Fluke 715 mA Loop
Calibrators
Fluke 715 Volt/mA Calibrator
Volt Milliamp Calibrator to troubleshoot process loops
The Fluke 715 Volt/mA Calibrator delivers outstanding performance,
durability and reliability. Offered in the rugged Fluke 80 Series DMM
package, the calibrator is compact, lightweight and easy to carry. With
a push button interface similar to the multifunction Fluke 740 Series
Documenting Process Calibrators, the 715 is easy to use. It is EMI toler
ant, dust/splash resistant and features a removable battery door for
quick battery changes. It features:
Measure loop current (0 mA to 20 mA, 4 mA to 20 mA) signals
•
with 0.01 %
Source precision voltage signals up to 20 V with 0.01 % accuracy
•
Measure voltage output process signals from PLCs, transmitters
•
Source or simulate 24 mA loop current
•
24 V loop supply with simultaneous current measurement
•
Fluke 707 Loop Calibrator
The fastest, one-handed tool for loop checks
The Fluke 707 Loop Calibrator is a high performance solution for
calibration, repair and maintenance of current loops. The “Quick Click”
knob makes it extremely fast and easy to use. It features:
One handed user interface with “Quick Click” detented knob
•
mA sourcing, simulation and measurement
•
0.015 % accuracy and 0.001 mA resolution
•
Simultaneous mA and % of span display
•
Selectable step or ramp outputs
•
24 V Loop supply with mA measure, including 250 Ω HART resistor
•
0 V dc to 28 V dc measurement to check Loop Voltage
•
Removable battery door for quick battery replacement
•
Compact case with protective holster
•
Three year warranty
•
accuracy and 0.001 mA resolution
Technical Data
-
Fluke 705 Loop Calibrator
Low cost calibration muscle
The Fluke 705 Loop Calibrator is a cost-effective, integrated solution for
calibration, repair and maintenance of current loops. It features:
mA sourcing, simulation and measurement
•
0.02 % accuracy and 0.001 mA resolution
•
Simultaneous mA and % of span display
•
Selectable step or ramp outputs
•
24 V Loop supply with mA measure
•
0 V dc to 28 V dc measurement to check Loop Voltage
•
Removable battery door for quick battery replacement
•
Compact case with protective holster
•
Three-year warranty
•
General Specifications:
Maximum voltage: 30 volts: Non-operating temperature: –4 °C to 60 °C
Operating temperature: –10 °C to 55 °C
Relative humidity: 95 % (10 °C to 30 °C); 75 % (30 °C to 40 °C); 45 % (40 °C to 50 °C); 35 % (50 °C to 55 °C)
Operating altitude: 3,000 m max
Shock: 1 m drop test
Vibration: Random, 2 g, 5 Hz to 500 Hz
Safety: CSA C22.2 No. 1010.1:1992
EMC: EN50082-1:1992 and EN55022:1994 Class B
Power: 9 V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two batteries in 718
Warranty
Functional Specifications: Fluke 705 and 707 Fluke 715
Voltage measurement
Range 0 V to 28 V 0 mV to 200 mV / 0 V to 25 V
Resolution 1 mV 10 µV / 1 mV
Accuracy 705: 0.025 % Rdg + 1 counts
Current measurement
Range 0 mA to 24 mA 0 mA to 24 mA
Resolution 0.001 mA 0.001 mA
Accuracy 705: 0.02 % Rdg + 2 counts
Current sourcing
Range 0 mA to 20 mA or 4 mA to 20 mA 0 mA to 20 mA or 4 mA to 20 mA
Accuracy 705: 0.02 % Rdg + 2 counts
Drive capability 705: 1000 Ω @ 24 mA
Loop power while measuring mA 24 V 24 V
Voltage sourcing n/a 0 mV to 200 mV or 0 V to 20 V
Display current and % of span Yes mA or %
Auto step, auto ramp Yes Yes
Span Check Yes Yes
Battery life 18 hours typical, at 12 mA 18 hours typical, at 12 mA
: Three-years
0.01 % Rdg + 2 counts
707, 707Ex: 0.015 % Rdg + 2 counts
0.01 % + 2 counts
707, 707Ex: 0.015 % Rdg + 2 counts
0.01 % Rdg + 2 counts
707, 707Ex: 0.015 % Rdg + 2 counts
1000 Ω @ 24 mA
707: 1200 Ω @ 24 mA
707Ex: 700 Ω @ 20 mA
Fluke. Keeping your world
up and running.
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800) 36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2006 Fluke Corporation. All rights reserved.
Printed in U.S.A. 2/2006 2574106 D-EN-N Rev A