To avoid possible electrical shock or personal injury, follow these guidelines:
⇒ Use the Meter only as specified in this manual or the protection provided by the Meter might be impaired.
⇒ Do not use the Meter or test leads if they appear damaged, or if the Meter is not operating properly. If in doubt, have the
Meter serviced.
⇒ Always use the proper terminals, switch position, and range for measurements.
⇒ Verify the Meter's operation by measuring a known voltage.
⇒ Do not apply more than the rated voltage, as marked on the Meter, between the terminals or between any terminal and earth
ground.
⇒ Use caution with voltages above 30 V ac rms, 42 V ac peak, or 60 V dc. These voltages pose a shock hazard.
⇒ Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or
capacitance.
⇒ Do not use the Meter around explosive gas or vapor.
⇒ When using the test leads, keep your fingers behind the finger guards.
⇒ Remove test leads from the Meter before opening the Meter case or battery door.
Underwriters Laboratories, Inc.
Meter in accordance with IEC 61010-1. 54CJ
Conforms to relevant Australian standards
.com
1.888.610.7664 info@Fluke-Direct.com
Page 4
The Fluke Model 77 Series IV is a battery-powered, average
Fluke-Direct
responding-rms indicating multimeter (hereafter "the Meter"), with
a 6000-count, 3 3/4-digit display, and a bar graph.
This meter meets CAT III and CAT IV IEC 61010 standards. The
IEC 61010 safety standard defines four measurement categories
(CAT I to IV) based on the magnitude of danger from transient
impulses. CAT III meters are designed to protect against transients
in fixed-equipment installations at the distribution level; CAT IV
meters are designed to protect against transients from the primary
supply level (overhead or underground utility service).
The Meter measures or tests the following:
♦ AC / DC voltage & current ♦ Diodes
♦ Resistance ♦ Continuity
♦ Voltage frequency ♦ Capacitance
Model 77 Series IV
Digital Multimeter
Warning and Caution Statements
A XWWarning identifies hazardous conditions and actions that
could cause bodily harm or death.
A Caution identifies conditions and actions that could damage the
Meter, the equipment under test, or cause permanent loss of data.
Unsafe Voltage
To alert you to the presence of a potentially hazardous voltage,
when the Meter detects a voltage ≥30 V or a voltage overload
Ysymbol is displayed.
(OL), the
Test Lead Alert
To remind you to check that the test leads are in the correct
terminals,
rotary switch to or from the mA or A position.
Attempting to make a measurement with a test lead in
an incorrect terminal might blow a fuse, damage the
Meter, and cause serious personal injury.
is momentarily displayed when you move the
LEAd
W Warning
1
.com
1.888.610.7664 info@Fluke-Direct.com
Page 5
Model 77 Series IV
Fluke-Direct
Users Manual
Battery Saver (Sleep Mode)
The Meter enters the "Sleep" mode and blanks out the display if
there is no function change or button press for 20 minutes. To
disable the Sleep mode, hold down the yellow button while turning
the Meter on. The Sleep mode is always disabled in the MIN MAX
AVG mode and the AutoHOLD mode.
Terminals
.
1
2
Item Description
Input terminal for AC and DC milliamp measurements to
1
400 mA.
Input terminal for AC and DC current measurements to
2
10 A.
Input terminal for voltage, continuity, resistance, diode
3
test, capacitance, and frequency measurements.
Common (return) terminal for all measurements.
4
V
FUSED
3
4
AIK01F.EPS
2
Rotary Switch Positions
Switch
Position
AC voltage from 0.001 to 1000 V.
K
Frequency from 2 Hz to 99.99 kHz.
Hz
DC voltage from 1 mV to 1000 V.
L
DC mV from 0.1 mV to 600 mV.
L
m
Ohms from 0.1 Ω to 50 MΩ.
e
Farads from 1 nF to 9999 μF.
E
Beeper turns on at <25 Ω and turns off at >250 Ω.
R
G
Diode test. Displays OL above 2.4 V.
