Fluke 754 Operating Manual

Page 1
754
HART Mode

Users Guide

July 2011 Rev. 1, 8/12 © 2011-2012 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies.
Page 2
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for three years from the date of purchas e. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend an y other warranty on Fluke’s behalf. To obtain service during the warranty period, contact your n earest F luke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A.
11/99
Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands
Page 3

Table of Contents

Title Page
Introduction .................................................................................................................... 1
How to Contact Fluke ..................................................................................................... 1
HART Channel Selection ............................................................................................... 2
Connect to a HART Transmitter ..................................................................................... 3
Supported vs. Generic Transmitters ............................................................................... 6
Communication Operations ............................................................................................ 7
See Process Variables .............................................................................................. 9
Setup Operations ....................................................................................................... 9
Basic ..................................................................................................................... 9
Sensor .................................................................................................................. 10
Device Identification .............................................................................................. 10
HART Output ........................................................................................................ 11
HART Information ................................................................................................. 11
Service Operations .................................................................................................... 11
Abort Softkey ............................................................................................................. 12
i
Page 4
754
Users Guide
Analog Mode and HART Mode Interaction .................................................................... 12
HART Calibration ........................................................................................................... 12
HART Mode Menus for Adjustment ............................................................................... 12
Calibrate a Supported HART Transmitter ...................................................................... 14
Loop Test .................................................................................................................. 15
Output Trim ............................................................................................................... 16
Sensor Trim .............................................................................................................. 16
Duplicate a Transmitters Basic Information ................................................................... 18
ii
Page 5

List of Tables

Table Title Page
1. Supported vs. Non-Supported Instruments ........................................................................... 6
iii
Page 6
754
Users Guide
iv
Page 7

List of Figures

Figure Title Page
1. HART Channel Selection....................................................................................................... 2
2. Active Device Screen ............................................................................................................ 4
3. HART Transmitter Connection .............................................................................................. 5
4. Supported Device Screen...................................................................................................... 7
5. 754 HART Mode Menu Tree ................................................................................................. 8
6. Device Variables Screen ....................................................................................................... 9
7. Basic Setup Screen ............................................................................................................... 9
8. Sensor Setup Screen ............................................................................................................ 10
9. Device Identification Screen .................................................................................................. 10
10. HART Output Screen ............................................................................................................ 11
11. HART Information Screen ..................................................................................................... 11
12. Block Diagram of a HART Transmitter .................................................................................. 13
13. Calibrator Mode Select Screen .............................................................................................. 14
14. Loop Test Screen .................................................................................................................. 15
15. Output Trim Screen ............................................................................................................... 16
v
Page 8
754
Users Guide
16. Sensor Trim Screen .............................................................................................................. 16
17. Sensor Trim Zero Screen ...................................................................................................... 17
18. Basic Screen ......................................................................................................................... 18
vi
Page 9

Introduction

 Warning
To prevent possible electrical shock, fire, or personal injury, read the Safety Information in the 753/754 Users Manual before you use the 754 Documenting Process Calibrator.
Adjustment is necessary for analog transmitters during calibration. With HART Transducer) transmitters, adjustments are by remote command. For these adjustments, use of a communication tool and a Calibrator is necessary. The 754 (Product or Calibrator) gives communication and calibration functions in one tool.
This guide tells you how to use the HART communication mode. HART mode is a procedure for the Product to communicate over its serial HART interface to a HART transmitter. See the 753/754 Users Manual for safety information, analog mode use instructions, specifications, and more general data. All of the Product functions shown in the 753/754 Users Manual are available and can be used with supported HART transmitters. The only transmitters addressed in this manual are HART transmitters.
Some specific HART transmitter commands are only available with supported devices.
(Highway-Addressable Remote
Note

How to Contact Fluke

To contact Fluke, call one of the following telephone numbers:
Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-6714-3114
Singapore: +65-6799-5566
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com. To register your product, visit http://register.fluke.com. To view, print, or download the latest manual supplement,
visit http://us.fluke.com/usen/support/manuals.
1
Page 10
754
Users Guide

