5606 and 5607
Full Immersion PRTs
for laboratory freezers,
autoclaves, and furnaces
• 5606: -200 °C to 160 °C
• 5607: 0 °C to 450 °C
• Calibration accuracy of ± 0.05 °C
• Transition junction and lead wires can
withstand full PRT temperature range
• Unique probe seal of 5606 prevents
ingress of moisture
• Calibration not included; NVLAP-accredited
calibration optional
Technical Data
The 5606 and 5607 Full
Immersion PRTs are designed
to perform in extreme environments where both the transition
junction and the lead wires are
required to withstand temperatures covering the entire
operating range of the probe.
(“Immersion” in this case simply
means exposure to temperatures other than ambient and
does not necessarily refer to the
medium to which the PRT and
its components are subjected.)
Applications where “full
immersion” of probe, transition junction, and lead wire is
required can include calibration
or validation of sensors used in
laboratory or bio freezers, walkin refrigerators, autoclaves,
ovens, stability test chambers,
furnaces or incubators. Other
applications might include IQ/
OQ/PQ qualification procedures,
temperature mapping, or data
logging in temperature controlled spaces or environmental
chambers.
The 5606 is just two inches
(50 mm) in length with a sheath
diameter of 1/8 inch (3.1 mm).
Since it can be fully immersed
over its entire temperature
range, you don’t have to worry
about calculating minimum
immersion depth—immerse the
entire probe, transition junction,
and lead wires, in either noncorrosive liquids or dry mediums
ranging from –200 °C to 160 °C.
The 100 W precision sensing
element (alpha = 0.00385) is
specially constructed to prevent
the ingress of moisture while
achieving calibration accuracies as good as ± 0.05 °C. You
get both accuracy and flexibility built into a small, durable
package.
The lead wires of the 5606
are made of single-conductor
enameled copper wire which
prevents moisture from wicking to the sensing element as
it would if a more common,
stranded lead wire were used.
The diameter of the lead wire is
no greater than 0.2 mm, making
it easy to run lead wires through
a freezer door to your readout or
transmitter with effectively zero
energy loss. The 5606 has been
proven in common heat transfer fluids such as silicone oil,
mineral oil, ethanol, and even
liquid nitrogen with no degradation of the transition junction
seal or lead wire insulation.
Are you performing calibrations or verifying the accuracy
of sensors used in ovens or
furnaces? The 5607 has been
designed specifically for your
application. Intended to be used
in dry mediums only, the 5607
can be fully immersed over a
temperature range of 0 °C to
450 °C. There isn’t another PRT
on the market that can be fully
immersed over this range and
remain as stable and accurate
(± 0.055 °C). The lead wire
of the 5607 is shielded with
braided stainless steel and connected to the transition junction
with a strain relief spring.
If you need longer lead wire
than the standard length, call
one of our Application Specialists today for a free quote.
They can also help you determine which readout and probe
termination are best suited for
your application or help you
select the right calibration for
your probe.
Specications
Specifications 5606 5607
Temperature range -200 °C to 160 °C 0 °C to 450 °C
Nominal resistance at 0.01 °C
Temperature coefficient
Accuracy
[1]
± 0.04 °C at 0.01 °C
± 0.06 °C at 160 °C
Short-term repeatability
[2]
± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C
Drift
[3]
± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C
Hysteresis ± 0.015 °C
Sheath length 50 mm ± 5 mm
(2 in ± 0.2 in)
Sheath diameter 3.1 mm ± 0.1 mm (1/8 in ± 0.004 in)
Sheath material 316 SST Inconel
Transition junction temperature
[4]
range
-200 °C to 160 °C 0 °C to 450 °C
Transition junction dimensions No transition
Sensor length 30 mm ± 3mm (1.2 in ±0.1 in)
Sensor location 3 mm ± 1 mm from tip (1.2 in ± 0.04 in)
Minimum insulation resistance
20 MW at 23 °C 100 MW at 23 °C
Minimum immersion length Full immersion Full immersion
Maximum immersion depth in
liquid medium
Maximum immersion depth in dry
medium
Response time
[5]
Full immersion
Full immersion Full immersion
12 seconds typical 9 seconds typical
Self heating (in 0 °C bath) ± 0.003 °C
Lead-wire cable type Enameled copper wire SST over-braid
Lead-wire length 2.4 meters (8 ft)
Lead-wire temperature range 160 °C 450 °C
Calibration
[1]
Includes calibration and 100 hr drift (k = 2).
[2]
Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k = 2).
[3]
After 100 hours at max temp, 95 % confidence (k = 2).
[4]
Temperatures outside this range will cause irreparable damage.
[5]
Per ASTM E 644.
Calibration not included. NVLAP accredited calibration available. See
ordering information.
100 W ± 0.1 W
0.00385 W/W/°C
± 0.04 °C at 0.01 °C
± 0.07 °C at 450 °C
± 0.020 °C at 0.01 °C
± 0.035 °C at 450 °C
± 0.02 °C at 0.01 °C
± 0.04 °C at 450 °C
40 mm ± 5 mm
(1.6 in ± 0.2 in)
®
35 mm x 7 mm
(1.4 in x 0.3 in)
Use SST or glass tube to protect
transition junction
600