B
AC mA from 0.01 mA to 400 mA.
F
mA
DC mA from 0.01 mA to 400 mA.
AC A from 0.001 A to 10 A.
F
BA
DC A from 0.001 A to 10 A
>10.00 display flashes.
>20 A, OL is displayed.
Measurement Function
.com
1.888.610.7664 info@Fluke-Direct.com
Page 6
Fluke-Direct
Display
Display
.
5
4
3
2
1
13
12
No. Symbol Meaning
1
s
2
O
3
K
4
Y
h
5
6
m
MAX , MIN,
AVG
7
nμ F, mVA,
Mke, kHz
6
11
14
Continuity test.
Diode test.
Negative readings.
Unsafe voltage. Voltage ≥30 V, or
voltage overload (OL)
AutoHOLD is enabled. Display holds
present reading until it detects new
stable input. Then the Meter beeps
and displays new reading.
MIN MAX AVG enabled.
Maximum, minimum, average, or
present reading.
Measurement units.
10
7
8
7
9
15
AIK02F.EPS
No. Symbol Meaning
bAtt
diSC
EEPr
Err
CAL
Err
8
9
10
11
12
13
14
15
DC, AC
b
610000mV
Bar graph
Auto Range
Manual Range
±
0L
LEAd
Replace the battery immediately.
In the capacitance function, too much electrical
charge is present on the capacitor being tested.
Invalid EEPROM data. Have Meter serviced.
Invalid calibration data. Calibrate Meter.
Direct current, alternating current.
Low battery. Replace battery.
All possible ranges.
Analog display.
The Meter selects the range with
the best resolution.
The user selects the range.
Bar graph polarity.
The input out of range.
WTest lead alert. Displayed when
the rotary switch is moved to or from
the mA or A position.
Error Messages
3
.com
1.888.610.7664 info@Fluke-Direct.com
Page 7
Model 77 Series IV
Fluke-Direct
Users Manual
MIN MAX AVG Recording Mode
The MIN MAX AVG recording mode captures the minimum and
maximum input values, and calculates a running average of all
readings. When a new high or low is detected, the Meter beeps.
For dc functions, accuracy is the specified accuracy of the
measurement function
275 ms in duration.
For ac functions, accuracy is the specified accuracy of the
measurement function
1.2 s in duration.
To use MIN MAX AVG recording:
⇒ Make sure that the Meter is in the desired measurement
function and range. (Autoranging is disabled in the MIN MAX
AVG mode.)
⇒ Press MIN MAX to activate MIN MAX AVG mode.
m and MAX light, and the highest reading detected
since entering MIN MAX AVG is displayed.
⇒ Press MIN MAX to step through the low (MIN), average
(AVG), and present readings.
⇒ To pause MIN MAX AVG recording without erasing stored
values, press HOLD. h is displayed.
To resume MIN MAX AVG recording, press HOLD again.
h turns off.
⇒ To exit and erase stored readings, press MIN MAX for 1
second or turn the rotary switch.
Note
±
12 counts for changes longer than
±
40 counts for changes longer than
4
AutoHOLD Modes
To avoid electric shock, do not use the AutoHOLD mode
to determine if a circuit is live. Unstable or noisy
readings will not be captured.
In the AutoHOLD mode, the Meter holds the reading on the display
until it detects a new stable reading. Then the Meter beeps, and
displays the new reading.
⇒ Press HOLD to activate AutoHOLD. h lights.
⇒ Press HOLD again or turn the rotary switch to resume normal
operation.
XWWarning
YELLOW Button
Press the yellow button to select alternate measurement functions
on a rotary switch setting, e.g., to select DC mA, DC A, Hz,
capacitance, or diode test.
Display Backlight
Press S to toggle the backlight on and off. The backlight
automatically turns off after 2 minutes.
.com
1.888.610.7664 info@Fluke-Direct.com
Page 8
Fluke-Direct
Manual Ranging and Autoranging
Manual Ranging and Autoranging
The Meter has both Manual range and Autorange modes.
