This Fluke product will be free from defects in material and
workmanship for three years from the date of purchase.
This warranty does not cover fuses, disposable batteries, or
damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling.
Resellers are not authorized to extend any other warranty
on Fluke’s behalf. To obtain service during the warranty
period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the
product to that Service Center with a description of the
problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS
NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES,
ARISING FROM ANY CAUSE OR THEORY. Since some
states or countries do not allow the exclusion or limitation of
an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
11/99
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
Anywhere in the world: +1-425-446-5500
Or visit Fluke’s Web site at: www.fluke.com
Register the Meter at: http://register.fluke.com
.
2
Page 9
HVAC Clamp Meter
Contacting Fluke
Safety Information
A “XW Warning” statement defines hazardous conditions
and actions that could cause bodily harm or death.
A “W Caution” statement identifies conditions and actions
that could damage the Meter or the equipment under test.
XWRead First: Safety Information
To ensure safe operation and service of the
Meter, follow these instructions:
•Read the Users Manual before use and
follow all safety instructions.
•Use the Meter only as specified in the
Users Manual; otherwise, the Meter's
safety features may be impaired.
•Avoid working alone so assistance can be
rendered.
•Never use the Meter on a circuit with
voltages higher than 600 V or a frequency
higher than 400 Hz fundamental. The Meter
may be damaged.
•Never measure ac current while the test
leads are inserted into the input jacks.
•Do not use the Meter or test leads if they
look damaged.
•Use extreme caution when working around
bare conductors or bus bars. Contact with
the conductor could result in electric
shock.
3
Page 10
902
Users Manual
•Use caution when working with voltages
above 60 V dc or 30 V ac rms or 42 V ac
peak. Such voltages pose a shock hazard.
•Clean the case with a damp cloth and mild
detergent only. Do not use abrasives or
solvents.
•To avoid false readings that can lead to
electrical shock and injury, replace the
batteries as soon as the low battery
indicator (B) appears. As the Meter gets
to the point where the low batteries affect
the readings, the Meter locks and no
measurements can be made until the
batteries are changed.
•Do not hold the Meter anywhere beyond
the tactile barrier, see Figure 1.
•Adhere to local and national safety codes.
Individual protective equipment must be
used to prevent shock and arc blast injury
where hazardous live conductors are
exposed.
4
Page 11
HVAC Clamp Meter
Symbols
Symbols
The following symbols are found on the Meter or in this
manual.
,
W
X Hazardous voltage. Risk of electric shock.
T Double insulation
B
)
P
J Earth ground
F DC (Direct Current)
B AC (Alternating Current)
~
;
N10140
s
May be used on hazardous live conductors
Risk of danger. Important information. See Users
Manual.
Battery
Complies with Canadian and US Standards
Conforms to relevant European Union directives
Do not dispose of this product as unsorted municipal
waste. Contact Fluke or a qualified recycler for
disposal.
Conforms to relevant Australian standards
Inspected and licensed by TÜV Product Services
5
Page 12
902
Users Manual
Getting Acquainted with the Meter
Refer to Figures 1 and 2 and Tables 1 and 2 to become
more acquainted with the Meter’s features.
CAT
COM
V
µ
A
600V
efu0001.eps
Figure 1. 902 HVAC Clamp Meter Features
6
Page 13
HVAC Clamp Meter
Getting Acquainted with the Meter
Table 1. 902 HVAC Clamp Meter Features
Number Description
A Backlight Button
B Jaw Release
C Tactile Barrier
D Jaws
E Hold Button
F Rotary Switch:
V DC and AC voltage
P Resistance and continuity
G DC microamps
F Degrees Fahrenheit / degrees Celsius
L Capacitance
K AC current
G LCD
H Min Max Button
I AC/DC, °F/°C Button
J Input Terminals
7
Page 14
902
Users Manual
B
C
D
A
Figure 2. Display Features
E
F
G
efu0006.eps
8
Page 15
HVAC Clamp Meter
Getting Acquainted with the Meter
Table 2. Display Features
Number Indication
Battery indicator -The batteries are low and need to
XWWarning: To avoid false
A
be changed.
readings, which could lead to possible
electric shock or personal injury, replace
the batteries as soon as the battery
indicator appears.
