Fluke 5606 Specifications

5606 and 5607
Full Immersion PRTs
for laboratory freezers,
autoclaves, and furnaces
5606: -200 °C to 160 °C
5607: 0 °C to 450 °C
Calibration accuracy of ± 0.05 °C
Transition junction and lead wires can
withstand full PRT temperature range
Unique probe seal of 5606 prevents ingress of moisture
Calibration not included; NVLAP-accredited calibration optional
Technical Data
The 5606 and 5607 Full Immersion PRTs are designed to perform in extreme environ­ments where both the transition junction and the lead wires are required to withstand tem­peratures covering the entire operating range of the probe. (“Immersion” in this case simply means exposure to tempera­tures other than ambient and does not necessarily refer to the medium to which the PRT and its components are subjected.)
Applications where “full immersion” of probe, transi­tion junction, and lead wire is required can include calibration or validation of sensors used in laboratory or bio freezers, walk­in refrigerators, autoclaves, ovens, stability test chambers, furnaces or incubators. Other applications might include IQ/ OQ/PQ qualification procedures, temperature mapping, or data logging in temperature con­trolled spaces or environmental chambers.
The 5606 is just two inches (50 mm) in length with a sheath diameter of 1/8 inch (3.1 mm). Since it can be fully immersed
over its entire temperature range, you don’t have to worry about calculating minimum immersion depth—immerse the entire probe, transition junction, and lead wires, in either non­corrosive liquids or dry mediums ranging from –200 °C to 160 °C. The 100 W precision sensing element (alpha = 0.00385) is specially constructed to prevent the ingress of moisture while achieving calibration accura­cies as good as ± 0.05 °C. You get both accuracy and flexibil­ity built into a small, durable package.
The lead wires of the 5606 are made of single-conductor enameled copper wire which prevents moisture from wick­ing to the sensing element as it would if a more common, stranded lead wire were used. The diameter of the lead wire is no greater than 0.2 mm, making it easy to run lead wires through a freezer door to your readout or transmitter with effectively zero energy loss. The 5606 has been proven in common heat trans­fer fluids such as silicone oil, mineral oil, ethanol, and even
liquid nitrogen with no degra­dation of the transition junction seal or lead wire insulation.
Are you performing calibra­tions or verifying the accuracy of sensors used in ovens or furnaces? The 5607 has been designed specifically for your application. Intended to be used in dry mediums only, the 5607 can be fully immersed over a temperature range of 0 °C to 450 °C. There isn’t another PRT on the market that can be fully immersed over this range and remain as stable and accurate (± 0.055 °C). The lead wire of the 5607 is shielded with braided stainless steel and con­nected to the transition junction with a strain relief spring.
If you need longer lead wire than the standard length, call one of our Application Special­ists today for a free quote. They can also help you deter­mine which readout and probe termination are best suited for your application or help you select the right calibration for your probe.
Specications
Specifications 5606 5607
Temperature range -200 °C to 160 °C 0 °C to 450 °C
Nominal resistance at 0.01 °C
Temperature coefficient
Accuracy
[1]
± 0.04 °C at 0.01 °C
± 0.06 °C at 160 °C
Short-term repeatability
[2]
± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C
Drift
[3]
± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C
Hysteresis ± 0.015 °C
Sheath length 50 mm ± 5 mm
(2 in ± 0.2 in)
Sheath diameter 3.1 mm ± 0.1 mm (1/8 in ± 0.004 in)
Sheath material 316 SST Inconel
Transition junction temperature
[4]
range
-200 °C to 160 °C 0 °C to 450 °C
Transition junction dimensions No transition
Sensor length 30 mm ± 3mm (1.2 in ±0.1 in)
Sensor location 3 mm ± 1 mm from tip (1.2 in ± 0.04 in)
Minimum insulation resistance
20 MW at 23 °C 100 MW at 23 °C
Minimum immersion length Full immersion Full immersion
Maximum immersion depth in liquid medium
Maximum immersion depth in dry medium
Response time
[5]
Full immersion
Full immersion Full immersion
12 seconds typical 9 seconds typical
Self heating (in 0 °C bath) ± 0.003 °C
Lead-wire cable type Enameled copper wire SST over-braid
Lead-wire length 2.4 meters (8 ft)
Lead-wire temperature range 160 °C 450 °C
Calibration
[1]
Includes calibration and 100 hr drift (k = 2).
[2]
Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k = 2).
[3]
After 100 hours at max temp, 95 % confidence (k = 2).
[4]
Temperatures outside this range will cause irreparable damage.
[5]
Per ASTM E 644.
Calibration not included. NVLAP accredited calibration available. See ordering information.
100 W ± 0.1 W
0.00385 W/W/°C
± 0.04 °C at 0.01 °C
± 0.07 °C at 450 °C
± 0.020 °C at 0.01 °C
± 0.035 °C at 450 °C
± 0.02 °C at 0.01 °C
± 0.04 °C at 450 °C
40 mm ± 5 mm (1.6 in ± 0.2 in)
®
35 mm x 7 mm (1.4 in x 0.3 in)
Use SST or glass tube to protect
transition junction
600
2 Fluke Calibration 5606 and 5607 Full Immersion PRTs