This Fluke product will be free from defects in material and workmanship for one year
from the date of purchase. This warranty does not cover fuses, disposable batteries,
or damage from accident, neglect, misuse, alteration, contamination, or abnormal
conditions of operation or handling. Resellers are not authorized to extend any other
warranty on Fluke’s behalf. To obtain service during the warranty period, contact your
nearest Fluke authorized service center to obtain return authorization information,
then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH
AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED.
FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY.
Since some states or countries do not allow the exclusion or limitation of an implied
warranty or of incidental or consequential damages, this limitation of liability may not
apply to you.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
11/99
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
1. The 9040/9040UK Phase Rotation Indicator............................................... 6
ii
Page 5
9040/9040UK
Introduction
The Fluke 9040 Phase Rotation Indicator (hereafter referred to, “the 9040”) is a handheld
instrument designed to detect the rotary field of three-phase systems.
Contacting Fluke
To contact Fluke, call one of the following telephone numbers:
USA: 1-888-44-FLUKE (1-888-443-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
USA Service: 1-888-99-FLUKE (1-888-993-5853)
Or, visit Fluke's Web site at www.fluke.com
To register your product, visit register.fluke.com
1
.
.
Page 6
9040/9040UK
Users Manual
Unpacking the 9040
The 9040 ships with the following items:
• 3 pieces self-retaining test probes (black) / Fused Test Probe (9040UK only)
• Alligator clip
• Users Manual
If an item is damaged or missing, contact the place of purchase immediately.
2
Page 7
Phase Rotation IndicatorSafety Information
Safety Information
WCaution identifies conditions and actions that may damage the 9040.
XWWarning identifies conditions and actions that pose hazard(s) to the user.
XW Read First: Safety Information
To avoid possible electric shock or fire, do the following:
•Read the following safety information carefully before using or
servicing the instrument.
• Adhere to local and national safety codes.
• Individual protective equipment must be used to prevent shock and
injury.
•Use of instrument in a manner not specified by the manufacturer may
impair safety features/protection provided by the equipment.
• Avoid working alone.
• Inspect the test leads for damaged insulation or exposed metal. Check
test lead continuity. Damaged leads must be replaced. Do not use the
9040 if it looks damaged.
•Be careful when working above 30 V ac rms, 42 V ac peak and 60 V dc.
Such voltages pose a shock hazard.
3
Page 8
9040/9040UK
Users Manual
•When using the probes, keep fingers away from probe contacts. Keep
fingers behind the finger guards on the probes.
•Measurements can be adversely affected by impedances of additional
operating circuits connected in parallel or by transient currents.
•Verify operation prior to measuring hazardous voltages (voltages above
30 V ac rms, 42 V ac peak and 60 V dc).
• Do not use the 9040 with any of the parts removed.
• Do not use the 9040 around explosive gas, vapor, or dust.
• Do not use the 9040 in a wet environment.
4
Page 9
Phase Rotation IndicatorSymbols
Symbols
The following symbols appear on the 9040 or in this manual.
Table 1. Symbols
Risk of electric shock
Y
Risk of Danger. Important information.
W
See manual.
Hazardous voltage. PConforms to EU directives.
X
Equipment protected by double or
T
reinforced Insulation
CAT III
5
Earth
J
AC or DC
D
OVERVOLTAGE (Installation)
CATEGORY III, Pollution Degree
2 per IEC1010-1 refers to the level
of Impulse Withstand Voltage
protection provided. Equipment of
OVERVOLTAGE CATEGORY III
is equipment in fixed installations
(e.g., electricity meter and primary
over-current protection equipment.
Page 10
9040/9040UK
Users Manual
Elements of the 9040
Indicators, buttons, and jacks are shown in Figure 1.
Figure 1. The 9040/9040UK Phase Rotation Indicator
6
Test lead input jack
L1, L2, L3 Indicators
Clockwise Rotation LCD Indicator
Counter-Clockwise Rotation LCD Indicator
Brief Instructions on instrument rear
bbx02f.eps
Page 11
Phase Rotation IndicatorDetermine the Rotary Field Direction
Determine the Rotary Field Direction
To determine the rotary field direction:
1. Connect the test probes to the end of the test leads.
2. Connect the test probes to the three mains phases.
3. The green ON indicator shows that the instrument is ready for testing.
4. Either the clockwise or counter-clockwise rotary indicator illuminates showing the type
of rotary field direction present.
XW Warning
The rotary indicator lights even if the neutral conductor, N, is connected
instead of L1, L2, or L3. Refer to the back of the 9040 for more information.
Note
The 9040 is powered from the installation under test.
7
Page 12
9040/9040UK
Users Manual
Maintaining the 9040
W Caution
To avoid damaging the 9040:
• Do not attempt to repair or service the 9040 unless qualified to do so.
• Make sure that the relevant calibration, performance test, and service
information is being used.
The only maintenance the 9040 requires is cleaning. Periodically wipe the case with a
damp cloth and mild detergent. Clean only with soap and water and remove any residue
afterwards.
W Caution
To avoid damaging the 9040:
•Do not use abrasives or solvents. Abrasives or solvents will damage
the 9040 case.
•Prior to cleaning, remove test leads from the 9040.
8
Page 13
Phase Rotation IndicatorReplacing the Fuse (9040UK only)
Replacing the Fuse (9040UK only)
XW Warning
To avoid probe damage or serious bodily injury, always use a replacement
fuse with the correct voltage and current as shown in the Specifications
section. Before you replace the fuse, disconnect the accessory (cable or
probe) at both ends.
1. Check the fuse using a simple continuity test.
2. Hold the probe in front of the finger guard and unscrew the tip in a counter-
clockwise direction.
3. Remove the defective fuse from the fuse holder.
4. Insert a new fuse in the fuse holder and reassemble the probe.
9
Page 14
9040/9040UK
Users Manual
Specifications
Environmental
Operating Temperature
0 °C to +40 °C
Pollution Degree
2
Type of Protection
IP 40
Mechanical Specifications
Size
124 x 61 x 27 mm (4.9 x 2.4 x 1.1 in)
Weight
200 g (0.44 lb)
Fuse
500 mA / 1000 V/ FF / 50 kA / 6.3 x 32
mm (0.25 x 1.26 in)
Safety Specifications
Electrical Safety
10
IEC 61010/EN 61010,
IEC 61557-7/EN 61557-7
Maximum Operating Voltage (Ume)
690 V
Protection Levels
CAT III, 600 V to ground
Electrical Specifications
Power Supply
From unit under test
Determine Rotary Field Direction
Nominal Voltage
40 to 690 VAC
Frequency Range (f
15 to 400 Hz
Current Pickup
1 mA
Nominal Test Current (in per phase)
1 mA
)
n