B Indicates the presence of high voltage
C
D Display Hold is active
E Volts
F Amps
G
Indicators for minimum and maximum recording
mode
°F - Degrees Fahrenheit
°C - Degrees Celsius
e - Ohms
M - Microamps
N - Microfarads
DC - Direct Current
AC - Alternating Current
9
Page 16
902
Users Manual
Using the Meter
AC and DC Voltage Measurement
To measure AC or DC voltage:
1. Insert the test leads into the Meter.
2. Turn the rotary switch to V.
3. Press A to choose AC or DC voltage. The display
reflects the chosen voltage mode.
4. Use the test leads to take the measurement. The Meter
reading appears on the display.
Note
When a measured voltage is above 30 V,
Zappears on the display. When the voltage drops
below 30 V, Zdisappears.
10
Page 17
HVAC Clamp Meter
Using the Meter
Resistance and Continuity
To measure resistance or continuity:
XW Warning
To avoid false readings that can lead to
electrical shock and injury, de-energize the
circuit before taking the measurement.
1. Insert the test leads into the Meter.
2. Turn the rotary switch to P.
3. Take the measurement. The resistance reading
appears on the display.
•If the resistance is shorted, the Meter beeps and
shows a reading < 30 Ω.
•If the resistance is open or exceeds the Meter’s
range, the display reads OL.
11
Page 18
902
Users Manual
Microamps µA Measurement
The µA dc (G) function on the Meter is primarily for HVAC
flame rod testing. To test a heating system flame rod (refer
to Figure 3):
1. Turn the heating unit off and locate the wire between
the gas-burner controller and the flame rod.
2. Break this connection.
3. Turn the rotary switch on the Meter to G.
4. Using alligator clips, connect test leads between the
flame sensor probe and control-module wire.
5. Turn heating unit on and check the reading on the
Meter.
Refer to the heating unit documentation for what the
6.
desired reading should be.
12
Page 19
HVAC Clamp Meter
Using the Meter
Burner
Figure 3. Testing a Flame Rod
efu0004.eps
Temperature
The Meter measures temperature in either Celsius (°C) or
Fahrenheit (°F).
To measure temperature (refer to Figure 4):
13
Page 20
902
Users Manual
1. Connect the 80BK Integrated DMM Temperature
Probe to the input jacks noting correct polarity of the
probe.
2. Turn the rotary switch to F.
3. Press A to select °C or °F. The display reflects the
chosen temperature mode.
4. Position the probe to take the measurement. The
reading appears on the display.
Note
To meet stated accuracy, the 80BK and Meter
must be at the same temperature.
XW Warning
To avoid possible electric shock DO NOT apply
the probe tip to any conductor that is greater
than 30 V ac, 42 V peak or 60 V dc to earth.
14
Page 21
HVAC Clamp Meter
Using the Meter
Figure 4. Temperature Measurement
efu0005.eps
15
Page 22
902
Users Manual
Capacitance
Turn off circuit power, then disconnect and discharge the
capacitor before measuring capacitance. Turn the Meter’s
rotary switch to capacitance (L).
If the capacitor requires more discharging, diSC is
displayed while the capacitor discharges. When measuring,
be sure to note the correct polarity of the capacitor.
AC Current Measurement
XW Warning
To avoid electrical shock and injury:
•Remove Test Leads before making current
measurements.
•Do not hold the Meter anywhere beyond
the tactile barrier, see Figure 1.
Turn the rotary switch to AC current (K). When measuring
AC current, it is necessary that the measured wire be
properly seated within the clamp jaws. The wire being
measured should be centered within the jaws, below the
horizontal line located on the clamp. Also note that currents
moving in different directions will cancel each other out, so
one wire must be measured at a time for a correct
measurement (see Figure 5).
16
Page 23
HVAC Clamp Meter
Using the Meter
Figure 5. Proper AC Current Measurement
efu0003.eps
17
Page 24
902
Users Manual
Backlight
Press C to toggle the backlight on and off. The backlight
automatically turns off after 2 minutes.
To disable the automatic 2-minute backlight timeout, hold
down
Cwhile turning the Meter on.
MIN MAX Recording Mode
The MIN MAX recording mode captures the minimum and
maximum input values. When a new high or low is
detected, the Meter beeps.
To use this feature:
1. Put the Meter into the desired measurement function
and range.
2. Press Jto enter MIN MAX Mode. MAX is displayed
and the highest reading detected since entering MIN
MAX is displayed.
3. Press Jto step through the minimum (MIN) and
present readings.
4. To pause MIN MAX recording without erasing stored
values, press I. K is displayed.
5. To resume MIN MAX recording, press I again.
6. To exit and erase stored readings, press Jfor at
least two seconds.
