76-804 & 76-807
Flood Phantoms
Users Manual

Fluke Biomedical
Radiation Management Services
6045 Cochran Road
Cleveland, Ohio 44139
440.498.2564
www.flukebiomedical.com/rms

Table of Contents
Section 1: General Information................................................................................... 1-1
1.1 Introduction .................................................................................................. 1-1
1.2 Specifications............................................................................................... 1-2
1.3 Filling the Phantom ...................................................................................... 1-2
1.4 Draining the Phantom .................................................................................. 1-3
1.5 Maintenance ................................................................................................ 1-3
i

General Information
Introduction
1
Section 1
General Information
1.1 Introduction
Flood Phantoms offer a simple, efficient means of obtaining optimum camera performance with respect to
uniformity of response over the entire crystal area. They consist of plastic forms that contain a watertight
central cavity into which a radioactive solution is introduced (via a filling port) and then mixed thoroughly.
After the activity has been distributed evenly, the camera’s uniformity of response can be checked.
Figure 1-1. Flood Phantom Models
1-1

Nuclear Associates 76-801,76-801-1926,76-804,76-807
Operators Manual
1.2 Specifications
Rectangular Flood Phantom (76-801)
Dimensions
Cavity Dimensions
Weight
Extra-Large Rectangular Flood Phantom (76-801-1926)
Dimensions
Cavity Dimensions
Weight
Jumbo Flood Phantom (76-804)
Dimensions
Cavity Dimensions
Weight
Gigantic Flood Phantom (76-807)
Dimensions
Cavity Dimensions
Weight
18” x 23” x 1” thick
16” x 21” x ½” deep
10 lbs
20½” x 28” x 1¼” thick
18½” x 26” x ½” deep
20 lbs
18” x 18” x 1” thick
17” x 17” x ½” deep
9 lbs
23” x 23” x 1” thick
22” x 22” x ½” deep
13 lbs
1.3 Filling the Phantom
The filler and vent plugs of these phantoms are located on the flat surface rather than on one of the
edges. Filling the phantom through openings on the edge requires that the phantom stand up on the
opposite edge. This causes water pressure to develop on the flat faces, making them bulge. These
bulges remain, even when the phantom is placed horizontally, causing non-uniformities in the distribution
of radioactivity. By filling the phantom in the flat or horizontal position (with the filling vent corner raised a
fraction of an inch, so as not to trap air), water pressure build-up due to depth is prevented, resulting in a
totally non-bulging phantom.
1. Remove the filler and vent plugs and the rubber “O” rings.
2. Partially fill the phantom with water to approximately 90% of capacity.
3. Spike the solution in the phantom with an appropriate isotope (5-10 mCi of Tc-99 m is more than
sufficient activity).
4. Replace the vent and filler plugs, and rubber “O” rings carefully so that the phantom is properly sealed.
(Do not over-tighten.)
5. Grasp the flood phantom by the edges and rotate it so that the trapped air serves to mix the spike and
carrier solution. Thirty seconds is sufficient to insure homogeneity of the mixture.
6. Remove the filler and vent plugs and carefully fill the flood phantom completely with degassed water.
Remove any trapped air by lightly tapping the face of the phantom.
7. Replace the filler and vent plugs in a controlled area and make sure that no leaks are present.
1-2

General Information
Filling the Phantom
NOTE
If a small bubble of air remains, raise the filling
corner of the phantom by tilting the camera detector
about 1/8” to ¼”, so the bubble will remain in the
side arm of the liquid-filled region.
The flood phantom is now ready to use.
1.4 Draining the Phantom
To empty the phantom, remove both the filler and vent plugs and the “O” rings, and invert over an
appropriate drain. Remove the drain vent plug now located on the top edge of the phantom.
Make sure that the radioactive solution being
drained is below the maximum permissible
concentration, which the N.R.C. (or your agreement
state) will allow to be discharged into public
drainage systems. If it is not, (a) hold the solution
in a separate container until it decays below this
value or (b) dilute the solution below the
maximum permissible level.
CAUTION
1
1.5 Maintenance
Cleaning
Use any mild detergent in warm water.
Care of Filler and Vent Plugs
Lubricate the threads of both plugs regularly with good silicone grease or a stopcock lubricant.
Do not use organic solvents, (they will attack the
phantom walls). As an algaecide, use 2 cc of
copper sulfate solution, made by dissolving 5
grams of copper sulfate in 100 cc of water.
Alternatively, use 5 to 10 drops of Lugol’s Solution
(iodine) or any common swimming
pool algaecide (chlorine).