HART Channel Selection

For HART communication, use the HART connector on the side of the Product or the mA jack on the front of the Product. The procedure you use depends on the application and preference.
Use Setup mode to choose the correct HART channel:
1. Push s. The first setup screen shows. See Figure 1.
2. Push  or  to select HART Channel.
3. Push .
4. Push  or  to select HART Port or mA Jack.
5. Push .
The selection is kept in Product memory. The Product uses this selection until it is changed in setup mode. Depending on the selection, HART Port or mA Jack shows at the top of the display when you push r.
Note
The figures in this guide show the mA Jack in use for HART communication.
Figure 1. HART Channel Selection
gnx19s.bmp
2
Page 11
HART Mode
Connect to a HART Transmitter

Connect to a HART Transmitter

To connect and start communication with a HART transmitter:
Note
If only a communication connection to a transmitter that is powered in a loop is necessary, attach the HART alligator clips to the loop power terminals on the transmitter and push r.
1. Connect the Product mA jacks to the loop power terminals of the transmitter. See Figure 3.
2. Connect the HART interface cable into the HART jack, and then connect the alligator clips to the terminals in step 1. There is no incorrect polarity. If the HART signal is directed through the mA jack in setup mode, this connection is not required.
3. Push r.
4. If the transmitter is not powered by a loop power supply, push the Loop Power softkey to start loop power.
Notes
The Product supplies loop power through an internal series resistance of 250
Ω
.
If the Product shows a measurement of 0 mA, make sure you have not reversed the current leads.
If an external loop power supply is used, a resistance of between 230
Ω
and 270 external loop supply and the transmitter is necessary.
5. The Product tries to connect to Poll Address 0 (single transmitter per loop). If no connection is made, push the Poll softkey to examine Poll Addresses 1 through 15 (multidrop).
6. When the Product establishes communication with the transmitter, the Active Device screen shows. See Figure 2. In a multidrop configuration, you must choose a transmitter from a list, and push .
connected in series with the
Ω
3
Page 12
754
Users Guide
The Active Device screen supplies this data for all transmitters:
Poll address (if not 0)
Model number and Tag ID
PV (Primary Variable)
PVAO (digital representation of the Analog Output)
PV LRV (PV Lower Range Value)
PV URV (PV Upper Range Value)
Softkeys for accessing HART operation menus
4
Figure 2. Active Device Screen
gnx06s.bmp
Page 13
HART Mode
Connect to a HART Transmitter
754
DOCUMENTING PROCESS CALIBRATOR
Figure 3. HART Transmitter Connection
+
gks61.eps
5
Page 14
754
Users Guide

Supported vs. Generic Transmitters

The Product communicates with most HART transmitters. “Supported Transmitters” are transmitters in which the Product is programmed to use their device-dependent commands and software version. “Generic transmitters” are transmitters that are not programmed in the Product. Table 1 shows the operations that are available for supported vs. generic transmitters in single-point and multidrop configurations.
Table 1. Supported vs. Generic Transmitters (Devices)
Menu Operation
Top Level Active Device screen
Basic (read/write, cloning capability) Sensor
Setup
Service
Process Detailed process information
(Temperature) Sensor (read/write) Device Identification (read/write) HART Output (read/write) HART Information (read only) Loop Test Pressure Zero Trim Output Trim Not Available Not Available Sensor Trim Not Available Not Available
Supported Transmitter Generic Transmitter
Single Point Multidrop Single Point Multidrop
Read only Read only
Not Available Not Available
6
Page 15
HART Mode
Communication Operations
Note
Sensor trim is supplied for supported transmitters, with some exceptions, as identified in the list of supported transmitters that you can see on the display.
1. To show a list of supported transmitters and software versions:
If the Product is connected to a HART
transmitter, push the Abort softkey, and then More Choices. Continue to Step 2.
If the Product is not connected to a HART
transmitter, push r and then More Choices. Continue to Step 2.
2. Push the Device Revs softkey. The screen in Figure 4 is shown.
gnx07s.bmp
Figure 4. Supported Device Screen
3. Push  or  to highlight the applicable manufacturer, and push . A list of model numbers is shown.
4. Push  or  to highlight the model number, and push . A list of software versions is shown.

