familiar with the safety guidelines and operating proce
-
dures of the calibrator as described in this user guide.
Hart Scientific Manual Rev. 180601 5
-
3 Specifications and Environment Conditions
3 Specifications and Environment
Conditions
3.1 Specifications
Temperature
Range
Accuracy
Stability ±0.1°C (±0.18°F)
Target Size 2.25" (57 mm)
Target Emissivity 0.95
Resolution 0.1
Heating Time 15 minutes (25°C to 150°C)
Cooling Time 15 minutes (25°C to -20°C)
Power 115 VAC (±10%), 3 A, or 230 VAC (±10%), 1.6 A, switch-
Size 6" H x 11.25" W x 10.5" H (152 x 286 x 267 mm)
Weight 10 lb. (4.6 kg)
When using a purge, the high and low range are reduced. This amount de-
pends on the flow and temperature of the purge.
Temperature difference between the IR target surface and the reference probe are calculated as shown in the graph in Figure 1 on page 8. Note: As the tem perature of the IR target surface increases, the temperature difference de creases.
–30°C to 150°C (-22°F to 302°F)
±0.4°C (±0.72°F) [using 1/8" PRT]
able, 50/60 Hz
-
-
Hart Scientific Manual Rev. 180601 7
3 Specifications and Environment Conditions
0
C
0.15
0.1
0.05
0
-0.05
delta T °
-0.1
-0.15
-0.2
-30 -10 10 30 50 70 90 110 130 15
Figure 1 Temperature Gradient Between Probe and IR Target Surface at Ambient of 30°C
Reference Probe °C
3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trouble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and cleaning recommendations can be found in the Mainte­nance section of this manual.
The instrument operates safely under the following conditions:
temperature range: 5–50°C (41–122°F)
ambient relative humidity: 15–60%
pressure: 75kPa–106kPa
mains voltage within ±10% of nominal
vibrations in the calibration environment should be minimized
altitudes less than 2,000 meters
3.3 Warranty
Hart Scientific, Inc. (Hart) warrants this product to be free from defects in material and workmanship under normal use and service for a period as stated in our current product catalog from the date of shipment. This war ranty extends only to the original purchaser and shall not apply to any
-
8 Manual Rev. 180601 9133
3 Specifications and Environment Conditions
product which, in Hart’s sole opinion, has been subject to misuse, alter
-
ation, abuse or abnormal conditions of operation or handling.
Software is warranted to operate in accordance with its programmed in
­structions on appropriate Hart products. It is not warranted to be error free.
Hart’s obligation under this warranty is limited to repair or replacement of a product which is returned to Hart within the warranty period and is de
­termined, upon examination by Hart, to be defective. If Hart determines that the defect or malfunction has been caused by misuse, alteration, abuse or abnormal conditions or operation or handling, Hart will repair the product and bill the purchaser for the reasonable cost of repair.
To exercise this warranty, the purchaser must forward the product after calling or writing Hart for authorization. Hart assumes NO risk for in-tran sit damage.
For service or assistance, please contact the manufacturer.
Hart Scientific, Inc.
799 East Utah Valley Drive
American Fork, UT 84003-9775
Phone: (801) 763-1600 – Fax: (801) 763-1010
E-mail: support@hartscientific.com
THE FOREGOING WARRANTY IS PURCHASER’S SOLE AND EXCLU­SIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EX­PRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OR MERCHANTABILITY, OR FITNESS FOR ANY PARTICULAR PURPOSE OR USE. HART SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OR LOSS WHETHER IN CONTRACT, TORT, OR OTHER
-
WISE.
-
Hart Scientific Manual Rev. 180601 9
Loading...
+ 32 hidden pages