Fluke 76-419-4150 User Manual

Nuclear Associates
76-419-4150
Adult/Pediatric CT Head and Body Dose Phantoms
March 2005 Manual No. 76-419-4150-1 Rev. 1 ©2004, 2005 Fluke Corporation, All rights reserved. Printed U.S.A. All product names are trademarks of their respective companies
Users Manual
Fluke Biomedical
Radiation Management Services
6045 Cochran Road Cleveland, Ohio 44139
440.498.2564
www.flukebiomedical.com/rms
Table of Contents
Section 1: General Information.......................................................................................................................... 1-1
1.1 Introduction ................................................................................................................................................................1-1
1.2 Applications.................................................................................................................................................................1-1
Section 2: Operation............................................................................................................................................... 2-1
i
(Blank Page)
General Information
Introduction
1
Section 1
General Information
1.1 Introduction
These phantoms can be used with any computed tomography (CT) system designed to image both adult and pediatric head and body. They can separate dose information for each. When performing dose profile measurements, the dose phantoms allow the user to collect information for the maximum, minimum and mid-range value of the nominal tomographic section thickness.
This essential phantom kit consists of three parts: an adult body phantom, an adult head phantom that doubles as a pediatric body phantom and a pediatric head phantom. All are made of solid acrylic, 15 cm thick, with diameters of 32 cm, 16 cm, and 10 cm respectively. Each part contains five probe holes, one in the center and four around the perimeter, 90° apart and 1 cm from the edge. The inside diameter of the holes is 1.31 cm. Each part includes five acrylic rods for plugging all the holes in the phantom. A storage and carrying case is available as an option.
1.2 Applications
The CT Dose Phantoms were designed in accordance with the Food and Drug Administration's performance standard for diagnostic x-ray systems, which includes regulations specifically applicable to CT systems (21 CFR
1020.33).