⇒ In the Autorange mode, the Meter selects the range with the
best resolution.
⇒ In the Manual Range mode, you override Autorange and select
the range yourself.
When you turn the Meter on, it defaults to Autorange and Auto Range is displayed.
1. To enter the Manual Range mode, press RANGE.
Manual Range is displayed.
2. In the Manual Range mode, press RANGE to increment the
range. After the highest range, the Meter wraps to the lowest
range.
You cannot manually change the range in the MIN MAX
AVG mode.
If you press RANGE while in MIN MAX
beeps, indicating an invalid operation, and the range does
not change.
3. To exit Manual Range, press RANGE for 1 second or turn the
rotary switch.
The Meter returns to Autorange and Auto Range is displayed.
5
Note
AVG, the Meter
Power-Up Options
To select a Power-Up Option, hold down the button indicated while
turning the Meter on.
Power-Up Options are cancelled when the Meter is turned OFF.
Button Power-Up Options
(YELLOW)
Turns on all display segments when in VAC switch
H
position.
Disables beeper. bEEP is diplayed when enabled.
M
Enables "Smoothing" mode. '___ is displayed when
enabled.
R
Dampens display fluctuations of rapidly changing
inputs by digital filtering.
Disables automatic power-down ("Sleep mode").
PoFF is displayed when enabled
B
Sleep mode is also disabled while the Meter is in a
MIN MAX AVG Recording mode, or the AutoHOLD
mode.
Disables automatic 2-minute backlight timeout.
S
LoFF is displayed when enabled.
.com
1.888.610.7664 info@Fluke-Direct.com
Page 9
Model 77 Series IV
Fluke-Direct
Users Manual
Making Basic Measurements
The figures on the following pages show how to make basic
measurements.
When connecting the test leads to the circuit or device, connect
the common (COM) test lead before connecting the live lead;
when removing the test leads, remove the live lead before
removing the common test lead.
XWWarning
To avoid electric shock or injury, or damage to the
Meter, disconnect circuit power and discharge all highvoltage capacitors before testing resistance, continuity,
diodes, or capacitance.
Measuring AC and DC Voltage
Volts AC
MIN MAX
RANGEHOLD
V
V
Volts DC
+
Millivolts DC
MIN MAX
RANGEHOLD
MIN MAX
RANGEHOLD
mV
_
+
AIK03F.EPS
6
Measuring Resistance
Measuring Capacitance
_
MIN MAX
RANGEHOLD
MIN MAX
RANGEHOLD
+
_
+
AIK04F.EPS
AIK05F.EPS
.com
1.888.610.7664 info@Fluke-Direct.com
Page 10
Fluke-Direct
Making Basic Measurements
Testing for Continuity
Testing Diodes
Good DiodeGood Diode
MIN MAX
MIN MAX
RANGEHOLD
RANGEHOLD
AIK06F.EPS
MIN MAX
RANGEHOLD
__
+
Single Beep
Forward Bias
Bad Diode
MIN MAX
RANGEHOLD
CAT
_
+
Open
MIN MAX
RANGEHOLD
+
Reverse Bias
Bad Diode
MIN MAX
RANGEHOLD
+
and
Shorted
_
AIK07F.EPS
7
.com
1.888.610.7664 info@Fluke-Direct.com
Page 11
Model 77 Series IV
Fluke-Direct
Users Manual
Measuring AC or DC Current
XWWarning
To avoid personal injury or damage to the Meter:
• Never attempt to make an in-circuit current measure-
ment when the open-circuit potential to earth ground
is >1000 V.
• Check the Meter's fuses before testing. (See “Testing
the Fuses”.)
• Use the proper terminals, switch position, and range
for your measurement.
• Never place the probes in parallel with a circuit or
component when the leads are plugged into the
current terminals.
Turn power OFF, break circuit, insert Meter in series, turn power
on.
MIN MAX
RANGEHOLD
mA
+
+
A
CATCAT
AIK08F.EPS
8
Measuring Frequency
XWWarning
To avoid electrical shock, disregard the bar graph for
frequencies >1 kHz. If the frequency of the measured
signal is >1 kHz, the bar graph is unspecified.
