Fluke 80PK-11 Operating Manual

80PK-11
Type-K Velcro Thermocouple
Instruction Sheet
Introduction
The 80PK-11 is a Type-K Velcro Thermocouple Temperature Probe designed for HVAC temperature measuring applications. The 20 inch PVC cable terminates with a Type-K thermocouple inside a 19.5 inch nylon Velcro cuff. The 80PK-11 can be used with any temperature measuring instrument that is designed to accept Type-K thermocouples and has a miniature connector input. The 80PK-11 is not suitable for liquid immersion but can work in humid to wet environments.
XWWarning
To avoid electrical shock, do not use this probe when voltages exceeding 24 V rms or 60 V dc are present. The probe tip is electrically connected to the output terminals.
®
Specifications
Type: K Special Grade (Chromel vs Alumel) Measurement Range: -30 ºC to 105 ºC (-22 ºF to 221 ºF) Display Accuracy: ± 0.1 ºC (0.1 ºF) Measurement Time (Time Constant): 2 seconds Maximum Voltage: 24 V ac rms or 60 V dc maximum for proper
safe operation.
Maximum Temperature of Bead: 150 ºC (302 ºF) Cable:
Total Length: 1 m (40 in), 20 in cable, 19.5 in Velcro cuff Insulation Material: Hytrel
Connector:
Type: Yellow mini-thermocouple with 0.500 in pin Material: Hytrel 4774 Maximum Temperature: 200 ºC (392 ºF)
PN 2724595 August 2006
© 2006 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies.
Measurement Considerations
Instrument Compatibility
The 80PK-11 is designed to be compatible with any temperature measuring instrument that accepts Type-K thermocouples, has a miniature thermocouple connector, and has cold reference junction compensation. Accuracy of the temperature measuring instrument must be considered along with the 80PK-11 accuracy specification in order to determine the overall accuracy of the combination.
Temperature Limitations
The 200 ºC (392 ºF) continuous temperature rating of the 80PK-11 is primarily determined by the Hytrel insulation. The bead alone may be momentarily subjected to higher temperatures without damage.
Media Limitations
Type-K Chromel-Alumel thermocouple wires are compatible with clean oxidizing atmospheres.
Operation
WCaution
Repeated sharp bending can break the 80PK-11 lead. To protect the lead, avoid sharp bends, especially near the connector.
Use the 80PK-11 as follows:
1. Connect the probe to a compatible Type-K measuring instrument using the miniature thermocouple connector (0.500 in pin spacing).
2. Turn on the measuring instrument and select the appropriate range and scale.
3. Check the readout on the measuring instrument. With no heat or cold source applied to the sensor, it should display the ambient (room) temperature.
4. Wrap the probe firmly around the object you wish to measure. The readout will give you the source temperature.