Manual Title: 71X Series Calibration Supplement Issue: 2 Part Number: 686540 Issue Date: 10/12 Print Date: January 1998 Page Count: 1 Revision/Date: 11, 3/12
This supplement contains information necessary to ensure the
accuracy of the above manual.
© 2012 Fluke Corporation. All rights reserved.
71X Series Calibration Manual Supplement
Change #1
On page 49, under mA Loop Power Verification, replace step 2 with:
2. Apply 1.2 k to the mA jacks from the decade box or other resistor source. The display must read
>19 mA but <OL.
On page 54, under mA Loop Power Verification (718 Only), replace step 2 with:
2. Apply 1.2 k to the mA jacks from the decade box or other resistor source. The display must read
>19 mA but <OL.
Change #2
On page 86, Table 71,
Change:
MP6 30G, 100G, 300G Pump Assembly 2558508 718 1
To:
MP6 Fluke-718, 30G/100G, Pump with clean out 2558508 718 1 MP6 Fluke-718, 300G, Pump with clean out 2632057 718 1