The Meter measures the frequency of a signal. The trigger level is
0 V ac for all ranges.
DC
⇒ To exit frequency, press yellow button or turn the rotary
switch.
⇒ In frequency, the bar graph shows the ac voltage accurately
up to 1 kHz.
⇒ Select progressively lower ranges using manual ranging for a
stable reading.
AC Voltage Frequency
MIN MAX
RANGEHOLD
V
+
Hz
EOM09F.EPS
.com
1.888.610.7664 info@Fluke-Direct.com
Page 12
Fluke-Direct
Using the Bar Graph
Using the Bar Graph
The bar graph is like the needle on an analog Meter. There is an
overload indicator (►) to the right, and a polarity indicator (±) to the
left.
Because the bar graph is much faster than the digital display, the
bar graph is useful for making peak and null adjustments, and for
observing rapidly changing inputs.
The bar graph is disabled when measuring capacitance. In
frequency, the bar graph accurately indicates the voltage or
current up to 1 kHz.
The number of lit segments indicates the measured value and is
relative to the full-scale value of the selected range.
For example, in the 60 V range (see below), the major divisions on
the scale represent 0, 15, 30, 45, and 60 V. An input of −30 V
lights the negative sign and the segments up to the middle of the
scale.
AIK11F.EPS
Cleaning
Wipe the case with a damp cloth and mild detergent. Do not use
abrasives or solvents. Dirt or moisture in the terminals can affect
readings.
Testing the Fuses
To avoid electrical shock or injury, remove the test
leads and any input signals before replacing the fuse.
Test fuses as shown below.
MIN MAX
RANGEHOLD
XWWarning
Ω
<12
V
<0.5
OK
11 A 440 mA
OKOK
OK
AIK12F.EPS
9
.com
1.888.610.7664 info@Fluke-Direct.com
Page 13
Model 77 Series IV
Fluke-Direct
Users Manual
Replacing the Battery and Fuses
To avoid shock, injury, or damage to the Meter:
• Use ONLY fuses with the amperage, interrupt,
voltage, and speed ratings specified. Disconnect test
leads before opening case.
XWWarning
I
F1
F1 Fuse, 440 mA, 1000 V, FAST Fluke PN 943121
F2 Fuse, 11 A, 1000 V, FAST Fluke PN 803293
B1 Battery, 9 V Alkaline,
NEDA 1604 / 1604A
10
F2
Fluke PN 614487
B1
AIK13F.EPS
b
.com
1.888.610.7664 info@Fluke-Direct.com
Page 14
Fluke-Direct
Specifications
Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy
specifications take the form of the following calculations:
Maximum voltage between any
terminal and earth ground: 1000
Surge Protection: 8 kV peak per IEC 61010
W Fuse for mA inputs:
W Fuse for A input:
Display: Digital: 6000 counts, updates 4/sec
Altitude: Operating: 2000 m; Storage: 12,000 m
Temperature: Operating: −10 °C to +50 °C;
Temperature coefficient: 0.1 X (specified accuracy) / °C
Electromagnetic Compatibility
(EN 61326-1:1997):
Relative Humidity: Maximum, noncondensing
Battery Life: Alkaline: 400 hrs typical
Size (H x W x L): 4.3 cm x 9 cm x 18.5 cm
Weight: 420 g
Safety Compliances: ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Measurement Category III, 600 V Measurement
In an RF field of 3 V/M, accuracy = specified accuracy.
90 % to 35 °C
75 % to 40 °C;
45 % to 50 °C
Category IV
± ( [ % of Reading ] + [ Counts ] )
; (N10140),VDE
11
.com
1.888.610.7664 info@Fluke-Direct.com
Page 15
Model 77 Series IV
Fluke-Direct
Users Manual
Function
AC Volts (Average
responding)
DC mV 600.0 mV 0.1 mV 0.3 % + 1
DC Volts 6.000 V
Continuity 600 Ω 1 Ω Meter beeps at <25 Ω, beeper turns off at >250 Ω;
Ohms
Diode test 2.400 V 0.001 V 1 % + 2
Capacitance
AC Amps (Average
[2]
responding)
[1] In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2.
