Port able Infrared Calibrators
Features
• Certify IR pyrometers from –30 °C
to 500 °C (–22 °F to 932 °F)
• Large 57 mm (2.25 in) blackbody
target
• RTD reference well for high
precision
• Small, compact design
Whether you’re using in-line or handheld infrared pyrometers, you need good calibration
standards to verify their accuracy. Hart Scientific’s portable IR calibrators provide stable
blackbody targets for calibrating noncontact IR thermometers from –30°C to 500°C.
These portable and accurate units feature a large, temperature controlled blackbody
target with a diameter of 2.25" (57 mm), which offers a large field of view area for optical
variations in infrared thermometers. The emissivity of the isothermal target is set at 0.95
(±0.02%), and the target temperature can be controlled in set-point increments of 0.1°
from –30°C to 500°C.
For even higher precision, a well is located directly behind the blackbody surface for
contact calibration of the blackbody.
Hart Scientific’s portable IR calibrators are as easy to use as “point and shoot.” Simply
set the desired blackbody temperature from the convenient front panel control buttons,
wait a few minutes for equilibrium, and point the gun at the target. The radiated energy
from the blackbody is measured by your IR thermometer. Simply compare its reading to
the display on the blackbody and record the difference.
PO Box 1073, 8 Leroy Road, Williston, VT 05495 USA • fax 802-863-1193 • ph 802-863-0085 • toll free 800-235-8367
Model 9132
For IR calibrations above normal ambient, the Hart Scientific Model 9132 provides a
stable blackbody target up to 500°C (932°F). With accuracy to ±0.5°C and stability to
±0.1°C, this portable IR unit can certify most handheld pyrometers.
Short heating and cooling times mean you won’t have to wait long to get your work done.
From room temperature to 500°C the Hart Scientific Model 9132 will be stable within 30
minutes. You won’t find a more compact IR calibrator.
Model 9133
If you’re calibrating IR guns at cold temperatures,
you’ll love the Hart Scientific Model 9133. With solidstate cooling technology, this new IR calibrator
reaches –30°C (22°F) in normal ambient conditions.
With a conveniently located dry gas fitting on the front
bezel, ice build up on the target can be avoided. At the
upper end of its range, the Model 9133 provides stable
temperatures to 160°C (320°F).
With heating and cooling times of about 15 minutes
from ambient to either extreme, the Hart Scientific
Model 9133 gets you to temperature quickly and
performs when it gets there. Compare your IR devices to the temperature display which
is factory calibrated to be within ±0.4°C (±0.7°F).
No other IR calibrators give you this level of precision in such compact packages.
Whatever your temperature application, trust a Hart Scientific portable infrared calibrator
to solve it.
Ordering Information
PO Box 1073, 8 Leroy Road, Williston, VT 05495 USA • fax 802-863-1193 • ph 802-863-0085 • toll free 800-235-8367
9132
9308
9133
9302
Portable IR Calibrator, 500 °C
Rugged Carrying Case, 9132
Portable IR Calibrator, –30 °C
Rugged Carrying Case, 9133