18
Page 25
HVAC Clamp Meter
Using the Meter
Display HOLD
XWWarning
To avoid possible electric shock or personal
injury, when Display HOLD is activated, be
aware that the display will not change when
you apply a different voltage.
In the Display HOLD mode, the Meter freezes the display.
The Meter also beeps every 4 seconds and Hflashes
to remind the user.
Press Ito activate Display HOLD; His displayed and
the reading is captured.
To exit and return to normal operation, press I.
Auto Off
The Meter automatically turns off after 20 minutes. The
rotary switch must be turned to “OFF” and then turned back
on for the Meter to restart. Auto Off is disabled during Min
Max mode. To disable Auto Off, hold J when turning
the Meter on.
19
Page 26
902
Users Manual
Maintenance
XWWarning
To avoid possible electric shock or personal
injury, repairs or servicing not covered in this
manual should be performed only by qualified
personnel.
Cleaning the Meter
XWWarning
To avoid electrical shock, remove any input
signals before cleaning.
W Caution
To avoid damaging the Meter, do not use
aromatic hydrocarbons or chlorinated solvents
for cleaning. These solutions will react with the
plastics used in the Meter.
Clean the instrument case with a damp cloth and mild
detergent.
20
Page 27
HVAC Clamp Meter
Maintenance
Battery Replacement
XWWarning
To avoid false readings that could lead to
possible electric shock or personal injury,
replace the batteries as soon as the low
battery indicator (B) appears.
Disconnect the test leads before replacing the
batteries.
To replace the batteries (refer to Figure 6):
1. Turn the rotary switch to “OFF” and remove the test
leads from the terminals.
2. Use a Phillips screwdriver to loosen the battery
compartment door screw, and remove the door from
the case bottom.
3. Remove the batteries.
4. Replace the batteries with two new AA batteries.
5. Reattach the battery compartment door to the case
bottom and tighten the screw.
21
Page 28
902
Users Manual
Figure 6. Battery Replacement
efu0002.eps
22
Page 29
HVAC Clamp Meter
Specifications
Specifications
Electrical Specifications
Function Range Resolution Accuracy
Voltage DC 0 – 600 V 0.1 V 1 % ± 5 counts
Voltage AC
(True Rms)
Current AC
(True Rms)
Current DC 0 - 200 µA 0.1 µA 1.0 % ± 5 counts
Resistance
Continuity < 30 Ω
Temperature -10 to 400 °C 0.1 °C 1 % ± 0.8 °C
Capacitance
0 – 600 V 0.1 V
0 – 600 A 0.1 A
0 – 999 Ω
0 – 9999 Ω
1-100 µF
100-1000 µF
0.1 Ω
1.0 Ω
0.1 µF
1 µF
1 % ± 5 counts
(50/60 Hz)
2.0 % ± 5 counts
(50/60 Hz)
1.5 % ± 5 counts
1.9 % ± 2 counts
23
Page 30
902
Users Manual
General Specifications
Operating Temperature -10 °C to +50 °C
Storage Temperature -40 °C to +60 °C
Operating Humidity Non condensing (< 10 °C)
90 % RH (10 °C to 30 °C)
75 % RH (30 °C to 40 °C)
45 % RH (40 °C to 50 °C)
(Without Condensation)
Operating Altitude 2500 meters above mean sea
level
Storage Altitude 12,000 meters above mean
sea level
IP Rating IP 30 per IEC 60529
Vibration Requirements MIL-PRF-28800F Class 2
random vibration
EMI, RFI, EMC EMI: instrument unspecified for
use in EMC field • 0.5 V /
Meter
EMC: Meets all applicable
requirements in EN61326-1
Temperature
Coefficients
0.1 x (specified accuracy)/ °C
(<18 °C or >28 °C)
24
Page 31
HVAC Clamp Meter
Specifications
Size (H X W X L) 9.1 x 3.8 x 1.7 inches
(240 x 80 x 40 mm)
Weight 1.1 lb
(310 g)
Design Standards and
Compliance
Agency Approvals
Over-voltage Category 600 V, CAT III per IEC 1010-1
Power Requirements Two AA Batteries, NEDA 15 A,
IEC 61010, IEC 61010-2032,CE
P )
N10140
CAT III equipment is designed
to protect against transients in
equipment in fixed-equipment
installations, such as
distribution panels, feeders
and short branch circuits, and
lighting systems in large
buildings.
IEC LR6
; s
25
Page 32
902
Users Manual
26
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.