Communication Operations

Figure 5 shows the HART Mode menu tree. Menu selections are determined by the type of transmitter that you use and the configuration connection type (current loop or multidrop).
7
Page 16
754
Users Guide
PROCESS
SETUP
SERVICE
ABORT
·
View process variables
·
View variable map
·
Re-map process
variables (Dual sensor
temperature devices)
Loop test Pressure zero trim Output trim Sensor trim
·
Basic
Sensor
Device Identification
HART Output
HART Information
Tag
·
PV units
·
LRV , URV
·
Damping
·
T ransfer function
·
Software Version
·
Final assembly number
·
Tag
·
Message
·
Date
·
Description
·
Manufacturer
·
Model
·
Device HART ID
·
Software revision
·
Hardware revision
·
Number of preambles
Figure 5. 754 HART Mode Menu Tree
·
Keypad input
·
Adjust URV, LRV to applied
values
Sensor 1
·
Sensor serial number
·
Sensor lower and upper limits
·
Sensor minimum span
Temperature devices only:
·
Change Sensor Type
·
Change Sensor Connections
Sensor 2
·
Change Sensor Type
·
Change Sensor Connections
(Dual sensor temperature devices)
·
Dual Sensor Config
(Dual sensor temperature devices)
· Write protect
· Alarm state
· HART poll address
· HART burst mode
· HART burst command
gnx03f.eps
8
Page 17
HART Mode
Communication Operations

See Process Variables

From the Active Device screen:
1. Push the Process softkey to see more device variables and their continuously updated values.
2. Select View Process Variables.
3. Push .
To see more data, push the Next Page softkey. See Figure 6.
Figure 6. Device Variables Screen
gnx08s.bmp

Setup Operations

The Setup softkey gives access to these five setup functions:
Basic
Sensor
Device Identification
HART Output
HART Information

Basic

Use the Basic setup screen to replicate a transmitter as described at the end of this manual.
From the Active Device screen, push the Setup and
Basic softkeys. See Figure 7.
Figure 7. Basic Setup Screen
gnx09s.bmp
9
Page 18
754
Users Guide

Sensor

This is where you can see data about the sensor in the transmitter, that includes serial number, limits, and span. The limits shown are the absolute limits for the sensor. (The Upper Range Value (URV) and Lower Range Value (LRV) are different, and are viewable and programmable through the Basic Setup screen.)
From the Active Device screen, push the Setup and Sensor softkeys to access the Sensor Setup screen. See Figure 8.
You can also program the Sensor Type and Sensor Connection registers in a supported temperature transmitter using this screen. Screen examples do not show Temperature device menu items.

Device Identification

You can program the Tag, Message, Date, and Descriptor registers in the transmitter when you use this screen.
From the Active Device screen, push the Setup and Device Identification softkeys to see data about the transmitter. See Figure 9.
gnx11s.bmp
Figure 9. Device Identification Screen
10
Figure 8. Sensor Setup Screen
gnx10s.bmp
Page 19
HART Mode
Communication Operations

HART Output

From the HART Output screen you can change the Poll Address (0 = single transmitter, any other address = multidrop), and control burst mode.
From the Active Device screen, push the Setup and HART Output softkeys to access the read/write HART Output screen. See Figure 10.
gnx12s.bmp
Figure 10. HART Output Screen

HART Information

This screen shows more complete data about the transmitter model, hardware and software revision numbers, and how many preambles it sends.
From the Active Device screen push the Setup and HART Information softkeys to access the read-only HART Information screen. See Figure 11.
Figure 11. HART Information Screen

Service Operations

The Service softkey gives you access to Loop Test, Pressure Zero Trim (where applicable), Output Trim, and Sensor Trim operations. For generic transmitters, only Loop Test, Output Trim, and Pressure Zero Trim are available (see Table 1.) The trim (adjustment) operations are later in this manual.
Note
Loop Test and Output Trim are not available if the transmitter is in multidrop mode.
gnx13s.bmp
11
Page 20
754
Users Guide

Abort Softkey

The Abort softkey stops the communication operation in process, and gives control to the previous screen. From the Active Device screen, Abort calls up the browser, in which you can see the list of connected transmitters.