[2] Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.
[3] 400.0 mA accuracy specified up to 600 mA overload.
[4] >10 A unspecified.
6.000 V
60.00 V
600.0 V
1000 V
60.00 V
600.0 V
1000 V
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
50.00 MΩ
1000 nF
10.00 μF
100.0 μF
9999 μF
60.00 mA
400.0 mA
6.000 A
10.00 A
Range Resolution Accuracy ± ( [ % of Reading ] + [ Counts ] )
0.001 V
0.01 V
0.1 V
1 V
0.001 V
0.01 V
0.1 V
1 V
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
1 nF
0.01 μF
[1]
[3]
[4]
0.1 μF
1 μF
0.01 mA
0.1 mA
0.001 A
0.01 A
2.0 % + 2
(45 Hz to 1 kHz)
0.3 % + 1
detects opens or shorts of 250 μs or longer.
0.5 % + 2
0.5 % + 1
0.5 % + 1
0.5 % + 1
0.5 % + 1
2.0 % + 1
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
2.5 % + 2
(45 Hz to 1 kHz)
12
.com
1.888.610.7664 info@Fluke-Direct.com
Page 16
Fluke-Direct
Specifications
Function
[3]
DC Amps
[1][ 2]
Hz
(ac voltage input )
MIN MAX AVG For dc functions, accuracy is the specified accuracy of the measurement function ±12 counts for changes longer than
[1] Frequency is specified from 2 Hz to 99.99 kHz.
[2] Below 2 Hz, the display shows zero Hz.
[3] Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.
[4] 400.0 mA accuracy specified up to 600 mA overload.
[5] >10 A unspecified.
60.00 mA
400.0 mA
6.000 A
10.00 A
99.99 Hz
999.9 Hz
9.999 kHz
99.99 kHz
275 ms in duration.
For ac functions, accuracy is the specified accuracy of the measurement function ±40 counts for changes longer than
1.2 s in duration.
Range Resolution Accuracy ± ( [ % of Reading ] + [ Counts ] )
[4]
[5]
0.01 mA
0.1 mA
0.001 A
0.01 A
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
1.5 % + 2
0.1 % + 1
13
.com
1.888.610.7664 info@Fluke-Direct.com
Page 17
Model 77 Series IV
Fluke-Direct
Users Manual
[1]
Function Overload Protection
Volts AC 1000 V >10 MΩ <100 pF >60 dB @ dc, 50 Hz or 60 Hz
Volts DC 1000 V >10 MΩ <100 pF >120 dB @ dc, 50 Hz or 60 Hz >60 dB @ 50 Hz or 60 Hz
mV 1000 V
Open Circuit Test
Ohms/Capacitance 1000 V
Continuity/Diode test 1000 V
7
[1] 10
V-Hz maximum.
[2] For circuits <0.3 A short circuit. 660 V for high energy circuits.
[2]
>10 MΩ <100 pF >120 dB @ dc, 50 Hz or 60 Hz >60 dB @ 50 Hz or 60 Hz
[2]
<8.0 V dc <660 mV dc <4.6 V dc <1.1 mA
[2]
<8.0 V dc 2.4 V dc <1.1 mA
Input Impedance
(Nominal)
Voltage
Common Mode Rejection Ratio
(1 kΩ Unbalanced)
Full Scale Voltage To:
6.0 MΩ 50 MΩ
Normal Mode Rejection
Short Circuit Current
mA Fused, 440 mA, 1000 V FAST Fuse 600 mA overload for 2 minutes maximum, 10
A Fused, 11 A, 1000 V FAST Fuse 20 A overload for 30 seconds maximum, 10
Function Overload Protection Overload
minutes rest.
minutes rest.
14
.com
1.888.610.7664 info@Fluke-Direct.com