Analog Mode and HART Mode Interaction

Analog mode is usual Product operation, as given in the 753/754 Users Manual. HART Communication mode starts when r is pushed.
Push r to change between HART and analog modes, or push to change to analog mode from HART mode. This lets the transmitter automatically set up analog mode for applicable measure and source functions, if necessary.
For supported transmitters, the change to analog mode goes to the MEASURE/SOURCE screen. This makes it easy to continue with an “as found” calibration.
For generic transmitters, the change to analog mode makes you choose the MEASURE or SOURCE screen, from which you select the applicable function.
For supported or generic transmitters, when you push r to return to HART mode, the Active Device screen is shown. The HART serial communication connection stays active as you change between HART and analog modes.

HART Calibration

An analog transmitter has one step of electrical conversion from a measured physical parameter to a 4-20 mA current loop output. A HART transmitter has the three blocks. See Figure 12.
It can be necessary to examine and adjust the Input block, the Output block, or both. For example, if your application requires the Primary Variable (PV) to be correct when read by a host computer, you must calibrate the Input stage.
If your application requires that the 4-20 mA current output value accurately show what the Input block is measuring, you must calibrate the Input and Output block.
Transmitters in multidrop systems, in which more than one is wired in parallel, do not use their Output blocks. Their analog outputs are all held at an idling level of 4 mA no matter what the Input block measures.

HART Mode Menus for Adjustment

Sensor Trim in HART mode is when you adjust the Input block. Output Trim is when you adjust the Output block. These adjustments are made from the Service menu.
For pressure transmitters, Pressure Zero Trim is an additional adjustment. This adjustment is the same as when you set the lower sensor point at zero. All three operations are energized from the HART mode Service menu.
12
Page 21
HART Mode
HART Mode Menus for Adjustment
754
DOCUMENTING PROCESS CALIBRATOR
754
Pressure zero trim
Physical
variable
Sensor trim
Input
PV PVAO
Computation
HART transmitter
Figure 12. Block Diagram of a HART Transmitter
Loop test Output trim
Output
4-20 mA loop
gnx02f.eps
13
Page 22
754
Users Guide

Calibrate a Supported HART Transmitter

An “as found” and “as left” transmitter calibration has an easier, more automated procedure for HART transmitters than for analog transmitters. The calibration procedure is the same as given in the 753/754 Users Manual. How to set up the calibration template, and how you adjust the transmitter are different.
Note
If you start a calibration procedure from an installed task, do not push Task until the Product is connected to the HART transmitter and communication in analog mode is established.
The subsequent procedure assumes that you know how to use a 754 Series Documenting Process Calibrator to calibrate analog transmitters, and are not running an installed Task.
1. Make the applicable measure, source, and HART interface connections between the Product and the transmitter.
2. Push r to establish communication.
3. Push r to change to analog mode.
4. You are shown a set of selections for analog mode. Use and to select one of the measure/source choices. This is where you select measure mA (analog output), or measure PV, which does not involve the Output block of the transmitter. See Figure 13.
gnx14s.bmp
Figure 13. Calibrator Mode Select Screen
5. Push .
6. Push the As Found softkey.
7. Push  or  followed by  to select an instrument calibration procedure. You can see that the calibration template is installed with applicable data. You can make changes to items if necessary. The Error% value must be entered before you continue.
8. Continue with calibration as given in the 753/754 Users Manual. When calibration is complete, do step 9.
9. Push the Adjust softkey to go back to HART mode and launch the Service menu for the transmitter.
14
Page 23
HART Mode
Calibrate a Supported HART Transmitter
10. Do an Output Trim, and/or a Sensor Trim. These are Service operations. See “Output Trim” and “Service Trim”. When you are done with the trim procedures, push the Done softkey.
11. Push the As Left procedure again to make sure that the transmitter passes.
softkey and do the calibration

Loop Test

The Loop Test feature sends a command to the transmitter to set its output block to a specified value. Use this to examine the calibration of the output block, or to verify a correct indication on an external loop reading device.
To do a Loop Test:
1. From the Active Device screen, push the Service softkey, followed by Loop Test. See Figure 14.
gnx15s.bmp
Figure 14. Loop Test Screen
2. Push a softkey to command the transmitter to set its analog output to the related value, or manually record a value with the numeric buttons.
The transmitter acknowledges the selected PVAO (digital representation of the Output stage) in the middle of the screen. The Product shows the measured value at the top of the screen so you can use Loop Test as a fast procedure to check the calibration of the transmitter’s Output block.
15
Page 24
754
Users Guide

Output Trim

An Output Trim adjusts the transmitter’s Output block. You can do an Output Trim on generic and supported transmitters.
To do an Output Trim:
1. From the Active Device screen, push the Service softkey and then Output Trim. See Figure 15.
gnx16s.bmp
Figure 15. Output Trim Screen
2. Push Fetch to put the mA value you measure with the Product into the dialog box.
3. Push Send to trim the low analog output point.
4. To trim the high analog output point, do steps 2 and 3 again. Follow the prompts on the display.

Sensor Trim

You can only do a Sensor Trim on a supported transmitter. A Sensor Trim adjusts the Input block of a HART transmitter. If your application does not use the Output block of the transmitter, Sensor Trim is the only adjustment necessary.
A Sensor Trim can involve one or more trim points. This is controlled by the transmitter’s software.
To do a Sensor Trim:
1. From the Active Device screen, push the Service softkey, and then Sensor Trim. See Figure 16.
Figure 16. Sensor Trim Screen
gnx17s.bmp
16
Page 25
HART Mode
Calibrate a Supported HART Transmitter
2. Push  or  to select the Sensor Trim operation, and push .
3. Follow the instructions on the display. For pressure, there is a message to connect a pressure module to the Product, and a message to push to zero the pressure module.
4. Push Continue. See Figure 17.
gnx 18s.bmp
Figure 17. Sensor Trim Zero Screen
5. Record the necessary trim value the Product must produce, and push . In the case of a pressure transmitter (as above), push Fetch to put the value that is measured by the Fluke pressure module into the dialog box.
6. Push Trim.
7. If asked, do steps 2 and 3 again to trim the remaining points. Follow the display prompts.
Note
Select trim points at the LRV or URV values. To review these values, push the Show LRV/URV softkey.
If you receive an error that shows excessive correction for the selected trim point, it is possible to repair the problem with a series of sensor trims. Start the sensor trims in small increments from the previous trim point to the necessary trim point.
17
Page 26
754
Users Guide

Duplicate a Transmitters Basic Information

The Product lets you duplicate the Basic Setup data from one transmitter to a different transmitter. You can copy generic and supported transmitters.
To use the Product to duplicate a transmitter:
1. Connect the Product to the transmitter you want to duplicate. Only a communication connection is necessary.
2. From the Active Device screen, push the Setup softkey.
3. Push  or  keys to select Basic from the list on the screen and push . See Figure 18.
gnx09s.bmp
Figure 18. Basic Screen
4. Push the Store Page softkey to put the settings into Product memory (not the transmitter).
5. Disconnect the Product from the transmitter and connect it to the transmitter that you will configure.
6. Push the Abort softkey to start communication.
7. Push the Setup softkey, and then select Basic to go back to the Basic Setup screen.
8. Push the Recall Page softkey to refresh the Product display with the duplicated parameter settings. The settings are not transmitted to the transmitter at this time.
9. Push  or  and  to select parameters individually, change them, or leave them as duplicated, and push the Send softkey to transmit them to the transmitter.
18
Loading...